Voluson E8 Expert BT13Data Sheet
Product descriptionThe Voluson* E8 Expert BT13 is an advanced imaging platform that combines extraordinary image quality with our superb volume ultrasound technology.
GE Healthcare
Highlights• HighResolutionFlatPanelDisplay19inches
• Advanced4D
• DynamicRenderEngine2ndGeneration
• HDlive*
• AdvancedVCI
• SonoBiometry†
• SonoAVC*follicle
• SonoVCAD*heart• SonoVCAD*labor• AdvancedSTIC
• AdvancedFetalEcho
• WideSector
• CodedContrastImaging† (not available in all countries)
• ScanAssistant
• SonoNT
• SonoIT†
• Elastography
• ElastographyAnalysis†
• ElastographyRatioMeasurement†
• AnatomicalM-Mode
• B-Flow
• SonoRenderStart
• MatrixArrayVolumeTechnology
• HighPerformanceTransvaginalProbe
• ElectricalHeightAdjustment
• FloatingUserInterface
• OnBoardArchiveincluding PreviewandPre-selection
General specificationsDimensions and weight
Height(maximum) 1393mm(54.8in)
Adjustable withelectronicalmotor
Width 580mm(22.8in)
Depth 920mm(36in)
Weight(noPeripherals) 131kg(289lb)
Power supply
Voltage 100,115-130,220-240VAC
Frequency 50/60Hz(+/-2%)
Power Max.1000VA withon-boardPeripherals
ThermalOutput 3446BTU/h
Console design
3ActiveProbePorts 3plus1non-imagingport
IntegratedHDD 500GB
IntegratedDVD+/-R(W)/CD-R(W)drive
On-boardstorageforPeripherals
Wheels Wheeldiameter150mm
Integratedcablemanagement
Frontandrearhandles
User interfaceOperator keyboard
FloatingKeyboard: Rotation:adjustable+/-38°fromcenter Heightadjustable+195mm
Full-sized,backlitalphanumerickeyboard
Ergonomichardkeylayout
Interactiveback-lighting
Integratedrecordingkeysforremotecontrolofupto4 PeripheralsorDICOMdevices,onededicatedDVDrecordingkey
Touch screen
10.4inHighResolutioncolorLCDscreen
Interactivedynamicsoftwaremenu
Brightnessadjustable
Monitor
19inHighResLCDmonitorwithDVIinterface
ResolutionSXGA1280x1024pixel
Highbrightnesswith220cd/m²
Tilt/RotateAdjustableMonitor TiltAngle:+40°/-90°
Hor.rotateAngle:+/-90°
Digitalbrightnessandcontrastadjustment.OSD,remote controlledbythesystem.Threedefaultsettings:DarkRoom, SemiDarkRoom,BrightRoom
System overviewApplications
Abdominal
Obstetrical
Gynaecological
SmallParts
Vascular
Pediatric
Urology
Cardiology
Neurology
Musculoskeletal(MSK)
Operating modes
BrightnessMode(B-Mode)(2D)
MotionMode–M-Mode(conventionalM-Mode)
AnatomicalM-Mode(AMM)
PulsedwaveDoppler(PW)withHPRF
ContinuousWaveDopplerImaging(CW)
ColorFlowDopplerMode(CFM)
PowerDopplerMode(PD)
HighDefinitionpowerDoppler(HD-Flow*)
TissueDopplerMode(TD)
B-Flow(BF)
Elastography
ContrastAgentMode† (not available in all countries)
Combinationmodes:M/CF,M/HDFlow,M/TDet
VolumeMode(3D/4D): •3DStatic •4DRealTime •VCI-A •VCI-OmniView •Spatio-TemporalImageCorrelation(STIC),STIC/HD-Flow, STIC/Contrast,STIC/B-Flow •4DBiopsy •ExtendedView(XTDView)
System overview (cont.)
Scanning methods
Electronic Sector
ElectronicConvex
ElectronicLinear
MechanicVolumeSweep
Transducer types
SectorArray
ConvexArray
MicroconvexArray
LinearArray
ActiveMatrixConvexArray(1.25,1.5D)
ActiveMatrixLinearArray(1.25,1.5D)
Volumeprobes`4D`: •ConvexArray •MicroconvexArray •ActiveMatrixConvexArray(1.25,1.5D) •ActiveMatrixLinearArray(1.25,1.5D) •LinearArray
PencilProbes(CW)
System standard features
State-of-the-artuserinterfacewithhighresolution10.4in LCDtouchpanel
AutomaticTissueOptimization
CodedHarmonicImagingwithPulseInversionTechnology
CodedExcitation(CE)
HD-Flow
B-Flow
Tissue Doppler
XTD
SRIII(SpeckleReductionImaging)
CrossXBeamCRI*(CompoundResolutionImaging)
SonoNT
SonoIT
SonoBiometry
SonoRenderStart
ScanAssistant
DICOM™
Static3DMode: •BModeonly •B+PowerDopplerMod •B+CFMDopplerMod •B+HD-FlowMod •B+CRI
FocusandFrequencyComposite(FFC)
HighResolutionZoom
PanZoom
Steering
VirtualConvex
WideAngleonendovaginalprobes
BetaView
Patientinformationdatabase
ImageArchiveonharddrive
3D/4Ddatacompression(lossy/lossless)
Inversion
Real-timeautomaticDopplercalcs
MeasurementandCalculationsincludingWorksheets/Reportfor: •OB •GYN •Vascular •Cardio •Abdominal
MultigestationalCalculations
System options
Advanced4D
VOCALII
AdvancedVCI(VolumeContrastImaging)
SonoVCADheart SonoVCADlabor DICOM
Elastography
4D–AdvancedSTIC: •STIC •STIC+PowerDopplerMode •STIC+CFMDopplerMode •STIC+HD-FlowMode •STIC+CRI •STIC+CRI+CFM
SonoAVC
CodedContrastImaging† (not available in all countries)
FootSwitch,withprogrammablefunctionality
Peripheral options
Integratedprinters: •B&Wthermalprinter •Colorthermalprinter
DVDRecorder
ExternalColordesktopprinter&connectionkits •B+CRI+CFM •B+CRI+PD •B+CRI+HD-Flow •B+B-Flow
•SmallParts •Urology •Paediatrics •Neurology
•STIC+CRI+PD •STIC+CRI+HD-Flow •STIC+B-Flow •STICM-Mode •STICflow
Display modes
SimultaneousCapabilityincombinationwithSRIand/orCRI: •B+PW •B+CFM,B+PD,B+TD, B+HD-Flow •B+M,B+AMM •B+3D,B+4D •B+CRI •B+SRI •B+CRI+SRI •Contrast†+SRI •B+CRI/3D+CRI •B+SRI/3D+SRI •B+CRI+SRI/3D+CRI+SRI •B+CRI/4D+CRI •B+SRI/4D+SRI •B+CRI+SRI/4D+CRI+SRI •B+CRI/STIC+CRI
Real-timeTriplexMode: •B/CFM/PW •B/PD/PW •B/HD-Flow/PW
SelectablealternatingModes: •B+PWorCW • B/HD-Flow+PWorCW •B/CFM+PWorCW • B+CFMorPDorHD-Flow •B/PD+PWorCW orCW
Multi-image(split,quad): •Liveand/orfrozen •Split:B+B,B/CFM+B/CFM,orB/PD+B/PDorB/TD+B/TD orB/HD-Flow+B/HD-FloworBF+BF,Contrast†+Contrast† •Splitsimultan:B+B/CFMorB+B/PDorB+B/HD-Flow •Split:B+PWorMorCW •Split:FrameReview/XTD-View •Quad:B+B+B+BorBForContrast†,B/CFM+B/CFM+B/CFM+B/ CFMorB/PDorB/TDorB/HD-Flow •IndependentCineplayback •Quad:A+B+C+3Dor4D •Quad:A+B+C+3Dor4D •TUI:1x1,1x2,2x2,3x2,3x3,3x4,4x4 •Segmentation:quad(A/B/C/Segm.Object),single (Segm.Object) •Split:TUIOverview+1slice
ZoomRead/Write(withorwithoutoverviewimage)
ColorizedImage: •ColorizedB •ColorizedM
Timelinedisplay: •IndependentDualB/PWDisplay •DisplayFormats:Top/Bottomselectableformat (Size:1/2:1/2;1/3:2/3;2/3:1/3)
Display annotation
PatientName: •Last:max32characters •First:max15characters •Middle:max15characters
ID:max32characters
SecondarypatientID(CitizenServiceNumber)
Accession#:max16characters
HospitalName:max30Characters
Sonographer(upto5charactersaredisplayeddepending onfontsize)
Gestationalage(OB)orLMP(GYN)
Birth date
Date:3Typesselectable: •MM/DD/YYYY •DD/MM/YYYY •YYYY/MM/DD
Time:2typesselectable: •24hours •12hours
ProbeName
ApplicationName
Gray Scale bar
Depth Scale
FocalZoneMarker
FrameRate
ZoomStart/Depth
B-Mode: •Userprogram • EdgeEnhance •ReceiverFrequency • Persistence •AcousticPower • SRI,CRI •Gain • FocalZoneMarkers •DynamicControl • DepthScaleMarker •GrayMap • ProbeOrientation
M-Mode/AMM-Mode: •Gain • Reject •DynamicControl • M-Cursor,AMM-Cursor •EdgeEnhance • TimeScale
DopplerMode: •AcousticPower • VelocityorFrequencyScale •Gain • SpectrumInversion •Angle • TimeScale •SampleVolumeDepth • PRF andWidth • HPRF •WallMotionFilter • DopplerFrequency
ColorFlowImagingModes(CFM,PD,TD,HD-Flow): •AcousticPower • ColorMap •ColorGain • ColorScale:kHz,cm/s,m/s •ColorBalance • PowerandSymmetrical •ColorBalanceMarker VelocityImaging •Quality • ColorVelocityRange •WallMotionFilter • SpectrumInversion •PRF
•B+SRI/STIC+SRI •B+CRI+SRI/STIC+CRI+SRI •B/B+CRI •B/B+SRI •B/B+SRI+CRI •B/CFM+CRI •B/CFM+SRI •B/CFM+CRI+SRI •B/PD+CRI •B/PD+SRI •B/PD+CRI+SRI •B/HD-Flow+CRI •B/HD-Flow+SRI •B/HD-Flow+CRI+SRI
•ColorizedPW •Colorized3D
System overview (cont.)
Display annotation (cont.)
3D/4DMode: •3D/4DSubProgram •Threshold •Quality •VolumeBoxAngle •Mix •AcquisitionMode •Compression
ElastographyMode: •AcousticOutput •TxFrequence •Transparency
VelocityRange
TGCCurve
CineFrameNumber
RecorderStatus
BodyPattern:117typesorganizedin10anatomicalgroups
MeasurementResults
DisplayedAcousticOutput: •TIS:ThermalIndexSoftTissue •TIC:ThermalIndexCranial(Bone) •TIB:ThermalIndexBone •MI:MechanicalIndex
Poweroutput
PredefinedBiopsyGuideLine
ECGLine
Trackballfunction(TrackballandTrackballbuttons)
GELogo
Zoomoverviewimage(zoomboxposition)
System parametersSystem setup
Pre-programmableCategoriesdateformat
UserProgrammablePresetCapability,Userprogrametc.
Languages:English,French,German,Spanish,Italian,Danish, Dutch,Finnish,Norwegian,Swedish,Russian,Japanese, SimplifiedChinese
EUMLanguages:English,German,Spanish,Italian,French
Upto400ProgrammableAnnotationsorganizedin10 anatomical groups
FreeprogrammableScanassistantlistsincludingAdd, Delete,EditandReorderofchecklistitems
FourprogrammablePxbuttonsfordocumentationpreferences likeSave,DICOMSend,Print,Check,Cinelength,jpeg,etc.
Severaluserconfigurablefunctions: •ClinicName •Display(TGCcurve,ScreenLock,Screensaver, AutoScanStop,Beeper,3D/4DScreenControls) •Trackballspeed •Dimfunction •Zoom:Overviewwindow •PatientInfodisplay •Titlebarsettings •StartExamandEndExamConfiguration
Measure setup
M&ASetupincludingAdd,Delete,EditandReorderof measure items
ApplicationSetupincludingseveralparametersof Measurement,DopplerTraceandCalculationpresets
GlobalSetupincludingseveralparametersofMeasurement, CursorandResultwindowpresets
Biopsy setup
Userprogrammableneedleguidelines
Pre-processing
WriteZoomupto8x
B/M-Mode: •Gain •TGC •DynamicRange •AcousticOutput •TransmissionFocusPosition •TransmissionFocusNumber •TransmissionFrequency
PW-Mode: •Gain •DynamicRange •AcousticOutput •TransmissionFrequency •PRF
ColorFlowImagingModes(CFM,PD,TD,HD-Flow) •Gain •AcousticOutput •PRF •WallMotionFilter •Linedensity •Ensemble •Dynamic
Post-processing
ReadZoom:0.8x–3.4xZoom(withHD-Zoomfunctionality upto22xZoom)
B-Mode: •2DGain •Dyn.Contr. •GrayMap
M-Mode: •GrayMap •ColorizedM
PWMode: •GrayMap •BaselineShift •AngleCorrection •ColorizedD
ColorFlowImagingModes(CFM,PD,TD,HD-Flow): •DisplayThreshold •DisplayMode(V,V-T,T,P,P-T) (CFMonly)
•OrientationMarkers •TUI:slicedistance (0.5-10mm) •TUI:slicepositionin overview image •SonoVCADheart
•ElastoMap •Persistance •LineDensity
•EdgeEnhancement •PersistenceControl •LineDensityControl •Reject •SweepSpeed •M-Cursorposition
•WallMotionFilter •SampleVolumeGate •Length,Depth,Pos •VelocityScale •SweepSpeed
•Smooth(RiseandFall) •Frequency •Balance •LineFilter •Quality •ArtifactSuppression
•ColorizedB •SRIII(SpeckleReduction Imaging)
•DisplayFormat •SweepSpeed
•Scale(KHz,m/s,cm/s) •Trace •Invert •SweepSpeed
•ColorMap •Scale(CFMandHD-Flow) •Baseline
Post-processing (cont.)
•B-Flow •GrayMap •ColorizedB-Flow •AdvancedSRI(SpeckleReductionImaging) •Dyn.Contr.
Image processing and presentation
Digital Beam former
1,979,578systemprocessingchanneltechnology
DisplayedImagingDepth:0–30cm
MinimumDepthofField:0–1cm(Zoom,probedependent)
MaximumDepthofField:0–36cm(probedependent)
TransmissionFocus
1-5FocusPointsselectable(probeandapplicationdependent)
FocalZoneposition,upto7steps
ContinuousDynamicReceiveFocus/Continuous DynamicReceiveAperture
256shadesofgray
16.8MillionColors24bit
Upto274dBDynamicRange
ImageReverse:Right/Left
Rotation:0°,180°
Cine features
CineFeatures: •Dual/QuadImageCINEDisplay •CINEGaugeandCINEImagenumberdisplay •CINEReviewLoop •SelectableCINESequenceforCINEReview (byStartFrameandEndFrame) •SideChangeindualCINEMode •Measurements/Calculations&AnnotationsonCINE
Length: •2D:512MB:upto10min(dependingonB-imagesize andFPS);typical:about3min/4000images(withcurved array:15cmdepth,angle81°,22FPS) •M-Mode:32MB:upto20minmotiontime (depending on sweep speed and depth) •Dop.–Mode:32MB:upto10minmotiontime (depending on sweep speed)
Cineoperation: •Manual:imagebyimage •Autorun:speed:25to200%ofreal-timerate,play repeatmode:forward–forward,forward-backward-forward
Image/volume storage (archive)
Imagedatastoredas: •RawDatafile(proprietaryformat) •DICOMfile(Single-orMulitframe)
Volumefilestoredas: •RawDatafile(proprietaryformat) •Size:typically:0.8–5MB(dependingonprobeand adjustedvolumesize)
Compression: •2D:JPEG,Lossless,high,midlow •3D/4D:Lossyandlosslesscompressionavailable.Typical compressionratesare50%withlosslesscompression,15% withlossycompressionbutmaximumqualityand5%with lossy compression and reduced quality (approximate values).
ReviewofcurrentExamandarchiveddatasets(SingleImages andCineClips).ViewFormat:Rawdata,DICOMdata.Display Formats:1x1,2x2,3x3
Reloadofcurrent/archiveddatasets:2DRawData(incl.Color Doppler,SpectralDopplerandM-Mode).3DRawData(Single Volumeincl.Calc.Cines).4DRawData(VolumeCine).
Exportas: •Bitmapfiles:BMP,TIFF,JPEG; •Rawfiles:RAW(2D),VOL(Volumedata),4DV (RAW,VOLincl.Patientdata) •SequenceofBitmaps:BMP,AVI,MOV; •DICOMFiles:DCM,DICOMFileswithDICOMDIR •3DRawData:conversiontoCartesianformatpossible
AVICodec:MPEG4,MSVideo1,FullFrames
Exportto:DVD+/-R(W),CD-R(W),Network,USBdevices
ExportAnonymousfunction:yes,availableforfollowing imagetypes:AVI,MOV,BMP,TIFF,JPEG
Backupfunctionto:DVD+/-R(W)/CD-R(W),Network, USBdevices
Reprofunction:Settingsrecall(e.g.Geometry,Gain, Colormap,etc.)fromastoredorreloadedpicture
ExamHistory:directaccesstoimagesfrompreviousexams; directaccesstoMeasureReportsimagesfromprevious exams;Imagecomparewindowonscreentocompareimages from previous exams with current exam image
HardDriveDataStoragesize:about450GB
Connectivity
Ethernetnetworkconnection
USBforUSBdevices
DICOMsupport: •Verify •Print •Store •ModalityWorklist •StructuredReporting •StorageCommitment •MPPS(Modalityperformedprocedurestep) •MediaExchange •Offnetwork/mobilestoragequeue
Query/Retrieve
Scanning parameters B-Mode
BAcousticPower 0-100
ScanAngle Probedependent
GAINrange +15to-15dB
Gray scale values 8 bit
SRI 6steps(0-5)
CRI 8steps(1-8)
CRIfilter 4steps:off,low,mid,high
CE On/Off(Probedependent)
FFC On/Off(Probedependent)
Persistencefilter 8steps(pre)
Linefilter 3steps(pre)off,low(12,5/75/ 12,5%),high(25/50/25%)
LineDensity 3steps(pre)low,norm,high
Reject 51steps(pre)from0to225
Enhance 6steps(pre)0,1,2,3,4,5
Graymaps 21(18basicmapsand3 User-definedmaps)
Tintmaps 15
Dynamic 12differentdynamiccurves C1–C12
DisplayModes B,XTD
ScreenFormats: •2DImaging:Single(B),Dual(B+B),Quad(B+B+B+B) •XTDView:Single(XTD),Dual(B+XTD)
M-Mode
WorkingModes M(conventionalM-Mode)l, AMM(AnatomicalM-Mode)
Powercontrolrange 1-100
Gainrange +15to-15dB
Msweepspeeds: •900/450/300/225/150/100pixels/sec; •26,44/13.22/8.81/6.61/4.40/2.94cm/sinrelation to system monitor
Review(memorytimes) >60s(32MB)
SignalprocessingM: •Dynamicrange:1to12 •Graymaps:18 •Reject:0to255 •Tintmaps:15 •Enhance:0to5
DisplayModes: •M:2D+M,2D+M/CFM,2D+M/HD-Flow,2D+M/PD,2D+M/TD •AMM:2D+AMM,2D/CFM+AMM/CFM,2D/HD-Flow+AMM/ HD-Flow,2D/TD+AMM/TD
ScreenFormats:(windowarrangement) •2D+Mand2D+AMM:up/down(horizontal):threedifferent subformats30/70,50/50,70/30%left/right(vertical):50/50% •2D+AMM+AMM:left/rt-up/rt-down:50/25/25%
M-Color Flow Mode
AcousticMCFMPower 1-100
MCFMColorMaps 8maps
CFMGain +/-16dBrange,1dBsteps
CFMVelocityScaleRange PRF:150Hzto13kHz
WallMotionFilter 8–3000Hz
Ensemble (color shots per line) 8-16,stepsize1
Gentle color filter
Smoothfilter: Rise:12steps Fall:12steps
CFMSpectrumInversion
CFMBaselineShift 17steps
Pre-settableandindependentlyadjustableB-,MandMCFMGain
CFMThreshold 1–255steps
Balance 25–225,stepsize5
Artifactsuppression On/Off
ColorDisplayMode: •V(Velocity) •T(Turbulence) •V-T(Velocity+Turbulence) •P-T(Power+Turbulence) •V-P(Velocity+Power)
Real-timeTriplexMode B+M+MCFMinanydepth
Spectral Doppler Mode (PW, CW)
OperatingModes PW(PulsedWaveDoppler, SingleGate),CW(Continuous WaveDoppler)
PulseRepetitionFrequency PW-Doppler:0.9...22kHz (PRF) CW-Doppler:1.3...40.0kHz
SampleVolume Length:0.7,1,2,3,4,5,6,7,8, (DopplerGate) 9,10,15mmPosition:5mmto B-scanendAnglecorrection: -85°...0°...+85°
Powercontrolrange 1–100
GAINrange +15to-25dB(PW) +15to-15dB(CW)
WMF(wallmotionfilter) PW:30...500Hz,CW:30...1000Hz
Zerolineshift ±PRF/2,±8steps
SpectrumAnalyzer FFT(FastFourierTransformation), max.256channels,256 amplitude levels
Spectral Doppler Mode (PW,CW) (cont.)
PWsweepspeeds Simplex(26,44/13.22/8.81/ 6.61/4.40/2.94cm/s),Duplex/ Triplex(8.81/6.61/4.40/2.94cm/s)
Review(memorytimes) >60s(32MB)
Measurableflowvelocities: •PW:1cm/s–8m/s(a=0°,2.0MHz,max.zeroshift) 1cm/s–16m/s(a=60°,2.0MHz,max.zeroshift) •CW:1cm/s–11.60m/s(a=0°,2.0MHz,max.zeroshift) 1cm/s–23.20m/s(a=60°,2.0MHz,max.zeroshift)) •Signalprocessing:Dynamicrange:15steps(10to40). Graymaps:18basiccurvesand3User-defined(pre,post), Tintmaps:15
Scale display Vert.:kHz,cm/s,m/s(selectable), Hor.:1smarker(big),½s marker(small)
ScreenFormats 2D/D:up/down(horizontal): three different sub formats 30/70,50/50,70/30%left/right (vertical):50/50%. D:pencilprobesonly
DisplayFormats D/D(duplexupdate, simultaneous);2D+CFM/D, 2D+HD-Flow/D,2D+PD/D, 2D+TD/D(triplexupdate, CWorPW).2D+CFM/PW, 2D+PD/PW,2D+HDFlow/PW, 2D+TD/PW,(triplex simultaneous,PWonly)
Audio-Modes Stereo(bothdirections separately in both channels)
AudioVolume Adjustable,controldigipots
Color Doppler Mode
ScreenFormats 2D+CFM(Single,Dual,Quad)
DisplayModes: •Simultaneousdualmode:2D/2D+CFM •Triplexmode:2D+CFM/PW,2D/M+MCFM •VolumeMode:3D+CFM
Colorcoding: •Steps:65536colorsteps •Displaymodes:V-T(velocity+turbulence),V(velocity), V-P(velocity+power),T(turbulence),P-T(power+turbulence)
Depthrange Axial:0toBscanrange Lateral:0toBscanrange
Baselineshift 17steps(independentfrom spectral Doppler)
Inversionofcolordirection Yes
WallMotionFilter: 7steps(low1,low2,mid1, mid2,high1,high2,max)
SmoothingFilter 12stepsrisingtime 12stepsfallingtime
Gaincontrol +15dBto-15dB,0.2dBstep
LineDensity(colorlinedensity) 10steps
Ensemble(colorshotsperline) CFM:7to31;MCFM:8to16
FlowResolution 4steps(low,mid1,mid2,high)
Pulserepetitionfrequency CFM:150Hzto20.5kHz, MCFM:150Hzto20.5kHz
ColorMap 8differentcolorcodesfor each probe
Frequencyrange 1to16MHzdependingonthe probe,adjustablein3steps (low,mid,high)
Balance From25to225
Max.meas.velocity 4.23m/sec
Min.meas.velocity 0.3cm/sec
Scale kHz,cm/s,m/s
Automaticmoving Yes tissue suppression
Power Doppler Mode (PD)
ScreenFormats 2D+PD(Single,Dual,Quad)
DisplayModes: •Simultaneousdualmode:2D/2D+PD •Triplexmode:2D+PD/PW •VolumeMode:3D+PD
PDcoding 256colorsteps
PDwindowsize Lateral:maximumto minimum B mode scan angle Axial:B-scanrange
Displaymode P(power)
WallmotionFilter 7steps(low1,low2,mid1, mid2,high1,high2,max)
SmoothingFilter Risingedge:12steps Fallingedge:12steps
Gaincontrol +15dBto-15dB,0.2dBsteps
PDEnsemble 7to31
PDLineDensity 10steps
Pulserepetitionfrequency 150Hzto20,5kHz
PDMap 8differentcolorcodes for each probe
Frequencyrange 1to16MHzdependingon theprobe,adjustablein3 steps(low,mid,high)
FlowResolution 4steps(low,mid1,mid2,high)
Balance From25to225in41steps
Frequencyrange 1-16Mhz(Dependingonthe probe,3stepshigh,mid,low)
Artefactsuppression Yes
Scanning parameters (cont.)
HD-Flow
ScreenFormats 2D+HDF(Single,Dual,Quad)
DisplayModes: •Simultaneousdualmode:2D/2D+HDF •Triplexmode:2D+HDF/PW;2D/M+MHDF •VolumeMode:3D+HDF
HD-FlowCodingSteps 256colorsteps
HD-Flowwindowsize:lateral MaximaltominimalBmode scanangle;axial:B-scanrange
Displaymode P(power)
WallMotionFilter 7steps(low1,low2,mid1, mid2,high1,high2,max)
SmoothingFilter 12stepsrisingedge 12stepsfallingedge
GainControl +15dBto-15dB,0.2dBsteps
HD-FlowEnsemble 7to31
HD-FlowLineDensity 10steps
PulseRepetitionFrequency 150Hzto20.5KHz
HD-FlowMap 8differentcolorcodes for each probe
FrequencyRange 1to16MHzdependingonthe probeadjustableinthree steps(low,mid,high)
FlowResolution 4steps(low,mid1,mid2,high)
Balance From25to225
Artefactsuppression Yes
Tissue Doppler Mode (TD)
ScreenFormats 2D+TD(Single,Dual,Quad)
DisplayModes Simultaneousdualmode: 2D/2D+TD;Triplexmode: 2D+TD/PW,2D/M+MTD;
TDcodingsteps 65536colorsteps
Depthrange Axial:0toB-scanrange Lateral:0toB-scan-range
Zerolineshift 17steps
Inversionofcolordirection Yes
SmoothingFilter 12stepsrisingtime, 12stepsfallingtime
Gaincontrol +15dBto-15dB,0.2dBsteps
LineDensity(colorlinedensity) 10steps
Ensemble (color shots per line) 3 to 31
FlowResolution 4steps(low,mid1,mid2,high)
Pulserepetitionfrequency 150Hzto20,5kHz
TDMap 4differentcolorcodesfor each probe
Frequencyrange 1to16MHzdependingonthe probe,adjustablein3steps (low,mid,high)
Balance From25to225
Max.meas.velocity 4.23m/sec
Min.meas.velocity 0.3cm/sec
DisplayMode V(velocity)
Scale kHz,cm/s,m/s
Volume Scan Module
Vol.scansize:max.64MBforgrayvolumes;max.90MBforcolor volumes;Therequiredmemoryspacedependsonscanpara- meters(VOL-boxsizeandquality(low,mid1,mid2,high1, high2,max).Typical:0.8-5MB
Lines/2D-image:max.1024(typ.80to350)
2D-images/volume:Upto4096(AcquisitionModedependent)
VOL-Frames/sec.:max.40(typ.4-8);Theframeratedependson scanparameters:VOL-Boxsize,qualityandprobe
4DVolumeCine:upto128volumes
Displayofsectionalplaneimages:synchronouswithcontrolse- ing,arbitrarymovementinvolume,monitoredpositioninvolume
Rotation:360°,1°or3°increments(X-,Y-andZ-axis)
Magnification:adjustablefrom0.3toafactorof4.00
AcquisitionModes: •3DStatic: –3D(2Dincl.CRI) –4DRealTime –3D/PD(incl.CRI) –4DBiopsy –3D/CFM(incl.CRI) –VCI-A –3D/HD-FLowincl.CRI –VCI-OmniView –3DB-Flow –STIC –3DContrast - FetalCardio - STICAngio:B/PowerDoppler(incl.CRI) - STICCFM:B/ColorDoppler(incl.CRI) - STICHD-Flow:B/HD-Flow(incl.CRI) - STICB-Flow
VisualizationModes: •3DRendering(diversesurfaceandintensityprojectionmodes) •SonoRenderStart •SectionalPlanes •Multiplanar •OmniView,actualandprojectedview •Niche •SonoVCADlabor •TUI(TomographicUltrasoundImaging (overviewimage+parallelslices) •TUIStandard •SonoVCADheart VolumeAnalysis •VOCAL:semi-auto/manualsegmentationtool(segmentation usingtouchscreen),(3DStaticonly) •ThresholdVolume:measurevolumebelowandabove a threshold •SonoAVCfollicle(SonoAutomatedVolumeCount) •SonoAVCgeneral •VCI(VolumeContrastImaging) •V-SRI
Volume Scan Module (cont.)
RenderModes: •HDlive •SurfaceTexture •SurfaceSmooth •SurfaceEnhanced •MixModeoftworendermodes
Displaygraphics: •Rotationaxis,centerpoint •ROIbox,3DFrame •Temporarydisplayofonscreencontrols(rotation,translation)
Graymaps:Slices:21(18basiccurvesand3User-defined (pre,post)3DImage:onegeneralmapadjustablewith bright(-50to+50)&contrast(-50to+50)
Tintmaps:Slices:15;3DImage:15
Depthrendermaps:3
BF (B-Flow)
ScreenFormats Single(BF),Dual(BF+BF), Quad(BF+BF+BF+BF)
DisplayModes BF,Update:BF/PW
Acc.Powerrange 1–100
Scanangle Takenfrom2D
GAINrange +15to-15dB
Gray scale values 8 bit
SRI Takenfrom2D
Persistencefilter 8steps(pre)
S./PRI 1.00,1.50,2.00,3.00,4..00,5.00
Quality 3steps(pre)low,norm,high
Enhance 6steps(pre)0,1,2,3,4,5
Graymaps 21(18basicmapsand3 User-definedmaps)
Tintmaps 15
Dynamic 12differentdynamic curvesC1–C12
Accumulation Off,0.20,0.35,0.50,0.75,1.00, 1.50,Infinite
Background 0,1,2
Contrast Imaging† (Not available in all countries)
Acc.Powerrange 1–100
Scanangle Takenfrom2D
GAINrange +15to-15dB
Gray scale values 8 bit
SRI Takenfrom2D
Persistencefilter 8steps(pre)
S./PRI 1.00,1.50,2.00,3.00,4..00,5.00
Quality 3steps(pre)low,norm,high
Enhance 6steps(pre)0,1,2,3,4,5
Graymaps 21(18basicmapsand 3User-definedmaps)
Tintmaps 15
Dynamic 12differentdynamiccurves C1–C12
Accumulation Off,0.20,0.35,0.50,0.75,1.00, 1.50,Infinite
Background 0,1,2
TimeDelay 0,0.5,1,2,3,………10
DisplayModes: •CodeHarmonicAngio • CodedPI:CIS •CodedPI • CodedPI:CCIS
ScreenFormats: •CodedHarm.Angio:Single(B),Dual(B+B),Quad(B+B+B+B) •CodePI:Single(B),Dual(B+B),Quad(B+B+B+B) •CIS:Dualsimultan(2D+CodedPI) •CCIS:Sngle(B),Dual(B+B),Quad(B+B+B+B)
Elastography
AcousticPowerrange:1–100
TxFrequency:3(penet/norm/resol)
Transparency:51steps90,5,10,…,255)
Softcompress: •Range:0-9 • StepSize:1
HardCompress: •Range0-9 • StepSize:1
PRF:10,15,25,40,60,85Hz
ElastoMaps:8
Persistance: •Range:1-9 • StepSize:1
LineDens.: •Range:1-2
FilterAxial: •Range:1-9 • StepSize:1
Filterlateral: •Range:1-21 • StepSize:2
Windowlength: •Range:8-25 • StepSize:1
ScreenFormats: •Single(2D/Elasto) •Dual(2D/Elasto+2D/Elasto) •Quad(2D/Elasto+2D/Elasto+2D/Elasto+2D/Elasto)
ElastographyAnalysis
ElastographyratioMeasurement
•Transparencymodes: max-min-andX-ray •Gradient •Inversion •GlassBody
Scanning featuresCoded Excitation (CE)
Availableonthefollowingprobes: •RIC5-9-D •AB2-7-D •RAB4-8-D •RIC6-12-D •M6C •11L-D •ML6-15-D •RM6C •RAM3-8 •C4-8-D
Coded Harmonic Imaging
Availableonallprobes,except: •P2D •P6D
Focus Frequency Composite (FFC)
Availableonthefollowingprobes: •4C-D •RIC5-9-D •RAB2-5-D •AB2-7-D •RAB4-8-D •RNA5-9-D •RRE6-10-D •RIC6-12-D •M6C •IC5-9-D •9L-D •RM6C •RAM3-8 •C1-5-D •RRE5-10-D •C4-8-D
Compound Resolution Imaging (CRI)
Availableonallprobes,except: •3SD •S4-10-D •P2D •P6D •3Sp-D •PA6-8-D
Speckle Reduction Imaging (SRI II)
Availableonthefollowingprobes: •RIC5-9-D •AB2-7-D •RAB4-8-D •RNA5-9-D •S4-10-D •RRE6-10-D •RIC6-12-D •RSM5-14 •M6C •11L-D •IC5-9-D •ML6-15-D •9L-D •RAM3-8 •C1-5-D •RRE5-10-D •C4-8-D •RM14L •3Sp-D •PA6-8-D
Speckle Reduction Imaging (V-SRI)
Availableonthefollowingprobes: •RIC5-9-D •RIC6-12-D •RM6C
Virtual Convex
Availableonthefollowingprobes: •3S-D •RSP6-16-D •SP10-16-D •S4-10-D •RSM5-14 •11L-D •ML6-15-D •9L-D •RM14L •3Sp-D •PA6-8-D
Wide Sector
Availableonthefollowingprobes: •4C-D •RIC5-9-D •RAB2-5-D •AB2-7-D •RAB4-8-D •RNA5-9-D •RIC6-12-D •M6C •IC5-9-D •RM6C •RAM3-8 •C1-5-D •RRE5-10-D •C4-8-D
Measurements tool Generic measurements
Distance •Distance(PointtoPoint) •2DTrace(PointLength) •Distance(LinetoLine) •Stenosis(%Dist) •2DTrace(TraceLength)
Area/Circumference •Ellipse •Stenosis(%Area) •Trace(Line) •Area(2Dist.) •Trace(Point)
Volume:followingMethods: •1Distance •3Distance •1Ellipse •Multiplane–Planimetric •1Dist.+Ellipse Volume(3Donly)
Angle: •Angle(3Point) •Angle(2Line)
M-Mode •Distance(PointtoPoint) •HR •Time •Stenosis(%Dist) •Slope
PWDopplerMode: •Auto&ManualTrace: –PS(PeakSystole) –ED(EndDiastole) –MD(Min.Diastole) –PS/ED(Ratio) –PI(PulsatilityIndex) –RI(ResistanceIndex) •TAmax(Timeavg.max.Velocity) •TAmean(Timeavg.meanVelocity) •VTI(VelocityTimeIntegral) •Ductusvenosus:S,D,a,PI,PLI,PVIV •HeartRate
SingleMeasurements: •Velocity •PS/ED •Acceleration •Time •RI •HR •PI
Abdomen calculations
Liver Gallbladder
Pancreas Spleen
Kidney(right/left) RenalArtery(right/left)
Aorta(Proximal,Mid,Distal) PortalVein
Vessel Bladder Volume
SummaryReports
Small part default calculations
Thyroid(right/left)
Testicle(right/left)
Vessel
SummaryReports
Small part breast calculations
Lesion1-5(right/left)
SummaryReports
Obstetrics Calculations
FetalBiometry
Early Gestation
FetalLongBones
FetalCranium
NTMethod:SonoNT/Manual
AFI
Uterus
Ovary
FetalDopplermeasurements(DuctusArt.,DuctusVen.,Ao, Carotid,MCA,CeliacArtery,SuperiorMesentericArtery, UmbilicalArt.,UmbilicalVein,UterineArt.,UmbilicalVein,FHR)
GestationalAgeCalculation
GestationalGrowthCalculation
FractionallimbVolume
FetalWeight(FW)Estimation
FetalTrendGraph
Multi-GestationalCalculation&FetalCompare
CalculationandRatios
FetalQualitativeDescription(Anatomicalsurvey)
FetalEnvironmentalDescription(Biophysicalprofile)
SummaryReports
Obstetrics Fetal Echo
4-Chamber-view
Thorax
OutflowTract,Aorticarch
Venous
Tricuspid valve
MitralValve
AorticValve
MainPulmonaryArtery
PulmonaryValve
Aorta,DuctusArt.
UmbilicalVein,DuctusVen.
FHR
AtrialFHR
LVOT
RVOT
PulmonaryVeins
Carotid
TEIIndex
Rt/Lt-UMA
IVC
SummaryReports
Obstetrics Z-scores
CalculationofZ-scoresfor: •LongAxis •Obl.Shortaxis •AorticArch •4Chambers •ShortAxis •SummaryReports
Cardiology
2DMode: •LVSimpson(Single&Bi-Plane) •Volume(AreaLength) •LV-Mass(Epi&EndoArea,LVLength) •LV(RVD,IVS,LVD,LVPW) •LVOTDiameter •RVOTDiameter •MV(DistA,DistB,Area) •TV(Diameter) •AV/LA(AorticValve/LeftAtrium) •PV(Diameter)
M-Mode: •LV(IVS,LVD,LVPW,RVD) •AV/LA(AoRootDiam,LADiam,AVCuspSep.,AoRootAmpl) •MV(D-E,E-FSlope,A-CInterval,EPSS) •HR(HeartRate),AtrialHR
D-Mode: •MV(MitralValve) •AV(AorticValve),TV(TricuspidValve) •PV(PulmonaryValve) •LVOT&RVOTDoppler(Left&RightVentricleOutflowTract) •PulmonicVeins •PAP(PulmonaryArteryPressuremeasurement) •HR(HeartRate)
C-Mode: •PISA •Tei-Index
Others: •Diast.Vol(Bi) •Syst.Vol.(Bi) •StrokeVolume •VolumeFlow •CardiacOutput •EjectionFraction •FractionalShortening •MyocardialThickness •LA/AoRatio •E/APeak •PeakGradientAcceleration
SummaryReports
Urology
Bladder
Prostate
Left/RightTesticle
Left/RightKidney
Left/RightRenalArtery
Left/RightDorsalPenileArtery
Vessel
SummaryReportsincl.PSAD,PPSA(1),PPSA(2)calculation
•MeanGradient •MeanGradientAcceleration •VTI •TVA •PG •PHT •MVA •AVA •ERO •CVP (CardioVascularProfile)Score
Measurements tool (cont.)
Vascular
Left/RightCCA(CommonCarotidArtery)
Left/RightICA(InternalCarotidArtery)
Left/RightECA(ExternalCarotidArtery)
Left/RightVertebralArtery
Left/RightSubclav.
Left/RightBulb
Vessels
SummaryReports
Gynaecology
Uterus
Right/LeftOvary
Right/LeftFollicle
Fibroid
Endometrialthickness(Dist,DoubleDist.)
CervixLength
Left/RightOvarianArtery
Left/RightUterineArtery
Vessels
PelvicFloor
FHR
SummaryReports
Paediatrics
Left/RightHipJoint
PericallosalArtery
SummaryReport
Neurology
Left/RightACA(AnteriorCerebralArtery)
Left/RightMCA(MiddleCerebralArtery)
Left/RightPCA(PosteriorCerebralArtery)
BasilarArtery
A-Com.A(AnteriorCom.Artery)
P-Com.A(PosteriorCom.Artery)
Left/RightCCA(CommonCarotidArtery)
Left/RightICA(InternalCarotidArtery)
Left/RightVertebralArtery
Vessels
SummaryReports
OB Tables
AgeTables: •AC:ASUM,CFEF,Hadlock_82,Hadlock_84,Hansmann, Hobbins,Jeanty,JSUM,Kurmanavicius,Merz,Nicolaides, Shinozuka,Siriraj,Tokyo •AD:Persson •APAD:Merz
•APTD:Hansmann •APTDxTTD:Shinozuka,Tokyo •BOD:Jeanty •BPD:ASUM,ASUM(old),Campbell,CFEF,Chitty(outer-outer) (outer-inner),Chitty,Hadlock_82,Hadlock_84,Hansmann, Hobbins,Jeanty,Johnsen,JSUM,Kurmanavicius,Kurtz, Persson,Merz,Nicolaides,OSAKA,Rempen,Sabbagha, Shinozuka,Siriraj,Tokyo,Verburg(outer-outer) •CLAV:YARKONI •CRL:ASUM,ASUM(old),DAYA,Hadlock,Hansmann,JSUM, Persson,Nelson,OSAKA,Rempen,Robinson,Shinozuka, Tokyo,Verburg •EFW:Hadlock,JSUM2001,Osaka,Shinozuka,Tokyo •FL:ASUM,ASUM_OLD,CFEF,Chitty,Hadlock_82,Hadlock_84, Hansmann,Hobbins,Hohler,Jeanty,JSUM,Kurmanavicius, Persson,Merz,Nicolaides,O´Brien,OSAKA,Shinozuka,Siriraj, Tokyo,WARDA,Johnsen •FTA:OSAKA •FIB:Jeanty •GS:Hansmann,Hellman,Holländer,Rempen,Tokyo •HC:ASUM,CFEF,Chitty,Hadlock_82,Hadlock_84, Hansmann,Jeanty,Johnsen,Kurmanavicius,Merz, Nicolaides,Siriraj,Johnsen •HL:ASUM,Hobbins,Jeanty,Merz,OSAKA •LV:Tokyo •MAD:EIK-NES,Kurmanavicius •OFD:ASUM,Chitty,Hansmann,Jeanty,Kurmanavicius, Merz,Nicolaides •RAD:Jeanty,Merz •TIB:Jeanty,Merz •TAD:CFEF,Merz,Chitty,Goldstein,HILL,Hobbins,Nicolaides, Hansmann •ULNA:Jeanty,Merz
GrowthTables: •AC:ASUM,CFEF,Chitty,Hadlock,Hansmann,Jacot-Uillarmod, Jeanty,JSUM,Kurmanavicius,Lessoway,Merz,Nicolaides, Shinozuka,Siriraj,Tokyo,Verburg,Johnsen •AD:Persson •AFI:Moore •AortaVmax:Rizzo •APTDxTTD:Shinozuka,Tokyo •BOD:Jeanty •BPD:ASUM,Campbell,CFEF,Chitty,Hadlock,Hansmann, Jacot-Uillarmod,Jeanty,JSUM,Kurmanavicius,Lessoway, Persson,Merz,Nicolaides,OSAKA,Sabbagha,Shinozuka, Siriraj,Tokyo,Verburg •CLAV:YARKONI •CM:Nicolaides •CRL:ASUM,Hadlock,Hansmann,JSUM,Persson,OSAKA, Robinson,Shinozuka,Tokyo,Pexters •DVPI,DVPLI,DVPVIV,DVS/a:Baschat •FL:ASUM,CFEF,Chitty,Chitty,Hadlock,Hansmann,Jacot- Uillarmod,Jeanty,JSUM,Kurmanavicius,Lessoway, Persson,Merz,Nicolaides,O´Brien,OSAKA,Shinozuka, Siriraj,Tokyo,Verburg,WARDA,Johnsen
OB tables (cont.)
•FTA:OSAKA •FIB:Chitty,Jeanty,Siriraj •FWg:Alexander •Foot:Chitty •GS:Hellman,Rempen,Tokyo •HC:ASUM,CFEF,Chitty,Hadlock,Hansmann,Jacot-Uillarmod, Jeanty,Kurmanavicius,Lessoway,Merz,Nicolaides,Siriraj, Verburg,Johnsen •HL:ASUM,Chitty,Jeanty,Merz,OSAKA,Siriraj •LV:Tokyo •MCAPI,RI:JSUM,Bahlman •MCAPV:Mari •MAD:EIK-NES,Kurmanavicius •MVE/A:HARADA •NBL:BUNDUKI,SONEK •OFD:ASUM,Chitty,Hansmann,Jeanty,Kurmanavicius,Merz, Nicolaides •MainPAVmax:Rizzo •RAD:Chitty,Jeanty,Merz,Siriraj •TAD:CFEF,JACOT-GUILLARMOD,Merz, •TC:Chitkara •TCD:Goldstein,HILL,JACOT-GUILLARMOD,Nicolaides, Verburg •TIB:Chitty,Jeanty,Merz,Siriraj •TTD:Hansmann •TVE/A:HARADA •ULNA:Chitty,Jeanty,Merz,Siriraj •UmbArtPI:JSUM,Merz •UmbArtRI:JSUM,Merz,Kurmanavicius •FractionalLimbAvol/Tvol:Lee
FetalweightExtimation(EFW) •Campbell(AC) •Hadlock(AC,BPD) •Hadlock1(AC,FL) •EFW •Hadlock2(BPD,AC,FL) •Hadlock3(HC,AC,FL) •Hadlock4(BPD,HC,AC,FL) •Hansmann(BPD,TTD) •Merz(AC,BPD) •Osaka(BPD,FTA,FL) •Persson(BPD,MAD,FL) •Persson2 •Schild(HC,AC,FL) •Shepard(AC,BPD) •Shinozuka1(BPD,APTD,TTD,FL) •Shinozuka2(BPD,FL,AC) •Shinozuka3(BPD,APTD,TTD,LV) •Tokyo(BPD,APTD,TTD,FL)
Fetal ratios
CI(BPD/OFD)(Hadlock)
FL/AC(Hadlock)
FL/BPD(Hohler)
FL/HC(Hadlock)
HC/AC(Campbell)
Va/Hem(Nicolaides)
Va/Hem(Hansmann)
Vp/Hem(Nicolaides)
Probes4C-D
WideBandConvexProbe
Applications Abdomen,OB,GYN
MaximumBandWidth(-20dB) 2–5MHz
NumberofElements 128
ConvexRadius 60.5mm
FOV 58°
FOV(WideSector) 81°
FootPrint 18.3x68.7mm
Depth Max.30cm
BiopsyGuideavailable Multi-Angle,disposable withreusablebracket
C1-5-D
WideBandConvexProbe
Applications Abdomen,OB,GYN
MaximumBandWidth(-20dB) 2–5MHz
NumberofElements 192
ConvexRadius 56.1mm
FOV 69°
FOV(WideSector) 113°
FootPrint 69.3x17.2mm
Depth Max.30cm
BiopsyGuideavailable Multi-Angle,disposable withreusablebracket
C4-8-D
WideBandConvexProbe
Applications Abdomen,OB,GYN,Urology, Paediatrics
MaximumBandWidth(-20dB) 2–8MHz
NumberofElements 192
ConvexRadius 39.1mm
FOV 75°
FOV(WideSector) 95°
FootPrint 55.2x17.6mm
Depth Max.26cm
BiopsyGuideavailable Multi-Angle,disposablewith reusablebracket
Probes (cont.)
AB2-7-D
WideBandConvexProbe
Applications Abdomen,OB,GYN,Urology, Paediatrics
MaximumBandWidth(-20dB) 2–8MHz
NumberofElements 192
ConvexRadius 41.2mm
FOV 80°
FootPrint 58.9x23.4mm
FOV(WideSector) 107°
Depth Max.28cm
BiopsyGuideAvailable Single-Angle,Reusable
M6C
WideBandConvexProbe(1.25DArray)
Applications Abdomen,OB,GYN,Paediatrics
MaximumBandWidth(-20dB) 1–7MHz
NumberofElements 960
ConvexRadius 50.7mm
FOV 60°
FOV(WideSector) 84°
FootPrint 62.8x24.8mm
Depth Max.26cm
BiopsyGuideavailable Multi-Angle,disposablewith reusablebracket
IC5-9-D
WideBandConvexProbe
NumberofElements 192
Applications OB,GYN,Urology
MaximumBandWidth(-20dB) 4–9MHz
ConvexRadius 10.8mm
FOV 146°
FOV(WideSector) 179°
FootPrint 21.2x17.2mm
Depth Max.16cm
BiopsyGuideAvailable Single-Angle,Reusable
SP10-16-D
WideBandLinearProbe
Applications SmallParts,Paediatrics,MSK PeripheralVascular,Breast
MaximumBandWidth(-20dB) 7–18MHz
NumberofElements 192
FOV 33.7mm
FootPrint 43.412.7mm
Depth Max.6cm
BiopsyGuideavailable PEC64
11L-D
WideBandLinearProbe
Applications SmallParts,Peripheral,Breast Vascular,Paediatrics,MSK
MaximumBandWidth(-20dB) 4–10MHz
NumberofElements 192
FOV 37.4mm
FootPrint 46.9x14.4mm
Depth Max.11cm
BiopsyGuideAvailable Multi-Angle,disposablewith reusablebracket
9L-D
WideBandLinearProbe
NumberofElements 192
Applications SmallParts,Paediatrics,MSK PeripheralVascular,OB
MaximumBandWidth(-20dB) 3–8MHz
FOV 43mm(width)
FootPrint 53x14.1mm
Depth Max.14cm
BiopsyGuideavailable 9L,Multi-Angle,disposable withreusablebracket
ML6-15-D
WideBandMatrixLinearProbe
NumberofElements 1008
Applications SmallParts,PeripheralVascular, Paediatrics,MSK,Breast
MaximumBandWidth(-20dB) 4–13MHz
FOV 49.6mm(width)
FootPrint 60.7x16mm
Depth Max.12cm
BiopsyGuideavailable Multi-Angle,disposablewith reusablebracket
RAB2-5-D
WideBandConvexVolumeProbe
Applications Abdomen,OB,GYN
MaximumBandWidth(-20dB) 1–4MHz
NumberofElements 192
ConvexRadius 46mm
VolumeSweepRadius 22.6mm
FOV 80°(B),80°x85°(Volumescan)
FOV(WideSector) 98°(B),98°x85°(Volumescan)
FootPrint 63.6x38.9
Depth Max.30cm
BiopsyGuideAvailable PEC74,Single-Angle,Reusable and disposable
RAB4-8-D
WideBandConvexVolumeProbe
Applications Abdomen,OB,GYN,Paediatrics, Urology
MaximumBandWidth(-20dB) 2–8MHz
NumberofElements 192
ConvexRadius 46mm
VolumeSweepRadius 22.6mm
FOV 70°(B),70°x85°(Volumescan)
FOV(WideSector) 90°(B),90°x85°(Volumescan)
FootPrint 63.6x37.8mm
Depth Max.26cm
BiopsyGuideAvailable: PEC74,Single-Angle,Reusable and disposable
RAB6-D
WideBandConvexVolumeProbe
Applications Abdomen,OB,GYN,Paediatrics, Urology
MaximumBandWidth(-20dB) 2–7MHz
NumberofElements 192
ConvexRadius 46.8mm
VolumeSweepRadius 24.11mm
FOV 63°(B),63°x85°(Volumescan)
FOV(WideSector) 90°(B),90°x85°(Volumescan)
FootPrint 62.2x34.0mm
Depth Max.26cm
BiopsyGuideAvailable Disposable
RM6C
WideBandConvexVolumeProbewith ActiveMatrixArrayTechnology
NumberofElements 960
Applications Abdomen,OB,GYN,Paediatrics, Urology
MaximumBandWidth(-20dB) 1–7MHz
ConvexRadius 58.8mm
VolumeSweepRadius 22.8mm
FOV 60°(B),60°x85°(Volumescan)
FOV(WideSector) 90°(B),90°x85°(Volumescan)
FootPrint 64.1x40.1mm
Depth Max.26cm
BiopsyGuideAvailable PEC81,Single-Angle,Reusable
RIC5-9-D
WideBandConvexVolumeProbe
Applications OB,GYN,Urology
BandWidth(-20dB) 4–9MHz
NumberofElements 192
ConvexRadius 11.6mm
VolumeSweepRadius 11.6mm
FOV 146°(B),146°x120° (Volume scan)
FOV(WideSector) 179°(B),179°x120° (Volume scan)
FootPrint 22.4x22.6mm
Depth Max.16cm
BiopsyGuideAvailable PEC63,Single-Angle,Reusable, Disposable,disposablewith latex cover
RIC6-12-D
WideBandConvexVolumeProbe
Applications OB,GYN,Urology
BandWidth(-20dB) 5–13MHz
NumberofElements 256
ConvexRadius 11.6mm
VolumeSweepRadius 11.6mm
FOV 149°(B),149x120° (Volume scan)
FOV(WideSector) 195°(B),195°x120° (Volume scan)
FootPrint 22.4(B)x22.6(V)mm
Depth Max.13cm
BiopsyGuideAvailable PEC63,Single-Angle,Reusable, Disposable,disposablewith latex cover
RSP6-16-D
WideBandLinearVolumeProbe
Applications SmallParts,Paediatrics,MSK PeripheralVascularBreast
MaximumBandWidth(-20dB) 6–18MHz
NumberofElements 192
VolumeSweepRadius 80.7mm
FOV 37.4mm(B);37.4mmx29° (Volume scan)
FootPrint 48.6x55.9mm
Depth Max.8cm
BiopsyGuideAvailable PEC75,Single-Angle,Reusable and Disposable
Probes (cont.)
RSM 5-14
WideBandLinearVolumeProbewith ActiveMatrixArrayTechnology
NumberofElements 960
Applications SmallParts,Paediatrics,MSK PeripheralVascular
MaximumBandWidth(-20dB) 5–13MHz
VolumeSweepRadius 61.1mm
FOV 37.5mm(B)x30°(Volumescan)
FootPrint 54.3mmx50.5mm
Depth Max.8cm
BiopsyGuideAvailable PEC80,Single-Angle,Reusable
RM14L
WideBandLinearVolumeProbewith ActiveMatrixArrayTechnology
NumberofElements 960
Applications SmallParts,Paediatrics,MSK PeripheralVascular,Breast
MaximumBandWidth(-20dB) 4–14MHz
VolumeSweepRadius 61.1mm
FOV 37.4mm(B)x30°(Volumescan)
FootPrint 54.3mmx50.5mm
Depth 12cm
BiopsyGuideAvailable PEC80,Single-Angle,Reusable
RNA5-9-D
WideBandConvexVolumeProbe
NumberofElements 192
Applications Abdominal,SmallParts,OB, Cardiology,Paediatrics
MaximumBandWidth(-20dB) 3–9MHz
ConvexRadius 15.4mm
VolumeSweepRadius 15.4mm
FOV 116°(B);116°x90° (Volume scan)
FOV(WideSector) 144°(B),144°x90° (Volume scan)
FootPrint 26.7x22.9mm
Depth Max.18cm
BiopsyGuideAvailable PEC76,Single-Angle, ReusableandDisposable
RRE6-10-D
WideBandConvexVolumeProbe
NumberofElements 192
Applications Gynaecology,Urology
MaximumBandWidth(-20dB) 4–9MHz
ConvexRadius 11.7mm
VolumeSweepRadius 11.7mm
FOV 144°(B),144°x135° (Volume scan)
FootPrint 23.0x22.6mm
Depth Max.12cm
BiopsyGuideAvailable PEC69,Single-Angle
RRE5-10-D
WideBandConvexVolumeProbe
NumberofElements 192
Applications Gynaecology,Urology
MaximumBandWidth(-20dB) 4–9MHz
ConvexRadius 12.2mm
VolumeSweepRadius 12.2mm
FOV 147°(B),147°x135° (Volume scan)
FOV(WideSector) 206°(B),206°x135° (Volume scan)
FootPrint 23.6x24.9mm
Depth Max.16cm
BiopsyGuideAvailable PEC84,Single-Angle
3S-D
WideBandPhasedArrayProbe
NumberofElements 64
Applications Abdominal,Cardiology, Paediatrics,Neurology
MaximumBandWidth(-20dB) 1–3MHz
FOV 90°
FootPrint 27.6x19.3mm
Depth Max.24cm
BiopsyGuideAvailable Multi-Angle,disposable withreusablebracket
3Sp-D
WideBandPhasedArrayProbe
NumberofElements 64
Applications Abdominal,Cardiology,OB, Paediatrics,Neurology
MaximumBandWidth(-20dB) 1–5MHz
FOV 90°
FootPrint 23.4x20.2mm
Depth Max.24cm
BiopsyGuideAvailable Multi-Angle,disposablewith reusablebracket
S4-10-D
WideBandPhasedArrayProbe
NumberofElements 128
Applications SmallParts,Cardiology, Paediatrics
MaximumBandWidth 4–9MHz
FOV 90°
FootPrint 20.0x15.0mm
Depth Max.14cm
BiopsyguideAvailable Notavailable
P2-D
CWDopplerPencilProbe
NumberofElements 2
Applications Cardiology,Peripheral Vascular,Neurology
CenterFrequency 2MHz
FootPrint Diam.:17mm
BiopsyGuideAvailable Notavailable
P6-D
CWDopplerPencilProbe
NumberofElements 2
Applications Cardiology,PeripheralVascular
CenterFrequency 6MHz
FootPrint Diam.:9mm
BiopsyGuideAvailable Notavailable
PA6-8-D
WideBandPhasedArrayProbe
NumberofElements 128
Applications SmallParts,Cardiology, Paediatrics
Maximumbandwidth(-20dB) 4–10MHz
FOV 90°
Footprint 22x11.8mm
Depth Max.14cm
External Inputs and Outputs Connectivityonrearpanel(directaccess) •VGAOut •Network(RJ45) •WirelessNetworkinterface(USB)(Option) •USB(6x) •S-VideoOut1
Connectivitybehindrearpanel(accessafteropening): •DVI-Dout •S-VideoOut2(VTR) •S-VideoIn(VTR) •S-VideoOut1 •AudioOut –Left/right •AudioIn –Left/right •USB(5xinternal) •RS232:Optional,USBtoRS232converter •ParallelPort •Ext.Device/RemoteConnections: –RemoteBWPrinterviaUSB –RemoteColorPrinter/DVRviaUSB –RemoteVCR(RS232)/DVRviaUSB –RemotePrinterviaBluetoothConnectionKit(Option) –FootswitchviaUSB –ECG
GE Healthcare9900InnovationDrive Wauwatosa,WI53226U.S.A.
www.gehealthcare.com
Europe GE Healthcare Beethovenstr.239 D–42655Solingen T4921228020 F49212280228
APACGE Healthcare Asia Pacific 4-7-127,Asahigaoka, Hino-shi,Tokyo191-8503JapanT+81425855111
©2012GeneralElectricCompany–Allrightsreserved.
GeneralElectricCompanyreservestherighttomakechangesinspecificationsandfeaturesshownherein,ordiscontinue the product described at any time without notice or obligation. ContactyourGERepresentativeforthemostcurrentinformation.
GE,GEMonogram,Voluson,CrossXBeam,HD-Flow,SonoAVCandSonoVCADaretrademarksoftheGeneralElectricCompany.
DICOMisaregisteredmarkoftheNationalElectrical ManufacturersAssociation
GEMedicalSystemsUltrasound&PrimaryCareDiagnostics,LLC, aGeneralElectriccompany,doingbusinessasGEHealthcare.
*TrademarkofGeneralElectricCompany.
DOC1022402
† NotforsaleintheUSA.NotapprovedorclearedbytheU.S.FDA.
PleasecontactyourGESalesRepresentative for information about availability in your area.
Safety ConformanceThe Voluson E8 Expert is:
NRTLcertifiedaccordingUL60601-1(TÜVPS)
CertifiedtoCSA22.2,60601.1byanSCCaccreditedTestLab
CB-TestReportbyNationalCertificationBody
CEMarkedtoCouncilDirective93/42/EEConMedicalDevices
Conformstothefollowingstandardsforsafety:
EN60601-1Generalsafetyrequirementsformedicalproducts
EN60601-1-1Particularrequirementsforelectricalmedical systems
EN60601-1-2Electromagneticcompatibility
EN60601-1-4Programmablemedicalsystems
EN60601-1-6Usabilityrequirementsformedicalproducts
EN60601-2-37Particularrequirementsforthesafetyof ultrasound medical diagnostic and monitoring equipment
IEC601157Declarationofacousticoutput
ISO10993Biologicalevaluationofmedicaldevices
NEMAUD3Acousticoutputdisplay(MI,TIS,TIB,TIC)
WEEE(WasteElectricalandElectronicEquipment)