Number of fi ltration columns
Number of fi lters
Number of IP44 fans
Average volume of treated air
Average face velocity
Internal volume of the cabinet
Renewals inside the cabinet / min
Total electrical power consumption Voltage-
Frequency
LED light intensity
Noise level
Packaging: 100% recycled wooden boxwith international phytosanitary certifi cate
TECHNICAL FEATURES
Volume
Weight
USESGeneral chemistry involving small volumes of reagents or chemical compounds at ambient/moderate temperature in all types of laboratories:
√ Research laboratories √ Quality control laboratories √ Clinical and hospital laboratories √ University and school laboratories
…In general, in any kind of laboratory.
DUCTLESS FUME HOOD
CERTIFIED –Made in Barcelona and certifi ed by an external laboratory accordingto international standards, and complying with the criteria of ISO 9001 standard.
PLUG&PLAY –It is sent from our warehouse assembled, and whenunpacked itjust needs to be connected into a plug.
FLEXIBLE –It can be used in hard-to-duct areas such as the center or bottomlevel of multi-level buildings.
TURNKEY –Installation expenses are far less than traditional hoods becauseno ductwork and remote blower are required.
GREEN & SUSTAINABLE –Unlike traditional fume hoods, costly tempered roomair is not exhausted from the laboratory, resulting in lower energy costs.
The new Cruma G-3 is perfect to remove low quantities of gaseous pollutingagents and/or solid particles/aerosols from its working area in a simple, secure,effi cient and cost effective way, protecting both the user and the environment.
Cruma G-3 ductless fume hood uses the patented* Cruma Filtration System,without any exterior duct connection. All molecular and dust particles areabsorbed and retained into the fi ltration system. *Invention Patent No.2397598
SIZES (mm)External Dimensions Internal Dimensions
Width1197 1176850 8001195 762
Depth Height Width Depth Height
www.cruma.es [email protected]
Because we are convinced of the qualityof our products.
It is not a typographical error,7 year warranty
7
Well done, well shipped.Our responsible packaging
Wood box 100% recyclable with international phytosanitary certifi cate.
*More informationwww.cruma.es
Do you need help ortechnical assistance?
Contact your distributor or call us if you have any questions or need technical support, spare parts,
maintenance service...
+34 93 370 61 62
2
2 to 6
2
157 m3/h
0,50 m/s
0,67 m3
4,2
174 W
110-220 V / 50-60 Hz
900 Lux
45 dB
1,24 m3
130 Kg
NEW FEATURES
More information on thenew LCD display
√ New size 127x34mm display √ Air speed continuously monitored √ Type of fi lter installed, working hours, expiration date and next
revision date √ Open door warning through electronic photocell √ Countdown timer √ Clock and calendar
New features and components √ Initial air fl ow cycle adequacy and fi nal purge cycle √ Fault LED √ Control of air fl ow through Microprocessor √ Activated carbon fi lters with electronic chip √ Internal temperature sensor √ LED illumination
New alarms and scheduled warnings √ Open door warning √ Open door in off mode warning √ 60h of fi lter use warning √ Next validation warning √ Few hours of fi lter life warning √ Countdown timer warning √ Expired fi lter alarm (by hours) √ Expired fi lter alarm (by date) √ Temperature alarm √ Equipment without fi lter alarm √ Low barrier alarm
New size 127x34mm display
USES
to international standards, and complying with the criteria of ISO 9001 standard.
agents and/or solid particles/aerosols from its working area in a simple, secure,effi cient and cost effective way, protecting both the user and the environment.
DUCTLESS FUME HOOD
SIZES (mm) 1197
1192
792
632
1176
304
210
850
1197
760 850
119
5 7
62
304
5
50
SERIAL EQUIPMENTElectronic circuit with large format LCD screen
Electronic anemometer device Photocell sensor devicefor open door detection Electronic control devicefor filters replacement
Illumination
Temperature sensor
Sampling system to analize the filtered air at the exhaust 60 hours alarm
Electronic cronometrewith audible alarm Clock and calendar Working surface 1
G4 Prefilter
Cable entry holes (2) Chemical Listing Warranty
Security levels: level 1 for users and level 2 for maintenance users Electronic sensor monitoring continuously air face velocity Electric device with open door alarm
The fi lters incorporate a microchip with miniUSB connection that identifi es the type of fi lter installed,the expiry date and the serial no. 96 LED Tube high light intensity and low power consumption - 16 Watts / 700 Lux Continuous monitoring of the temperature insidethe cabinet
To detect the level of fi lter saturation
Countdown timer according to French NF X 15-211:2009 To program the work inside the fume hood
Display of date and time Spill retention tray (2-10 liters) with working surface in white tempered glass G4 class pre-fi ltering blanket of synthetic biofi bres (according to EN-779) for the retention of atmospheric dust Access to the rear wall for cables and / or services entry Guide of retained products by type of fi lter 7 years
OPTIONAL EQUIPMENTMolecular detector
Movilair Tubular steel stand Working surface 2
Working surface 3
Transparent rear back pannel
Voltage / Frequency Filter test kit
Junction frame
Automatic Alarm Device for detection of saturation in organic vapors fi lters (required for Class I according to standard NF X 15-211:2009) Stand with wheels and internal tray in Epoxy coated steel Support stand in Epoxy coated steel Spill retention tray (2-10 liters) with working surfacein phenolic resin Spill retention tray (2-10 liters) with working surfacein inox steel Transparent polymethylmethacrylate rear pannel8 mm thick (light transmission of 93%).Ideal for teaching sessions 125 V / 50 Hz Dräger pump with reactive colorimetric tubes (pack 10u)
Allows to join two units of the same model without internal divisions
MAIN STRUCTURE
Metal parts: baseframe, rear wall and head
1.2 mm galvanized coated steel with anti acidpolymer resin powder heat-hardened at 200 °C
Transparent polymethylmethacrylate 6 mm thick (light transmission of 93%)Front and side panels
ACCORDING TO STANDARDSCabinets / Fume Hoods
Filters
Quality
AFNOR NF X 15-211:2009 (France)BS 7258: 94 (UK)BS 7989: 2001 (UK)ANSI/ASHRAE 110-1995 (USA)
CSA Z 316.5-94 (Canada)EN-779: 1996 (HEPA & ULPA Filters)EN-1822:1998 (HEPA & ULPA Filters)EN-141:2001 (Gas Filters)
UNE EN ISO 9001:2008
DUCTLESS FUME HOOD
MODULAR FILTRATION COLUMN FOR GASES AND PARTICLES (according to NFX 15-211:2009)CLASS 2 CLASS 1
Type GHandling of liquid compounds/products
Type 2GLiquid compounds/products handling with security molecular fi lter
FILTRATION COLUMN FOR POWDERS
Type DHandling of powder compounds
Fan
Molecular Filter
HEPA-H14 Filter
Type GSHandling of liquid and particles compounds/products
Type 2GDHandling of liquid compounds in cleanroom with molecular security fi lter
Type 2GSLiquid and particles compounds/products handling with molecular security fi lter
Type 2DDHandling of powder and molecular compounds in clean room with molecular security fi lter
Type DDHandling of powder compounds in clean room
FILTER TYPES
Type A Type K
Type F
Type BE
Type D
Type ABEK
For organic vapors such as ketones, ethers, alcohols, xylenes ... Eventually it can be used for inorganic acids, but only if used in small quantities because this activated carbon is not impregnated and the excess of acid vapors could saturate it quickly.
For NH3 vapors and amines; also good for other organic compounds. Carbon with metal salt complexes impregnation.
For formaldehyde vapors and derivatives; also good for other organic compounds. Carbon impregnated with Cu leads, so that it shouldnever be used with inorganic acid vapors.
For inorganic acid vapors as H2SO4, HCl, HNO3, and volatile sulfur compounds such as H2S, SO3, ... It can be used with organic vapors because the activated carbon incorporates impregnation of metal compounds and neutralizing salts. It is also suitable to fi lter organic and inorganic compounds when they are in similar proportions.
HEPA H-14 fi lter (High Effi ciency Particulate Air, according toEN-1822: 1998) for fi ltering dust and smoke particles.
Mixed type to be used when the ratios between organic, inorganic and NH3/amines are similar.
Information storage and interactionwith the alarm system.
Filters with microchip and USB connection,intelligent fi ltering
we recognise our responsibilityand dependence towards a healthy environment
and, therefore, donate 1% of our annual sales to environmental organisations
around the world.A. Canals
DISTRIBUTED BY:
Pol. Ind. Estruch - Carrer Moreres, 51 - 08820 El Prat de Llobregat - BarcelonaT. +34 93 370 61 62 - [email protected] - www.cruma.es
&