1 ATS Laboratories, Inc. ATS Laboratories, Inc. 1 Ultrasound QC Workshop Ultrasound QC Workshop Doppler Doppler AAPM AAPM Orlando, FL Orlando, FL August, 2006 August, 2006 2 ATS Laboratories, Inc. ATS Laboratories, Inc. Doppler QC Doppler QC Different than ultrasound imaging Different than ultrasound imaging – able to able to measure physical parameters or qualities related measure physical parameters or qualities related to blood flow to blood flow Doesn’t judge image quality but rather the Doesn’t judge image quality but rather the accuracy of measured accuracy of measured hemodynamic hemodynamic values values Volumetric flow rate Volumetric flow rate Peak velocity Peak velocity Pressure Pressure Is Doppler system performance up to par for Is Doppler system performance up to par for clinical use…what are the effects of transducer clinical use…what are the effects of transducer degradation? degradation? 3 ATS Laboratories, Inc. ATS Laboratories, Inc. Some difficulties Some difficulties Great variability among different Great variability among different machines machines Measurement differences and therefore Measurement differences and therefore ultimate accuracy is machine dependent ultimate accuracy is machine dependent Operator variability Operator variability Recognizing system performance change Recognizing system performance change due to component degradation due to component degradation 4 ATS Laboratories, Inc. ATS Laboratories, Inc. Doppler system performance Doppler system performance Detectability Detectability of flow of flow – single most critical aspect single most critical aspect of Doppler system performance of Doppler system performance Sensitivity Sensitivity – minimum detectable signal level (system minimum detectable signal level (system signal to noise ratio) signal to noise ratio) Relates directly to depth of penetration Relates directly to depth of penetration Dynamic range Dynamic range – clutter to noise ratio clutter to noise ratio Major determinant of Major determinant of detectability detectability when solid structures are when solid structures are present present Frequency range Frequency range – range range of detectable Doppler shift of detectable Doppler shift frequencies frequencies Relates to minimum and maximum detectable velocities Relates to minimum and maximum detectable velocities Defective transducer is a major determinant of Defective transducer is a major determinant of accurate accurate detectability detectability of flow of flow

Ultrasound QC Workshop Doppler - AAPM

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

1

ATS Laboratories, Inc.ATS Laboratories, Inc. 11

Ultrasound QC WorkshopUltrasound QC Workshop

DopplerDoppler

AAPMAAPM

Orlando, FLOrlando, FLAugust, 2006August, 2006

22ATS Laboratories, Inc.ATS Laboratories, Inc.

Doppler QCDoppler QC

Different than ultrasound imaging Different than ultrasound imaging –– able to able to measure physical parameters or qualities related measure physical parameters or qualities related to blood flowto blood flow

Doesn’t judge image quality but rather the Doesn’t judge image quality but rather the accuracy of measured accuracy of measured hemodynamichemodynamic valuesvalues�� Volumetric flow rateVolumetric flow rate

�� Peak velocityPeak velocity

�� PressurePressure

Is Doppler system performance up to par for Is Doppler system performance up to par for clinical use…what are the effects of transducer clinical use…what are the effects of transducer degradation?degradation?

33ATS Laboratories, Inc.ATS Laboratories, Inc.

Some difficultiesSome difficulties

Great variability among different Great variability among different

machinesmachines

Measurement differences and therefore Measurement differences and therefore

ultimate accuracy is machine dependentultimate accuracy is machine dependent

Operator variabilityOperator variability

Recognizing system performance change Recognizing system performance change

due to component degradationdue to component degradation

44ATS Laboratories, Inc.ATS Laboratories, Inc.

Doppler system performanceDoppler system performance

DetectabilityDetectability of flow of flow –– single most critical aspect single most critical aspect of Doppler system performanceof Doppler system performance�� Sensitivity Sensitivity –– minimum detectable signal level (system minimum detectable signal level (system

signal to noise ratio)signal to noise ratio)Relates directly to depth of penetrationRelates directly to depth of penetration

�� Dynamic range Dynamic range –– clutter to noise ratioclutter to noise ratioMajor determinant of Major determinant of detectabilitydetectability when solid structures are when solid structures are presentpresent

�� Frequency range Frequency range –– rangerange of detectable Doppler shift of detectable Doppler shift frequenciesfrequencies

Relates to minimum and maximum detectable velocitiesRelates to minimum and maximum detectable velocities

Defective transducer is a major determinant of Defective transducer is a major determinant of accurate accurate detectabilitydetectability of flowof flow

2

55ATS Laboratories, Inc.ATS Laboratories, Inc.

Doppler shift Doppler shift

frequency measurementfrequency measurement

Doppler sample volume Doppler sample volume guided by realguided by real--time Btime B--mode imaging to proper mode imaging to proper locationlocation

Accuracy dependent Accuracy dependent upon:upon:�� Machine componentsMachine components

�� Signal processingSignal processing

�� Doppler angleDoppler angle

�� Sample volume and Sample volume and locationlocation

�� Automatic vs. manual Automatic vs. manual control of displayed control of displayed parametersparameters

66ATS Laboratories, Inc.ATS Laboratories, Inc.

Overestimation of velocityOverestimation of velocity

If >60 deg, minor If >60 deg, minor

errors in Doppler errors in Doppler

angle determination angle determination

results in large results in large

velocity errorsvelocity errors

Preferred Doppler Preferred Doppler

angle range is 20angle range is 20--60 60

degreesdegrees

49.149.1149.1149.110010080807575

18.318.3118.3118.310010065656060

10.010.0110.0110.010010050504545

5.75.7105.7105.710010035353030

0.40.4100.4100.41001005500

% Error% Error

Estimated Estimated

velocityvelocity

(cm/sec)(cm/sec)

Actual Actual

VelocityVelocity

(cm/sec)(cm/sec)

SetSet

AngleAngle

(degrees)(degrees)

DopplerDoppler

Angle Angle

(degrees)(degrees)

77ATS Laboratories, Inc.ATS Laboratories, Inc.

AliasingAliasing

Wrapping of spectral Wrapping of spectral

waveform to opposite waveform to opposite

direction of flowdirection of flow

Caused by sampling at Caused by sampling at

less than the less than the NyquistNyquist rate rate

(PRF must be at least 2x (PRF must be at least 2x

the maximum Doppler the maximum Doppler

shift frequency)shift frequency)

To eliminate, increase the To eliminate, increase the

PRF (velocity scale, PRF (velocity scale,

sample volume depth)sample volume depth)

88ATS Laboratories, Inc.ATS Laboratories, Inc.

Spectral broadeningSpectral broadeningDetermined byDetermined by

�� Velocity profile in vesselVelocity profile in vessel

�� Size of sample volumeSize of sample volume

Reynolds numberReynolds number

�� <2000 will insure laminar <2000 will insure laminar

flowflow

Parabolic flowParabolic flow

�� Broad bandwidth of Broad bandwidth of

Doppler shift frequenciesDoppler shift frequencies

�� Broad range of velocitiesBroad range of velocities

Defective transducer willDefective transducer will

�� Degrade displayed velocity Degrade displayed velocity

accuracyaccuracy

�� Increase noiseIncrease noise

3

99ATS Laboratories, Inc.ATS Laboratories, Inc.

Types of Doppler test objectsTypes of Doppler test objects

Flow (fluid)Flow (fluid)�� Tubes carrying a bloodTubes carrying a blood--mimicking fluid in a mimicking fluid in a

scattering and attenuating mediumscattering and attenuating medium

StringString�� A vessel containing a string or belt moved by A vessel containing a string or belt moved by

a motor and pulleysa motor and pulleys

Electronic/acousticElectronic/acoustic�� Direct injection of a signal into the signal Direct injection of a signal into the signal

processing chain or direct coupling to a processing chain or direct coupling to a transducer inputtransducer input

1010ATS Laboratories, Inc.ATS Laboratories, Inc.

Fundamental QC testsFundamental QC tests

Measurement of flow volume and velocityMeasurement of flow volume and velocity

Maximum penetrationMaximum penetration

Sensitivity at varying depthsSensitivity at varying depths

Flow locationFlow location

Directional discriminationDirectional discrimination

1111ATS Laboratories, Inc.ATS Laboratories, Inc.

Typical clinical applicationsTypical clinical applications

11--1010<2<2<100<100<1<1<5<5Small vesselSmall vessel

11--1010<2<2100100--20002000>100>100<50<50Arterial Arterial stenosisstenosis

11--551010--2020<50<50<20<20200200--10001000Systemic venousSystemic venous

33--8822--1010100100--100010002020--50505050--200200Cerebral arterialCerebral arterial

11--5522--1010100100--100010005050--100100200200--10001000Systemic arterialSystemic arterial

55--1010>20>20>1000>1000>100>100>1000>1000CardiacCardiac

Depth (cm)Depth (cm)Size (mm)Size (mm)AccelerationAcceleration

cm/seccm/sec22

VelocityVelocity

cm/seccm/sec

VolumeVolume

cmcm33/min/min

TARGET VESSELTARGET VESSELFLOWFLOW

APPLICATIONAPPLICATION

1212ATS Laboratories, Inc.ATS Laboratories, Inc.

Flow Doppler test object and Flow Doppler test object and

pumping systempumping system

4

1313ATS Laboratories, Inc.ATS Laboratories, Inc.

HandsHands--on demonstrationon demonstration

Set up flow Doppler test object and pumping system Set up flow Doppler test object and pumping system

variablesvariables

Choose region of interest using BChoose region of interest using B--mode imagemode image

Manipulate “gate”Manipulate “gate”

�� Angle Angle –– the Doppler anglethe Doppler angle

�� Size Size -- % of vessel sampled% of vessel sampled

�� Position Position –– location of gate within the vessellocation of gate within the vessel

Compare performance of functional vs. degraded Compare performance of functional vs. degraded

transducertransducer

1414ATS Laboratories, Inc.ATS Laboratories, Inc.

Typical results with defective Typical results with defective

transducertransducer

Errors in velocity Errors in velocity measurementmeasurement

Spectral broadeningSpectral broadening

Increased noiseIncreased noise

Phase ambiguityPhase ambiguity