1
E-Pure Low Purity Checks 275 Aiken Road Phone 800-438-4851 www.thermofisher.com Asheville, NC 28804 1. Check the cartridges and make sure they are Thermo Scientific Brand, that they are the correct cartridges for the unit, that they are in the correct order, and that they are not expired. Please note the catalog number and lot code on each cartridge. Also check for "spaces” (empty space with no resin material) or an exposed spring (extended 2-3”) at the top or bottom of each cartridge. If a space exists, it indicates that the resin beads inside have shrunk and are exhausted or nearly exhausted. Make sure that each cartridge can “hang” on its own in the canister head. To do this, remove the canister and re-install the cartridge. If it cannot hang in place on its own, the small GSX27 o-ring may need to be replaced. 2. If there is an organic free cartridge, remove it and move the third cartridge over to the last position (farthest on the right; the third canister will now be empty…alternatively you can switch cartridges 3 & 4). If the purity goes up above 18, the organic cartridge may be faulty, and it may need to be replaced. 3. Make sure nothing else was changed in the unit - adding/removing a remote dispenser, re-routing any tubing, etc. 4. Check the flow rate from the system: it should be close to 2 liters per minute. If the flow is low, try dispensing without the final filter. If the flow rate is still low, even when dispensing without the final filter, check for obstructions in the system (especially in the pressure regulator), verify the feed water pressure is adequate, or consider replacing the pump/motor. 5. Examine the check valve (part number 02214; see attached procedure). If it is obviously clogged or leaking, attempt to free any clogged debris or replace the valve. 6. Check the resistivity meter (see D2770 test procedure). If it is not reading correctly, there are two options: a. To purchase a new meter, please contact our Customer Service department at 866-984-3766. The part number for the meter is D2770. b. If you want to repair or calibrate the meter, please contact Thornton (Mettler-Toledo) at 888-687-7687 or [email protected] .

Troubleshooting Guide - Barnstead - E-Pure - Low Resistivity

Embed Size (px)

DESCRIPTION

Barnstead E-Pure Low Resistivity Troubleshooting Guide

Citation preview

  • E-Pure Low Purity Checks

    275 Aiken Road Phone 800-438-4851 www.thermofisher.com

    Asheville, NC 28804

    1. Check the cartridges and make sure they are Thermo Scientific Brand, that they are the correct cartridges for the unit, that they are in the correct order, and that they are not expired. Please note the catalog number and lot code on each cartridge. Also check for "spaces (empty space with no resin material) or an exposed spring (extended 2-3) at the top or bottom of each cartridge. If a space exists, it indicates that the resin beads inside have shrunk and are exhausted or nearly exhausted. Make sure that each cartridge can hang on its own in the canister head. To do this, remove the canister and re-install the cartridge. If it cannot hang in place on its own, the small GSX27 o-ring may need to be replaced.

    2. If there is an organic free cartridge, remove it and move the third cartridge over to the last position (farthest on the right; the third canister will now be emptyalternatively you can switch cartridges 3 & 4). If the purity goes up above 18, the organic cartridge may be faulty, and it may need to be replaced.

    3. Make sure nothing else was changed in the unit - adding/removing a remote dispenser, re-routing any tubing, etc.

    4. Check the flow rate from the system: it should be close to 2 liters per minute. If the

    flow is low, try dispensing without the final filter. If the flow rate is still low, even when dispensing without the final filter, check for obstructions in the system (especially in the pressure regulator), verify the feed water pressure is adequate, or consider replacing the pump/motor.

    5. Examine the check valve (part number 02214; see attached procedure). If it is obviously clogged or leaking, attempt to free any clogged debris or replace the valve.

    6. Check the resistivity meter (see D2770 test procedure). If it is not reading correctly, there are two options:

    a. To purchase a new meter, please contact our Customer Service department at 866-984-3766. The part number for the meter is D2770.

    b. If you want to repair or calibrate the meter, please contact Thornton (Mettler-Toledo) at 888-687-7687 or [email protected].