ALD MCP 116 Life Test Update • MCP 116 - 20µm pore, ANL functionalized, R = 31 MΩ • Single plate in phosphor screen detector on life test system • Pre-scrub imaging vs. voltage • Baked at 145°C for 48 hours (peak pressure 1e -4 torr, final pressure 4e -7 torr) • Post-bake imaging and “gain” vs. voltage • Start scrub 9/10/10, has been going > 95 days (~3 C/cm 2 ) • Scrubbed down to about 20% relative gain, recovered to about 60%, then started a slow decrease which is gradually leveling. • Intermediate imaging and gain vs. voltage at 0.43 C/cm 2 , 1.74 C/cm 2 , 3 C/cm 2 16 November 2010 J. McPhate – MCP 116 Life Test Update 1

UCB-20101213 MCP116-life-test.pptx (Read-Only)• Single plate in phosphor screen detector on life test system • Pre-scrub imaging vs. voltage • Baked at 145°C for 48 hours (peak

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

ALD  MCP  116  Life  Test  Update  

•  MCP 116 - 20µm pore, ANL functionalized, R = 31 MΩ •  Single plate in phosphor screen detector on life test system •  Pre-scrub imaging vs. voltage •  Baked at 145°C for 48 hours (peak pressure 1e-4 torr, final pressure 4e-7 torr) •  Post-bake imaging and “gain” vs. voltage •  Start scrub 9/10/10, has been going > 95 days (~3 C/cm2) •  Scrubbed down to about 20% relative gain, recovered to about 60%, then started a slow decrease which is gradually leveling. •  Intermediate imaging and gain vs. voltage at 0.43 C/cm2, 1.74 C/cm2, 3 C/cm2

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   1  

RelaCve  Gain  vs.  Extracted  Charge  

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   2  

0%  

20%  

40%  

60%  

80%  

100%  

0.0   0.5   1.0   1.5   2.0   2.5   3.0   3.5  

RelaCve  Ga

in  

Extracted  Charge  (C/cm2)  

ALD  MCP  116  Scrub  RelaCve  Gain  vs.  Extracted  Charge  

Photocurrent  vs.  Voltage  (at  different  scrub  epochs)  

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   3  

0.01  

0.1  

1  

10  

100  

1000  

10000  

0   200   400   600   800   1000   1200   1400   1600  

Photocurrent  (n

A)  

MCP  HV  (volts)  

"Gain"  vs.  Voltage,  ALD  MCP  116  Life  Test  

0.00  C/cm2  (nA)  

0.43  C/cm2  (nA)  

1.74  C/cm2  (nA)  

3.05  C/cm2  (nA)  

Values  below  0.1nA  and  above  2000  nA  are  subject  to  measurement  errors  

UV  Flats  Overview  

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   4  

Post-­‐Bake  900V  

1100V  

@  0.43  C/cm2  

1100V  

1300V  

@  1.74  C/cm2  

1100V  

1300V  

UV  Flat  Closeups  @  1.74  C/cm2  

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   5  

1100V   800V  

Spots  are  likely  to  be  areas  where  the  ALD  overlayer  has  been  removed  by  cleaning  off  dust,  Very  bright  spots  are  field  emission,  worst  one  is  in  a  triple  point  –  debris???  

ALD  MCP  116  Life  Test  Update  

16  November  2010   J.  McPhate  –  MCP  116  Life  Test  Update   6  

Current  spikes  likely  due  to  bright  spots  in  image  flaring  up  

1µA  

0  µA