Verigy V93000Service Training
Module 3:Support Rack &
Power Subsystem
03 - Support RackPage 2
Verigy Restricted
Support RackPower
Backplane
Test Head
355V DC
Trim DC out Voltage
200-240V AC
To• Controller (RS232)• Cooling Unit• Manipulator MotorLeakage Sensor
Pressure Sensor
DIS
Temp.Sensors
Sense
DC output to each card slot
Temp / DCMonitor
DIS
Power Control
From
Mains Power/ Auto Transformer
208-400V AC
ACDCControl
Primary Power Box ,AC Dist Box
AC/DC Power Unit
Rack Interface
Board
ON/OFF Unit
DC/DC Power Unit
Power Control Bd
Power Subsystem Key Components
Power Block Diagram – LTH & STH
K1
03 - Support RackPage 3
Verigy Restricted
Compact Test HeadPower
Backplane
355V DC
Trim DC out Voltage
200-240V AC
To• Controller (RS232)• Cooling Unit
Pressure Sensor
DIS
Temp.Sensors
Sense
DC output to each card slot
Temp/ DCMonitor
DIS
Power Control
From
Mains Power
208-400V AC
ACDCControlPrimary
Power Box ,AC Dist Box
AC/DC Power Unit
Rack Interface
Board
ON/OFF Unit
DC/DC Power Unit
Power Control Bd
Power Subsystem Key Components
Power Block Diagram - CTH
K1
03 - Support RackPage 4
Verigy Restricted
On/Off UnitEmergency Off button, illuminated ON and STANDBY switches, and an additional warning lamp. The unit is integrated into the front panel of the support rack. A ribbon cable connects it to the power back plane
Air Intake+Cooling Fan for Rack Interface Board & Power Backplane
Rack Interface Board – controls the system start and shutdown procedures and contains vital parts of the diagnostic subsystem (DIS). Also contains power supplies for the DIS. Rack interface board is accessible from front side.
Air Intake for Support Rack
AC/DC Power UnitsUp to eight water cooled AC/DC units, which supply the test head with 355V DC. AC/DC units are accessible from front side.
Test Head Cooling Water Hoses
Test Head Power & Fiber-Optic Cables
Front View
03 - Support RackPage 5
Verigy Restricted
Space for additional equipment
Primary Power BoxConnection to mains power, the system’s line switch, and the line breaker
AC Distribution Box and Power BackplaneAC Distribution Box – Fuses and individual breakers for switched outlets, non-switched outlets, additional support racks, and the rack interface board.The switched and non-switched outlets for auxiliary equipment provide output 200-230V AC.
Power Backplane – A board placed behind the AC Distribution Box. There are one DC-out connector and two fuses per card cage. There are several other connectors for WS, Cooler, Manipulator, EMO, DIS, etc. Temperature sensors also exist.
Fiber-Optic Cables
Compressed Air Connector
Water Hose Inlet/Outlet to Cooling Unit
Rear View
03 - Support RackPage 6
Verigy Restricted
CTH - Bottom/Maintenance View
Active Control BoardCC 1
Passive Control BoardCC2
Rack Interface Board
Jumper Board
Optical link
AC/DC Power Units
On/Off Unit AC/DC Power Box
03 - Support RackPage 7
Verigy Restricted
CTH – End View Connections
Mains power inlet
Optical link
Cooling water in
Cooling water out
Compressed air
Mains power out(Cal Robot)
Main breaker
Low Level Diagnostic Cable RS232
Cooler Control
ON/OFF Unit cable
Nitrogen
Air PressureGovernors &
Guages
03 - Support RackPage 8
Verigy Restricted
Primary Power Box (PPB)
ContactorDisconnect mains power controlled by the signal from rack interface board in case that the system detects hardware failure
Main Breaker
Connection Point of Main Power Cable (3-phase)
Connection Point of Main Power Cable (PE/Neutral)
AC Distribution Box
Non-switched Power Outlets
Switched Power Outlets
Breakers for Additional Support Racks
Primary Power Box and AC Distribution
03 - Support RackPage 9
Verigy Restricted
Primary Power Box
Power Backplane
AC Dist. BoxFrom
Mains (3 phase)• Delta 208 V• Star 230/400 V
From Rack Interface Board
To Non-switched Power Outlets
200 – 230 V AC
To Rack Interface Board200 – 230 V AC
To Additional Support Racks200 – 230 V AC
To Switched Power Outlets200 – 230 V AC
To Card Cages355 V DC
Main Breaker
ON/OFF
Contactor
AC/DC
Mains power is distributed to:• AC/DC Power Units• 3 switched and 3 non-switched power sockets• Additional support racks• Rack Interface Board
Primary Power Box is configurable to 3-phase delta 208V and to 3-phase star 400/230V. As shipment default, configured to the European power option of star 400/230V.
Primary Power Distribution
03 - Support RackPage 10
Verigy Restricted
Power Options with Autotransformer
The Autotransformer E8005A is required for systems with:
Small or Large test head, if the primary voltage is 440V or 480V.
Large test head, if the primary voltage is 200V/208V AND power consumption exceeds 36kVA. Currently this is the case if more than 42 digital channel boards (of any type) are installed.
• Analog boards can be neglected.• 1 Additional Support Rack is equivalent to 5 channel boards. • Example: For systems with 2 Additional Support Racks up to 32 channel boards are
supported without the transformer.
The E8005A can be pro-actively ordered for large test head systems to allow for later upgrades.
03 - Support RackPage 11
Verigy Restricted
Autotransformer Front and Back Views
Power Output Cable
Power On Lamp
Output Breaker ST2
Emergency Off Button
Power Cable Outlet
Cooling Fan
Control Transformer
Fan Power Supply
FRONT BACK
Primary PowerCable Inlet
Cooling Fan
PE
Primary PowerConnections
Input Breaker ST1
03 - Support RackPage 12
Verigy Restricted
Breaker S1 Contactor K1
AC Distribution Box
PE
L3
L2
L1
T3
T2
T1
S2
S4
S5
S6
to switched Power Outlets (J 2)
S3
to Analog Rack #3
to Analog Rack #1
to Analog Rack #2
to non-switched Power Outlets (J 1)
from Rack Interface Board
Power Backplane
AC/ DC TDI #6
AC/ DC TDI #3
AC/ DC TDI #8
AC/ DC TDI #5
AC/ DC TDI #2
AC/ DC TDI #7
AC/ DC TDI #4
AC/ DC TDI #1
F10
F11
F12
F13
F14
F15
F16
F17
F26
F27
F20
F21
F22
F23
F24
F25
F34
F35
F36
F37
F30
F31
F32
F33
2×355 V to Card Cage #1
2×355 V to Card Cage #2
2×355 V to Card Cage #3
2×355 V to Card Cage #4
2×355 V to Card Cage #5
2×355 V to Card Cage #6
2×355 V to Card Cage #7
2×355 V to Card Cage #8
LineFilter
to Rack Interface BoardF5
F4
N
Primary Power Box
Primary Power – Star 400/230V – LTH & STH
NOTE: LTH systems require an autotransformer (E8005A) if the primary voltage is 200V/208V AND power consumption exceeds 36kVA (Currently this is the case if more than 42 digital channel boards (of any type) are installed).
Line Filter PowerBackplane
RackInterface
F50
F51
NOTE: Newer HD support racks will have the new lineConfiguration going to the RIB. This will include the additionOf fuses F50 and F51 (PN 2110-1377), bypassing the S3 breaker.
03 - Support RackPage 13
Verigy Restricted
Breaker S1 Contactor K1
AC Distribution Box
PE
L3
L2
L1
T3
T2
T1
S2
S4
S5
S6
to switched Power Outlets (J 2)
S3
to Analog Rack #3
to Analog Rack #1
to Analog Rack #2
to non-switched Power Outlets (J 1)
from Rack Interface Board
Power Backplane
AC/ DC TDI #6
AC/ DC TDI #3
AC/ DC TDI #8
AC/ DC TDI #5
AC/ DC TDI #2
AC/ DC TDI #7
AC/ DC TDI #4
AC/ DC TDI #1
F10
F11
F12
F13
F14
F15
F16
F17
F26
F27
F20
F21
F22
F23
F24
F25
F34
F35
F36
F37
F30
F31
F32
F33
2×355 V to Card Cage #1
2×355 V to Card Cage #2
2×355 V to Card Cage #3
2×355 V to Card Cage #4
2×355 V to Card Cage #5
2×355 V to Card Cage #6
2×355 V to Card Cage #7
2×355 V to Card Cage #8
LineFilter
to Rack Interface Board
F4
F5
Primary Power Box
Primary Power – Delta 208V – LTH & STH
Line Filter PowerBackplane
RackInterface
F50
F51
NOTE: Newer HD support racks will have the new lineConfiguration going to the RIB. This will include the additionOf fuses F50 and F51 (PN 2110-1377), bypassing the S3 breaker.
03 - Support RackPage 14
Verigy Restricted
Primary Power – Star 400/230V – CTH
Pow er D istribution U nit (star connection for 400V /230V)
T1
T2
T3
L1
L2
L3S1 K1
S2 AC /D C #1
S3
L2/N
L3/N AC /D C #2
S4 O utle t
PE
F1
F2L1/N
N
PE
Line
F3
F4
Filter
T1/N
R ack In terfaceBoard
N
N
N
L1
L2
L3
03 - Support RackPage 15
Verigy Restricted
Primary Power – Delta 208V – CTH
T1
T2
T3
L1
L2
L3S1 K1
S2 AC/D C #1
S3
L2/N
L3/N AC/D C #2
S4 O utlet
PE
F1
F2L1/N
N
PE
Line
F3
F4
Filter
T1/N
R ack In terfaceBoard
N
N
N
L1
L2
L3
Pow er D istribution Unit (delta connection for 200V/115V)
03 - Support RackPage 16
Verigy Restricted
L1
L2
L3
N
PE
Note• Select suitable cable clamp according to cable diameter• Torque Specification at Breaker: 15Nm• Star to Delta reconfiguration is described in Installation Guide
Star Configuration 400/230V Delta Configuration 208V
Primary Power Configuration – LTH & STH
03 - Support RackPage 17
Verigy Restricted
Primary Power Configuration – CTH
Phase terminals L1
L2L3
N terminal(Star config. only)
PE terminal
03 - Support RackPage 18
Verigy Restricted
The left side terminal of the breaker always connects two wires.
The right side terminal connects one wire to Neutral in Star configuration or to another phase in Delta configuration.
The supply voltage for the additional racks is: 200V..230V
Cable number are different between old and new racks:
E9695A (old)
E9695B (new)
Left 2 & 3 2 & 6Right 1 4
Power cable clamps for additional support racks
Breakers for additional Rack #3 #1 #2
Additional Support Rack Connections
03 - Support RackPage 19
Verigy Restricted
F17F15F11F16F14F12F10
F31 F33 F35F36F34F32 F37F30
S3
S2
F 2 5 F 2 7F 2 3F 2 1F 2 6F 2 4F 2 2F 2 0 S 4 S 5 S6
gray: Used for 8CC test head only
Fuses and Breakers – LTH & STH
Breaker / Fuse Location (ID) Purpose
Main Breaker Primary Power Box (S1) For disconnect mains power completely.
Contactor Primary Power Box (K1) For remotely disconnect mains power controlled by the signal from rack interface board in case that the system detects hardware failure.
Switched power line breaker and fuses
AC Distribution Box (S2/F10/F11) 1 breaker 13A and 2 fuses 45A for switched power sockets
Non-switched power line breaker and fuses
Primary Power Box (S3)AC Distribution Box (F4/F5)
1 breaker 13A and 2 fuses 45A for non-switched power sockets, power back plane and rack interface board
Additional support rack breakers and fuses
AC Distribution Box (Rack 1: S5/F34/F35) (Rack 2: S6/F36/F37) (Rack 3: S4/F26/F27)
1 breaker 16A and 2 fuses 45A for each additional support racks (max. 3 racks)
AC/DC fuses AC Distribution Box (AC/DC 1: F12/F13, AC/DC 2: F20/F21) (AC/DC 3: F30/F31, AC/DC 4: F14/F15) (AC/DC 5: F22/F23, AC/DC 6: F32/F33) (AC/DC 7: F16/F17, AC/DC 8: F24/F25)
2 fuses 45A for each AC/DC unit
DC Out fuses Power Backplane 2 fuses 10A for each Card Cage
03 - Support RackPage 20
Verigy Restricted
Fuses and Breakers – CTH
Breaker / Fuse Location (ID) Purpose
Main Breaker Power Distribution Unit (S1) For disconnect mains power completely.
Contactor Power Distribution Unit (K1) For remotely disconnect mains power controlled by the signal from rack interface board in case that the system detects hardware failure.
Switched power line breaker and fuses Power Distribution Unit (S4/F1/F2) 1 breaker 16A and 2 fuses 10A for switched power sockets
Rack Interface Board line breakers Power Distribution Unit (F3/F4)) 2 fuses 3A for rack interface board
AC/DC line breakers Power Distribution Unit (AC/DC 1: S2, AC/DC 2: S3)
1 breaker 50A for each AC/DC unit
03 - Support RackPage 21
Verigy Restricted
Fuses and Breakers Maintenance
Indicator : Blown fuse : Good fuse Fuse ExchangeThere is a Fuse Extractor available to remove the thermal fuse. Part number: E8001-63851
S3
S2
GFCI test button (GFCI – Ground failure Current Interrupter)Perform GFCI test with appropriate procedures (see speaker note) every 6 months. Incorrect procedures damage the components in the test head.
For Exchange:Use a flat screwdriver to pull out tab under the breaker
Check continuity by using multimeter. Blown fuse shows open connection.
Test Point
03 - Support RackPage 22
Verigy Restricted
AC/DC Power Unit
Output Power3kW 6kW
350V355V360V
Uout
Control
DC/DC-Converter
PFCAC/DC-
Converter LC Line
Filter
PWR-OK
AC-OKTEMP-OK
180-264 Vac
Disable
EEPROM
AC1
AC2
DIS
Uout = 350 - 360 Vdc
DC -
DC+
Disable PE
100K
100msTemp. Monitor
INSTALL*
Block Diagram
03 - Support RackPage 23
Verigy Restricted
• Output power from all AC/DC units are connected in parrallel.
• Filling Order of AC/DC unit:
• Minimum configuration:
• Required number of AC/DC units is determined by the power calculation. (See below)
(Number of ACDC) = (Power Summation of HW Resources) / 6000
Example: 8CC test head with 24 Pin Scale 3600 + 4 DPS
(8 x 414 + 4 x 457 + 24 x 695 ) / 6000 = 3.63 Requires 4 AC/DC’s
AC/DC Power Unit Filling Order
8CC 4CC
#1
#5
#2
#6
#3
#7
#4
#8
#1 #2 #3 #4
8CC: 5 and 14CC: 1
8CC: 516273844CC: 1234
Minimum Configuration
Installable Slot
HW Resource Power per unit
Clock(214) + CC power margin 414 WDPS (GP, HV, HC) 457 WDC Scale DPS, VI 650 WUtility DPS 30 WDigital I/O (PS800) 600 WDigital I/O (PS400, PS3600) 650 WRF Card (Griffin) 124 WAV8 (MCA, MCB) 298 WPinScale HX 300 W
03 - Support RackPage 24
Verigy Restricted
Support the weight ofthe AC/DC power unitwhile pulling it out.
Use the hose extraction tool for the cooling hoses
15kg / unit
Unit Exchange
LTH & STH CTH
03 - Support RackPage 25
Verigy Restricted
Rack Interface State Machine and error latches
Residual charge warning LEDsThermal (Bimetal) switch with release button
Connectors to Power Backplane
Locating pin
DIS Controller and CAN bus interface
Rack interface board is the heart of the power subsystem. It contains a part of the diagnostic subsystem (DIS).
Accessible from front side of the support rack
RIB Location and Components – LTH & STH
03 - Support RackPage 26
Verigy Restricted
Rack Interface State Machine and error latches
Residual charge warning LEDs
Thermal (Bimetal) switch with release button
Connectors to Power Backplane
Locating pin
DIS Controller and CAN bus interface
Accessible from the back side of the CTH
RIB Location and Components – CTH
Fan
Water cooled heat sink
Prototype
03 - Support RackPage 27
Verigy Restricted
To (+24VC)Contactor
Rack Interface Board
RS232Con
nect
ions
to e
xter
nal u
nits
Connections to test head
Pressure Sensors
Leakage Sensor
ON/OFF Unit
Cooling Unit
External EMO
Signal Lamps
Manipulator
System Controller (RS232)
CAN Bus DIS Power +5V, ±15 V , - 12 V
AC/DC MonitorTemperature Monitor
Support Voltage +30 V +24 VC +15 VC +5 VC -15 VC +24V
DISController
230 VACfromMainBreaker
Bi-metal Switch
• Functions of the board:• Executing the system start procedure• ON/OFF Unit interface• Monitoring the system status• Controlling the contactor K1• Providing the diagnostic interface to the
controller (via RS232)• Controlling the manipulator, soft and
hard docking• Generating support voltage• Monitoring pressure, leakage, and
cooling fan• Monitoring temperature for AC/DC,
Rack Interface Board, and Power Backplane
• Monitoring AC/DC supply voltage• DIS controller with RAM banks and
CAN Bus interface• DIS and RS232 interface
Rack Interface
State Machine
& support circuits
Block Diagram
To (+24V)Manipulator and Docking
03 - Support RackPage 28
Verigy Restricted
Cooling fan for Rack Interface Board and Power Backplane
Connector to Power Backplane
The Fan operation is monitored by the Rack Interface Board
Support Rack Cooling Fan
03 - Support RackPage 29
Verigy Restricted
System Status by LEDStatus Green Yellow Red
System stand-by Off Off OffStarting cooler, connecting AC/DC power units, and switched/non-switched power outlets
Blinking Slow Off Off
Starting DC/DC power units Blinking Fast Off Off
System OK On Off OffSystem running, cooler warning On On OffSystem shutdown, hardware failure Off Off On
Emergency Off (EMO) Button
ON Button / SYS OK LED
Warning LED
STANDBY Button / SYS FAIL LED
EMO CircuitInterlock(jumper as default)
EMO Button
EMO to Cooler
ContactorCoolerOn/Off
External EMO
+24 V
At Front Panel
External EMO connector on Power Backplane
Cooler control connector on Power Backplane
From Rack Interface Bd.
ON/OFF and EMO – LED and Circuit
03 - Support RackPage 30
Verigy Restricted
State State Name Description LED’s
0 state_off• System in stand-by• AC/DC power units disconnected from the mains power supply
1 state_start_1
• Cooling unit is started• Timeout counter for the cooling unit startup is set for 20 sec.• Green LED blinks with low frequency
2 state_half_on
• Contactor closed. AC/DC power units connected to mains power• DC/DC power unit still inactive• Timeout counter for the AC/DC power units set• Green LED blinks with low frequency
3 state_start_2
• Enable DC/DC power units by setting Start_PWR• Timeout counter for DC/DC power units set to 1.3 sec.• Green LED blinks with high frequency
4 state_on • Green LED on
5 state_half_off
• DC/DC power units disconnected in sequence• AC/DC power units disconnected from mains power supply• Cooling unit is shutdown
‘EMO’ Engaged
Leakage detected
Cooling failure
6 state_fail • Turn on red LED
System Start-Up States
03 - Support RackPage 31
Verigy Restricted
DC Out Connectors
Fuses for DC Out
Cables toAC Fuses
Power Backplane - Location
03 - Support RackPage 32
Verigy Restricted
EEPROM
Over-temperature Sensors
LED’s, Connectors and Fuses for DC Out
Rear Side Front SideLED & Connectors(See next slide)
Rack interface Board Connectors
AC/DC Power Unit Connectors
Cooling Loop 1 Cooling Loop 2 Leakage Sensor(J140) (J141) (J143)
TH Power Control(J113)
External EMO Sense / Interlock(J130)
24V2, 24VC2 for TH(J135)
+5VC for TH(J114)
* Must be connected for system installation.
Manipulator Motor(J142) *
Cooler Control (EMO and On/Off)(J107) *
Power Backplane – LTH & STH
03 - Support RackPage 33
Verigy Restricted
Installation:Connect the RS232 cable to the system controller for low level diagnostic.
Connectors and LEDs (front side)
03 - Support RackPage 34
Verigy Restricted
Power Backplane - CTH
Cooler Control RS232 for low level diagnostic
ON/OFF Unit with EMO
Rack Interface Board Connectors
AC/DC Power Unit Connectors
LED’s(See Next Slide)
03 - Support RackPage 35
Verigy Restricted
LEDs (front side)
Remove cover to check LEDs
03 - Support RackPage 36
Verigy Restricted
Support Rack to Test Head Connections
03 - Support RackPage 37
Verigy Restricted
Beeper Board
Hard DockingSensor
Rack Interface Board
ManipulatorMotor
Magnetic Valve(Interface)
+24V when undocked
+24V
OpenClose
+5V when undocked
GNDInterface dockInterf. undock
Hard dockHard undock
Motor onUp / DownIdleOverride
Support Rack
Test Head
Up / D
own
Motor on
24VDC
Power Backplane
Interface Micro Switch
Rem
ote
Con
trol
Con
nect
or
+24V
OpenClose
Magnetic Valve(Hard Docking)
Up -
Up+
Override Enable
Test Head Up/Down
DUT UndockDUT Dock
Hard UndockHard NeutralHard Dock
Manipulator and Docking Control
03 - Support RackPage 38
Verigy Restricted
• 8 DC/DC Power Units are in a 4CC TH • 16 DC/DC Power Units are in a 8CC TH • The DC/DC Power Units step down the 355 VDC from the
AC/DC boards to different low voltages required by the TH boards. The DC/DC Power Unit modules are mounted to water-cooled heat sinks.
• Each card cage contains two identical DC/DC Power Units.
• A DC/DC Power Unit consists of two DC/DC conversion units which are screwed onto both sides of a common heat sink.
• DC/DC Power Unit generates voltages for the following modules: Channel boards, Analog module, Clock boards, Device power supplies and Control boards.
• Two types of DC/DC Power Units are available:• Digital DC/DC Power Unit,• Analog DC/DC Power Unit(generate 16V for analog
channel boards).
DC/DC Power UnitBack plane address, data , control bus interface.
355 VDC from the AC/DC boards, low voltages for the following modules: Channel boards, Analog module, Clock boards, Device power supplies and Control boards.
03 - Support RackPage 39
Verigy Restricted
There are 2 DC/DC power units per card cage in the test head. Each DC/DC power unit occupies 2 slots.
Each DC/DC power unit is composed of two boards. Each board has 5 DC/DC converters mounted to one water-cooled heat sink. These DC/DC converters supply required voltages to all test head boards.
Each converter has sense lines that adjust the output level roughly to its nominal level (without control circuitry).
Output levels are monitored and regulated by monitoring circuits to nominal levels in case of load variations.
The temperature of the heat sink is measured. The contactor K1 is disabled as soon as the temperature exceeds 57°C.
Each DC/DC converter has an enable input, which is used to switch on/off the converter. A certain switching sequence is followed in order not to damage sensitive components (ADC and bus driver).
Universal DC/DC only powers up, if PinScale or BIST Assist cards are detected.
CAUTION: Putting the analog DC/DC in a digital card cage might damage the channel boards. See supply voltage difference in the table.
Universal DC/DC (blue coding key) Analog DC/DC (pink coding key)
Single Density E6989-68591, +3.3V, For P/C/Ce/NP/BAA/PPS, Strap: blackE6989-68597, ±16V, For all analog, Strap: blue
High Density E8001-68515, +6V +12V, For Pin Scale/BAA, Strap: orange stripe
Backplane Connectors
Frontplane Connectors
DC/DC Power Unit
03 - Support RackPage 40
Verigy Restricted
• 4 Power Control boards in a 4CC test head • 8 Power Control boards in a 8CC test head•• The task of the Power Control board are to:
• Provide a defined power ON/OFF sequence• Measure the DC output voltages and generates trim
voltages, which are applied to the converters.• Monitor the trim voltages• Monitor the temperatures of all boards in its card cage• Contain a DIS controller and DIS interface
DC/DC Power Unit output voltages, trim voltages, temperatures of all boards in its quadrant, DIS controller and DIS interface.
Power Control Board
03 - Support RackPage 41
Verigy Restricted
Each card cage has a Power Control Board, which is plugged into the front plane board.
The power control board controls and monitors the DC/DC power units.
LTH STH
Power Control Board - Location
CTH
03 - Support RackPage 42
Verigy Restricted
Functions of the Power Control Board:
Contains a DIS controller which is connected to all other DIS controllers in the system via the CAN Bus.
Powered by +5 V / ±15 V / - 12 V from the Rack Interface Board
The supply voltages are present as soon as the main breaker is ON. Thus, the diagnostic hardware is accessible even if the system was shut down by a hardware failure.
The State Machine provides a defined power ON/OFF sequence.
Controls and monitors the DC/DC converters on the DC/DC Power Units.
Monitors the temperatures (reference: 57 degree C) of all cards in each card cage.
For PinScale channel cards only:Detects the coding resistors and monitors the DC/DC converters on the PinScale channel cards.
Power Control Board
Error Logic &
State Machine
DIS Controller
Rack Interface
Board
DIS
Status LEDs
Power Control
CAN Bus
Channel Cards
Select
Temp DC
Voltages & Temp. Signals
Voltages & Temp. Signals
Coding
Trim DC out Voltage
Enable DC/DC
DC/DC Power Units
+ 5 V ±15 V - 12 V
Block Diagram
03 - Support RackPage 43
Verigy Restricted
301/302 PS3600
303/304 PS3600
305/306 PS800
307/308 PS800
101/102 PS3600
107/108 PS800
105/106 PS800
103/104 PS3600301/302 PS800
303/304 PS800
305/306 PS800
307/308 PS800
101/102 PS3600
107/108 PS3600
105/106 PS3600
103/104 PS3600
Possible Configuration:DC/DC unit hosts a pair of same card
Impossible Configuration:DC/DC unit doesn’t host Pin Scale 3600 and 800 pair.
301/302 BAA
303/304 BAA
305/306 BAA
307/308 BAA
101/102 PS3600
107/108 PS800
105/106 PS800
103/104 PS3600
Possible Configuration:BIST Assist 6.4 can co-exist with any of Pin Scale card.PS3600 has priority to PS800.
301/302 Digital
303/304 Digital
305/306 Digital
307/308 Digital
101/102 Digital
107/108 Digital
105/106 Digital
103/104 Digital
DC/DC E8003A
DC/DC E8003A
Power Distribution in a Cage
8
7
6
5
4
3
2
1
For slot 7&8For slot 5&6For slot 3&4For slot 1&2
Examples at Group1, CC 3
DC/DC vs. PinScale IO Type Compatibility
03 - Support RackPage 44
Verigy Restricted
Card compatibility table
PS400 PS800 PS3600 PS HXBist-
AssistDPS32 VI32 MB AV8 6G QS Port Scale RF MB AV8 6G QS AV 8 Port Scale RF Griffin
PS400PS800
PS3600PS HX
Bist-AssistDPS32
VI32MB AV8
6G QSAV8
Port Scale RFGriffin
cards can sit side-by-side cards cannot sit side-by-side
analog DCDCuniversal DCDC
Which board can pair up with which other board
The following board types can be operated with both kind of DC/DC (analog or universal):
- MBAV8- 6GHz Sampler (Multi-site Sampler)- Port Scale RF
03 - Support RackPage 45
Verigy Restricted
During power-up and power-down the state machine enables/disables the DC/DC converters in a defined sequence by activating the enable signal for the DC/DC converters.
Single DensityVoltage Analog
Voltage Digital
Distribution Route
-12 V - 12 V Frontplane+14 V + 14 V Frontplane±16 V +3.3 V Backplane+5 V + 5 V Frontplane- 6 V - 6 V Frontplane
+ 10 V +10 V Frontplane+ 5 V + 5 V Backplane
±16 V + 3.3 V Frontplane (for DPS)
PinScaleVoltage Analog
Voltage Digital PS800
Voltage Digital PS3600
Distribution Route
- 12 V - 12 V - 12 V Frontplane+ 14 V + 14 V + 14 V Frontplane+12V +12V +12V Backplane±16 V + 12 V + 12↓6 V Backplane+ 5 V +5↑6.5 V +5↑6.5 V Frontplane- 6 V - 6 V - 6 V Frontplane+ 10 V + 10 V + 8 V Frontplane+ 5 V + 5 V + 5 V Backplane±16 V + 3.3 V + 3.3 V Frontplane (for DPS)
Enab
le
Disa
ble
NOTE: The red numbers indicates voltage differences.
Power-up Sequence
03 - Support RackPage 46
Verigy Restricted
LED Descriptions
OK System is switched on and power is ok
UVPb UVPd Under voltage detected (b: channel bd 5 – 8, d: -12V for DPS)
UVPa UVPc Under voltage detected (a: channel bd 1 – 4, c: +12V for DPS)
OVP Over voltage detected
SPWR Start power (on for 1.3 second during start up)
OVT1 OVT2 Over temperature detected (OVT1 is on and OVT2 is off)
DC/DC failure on channel board (OVT1 and OVT2 are on)
OFF DC off (system switched off, mains present)
MISMATCH A Incompatible Channel Boards in Slots 1 / 2
MISMATCH B Incompatible Channel Boards in Slots 3 / 4
MISMATCH C Incompatible Channel Boards in Slots 5 / 6
MISMATCH D Incompatible Channel Boards in Slots 7 / 8
Power Control Board – LED’s
03 - Support RackPage 47
Verigy Restricted
RS232 and CAN Bus System (CTH)
Rack interfaceboard
RS232cable
Clockboard
Active controlboard
Card Cage 2 Card Cage 1
Powercontrol board
CTH
Systemcontroller
Power backplane
To M
anip
ulat
orPowercontrol board
Termination
Con
trol
Cab
le 2
Con
trol
Cab
le 1
Wiring Board
03 - Support RackPage 48
Verigy Restricted
Standard support rack
Rack interfaceboard
RS232cable
Clockboard
Clockboard
Active controlboard
Active controlboard
Card Cage 2
Card Cage 4
Card Cage 3
Card Cage 1
Powercontrol board
Powercontrol board
Power control board
Powercontrol board
Connectionboard
Termination
STHSystem
controller
Control cable
Power backplane
To M
anip
ulat
or
RS232 and CAN Bus System (STH)
03 - Support RackPage 49
Verigy Restricted
Clockboard
CardCage 1
CardCage 5
LTH
Active controlboard
Connectionboard
CardCage 7
Card Cage 3
CardCage 6
CardCage 2
CardCage 4
Card Cage 8
Clockboard
Clockboard
Clockboard
Active controlboard
Active controlboard
Active controlboard
Control cable
Standard support rack
Rack interface
board
RS232cable
Systemcontroller
Power backplane
Power controlboard
Powercontrolboard
Powercontrolboard
Power controlboard
Termination
Supply only
To M
anip
ulat
or
RS232 and CAN Bus System (LTH)