VPT 607 Veterinary Chemotherapy Lab Manual

  • Upload
    sunil

  • View
    231

  • Download
    0

Embed Size (px)

Citation preview

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    1/26

    Ex No: 1

    GENERAL METHODS FOR ASSAY OF CHEMOTEHRAPEUTICAGENTS

    Unlike tests for the evaluation of disinfectants wheredetermination of cidal activity is of paramount importance, tests

    involving potential chemotherapeutic agents (antibiotics) invariably

    have as their main focus determination of MIC.

    Tests for bacteriostatic activity

    Disc tests

    These are modifications of the earlier cup or ditch-plate procedures

    where filter-paper discsimpregnated with the antimicrobial replace the

    antimicrobial-filled cups or wells. For disc tests, standard suspensions

    (e.g. 0.5 McFarland standard) of log phase growth cells are prepared

    and inoculated onto the surface of appropriate agar plates to form a

    lawn. Commercially available ilter-paper discs containing known

    concentrations of antimicrobial agent (it is possible to prepare your

    own discs for use with novel drugs) are then placed on the dried lawn

    and the plates are incubated aerobically at 35C for 18 hours.

    Disc tests are basically qualitative, although it is possible to get

    some information on the degree of activity depending on the zone

    size . Although there are subtle variations of the disc test used in somecountries, the basic principles behind the tests remain similar and are

    based on the original work of Bauer and colleagues. Some techniques

    employ a control bacterial isolate on each plate so that comparisons

    between zone sizesaround the test and control bacterium can be

    ascertained (i.e. a disc potency control).

    Control strains of bacteria are available which should have

    inhibition zones of a given diameter with stipulated antimicrobial discs.

    Use of such controls endorses the suitability of the methods (e.g.

    medium, inoculum density, incubation conditions) employed. For slow-

    growing microorganisms, the incubation period can be extended.

    Dilution tests

    These usually employ liquid media but can be modified to involve

    solid media. Doubling dilutions usually in the range 0.12256mg/L

    of the antimicrobial under test are prepared in a suitable broth

    medium, and a volume of log phase cells is added to each dilution to

    1

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    2/26

    result in a final cell density of around 5 \ 105 CFU/ml. After incubation

    at 35C for 18 hours, the concentration of antimicrobial contained in

    the first clear tube is read as the MIC..

    Endpoints with dilution tests are usually sharp and easily defined,

    although skipped wells (inhibition in a well with growth either side)

    and trailing (a gradual reduction in growth over a series of wells) may

    be encountered. The latter is especially evident with antifungal tests

    Dilution tests can also be carried out using a series of agar plates

    containing known antimicrobial concentrations. Appropriate bacterial

    suspensions are inoculated onto each plate and the presence or

    absence of growth is recorded after suitable incubation.

    E-tests

    The most convenient and presently accepted method of determining

    bacterial MICs, however, is the E (Epsilometer)-test.Basically this is performed in a similar manner to the disc test

    except that nylon strips that have a linear gradient of antimicrobial

    lyophilized on one side are used instead of the filter-paper

    impregnated antimicrobial discs.

    On the other side of the nylon strip are a seriesof lines and

    figures denoting MIC values (the nylon strips are placed antimicrobial

    side down on the freshly prepared bacterial lawn and, after incubation,

    the MIC is determined by noting where the ellipsoid (pear-shaped)

    inhibition zone crosses the strip . For most microorganisms, there

    appears to be excellent correlation between dilution and E-test MIC

    results.

    Tests for bactericidal activity

    The MBC is the lowest concentration (in mg/L) of antimicrobial that

    results in 99.9% killing of the bacterium under test. The 99.9% cutoff

    is an arbitrary in vitro value with 95% confidence limits that has

    uncertain clinical relevance.

    MBCs are determined by spreading 0.1-ml(100-ml) volumes of all clear

    (no growth) tubes from a dilution MIC test onto separate agar plates(residual antimicrobial in the 0.1-ml sample is diluted out over the

    plate).

    After incubation at 35C overnight (or longer for slow-growing

    bacteria), the numbers of colonies growing on each plate are

    recorded. The first concentration of drug that produces

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    3/26

    with MICs, the initial bacterial inoculum should result in about 5 \ 105

    CFU/ml. Inhibition, but not killing of this inoculum, should therefore

    result in the growth of 50 000 bacteria from the 0.1-ml sample. A

    99.9% (3-log) kill would result in no more than 50 colonies on the

    subculture plate. With most modern antibacterial drugs, the

    concentration that inhibits growth is very close to the concentration

    that produces death, e.g. within one or two dilutions. In general, only

    MICs are determined for such drugs.

    Ex No: 2

    ANTIMICROBIAL SUSCEPTIBILITY TESTING

    PrincipleThe principles of determining the effectivity of a noxious agent to a

    bacterium were well enumerated by Rideal ,Walker and others at theturn of the century, the discovery of antibiotics made these tests(or

    their modification)too cumbersome for the large numbers of tests

    necessary to be put up as a routine. The ditch plate method of agar

    diffusion used by Alexander Fleming was the forerunner of a variety of

    agar diffusion methods devised by workers in this field .The Oxford

    group used these methods initially to assay the antibiotic contained in

    blood by allowing the antibiotics to diffuse out of reservoirs in the

    medium in containers placed on the surface.

    With the introduction of a variety of antimicrobials it became

    necessary to perform the antimicrobial susceptibility test as a routine.

    For this, the antimicrobial contained in a reservoir was allowed to

    diffuse out into the medium and interact in a plate freshly seeded with

    the test organisms. Even now a variety of antimicrobial containing

    reservoirs are used but the antimicrobial impregnated absorbent paper

    disc is by far the commonest type used. The disc diffusion method of

    AST is the most practical method and is still the method of choice for

    the average laboratory. Automation may force the method out of the

    diagnostic laboratory but in this country as well as in the smaller

    3

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    4/26

    laboratories of even advanced countries, it will certainly be the most

    commonly carried out microbiological test for many years to come. It

    is, therefore, imperative that microbiologists understand the principles

    of the test well and keep updating the information as and when

    necessary. All techniques involve either diffusion of antimicrobial agent

    in agar or dilution of antibiotic in agar or broth.

    Even automated techniques are variations of the above methods.

    Methods of Antimicrobial Susceptibility Testing

    Antimicrobial susceptibility testing methods are divided intotypes based on the principle applied in each system. They

    include:

    Diffusion DilutionDiffusion&Dilution

    Stokes method Minimum Inhibitory Concentration

    E-Test method

    Kirby-Bauer method i) Broth dilution

    ii)Agar Dilution

    Quantification of antimicrobial concentration in plasma/tissues is thebasic prerequisite to determine the bioavailability / bioequivalence(for new commercial formulation), pharmacokinetic parameters andfixation of dose/dosing schedule. The different methods available forassay of antimicrobial agents are:

    1. Microbiological assay ( disc-diffusion method)2. Colorimetric/ Spectrophotometric : eg. Sulphonamides3. Fluorimetric : Tetracycline/ Doxycycline4. HPLC: any antimicrobial at residue levels can be quantified by

    using suitable detection electrode

    DISK DIFFUSION

    Reagents for the Disk Diffusion Test1. Meller-Hinton Agar MediumOf the many media available, Meller-Hinton agar is considered to bethe best for routine susceptibility testing of nonfastidious bacteria forthe following reasons:* It shows acceptable batch-to-batch reproducibility for susceptibility

    testing.* It is low in sulphonamide, trimethoprim, and tetracycline inhibitors.* It gives satisfactory growth of most nonfastidious pathogens.

    4

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    5/26

    * A large body of data and experience has been collectedconcerning susceptibility tests performed with this medium.

    Although Meller-Hinton agar is reliable generally for susceptibilitytesting, results obtained with some batches may, on occasion, varysignificantly. If a batch of medium does not support adequate growth

    of a test organism, zones obtained in a disk diffusion test will usuallybe larger than expected and may exceed the acceptable qualitycontrol limits. Only Meller-Hinton medium formulations that havebeen tested according to, and that meet the acceptance limitsdescribed in, NCCLS document M62-A7- Protocols for EvaluatingDehydrated Meller-Hinton Agar should be used.

    Preparation of Meller-Hinton AgarMeller-Hinton agar preparation includes the following steps.1. Meller-Hinton agar should be prepared from a commercially

    available dehydrated base according to the manufacturer's

    instructions.2. Immediately after autoclaving, allow it to cool in a 45 to 50C water

    bath.3. Pour the freshly prepared and cooled medium into glass or plastic,

    flat-bottomed petri dishes on a level, horizontal surface to give auniform depth of approximately 4 mm. This corresponds to 60 to70 ml of medium for plates with diameters of 150 mm and 25 to30 ml for plates with a diameter of 100 mm.

    4. The agar medium should be allowed to cool to room temperatureand, unless the plate is used the same day, stored in a refrigerator

    (2 to 8C).

    5. Plates should be used within seven days after preparation unless

    adequate precautions, such as wrapping in plastic, have been

    taken to minimize drying of the agar.

    6. A representative sample of each batch of plates should be

    examined for sterility by incubating at 30 to 35C for 24 hours orlonger.

    Preparation of antibiotic stock solutions

    Antibitiotics may be received as powders or tablets. It is recommended

    to obtain pure antibiotics from commercial sources, and not use

    injectable solutions. Powders must be accurately weighed and dissolved

    in the appropriate diluents (Annexure III) to yield the required

    concentration, using sterile glassware. Standard strains of stock cultures

    should be used to evaluate the antibiotic stock solution. If satisfactory,

    the stock can be aliquoted in 5 ml volumes and frozen at -20C or -60C.

    5

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    6/26

    Stock solutions are prepared using the formula (1000/P) X V X C=W,

    where P+potency of the anitbiotic base, V=volume in ml required,

    C=final concentration of solution and W=weight of the antimicrobial to

    be dissolved in V.

    Preparation of dried filter paper discs

    Whatman filter paper no. 1 is used to prepare discs approximately 6 mm

    in diameter, which are placed in a Petri dish and sterilized in a hot air

    oven.

    The loop used for delivering the antibiotics is made of 20 gauge wire and

    has a diameter of 2 mm. This delivers 0.005 ml of antibiotics to each

    disc.

    Storage of commercial antimicrobial discs

    Cartridges containing commercially prepared paper disks specifically forsusceptibility testing are generally packaged to ensure appropriate

    anhydrous conditions. Discs should be stored as follows:* Refrigerate the containers at 8C or below, or freeze at -14C orbelow, in a nonfrost-free freezer until needed. Sealed packages ofdisks that contain drugs from the -lactam class should be storedfrozen, except for a small working supply, which may berefrigerated for at most one week. Some labile agents (e.g.,imipenem, cefaclor, and clavulanic acid combinations) may retaingreater stability if stored frozen until the day of use.

    * The unopened disc containers should be removed from therefrigerator or freezer one to two hours before use, so they mayequilibrate to room temperature before opening. This procedure

    minimizes the amount of condensation that occurs when warm aircontacts cold disks.

    * Once a cartridge of discs has been removed from its sealedpackage, it should be placed in a tightly sealed, desiccatedcontainer. When using a disc-dispensing apparatus, it should befitted with a tight cover and supplied with an adequate desiccant.The dispenser should be allowed to warm to room temperaturebefore opening. Excessive moisture should be avoided byreplacing the desiccant when the indicator changes color.

    * When not in use, the dispensing apparatus containing the discsshould always be refrigerated.

    * Only those discs that have not reached the manufacturer'sexpiration date stated on the label may be used. Discs should bediscarded on the expiration date.

    Turbidity standard for inoculum preparationTo standardize the inoculum density for a susceptibility test, a BaSO4turbidity standard, equivalent to a 0.5 McFarland standard or its opticalequivalent (e.g., latex particle suspension), should be used. A BaSO4 0.5McFarland standard may be prepared as follows:

    6

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    7/26

    1. A 0.5-ml aliquot of 0.048 mol/L BaCl2 (1.175% w/v BaCl2 . 2H2O) isadded to 99.5 ml of 0.18 mol/L H2SO4 (1% v/v) with constantstirring to maintain a suspension.

    2. The correct density of the turbidity standard should be verified byusing a spectrophotometer with a 1-cm light path and matched

    cuvette to determine the absorbance. The absorbance at 625 nmshould be 0.008 to 0.10 for the 0.5 McFarland standard.3. The Barium Sulfate suspension should be transferred in 4 to 6 ml

    aliquots into screw-cap tubes of the same size as those used ingrowing or diluting the bacterial inoculum.

    4. These tubes should be tightly sealed and stored in the dark atroom temperature.

    5. The barium sulfate turbidity standard should be vigorouslyagitated on a mechanical vortex mixer before each use andinspected for a uniformly turbid appearance. If large particles

    appear, the standard should be replaced. Latex particlesuspensions should be mixed by inverting gently, not on a vortexmixer

    6. The barium sulfate standards should be replaced or their densitiesverified monthly.

    DISC DIFFUSION METHODS The Kirby-Bauer and Stokes' methods are usually used forantimicrobial susceptibility testing, with the Kirby-Bauer method beingrecommended by the NCCLS. The accuracy and reproducibility of thistest are dependent on maintaining a standard set of procedures asdescribed here.

    NCCLS is an international, interdisciplinary, non-profit, non-governmental organization composed of medical professionals,government, industry, healthcare providers, educators etc. It promotesaccurate antimicrobial susceptibility testing (AST) and appropriatereporting by developing standard reference methods, interpretativecriteria for the results of standard AST methods, establishing qualitycontrol parameters for standard test methods, provides testing andreporting strategies that are clinically relevant and cost-effectiveInterpretative criteria of NCCLS are developed based on internationalcollaborative studies and well correlated with MICs and the resultshave corroborated with clinical data. Based on study results NCCLS

    interpretative criteria are revised frequently. NCCLS is approved byFDA-USA and recommended by WHO.Procedure for Performing the Disc Diffusion Test

    Inoculum Preparation

    Growth Method

    1. At least three to five well-isolated colonies of the same

    morphological type are selected from an agar plate culture. The

    7

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    8/26

    top of each colony is touched with a loop, and the growth is

    transferred into a tube containing 4 to 5 ml of a suitable broth

    medium, such as tryptic soy broth.

    2. The broth culture is incubated at 35C until it achieves or exceedsthe turbidity of the 0.5 McFarland standard (usually 2 to 6 hours)

    3. The turbidity of the actively growing broth culture is adjusted with

    sterile saline or broth to obtain a turbidity optically comparable to

    that of the 0.5 McFarland standard. This results in a suspension

    containing approximately 1 to 2 x 108 CFU/ml for E.coli ATCC

    25922. To perform this step properly, either a photometric device

    can be used or, if done visually, adequate light is needed to

    visually compare the inoculum tube and the 0.5 McFarland

    standard against a card with a white background and contrasting

    black lines.

    Direct Colony Suspension Method

    1. As a convenient alternative to the growth method, the inoculum

    can be prepared by making a direct broth or saline suspension of

    isolated colonies selected from a 18- to 24-hour agar plate (a

    nonselective medium, such as blood agar, should be used). The

    suspension is adjusted to match the 0.5 McFarland turbidity

    standard, using saline and a vortex mixer.

    2. This approach is the recommended method for testing the

    fastidious organisms, Haemophilus spp., N. gonorrhoeae, and

    streptococci, and for testing staphylococci for potential methicillin

    or oxacillin resistance.

    Inoculation of Test Plates

    1. Optimally, within 15 minutes after adjusting the turbidity of the

    inoculum suspension, a sterile cotton swab is dipped into the

    adjusted suspension. The swab should be rotated several times

    and pressed firmly on the inside wall of the tube above the fluid

    level. This will remove excess inoculum from the swab.

    2. The dried surface of a Meller-Hinton agar plate is inoculated bystreaking the swab over the entire sterile agar surface. This

    procedure is repeated by streaking two more times, rotating the

    plate approximately 60 each time to ensure an even distributionof inoculum. As a final step, the rim of the agar is swabbed.

    8

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    9/26

    3. The lid may be left ajar for 3 to 5 minutes, but no more than 15

    minutes, to allow for any excess surface moisture to be absorbed

    before applying the drug impregnated disks.

    NOTE: Extremes in inoculum density must be avoided. Never use

    undiluted overnight broth cultures or other unstandardized inocula for

    streaking plates.

    Application of Discs to Inoculated Agar Plates

    1. The predetermined battery of antimicrobial discs is dispensed onto

    the surface of the inoculated agar plate. Each disc must be

    pressed down to ensure complete contact with the agar surface.

    Whether the discs are placed individually or with a dispensing

    apparatus, they must be distributed evenly so that they are no

    closer than 24 mm from center to center. Ordinarily, no more than12 discs should be placed on one 150 mm plate or more than 5

    discs on a 100 mm plate. Because some of the drug diffuses

    almost instantaneously, a disc should not be relocated once it has

    come into contact with the agar surface. Instead, place a new disc

    in another location on the agar.

    2. The plates are inverted and placed in an incubator set to 35Cwithin 15 minutes after the discs are applied. With the exception

    ofHaemophilus spp., streptococci and

    N. gonorrhoeae, the plates should not be incubated in an

    increased CO2 atmosphere, because the interpretive standards

    were developed by using ambient air incubation, and CO2 will

    significantly alter the size of the inhibitory zones of some agents.

    Reading Plates and Interpreting Results

    1. After 16 to 18 hours of incubation, each plate is examined. If the

    plate was satisfactorily streaked, and the inoculum was correct,

    the resulting zones of inhibition will be uniformly circular and there

    will be a confluent lawn of growth. If individual colonies areapparent, the inoculum was too light and the test must be

    repeated. The diameters of the zones of complete inhibition (as

    judged by the unaided eye) are measured, including the diameter

    of the disc. Zones are measured to the nearest whole millimeter,

    using sliding calipers or a ruler, which is held on the back of the

    inverted petri plate. The petri plate is held a few inches above a

    9

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    10/26

    black, nonreflecting background and illuminated with reflected

    light. If blood was added to the agar base (as with streptococci),

    the zones are measured from the upper surface of the agar

    illuminated with reflected light, with the cover removed. If the test

    organism is a Staphylococcus or Enterococcus spp., 24 hours of

    incubation are required for vancomycin and oxacillin, but other

    agents can be read at 16 to 18 hours. Transmitted light (plate

    held up to light) is used to examine the oxacillin and vancomycin

    zones for light growth of methicillin- or vancomycin- resistant

    colonies, respectively, within apparent zones of inhibition. Any

    discernable growth within zone of inhibition is indicative of

    methicillin or vancomycin resistance.

    2. The zone margin should be taken as the area showing no obvious,

    visible growth that can be detected with the unaided eye. Faintgrowth of tiny colonies, which can be detected only with a

    magnifying lens at the edge of the zone of inhibited growth, is

    ignored. However, discrete colonies growing within a clear zone of

    inhibition should be subcultured, re-identified, and retested.

    Strains ofProteus spp. may swarm into areas of inhibited

    growth around certain antimicrobial agents. With Proteus spp.,

    the thin veil of swarming growth in an otherwise obvious zone of

    inhibition should be ignored. When using blood-supplemented

    medium for testing streptococci, the zone of growth inhibition

    should be measured, not the zone of inhibition of hemolysis. With

    trimethoprim and the sulfonamides, antagonists in the medium

    may allow some slight growth; therefore, disregard slight growth

    (20% or less of the lawn of growth), and measure the more

    obvious margin to determine the zone diameter.

    3. The sizes of the zones of inhibition are interpreted by referring to

    Tables 2A through 2I (Zone Diameter Interpretative Standards and

    equivalent Minimum Inhibitory Concentration Breakpoints) of the

    NCCLS M100-S12: Performance Standards for AntimicrobialSusceptibility Testing: Twelfth Informational Supplement, and the

    organisms are reported as either susceptible, intermediate, or

    resistant to the agents that have been tested. Some agents may

    only be reported as susceptible, since only susceptible breakpoints

    are given.

    10

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    11/26

    DILUTION METHODS

    Dilution susceptibility testing methods are used to determine theminimal concentration of antimicrobial to inhibit or kill themicroorganism. This can be achieved by dilution of antimicrobial ineither agar or broth media. Antimicrobials are tested in log2 serialdilutions (two fold).

    Minimum Inhibitory Concentration (MIC)Diffusion tests widely used to determine the susceptibility of organismsisolated from clinical specimens have their limitations; when equivocalresults are obtained or in prolonged serious infection e.g. bacterial

    endocarditis, the quantitation of antibiotic action vis-a-vis thepathogen needs to be more precise. Also the terms Susceptible andResistant can have a realistic interpretation. Thus when in doubt, theway to a precise assessment is to determine the MIC of the antibioticto the organisms concerned.There are two methods of testing for MIC:(a) Broth dilution method(b) Agar dilution method.

    Broth Dilution Method The Broth Dilution method is a simpleprocedure for testing a small number of isolates, even single

    isolate. It has the added advantage that the same tubes can betaken for MBC tests also:

    MaterialsSterile graduated pipettes of 10ml, 5ml, 2ml and 1ml Sterile capped7.5 x 1.3 cm tubes / small screw-capped bottles, Pasteur pipettes,overnight broth culture of test and control organisms ( same as for discdiffusion tests), required antibiotic in powder form (either from themanufacturer or standard laboratory accompanied by a statement ofits activity in mg/unit or per ml. Clinical preparations should not beused for reference technique.), required solvent for the

    antibiotic, sterile Distilled Water - 500ml and suitable nutrient brothmedium.Trimethoprim and sulphonamide testing requires thymidine free mediaor addition of 4% lysed horse blood to the mediaA suitable rack to hold 22 tubes in two rows i-e 11 tubes in each row.

    11

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    12/26

    Stock solutionStock solution can be prepared using the formula

    1000------- x V x C= WP

    Where P=Potency given by the manufacturer in relation to the baseV= Volume in ml required; C=Final concentration of solution (multiplesof 1000)W= Weight of the antimicrobial to be dissolved in the volume VExample: For making 10 ml solution of the strength 10,000mg/l frompowder base whose potency is 980 mg per gram,the quantities of theantimicrobials required is

    W = 1000------- x 10 x 10=102.04mg

    980

    Note:the stock solutions are made in higher concentrations to maintaintheir keeping qualities and stored in suitable aliquots at -20oC .Oncetaken out,they should not be refrozen or reused.MethodPrepare stock dilutions of the antibiotic of concentrations 1000 and100 g/L as required from original stock solution (10,000mg/L).Arrange two rows of 12 sterile 7.5 x1.3 cm capped tubes in the rack. Ina sterile 30ml (universal) screw capped bottle, prepare 8ml of brothcontaining the concentration of antibiotic required for the first tube ineach row from the appropriate stock solution already made. Mix thecontents of the universal bottle using a pipette and transfer 2ml to the

    first tube in each row. Using a fresh pipette ,add 4 ml of broth to theremaining 4 ml in the universal bottle mix and transfer 2ml to thesecond tube in each row. Continue preparing dilutions in this way butwhere as many as 10 or more are required the series should be startedagain half the way down. Place 2ml of antibiotic free broth to the lasttube in each row. Inoculate one row with one drop of an overnightbroth culture of the test organism diluted approximately to 1 in 1000 ina suitable broth and the second row with the control organism ofknown sensitivity similarly diluted. The result of the test is significantlyaffected by the size of the inoculum.The test mixture should contain106 organism/ml.If the broth culture used has grown poorly,it may be

    necessary to use this undiluted. Incubate tubes for 18 hours at 37oC.Inoculate a tube containing 2ml broth with the organism and keep at+4oC in a refrigerator overnight to be used as standard for thedetermination of complete inhibition.Calculations for the preparation of the original dilution.

    This often presents problems to those unaccustomed to

    performing these tests. The following method advocated by Pamela M

    12

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    13/26

    Waterworth is presented. Calculate the total volume required for the

    first dilution. Two sets of dilution are being prepared (one for the test

    and one for the control), each in 2ml volumes i-e a total of 4 ml for

    each concentration as 4ml is required to make the second dilution, the

    total requirement is 8ml. Now calculate the total amount of the

    antibiotic required for 8ml. For 64 g/l concentration, 8x64mg/l =512g

    in 8 ml. Place a decimal point after the first figure (5.12) and take this

    volume in ml (i.e 5.12 ml) of the dilution below 512mg/l and make upto

    8ml with broth. In this example given above, the series has to be

    started again mid way at 2 mg/l which would be obtained in the same

    way:

    8ml of 2mg/l=16g in 8ml.1.6 ml of 10 mg/ l + 6.4 ml of broth.

    READING OF RESULTMIC is expressed as the lowest dilution, which inhibited growth judgedby lack of turbidity in the tube.Because very faint turbidity may begiven by the inoculum itself, the inoculated tube kept in therefrigerator overnight may be used as the standard for thedetermination of complete inhibition.Standard strain of known MICvalue run with the test is used as the control to check the reagents andconditions.

    Micro-broth dilution test This test uses double-strength Meller-Hinton broth, 4X strengthantibiotic solutions prepared as serial two-fold dilutions and the test

    organism at a concentration of 2x106/ml. In a 96 well plate, 100 l ofdouble-strength MHB, 50 l each of the antibiotic dilutions and theorganism suspension are mixed and incubated at 35C for 18-24 hours.The lowest concentration showing inhibition of growth will be consideredthe MIC of the organism.Reading of resultMIC is expressed as the highest dilution which inhibited growth judgedby lack of turbidity in the tube. Because very faint turbidity may begiven by the inoculum itself,the inoculated tube kept in the refrigeratorovernight may be used as the standard for the determination of

    complete inhibition. Standard strain of known MIC, run with the test isused as the control to check the reagents and conditions.

    The Agar dilution MethodAgar dilutions are most often prepared in petri dishes and haveadvantage that it is possible to test several organisms on each plate.If only one organism is to be tested e.g M.tuberculosis,the dilutionscan be prepared in agar slopes but it will then be necessary to preparea second identical set to be inoculated with the control organism.The

    13

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    14/26

    dilutions are made in a small volume of water and added to agar whichhas been melted and cooled to not more than 60oC.Blood may beadded and if chocolate agar is required,the medium must be heatedbefore the antibiotic is added.It would be convenient to use 90 mm diameter petri dishes and add

    one ml of desired drug dilutions to 19 ml of broth.The factor of agardilution must be allowed for in the first calculation as follows.final volume of medium in plate = 20 ml Top antibiotic concentrations = 64mg/l Total amount of drug = 1280g to beadded to 1 ml ofwater2ml of 1280 g /ml will be required to start the dilution = 2560gin 2 ml

    = 1.28ml of 2000g /ml

    0.72 ml of water.1 ml of this will be added to 19 ml agar.

    (Note stock dilution of 2000g /ml is required for this range of MIC)

    Reading of results: The antibiotic concentration of the first plate

    showing 99% inhibition is taken as the MIC for the organism.

    DILUTION AND DIFFUSION

    14

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    15/26

    E test also known as the epsilometer test is an exponential gradienttesting methodology where E in E test refers to the Greek

    symbol epsilon ( ).The E test(AB Biodisk) which is a quantitativemethod for antimicrobial susceptibility testing applies both the

    dilution of antibiotic and diffusion of antibiotic into the medium.. Apredefined stable antimicrobial gradient is present on a thin inertcarrier strip. When this E test strip is applied onto an inoculatedagar plate, there is an immediate release of the drug. Followingincubation , a symmetrical inhibition ellipse is produced. Theintersection of the inhibitory zone edge and the calibrated carrierstrip indicates the MIC value over a wide concentration range(>10 dilutions) with inherent precision and accuracy .

    E test can be used to determine MIC for fastidious organismslike S. pneumoniae,

    -hemolytic streptococci, N.gonorrhoeae, Haemophilus sp. andanaerobes. It can also be used for Nonfermenting Gram Negativebacilli (NFGNB) for eg-Pseudomonas sp. and Burkholderiapseudomallei.

    Resistance of major consequence may be detected for e.g., the test isvery useful in detecting glycopeptide resistant Enterococci (GRE) andglycopeptide intermediate S.aureus (GISA) and slow growingpathogens such as Mycobacterium tuberculosis. Further it can be usedfor detection of extended spectrum beta lactamases (ESBL). Inconclusion E test is a simple, accurate and reliable method to

    determine the MIC for a wide spectrum of infectious agents.

    15

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    16/26

    Ex No: 3

    ESTIMATION OF SULPHONAMIDES IN BIOLOGICAL FLUIDS

    Aim: To estimate sulphonamide concentration in plasma

    Principle: Sulphonamides react with nitrite in acid medium to formdiazonium salts. The excess nitrite is destroyed by ammoniumsulphamate and the diazonium salts couples with N-(1-naphthyl)ethylene diamine to form a stable dye (purple colour) and thecorresponding concentration can be quantifiedspectrophotometrically(545 nm).Requirements:Drugs/chemicals:Trichloroaceticacid(TCA,15%sol.),Sodium nitrite (0.05%),Ammonium sulphamate(0.5%),N-1-naphthyl-ehhylene-diamine dihydrochloride(0.5% sol.), Hydrochloric acid (4.0N),Sulphonamide(10% sol.)

    Instruments required: Spectrophotometer, Water bathProcedure: (Bratton & Marshal method, 1939)(i) Quantification of free- sulphonamide:(a) Sample:1. Take one ml of sample (plasma/ urine) in a test tube and add 14 mlof distilled water and 5ml of TCA(15%). Mix the contents thoroughly.2. Keep the tube(s) at room temperature for 10min. for completedeproteinization. Filter the contents using Whatman filter paper # 1 toget the protein free filtrate.3. Take 5 ml of protein free filtrate in a test tube and add 0.5 ml ofsodium nitrite (0.05%). Thoroughly mix the contents and wait for 5

    min.4. Add 0.5 ml of ammonium sulphamate(0.5%). Mix thoroughly andwait for 2 min.5. Add 0.5 ml of N-1-naphthyl-ehhylene-diamine dihydrochloride(0.5%sol.) and shake the tubes thoroughly and wait for 5 min.6. Measure the optical density (O.D) spectrophotometrically at 545nmagainst the blank within 30min of development of colour.(b) Blank:

    Prepare blank as described above except that the sample isreplaced by 1ml of distilled water.(c) Standard: Prepare working standard solutions of sulphonamide

    to be estimated from the stock solution (10%) to get aconcentration in the range of 2-50 g/ml. Take one ml of thesesolutions and add 14ml of distilled water and 5ml TCA (15%). Addother reagents as described above, but the filtration step may beomitted.

    (ii) Quantification of total- sulphonamide: This includes free andacetylated sulphonamide present in the blood1. Make protein free filtrate as described above (step 1& 2)

    16

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    17/26

    2. Take 5ml of protein free filtrate into a graduated glass tube and add0.5ml of 4NHCl

    3. Cover the tube with a marble and place it in a water bath at 100 Cfor 1hour.4. Cool the tube at room temperature and make up the volume to 5ml

    by adding dist.water5. Follow the same procedure as detailed in the case of freesulphonamide

    (iii) acetyl-sulphonamide:he difference between the concentrationof free and total sulphonamide gives the concentration -acetylcompound.

    Calculation: Plot the standard curve (sulphonamide concentrationv/s O.D) and the read the values of sulphonamide present in thesamples.

    Note: Strictly follow the time schedule as mentioned in the procedure, The colour development is sensitive to sunlight, therefore wrap thetest tubes with black paper & switch off light

    17

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    18/26

    Ex No: 4

    ESTIMATION OF OXYTETRACYCLINE IN BIOLOGICAL FLUIDS

    Aim: To estimate the concentration of Oxytetracycline in plasma

    Requirements:Drugs/ chemicals:Trichloroaceticacid(TCA,15%sol.), Para-nitroaniline ,

    Conc. HCl,Glacial acetic acid,Sodiun nitrite (15%) freshly preparedOxytetracycline standardNote: Para-nitroaniline colour reagent can be prepared by mixing 0.2gp-nitroaniline +

    3ml con. HCl + 5ml glacial acetic acid + 2ml 15% sodium nitriteInstruments required: Spectrophotometer

    Procedure:( Snell and Snell, 1971)(a) Sample:1. Take I ml of sample (plasma) and add 2ml of 15% TCA. Mixthoroughly and centrifuge at 4000rpm for 15 minutes.2. To 2ml of supernatant, add 0.2ml of p-nitroaniline colour reagent

    and 2.0 ml of glacial acetic acid. Mix them and incubate at 65C for45min.3. Measure the resultant colour intensity at 435nm in thespectrophotometer against blank4. Derive the concentration of Oxytetracycline from the standardcurve.

    (b) Blank: Prepare the blank in plasma and run along with the sample.

    (c) Standard: Prepare the working standard solutions from the stocksolution of Oxytetracycline having concentrations in the range of 5-

    200 g/ml in plasma and run along with the sample.Note: The limit of detection of Oxytetracycline is 0.6 g/ml of

    plasma

    18

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    19/26

    Ex No: 5

    Introduction to High performance liquid chromatography

    (HPLC)

    High-performance liquid chromatography [HPLC] is now one of

    the most powerful tools in analytical chemistry. It has the ability to

    separate, identify, and quantitate the compounds that are present in

    any sample that can be dissolved in a liquid. Today, compounds in

    trace concentrations as low as parts per trillion [ppt] may easily be

    identified. HPLC can be, and has been, applied to just about any

    sample, such as pharmaceuticals, food, nutraceuticals, cosmetics,

    environmental matrices, forensic samples, and industrial chemicals.

    The components of a basic high-performance liquid chromatography

    [HPLC] system are shown in the simple diagram in Figure E.

    19

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    20/26

    High performance (pressure) liquid chromatography (HPLC) is an

    efficient technique to separate, characterise and quantify organic

    molecules in biological fluids or drug formulations. The essential

    principle of HPLC is the partitioning (or distribution) of substances

    between two liquid phases otherwise called partition chromatography.

    In gas-liquid chromatography (GLC) high temperature are used to

    volatilise the compounds so that separation can be done, however

    elevated temperature may cause thermal degradation of some

    compounds/drugs.

    HPLC analysis of drugs is non-destructive in nature since: a.

    there is no need for high temperature and b. the mobile phase is liquid

    instead of gas (in GLC) and solubility is the limiting factor and not

    volatility. Therefore, appropriate solvent (i.e. inert to the compound)

    can be chosen so that no chemical changes will occur to the compound

    /or its metabolite of our interest during separation.The major setback of the HPLC is sensitivity, but this limitation can beovercome by adopting good sample clean-up procedures. Thesensitivity of HPLC is generally in the nanogram range but in few casesit can measure picogram level

    The various components of HPLC are: a. Pump b. Injection

    system c. Column stationary phase) d. Mobile phase and e. Detector

    Pump: it is used to deliver the solvent mobile phase through the

    column. There are two types of column: a. Isocratic pump- means if

    one or fixed solvent (mobile phase) at a set rate is used b. Gradient

    pump- means two or more solvents in varying concentrations at adifferent flow rate are used (eg. in peptide purification).

    Injector: it is to introduce sample into the column by syringe through

    a spectrum of injection system (rheodune injector)

    Column: it is the basic unit for stationary phase and it can be

    described under three headings.

    a. Column dimension or size- column are tubular structure and

    usually made with stainless steel with internal diameter x length of:

    4.6mm x 250mm. The size may vary if (600-1000mm in length) you

    are performing a large scale preparative chromatography or shorter

    (50 or 30mm length) if you want to do fast chromatography at higher

    flow rate than normal speed.

    b. Column particle and pore size- most column are packed with

    inert hydrocarbon usually spherical shaped silica beads/particles of 3-

    50m size, with 5m particles are most popular for peptides. Larger

    particles will generate less pressure while smaller particles create high

    20

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    21/26

    system pressure as well as higher separation efficiency. The pore size

    of these silica particles is expressed as A and generally range

    between 100-1000 A. A pore size of 100A is suitable for most of the

    small molecules while 300A is ideal for proteins or peptides.

    Stationery phase- it is generally made up of hydrophobic alkyl

    chains (-CH2-CH2-CH2-CH3) that interacts with the analyte(drug

    molecules) and there are three chain lengths. C8and C18(octadecyl

    silane, ODS) is used for smaller hydrophilic peptides and drug

    molecules respectively. C8 is used for normal phase while C18(non-

    polar) is used for RPHPLC. C4 column is generally used for proteins

    since larger protein molecules are likely to have more hydrophobic

    moieties to interact with the column.

    C18, 4.6 x 250mm, 5m, 300A

    chain length column size particle size pore

    size

    Eg. Column specifications:

    Mobile phase: The mobile phase is used to control retention and

    selectivity. There are two types of HPLC assay.

    A. Normal phase: Here the stationary phase is polar in

    physiochemical terms and the retention of molecules is governed by

    polar interaction between stationary and mobile base. In normal phase

    HPLC the most popular mobile phase is either hexane or heptane to

    which varying concentrations of more polar solvents like propanol or

    ethyl acetate are added. The mobile phase containing proton donor

    acceptors will interact with basic solutes. Conversely, mobile phases

    containing proton donar acceptors will interact strongly with acidic

    solvents. When excessive interaction occurs between solute and the

    mobile phase peak tailing can result. Addition of basic modifier

    (triethylamine) can overcome such strong interactions and thereby

    improve peak shape. Normally increase in temperature results in

    decrease in retention and this effect dependent on the specific polarmodifier in the solvent, for example ethanol increases retension.

    B. Reverse phase (RPHPLC): Here mobile phase is more polar than

    stationary phase (non-polar) and compounds are separated based on

    their hydrophobic character. Normally, isocratic pump is the choice

    when ever RPHPLC is employed.

    21

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    22/26

    The reverse phase HPLC solvents are by convention installed in

    the system channels A and B. The solvent A is the aqueous solvent

    (HPLC grade water + 0.1% v/v acetic acid) while solvent B is the

    organic solvent (HPLC grade acetonitrile, methanol or proponol + 0.1%

    v/v acetic acid). The acid is added usually to improve the

    chromatography peak shape and serve as a source protons in reverse

    phase HPLC or LC/MS. The acids most commonly used are formic acid,

    acetic acid or triflouroacetic acid. Usually, 0.1 %v/v solution is made by

    adding 1ml/liter of solvent.

    Quality of solvents: Solvents should be chosen to be compatible

    with the column employed and the solubility and chemistry of the

    analyte (drug). While formulating mixed solvents, they should be fully

    miscible with each other; otherwise droplets of a second phase may be

    trapped in the column and ruin separation. None of the mobile phase

    constituents should interact with stationary phase and affect itadversely.Always one should use HPLC grade.

    Sample cleanup: Biological samples may contain proteins, salts

    and a host of organic compounds with widely differing polarities.

    Pharmaceutical samples also contain a wide range of soluble and

    insoluble excipients. The sample clean-up procedures can be designed

    tentatively on the basis of extraction and polarity profiles.

    Conventionally, solvent extraction is most commonly used to

    extract organic molecules from biological fluids or tissues. If the

    compound(s) are extracted in polar solvents then reverse phase HPLC

    system would be the logical choice.

    sample + Solvent to precipitate protein

    Centrifuge and decant

    supernatant

    (Repeat twice /thrice for maximum

    extraction)

    Evaporate the solvent

    Reconstitute with mobile phase

    Inject 10/20l to HPLC

    Fig. Steps involved in solvent

    extraction

    22

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    23/26

    Stationary Phase

    The stationary phase in HPLC refers to the solid support contained

    within the column over which the mobile phase continuously flows. The

    sample solution is injected into the mobile phase of the assay through

    the injector port. As the sample solution flows with the mobile phase

    through the stationary phase, the components of that solution will

    migrate according to the non-covalent interactions of the compounds

    with the stationary phase. The chemical interactions of the stationary

    phase and the sample with the mobile phase, determines the degree of

    migration and separation of the components contained in the sample.

    For example, those samples which have stronger interactions with the

    stationary phase than with the mobile phase will elute from the column

    less quickly, and thus have a longer retention time, while the reverse isalso true. Columns containing various types of stationary phases are

    commercially available. Some of the more common stationary phases

    include: Liquid-Liquid, Liquid-Solid (Adsorption), Size Exclusion, Normal

    Phase, Reverse Phase, Ion Exchange, and Affinity.

    Liquid-Solid operates on the basis of polarity. Compounds that

    possess functional groups cabable of strong hydrogen bonding will

    adhere more tightly to the stationary phase than less polar

    compoounds. Thus, less polar compounds will elute from the column

    faster than compounds that are highly polar.

    Liquid-Liquid operates on the same basis as liquid-solid. However,

    this technique is better suited for samples of medium polarity that are

    soluble in weakly polar to polar organic solvents. The separation of

    non-electrolytes is achieved by matching the polarities of the sample

    and the stationary phase and using a mobile phase which possesses a

    markedly different polarity.

    Normal Phase operates on the basis of hydrophilicity and lipophilicity

    by using a polar stationary phase and a less polar mobile phase. Thus

    hydrophobic compounds elute more quickly than do hydrophilic

    compounds.

    Reverse Phase operates on the basis of hydrophilicity and

    lipophilicity. The stationary phase consists of silica based packings with

    n-alkyl chains covalently bound. For example, C-8 signifies an octyl

    23

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    24/26

    chain and C-18 an octadecyl ligand in the matrix. The more

    hydrophobic the matrix on each ligand, the greater is the tendancy of

    the column to retain hydrophobic moieties. Thus hydrophilic

    compounds elute more quickly than do hydrophobic compounds.

    HPLC PUMPSThee several types of pumps available for use with HPLC

    analysis are: Reciprocating Piston Pumps, Syringe Type Pumps, and

    Constant Pressure Pumps.

    Reciprocating Piston Pumps consist of a small motor driven piston

    which moves rapidly back and forth in a hydraulic chamber that may

    vary from 35-400 L in volume. On the back stroke, the separation

    column valve is closed, and the piston pulls in solvent from the mobile

    phase reservoir. On the forward stroke, the pump pushes solvent out

    to the column from the reservoir. A wide range of flow rates can beattained by altering the piston stroke volume during each cycle, or by

    altering the stroke frequency. Dual and triple head pumps consist of

    identical piston-chamber units which operate at 180 or 120 degrees

    out of phase. This type of pump system is significantly smoother

    because one pump is filling while the other is in the delivery cycle.

    Syringe Type Pumps are most suitable for small bore columns

    because this pump delivers only a finite volume of mobile phase before

    it has to be refilled. These pumps have a volume between 250 to 500

    mL. The pump operates by a motorized lead screw that delivers mobilephase to the column at a constant rate. The rate of solvent delivery is

    controlled by changing the voltage on the motor.

    Constant Pressure Pumps- the mobile phase is driven through the

    column with the use of pressure from a gas cylinder. A low-pressure

    gas source is needed to generate high liquid pressures. The valving

    arrangement allows the rapid refill of the solvent chamber whose

    capacity is about 70 mL. This provides continuous mobile phase flow

    rates.

    MOBILE PHASE:The mobile phase in HPLC refers to the solvent being

    continuously applied to the column, or stationary phase. The mobile

    phase acts as a carrier for the sample solution. A sample solution is

    injected into the mobile phase of an assay through the injector port. As

    a sample solution flows through a column with the mobile phase, the

    components of that solution migrate according to the non-covalent

    24

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    25/26

    interactions of the compound with the column. The chemical

    interactions of the mobile phase and sample, with the column,

    determine the degree of migration and separation of components

    contained in the sample. For example, those samples which have

    stronger interactions with the mobile phase than with the stationary

    phase will elute from the column faster, and thus have a shorter

    retention time, while the reverse is also true.

    COLUMNS The are various columns that are secondary to the

    separating column or stationary phase. are: Guard, Derivatizing,

    Capillary, Fast, and Preparatory Columns.Guard Columns are placed

    anterior to the separating column. This serves as a protective factor

    THE DETECTOR

    The detector for an HPLC is the component that emits a response dueto the eluting sample compound and subsequently signals a peak on

    the chromatogram. It is positioned immediately posterior to the

    stationary phase in order to detect the compounds as they elute from

    the column. The bandwidth and height of the peaks may usually be

    adjusted using the coarse and fine tuning controls, and the detection

    and sensitivity parameters may also be controlled (in most cases).

    There are many types of detectors that can be used with HPLC. Some

    of the more common detectors include: Refractive Index (RI), Ultra-

    Violet (UV), Fluorescent, Radiochemical, Electrochemical, Near-Infra

    Red (Near-IR), Mass Spectroscopy (MS), Nuclear Magnetic Resonance(NMR), and Light Scattering (LS).

    Ultra-Violet (UV) detectors measure the ability of a sample to absorb

    light. This can be accomplished at one or several wavelengths:

    Fluorescent detectors measure the ability of a compound to absorb

    then re-emit light at given wavelengths. Each compound has a

    characteristic fluorescence.Radiochemical detection involves the use

    of radiolabeled material, usually tritium (3H) or carbon-14 (14C). It

    operates by detection of fluorescence associated with beta-particle

    ionization, and it is most popular in metabolite research. Two detector

    typesomogeneous- Where addition of scintillation fluid to column

    effluent causes fluorescence; Heterogeneous- Where lithium silicate

    and fluorescence caused by beta-particle emission interact with the

    detector cell

    25

  • 8/3/2019 VPT 607 Veterinary Chemotherapy Lab Manual

    26/26

    INJECTORS : Samples are injected into the HPLC via an injection

    port. The injection port of an HPLC commonly consists of an injection

    valve and the sample loop. The sample is typically dissolved in the

    mobile phase before injection into the sample loop. The sample is then

    drawn into a syringe and injected into the loop via the injection valve.

    A rotation of the valve rotor closes the valve and opens the loop in

    order to inject the sample into the stream of the mobile phase. Loop

    volumes can range between 10 l to over 500 l. In modern HPLC

    systems, the sample injection is typically automated.

    Stopped-flow Injection is a method whereby the pump is turned off

    allowing the injecion port to attain atmospheric pressure. The syringe

    containing the sample is then injected into the valve in the usual

    manner, and the pump is turned on. For syringe type and reciprocation

    pumps, flow in the column can be brought to zero and rapidly resumedby diverting the mobile phase by means of a three-way valve placed in

    front of the injector. This method can be used up to very high

    26