2
Designed specifically for hospital use, the Volumetric Diffusive Respiration unit (VDR®-4) is a pressure-limited and time-cycled ventilator which provides high frequency percussive ventilation (HFPV). This HFPV system supports both diffusive and convective flow by stacking breaths in cumulative sub-tidal volumes, allowing for air exchange, airway clearance and lung recruitment for most hospital patient populations, neonatal to adult. The VDR®-4 can be used to treat respiratory conditions including hypoxemia, hypercapnia, aspiration, smoke inhalation and ARDS. This unit is paired with the Monitron® II Waveform Analyzer which provides accurate, real-time visual waveform, pressures and alarms making the VDR®-4 a versatile tool for critical and intensive care respiratory ventilation. Accessories VDR ® - 4 HIGH FREQUENCY PERCUSSIVE VENTILATION VDR®- 4 WITH MONITRON® II K00008-1 VDR ® -4 Phasitron ® Breathing Circuit Kit A50094-D Single Patient Use Only

VDR - 4B)_Product_Specification,_VDR4,_English.pdfthe Volumetric Diffusive Respiration unit (VDR®-4) is a pressure-limited and time-cycled ventilator which provides high frequency

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: VDR - 4B)_Product_Specification,_VDR4,_English.pdfthe Volumetric Diffusive Respiration unit (VDR®-4) is a pressure-limited and time-cycled ventilator which provides high frequency

Designed specifically for hospital use, the Volumetric Diffusive Respiration unit (VDR®-4) is a pressure-limited and time-cycled ventilator which provides high frequency percussive ventilation (HFPV). This HFPV system supports both diffusive and convective flow by stacking breaths in cumulative sub-tidal volumes, allowing for air exchange, airway clearance and lung recruitment for most hospital patient populations, neonatal to adult. The VDR®-4 can be used to treat respiratory conditions including hypoxemia, hypercapnia, aspiration, smoke inhalation and ARDS. This unit is paired with the Monitron® II Waveform Analyzer which provides accurate, real-time visual waveform, pressures and alarms making the VDR®-4 a versatile tool for critical and intensive care respiratory ventilation.

Accessories

VDR®- 4HIGH FREQUENCY PERCUSSIVE VENTILATION

VDR®- 4 WITH MONITRON® II K00008-1

VDR®-4 Phasitron® Breathing Circuit Kit A50094-D

Single Patient Use Only

Page 2: VDR - 4B)_Product_Specification,_VDR4,_English.pdfthe Volumetric Diffusive Respiration unit (VDR®-4) is a pressure-limited and time-cycled ventilator which provides high frequency

130 McGhee Road, Suite 109Sandpoint, ID 83864800-850-7205 | 208-263-2549www.percussionaire.comP20060 Rev B

VDR®- 4Technical Specifications

See Waveform Analyzer Monitron® II for Display output

Accesories

Phasitron® Kit

Settings

Liquid Consumption

Percussion Frequency

Pulse/Interval Ratio

Operating Conditions

Operating Range

Storage and Transport Range

Power Input

Run Time

Digital Multimeter

Pulse Frequency

Mean Airway Pressure (MAP)

Amplitude Bar Graph

Average Inhalation Pressure (AIP)

Average Exhalation Pressure (AEP)

Alarm Condition

High Airway Pressure Limit

Electronic High/Low Pressure Alarm

Disconnect Alarm

Blender Gas Supply Alert

Gas Source

Wall Gas

Flow

Dimensions/Weights

Dimensions (WxHxD)

Weight

Required Maintenance

Part No.

A50094-D

Specifications

1cc minute

50-900 pulses per minute

Adjustable

Specifications

Temp., 0° C to 49° C (32° F to 120° F),humidity 5%-95%

Temp., -20° C to 60° C (-4° F to 140° F), humidity <93% non-condensing

100-240 VAC, 50/60 Hz, 3 amp

Continuous

Specifications

Digital display, 50-999 pulses per minute

Digital display, 0-99 cmH2O/hPa, accurate to +/-1 cmH2O/hPa

Digital display, 0-50 cmH2O/hPa

Digital display, 0-100 cmH2O/hPa

Digital display, 0-100 cmH2O/hPa

Indication

Audible indicator (Pneumatic Pressure failsafe) w/ Automatic Pressure Relief

Visual and audible indicators

Audible indicator

Audible indicator

Specifications

50-80 PSI air and O2

25 LPM per gas source-sustained

Specifications

40.64 cm x 33 cm x 25.4 cm (16" x 13" x 10" )

8.98 kg (19.8 lbs)

Every 3 years

Average Inhalation Pressure (AIP)

Pulse Frequency Rate Average Exhalation Pressure (AEP)

Mean Airway Pressure Bar graph

measuring pulse amplitude

(0-50 cmH20/hPa) Increments 2.5-3

cmH20/hPa

VDR®-4 display

50

0

30

VOLUMETRIC DIFFUSIVE RESPIRATION