For Oval Gear Meter Models:

Stealth Electronic Manual ZTM1010Installation & Maintenance

TS10TS15TS20TS30

Meter Model No.: TS _______________

including :

CC56 PIA EDB 180PC58 PCDT58 PDB 180DU53 PRT PDMB 180ELNC PAC PMFC

TUTHILL

Page 2: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two

FPP Meters’ Superior Meter Accuracy and Reliability Results from TwoClose-Fitting Oval Gears within A Precision Measuring Chamber

About FPP Meters

FPP Meters, formerly Fluid Power Products, is now atradename of Tuthill Transfer Systems. FPP wasestablished in 1980. Since its inception, the companyhas been dedicated to manufacturing cost-effectivehigh performance flow metering devices for petroleum,industrial, commercial, and municipal service.

Our facilities include computerized order entry andinventory control so that you are assured accurate andprompt deliveries. Furthermore, our productionpersonnel ensure that each order, regardless of size,receives individual attention.

Constant attention to new product development andproduction design, our high standards of manufactureand final testing are the reasons why FPP Meters meetyour most demanding requirements.

Our meters are used in batching, blending, processcontrol and dispensing of liquid products in all liquidhandling industries.

Principle of Operation

Elliptical-shaped rotors are used to accurately measureliquid flow by precisely sweeping a known volume ofliquid through a close tolerance measuring chamber.The meshed rotors seal the inlet from the outlet flow,developing a small pressure differential. As the rotorsrotate, they trap a precise quantity of liquid in the areabetween the rotor and the measuring chamber wall.The total quantity of flow for one rotation of the rotorpair is four times the volume trapped between the rotorand the measuring chamber wall. Close tolerancemachining of the rotors and the measuring chambermeans minimum slippage and high accuracy which isessentially unaffected by changes in viscosity. Therate of flow through the meter is proportional to therotational speed of the rotors. Therefore, the meteroutput is directly proportional to the rate of flow. Theoutput for Stealth meters can be either mechanical or

In the electronic version of the meter, magnetsmounted in the rotors actuate a switch. Eachswitch closure represents a precise increment ofvolume. The switch output can then be used as aninput to a variety of electronics including scalingcounters, batch controllers, computers, analogconverters, etc. The combination of accuracy, low costand few moving parts makes the FPP Meters’ minioval gear meters a natural for your meteringapplications.

Fluid Compatibility

If in doubt about compatibility of a specified fluid,contact our manufacturing plant or see ChemicalCompatibility Chart to check compatibility of chemicalsto the wetted materials.

FPP Meters and Principle of OperationFPPMeters and Our Principle of Operation

If You Have AnyQuestions About

Installation orOperation,

Please Consult theFactory Prior to

Installation(888) 578-3258.

SHAFTS

SHAFTS

SHAFTS SHAFTS

electronic. The output for mini oval gear meters iselectronic.

In the mechanical meter, the output shaft drives a geartrain that provides a registration in engineering units.

1

Page 3: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two

Installation & Operation

Make sure that all necessary safety precautionshave been taken including proper clothing,personal safety equipment and fire safetyequipment if required.

Before Start-up of Meter Make Certain:1. The meter is properly mounted, secured and

piped.2. All connections are tight.3. All bleed and drain valves are closed.4. Do NOT smoke near meter or use meter near an open flame when metering flammable fluids. Fire could result.5. This meter is not intended for use with fluids for human consumption.

Install the Meter and Accessories inCompliance with All Applicable Local, State andFederal Construction, Electrical and SafetyCodes.

InstallationPositive displacement meters are designed to operate full ofliquid. The meter should be installed in a manner such thatit remains full of liquid at all times.

Protective caps installed in each meter at the time ofshipment should remain in the openings until you are readyto install in the piping system.

Prior to meter installation, the piping system should beflushed to remove all debris. Apply pipe compound to themale threads, to install the companion flanges.

The meter must be installed in the system free of pipestress. Pipe stress can deform the meter chamber,resulting in loss of accuracy. The meter should alwaysbe supported by bolting to a platform. Never use theconnecting pipe as the means of support.

When installing the meter consider future maintenance, andinstall in the best location available to facilitate future metermaintenance. The meter can be disassembled and servicedin place, and provisions for service should be consideredduring installation.

For the best accuracy, install the meter so that the gearshafts are positioned in a horizontal plane, rotating withoutthe weight of the gear resting on the body or cover of themeter as shown on page 1. In critical installations blockvalves and by-pass lines are recommended. This allows themeter to be serviced without interruption of flow in criticalprocess application.

SAFETY INSTRUCTIONS

It is recommended that a strainer be installedupstream of each meter to prevent damage from foreignmatter, such as weld slag, pipe scale, etc.

Calibration means should be provided during installation. Aneasy means for diverting flow into a calibration vessel shouldbe considered.

Hydraulic shock like thermal expansion can be harmful tometer components. Consideration should be given todesigning pumping piping systems to eliminate hydraulicshock.

OperationFill the system slowly to avoid operation on air or vapor. Thiscan be accomplished in the following manner:

1. Throttle the meter inlet valve, and allow to fill slowly bygravity.

2. Crack open the outlet valve, start the pump, then slowlycrack open the inlet valve and fill the meter slowlybefore fully opening the inlet and outlet valves.

The meter is not designed to operate on air, but the designand materials of construction allow for operation on vapor forshort periods of time without damage to the elliptical gears ormeter internals.

Note: Over-speed and hammer caused by the presenceof vapor in the system can cause internal damage to themeter.

Operating the meter in excess of its maximum design flowcan cause excessive wear or premature failure. However,the meter can be calibrated to operate below the minimumdesign flow rate as indicated on the meter name plate if theflow remains constant, or the product is viscous.

Thermal relief valves are recommended and should beinstalled whenever it is possible to block the meterbetween two valves. Thermal pressures many times theoperating pressure are possible with only a small rise intemperature.

The meter’s non-shock Maximum OperatingPressure is indicated on the meter name plate.The meter should never be operated in excess ofthis pressure. Care should be taken to eliminatethermal and hydraulic shock pressures so thatthey do not exceed the meter’s MaximumWorking Pressure design.

SAFETY INSTRUCTIONS

2

Page 4: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two
Page 5: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two
Page 6: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two
Page 7: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two
Page 8: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two

Parts List for Electronic Meter Model TS15CSerial no.s > B8360

Torque SpecificationsPart # Description Torque ToolFS9830 1/4-20 X 3/4” SS SHCS 78 in.lbs. 3/16 AllenFS9651 10-32 X 3/8” SS SHCS 24 in.lbs. 5/32 AllenFS9901 5/16-18 X 1” SS SHCS 24 ft.lbs. 1/4 Allen

Part # Description Torque ToolFS9515 8-32 X 3/8” SS SHCS Snug tite 9/64 AllenFS9813 1/4-20 X 5/8” FLHD SHCS 30 ft.lbs. 5/32 AllenMS4019 1/4” SS Drain plug Snug tite 1/4 Allen

Item Qty. Description Part #

1 1 TS15C Meter Body MB85802* 1 Gear Set, RytonTM, no bearings GS588R

1 Gear Set, 34 PPG GS580R341 Gear Set, 67 PPG GS580R671 Gear Set w/carbons, 34 PPG GS580RC341 Gear Set w/carbons, 67 PPG GS580RC671 Gear Set, TeflonTM, high viscosity GS580RT34-HV

3** 1 Post Plate assembly CP985213a 1 Post plate CP98523b 2 Gear Posts MP90863c 2 1/4-20 X 3/4” SS SHCS FS98303d 2 Lockwashers FS93024 1 Gear Plate CP98535 4 Dowel pin MS90216 8 10-32 X 3/8” SS SHCS FS9651

Item Qty. Description Part #

7 1 Front cover CP9560 8 1 Rear cover CP9561 9* 2 “O” Ring, TeflonTM, cover assembly SL215710 24 5/16-18 X 1” SS SHCS FS9901

11 1 “O” ring, TeflonTM, sensor SL221212 1 Retaining plate MP906113 3 8-32 X 3/8” SS SHCS FS951514 2 “O” ring, TeflonTM, comp. flanges SL223315 2 1 1/2 ” Companion flange, NPT MP9511

2 1 1/2 ” Companion flange, BSP MP951216 1 Meter pedestal MP911017 2 1/4-20 X 5/8” SS FLHD SHCS FS981318 2 1/4” SS drain plug MS401919* 1 Reed switch assembly EL5599-RS20* 1 Hall effect assembly EL5599-HE

* Recommended Spare Parts** Part of Post plate assembly, Part No.: CP98521

ZLT188 9/00Printed in the USA

Page 9: TUTHILL - Controls Warehouse · Stealth Electronic Manual ZTM1010 ... DU53 PRT PDMB 180 ELNC PAC PMFC TUTHILL. FPP Meters’ Superior Meter Accuracy and Reliability Results from Two

Warranty

Tuthill Transfer Systems (“Manufacturer”) warrants to each buyerof its FPP Meters products (the “Buyer”) for a period of 12 monthsfrom date of invoice or sales receipt, but in no event more than 18months from date of manufacture, that goods of its manufacture(“Goods”) will be free from defects of material and workmanship.Manufacturer’s sole obligation under the foregoing warranties willbe limited to either, at Manufacturers’ option, replacing or repairingdefective Goods (subject to limitations hereinafter provided) orrefunding the purchase price for such Goods theretofore paid bythe Buyer, and Buyer’s exclusive remedy for breach of any suchwarranties will be enforcement of such obligations of Manufacturer.If Manufacturer so requests the return of the Goods, the Goods willbe redelivered to Manufacturer in accordance with Manufacturer’sinstructions F.O.B. Factory. The remedies contained herein shallconstitute the sole recourse of the Buyer against Manufacturer forbreach of warranty. IN NO EVENT SHALL MANUFACTURER'S

Please have the following information available when you makeinquiries, order replacement parts, or schedule service:

Your meter’s serial number: _____________________________

Your meter’s model number: _____________________

Your full service Distributor: ________________________

Your full service Distributor’s phone number: _______________

LIABILITY ON ANY CLAIM FOR DAMAGES ARISING OUT OFTHE MANUFACTURE SALE, DELIVERY OR USE OF THEGOODS EXCEED THE PURCHASE PRICE OF THE GOODS.The foregoing warranties will not extend to Goods subjected tomisuse, neglect, accident or improper installation or maintenance,or which have been altered or repaired by anyone other thanManufacturer or its authorized representative. THE FOREGOINGWARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHERWARRANTIES OF MERCHANTABILITY, FITNESS FORPURPOSE OF ANY OTHER TYPE, WHETHER EXPRESS ORIMPLIED. No person may vary the foregoing warranties andremedies except in writing signed by a duly authorized officer ofManufacturer. Warranties or remedies that differ from theforegoing shall not otherwise be binding on Manufacturer. TheBuyer’s acceptance of delivery of the Goods constitutesacceptance of the foregoing warranties and remedies, and allconditions and limitations thereof.