Upload
trandiep
View
220
Download
5
Embed Size (px)
Citation preview
一、 OverviewThank you for choosing this series incubator, this series incubator stable performance, high reliability, using the latest type of microprocessor and electrical element. Imported microcomputer chip, high interference rejection, high working stability, using imported ultra-precision sensor, high acquisition accuracy,wide test range. It ensure the properties of products are stability and reliability. Generous and fashion appearance,one –piece injection molding technology, corrosion resistant, aging resistant. It is high function and convenient home-using micro incubator.
二.Performance parameters of the incubator1. 1.Range of temperature display: 0~99℃
2.Precision of temperature measurement: ±0.1℃
3.Range of humidity display: 0~99 RH﹪
4.Precision of humidity control:±3%RH
5.Output channels: 2 channels(AC 220V electric heater/DC 12V electric heater)
6.Output cuurent:3A
7.Machine power: 220V operating power<80W,12v operating power<50W.
三、Working conditions:
1.Working voltage: AC 180V ~240V; Frequency 50Hz or DC 12V.
2.Relative humidity: less than 85%
3. Environment temperature: 0℃~40℃
四、Components and function
1. Power supply inlet 2. Controller display /operation interface 3. Uniformity temperature fin 4.Egg turning device 5.Egg-holder clapboard 6.Egg-holder 7.Hatching gridding 8.Chassis 9.Loophole 10. Humidification gutter 11. Water injection hole
五.Temperature setting method
Press SET button(short press) under normal working status,humidity window show code PP, blink on on the window show the temperature valve
under current setting,pres “+”button or “-”button to adjust temperature to
using one. After finish setting, press “SETTING” button exit or no operation on button more than 10s, the system can automatic exit and changed parameter can be auto-saved.
六.Eliminate Alarming
When the incubator alarming, press “–”button, the alarming sound disappear, if the alarming light is still on, Press “–”button again, alarming sound recover.
七、 Open illuminate method
(1) Automatic illuminate function: press LIGHT button(short press),it can
realise open/close illuminate function,and if nobody operation,it can
automatic close after LED light on 5 minutes.
(2) Constraint illuminate function: Long press LIGHT press for 3s (long
press),it can realize open/close illuminate function, and if nobody
operation, the LED light will on for long time and repetitive operation once
can close.
八.Restore Factory Setting method
Press “+” and “- ” meanwhile for more than 5s (long press) under the normal working status,buzzer call a sound , after the screen show “DAY. DAY. DAY DAY .DAY ”restore normal working status, all the paramater will restore factory defaults---temperature value is 37.8℃ and incubation days will restore to 1 day.
九 、 Checking and correction incubation
days method
(1) Query incubation days method: long press SET button more than 3s,and the
screen show code P1 and then press SET button,switch code to F3 option,the
value of number on show is current incubation days.
(2)Press+/- can adjust incubation days for F3 value and when the incubation
days stop the date can hold.
(3) Incubation days of factory default is 1, from the incubator on work ,the
timing unit is 24 hours /day.
(4) The Maximum days for incubation is 99 days and it will not accumulate more
than 99days.
(5) Long Pressing “+” button more than 3s under the normal working status,
incubation days will restore to 1day.
十 . Temperature and humidity manual
calib method
It will appear deviation for long time using the temperature and humidity
sensor, we can calibrate according the exactly thermometer and hygrometer.
Temperature calibration: long pressing SET more than 3s, screen show the
code, press SET button switch code to F1 option,F1 will be the current
temperature value and press “+”or“-”button,adjust the temperature as
same as the thermometer.
Humidity calibration: same as operation above,when the screen show F2
press “ + ” or“ - ” button, adjust the humidification as same as the
hygrometer..
十一. Sensor error
When the temperature window shoe EEE and temperature sensor break down or no connection, meanwhile the electric heator stop working. When the humidity window shoe EE and humidity sensor break down or no connection,when any sensor break down,the buzzer will chirp.
十 二 、 Dual power automatic
identification function
This incubator has dual power automatic identification function,when the 12V
battery and 220V AC access system at the same time , the system preference for
220V city power,and 220V power light on work;when city power outage, the
system switch to 12V battery DC power and 12V power light will on work.
Remark : Under setting status,setting paramter blinks diplay and no press button more than 10s ,the system will exit setting status and revised parameter auto-saved.
十 三 . Pressing setting button (short press) under normal working status, the
controller will enter fiducial temperature parameter setting.(After the system
leave factory,user only an setted fiducial parameter and temperature parameter
will change with setting of the fiducial parameter.)
No Parameter Name Code Range Factory Default
1 Fiducial temperature
paramter setting
PP (P2) 0-99.9℃ 37.8
Long pressing SETTING button more than 3s under normal working status,the
controller enter interval parameter setting. ( befor system leave factory,interval
parameter has been setted and no adjust to the system or will affect operating of
incubator and incucating result.if operating wrong, please refer to restore factory
default setting method chapter)
No Parameter Name Code Range Factory Default
1 High temperature alarm
setting parameter
P1 0-99.9℃ 38.8
2 Stop heating setting
parameter
P2 0-99.9℃ 37.8
3 Start heating setting
parameter
P3 0-99.9℃ 37.7
4 Low temperature alarm
setting parameter
P4 0-99.9℃ 36.5
5 High humidity alarm
setting parameter
H1 0-99% 85%
6 Low humidity alarm
setting parameter
H2 0-99% 30%
7 Temperature calibration
setting parameter
F1 As the thermometer
8 Humidification
calibration setting
parameter
F2 As the hygrometer
9 Incubation days F3
十四. Attentions and maintenance
1,Regular Checking water lever in the water tank , no water shortage for the incubator and low humidity affect incubation effect.
1. Do not using this machine for long time, stop using for 1-2 days when finish one batch, increased the lifetime of machine.
2. Regular checking the work status of temperature and humidity display screen ,fin turning and egg turning motor ,ensure the incubator normal running and finish incubation.
3. Using dry cloth scrub every parts when the incubation finish, clear the villus of young chicks and put away to use next time.
六. Accessory
4. Incubator 1 pcs, Manual 1 pcs, power line 2 pcs , Kettle 1 pcs ,spare fuse 1 pcs.
十六、Incubation technology knowledge Birds Incubating Parameter
Name Incubation days
Incubation Period Hatching PeriodHatching egg weight
Tem±0.5℃ Hum% Tem±0.5℃ Hum% g
Chicken 21 37.6 55-70 37.5 65-85 50-60Peacock 28 37.6 60-75 37.3 65-85 100-120Turkey 28 37.6 60-75 37.2 65-85 75-80Duck 28 37.8 60-75 37.3 65-85 80-100Goose 30 37.6 65-80 37.1 65-85 100-120Pigeon 18 38 55-70 37.8 65-85 30-40Pheasant 24 37.8 55-70 37.7 65-85 30-40Local chicken
21 37.6 55-70 37.5 65-85 45-55
Certification,Service CARD
CERTIFICATION
Model: Date In produced :
Inspector:
SERVICE CARD
USER NAME: TEL:ADDRESS: POST CODE:PURCHASE PLACE: PURCHASE DATE:
Incubation progress record card
NO QTY TIME LIGHTING TIME
Wind egg QTY
Hatching time
Hatching time
Hatching percent