15
Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members: Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris (ChBE) Mentor: Prof. Baudenbacher

Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Embed Size (px)

Citation preview

Page 1: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Vanderbilt Senior Design 2010:Non-Electronic Blood Pressure Assist Device

Members: Laura Allen (ChBE)James Berry (BME)Casey Duckwall (BME)David Harris (ChBE)

Mentor: Prof. Baudenbacher

Page 2: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Blood Pressure Assist Device

•The Engineering World Health (EWH) Organization proposed the challenge:▫Can a mechanical adjunct for a

sphygmomanometer be designed to amplify the oscillatory pressure signal?

▫Can the design be used by minimally trained users to identify at least systolic pressure? (Ideally diastolic too)

Page 3: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Developing World Usability

•Self-sustainable▫Power supply from mechanical or solar means

•Low Cost: Less than $10▫When produced in quantities of 5000+

•Mechanical adjunct to sphygmomanometer▫Eliminating the need for Korotkoff sound

identification• Identify at least systolic

▫Pictorial instructions for all users

Page 4: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Rationale•Cardiovascular diseases kill more people

worldwide than any other disease1

▫Even a leading cause in developing nations, where AIDS and malaria receive greater attention

•Hypertension is a primary risk factor for cardiovascular illnesses▫Managing blood pressure is vital for long-

term health of cardiovascular disease patients2

▫Diagnosing hypertension can be challenging in developing countries

Page 5: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Literature Review

•Patent search did not reveal any useful information

•Researched the science behind blood pressure

•Investigated different methods of measurement

•Identified a range for oscillation frequencies

Page 6: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Preliminary Data Collection

•Findings indicate that measuring blood pressure by observation of sphygmomanometer alone is unreliable

•Our final design should be within ±10mmHg per reading, taking into account that normal blood pressures vary by ±3mmHg with each beat

Page 7: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Results from Commercial BP Device•Commercially available electronic blood

pressure devices in the range of $40 yielded extreme variations in blood pressure readings, going so far as to place some individuals in both stage 2 hypertension as well as hypotension.

•Furthermore, these readings were extremely variable between measurements on the same individual, varying by more than 20mmHg for systolic.

Page 8: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Project BudgetElectronic blood pressure cuff $45Standard blood pressure cuff $203 solar cells $15Pressure transducer $40Batteries (AAA) $8Rubber tubing (8mm) $10T-junction $10Total $148Shipping $30

Page 9: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Circuit Specification

•An electronic approach yields 5 new design elements▫Power Supply▫Filtering▫Amplification▫Readout▫Transducer

Page 10: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Circuit

Page 11: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Power Supply

•Reusable▫Solar cell - $3.45 in bulk▫3.4V, 25mA

•Sufficient voltage output▫±3V operational amplifier rails▫~2V LED, 20mA

Page 12: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Filtering and Amplification

•Literature suggests frequency range of 20-80Hz

•First-order active filtering using▫High-pass filter > 10Hz▫Low-pass filter < 100Hz

•Amplify signal to approximately 2V in passed range

Page 13: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Output

•Two monochromatic LED▫1) Power light - Demonstrate sufficient

power to the device▫2) Indicator light – Light up during pass-

band frequencies

Page 14: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

Transducer• Pressure Sensor: $3.65 in bulk

▫NovaSensor NPC-100▫Developed for usage in biomedical diagnostics▫Sensitivity ±1%

5µV/V/mmHg▫Linearity ±1%

For physiological range▫Full range -30mmHg to 300mmHg

▫Test operating resistance to compute peripheral resistor values

Page 15: Vanderbilt Senior Design 2010: Non-Electronic Blood Pressure Assist Device Members:Laura Allen (ChBE) James Berry (BME) Casey Duckwall (BME) David Harris

References(1) WHO. “Fact Sheet: The Top Ten Causes of Death.” WHO.

November 2008. Accessed October 28, 2009 http://www.who.int/mediacentre/factsheets/fs310_2008.pdf

(2) Pickering TG. , Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T, Sheps SG, Roccella EJ; Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension. 2005 Jan;45(1):142-61.