68
User’s Manual Yokogawa Electric Corporation AQ2200-136 TLS Module Optical Passive Device Measurement and Analysis IM 810518904-03E 1st Edition

User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

User’sManual

Yokogawa Electric Corporation

AQ2200-136 TLS ModuleOptical Passive DeviceMeasurement and Analysis

IM 810518904-03E1st Edition

Page 2: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E i

Introduction Thank you for your purchasing of this AQ2200-136 TLS Module. This user’s manual describes operating procedures necessary to measure the performance of an optical passive device with the TLS module combined with other modules, which can be mounted on the AQ2201/AQ2202.

Two kinds of manuals including this manual shown below are provided for this module. Therefore, you must thoroughly read both the manuals before starting the measurement. Manual Item Manual No. Description AQ2200-136 TLS Module User’s Manual [Read this first.]

IM810518904-01E This manual describes functions of the TLS module and proper operating procedures (except for application functions).

AQ2200-136 TLS Module Optical Passive Device Measurement and Analysis User’s Manual

IM810518904-03E This user’s manual. This manual describes how to operate application functions.

Notes The contents of this manual are subject to change without prior notice as a result of continuing improvements to the instrument's performance and functions. The figures given in this manual may differ from the actual screen. Every effort has been made in the preparation of this manual to ensure the accuracy of its contents. However, should you have any questions or find any errors, please contact your nearest YOKOGAWA dealer. Copying or reproducing all or any part of the contents of this manual without the permission of Yokogawa Electric Corporation is strictly prohibited.

Trademarks This unit uses MontavistaTM Linux®. MontavistaTM is a registered trademark of Montavista Software, Inc. Linux and Linux® is a registered trademark of Linus Torvalds.

Revisions • 1st Edition: October 2004

Page 3: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

ii IM 810518904-03E

Product Versions Covered by This Manual This manual is applicable to the firmware version “1.00” or more of the AQ2200-136 TLS Module.

Page 4: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E iii

1 1

2 2

3 3

4 4

Contents Introduction................................................................................................................. i Product Versions Covered by This Manual ............................................................... ii

Chapter 1 About Application Function 1.1 TLS Application.............................................................................................1-1

Chapter 2 Swept Measurement 2.1 Functional Description ..................................................................................2-1 2.2 Flow of Measurement Steps .........................................................................2-2 2.3 Starting Up the Application............................................................................2-3 2.4 Operating the Wavelength Calibration and ZERO-SET................................2-5 2.5 Setting the Sweep Conditions.......................................................................2-6 2.6 Setting the Measurement Dynamic Range ...................................................2-7 2.7 Starting the Reference Measurement ...........................................................2-8 2.8 Starting the Measurement...........................................................................2-11 2.9 Displaying the Graph ..................................................................................2-13 2.10 Saving the Measurement Results...............................................................2-14 2.11 Analyzing the Performance of the WDM Filter............................................2-16 2.12 Analyzing the Performance of the Flat-type Device....................................2-31

Chapter 3 Manual Measurement 3.1 Functional Description ..................................................................................3-1 3.2 Flow of Measurement Steps .........................................................................3-2 3.3 Starting Up the Application............................................................................3-3 3.4 Operating the Wavelength Calibration and ZERO-SET................................3-5 3.5 Starting the Manual Measurement................................................................3-6

Chapter 4 Remote Control 4.1 General Description of Commands...............................................................4-1 4.2 Commands for Starting Up and Exiting of the Swept Function.....................4-2 4.3 Commands for Specifying of Sweep Area Only by Wavelength ...................4-3 4.4 Commands for Setting of Sweep Area by Center Wavelength,

Span, and Number of Samples.....................................................................4-5 4.5 Commands for Setting of the Sweep Mode ..................................................4-7 4.6 Commands for Setting of Sweep Power Conditions.....................................4-9 4.7 Commands for Setting of OPM Conditions in the Swept Functions ...........4-11 4.8 Commands for Reference Measurement....................................................4-12 4.9 Commands for Measurement .....................................................................4-14 4.10 Commands for Data Save and Analysis .....................................................4-16 4.11 Commands To Be Set Disabled during Execution of Swept Function ........4-18

Page 5: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

Chapter 1 About Application Function

IM 810518904-03E 1-1

2 1

About Application Function

1.1 TLS Application This TLS application function measures the insertion loss (IL) of the performance of a passive component used for the optical network, such as WDM network with this unit to be mounted on the AQ2201/AQ2202 Frame Controller combined with OPM module. Measurement result display, analysis, and data save functions of each measurement are provided. The application using the TLS module provides two measurement functions as described below.

Swept: By synchronizing with the OPM module, an optimal sweep type is selected from three sweep types to measure the IL. By setting up a threshold level, the OK or NG judgement of the analysis item can be performed easily.

Manual: The wavelength sensitivity correction value of the OPM module can be changed automatically while linking with the wavelength set using the TLS module. Use of this function makes it possible to perform the measurement with the wavelength display linked with the measurement result of the OPM module.

Pressing the "APPLI" key on the frame controller will start up this TLS application function.

ENTER CANCEL

AQ2201 FRAME CONTROLLER

POWER

OFF ON

0

1 2 3

4 5 6

7 8 9

FRAME DETAIL CHAN

USER1

USER2

USER3

HOLD

SYSTEM

APPLI

PRESET SHIFT

DISP TOP

BS -

ENTER CANCEL

AQ2201 FRAME CONTROLLER

POWER

OFF ON

0

1 2 3

4 5 6

7 8 9

.

FRAME DETAIL CHAN

USER1

USER2

USER3

HOLD

SYSTEM

APPLI

PRESET SHIFT

DISP TOP

BS -

Select a measurementfunction.

Press this "APPLI" key to call up the application screen.

Page 6: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

Chapter 2 Swept Measurement

IM 810518904-03E 2-1

2

Swept Measurement

2.1 Functional Description This Swept function provides wide wavelength band, high-speed sweep, and high-speed time measurement of the IL by combining the TLS module with the optical power meter (OPM) module. Use of this function makes it possible to measure the IL characteristics of the single port having wavelength characteristics of multi-port device or filter, such as MUX/DEMUX for CWDM or splitter within a short time.

TLS Module + OPM Module

Object device to be measured

This measurement function is used with the TLS and OPM modules mounted inside the frame controller.

IL measurement Measurement results can be displayed on the screen or output to the outside using a CSV file.

Page 7: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2-2 IM810518904-03E

2.2 Flow of Measurement Steps The following shows the outline flow necessary to operate this unit.

Setting of analysis conditions

Displaying of graph

Saving of measurement results

• Analysis of multi-port device

Marker display, zoom-in function, and graph screen save

• Auto data save function

Starting up of application screen

Setting of combination of TLS and OPM

Setting of wavelength range and sweep span

Flow of Measurement Steps

Calibration and Zero-set

Setting of output level and line width

Setting of sweep type Setting of measurement

dynamic range

Setting of other parameters

Reference measurement Starting of measurement

Summarization of Measurement Data

Completion of measurement

Analysis

• Analysis of single-port device Convenient Functions when Set Enabled.

Measurement data auto save Auto analysis after measurement Auto generation of file name to be saved Monitor function

• Saving of the measurement results into compact flash memory in the CSV format

Page 8: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-3

2

Swept Measurement

2.3 Starting Up the Application This section describes how to start up the screen necessary to operate the SWEPT function.

Operating Procedures

[Operation 1] Select a desired item.

[Operation 2] Select the Swept item.

(To the next page.)

Note Do not press the [APPLI] key while the WL Cal or ZERO-SET is

running. If the [APPLI] key is pressed while the WL Cal or ZERO-SET is running on the Detail screen, the operation is forcibly aborted and the APPLICATION screen is then started up. Therefore, WL Cal and ZERO-SET become invalid.

Press the [APPLI] key to start up the "APPLICATION" screen. Move the cursor to "TLS APPLICATION" with the [ ] or [ ] cursor key, and then press the [ENTER] key or <OK> function button. The "TLS APPLICATION" popup screen will appear. Press the [CANCEL] key or <Close> function button to exit the screen.

Press [ENTER] or Press<OK>.

Move the cursor to "Swept" with the [ ] or [ ] cursor key, and then press the [ENTER] key or <OK> function button. The next screen (MODULE SELECT for SWEPT) will appear. Press the [CANCEL] key or <Close> function button to close the popup screen.

Press [ENTER] or Press<OK>.

Page 9: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.3 Starting Up the Application

2-4 IM810518904-03E

[Operation 3] Select a module you want to combine. Select a module you want to combine with the [ ] or [ ] cursor key, and then press the <Select> function button. When a TLS module is selected, the color of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the color of this check box is changed from white to black.

An optimal OPM module can be selected from two types, that is, AQ2200-211 and AQ2200-231. If one TLS module and one OPM module are not selected, the next screen is not displayed even though the <OK> function button or [ENTER] key is pressed.

[Completion of Operation] Note Since the AQ2201 Frame Controller is a type of 3-slot model, select all

modules to change the color of the check box to green (black for AQ2200-211 and AQ2200-231).

When checking on relevant check box, the dimmed <OK> function button is changed to the normal button display. At this time, when pressing the <OK> button or [ENTER] key, the main screen of the Swept function, "SWEPT APPLICA- TION", will appear.

Press [ENTER] or Press<OK>.

Swept main screen of Application function

Page 10: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-5

2

Swept Measurement

2.4 Operating the Wavelength Calibration and ZERO-SET

When operating this unit, you must calibrate the wavelength every time the laser output is turned ON. This section describes how to calibrate the wavelength. The Swept measurement cannot be started unless the wavelength calibration (WL Cal) is performed. Therefore, you must always calibrate the wavelength before staring the measurement.

Operating Procedures Note When canceling the operation during execution of the wavelength

calibration, the wavelength calibration, which has been performed until that time, becomes invalid. Therefore, before starting the operation, perform the wavelength calibration again to complete it.

This indication is shown if “WL Cal” is not completed. When “WL Cal” is completed, this indication disappears.

Press the <Zero> button to perform ZERO-SET of the OPM module. For details about ZERO-SET, see Chapter 3 in the User’s Manual for AQ2201 Frame Controller.

Press the <WL Cal> button to start the wavelength calibration. The execution screen appears while the wavelength calibration is running. The execution screen will disappear when the wavelength calibration is completed. Therefore, you need to wait until the execution screen disappears. For details about wavelength calibration, see Chapter 5 in the User’s Manual for AQ2200-136 TLS Module.

Screen during Execution of Wavelength Calibration

Screen during Execution of ZERO-SET

Page 11: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2-6 IM810518904-03E

2.5 Setting the Sweep Conditions This section describes how to set up the sweep wavelength parameters, such as the sweep start wavelength, stop wavelength, and interval.

Setting the Sweep Area, Sweep Mode, Line Width, and Power Operating Procedures

Press the <TLS PRMTR> button. Select a sweep item on the popup screen to set it. For details about sweep, see the following Chapters in the “AQ2200-136 TLS Module User’s Manual” . • Start, WL, Stop WL, StepWL, Sweep Mode: Chapters 9 and 10 • Power: Chapter 8 • Linewidth: Chapter 7 • WLMC Settle: Chapter 12

Explanation Appropriate sweep conditions are used corresponding to the applications shown below.

Overall wavelength needs to be measured. : StartWL/StopWL/StepWL One wavelength and its surrounding need to be measured. :

CenterWL/Span/Samples Measurement needs to be performed by increasing the measurement dynamic range. : Select “Step” for SweepMode setup.

Measurement results need to be known early. : Select “Continuous” for SweepMode setup. Note A sweep step of 0.001nm or less cannot be set.

20,000 or more measurement points cannot be set. Since the sweep wavelength range is determined by the sweep span and the number of measurement points, make the setting so that the wavelength range you want to measure is satisfied.

example: When the sweep span is set at “0.001nm” and the number of measurement points is set at “20000”, the sweep wavelength range becomes as shown below. 0.001nm × 20000 = 20nm

TLS PRMTRStart WLStop WLStep WLSweep ModeCenter WLSpanSamplesPowerLinewidthWLMC Settle

TLS PRMTRStart WLStop WLStep WLSweep ModeCenter WLSpanSamplesPowerLinewidthWLMC Settle

Two kinds of sweep wavelength settings are provided and either setting becomes valid. When selecting a desired setting, it becomes valid and other setting indication is dimmed showing that it is invalid.

Page 12: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-7

2

Swept Measurement

2.6 Setting the Measurement Dynamic Range This section describes how to set up the measurement dynamic range for the sweep. Applicable dynamic range may vary depending on the sweep mode. Auto1 Auto2 Hold Stepped Manual Continuous

Operating Procedures For details about "Average" and "Range Mode", see Chapter 3 in the User’s Manual for AQ2201/AQ2202 Frame Controller.

10dBm - 10dBm - 30dBm - 50dBm - 70dBm

AUTO1

AUTO2

HOLD(High)

Range type

HOLD(Middle)

HOLD(Low)

10dBm - 10dBm - 30dBm - 50dBm - 70dBm

AUTO1

AUTO2

HOLD(High)

Range

HOLD(Middle)

HOLD(Low)

Page 13: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2-8 IM810518904-03E

2.7 Starting the Reference Measurement Before starting the Swept measurement, you must always perform the reference measurement. Since the output characteristics of the TLS module cannot output the constant level to the wavelength axis, it is absolutely necessary to perform the reference measurement before measuring the device. This section describes how to operate the reference measurement.

Operating Procedures

When pressing the <Ref Meas> button, the "Ref Measurement" popup screen will appear, prompting you to check the execution. To start the reference measurement, press the <OK> button.

The progress bar showing the progress status is shown during measurement. Pressing the <Stop> button will abort the measurement. When the measurement is completed successfully, the "NO REF" indication will disappear. Note The reference measurement cannot be started unless the

wavelength calibration (WL Cal) is completed.

Press [ENTER] or Press<OK>.

Page 14: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.7 Starting the Reference Measurement

IM 810518904-03E 2-9

2

Swept Measurement

Explanation The following Table shows Ref valid/invalid when each parameter setting is changed.

Parameter Settings Valid, Invalid, Conditional

Item Ref Valid/Invalid Remarks Start WL Valid when the value is within the specified range of “Ref”.Stop WL Valid when the value is within the specified range of “Ref”.Step WL Valid when the value is within the specified range of “Ref”.Center WL Valid when the value is within the specified range of “Ref”.Span Valid when the value is within the specified range of “Ref”.Data Points Valid when the value is within the specified range of “Ref”.Sweep Mode Power Linewidth Range Mode WLMC Settle Average Sweep Cycles WL Cal Re-execution of WL Cal is invalid.

Module Settings

When using multiple OPM modules, the reference measurement data is valid even though any module is unmounted after completion of the reference measurement. (When at least one OPM module remains, which has been set during reference measurement, the reference measurement data is valid.)

However, if a new OPM module is mounted after completion of the reference measurement, the reference measurement data becomes invalid. At this time, you must perform the reference measurement again. Even though the OPM module is unmounted, which has been measured actually with "ALL OPM" during reference measurement, the reference measurement data is valid.

Page 15: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.7 Starting the Reference Measurement

2-10 IM810518904-03E

About Invalid Reference Measurement Data If any of the following conditions arises, the reference measurement data becomes invalid.

• Power is turned ON again. • Module is added. • Reference measurement is aborted. • Conditions for reference measurement are not satisfied by setting the

parameters. • Wavelength calibration becomes invalid. • Laser output is turned OFF. • Application is changed. • <Ref Clear> is executed.

Conditions for Execution of Reference Measurement • Wavelength calibration is completed.

Page 16: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-11

2

Swept Measurement

2.8 Starting the Measurement Before starting the Swept measurement, it is absolutely necessary to complete the reference measurement. If the reference measurement is not performed or if the reference measurement is disabled, the Swept measurement cannot be started on the screen. Therefore, you must check that the reference measurement is completed and the "NO REF" indication disappears. The measurement can be started from any of the following screens.

Operating Procedures When the reference measurement becomes valid, the indication of the <Meas> button is changed from the dimmed display to the normal display. In this case, when pressing this <Meas> button, the Swept measurement screen is started up. On the Swept measurement screen, you can start the measurement, the graph screen, and/or the analysis screen. If there is no Swept measurement data, "NO DATA" is shown. Additionally, the screen is transited to the Swept measurement screen even when the reference measurement is completed successfully.

When the indication is dimmed, this shows that the reference measurement is not completed.

The popup screen showing the confirmation message appears. After checking the message, press the <OK> button. The measurement is then started.

SWEPT Measurement Main Screen

Screen Immediately after Completion of Reference Measurement

(Measurement Start Screen)

Measurement Result Display Screen

Page 17: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.8 Starting the Measurement

2-12 IM810518904-03E

To perform the swept measurement continuously, set “0” in the Sweep Cycles field. For details about how to set Sweep Cycles, see section 9.4 in the User’s Manual for AQ2200-136 TLS Module. To exit the continuous swept measurement, press the <Stop> button.

The progress bar showing the progress status is displayed during measurement. Pressing the <Stop> button will abort the measurement.

When the measurement is completed successfully, the "NO DATA" indication will disappear.

1. Press the [ENTER] key. 2. Input “0” and press the <OK> button.

Completion of Continuous Swept Measurement Setting

Page 18: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-13

2

Swept Measurement

2.9 Displaying the Graph You can display the measured data on the graph screen or save the graph screen. For details, see section 7.1.9 in the “AQ2201/AQ2202 Frame Controller USER'S MANUAL”.

Operating Procedures

Starting Up the Graph Screen

Press the <Graph> button. The graph screen will appear.

Press <Graph>

Graph Screen

For details about operation, see section 7.1.9 in the “AQ2201/AQ2202 Frame Controller USER'S MANUAL”.

Page 19: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2-14 IM810518904-03E

2.10 Saving the Measurement Results You can save the measurement results into the compact flash memory mounted on the rear of the AQ2201/AQ2202 Frame Controller. The measurement results are saved in the CSV format.

Operating Procedures

[Operation 1] Select a desired item.

[Operation 2] Specify a data format.

Press the <Save Data> button after the compact flash memory has been mounted on the rear of the frame controller. If the memory is not mounted, relevant error message appears.

Press <Save Data>

[Measurement Data only (Analysis Is Not Executed.)]

[Measurement Data and Analysis Result Data Exist.]

The screen on the left shows the status that the data is being saved. You should wait until the "DATA SAVE" popup screen disappears. If the compact flash memory is removed from the main unit before the popup screen disappears, the data may be corrupted.

To start the analysis, press the <Analyze> button.Press [ENTER] or Press<OK>

Page 20: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.10 Saving the Measurement Results

IM 810518904-03E 2-15

2

Swept Measurement

Setting the Auto Save When saving the measurement results, make the following settings to automatically save the data with a desired file name put.

<Auto Save> If “Auto Save” is turned ON, the results are saved automatically after the measurement has been completed.

<Input S/N> If “Input S/N” is turned ON, S/N character string you have set is used as a file name when saving the results. If “Input S/N” is turned OFF, a file name is generated when saving the results

Execution of Auto Save Execution of Input S/N

Page 21: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2-16 IM810518904-03E

2.11 Analyzing the Performance of the WDM Filter This section describes how to analyze the CWDM filter using an operation example. System parameter settings, device analysis items, and analysis threshold values are provided for the analysis settings. The dynamic range to be measured by this system depends on an object device to be measured and the performance of the TLS module. Therefore, you need to input appropriate judgement criteria (threshold values). The analysis method conforms to ITU-T G-671.

♦ System Parameter Settings

The following Table shows the parameter items for multi-port device and single-port device. Multi-port Device Single-port Device CHAN: Number of channels CHAN: Number of channels (1:Fixed) CHAN Space: Channel space WL Range: Wavelength range WL Range: Wavelength range Center WL: Center wavelength

♦ Device Analysis Items

The following Table shows the analysis items for multi-port device and single-port device. Multi-port Device Single-port Device IL: Insertion loss IL: Insertion loss Ripple Ripple Pass band Pass band Stop band Stop band CHAN ISLN ADJ − CHAN ISLN Non-ADJ −

Page 22: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-17

2

Swept Measurement

Step 1 Setting Up the Conditions for the Filter Type To Be Analyzed

Operating Procedures

[Operation 1] Start up the analysis setup.

[Operation 2] Select the “Type” item.

[Operation 3] Select a filter type.

Press the <ANAL PRMTR> button to start up the "ANALYSIS SETUP" screen.

Press <ANAL PRMTR>.

Press the <TYPE> button to display the "TYPE" popup screen. Or, move the cursor to "Type"

with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Press <TYPE> or Press[ENTER].

Move the cursor to "WDM Peak" with the [ ] or [ ] cursor key, and then press the <OK> button or [ENTER] key.

Press [ENTER] or Press<OK>.

Page 23: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-18 IM810518904-03E

[Operation 4] Set the number of system channels. When the value of "CHAN" is "1", the "CHAN Space" item becomes "Center WL". Set a wavelength value you want to analyze.

* System channel: The system channel is determined by the wavelength division multiplexing system of the device to be measured.

Example CWDM has 8 system channels.

Press the <CHAN> button to display the "CHAN" popup screen.

Or, move the cursor to "CHAN" with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the <OK> button or [ENTER] key. The setting range is “1” to “50”.The default value is “8”.

Press [ENTER] or Press<OK>.

Press <CHAN> or Press[ENTER].

Display item is changed.

Page 24: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-19

2

Swept Measurement

[Operation 5] Set a valid wavelength band (wavelength range) of the system channel.

Press the <WL Range> button to display the "WL Range" popup screen. Or, move the cursor to "WL Range" with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Press <WL Range> or Press[ENTER].

Input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the <OK> button or [ENTER] key. The setting range is “0.001” to “100.000nm”. The default value is “13.000nm”.

Press [ENTER] or Press<OK>.

The value you have set is then set.

Page 25: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-20 IM810518904-03E

[Operation 6A] Set a channel spacing . (The number of channels is other than "1".) For subsequent operations, read [Operation 7].

Press the <CHAN Space> button to display the "CHAN Space" popup screen.

Or, move the cursor to "CHAN Space" with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Press <CHAN Space> or Press[ENTER].

Input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the <OK> button or [ENTER] key. The setting range is “0.001” to “100.000nm”. The default value is “20.000nm”.

Press [ENTER] or Press<OK>.

The value you have set is then set.

Page 26: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-21

2

Swept Measurement

[Operation 6B] Set a channel space (interval). (The number of channels is "1".) For subsequent operations, read [Operation 7].

Press the <Center WL> button to display the "Center WL" popup screen. Or, move the cursor to "Center WL" with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Press <Center WL> or Press[ENTER].

Input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the <OK> button or [ENTER] key. The setting range is “1440.000” to “1640.000nm”. The default value is “1471.000nm”.

Press [ENTER] or Press<OK>.

The value you have set is then set.

Page 27: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-22 IM810518904-03E

[Operation 7] Set or change analysis conditions.

Press the <Next> button to display the "ANALYSIS SETUP" screen.

Press [Next].

When pressing <Next>, the analysis item screen, "ANALYSIS ITEM", will appear. For details about how to operate the analysis item, see "Operation 8".

When pressing the <EDIT> button, you can change the wavelength of CHAN at the cursor position.

Input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the <OK> button or [ENTER] key.

The setting range is “1440.000” to “1640.000nm”. The default wavelength set value is the basic wavelength of the CWDM.

* Change the value as required.

<Auto Input>: This function automatically inputs the value of each CHAN. By setting the Start wavelength or Stop wavelength, the value is input to other CHAN items at intervals set in "CHAN Space".

Page 28: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-23

2

Swept Measurement

Explanation "ANALYSIS SETUP" item is used to set up a type of the wavelength you want to analyze. The following Fig. shows the relationship between the setup item and wavelength. This example shows the filter for the CWDM. You must input a grid and an interval of the determined wavelength. You may select a filter you want to measure the performance and set each value.

CHAN1 CHAN2 CHAN3 CHAN4 CHAN5 CHAN6 CHAN7 CHAN8 Wavelengthλ [nm] 1471 1491 1511 1531 1551 1571 1591 1611

CHAN Space

WL Range

CHAN1 CHAN2 CHAN3 CHAN4 CHAN5 CHAN6 CHAN7 CHAN8 [nm] Example: 1471 1491 1511 1531 1551 1571 1591 1611

CHAN Space

WL Range

Page 29: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-24 IM810518904-03E

[Operation 8] Select an item you want to analyze.

[Operation 9] Set a threshold level. (1) Next, set a threshold level for the judgement during analysis. Press the <THR> button to open the Threshold Setup screen. Items displayed on the screen are those specified for the analysis. Items, which have not been specified, are not displayed.

Move the cursor to an item you want to analyze with the [ ] or [ ] cursor key, and then press the <Select> button. When the analysis item is specified, the check box on the left is indicated in black.

When the value in "CHAN" is “1”, the "CHAN ISLN ADJ" and "CHAN ISLN Non-ADJ" items are not displayed.

Press this button.

Press this button.

Page 30: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-25

2

Swept Measurement

[Operation 10] Set a threshold level. (2)

Move the cursor to an item you want to set a threshold level with the [ ] or [ ] cursor key, and then press the [ENTER] key or <Edit> button.

Press [ENTER] or Press<Edit>

The "IL" popup screen will appear. Input a value with the [ ] or [ ] key or through the ten-key pad, and then press the [ENTER] key or <OK> button.

Press [ENTER] or Press<OK>

The threshold level you have set is then reflected.

Page 31: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-26 IM810518904-03E

Explanation "ANALYSIS ITEM" shows parameters used to analyze whether or not the band of the WDM filter satisfies the specification. The parameters show the values of the following portions.

Center wavelength

Grid specified with ITU-T.

For the IL value, select a wavelength having larger insertion loss.

Ripple valueLevel lowered from the point intersecting with the grid.

IL value in this case. Passband value

Stopband value

Page 32: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-27

2

Swept Measurement

CHAN ISLN ADJ Value This value shows a difference in IL between the channel to be measured and the adjacent channel. A wavelength having larger insertion loss is selected for the IL value of the channel to be measured. A wavelength having smaller insertion loss is selected for the IL value of the adjacent channel.

CHAN ISLN Non-ADJ Value This value shows a difference in IL between the channel to be measured and the non-adjacent channel. A wavelength having larger insertion loss is selected for the IL value of the channel to be measured. A wavelength having smaller insertion loss is selected for the IL value of the non-adjacent channel.

Non-adjacent grid

Applicable grid

Adjacent grid

Non-adjacent grid

Adjacent grid

CHAN ISLN ADJ value

CHAN ISLN Non-ADJ value

Center wavelength

Page 33: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-28 IM810518904-03E

Step 2 Displaying the Judgment Results

Operating Procedures

[Operation 1] Execute the analysis. At this time, when <Auto Analysis> is turned ON, the analysis is executed automatically after completion of the measurement. As a result, the following analysis result judgement screen will appear.

[Operation 2] Display the analysis result (Summary).

Indication of synthetic judgment:

If at least one CHAN judgement of each OPM is determined as "NG", “NG” is shown. The background color is changed to red.

Indication of each OPM module: If at least one CHAN judgement is determined as "NG", “NG” is shown. The background color is changed to red. If the Peak value of OPM is not measured, "−" is shown on both the wavelength and result.

When pressing the <Analyze> button, the analysis is performed from the measurement results and the analysis results (ANALYSIS RESULT) are displayed.

Indication of synthetic judgement

Indication of each OPM module

Move the cursor to OPM you want to check the detailed measurement results on the Summary screen with the [ ] or [ ] key, and then press the <Detail> button. The Detail screen will appear. Pressing the <Close> button will close the ANALYSIS RESULT screen.

Analysis Result Judgment Screen[Summary Screen]

Page 34: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

IM 810518904-03E 2-29

2

Swept Measurement

[Operation 3] Display the analysis results (Detail). Judgment indication:

The background color of the item determined as NG becomes red. Item indication:

Only items you have set on the ANALYSIS ITEM screen are shown.

To check each threshold value on the Detail screen, move the cursor to a desired item with the [ ] or [ ] key, and then press the <THR> button to display the THR screen. When pressing the <SMY> button, the screen is returned to the Summary screen. Pressing the <Close> button will close the ANALYSIS RESULT screen.

Analysis Result Judgment Screen[Detail Screen]

Page 35: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.11 Analyzing the Performance of the WDM Filter

2-30 IM810518904-03E

[Operation 4] Set a file name when saving the data.

When pressing the <S/N> button, the serial No. setup popup screen will appear. On this popup screen, input a desired character string. S/N you have input can be used as a file name for data saving using a function to be described in the later part.

Press<S/N> Select characters you want to

input with the [ ] or [ ] key and press the [ENTER] key to set the character string you have input. When the character string is input, press the <OK> button to set the S/N indication.

The S/N No. is then set. When the setup is completed, press the <Close> button to close the screen.

Returns the cursor one character.

Advances the cursor one character.

Clears one character.Press<OK>

Page 36: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 2-31

2

Swept Measurement

2.12 Analyzing the Performance of the Flat-type Device

This section describes how to analyze the flat type device using an operation example. System parameter settings, device analysis items, and analysis threshold level are provided for the analysis settings.

♦ System Parameter Settings • Lower: Wavelength value at sweep start • Upper: Wavelength value at sweep end

♦ Device Analysis Items • IL Max: Maximum value of insertion loss • IL Min: Minimum value of insertion loss • Ripple: Ripple value

Explanation In the analysis of the flat filter, values shown in the Fig. below are measured, and then measurement results and judgement results are displayed.

[nm]

Power

Lower Upper ( (Wavelength at end)

λ [nm]

Power

Upper

IL Max value

(Wavelength at start)

Ripple value

IL Min value

Page 37: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

2-32 IM810518904-03E

Step 1 Setting Up the Conditions for the Filter Type To Be Analyzed

Operating Procedures

[Operation 1] Start up the analysis setup.

[Operation 2] Select the “Type” item.

[Operation 3] Select a filter type.

Press the <ANAL PRMTR> button to start up the "ANALYSIS SETUP" screen.

Press <ANAL PRMTR>.

Press the <TYPE> button to display the "TYPE" popup screen.Or, move the cursor to "Type" with the [ ] or [ ] cursor key, and then press the [ENTER] key.

Press <TYPE> or Press[ENTER].

Move the cursor to "Flat" with the [ ] or [ ] cursor key, and then press the <OK> button or [ENTER] key.

Press [ENTER] or Press<OK>.

Page 38: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

IM 810518904-03E 2-33

2

Swept Measurement

[Operation 4] Set a wavelength range.

The “Type” setting becomes "Flat".

To change the value in "Lower" or "Upper", press the <Lower> or <Upper> button to display the popup screen, on which you may input a desired value.

Press<Lower> or Press<Upper>

Page 39: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

2-34 IM810518904-03E

Step 2 Setting the Analysis Conditions

Operating Procedures

[Operation 1] Display the analysis item setup screen.

[Operation 2] Select an analysis item.

Press the <Next> button. The "ANALYSIS ITEM" screen will appear.

Press<Next>

Move the cursor to an item you want to analyze with the [ ] or [ ] cursor key, and then press the <Select> button. When the analysis item is specified, the check box on the left is indicated in black.

* "IL Max" and "IL Min" cannot be selected at the same time.

Page 40: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

IM 810518904-03E 2-35

2

Swept Measurement

[Operation 3] Set a threshold level.

When pressing the <THR> button, the Threshold setup screen will appear.

Press<THR>

Press [ENTER] or Press<Edit>

Items you have set for the analysis items are shown on the screen. Items, which have not been specified, are not shown. Move the cursor to an item you want to set a threshold level with the [ ] or [ ] cursor key, and then press the [ENTER] key or <Edit> button.

The popup screen will appear. Input a value with the [ ] or [ ] cursor key or through the ten-key pad. Press the [ENTER] key or <OK> button to set the value.

Page 41: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

2-36 IM810518904-03E

Step 3 Displaying the Judgment Results

Operating Procedures

[Operation 1] Execute the analysis. At this time, when <Auto Analysis> is turned ON, the analysis is executed automatically after completion of the measurement. As a result, the following analysis result judgment screen will appear.

[Operation 2] Display the analysis result (Summary). Indication of synthetic judgment:

If at least one CHAN judgment of each OPM is determined as "NG", “NG” is shown. The background color is changed to red.

Indication of each OPM module: If at least one CHAN judgment is determined as "NG", “NG” is shown. The background color is changed to red. If the Peak value of OPM is not measured, "−" is shown on both the wavelength and result.

When pressing the <Analyze> button, the analysis is performed from the measurement results and the analysis results (ANALYSIS RESULT) are displayed.

Indication of synthetic judgment

Indication of each OPM module

Move the cursor to OPM you want to check the detailed measurement results on the Summary screen with the [ ] or [ ] key, and then press the <Detail> button. The Detail screen will appear. Pressing the <Close> button will close the ANALYSIS RESULT screen.

Analysis Result Judgment Screen[Summary Screen]

Page 42: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

2.12 Analyzing the Performance of the Flat-type Device

IM 810518904-03E 2-37

2

Swept Measurement

[Operation 3] Display the analysis results (Detail). Judgment indication:

The background color of the item determined as NG becomes red. Item indication:

Only items you have set on the ANALYSIS ITEM screen are shown.

[Operation 4] Display the analysis results (Detail). Set a file name when saving the data. See (4) in section 2.11

[Operation 5] Saving the Judgment and Measurement Results See section 2.10.

To check each threshold value on the Detail screen, move the cursor to a desired item with the [ ] or [ ] key, and then press the <THR> button to display the THR screen. When pressing the <SMY> button, the screen is returned to the Summary screen. Pressing the <Close> button will close the ANALYSIS RESULT screen. Analysis Result Judgment Screen[Detail Screen]

Page 43: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

Chapter 3 Manual Measurement

IM 810518904-03E 3-1

2

Manual Measurement

3

3.1 Functional Description This Manual function automatically changes the wavelength correction value of the OPM corresponding to the set wavelength of the TLS.

TLS Module + OPM Module

Object device to be measured

Page 44: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

3-2 IM810518904-03E

3.2 Flow of Measurement Steps The following shows the outline flow necessary to operate this unit.

Starting up of application screen

Setting of wavelength Setting of measurement

dynamic range

Starting of measurement

Displaying of measurement results

Numeric values are displayed.

Flow of Measurement Steps

Calibration and Zero-set

Setting of combination of TLS and OPM

Numeric values are set. Conditions, such as Range and Avg, etc.

Page 45: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 3-3

2

Manual Measurement

3

3.3 Starting Up the Application This section describes how to start up the screen necessary to operate the Manual function.

Operating Procedures

[Operation 1] Select a desired item.

[Operation 2] Select the Manual item.

(To the next page.) Note Do not press the [APPLI] key while the waveform calibration or

ZERO-SET is running. If the [APPLI] key is pressed while the waveform calibration or ZERO-SET is running on the Detail screen, the operation is forcibly aborted and the APPLICATION screen is then started up. Therefore, waveform calibration and ZERO-SET become invalid.

Press the [APPLI] key to start up the "APPLICATION" screen. Move the cursor to "TLS APPLICATION" with the [ ] or [ ] cursor key, and then press the [ENTER] key or <OK> function button. The "TLS APPLICATION" popup screen will appear. Press the [CANCEL] key or <Close> function button to exit the screen.

Press [ENTER] or <OK>.

Move the cursor to "Manual" with the [ ] or [ ] cursor key, and then press the [ENTER] key or <OK> function button. The next screen (MODULE SELECT for MANUAL) will appear. Press the [CANCEL] key or <Close> function button to close the popup screen.

Press [ENTER] or <OK>.

Page 46: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

3.3 Starting Up the Application

3-4 IM810518904-03E

[Operation 3] Select a module you want to combine. Select a module you want to combine with the [ ] or [ ] cursor key, and then press the <Select> function button. When a TLS module is selected, the color of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the color of this check box is changed from white to black.

An optimal OPM module can be selected from two types, that is, AQ2200-211 and AQ2200-231. If one TLS module and one OPM module are not selected, the next screen is not displayed even though the <OK> function button or [ENTER] key is pressed.

[Completion of Operation]

Explanation The wavelength correction value of the OPM is changed while linking with the wavelength setting of the TLS module. This function is convenient for the system, in which multiple OPM modules are mounted on the frame controller having three or more slots. Note Since the AQ2201 Frame Controller is a type of 3-slot model, select all

modules to change the color of the check box to green (black for AQ2200-211 and AQ2200-231).

When checking on relevant check box, the dimmed <OK> function button is changed to the normal button display. At this time, when pressing the <OK> button or [ENTER] key, the main screen of the Manual function will appear.

Press [ENTER] or <OK>.

Manual main screen of Application function

Page 47: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 3-5

2

Manual Measurement

3

3.4 Operating the Wavelength Calibration and ZERO-SET

When operating this unit, you must calibrate the wavelength every time the laser output is turned ON. This section describes how to calibrate the wavelength.

Operating Procedures Note When canceling the operation during execution of the wavelength

calibration, the wavelength calibration, which has been performed until that time, becomes invalid. Therefore, before starting the operation, perform the wavelength calibration again to complete it.

This indication is shown if “WL Cal” is not completed. When “WL Cal” is completed, this indication disappears.

Press the <Zero> button to perform ZERO-SET. For details about ZERO-SET, see Chapter 3 in the User’s Manual for AQ2201/AQ2202 Frame Controller.

Press the <WL Cal> button to start the wavelength calibration. The execution screen appears while the wavelength calibration is running. The execution screen will disappear when the wavelength calibration is completed. Therefore, you need to wait until the execution screen disappears. For details about wavelength calibration, see Chapter 5 in the User’s Manual for AQ2200-136 TLS Module.

Screen during Execution of Wavelength Calibration

Screen during Execution of ZERO-SET

Page 48: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

3-6 IM810518904-03E

3.5 Starting the Manual Measurement This section describes how to measure the wavelength level manually.

Operating Procedures

[Operation 1] Call up the screen used to set a wavelength you want to measure.

[Operation 2] Set a wavelength you want to measure.

[Operation 3] Complete operation. ***** page 3-7

Move the cursor to a slot where the OPM module to be measured is mounted, and then press the <Wave length> button to display the wavelength setup popup screen.

Press the <Wavelength> button.

The popup screen will appear. To set a numeric value, input a value with the [ ] or [ ] cursor key or through the ten-key pad, and then press the [ENTER] key or <OK> button.

Press the [ENTER] key or <OK> button.

Set an output power level with the <Power> button. For details about the functionality, see Chapter 8 in the “AQ2200-136 TLS Module User’s Manual”.

Page 49: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

3.5 Starting the Manual Measurement

IM 810518904-03E 3-7

2

Manual Measurement

3

Changing the Screen Display Every time the [Detail] key is pressed, the screen is changed. For details about function keys on the Detail screen, see Chapter 3 in the “AQ2201/AQ2202 Frame Controller USER'S MANUAL” .

Changing the Screen Display of the TLS Module Move the cursor to the "TLS Module: Slot" item and press the [Detail] key. The Detail screen will appear. Every time the [Detail] key is pressed, the screen is alternately switched between the Summary screen and Detail screen.

Summary Screen

Press the [Detail] key.

Detail Screen

Page 50: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

3.5 Starting the Manual Measurement

3-8 IM810518904-03E

Changing the Screen Display of the OPM Module Move the cursor to the "Sensor Module: Slot" item and press the [Detail] key. The Detail screen will appear. Every time the [Detail] key is pressed, the screen is alternately switched between the Summary screen and Detail screen.

Summary Screen

Press the [Detail] key.

Detail Screen

Page 51: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

Chapter 4 Remote Control

IM 810518904-03E 4-1

2 4

Remote Control

4.1 General Description of Commands The following shows the list of commands. For details about command syntax description rules, see the “AQ2200 Multi-Application Test System Remote Command Reference MANUAL” . Command Description APPLication:MSELect:TLSApplication:SWEPt:TLS Selecting or reading the TLS module

APPLication:MSELect:TLSApplication:SWEPt:SENSe Selecting or reading the OPM module

APPLication:TLSApplication:SWEPt:STOP Exiting the Swept function

APPLication:PARameter:TLSApplication:SWEPt:STARt Setting or reading the sweep start wavelength

APPLication:PARameter:TLSApplication:SWEPt:STOP Setting or reading the sweep stop wavelength

APPLication:PARameter:TLSApplication:SWEPt:STEP Setting or reading the sweep step wavelength

APPLication:PARameter:TLSApplication:SWEPt:CENTer Setting or reading the sweep center wavelength

APPLication:PARameter:TLSApplication:SWEPt:SPAN Setting or reading the sweep span

APPLication:PARameter:TLSApplication:SWEPt:SAMPles Setting or reading the number of sweep samples

APPLication:PARameter:TLSApplication:SWEPt:MODE Setting or reading the sweep mode

APPLication:PARameter:TLSApplication:SWEPt:RANGe Setting or reading the sweep dynamic range

APPLication:PARameter:TLSApplication:SWEPt:CYCLes Setting or reading the number of sweep cycles

APPLication:PARameter:TLSApplication:SWEPt:WLMC:MODE Setting or reading settling wavelength function (with a Wavelength Meter)

APPLication:PARameter:TLSApplication:SWEPt:POWer Setting or reading the sweep sweep power

APPLication:PARameter:TLSApplication:SWEPt:PMAX Setting or reading the level to the maximum power within the sweep area

APPLication:PARameter:TLSApplication:SWEPt:LINewidth Setting or reading the sweep line width

APPLication:PARameter:TLSApplication:SWEPt:ATIMe Setting or reading the averaging time

APPLication:TLSApplication:SWEPt:REFMeasure:STATe ,STARt Starting the reference measurement

APPLication:TLSApplication:SWEPt: REFMeasure:STATe Reading the reference measurement status

APPLication:TLSApplication:SWEPt: REFMeasure: RESult Reading the reference measurement result

APPLication:TLSApplication:SWEPt:REFMeasure:MODE Setting or reading the reference measurement method

APPLication:TLSApplication:SWEPt:REFMeasure:CLEar Clearing the reference measurement result

APPLication:TLSApplication:SWEPt[:MEASure]:STATe,STARt Starting the measurement

APPLication:TLSApplication:SWEPt[:MEASure]:STATe Reading the measurement status

APPLication:TLSApplication:SWEPt[:MEASure]:NEXT Moving to the sweep wavelength during measurement (in Manual mode)

APPLication:TLSApplication:SWEPt:[MEASure]: RESult Reading the measurement result

APPLication:PARameter:TLSApplication:SWEPt:AUTO:SAVE Setting or reading the auto data save function

APPLication:PARameter:TLSApplication:SWEPt:AUTO:ANALysis Setting or reading the auto analysis function

APPLication:PARameter:TLSApplication:SWEPt:INPut:SN Setting or reading the serial No. of the device to be measured.

Module selection

Wavelength Calibration (WL Cal)

Setting of each parameter

Reference measurement Measurement

Always execute the measurement commands in the order of the following flow. If the measurement commands are executed in incorrect order, the command setup cannot be performed.

Page 52: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4-2 IM810518904-03E

4.2 Commands for Starting Up and Exiting of the Swept Function

To use the Swept function, always select a module having at least one TLS and one sensor.

Selecting a TLS Module To Be Used syntax: [:FRAMe[F]]:APPLication:MSELect:TLSApplication:SWEPt:TLS[<wsp> <para1>] parameter: <para1>: <Module No.> <Device No.> None: Do not select TLS module within relevant frame. example: :frame0:application:mselect:tlsapplication:swept:tls 1 1 example of abbr.: :APPL:MSEL:TLSA:SWEP:TLS 1 1 * When executing this command, the screen display of this unit is changed to "MODULE

SELECT for SWEPT".

Reading the TLS Module in Use syntax: [:FRAMe[F]]:APPLication:MSELect:TLSApplication:SWEPt:TLS? parameter: None response: <para1>: TLS module at the position shown by para1 is used. None: TLS module in use does not exist in relevant frame. example: :frame0:application:mselect:tlsapplication:swept:tls? → 1 1 example of abbr.: :APPL:MSEL:TLSA:SWEP:TLS? → 1 1

Selecting an OPM Module To Be Used syntax: [:FRAMe[F]]:APPLication:MSELect:TLSApplication:SWEPt:[SENSe]<wsp> [|<para1>|all|] parameter: <para1>: <Module No.> <Device No.> It is possible to specify multiple modules with they separated

by a comma “,”. example: <para1>,<para2> None: Do not select OPM module within relevant frame. example: :frame0:application:mselect:tlsapplication:swept:sense 1 1,2 1,3 1 example of abbr.: :APPL:MSEL:TLSA:SWEP:SENS 1 1,2 1,3 1 * When executing this command, the screen display of this unit is changed to "MODULE

SELECT for SWEPT".

Reading the OPM Module To Be Used syntax: [:FRAMe[F]]:APPLication:MSELect:TLSApplication:SWEPt:[SENSe]? parameter: None response: <para1>: OPM module at the position shown by para1 is used. None: OPM module in use does not exist in relevant frame. example: :frame0:application:mselect:tlsapplication:swept:sense? → 1 1, 2 1, 3 1 example of abbr.: :APPL:MSEL:TLSA:SWEP:SENS? → 1 1, 2 1, 3 1

Exiting the Swept Function syntax: [:FRAMe[F]]:APPLication:TLSApplication:SWEPt:STOP parameter: None example: :frame0:application:tlsapplication:swept:stop example of abbr.: :APPL:TLSA:SWEP:STOP * When executing this command, the Swept function is exited.

Page 53: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 4-3

2 4

Remote Control

4.3 Commands for Specifying of Sweep Area Only by Wavelength

Setting the Sweep Start Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:STARt<wsp> <value>[ PM|NM|UM|MM|M] parameter: <value>PM or NM or UM or MM or M: Sets the wavelength in [m]. example: :application:parameter:tlsapplication:swept:start 1440nm example of abbr.: :APPL:PAR:TLSA:SWEP:STAR 1440nm

Reading the Sweep Start Wavelength syntax: APPLication:PARameter:TLSApplication:SWEPt:STARt? parameter: None response: <value>: Returns the float-type wavelength set value in [m]. example) “1440nm” → +1.44000000E-006 example: :application:parameter:tlsapplication:swept:start? → +1.44000000E-006 example of abbr.: :APPL:PAR:TLSA:SWEP:STAR? → +1.44000000E-006

Setting the Sweep Stop Wavelength syntax: APPLication:PARameter:TLSApplication:SWEPt:STOP<wsp> <value>[ PM|NM|UM|MM|M] parameter: <value>PM or NM or UM or MM or M: Sets the wavelength in [m]. example: :application:parameter:tlsapplication:swept:stop 1640nm example of abbr.: :APPL:PAR:TLSA:SWEP:STOP 1640nm

Reading the Sweep Stop Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:STOP? parameter: None response: <value>: Returns the float-type wavelength set value in [m]. example) “1640nm” → +1.64000000E-006 example: :application:parameter:tlsapplication:swept:stop? → +1.64000000E-006 example of abbr.: :APPL:PAR:TLSA:SWEP:STOP? → +1.64000000E-006

Page 54: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.3 Commands for Specifying of Sweep Area Only by Wavelength

4-4 IM810518904-03E

Setting the Sweep Step Wavelength syntax: [:FRAME[F]]:APPLication:PARameter:TLSApplication:SWEPt:STEP<wsp> <value>[ PM|NM|UM|MM|M] parameter: <value>PM or NM or UM or MM or M: Sets the wavelength in [m]. example: :frame0:application:parameter:tlsapplication:swept:step 0.5nm example of abbr.: :APPL:PAR:TLSA:SWEP:STOP 0.5nm

Reading the Sweep Step Wavelength syntax: [:FRAME[F]]:APPLication:PARameter:TLSApplication:SWEPt:STEP? parameter: None response: <value>: Returns the float-type wavelength set value in [m]. example) “0.5nm”→ +5.00000000E-010 example: :frame0:application:parameter:tlsapplication:swept:step?

→ +5.00000000E-010 example of abbr.: :APPL:PAR:TLSA:SWEP:STEP? → +5.00000000E-010

Page 55: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 4-5

2 4

Remote Control

4.4 Commands for Setting of Sweep Area by Center Wavelength, Span, and Number of Samples

Setting the Sweep Center Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:CENTer<wsp> <value>[ PM|NM|UM|MM|M] parameter: <value>PM or NM or UM or MM or M: Sets the wavelength in [m]. example: :application:parameter:tlsapplication:swept:center 1540nm example of abbr.: :APPL:PAR:TLSA:SWEP:CENT 1540nm

Reading the Sweep Center Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:CENTer? parameter: None response: <value>: Returns the float-type wavelength set value in [m]. example) “1540nm” → +1.54000000E-006 example: :application:parameter:tlsapplication:swept:center? → +1.54000000E-006 example of abbr.: :APPL:PAR:TLSA:SWEP:CENT? → +1.54000000E-0063

Setting the Sweep Span Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:SPAN<wsp> <value>[ PM|NM|UM|MM|M] parameter: <value>PM or NM or UM or MM or M: Sets the wavelength in [m]. example: :application:parameter:tlsapplication:swept:SPAN 200nm example of abbr.: :APPL:PAR:TLSA:SWEP:SPAN 200nm

Reading the Sweep Span Wavelength syntax: :APPLication:PARameter:TLSApplication:SWEPt:SPAN? parameter: None response: <value>: Returns the float-type wavelength set value in [m]. example) “200nm” → +2.00000000E-007 example: :application:parameter:tlsapplication:swept:span? → +2.00000000E-007 example of abbr.: :APPL:PAR:TLSA:SWEP:SPAN? → +2.00000000E-007

Page 56: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.4 Commands for Setting of Sweep Area by Center Wavelength, Span, and Number of Samples

4-6 IM810518904-03E

Setting the Number of Sweep Samples syntax: :APPLication:PARameter:TLSApplication:SWEPt:SAMPles<wsp> <value> parameter: <value>: Number of samples example: :application:parameter:tlsapplication:swept:samples 10000 example of abbr.: :APPL:PAR:TLSA:SWEP:SAMP 10000

* The maximum number of samples you can set is “20000”.

Reading the Number of Sweep Samples syntax: :APPLication:PARameter:TLSApplication:SWEPt:SAMPles? parameter: None response: <value>: Number of samples example: :application:parameter:tlsapplication:swept:samples? → +10000 example of abbr.: :APPL:PAR:TLSA:SWEP:SAMP? → +10000

Page 57: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 4-7

2 4

Remote Control

4.5 Commands for Setting of the Sweep Mode Setting the Sweep Mode

syntax: :APPLication:PARameter:TLSApplication:SWEPt:MODE<wsp> STEPped|MANual|CONTinuous parameter: STEPped: Step sweep mode MANual: Manual sweep mode CONTinuous: Continuous sweep mode example: :application:parameter:tlsapplication:swept:mode stepped example of abbr.: :APPL:PAR:TLSA:SWEP:MODE STEP

Reading the Sweep Mode syntax: :APPLication:PARameter:TLSApplication:SWEPt:MODE? parameter: None response: STEP: Step sweep mode MAN: Manual sweep mode CONT: Continuous sweep mode example: :application:parameter:tlsapplication:swept:mode? → STEP example of abbr.: :APPL:PAR:TLSA:SWEP:MODE? → STEP The sweep Dynamic range you can set may vary depending on the sweep mode. For details, see section 2.6, Setting the Measurement Dynamic Range.

Setting the Sweep Dynamic Range syntax: :APPLication:PARameter:TLSApplication:SWEPt:RANGe<wsp> 0|1|2|3|4 parameter: 0: Auto1 range (full-range) 1: Auto2 range (middle-range) 2: High range 3: Middle range 4: Low range example: :application:parameter:tlsapplication:swept:range 1 example of abbr.: :APPL:PAR:TLSA:SWEP:RANG 1

Setting the Sweep Dynamic Range syntax: :APPLication:PARameter:TLSApplication:SWEPt:RANGe? parameter: None response: +0: Auto1 range (full-range) +1: Auto2 range (middle-range) +2: High range +3: Middle range +4: Low range example: :application:parameter:tlsapplication:swept:range? → +1 example of abbr.: :APPL:PAR:TLSA:SWEP:RANG? → +1

Page 58: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.5 Commands for Setting of the Sweep Mode

4-8 IM810518904-03E

Setting the Number of Sweep Cycles syntax: :APPLication:PARameter:TLSApplication:SWEPt:CYCLes<wsp> <value> parameter: <value>: Sets the number of sweep cycles. Setting range: 0 to 99999, 1-step (Continuos sweep when

set at "0".) Default: 1 * When a value other than "1" is set, "Auto Save", "Auto

Analysis", and "Input S/N" are turned OFF. example: :application:parameter:tlsapplication:swept:cycles 10 example of abbr.: :APPL:PAR:TLSA:SWEP:CYCL 10

Reading the Number of Sweep Cycles syntax: :APPLication:PARameter:TLSApplication:SWEPt:CYCLes? parameter: None response: <value>: Returns the number of sweep cycles. example) “10” → +10 example: :application:parameter:tlsapplication:swept:cycles? → +10 example of abbr.: :APPL:PAR:TLSA:SWEP:CYCL? → +10

Setting the Wavelength settling functon (with a Wavelength Meter) syntax: :APPLication:PARameter:TLSApplication:SWEPt:WLMC:MODE <wsp>OFF|0|ON|1 parameter: OFF or 0: Sets the convergence function of the Wavelength Meter to “OFF”. ON or 1: Sets the convergence function of the Wavelength Meter to “ON”. example: :application:parameter:tlsapplication:swept:wlmc:mode on example of abbr.: :APPL:PAR:TLSA:SWEP:WLMC :MODE ON * This setting cannot be made when the sweep mode is set at "Continuous".

Reading the Wavelength settling function (with a Wavelength Meter) syntax: :APPLication:PARameter:TLSApplication:SWEPt:WLMC:MODE? parameter: None response: 0: Convergence function of the Wavelength Meter is set at “OFF”. 1: Convergence function of the Wavelength Meter is set at “ON”. example: :application:parameter:tlsapplication:swept:wlmc:mode? → 1 example of abbr.: :APPL:PAR:TLSA:SWEP:WLMC :MODE? → 1

Page 59: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 4-9

2 4

Remote Control

4.6 Commands for Setting of Sweep Power Conditions

Setting the Sweep Power syntax: :APPLication:PARameter:TLSApplication:SWEPt:POWer<wsp> <value>[DBM]|MIN|MAX|DEF parameter: value: Power level value in dBm. MIN: Minimum power level MAX: Maximum power level DEF: Default power level example: :application:parameter:tlsapplication:swept:power MAX example of abbr.: :APPL:PAR:TLSA:SWEP:POW MAX

Reading the Sweep Power syntax: :APPLication:PARameter:TLSApplication:SWEPt:POWer? parameter: MIN: Returns the minimum power level. MAX: Returns the maximum power level. DEF: Returns the default power level. response: <value>: Returns the float-type level set value in [dBm]. example) “3dBm” → +3.00000000E+000 example: :application:parameter:tlsapplication:swept:power?

→ +3.00000000E+000 example of abbr.: :APPL:PAR:TLSA:SWEP:POW? → +3.00000000E+000

Setting the Level to the Maximum Power within the Sweep Area syntax: :APPLication:PARameter:TLSApplication:SWEPt:PMAX parameter: None example: :application:parameter:tlsapplication:swept:pmax example of abbr.: :APPL:PAR:TLSA:SWEP:PMAX

Reading the Level to the Maximum Power within the Sweep Area syntax: :APPLication:PARameter:TLSApplication:SWEPt:PMAX? parameter: None response: <value>: Returns the float-type level set value in [dBm]. example) “3dBm” → +3.00000000E+000 example: :application:parameter:tlsapplication:swept:pmax?

→ +3.00000000E+000 example of abbr.: :APPL:PAR:TLSA:SWEP:PMAX? → +3.00000000E+000

Note The output level may vary depending on each TLS module. For

details, see section 4.3 in the “AQ2200-136 TLS Module User’s Manual”.

Page 60: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.6 Commands for Setting of Sweep Power Conditions

4-10 IM810518904-03E

Setting the Line Width in the Swept Functions syntax: :APPLication:PARameter:TLSApplication:SWEPt:LINewidth<wsp> NARRow|0|WIDE|1 parameter: NARRow or 0: Sets the spectrum line width to “Narrow”. WIDE or 1: Sets the spectrum line width to “Wide”. example: :application:parameter:tlsapplication:swept:linewidth narrow example of abbr.: :APPL:PAR:TLSA:SWEP:LIN NARR

Reading the Line Width in the Swept Functions syntax: :APPLication:PARameter:TLSApplication:SWEPt:LINewidth? parameter: None response: +0: Spectrum line width is set at “Narrow. +1: Spectrum line width is set at “Wide”. example: :application:parameter:tlsapplication:swept:linewidth? → +0 example of abbr.: :APPL:PAR:TLSA:SWEP:LIN? → +0

Page 61: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

IM 810518904-03E 4-11

2 4

Remote Control

4.7 Commands for Setting of OPM Conditions in the Swept Functions

Setting the Averaging Time in the Swept Functions syntax: :APPLication:PARameter:TLSApplication:SWEPt:ATIMe<wsp> <value> parameter: <value>: Sets the averaging time in “sec”. example: :application:parameter:tlsapplication:swept:atime 100us example of abbr.: :APPL:PAR:TLSA:SWEP:ATIM 100us * When the sweep mode is set at "Continuous", this setting cannot be made.

Reading the Averaging Time in the Swept Functions syntax: :APPLication:PARameter:TLSApplication:SWEPt:ATIMe? parameter: None response: <value>: Returns the set value of the float-type averaging time in

“sec”. example) “100us” → +1.00000000E−004 example: :application:parameter:tlsapplication:swept:ATIMe? → +1.00000000E−004 example of abbr.: :APPL:PAR:TLSA:SWEP:ATIM? → +1.00000000E−004

Page 62: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4-12 IM810518904-03E

4.8 Commands for Reference Measurement Before starting the reference measurement, always set "Reference measurement method".

Setting the Reference Measurement Method syntax: :APPLication:TLSApplication:SWEPt:REFMeasure:MODE <wsp><para1>| INDividual parameter1: <para1>: Selects a module you want to perform the reference

measurement. <Frame No.>, <Module No.>, <Device No.> INDividual: Reference measurement of the sensor module you have

selected is performed individually. example: :application:tlsapplication:swept:refmeasure:mode 0,1,1 example of abbr.: :APPL:TLSA:SWEP:REFM:MODE 0,1,1

Reading the Reference Measurement Method syntax: :APPLication:TLSApplication:SWEPt: REFMeasure:MODE? parameter: None response: <para1>: Returns the module No., the reference measurement of

which is performed. example: +0, +1, +1 IND: Individual reference measurement is running. NOT SELECT: Reference measurement method is not selected. example: :application:tlsapplication:swept: REFMeasure:mode? → +0, +1, +1 example of abbr.: :APPL:TLSA:SWEP:REFM:MODE? → +0, +1, +1

Starting the Reference Measurement syntax: [:FRAMe[F]]:APPLication:TLSApplication:SWEPt:REFMeasure:STATe <wsp><para1>,STARt parameter1: <para1>: Selects a module you want to perform the reference

measurement. <Module No.> <Device No.> It is possible to specify multiple modules with they separated

by a comma ",". example: <para1>, <para2> parameter2: STARt: Starts the reference measurement. example: :frame0:application:tlsapplication:swept:refmeasure:state 1 1,start example of abbr.: :APPL:TLSA:SWEP:REFM:STAT 1 1,STAR

Page 63: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.8 Commands for Reference Measurement

IM 810518904-03E 4-13

2 4

Remote Control

Reading the Reference Measurement Status syntax: [:FRAMe[F]]:APPLication:TLSApplication:SWEPt: REFMeasure:STATe <wsp><para1>? parameter: <para1>: Selects a module you want to perform the reference

measurement. <Module No.> <Device No.> It is possible to specify multiple modules with they separated

by a comma ",". example: <para1>, <para2> response: COMPLETE: Reference measurement is completed (reference data

exists.). PROGRESS: Reference measurement is being executed. SETUP: Module is being selected or parameter is being set up. NONE: No reference data exists. example: :frame0:application:tlsapplication:swept: REFMeasure:STATe? → COMPLETE example of abbr.: :APPL:TLSA:SWEP:REFM:STAT? → COMPLETE

Reading the Reference Measurement Result syntax: [:FRAMe[F]]:APPLication:TLSApplication:SWEPt: REFMeasure: RESult? <wsp><para1>? parameter: <para1>: Selects a module you want to read the reference

measurement result. <Module No.> <Device No.> example: 1 1 response: <value>: Returns the float-type measurement results separated by a

comma (,) in [dBm]. example) "3dBm" → +3.00000000E+000 NO DATA: No reference measurement data exists. example: :frame0:application:tlsapplication:swept: REFMeasure:result 1 1? → +3.00000000E+000, +3.10000000E+000 example of abbr.: :APPL:TLSA:SWEP:REFM:RES 1 1? → +3.00000000E+000, +3.10000000E+000

Clearing the Reference Measurement Data syntax: :APPLication:TLSApplication:SWEPt:REFMeasure:CLEar parameter1: None example: :application:tlsapplication:swept:refmeasure:clear example of abbr.: :APPL:TLSA:SWEP:REFM:CLE

Page 64: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4-14 IM810518904-03E

4.9 Commands for Measurement Starting the Measurement

syntax: :APPLication:TLSApplication:SWEPt[:MEASure]:STATe <wsp>STARt parameter: STARt: Starts the measurement. example: :application:tlsapplication:swept:measure:state start example of abbr.: :APPL:TLSA:SWEP:STAT STAR * The measurement start cannot be executed unless the reference measurement is

completed.

Reading the Measurement Status syntax: :APPLication:TLSApplication:SWEPt[:MEASure]:STATe? parameter: None response: COMPLETE: Measurement is completed. PROGRESS: Measurement is being executed. SETUP: Module is being selected or parameter is being set. NONE: Swept function is not started up. example: :application:tlsapplication:swept: measure:STATe? → COMPLETE example of abbr.: :APPL:TLSA:SWEP:STAT? → COMPLETE

Moving to the Sweep Wavelength during Measurement (in Manual Mode) syntax: :APPLication:TLSApplication:SWEPt[:MEASure]:NEXT parameter: None example: :application:tlsapplication:swept:measure:next example of abbr.: :APPL:TLSA:SWEP:NEXT

Page 65: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.9 Commands for Measurement

IM 810518904-03E 4-15

2 4

Remote Control

Reading the Measurement Result syntax: [:FRAME[F]]:APPLication:TLSApplication:SWEPt:[MEASure]: RESult ? <wsp><para1> parameter: <para1>: Selects a module you want to read the measurement result. <Module No.> <Device No.> example: 1 1 response: <value>: Returns the float-type measurement results separated by a

comma (,) in [dBm]. example) "3dBm" → +3.00000000E+000 NO DATA: No measurement data exists. example: :frame0:application:tlsapplication:swept: measure:result ? 1 1 → +3.00000000E+000, +3.10000000E+000 example of abbr.: :APPL:TLSA:SWEP:RES ? 1 1 → +3.00000000E+000, +3.10000000E+000

Page 66: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4-16 IM810518904-03E

4.10 Commands for Data Save and Analysis Setting the Auto Data Save Function

syntax: :APPLication:PARameter:TLSApplication:SWEPt:AUTO:SAVE <wsp>OFF|0|ON|1 parameter: OFF or 0: Sets the auto data save function to “OFF”. ON or 1: Sets the auto data save function to “ON”. example: :application:parameter:tlsapplication:swept:auto:save on example of abbr.: :APPL:PAR:TLSA:SWEP:AUTO:SAVE ON

Reading the Auto Data Save Function Setting syntax: :APPLication:PARameter:TLSApplication:SWEPt:AUTO:SAVE? parameter: None response: 0: Auto data save function is set at “OFF”. 1: Auto data save function is set at “ON”. example: :application:parameter:tlsapplication:swept:auto:save? → 1 example of abbr.: :APPL:PAR:TLSA:SWEP:AUTO:SAVE? → 1

Setting the Auto Analysis Function syntax: :APPLication:PARameter:TLSApplication:SWEPt:AUTO:ANALysis <wsp>OFF|0|ON|1 parameter: OFF or 0: Sets the auto analysis function to “OFF”. ON or 1: Sets the auto analysis function to “ON”. example: :application:parameter:tlsapplication:swept:auto:analysis on example of abbr.: :APPL:PAR:TLSA:SWEP:AUTO:ANAL ON * There is no command to individually set an analysis parameter. Before starting

the analysis, always turn ON the auto analysis function using this command.

Reading the Auto Analysis Function Setting syntax: :APPLication:PARameter:TLSApplication:SWEPt:AUTO:ANALysis? parameter: None response: 0: Auto analysis function is “OFF”. 1: Auto analysis function is “ON”. example: :application:parameter:tlsapplication:swept:auto:analysis? → 1 example of abbr.: :APPL:PAR:TLSA:SWEP:AUTO:ANAL? → 1

Page 67: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4.10 Commands for Data Save and Analysis

IM 810518904-03E 4-17

2 4

Remote Control

Inputting the Serial Number of the Device To Be Measured syntax: :APPLication:PARameter:TLSApplication:SWEPt:INPut:SN <wsp>OFF|<value> parameter: OFF: Serial number is not input ("Input S/N" is set at “OFF”.). <value>: Inputs 13 or less alphanumeric characters for the serial

number. ("Input S/N" is set at ”ON”.) example: :application:parameter:tlsapplication:swept:input:sn 1234ABC example of abbr.: :APPL:PAR:TLSA:SWEP:INP:SN 1234ABC

Reading the Serial No. of the Device To Be Measured syntax: :APPLication:PARameter:TLSApplication:SWEPt:INPut:SN? parameter: None response: OFF: Serial number is not input. ("Input S/N" is set at “OFF”.) <value>: Returns the serial number expressed by alphanumeric

characters. ("Input S/N" is set at “ON”.) example: :application:parameter:tlsapplication:swept:input:sn? → 1234ABC example of abbr.: :APPL:PAR:TLSA:SWEP:INP:SN? → 1234ABC

Page 68: User’s AQ2200-136 TLS Module Manual Optical Passive Device ... · of the check box on the left is changed from white to green. On the contrary, when an OPM module is selected, the

4-18 IM810518904-03E

4.11 Commands To Be Set Disabled during Execution of Swept Function

Individual commands of each module become invalid while the Swept function is running. At this time, only the following commands for the system become valid.

Command Description SYST:PRES Initializes the system. (Swept function is aborted.)SYST:ERR? Reads the error status.