UoM activities The Usual (and More) Suspects New Arrival Time Fit Function AAT0002 gas performance Initial tests with Br doped gases

UoM activities The Usual (and More) Suspects

Embed Size (px)

DESCRIPTION

UoM activities The Usual (and More) Suspects. New Arrival Time Fit Function AAT0002 gas performance Initial tests with Br doped gases. New Arrival Time Fit Function. AAT0001 HV= 1150V. AAT0002 Voltage Scan (Zoe). Arrival Time AAT0002 at 1000V. Arrival Time AAT0002 at 1020V. - PowerPoint PPT Presentation

Citation preview

Page 1: UoM activities The Usual (and More) Suspects

UoM activitiesThe Usual (and More) Suspects

New Arrival Time Fit FunctionAAT0002 gas performance

Initial tests with Br doped gases

Page 2: UoM activities The Usual (and More) Suspects

New Arrival Time Fit Function

6/24/14 UoM activities 2

AAT0001HV=1150V

Page 3: UoM activities The Usual (and More) Suspects

AAT0002 Voltage Scan (Zoe)

6/24/14 UoM activities 3

Page 4: UoM activities The Usual (and More) Suspects

Arrival Time AAT0002 at 1000V

6/24/14 UoM activities 4

Page 5: UoM activities The Usual (and More) Suspects

6/24/14 UoM activities 5

Arrival Time AAT0002 at 1020V

Page 6: UoM activities The Usual (and More) Suspects

UoM activities 6

Future Microcavity Experiments

• Repeat voltage scan for the same pixel (for reproducibility) while extending the HV range

• Arrival time at higher HV• Try as AAT0003 Ne + 0.4% Ar + 1% SF6 :

replacing CF4 with SF6 in AAT0002 to see if it works at higher HV and so shrinking the arrival time distribution.

6/24/14

Page 7: UoM activities The Usual (and More) Suspects

UoM activities 7

New Gas Mixing Station

6/24/14

MixingVolumeValve

Panel Valve Shutoff

Gauge

Bypass

Vacuum

Gas AGas B

Needle

Page 8: UoM activities The Usual (and More) Suspects

UoM activities 8

Initial Tests with Br-gases on VPIAr + 0.5% Br (bottle) at 747 Torr1. No pulses observed up to HV=1000V (mV threshold)2. At around 500 V, not null current:

a) onset of current is sharp and repeatable at 495-500 V.b) The current is at the level of a microamps, but increasing

nonlinearly with increasing voltage. At 600 V a ~15% increase in HV leads to about 1.7-1.8 increase in current.

3. HV polarity reversal attempt: briefly bipolar pulses, unrelated to source.

Pure Argon at 741 Torr4. Large amplitude pulses (>30V) at ~740 V and above.

Some lines are much more active than others, but not visible response to the source. Easily in regeneration.

6/24/14

Page 9: UoM activities The Usual (and More) Suspects

UoM activities 9

Initial Tests with Br-gases on VPI (2)Ar + 0.05% Br at 734 Torr• Seen a few pulses (~6V amplitude) for a minute or so

around -460V, then nothing.• After polarity reversal again a few pulses and then

nothing.• Interestingly with HV=+1kV bipolar signal visible.• Also a unipolar signal 5-8 ns wide is observable, although

unclear if it was in response to the source.Ar + 0.01% Br at 730 Torr• Still to be tested (filled yesterday)If it does not work, we will move to 3 components gas starting with Ar + 0.5% Br + 0.5% Xe.6/24/14