288
U-Spin Symmetry Test of the Σ *+ Electromagnetic Decay A dissertation presented to the faculty of the College of Arts and Sciences of Ohio University In partial fulfillment of the requirements for the degree Doctor of Philosophy Dustin M. Keller November 2010 © 2010 Dustin M. Keller. All Rights Reserved.

U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

U-Spin Symmetry Test of the Σ∗+ Electromagnetic Decay

A dissertation presented to

the faculty of

the College of Arts and Sciences of Ohio University

In partial fulfillment

of the requirements for the degree

Doctor of Philosophy

Dustin M. Keller

November 2010

© 2010 Dustin M. Keller. All Rights Reserved.

Page 2: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

2

This dissertation titled

U-Spin Symmetry Test of the Σ∗+ Electromagnetic Decay

by

DUSTIN M. KELLER

has been approved for

the Department of Physics and Astronomy

and the College of Arts and Sciences by

Kenneth H. Hicks

Professor of Physics and Astronomy

Benjamin M. Ogles

Interim Dean, College of Arts And Sciences

Page 3: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

3

ABSTRACT

KELLER, DUSTIN M., Ph.D., November 2010, Physics

U-Spin Symmetry Test of the Σ∗+ Electromagnetic Decay (288 pp.)

Director of Dissertation: Kenneth H. Hicks

This dissertation presents analysis for electromagnetic decay of the Σ0(1385) from

the reaction γ p→ K+Σ∗0. Also presented is the first ever measurement of the

electromagnetic decay of the Σ+(1385) from the reaction γ p→ K0Σ∗+. Both results are

extracted from the g11a data set taken using the CLAS detector at Thomas Jefferson

National Accelerator Facility. A real photon beam with a maximum energy of 3.8 GeV

was incident on a liquid hydrogen target during the experiment resulting in the

photoproduction of the kaon and Σ∗ hyperons. Kinematic fitting is used to separate signal

from background in each case. For the first time, a method to kinematically fit the neutron

in the Electromagnetic Calorimeter (EC) of CLAS was performed, leading to a high

statistics study of the neutron resolutions in the EC. New techniques in neutron resolution

matching for Monte Carlo simulation using dynamic variable smearing are also

developed. The results from the Σ0(1385) electromagnetic decay lead to smaller statistical

and systematic uncertainties than the previous measurement by Taylor et al. A U-spin

symmetry test using the U-spin SU(3) multiplet representation gave a prediction for the

Σ∗+→ Σ+γ partial width and the Σ∗0→ Λγ partial width. The latter agrees, within the

experimental uncertainties, with the prediction from U-spin symmetry, but the former

reaction is much smaller than its prediction.

Approved:

Kenneth H. Hicks

Professor of Physics and Astronomy

Page 4: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

4

TABLE OF CONTENTS

Abstract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

List of Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

List of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201.1 Quantum Chromodynamics . . . . . . . . . . . . . . . . . . . . . . . . . . 201.2 Spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221.3 The Constituent Quark Model . . . . . . . . . . . . . . . . . . . . . . . . 221.4 Electromagnetic decays . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231.5 The CLAS detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 271.6 Previous Experimental Results . . . . . . . . . . . . . . . . . . . . . . . . 281.7 The Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

2 Theoretical Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342.1 Theoretical Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

2.1.1 The NRQM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 342.1.2 The RCQM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 362.1.3 The χCQM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 372.1.4 The MIT Bag Model . . . . . . . . . . . . . . . . . . . . . . . . . 382.1.5 The Soliton Model . . . . . . . . . . . . . . . . . . . . . . . . . . 392.1.6 The Skyrme Model . . . . . . . . . . . . . . . . . . . . . . . . . . 402.1.7 The Algebraic Model . . . . . . . . . . . . . . . . . . . . . . . . . 42

2.2 U-Spin Symmetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 442.2.1 Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 452.2.2 U-Spin prediction . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

2.3 Meson Cloud effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

3 Jefferson Lab, CEBAF and CLAS detector . . . . . . . . . . . . . . . . . . . . . 533.1 Continuous Electron Beam Accelerator Facility . . . . . . . . . . . . . . . 543.2 The Bremsstrahlung Photon Tagger . . . . . . . . . . . . . . . . . . . . . 58

3.2.1 The Radiator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 593.2.2 The Magnetic Spectrometer . . . . . . . . . . . . . . . . . . . . . 603.2.3 The Hodoscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . 603.2.4 The Tagger Readout . . . . . . . . . . . . . . . . . . . . . . . . . 61

3.3 The CLAS Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 633.3.1 The g11a Cryotarget . . . . . . . . . . . . . . . . . . . . . . . . . 643.3.2 The Start Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . 653.3.3 The Superconducting Toroidal Magnet . . . . . . . . . . . . . . . 663.3.4 The Drift Chambers . . . . . . . . . . . . . . . . . . . . . . . . . 68

Page 5: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

5

3.3.5 The Time-of-Flight Detector . . . . . . . . . . . . . . . . . . . . . 703.3.6 The Forward Electromagnetic Calorimeter . . . . . . . . . . . . . . 72

3.4 The Beamline Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 753.5 The g11a Trigger and Data Acquisition . . . . . . . . . . . . . . . . . . . 753.6 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

I Electromagnetic decay of the Σ∗0 78

4 Event Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 804.1 Run Inclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 804.2 Energy Loss Corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . 804.3 Tagger Corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 824.4 Momentum Corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 834.5 Effectiveness of Corrections . . . . . . . . . . . . . . . . . . . . . . . . . 844.6 Particle Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 854.7 Vertex Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 944.8 Beam Photon Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 964.9 Detector Performance Cuts . . . . . . . . . . . . . . . . . . . . . . . . . . 96

5 Analysis Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100

6 Kinematic fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

7 Simulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1217.1 Monte Carlo Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1217.2 Monte Carlo Smearing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1237.3 Monte Carlo Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1257.4 Trigger Simulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1277.5 Matching Data and Tuning . . . . . . . . . . . . . . . . . . . . . . . . . . 127

8 Extraction Methods and Constraints . . . . . . . . . . . . . . . . . . . . . . . . 1398.1 Method-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141

8.1.1 Acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1458.1.2 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158

8.2 Method-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1648.2.1 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172

8.3 Method-3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1758.3.1 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

9 Systematic Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1789.1 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

Page 6: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

6

II Electromagnetic decay of the Σ∗+ 187

10 Event Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18810.1 Particle Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

11 Neutron Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19411.1 Neutron Detection Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19511.2 Neutron Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19611.3 Neutron Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19611.4 Neutron Momentum Correction . . . . . . . . . . . . . . . . . . . . . . . 19711.5 Neutron Resolutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20111.6 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214

12 Analysis Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215

13 Neutron Kinematic Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22313.1 Neutron EC Covariance Matrix . . . . . . . . . . . . . . . . . . . . . . . . 223

13.1.1 Diagonal Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22413.1.2 Off-Diagonal Terms . . . . . . . . . . . . . . . . . . . . . . . . . 225

13.2 Neutron Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226

14 Simulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230

15 Extraction Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24115.1 Acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24115.2 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24815.3 Systematic Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25015.4 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263

16 Overall Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

Appendix A: Ratio Calculation Details . . . . . . . . . . . . . . . . . . . . . . . . . 273

Appendix B: BOS CLAS Data Structure . . . . . . . . . . . . . . . . . . . . . . . . 278

Page 7: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

7

LIST OF TABLES

1.1 Theoretical predictions of the decay widths for the model shown of electro-magnetic decays of various baryons (all in units of keV). . . . . . . . . . . . . 27

4.1 Bad Time-of-flight scintillators. . . . . . . . . . . . . . . . . . . . . . . . . . 984.2 The list of run numbers that have problems with the forward part of the TOF. . 99

7.1 The set of Monte Carlo channel generated for acceptance studies. . . . . . . . . 130

8.1 Counts for nΛ found though the different methods; the raw counts rejectedfrom the π0 hypothesis, and the estimated number of nΛ from Monte Carlo.The uncertainties are fit uncertainties combined with statistical uncertainties. . . 153

8.2 Counts for nK∗ found though different methods. Uncertainties are fituncertainties combined with statistical uncertainties. . . . . . . . . . . . . . . 156

8.3 Acceptances (in units of 10−3) for the channels used in the calculation of thebranching ratios. Here there is a 10% confidence level used as upper and lowerP(χ2) cuts; the Pxy cut was 0.03 GeV. The uncertainties are statistical only.The three columns contain the acceptance for each hypothesis Aγ , Aπ , and thecounts that made all other cuts but did not satisfy either γ or π0 hypothesisdenoted as Aγπ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157

8.4 Breakdown of statistics for each term in Eq. 15.5 for the Λ(γ) and Λ(π0)hypothesis. Each listed channel is subtracted from the raw counts directly fromthe kinematic fit to obtain the final ratio. The uncertainties listed are statisticalonly. The K∗+ counts are not used in method-1 in order to follow Taylor [10]. . 159

8.5 Dependence of the corrected branching ratio on the confidence level cuts. . . . 1618.6 Optimization points for each Pa

π0 and Pbγ from Ref. [71]. . . . . . . . . . . . . . 168

8.7 Acceptances (in units of 10−3) for the channels used in the calculation of thebranching ratios. Here Pb

γ (χ2) > 10% and Paπ (χ2) < 1% while Pb

π (χ2) > 10%and Pa

γ (χ2) < 1% and the Pxy cut was 0.03 GeV. The uncertainties are statisticalonly. The three columns contain the acceptance for each hypothesis Aγ , Aπ ,and the counts that made all other cuts but did not satisfy either the γ or π0

hypothesis denoted as Aγπ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1728.8 Counts for nΛ found through the different methods; the raw counts rejected

from the π0 hypothesis, and the estimated number of nΛ from Monte Carlo.The uncertainties are fit uncertainties combined with statistical uncertainties. . . 173

8.9 Counts for nK∗ found through the different methods. Uncertainties are fituncertainties combined with statistical uncertainties. . . . . . . . . . . . . . . 173

9.1 Branching ratio for excluded channels in (%). . . . . . . . . . . . . . . . . . . 1829.2 Dependence of the corrected branching ratio on the confidence level cuts for

the selected systematic range. . . . . . . . . . . . . . . . . . . . . . . . . . . . 184

Page 8: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

8

9.3 Ranges of systematic variation in resulting ratio in (%) showing L(Low)-contribution and H(High)-contribution and rang in each case. . . . . . . . . . . 185

14.1 The set of Monte Carlo channels and amount of events generated for theacceptance studies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 232

15.1 The cuts used to extract the final radiative and π0 counts. (See text for details.) 24415.2 Acceptances (in units of 10−3) for the channels used in the calculation of the

branching ratios. All the cuts used to obtain the acceptance values are listed inTable 15.1. The uncertainties are statistical only. The two columns contain theacceptance for each hypothesis Aγ , Aπ . In some cases the values are roundedup to 0.0001. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244

15.3 Breakdown of statistics for each term in Eq. 15.4 for the Σ(γ) and Σ(π0)hypothesis. Each counts for each hypothesis is subtracted accordingly asshown in Eq. 15.4. Th raw count are taken directly from the kinematic fitto use in the final ratio calculation. The uncertainties listed are statistical only. . 249

15.4 The resulting contribution to the ratio for a f value of 10%. Some values arerounded up. Uncertainty is statistical only. . . . . . . . . . . . . . . . . . . . . 254

15.5 Ranges of systematic variation in resulting ratio in (%) showing L(Low)-contribution and H(High)-contribution and rang in each case. . . . . . . . . . . 262

16.1 Comparison of theorectical model predictions for the radiative decay widthswith the present results, all in keV. . . . . . . . . . . . . . . . . . . . . . . . . 265

Page 9: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

9

LIST OF FIGURES

1.1 The missing mass of the γ p→ K+X reaction in the photon energy range (a)1.5 < Eγ < 2.0 GeV and (b) 2.0 < Eγ < 2.4. The data points are shownin closed circles. The strengths of the Σ∗0 production were obtained fromthe Λπ0 decay mode and the other reactions were obtained by fitting. Thesolid fit line, open circle, dashed lines, and dot-dashed indicate the spectrafrom K+Λ(1405), K+Σ(1385), nonresonant K+Σπ and K∗0Σ+ production,respectively. Source [46]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

1.2 DESY Bubble Chamber Group total cross section [48] in with CLAS totalcross section with parameterization fit for γ p→ K+Σ(1385). Source [49]. . . . 30

2.1 Baryon decuplet, hypercharge (Y ) versus isospin (I3). . . . . . . . . . . . . . . 452.2 Baryon decuplet, plotted for U-spin multiplets with charge (Q) versus u-spin

(U3). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 452.3 The diagrams for the dressed γN → ∆ vertex. The meson cloud diagrams are

in brackets. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 502.4 The magnetic dipole transition form factor G∗(Q2) for γ∗∆→ ∆(1232). The

experimental points (empty circles) are for the inclusive data from pre-1990experiments at DESY and SLAC [42], and exclusive data (filled squares) arefrom BATES [43], MAMI [44], and JLAB [45]. The Solid curve indicates thedressed calculation, while the dotted line is without the meson cloud effect. . . 52

3.1 The aerial view of CEBAF at Thomas Jefferson National Labs. The“racetrack”-shaped area indicates where the accelerator ring lies underground.Each experimental hall is underneath the three grassy sectors near the bottomof the picture. Hall B is the middle hall. Image source:[35]. . . . . . . . . . . . 53

3.2 A schematic diagram of the Continuous Electron Beam Accelerator Facility.The linear red tubes represent the LINACs made of superconducting RFcavities grouped into 20 cryomodules. A magnification of the recirculationarcs is shown in the top right. The experimental halls are shown on the bottomleft. Image source:[35]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

3.3 A picture of a pair of superconducting niobium Radio Frequency (RF) cavities.CEBAF uses 338 uperconduction cavities, like the RF cavity shown. Imagesource:[36]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

3.4 A schematic diagram of the superconducting RF cavity in operation. Theacceleration gradient is provided by establishing a standing wave, leading toa continuous positive electric force on the electron. The phase of the wavesadvances the position of the electron bunch in the cavity creating the gradient.Image source:[35]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

Page 10: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

10

3.5 A schematic of the photon tagging system. The tagger is setup to allow indirectmeasurements of the photon beam energy. The recoil electron is directed intothe tagger spectrometer to that is energy can be measured to deduce the photonenergy produced. Image source: [56]. . . . . . . . . . . . . . . . . . . . . . . 59

3.6 A schematic of the tagger magnet and hodoscope used in the taggingsystem. The trajectories of the recoil electrons are depicted by the dashedlines. Electrons from various trajectories in the spectrometer correspond tobremsstrahlung photons of a given energy. Image source: [56]. . . . . . . . . . 61

3.7 A schematic of the tagger logic setup. The T-counter hits are also used to setthe event trigger. The common stop to the E-counter TDC array is controlledby the the CLAS Level 1 trigger. Image source: [56]. . . . . . . . . . . . . . . 62

3.8 A photograph of the CLAS detector from inside Hall B. The Time-of-flightscintillators are pulled away from the drift chambers showing inside. Imagesource: [35]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

3.9 A schematic diagram of the full CLAS detector. The drift chambers areshown in violet, the toroidal magnet is shown in light blue, the Time-of-flightscintillators are shown in red, and the electromagnetic calorimeter is shown ingreen. Image source: [35]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

3.10 A diagram of the g11a Cryotarget. The dimensions of the target cell used were40 cm long and 4 cm in diameter. The cell was filled with liquid H2. Imagesource: [35]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

3.11 A diagram of the new g11a Start Counter. The detector is a six-sectorscintillation device with four mounted photomultiplier tubes for each sector.One sector is cut away in the diagram to show the target space. Image source:[58]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

3.12 A photograph of the coils of the CLAS toroidal magnet before installation ofthe rest of the detector subsystem. Image source: [59]. . . . . . . . . . . . . . 67

3.13 (a) The contours of constant absolute magnetic field of the CLAS toroid in themidplane between two of the coils. (b)The field vectors for the CLAS toroidtransverse to the beam. The field lines represent the field strength. The sixcoils are shown in the cross section. Image source: [52]. . . . . . . . . . . . . 68

3.14 A diagram look down showing the CLAS drift chamber region relative to theother subsystems. The dashed lines outline the location of the toroidal magnetcoils. Image source: [59]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

3.15 (a) An isolated view of the time-of-flight paddles in one sector. The setdesignated for forward angle detection are on the right. (b) The schematic ofthe light guide used to connect the backward angle TOF paddles to the PMTs. . 71

3.16 A vertical slice of the EC light readout system. PMT - Photomultiplier TubeLG - Light Guide, FOBIN - Fiber Optic Bundle Inner, FOBOU - Fiber OpticBundle Outer, SC - Scintillators, Pb - 2.2 mm Lead sheets, IP - Inner Plate(closest to target or face of EC) Image source: [55]. . . . . . . . . . . . . . . . 73

Page 11: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

11

3.17 (a) View of one of the six CLAS electromagnetic calorimeter triangularmodules showing the three projection planes. (b) The diagram of eventreconstruction in the EC. Energy deposition profile is shown along each stereoview. Image source: [55]. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

3.18 A schematic layout of the beamline and flux monitoring devices. The beamlineenters from the left. Image source: [56]. . . . . . . . . . . . . . . . . . . . . . 76

4.1 ∆E with respect to momentum for left: proton, and right: K+ . . . . . . . . . . 814.2 Left: ∆Eγ/Ebeam vs. E-counter for the reaction (γ)p→ pπ+π− used to find the

tagger correction. Right: The extracted Gaussian mean from the ∆Eγ/Ebeam vs.E-counter fits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

4.3 Left: ∆p+ found for the K+ in the topology γ p→ K+Λ→ K+pπ− for Sector1, θ ∈ (20,25) and φ ∈ (−15,−10). Right: The Gaussian fit for ∆p+. . . . 84

4.4 Left: Invariant mass of (pπ−), Right: Missing mass off the K+. Beforecorrections is shown in black and after is shown in red. . . . . . . . . . . . . . 85

4.5 Left: Invariant mass of (pπ−) with Gaussian fit before corrections, Right:Invariant mass of (pπ−) with Gaussian fit after corrections. . . . . . . . . . . . 86

4.6 Left: Missing mass off the K+ with Gaussian fit before corrections, Right:Missing mass off the K+ with Gaussian fit after corrections. . . . . . . . . . . 86

4.7 ∆t giving the PID timing cut. Clusters of events at (±2 ns,±2 ns),(±4 ns,±4ns), etc. are from photons from a different beam bucket. The dashed linesrepresent the timing cut used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

4.8 Calculated mass based on the PART bank information with no cuts used.Visible accidentals result from other RF buckets that are present. . . . . . . . . 89

4.9 Left: K+ β distribution before any cuts, Right:K+ β distribution after timing cut. 904.10 ∆β for the K+ before any cuts. . . . . . . . . . . . . . . . . . . . . . . . . . . 914.11 ∆β for the K+ after a ±1 ns timing cut. . . . . . . . . . . . . . . . . . . . . . 914.12 K+ calculated mass versus momentum after a ±1 ns timing cut. . . . . . . . . 924.13 π− calculated mass versus momentum after a ±1 ns timing cut. . . . . . . . . . 924.14 Proton calculated mass versus momentum after a ±1 ns timing cut. . . . . . . . 934.15 Left: Invariant mass of p,π− before DOCA cut. Right: Invariant mass of p,π−

after DOCA cut. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 944.16 Kaon vertex distribution. The dashed lines are the implemented cuts. . . . . . . 954.17 Distance of closest approach; the dashed line is the implemented cut. . . . . . . 954.18 Angular distribution of the kaons before fiducial cuts. . . . . . . . . . . . . . . 974.19 Angular distribution of the kaons after fiducial cuts. . . . . . . . . . . . . . . . 98

5.1 Pion contamination: Mass squared (M2X ) of any missing particle for the

γ p→ pπ+π−(X) reaction where the π+ was a potentially mis-identified kaon.Events with M2

x < 0.01 GeV2 were removed. . . . . . . . . . . . . . . . . . . 1015.2 Pion contamination: Mass squared (M2

X ) of any missing particle for the γ p→π+π−(X) reaction, where the π+ was a potentially mis-identified kaon. . . . . 101

5.3 Λ peak from the invariant mass of the π− and proton. . . . . . . . . . . . . . . 102

Page 12: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

12

5.4 No kinematic cuts applied. Top left: Invariant mass of the p-π−. Top right:Missing mass squared of all detected particles. Bottom left: Missing mass offthe p-π−. Bottom right: Missing mass off the K+. All units are (GeV). . . . . . 104

5.5 Plots after Λ cut only. Top left: Missing energy of all particles detected. Topright: Missing mass squared of all detected particles. Bottom left: Missingmass off the Λ. Bottom right: Transverse missing momentum. All units are(GeV). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105

5.6 Missing mass off the K+ after the cut on the Λ. The fit to the Σ∗ uses arelativistic Breit-Wigner and quadratic background. The dashed line indicatesthe cut around the Σ∗0. The Λ(1520) is also clearly visible. . . . . . . . . . . . 106

5.7 Plots after Σ∗ cut only. Top left: Missing energy of all particles detected. Topright: Missing mass squared of all detected particles. Bottom left: Missingmass off the Λ. Bottom right: Transverse missing momentum. All units are(GeV). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107

5.8 Plots after Σ∗ and Λ restrictions. Top left: Missing energy of all particlesdetected. Top right: Missing mass squared of all detected particles. BottomLeft: Missing mass off the Λ (counts below the dashed line are cut). Bottomright: Transverse missing momentum. All units are (GeV). . . . . . . . . . . . 108

5.9 Missing energy produced from simulations for the reaction γ p→ K+Λπ0. . . . 1095.10 Left: Transverse missing momentum Pxy after only one tagged photon is

selected. Right: P2xy after only one tagged photon is selected with cut

implemented, seen as the dotted line. . . . . . . . . . . . . . . . . . . . . . . . 1105.11 Simulations for the perpendicular momentum Pxy of missing mass candidates.

The blue distribution is for the simulated γ p → K+Λ(π0) reaction and theyellow is for the γ p→ K+Λ(γ) reaction. . . . . . . . . . . . . . . . . . . . . . 111

5.12 Magnification of the simulation of the perpendicular momentum Pxy of missingmass candidates and the implemented cut showing little effect on signals. . . . 112

5.13 Plots after all mentioned cuts. Top left: Missing energy of all particles detected.Top right: Missing mass squared of all detected particles. Bottom left: Missingmass off the Λ. Bottom right: Transverse missing momentum. All units are(GeV). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

7.1 Left: momentum resolution from data used to match Monte Carlo. Right: thematching of Monte Carlo to data by smearing out the measured resolution, redis Monte Carlo and blue is data. . . . . . . . . . . . . . . . . . . . . . . . . . 123

7.2 Comparison between data and Monte Carlo width of the missing mass squareddistribution. The left plot show the data with a Gaussian fit, while the middle isthe Monte Carlo before smearing and the right is the Monte Carlo after smearing.125

7.3 Decay time for the Λ from the Monte Carlo generated for γ p→K+Σ∗→K+Λπ0.1287.4 Monte Carlo distributions for the left: K∗, middle: Λ(1405), and right: Σ∗0. A

Breit-Wigner fit is used to demonstrate accurate width and mass in each case. . 129

Page 13: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

13

7.5 Left: Photon energy distribution for data using unskimed data without anycuts. Right: Photon energy distribution taken from the Monte Carlo generatorshowing the bremsstrahlung distribution. . . . . . . . . . . . . . . . . . . . . . 130

7.6 Left: Cross section and fit function used in Monte Carlo generation for thereaction Σ∗ → Λπ0. Right: Comparison between data and generated crosssection after using correction to the photon energy distribution. . . . . . . . . . 131

7.7 Approximated differential cross section for γ p→ K+Σ∗0→ K+Λπ0 in cosθKin the center-of-mass frame. . . . . . . . . . . . . . . . . . . . . . . . . . . . 132

7.8 Final comparison between Monte Carlo (lines) and data (points with errors)for the reaction γ p→ K+Σ∗0 → K+Λπ0. Top left: cosθ of kaon in the labframe. Top right: kaon momentum distribution. Bottom left: pion momentumdistribution. Bottom right: proton momentum distribution. . . . . . . . . . . . 134

7.9 Monte Carlo for the γ p→K+Σ∗0→K+Λγ channel. Left: the K+pπ− missingmass squared. Right: the missing mass off the Λ. . . . . . . . . . . . . . . . . 135

7.10 Monte Carlo for the γ p → K+Σ∗0 → K+Λπ0 channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . . 1357.11 Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ0π0 channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . . 1357.12 Monte Carlo for the γ p→ K+Λ(1405)→ K+Λγ channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . 1367.13 Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ0γ channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . 1367.14 Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ+π− channel (Before Cuts).

Left: the K+pπ− missing mass squared. Right: the missing mass off the Λ. . . 1367.15 Monte Carlo for the γ p → K+Σ∗ → K+Σ+π− channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . 1377.16 Monte Carlo for the γ p→ K∗+Λ→ ΛK+γ channel. Left: the K+pπ− missing

mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . . . . . . 1377.17 Monte Carlo for the γ p→K∗+Λ→ΛK+π0 channel. Left: the K+pπ− missing

mass squared. Right: the missing mass off the Λ. . . . . . . . . . . . . . . . . 1377.18 Left: Monte Carlo missing energy for the γ p→ K∗+Λ→ ΛK+π0 channel.

Right: Monte Carlo missing energy for the γ p → K+Λ(1405) → K+Σ0π0

channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138

8.1 Left: χ2 distribution for the π0 hypothesis with a fit using Eq. 8.2. Right:confidence level distribution for the π0 hypothesis. . . . . . . . . . . . . . . . 143

8.2 Left: χ2 distribution for the γ hypothesis with a fit using Eq. 8.2. Right:confidence level distribution for the γ hypothesis. . . . . . . . . . . . . . . . . 143

8.3 Left: χ2 distribution for γ hypothesis with a fit using Eq.8.2. Right confidencelevel distribution for γ hypothesis. . . . . . . . . . . . . . . . . . . . . . . . . 145

Page 14: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

14

8.4 Missing mass squared distribution for the events that are going into the secondstep of the kinematic fitting procedure. The kinematic fit to π0 satisfyingP(χ2) < 10% shows the radiative candidates (yellow), as well at the rejectedbackground from the Λ(1405) (green). The white region shows the π0

candidates from a P(χ2)≥ 10% cut. . . . . . . . . . . . . . . . . . . . . . . . 1468.5 Left: data (error bars) with Monte Carlo (line) from the Λπ0 channel and the

full spectrum of Λ(1405) filled to match the data. Right: data (error bars) withMonte Carlo (line) from the Λ(1405) only. . . . . . . . . . . . . . . . . . . . . 152

8.6 Left: fit with a Gaussian and quadratic background for the hyperon mass range1.34-1.43 GeV. Right: fit with polynomial to the point derived from Gaussianfits of various mass windows. . . . . . . . . . . . . . . . . . . . . . . . . . . . 155

8.7 Missing mass off of the Λ for the match of data and Monte Carlo to obtain thecounts from K∗. The lines is data and the points are from Monte Carlo afteradding in the K∗ and Λ(1405). . . . . . . . . . . . . . . . . . . . . . . . . . . 155

8.8 A study over a variation in confidence level; each cut corresponds to Pπ0(χ2)= Pγ(χ2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162

8.9 Only the selected confidence level cuts with the best quality signal based on theP1 parameter in the χ2 fit (values seen in Table 8.5). Pπ0(χ2)/Pγ(χ2) is used onthe x-axis to obtain a distinguishable point for each ratio value. . . . . . . . . . 163

8.10 The fit to the χ2 distribution from Monte Carlo for various mixtures ofradiative signal and Λπ0 events for small statistics, after cutting events withPa

π (χ2) < 1%. The amount of π0 background still present in each case can bedetermined by looking at the number of events added from the pure singal casein the top left plot. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167

8.11 Left: χ2 distribution and fit with two degrees of freedom for the π0 candidatesfrom the g11a data after the Pa

γ (χ2) < 1% cut and before the Pbπ (χ2) > 10%

cut; Right: the corresponding confidence level distribution for the π0 candidates.1708.12 Left: χ2 distribution and fit with two degrees of freedom for the γ candidates

from the g11a data before the Paπ (χ2) < 1% cut; Right: the corresponding

confidence level distribution for the γ candidates. . . . . . . . . . . . . . . . . 1708.13 Left: χ2 distribution and fit with two degrees of freedom for the γ candidates

from the g11a data after the Paπ (χ2) < 1% cut and before the Pb

γ (χ2) > 10%cut; Right: the corresponding confidence level distribution for the γ candidates. 171

8.14 Left: the nπ counts extracted using the confidence level cuts Paγ < 0.01 and

Pbπ > 0.1. Middle: the nγ counts extracted using the confidence level cuts

Paπ < 0.01 and Pb

γ > 0.1. Right: the counts nπ and nγ shown in the spectrumbefore any kinematic fit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174

8.15 The confidence level distribution with the additional constraint on the invariantmass of the Σ∗0 using a missing π0 hypothesis. . . . . . . . . . . . . . . . . . 176

8.16 The χ2 and confidence level distribution with the additional constraint on theinvariant mass of the Σ∗0 with the missing γ hypothesis. . . . . . . . . . . . . . 177

Page 15: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

15

9.1 The variation in the DOCA cuts in centimeters with the acceptance correctedbranching ratio shown in (%). . . . . . . . . . . . . . . . . . . . . . . . . . . . 179

9.2 The variation in the perpendicular momentum cuts with the acceptancecorrected branching ratio shown in (%). . . . . . . . . . . . . . . . . . . . . . 180

9.3 Left: Variation in the acceptance corrected branching ratio for various t-slopesetting in the generator for the π0 channel; right: for the radiative channel. . . . 181

10.1 The ∆β distribution for the π− and the cut applied to clean up identification. . . 18910.2 Invariant mass of the π+-π− combination for the two different π+ detected,

prior to any π+ organization to optimize the K0 cut. . . . . . . . . . . . . . . . 19010.3 Invariant mass of the π+-π− combination for the two different π+ after π+

organization to optimize the K0 peak. . . . . . . . . . . . . . . . . . . . . . . 19010.4 The confidence level distribution for the (2-C) kinematic fit with constraint on

π+ and π− to be the mass of the K0 with a missing mass off the K0 of the Σ∗+. 19110.5 (a) The invariant mass of the π+ π− after the best π+ is selected. (b) The

invariant mass of the π+ and π− after the P(χ2) > 0.1% confidence level cut. . 19210.6 (a) The missing mass off the K0 before the confidence level cut. (b) The

missing mass off the K0 after the P(χ2) > 0.1% confidence level cut. . . . . . . 193

11.1 (a) ∆p before the correction, showing EC counts for inner and outer blocklayers. (b) The pk f it vs pmeas for EC counts for inner and outer block layersbefore corrections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198

11.2 (a) ∆p after correction showing EC counts for inner and outer block layerstogether. (b) The pk f it vs pmeas for EC counts for inner and outer block layersafter corrections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198

11.3 Demonstration of cosθ vs φ for the neutron candidates. . . . . . . . . . . . . . 19911.4 Demonstration of cosθ vs φ for the detected neutrons. . . . . . . . . . . . . . . 20011.5 The g11a data set is used to obtain ∆p over a range of p (upper left), ∆p over a

range of φ (upper right), ∆p over a range of θ (lower left), ∆p over a range ofnuetron path length in (cm) (lower right). . . . . . . . . . . . . . . . . . . . . 202

11.6 Monte Carlo is used to obtain ∆p over a range of p (upper left), ∆p over a rangeof φ (upper right), ∆p over a range of θ (lower left), ∆p over a range of neutronpath length in (cm) (lower right). . . . . . . . . . . . . . . . . . . . . . . . . . 203

11.7 The g11a study of σ(p). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required rangeof p. The lower right plot shows the χ2 from each Gaussian fit for every binused. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204

11.8 The g11a study of σ(θ). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required rangeof θ . The lower right plot shows the χ2 from each Gaussian fit for every binused. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

Page 16: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

16

11.9 The g11a study of σ(φ). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required rangeof φ . The lower right plot shows the χ2 from each Gaussian fit for every binused. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206

11.10The Monte Carlo study of σ(p). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussianfits. The lower left plot shows the values of σ from the Gaussian fits over therequired range of p. The lower right plot shows the χ2 from each Gaussian fitfor every bin used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207

11.11The Monte Carlo study of σ(θ). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussianfits. The lower left plot shows the values of σ from the Gaussian fits over therequired range of θ . The lower right plot shows the χ2 from each Gaussian fitfor every bin used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208

11.12The Monte Carlo study of σ(φ). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussianfits. The lower left plot shows the values of σ from the Gaussian fits over therequired range of φ . The lower right plot shows the χ2 from each Gaussian fitfor every bin used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209

11.13The g11a data study of σ(θ). The upper left plot shows ∆θ over a range ofmomentum. The upper left shows the resolutions from the Gaussian fits of ∆θ

as a function of momentum. The middle right shows the ∆θ over a range ofφ . The middle left shows the resolutions from the Gaussian fits of ∆θ as afunction of φ . The bottom left plot shows ∆θ over a range of θ . The bottomleft shows the resolutions from the Gaussian fits of ∆θ as a function of θ . A cuton the cosine of the angle between the missing neutron vector and the detectedneutron vector of 3 is used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 210

11.14The g11a data study of σ(φ). The upper left plot shows ∆φ over a range ofmomentum. The upper left shows the resolutions from the Gaussian fits of ∆φ

as a function of momentum. The middle right shows the ∆φ over a range ofφ . The middle left shows the resolutions from the Gaussian fits of ∆φ as afunction of φ . The bottom left plot shows ∆φ over a range of θ . The bottomleft shows the resolutions from the Gaussian fits of ∆φ as a function of θ . A cuton the cosine of the angle between the missing neutron vector and the detectedneutron vector of 3 is used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 211

Page 17: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

17

11.15The Monte Carlo study of σ(θ). The upper left plot shows ∆θ over a rangeof momentum. The upper left shows the resolutions from the Gaussian fits of∆θ as a function of momentum. The middle right shows the ∆θ over a rangeof φ . The middle left shows the resolutions from the Gaussian fits of ∆θ as afunction of φ . The bottom left plot shows ∆θ over a range of θ . The bottomleft shows the resolutions from the Gaussian fits of ∆θ as a function of θ . A cuton the cosine of the angle between the missing neutron vector and the detectedneutron vector of 3 is used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 212

11.16The Monte Carlo study of σ(φ). The upper left plot shows ∆φ over a rangeof momentum. The upper left shows the resolutions from the Gaussian fits of∆φ as a function of momentum. The middle right shows the ∆φ over a rangeof φ . The middle left shows the resolutions from the Gaussian fits of ∆φ as afunction of φ . The bottom left plot shows ∆φ over a range of θ . The bottomleft shows the resolutions from the Gaussian fits of ∆φ as a function of θ . A cuton the cosine of the angle between the missing neutron vector and the detectedneutron vector of 3 is used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 213

12.1 The invariant mass of the π+1 -π−, (upper left), missing mass off the π

+1 -

π− (upper right), the n-π+2 invariant mass (lower left), and the missing mass

squared of all the detected particles (lower right). All distributions are beforeany kinematic constraints. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218

12.2 The invariant mass of the n-π+2 , (upper left), missing mass off the K0 (upper

right), the missing mass off the n-π+2 combination (lower left), and the missing

mass squared of all the detected particles (lower right), after the±0.005 K0 peak.21912.3 The missing mass off the K0 (upper left), and the missing mass off the Σ+

(upper right), the missing mass squared of all detected particles (lower left),and a magnification of the missing mass squared with Gaussian fit to the π0

region (lower right) . These distribution are made after take a±0.02 cut aroundthe Σ+ peak as well as the prior cut of ±0.005 on the K0 peak. . . . . . . . . . 220

12.4 The comparison between the final missing mass squared distribution from theset of cuts discribed after a ±0.05 cut around the Σ∗+ mass (upper left) andfrom the P(χ2) > 0.1% confidence level cut used obtain the previous K0 andΣ∗+ candidates. The same ±0.02 cut around the Σ+ peak is used in both. . . . . 221

12.5 The missing mass off the Σ+ after all cuts described (no kinematic fit). Thedistribution is similar for the the kinematic fit (not shown). . . . . . . . . . . . 222

13.1 Confidence level distribution for a (4-C) kinematic fit π+π−π+n. . . . . . . . . 22713.2 Left:Showing the (4-C) kinematic fit to get the invariant mass of π+n, Right:

improvement in missing mass off the Σ+ (both in GeV). . . . . . . . . . . . . . 22813.3 Left:Showing the (5-C) kinematic fit to the invariant mass of π+n, Right:

improvement to the missing mass off the Σ+ (both in GeV). . . . . . . . . . . . 22813.4 Pull distributions for a (4-C) kinematic kit π+π−π+n. . . . . . . . . . . . . . . 229

Page 18: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

18

14.1 The missing mass off the N∗ using a ±0.05 invariant mass cut on the π+ncombination at a mass of 1440 MeV (left) demonstrating a clear ω(782) peak;(right) at a mass of the Σ+ demonstrating that the ω(782) peak is still present. 231

14.2 The γ p→ K0Σ∗+→ K0Σ+γ Monte Carlo distibutions for the π+1 -π− invariant

mass, the missing mass off the π+1 -π− combination, the missing energy from

all detected particle, and the missing mass squared of all detected particles. . . 23514.3 The γ p→K0Σ∗+→K0Σ+π0 Monte Carlo distibutions for the π

+1 -π− invarinat

mass, the missing mass off the π+1 -π− combination, the missng energy from

all detected particle, and the missing mass squared of all detected particles. . . 23614.4 The ωN(1440) → π+π−π0nπ+ Monte Carlo distributions for the π

+1 -π−

invarinat mass, the missing mass off the π+1 -π− combination, the missing

energy from all detected particle, and the missing mass squared of all detectedparticles, no cuts yet applied. . . . . . . . . . . . . . . . . . . . . . . . . . . . 237

14.5 The ρN(1520) → π+π−π0π+n Monte Carlo distibutions for the π+1 -π−

invariant mass, the missing mass off the π+1 -π− combination, the missing

energy from all detected particle, and the missing mass squared of all detectedparticles, no cuts yet applied. . . . . . . . . . . . . . . . . . . . . . . . . . . . 238

14.6 The ηnπ+ → π+π−π0π0nπ+ Monte Carlo distibutions for the π+1 -π−

invarinat mass, the missing mass off the π+1 -π− combination, the missng

energy from all detected particle, and the missing mass squared of all detectedparticles, no cuts yet applied. . . . . . . . . . . . . . . . . . . . . . . . . . . . 239

14.7 The K∗0Σ+ → K0Σ+π0 Monte Carlo distributions for the π+1 -π− invariant

mass, the missing mass off the π+1 -π− combination, the missing energy from

all detected particle, and the missing mass squared of all detected particles, nocuts yet applied. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240

15.1 The variation in the ratio due to the confidence level cut from the kinematic fitof π

+1 -π− combination to the invariant mass of K0 with missing mass of Σ∗0. . 251

15.2 Variation in the acceptance corrected branching ratio for various t-slope settingin the generator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253

15.3 Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts are equal for eachhypothesis such that Pγ(χ2) = Pπ(χ2). . . . . . . . . . . . . . . . . . . . . . . 256

15.4 Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts for the π0 hypothesisis kept at 10% while the radiative hypothesis confidence level cut, Pγ(χ2), isvaried. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257

15.5 Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts for the radiativehypothesis is kept at 10% while the π0 hypothesis confidence level cut,Pπ0(χ2), is varied.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258

15.6 Variation in the ratio for a range of missing energy cuts on the final extractionof the radiative signal. Here the cuts are only applied to the radiative hypothesis. 260

Page 19: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

19

15.7 The variation in ratio from the cut on the Σ+ invariant mass. . . . . . . . . . . 261

Page 20: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

20

1 INTRODUCTION

The quark model tells us that the properties of hadrons that arise from the quantum

numbers are determined by their valence quarks. For example consider the proton, which

is made of two up quarks of charge +2/3 and one down quark of charge −1/3. Adding

these together make a total charge of +1. However, quarks also carry color charge,

hadrons have zero total color charge because of a phenomenon known as color

confinement. This means that hadrons are always colorless. To achieve a colorless

combinations consider a Baryon with three quarks, each with different colors, or a meson

with a quark of one color and an antiquark with the required anticolor.

The hadrons are held together by the strong interaction. The theory that encompasses

the physics that bind the quarks into hadrons is Quantum Chromodynamcs (QCD). QCD

dictates the behavior of the strong force and the interaction between the quarks and gluons.

It is a non-linear theory and there are at present no analytic solutions. This means that

only an approximate modeling of the nucleon spectrum is possible. There are a variety of

theories which attempt to incorporate empirical information into functional models.

There has been a great deal of attention given to excited states of the nucleon. The

transition form factors posses characteristic information on such things as the exited state

wave function and the spin structure of the transition. Precise measurements can test

baryon structure models and lead to a greater understanding of the strong interaction.

1.1 Quantum Chromodynamics

In the representation of the color SU(3) gauge group, quarks are the fermionic flavor

carriers. Gluons, with field four-vector Aaµ with a = 1, . . .8 are the exchange particles (or

gauge bosons) for the color force, and represent the gauge fields in the octet

representation. Flavor is a degree of freedom independent of color.

Page 21: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

21

The QCD Lagrangian is expressed as

LQCD =−12

trGµνGµν + ∑f =u,...,b

q f (i /D−m)q f (1.1)

where the notation is defined by

Gµν = ∂µAν −∂νAµ − ig[Aµ ,Aν ],

Dµ = ∂ − igAµ

Aµ =8

∑a=1

λ a

2Aa

µ ,

where the λ a are the Gell-Mann matrices. For the strong interaction, the mass matrix m is

diagonal in flavor space. The quintessential characteristics of QCD are found in the

behavior of asymptotic freedom and infrared slavery. Asymptotic freedom is the property

that causes the interaction between quarks to become arbitrarily weak at very small length

scales (sub-femtometer) or, equivalently, large energy scales (tens of GeV). Infrared

slavery, or quark confinement, is the phenomenon that color quarks cannot be isolated

singularly, and ultimately cannot be directly observed. Both can be characterized by the

momentum dependent coupling constant from the theory of renormalization. The

coupling of each QCD vertex is therefore energy dependent and is on the order of unity at

low energies. At this energy, more complicated diagrams may contribute on the same

order as the leading-order diagrams. The opposite is true for high energy, making higher

order diagrams essentially negligible in most cases.

From experiments, it is seen that quark-antiquark pairs (mesons) can be found in

nature such as the π− (du) and K+ (us). The three-quark combinations (baryons) are

prominent such as the proton (uud), neutron (udd) or the Λ (uds). The masses of the up

and down quarks in the proton make up only ∼ 1% of the total proton mass. Clearly there

is additional phenomena and/or highly-relativistic quarks which are not contained in many

models. Ultimately, at the energy scale of nuclear physics, phenomenological tools and

models with degrees of freedom different than true QCD are the most successful.

Page 22: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

22

1.2 Spectroscopy

The central focus of hadronic physics is to develop more detailed descriptions of the

transition between quark-gluon and hadronic degrees of freedom. Much effort in recent

years has taken place on improving the picture of the nucleon spectrum. Atomic

spectroscopy uses excitations of atoms to investigate the governing interactions. Studying

the nucleon excitations or resonances (N∗) provides essential information on the behavior

of quarks and the baryon structures. Quantum mechanics describes energy transitions in

discrete intervals leading to descriptions of excited baryons states that also exist in

discrete intervals. This corresponds to observation. Using the experimental results,

theories are able to derive highly successful models that are only approximations of full

QCD, but yield practical predictions.

1.3 The Constituent Quark Model

The Constituent Quark Model (CQM) has been reasonably successful in its

predictions. The CQM uses only the valence quarks as relevant degrees of freedom. A

simplified picture is employed such that excitations can be expressed as the result of

changes to the angular momentum of these quarks. This type of simplification makes

predictions such as the masses of baryon resonances possible to calculate. It is found that

the number of states predicted by the CQM calculation compared with those actually

observed is approximately 4:1 [60]. Because of the many successes of the CQM, these

missing states have been the central focus and motivation of many experiments and

analysis. The Missing Baryons Problem may be an indication that the model, though very

useful in some cases, does not reflect nature well enough to provide reliable predictions.

Page 23: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

23

1.4 Electromagnetic decays

A nice demonstration of the CQM predictive capabilities is the calculation of the

magnetic moments of the low-mass baryons, using the SU(6) wave functions [1, 2]. These

calculations [3] are within ∼10% of the current measured values [60]. In the CQM,

quarks are treated as point-like Dirac dipoles. However, nucleons are known to be far

more structured. For example a third of the of the spin of the proton comes from the

valence quarks [4], with the rest of the spin coming from a combination of the gluon spins

and orbital motion of the quarks [5, 6]. Despite these complexities, the CQM captures the

essential degrees of freedom that are relevant to the magnetic moments. Each new high

precision measurement of a baryon magnetic moment leads to new model restrictions and

a deeper understanding of the nature of the baryon wave-functions.

Measurements of electromagnetic (EM) decays can be challenging as they are often

quite small and the EM transitions of decuplet-to-octet baryons can be overwhelmed by

the strong decays. For example the branching ratio of the Σ∗0→ Λγ has been shown [10]

to be on the order of 1% but the signal is buried by the stronger Σ∗0→ Λπ0 decay.

The resulting width from the previous measurement of the Σ∗0→ Λγ [10] is much

larger than most current theoretical predictions. This could be due to meson cloud effects,

which were not included in these calculations. There is a theoretical basis [9] that suggests

that pion cloud effects contribute on the order of ∼40% to the γ p→ ∆+ magnetic dipole

transition form factor, G∆M(Q2), for low Q2. The CQM [11] indicates that the value of

G∆M(0) is directly proportional to the proton magnetic moment [9], and measurements of

G∆M for low Q2 are rationalized in the frame work of the model if the experimental

magnetic moment is lowered by about 25%.

Hyperons produced through low-rate strangeness-conserving reactions can be used to

measure EM transitions to other decuplet baryons [10]. These small electromagnetic

Page 24: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

24

decay branching ratios are, as stated above, difficult to measure directly. Ultimately these

measurements are needed to understanding the nature of the baryon wave-functions.

If the EM transition form factors for decuplet baryons with strangeness are also

sensitive to meson cloud effects, models attempting to make prediction of the decuplet

radiative decay widths will need revisions to incorporate this effect. Comparison of data

for the EM decay of decuplet hyperons, Σ∗, to the present predictions of quark models

provides a measure of the importance of meson cloud diagrams in the Σ∗→ Y γ transition.

Establishing experimental results for all three charged states has not yet been achieved.

With theoretical predictions for the degree at which the meson cloud effect plays a

role, it is then possible to test SU(3) flavor symmetry breaking (and the degree at which it

is broken) with a measurement of Σ∗→ Y γ decays. This can be best achieved by

measuring both Γ(Σ∗−→ Σ−γ) and Γ(Σ∗+→ Σ+γ) decay widths.

U-spin symmetry forbids radiative decay of specific decuplet baryons. U-spin is

analogous to isospin in the sense that it is a symmetry in the exchange of the d and s

quarks rather than the u and d quarks, see Section 2.2. A value U of U-spin can be

assigned to each baryon based on the quark composition. The Σ∗− and the Ξ∗− of the

baryon decuplet have U = 3/2 whereas the octet baryons Σ− and Ξ− have U = 1/2. Since

the photon has U = 0, this implies that

Σ∗−→ Σ

−γ and Ξ

∗−→ Ξ−

γ

have zero amplitude in the equal-mass limit due to U-spin symmetry. This can also be

understood in the context of the SU(6) wavefunctions for these baryons. The M1

transition operator is written between initial and final states :

〈Σ−SU(6)|∑q

Qq

2mqσq ·(

k× ε∗λ)|Σ∗−SU(6)〉= 0.

Page 25: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

25

Here the sum is over all q constituent quarks, and mq is the mass of the q quark. One can

also show that doing the same for the Σ∗+ gives a non-zero amplitude. U-spin invariance

implies a large difference in the radiative decay widths of Σ∗− and Σ∗+.

The chiral symmetry for U-spin is strongly broken because the constituent mass of

the strange quark, ms, is approximately 1.5 times greater than the non-strange quarks, m.

The magnetic moment is inversely proportional to the mass, and so there is no

cancellation like in the equal-mass SU(6) case above. From Ref. [12], an estimate of the

correction givesΓ(Σ∗−→ Σ−γ)Γ(Σ∗+→ Σ+γ)

=19

(1− m

ms

)2

which is a ratio of about 1%, suggesting that U-spin symmetry breaking for radiative

decays is at the level of only a few percent. This implies that U-spin is an effective tool,

even considering the quark mass difference.

Detailed corrections have been carried out by several groups [13, 14], all of which

come up with corrections that are of a few percent. In lattice QCD, the quarks have much

different interactions with the photon than for the CQM, but these too have ratios (for the

above equation) within a few percent [15]. This makes a stronger case for the usefulness

of U-spin symmetry.

There has been much theoretical interest in radiative baryon decays. However, there

are only a few measurements. Recently, a measurement the radiative decay of the Σ∗−

been attempted by the SELEX collaboration [16], resulting in only an upper limit. The

90% confidence level upper bound of Γ = 9.5 keV was reported, however most models

predict a value of less than 4 keV. Ultimately this result is not very useful in constraining

theoretical estimates. More experimental measurements are necessary with better

experimental constraints.

A set of steps have been carried out to investigate the various Σ∗ electromagnetic

decay at CLAS. First, a check on the previous CLAS measurement of the EM decay of the

Page 26: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

26

Σ∗0 was done. The previous measurement is quite high compared to nearly all available

calculations, and confirmation of the result is required to determine if the result is reliable

and hence if meson cloud effects are indeed contributing significantly. The next step was

to measure the Σ∗+ electromagnetic decay which has not been done before. The topology

of the reaction γ p→ K0Σ∗+→ K0Σ+γ in its final set of decay products is

γ p→ π+π−nπ+γ , compared to the Σ∗− case γ p→ K+Σ∗−→ K+nπ−γ . The advantages

of the Σ∗+ is that its rate is expected to be much larger, at nearly an order of magnitude

larger than the Σ∗− electromagnetic decay. Also, the CLAS data set g11a, available for the

positive channel, is quite large.

The first part of this thesis is dedicated to developing a method to extract the small

radiative signal out from under a large π0 background. The result from the first of the

present measurements is the EM decay width Σ∗0→ Λγ normalized to the strong decay

Σ∗0→ Λπ0. These results can be compared to previous measurements of the Σ∗0 EM

decay [10] that had a larger uncertainty (∼25% statistical and ∼15% systematic

uncertainty). The smaller uncertainties here are due to a larger data set (more than 10

times bigger) and subsequently a better control over systematic uncertainties. The reduced

uncertainty is important because, as mentioned above, meson cloud effects are predicted

to be on the order of ∼25-40%. In order to know quantitatively the effect of meson clouds

for baryons with non-zero strangeness, it is desirable to keep measurement uncertainties

below ∼10%.

There are many calculations of the EM decays of decuplet hyperons such as: the

non-relativistic quark model (NRQM) [17, 23], a relativistic constituent quark model

(RCQM) [24], a chiral constituent quark model (χCQM) [25], the MIT bag model[26],

the bound-state soliton model[27], a three-flavor generalization of the Skyrme model that

uses the collective approach[28, 29], an algebraic model of hadron structure[30], among

Page 27: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

27

others. Table 1.1 summarizes the theoretical predictions and experimental branching

ratios for the EM transitions of interest.

Table 1.1: Theoretical predictions of the decay widths for the model shown ofelectromagnetic decays of various baryons (all in units of keV).

Model ∆→ Nγ Σ→ Λγ Σ∗0→ Λγ Σ∗+→ Σ+γ

NRQM[17, 23, 26] 360 8.6 273 104

RCQM[24] 4.1 267

χCQM[25] 350 265 105

MIT Bag[26] 4.6 152 117

Soliton[27] 243 91

Skyrme[28, 29] 309-326 157-209 47

Algebraic model[30] 341.5 8.6 221.3 140.7

Experiment[60] 660±47 9.1±0.9 470±160

1.5 The CLAS detector

A comprehensive investigation of electromagnetic strangeness production has been

carried out using the CLAS detector at Thomas Jefferson National Accelerator Facility.

Many data on ground-state hyperon photoproduction have already been published

[32, 33, 34] using data from the so-called g1 and g11 data sets. The g1 experiment had an

open trigger [32] and lower data acquisition speed, whereas the g11 experiment required

at least two particles to be detected [34], lowering the trigger rate, yet had a much higher

data acquisition speed. The g11 data had a higher average beam current, resulting in over

20 times more useful triggers than for the g1 data. The present results used the g11 data

set whereas the previous CLAS measurement used the g1 data set. Previous CLAS results

Page 28: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

28

for other known reactions give confidence to the corresponding calibration of these data

sets [34].

As previously mentioned, the EM decay of the Σ∗0 is only about 1% of the total

decay width. To isolate this signal from the dominant strong decay Σ∗0→ Λπ0, the

missing mass of the detected particles, γ p→ K+Λ(X) is calculated. Because of its

proximity to the π0 peak in the mass spectrum from strong decay, the EM decay signal is

difficult to separate using simple peak-fitting methods. The strategy here is to understand

and eliminate as much background as possible using standard kinematic cuts, and then use

a kinematic fitting procedure for each channel. As described later, by varying the cut

points on the confidence levels of each kinematic fit, the systematic uncertainty associated

with the extracted branching ratio for EM decay can be quantitatively determined. The

increased statistics for the g11 data also helps greatly to reduce the systematic uncertainty.

The second half of this thesis is dedicated to development of a method that can be

used to extract the electromagnetic decay signal from the more difficult topology of the

Σ∗+. The Σ∗+→ Σ+γ is estimated to be event smaller than that of the Σ∗0→ Λγ but with

the same proximity to the π0 peak in the mass spectrum. The same strategy (to understand

and eliminate as much background as possible using standard kinematic cuts and then use

a kinematic fitting procedure) is used for each channel.

1.6 Previous Experimental Results

The total and differential cross section are analyzed in the present analysis to improve

the Monte Carlo simulations of the γ p→ K+Σ∗0 reaction. The differential cross section

has been measured in the photon energy range from 1.5 to 2.4 GeV and the angular range

of 0.8 < cosθK < 1.0 in the K+ center-of-mass angle at SPing-8 from the LEPS

collaboration [46].

Page 29: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

29

In the LEPS experiment, forward going K+’s from the γ p→ K+X reaction were

detected in the spectrometer. A time projection chamber was used together with the

spectrometer to assist in the detection of the decay products. The main decay mode was

identified by detecting the Λ and the K+. The decay to Σπ is also measured. The missing

mass spectrum of the γ p→ K+X reaction for events with K+Σπ is shown if Figure 1.1.

The experimental results are shown as closed circles. The data was fit with a shape

determined by MC simulations. The solid histograms show the fit results. The strengths of

the Σ∗0 production were obtained from the Λπ0 decay mode and the other reaction

channels were obtained by fitting. The solid fit line, open circle, dashed lines, and

dot-dashed indicate the spectra of the K+Λ(1405), K+Σ(1385), nonresonant K+Σπ and

K∗0Σ+ production, respectively.

Figure 1.1: The missing mass of the γ p→ K+X reaction in the photon energy range (a)1.5 < Eγ < 2.0 GeV and (b) 2.0 < Eγ < 2.4. The data points are shown in closed circles.The strengths of the Σ∗0 production were obtained from the Λπ0 decay mode and the otherreactions were obtained by fitting. The solid fit line, open circle, dashed lines, and dot-dashed indicate the spectra from K+Λ(1405), K+Σ(1385), nonresonant K+Σπ and K∗0Σ+

production, respectively. Source [46].

Page 30: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

30

CLAS has also looked at the production cross section for γ p→ K+Σ∗0 in a

preliminary analysis with the g11a data set [47]. The two competing processes

γ p→ K+Σ∗0 and γ p→ K∗+Λ are studied. Simulations assuming a t-channel process

where used to correct for the detector acceptance. The Σ(1385) yield was achieved using a

p-wave Breit-Wigner function with background shape generated using K∗+Λ simulations.

Only total cross sections are presented.

There has also been some earlier work by the DESY Bubble Chamber Group [48]

that measured a total cross section. The CLAS and DESY results are shown in Figure 1.2.

Figure 1.2: DESY Bubble Chamber Group total cross section [48] in with CLAS total crosssection with parameterization fit for γ p→ K+Σ(1385). Source [49].

There has been previous work on the branching ratio of Σ∗0→ Λγ to Σ∗0→ Λπ0 by

Taylor [10] using g1c data set at CLAS. The result with its statistical uncertainty is

Page 31: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

31

1.53±0.39. The g11a data set has many more triggered events, ultimately leading to a

reasonable reduction of statistical uncertainty. Considerable effort is made to reduce

systematic uncertainty and to include systematic studies not performed in Taylor [10].

Here, other methods are used with better control over various systematic effects. The

essential differences in the steps taken by Taylor [10] include: (a) using a kinematic fit to

the Λ hypothesis prior to the final missing π0 or radiative hypothesis, (b) the method of

extracting background counts, and (c) using a Gaussian fit after the final kinematic fit to

obtain the counts and uncertainties. In computer simulations, Taylor used a model to

approximate the angular dependence of the γ p→ K+Σ∗0→ K+Λπ0 cross section

distribution. In the present analysis the angular distributions from the data are used.

Various degrees of improvement to the simulations and extraction of the radiative signal

are discussed in the analysis section of this thesis.

For completeness, some earlier work is mentioned that made the an attempt to

measure the radiative decay of the Σ∗0. Meisner [21] observed a single event consistent

with K0 p→ Σ(1385)π+→ Λγπ+. From this measurement a branching ratio of

0.17±0.17 was obtained. Colas et al. [22] determined the limits on the Σ(1385) radiative

decay of

Γ(Σ(1385)→ Λγ)Γ(Σ(1385)→ Λπ0)

< 0.06 (1.2)

Γ(Σ(1385)→ Σ0γ)Γ(Σ(1385)→ Λπ0)

< 0.05. (1.3)

The Σ∗+ electromagnetic decay has not been observed before and there are no

published cross sections. However, the CLAS collaboration is presently working on

differential cross section for the γ p→ K0Σ∗+ reaction.

Page 32: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

32

1.7 The Experiment

Here, a brief description is given, with more details in chapters to follow. The present

experiment was carried out in Hall B of the Thomas Jefferson National Accelerator

Facility. The particle physics detector is described in detail in Chapter 2. Below, the main

points of the present measurement are given.

In this experiment, a proton at rest is hit with a photon of known energy. The reaction

γ p→ KY ∗ is detected, where Y ∗ denotes the excited state hyperon of interest. The trigger

and detector timing information is used to select events. Each channel of interest has its

own set of competing background channels that must be understood. For example, the

γ p→ KY ∗ reaction can have the same topology as the γ p→ K∗Y . Other vector mesons

can have similar contaminating effects such as the γ p→ ωN∗, which can have the same

final decay products as the K0Σ∗+→ π+π−π+nπ0. The K+Σ∗0 is very close in the mass

sepctrum to the K+Λ(1405), complicating the separation of the two without signal loss.

The photon from the electromagnetic decay is not directly measured in the detectors

but is detected through the missing mass of the other final decay products. For example

the reaction γ p→ K+Σ∗0 has a possible final state of K+Σ∗0→ K+pπ−γ for its

electromagnetic decay. All the charged particle can be detected in the drift chambers and

the γ can be detected using conservation of energy and momentum such that

Pµx = Pµ

tot−Pµ

K −Pµp −Pµ

π , where Pµ

tot is the total incoming four-momentum, and Pµx is the

missing four-momentum. More explicitly the missing energy and momentum are,

Ex = Eγ +Mp−EK−Ep−Eπ

~Px = ~Pγ −~PK−~Pp−~Pπ

leading to a missing mass of,

Mx =√

E2x −~P2

x .

Page 33: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

33

This technique improves resolution and statistics compare to detecting all final decay

products. Additionally kinematic checks and constraints can be used by studying the

missing mass off a known decay product. For example in this case the missing mass off

the K+ will give the Σ∗0 mass.

For each γ p→ KΣ∗ reaction measured, both a radiative decay and a strong decay to a

π0 are possible. The work presented here is rooted in the development of the extraction

techniques of small radiative signals among other backgrounds. Its is only through

progress in event-weighting analysis that small signals are able to be observed in the

presence of overwhelming background.

By extracting these electromagnetic decay signals from the Σ∗0→ Λγ and

Σ∗+→ Σ+γ final states, a test for U-spin invariance is achieved by comparing the U-spin

SU(3) prediction for the Σ∗+→ Σ+γ partial width and the Σ∗0→ Λπ0 partial width to

same widths measured experimentally.

Through the results of this work it is possible to more accurately determine which

theoretical models provide reasonable predictions, and the approximate degree of

contribution from the meson cloud effects in the wave-functions of baryons.

Page 34: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

34

2 THEORETICAL OVERVIEW

Theoretical investigations of electromagnetic decays provide a framework to test

against. This provides a language to describe the possible distribution of different quark

configurations in hadronic matter. Modeling electromagnetic decays can indicate where

the language has become too simple. The interaction of photons with electric charges of

the quark fields provides an invaluable probe into strong interactions and helps to

eliminate models. The radiative processes allow a more complete theoretical

interpretation than that of the purely hadronic interactions. Some highlights of each model

listed in Table 1.1 is now discussed.

2.1 Theoretical Background

2.1.1 The NRQM

In the non-relativistic quark model (NRQM) [3], the force binding quarks together is

approximated to be linear defined by their separation distance, resulting in simple

harmonic oscillator wave-function at zeroth-order. The quarks naturally carry spin and

interact with each under the hyperfine interaction

H i jhyp =

2αS

3mim j

[8π

3(~ri j)~si ·~s j +1r3

i j(3~si · ri j~s j · ri j−~si ·~s j)

](2.1)

where the mass of quark i ( j) is mi (m j), the spin is si (s j) and the distance between quarks

is~ri j, and αS is a free parameter. Spin-orbit coupling is ignored, as the effects are

estimated to be relatively small.

Page 35: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

35

The wave-functions presented in Ref. [23] for the hyperons is∣∣∣Λ(12+)⟩

= 0.93∣∣Λ8

2SS⟩−0.30

∣∣Λ82S′S⟩−0.20

∣∣Λ82SM

⟩−0.05

∣∣Λ12SM

⟩, (2.2)∣∣∣Σ(1

2+)⟩

= 0.97∣∣Σ8

2SS⟩−0.18

∣∣Σ82S′S⟩−0.16

∣∣Σ82SM

⟩−0.02

∣∣Σ102SM

⟩, (2.3)∣∣∣Σ(3

2+)⟩

=∣∣Σ10

4SS⟩, (2.4)∣∣∣Λ(1

2−)⟩

= 0.90∣∣Λ1

2PM⟩+0.43

∣∣Λ82PM

⟩−0.06

∣∣Λ84PM

⟩, (2.5)∣∣∣Λ(3

2−)⟩

= 0.91∣∣Λ1

2PM⟩+0.40

∣∣Λ82PM

⟩+0.01

∣∣Λ84PM

⟩. (2.6)

The first term in the “ket” notation gives the baryon with a subscript of the SU(3)

multiplet, then a superscript 2J +1 where J is the total spin. The orbital angular

momentum is S or P with a subscript representing the SU(6) permutation symmetry. S′S

indicates the baryon excitation corresponding to the n = 2 harmonic oscillator level.

The Isgur and Karl model is used to calculate the radiative widths and amplitudes by

Darewych, Horbatsch, and Koniuk in Ref. [17]. In this calculation, the photoemission is

assumed to come from the de-excitation of a single quark. A nonrelativistic reduction of

the quark-photon interaction is put between the initial and final baryon states to obtain the

T-matrix element,

〈B′(p′,s′)γ(K,λ )|T |B(p,s)〉 =

−3ie(2π)3/2 〈B′(p′,s′)| e3

e

[~σ3 · (

~K×~ε∗)2m3

+ i~p′3·~ε

m3

]× e−i~K·~r3 |B(p,s)〉 (2.7)

for the B(p,s)→ B′(p′,s′)γ process with momentum p and spin s dependence. Here,

ε(K,λ ) is the photon polarization vector and e3,~r, m3, and 1/2~σ3 are the third quark

charge, position, mass and spin. The momentum of the third quark in B′ is denoted as ~p3′.

Three possible independent helicity amplitudes can be obtained using all photon

polarizations and spin states. The radiative widths are calculated as

Γγ =1

(2J +1)MB′

MB

q2

2π4∑

Jz

|AJz(q)|2, (2.8)

where J is the angular momentum and MB is the mass of the hyperon.

Page 36: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

36

2.1.2 The RCQM

The relativistic constituent quark model (RCQM) calculation in Table 1.1 comes

from the work of Warns, Pfeil, and Rollnik [24]. The model specifies using hyperons that

are built up of three massive point-like quarks in the RCQM framework. The amplitudes

for a process are related to the matrix elements

〈ΨY ′|HEM |ΨY 〉 , (2.9)

for the process Y ′→ Y + γ∗ where Y is the initial hyperon state and Y ′ is the final and the

photon γ∗ is either real or virtual. The RCQM uses electromagnetic transition amplitudes

that also assume a single quark transition hypothesis such that the incoming photon is

coupled to only one of the constituent quarks. After taking additional relativistic effects

into account and the center-of-mass motion of the three-quarks the electromagnetic

Hamiltonian H EM can be expressed as,

H EM = HnrSQ +H(2)

SQ +H(3)SQ +H(2)

I +H(3)I +H(2)

CM +H(3)CM +O

(1c4

), (2.10)

where HnrSQ is the non-relativistic Hamiltonian from the coupling between the photon field

and a single quark, and the second and third order relativistic corrections in 1/c are H(2)SQ

and H(3)SQ , similarly for the interaction potential between the quarks, H(2)

I and H(3)I . Finally

H(2)CM and H(3)

CM are the adjustments needed for the center-of-mass motion correction of the

three quark system.

The Isgur-Karl wave-functions (equations 2.2-2.6) were used with the u and d quark

masses at 350 MeV each and with the strange quark mass of 580 MeV. Numerical

calculations of the electromagnetic transition amplitudes with both transversly and

longitudinally polarized photons. Calculations of the form factors and electoexcitation

amplitudes were achieved in the equal velocity frame, valid only for a momentum transfer

of Q2 ≤ 3 GeV2. Only calculations for the neutral channel were done.

Page 37: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

37

2.1.3 The χCQM

In the chiral constituent quark model, spontaneously broken chiral symmetry of

low-energy QCD leads to constituent quarks and the pseudoscalar mesons as pertinent

degrees of freedom. The Hamiltonian, using three different quark masses mi is,

H =3

∑i=1

(mi +

Pi2

2mi

)− P2

2M−ac

3

∑i< j

λC ·λ C(ri− rj)2 +

3

∑i<j

Vres(ri,rj) (2.11)

where λ C ·λ C is the color space scalar product of the Gell-Mann matrices, P and M are

the center of mass momentum and mass respectively, ac is the confinement strength and

ri j is the distance between quarks i and j. This expression uses a quadratic confinement

potential. The V res term is the residual interaction potential which contains the one-gluon

exchange contribution as well as the chiral interaction potential from the pseudoscalar

meson exchange. The electromagnetic currents are built from the nonrelativistic reduction

of the Feynman diagrams that consist of the impulse approximation, the pseudoscalar

meson-pair current, the pseudoscalar meson in-flight current, the gluon pair current and

the scalar-exchange current. The helicity amplitudes as a function of the square of the

photon momentum are denoted as A3/2(q2) and A1/2(q2) which enables one to write the

expression for the partial decay width,

ΓM1 =ω2

π

Moct

Mdec

22J +1

∣∣A3/2(q

2)+A1/2(q2)∣∣. (2.12)

Here the total spin J = 3/2 is from the excited state baryon and ω is the resonance

frequency in the center of mass of the decaying hyperon. The E1 transition follows from

the same expression using the corresponding helicity amplitudes. The decay widths are

primarily governed by M1 transitions, which are determined by the impulse

approximation. This is due to a predominant two-body current cancellation.

Page 38: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

38

2.1.4 The MIT Bag Model

The concept of the bag model originated from the central idea of confining the colors

of quarks and gluons inside the hadron. The color electric fields can only exist within the

hadron interior where the vacuum is restored to the perturbative vacuum. The region

where the quarks and gluons are allowed is called the “bag.” The Lagrangian of the MIT

bag model is,

L =i2

ψ(←−/∂ −−→/∂ )ψθ(R− r)− 1

2ψψδ (r−R), (2.13)

where θ(R− r) is the Heaveside’s step function, ψ is the quark field, R is the bag radius,

with exterior mass M. The quark mass inside the bag is zero.

The starting point is to consider the Λ(12−) wave function and perform the SU(3)

symmetry breaking. Taking a non-zero strange quark mass allows for mixing of the

flavor-singlet∣∣Λ1

2⟩ and flavor-octet∣∣Λ8

2⟩. The basis states can be represented using the

SU(3) flavor multiplet, using again spin and parity of the excited quark. Using standard

MIT bag parameters, the first lowest-lying Λ(12−) state wavefunctions at 1364 MeV is,

|Λ〉1 = 0.39∣∣∣∣Λ1,

12

−⟩+0.42

∣∣∣∣Λ8,12

−⟩A+0.46

∣∣∣∣Λ8,12

−⟩B+0.67

∣∣∣∣Λ8,32

−⟩,(2.14)

where index A indicates that the state is totally symmetric while B indicate a mixed

symmetric wave-function in pseudospin for the octet states. The large (8, 32−

) component

is indicative of the low eigenfrequency of the p 32

mode in the bag. There is coupling of the

s 12

quarks to S = 1 and so it can couple to the Σ state in the radiative decay. The next state

predicted at 1446 MeV is,

|Λ〉2 = 0.93∣∣∣∣Λ1,

12

−⟩−0.11

∣∣∣∣Λ8,12

−⟩A−0.21

∣∣∣∣Λ8,12

−⟩B−0.26

∣∣∣∣Λ8,32

−⟩,(2.15)

which is closer to the Λ(12−) ground state in the SU(3) limit. An evaluation of the

radiative decay width is made for both Λ1 and Λ2 to gain a crude measure of the

sensitivity of the bag wave function.

Page 39: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

39

The MIT bag model radiative width calculation is then determined as,

ΓJ f Ji = 2k 12Ji+1 ∑Mi,M f ∑λ=±1

∣∣∣⟨J f M f∣∣∫ d3r~ε ∗

λ(~k) ·~j(~r) e−i~k·~r |JiMi〉

∣∣∣2 , (2.16)

~j(~r) = e f ψ†(~r)~αψ(~r),~α =

0 ~σ

~σ 0

, (2.17)

where Ji, Mi are the initial state total angular momentum and the z component of the

angular momentum and J f , M f are for the final state. The ~σ are just the Pauli spin

matrices while the quark associated with the radiative transition is the Dirac spinor ψ(~r).

The momentum vector of the photon is represented by~k ≡ kz. The inital and final mass of

the baryon is mi and m f respectively. The flavor-dependent quark charge is e f . The photon

polarization vector is~ε ∗λ(~k). Numerical calculations for the radiative widths for U-spin

allowed transitions are done as well as the widths of the lower-lying neutral hyperons.

2.1.5 The Soliton Model

The bound state soliton model is used to obtain total widths and E2/M1 ratios

corresponding to decuplet-to-octet electromagnetic transitions in the paper by Schat,

Gobbi, and Scoccola [27]. The bound state soliton model uses the effective SU(3) chiral

action with a symmetry breaking term,

Γ =∫

d4x− f 2

π

4Tr(LµLµ)+

132e2 Tr[Lµ ,Lν ]

+ΓWZ +Γsb. (2.18)

Here, ΓWZ is the non-local Wess-Zumino action and Γsb is the symmetry breaking term.

The left current is Lµ = U†∂µU , where U is the chiral field. The Callan-Klebanov ansatz

is implemented and then an expansion up to second order is made in the kaon field. The

Lagrangian density is expressed as the sum of a pure SU(2) Lagrangian (dependent on the

chiral field) and the effective Lagrangian, which contains the interaction of the soliton and

the kaon fields. A solution set can be achieved by minimizing the classical SU(2) energy

Page 40: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

40

for the equation,[− 1

r2ddr

(r2h

ddr

)+m2

k +V Λ,le f f − f ω

2Λ,l−2λωΛ,l

]kΛ,l(r) = 0 (2.19)

where the bound state energy for quantum numbers Λ = L+T and l is ωΛ,l . The quantum

numbers are expressed in a mode of decomposition of the kaon field where L is the

angular momentum operator and T is the isospin operator. The chiral angle produces the

terms h, f ,λ , and V Λ,le f f .

To calculate the radiative widths of the decuplet hyperons for the M1 transition a

multipole expansion of the electromagnetic field is implemented. The partial width is

expressed as,

ΓM1 = 18αq|⟨M3(q)

⟩|2. (2.20)

The matrix element is between the initial state octet hyperon and the final state deculpet

hyperon. The fine structure constant is α = 1/137 and q is the photon momentum. The

M3(q) operator is defined by

M3(q) =12

ε3i j

∫d3r

j1(qr)r

riJemj . (2.21)

Here j1(qr) is from the l = 1 spherical Bessel function and Jemj is simply the spatial

component of the electromagnetic current.

2.1.6 The Skyrme Model

The Skyrme model for the nucleon is based on an extended version of the non-linear

sigma model that unifies the description of mesons and nucleons at low energies. The

framework was developed by treating the nucleon as a collective excitation of a meson

field rather than a single particle state of quarks. From topological analysis the nucleon is

considered in 1+3 dimensions leading to a solution of a non-linear theory of the three

component pion. In the work by Abada, Weigel, and Reinhard [28], the hyperons are

Page 41: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

41

considered as kaons bound in the background of the static soliton field. Strange degrees of

freedom are treated as SU(3) collective excitations of the non-strange soliton. This

collective approach starts with the non-linear representation of the pseudoscalar nonet.

The Skyrme model contains the Lagrangian component of the non-linear σ model as well

as the flavor symmetric fourth-order stabilizing term,

LS = Tr(− f 2

π

4αµα

µ +1

32e2 [αµ ,αν ][αµ ,αν ])

. (2.22)

Here the αµ = U†∂U for the field U = exp(iΦ) of the pseudoscalar nonet Φ. The physical

pion decay constant fπ = 93 MeV and f 2π = f 2

π +8β ′, whre β ′ =−26.4 MeV2. The SU(3)

symmetry breaking component to the Lagrangian is,

LSB = Tr(T + xS)[β′(Uαµα

µ +αµαµU†)+δ

′(U +U†−2)], (2.23)

where T projects onto the non-strange degrees of freedom and similarly S projects onto

strange degrees of freedom. The masses of the pion and kaon along with their decay

constants determine the needed parameters. To take into account the axial anomaly, the

Wess-Zumino term is used and written as,

ΓWZ =− iNc

240π2

∫d5xε

µνρσκTr(αµαναρασ ακ). (2.24)

To include the electromagnetic properties of baryons at finite momentum transfer a direct

derivative coupling to the photon field, Aµ , is used,

L9 = iL9(∂µAν −∂νAµ)Tr(

ξ†[

λ3 +1√3

λ8

]ξ [ξ †

αµ

ανξ +ξ α

µα

νξ

†])

. (2.25)

Here the Gell-Mann matrices are used with the square root of the chiral field such that

ξ = U1/2. The L9 is a dimensionless coefficient with value of 6.9×10−3. The total action

can then be expressed as,

Γ =∫

d4xLS +LSB +L9 +ΓWZ. (2.26)

Page 42: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

42

To obtain information about the radiative decays of the 32+

baryons, the quadrupole and

monopole pieces of the electric and magnetic from factors are required. E(q) denotes the

electric quadrapole operator that excludes contributions with a total angular momentum of

zero. The magnetic monopole operator is denoted as M(q) for momentum q. The decay

width for the radiative transition from 32+

baryons to 12+

baryons can be calculated from

the matrix elements of E and M,

ΓE2 =6758

αh f q∣∣∣∣⟨B(

12

+)∣∣∣∣ E(q)

∣∣∣∣B′(32

+)⟩∣∣∣∣2 , (2.27)

ΓM1 = 18αh f q∣∣∣∣⟨B(

12

+)∣∣∣∣M(q)

∣∣∣∣B′(32

+)⟩∣∣∣∣2 , (2.28)

where q is the momentum of the photon in the rest frame of the 32+

baryon, and αh f is the

fine structure constant. The total decay widths in the Skyrme model calculation are found

to be strongly dominated by the M1 contribution leading to E2/M1 ratios of a few percent.

2.1.7 The Algebraic Model

The algebraic model of hadron structure, presented by Bijker, Iachello, and Leviatan

[30], introduces a spectrum generating algebra for the radial excitations. The algebra for

mesons is taken to be U(4), and for baryons, U(7). The method unifies the harmonic

oscillator quark model, U(4)⊃ U(3) for mesons and U(7)⊃ U(6) for baryons using a

formalism that treats the orbital excitations as rotations and vibrations of string structures.

The mass spectrum is presented along with the strong and electromagnetic decay widths.

This is done using a dynamical symmetry framework where the spin-flavor symmetry is

broken by the masses.

The Algebraic Model builds baryons using three constituent parts which have

internal and spatial degrees of freedom. The internal degrees of freedom are chosen to be

flavor-triplet, spin-doublet, and color triplet. The motion degrees of freedom of the three

Page 43: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

43

quarks are given by the relative Jaccobi,

ρ =1√2(r1− r2), (2.29)

λ =1√

m21 +m2

2 +(m1 +m2)2(m1r1 +m2r2− (m1 +m2)r3), (2.30)

where mi and r are the mass and coordinates of the ith quark. As in the constituent quark

model, the electromagnetic interaction comes from the coupling of the constituent parts to

the electromagnetic field. The nonrelativistic portion of the transverse electromagnetic

coupling for the left-handed photon emission from B→ B′+ γ can be described as,

HEM = 2√

π

k0

3

∑j=1

µ je j

[ks je−ik·rj +

12g j

(p je−ik·rj + e−ik·rj p j)]

(2.31)

for the e−ik·rj wave with rj, pj, and sj for the coordinate, momentum, and spin for the jth

constituent, respectively. Also, k0 is the photon energy and k = kz is the momentum

carried by the emitted photon and g j is the strength coefficient. The transverse helicity

amplitude for the excited state of the baryon resonance is

Aν =∫

dβg(β )⟨Ψ′(0,S′,S′,ν−1)

∣∣HEM |Ψ(L,S,J,ν)〉 (2.32)

where ν gives the helicity. L,S,and J are the orbital angular momentum, the spin, and the

total angular momentum respectively. The g(β ) is the charge and magnetization

distribution. The radiative hyperon decay widths are calculated using the expression,

Γ(B→ B′+ γ) = 2πρ1

(2π)32

2J +1 ∑ν>0|Aν(q)|2 . (2.33)

The electromagnetic decay widths are calculated assuming SUs f (6) spin-flavor symmetry

while using the rest frame of the excited state baryon.

Page 44: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

44

2.2 U-Spin Symmetry

It is possible to show that in SU(3) flavor space, the quark content of the non-strange

sector can be found by combining the (d,u) I-spin doublets. In the same way, other states

can be found such that the two (s,d) quarks are used for the U-spin doublets, and the (u,s)

quarks are used for the V-spin doublets. States of a given multiplet are transformed into

each other by means of the X-spin operators. Within an I-spin multiplet, the mass

differences are of the order of magnitude of 1 MeV, or about 1% of the particle mass.

Hence, isospin symmetry is weakly broken. I-spin respects chiral symmetry but is broken

by the electromagnetic interaction. The photon couples to the charge (and magnetic

moment) of the quarks, which are different for u and d quarks. U-spin is expected to be

strongly broken by chiral symmetry, however it respects charge symmetry. The other

subgroup, V-spin, is broken by both chiral and charge symmetry.

Baryons can be represented in symmetric spin multiples of I-spin versus hypercharge,

as shown Figure 2.1. Particles of nearly-equal mass are put in horizontal rows. The same

representation for U-spin versus charge can be shown as in Figure 2.2. In this case,

particles of equal charge form horizontal rows. This representation of the baryon decuplet

can be useful when considering electromagnetic transitions, such as radiative decays.

SU(3) f can be written in terms of its subgroups as SU(2)I × U(1)Y or equivalently as

SU(2)U × U(1)Q.

To look closer at how U-spin symmetry is useful, first consider two examples of

U-spin symmetry applied to baryon decays. It is possible to show that U-spin conservation

forbids some baryon radiative decays, and can in principle suggest an experimental test of

U-spin symmetry. A calculation of the corrections, based on the known mass difference of

the s and d quarks, gives a prediction of the degree of U-spin symmetry breaking expected

in the present measurements, as shown next.

Page 45: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

45

Figure 2.1: Baryon decuplet, hyper-charge (Y ) versus isospin (I3).

Figure 2.2: Baryon decuplet, plottedfor U-spin multiplets with charge (Q)versus u-spin (U3).

2.2.1 Examples

To test for U-spin invariance, consider the example of comparing the decays of

∆−→ nπ− and Σ∗−→ Λπ−. Both the ∆− (U3 =−3/2) and the Σ− (U3 =−1/2) are

members of the U = 3/2 multiplet [37]. The mesons can also be arranged in U-spin

multiplets, with the π− having U = 1/2 and U3 =−1/2. The neutron has definite U-spin,

with U = 1 and U3 =−1. However the Λ mixes with the Σ0, so each one has both U = 1

and U = 0 components [38]. Using the U-spin unitary rotation in SU(2) leads to

|ΛU〉=√

32|U = 1〉+ 1

2|U = 0〉 .

As with I-spin, the usefulness of U-spin comes from its ability to predict transition

probabilities using just Clecsh-Gorden (CG) coefficients. A decuplet decay from an initial

state of U = 3/2 to a final state with one particle having U = 1/2, requires a ∆U = 1

transition. For the decay of the ∆−, the U-spin (CG) is unity. The amplitude is denoted as

M (∆−→ nπ−)≡M1 .

Page 46: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

46

The Σ∗− decay has a CG of 〈32 −

12 |1 0 1

2 −12〉=

√3/2 where only the U = 1 component

of the Λ contributes. Under U-spin symmetry the amplitude is,

M (Σ∗−→ Λπ−) =

1√2M1.

Taking into account the phase space factors, which are proportional to

|p3|(E ′+M′)/(MRM′2), for a P-wave (J = 3/2→ J = 1/2) decay it is possible to

compare the decay widths. The p in the phase space factor is the center-of-mass

momentum of each decay particle, E ′ is the energy of the decay baryon, and MR the mass

of the initial baryon resonance. U-spin symmetry gives,

Γ(∆−→ nπ−)Γ(Σ∗−→ Λπ−)

= 1.769× 2|M1|2

|M1|2= 3.54 .

The experimental results lead to a ratio of,

Γ(∆−→ nπ−)Γ(Σ∗−→ Λπ−)

=118 MeV

0.87×39.4 MeV= 3.44±0.20

from Ref. [60], where 0.87±0.015 is the branching ratio and 39.4±2.0 MeV is the full

width of the Σ∗. Here U-spin is very much in agreement with the physical result.

Note that U-spin is not necessarily a symmetry of the strong interaction, since there is

a big difference in the masses of the d and s quarks, so this good agreement with

experiment is somewhat surprising. However, this is only one case, and more tests of

U-spin symmetry (cases which are not already constrained by I-spin, charge or other

symmetry laws) are needed before any conclusion can be reached.

2.2.2 U-Spin prediction

It is possible using the U-spin SU(3) multiplet representation to obtain a prediction

for the ratio of the ∆0→ nγ partial width to the Σ∗0→ Λγ partial width. This implies that

the experimental partial width of the ∆0→ nγ reaction can be used with the U-spin

Clebsh-Gordon coefficients along with the corresponding phase space factors to obtain the

Page 47: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

47

expected partial width of the Σ∗0→ Λγ . In the strict limit of SU(3) symmetry, U-spin is

conserved for all processes. Only radiative transition between states with the same value

of U-spin can occur within this limit.

In U-spin space, the unitary rotation in the SU(2) symmetry space can be used to

describe the mixing between the neutral ground state SU(3) multiplet members Λ and Σ0

such that, ∣∣Σ0U⟩

=12|U = 1〉+

√3

2|U = 0〉

|ΛU〉 =12|U = 0〉−

√3

2|U = 1〉 (2.34)

The radiative decays B(32+)→ B′(1

2+)γ are M1 and E2 electromagnetic transitions with

the spin flip of one of the quarks in B. The E2 amplitudes are very small and are

negligible. The amplitude requires the Clebsch-Gordon coefficients, which can be found

by contraction of the initial excited state baryon with the final state baryon and the emitted

photon, where the photon is a U-spin scalar with U = 0, resulting in⟨∆

0|nγ⟩

= 〈1−1|1−1 0 0〉= 1⟨Σ∗0|Λγ

⟩= −

√3

2〈1 0|1 0 0 0〉=−

√3

2. (2.35)

The ratio strictly based on this rotation is then

|〈∆0|nγ〉|2

|〈Σ∗0|Λγ〉|2=

43. (2.36)

The phase space factors required are presented in Ref. [17] and use a radiative width

Γγ ∝M′BMB

q2|AB→B′γ(q)|2 (2.37)

where q is the center of mass momentum, A is the amplitude of the decay (which contains

a factor of√

q), and MB (MB′) is the mass of the decaying (final state) hyperon. The ratio

can then be expressed as

〈∆0|nγ〉2

〈Σ∗0|Λγ〉2=(

Mn

M∆

)(MΛ

MΣ∗0

)−1( qn

)3 43

= 1.56. (2.38)

Page 48: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

48

The values for the center of mass momentum are qn = 0.259 GeV/c and qΛ = 0.241

GeV/c, [60].

This implies that the U-spin prediction for the partial width of the electromagnetic

decay using the measured width of the ∆0→ nγ is,

1.56−1×Γ(∆0→ nγ) = 1.56−1×660±60 = 423±38 keV,

where the partial width from the ∆→ nγ comes from reference [60].

Similarly a U-spin prediction can be calculated for the ∆+→ pγ partial width to the

Σ∗+→ Σ+γ partial width.

⟨∆

+|pγ⟩

=⟨

12− 1

2|12− 1

20 0⟩

= 1

⟨Σ∗+|Σ+

γ⟩

=⟨

12

+12|12− 1

20 0⟩

= 1,

leading to a ratio of,

Γ(∆+→ pγ)Γ(Σ∗+→ Σ+γ)

=(

Mp

M∆

)(MΣ+

MΣ∗+

)−1( qp

qΣ+

)3

= 2.96.

The values for the center of mass momentum are qp = 0.259 GeV/c and q+Σ

= 0.173

GeV/c, [60].

This implies that the U-spin prediction for the partial width of the electromagnetic

decay using the width of the ∆+→ pγ decay is,

2.96−1×Γ(∆+→ pγ) = 2.96−1×660±60 = 223±20 keV. (2.39)

It is important to note that the U-spin predictions contain no direct meson cloud

calculation but instead implements the empirical information from the ∆ electromagnetic

decays. It is assumed for the sake of these predictions that all meson cloud contributions

are contained in this empirical value. It is not clear yet if the meson cloud contributions

are the same for both the Σ∗0 and Σ∗+ decays. More theoretical calculations are necessary

before this issue can be settled.

Page 49: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

49

2.3 Meson Cloud effect

In the work by Sato and Lee [39], an effective hamiltonian of the γN→ ∆ vertex

interaction is developed by applying a unitary transformation to the Lagrangian with

various field components N,∆,π,ρ,ω and γ . It is found that the helicity amplitudes

calculated from the dressed γN→ ∆ vertex are in better agreement to the empirical values

(listed in the Particle Data group) than the bare amplitude. The differences in the bare and

dressed amplitude are due to the nonresonant meson exchange mechanisms refered to as

“meson cloud” effects. The Sato and Lee (SL) model attempts to have a consistent

description of both the πN scattering and the electromagnetic pion production reactions.

The SL model expresses the γN→ πN amplitude as

TπN,γN(E) = tπN,γN(E)+ tRπN,γN(E). (2.40)

The nonresonant component of the amplitude is

tπN,γN(E) = [tπN,γN(E)GπN(E)+1]νπN,γN . (2.41)

where the nonresonant πN scattering amplitude is

tπN,πN(E) = νπN,πN [1+GπN(E)tπN,πN(E)]. (2.42)

Here νπN,πN is the πN potential and the GπN(E) is the πN propagator. The amplitude in

terms of the dressed vertex interactions is then

tRπN,γN(E) =

Γ†∆,πNΓ∆,γN

E−m∆−Σ∆(E), (2.43)

where m∆ is the ∆ mass and Σ∆(E) is the energy shift. The dressed components of the

vertex interaction, described in terms of the bare vertex interaction, are

Γ†∆,πN = [tπN,γN(E)GπN(E)+1]Γ†

∆,πN (2.44)

Γ∆,γN = Γ†∆,γN +Γ∆,πNGπN(E)tπN,γN . (2.45)

Page 50: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

50

Using Eq. 2.41 and 2.42, the expression in Eq. 2.45 can be rewritten as

Γ∆,γN = Γ†∆,γN +δΓ∆,γN , (2.46)

where

δΓ∆,γN = Γ∆,πNGπN(E)νπN,γN (2.47)

and

Γ∆,πN = Γ∆,πN [1+GπN(E)tπN,πN ] . (2.48)

Figure 2.3: The diagrams for the dressed γN → ∆ vertex. The meson cloud diagrams arein brackets.

The πN potential contains the direct and crossed nucleon terms, ρ exchange, and the

crossed ∆ term. Each interaction vertex is regularized with a dipole form factor. The

parameters of the model are adjusted along with m∆ to fit the empirical πN scattering

phase shift. The dressed γN→ ∆ term contains the meson loops illustrated in Figure 2.3.

To analyze the N∆ form factors, the model fits all of the available pion

electroproduction data at energies close to the ∆ position. The results, using a simplified

Page 51: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

51

parameterization, are used to obtain the dressed M1 form factors G∗(Q2). The magnetic

dipole transition form factor for γ∗∆→ ∆(1232) are normalized to the proton dipole form

factor and compared to experimental data, shown in Figure 2.4. The data is from DESY

and SLAC [42], BATES [43], MAMI [44], and JLAB [45]. The solid curve indicates the

dressed calculation, while the dotted line is without the meson cloud effect.

The results in Ref. [41] reveal that the meson cloud effect can contribute significantly

(∼ 40%) to the overall electromagnetic decay width of the ∆→ Nγ . There are no

calculations yet available for the Σ∗→ Y γ decay amplitude. Assuming the meson cloud

effects are important, more sophisticated calculations are necessary to probe the structure

of the baryon resonances. Few models actually include meson cloud contributions for

electromagnetic decay.

Page 52: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

52

Figure 2.4: The magnetic dipole transition form factor G∗(Q2) for γ∗∆→ ∆(1232). Theexperimental points (empty circles) are for the inclusive data from pre-1990 experimentsat DESY and SLAC [42], and exclusive data (filled squares) are from BATES [43], MAMI[44], and JLAB [45]. The Solid curve indicates the dressed calculation, while the dottedline is without the meson cloud effect.

Page 53: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

53

3 JEFFERSON LAB, CEBAF AND CLAS DETECTOR

The g11a data were extracted from an experiment conducted in Hall B at Thomas

Jefferson National Accelerator Facility (TJNAF), in Newport News, Virginia. The g11a

data set was collected in 2004 as part of the E04021 experiment Spectroscopy of Excited

Baryons with CLAS: Search for Ground and First Excited States.The initial purpose for

this run was a high-statistics search for the Θ+ pentaquark [54]. A loose trigger was used,

allowing for many other photoproduction studies.

Figure 3.1: The aerial view of CEBAF at Thomas Jefferson National Labs. The“racetrack”-shaped area indicates where the accelerator ring lies underground. Eachexperimental hall is underneath the three grassy sectors near the bottom of the picture.Hall B is the middle hall. Image source:[35].

Page 54: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

54

The run conditions for g11a included a tagged photon beam incident on a liquid

Hydrogen target. The CLAS (CEBAF Large Acceptance Spectrometer) detector was used

in the acquisition of multi-particle final states at approximately 60% coverage of the full

4π solid angle. The data used in the following analysis was produced with tagged

bremsstrahlung from a electron beam at 4.023 GeV. The CLAS tagger hodoscope can

measure photons between the energies of 20% and 95% of the electron beam energy. The

proton target was came from liquid Hydrogen maintained by a cryogenic system. The

g11a run resulted in roughly 20 billion triggers stored as 21 TB of raw data.

In the follow chapter details of the experimental apparatus and setup used in data

collection for the g11a experiment are covered. An overview of the Continuous Electron

Beam Accelerator Facility (CEBAF) is seen in Figure 3.1.

3.1 Continuous Electron Beam Accelerator Facility

CEBAF is a 6 GeV end-point energy electron beam accelerator. The facility delivers

electron beam to the experimental halls A,B, and C at Thomas Jefferson Lab. CEBAF

uses radio-frequency (RF) cavities for electron acceleration. The superconductors

employed in the cavities provides three times more power than without the

superconductors in the RF design. Greater efficiency using the superconductors is due to

zero energy lost by electrons in the RF cavity. In addition, the cavity temperature is not

increased by transmission of electrons. This non-resistive superconducting nature of the

cavities allows CEBAF to attain a 100% duty factor being that there in no down-time

required to cool the conduction elements of the accelerator [36]. The continuous delivery

of electrons allows rapid acquisition of high statistics datasets. A schematic diagram of

CEBAF can be seen in Figure 3.2.

Injectors are used to input the electron beam. The injector consists of a polarized

photo-emission electron gun. The three diode lasers of the electron gun, one for each

Page 55: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

55

Figure 3.2: A schematic diagram of the Continuous Electron Beam Accelerator Facility.The linear red tubes represent the LINACs made of superconducting RF cavities groupedinto 20 cryomodules. A magnification of the recirculation arcs is shown in the top right.The experimental halls are shown on the bottom left. Image source:[35].

experimental hall, allows each hall to independently control its current and beam

polarization. The lasers are pulsed independently at 499 MHz, incident on a strained

GaAs photocathode. The lasers are pulsed 120 out of phase, matching the 1497 MHz

frequency of the accelerator. Each experimental hall receives electron bunches at intervals

of 2 ns. After extraction from the photo-cathode the electrons are accelerated to 45 MeV

with the 2 1/4 superconducting Radio frequency (RF) cavities. An optical chopper is then

used in the injector system to cleanly separate the bunches prior to sending them to the

CEBAF’s recirculating linear accelerators (LINACs).

Page 56: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

56

Figure 3.3: A picture of a pair of superconducting niobium Radio Frequency (RF) cavities.CEBAF uses 338 uperconduction cavities, like the RF cavity shown. Image source:[36].

Figure 3.4: A schematic diagram of the superconducting RF cavity in operation. Theacceleration gradient is provided by establishing a standing wave, leading to a continuouspositive electric force on the electron. The phase of the waves advances the position of theelectron bunch in the cavity creating the gradient. Image source:[35].

Each recirculating LINAC contains 168 superconducting RF Niobium cavities.

Figure 3.3 is a picture of the RF cavity assembly. The superconducting cavity temperature

is cooled to 2 K with liquid Helium. Radio frequency standing waves in the cavities are

used to produce the acceleration gradient for the electron beam, see Figure 3.4. A

Page 57: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

57

continuous positive force is made on the beam electrons in the cavity by the use of

standing waves kept in phase with the beam bunches. There are two LINACs located along

the straight portion of the 7/8-mile recirculating accelerator path. Recirculation occurs by

bending magnets in the curved portions of the tracks. Electrons can pass through the pair

of LINACs up to five times before being delivered to an experimental hall. Each pass

through a LINAC adds up to 600 MeV to the beam energy. The maximum energy

presently attainable by five passes is approximately 6 GeV. The beam is extracted by the

experimental halls by using RF separator cavities. Each hall can control the beam energy

delivered by extracting the beam after a given number of passes (no greater than five).

Before delivery, the beams are fanned out according to energy at the end of the south

LINAC. To divert specific electron beam bunches to the desired hall, a 120 phase

separation is used by the RF separator cavities. This style of separation allows

experiments that require different beam to run at the same time. CEBAF is capable of

delivering a five-pass beam to each hall simultaneously, but it cannot provide a single low

energy beam to two halls at the same time.

Page 58: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

58

3.2 The Bremsstrahlung Photon Tagger

For this and other photoproduction experiments at CLAS, the CEBAF electron beam

is used to produce bremsstrahlung photons from the beam incident on a target. Photons

are created by passing electrons through a thin radiator, creating bremsstrahlung radiation.

The bremsstrahlung cross section is proportional to Z(Z +1), with Z being the atomic

number of the radiator. Because of the spectral nature of the photon energy produced, a

tagger system is required to measure the recoiling beam electron energy. The energy of

the bremsstrahlung photon is then deduced through conservation of energy. Because the

produced number of bremsstrahlung photons is nγ ∝ 1/Eγ , the production is dominated by

low energy photons.

The high mass and density of the gold radiator is ideal for the reaction

ebeam +Au→ e′beam +Au′+ γbrem. For high momentum incident electrons, the production

photon and the recoil electron will continue in the same relative direction, in the lab

frame, sharing the energy of the incident electron. The beam line then passes through a

dipole magnet which separates out the electrons from the photons. Electrons that have not

lost energy and have not reacted with the radiator are disposed of in a beam dump. The

reacted electrons experience a greater change in direction through the field. Electrons with

energies between 20% and 95% of the original beam energy are passed into the

hodoscope. Two layers of scintillation paddles to determine the electrons energy and

timing. The bremsstrahlung photons, γbrem, associated with the tagged electrons pass

through a collimator, and continue down the beam line to interact with the target. Beam

buckets are delivered every 2 ns with a width of ∼0.02 ns. A schematic diagram of the

photon tagging system is shown in Figure 3.5.

Page 59: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

59

Figure 3.5: A schematic of the photon tagging system. The tagger is setup to allow indirectmeasurements of the photon beam energy. The recoil electron is directed into the taggerspectrometer to that is energy can be measured to deduce the photon energy produced.Image source: [56].

3.2.1 The Radiator

The radiator for the tagger consists of several different foils and a “harp” made of

two perpendicular wires. These are mounted on a “ladder” that can move various radiators

into the beam line. For the g11a data-taking runs, the thickest of the radiators used was a

gold foil with a thickness of 10−4 radiation lengths (646 µg/cm2). A thinner foil (10−5

radiation lengths) was used for the normalization run.

The “harp” was used to measure the electron beam profile and position at the

radiator. During the scan, two perpendicular wires pass through the beam, while

scintillators downstream are used measure the intensity of the beam at the wire position.

Page 60: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

60

3.2.2 The Magnetic Spectrometer

The tagging system used a magnetic spectrometer to measure the energy of the

recoiling electron after the interaction in the radiator. A 1.75 T maximum field,

normal-conducting dipole magnet is used to direct both recoiling and non-interacting

electrons. The dipole magnet was designed to enable electron beams up to 6.1 GeV to be

used to produce real photons and still bend the non-radiating electrons into the beam

dump. The energy resolution, for operating below 4 GeV, is ∼ 0.2%.

3.2.3 The Hodoscope

The hodoscope is the scintillation device used to measure the energy of the electrons

that radiated, to obtain the photon energy, to determine the timing of the event with

sufficient precision, and to act as part of the trigger. The hodoscope is made with two rows

of scintillating paddles, the E-counters and T-counters for “Energy” and “Timing”. The

top plane of 384 scintillators that are used to determine the momentum of the recoiling

electron, called the E-plane. The E-plane scintillators are 20 cm long, 4 mm think and

between 8 cm - 16 cm wide with a photomultiplier tube (PMT) on one end. These

E-counters are designed to cover areas that represent approximately equal sized energy

bins. The placement of each E-counter improves resolution through the use of an

overlapping formation, increasing the effective number of E-counters to 767, giving to an

energy resolution of ∼ 10−3×Ebeam. The lower scintillator plane determines precise

timing information of the recoiling electrons, called the T-plane. The T-plane sits 20 cm

below the E-plane and consists of 61 scintillators with a PMT on each end. The T-counters

are 2 cm thick, leading to a timing resolution of 110 ps. Both the E-counters and

T-counters are arranged so that the scintillators are normal to the electron trajectory as

they pass through the focal plane. A schematic diagram of the tagger magnet and

hodoscope is provided in Figure 3.6.

Page 61: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

61

Figure 3.6: A schematic of the tagger magnet and hodoscope used in the tagging system.The trajectories of the recoil electrons are depicted by the dashed lines. Electrons fromvarious trajectories in the spectrometer correspond to bremsstrahlung photons of a givenenergy. Image source: [56].

3.2.4 The Tagger Readout

The signals from each E-counter PMT are passed to a discriminator. The PMTs from

each end of the T-counters are fed into a constant fraction discriminator (CFD). The

signals that pass the CFD are sent to a Master Or (MOR) and then the time-to-digital

converter (TDC) array. The TDC is stored in raw data bank preserving precise timing

information for each T-plane hit as well as the total number of hits recorded in the tagger.

The total number of recoil electron hits is needed to derive the energy-dependent photon

flux. The electron timing information is used to find the photon interaction time for each

event recorded. The MOR is used to set the trigger and controls the stop signal to the

E-counter TDC array. The E-counter readout in the trigger is written to the data stream

along with the T-counter readout that set the trigger. A schematic of the Hall B tagger

Page 62: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

62

logic setup is shown in Figure 3.7. More information on the Photon Tagger can be found

in [56].

Figure 3.7: A schematic of the tagger logic setup. The T-counter hits are also used to setthe event trigger. The common stop to the E-counter TDC array is controlled by the theCLAS Level 1 trigger. Image source: [56].

Page 63: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

63

3.3 The CLAS Detector

The CEBAF Large Acceptance Spectrometer (CLAS) Detector is comprised of

various detector subsystems. For the present analysis the subsystems required are the start

counter, the drift chambers, time-of-flight counters, the toroidal magnet, and the

electromagnetic calorimeter. The gas Cerenkov detector is primarily used in

electroproduction experiments and is not discussed here. A photograph and schematic are

shown in Figures 3.8 and 3.9 respectively.

Figure 3.8: A photograph of the CLAS detector from inside Hall B. The Time-of-flightscintillators are pulled away from the drift chambers showing inside. Image source: [35].

Page 64: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

64

Figure 3.9: A schematic diagram of the full CLAS detector. The drift chambers are shownin violet, the toroidal magnet is shown in light blue, the Time-of-flight scintillators areshown in red, and the electromagnetic calorimeter is shown in green. Image source: [35].

3.3.1 The g11a Cryotarget

To study reactions of the type γ p→ K+Y , a target of high density atomic protons is

required. Liquid hydrogen was used as the target material in the cell. Many different

target geometries and materials are used during various production runs. The cryotarget

cell geometry used during the g11a run period was a cylindrical Kapton chamber with

dimensions of 40 cm in length and 4 cm in diameter [57]. The density of the liquid

hydrogen is determined using the temperature and pressure inside the cell which were

Page 65: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

65

monitored on an hourly basis. The target density averaged over all g11a runs is 0.07177

g/cm3. A diagram of the g11a cryotarget is shown in Figure 3.10.

Figure 3.10: A diagram of the g11a Cryotarget. The dimensions of the target cell usedwere 40 cm long and 4 cm in diameter. The cell was filled with liquid H2. Image source:[35].

3.3.2 The Start Counter

A new start counter was installed for use in the g11a run period. The start counter

was designed to achieve full acceptance coverage using the 40 cm long cryotarget. The

start counter is a segmented scintillation detector surrounding the cryotarget. The start

counter is segmented into six sectors corresponding to each sector of CLAS. Each sector

contains four scintillator strips instrumented with PMTs. The start counter timing

resolution is roughly 400 ps. The number of scintillator paddles was chosen based on the

estimated integrated rate load at the anticipated luminosity. The start counter timing

information is not directly used in the present analysis. However, the start counter is an

integral part of the g11a trigger. Other details on the CLAS start counter can be found in

[58]. A diagram of the start counter used in the g11a run period is shown in figure 3.11.

Page 66: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

66

Figure 3.11: A diagram of the new g11a Start Counter. The detector is a six-sectorscintillation device with four mounted photomultiplier tubes for each sector. One sectoris cut away in the diagram to show the target space. Image source: [58].

3.3.3 The Superconducting Toroidal Magnet

CLAS is assembled around six superconducting magnetic coils separated in the

azimuthal angle by 60 around the beam line. The geometry of many of the CLAS

detector elements are determined by the toroidal magnet. The six superconducting coils

required to generate the field create shadow regions in which no particles can be detected,

making a natural segmentation for the other assemblies. The electronics are mounted in

the shadow regions of the detector created by the torus cryostats. The beam line passes

through the center of the coil configuration. A photograph of the toroidal magnet is shown

in Figure 3.12. The toroidal magnet has a maximum operating current of 3861 A, leading

to a maximum magnetic field of 3.5 T. The value of∫

~Bd~l varies from 2 Tm for high

momentum tracks in the forward angles to 0.5 Tm for trajectories where θ > 90. During

Page 67: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

67

Figure 3.12: A photograph of the coils of the CLAS toroidal magnet before installation ofthe rest of the detector subsystem. Image source: [59].

the g11a run period the current was set to 1920 A, leading to a maximum field of ∼ 1.8 T

in the anti-clockwise direction about the beam line when looking upstream. The magnetic

coils generate a toroidal magnetic field ~B, of six-fold cylindrical symmetry around the

beam line. The charged particles in CLAS pass through this field and bend in θ , toward

(away) from the beam line if the particle is negatively (positively) charged in the default

magnet configuration. The radius of curvature of a charged particles can be expressed as

r = pprep/qB, where pprep is the component of the particle’s momentum perpendicular to

the magnetic field. The contour and cross section of the CLAS toroid magnetic field is

shown in Figure 3.13. The superconducting magnet has a cryostat temperature regulation

system using liquid helium delivered from the central CEBAF helium refrigerator (CHL).

Page 68: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

68

(a) (b)

Figure 3.13: (a) The contours of constant absolute magnetic field of the CLAS toroid in themidplane between two of the coils. (b)The field vectors for the CLAS toroid transverse tothe beam. The field lines represent the field strength. The six coils are shown in the crosssection. Image source: [52].

The coils are regulated to the operate at a temperature of 4.5K. Other details on the

CLAS superconduction toroidal magnet can be found in [59].

3.3.4 The Drift Chambers

The drift chambers are the core of the charged particle tracking system. Effective

charged particle tracking can be achieved by measuring the location of a particle at several

points along its trajectory while minimizing its multiple scattering. The CLAS drift

chambers enable position measurements of charged particles with a precision of a few

hundred microns, while limiting the interference of background radiation. Momentum is

determined by tracking the particles as they travel through the field generated by the

toroidal magnet. Particles are tracked with three drift chamber regions. Region 1 is

mounted to the magnet’s cryostats and lies between the start counter and the innermost

part of the toroidal magnet coils. Region two sits in the middle of CLAS in the area of the

Page 69: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

69

Figure 3.14: A diagram look down showing the CLAS drift chamber region relative to theother subsystems. The dashed lines outline the location of the toroidal magnet coils. Imagesource: [59].

strongest magnetic field. Region 3 sit closest to the Time-of-flight paddles outside the

magnet coils, see Figure 3.14.

Drift chambers should have a one-to-one relationship between the time measured and

the distance of closest approach (DOCA) of the track to the wire. This implies that the

electric field should be cylindrically symmetric about the high-voltage sense wire. A

quasi-hexagonal geometry made of six field wires are used to create the needed field

around each of the 35,148 sense wires.

Each drift chamber region has six sectors, each which span roughly 60 of the plane

perpendicular to the beamline. Each region consists of two sublayers or “super-layers.”

The super-layers contain an array of drift cells that have six 140 µm gold-plated field

Page 70: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

70

wires around the 20 µm gold-plated tungsten sense wire. The super-layers which are

oriented such that the sense wires are perpendicular to the mid-plane of each sector are

called the “axial” layers. The super-layers which are oriented 6 to the axial wires, to

improve resolution in the φ -direction, are known as the “stereo” layers. There is also a set

of “guard” wires that sit on the edges of each super-layer that hold a low voltage to

simulate a continuous electric field configuration in each cell. This allows the field in each

cell to by roughly independent of the cell’s position in the chamber.

The drift chambers are filled with a 90% Argon and 10% CO2 gas mixture. A

potential difference between the sense wires and the field wires in maintained by

operating the field wires at a negative high voltage. During detection, charged tracks

ionize the gas in the cell around the positively-charged sense wires. As a particle passes

through the cell and ionizes the gas electrons along the particle’s path, these electrons

begin to accelerate towards the sense wire. Electron collisions with the other molecules in

the gas lead to the a fairly constant “drift” velocity. The collection of these electrons

registers as a voltage pulse in the sense wire. The signals from the sense wires pass

through a preamplifier and discriminator before being recorded in the data stream. More

details on the CLAS drift chambers can be found in [53]

3.3.5 The Time-of-Flight Detector

The time-of-flight (TOF) subsystem is an essential component of the CLAS detector.

The TOF detector provides the required timing information for charged tracks that can be

used to develop the particle identification schemes. The typical path length that particle

will travel from the target to the TOF paddles is ∼ 5 m. The TOF shell is a six-fold

segmented array of scintillator strips that covers the outside shadow area of the torus coils

in each sector from 8 to 142. Each sector of the TOF shell is composited of 57

Page 71: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

71

scintillator paddles made of Bicron BC-408 scintillation material. Each paddle is 5.08 cm

in thick and has a PMT instrumented at either end.

There are four scintillator panels in each sector, as seen in Figure 3.15 (a). In the first

panel there are 23 paddles that detect particles in the range of 0−45. Each of the 23

paddles is 15 cm wide. The backward angle paddles are 22 cm wide and are

photo-coupled to 3 PMTs using light guides, as seen in Figure 3.15 (b). The length of the

paddles varies from 32 cm to 445 cm as required by the shape of the sectors. The eighteen

least forward paddles are coupled in pairs. This leads to a total of 48 logical counters. The

timing resolution of the TOF detector is between 80 and 160 ps, in which the spread is

defined by the variation in scintillator length.

The TOF subsystem is a critical part of the g11a trigger. Details concerning the

design, construction, and testing of the TOF detector are discussed in [51].

(a) (b)

Figure 3.15: (a) An isolated view of the time-of-flight paddles in one sector. The setdesignated for forward angle detection are on the right. (b) The schematic of the lightguide used to connect the backward angle TOF paddles to the PMTs.

Page 72: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

72

3.3.6 The Forward Electromagnetic Calorimeter

The detection of neutral particle like photons and neutrons is made possible during

the g11a run period by used the the Electromagnetic Calorimeter (EC). The EC sits about

5 m from the target on the outside of the TOF system in CLAS. There are six nearly

equilateral triangle EC volumes, one for each CLAS sector. Each EC volume is made of a

sandwich of 39 alternating layers of plastic Bicron BC412 scintillator strips that have

dimension of 1 cm × 10 cm that run between 0.15 to 4.2 m in length, and lead sheets 2.2

mm thick. The calorimeter area of each successive layer increases, minimizing the shower

leakage at the EC edges.

Each scintillator layer consists of 36 scintillator strips that run parallel to one side of

the triangle. Each successive layer is oriented 120 relative to the previous layer,

following the orientation of the triangle. This structure results in three stereo views

labeled as the U-plane, V-plane and the W-plane, see Figure 3.17. Each stereo view is

subdivided into an inner (5 layers) and outer (8 layers). The inner set of layers sit closest

to the target. There are 216 PMTs required for each stereo view, leading to a total of 1296

PMTs and 8424 scintillator strips in the whole CLAS EC.

The signal from the scintillator is transmitted to the PMTs through green wavelength

shifting fibers, see Figure 3.16 (a). All scintillators from one vertical line, see Figure 3.17

(a), are connected to a single PMT so that a shower created by a particle passing through

has three such lines (U,V,W) associated with its cluster reconstruction position. The

reconstruction of a typical electromagnetic shower is determined by identifying groups of

strips in the three stereo views. The reconstruction algorithm collects groups that have a

PMT response above an energy deposition threshold for contiguous strips. The algorithm

sorts the groups according to the sum of the scintillator strip energy. An electromagnetic

shower reconstruction event is shown if Figure 3.17 (b).

Page 73: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

73

Figure 3.16: A vertical slice of the EC light readout system. PMT - Photomultiplier TubeLG - Light Guide, FOBIN - Fiber Optic Bundle Inner, FOBOU - Fiber Optic Bundle Outer,SC - Scintillators, Pb - 2.2 mm Lead sheets, IP - Inner Plate (closest to target or face of EC)Image source: [55].

Page 74: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

74

(a)

(b)

Figure 3.17: (a) View of one of the six CLAS electromagnetic calorimeter triangularmodules showing the three projection planes. (b) The diagram of event reconstructionin the EC. Energy deposition profile is shown along each stereo view. Image source: [55].

Page 75: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

75

3.4 The Beamline Devices

There are additional instruments used in the beamline for diagnostic studies. The

Beam Position Monitors (BPM) system uses RF-cavity beam-position monitors located

36.0 and 24.6 m from the target. The beam profile was measured with the Hard Scanners

which use two orthogonal tungsten and iron wires that pass through the beam. The

electrons scatter off of the harp wires and are detected by the PMT arrays upstream. The

harp scanners are located at 36.7, 22.1 and 15.5 m upstream from the target. The Total

Absorption Shower Counter (TASC) sits at the end of the beamline and is used for

measuring real photon flux. The TASC is a lead glass scintillator array, with close to 100%

photon detection efficiency. This device is only used during the g11a “normalization”

runs, when beam current was reduced. The Pair Spectrometer (PS) is a diagnostic device

used to measure the rate of production of e+e− pairs. It consists of a dipole magnet and a

thin aluminum radiator located within the magnet and an array of eight scintillator

paddles. When photons hit the radiator e+e− pairs are produced. The magnetic field

deflects the pair in opposite directions as they are directed into the scintillators. The

photon can then be reconstructed from the left-right coincidence from the scintillators.

The Pair Counter (PC) is used to study the photon beam near the beam dump. The PC is

made of a four-scintillator array used to detect e+e− pairs that are created when the beam

photons strike a thin aluminum foil. Figure 3.18 shows a schematic of the general layout

of these devices. More information about these devices can be found in Ref. [56].

3.5 The g11a Trigger and Data Acquisition

Each component of CLAS is setup to actively run and monitor the signals created in

the subsystem. Naturally, not all signals are worth recording for analysis. The trigger is

designed to determine which sets of signals pertain to the physics of interest and to turn on

Page 76: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

76

Figure 3.18: A schematic layout of the beamline and flux monitoring devices. The beamlineenters from the left. Image source: [56].

and off data recording. Once triggered, the data acquisition system (DAQ) writes them to

magnetic tape to be analyzed.

The trigger for the g11a run period was setup to optimize the data collection for

studies on the existence of the pentaquark state. During the g11a data collection, events

are recorded when both the tagger Master OR (MOR) and the CLAS Level 1 trigger fall

within a coincidence window of 15 ns. The full tagger focal place was kept functional and

recorded data. However, only the first 40 of the total 61 T-counters were included in the

MOR. This effectively keeps event that are below the desired center-of-mass range from

being triggered by taking low energy photons out of the trigger. Still some low energy

photon events were recorded when a high energy photon accidentally fell in the same

timing window. The CLAS Level 1 trigger required a coincidence between any of the four

start counter paddles and the TOF paddles, from two separate sectors, in a window of 150

ns. This requirement leads to data collection of events that all have at least two charged

tracks. The tagger, the start counter, and the TOF paddles all have multiple detection

Page 77: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

77

elements so the logic requires a pre-trigger OR of the discriminated signal in each system

to generate one signal from each control system that could be used in the trigger module

coincidence. Prior to the pre-trigger, the signals in each detector system are split to go to

the analog-to-digital (ADC) and the time-to-digital-converter (TDC) boards. During a

trigger the ADC and TDC from all detector systems are read into the data stream and the

data banks are assembled into an event and recored.

During the reconstruction of g11a, the cooking scripts used a prlink of

prlink g11 1920.bos and the hybrid magnetic field map bgrid T67to33.fpk resulting in

∼67% of the “new” or cold magnetic field map and 33% of the “old” or warm magnetic

field map. Approximately 20 billion event triggers were recorded by the data acquisition

system.

3.6 Summary

A general description has been given of the Continuous Electron Beam Accelerator

Facility and the CEBAF Large Acceptance Spectrometer at the Thomas Jefferson National

Accelerator Facility. The subsystems specific to this analysis are covered along with some

minor discussion of how they are used in each case. Though initially intended for other

purposes, the g11a dataset collected by the CLAS detector in Hall B can be used to study

other types of hadron photoproduction. In the next several chapters, the details of the

extraction of the Σ∗0, and Σ∗+ radiative decay branching ratios are given.

Page 78: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

78

Part I

Electromagnetic decay of the Σ∗0

Page 79: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

79

The details of the present study to extract the electromagnetic (EM) decay to the Σ∗0

are now presented. As a check on previous results, an attempt is made to reproduce the

original CLAS analysis of Taylor [10]. A rigorous approach is undertaken to understand

the method of kinematic fitting and how best to apply this method to the general

investigation of hadronic electromagnetic decays. From this, an analysis method for the

more challenging topology γ p→ K0Σ∗+ is proposed.

A detailed study of the Monte Carlo used for the Σ∗0 electromagnetic decay is also

presented. This provides guidance on the nature of the corrections and the systematic

effects of the EM decay ratio in the CLAS data.

Some critical details on the overall systematic uncertainties for the physics and

method of extraction of the EM decay ratio are presented in the next chapters. These

studies lead to a relatively good procedure to optimize the statistical and systematic

uncertainties in the final results.

Page 80: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

80

4 EVENT SELECTION

Using the g11a data set, it is necessary to select events of the channel γ p→ K+Σ∗0.

The Σ∗0 resonance decays 87.0±1.5% of the time into Λπ0 and about 1% into the radiative

decay Λγ [60]. Roughly 63.9±0.5% of the time the Λ decays to pπ−, leading to the final

states γ p→ K+pπ−π0 and γ p→ K+pπ−γ , respectively [60]. The charged particles can

easily be detected with use of the CLAS drift chambers and Time-of-flight scintillators,

whereas the π0 and γ must be deduced using conservation of energy and momentum. The

analysis was done using the previously skimmed g11a data set for two positively charged

tracks and one negatively charged track with no other skim conditions. For checking and

testing corrections, the particle identification and event selection of Section 4.6 is used.

4.1 Run Inclusion

The CLAS runs for g11a include 43490 to 44133, of which only 43490 to 44107

were taken at beam energy 4.019 GeV. The smaller portion at beam energy 5.021 GeV

would need to be analyzed separately, and was excluded. The commissioning runs 43490

to 43525 were excluded as well. Runs 43675, 43676, 43777, 43778, and 44013 were

taken to test different trigger configurations and were also not included. Runs 43981 and

43982 had drift chamber issues, while runs 43989-43991 had data acquisition problems.

These runs were not included to minimize systematic differences in the sequence of runs

analyzed.

4.2 Energy Loss Corrections

Energy loss for charged particles as they pass through various materials in the CLAS

detector occurs, and a systematic adjustment to the particle’s energy is necessary. Energy

losses occur in the target material, target walls, the scattering chamber, the start counter,

and the various air gaps between the detector elements. These corrections are dealt with

Page 81: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

81

by using the eloss software package, which can be set for the target parameters and start

counter used in g11 [67].

Figure 4.1: ∆E with respect to momentum for left: proton, and right: K+

In the three-track topology for the K+Λ final state, the detached vertex of the Λ is

reconstructed using tracking information from the p and π−. The Λ decay vertex is the

starting point for the charged tracks to which an energy loss correction would be applied.

The energy loss package was tested, to ensure it was making reasonable corrections,

by studying the energy of the π−, K+, and proton before the correction and then after. The

difference ∆E = Ea f ter−Ebe f ore for the proton and K+ with respect to particle momentum

is shown in Figure 4.1. The program is made to work for all charged particles with charge

equal to 1 (except for electrons). The valid momentum range is 0.05 < p/m < 50. It was

observed that for high momentum π+ the energy loss is quite small, and for low

momentum the correction reaches up to ∼50 MeV in accordance with dE/dx ∼ 1/β 2. For

heavier particles, such as the proton, the effects are more easily demonstrated as seen in

Figure 4.1. These plots were made using the particle id given in the PART bank (described

in Section 4.6, also see Appendix B for BOS structure) leading to some irregular bands

and edges for a small number of counts. The vast majority of the counts lead to the

expected trend. This indicates that the energy loss correction is working correctly.

Page 82: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

82

4.3 Tagger Corrections

The tagged-photon energy comes from the E-counters of the CLAS photon tagger.

There are known inaccuracies in the tagger energy found in the focal plane. This effect

was originally studied using inclusive (γ)p→ pπ+π− events with kinematic fitting. The

three charged particles are detected and kinematically fit to the missing photon. The

events that pass a 10% confidence level cut were used to find the difference between the

measured and the kinematically fit photon energy.

∆Eγ = Ek f itγ −Emeas

γ

The corrections are binned for each tagger E-counter, and from each bin a Gaussian mean

is extracted. For each run the beam offset Br was calculated and used in the correction.

For an event from run number r with photon from E-counter e, the correction applied to

the photon energy becomes,

∆Eγ,e,r = ∆Ek f itγ,e +Br,

with ∆Eγ,e being the Gaussian mean from E-counter e. This correction has been

previously derived for g11a [64], with additional studies on the need for the correction due

to the sagging of the focal plane between its four support yokes provided in [56]. The sag

shifts the narrow E-counters from the set locations, which can lead to detecting electrons

with slightly distorted energies. To confirm the results of these previous studies, the same

topology (γ)p→ pπ+π− was used to kinematically fit the missing photon, giving the

same trend in ∆E as in the original work [64]. The results from this check are shown in

Figure 4.2, which shows the relative tagger correction needed for the set of E-counters

found by fitting the ∆Eγ = Ek f itγ −Emeas

γ distribution with a Gaussian for each E-counter

bin. The Gaussian mean over the full range is used to apply a systematic correction for

each E-counter. There are several points seen in Figure 4.2 (Left) that fall off the general

Page 83: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

83

trend. These are due to swapped cables [64]. A correction can still be applied to these

E-counter so that no E-counters are excluded. There are also no T-counters excluded.

Figure 4.2: Left: ∆Eγ/Ebeam vs. E-counter for the reaction (γ)p→ pπ+π− used to find thetagger correction. Right: The extracted Gaussian mean from the ∆Eγ/Ebeam vs. E-counterfits.

4.4 Momentum Corrections

The actual momentum of charged particles can differ from the corresponding CLAS

tracking information. These differences can be due to faulty drift chambers that may be

misaligned or distorted, as well as inaccuracies in the magnetic field map [56].

The g11a momentum correction for all runs has been developed using the test

channel γ p→ pπ+π−. The pπ+π− are detected and the energy loss and tagger

corrections are applied. Three kinematic fits done, each having a different final-state

particle as the “missing” (or undetected) particle from energy and momentum

conservation. The hypotheses for the corresponding fits are γ p→ pπ+(π−),

γ p→ pπ−(π+), γ p→ π+π−(p). The measured magnitude of the momentum and the

directional components were then compared with the “missing” momentum vector from

the kinematic fit such that ∆p = pk f it− pmeas, ∆λ = λ k f it−λ meas, and

∆φ = φ k f it−φ meas. These distributions were binned in the relevant kinematic variables

Page 84: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

84

and fit with Gaussians. Here λ and φ are the directional tracking parameters. An

expression for each correction needed is found as a function of momentum and orientation

(θ ,φ) for each particle species. Corrections were found for each CLAS sector, with each

sector broken up into twelve 5 degree bins in the azimuthal angle φ , and then each φ bin

into fifteen polar angle θ bins. The momentum correction is tabulated after all kinematic

regions have been considered [64].

Figure 4.3: Left: ∆p+ found for the K+ in the topology γ p→ K+Λ→ K+pπ− for Sector1, θ ∈ (20,25) and φ ∈ (−15,−10). Right: The Gaussian fit for ∆p+.

To check the corrections, the topology γ p→ K+Λ→ K+pπ− is used, with the tagger

and energy loss corrections previously implemented. Because the momentum correction is

small, the bin for Sector 1, θ in (20,25), and φ in (−15,−10) is shown in Figure 4.3,

which is known to have a visible correction. This result compares well with the work

previously done [64]. It is found that the magnitude and trend of the correction are the

same for all positively charged particles, and differ from all negatively charged particles

but mostly just by a sign change, such that ∆p+ ≈−∆p−.

4.5 Effectiveness of Corrections

Some studies were performed to check the effectiveness of the corrections using the

channel γ p→ K+Λ from the final state decay products K+pπ−. The variables are the

invariant mass of the pπ− and the missing mass off the K+. The PDG value of the Λ mass

Page 85: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

85

is ∼ 1.1156 GeV. The particle identification outlined in the following section is employed

to isolate the decay products with no other kinematic cuts applied. Only the detected

particles that fall within |∆t|< 2 ns are considered, where ∆t is the difference between the

time the particle struck the start counter and the time at which the photon was at the

interaction vertex for that particle. The energy loss correction, the tagger correction, and

the momentum correction are applied and the spectrum is studied before and after these

corrections. A comparison for each distribution is seen in Figure 4.4. A Gaussian fit to the

invariant mass of pπ− before the corrections gives a mean of 1.115 GeV with a σ of 1.4

MeV, which is then improved to a mean of 1.116 GeV and a σ of 1.35 MeV, as seen in

Figure 4.5. The mass off the K+ in the region 1.1 to 1.21 GeV is used to find the Λ

candidates and also fit with a Gaussian. The mean of the fit improved from 1.110 GeV to

1.116 GeV as seen in Figure 4.6.

Figure 4.4: Left: Invariant mass of (pπ−), Right: Missing mass off the K+. Beforecorrections is shown in black and after is shown in red.

4.6 Particle Identification

In the present analysis, the PART identification scheme was used as an initial starting

point. The PART id uses the start counter to find an interaction vertex time for each

Page 86: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

86

Figure 4.5: Left: Invariant mass of (pπ−) with Gaussian fit before corrections, Right:Invariant mass of (pπ−) with Gaussian fit after corrections.

Figure 4.6: Left: Missing mass off the K+ with Gaussian fit before corrections, Right:Missing mass off the K+ with Gaussian fit after corrections.

charged particle and matches it up with photons in the tagger, where there are up to 10

photons for a given event. The photon with the closest time to any track is selected as the

photon that caused the event. Specifically, the time of interaction is acquired using the

electron beam bucket (RF time) that produced the event. To correlate the interaction time

with the photon production time, a coincidence of the tagger T-counter with the Start

Counter is used. The RF time for the photon is then used to get the vertex time (photon

interaction time) for the event. Using the time of flight from the event vertex to the

scintillator counter, the velocity β is calculated for each particle. From β and the

particle’s measured momentum, a mass is calculated. Each track does not need to have a

Page 87: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

87

hit registered in the start counter for its mass to be calculated, only one track in the event

needs a Start Counter hit. The PART id is made during the “cooking” of the data.

The restrictions for PART from the mass time of flight for kaons is

0.35≤MK+ ≤ 0.65 GeV, for pions is 0.0≤Mπ− ≤ 0.3 GeV, and for protons is

0.8≤Mp ≤ 1.2 GeV. The mass calculated from time of flight is

mcal =

√p2(1−β 2)

β 2c2 , (4.1)

where

β = L/ctmeas. (4.2)

From this initial identification it is possible to incorporate additional timing information to

improve event selection with quality constraints. The measured time-of-flight and

calculated time-of-flight can be used for an additional constraint. The measured

time-of-flight is tmeas = tsc− tγ , where tsc is the time at which the particle strikes the

time-of-flight scintillator wall and tγ is the time at which the photon was at the interaction

vertex. ∆t is then,

∆t = tmeas− tcal, (4.3)

where tcal is the time-of-flight calculated for an assumed mass such that

tcal =Lc

√1+(

mp

)2

, (4.4)

where L is the path length from the target to the scintillator, c is the speed of light, m is the

assumed mass for the particle of interest and p is the momentum magnitude. Cutting on

∆t or mcal should be effectively equivalent.

Using ∆t for each particle it is possible to reject events that are not associated with

the correct RF beam bucket. This is done by accepting only |∆t| ≤2 ns in the initial

analysis. This cut is loose enough to minimize signal loss, while providing a large enough

range in ∆t to study systematic variations down to |∆t| ≤1 ns. Figure 4.7 shows ∆t for the

Page 88: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

88

protons versus the ∆t for the K+. The main cluster of events at (0,0) comes from the

events that best satisfy the mass hypothesis for that particle. In all of the plots shown, the

bad TOF paddles have already been removed, see Section 4.9. The clusters of events at

(±2 ns,±2 ns),(±4 ns,±4 ns), etc. indicate the photon was from the wrong beam bucket.

The dashed lines represent the time cut used in the initial look at the data. For the final

results reported, a |∆t| ≤1 ns cut is used.

Figure 4.7: ∆t giving the PID timing cut. Clusters of events at (±2 ns,±2 ns),(±4 ns,±4ns), etc. are from photons from a different beam bucket. The dashed lines represent thetiming cut used.

The kaon mass distribution directly acquired from the PART bank is seen in Figure

4.8. The pion background from other beam bunches, seen as diagonal bands, is reduced

after the timing cut. It is also possible to see this reduction in accidentals in the β vs. p in

Page 89: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

89

Figure 4.6 (left), which shows the β distribution before any cuts. Figure 4.6 (right) shows

the same distribution after the timing cut implemented in Figure 4.7.

Figure 4.8: Calculated mass based on the PART bank information with no cuts used.Visible accidentals result from other RF buckets that are present.

There are also some counts seen at low momentum in the kaon band that are reduced

by the restriction on the lower bound of the kaon momentum, along with a ∆β cut

described next. There is also a pion and proton low-momentum restriction implemented.

A ∆β cut is used to clean up the identification scheme. ∆β is the difference between

time-of-flight β = L/(ctmeas), and the calculated β = p/√

p2 +m2, where L is the particle

path length and tmeas = tsc− tγ is the difference in the time at which the particle strikes the

time-of-flight scintillator counter and the time at which the photon was at the interaction

vertex. p is the particle momentum and m is the known PDG particle mass. The ∆β

Page 90: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

90

Figure 4.9: Left: K+ β distribution before any cuts, Right:K+ β distribution after timingcut.

distribution for the kaon is shown in Figure 4.10 before any cuts or photon time restriction.

Figure 4.11 shows the ∆β for the K+ after a ±1 ns timing cut, showing a much cleaner

distribution. The good events were taken within a cut of −0.02≤ ∆β ≤ 0.02 for the kaon.

For the sake of demonstrating how well a ±1 ns cut can clean up the particle

selection, the calculated mass without any ∆β cut is shown for the kaon, pion, and proton

in Figures 4.12, 4.13, and 4.14. For the kaon, Figure 4.12 can be compared to Figure 4.8

that has no timing cut yet implemented.

Page 91: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

91

Figure 4.10: ∆β for the K+ before any cuts.

Figure 4.11: ∆β for the K+ after a ±1 ns timing cut.

Page 92: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

92

Figure 4.12: K+ calculated mass versus momentum after a ±1 ns timing cut.

Figure 4.13: π− calculated mass versus momentum after a ±1 ns timing cut.

Page 93: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

93

Figure 4.14: Proton calculated mass versus momentum after a ±1 ns timing cut.

Page 94: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

94

4.7 Vertex Information

The reaction of interest γ p→ K+Σ∗0→ K+Λ(X) has two detached vertices: one

from the Σ∗0 and one from the Λ. Using tracking information from the TBER bank (see

Appendix B) it is possible to find a distance of closest approach (DOCA) from the p track

and π− track to obtain a vertex and cut at a reasonable distance. The cut can help to

achieve a good Λ. An initial cut of 5 cm is used as a starting point for the DOCA cut as is

seen in Figure 4.17. Figure 4.15 shows the p,π− invariant mass before and after the

DOCA cut, showing a reduction in background and improvement in the peak width. This

loose DOCA cut is used only to reduce the background before the kinematic fitting

procedure. A study of the effects of this cut on the final result is performed later in Section

9.

Figure 4.15: Left: Invariant mass of p,π− before DOCA cut. Right: Invariant mass ofp,π− after DOCA cut.

It is expected that of the final state decay products, only the kaons are produced at the

primary vertex. To reduce the likelihood of considering accidentals, a cut is applied to the

kaon vertex so that only kaons produced in the target volume are considered. The kaon

vertex z-coordinate distribution is shown in Figure 4.16. The kaon vertex was also

obtained using the kaon track TBER vertex information and the kaon track with the

Page 95: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

95

Figure 4.16: Kaon vertex distribution. The dashed lines are the implemented cuts.

closest approach to the beam line. The dashed lines in Figure 4.16 indicate the cuts at -30

cm and 10 cm on the z-coordinate of the vertex position.

Figure 4.17: Distance of closest approach; the dashed line is the implemented cut.

Page 96: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

96

4.8 Beam Photon Selection

The status of the tagger hits was obtained directly from the TARG bank (see

Appendix B) by checking for good events with tagger bank status 7 or 15. For bank status

7 and 15, tagger hits with no more than one E-counter hit is associated with a T counter

hit. These tagger events were accepted unless there were two type 15 at once, which were

eliminated. The resulting good photons were kept until the end of the analysis. If there

was more than one photon tagged with the same event number that survived all cuts, the

event would also be eliminated. The number of events with more than one photon tagged

after all cuts is negligible at less than 0.001%.

4.9 Detector Performance Cuts

Cuts were applied to take into account both the regions where there are known

obstructions to the acceptance and regions of CLAS that are not well simulated. This

includes tracks at extremely forward or backward angles, areas near the torus coils, and

regions where the drift chambers and scintillator counter efficiencies were not well

understood. Some minimal fiducial cuts are applied to account for some angular regions

that are shadowed by the toroidal coils and the Region 2 end-plates. As θ decreases, the

shadowed region of detection increases as seen from the center of the target. Tracks that

point near these shadow regions are less likely to reflect an accurate reconstruction. These

regions can be eliminated by identifying the tracks affected and applying a general fiducial

volume cut. Figure 4.18 shows the angular distribution for the kaon before the fiducial

cuts and Figure 4.19 shows the same distribution after the fiducial cuts.

The azimuthal angle φ is within the range of ±30 relative to the sector mid-plane.

The restriction cutting out the edge of the acceptance was developed using the engineering

drawings for CLAS showing the width of the shadow region due to the coils as a function

of θ found after initial installation. There is an azimuthal cut of |φ |< 26 implemented

Page 97: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

97

with the θ restriction limiting the fiducial volume. The functional form of the cut used to

define the fiducial volume is

θ > 4.0+510.58

(30−φ)1.5518 . (4.5)

A minimum momentum of 0.125 GeV, after the energy loss correction, was

implemented for both positive and negative particles. For cleaner momentum

reconstruction, tracks with a minimum proton momentum of 0.4 GeV were required as

well. These restrictions were studied previously for the g11 run period [63].

Some dead or unreliable Time-of-flight (TOF) scintillators were removed. These

TOF’s were originally identified by the g11 run group by examining occupancy plots of

both the data and Monte Carlo. All TOF’s with noticeable discrepancies for pions or

protons were flagged as unreliable [63]. Table 4.1 shows the list of bad TOF’s in each

sector. Particles hitting these TOF’s are rejected in the analysis.

Figure 4.18: Angular distribution of the kaons before fiducial cuts.

Page 98: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

98

Figure 4.19: Angular distribution of the kaons after fiducial cuts.

Table 4.1: Bad Time-of-flight scintillators.

Sector BAD TOF’s

1 33

2 8,17,27

3 18,19

4 8,13,14

5 23

6 30

Several runs in the g11a data set have been identified in which the forward part of the

TOF array in sectors 2 and 3 have been determined to be either not operating or running at

Page 99: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

99

very low voltage. Rather than setting additional cuts to the analysis, these runs were

simply eliminated. A list of these runs is given in Table 4.2.

Table 4.2: The list of run numbers that have problems with the forward part of the TOF.

Sector 2 Sector 3

43989 43586

43990 43587

43991 43588

44000 43589

44001 43590

44002 43591

44007 43592

44008 43593

44010 43594

44011 43595

44012 43596

Page 100: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

100

5 ANALYSIS PROCEDURE

The approach used in this analysis is to remove as much identifiable background as

possible while preserving the π0 and radiative signals. Because of the closeness of the

radiative signal to the π0 peak in the mass spectrum from Σ∗0→ Λπ0 decay, the radiative

signal extraction requires a certain degree of finesse with a kinematic fitting procedure.

To clean up the kaon for the analysis, there is a cut made on identified kaon

candidates that are truly π+. By reestablishing the energy of the PART identified kaon

with the mass of the pion, one can test for possible contamination. The missing mass

squared is studied for the reaction γ p→ pπ+π−(X). A spike at zero mass squared

indicates the reaction γ p→ pπ+π−. The missing mass squared (see Section 1.7) is shown

for the reaction γ p→ pπ+π−(X) in Figure 5.1. This particle identification contamination

can be removed by cutting slightly above zero but not over the region where the π0 peak

should be. A cut at 0.01 GeV2 is chosen so as to not cut into the good K+ events.

Reactions such as ρ → π+π− are also eliminated by this cut.

It is also possible to do a similar study using γ p→ π+π−(X). A peak below zero

would indicate extra pion contamination that could be cut out, as well as anything around

the mass of the proton. The γ p→ π+π−(X) spectrum after the cut on γ p→ pπ+π− is

seen in Figure 5.2. With no peak at the proton and nothing to speak of below zero, a clean

kaon is achieved from the cut in Figure 5.1.

Page 101: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

101

Figure 5.1: Pion contamination: Mass squared (M2X ) of any missing particle for the

γ p→ pπ+π−(X) reaction where the π+ was a potentially mis-identified kaon. Eventswith M2

x < 0.01 GeV2 were removed.

Figure 5.2: Pion contamination: Mass squared (M2X ) of any missing particle for the

γ p→ π+π−(X) reaction, where the π+ was a potentially mis-identified kaon.

Page 102: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

102

For the hyperon excited-state region, the Λ(1116) can be seen from the invariant

mass of the p-π− pair in Figure 5.3. The instrumental resolution (σ ) of the peak is about

1.3 MeV, see Figure 5.3. The p-π− pairs with invariant masses between ±0.005 GeV of

the PDG Λ mass are kept for analysis. This is intended to be a loose cut at around ±3σ to

keep as many candidates as possible that will be further cleaned up with the kinematic

fitting procedure discussed later. For this same reason it is not necessary to be concerned

about the background the peak sits on.

Figure 5.3: Λ peak from the invariant mass of the π− and proton.

The various spectra are studied before and after each cut. Before any kinematic cuts

are included, the plots in Figure 5.4 show the invariant mass of the p-π−, the missing mass

spectrum for all detected particles, the mass off of the p-π−, and the missing mass off the

Page 103: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

103

K+. The next set of plots shown in Figure 5.5 include the cut on the invariant mass of the

p-π− (shown in Figure 5.3) at ±0.005 GeV around the Λ. There is a plot added that shows

the range of missing energy of the detected particles. There is clearly additional structure

around 0 and 0.08 GeV that is removed by the restriction on the excited state hyperon

mass discussed next. There is also a plot of the transverse component of the missing

momentum. Pxy close to zero indicates that the missing particle is traveling nearly straight

down the beam line.

After the Λ cut, the missing mass off the K+ is investigated. The Σ∗(1381) and

Λ(1520) are clearly seen in Figure 5.6. The Λ(1405), which decays almost 100% to Σπ ,

would show up as a shoulder on the right of the Σ∗ peak, however no evidence is seen in

the spectrum. The plot in Figure 5.6 shows a fit of a relativistic Breit-Wigner to the Σ∗0

with a quadratic background. The mass and width from the fit can be seen from the fit

parameters. The mass from the fit without any constraints on the parameters is 1.386 GeV.

The width from the fit is 39.8 MeV. These values compares well with the Particle Data

Group value centroid and width of 1.3837 and 36 ±5 MeV respectively. The cut on the

missing mass of the K+ is also shown in the plot.

To study the excited hyperon region of interest before the Λ cut, an additional set of

plots was made without the ±0.005 GeV cut around the Λ, but instead with a cut from

1.34 GeV to 1.43 GeV on the missing mass of the K+. This cut selects the potential Σ∗

candidates. Figure 5.7 shows the same plots but only with the hyperon mass restriction.

Figure 5.8 show these distributions after both kinematic cuts are implemented. In the

missing mass off the p-π−, the visible peak at the mass of the K∗(892) implies that there

will be a presence of a background that can be studied using Monte Carlo simulations.

The immediate issue is the presence of a peak at the mass of the K+; this is due to leakage

of ground-state hyperon production into the excited-state region. To eliminate this

background the requirement of MX > 0.55 GeV is set.

Page 104: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

104

Figure 5.4: No kinematic cuts applied. Top left: Invariant mass of the p-π−. Top right:Missing mass squared of all detected particles. Bottom left: Missing mass off the p-π−.Bottom right: Missing mass off the K+. All units are (GeV).

From Figure 5.8 (upper right plot) one can see the missing mass squared of the

reaction γ p→ K+Λ(X) with all the afore mentioned cuts applied. A peak is visible for the

M2π0 as well as at zero missing mass squared. The counts above the π0 peak are mostly

Page 105: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

105

Figure 5.5: Plots after Λ cut only. Top left: Missing energy of all particles detected. Topright: Missing mass squared of all detected particles. Bottom left: Missing mass off the Λ.Bottom right: Transverse missing momentum. All units are (GeV).

due to γ p→ K+Σ0(X), which can also be taken into consideration with the Monte Carlo.

As seen in the progressive cuts, the peak at zero missing mass squared is greatly

diminished; this is primarily due to the elimination of the γ p→ K+Λ channel. In principle

one can now use the distribution of the missing energy and the perpendicular component

of the missing momentum to check against what is expected. In the missing energy all of

Page 106: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

106

Figure 5.6: Missing mass off the K+ after the cut on the Λ. The fit to the Σ∗ uses arelativistic Breit-Wigner and quadratic background. The dashed line indicates the cutaround the Σ∗0. The Λ(1520) is also clearly visible.

the other structures have more or less disappeared, leaving a peak at the missing energy

between 0.2 and 0.3 GeV. The counts at lower energy are from the K∗Λ and Λ(1405),

which will also be managed though the simulation’s acceptance of these channels, see

Figure 7.18. The missing energy is not used to separate the radiative signal from the π0,

but it can be used qualitatively to indicate that other background channels have been

reduced since the expected missing energy for the π0 channel is ∼0.2 GeV as seen from

the simulations in Figure 5.9. Clearly, the remaining peak in Figure 5.8 (upper left)

primarily comes from the missing energy from the π0.

The missing transverse momentum is defined as P2xy = P2

x +P2y , where Px and Py are

the x and y components of the missing momentum. One would expect a relatively rounded

Page 107: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

107

Figure 5.7: Plots after Σ∗ cut only. Top left: Missing energy of all particles detected. Topright: Missing mass squared of all detected particles. Bottom left: Missing mass off the Λ.Bottom right: Transverse missing momentum. All units are (GeV).

and featureless distribution for Pxy with no sharp peaks for either the radiative or π0

channels, such as the distributions seen in Figure 5.11 from the simulations of these

channels. The small peak close to zero in the Pxy distribution in Figure 5.8 (bottom right)

is reduced once only one tagged photon is required, but as seen in Figure 5.10, there is

still a spike in the distribution close to zero that can be seen in the P2xy distribution that is

Page 108: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

108

Figure 5.8: Plots after Σ∗ and Λ restrictions. Top left: Missing energy of all particlesdetected. Top right: Missing mass squared of all detected particles. Bottom Left: Missingmass off the Λ (counts below the dashed line are cut). Bottom right: Transverse missingmomentum. All units are (GeV).

important to handle. This peak is likely caused by double bremsstrahlung in the radiator

such that γ1 + γ2 p→ K+Λ+ γ1. The ground-state Λ production events for which a photon

readout from the tagger can contain an extra photon in the event. These extra photons will

be directed straight down the beam line and show up as a spike at P2xy = 0. This may also

Page 109: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

109

Figure 5.9: Missing energy produced from simulations for the reaction γ p→ K+Λπ0.

occur if the event is an accidental or the wrong electron is selected due to inefficiencies in

the tagger plane. A final cut is then used to eliminate this contamination. Based on the

plot seen in Figure 5.10, candidates with P2xy > 0.0009 GeV2 were preserved for further

analysis. It is important to remove 100% of this forward peak because any leakage could

show up as a radiative signal. The cut is clearly seen to be far enough from the peak that

essentially no tail can pass through. Simulations were used to make sure that the

γ p→ K+Λγ channel was not significantly reduced. The simulations shown in Figure 5.11

were used to estimate the amount of radiative signal lost by assuming equal counts in both

the π0 and radiative channel and finding the percent of signal lost with a P2xy > 0.0009

GeV2 cut. About 1% of the simulated signal events lost by this cut. Figure 5.11 shows the

transverse missing momentum distribution for the hypothetical radiative decay and the π0.

Figure 5.12 shows a magnification of the area of interest and the cut implemented. Neither

channel is greatly affected by this cut, whereas the γ1 + γ2 p→ K+Λ+ γ1 contribution is

Page 110: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

110

eliminated. Systematic variation of this cut is studied in Section 9. Full details of the

Monte Carlo are provided in Section 7.

Figure 5.10: Left: Transverse missing momentum Pxy after only one tagged photon isselected. Right: P2

xy after only one tagged photon is selected with cut implemented, seen asthe dotted line.

The missing mass spectrum can be seen after all required cuts, including the Pxy cut,

in Figure 5.13 (top right). The radiative signal is small and not directly visible. Fitting

with a double Gaussian is not a viable option because the tail of the π0 bleeds into the

space for the radiative signal. The extraction of such a signal prompts a challenge

specifically because the phase space of the γ p→ K+Λγ and the γ p→ K+Λπ0 reactions

are so similar and the radiative decay branching ratio is so small relative to the π0 branch.

A decay of π0→ γγ can easily look like a radiative decay directly from the Σ∗,

overwhelming the channel of interest. Next, a detailed understanding of the uncertainties

in tracking of all the detected particles will be used in order to establish a well-defined

confidence level that the missing particle is either a π0 or a γ .

Page 111: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

111

Figure 5.11: Simulations for the perpendicular momentum Pxy of missing mass candidates.The blue distribution is for the simulated γ p→ K+Λ(π0) reaction and the yellow is for theγ p→ K+Λ(γ) reaction.

Page 112: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

112

Figure 5.12: Magnification of the simulation of the perpendicular momentum Pxy ofmissing mass candidates and the implemented cut showing little effect on signals.

Page 113: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

113

Figure 5.13: Plots after all mentioned cuts. Top left: Missing energy of all particlesdetected. Top right: Missing mass squared of all detected particles. Bottom left: Missingmass off the Λ. Bottom right: Transverse missing momentum. All units are (GeV).

Page 114: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

114

6 KINEMATIC FITTING

A kinematic fitting package was written to improve the resolving tools for working

with CLAS measured variables. Considerable effort had been taken to improve TBER

information for the drift chamber tracking errors for G11 [66]. Additional energy loss as

well as multiple scattering effects were also considered in this package.

Track reconstruction in CLAS is done in a sector-dependent coordinate system. This

same coordinate system is used in developing the kinematic fitting and performing the fits.

During tracking, a covariance matrix is produced, containing the resolution uncertainties

and correlating coefficients of the tracking parameters for each track, which are located in

the BOS TBER bank (see Appendix B). After achieving an accurate covariance matrix,

the process of kinematic fitting can be used to greatly improve an analysis procedure.

The kinematic fitting technique can take advantage of a number of types of

constraints such as energy and momentum conservation, common vertices, or physical

limits to improve the measured quantities that are used in the analysis. The method of

Lagrange multipliers is a common way to handle the constraints with a least squares

criteria. Here some highlights are discussed from the methods outlined in [68], and [69].

Assume there are n independently measured data values y, which in turn are

functions of m unknown variables qi, with m≤ n. The condition that y = fk(qi) is

introduced where fk is a function dependent on the data points that are being tested for

each k independent variable at each point.

Because each yk is a measurement with corresponding standard deviation σk, the

equation yk = fk(qi) cannot be satisfied exactly for m < n. It is possible to require that the

relationship be closely numerically satisfied by defining the χ2 relation such that

χ2 = ∑

k

(yk− fkQ−∑k Qkiηi)2

σ2k

≡∑k

(∆yk−∑i Qkiηi)2

σ2k

. (6.1)

Page 115: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

115

The resulting values ηi = qi−q−Q found through the minimization of the linearized χ2

function are unbiased and have minimum variance σ2qi

. This implies the best possible

parametrization.

The linearized minimization conditions ∂ χ2/∂qi = 0 can then be expressed as

∑k

QkiQk j

σ2k

η j = ∑k

Qki

σ2k

∆yk,

where 1≤ i, j ≤ m, defining the covariance matrix

VQ i j =

(∑k

QkiQk j/σ2k

)−1

,

with solution

ηi = ∑j,k

VQ i jQk j∆yk

σ2k

.

The linearized χ2 equation can now be written in matrix form:

χ2 = (∆y−Qη)T V−1

y (∆y−Qη),

where Q is the matrix of coefficients and V−1y is the inverse of the covariance matrix.

The partial derivative with respect to the η parameters leads to the m equations

QT V−1y (∆y−Qη) = 0.

resulting in

QT V−1y Qη = QT V−1

y ∆y.

Using the relation VQ = (QT V−1y Q)−1, one obtains the m equations

η = VAQT V−1y ∆y.

The parameters q are then determined from q = qQ +η . The matrix V−1y necessarily

has dimensions of n×n, while Q is n×m, and VQ has the size m×m.

Page 116: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

116

To impose constraints a more generalized expression for χ2 is required. First the set

of r constraints are described in the from M(q) = 0. A set of linearized equations are

obtained by expanding around qQ, M(qQ)+(q−qQ)∂M(qQ)/∂q≡ Dη +d = 0. Here D

is a r×m matrix and d is a vector of length r. This matrix form of the constraints are then

adding to the matrix form of the χ2 equation resulting in,

χ2 = (∆y−Qη)T V−1

y (∆y−Qη)+2L T (Dη +d).

The vector L is of length r and is the corresponding set of Lagrange multipliers. Again a

solution for η is found by taking the partial derivatives with respect to η and L and

setting them equal to 0. The ∂/∂L = 0 equation provides all the constraint conditions.

Solving the equations lead to an expression of η , where η is equal to the initial

unconstrained values η0 added to a term proportional to the Lagrange multiplier. The

unmeasured variables are then improved by stretching the values of η from the initial

unconstrained values η0, within the kinematically defined resolutions, to meet the

imposed constraints.

For the purpose of implementation one can now simplify the notation to build an

algorithmic outline. In the case of a linear nature in the constraint equations, it is possible

to use an iterative approach.

The unknowns again are divided into a set of measured (~η) and unmeasured (~u)

variables. The Lagrange multipliers, Li, are introduced to be used in each constraint

equation. These terms can be used to rewrite the equation for χ2,

χ2(~η ,~u, ~L ) = (~η0−~η)TV−1(~η0−~η)+2 ~L T ~F . (6.2)

The χ2 differentiation with respect to all the other variables linearizes the constraint

equations. The new constraint can then be used in multiple iterations. The vector ~η0 is the

content of the initial tracking information for the measured variables. V is the covariance

matrix containing all the kinematic resolutions and correlation of the measured variables.

Page 117: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

117

Naturally the result for each new iteration ν +1 depends on the numerical values found in

the previous iteration (ν):

~uν+1 =~uν − (ATu S−1Au)−1AT

u S−1~r, (6.3)

~L ν+1 = S−1[~r +Au(~uν+1−~uν)], (6.4)

~ην+1 =~ηo−VATη

~L ν+1, (6.5)

with definitions of,

(Aη)i j ≡∂Fi

∂η j, (6.6)

(Au)i j ≡∂Fi

∂u j, (6.7)

~r ≡ ~Fi +(Aη)ν(~η0−~ην), (6.8)

S≡ (Aη)νV (ATη)ν , (6.9)

for the ν th iteration. The change in χ2 from each iteration ∆χ2 = |χ2ν −χ2

ν+1| is used to

find the termination in the iteration procedure such that ∆χ2 > 0.001 for each recursion.

There is also an exit in the loop for any confidence level below 1×10−6 to which these

events are not expected to improve by a measurable value and can be passed. The

iterations were counted and it was found that this convergence occurs within less than 15

iterations.

For clarity it is important to discuss which coordinate system was used for the

constraint equations and applied covariance corrections. There are three main coordinate

systems used for CLAS analysis. One can use the lab system, the sector system and the

tracking system. For analysis the lab system is most useful. In this system the beam line

sits along the zlab-axis; the xlab-axis goes right through sector 1 and the ylab-axis points

vertically up in between sector 2 and 3.

The track reconstruction in CLAS is done in a sector-dependent scheme. The tracking

coordinates in each sector are defined so that the xtrack-axis lies along the beam line

Page 118: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

118

whereas the ztrack-axis is aligned with the average Poynting vector of the magnetic field in

the given sector. Then ytrack simply passes through the center of that sector orthogonal to

the other axes. The tracking coordinates transform into the lab coordinates likextrack

ytrack

ztrack

=

zlab

cos(α)xlab + sin(α)ylab

−sin(α)xlab + cos(α)ylab

(6.10)

and α is sector dependent such that α = π

3 (Nsector−1).

The resulting tracking parameters of interest are given by q/p, λ , φ , d0 and z0. The

parameters d0 and z0 are variables giving the vertex information in the projection of the

“vertex” plane. Momentum in the laboratory frame for each track can be described as

px = p(cosλ sinφ cosα− sinλ sinα), (6.11)

py = p(cosλ sinφ sinα + sinλ cosα), (6.12)

pz = pcosλ cosφ , (6.13)

where α is the same as in Eq 6.10.

The covariance matrix V from tracking found in the TBER bank is limited and does

not contain the needed effects of multiple scattering and the energy loss in the target cell,

the carbon epoxy pipe, or the start counter. Corrections to the diagonal elements of the

covariance matrix V are systematically applied, as well as the correction needed for the

tracking uncertainties found previously for g11 [66].

Using the tracking variables and the geometry of CLAS one can take reasonable

steps in the determination of uncertainties due to multiple scattering. Following the

observationally derived approach employed by Lynch and Dahl [70], the RMS scattering

angle in a material can be described in terms of material density ρ , mean atomic number

Z, mean atomic weight A, speed of traversing particle βc, momentum p, and thickness x.

Page 119: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

119

The expression for the RMS scattering angle is then,

⟨θ

2⟩= 2χ2

c1+F2

[1+ v

vln(1+ v)−1

], (6.14)

χ2c = 0.157

Z(Z +1)A

ρxp2β 2 , (6.15)

χ2a = 2.007×10−5 Z2/3

p2

(1+3.34

Z2α2

β 2

), (6.16)

Ω =χ2

cχ2

a, (6.17)

v =Ω

2(1−F). (6.18)

The α here is the fine structure constant. F is the fraction of the full multiple-scattering

distribution that is estimated for the process of interest. The value of F = 0.99 was chosen

based on the notion that F varies anywhere in the range of 90% to 99.5% for the material

of interest. It is estimated that the formula yields results within 2% of experimental values

[70]. This expression can then be used to approximate the variance of the dip angle λ due

to multiple scattering in the target, scattering chamber, and Start Counter such that,

∆V (λ ,λ ) =⟨θ

2⟩target +

⟨θ

2⟩scattering chamber +

⟨θ

2⟩start counter . (6.19)

This term can then be used to find the correction for the φ diagonal element,

∆V (φ ,φ) =∆Vλλ

61+ sin2

λ

cos2 λ. (6.20)

Taking into consideration the effects of the energy loss when charged particles

traverse these materials, a standard thick absorber particle approximation was used. This

method is also found in Leo [70]. In this method the energy spread is Gaussian with a

variance written as,

σ2E = 4πNAr2

em2ec4

ρxZA

γ2(

1− β 2

2

), (6.21)

where NA is Avogadro’s number, re represents the classical electron radius, and me is the

electron mass. The adjustment to the momentum term in the covariance matrix can then

Page 120: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

120

be described as,

∆V (1/p,1/p) =σ2

Ep4β 2 . (6.22)

The vertex information is extracted after vertex corrections are made for the K+. The

vertex information for the proton and π− are from the secondary vertex and are not

applied until the Λ decay. In general if a channel is to be kinematically fit that contains a

neutral particle that decays, the energy and multiple scattering effects will be based on the

position of the secondary vertex.

In completing construction of the kinematic fitting program and the development of

an accurate covariance matrix, it is necessary to test and check usability. Various

techniques were studied to investigate procedure efficiency and systematic uncertainties.

This work is discussed in reference [71].

After developing and testing the software one is able to produce a flat Confidence

Level distributions after fitting Monte Carlo and data from g11 for reactions such as

γ p→ K+π−p(π0) channel. The pull distributions obtained from fitting are approximated

by a Gaussian with σ ranging from 0.966 to 1.064. Kinematic quantities, such as missing

mass spectra calculated from the fits, are significantly improved, while background is

significantly reduced. These studies are discussed in reference [71].

Page 121: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

121

7 SIMULATIONS

For the Monte Carlo setup, the relative contribution of different reaction channels

were taken into account. The experimental photon energy distribution for the subset of

g11a at 4.0186 GeV was used to determine the energies of the incident photons in the

simulation. The weighting of each channel and its contribution to the variation in the final

result is discussed in Section 9.

7.1 Monte Carlo Generator

The event generator FSGEN (Full Spectrum Generator) was used with a variable

t-dependence such that a channel with a kaon is generated uniformly in the center-of-mass

frame in φ with a t-dependent distribution in θcm according to P(t) ∝ e2.0t where t < 0.

The decay products of excited state hyperons are produced isotropically in the rest frame

of the hyperon and boosted back to the lab frame.

The incident electron energy was set to 4.0186 GeV and the real photon energy range

setting used was 1.0-4.2 GeV. Gaussian distributions in x and y with σ = 0.5 cm were used

to approximate the beam width in the target. A target length of 40 cm, from 30 cm to -10

cm, was used to generate events uniformly along the length of the target on the beam axis.

FSGEN produces the BOS MCTK track bank and the MCVX vertex bank (see

Appendix B) that contain the needed information to be thrown into CLAS. The generated

events were then fed into GSIM, the CLAS Geant-based Monte Carlo program. GSIM

simulates each CLAS detector portion and creates the reconstructed information of the

simulated hits. The hit information is then written out in the same BOS bank format as for

the real data.

Page 122: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

122

The generated events were then fed into GSIM with the settings for the ffread card

for processing as:

AUTO 1

LIST

KINE 1

BEAM 4.016

MAGTYPE 2

MAGSCALE 0.4974 0.0

FIELD 2

GEOM ’ALL’

NOGEOM ’EC1’ ’CC’ ’MINI’ ’PTG’

NOSEC ’OTHE’

TARGET ’g11a’

TGPOS 0.0 0.0 4.06

STZOFF -14.06

STTYPE 1

RUNG 43582

CUTS 5.e-3 5.e-3 5.e-3 5.e-3 5.e-3

DCCUTS 1.e-4 1.e-4 1.e-4 1.e-4 1.e-4

ECCUTS 5.e-4 5.e-4 5.e-4 5.e-4 5.e-4

SCCUTS 1.e-4 1.e-4 1.e-4 1.e-4 1.e-4

STCUTS 5.e-5 5.e-5 5.e-5 5.e-5 5.e-5

FASTCODE

TRIG 500000

STOP

The RUNG flag specifies a known good run with correct parameter configurations.

Page 123: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

123

7.2 Monte Carlo Smearing

The GSIM output files were then fed into GPP. GPP (GSIM post-processor) is

primarily used to smear timing and remove some bad wires using the data base wire map.

The GPP command and flag used is:

gpp -P0x1f -R43582 -Y -f0.50 -a1.0 -b1.0 -c1.0 -o<output> <input>

where the GPP parameters used were:

• f = 0.50 (time smearing)

• a = 1.0 (DOCA smearing Region 1)

• b = 1.0 (DOCA smearing Region 2)

• c = 1.0 (DOCA smearing Region 3)

Figure 7.1: Left: momentum resolution from data used to match Monte Carlo. Right: thematching of Monte Carlo to data by smearing out the measured resolution, red is MonteCarlo and blue is data.

The scintillator times are smeared according to the length of the scintillator. The time

smearing from flag -f uses a 0.50 scale factor for SC tdcl/tdcr smearing and comes directly

Page 124: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

124

from the work in reference [72]. The 1.0 scale factor for DOCA smearing for Regions 1, 2,

and 3 are default values that introduce only minimal smearing. The wire map of run

number 43582 is uses are the wire map for the Monte Carlo. This run has a set of bad

wires consistently down in the g11a run period, this run is indicated with the -R flag. The

-Y flag drops the DC hits according to the efficiency in the GPP map, and the DC wire

map in the data base. Dead wires are removed with the -P Bitwise Process flag.

The GPP settings are selected by studying and comparing resolutions of data and

Monte Carlo and implementing a degree of smearing that leaves space for additional

kinematic dependent smearing. To more accurately reflect the resolution of the data for

each measured variable empirical smearing was also applied to each reconstructed vector

in the simulations. During the construction of the Covariance Matrix it was necessary to

obtain simulations with near identical resolutions to data so that the same Covariance

Matrix could be used for both. This was done by first applying the minimal smearing

mentioned, from GPP, to the Monte Carlo then sampling a Gaussian distribution with

mean value λ , and φ obtained from the recovered tracking angles. The width of the

Gaussian was provided by the resolutions in λ , and φ in the kinematic range of the data.

This effectively smears out the directional tracking variables by an additional factor of

∼1.91 of the resolution obtained from the tracking code. The value 1.91 is averaged over

all kinematic ranges. Momentum was smeared in a similar fashion to match data. Figure

7.1 shows the δ p as a function of momentum, where δ p = pk f it− pdet . The pk f it is the

momentum from the missing proton using the test channel γ p→ K+Λ. The pdet is the

proton detected momentum. This distribution is then sliced up and fit with a Gaussian in

several momentum bins and plotted and shown in Figure 7.1 (right). For each momentum

bin a different width of the smear term is applied. Without this type of careful smearing to

all tracking parameters, the confidence level obtained from any kinematic fit of the Monte

Carlo could not have a flat distribution for the appropriate hypothesis.

Page 125: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

125

To check that the smearing is working appropriately, the missing mass peak is

checked against data. A peak that is too narrow could lead to a misrepresented acceptance

in both the Λπ0 and Λγ channels. Figure 7.2 shows the change in the missing mass peak

width from the Monte Carlo reaction γ p→ K+Σ∗0→ K+Λ(X).

Figure 7.2: Comparison between data and Monte Carlo width of the missing mass squareddistribution. The left plot show the data with a Gaussian fit, while the middle is the MonteCarlo before smearing and the right is the Monte Carlo after smearing.

7.3 Monte Carlo Processing

After GPP, the output files were passed through RECSIS for processing and

reconstruction to prepare for analysis.

The RECSIS tcl configuration file was set such that:

source /u/group/clas/builds/release-4-8/packages/tcl/recsis proc.tcl;

turnoff ALL;

global section off;

Page 126: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

126

turnon seb trk tof egn user pid;

setc outbanknames(1) ”all”;

outputfile clas.out PROC 2047;

setc prlink file name ”prlink g11 1920.bos”;

setc bfield file name ”bgrid T67to33.fpk”;

set torus current 1920;

set mini torus current 0;

set poltarget current 0;

set TargetPos(3) -10.;

set trk maxiter 8;

set trk minhits(1) 2;

set trk lrambfit chi2 50.;

set trk tbtfit chi2 70.;

set trk prfit chi2 70.;

set trk statistics 3 ;

set dc xvst choice 0;

set def adc -1;

set def tdc -1;

set def atten -1;

set def geom -1;

set st tagger match 15.;

set lst do -1;

set lpid make trks 0;

set trigger particle 2212;

where the frozen version of recsis proc.tcl for g11 processing was sourced.

Page 127: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

127

7.4 Trigger Simulations

The GSIM package helps to systematically account for inefficiencies in the various

CLAS detection hardware for a given set of tracks triggers an event. Extensive efforts have

been done to develop a procedure to correctly simulate the g11a trigger. The investigation

by Krahn [72] utilized the trigger word written into the data stream for each recored event.

The CLAS Level 1 trigger requires a coincidence between the start counter time and TOF

scintillator time within a time window for two charged tracks in two different sectors. The

trigger word contains information on which sectors met the CLAS Level 1 trigger

requirements. The study uses the test channel γ p→ pπ+π− to check how many times one

of the three final state particles did not meet the trigger condition. This work lead to a

“trigger map” for each particle type (proton, π+, π−) as a function of sector, TOF paddle

and azimuthal angle φ . The same map is used for both the K+ and the π+ [72].

To simulate the trigger in Monte Carlo, a random number between 0 and 1 is

generated and if this number is less than the efficiency in the map, then the trigger word is

set to positive for that track. The event is accepted if two or more tracks are triggered by

this method.

Because of the nature of the present analysis, no effect is seen from the trigger

efficiency correction. The change in effective triggers in the Monte Carlo decreases

equally in the dominating acceptance terms in the numerator and denomination of the final

results, due to similar topology. The procedure is reproduced only as a check to verify that

the final ratio is not changed.

7.5 Matching Data and Tuning

The FSGEN generator is based off the CERN generator GENBOD, with refined

modification to do physics at the CLAS energy scale and the added features for generating

t-channel production events. These type of generators make multi-particle weighted

Page 128: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

128

events according to Lorentz-invariant Fermi phase space. The total center-of-mass energy,

as well as the number and masses of the outgoing particles are specified by the user, but

may be changed from event to event. The center-of-mass vector momenta (and energies)

of the outgoing particles are generated with the weight that must be associated with each

event. The average proper lifetime can be specified by the user for each ground state

decay product. To demonstrate, the lifetime of the λ is reconstructed from the Monte

Carlo channel γ p→ K+Σ∗→ K+Λπ0 using the distance between the primary interaction

vertex and the reconstructed vertex from the proton and π−. The mean life from the PDG

[60] is entered into the generator as cτ = 7.89 cm, while the decay time from the fit is

cτ = 7.82 cm, as seen in Figure 7.3, which is within reason.

Figure 7.3: Decay time for the Λ from the Monte Carlo generated for γ p→ K+Σ∗ →K+Λπ0.

The generator uses a relativistic Breit-Wigner shape for all resonance mass

distributions. The width of the resonance needs to by specified for each mass. The K∗,

Λ(1405), and Σ∗ Monte Carlo mass distributions are shown if Figure 7.4. These

Page 129: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

129

distributions are from the γ p→ K∗+Λ, the γ p→ K+Λ(1405), and the γ p→ K+Σ∗0,

respectively.

The analysis relies on small corrections from the acceptance of the detector and

leakage of various channels into the signal being extracted. Because of the minimal size of

the leakages of the background, as shown by the acceptance tables in section 7, the effect

of individual cross sections on the final result is very small. The channels studied here

with Monte Carlo are listed in Table 8.4.

Figure 7.4: Monte Carlo distributions for the left: K∗, middle: Λ(1405), and right: Σ∗0. ABreit-Wigner fit is used to demonstrate accurate width and mass in each case.

To tune the Monte Carlo for the Σ∗→ Λπ0 channel, first a 1/Eγ photon energy

distribution was used. Figure 7.5 show the bremsstrahlung distribution for both data and

Monte Carlo. The photon energy distribution for the data in the figure was obtained using

unskimed data without any cuts and iterating over all photons in the TAGR bank. The

Monte Carlo photon energy distribution in the figure come directly from the generator,

also without any cuts. The acceptance for the Σ∗→ Λπ0 channel was determined by the

ratio of accepted events to thrown events in each Eγ bin. The data and Σ∗→ Λπ0 Monte

Page 130: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

130

Table 7.1: The set of Monte Carlo channel generated for acceptance studies.

Reaction Generated number of counts produced

Λ(1405)→ Σ0π0 4×106

Λ(1405)→ Σ+π− 4×106

Λ(1405)→ Λγ 4×106

Λ(1405)→ Σ0γ 4×106

Σ(1385)→ Λπ 1.8×107

Σ(1385)→ Σ+π− 4×106

Σ(1385)→ Λγ 1.8×107

Σ(1385)→ Σ0γ 4×106

ΛK∗+→ K+π0 4×106

ΛK∗+→ K+γ 4×106

Figure 7.5: Left: Photon energy distribution for data using unskimed data without anycuts. Right: Photon energy distribution taken from the Monte Carlo generator showing thebremsstrahlung distribution.

Carlo were cut on the Y ∗ mass range of 1.34-1.43 GeV, the Λ invariant mass at ±0.005

GeV around the PDG value of the Λ mass, and the missing mass squared of the reaction at

Page 131: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

131

0.017-0.022 GeV2, to isolate the Σ∗→ Λπ0 channel in the data. The yield was determined

by the ratio of the raw Λπ0 events to the number of incident photons in each Eγ bin, so as

to normalize with the Bremsstrahlung spectrum. Corrections were made for each bin with

the newly obtained acceptance. The density of the target was assumed to be constant, and

no background subtraction was preformed. The resulting cross section was then fit with a

sixth-order polynomial. The polynomial curve was then implemented in the generator

using a Von Neumann rejection method to improve the photon energy distribution to

better match the form of the Λπ0 cross section from the data. This new Λπ0 Monte Carlo

is then used to find a more accurate acceptance and the process is done again. Using the

momentum distributions of each decay product, shown in Figure 7.8, no further

improvement in Monte Carlo to data matching is seen after one iteration. Figure 7.5

Figure 7.6: Left: Cross section and fit function used in Monte Carlo generation for thereaction Σ∗ → Λπ0. Right: Comparison between data and generated cross section afterusing correction to the photon energy distribution.

shows the cross section with respect to photon energy from the data channel Λπ0 used to

obtain the line function used in event weighting. The last point at high photon energy was

not used in the fit. A comparison between Monte Carlo cross section and data cross

Page 132: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

132

section with respect to photon energy is shown in Figure 7.5 (right) once the modification

to the Monte Carlo cross section was made.

From the adjusted Monte Carlo, acceptance corrections are found for bins in the kaon

cosine center-of-mass angle and the approximated differential cross sections are used to

adjust the Λπ0 generator to fill each corresponding bin according to the form of the

distributions from the data. Each angle bin is broken in to Eγ bins and represented

accordingly in the new event weighting scheme of the generator. The differential cross

section used in this modification are shown in Figure 7.7 for energies 1.5-3.1 GeV.

Figure 7.7: Approximated differential cross section for γ p→ K+Σ∗0→ K+Λπ0 in cosθKin the center-of-mass frame.

Page 133: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

133

After these modifications were made the resulting Monte Carlo can be checked with

the data using the momentum distributions for the kaon, pion, and proton tracks as well as

the kaon lab frame angle distribution, see Figure 7.8. The change in the branching ratio

from these corrections is very small and discussed in Section 9.

The form of the differential cross sections found for the reaction

γ p→ K+Σ∗0→ K+Λπ0 was also used to generate the other Σ∗0 channels. This includes

the radiative channel as well as the Σ+π− channel. It is shown later in Section 9 that

because of the low acceptance of the Σ+π− channel, these corrections are very small, and

make no difference in the resulting ratio.

In method-1 (section 8.1), a flat distribution with no t-dependence was used to

produce the K∗Λ and Λ(1405) decays, which was also used in previous studies [10]. In

method-2 (section 8.2), a t-slope of 2.0 GeV2 is used for all the Λ(1405) channels. The

corrections based on the acceptances of these contributions are small and tuning the

Monte Carlo generator make changes that are negligible in the final branching ratio. This

is discussed further in Section 9.

The simulations for γ p→ K+Σ∗0→ K+Λγ and γ p→ K+Σ∗0→ K+Λπ0 were

produced and studied. The same cuts as used for data are applied, and the missing mass

spectrum is shown along with the missing mass off the p-π−, seen in Figure 7.9 for the

radiative channel and Figure 7.10 for the π0 channel. The same distributions are studied

for the various Λ(1405) decays, as well as for the γ p→ K+Σ∗0→ K+Σ+π− channel (see

Figures 7.11-7.14). For the Λ(1405)→ Σ+π− decay, very few events should pass all the

cuts, so the distribution is shown before kinematic constraints are used. The various

Monte Carlo distributions for the γ p→ K+Σ∗0→ K+Λ(X) give a basis for comparison

when studying the missing mass for the data. One can also gain perspective on the form of

the anticipated background using the Monte Carlo simulations. For example, the

K∗→ K+Λ channel is expected to add a peak to the missing mass off the Λ around 0.9

Page 134: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

134

GeV, as seen in Figure 7.17. The radiative decay of the K∗ is also considered, though that

signal is very small.

Figure 7.8: Final comparison between Monte Carlo (lines) and data (points with errors) forthe reaction γ p→ K+Σ∗0→ K+Λπ0. Top left: cosθ of kaon in the lab frame. Top right:kaon momentum distribution. Bottom left: pion momentum distribution. Bottom right:proton momentum distribution.

To help demonstrate the possible effects of the various backgrounds to the missing

mass distribution and the missing mass off the Λ, the Monte Carlo distributions are shown

in Figures 7.11-7.17 for the γ p→ K+Λ(1405), the γ p→ K+Σ∗0→ K+Σ+π−, and the

γ p→ K∗+Λ channels. Each channel is used to obtain a correction term for the

corresponding background channel that can be used to correct for the contamination of

that channel in the final ratio. This is discussed in Section 8.1.

Page 135: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

135

Figure 7.9: Monte Carlo for the γ p→ K+Σ∗0→ K+Λγ channel. Left: the K+pπ− missingmass squared. Right: the missing mass off the Λ.

Figure 7.10: Monte Carlo for the γ p→ K+Σ∗0 → K+Λπ0 channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Figure 7.11: Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ0π0 channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Page 136: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

136

Figure 7.12: Monte Carlo for the γ p→ K+Λ(1405)→ K+Λγ channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Figure 7.13: Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ0γ channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Figure 7.14: Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ+π− channel (Before Cuts).Left: the K+pπ− missing mass squared. Right: the missing mass off the Λ.

Page 137: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

137

Figure 7.15: Monte Carlo for the γ p→ K+Σ∗ → K+Σ+π− channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Figure 7.16: Monte Carlo for the γ p→K∗+Λ→ΛK+γ channel. Left: the K+pπ− missingmass squared. Right: the missing mass off the Λ.

Figure 7.17: Monte Carlo for the γ p → K∗+Λ → ΛK+π0 channel. Left: the K+pπ−

missing mass squared. Right: the missing mass off the Λ.

Page 138: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

138

Figure 7.18: Left: Monte Carlo missing energy for the γ p→ K∗+Λ→ ΛK+π0 channel.Right: Monte Carlo missing energy for the γ p→ K+Λ(1405)→ K+Σ0π0 channel.

Page 139: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

139

8 EXTRACTION METHODS AND CONSTRAINTS

In the following sections, three different methods are laid out based on different

constraints required in the kinematic fitting procedure. The steps required to obtain the

radiative branching ratio for the excited-state hyperon Σ(1385) region relative to the

well-known Λπ0 decay channel of the Σ0(1385) are described. The procedure for

obtaining the Λ(1405) counts and K∗ counts that are used to correct the branching ratio is

outlined in the discussion of method-1 in Section 8.1. The procedure for obtaining these

background contributions is exactly the same as for the other two methods using the

acceptance terms for the corresponding method.

The tactic behind the use of constraints in an investigation of the branching ratio is to

maximize the usage of the known physics contained in the measured values without

significantly depleting statistics. In the original CLAS analysis [10] of this ratio, the p-π−

was constrained to be a Λ in a separate (2-C) fit prior to the final kinematic fitting

procedure used to separate the radiative signal from the π0. Using cuts based on

covariance early in the analysis can introduce biases if there is further reliance on the

covariance matrix at later stages in the analysis, as is discussed in Ref. [71]. Method-1 in

Section 8.1 attempts to reproduce the previous work by Simon Taylor [10] without a prior

kinematic fit of the p-π− to the invariant mass of the Λ. The Monte Carlo generation in

method-1 is also handled in attempt to be consist with this previous study.

Even with an excellent quality covariance matrix, the size of the radiative signal to

the π0 and the similar topology makes it difficult to cleanly separate them with a single

kinematic fit. This lead to the use of a two-step kinematic fitting procedure that first fits

with a missing π0 hypothesis, preserving only the low confidence level candidates, then

fits these remaining events with a radiative hypothesis and preserves the high confidence

level candidates. This two-step kinematic fitting procedure requires all other background

channels to be previously minimized so that there is a high probability that the low

Page 140: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

140

confidence levels event from the missing π0 fit are from the radiative decay of the

Σ0(1385). The resulting counts after the final fit are then corrected by the acceptance and

background subtraction to result in the true radiative counts. The same is done for the π0

channel. Once the counts for each channel and all the acceptances for each leakage

channel are known, a branching ratio can be calculated. This two-step kinematic fitting

technique is used in all three methods.

After the first kinematic fit, it is necessary to test the quality of the radiative signal

going into the final kinematic fit. This is done by studying the χ2 probability density

function of the events in the second kinematic fit. If the fit χ2 distribution emulates the

theoretical distribution for that particular set of constraints then the majority of events are

matching the radiative hypothesis. This “quality test” can be used to place the numerical

value of the first confidence level cut to get the best χ2 distribution. When the final

kinematic fit is performed, the resulting events will have a maximized probability of

satisfying the radiative hypothesis. There should then be little systematic variation in the

ratio based on the final confidence level cut as long as the cut is beyond the noticeable rise

around zero in the confidence level distribution from the poor quality candidates.

In the first method (Section 8.1), the χ2 probability density function shape is used to

regulate the confidence level cuts in the (1-C) fits. Due to the nature of the singularity in

the χ2 distribution of the (1-C) fit, only qualitative information can be deduced, leading to

a more arbitrary placement of the first confidence level cut. One can reduce this

uncertainty by using a (2-C) fit in the separation procedure, which has a χ2 distribution

that is easier to fit, putting a more quantitative regulation on the confidence level cuts. To

do this, a constraint is placed on the invariant mass of the Λ in the final kinematic fitting

procedure taking it from a (1-C) to a (2-C) fit. This is shown in method-2 and discussed in

Section 8.2.

Page 141: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

141

Kinematically fitting the p-π− and the missing particle all together to the mass of the

Σ∗0 is also a (2-C) fit. This is the approach of method-3. However, a greater number of

good candidates would be excluded from the same confidence level cuts due to the

Breit-Wigner shape of the resonance (discussed in reference [71]) leading to an increase in

systematic uncertainty at the same confidence level cuts or a increase in statistical

uncertainty at more effectively placed confidence level cuts. Method-3 is completed only

as a consistency check and discussed in Section 8.3.

Method-2 is then the most suitable approach, with reasonable constraints that

preserve the highest number of good quality candidates. This method is used in the final

reporting of the branching ratio as well as the systematic studies discussed in Section 9.

Each method assumes the cuts previously outlined to minimize background before

kinematic fitting.

8.1 Method-1

The first method shown is meant to emulate the original analysis for this channel as

done by Taylor [10]. Some differences in the constraints and how they are applied are

introduced to reduce any systematic bias in the fitting procedure. Details in kinematic and

vertex fitting that can introduce systematic bias are covered in Ref. [71]. A suitable

kinematic fitting procedure is used to fit each track of all detected particles to a particular

missing particle hypothesis. This is done by using the undetected particle mass in the

constraint equation. The detected particle tracks are kinematically fit as a final stage of

analysis and filtered with a confidence level cut. There are three unknowns (~px) and four

constraint equations from conservation of momentum: this is a simple (1-C) kinematic fit.

In the attempt to separate the contributions of the Σ∗0 radiative decay and the decay to

Λπ0, the events were fit using different hypotheses for the topology:

Page 142: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

142

γ p→ K+pπ−(π0) (1-C)

γ p→ K+pπ−(γ) (1-C).

The constraint equations are

F =

Ebeam +Mp−EK−Ep−Eπ −Ex

~pbeam−~pK−~pp−~pπ −~px

=~0. (8.1)

~Px and Ex represent the missing momentum and energy of the undetected π0 or γ .

Because of such similar topologies and the size of the radiative signal, the kinematic

fitting procedure cannot be expected to cleanly separate the “Λγ” events from the “Λπ0”

events in just one fit. This is handled with a two-step kinematic fitting procedure, making

first a kinematic fit to a missing π0 hypothesis and then checking the quality of the fit of

the low confidence level candidates in a second kinematic fit to the actual radiative

hypothesis.

One way to check the quality of the fit to a particular hypothesis is to use a fit to the

χ2 function from the fitting procedure. Because a (1-C) fit is used, the χ2 distribution for

one degree of freedom is used [71] and a fit function with a flat background (represented

by P2) can be applied,

f (χ2) =

P0√2Γ(1/2)

e−P1χ2/2√(χ2)

+P2. (8.2)

By setting an upper bound on parameter P0 it is possible to fit around the singularity

and get a quantitative measure for the quality of the fit from parameter P1. The bound on

P0 is chosen so that χ2 > 1.0×10−6, which is reasonably close to zero. Because of this

restriction around the singularity, it is difficult to get a very good fit, but in principle the

closer P1 is to unity, the closer the data matches the ideal χ2 distribution for one degree of

freedom. Naturally the distribution is not expected to be ideal due to the present

background, but it is possible to use the χ2 distribution fit in conjunction with the

confidence level from the kinematic fit as an additional tool for testing the resolving

capacity of the fitting procedure.

Page 143: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

143

Figure 8.1: Left: χ2 distribution for the π0 hypothesis with a fit using Eq. 8.2. Right:confidence level distribution for the π0 hypothesis.

Figure 8.2: Left: χ2 distribution for the γ hypothesis with a fit using Eq. 8.2. Right:confidence level distribution for the γ hypothesis.

Considering the π0-hypothesis for the data in the fit, the confidence level in Figure

8.1 is flat (right) and the χ2 distribution (left) looks as expected for one degree of freedom.

The P1 parameter is close to 1 and the expected shape is quite reasonable.

For the γ-hypothesis the confidence level also seems relatively flat in Figure 8.2

(right), but the χ2 distribution (left) has a different shape than expected. The fit to a

distribution for one degree of freedom shows a P1 parameter that is particularly small at

0.15. This is an indication that the vast majority of data being kinematically fit at this

stage are not satisfying the base assumption that the missing particle is massless. This

Page 144: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

144

implies that even with a high confidence level cut there will still be an overwhelming

amount of π0 contamination passing through. It is possible to take an additional step in

the kinematic fitting procedure for cleaner separation.

The two-step kinematic fitting procedure is used to first fit to a π0-hypothesis and

take all low confidence level candidates, and then fit to a γ-hypothesis and take a high

confidence level cut. Because of the previous cuts outlined earlier in the analysis, there

should now be primarily π0 background and the true radiative signal. By first fitting to π0

and taking the low confidence level candidates, one reduces the measured probability that

the surviving candidates will have a missing mass of the π0 before they are finally fit to a

γ-hypothesis.

The initial selection of the confidence level cut was chosen with the intention of

separating the γ p→ K+Σ∗0→ K+Λπ0 and γ p→ K+Σ∗0→ K+Λγ channels, while

maximizing the statistics. A confidence level P(χ2) of 10% is estimated using Figure 8.1

(left) to be able to minimize contamination and is used as a starting point. It is important

to start at a point before the distribution begins to rise on the low end of the confidence

level distribution. These events represent events that do not meet the hypothesis in the fit

at all. Various confidence level cuts can be used to study the systematic variation of the

ratio, see Section 15.2.

The two-step kinematic fit is implemented by a requirement of Pπ0(χ2) ≤ 10% and

Pγ(χ2) > 10% to isolate the radiative channel, where Pπ0 and Pγ are the confidence levels

of the π0 missing mass hypothesis and the γ missing mass hypothesis, respectively. A

requirement of Pγ(χ2) ≤ 10% and Pπ0(χ2) > 10% is then used to isolate the π0 events.

The candidates that pass all other cuts, but do not pass both secondary cuts for each

hypothesis Pγ(χ2) > 10% and Pπ0(χ2) > 10%, are used in the calculation of how much

background is present from such channels as the Λ(1405). An acceptance from Monte

Page 145: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

145

Carlo is needed for all background channels to be able to determine the leakage and

corrections needed.

Figure 8.3: Left: χ2 distribution for γ hypothesis with a fit using Eq.8.2. Right confidencelevel distribution for γ hypothesis.

After the two-step kinematic fitting procedure, one can again study the γ-hypothesis

confidence level, which seems relatively flat in Figure 8.3, as well as the χ2 fit, which now

looks more like a standard distribution for one degree of freedom. The fit parameter P1 has

increased much closer to 1. This is an indication that an improvement has been made on

the quality of candidates going into the fit. This gives some assurance that the candidates

that come out of the secondary fit can accurately be filtered with the confidence level cut.

Figure 8.4 shows the missing mass squared distribution after all cuts and the first

kinematic fit. The radiative signal defined by the first confidence level cut is seen around

zero, separated from the π0 peak. The π0 peak as defined by the first confidence level cut

is seen as the predominant peak separated from the radiative signal to the left of the peak

and other rejected background to the right of the peak.

8.1.1 Acceptance

The π0 leakage into the γ channel is the dominant correction to the branching ratio.

To properly calculate the ratio the leakage into the π0 region from the γ channel is also

Page 146: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

146

Figure 8.4: Missing mass squared distribution for the events that are going into the secondstep of the kinematic fitting procedure. The kinematic fit to π0 satisfying P(χ2) < 10%shows the radiative candidates (yellow), as well at the rejected background from theΛ(1405) (green). The white region shows the π0 candidates from a P(χ2)≥ 10% cut.

used. Taking just these two channels into consideration the number of true counts can be

represented as N(Λγ) for the Σ∗→ Λγ channel, and N(Λπ) for the Σ∗→ Λπ0 channel.

The acceptance under the Σ∗→ Λγ hypothesis can be written as Aγ(X), with the subscript

showing the hypothesis type and the actual channel of Monte Carlo input to obtain the

acceptance value is indicated in the parentheses. For the calculated acceptance of the

Σ∗→ Λγ channel, under the Σ∗→ Λγ hypothesis, the acceptance is Aγ(Λγ), and for the

Σ∗→ Λπ0 hypothesis it is Aπ(Λγ). It is now possible to express the measured values for

each channel nγ and nπ as

nγ = Aγ(Λγ)N(Λγ)+Aγ(Λπ)N(Λπ) (8.3)

Page 147: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

147

nπ = Aπ(Λπ)N(Λπ)+Aπ(Λγ)N(Λγ). (8.4)

The desired branching ratio of the radiative channel to the π0 channel using the true

counts is then R = N(Λγ)/N(Λπ). This can be obtained by dividing Equation 15.2 by 8.4

and expressing in terms of R such that;

=RAγ(Λγ)+Aγ(Λπ)Aπ(Λπ)+RAπ(Λγ)

, (8.5)

then solving for R to get the branching ratio expressed in terms of measured values and

acceptances,

R =nγAπ(Λπ)−nπAγ(Λπ)nπAγ(Λγ)−nγAπ(Λγ)

. (8.6)

Equation 8.6 is under the assumption that contributions from the Σ(1385) will only

show up as Λγ or Λπ0, neglecting the Σ(1385)→ Σπ channel. The estimate of the total

number of Σ(1385)’s produced using the Λπ0 channel is:

N(Σ∗) =N(Σ∗→ Λπ0)

R(Σ∗→ Λπ0)A(Σ∗→ Λπ0), (8.7)

where R(Σ∗→ Λπ0) is the branching ratio of the Σ(1385) decay to Λπ0 and A(Σ∗→ Λπ0)

is the acceptance for that channel. It is then possible to deduce an estimate of the number

of Σ(1385)→ Σ+π− counts that would contribute to the π0 peak:

N(Σ∗→ Σ+

π−) = R(Σ∗→ Σ

+π−)A(Σ∗→ Σ

+π−)N(Σ∗) (8.8)

=R(Σ∗→ Σ+π−)A(Σ∗→ Σ+π−)

R(Σ∗→ Λπ0)A(Σ∗→ Λπ0)N(Σ∗→ Λπ

0), (8.9)

where R(Σ∗→ Σ+π−) is the branching ratio of the Σ(1385) to decay into Σ+π− and

A(Σ∗→ Σ+π−) is the corresponding acceptance after all cuts. It is possible to simplify the

expression by using,

RΣπΛπ =

R(Σ∗→ Σ±π∓)R(Σ∗→ Λπ0)

= 0.135±0.011,

using the PDG average value [60]. The two charged combinations of the Σπ have equal

probability of decay. The Clebsch-Gordon coefficient for the Σ∗→ Σ0π0 decay is zero,

Page 148: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

148

assuming isospin symmetry. The π0 peak counts can then be more accurately depicted,

while the γ peak counts remain unchanged. The observed counts expressed in terms of

true counts and corresponding acceptances for each hypothesis can then be expressed as

nγ = Aγ(Λγ)N(Λγ)+(Aγ(Λπ)+RΣπ

Λπ

2Aγ(Σπ))N(Λπ) (8.10)

and

nπ = (Aπ(Λπ)+RΣπ

Λπ

2Aπ(Σπ))N(Λπ)+(Aπ(Λγ))N(Λγ). (8.11)

Solving for R will result in a branching ratio that includes all needed information from the

Σ(1385). Though the corrections to R should be small for other contamination it is

necessary to include them in the calculation. The variation in the ratio based on the

background contamination is studied in Section 9 and shown to be quite small. The

distributions seen in Figures 7.11-7.17 indicate that there is some probability that

contamination for these other channels can leak through and acceptance studies need to be

done for all channels under both the Λγ and Λπ0 hypotheses. Results from the acceptance

for each hypothesis can be seen in Table 8.3. The branching ratio must include corrections

for the K∗+→ K+X and the Λ(1405)→ Σ+π− contamination, as well as the contribution

to the numerator from the Λ(1405)→ Λγ decay. The leakage of the Σγ channel is

assumed to be small relative to the Λγ signal. However, this channel is still considered in

the acceptance studies, see Table 8.3.

The formula for the branching ratio to take these backgrounds into consideration is

R =∆nγ

(AΣ

π(Λπ)+ RΣπΛπ

2 AΣπ(Σπ)

)−∆nπ

(AΣ

γ (Λπ)+ RΣπΛπ

2 AΣγ (Σπ)

)∆nπAΣ

γ (Λγ)−∆nγAΣπ(Λγ)

,

(8.12)

∆nπ = nπ −Nπ(Λ∗→ Σ+

π−)−Nπ(Λ∗→ Σ

0)−Nπ(Λ∗→ Σ0γ)

−Nπ(Λ∗→ Λγ)−Nπ(K∗→ Kπ0),

(8.13)

Page 149: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

149

∆nγ = nγ −Nγ(Λ∗→ Σ+

π−)−Nγ(Λ∗→ Σ

0)−Nγ(Λ∗→ Σ0γ)

−Nγ(Λ∗→ Λγ)−Nγ(K∗→ Kγ). (8.14)

The nγ (nπ ) terms come directly from the yield of the kinematic fits and represent the

measured number of photon (pion) candidates. In the notation used, lower case n

represents the measured counts, while upper case N represents the acceptance corrected or

derived quantities. The Nγ,π terms are corrections needed for the leakage from the

Λ(1405) and K∗ channel. These corrections are necessary to take into account due to the

fact that the structure of the background underneath the Σ(1385) is not flat, which could

lead to over estimating the Σ(1385) contribution. The structure of the various

contributions can be seen in the simulation plots in Figures 7.11-7.17. The notation

utilized is such that the pion (photon) channel identifications are denoted AΣπ(Σ+π−)

(AΣγ (Σ+π−)) so that AΣ

γ (Λπ) denotes the relative leakage of the Λπ channel into the Λγ

extraction and AΣπ(Λγ) denotes the relative leakage of the Λγ channel into the Λπ

extraction, where AΣ denotes acceptance strictly for Σ(1385) compared to the AΛ which is

the acceptance for the Λ(1405).

Table 8.3 lists all channels taken into consideration and the value of the acceptance

for the Pπ0(χ2) ≤ 10% with Pγ(χ2) > 10% and the Pγ(χ2) ≤ 10% with Pπ0(χ2) > 10%

cuts used. The table list three columns sorted by hypothesis Aγ , Aπ , and the counts that

made all other cuts but did not satisfy either the γ or π0 hypothesis denoted as Aγπ .

The number of N(Λ∗) is dependent on the number of observed counts n(Σ0π0) above

the π0 peak seen in green in Figure 8.4, which can be expressed as

N(Λ∗) =nΛ

R(Λ∗→ Σ0π0)AΛ(Σ0π0). (8.15)

The notation nΛ here is short hand for n(Σ0π0) while R(Λ∗→ Σ0π0) is the probability that

the Λ(1405) will decay to the Σ0π0 and AΛ(Σ0π0) is the probability that this decay

channel will be observed after all the applied cuts. Isospin symmetry is assumed so that

Page 150: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

150

R(Σ0π0) = R(Σ+π−) = R(Σ−π+)≈ 1/3 for the Λ(1405) decay channels. An estimate of

the number of counts in the π0 peak coming from the reaction Λ∗→ Σ+π−, using Eq.

8.15, can be calculated;

Nπ(Λ∗) = R(Λ∗→ Σ+

π−)AΛ

π (Σ+π−)N(Λ∗) =

AΛπ (Σ+π−)nΛ

AΛγπ(Σ0π0)

. (8.16)

A small adjustment can be made to ensure that the Λ∗→ Σ+π− contributions to the

green mound in Figure 8.4 are also included by adding in the relative acceptance

AΛ(Σ+π−) to the denominator. These acceptance terms are achieved by independently

using Monte Carlo for the γ p→ K+Λ(1405)→ K+Σ0π0 and

γ p→ K+Λ(1405)→ K+Σ+π− reactions and using the counts that survive all cuts but did

not satisfy either the γ or π0 hypothesis also denoted by Aγπ . The leakage for the

γ p→ K+Λ(1405)→ K+Σ+π− channel is very small but is included for completeness.

The final result is,

Nπ(Λ∗) =AΛ

π (Σ+π−)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−). (8.17)

From this example it becomes transparent how to express all other associated Λ(1405)

corrections using only the observed nΛ counts and the corresponding acceptance for that

channel. The corrections for the γ channel are obtained from,

Nγ(Λ∗→ Λγ) =AΛ

γ (Λγ)R(Λ∗→ Λγ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−), (8.18)

Nγ(Λ∗→ Σ0γ) =

AΛγ (Σ0γ)R(Λ∗→ Σ0γ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−), (8.19)

Nγ(Λ∗→ Σ0π

0) =AΛ

γ (Σ0π0)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−),

Nγ(Λ∗→ Σ+

π−) =

AΛγ (Σ+π−)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−). (8.20)

Page 151: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

151

For the π0 channel they take the form,

Nπ(Λ∗→ Λγ) =AΛ

π (Λγ)R(Λ∗→ Λγ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−), (8.21)

Nπ(Λ∗→ Σ0γ) =

AΛπ (Σ0γ)R(Λ∗→ Σ0γ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−),

Nπ(Λ∗→ Σ0π

0) =AΛ

π (Σ0π0)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−), (8.22)

Nπ(Λ∗→ Σ+

π−) =

AΛπ (Σ+π−)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−), (8.23)

where R is used for the corresponding branching ratio in each case. The value of 27 keV is

used for the width of the Λ(1405)→ Λγ decay, and the value of 10 keV is used for the

width of the Λ(1405)→ Σ0γ decay which come from reference [74]. The choice in these

values is discussed in Section 9. The terms for Aγ (Aπ ) are indicative of the acceptance

terms under a radiative (π0) hypothesis in the kinematic fit.

In order to find nΛ, one can look at the events for which neither the γ nor the π0

hypothesis is satisfied, seen in green in Figure 8.4. The value of nΛ is difficult to

determine due to the non-Breit-Wigner shape of the Λ(1405) decay. Because the

contribution of the Λ∗ to the overall ratio is small, an exact fit is not required. To establish

an upper and lower limit of the contamination affects on the ratio, two methods of

extracting the nΛ counts are employed and the variation in the ratio based on these two

methods is discussed later in Section 9. To obtain the maximum possible counts for nΛ the

raw counts rejected from the kinematic fit can be used. Because it is assumed that the

counts that survive all other cuts but do not satisfy the radiative hypothesis or the π0

hypothesis are primarily from the Λ(1405)→ Σ0π0 channel it is possible to use these

counts directly as an upper limit of the value of nΛ. This should be the maximized value of

nΛ because there should still be a notable amount π0 events that did not pass the high

confidence level cut of the π0 hypothesis.

Page 152: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

152

A second method was also employed. Monte Carlo was used to fill the background

according to internal decay kinematics and normalized to the data so that the Monte Carlo

mass spectrum level matches the that of the data providing an estimate for nΛ count. This

was done by using the Λπ0 channel Monte Carlo and normalizing it to the data, then

adding the full spectrum of the Λ(1405) Monte Carlo so that it would decay to all known

channels with appropriate branching ratios, under the same normalization as the Λπ0

channel Monte Carlo. The Λ(1405) reaction Monte Carlo was added until it matched the

data distribution relative to the Λπ0 channel.

Figure 8.5: Left: data (error bars) with Monte Carlo (line) from the Λπ0 channel and thefull spectrum of Λ(1405) filled to match the data. Right: data (error bars) with Monte Carlo(line) from the Λ(1405) only.

Figure 8.5 (left) shows the matching of the data and the Monte Carlo. The Monte

Carlo used is γ p→ K+Σ0→ K+Λπ0 mixed with the full spectrum of the

γ p→ K+Λ(1405) according to PDG [60]. By adding more and more Λ(1405) until the

region M2x (K+pπ−) > 0.027 GeV begins to match the data, an estimate of the amount of

Λ(1405) present relative to the Λπ0 channel is found. Figure 8.5 (right) shows the added

counts from the Λ(1405) without the Λπ0 channel, matched the data points in that region.

These counts are used to correct for all other contamination except for the K∗.

Page 153: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

153

Table 8.1: Counts for nΛ found though the different methods; the raw counts rejected fromthe π0 hypothesis, and the estimated number of nΛ from Monte Carlo. The uncertaintiesare fit uncertainties combined with statistical uncertainties.

Method Counts

Raw Counts 5429±82.45

MC Data Match 5112±69.95

Table 8.1 show the results of the nΛ counts for the two methods. The counts from the

Monte Carlo data matching are used in the final result, however the variation in the final

branching ratio is small based on this background as seen is Section 9.

Few counts from the K∗+ survive all the cuts, but to include corrections that consider

the few that do, an estimate of the K∗+ counts present must be established. With this

estimate it is possible to find the acceptance corrected number of K∗+ events,

N(K∗+) =n(K∗+→ K+π0)

R(K∗+→ K+π0)A(K∗+→ K+π0), (8.24)

where n(K∗+→ K+π0) is the estimated number of K∗+→ K+π0 events present.

R(K∗+→ K+π0) = 2/3 is the probability for the decay and A(K∗+→ K+π0) is the

acceptance of the K∗+→ K+π0 channel. A(K∗+→ K+π0) is determined by developing

an extraction method to obtain the K∗+→ K+π0 counts and observing how many thrown

events survive all cuts for the extraction method used. Three different extraction methods

of obtaining the K∗+→ K+π0 counts are discussed next. Once N(K∗+) is established, the

number of K∗+ that would be present under the π0 hypothesis could be expressed as,

Nπ(K∗+→ K+π

0) =Aπ(K∗+→ K+π0)n(K∗+→ K+π0)

A(K∗+→ K+π0), (8.25)

where Aπ(K∗+→ K+π0) is the acceptance for the K∗+→ K+π0 channel under the Λ(π0)

hypothesis. It is also possible to consider the radiative decay of the K∗ using N(K∗+) such

Page 154: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

154

that,

Nγ(K∗+→ K+γ) = R(K∗+→ K+

γ)Aγ(K∗+→ K+π

0)N(K∗+), (8.26)

Nγ(K∗+→ K+γ) =

32

R(K∗+→ K+γ)

Aγ(K∗+→ K+γ)A(K∗+→ K+π0)

n(K∗+→ K+π

0). (8.27)

Here the branching ratio R(K∗+→ K+γ) is small at 9.9×10−4 [60]. The change in the

ratio from the N(K∗+) events is quite small and was neglected in the previous analysis

[10], so the procedure to obtain the N(K∗+) counts is laid out here but adjustment to the

ratio is used only in method-2.

To acquire an estimate of the K∗ events present, again more than one method is

employed. The methods include: a straight Gaussian fit with quadratic background, an

estimate from MC matching to data, and an extrapolation using a series of Gaussian fits at

various excited state hyperon mass ranges.

Relatively few K∗ candidates survive the 1.34-1.43 GeV hyperon mass cut used in

the analysis as seen in Figure 8.6 (left). This make it difficult to get an accurate Gaussian

fit on the bump is the missing mass off the Λ spectrum. A simple Gaussian fit with

quadratic background is used to obtain a crude estimate of counts. The Gaussian fit,

shown in green is seen Figure 8.6 (left), with a quadratic background, shown in red, and

the final Gaussian, shown in blue.

It is much easier to get a minimized χ2 fit over the range that allows the higher part

of the excited state mass spectrum to pass through. Fits are done at ranges 1.34-1.5 GeV,

1.34-1.55 GeV, 1.34-1.6 GeV, 1.34-1.65 GeV, 1.34-1.7 GeV, 1.34-1.75 GeV, and 1.34-1.8

GeV. A Gaussian fit is acquired at each hyperon mass range. These resulting counts from

each Gaussian fit are then plotted for each mass window to produce the trend seen in

Figure 8.6 (right). The mass window is defined as the difference from the starting hyperon

mass range to the cut off. The range of interest from 1.34-1.43 GeV is 0.09 GeV. A

polynomial fit to the mass window bins is applied and then a value at the low range of

interest is extrapolated. Figure 8.6 (right) is used to extract a value of 1983 for the K∗

Page 155: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

155

contribution for the mass window of 0.09 GeV. The data matching method for obtaining

Figure 8.6: Left: fit with a Gaussian and quadratic background for the hyperon mass range1.34-1.43 GeV. Right: fit with polynomial to the point derived from Gaussian fits of variousmass windows.

Figure 8.7: Missing mass off of the Λ for the match of data and Monte Carlo to obtain thecounts from K∗. The lines is data and the points are from Monte Carlo after adding in theK∗ and Λ(1405).

the K∗ counts is the same as for the data matching for nΛ. After using the Monte Carlo to

first match the π0 peak, and then the Λ(1405) spectrum, the missing mass off of the Λ is

used to fill the full spectrum of the K∗ to get the Monte Carlo mass spectrum to

appropriately match that of data, see Figure 8.7.

Page 156: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

156

The estimated counts from various methods are reported in Table 8.2. The

extrapolated value, being the median of the three methods, is used in the final result, again

the variation in the final reported ratio due to this contribution is discussed in Section 9.

Table 8.2: Counts for nK∗ found though different methods. Uncertainties are fituncertainties combined with statistical uncertainties.

Method Counts

Gaussian Fit 2121±60.57

Extrapolation 1983±55.45

Data Match 1888±43.95

Once the estimated K∗ and Λ∗ counts are determined, all of the contamination can be

subtracted out of the calculated branching ratio in Eq. 15.5 using the form in Eq. 8.13 for

contamination in the Λπ0 channel and Eq. 8.14 for the contamination to the radiative

signal. The acceptance values required for the calculation of each contamination term in

Equations 15.12-8.23 are listed in Table 8.3. The table includes the acceptance values for

each Monte Carlo channel taken into consideration for the radiative hypothesis Aγ , the

missing π0 hypothesis Aπ0 , and the counts that survive all other cuts but do not satisfy

either of the radiative or missing π0 hypotheses Aγπ . The confidence level cuts used for

the two-step kinematic fitting procedure to produce the table and the final result where

Pγ = Pπ0 = 10%. Table 8.3 shows more Aγπ acceptance terms than are required. Only the

AΛγπ(Σ0π0) and the AΛ

γπ(Σ+π−) terms in the Aγπ column are needed for the branching ratio

calculation. The rest of the column is produced for completeness. The statistical

uncertainty reported for each term is based on the uncertainty from the counts

δnΛ ∼√

nΛ, the acceptance terms and uncertainties and ratio and associate uncertainty all

propagated accordingly.

Page 157: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

157

Table 8.3: Acceptances (in units of 10−3) for the channels used in the calculation of thebranching ratios. Here there is a 10% confidence level used as upper and lower P(χ2) cuts;the Pxy cut was 0.03 GeV. The uncertainties are statistical only. The three columns containthe acceptance for each hypothesis Aγ , Aπ , and the counts that made all other cuts but didnot satisfy either γ or π0 hypothesis denoted as Aγπ .

Reaction Aπ Aγ Aγπ

Λ(1405)→ Σ0π0 0.0495±0.0010 0.0009±0.0004 1.491±0.029

Λ(1405)→ Σ+π− 0.021±0.0008 0.001±0.0004 0.0078±0.0009

Λ(1405)→ Λγ 0.0051 ±0.002 1.61±0.025 0.0632±0.0012

Λ(1405)→ Σ0γ 0.291±0.009 0.190±0.007 0.731±0.022

Σ(1385)→ Λπ 1.98±0.0396 0.0561±0.002 0.0442±0.002

Σ(1385)→ Σ+π− 0.191±0.0004 0.0058±0.001 0.0018±0.0008

Σ(1385)→ Λγ 0.0179±0.002 2.131±0.042 0.0728±0.006

Σ(1385)→ Σ0γ 0.407±0.012 0.162±0.095 0.282±0.017

ΛK∗+→ K+π0 0.201±0.015 0.001±0.0004 3.261±0.051

ΛK∗+→ K+γ 0.0028±0.0001 0.162±0.003 2.230±0.046

Page 158: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

158

8.1.2 Results

To calculate the branching ratio Eq. 15.5 is employed. The terms that take the

Σ∗→ Σ+π− into account are relatively small. Using the acceptance values in Table 8.3,

these terms become;RΣπ

Λπ

2AΣ

π(Σπ) = 1.29×10−5, (8.28)

andRΣπ

Λπ

2AΣ

γ (Σπ) = 3.92×10−7. (8.29)

The numerator of Eq. 15.5 can then be expressed as,

∆nγ(1.98×10−3 +1.29×10−5)−∆nπ(0.0561×10−3 +3.92×10−7), (8.30)

and the denominator,

∆nπ(2.13×10−3)−∆nγ(0.0179×10−3). (8.31)

Each ∆n contains the raw counts extracted from the kinematic fit and the background

counts that should be removed.

The results for the various background contributions are listed in Table 8.4. Though

the largest background in the γ channel is from the contribution of the π0 tail, it is not

listed here because the correction is included in the acceptance terms seen if Eq. 8.30.

This is also true for the γ background leakage into the π0 channel, seen in Eq. 8.31. The

final values for the subtracted contamination terms come from Equations 15.12-8.23. The

raw counts for the radiative and π0 extraction using Pγ = Pπ0 = 10% are also listed.

After subtracting out the values from the table the true signal counts become

∆nγ = 579.42 and ∆nπ = 13281.38.

The final ratio, using Eq. 15.5, RΛγ

Λπfor the Σ(1385) for this method is then,

RΛγ

Λπ=

Γ[Σ0(1385)→ Λγ]Γ[Σ0(1385)→ Λπ0]

= 1.43±0.11% (8.32)

Page 159: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

159

Table 8.4: Breakdown of statistics for each term in Eq. 15.5 for the Λ(γ) and Λ(π0)hypothesis. Each listed channel is subtracted from the raw counts directly from thekinematic fit to obtain the final ratio. The uncertainties listed are statistical only. TheK∗+ counts are not used in method-1 in order to follow Taylor [10].

Hypothesis Λ(γ) Λ(π0)

Raw counts 589±24.27 13522±116.28

Λ(1405) : N(Σ0π0) 3.07± 1.37 168.8± 5.28

Λ(1405) : N(Σ+π−) 3.41± 1.36 71.62± 4.26

Λ(1405) : N(Λγ) 2.97± 0.60 0.01± 0.07

Λ(1405) : N(Σ0γ) 0.13± 0.053 0.19± 0.054

K(892) : N(ΛK∗+) 0.100± 0.011 14.8± 3.7

Adjusted counts 579.42±24.35 13281.38±116.48

The uncertainty presented is statistical only. The statistical uncertainty in the counts nπ

and nγ used is the square root of the counts. The overall statistical uncertainty is

propagated accordingly through Eq. 15.5 using the corresponding uncertainty for each

acceptance term that is listed in Table 8.3. The branching ratio 1.43±0.11% is

comparable within the statistical uncertainty to previous results 1.53±0.39% from Taylor

[10]. The essential differences in Taylor’s method using the G1C data set included using a

kinematic fit to the Λ hypothesis prior to the final missing π0 or radiative hypothesis, the

method of extracting background counts, and using a Gaussian fit after the final kinematic

fit to obtain the counts and uncertainties. In simulations, Taylor used an initial angular

dependence of 5−3cos2θ for the γ p→ K+Σ∗0→ K+Λπ0 channel. In the present analysis

the angular distribution from the data are used. Like Taylor, the present analysis uses a

zero t-slope for the Λ(1405) and the γ p→ K+Σ∗0→ K+Λγ channel; a more physical

representation is used in section 8.2.

Page 160: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

160

The method of using a two-step kinematic fitting procedure to extract the ratio has a

great deal of systematic dependency on the choice of the confidence level cut. This is first

studied by varying the overall range of the confidence level while keeping the relative

probability from the higher and lower bound consistent with each other such that Pπ0(χ2)

= Pγ(χ2) for isolating both the π0 and radiative signal. This is done in a range from 0.1%

to 40%, see Figure 8.8. At high confidence levels too much π0 background will pass the

first kinematic fit and the χ2 distribution will again be distorted. In addition to this issue,

for very low confidence level cuts in the initial fit, the π0 background is minimized, but

the same cut in the second kinematic fit would allow almost all candidates to pass, again

letting too much background in.

To correctly study the systematic variation of the confidence level cut, the

distribution of the χ2 should be first tested for each variation to make sure the cuts to

Pπ0(χ2) and Pγ(χ2) correspond to a good signal. To find the best range of variation, the

cuts are staggered so that Pπ0(χ2) and Pγ(χ2) are no longer equivalent. Table 8.5 shows

the selected range after an attempt to achieve the best quality of the radiative signal based

on a qualitative analysis of the χ2 distribution fit for one degree of freedom and the

resulting P1 fit parameter. Cuts are selected to keep P1 within 20% of P1 = 1. Also in the

table is a column for Pπ0(%)/Pγ(%) enabling an independent point for each value of R to

be plotted against for each cut selection. This is done over a range of confidence level

cuts, see Figure 8.9. In conclusion, the cuts 10% = Pπ0(χ2) = Pγ(χ2) used to obtain the

branching ratio is a reasonable choice.

Further systematic studies are not performed for this method. The χ2 probability

density function shape used to regulate the confidence level cuts for the (1-C) fits is too

difficult to fit, leading to no more than qualitative conclusions drawn from the fit

parameter P1. Because the uncertainty of the systematics in this approach can not be well

defined, only a statistical uncertainty is reported for the first method. Method-2 attempts

Page 161: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

161

to provide a clearly defined systematic uncertainty by using a (2-C) fit in the separation

procedure, which has a χ2 distribution that is easier to fit, putting a more quantitative

regulation on the confidence level cuts.

Table 8.5: Dependence of the corrected branching ratio on the confidence level cuts.

Pπ0(%) Pγ(%) R(%) Pπ0(%)/Pγ(%) P1

0.1 5 1.57± 0.13 0.02 1.128

1 5 1.52± 0.12 0.2 1.166

1 10 1.45± 0.11 0.1 1.166

10 10 1.43± 0.11 1.0 1.187

0.01 1 1.44± 0.14 0.01 1.101

Page 162: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

162

Figure 8.8: A study over a variation in confidence level; each cut corresponds to Pπ0(χ2) =Pγ(χ2).

Page 163: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

163

Figure 8.9: Only the selected confidence level cuts with the best quality signal based on theP1 parameter in the χ2 fit (values seen in Table 8.5). Pπ0(χ2)/Pγ(χ2) is used on the x-axisto obtain a distinguishable point for each ratio value.

Page 164: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

164

8.2 Method-2

It is possible to reduce the uncertainty in the ratio that comes from the initial Λ cut of

±0.005 GeV relative to the PDG value by using an additional constraint in the kinematic

fit. The additional constraint is introduced into the kinematic fitting procedure already

outlined in Section 8.1 but with the further requirement that the proton and π− track have

an invariant mass of the Λ in the hypothesis. To ensure that no systematic bias is

introduced, it is best to fit the Λ and missing mass hypothesis together, see reference [71].

This fitting procedure involves fitting each track of all detected particles to a particular

missing particle hypothesis, while requiring the proton and π− be constrained to have the

invariant mass of the Λ. This is done by using the undetected particle mass in the

constraint equation, while the invariant mass of the Λ constraint is also met. The detected

particle tracks are kinematically fit as a final stage of analysis and again filtered with the

confidence level cut. In this fit, there are three unknowns (~px) and five constraint

equations, four from conservation of momentum and then the additional invariant mass

condition. This makes a (2-C) kinematic fit. In the attempt to separate the various

contributions of the Σ∗0 radiative decay and the decay to Λπ0, the events were again fit

using the hypotheses for the topology:

γ p→ K+pπ−(π0) (2-C)

γ p→ K+pπ−(γ) (2-C).

The constraint equations are

F =

(Eπ +Ep)2− (~pπ +~pp)2−M2

Λ

Ebeam +Mp−EK−Ep−Eπ −Ex

~pbeam−~pK−~pp−~pπ −~px

=~0. (8.33)

Again, ~Px and Ex represent the missing momentum and energy of the undetected π0 or γ .

The same two-step kinematic fitting procedure is used to resolve the radiative signal

from the π0. An additional advantage of this approach is that it is possible to get a much

Page 165: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

165

better fit to the χ2 distribution for a 2-C fit because there is no singularity in the form of

f (χ2), which was the case for a 1-C fit. The distribution for a 2-C fit follows the from Ref.

[71],

f (χ2) =

P0

2e−P1χ2/2 +P2. (8.34)

This fit function still has a flat background term, P2, but P1 now becomes the measure of

how close the distribution in the histogram is to the ideal (P1 = 1.0) theoretical χ2

distribution for two degrees of freedom.

The more accurate nature of the fit to the theoretical probability density function for

two degrees of freedom enables a more systematic approach to defining the placement of

the confidence level cuts. Because there are two kinematic fits for both the π0 channel and

radiative signal, some new notation is introduced. The first confidence level cut used to

filter out the larger π0 signal from the radiative signal by using a kinematic fit to Λ(π0)

and taking only the low confidence level candidates is denoted as Paπ (χ2). The final

kinematic fit used to isolate the radiative signal, using a Λ(γ) hypothesis has a confidence

level cut denoted as Pbγ (χ2), taking only the high confidence level candidates. These two

confidence level cuts need not be at the same value, as was assumed in method-1

(following Taylor).

The key question here is how to select the cut points for Paπ (χ2) and Pb

γ (χ2). The

Λπ0 channel will be reduced for a lower value of Paπ (χ2), which is desirable for extracting

the radiative decay signal. On the other hand, this cut cannot be made arbitrarily small,

since it reduces the statistics (i.e., increases the statistical uncertainty). Similarly, the Λγ

signal will be purified by a higher cut on Pbγ (χ2), but again the higher the cut, the lower

the statistics. Of course, Monte Carlo can be used to examine the acceptance of these cuts

for various branching ratios (Λγ/Λπ0). In the end, the branching ratio extracted from the

data should not depend on the cut points chosen (assuming the Monte Carlo gives accurate

cut acceptances). As was shown for method-1, this was only the case after careful

Page 166: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

166

consideration of the P1 parameter. In method-1 the P1 parameter can only be studied

qualitatively. Furthermore, for method-2, the Monte Carlo can be used to optimize the

trade-off between statistical uncertainty and systematic uncertainty (due to the choice of

confidence level cuts based on a more quantitative analysis of P1).

Details of the optimization method of the confidence level cuts, using the Monte

Carlo, are described in [71]. Here, an overview of the optimization method is described.

In the Monte Carlo, only Λγ events can be generated, and the value for “pure” P1

signal is obtained with a fit to equation (8.34). This pure P1 can be compared with the P1

value from the fit of signal and Λπ0 contaminated events. The degree of contamination can

then be regulated with a low Paπ cut, giving the capacity to find an ideal Pa

π based on the

change in P1. This step is essential because the resolving capacity of the final kinematic fit

to Λγ is limited depending on the amount of signal to background present in the fit.

Examples of fits to the χ2 distributions for different amounts of Λγ and Λπ0 Monte

Carlo events present in the final kinematic fit to Λγ are shown in Figure 8.10. The low

statistics of this Monte Carlo study is intentional to emulate the statistics of the radiative

signal in the g11a data. A cut of Paπ (χ2) < 1% has been implemented before the final

kinematic fit in each case.

At the top (far left), the unphysical case of Λγ dominance is shown to demonstrate

that the fit gives a P1 value higher than those fits including events from the Λπ0 channel.

With an increasing number of Λπ0 events in the kinematic fit, the P1 parameter decreases

from the initial P1 values. In other words, the change of the P1 parameter from its initial

pure value is an indication of how much π0 background is present. The fit to the χ2

distributions for the different amount of Λπ0 background gives a basis of comparison

when analyzing the g11 data while using a Paπ (χ2) < 1% cut.

Page 167: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

167

A quantitative measure of the deviation of the P1 parameter can be defined (notation

σγ in [71]) and the dependence of this P1 deviation has been mapped out as a function of

these cut points [71].

Figure 8.10: The fit to the χ2 distribution from Monte Carlo for various mixtures ofradiative signal and Λπ0 events for small statistics, after cutting events with Pa

π (χ2) < 1%.The amount of π0 background still present in each case can be determined by looking atthe number of events added from the pure singal case in the top left plot.

At the heart of the optimization method, the Monte Carlo is used to determine both

the statistical uncertainty and the “recovery” uncertainty. The recovery uncertainty (given

Page 168: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

168

by δR in [71]) is given by the deviation of ratio R (see equation 8.6) from the full analysis

procedure as compared to the known ratio used in the Monte Carlo. As the confidence

level cuts are made tighter, the amount of Λπ0 events are reduced, which reduces the ratio

deviation (δR) and the deviation of the P1 parameter from its initial pure radiative value.

On the other hand, tighter cuts increases the statistical uncertainty in RΛγ

Λπ, so the

confidence level cuts are placed independently to improve statistics without introducing

uncertainty from δR.

The optimum cuts occur when the fractional deviation in the ratio δR equals the

fractional error due to statistics. The optimum cuts occur at different values for a given

mixture of Monte Carlo Λπ0 and Λγ events. Using the g11 data, the Monte Carlo has been

tuned to have approximately the same ratio and the same statistics as the real data, and the

optimum cuts are thereby determined quantitatively. Again, the details are given in Ref.

[71]. The optimum cuts, along with the value of δR, are given in Table 8.6. Note that the

value of δR is not the systematic uncertainty in RΛγ

Λπ. Rather, the systematic uncertainty

comes from the data, based on the variation in the extracted ratio of Eq. (15.5) after all

Monte Carlo corrections for each set of optimized cuts. The systematic uncertainty is

presented in Section 9.

Table 8.6: Optimization points for each Paπ0 and Pb

γ from Ref. [71].

Paπ0(%) Pb

γ (%) δR

7.5 25 0.097

6.25 20 0.09

5.0 15 0.087

1.0 10 0.085

0.075 5 0.088

Page 169: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

169

Equally important to extracting the true radiative signal is extracting the true Λπ0

signal, which is the leading term in the denominator of Eq. (15.5). To do this, we reverse

the above procedure, and first take the low confidence level events for the Λ(γ)

hypothesis, denoted by the Paγ (χ2) cut, and then take the high confidence level events for

the Λ(π0) hypothesis, denoted by the Pbπ (χ2) cut. (Note the reversal in the use of a and b

as compared with the radiative signal given previously.)

As a starting point, the cuts Pbπ (χ2) > 10% and Pa

γ (χ2) < 1% are used to extract the

Λπ counts. Variations in these cuts are studied in Section 9. Figure 8.11 shows the χ2 and

confidence level distributions of the π0 candidates after the Paγ (χ2) < 1% cut and before

any cut on Pbπ (χ2). The fit to the χ2 distribution shows a P1 parameter of 1.117, close to

the ideal of unity, indicating a very small portion of background from the radiative decay

is present in the kinematic fit. In fact, the cut on Paγ (χ2) removes very few events; the

plots in Figure 8.11 would look very similar even without this cut, since there is so little

Λγ background.

For comparison, the corresponding χ2 and confidence level distributions for the

radiative hypothesis are shown in Figure 8.12 before any cuts are applied. Here, the χ2

distribution (left plot in Fig. 8.12) is far from the ideal, indicating that the vast majority of

events are not consistent with the radiative hypothesis, as expected since the Λπ0 events

are dominant.

After regulating the background of the π0 with the Paπ (χ2) < 1% cut, the χ2 and

confidence level distributions improve as seen in Figure 8.13. The fit to the χ2 distribution

shows a P1 parameter of 0.868 indicating that a much greater percentage of events in the

fit are now consistent with the radiative hypothesis. The value P1 can be compared to the

values seen in Figure 8.10 giving an estimate of the amount of Λπ0 background still

present in the final Λγ fit. The placement of the Pbγ is determined within 10% by using δR

from Table 8.6.

Page 170: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

170

Figure 8.11: Left: χ2 distribution and fit with two degrees of freedom for the π0 candidatesfrom the g11a data after the Pa

γ (χ2) < 1% cut and before the Pbπ (χ2) > 10% cut; Right: the

corresponding confidence level distribution for the π0 candidates.

Figure 8.12: Left: χ2 distribution and fit with two degrees of freedom for the γ candidatesfrom the g11a data before the Pa

π (χ2) < 1% cut; Right: the corresponding confidence leveldistribution for the γ candidates.

Page 171: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

171

Figure 8.13: Left: χ2 distribution and fit with two degrees of freedom for the γ candidatesfrom the g11a data after the Pa

π (χ2) < 1% cut and before the Pbγ (χ2) > 10% cut; Right: the

corresponding confidence level distribution for the γ candidates.

Page 172: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

172

8.2.1 Results

The branching ratio is calculated using Eq. 15.5 as outlined in method-1. ∆β is

selected using a 1 ns timing cut as described while keeping ∆β < 0.02 for the kaon but

using a looser cut of the ∆β < 0.05 for the proton and pion. This choice is discussed in

Section 9. All other cuts remain the same as previously outlined. The procedure to obtain

the acceptance terms is the same but must be done for the new constraints and cuts

implemented. The acceptance used in listed in Table 8.7.

Table 8.7: Acceptances (in units of 10−3) for the channels used in the calculation ofthe branching ratios. Here Pb

γ (χ2) > 10% and Paπ (χ2) < 1% while Pb

π (χ2) > 10% andPa

γ (χ2) < 1% and the Pxy cut was 0.03 GeV. The uncertainties are statistical only. The threecolumns contain the acceptance for each hypothesis Aγ , Aπ , and the counts that made allother cuts but did not satisfy either the γ or π0 hypothesis denoted as Aγπ .

Reaction Aπ Aγ Aγπ

Λ(1405)→ Σ0π0 0.0495±0.0031 0.001±0.0001 1.189±0.019

Λ(1405)→ Σ+π− 0.029±0.002 0.0013±0.0001 0.0078±0.001

Λ(1405)→ Λγ 0.0011±0.0001 1.65±0.031 0.0223±0.002

Λ(1405)→ Σ0γ 0.170±0.012 0.191±0.009 0.437±0.013

Σ(1385)→ Λπ 1.421±0.0278 0.0321±0.002 0.0312±0.002

Σ(1385)→ Σ+π− 0.161±0.01 0.00254±0.001 0.00138±0.0006

Σ(1385)→ Λγ 0.0184±0.002 2.335±0.039 0.0704±0.005

Σ(1385)→ Σ0γ 0.191±0.011 0.058±0.0001 0.225±0.015

ΛK∗+→ K+π0 0.213±0.010 0.010±0.006 2.931±0.051

ΛK∗+→ K+γ 0.0022±0.0001 0.158±0.003 2.351±0.046

The methods used for obtaining the background counts for the γ p→ K+Λ(1405) and

the γ p→ K∗+Λ outlined in method-1 are the same, but with the exception of using a more

Page 173: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

173

realistic t-slope of 2 GeV2 for the Λ(1405) channels [75]. The differential cross section

from the data from the γ p→ K+Σ∗0→ K+Λπ0 are used to modify the generator for all

Σ∗0 channels.

Because the amount γ p→ K+Λ(1405) and γ p→ K∗+Λ background counts depend

on the hypothesis and confidence level cuts used, it is necessary to obtain new values.

Table 8.8 shows the result for the three methods used to obtain the Λ(1405) counts and

Table 8.9 shows the result for the three methods used to obtain the K∗+Λ counts.

Table 8.8: Counts for nΛ found through the different methods; the raw counts rejected fromthe π0 hypothesis, and the estimated number of nΛ from Monte Carlo. The uncertaintiesare fit uncertainties combined with statistical uncertainties.

Method Counts

Raw Counts 4172±64.59

MC Data Match 4085±63.91

Table 8.9: Counts for nK∗ found through the different methods. Uncertainties are fituncertainties combined with statistical uncertainties.

Method Counts

Gaussian Fit 1352±36.77

Extrapolation 1289±35.90

Data Match 1207±34.74

The counts nγ and nπ are again taken directly from the kinematic fit, see Figure 8.14.

The newly calculated acceptance terms from Table 8.7 are used with the nΛ and

Page 174: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

174

Figure 8.14: Left: the nπ counts extracted using the confidence level cuts Paγ < 0.01 and

Pbπ > 0.1. Middle: the nγ counts extracted using the confidence level cuts Pa

π < 0.01 andPb

γ > 0.1. Right: the counts nπ and nγ shown in the spectrum before any kinematic fit.

n(K∗+→ K+π0) estimates from Table 8.8 and Table 8.9 to obtain a new ratio calculated

again using Eq. 15.5,

RΛγ

Λπ=

Γ[Σ0(1385)→ Λγ]Γ[Σ0(1385)→ Λπ0]

= 1.42±0.12(stat)% (8.35)

The details of the calculation of R and statistical uncertainty are presented in

Appendix A. The systematic uncertainty for this method is reported in Section 9.

Page 175: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

175

8.3 Method-3

It is also possible to put a constraint on the missing particle in combination with the

proton and π− tracks. It is possible to kinematically fit each track of all detected particles

to a particular missing particle hypothesis, while requiring the proton, π−, and missing

particle be constrained to have an invariant mass of the Σ∗0. The fit is achieved by using

the undetected particle mass in the constraint equation, while the decay products from the

Λ in combination with the missing particle are confined to be the mass of the Σ∗0 for each

hypothesis. The detected particle tracks are kinematically fit as a final stage of analysis

and again filtered with the confidence level cuts. In this fit there are three unknowns (~px),

five constraint equations from conservation of momentum, and the additional invariant

mass condition similar to method-2. This type of fit is also a (2-C) kinematic fit. The

constraint equations are

F =

(Eπ +Ep +Ex)2− (~pπ +~pp +~px)2−M2

Σ∗0

Ebeam +Mp−EK−Ep−Eπ −Ex

~pbeam−~pK−~pp−~pπ −~px

=~0. (8.36)

Again ~Px and Ex represent the momentum vector and energy for the missing particle in the

hypothesis.

8.3.1 Results

The same two-step kinematic fitting procedure is used to resolve the radiative signal

from the π0. The χ2 distribution should again follow the form of the distribution for two

degrees of freedom. This method not only restricts the proton and π− in the constraint, but

also effectively uses the additional restriction of the missing energy to achieve a

separation of the two topologies. The disadvantage of this method is that the three part

invariant mass of Σ∗0 should be a 36 MeV wide Breit Wigner shaped peak which means

that an invariant mass from poorly measured events can look the same as a well measured

Page 176: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

176

event in the tail of the peak. In other words the invariant mass constraint is sensitive only

to detector resolutions and not peak width. All events from the outer edges of a wide peak

receive a χ2 low weighting in the probability distribution of the measured events in the

kinematic fit. The additional restriction ultimately makes separation of signal and

background more challenging by removing the freedom in the three invariant mass to be

the form of a resonance. This can be seen in the confidence level distribution of the π0

channel in Figure 8.15.

Figure 8.15: The confidence level distribution with the additional constraint on the invariantmass of the Σ∗0 using a missing π0 hypothesis.

Notice that the confidence level distribution in Figure 8.15 is no longer flat around

10%. This means that a cut at this point would have an indeterminate systematic error

introduced. In fact the distribution does not flatten out until much higher in the

distribution.

Instead of proceeding with a cut on the distribution in Figure 8.15 the previous

method is used in the initial step of the two step kinematic fitting procedure. By using the

kinematic fit to a missing π0 with only the Λ constraint it is possible to use the previous

initial cut of 1% to filter out the π0 background. This allows the radiative signal to pass

Page 177: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

177

through to be tested with the Σ∗0 invariant mass constraint. The χ2 distribution and

confidence level distribution for the second kinematic fit are shown in Figure 8.16.

Figure 8.16: The χ2 and confidence level distribution with the additional constraint on theinvariant mass of the Σ∗0 with the missing γ hypothesis.

Figure 8.16 indicates the correct general shape for both distributions, however the

same issue arises: the confidence level distribution is not flat in the lower region. In

addition, the form of the χ2 distribution would be very difficult to investigate due to low

statistics. There is no direct way to optimize with the choice of confidence level cuts in

this method. To obtain a result the confidence level cuts are taken from Method 2, using

the Σ∗0 invariant mass constraint only with the missing γ hypothesis.

The ratio achieved is

RΛγ

Λπ=

Γ[Σ0(1385)→ Λγ]Γ[Σ0(1385)→ Λπ0]

= 1.37±0.20(stat)%. (8.37)

This result is with Pγ = Pπ0 < 1% on the initial confidence level cut and Pγ = Pπ0 >

10% on the the second cut.

Page 178: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

178

9 SYSTEMATIC STUDIES

Method-2 is selected for a set of systematic studies. The value of each cut is varied to

study the effect on the final acceptance corrected ratio. For each variation the new

acceptance terms in Equation 15.5 are recalculated with the corresponding Monte Carlo.

Each major systematic uncertainty contribution is collected together in a table at the end

of this section. Each term is numbered as it is discussed to tie the discussion to the

corresponding term listed Table 9.3.

Several ∆β cut variations are checked starting with ∆β < 0.02 for all charged

particles leading to a branching ratio of 1.49±0.15%. There is also a check at ∆β < 0.1

giving a ratio of 1.38±0.11%. The ∆β selected is using a 1 ns timing cut while keeping

the ∆β < 0.02 for the kaon but a looser cut of ∆β < 0.05 for the proton and π−. This

variation is presented at the top of Table 9.3 as number (1).

To clean up the selected kaon events there is a cut made on identified kaon candidates

that are truly π+. The missing mass squared is studied for the reaction γ p→ π+π−(X).

This contamination is easily removed with a cut slightly above zero but below the mass

region of the π0. A cut of 0.01 GeV2 was used with the intention of not cutting into the

good K+ events. This cut is seen in Figure 5.1. The ratio was calculated with and without

this cut, and there is no measurable difference in the resulting branching ratio or statistical

uncertainty. The ratio with and without this cut is presented in Table 9.3 as number (2).

The distance of closest approach cut for the proton and π− is varied from a cut of

DOCA < 1 cm to DOCA < 20 cm and it is found that there is some variation, however the

statistical error increases significantly for smaller cut values. The cut at DOCA < 5 cm

marks the beginning of the most stable region in Figure 9.1. The various DOCA cuts

shown in Figure 9.1 are in centimeters while the acceptance corrected branching ratio is

shown in (%). The cut at DOCA < 5 discussed in the analysis is preserved and used in the

Page 179: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

179

final quoted ratio. The variation based on the DOCA cut is presented in Table 9.3 as

number (3).

Figure 9.1: The variation in the DOCA cuts in centimeters with the acceptance correctedbranching ratio shown in (%).

The perpendicular momentum cut is required in the analysis due to its exact

topological similarity to the radiative signal. To find the most efficient Pxy cut, the ratio is

calculated at various cut values until the ratio stabilizes. The lowest stable cut is selected

to preserve statistics. The final cut is the cut reported in the analysis previously at

Pxy < 0.03 GeV. Figure 9.2 shows the acceptance corrected ratio for various Pxy cuts with

a flat line fit to the stable points used to obtain the uncertainty in the Pxy cut. This variation

is presented as number (4).

Page 180: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

180

Figure 9.2: The variation in the perpendicular momentum cuts with the acceptancecorrected branching ratio shown in (%).

There is no direct way to estimate the systematic uncertainty in the Monte Carlo’s

capacity to match the data. However, it is possible to check the affects of various tuning

parameters to see the variation in the branching ratio. The systematic effects of the Monte

Carlo are studied by making adjustments to the generator, producing a new simulation and

then re-evaluating the acceptance terms. It is possible to check variation in the ratio (based

on the new acceptance) by modifying only the generator to match the production cross

section seen in the photon energy distribution from the data and then checking various

exponential t-dependences. Prior to using the approximated differential cross section, the

best data matching t-slope value in the Monte Carlo is found to be 0.65 GeV. To study the

generator’s effect on the ratio a t-slope in the range of 20% lower and 20% higher than the

Page 181: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

181

ideal matching value of 0.65 GeV is selected to test the effects on the resulting acceptance

corrected branching ratio. The change in t-slope is done simultaneously for the radiative

and π0 channels. The acceptance for both channels are recalculated and the corresponding

ratio at various t-slopes is plotted in Figure 9.3. There is very little variation seen. The

branching ratio using a t-slope of 0.65 is 1.422%, and using the differential cross sections

to match the data, the branching ratio is 1.421%. In the final ratio reported, the

modification in the generator used to match the differential cross sections is implemented

after the production cross section modification to the photon energy distribution. The

t-slope is implemented only in these variation studies. This is listed as number (5) in the

table.

Figure 9.3: Left: Variation in the acceptance corrected branching ratio for various t-slopesetting in the generator for the π0 channel; right: for the radiative channel.

Page 182: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

182

To demonstrate the effect of each acceptance term on the final branching ratio, all

Monte Carlo channels that are included in the branching ratio are tested by setting the

yield for the contamination found for that channel to zero in Eq. 15.5. This is done for

every included contamination one at a time. The result for each channel is shown in Table

9.1. The branching ratio with all channels included is 1.420%.

Table 9.1: Branching ratio for excluded channels in (%).

Counts Excluded Branching Ratio

Λ(1405) : N(Σ0π0) 1.448±0.116

Λ(1405) : N(Σ+π−) 1.424±0.116

Λ(1405) : N(Λγ) 1.446±0.116

Λ(1405) : N(Σ0γ) 1.422±0.116

Σ(1385) : N(Σ+π−) 1.425±0.116

Σ(1385) : N(Σ0γ) 1.421±0.116

Each term is Table 9.1 is not separately included in the systematic uncertainty

calculation. Instead terms (6) and (7) in Table 9.3 show the branching ratio with and

without the total background subtraction taken into account for the Λ(1405) and the K∗

respectively. These variation are found by simply setting nΛ = 0 or n(K∗+→ K+π0) = 0

and comparing the resulting ratio with the nΛ and n(K∗+→ K+π0) previously extracted.

Prior analysis using the g11a data set shows that there are substantial inefficiencies in

the production trigger used to collect the data [72]. The inefficiencies are deduced to be

caused by inefficiencies in the TOF array and timing in the start counter in the trigger.

This leads to sector, θ , φ and particle dependent efficiencies in the trigger that need to be

accounted for when extracting finite cross sections from the data. A study was preformed

Page 183: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

183

to test the impact on the ratio due to the trigger efficiency corrections to the acceptance

from the Monte Carlo. The change seen in the ratio is of the order of ∼ 0.05% of the ratio.

This is a negligible uncertainty relative to the other uncertainty discussed in the Monte

Carlo.

To look at the systematic dependency on the choice of the confidence level cuts, the

range defined by the σγ [71] uncertainty for Pbπ (χ2) should be checked. As previously

described the Monte Carlo studies leads to the set of optimal Paπ0 cuts for a given Pb

γ , as

seen in Table 8.6. To maximize statistics and minimize δR the optimal cuts chosen for the

analysis were Pbγ > 1% and Pb

γ > 10%. The variation in the branching ratio is studied by

selecting the confidence level cuts that lie slightly outside the optimization region in

Figure 27 in the CLAS note [71]. In this way the largest range for Pbγ and Pb

γ can be tested

while still respecting the information derived in the optimization map. This leads to a

systematic uncertainty that can be assigned to the resolving procedure used in method-2.

Table 9.2 lists the cut defined by the Monte Carlo derived optimization map in CLAS note

[71]. Three points in Paπ are used in the optimization region for the Pb

γ > 10% cut. There

are then two cuts used for both Pbγ > 5% and Pb

γ > 15% that lie just outside the

optimization region.

Using the full range of ratios in Table 9.2, the largest and smallest values are used to

find the deviation of the ratio defined by the new set of confidence level cuts and the

quoted ratio R = 1.42% in method-2. This deviation is used as the uncertainty in the

resolving procedure and added in with the other systematic uncertainties. The variation

from the choice of Confidence level cut is listed in Table 9.3 as (8).

The exponential t-slope is also varied for the Λ(1405) and found to have negligible

effects on the branching ratio. The variation in t-slope tested is 0-2 GeV2, resulting in a

change of less than 0.1% of the branching ratio.

Page 184: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

184

Table 9.2: Dependence of the corrected branching ratio on the confidence level cuts for theselected systematic range.

Pbγ (%) Pa

π (%) R(%)

15 7.5 1.388± 0.12

15 5 1.390± 0.12

10 5 1.422± 0.12

10 1 1.420± 0.12

10 0.5 1.421± 0.12

5 0.1 1.448± 0.12

5 0.05 1.436± 0.12

The branching ratio of the electromagnetic decays of the Λ(1405) are not known.

There has been experimental work by Burkhardt and Lowe [74] which are the values used

in the present analysis of 27 keV for the Λ(1405)→ Λγ channel and 10 keV for the

Λ(1405)→ Σ0γ . These values are not included in the PDG official listing. The

Λ(1405)→ Σ0γ has such small acceptance that little effect is seen in the ratio. However,

the Λ(1405)→ Λγ acceptance can lead to a noticeable affect on the ratio.

There is some work by Workman and Fearing [76], which is a very similar analysis

to Burkhardt and Lowe, but before the experiment on kaonic atoms had a result. They also

find an interference between Born and resonance diagrams, giving a value of 23 keV for

the EM decay width to Λ(1405)→ Λγ .

The Chiral bag model work of Umino and Myhrer [77], presented a “cloudy bag”

model, where the bag parameters have been tuned using other data (mass spectrum of the

negative parity hyperons). They also find an interference (cancellation), which makes a

smaller EM decay width, but from a different set of diagrams. They get 75 keV for the

EM decay to Λ(1405)→ Λγ .

Page 185: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

185

Other work by Darewych, Horbatsch and Koniuk, [78], use a straight-forward mixing

model based on the SU(6) wave-functions. No interference between diagrams is found,

leading to a prediction of 143 keV for the EM decay to Λ(1405)→ Λγ . Using this width

in the calculation of the present analysis would lead to a ratio of 1.33±0.12.

The width from the Darewych calculation is large in comparison to all other models,

and especially large compared with the Burkhardt and Lowe result, so it is excluded from

the systematic studies. This leads to the largest width of 75 keV which leads to a ratio in

the present analysis of 1.39±0.12. This result is included in line (9) in Table 9.3.

Table 9.3 shows a summary of the systematic studies and the higher and lower range

of the ratio based on the variations mentioned for each type of uncertainty.

Table 9.3: Ranges of systematic variation in resulting ratio in (%) showing L(Low)-contribution and H(High)-contribution and rang in each case.

Type of Error Low Range L-Contribution High range H-Contribution

(1)∆β 1.380±0.12 -0.040 1.490± 0.15 +0.070

(2)Miss ID 1.420±0.12 -0.000 1.422±0.12 +0.002

(3)DOCA 1.350±0.12 -0.007 1.480± 0.12 +0.060

(4)Pxy 1.415±0.12 -0.005 1.433± 0.12 +0.013

(5)t-slop 1.380±0.12 -0.040 1.440± 0.12 +0.020

(6)Λ∗-BG 1.420±0.12 -0.000 1.470± 0.12 +0.050

(7)K∗-BG 1.420±0.12 -0.000 1.431± 0.12 +0.011

(8)P(χ2) cuts 1.388±0.12 -0.032 1.448± 0.12 +0.028

(9)Λγ cuts 1.390±0.12 -0.030 1.420± 0.12 +0.000

Total Uncertainty -0.072 +0.112

Page 186: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

186

To calculate the final systematic uncertainty for method-2, the difference in the ratio

R = 1.42 and the high range of the ratio for each case in Table 9.3 is added in quadrature

to obtain a value for the uncertainty of 0.11 greater than the ratio. The lower systematic

uncertainty is based on the difference between the ratio R = 1.42 and the low range of the

ratio for each case, resulting in a value of 0.07 less than the ratio.

9.1 Conclusion

Because of the increase in the statistical uncertainty in method-3, the branching ratio

reported in method-2 of 1.42±0.12(stat)+0.11−0.07(sys) is favored. Previously published work

on this channel yielded a ratio of 1.53±0.39(stat)+0.15−0.17 [10]. The value in method-2 is

consistent within uncertainties.

Page 187: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

187

Part II

Electromagnetic decay of the Σ∗+

Page 188: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

188

10 EVENT SELECTION

Using the g11a data set, events are selected for the channel γ p→ K0Σ∗+. The present

PDG branching ratios lists the decay Σ∗→ Σπ to be 11.7±1.5% and assuming isospin

symmetry this leads to a branching ratio of 5.85±0.75% for the Σ∗+→ Σ+π0 decay [60].

This channel will be used to normalized the radiative signal that comes from the channel

Σ∗+→ Σ+γ . The radiative signal is assumed to be quite small, in the range of 0.1−0.5%

of the full Σ∗ width. However, there have yet to be any measurements. For both channels

the topology of the decay is γ p→ K0Σ+(X) where X is not directly measured such that

the π0 and γ are differentiated using conservation of energy and momentum. This

topology leads to the final set of decay products γ p→ K0Σ+(X)→ π+π−π+n(X). The

charged particles can easily be detected with the use of the CLAS drift chambers and

Time-of-flight. The neutron must be detected with the CLAS electromagnetic calorimeter.

The analysis was done using the previously skimmed g11a data set for two positively

charged tracks and one negatively charged track with no other skim conditions.

The previously listed CLAS runs form Part-1 of this thesis were used for the present

analysis. The same energy loss, tagger corrections, and momentum corrections were also

used. The same beam photon selection and detector performance cuts for all charged

particle were used as previously described. The same timing cut of 1 ns is used. All good

photons are kept until the final cuts are implemented and then a check is done to ensure

only one photon survives.

Only pions above 0.125 GeV/c momentum are studied. The beam photon is required

to be between 1.5 and 3.8 GeV.

Page 189: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

189

10.1 Particle Identification

Figure 10.1: The ∆β distribution for the π− and the cut applied to clean up identification.

Again the PART bank is used to make the initial selection of the charged particles.

Using the time of flight from the event vertex to the scintillator counter, β is calculated for

each pion. The ∆β can then be calculated using the difference between the time-of-flight

β and calculated β . Figure 10.1 shows an example of the ∆β distribution for the π− and

the cut used. The same ∆β < 0.025 is cut used for all pions.

In the reaction of interest, γ p→ K0Σ∗+(X), it is necessary to determine which π+

comes from the K0. It is possible to check both π+ with the detected π− to see the kaon

candidates in each case by using the invariant mass. A preliminary invariant mass plot for

each combination is shown is Figure 10.2.

Page 190: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

190

Figure 10.2: Invariant mass of the π+-π− combination for the two different π+ detected,prior to any π+ organization to optimize the K0 cut.

Figure 10.3: Invariant mass of the π+-π− combination for the two different π+ after π+

organization to optimize the K0 peak.

Page 191: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

191

Figure 10.4: The confidence level distribution for the (2-C) kinematic fit with constraint onπ+ and π− to be the mass of the K0 with a missing mass off the K0 of the Σ∗+.

After the K0 selection, a kinematic fit can be used on just the final K0 pions with the

constraint as the invariant mass of the K0. This ensures a quality K0 for a certain

confidence level cut. Prior to the kinematic fit for each event the invariant mass is checked

for each π+ pair. Whichever π+ leads to the invariant mass that is closest to the PDG

mass for the K0 is used in the kinematic fit. Figure 10.3 shows the invariant mass plot for

each combination after the π+ selection.

A (2-C) fit is used with only the selected π+ and π− tracks and an additional

constraint of the missing mass of the Σ∗+. This allows systematic control on the quality of

the K0 and Σ∗+ identification through a single confidence level cut. It is shown in a CLAS

Note [71] that using resonances in missing mass constraints in a kinematic fit can lead to a

significant loss of counts around the tail. Wide resonances can suffer greatly from this

type of fit. However, these types of cuts are only problematic when studying structure

dependent physics such as cross sections. Because the study presented here is on a

branching ratio, the preservation of the Σ∗+ shape is not required. In other words the Σ∗+

counts in the tail of the peak will be lost in the both the numerator and denominator. The

Page 192: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

192

kinematic fit then helps to ensure only good quality Σ∗+ events survive. An initial

confidence level cut of P(χ2) > 0.1% is chosen, which is quite small, to let in as many

reasonable counts as possible. The systematic variation of this cut is then studied in

Section 15.3. Figure 10.4 shows the confidence level distribution from the kinematic fit

with constraint on π+ and π− to be the mass of the K0 with a missing mass constraint off

the K0 to be the Σ∗+ mass. The distribution is not expected to be flat since there is a lot of

background present. Figure 10.5 shows the invariant mass of the π+, π− combination

before and after the P(χ2) > 0.1%confidence level cut. Figure 10.6 shows mass of the π+,

π− combination before and after the P(χ2) > 0.1% confidence level cut.

(a) (b)

Figure 10.5: (a) The invariant mass of the π+ π− after the best π+ is selected. (b) Theinvariant mass of the π+ and π− after the P(χ2) > 0.1% confidence level cut.

Page 193: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

193

(a) (b)

Figure 10.6: (a) The missing mass off the K0 before the confidence level cut. (b) Themissing mass off the K0 after the P(χ2) > 0.1% confidence level cut.

Page 194: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

194

11 NEUTRON IDENTIFICATION

Neutral particles are detected in CLAS as clusters in the EC not associated with any

reconstructed charged track from the drift chambers. Because the g11a run period was not

intended for neutral detection, precision neutron momentum reconstruction requires some

additional steps. The EC timing calibration was done using fast pions such that β ∼ 1

using the TOF as a reference time. This is far from ideal for neutrons where β < 1. The

directional components of the neutral track are found by using the neutron vertex and the

cluster position on the EC for that hit. The momentum of the neutral particle is calculated

from the EC time-of-flight and the path length from the neutron vertex to the cluster

position. The neutrons are differentiated from photons using a β < 0.9 cut eliminating fast

neutrals.

The neutron detection is essential for the reaction of interest. The neutron momentum

can be obtained and then used in combination with the π+ to study the kinematics of the

Σ+. Having clean constraints on the K0 and Σ+ is important when considering the event

topology γ p→ K0Σ∗+→ K0Σ+(X).

A thorough study of the accuracy of the EC for neutron momentum reconstruction in

all kinematic ranges has not been achieved previously; it is an essential part of the present

analysis. Correlations between each measured variable in the EC have also not been

studied well. The covariance of the neutron can give a lot of information about the quality

of the kinematic variables in various areas of the EC. These values can then be used to

weight the neutron measurements appropriately in kinematic constraints that depend on

maximum likelihood methods, see Section 13.

There are resolution changes that are related to the acceptance of the EC. Hits from

the center of each triangular sector have better measurements over those on the edges.

There are also effects from the torus coils and forward angles to consider. The EC consists

Page 195: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

195

of two main blocks. The inner and outer blocks can have different resolution parameters

for each measured variable.

Finally, there are some kinematic ranges that have better resolution than others due to

geometry and the phase space of the reaction. All possible combinations of momentum

and position are studied to develop a complete understanding of the neutron variance and

covariance in the electromagnetic calorimeter.

11.1 Neutron Detection Test

The test reaction γ p→ π+π−π+n is isolated in the g11a data set by selecting a π−

and two π+ and kinematically fitting to a missing neutron hypothesis and then taking a

10% confidence level cut. This channel is selected because the final decay products are

identical to the reaction of interest γ p→ K0Σ∗+→ K0Σ+(X). In addition the momentum

range of the detected particles is the same. The simplification made by working with the

test channel is that in the γ p→ π+π−π+n reaction, there is only one interaction vertex.

This implies that the neutron comes from the primary interaction vertex, which can be

easily determined using the charged pions.

It is possible to study the neutron momentum reconstruct and resolutions by requiring

each event to have one detected neutron and then compare this with the missing

momentum. Assuming a high quality missing neutron four vector, this procedure can be

used to find the dynamic resolution over the EC face.

The neutron vertex is found from the MVRT bank (see Appendix B), which uses a

multi-track-vertex fitting routine giving a very accurate vertex for multiple final state

particles, all coming from the same vertex [32]. Because the neutron comes from the

primary interaction vertex in this study, the vertex of the neutron is accurately know.

However, for any tracks in which the neutron comes from secondary vertices, this neutron

vertex is not a easily obtained. Because the neutron vertex information can effect the

Page 196: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

196

direction vector of the neutron and its momentum, these differences can be important

when studying resolutions.

11.2 Neutron Path

The distance that the neutron travels in CLAS is used with the time-of-flight to

determine the momentum of the neutron. The distance dependent on which EC block

layer and the position of the cluster reconstruction. This is determined using the ECHB

bank (see Appendix B) by looking at the integer value of matchid1 and matchid2. If

matchid2 is between 1 and 5 while matchid1 is zero, then the hit is only from the outer

block of the EC. The other hits, with matchid1 and matchid2 being between 0 and 5

(excluding 0 for matchid1), means that the hit was either in the inner block or both. If

there is a hit in both, the cluster reconstruction position in the inner block is used. If there

is a hit only in the outer EC block the first layer (layer closest to the target) of the outer

block gives the plane of the EC cluster coordinates. If there is a hit in the inner block or

both, the first layer of the inner block is used as the plane of the EC cluster coordinates.

The distance in which the neutron travels inside each block before a cluster is formed

cannot be determined, so there is naturally some uncertainty in path length associated with

each hit of the order of the width the EC block. Because of differences for the inner and

outer EC, the resolutions of each block are studied separately.

11.3 Neutron Time

The TOF for the neutron comes from the start time of that event to the EC cluster

time. Because the EC timing calibration was done using high β particles, there is

additional uncertainty introduced to the reconstructed neutron momentum associated with

the poor time-of-flight. The time-of-flight must be corrected to achieve an accurate

magnitude of the neutron momentum.

Page 197: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

197

11.4 Neutron Momentum Correction

It is possible to make the corrections needed for the neutron path and time-of-flight

by empirically adjusting the magnitude of the neutron momentum using a correction

function found by studying the trend in the momentum resolution over a kinematic range.

This is done by plotting ∆p = pk f it− pmeas against p for its measured range. The value of

pk f it is the magnitude of the missing neutron momentum from a (1-C) kinematic fit of the

pions detected in the drift chambers. The pmeas is the magnitude of the momentum

reconstructed using the EC cluster position and EC time-of-flight. The trend should be

evenly distributed around ∆p = 0. If it is not then the distribution will display a trend that

can be used to correct the neutron momentum. Once the neutron momentum resolution is

evenly distributed around zero the plot of the neutron detected momentum and the missing

momentum should be approximately one to one. This implies that for the majority of

events, the detected neutron momentum is comparable to the missing neutron momentum

from the kinematic fit. Figure 11.1 shows the magnitude of the momentum distribution of

∆p and the plot of pk f it vs pmeas for both EC block layers, where the momentum

correction has not yet been applied for the outer EC block. Figure 11.2 shows the same

distributions where the outer EC block momentum correction is now applied.

Page 198: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

198

(a) cosθ vs φ (missing n) (b) cosθ vs φ (detected n)

Figure 11.1: (a) ∆p before the correction, showing EC counts for inner and outer blocklayers. (b) The pk f it vs pmeas for EC counts for inner and outer block layers beforecorrections.

(a) cosθ vs φ (missing n) (b) cosθ vs φ (detected n)

Figure 11.2: (a) ∆p after correction showing EC counts for inner and outer block layerstogether. (b) The pk f it vs pmeas for EC counts for inner and outer block layers aftercorrections.

Page 199: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

199

To ensure a high quality missing neutron vector to study the corrections and

resolutions of the neutron, a 10% confidence level cut on the kinematic fit is used to obtain

pk f it . Only the detected neutrons found in a direction less than 3 from the missing

neutron are used. A cut in polar lab angle θ is also used between 10-40. Figure 11.3

shows the cosθ vs φ form the missing neutron vector from the kinematic fit. Figure 11.3

shows the same for the reconstructed neutron vector. This same procedure is repeated with

the Monte Carlo of the test channel.

Figure 11.3: Demonstration of cosθ vs φ for the neutron candidates.

Page 200: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

200

Figure 11.4: Demonstration of cosθ vs φ for the detected neutrons.

Page 201: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

201

11.5 Neutron Resolutions

The structure of CLAS prevents neutrons from being detected with the same

precision in all directions. The six superconduction coils in front of the EC limit the

capability of the calorimeter detection. The resolution for the neutron around these

blocked regions can be irregular. The EC neutron resolutions give an indication as to the

quality of the reconstructed four-vector for both Monte Carlo and data. The variance and

covariance will be used in the kinematic fitting procedure discussed in Section 13.

Again the missing neutron four-vector is used to study ∆p with respect to p, φ , and θ

as defined in the target frame. Figure 11.5 shows ∆p over a range of the momentum

magnitude, θ , φ and the path used in the momentum reconstruction for the g11a data.

A similar study of ∆p is done for Monte Carlo as well. Figure 11.6 shows ∆p over a

range of the momentum magnitude, θ , φ and the path used in the momentum

reconstruction for simulations.

Once ∆p is obtained for data and Monte Carlo, the distributions with respect to p, θ

and φ are sliced in binned in each dynamic variable. Each bin is projected on to the ∆p

axis and fit with a Gaussian to find the σp resolution for that bin. The Gaussian mean for

each fit is also checked to be centered around zero implying good neutron reconstruction

and correction. Figures 11.7-11.9 shows the results of the resolution extraction for the

g11a data. Figures 11.10-11.12 shows the results of the resolution extraction for the

Monte Carlo. In each case the upper left plot shows the number of events per bin used.

The upper right plot shows the mean from the Gaussian fits. The lower left plot shows the

values of σ from the Gaussian fits over the required a dynamical range. The lower right

plot shows the χ2 from each Gaussian fit for every bin used.

Changes in the momentum resolution also occur in the directional components and

can be studied in a similar manner. A ∆θ and ∆φ can also be obtained using the deviation

of the detected neutron direction from the missing neutron direction. Once ∆θ (∆φ ) is

Page 202: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

202

obtained for data and Monte Carlo, the distributions with respect to p, θ and φ are sliced

and binned in each dynamic variable. Each bin is projected on to the ∆θ (the ∆φ axis and

fit with a Gaussian to find the σθ (σφ ) resolution for that bin). Figures 11.13 (11.14) show

the results of the resolution extraction for the g11a data. Figures 11.15 (11.16) show the

results of the resolution extraction for the Monte Carlo.

Figure 11.5: The g11a data set is used to obtain ∆p over a range of p (upper left), ∆p overa range of φ (upper right), ∆p over a range of θ (lower left), ∆p over a range of nuetronpath length in (cm) (lower right).

Page 203: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

203

Figure 11.6: Monte Carlo is used to obtain ∆p over a range of p (upper left), ∆p over arange of φ (upper right), ∆p over a range of θ (lower left), ∆p over a range of neutron pathlength in (cm) (lower right).

Page 204: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

204

Figure 11.7: The g11a study of σ(p). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lower left plotshows the values of σ from the Gaussian fits over the required range of p. The lower rightplot shows the χ2 from each Gaussian fit for every bin used.

Page 205: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

205

Figure 11.8: The g11a study of σ(θ). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lower left plotshows the values of σ from the Gaussian fits over the required range of θ . The lower rightplot shows the χ2 from each Gaussian fit for every bin used.

Page 206: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

206

Figure 11.9: The g11a study of σ(φ). The upper left plot shows the number of events perbin used. The upper right plot show the mean from the Gaussian fits. The lower left plotshows the values of σ from the Gaussian fits over the required range of φ . The lower rightplot shows the χ2 from each Gaussian fit for every bin used.

Page 207: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

207

Figure 11.10: The Monte Carlo study of σ(p). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required range of p. Thelower right plot shows the χ2 from each Gaussian fit for every bin used.

Page 208: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

208

Figure 11.11: The Monte Carlo study of σ(θ). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required range of θ . Thelower right plot shows the χ2 from each Gaussian fit for every bin used.

Page 209: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

209

Figure 11.12: The Monte Carlo study of σ(φ). The upper left plot shows the number ofevents per bin used. The upper right plot show the mean from the Gaussian fits. The lowerleft plot shows the values of σ from the Gaussian fits over the required range of φ . Thelower right plot shows the χ2 from each Gaussian fit for every bin used.

Page 210: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

210

Figure 11.13: The g11a data study of σ(θ). The upper left plot shows ∆θ over a range ofmomentum. The upper left shows the resolutions from the Gaussian fits of ∆θ as a functionof momentum. The middle right shows the ∆θ over a range of φ . The middle left showsthe resolutions from the Gaussian fits of ∆θ as a function of φ . The bottom left plot shows∆θ over a range of θ . The bottom left shows the resolutions from the Gaussian fits of ∆θ

as a function of θ . A cut on the cosine of the angle between the missing neutron vector andthe detected neutron vector of 3 is used.

Page 211: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

211

Figure 11.14: The g11a data study of σ(φ). The upper left plot shows ∆φ over a range ofmomentum. The upper left shows the resolutions from the Gaussian fits of ∆φ as a functionof momentum. The middle right shows the ∆φ over a range of φ . The middle left showsthe resolutions from the Gaussian fits of ∆φ as a function of φ . The bottom left plot shows∆φ over a range of θ . The bottom left shows the resolutions from the Gaussian fits of ∆φ

as a function of θ . A cut on the cosine of the angle between the missing neutron vector andthe detected neutron vector of 3 is used.

Page 212: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

212

Figure 11.15: The Monte Carlo study of σ(θ). The upper left plot shows ∆θ over a rangeof momentum. The upper left shows the resolutions from the Gaussian fits of ∆θ as afunction of momentum. The middle right shows the ∆θ over a range of φ . The middle leftshows the resolutions from the Gaussian fits of ∆θ as a function of φ . The bottom left plotshows ∆θ over a range of θ . The bottom left shows the resolutions from the Gaussian fitsof ∆θ as a function of θ . A cut on the cosine of the angle between the missing neutronvector and the detected neutron vector of 3 is used.

Page 213: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

213

Figure 11.16: The Monte Carlo study of σ(φ). The upper left plot shows ∆φ over a rangeof momentum. The upper left shows the resolutions from the Gaussian fits of ∆φ as afunction of momentum. The middle right shows the ∆φ over a range of φ . The middle leftshows the resolutions from the Gaussian fits of ∆φ as a function of φ . The bottom left plotshows ∆φ over a range of θ . The bottom left shows the resolutions from the Gaussian fitsof ∆φ as a function of θ . A cut on the cosine of the angle between the missing neutronvector and the detected neutron vector of 3 is used.

Page 214: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

214

Each Gaussian mean from these studies is centered around zero, which indicates the

corrections used on the neutron momentum are working reasonably well. A non-zero

mean would indicate an unreliable resolution measurement. The resolutions σ seen for

data match the trend and magnitude of those seen in the Monte Carlo. This is a

confirmation that the corrections for the Monte Carlo are working well and that the trends

seen in the σ distribution are indeed based on geometric factors. The resolutions also

match the data, which is a confirmation that the neutron in the Monte Carlo is well

represented and no additional smearing to the neutron is required.

11.6 Summary

The g11a run period was not intended for neutral detection leading to a momentum

correction giving a more precise neutron momentum. A study of the accuracy of the EC

for neutron momentum reconstruction with kinematic dependence has been done. Some

kinematic ranges have better resolution than others due to geometry and the phase space

of the reaction. All possible combinations of momentum and position have been studied to

develop a complete understanding of the neutron variance and covariance in the

electromagnetic calorimeter. The test reaction γ p→ π+π−π+n is isolated in the g11a

data set by selecting a π− and two π+ and then kinematically fitting to a missing neutron

hypothesis with a 10% confidence level cut. The resulting residuals are studied by slicing

each distribution and fitting the bins with a Gaussian. Each Gaussian mean from these

studies is centered around zero, as expected. The resolutions σ seen for the data match

those seen in the Monte Carlo, so no additional smearing to the momentum is required.

Page 215: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

215

12 ANALYSIS PROCEDURE

In the following analysis, an attempt is made to remove as much identifiable

background as possible while preserving the counts from the channels

γ p→ K0Σ∗+→ K0Σ+(π0) and γ p→ K0Σ∗+→ K0Σ+(γ). As in Part 1, the radiative

signal is buried by the π0→ γγ decay. The radiative signal extraction again requires a

kinematic fitting procedure. This requires detection of all other particles to differentiate

between the missing π0 and missing γ . This can be done with a kinematic fitting

procedure using the neutron covariance in the EC along with the standard DC covariance

for the detected pions.

The immediate goal is then to achieve clean hadron identification before using the

kinematic fitting procedure for the competing π0 and radiative signal. The best way to

achieve a clean K0 and Σ∗+ identification is with the initial kinematic fit using two pion

tracks to a K0 hypothesis with total missing mass of the Σ∗+. Before proceeding with this

step in the analysis, a preliminary look at the mass spectrum is required to check that the

neutron provides useful kinematics constraints. This check is three fold. First, it allows

testing of the corrected neutron momentum. Second, the final distribution of the missing

mass squared of all particles detected can be compared with the results of the initial

kinematic fit (to the K0 mass). Third, the distribution can be compared with the Monte

Carlo distribution, in order to identify background leakage.

For the sake of notation, let π+1 indicate the π+ used in the K0 invariant mass

selection. Hence π+2 is the other π+. For an initial investigation of the hyperon

excited-state region, ∼ 25% of the g11a is used. Figure 12.1 shows the invariant mass of

the π+1 -π−, (upper left), missing mass off the π

+1 -π− (upper right), the n-π+

2 invariant

mass (lower left), and the missing mass squared of all the detected particles (lower right).

The distributions in Figure 12.1 are before any kinematic constraints and after the π+

organization is done. The K0 is cut on at ±0.005 GeV of the PDG K0 mass to eliminate

Page 216: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

216

the π+-π− background and the spectra are plotted again. Figure 12.2 shows the invariant

mass of the n-π+2 , (upper left), missing mass off the K0 (upper right), the missing mass off

the n-π+2 combination (lower left), and the missing mass squared of all the detected

particles (lower right), after the K0 cut. Clearly the Σ+ peak is visible and in the correct

place shown in the (upper left) plot. The clear visible peak in the missing mass squared

where the π0 should be is also an indication that the neutron measurement is effective. A

cut on the invariant mass of the n-π+2 combination cleans up the excited-state hyperons,

making the Σ∗+ quite visible. Figure 12.3 shows the missing mass off the K0 (upper left),

and the missing mass off the Σ+ (upper right) after a ±0.05 cut around the Σ∗+ peak. The

lower plots are the missing mass of all detected particles and a magnified Gaussian fit to

the π0 peak.

Finally, Figure 12.4 shows a comparison between the final missing mass squared

distribution from the set of cuts described with a ±0.05 cut around the Σ∗+ (upper left)

compared with a P(χ2) > 0.1% confidence level cut used to obtain the K0 and Σ∗+

candidates (upper right). The same ±0.02 cut around the Σ+ peak is used in both. Also

shown is the missing energy distribution in the Σ∗+ frame, and the P2xy distribution. The

Pxy is shown only as a check against double bremsstrahlung. Clearly there is no peak near

zero, so no further check is required. This final comparison between the two missing mass

squared distributions give a consistency check. In principle the kinematic fit has a well

define confidence level associated with the candidates that survive the cuts. Hence, the

kinematic fit will be used for the rest of the analysis.

The distributions in Figure 12.1-12.4 can be compared with simulations. After

producing the Monte Carlo for each channel, the structure of the distribution in each case

were compared with the data distributions to determine which channels are relevant.

Figure 12.5 shows the missing mass off the Σ+. In the case of the

γ p→ K0Σ∗+→ K0Σ+(π0) and γ p→ K0Σ∗+→ K0Σ+(γ) channels, there should be no

Page 217: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

217

definite peak or structure present. However, a reaction like γ p→ K∗0Σ+→ K0Σ+(π0)

would show a peak at the K∗ mass.

The distribution in Figure 12.1-12.4 and the result from the constraint on the π+2 -n

indicates that the neutron detection and the corrections to its momentum are working.

From Figure 12.4, it is not possible to resolve the radiative single from a standard fitting

technique. The procedure of kinematically fitting of all the detected particles will be used

to again make this separation. The necessary neutron resolutions have already been

outlined in Section 11.5. Now it is possible to use these resolutions to build a full

covariance matrix for this analysis.

Page 218: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

218

Figure 12.1: The invariant mass of the π+1 -π−, (upper left), missing mass off the π

+1 -π−

(upper right), the n-π+2 invariant mass (lower left), and the missing mass squared of all the

detected particles (lower right). All distributions are before any kinematic constraints.

Page 219: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

219

Figure 12.2: The invariant mass of the n-π+2 , (upper left), missing mass off the K0 (upper

right), the missing mass off the n-π+2 combination (lower left), and the missing mass

squared of all the detected particles (lower right), after the ±0.005 K0 peak.

Page 220: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

220

Figure 12.3: The missing mass off the K0 (upper left), and the missing mass off theΣ+ (upper right), the missing mass squared of all detected particles (lower left), and amagnification of the missing mass squared with Gaussian fit to the π0 region (lower right) .These distribution are made after take a ±0.02 cut around the Σ+ peak as well as the priorcut of ±0.005 on the K0 peak.

Page 221: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

221

Figure 12.4: The comparison between the final missing mass squared distribution fromthe set of cuts discribed after a ±0.05 cut around the Σ∗+ mass (upper left) and from theP(χ2) > 0.1% confidence level cut used obtain the previous K0 and Σ∗+ candidates. Thesame ±0.02 cut around the Σ+ peak is used in both.

Page 222: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

222

Figure 12.5: The missing mass off the Σ+ after all cuts described (no kinematic fit). Thedistribution is similar for the the kinematic fit (not shown).

Page 223: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

223

13 NEUTRON KINEMATIC FITTING

Neutral particles are identified in CLAS as clusters in the Electromagnetic

Calorimeter (EC) that are not associated with any charged track reconstructed from the

drift chambers. It is then possible to differentiate between photons and neutrons using the

particle velocity β . Photons will naturally be very close to β = 1. If the resolution of the

neutral particle is well understood it can be used to develop a covariance matrix for that

particle for use in a kinematic fitting procedure. The covariance matrix for photons and

neutrons in the EC is quite different; however, the standard (target frame) CLAS variables

can be used for both.

13.1 Neutron EC Covariance Matrix

The neutron covariance matrix is developed using the neutron resolutions in p, θ , and

φ as outlined in Section 11.5. Because the full covariance matrix has already been done

for the charged particles it is only the neutron EC submatrix that is required to purse this

technique. The neutron subcovariance matrix can be expressed in the same coordinates;

however, the neutron momentum is weakly correlated with the other measured variables

leading to a matrix of the form,

Vni =

V pp

i V pθ

i V pφ

i

V pθ

i V θθi V φθ

i

V pφ

i V φθ

i V φφ

i

. (13.1)

The momentum resolution for the the neutron depends on how the neutron momentum

was achieved. The term V pp is strictly depends on the velocity β of the neutron and the

calibration of the EC with regard to the EC time of flight (EC time). For many

photoproduction experiments with CLAS, the electromagnetic calorimeter is calibrated

using fast pions. This leads to poor time resolution for neutrals. For neutrons this can

make a big difference in momentum resolution. There is usually a correction to neutron

Page 224: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

224

path and EC time of flight to get accurate momentum reconstruction. The inner and outer

layers can have different neutron momentum resolutions and must also be taken into

account.

The position and angular resolution for neutrons have dynamic dependencies on

momentum that can be studied using a prominent channel for a particular data set, as well

as Monte Carlo. Studying the various measured variable resolutions for all EC sector at

various θ , φ , and p dependence leads to functional relations that can be used as error

estimates to fill in the angular elements of the subcovariance matrix.

Once the covariance matrix is constructed and tested for all particle types, the

kinematic fitting procedure can be used.

13.1.1 Diagonal Terms

There is often dynamic dependence on the various measured parameters as well a

particle dependence for each resolution term. The residual quantities δvi in Vii should be

accurately depicted for all kinematic ranges and experimental configuration changes.

When investigating resolutions it is necessary to break up the study into all variables and

as many kinematic ranges as possible and analyze the effects on

δv = vtrue− vmeas,

which is the difference in the measured quantity vmeas and the true vtrue, known quantity

for that variable. With simulations it is relatively easy to extract the vtrue from what is

generated. Experimentally it is necessary to use the other known variables to extract the

vtrue. One example is to use a missing 4-vector with all other decay products detected. In

the case of neutrons it is possible to use a kinematic fit to the missing neutron using the

other charged decay products in the drift chambers that already have well established

resolution parameters. This gives a true value for the position and momentum of the

Page 225: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

225

neutron that can be compared to the reconstructed values from the Electromagnetic

Calorimeter.

The residuals over various ranges and variables for each δvi can then be sliced up and

binned for fitting with a Gaussian. These fit results are then used to develop trends that

map out and express each component as functionals with a designated four-vector going in

and the error estimate coming out.

13.1.2 Off-Diagonal Terms

In order to construct the full covariance matrix for any particle it is necessary get all

of the correlations between the measured variables represented in the off-diagonal terms.

One method to obtain these correlations is to find the Pearson product-moment correlation

which is a measure of dependence between any two observed quantities. The population

correlation coefficient between two observed data values vi and v j and with standard

deviations σi and σ j is defined as,

ρi j =Vi j

σiσ j=

Vi j√ViiVj j

=〈(vi− vi)(v j− v j)〉

σiσ j,

for finite standard deviations and non-zero vi and v j such that the off-diagonal terms to the

covariance matrix become,

Vi j = ρi jσiσ j,

for ρi j where (i , j). It is a corollary of the Cauchy-Schwarz inequality that the |ρ| cannot

exceed 1. The correlation coefficient is also symmetric such that ρi j = ρ ji.

For a correlation of +1 there is a ideal positive or increasing linear relationship

showing a slope of 1 between the two variables, and for a correlation of -1 there is a ideal

decreasing or negative linear relationship. I linear relation of some slope between the two

indicates the degree of linear dependence between the measured parameters. Naturally a

value close to zero indicates a small correlation. Zero correlation implies ideal

Page 226: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

226

independence between the measured parameters; however, the Pearson correlation

coefficient can only detect linear dependencies between two variables.

For a set of n measurements of variables x and y, the Pearson correlation can be

obtained from the sample correlation coefficient. The correlation coefficient of the

measured variables are,

ρxy =1

(n−1)σxσy

n

∑l=1

(xl− x)(yl− y),

where x and y are the sample means for each measured variable, and σx and σy are the

corresponding standard deviations. This value is then used to acquire Vxy for the

covariance matrix.

As an example, assume that for the neutron in the EC the φ resolutions were found to

be dependent on θ and momentum in the target frame. The θ resolution parameters are a

function of θ as well so that

σ2i j = ρi jσiσ j = Vi j,

or

σ2θφ = ρθφ σφ (θ , p)σθ (θ) = Vθφ .

The resulting value would then fill two off-diagonal terms Vθφ and Vφθ in the covariance

matrix.

13.2 Neutron Fitting

For the neutron, the development of the neutron covariance matrix is the main

challenge for kinematic fitting. There are other issues of concern, such as detection

efficiency and misidentifying photons as neutrons, but once the neutron resolution

functionals are found for a data set, kinematic fitting can make a big difference in analysis

without introducing extra systematic uncertainty. As an example using simulations,

consider the reaction γ p→ K0Σ+→ π+π−π+n. After identifying the K0, it is possible to

Page 227: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

227

fit the decay products in a (4-C) fit with nothing missing, Figure 13.1 shows the

confidence level distribution and pull distributions are shown in Figure 13.4.

To demonstrate some adjustments in measured values, a (4-C) kinematic fitting with

a neutron is used and an invariant mass is obtained after the correction to the 4-vectors are

made. Clearly, a nice improvement is seen in Figure 13.2. It is also possible to make a

(5-C) fit to utilize an additional constraint on the π+n to have the invariant mass of the Σ+

making the distribution essentially a delta function. This can improve the missing mass off

the Σ+.

Figure 13.1: Confidence level distribution for a (4-C) kinematic fit π+π−π+n.

Page 228: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

228

Figure 13.2: Left:Showing the (4-C) kinematic fit to get the invariant mass of π+n, Right:improvement in missing mass off the Σ+ (both in GeV).

Figure 13.3: Left:Showing the (5-C) kinematic fit to the invariant mass of π+n, Right:improvement to the missing mass off the Σ+ (both in GeV).

Page 229: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

229

Figure 13.4: Pull distributions for a (4-C) kinematic kit π+π−π+n.

Page 230: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

230

14 SIMULATIONS

As in Part 1, the relative contribution of different reaction channels have been studied

using Monte Carlo simulations. The experimental photon energy distribution for the

subset of g11a at 4.0186 GeV was again used to determine the energies of the incident

photons in the simulation. The weighting of each channel and its contribution to the

variation in the final result is discussed in Section 15.3.

The event generator FSGEN was used with a variable t-dependence such that a

channel with a kaon is generated uniformly in the center-of-mass frame in φ with a

t-dependent distribution in θcm according to P(t) ∝ e2.0t . The real photon energy range

setting used was 1.0-4.2 GeV. Gaussian distributions in x and y with σ = 0.5 cm were

used to approximate the beam width in the target. The target length of the g11a target at

40 cm from (30 cm,-10 cm) was again used to generate events uniformly along the length

of the target along the beam axis. The generated events were fed into GSIM to simulate

the CLAS detector in the Monte Carlo.

This analysis relies on an understanding of the contributing leakage of background

channels into the π0 and radiative signal counts used in the ratio calculation. For example

π0 leakage from a background channel such as γ p→ ωN∗→ π+π−π0nπ+ will make the

ratio of Σ∗+→ Σ+γ to Σ∗+→ Σ+π0 smaller then it should be. The Monte Carlo of various

possible contribution has been used to study the possible background leakage at various

stages of the analysis. The acceptance of each possible background have been used to

estimate the counts that survive all cuts and contribute in the final signal counts.

One likely background reaction is γ p→ ωN(1440) followed by N(1440)→ nπ+

decay. The ω decays primarily to π+π−π0 followed by π0→ 2γ . The full reaction

γ p→ ωN(1440) then has the same final state as the K0Σ∗+→ Σ+π0 and should be

carefully considered. The N(1440) has a relatively large decay width at 250-450 MeV

[60] which implies that almost any cut around the π+n at the invariant mass of the Σ+ can

Page 231: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

231

still have contributions from the N(1440) in this spectrum. Figure 14.1 shows the

spectrum of the mass off the invariant mass of the π+n. A cut of ±0.05 from the N∗ mass

is used in each case. A cut near a mass of 1440 MeV is shown in the left plot, giving a

clear ω(782) peak. Next, a cut near the mass of the Σ+ is shown in the right plot,

demonstrating that the ω(782) peak is still present.

Figure 14.1: The missing mass off the N∗ using a ±0.05 invariant mass cut on the π+ncombination at a mass of 1440 MeV (left) demonstrating a clear ω(782) peak; (right) at amass of the Σ+ demonstrating that the ω(782) peak is still present.

Another source of possible background it the reaction γ p→ m∗N∗ where m∗ is any

meson that can decay to ρπ0, and the N∗→ Nπ+ provides the detected pion. Similarly,

γ p→ ρN∗ where the N∗ decay to nπ+π0 is also a possible contaminant. In addition to the

kinematic constraints previously mentioned, these backgrounds cannot contribute for low

W . For testing purposes of these types of reaction, the channel ρN(1520) is considered.

The ρ(770) has a width of Γ = 150.3 MeV and decays almost 100% to ππ so it is

possible to leak under the K0 invariant mass cut.

The K∗ is also checked. The constraints on the K0 combined with the missing mass

constraint off the K0 to be the mass of the Σ∗+ should minimize any K∗ contribution.

However, because the reaction γ p→ K∗0Σ+ has the same possible final states that are

being analyzed, it should be investigated.

Page 232: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

232

Based on the possible final state decay products, the reactions γ p→ ηnπ+,

γ p→ K0Σ0π+, and γ p→ K0Σ+ are also considered. While these backgrounds cannot be

dismissed, careful comparison of the mass spectra from the data and the Monte Carlo

mass spectrum provide an indication of what channels are relevant. The Monte Carlo

channels generated and studied are listed in Table 14.1.

Table 14.1: The set of Monte Carlo channels and amount of events generated for theacceptance studies.

Reaction Generated Generated Events

K0Σ∗+→ K0Σ+γ 4×106

K0Σ∗+→ K0Σ+π0 4×106

K∗0Σ+→ K0Σ+γ 4×106

K∗0Σ+→ K0Σ+π0 4×106

ρN(1520)→ π+π−π0π+n 4×106

ωN(1440)→ π+π−π0nπ+ 4×106

ωN(1440)→ π+π−γnπ+ 4×106

ηnπ+→ π+π−π0π0nπ+ 4×106

γ p→ K0Σ0π+→ π+π−nπ0γπ+ 4×106

γ p→ K0Σ+→ π+π−nπ+ 4×106

To tune the Σ∗+ Monte Carlo, an approximate differential cross section for the

reaction γ p→ K0Σ∗+→ K0Σ+π0 was generated. A 1/Eγ photon energy distribution was

used in the generator. The acceptance for the Σ∗→ Σ+π0 channel was determined by the

ratio of accepted events to thrown events in each Eγ bin. The data and the Σ∗+→ Σ+π0

Monte Carlo were cut on the Y ∗ mass range of 1.34-1.43 GeV, the Σ+ invariant mass was

cut at ±0.005 GeV around the PDG value of the Σ+ mass, and a cut around the missing

Page 233: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

233

mass squared of 0.017-0.022 GeV2, in order to isolate the Σ∗+→ Σ+π0 channel in the

data. The yield was determined by the ratio of the raw Σ+π0 events to the number of

incident photons in each Eγ bin, so as to normalize with the Bremsstrahlung spectrum.

Corrections were made for each bin with the newly obtained acceptances. The density of

the target was assumed to be constant, and no background subtraction was performed. The

cross sections are used to modify the data similar to the discussion in Method 1 of Part 1.

From the adjusted Monte Carlo, acceptance corrections are found for bins in the kaon

cosine center-of-mass angle and the approximated differential cross sections are used to

adjust the Σ∗+ generator. Each corresponding bin was filled according to the distributions

of the data. Each angle bin is broken into Eγ bins and represented accordingly in the new

event weighting scheme of the generator. After these modifications were made, the

resulting Monte Carlo was compared with the data, using the momentum distributions for

the kaon, pion, and proton tracks (as well as the kaon lab frame angle distribution), and

found to be reasonable.

The same differential cross sections for the reaction γ p→ K0Σ∗+→ K+Σ+π0 was

also used to generate the radiative channels. A 1 GeV2 t-dependence was used to produce

the other Monte Carlo background channels.

The simulations for γ p→ K0Σ∗+→ K0Σ+γ and γ p→ K+Σ∗+→ K0Σ+π0 were

produced and studied. Figure 14.2 shows the γ p→ K0Σ∗+→ K0Σ+γ channel distributions

for the π+1 -π− invariant mass, the missing mass off the π

+1 -π− combination, the missing

energy from all detected particle, and the missing mass squared of all detected particles.

Figure 14.3 shows the same for the γ p→ K+Σ∗+→ K0Σ+π0 channel. For comparison the

same distributions for the ωN(1440)→ π+π−π0nπ+, ρN(1520)→ π+π−π0π+n,

ηnπ+→ π+π−π0π0nπ+, and K∗0Σ+→ K0Σ+π0 channels are shown in Figure 14.4,

14.5, 14.6, and 14.7 respectively. The background channels are shown without any cuts

applied but with the selection of π+1 and π

−2 made as described previously for the K0.

Page 234: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

234

As expected many of the background channel lead to a similar total missing mass

squared spectrum as for the γ p→ K0Σ∗+→ K0Σ+π0 channel. The most important thing

to notice from the distributions in Figure 14.2-14.7 is that the missing mass off of the π+2 n

combination (middle right plot) in Figure 14.3 compares well with the distribution of the

data in Figure 12.5. Had it looked more like the ω distribution in Figure 14.4 (middle

right), then this would be indicative that the events selected do not come from the channel

of interest. This would be cause for concern even if the acceptance for the ω (or other

background) is very small. The conclusion that can be drawn by this study of the Monte

Carlo is that the required channels have indeed been isolated as required in the data. Any

background from the γ p→ K0Σ∗+→ K0Σ+γ and γ p→ K0Σ∗+→ K0Σ+π0 channels

appear to be only minor contributions.

To calculate the acceptance of the signal and background reactions, the extraction

method used to resolve the radiative and π0 channels will be discussed next.

Page 235: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

235

Figure 14.2: The γ p→K0Σ∗+→K0Σ+γ Monte Carlo distibutions for the π+1 -π− invariant

mass, the missing mass off the π+1 -π− combination, the missing energy from all detected

particle, and the missing mass squared of all detected particles.

Page 236: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

236

Figure 14.3: The γ p → K0Σ∗+ → K0Σ+π0 Monte Carlo distibutions for the π+1 -π−

invarinat mass, the missing mass off the π+1 -π− combination, the missng energy from all

detected particle, and the missing mass squared of all detected particles.

Page 237: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

237

Figure 14.4: The ωN(1440)→ π+π−π0nπ+ Monte Carlo distributions for the π+1 -π−

invarinat mass, the missing mass off the π+1 -π− combination, the missing energy from all

detected particle, and the missing mass squared of all detected particles, no cuts yet applied.

Page 238: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

238

Figure 14.5: The ρN(1520) → π+π−π0π+n Monte Carlo distibutions for the π+1 -π−

invariant mass, the missing mass off the π+1 -π− combination, the missing energy from

all detected particle, and the missing mass squared of all detected particles, no cuts yetapplied.

Page 239: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

239

Figure 14.6: The ηnπ+ → π+π−π0π0nπ+ Monte Carlo distibutions for the π+1 -π−

invarinat mass, the missing mass off the π+1 -π− combination, the missng energy from all

detected particle, and the missing mass squared of all detected particles, no cuts yet applied.

Page 240: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

240

Figure 14.7: The K∗0Σ+ → K0Σ+π0 Monte Carlo distributions for the π+1 -π− invariant

mass, the missing mass off the π+1 -π− combination, the missing energy from all detected

particle, and the missing mass squared of all detected particles, no cuts yet applied.

Page 241: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

241

15 EXTRACTION METHOD

The missing energy boosted in the frame of the Σ∗+ is analyzed to help separate the

γ p→ K0Σ∗+→ K0Σ+γ signal and the γ p→ K+Σ∗+→ K0Σ+π0 background. The

missing energy of the Σ∗+ should be the difference in mass energy between the Σ∗+ and

Σ+ (∼ 0.193 GeV). The two step kinematic fitting procedure of Part 1 is not used in

attempt to maximize statistics. A missing energy restriction can help reduce the

overwhelming amount of π0 background before applying a kinematic fit to the radiative

signal. The intial energy restriction for the radiative signal used is a cut a 0.01 GeV below

the expected missing energy of 0.193 GeV.

Similar to Part 1, a kinematic fit is used to fit each track of all detected particles to a

particular missing particle hypothesis. This is again done by using the undetected particle

mass in the constraint equation. The (1-C) kinematic fit is use to kinematically fit at the

final stage of analysis. In the attempt to separate the contributions of the Σ∗+ radiative

decay and the decay to Σ+π0, the events were fit using different hypotheses for the

topology:

γ p→ π+π−π+n(π0) (1-C)

γ p→ π+π−π+n(γ) (1-C).

The constraint equations are

F =

Ebeam +Mp−Eπ+−Eπ−−Eπ+−En−Ex

~pbeam−~pπ+−~pπ−−~pπ+−~px

=~0. (15.1)

~Px and Ex represent the missing momentum and energy of the undetected π0 or γ .

15.1 Acceptance

The π0 leakage into the γ channel is still the dominant correction to the radiative

branching ratio. To properly calculate the ratio, the leakage into the π0 region from the γ

Page 242: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

242

channel is also used. The same style of notation is use from Part 1 of this thesis. Taking

just these two channels into consideration, the number of true counts can be represented

as N(Σγ) for the Σ∗+→ Σγ channel and N(Σπ) for the Σ∗+→ Σ+π0 channel. The

acceptance under the Σ∗→ Σ+γ hypothesis can be written as Aγ(X), with the subscript

showing the hypothesis type and the actual channel of Monte Carlo input to obtain the

acceptance value is indicated in the parentheses. For the calculated acceptance for the

Σ∗+→ Σγ channel under the Σ∗+→ Σ+γ hypothesis the acceptance is Aγ(Σγ), and for the

Σ∗+→ Σ+π0 hypothesis it is Aπ(Σ+γ). It is now possible to express the measured values

for each channel nγ and nπ as

nγ = Aγ(Σγ)N(Σγ)+Aγ(Σπ)N(Σπ) (15.2)

nπ = Aπ(Σπ)N(Σπ)+Aπ(Σγ)N(Σγ). (15.3)

The desired branching ratio of the radiative channel to the π0 channel using the true

counts is then R = N(Σγ)/N(Σπ). Solving for R to get the branching ratio expressed in

terms of measured values and acceptances,

R =nγAπ(Σπ)−nπAγ(Σπ)nπAγ(Σγ)−nγAπ(Σγ)

. (15.4)

Equation 15.4 is under the assumption that there are no further background contributions.

The formula for the branching ratio to take into account background from the ωN(1440)

can be expressed as

R =∆nγAΣ

π(Σπ)−∆nπAΣγ (Σπ)

∆nπAΣγ (Σγ)−∆nγAΣ

π(Σγ),

(15.5)

∆nπ = nπ −Nπ(ω → π+

π−

π0)−Nπ(ω → π

+π−

γ) (15.6)

∆nγ = nγ −Nγ(ω → π+

π−

γ)−Nγ(ω → π+

π−

π0). (15.7)

Page 243: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

243

The nγ (nπ ) terms come directly from the yield of the kinematic fits and represent the

measured number of photon (pion) candidates. In the notation used, lower case n

represents the measured counts, while upper case N represents the acceptance corrected or

derived quantities. The Nγ,π terms are corrections needed for the leakage from the ωN∗

channel (this background is used as an example). The notation utilized is such that the

pion (photon) channel identifications are denoted AΣπ(Σ+π−) (AΣ

γ (Σ+π−)) so that AΣγ (Σπ)

denotes the relative leakage of the Σπ channel into the Σ+γ extraction and AΣπ(Σγ) denotes

the relative leakage of the Σ+γ channel into the Σ+π extraction.

Each Monte Carlo channel listed in Table 14.1 is run through the analysis with the

same cuts as used for the data. These cuts for the extraction of the radiative and π0 signal

are listed is Table 15.1. The cuts are listed in the order implemented. The first cut, (1), is

the confidence level cut taken on the π+1 and π− kinematic fit to a K0 invariant mass with

total missing mass off the K0 of the Σ∗+. The second cut, (2), lists the invariant mass cut

used on the π+-n combination around the value of the Σ+. The energy restriction used

only for the radiative signal is number (3). Then number (4) and (5) list the final

confidence level cuts used from the kinematic fit to the missing π0 and radiative

hypothesis. The second column lists whether the cut was applied to just one channel or

both.

The acceptances are found for each channel in Table 14.1 with the set of cuts listed in

Table 15.1. Table 15.2 lists all channels taken into consideration and the value of the

acceptance for the Pπ0(χ2) > 10% and Pγ(χ2) > 10%. The table lists two columns sorted

by hypothesis Aγ , Aπ . The uncertainty is statistical only.

Page 244: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

244

Table 15.1: The cuts used to extract the final radiative and π0 counts. (See text for details.)

Cut Used (Applied)

(1)PK0(χ2) > 0.1% (both)

(2)|M(π+2 n)−MΣ+|< 0.02 (both)

(3)Ex(γ) < 0.183 (γ)

(4)Pπ0(χ2) > 10% (π0)

(5)Pγ(χ2) > 10% (γ)

Table 15.2: Acceptances (in units of 10−3) for the channels used in the calculation of thebranching ratios. All the cuts used to obtain the acceptance values are listed in Table 15.1.The uncertainties are statistical only. The two columns contain the acceptance for eachhypothesis Aγ , Aπ . In some cases the values are rounded up to 0.0001.

Reaction Aπ Aγ

K0Σ∗+→ K0Σ+γ 0.01130±0.0020 1.5510±0.0140

K0Σ∗+→ K0Σ+π0 1.3530±0.0120 0.0337±0.0021

K∗0Σ+→ K0Σ+γ 0.0055±0.0008 0.0009±0.0005

K∗0Σ+→ K0Σ+π0 0.0001±0.0000 0.0002±0.0000

ρN(1520)→ π+π−π0π+n 0.0001±0.0000 0.0001±0.0000

ωN(1440)→ π+π−π0nπ+ 0.0035±0.0009 0.0008±0.0003

ωN(1440)→ π+π−γnπ+ 0.0001±0.0000 0.0001±0.0000

ηnπ+→ π+π−π0π0nπ+ 0.0001±0.0000 0.0001±0.0000

γ p→ K0Σ0π+→ π+π−nπ0γπ+ 0.0001±0.0010 0.0001±0.0000

γ p→ K0Σ+→ π+π−nπ+ 0.0001±0.0000 0.0001±0.0000

Page 245: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

245

The acceptance values indicate that very little background gets in to the final

radiative and π0 counts. Had there been a structure seen is Figure 12.5 indicating some

background present it would be necessary to obtain a measure of the surviving counts

from that channel and calculate the contribution to the ratio. This is true even for the small

values of the acceptances due to the sensitive nature of the signal extraction. Based on the

simulation distributions in Figures 14.2-14.7 in comparison to that of the data in Figure

12.1-12.5, no background will be taken into account in the initial calculation of the

branching ratio. However it is possible to study the effect of significant background as part

of the systematic studies (next section). Due to the fact that the ωN(1440) is very difficult

to differentiate from the π0 channel, there could be a large amount of systematic

uncertainty.

To proceed with the background calculation, consider the background of the

ωN(1440). In order to obtain an estimate for the ωN(1440) counts present in the final

singal extraction a relationship between the number of counts from the K0Σ∗+→ K0Σ+π0

to the number of ωN(1440)→ π+π−π0nπ+ present in the peak seen in the full missing

mass squared in the final stage of the analysis before taking the confidence level cuts for

each of the radiative and π0 hypothesis (the mass of the Σ+ could also be used). For this

example calculation, it is assumed that the radiative and all other contribution to the

missing mass squared peak are negligible. The relationship between the π0 counts from

the Σ∗0 and the ω is then

nω = f nΣ∗. (15.8)

Because this calculation of background is used only in a test, the value of f is specified for

a particular systematic study, see Section 15.3 for details.

Page 246: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

246

The total number of events from ωN(1440)→ π+π−π0nπ+ present (N(ω)) at the

final stage of the analysis is calculated as,

N(ω) =nω

R(ω∗→ π+π−π0)Aω(π+π−π0). (15.9)

Here R(ω → π+π−π0) is the probability that the ω will decay to the π+π−π0 and

Aω(π+π−π0) is the probability that this decay channel will be observed after all the

applied cuts. An estimate of the number of counts in the π0 peak coming from some

decay mode of the ω using Eq. 15.9, can be calculated;

Nπ(ω) = R(ω → π+

π−

π0)Aω

π (X)N(ω) =Aω

π (X)nω

Aωγπ(π+π−π0)

. (15.10)

The notation here Aγπ again denote the counts that survive all other cuts but do not

satisfy either the γ or π0 hypothesis in the final confidence level cuts and still fit in the π0

missing mass squared peak.

To look at the contribution form the ωN(1440)→ π+π−γnπ+ channel an estimate

can be obtained from,

Nπ(ω) = R(ω → π+

π−

π0)Aω

π (π+π−

γ)N(ω)R(ω → π+

π−

γ) =

Aωπ (π+π−γ)R(ω → π+π−γ)nω

Aωγπ(π+π−π0)

(15.11)

where R(ω → π+π−γ) is the branching ratio for the radiative decay of the ω with a value

of 3.6×10−3. It is now possible to express all other associated ω corrections used in a

systematic study for a given value of f , along with the acceptance terms for that particular

channel. For example the corrections for the γ channel can be written as

Nγ(ω → ηγ) =Aω

γ (ηγ)R(ω → ηγ)nω

Aωγπ(π+π−π0)

, (15.12)

Page 247: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

247

Nγ(ω → π+

π−

π0π

0) =Aω

γ (π+π−π0π0)R(ω → π+π−π0π0)nω

Aωγπ(π+π−π0)

, (15.13)

Nγ(ω → π+

π−

γ) =Aω

γ (π+π−γ)R(ω → π+π−γ)nω

Aωγπ(π+π−π0)

,

Nγ(ω → π+

π−

π0) =

Aωγ (π+π−π0)nω

Aωγπ(π+π−π0)

. (15.14)

For the π0 channel they take the form,

Nπ(ω → ηγ) =Aω

π (ηγ)R(ω → ηγ)nω

Aωγπ(π+π−π0)

, (15.15)

Nπ(ω → π+

π−

π0π

0) =Aω

π (π+π−π0π0)R(ω → π+π−π0π0)nω

Aωγπ(π+π−π0)

,

Nπ(ω → π+

π−

γ) =Aω

π (π+π−γ)R(ω → π+π−γ)nω

Aωγπ(π+π−π0)

,

Nπ(ω → π+

π−

π0) =

Aωπ (π+π−π0)nω

Aωγπ(π+π−π0)

. (15.16)

where R is used for the corresponding branching ratio in each case. The value R(ω → ηγ)

is listed as 4.9×10−4 and the value of R(ω → π+π−π0π0) is listed as < 2% [60].

The same sort of calculation can be used to check the variation in the ratio for the K∗

contribution and the ρN(1520). The possible contributions these channels is discussed in

Section 15.3.

Page 248: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

248

15.2 Results

To calculate the branching ratio Eq. 15.4 is employed. All terms that take into

account any channel other than the π0 and radiative signal are for the time being ignored.

The acceptance values are taken from Table 15.2. The raw values obtained out of the final

kinematic fit are nγ = 63 and nπ = 1798.

The ratio of the K0Σ∗+→ K0Σ+γ channel to the K0Σ∗+→ K0Σ+π0 is then,

R =nγAπ(Σπ)−nπAγ(Σπ)nπAγ(Σγ)−nγAπ(Σγ)

. (15.17)

Plugging in values for the numerator,

nγAπ(Σπ)−nπAγ(Σπ) = (63)(1.353)− (1798)(0.0337). (15.18)

The denominator is then,

nπAγ(Σγ)−nγAπ(Σγ) = (1798)(1.551)− (63)(0.0113). (15.19)

This leads to the final value of R to be,

RΣ∗+→Σ+γ

Σ∗+→Σ+π0 =nγAπ(Σπ)−nπAγ(Σπ)nπAγ(Σγ)−nγAπ(Σγ)

= 0.88402±0.39%. (15.20)

The uncertainty is statistical only propagated from δnγ = √nγ and δnπ =√

nπ .

The results for the extracted values after the acceptance corrections are listed in Table

15.3. The raw counts for the radiative and π0 extraction using Pγ = Pπ0 = 10% and the

cuts from Table 15.1 are also present in Table 15.3. After adjusting for acceptance the

true signal counts become ∆nγ = 24.65 and ∆nπ = 2787.99.

To calculate the width from the branching ratio achieved here one uses the full width

of the Σ∗+ which is ΓFull = 35.8±0.8 MeV with the the branching ratio that the radiative

signal is being normalized to, which is the R(Σ∗+→ Σ+π0) = 5.85±0.75%. The partial

width calculation is then

Page 249: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

249

Table 15.3: Breakdown of statistics for each term in Eq. 15.4 for the Σ(γ) and Σ(π0)hypothesis. Each counts for each hypothesis is subtracted accordingly as shown in Eq. 15.4.Th raw count are taken directly from the kinematic fit to use in the final ratio calculation.The uncertainties listed are statistical only.

Hypothesis Σ(γ) Σ(π0)

Raw counts 63±7.94 1798±42.40

Σ(1385)→ Σ+γ 85.24±10.77 60.59± 3.87

Σ(1385)→ Σ+π 2788.70±70.42 0.712± 0.155

Adjusted counts 24.65 2787.99

ΓΣ∗+→Σ+γ = RΣ∗+→Σ+γ

Σ∗+→Σ+π0R(Σ∗+→ Σ+

π0)ΓFull = 18.5±8.6 keV. (15.21)

Again only statistical uncertainty is presented. The issue to determine is exactly how

reliable are the presented numbers. As with the previous methods used in Part 1, some

systematic variation in the ratio based on the choice of confidence level is expected.

Unlike the previous method, there is a cut introduced on the extraction of the radiative

signal to restrict the missing energy. The choice to used an energy restriction comes from

the attempt to maximize statistics rather then using the two step kinematic fit. However a

simple energy restriction can not be expected to filter out that much of the π0 background

due to the fact that the radiative signal is so small compared to the π0. This variation and

all other systematic studied are considered in next section.

Page 250: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

250

15.3 Systematic Studies

The value of each cut is varied to study the effect on the final acceptance corrected

ratio. For each variation the new acceptance terms in Equation 15.4 are recalculated with

the corresponding Monte Carlo. Each major systematic uncertainty contribution is

collected together in a table at the end of this section. Each term is numbered as it is

discussed to tie the discussion to the corresponding term listed in Table 15.5.

Several ∆β cut variations are checked starting with ∆β < 0.022 for all charged

particles leading to a ratio of 1.11±0.44%. There is also a check at ∆β < 0.1 giving a

ratio of 0.84±0.39%. The ∆β selected uses a ±1 ns timing cut, while keeping ∆β < 0.025

for both the π+ with only a loose cut of ∆β < 0.1 for the π−. This variation is presented

at the top of Table 15.5 as number (1).

The first kinematic fit used in this analysis is applied only to the π+ and π−

associated with the K0. To investigate the systematic variation in the ratio due to the

confidence level cut applied the ratio signals are re-extracted and new acceptance terms

are calculated over a set of confidence level cuts.

Page 251: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

251

Figure 15.1: The variation in the ratio due to the confidence level cut from the kinematicfit of π

+1 -π− combination to the invariant mass of K0 with missing mass of Σ∗0.

Page 252: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

252

It is again difficult to obtain an estimate of the systematic uncertainty in the Monte

Carlo’s capacity to match the data. The method used in Part 1 is again employed by

checking the effects of various tuning parameters on the final ratio. The systematic effects

of the Monte Carlo are studied by making adjustments to the generator, producing a new

simulation and then re-evaluating the acceptance terms. It is possible to check the

variation in the ratio (based on the new acceptances) by modifying only the generator to

match the production cross section seen in the photon energy distribution from the data

and then checking various exponential t-dependences. Prior to using the approximated

differential cross section, the best data matching t-slope value in the Monte Carlo is found

to be ∼ 1.8 GeV2. To study the generator’s effect on the ratio, a t-slope in the range of

∼±20% the ideal matching value of 1.8 GeV2 is selected to test the effects on the

resulting acceptance corrected ratio. The change in t-slope is varied simultaneously for the

radiative and π0 channels. The acceptances for both channels are recalculated and the

resulting ratio for various t-slopes is shown in Figure 15.2. The branching ratio using a

t-slope of 1.8 GeV2 is0.891%, and using the differential cross sections to match the data,

the branching ratio is 0.884%. In the final ratio reported, the modification in the generator

used to match the differential cross sections seen in the data is used. The t-slope is only

used as a way to study the variation in the ratio. Only the flat region in Figure 15.2 is used

in the systematic uncertainty contribution. The values are between 0.76-0.99. The

variation is listed as number (2) in the table.

Page 253: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

253

Figure 15.2: Variation in the acceptance corrected branching ratio for various t-slope settingin the generator.

Page 254: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

254

To demonstrate the possible effect of background on the ratio at values of f from

Equation 15.8 can be used to test the variation on the ratio for a certain fraction of the final

missing mass squared distribution. For this study the value of f is taken to be 10% which

implies for the case of the ωN(1440) that the number of count present in the missing mass

squared peak of the π0 are 10% from ω → π+π−π0. The choice of f is based on the

estimation of the ω counts seen in Figure 14.1 that survive all the cuts except the

kinematic fitting cuts to the missing π0 or radiative signal. After calculating the resulting

effects for each decay mode for the ω , the contribution is shown in Table 15.4.

Table 15.4: The resulting contribution to the ratio for a f value of 10%. Some values arerounded up. Uncertainty is statistical only.

Contribution Counts

Nγ(ω → π+π−π0) 0.001±0.0001

Nγ(ω → π+π−γ) 0.004±0.0002

Nγ(ω → π+π−π0π0) 0.0001±0.0001

Nγ(ω → ηγ) 0.001±0.0001

Nπ(ω → π+π−π0) 20.31±4.89

Nπ(ω → π+π−γ) 1.42±0.51

Nπ(ω → π+π−π0π0) 0.11±0.027

Nπ(ω → ηγ) 0.002±0.0001

Clearly very few counts will contribute to the ratio however due to the sensitive

nature of the ratio still some change can be seen. The ρN(1520) and K∗0Σ+ are

determined to be negligible due to the lack of structure seen in Figure 12.2 and 12.3 in the

missing mass off the π+-n. The acceptances for all other possible background

Page 255: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

255

contribution are too small to make an impact. The resulting ratio including the ω

contribution (with f =10%) is 0.923±0.39.

An additional method of checking the effects of background come from cutting

directly on the peaks that can be identified as background. For example using the missing

mass off all detected particles and the π+ and π− from the K0 it is possible to make a

three particle invariant mass. A cut can then be applied around the mass of ω and the

analysis procedure continued. By using a cut of ±0.01 around the PDG value of the ω in

the three particle invariant mass spectrum a new ratio is found to be R = 0.90±42.

Because the variation in method using f =10% is larger, its value is used in the systematic

uncertainty calculation. This values is listed in Table 15.5 as number (3).

To look at the systematic dependency on the choice of the confidence level cuts in

both the radiative and π0 hypothesis the ratio is recalculated from the resulting raw counts

in each case with the new acceptance terms for a set of equal cuts such that

Pπ(χ2) = Pγ(χ2). A range of confidence level cuts from 0.5%-30% is used. Figure 15.3

show the ratio at each tested confidence level.

The variation in confidence level is also independently tested for each channel. The

confidence level cut for the π0 hypothesis is kept at 10% while the confidence level cut for

the radiative hypothesis is varied. The resulting ratio is shown in Figure 15.4.

Finally the variation in the confidence level for the radiative hypothesis is kept at

10% while the confidence level cut for the π0 hypothesis is varied. The resulting ratio is

shown if Figure 15.5.

Page 256: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

256

Figure 15.3: Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts are equal for each hypothesis suchthat Pγ(χ2) = Pπ(χ2).

Page 257: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

257

Figure 15.4: Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts for the π0 hypothesis is kept at10% while the radiative hypothesis confidence level cut, Pγ(χ2), is varied.

Page 258: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

258

Figure 15.5: Variation in the ratio for a range of confidence level cuts on the final extractionof radiative and π0 signal. Here the confidence level cuts for the radiative hypothesis is keptat 10% while the π0 hypothesis confidence level cut, Pπ0(χ2), is varied..

Page 259: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

259

The change in the ratio over the variation in Pπ0(χ2) is a little bit more stable. This is

expected because the challenge of this analysis is to isolate the radiative signal not the π0.

The largest change is seen in Figure 15.3 over P(χ2) showing a variation in the ratio from

starting a 1.62% however once ratio come down to the P(χ2)∼ 0.90% range and the

numbers stabilize, for this reason only the points in the linear fit are used in the systematic

uncertainty contribution. The values the ratio including all confidence level plots range

from 0.99%-0.79% and are listed in Table 15.5 as number (4).

The missing energy cut on the radiative candidates is also studied. This cut is used to

reduce the amount of π0 background going into the final kinematic fit used to resolve the

radiative signal without greatly reducing statistics. A set of cuts is chosen to remove the

majority of the π0 events to prepare for the final kinematic fit. The range chosen runs from

0.16GeV-0.19GeV. This range is based on simulation studies that suggest that ∼ 0.01

lower than the ∼ 0.193 GeV centroid (0.18 GeV)gives the optimal signal to background

ratio. Again the acceptance for each cut is determined and the ratio is recalculated. Figure

15.6 shows the resulting ratio for each cut.

Clearly the cut is not as stable as the previous studied systematics, however to obtain

a range of uncertainty in the ratio from this cut only the point shown in Figure 15.6 within

the linear fit are used. Two points to the left and two point to the right of the 0.18 GeV cut

used in the final ratio are taken as the systematic range. The resulting variation is from

0.69%-1.22%. These numbers are the largest contributors to the systematic uncertainty

and listed in the table as (5).

There is also the variation from the cut on the π+2 -n combination. This can be studied

by taking various cut in |M(π+2 n)−MΣ+|< Mcut and again analyzing the change in the

ratio. Figure 15.7 show the variation over a range in the cut value from 0.005-0.06 GeV.

Again only the range in the linear fit are used in the systematic uncertainty calculation.

Page 260: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

260

This rage is chosen qualitatively to be the most stable leading to a variation from

0.66%-1.15%. This variation is listed the Table 15.5 as number (6).

Figure 15.6: Variation in the ratio for a range of missing energy cuts on the final extractionof the radiative signal. Here the cuts are only applied to the radiative hypothesis.

Page 261: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

261

Figure 15.7: The variation in ratio from the cut on the Σ+ invariant mass.

Page 262: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

262

Table 15.5 shows a summary of the systematic studies and the higher and lower range

of the ratio based on the variations mentioned for each type of uncertainty.

Table 15.5: Ranges of systematic variation in resulting ratio in (%) showing L(Low)-contribution and H(High)-contribution and rang in each case.

Source Low Range Low Contribution High range High Contribution

(1)∆β 0.840±0.39 -0.044 1.110± 0.440 +0.226

(2)t-slope 0.760±0.39 -0.124 0.99±0.39 +0.106

(3)Background 0.884±0.39 -0.000 0.920± 0.39 +0.036

(4)P(χ2) 0.790±0.39 -0.094 0.990± 0.33 +0.106

(5)Ex 0.690±0.45 -0.194 1.22± 0.48 +0.336

(6)MΣ+ 0.884±0.39 -0.000 1.150± 0.66 +0.266

Total Uncertainty -0.253 +0.508

To calculate the final systematic uncertainty the difference in the ratio R = 0.884%

and the high range of the ratio for each case in Table 9.3 is added in quadrature to obtain a

value for the uncertainty of 0.508 greater than the ratio. The lower systematic uncertainty

is based on the difference between the ratio R = 0.884% and the low range of the ratio for

each case, resulting in a value of 0.253 less than the ratio.

Page 263: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

263

15.4 Conclusion

The increase in the statistical uncertainty in ratio compared to Part 1 is due to the

detection of an additional particle, and the requirement of using the EC. The increase in

systematic uncertainty is due mostly to the method used in extracting the radiative signal

out from under the π0. Using the missing energy restriction can only help this filtration in

a very mild way. The branching ratio reported is 0.884±0.39(stat)+0.51−0.25(sys)%.

Page 264: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

264

16 OVERALL RESULTS

From the U-spin SU(3) multiplet representation the prediction for the ratio of the

∆→ nγ partial width to the Σ∗0→ Λγ partial width from Section 2.2 is,

Γ(∆0→ nγ)Γ(Σ∗0→ Λγ)

=(

Mn

M∆

)(MΛ

MΣ∗0

)−1( qn

)3 43

= 1.56,

leading to the U-spin prediction of the partial width of the electromagnetic decay using the

wdith of the ∆0→ nγ decay of,

1.56−1×Γ(∆0→ nγ) = 1.56−1×660±60 = 423±38 keV.

Using the results from method-2 of Part 1 the ratio of the Σ∗0→ Λ+γ partial width to

the Σ∗0→ Λπ0 partial width is,

RΛγ

Λπ=

Γ[Σ0(1385)→ Λγ]Γ[Σ0(1385)→ Λπ0]

= 1.42±0.12(stat)+0.11−0.07(sys)%

The width for the branching ratio achieved for Part 1 comes from the full width of the Σ∗0

which is Γ(Σ∗0)Full = 36.0±5 MeV with the the branching ratio that the radiative signal

is being normalized to, which is the R(Σ∗0→ Λπ0) = 87.0±1.5% [60]. The partial width

calculation is then

ΓΣ∗+→Σ+γ = RΣ∗0→Λγ

Σ∗0→Λπ0R(Σ∗0→ Λπ0)Γ(Σ∗0)Full = 444.74±72.87(stat)+71.3

−66.0(sys) keV.

(16.1)

This can be compared to the previously published work on this channel which yielded a

ratio of 1.53±0.39(stat)+0.15−0.17% [73]. Again calculating a width from this ratio gives,

Γ = 479±120(stat)+81−100(sys) keV.

The final result from Part 2 with systematic uncertainties is

RΣ∗+→Σ+γ

Σ∗+→Σ+π0 =nγAπ(Σπ)−nπAγ(Σπ)nπAγ(Σγ)−nγAπ(Σγ)

= 0.88402±0.39+0.51−0.25(sys)%, (16.2)

Page 265: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

265

To calculate the partial width from the branching ratio achieved in Part 2 the full

width of the Σ∗+ which is ΓFull = 35.8±0.8 MeV with the the branching ratio that the

radiative signal is being normalized to, which is the R(Σ∗+→ Σ+π0) = 5.85±0.75%.

The partial width calculation including systematic uncertainties leads to,

ΓΣ∗+→Σ+γ = RΣ∗+→Σ+γ

Σ∗+→Σ+π0R(Σ∗→ Σ+

π0)ΓFull = 18.51±8.58(stat)+11.04

−5.81 (sys) keV. (16.3)

Table 16.1 shows the previous model predictions along with the U-spin prediction

and the final results from this analysis in each case. Statistical and systematic

uncertainties are combined in the present partial widths.

Table 16.1: Comparison of theorectical model predictions for the radiative decay widthswith the present results, all in keV.

Model Σ(1385)0→ Λγ Σ+(1385)→ Σ+γ

NRQM 273 104

RCQM 267

χCQM 265 105

MIT Bag 152 117

Soliton 243 91

Skyrme 157-209 47

Algebraic model 221.3 140.7

U-spin 423±38 223±20

Present Results 445±102 18.5±14

Page 266: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

266

Interestingly enough, the partial width of the Σ∗0→ Λγ is still larger than any

prediction listed except for the U-spin prediction. In contrast the partial width of the

Σ∗+→ Σ+γ is smaller than all predictions.

The U-spin prediction for the Σ∗0→ Λγ partial width is well validated by the

experimental results here. In this case U-spin symmetry is confirmed within the

experimental uncertainties. However, the study in Part 2 suggests that within the

experimental uncertainties the Σ∗+ EM decay partial width does not agree with U-spin

symmetry and the other model predictions. It is important to note that the U-spin

prediction for the Σ∗+ EM decay partial width ignores the effects of the interference of the

isovector and isoscaler components of the photon. If the isoscaler component interfered

destructively the resulting prediction could indeed be much smaller.

The results in Ref. [41] reveal that the meson cloud effect can contribute significantly

(∼ 40%) to the overall electromagnetic decay width of the ∆→ Nγ . This puts the

prediction from the model at about 80% of the experimental measurement. As stated

previously it has not yet been determined from a theoretical standpoint if the meson cloud

effects contribute and if so to what degree for the radiative decay of the Σ∗0 and Σ∗+. This

may be the reason for such a difference in the predictions seen from experiment and why

the deviation from the experimental result does not indicate a similar trend for each of the

Σ∗0 and Σ∗+ results. Because the U-spin prediction for Σ∗0 EM decay width uses

empirical information from the ∆ EM decay, contributions from phenomena like the

meson cloud effect should be inherent. The agreement between experiment and the U-spin

prediction, and the disagreement with the other models listed in Table 16.1, suggests that

meson cloud effects are important for calculations of Σ∗0→ Λγ .

Perhaps this work can prompt more encompassing calculations that are necessary to

probe the structure of the baryon resonances and motivate consideration of the meson

cloud contributions for electromagnetic decay predictions.

Page 267: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

267

REFERENCES

[1] M.A.B. Beg, B.W. Lee, and A. Pais, Phys. Rev. Lett. 13, 514 (1964).

[2] H.R. Rubinstein, F. Scheck, and R.H. Socolow, Phys. Rev. 154, 1608 (1967).

[3] N. Isgur, and G. Karl, Phys. Rev. D 21, 3175 (1980).

[4] J. Ashman et al., Phys. Lett. B 206, 364 (1988).

[5] S.J. Brodsky, J. Ellis, and M. Karliner, Phys. Lett. B 206, 309 (1988).

[6] A. Deshpande, R. Milner, R. Venugopalan, and W. Vogelsong, Ann. Rev. Nucl. Part.

Sci. 55, 165 (2005).

[7] R.H. Dalitz, and D.G. Sutherland, Phys. Rev. 146, 1180 (1966).

[8] T. Sato, and T.S.-H. Lee, Phys. Rev. C 54, 2660 (1996).

[9] B. Julia-Dıaz, T.-S.H. Lee, T. Sato, and L.C. Smith, Phys. Rev. C 75:015205 (2007).

[10] S. Taylor et al. , Phys. Rev. C 71:054609 (2005); ibid. 72:039902 (2005).

[11] N. Isgur, and G. Karl, Phys. Rev. D 20, 1191 (1979).

[12] H. Lipkin, Phys. Rev. D 7, 846 (1973).

[13] H. Lipkin, and M. Moinester, Phys. Rev. B287, 179 (1992).

[14] R. Dashen, A.V. Manohar, Phys. Lett. B315, 425 (1993).

[15] D. Leinweber, T. Draper, and R. Woloshyn, Phys. Rev. D 48, (1993).

[16] V. V. Molchanov, (SELEX collaboration), Phys. Rev. B 590, (2004).

[17] J. W. Darewych, M. Horbatsch, and R. Koniuk, Phys. Rev. D 28 1125 (1983).

Page 268: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

268

[18] R. J. Hemingway, “Production of Λ(1405)inK−p Reactions at 4.2 GeV/c,” Nucl.

Phys. B253, (1985) 742-752.

[19] D. W. Thomas, A. Engler, H. E. Fisk, and R. W. Kraemer, “Strange Particle

Production From π−p Interactions at 1.69 GeV/c,” Nucl. Phys. B56, (1973) 15-45.

[20] D. A. Whitehouse, et al. “Radiative Kaon Capture at Rest in Hydorgen,” Phys. Rev.

Lett. 63.13 (1989) 1352-1355.

[21] G. W. Meisner, “Search for γ Production from K0 p Interactions,” Nuovo Cimento

12A.1 (1972) 62-70.

[22] J. Colas, “Search for Electromagnetic Decay of Neutral Hyperons in 1210-1500

MeV Mass Range and of Neutral Meson in the 150-600 MeV Mass Range, ” Nucl.

Phys. B56, (1973) 15-45.

[23] R. Koniuk and, N. Isgur, Phys. Rev. D 21, 1868 (1980); Erratum: ibid. 23, 818

(1981).

[24] M. Warns, W. Pfeil, and H. Rollnik, Phys. Lett. B 258, 431 (1991).

[25] G. Wagner, A. J. Buchmann, and A. Faessler, Phys. Rev. C 58, 1745 (1998).

[26] E. Kaxiras, E.J. Moniz, and M. Soyeur Phys. Rev. D 32, 695 (1985).

[27] C. L. Schat, C. Gobbi, and N. B. Scoccola, Phys. Lett. B 356, 1 (1995).

[28] A. Abada, H. Weigel, and H. Reinhardt, Phys. Lett. B 366, 26 (1996).

[29] T. Haberichter, H. Reinhardt, N. N. Scoccola, and H. Weigel, Nucl. Phys. A 615, 291

(1997).

[30] R. Bijker, F. Iachello, and A. Leviatan, Annals Phys. 284, 89 (2000).

Page 269: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

269

[31] M. N. Butler, M. J. Savage, and R. P. Springer, Nucl. Phys. B 399, 69 (1993).

[32] J. W. C. McNabb et al. , Phys. Rev. C 69:042201 (2004).

[33] R. Bradford et al. , Phys. Rev. C 73:035202 (2006).

[34] M. E. McCracken et al. , Phys. Rev. C 81:025201 (2010).

[35] Jefferson Laboratory. The JLab picture Exchange. http://www1.jlab.org/ul/jpix

[36] G. Gagnon. Jefferson Lab Site Tour,

http://education.jlab.org/sitetour/guidedtour05.l.alt.html

[37] W. Greiner, and B. Miller, Quantum Mechanics: Symmetries, Springer, (1994).

[38] S. Gasiorowicz, Elemetary Particle Physics, Wiley, (1967).

[39] B. Juliaa-Diaz, T.-S. H. Lee, T. Sato, and L. C. Smith Phys. Rev. C 75:015205

(2007).

[40] C .A. Levinson, H. T. Lipkin, and S. Meshov, Phys. Rev. Lett. 7 (1963) 81.

[41] T. Sato, and T.-S. H. Lee

Phys. Rev. C 54:2660 (1996).

[42] W. Bartel et al. ,Phys. Lett. 28, 148 (1968); J. C. Adler et al. , Nucl. Phys. 46, 573

(1972) S. Stein et al. , Phys. Rev. D 12, 1884 (1975); L. M. Stuart et al. Phys. Rev.

D 58 032003 (1998).

[43] C. Mertz et al. , Phys. Rev. Lett. 86, 2963 (2001); C. Kunz et al. , Phys. Lett. B564,

21 (2003); N. F. Sparveris et al. , Phys. Rev. Lett. 94, 022003 (2005).

[44] S. Stave et al., Eur. Phys. J. A 30, 471 (2006).

Page 270: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

270

[45] K. Joo et al. (CLAS Collaboration), Phys. Rev. Lett. 88, 122001 (2002);V. V. Frolov

et al., Phys. Rev. Lett. 82, 45 (1999);J. J. Kelly, Phys. Rev. C 72, 048201 (2005); J. J.

Kelly et al., Phys. Rev. Lett.95, 102001 (2005); M. Ungaro et al. (CLAS

Collaboration), Phys. Rev. Lett.97, 112003 (2006).

[46] M. Niiyama Nucl. Phy. A 827 261c-263c (2009).

[47] L. Guo, and D. P. Weygand for CLAS collaboration arxiv:hep-ex/0601010v (2006).

[48] DESY Buble Chamber Group, R. ERbe et al. , Nuovo Cimento 49A, 504 (1967).

[49] Y. Oh, C. M. Ko, K. Nakayama arxiv:nucl-th/0712.4285v2 (2008).

[50] T. S. Mast et al., “Electromagnetic decay of the Y ∗0 (1520)”, Phys. Rev. Lett. 21

1715-1717 (1968).

[51] E. S. Smith et al.The time of flight system for CLAS. Nucl. Inst. Meth., A 432, 265

(1999).

[52] B. A. Mecking et al.The CEBAF Large Acceptance Spectrometer (CLAS). Nucl.

Instrum. Meth., A503:513-553, 2003.

[53] M. D. Mestayer et al. The CLAS drift chamber system. Nucl. Instrum. Meth., A 449,

81 (2000).

[54] M. Battaglieri et al. Spectroscopy of Exotic Baryons with CLAS: Search for ground

and first excited states. CLAS Analysis Proposal PRO4-021 (2004). Nucl. Inst.

Meth. A 440, 263 (2000).

[55] M. Amarian et al. The CLAS Forward Electromagnetic Calorimeter. Nucl. Inst.

Meth. A 460, 239 (2001).

Page 271: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

271

[56] D. I. Sober et al. The bremsstrahlung tagged photon beam in Hall B at JLab. Nucl.

Inst. Meth. A 440, 263 (2000).

[57] S. Christo. The g11a Target Cell, http://www.jlab.org/∼christo/g11a target.html

[58] Y. G. Sharabian et al. A new highly segmented start counter for the CLAS detector.

Nucl. Inst. Meth. A 556, 246 (2006).

[59] Alan J. Street et al. Final Site Assembly and Testing of the Superconduction Toroidal

Magnet for the CEBAF Large Acceptance Spectrometer (CLAS). IEEE Trans. Mag.

32, No. 4, 2074 (1996).

[60] Review of Particle Properties, J. Phys. G 33, 1 (2009).

[61] B. Julia-Diaz et al., Phys. Rev. C 75, 015205 (2007).

[62] G. W. Meisner, Nuovo Cimento A 12, 62 (1972).

[63] M. Battaglieri, De Vita, “The g11 fiducial cuts”,

http://www.jlab.org/Hall-B/secure/g11/analysis

archive/genova/notes/g11fiducial note.ps

[64] M. Williams, Ph.D. thesis, Carnegie Mellon University 2007.

[65] M. McCracken, Ph.D. thesis, Carnegie Mellon University 2008 (and references

therein).

[66] M. Williams, C.A. Meyer,“Kinematic Fitting in CLAS,” CLAS-NOTE 2003-017

2003.

[67] E. Pasyuk, “The CLAS ELOSS package”, Software available via CLAS CVS

repository. CLAS-NOTE 2007-016.

Page 272: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

272

[68] Brandt, Siegmund, Data Analysis, 3rd Edition, Springer 1999.

[69] P. Avery, “General Least Squares Fitting Theory,” CBX-91-72.

[70] W. R. Leo, Techniques for Nuclear and Particle Physics Experiments:

A How-to Approach, 2nd. edition (Berlin: Springer-Verlag, 1994).

[71] D. Keller “Techniques in Kinematic Fitting”, CLAS-NOTE, Jefferson Lab,

2010-014.

[72] D. Applegate, M. Bellis, Z. Krahn, C. Meyer, and M. Williams.“ A Detailed Study of

the Sources of Systemtic Errors in the CLAS g11 Data”, CLAS-NOTE 2006-017.

[73] S. Taylor, and G. Mutchler,“Analysis update for E89-024 (radiative decays of

exited-state hyperons)”, CLAS-NOTE 2003-108.

[74] H. Burkhardt and, J. Lowe, “Decays of the K−p atom and the Λ(1405),” Physical

Review C44.2 (1991) 607-614.

[75] J. C. Nacher, E. Oset, H.. Toki, and A. Ramos, “Photoproduction of the Λ(1405) on

the Proton and Nuclei,” Phys. Lett. B 455, 55 (1999), nucl-th/9812055.

[76] Workman, and Fearing, “Radiative cature of stopped kaons: K−p→ Λγ and

K−p→ Σ0γ ,” Physical Review D37 11 (1988) 3117.

[77] Umino and, Myhrer,“Λ(1405) in the bound-state soliton model,” Nuclear Physics

A554 (1993) 593.

[78] Darewych, Horbatsch and, Koniuk,“Λ(1405) in the bound-state soliton model,”

Physical Review D28 5 (1983) 1125.

Page 273: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

273

APPENDIX A: RATIO CALCULATION DETAILS

A.1 Full Ratio Calculation

In this section the branching ratio and uncertainty is calculated for Method-2 using

nγ = 635 and nπ = 13950. The values from Table 8.7 are used for each acceptance term in

the following calculations.

The ratio is calculated using the expression in Eq. 15.5 to take all backgrounds into

account such that,

R =∆nγ

(AΣ

π(Λπ)+ RΣπΛπ

2 AΣπ(Σπ)

)−∆nπ

(AΣ

γ (Λπ)+ RΣπΛπ

2 AΣγ (Σπ)

)∆nπAΣ

γ (Λγ)−∆nγAΣπ(Λγ)

,

where the adjusted counts considering the background are,

∆nπ = nπ −Nπ(Λ∗→ Σ+

π−)−Nπ(Λ∗→ Σ

0)−Nπ(Λ∗→ Σ0γ)

−Nπ(Λ∗→ Λγ)−Nπ(K∗→ Kπ0),

∆nγ = nγ −Nγ(Λ∗→ Σ+

π−)−Nγ(Λ∗→ Σ

0)−Nγ(Λ∗→ Σ0γ)

−Nγ(Λ∗→ Λγ)−Nγ(K∗→ Kγ).

The radiative background term derived from the Λ(1405) counts is, using Γγ(Λ∗) = 27±8

keV and Γtot(Λ∗) = 50 MeV,

Nγ(Λ∗→ Λγ) =AΛ

γ (Λγ)R(Λ∗→ Λγ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 3.0409.

(A.1)

Page 274: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

274

To carry out an example error calculation for this term,

δN2γ (Λ∗→ Λγ) =

(RnΛδAγ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)

)2

+

(AγRδnΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)

)2

+

(AγnΛδR

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)

)2

+

(AγRnΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)δAΛ

γπ(Σ0π

0)

)2

+

(AγRnΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)δAΛ

γπ(Σ+π−)

)2

= 0.592.

Where R(Λ∗→ Λγ) = 5.4×10−4±1.62×10−4 is from Ref. [74]. This leads to the

statistical uncertainty in Nγ(Λ∗→ Λγ) of 0.769. Each set of counts and uncertainties is

calculated accordingly,

Nγ(Λ∗→ Σ0γ) =

AΛγ (Σ0γ)R(Λ∗→ Σ0γ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 0.1297±0.0442,

Nγ(Λ∗→ Σ0π

0) =AΛ

γ (Σ0π0)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 3.413±0.355,

Nγ(Λ∗→ Σ+

π−) =

AΛγ (Σ+π−)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 4.437±1.087.

Where R(Λ∗→ Σγ) = 2−4±8.05×10−5 is from Ref. [74]. For the π0 channel the

background terms derived from the Λ(1405) counts are,

Nπ(Λ∗→ Λγ) =AΛ

π (Λγ)R(Λ∗→ Λγ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 0.002±0.0005,

Nπ(Λ∗→ Σ0γ) =

AΛπ (Σ0γ)R(Λ∗→ Σ0γ)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 0.116±0.04,

Nπ(Λ∗→ Σ0π

0) =AΛ

π (Σ0π0)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 168.943±11.649,

Nπ(Λ∗→ Σ+

π−) =

AΛπ (Σ+π−)nΛ

AΛγπ(Σ0π0)+AΛ

γπ(Σ+π−)= 98.977±7.797.

Page 275: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

275

The background terms derived from the K∗ counts subtracted from the radiative signal

come from an estimate of the total K∗ present (see Table 12).

N(K∗+) =n(K∗+→ K+π0)

R(K∗+→ K+π0)A(K∗+→ K+π0)

= 1287/0.66666(0.01)

= 1.93×105±0.06×105,

where the term A(K∗+→ K+π0) = 0.010±0.0003 is the acceptance for the K∗+→ K+π0

using the extrapolation method to achieve the K∗+ counts. The counts 1287 come from

Table 8.9. To find the counts to the radiative contribution, using

R(K∗→ Kγ) = 9.9×10−4,

Nγ(K∗+→ K+γ) = R(K∗+→ K+

γ)Aγ(K∗+→ K+π

0)N(K∗+)

= 0.03±0.0015.

The background counts from the K∗ to the π0 contribution is

Nπ(K∗+→ K+π

0) =Aπ(K∗+→ K+π0)n(K∗+→ K+π0)

A(K∗+→ K+π0)

= 27.41±1.71.

All of these terms are subtracted out of the initial nγ and nπ to obtain

∆nγ = 623.92±25.23 and ∆nπ = 13654.56±118.95. The term that takes the Σ∗→ Σ+π−

counts into consideration is calculated as

RΣπΛπ

2AΣ

π(Σπ) = (0.135±0.011

2)0.161×10−3±0.01×10−3

= 1.087×10−5±8.888×10−6

Page 276: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

276

then added to the other dominant term

AΣπ(Λπ)+

RΣπΛπ

2AΣ

π(Σπ) = 1.421×10−3±0.0278×10−3 +1.087×10−5±8.88×10−6

= 1.432×10−3±2.91×10−5

and

AΣγ (Λπ)+

RΣπΛπ

2AΣ

γ (Σπ) = 0.0321×10−3±0.002×10−3 +1.715×10−7±7.68×10−8

= 3.227×10−5±2.0×10−6,

which can then be used in the numerator to calculate R,

R = 623.92(1.432×10−3)−13654.56(0.03227×10−3)13654.56(2.335×10−3)−623.92(0.0184×10−3) = 0.0142. (A.2)

A.2 Statistical Uncertainty Calculation

To calculate the statistical uncertainty, the ratio derivation is differentiated with

respect to every respective contributing variable to obtain,(δRR

)2

=((

Aπ(Λγ)nπAγ(Λγ)−nγAπ(Λγ)

+Aπ(Λπ)

nγAπ(Λπ)−nπAγ(Λπ)

)δ∆nγ

)2

+((

Aγ(Λγ)nπAγ(Λγ)−nγAπ(Λγ)

+Aγ(Λπ)

nγAπ(Λπ)−nπAγ(Λπ)

)δ∆nπ

)2

+(

nγAπ(Λπ)−nπAγ(Λπ)δAπ(Λπ)

)2

+(

nγAπ(Λπ)−nπAγ(Λπ)δAγ(Λπ)

)2

+(

nπAγ(Λγ)−nγAπ(Λγ)δAγ(Λγ)

)2

+(

nπAγ(Λγ)−nγAπ(Λγ)δAπ(Λγ)

)2

,

where the terms containing the acceptance statistical uncertainties turn out to be small.

The dominating contributions come from the δ∆nπ and δ∆nγ terms. To evaluate, use the

Page 277: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

277

previously given terms in Eq. A.2 to fill in the acceptance and denominators,(∆RR

)2

=((

Aπ(Λγ)31.88

+Aπ(Λπ)0.443

)δ∆nγ

)2

+((

Aγ(Λγ)31.88

+Aγ(Λπ)0.443

)δ∆nπ

)2

=((

0.0184×10−3

31.88+

1.432×10−3

0.443

)25.23

)2

+((

2.33×10−3

31.88+

0.03227×10−3

0.443

)118.95

)2

,

resulting in the final fractional uncertainty of,

δRR

= 0.08228

which means δR = 0.08228×0.0142 = 0.00117.

Page 278: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

278

APPENDIX B: BOS CLAS DATA STRUCTURE

The data structure management of CLAS is a dynamical storage bank knows as BOS

banks. These data structures contain information on raw events, detector status, geometry

and calibration constants, reconstructed event information, tracking information, cluster

reconstruction information and in essence all required CLAS detector information needed

for physics analysis. The recorded units of information are contained in each bank which

is identified by a name consisting of up to four characters to identify the data structure

associate with it.

The communication between detector modules and the acquisition of the CLAS data

stream uses only the BOS bank data structures. Each bank structure type is constructed

using a readable DDL definition. The DDL definition for several BOS banks mentioned in

this analysis are listed here.

Page 279: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

279

PART:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE PART ! create write display delete ! Hit Based Tracking Result bank

!

!CLAS COORDINATE SYSTEM USDED.

!

!ATTributes:

!———–

!COL ATT-name FMT Min Max ! Comments

!

1 pid I 0 10 ! particle id (GEANT)

2 x F -100. 100. ! vector3 t vert; Vertex position x,y,z

3 y F -100. 100. ! y

4 z F -500. 500. ! z

5 E F 0. 16. ! vector4 t p; Energy

6 px F -16. 16. ! momentum x,y,z

7 py F -16. 16. ! py

8 pz F -16. 16. ! pz

9 q F -16. 16. ! charge

0 trkid I -16. 16. ! index to TBID bank, counting from 1

11 qpid F -100.0 100.0 ! quality factor for the pid

12 qtrk F -100. 100.0 ! quality factor for the trk

13 flags I 0 10000 ! set of flags defining track (ie, BEAM)

!

Page 280: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

280

TAGR:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE TAGR ! create write display delete ! Tagger result bank

!

! ATTributes:

! ———–

!COL ATT-name FMT Min Max ! Comments

!

1 ERG F 0. 10. ! Energy of the photon in GeV

2 TTAG F -20. 200. ! Time of the photon has reconstructed in the Tagger

3 TPHO F -20. 200. ! Time of the photon after RF correction

4 STAT I 0 4096 ! Status ( 7 or 15 are Good)

5 T id I 1 121 ! T counter Id

6 E id I 1 767 ! E counter Id

!

Page 281: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

281

MVRT:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE MVRT ! create write display delete ! vertex Result bank

!

! THE DETECTOR COORDINATE SYSTEM IS USED.

!

! ATTributes:

! ———–

!COL ATT-name FMT Min Max ! Comments

!

1 v id I -1000 1000 ! info about track ids

2 ntrk F -100. 100. ! number of tracks used to make vertex

3 x F -1000. 1000. ! x vector3 t vertx,y,z

4 y F -1000. 1000. ! y

5 z F -1000. 1000. ! z

6 chi2 F -1000. 1000. ! chi2

7 cxx F -1000. 1000. ! Covariance matrix array element

8 cxy F -1000. 1000. ! Covariance matrix array element

9 cxz F -1000. 1000. ! Covariance matrix array element

10 cyy F -1000. 1000. ! Covariance matrix array element

11 cyz F -1000. 1000. ! Covariance matrix array element

12 czz F -1000. 1000. ! Covariance matrix array element

13 stat I -1000. 1000. ! status integer, not used yet

!

Page 282: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

282

! note v id is based upon the track id used to make the

! vertex. v id = (summed over all tracks used) 2(tber id of track(1-10))

! + 1 if beamline info used

Page 283: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

283

ECHB:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE ECHB ! create write display delete ! Forward calorimeter result bank !

! ATTributes:

! ———– !COL ATT-name FMT Min Max ! Comments

!

1 Sect I 0 0xFFFF ! Sector number & Layer number

2 E hit F 0.0 6.0 ! energy found

3 dE hit F 0.0 6.0 ! error on the energy found

4 t hit F 0.0 9999.0 ! time found

5 dt hit F 0.0 9999.0 ! error time found

6 i hit F 0.0 9999.0 ! sector rectangular coordinate

7 j hit F 0.0 9999.0 ! sector rectangular coordinate

8 di hit F 0.0 9999.0 ! sector rectangular coordinate error,

9 dj hit F 0.0 9999.0 ! sector rectangular coordinate error,

10 x hit F 0.0 9999.0 ! lab coordinate,

11 y hit F 0.0 9999.0 ! lab coordinate,

12 z hit F 0.0 9999.0 ! lab coordinate,

13 dx hit F 0.0 9999.0 ! lab coordinate error,

14 dy hit F 0.0 9999.0 ! lab coordinate error,

15 dz hit F 0.0 9999.0 ! lab coordinate error,

16 u2 hit F 0.0 9999.0 ! second moment of u hit pattern

17 v2 hit F 0.0 9999.0 ! second moment of v hit pattern

18 w2 hit F 0.0 9999.0 ! second moment of w hit pattern

Page 284: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

284

19 u3 hit F -9999.0 9999.0 ! third moment of u hit pattern

20 v3 hit F -9999.0 9999.0 ! third moment of v hit pattern

21 w3 hit F -9999.0 9999.0 ! third moment of w hit pattern

22 u4 hit F -9999.0 9999.0 ! forth moment of u hit pattern

23 v4 hit F -9999.0 9999.0 ! forth moment of v hit pattern

24 w4 hit F -9999.0 9999.0 ! forth moment of w hit pattern

25 centr U F 0.0 9999.0 ! peak position on U axis

26 centr V F 0.0 9999.0 ! peak position on V axis

27 centr W F 0.0 9999.0 ! peak position on W axis

28 path U F 0.0 9999.0 ! path length from hit position to U axis

29 path V F 0.0 9999.0 ! path length from hit position to V axis

30 path W F 0.0 9999.0 ! path length from hit position to W axis

32 Nstrp V I 0 36 ! Number of V strips in the hit

31 Nstrp U I 0 36 ! Number of U strips in the hit

32 Nstrp V I 0 36 ! Number of V strips in the hit

33 Nstrp W I 0 36 ! Number of W strips in the hit

34 MatchID1 I 0 30 ! Id of matched hit in the layer1

35 CH21 F 0. 999. ! Quality measure of matching with layer1

36 MatchID2 I 0 30 ! Id of matched hit in the layer2

37 CH22 F 0. 999. ! Quality measure of matching with layer2

38 istat I 0 0xFFFF ! Number of hits & hit ID

!!

! For matching if current layer is WHOLE then layer1=INNER and layer2=OUTER

! if current layer is INNER then layer1=WHOLE and layer2=OUTER

! if current layer is OUTER then layer1=WHOLE and layer2=INNER

Page 285: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

285

TBER:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE TBER ! create write display delete ! Time Based Tracking ERror bank

! record no=0

! ! Fit parameter and Covariance matrix: (Cij)

! ! Track# = row# (cf. TBTR bank)

! note these are in the sda tracking coordinate system

! (x=beamline, y=radially outward, z=parallel to axial wires)

! ATTributes:

! ———–

!COL ATT-name FMT Min Max ! Comments

!

1 q over p F 0. 100. ! q/p

2 lambda F -10. 10. ! dip angle (pi/2 - theta)

3 phi F -60. 60. ! phi

4 d0 F -100. 100. ! min.distance from (x=0,y=0,z=?) [cm]

5 z0 F -100. 100. ! z position of starting point [cm]

6 c11 F -10. 10. ! element C1,1

7 c12 F -10. 10. ! element C1,2

8 c13 F -10. 10. ! element C1,3

9 c14 F -10. 10. ! element C1,4

10 c15 F -10. 10. ! element C1,5

11 c22 F -10. 10. ! element C2,2

12 c23 F -10. 10. ! element C2,3

Page 286: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

286

13 c24 F -10. 10. ! element C2,4

14 c25 F -10. 10. ! element C2,5

15 c33 F -10. 10. ! element C3,3

16 c34 F -10. 10. ! element C3,4

17 c35 F -10. 10. ! element C3,5

18 c44 F -10. 10. ! element C4,4

19 c45 F -10. 10. ! element C4,5

20 c55 F -10. 10. ! element C5,5

21 chi2 F 0. 50. ! Chisquare for this Track

22 layinfo1 I 0. 0. ! layerhit info

23 layinfo2 I 0. 0. ! layerhit info&sector&track#in hber

! the layer hit info is stored in the following way

! for layinfo1= sum over each layer used in track(layers 1-30) Of 2(layer#-1)

! for layinfo2 = sum of 2(layer#-31) for (layers 31-36)

! + 256 * track# in sector+2562*track# in hber

! + 2563 * sector

Page 287: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

287

MCTK:

!********************************************************************

! BANKname BANKtype ! Comments

TABLE MCTK ! create write display delete ! GSIM Monte Carlo track bank

! ! ATTributes:

! ———–

!COL ATT-name FMT Min Max ! Comments

! 1 cx F -1. 1. ! x dir cosine at track origin

2 cy F -1. 1. ! y dir cosine

3 cz F -1. 1. ! z dir cosine

4 pmom F 0. 20. ! momentum

5 mass F 0. 10. ! mass

6 charge F -1. 1. ! charge

7 id I -5000 5000 ! track Particle Data Group id

8 flag I 0 0xFFFF ! track flag

9 beg vtx I 0 65536 ! beginning vertex number

10 end vtx I 0 65536 ! ending vertex number

11 parent I 0 65536 ! parent track

Page 288: U-Spin Symmetry Test of the *+ Electromagnetic Decay · photoproduction of the kaon and Σ∗hyperons. Kinematic fitting is used to separate signal from background in each case

288

MCVX

!********************************************************************

! BANKname BANKtype ! Comments

TABLE MCVX ! create write display delete ! GSIM Monte Carlo vertex parameters

!

! ATTributes:

! ———–

!COL ATT-name FMT Min Max ! Comments

!

1 x F -1000. 2000. ! x of vertex

2 y F -1000. 2000. ! y

3 z F -1000. 2000. ! z

4 tof F 0.0 999999. ! secs of flight

5 flag I 0 65536 ! vertex flag

!