Validation2_0506_1_

Embed Size (px)

Citation preview

  • 8/4/2019 Validation2_0506_1_

    1/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 1/33

    Supplementary Training Workshop on Good

    Manufacturing Practices (GMP)

    QUALIFICATION ofEQUIPMENT

    Jnos Pogny, pharmacist, PhD,consultant to WHO

    Pretoria, South Africa, 28 June 2005

    E-mail:[email protected]

  • 8/4/2019 Validation2_0506_1_

    2/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 2/33

    WHO GMP and Guidelines WHO good manufacturing practices (GMP):

    main principles for pharmaceutical products

    Section 4. Qualification and validation

    Section 13. Equipment (see notes page)

    Supplementary guidelines on good

    manufacturing practices (GMP): Validation

    (2003) Draft.

  • 8/4/2019 Validation2_0506_1_

    3/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 3/33

    Causes of variation Man (different operators - lack of proper training)

    Machine / equipment

    Measurement (lack of calibration)

    Method (accuracy of validated processes/methods)

    Material (batch-to-batch variation of APIs/excipients)

    Manufacturing environment (T and RH)

  • 8/4/2019 Validation2_0506_1_

    4/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 4/33

    Qualification QUALIFICATION is the Action of proving that any

    premises, (pharmaceutical utility) systems and items ofequipment (e.g., machines, stability chamber)work

    correctly and actually lead to the expected results. Themeaning of the word validation is sometimesextended to incorporate the concept of qualification.

    REQUALIFICATION is the main part of thepreventive

    maintenance programme of proving that anypremises,(pharmaceutical utility) systems and items of equipmentwork correctly and keep on leading to the expected

    results. (normal wear and tear)

  • 8/4/2019 Validation2_0506_1_

    5/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 5/33

    Validation VALIDATION is the Action of proving, in

    accordance with the principles of GMP, that anyprocedure,process, equipment, material, activity

    or (pharmaceutical utility) system actually leadsto the expected results (see also qualification).

    REVALIDATION is a part of the change

    control system of proving that any procedure,process, equipment, material, activity or (pharmaceutical utility) system actually keepson leading to the expected results.

  • 8/4/2019 Validation2_0506_1_

    6/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 6/33

    Qualification - ValidationRegulatory

    requirements

    DQ,IQ,OQ

    inputs

    DQ,IQ,OQ

    process

    DQ,IQ,OQ

    outputsProcess

    Qualification

    Validation

    Premises, equipment and supporting utilities must be

    qualified to operate in a validated process.

  • 8/4/2019 Validation2_0506_1_

    7/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 7/33

    Relationship between qualification

    and validation

    DESIGN SPECIFICATION/QUALIFICATION

    INSTALLATION QUALIFICATION

    OPERATIONAL QUALIFICATION

    PROCESS VALIDATION

    CHANGE CONTROL

    Qualification is a sub-set of validation.

  • 8/4/2019 Validation2_0506_1_

    8/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 8/33

    CalibrationThe set of operations that establish, under specified

    conditions, the relationship between values indicated

    by an instrument or system for measuring(especially weighing), recording, and controlling, or

    the values represented by a material measure, and

    the corresponding known values of a reference

    standard. Limits for acceptance of the results ofmeasuring should be established. (see also the

    notes page) Verification.

  • 8/4/2019 Validation2_0506_1_

    9/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 9/33

    Calibration test equipmentInstrument or device used to calibrate other

    instruments, the calibration of which is traceable

    back to accepted national standards. The testequipment should have greater precision, accuracy,

    and repeatability, than the instruments being

    calibrated. These instruments should have

    documentary evidence of traceable calibration to atleast the same standard as the site calibration.

  • 8/4/2019 Validation2_0506_1_

    10/33

    DESIGN

    QUALIFICATION

  • 8/4/2019 Validation2_0506_1_

    11/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 11/33

    4.3 Documentary evidence(a) the premises, supporting utilities, equipment and

    processes have been designed in accordance with

    the requirements for GMP (design qualification orDQ);

    (b) the premises, supporting utilities and equipment

    have been built and installed in compliance with

    their design specifications (installation qualificationor IQ);

  • 8/4/2019 Validation2_0506_1_

    12/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 12/33

    4.3 Documentary evidence(c) the premises, supporting utilities and equipment

    operate in accordance with their design

    specifications (operational qualification or OQ);(d) a specificprocess will consistently produce a

    product meeting its predetermined specifications

    and quality attributes (process validation or PV,

    also calledperformance qualification or PQ).(PQ is frequently applied to utility systems)

  • 8/4/2019 Validation2_0506_1_

    13/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 13/33

    Basic principles Throughout this presentation the term equipment is

    used to represent both manufacturing machinery,equipment and QC instruments.

    The equipment should be correctly installed inaccordance with an installation plan, as persupplierand any special (purchaser) requirements.

    The requirements forcalibration, maintenance andcleaning developed as draft procedures should bereviewed and finally issued as authorised standardoperating procedures (SOPs).

  • 8/4/2019 Validation2_0506_1_

    14/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 14/33

    Basic principles Operating requirements should be established and

    tests conducted to assure equipment is operating

    correctly, undernormal and worst caseconditions ( failure mode analysis)

    Operator training requirementspertaining to the

    new equipment should be finalised and

    documented.

  • 8/4/2019 Validation2_0506_1_

    15/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 15/33

    DQ andIQ DQ protocols and reports

    GMP

    VMP

    National law

    Engineering design and construction documents

    Do not start IQ before DQ has been completed!

    IQ/OQ protocols and reports Above inputs + machine manuals

    Separate VMPs forHVAC and watersystems

    Do not start OQ before IQ has been completed!

  • 8/4/2019 Validation2_0506_1_

    16/33

    INSTALLATION

    QUALIFICATION

  • 8/4/2019 Validation2_0506_1_

    17/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 17/33

    Prequalification of vendor Vendor specification

    familiarity with GMP

    installation, training and start-up support performance testing at the vendors site

    experience ofcurrent users

    cost analysis

    Once selected, the vendor may assist in preparingthe qualification protocols

  • 8/4/2019 Validation2_0506_1_

    18/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 18/33

    Pre-delivery checkForcomplicated orlarge pieces of equipment, a

    pharmaceutical manufacturer may elect to undertake a pre-

    delivery check of the equipment at the supplier's assembly

    facility. This pre-delivery check cannot substitute for the

    Installation Qualification. However, it is acknowledged that

    the checks conducted and documented at this stage may

    duplicate a number of the checks conducted at the

    Installation Qualification stage, hence, there could be areduction in the scope of the Installation Qualification

    checks.

  • 8/4/2019 Validation2_0506_1_

    19/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 19/33

    Checking at usersInstallation Qualification requires a formal andsystematic check of all installed equipment against theequipment supplier's specifications and additionalcriteria identified by the user as part of the purchasespecifications. At the Installation Qualification, allequipment, gauges and services should be given a serial(or other reference) numberand a check conducted that

    the installed equipment (or plant) has been installed inaccordance with the current (approved) version of thePiping & Instrument Diagram (P&ID).

  • 8/4/2019 Validation2_0506_1_

    20/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 20/33

    IQ - Overview StatementIdentification and documenting ofmaintenance

    requirements for each installed item and the

    collection and collation ofsupplier operating andworking instructions, maintenance and cleaning

    requirements, should form the minimum

    documentation for a satisfactory Installation

    Qualification.

  • 8/4/2019 Validation2_0506_1_

    21/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 21/33

    Installation of equipmentAn effective Change Managementprocedure should

    be in place. All changes to the original design

    criteria should be documented and after that,appropriate modifications made to Equipment

    Specifications, Plant Functional Specifications and

    Piping & Instrument Diagrams (P&IDs).

  • 8/4/2019 Validation2_0506_1_

    22/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 22/33

    CalibrationRequirements Confirmation of calibration with reference to the

    appropriate national standard.

    Calibration of measuring devices utilised in the

    Operational Qualification stage, where confirmation of

    calibration is unavailable.

    Calibration of measuring devices related to installed

    equipment.

    Identification ofcalibration requirements for measuring

    devices for the future use of the equipment

  • 8/4/2019 Validation2_0506_1_

    23/33

    OPERATION

    QUALIFICATION

  • 8/4/2019 Validation2_0506_1_

    24/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 24/33

    OQ - Overview StatementOperational Qualification is an exercise oriented tothe engineering function, generally referred to ascommissioning. Studies on the critical variables(parameters) of the operation of the equipment orsystems will define the critical characteristics foroperation of the system or sub-system. All testingequipment should be identified and calibrated before

    use. Test methods should be authorised,implemented and resulting data collected andevaluated.

  • 8/4/2019 Validation2_0506_1_

    25/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 25/33

    OQ - Overview Statement It is important at this stage to assure all

    operational test data conform with pre-determinedacceptance criteria for the studies undertaken.

    It is expected that during the OperationalQualification stage the manufacturer shoulddevelop draft standard operating procedures(SOPs) for the equipment and services operation,

    cleaning activities, maintenance requirements andcalibration schedules.

  • 8/4/2019 Validation2_0506_1_

    26/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 26/33

    OQ - Essential Elements The conduct of an Operational Qualification should follow an

    authorised protocol.

    The plans for the Operational Qualification should identify the

    studies to be undertaken on the critical variables, the sequence

    of those studies and the measuring equipment to be used and

    the acceptance criteria to be met.

    Studies on the critical variables should incorporate specific

    details and tests that have been developed from specialistknowledge of the process and how the equipment will work

    (defined in design criteria and specifications).

  • 8/4/2019 Validation2_0506_1_

    27/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 27/33

    OQ - Essential Elements Where applicable, simulated product may be used to conduct

    the OQ. Studies on the critical variables should include acondition or a set of conditions encompassing upper and

    lower processing or operating limits and circumstances;commonly referred to as "worst case" conditions. Suchconditions should not necessarily induce product or processfailure.

    The completion of a successful OQ should allow the

    finalisation of operating procedures and operator instructionsdocumentation for the equipment. This information should beused as thebasis for training of operators in the requirementsfor satisfactory operation of the equipment.

  • 8/4/2019 Validation2_0506_1_

    28/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 28/33

    OQ - Essential Elements Draft cleaning procedures developed at the IQ stage

    should be finalised. Where applicable, these procedures

    should be validated as part of the PQ phase. The completion of satisfactory IQ and OQ exercises

    should permit a formal "release" of the equipment/plant

    for the next stage in the validation exercise (Process

    Validation). The release should take the form of writtenauthorisations for both Installation Qualification and

    Operational Qualification.

  • 8/4/2019 Validation2_0506_1_

    29/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 29/33

    OQ - Essential Elements Some OQ testing should be performed following

    equipment maintenance

    Current regulations require the validation ofcomputers and software that are used to control

    laboratory equipment or process data.

    The overall program should be evaluated

    periodically to ensure that current requirementsare met.

  • 8/4/2019 Validation2_0506_1_

    30/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 30/33

    Re-Qualification Modifications to, orrelocation of, equipment

    should only follow satisfactory review andauthorisation of the documented change proposalthrough the change control procedure. Part of thereview procedure should include consideration ofre-qualification of the equipment. Minor changesor changes having no direct impact on final or in-

    process product quality should be handledthrough the documentation system of the

    preventative maintenanceprogramme.

  • 8/4/2019 Validation2_0506_1_

    31/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 31/33

    Qualification of in-use equipmentWhile it is not possible to undertake the details of anIQ for established equipment nor the detailedapproach for an OQ, nevertheless there should bedata available that support and verify the operating

    parameters and limits for the critical variables of theoperating equipment. Additionally, the calibration,cleaning, preventative maintenance, operating

    procedures and operator trainingprocedures for theuse of the equipment should be documented and inuse as standard operating procedures (SOPs).

  • 8/4/2019 Validation2_0506_1_

    32/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 32/33

    Main points again Equipment is one of the potential causes of

    variation.

    DQ, IQ, OP and PQprotocols and reports arerequired that equipment was designed, installed and

    operated in accordance with suppliers and users

    requirements.

    Effective change control keeps documentation upto date.

    Qualification is a sub-set of validation

  • 8/4/2019 Validation2_0506_1_

    33/33

    2005.06.28. Dr. Pogny - WHO, Pretoria 33/33

    Literature Validation Master Plan, Installation and

    Operational Qualification, Non-sterile ProcessValidation, Cleaning Validation, Pharmaceutical

    Inspection Convention (PIC/S), August 2001

    Model IQ/OQ protocol and report (distributed amongparticipants of the training course)

    Comparison ofWHO, EU and US-FDA GMPRequirements for Pharmaceutical Equipment(distributed among participants of the training course)