Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
TriLIN
Linearization
Electronics
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 2
Table of contents
1. General information ................................................................................... 3
2. Safety regulations ...................................................................................... 4 2.1 Marking of important information ......................................................... 4 2.2 General safety regulations ................................................................... 4
3. Description ................................................................................................. 5
3.1 Design and measuring principle .......................................................... 5 3.2 Technical data ..................................................................................... 6
3.3 Type key .............................................................................................. 8
4. Installation / commissioning ....................................................................... 9 4.1 Safety regulations ................................................................................ 9 4.2 Checks after delivery ........................................................................... 9
4.3 Principle of operation ........................................................................... 9 a. Impact of the viscosity ..................................................................... 9
b. Compensation of the effect of viscosity ......................................... 10 c. Calculation method ....................................................................... 10
4.4 Programming the electronics ............................................................. 11 a. Tool necessary .............................................................................. 11
b. Read-out from the electronics ....................................................... 12 c. Importing the calibration curve ...................................................... 12 d. Adjusting the electronics ............................................................... 13
e. Liquid data changes ...................................................................... 14 f. Scaling the outputs ....................................................................... 15
g. Programming ................................................................................ 16 h. Saving data ................................................................................... 16
i. Printing the data ............................................................................ 17 4.5 Adjusting the analogue outputs ......................................................... 18 4.6 Adjusting the PT100 input .................................................................. 19
4.7 TriLIN pin assignments ...................................................................... 20 4.8 Trouble shooting ................................................................................ 21
5. Maintenance ............................................................................................ 22
6. Declaration of conformity ......................................................................... 22
7. Warranty .................................................................................................. 23
8. Customer service ..................................................................................... 23
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 3
1. General information Thank you for selecting a TrigasDM product for your flow measurements.
Flow meter manufacture
As a specialist in flow measurement technology, TrigasDM supplies high-quality measuring instruments, electronics and calibrators for liquids and gases.
Made in Germany
Our products are exclusively developed and manufactured in the municipality of Neufahrn, 20 km north of Munich, ensuring world-class technical know-how for our customers.
Contact
We are proud of our high-quality products and friendly customer service and welcome you as a valued customer to our growing family. You can benefit from our long-standing experience and our comprehensive technical support.
TrigasDM GmbH Tel.: +49 8165 9999 300 Erdinger Str. 2b Fax: +49 8165 9999 369
Email: [email protected] 85375 Neufahrn, Germany www.trigasdm.com
This user manual contains information on the description, operation / commissioning and maintenance of the TrigasDM linearization electronics TriLIN. For special applications, repair or further information on this or other products, please contact TrigasDM directly.
The manufacturer may modify this document without prior notice. Prior to use, request or enquire about valid documentation from the manufacturer if necessary. Warranty claims against the manufacturer may become void if invalid documentation is used.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 4
2. Safety regulations
2.1 Marking of important information
Important information is specially highlighted in this user manual. WARNING To avoid danger to persons, this information is marked with WARNING. CAUTION To avoid danger to equipment, this information is marked with CAUTION. NOTE Special information for operation / commissioning and maintenance is marked with NOTE.
2.2 General safety regulations
Before using the TrigasDM flow meter, this user manual and all safety instructions must be carefully read in their entirety and understood.
Take all necessary precautions to ensure the safety of personnel and equipment. These precautions include, but are NOT limited to, the following examples:
Mechanical and electrical installations must only be carried out by qualified and authorized personnel.
It must be ensured that the measuring range of the flow meter is not exceeded.
Do not install measuring instruments and cables in the vicinity of strong magnetic sources, such as electrical cables, electric motors, transformers, welding equipment, relays or high-voltage cables. These sources can cause electrical noise, resulting in incorrect pulse signals.
Flow meters, which are designed for applications in liquids are not suitable for applications in gas.
The valid safety standards (for example, in accordance with the German Occupational Safety and Health Act) must be observed for the installation and/or operation of the flow meter. Non-observance can result in DANGER to personnel.
A flow meter is a precision instrument. Do not use compressed air to clean the flow meter or check its function.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 5
3. Description
3.1 Design and measuring principle
The TriLIN electronics amplifies, linearises and scales the frequency signals of almost all commercially available flow meters.
The offsetting and temperature compensation is done in accordance with a principle designed for turbine flow meters.
The output after scaling takes place both as an analogue signal 0-10 V (optionally 4-20mA) and also as a frequency signal 0-4800 Hz with a calculation time of about 1.5 to 2.5 milliseconds.
The software that is easy to operate and work with permits programming and scaling.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 6
3.2 Technical data
Input signals There are three standard signal inputs supported, which can be selected by easily setting jumpers:
Carrier frequency signal of carrier frequency signal sensors in the range of 1 to 4000 Hz. Inductance 1 mH, Ohmic resistance 10-13 Ohms, Carrier frequency 45 to 55 kHz
Sinusoidal signal of magnetically inductive signal sensors. Input voltage from 10 mV to 10 V
Pulse input of standard frequency outputs ULOW < 1.5 volts UHIGH > 3-30 volts Range 0.4-32000 Hz optionally also for lower frequencies. Input impedance > 10 kOhm
Linearization Up to 20 points in the assignments
Frequency/Viscosity to pulses/litre or
Frequency to pulses/litre
Linear interpolation between the points lying closest to the frequency.
Temperature input PT100 4-wire or 0-10 volts (scalable)
Refresh time (Mean value) 3 milliseconds
Leak flow volume suppression 3 milliseconds
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 7
Output signals
Frequency Linearised and scaled frequency 5V NPN related to analogue mass Scalable end value up to 4800 Hz Output impedance < 2.2 kOhms Accuracy 0.01% of the measured value Resolution 0.018 Hz
Analogue 0 – 10 volts linearised and scaled Zero point offset <= 2.5 mV for analogue mass Accuracy 0.003% of the measured value Resolution 16-bit (~0.15 mV) Optionally 4 - 20 mA extension
Data transmission via serial port with a Windows Operating System (We recommend Windows 7 or higher). Data can be transmitted with the help of the optional programming cable with standard USB port or with 9-pin Sub-D plug-in connector.
Data entry Manually or as data import
Data output and storage is done completely on the electronics and on storage media
Ambient conditions Operating temperature: -40 … 85°C Storage temperature: -55…+125 °C Humidity: 0 – 85% relative, non-condensing
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 8
3.3 Type key
TriLIN-Series (e.g. LIN-T3-01-R3-O-U1-V2)
LIN- T3- 01- R3- O- U1- V2
Power Supply
V1 = ODU MINI-SNAP Socket 2-Pin (G11L0C-P02LPH0-0000)
Pin 1 = 9-32 VDC ca. 900mW Ansicht: Lötseitig
Pin 2 = 0 V (GND)
V2 = M12 Industrie Plug (SACC-E-MS-5CON-M16/0,5 SCO) (Standard)
Pin 1 = 9-32 VDC ca. 900mW
Pin 2 = 0 V (GND)
V3 = M12 Industrie Plug (SACC-E-MS-5CON-M16/0,5 SCO)
Pin 1 = 0 V (GND)
Pin 2 = 9-32 VDC ca. 900mW
V4 = M12 Industrie Plug (SACC-E-MS-5CON-M16/0,5 SCO)
Pin 1 = 9-32 VDC ca. 900mW
Pin 3 = 0 V (GND)
Voltage, frequency and temperature output
U1 = 3x BNC Socket, insolated from housing (Standard for LIN-E and LIN-T)
Middle contact = Signal +
Outer contact = Signal –
I1 = 4-20mA Current Output, 3x BNC Socket like U1
U2 = LEMO Socket 6-Pin (EGA.1B.306.CLL) (Standard for LIN-T with Display)
Pin 1 = Durchfluss-Ausgang (0-10V)
Pin 3 = Frequenz-Ausgang (TTL)
Pin 5 = Temperatur-Ausgang (0-10V) [Optional]
Pin 2, 4, 6 = Analog GND (0V)
I2 = 4-20mA Current Output, LEMO Socket 6-Pin like U2
U3= 2x ODU MINI-SNAP Plug 5-Pin (GB1F1C-P05MJG0-0000)
Pin 1 = cable shield
Pin 2 = Signal -
Pin 5 = Signal +
Wall holder
0 = Without Holders (Standard)
W = Wall holder
Flowmeter Input
N1 = M12 NPN Pulse Input (Special version)
R1 = ODU MINI-SNAP Socket 5-Pin Carrier frequency (RF) (GB1F1C-P05LJG0-0000)
Pin 1 = cable shield Ansicht: Lötseitig
Pin 2 = carrier frequency -
Pin 5 = carrier frequency +
R2 = M12 Industrie Socket Carrier frequency (RF) (SACC-DSI-M12FS-5CON-M16/0,5) (Standard for LIN-E)
Pin 2 = cable shield
Pin 3 = carrier frequency +
Pin 4 = carrier frequency -
R3 = M12 Industrie Socket Carrier frequency (RF) with integrated Temperature Input (SACC-DSI-M12FS-8CON-M16/0,5) (Standard for LIN-T)
Pin 2 = cable shield
Pin 3 = carrier frequency +
Pin 4 = carrier frequency -
Pin 5 = RTD PT100 Supply +
Pin 6 = RTD PT100 Sensor +
Pin 7 = RTD PT100 Sensor -
Pin 8 = RTD PT100 Supply -
M1 = M12 Industrie Socket (SACC-DSI-M12FS-5CON-M16/0,5)
Pin 2 = cable shield
Pin 3 = Sinus +
Pin 4 = Sinus –
P1 = M12 Industrie Socket (SACC-DSI-M12FS-5CON-M16/0,5)
Pin 1 = Supply +
Pin 2 = Supply -
Pin 5 = Pulse input
Housing
01 = Standard housing
02 = Mini housing
03 = Housing with Display
04 = Mini housing, SIL
ZZ = Special version
Temperature compensation
E = Without
T1 = Temperature Input ODU MINI-SNAP Socket 5-Pin (GB1F1C-P05LJG0-0000)
Pin 2 = RTD PT100 supply + Ansicht: Lötseitig
Pin 3 = RTD PT100 Sensor +
Pin 4 = RTD PT100 Sensor -
Pin 5 = RTD PT100 supply -
T2 = Temperature Input LEMO Socket 6-Pin (EGB.1B.306.CLL)
Pin 1-4 = PT100 input
Pin 5 = Temperature input +
Pin 6 = Temperature input -
T3 = Temperature Input integated in Flowmeter Input R3 (Standard)
Product name
Ca
rrie
r F
req
ue
nc
y f
rom
RF
-Pic
ko
ff
Mag
neti
c
Ind
ucti
ve
Pu
lse
s f
rom
Am
plifi
ed
-
Pic
ko
ff
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 9
4. Installation / commissioning
4.1 Safety regulations
Mechanical and electrical installations must only be carried out by qualified and authorized personnel.
When working on electrical systems, the 5 rules on safety from the DIN VDE 0105 series of standards should be applied diligently and carefully: - De-energise
- Secure against being switched on again
- Ensure absence of voltage
- Connect to earth and short-circuit
- Cover or fence off live adjacent parts
4.2 Checks after delivery
Carefully unpack the product and check for cleanliness (e.g. packaging residues).
Check its condition and check for damage.
4.3 Principle of operation
a. Impact of the viscosity A turbine flow meter changes its frequency output with a change in the viscosity. In other words, the higher the viscosity, the lower is its output frequency with the same flow rate. A good display method for assessment is generally the ratio of the frequency to the factor pulses/litre (pul/lit).
5500
6500
7500
8500
9500
10500
11500
12500
13500
1 10 100 1000 10000
Hz
Pu
l/lit 1.3 cSt
10 cSt
20 cSt
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 10
b. Compensation of the effect of viscosity A turbine flow meter has the characteristic of behaving consistently in the linear measurement range in the factor of frequency to viscosity. This is represented graphically as follows.
You can clearly identify a deviation from the ideal line in the blue line (=1.3 cSt or mm²/sec.). These calibration points of the turbine are not used for subsequent linearization and temperature compensation. c. Calculation method The electronics contains an assignment of the temperature to viscosity. This table is provided to the customer for programming or he can program it himself. This table must be equivalent to the current operating medium. The current viscosity is determined continuously from the current temperature and this table.
Determining the viscosity (Assignment °C to mm²/sec.)
Measurement of the current input frequency and calculation of the current factor (f/V = Frequency to viscosity)
Determining the current K factor (pul/lit) from the programmed table
Conversion of the K factor with the help of the current frequency in the current throughput (e.g. litres/min.)
Optional: Conversion of the throughput with the help of a programmed temperature/density table and the current temperature in a mass throughput (e.g. kg/min.)
Scaling and output of the current temperature at the analogue output (e.g. -50…+150°C = 0…10 volts)
Scaling and output of the current throughput at the analogue and frequency output. (e.g. 0…10 l/min. = 0…10 V = 0….2 kHz)
UVC Graph
5500
6500
7500
8500
9500
10500
11500
12500
13500
0 1 10 100 1000 10000
f/V
pu
l/lit 1.3 cSt
10 cSt
20 cSt
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 11
4.4 Programming the electronics
a. Tool necessary
Power supply 12-32 V DC
Computer with appropriate USB port and a Windows Operating System (We recommend Windows 7 or higher)
LPU software
Programming cable
Possibly a voltmeter (Ammeter) for the new comparison of the analogue outputs and functional test
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 12
b. Read-out from the electronics
Supplying the electronics with voltage, connecting the programming cable and opening the software.
"COM Port" tab and then choose an appropriate port from the list
Reading out from the electronics with the help of the "Read EE" button.
If the data is not read out correctly, the operation must be repeated.
c. Importing the calibration curve A Trigas calibration file "*.sav" is necessary for the programming
Choose the option "Load F/V kCurve" in the "File" menu – for temperature-compensated programming (Frequency/Viscosity vs.. K-factor) / – Refer to Section 4.3.
Choose optional "Load F kCurve" – for non-temperature compensated programming (Frequency vs. K-factor). After the data has been reloaded, the serial number must be compared with that of the calibrated flow meter.
Optional: Manual entry of the calibration curve is necessary only if the data is not available to you. NOTE The values must be entered with a dot as the decimal point separator. e.g. 15.45 instead of 15,46
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 13
d. Adjusting the electronics
Cut-off (ms) Maximum waiting time for completing a signal period. If this time gets exceeded, no input signal gets detected and all outputs are set to zero.
Update rate (ms) Measurement period before a measurement result is obtained and calculated. Special case: "0" means that the calculation is already taking place after one pulse.
Recommendation: Cut Off = 2 × Update Rate
Timebase (sec.) Time constant for conversion from the K-factor and flow units.
o 1 = xxx/sec. o 60 = xxx/min. o 3600 = xxx/hour
Minimum frequency (Hz) Lowest frequency (Leakage flow rate). Below the frequency, the TriLIN displays zero at all outputs.
Crystal factor Correction factor for internal time base.
Temperature offset Additional compensation factor for temperature correction with normal use (should be 0 for normal use).
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 14
Mode
Operating mode of the electronics.
Mode Compensation Analogue output Frequency output
0 UVC Volumetric Volumetric
1 None Volumetric Volumetric
2 UVC Mass Volumetric
3 None Mass Volumetric
4 UVC Volumetric Mass
5 None Volumetric Mass
6 UVC Mass Mass
7 None Mass Mass
Version Should not be changed.
Pt100 K-factor (bits to Ohm) K-factor of the PT100 input. Should be corrected only when determining deviations of temperature input.
e. Liquid data changes The data pertaining to the media can be corrected whenever desired. To do this, choose the option "Load Fluid" in the "File" menu in order to load the new data. The table can also be edited manually as an option. NOTE The values must be entered with a dot as the decimal point separator. e.g. 15.45 instead of 15,46
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 15
f. Scaling the outputs
Frequency output
You only assign the desired scaling in the fields. You can set the units with "Mode" and "Time Base".
e.g.: Mode – 0: Frequency output, volumetric
Time Base – 60 per minute Flow / Frequency
0-100 l/min. = 0-1000 Hz
Flow, analogue output Mode – 0: Analogue output, volumetric Time Base – 60 per minute
0-100 l/min. = 0-10 V (possibly 4-20 mA) Volts (bits) – A/D value is calculated automatically
Temperature, analogue output -50 – 150°C = 0-10 V (possibly 4-20 mA)
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 16
g. Programming Press the "Write EE" button to program the electronics. After successful programming, the message appears:
Attention! The period of time after pressing the "Write ..." and the "Succes" message should be 3sec. If not, the process should be performed again. Run Live After programming, the program enables the data to be checked. The Live mode is started by pressing the "Run Live" button.
h. Saving data Save the settings in the "File" menu option with "Save All" (*.lin)
If you need further information, please contact us.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 17
i. Printing the data You need Excel 2010 or later to print the data. Open the Excel table enclosed.
Confirm the security warning with "Activate contents" Load the saved data with the "Load LIN" button. Enter the serial number of the electronics. Print out the report.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 18
4.5 Adjusting the analogue outputs
NOTE The adjustment should be carried out only with a calibrated voltmeter or ammeter.
Connect the electronics as you do for programming with the data cable and
power supply.
Connect the voltmeter (or ammeter) to the analogue output to be calibrated.
Read out from the electronics with "Read EE".
There are arrow keys and scroll bars located to the right side of the scaling fields
with the help of which the bit values of the respective analogue outputs can be
changed.
With the upper scroll bar, you can set the minimum value on the multimeter (e.g.
0 V for 0-10 V), and then with the lower scroll bar you can set the maximum
value (e.g. 10 V for 0-10 V). Press STOP at the end -> repeat this for the second
output.
Press the "Write EE" button to program the electronics.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 19
4.6 Adjusting the PT100 input
NOTE You should carry out this adjustment only with a calibrated resistance decade
Connect the electronics as you do for programming with the data cable and
power supply.
Connect the decade at the temperature input and set 100 Ω.
Read out from the electronics with "Read EE"
Enter 100 in the calibration field and press the "Calculate" button – The PT100 K-
factor is recalculated.
Press the "Write EE" button to program the electronics.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 20
4.7 TriLIN pin assignments
Terminal block assignment:
TERMINAL DESCRIPTION TRILIN TRILIN +TEMP.
1 +9-32 V DC supply input
2 0 V
3 Sensor input +
4 Sensor input -
5 Analogue flow output
6 GND
7 Analogue temperature output
8 RTD PT100 supply +
9 RTD PT100 sensor +
10 RTD PT100 sensor -
11 RTD PT100 sensor -
12 0-10 V DC temperature input
13 5 V (TTL pulse) frequency output
14 GND
15 NC 16 NC
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 21
4.8 Trouble shooting
4.8.1 Electronics has no output signal
a. Check supply voltage (9-32V DC)
i. Standard current consumption is (depending on the configuration) up to
40mA.
- Current > 100mA - board defective.
- Current = 0 mA - electronics connected incorrectly.
b. Flowmeter functionality check
i. Attention! The turbine should not be tested with compressed air - turbine may be damaged. ii. Flow simulation: turbines pickoff evaluation – Remove pickoff from the turbine and move an iron element (not stainless steel) under the pickoff back and forth.
iii. Pickup Resistance Measuring - between pin 1 and 8 RF Sensor: 10Ω +/- 15%
c. Check the condition of the electronics (if available) connect electronics
with PC and RUN LIVE Switch Mode
→ Programming - Point 4.4
4.8.2 Baseline value appears to be wrong
a. If a problem is noted - RUN LIVE check mode.
Compare Analog temperature output versus medium (liquid, gas) temperature.
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 22
5. Maintenance The TrigasDM TriLIN is maintenance free. Depending on e.g.:
Type,
System design,
Environmental or operating conditions,
Measuring liquid and
age However, flow meters must be recalibrated and replaced if required. NOTE Independent consulting and calibration services, among others, are provided by TrigasFI. TrigasFI has a certified recording system and an in-house calibration laboratory. The calibration results, which are recorded in a database, form the basis for the determination of customer-specific and application-specific calibration intervals.
6. Declaration of conformity The TriLIN from TrigasDM is not subject to the WEEE directive and complies with the RoHS directive. The TriLIN complies with the EU Directives applicable (EC Declaration of Conformity).
CE: EN50081-1, EN50082-1 and EN61010
TriLIN Electronic linearization with viscosity compensation User manual
TrigasDM GmbH – September 2016 23
7. Warranty
TrigasDM GmbH (supplier) guarantees that all the equipment supplied hereunder is flawless with regard to materials and workmanship, provided that the equipment was selected in accordance with its intended purpose, installed properly and not operated incorrectly. Only the current "General Terms and Conditions" of TrigasDM apply as the T&Cs. You can either request a copy of the terms and conditions by calling the telephone number +49 8165 9999 300, or visit our website at www.trigasdm.com for information.
8. Customer service
If you should require customer service for your TrigasDM products, please contact our customer service department. All requests for information concerning a specific measuring instrument must include the type and serial number of the measuring instrument. We will provide you with all possible assistance over the phone. If your equipment has to be examined or repaired at our plant, whether within the warranty period or after its expiration, our customer service department will issue you with an authorisation number, which is used to initiate our quick and efficient customer service processing. On receipt at our plant, your equipment will be repaired or replaced without delay and returned to you within the shortest period possible.
Please do not return any products without an authorisation number. TrigasDM GmbH Tel.: +49 8165 9999 300 Erdinger Str. 2b Fax: +49 8165 9999 369
Email: [email protected] 85375 Neufahrn, Germany www.trigasdm.com