Upload
sksingh82
View
215
Download
0
Embed Size (px)
Citation preview
8/4/2019 Validation2_0506_1_
1/33
2005.06.28. Dr. Pogny - WHO, Pretoria 1/33
Supplementary Training Workshop on Good
Manufacturing Practices (GMP)
QUALIFICATION ofEQUIPMENT
Jnos Pogny, pharmacist, PhD,consultant to WHO
Pretoria, South Africa, 28 June 2005
E-mail:[email protected]
8/4/2019 Validation2_0506_1_
2/33
2005.06.28. Dr. Pogny - WHO, Pretoria 2/33
WHO GMP and Guidelines WHO good manufacturing practices (GMP):
main principles for pharmaceutical products
Section 4. Qualification and validation
Section 13. Equipment (see notes page)
Supplementary guidelines on good
manufacturing practices (GMP): Validation
(2003) Draft.
8/4/2019 Validation2_0506_1_
3/33
2005.06.28. Dr. Pogny - WHO, Pretoria 3/33
Causes of variation Man (different operators - lack of proper training)
Machine / equipment
Measurement (lack of calibration)
Method (accuracy of validated processes/methods)
Material (batch-to-batch variation of APIs/excipients)
Manufacturing environment (T and RH)
8/4/2019 Validation2_0506_1_
4/33
2005.06.28. Dr. Pogny - WHO, Pretoria 4/33
Qualification QUALIFICATION is the Action of proving that any
premises, (pharmaceutical utility) systems and items ofequipment (e.g., machines, stability chamber)work
correctly and actually lead to the expected results. Themeaning of the word validation is sometimesextended to incorporate the concept of qualification.
REQUALIFICATION is the main part of thepreventive
maintenance programme of proving that anypremises,(pharmaceutical utility) systems and items of equipmentwork correctly and keep on leading to the expected
results. (normal wear and tear)
8/4/2019 Validation2_0506_1_
5/33
2005.06.28. Dr. Pogny - WHO, Pretoria 5/33
Validation VALIDATION is the Action of proving, in
accordance with the principles of GMP, that anyprocedure,process, equipment, material, activity
or (pharmaceutical utility) system actually leadsto the expected results (see also qualification).
REVALIDATION is a part of the change
control system of proving that any procedure,process, equipment, material, activity or (pharmaceutical utility) system actually keepson leading to the expected results.
8/4/2019 Validation2_0506_1_
6/33
2005.06.28. Dr. Pogny - WHO, Pretoria 6/33
Qualification - ValidationRegulatory
requirements
DQ,IQ,OQ
inputs
DQ,IQ,OQ
process
DQ,IQ,OQ
outputsProcess
Qualification
Validation
Premises, equipment and supporting utilities must be
qualified to operate in a validated process.
8/4/2019 Validation2_0506_1_
7/33
2005.06.28. Dr. Pogny - WHO, Pretoria 7/33
Relationship between qualification
and validation
DESIGN SPECIFICATION/QUALIFICATION
INSTALLATION QUALIFICATION
OPERATIONAL QUALIFICATION
PROCESS VALIDATION
CHANGE CONTROL
Qualification is a sub-set of validation.
8/4/2019 Validation2_0506_1_
8/33
2005.06.28. Dr. Pogny - WHO, Pretoria 8/33
CalibrationThe set of operations that establish, under specified
conditions, the relationship between values indicated
by an instrument or system for measuring(especially weighing), recording, and controlling, or
the values represented by a material measure, and
the corresponding known values of a reference
standard. Limits for acceptance of the results ofmeasuring should be established. (see also the
notes page) Verification.
8/4/2019 Validation2_0506_1_
9/33
2005.06.28. Dr. Pogny - WHO, Pretoria 9/33
Calibration test equipmentInstrument or device used to calibrate other
instruments, the calibration of which is traceable
back to accepted national standards. The testequipment should have greater precision, accuracy,
and repeatability, than the instruments being
calibrated. These instruments should have
documentary evidence of traceable calibration to atleast the same standard as the site calibration.
8/4/2019 Validation2_0506_1_
10/33
DESIGN
QUALIFICATION
8/4/2019 Validation2_0506_1_
11/33
2005.06.28. Dr. Pogny - WHO, Pretoria 11/33
4.3 Documentary evidence(a) the premises, supporting utilities, equipment and
processes have been designed in accordance with
the requirements for GMP (design qualification orDQ);
(b) the premises, supporting utilities and equipment
have been built and installed in compliance with
their design specifications (installation qualificationor IQ);
8/4/2019 Validation2_0506_1_
12/33
2005.06.28. Dr. Pogny - WHO, Pretoria 12/33
4.3 Documentary evidence(c) the premises, supporting utilities and equipment
operate in accordance with their design
specifications (operational qualification or OQ);(d) a specificprocess will consistently produce a
product meeting its predetermined specifications
and quality attributes (process validation or PV,
also calledperformance qualification or PQ).(PQ is frequently applied to utility systems)
8/4/2019 Validation2_0506_1_
13/33
2005.06.28. Dr. Pogny - WHO, Pretoria 13/33
Basic principles Throughout this presentation the term equipment is
used to represent both manufacturing machinery,equipment and QC instruments.
The equipment should be correctly installed inaccordance with an installation plan, as persupplierand any special (purchaser) requirements.
The requirements forcalibration, maintenance andcleaning developed as draft procedures should bereviewed and finally issued as authorised standardoperating procedures (SOPs).
8/4/2019 Validation2_0506_1_
14/33
2005.06.28. Dr. Pogny - WHO, Pretoria 14/33
Basic principles Operating requirements should be established and
tests conducted to assure equipment is operating
correctly, undernormal and worst caseconditions ( failure mode analysis)
Operator training requirementspertaining to the
new equipment should be finalised and
documented.
8/4/2019 Validation2_0506_1_
15/33
2005.06.28. Dr. Pogny - WHO, Pretoria 15/33
DQ andIQ DQ protocols and reports
GMP
VMP
National law
Engineering design and construction documents
Do not start IQ before DQ has been completed!
IQ/OQ protocols and reports Above inputs + machine manuals
Separate VMPs forHVAC and watersystems
Do not start OQ before IQ has been completed!
8/4/2019 Validation2_0506_1_
16/33
INSTALLATION
QUALIFICATION
8/4/2019 Validation2_0506_1_
17/33
2005.06.28. Dr. Pogny - WHO, Pretoria 17/33
Prequalification of vendor Vendor specification
familiarity with GMP
installation, training and start-up support performance testing at the vendors site
experience ofcurrent users
cost analysis
Once selected, the vendor may assist in preparingthe qualification protocols
8/4/2019 Validation2_0506_1_
18/33
2005.06.28. Dr. Pogny - WHO, Pretoria 18/33
Pre-delivery checkForcomplicated orlarge pieces of equipment, a
pharmaceutical manufacturer may elect to undertake a pre-
delivery check of the equipment at the supplier's assembly
facility. This pre-delivery check cannot substitute for the
Installation Qualification. However, it is acknowledged that
the checks conducted and documented at this stage may
duplicate a number of the checks conducted at the
Installation Qualification stage, hence, there could be areduction in the scope of the Installation Qualification
checks.
8/4/2019 Validation2_0506_1_
19/33
2005.06.28. Dr. Pogny - WHO, Pretoria 19/33
Checking at usersInstallation Qualification requires a formal andsystematic check of all installed equipment against theequipment supplier's specifications and additionalcriteria identified by the user as part of the purchasespecifications. At the Installation Qualification, allequipment, gauges and services should be given a serial(or other reference) numberand a check conducted that
the installed equipment (or plant) has been installed inaccordance with the current (approved) version of thePiping & Instrument Diagram (P&ID).
8/4/2019 Validation2_0506_1_
20/33
2005.06.28. Dr. Pogny - WHO, Pretoria 20/33
IQ - Overview StatementIdentification and documenting ofmaintenance
requirements for each installed item and the
collection and collation ofsupplier operating andworking instructions, maintenance and cleaning
requirements, should form the minimum
documentation for a satisfactory Installation
Qualification.
8/4/2019 Validation2_0506_1_
21/33
2005.06.28. Dr. Pogny - WHO, Pretoria 21/33
Installation of equipmentAn effective Change Managementprocedure should
be in place. All changes to the original design
criteria should be documented and after that,appropriate modifications made to Equipment
Specifications, Plant Functional Specifications and
Piping & Instrument Diagrams (P&IDs).
8/4/2019 Validation2_0506_1_
22/33
2005.06.28. Dr. Pogny - WHO, Pretoria 22/33
CalibrationRequirements Confirmation of calibration with reference to the
appropriate national standard.
Calibration of measuring devices utilised in the
Operational Qualification stage, where confirmation of
calibration is unavailable.
Calibration of measuring devices related to installed
equipment.
Identification ofcalibration requirements for measuring
devices for the future use of the equipment
8/4/2019 Validation2_0506_1_
23/33
OPERATION
QUALIFICATION
8/4/2019 Validation2_0506_1_
24/33
2005.06.28. Dr. Pogny - WHO, Pretoria 24/33
OQ - Overview StatementOperational Qualification is an exercise oriented tothe engineering function, generally referred to ascommissioning. Studies on the critical variables(parameters) of the operation of the equipment orsystems will define the critical characteristics foroperation of the system or sub-system. All testingequipment should be identified and calibrated before
use. Test methods should be authorised,implemented and resulting data collected andevaluated.
8/4/2019 Validation2_0506_1_
25/33
2005.06.28. Dr. Pogny - WHO, Pretoria 25/33
OQ - Overview Statement It is important at this stage to assure all
operational test data conform with pre-determinedacceptance criteria for the studies undertaken.
It is expected that during the OperationalQualification stage the manufacturer shoulddevelop draft standard operating procedures(SOPs) for the equipment and services operation,
cleaning activities, maintenance requirements andcalibration schedules.
8/4/2019 Validation2_0506_1_
26/33
2005.06.28. Dr. Pogny - WHO, Pretoria 26/33
OQ - Essential Elements The conduct of an Operational Qualification should follow an
authorised protocol.
The plans for the Operational Qualification should identify the
studies to be undertaken on the critical variables, the sequence
of those studies and the measuring equipment to be used and
the acceptance criteria to be met.
Studies on the critical variables should incorporate specific
details and tests that have been developed from specialistknowledge of the process and how the equipment will work
(defined in design criteria and specifications).
8/4/2019 Validation2_0506_1_
27/33
2005.06.28. Dr. Pogny - WHO, Pretoria 27/33
OQ - Essential Elements Where applicable, simulated product may be used to conduct
the OQ. Studies on the critical variables should include acondition or a set of conditions encompassing upper and
lower processing or operating limits and circumstances;commonly referred to as "worst case" conditions. Suchconditions should not necessarily induce product or processfailure.
The completion of a successful OQ should allow the
finalisation of operating procedures and operator instructionsdocumentation for the equipment. This information should beused as thebasis for training of operators in the requirementsfor satisfactory operation of the equipment.
8/4/2019 Validation2_0506_1_
28/33
2005.06.28. Dr. Pogny - WHO, Pretoria 28/33
OQ - Essential Elements Draft cleaning procedures developed at the IQ stage
should be finalised. Where applicable, these procedures
should be validated as part of the PQ phase. The completion of satisfactory IQ and OQ exercises
should permit a formal "release" of the equipment/plant
for the next stage in the validation exercise (Process
Validation). The release should take the form of writtenauthorisations for both Installation Qualification and
Operational Qualification.
8/4/2019 Validation2_0506_1_
29/33
2005.06.28. Dr. Pogny - WHO, Pretoria 29/33
OQ - Essential Elements Some OQ testing should be performed following
equipment maintenance
Current regulations require the validation ofcomputers and software that are used to control
laboratory equipment or process data.
The overall program should be evaluated
periodically to ensure that current requirementsare met.
8/4/2019 Validation2_0506_1_
30/33
2005.06.28. Dr. Pogny - WHO, Pretoria 30/33
Re-Qualification Modifications to, orrelocation of, equipment
should only follow satisfactory review andauthorisation of the documented change proposalthrough the change control procedure. Part of thereview procedure should include consideration ofre-qualification of the equipment. Minor changesor changes having no direct impact on final or in-
process product quality should be handledthrough the documentation system of the
preventative maintenanceprogramme.
8/4/2019 Validation2_0506_1_
31/33
2005.06.28. Dr. Pogny - WHO, Pretoria 31/33
Qualification of in-use equipmentWhile it is not possible to undertake the details of anIQ for established equipment nor the detailedapproach for an OQ, nevertheless there should bedata available that support and verify the operating
parameters and limits for the critical variables of theoperating equipment. Additionally, the calibration,cleaning, preventative maintenance, operating
procedures and operator trainingprocedures for theuse of the equipment should be documented and inuse as standard operating procedures (SOPs).
8/4/2019 Validation2_0506_1_
32/33
2005.06.28. Dr. Pogny - WHO, Pretoria 32/33
Main points again Equipment is one of the potential causes of
variation.
DQ, IQ, OP and PQprotocols and reports arerequired that equipment was designed, installed and
operated in accordance with suppliers and users
requirements.
Effective change control keeps documentation upto date.
Qualification is a sub-set of validation
8/4/2019 Validation2_0506_1_
33/33
2005.06.28. Dr. Pogny - WHO, Pretoria 33/33
Literature Validation Master Plan, Installation and
Operational Qualification, Non-sterile ProcessValidation, Cleaning Validation, Pharmaceutical
Inspection Convention (PIC/S), August 2001
Model IQ/OQ protocol and report (distributed amongparticipants of the training course)
Comparison ofWHO, EU and US-FDA GMPRequirements for Pharmaceutical Equipment(distributed among participants of the training course)