Upload
luroguita
View
17
Download
0
Embed Size (px)
DESCRIPTION
Barnstead Nanopure Infinity Troubleshooting Guide
Citation preview
NANOpure Infinty Low Purity System Checks:
275 Aiken Road Phone 800-438-4851 www.thermofisher.com
Asheville, NC 28804
1. Open the front door of the Nanopure and examine the cartridges. Note the cartridge catalog number, lot code, and expiration date on each cartridge. If the cartridges have
been installed longer than one year, replace the cartridges first. If you have an organic
removal cartridge installed in position four (far right), try switching this cartridge with
the cartridge in position three. If the purity rises to acceptable levels, the organic
cartridge is most likely at fault.
2. Measure the flow rate from the remote dispenser, which should be approximately 1.5 liters per minute. If the flow is significantly lower, open the front door and locate the
pump in the unit. Place your hand on the pump or listen for vibration (the pump
should be on/vibrating whenever the purity is displayed on the screen). If the pump is
not running, locate a screwdriver, grab it by the blade, and tap repeatedly on the
pump. The pump may start running again. If the pump does not continue to run, order
a new pump, part number PU896X3. If the pump is running and the flow is still low,
check the following:
a. Ensure feed water is of adequate pressure (Gravity feed 100 psi acceptable, 30-50 psi preferred)
b. Ensure no other filters are clogged (0.2 micron final filter or Ultrafilter)
c. Ensure there are no kinks or clogs in the tubing or other component (most notably the pressure regulator)
3. Purchase a NIST calibration module, part number E896X5. After you receive the module, first perform a calibration per the users manual. Then, leaving the module attached, press start on the display, and note the purity value. With the module plugged in, the display should read 15.0 Mohm-cm. If it does read 15.0 (+/- 0.3
Mohm), your boards are reading the purity properly. If it does not read 15.0 when
plugged into J2 of the calibration board, try plugging the module into P2 of the meter
board (slightly larger board, also mounted on the side door of the Nanopure). If the
reading does change to 15.0 when plugged into P2, the calibration board is bad. If it
does not change to 15.0 when plugged into P2, the meter board is most likely bad.
4. Examine the check valve by removing the tubing from its end. If the valve allows water to leak out the end, attempt to free any clogged debris or replace the valve.