Vinyl Chloride CH2:CHCI No.131L 0.2 0.5 1.0 1.5 2.0 2.5 3.0 ppm n=2 131L 0.02 ppm4 pump strokesYellow Reddish brown Temperature 0 to 40 (32 to 104 ) correction used Relative humidity 0 to 90 % correction not used 10 % for 0.2 to 1 ppm, 5 % for 1 to 3 ppm10 tubes for 5 tests 30 monthsin the refrigeratorDetecting limit : Colour change : Operating conditions : Relative standard deviation : Tube quantity and number of tests per box : Shelf life : Performance 0.1 to 0.2 ppm 0.2 to 3.0 ppm 3.0 to 6.9 ppm 4400 mL2200 mL1100 mL1/2 1 2.3 6 min 3 min 1.5 min Measuring range Number of pump strokes Correction factor Sampling time CH2:CHCl Cr 6 H2SO4 HCI HCl Base Chloride Reaction principle Other substances measurable with this detector tube Allyl chloride Factor : 16 2 3.2 to 48 ppm 1,1,2,2-Tetrachloroethane Factor : 10 2 2 to 30 ppm Substance Correction No. of pump strokes Measuring range Permeation tube method Calibration gas generation When used, these tubes are to be connected. See page 2-3. Possible coexisting substances and their interferences Tetrachloroethylene 1/3 Reddish brown Trichloroethylene 1/5 Benzene, Toluene 200 ppm No Ethylene 200 ppm Substance Concentration Interference Changes colour by itself to Water vapour is trapped in the white layer of the pretreatment tube. Scale

Vinyl Chloride CH :CHCI No

  • Upload
    others

  • View
    12

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Vinyl Chloride CH :CHCI No

Vinyl Chloride CH2:CHCI No.131L

0.2

0.5

1.0

1.5

2.0

2.5

3.0

ppm

n=2

131L

0.02 ppm(4 pump strokes)Yellow → Reddish brownTemperature 0 to 40 ℃ (32 to 104 ℉) correction usedRelative humidity 0 to 90 % correction not used10 %(for 0.2 to 1 ppm), 5 %(for 1 to 3 ppm)10 tubes for 5 tests30 months(in the refrigerator)

Detecting limit :Colour change :Operating conditions :

Relative standard deviation :Tube quantity and number of tests per box :Shelf life :

Performance

0.1 to 0.2 ppm 0.2 to 3.0 ppm 3.0 to 6.9 ppm 4(400 mL) 2(200 mL) 1(100 mL) 1/2 1 2.3 6 min 3 min 1.5 min

Measuring rangeNumber of pump strokesCorrection factorSampling time

CH2:CHCl + Cr6 + + H2SO4 → HCIHCl + Base → Chloride

Reaction principle

Other substances measurable with this detector tube

Allyl chloride Factor : 16 2 3.2 to 48 ppm1,1,2,2-Tetrachloroethane Factor : 10 2 2 to 30 ppm

Substance Correction No. of pump strokes Measuring range

Permeation tube methodCalibration gas generation

When used, these tubes are to be connected. See page 2-3.

Possible coexisting substances and their interferences

Tetrachloroethylene ≧ 1/3 + Reddish brownTrichloroethylene ≧ 1/5 + Benzene, Toluene ≧ 200 ppm − NoEthylene ≧ 200 ppm −

Substance Concentration Interference Changes colour by itself to

Water vapour is trapped in the white layer of the pretreatment tube.