Troubleshooting: Addressing oral leak For U.S. distribution only. ©2018 Vyaire. Vyaire, the Vyaire Logo and SuperNO2VA are trademarks of Vyaire Medical, Inc. | CF/801/18/0005 (0118) For training videos and tips visit vyaire.com CORPORATE HEADQUARTERS Vyaire Medical, Inc. 26125 North Riverwoods Blvd Mettawa, IL 60045 USA Use reservoir bag motion to monitor ventilation 4 To ensure an air-tight seal, apply firm downward pressure to the mask and use the headstrap to secure in place. 1 Adjust APL valve to 10 cmH2O 3 Set FGF to 10 lpm 2 Sub-mental pressure on the tongue 3 Slight head flexion 2 Manual mouth closure 1 Is your SuperNO 2 VA device sealed properly on the patient’s face? If the reservoir bag doesn’t pressurize: • Ensure SuperNO 2 VA device seal is air-tight. • Address any leaks from the mouth. • Check for leaks and close any open ports. HYPOXEMIA FROM UPPER AIRWAY OBSTRUCTION? SuperN O 2 V A T M nasal PAP ventilation device

Vyaire Super Nova 2 Workroom Poster 11x17 Final · Vyaire_Super Nova 2_Workroom Poster_11x17_Final Created Date: 1/25/2018 10:36:28 AM

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Vyaire Super Nova 2 Workroom Poster 11x17 Final · Vyaire_Super Nova 2_Workroom Poster_11x17_Final Created Date: 1/25/2018 10:36:28 AM

Troubleshooting: Addressing oral leak

For U.S. distribution only. ©2018 Vyaire. Vyaire, the Vyaire Logo and SuperNO2VA are trademarks of Vyaire Medical, Inc. | CF/801/18/0005 (0118)

For training videos and tips visit vyaire.com

CORPORATE HEADQUARTERS

Vyaire Medical, Inc.

26125 North Riverwoods Blvd

Mettawa, IL 60045 USA

Use reservoir bag motion to monitor ventilation4

To ensure an air-tight seal, apply firm downward pressure to the mask and use the headstrap to secure in place.1

Adjust APL valveto 10 cmH2O3Set FGF to 10 lpm2

Sub-mental pressure on the tongue3Slight head flexion2Manual mouth

closure1

Is your SuperNO2VA device sealed properly on the patient’s face? If the reservoir bag doesn’t pressurize: • Ensure SuperNO2VA device seal is air-tight. • Address any leaks from the mouth. • Check for leaks and close any open ports.

HYPOXEMIA FROM UPPER AIRWAY OBSTRUCTION?

SuperNO2VATM

nasal PAP ventilation device