Transducer Mounting and Test Setup Configurations 1 Rick Bono The Modal Shop

Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Embed Size (px)

Citation preview

Page 1: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Transducer Mounting

and Test Setup

Configurations

1

Rick Bono

The Modal Shop

Page 2: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Transducer Mounting

• Mechanical connection method

– Stud mount

– Adhesive mount

– Magnetic mount

– Press-fit friction mount

• Test parameter considerations

– Frequency range

– Mass loading

2

Page 3: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Mechanical Mounting:

Impact on Frequency Range

3

Sen

sitiv

ity D

evia

tio

n

(dB

) ~

Ref.

100

Hz

Log Frequency (Hz)

StudMount

Adhesive

Mount

Mounting

Pad

Flat

Magnet

Dual Rail

Magnet

Hand

Probe

+40

+30

+20

+10

0

-20

-10

1.0 10 100 1000 10 000 100 000

Page 4: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Stud Mount Transducers

• Best frequency response characteristics –

just like the manufacturer’s cal labs

• Apply silicon grease at mating surface

• Requires surface preparation

• Proper torque recommended

4

Page 5: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Adhesive Mounting Supplies

5

Page 6: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Adhesive Mount Transducers

• Cyanoacrylate (superglue)

– “Instant” adhesive; strong, but still removable

– Gel vs liquid – depends upon surface flatness

– Excellent frequency response characteristics

6

Page 7: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Adhesive Mount Transducers

• Petro wax (bees wax)

– Ultra convenient and simple

– Good for short term testing only

– Frequency response characteristics highly

dependent upon surface prep and amount

7

Page 8: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Adhesive Mount Transducers

• Hot glue

– Allows attachment to poorly-mated surfaces

– Good for short term to mid term testing

– Frequency response characteristics poor, but

generally good enough for modal apps

8

Page 9: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Adhesive Mount Transducers

• Dental cement / fast-cure epoxy

– Allows attachment to poorly-mated surfaces

– Pseudo-permanent attachment for reference

transducer at shaker input location

– Use “disposable” mounting pad with stud

9

Page 10: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Magnetic Mount Transducers

• Extremely convenient

• High attraction forces allow for

reasonable high frequency characteristics

• Available in dual-rail style for attachment

to curved surfaces

10

Page 11: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Press-fit Mount Transducers

• Extremely convenient and efficient

• Designed specifically for low frequency

(<1000 Hz) laboratory modal applications

• Cable base mounts adhesively, modal sensor

mechanically attaches using electrical pins

11

Page 12: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Mounted Accelerometer

Frequency Response Calibration

Mounted Accelerometer Frequency Response

0.0

50.0

100.0

150.0

200.0

250.0

300.0

0 2000 4000 6000 8000 10000 12000

Frequency (Hz)

Sen

sit

ivit

y (

mV

/g)

Cyanoacrylate

Hot Glue

Putty

Petrowax

Stud (handtight w/out grease)

Stud (w/ grease)

12

Page 13: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Mass Loading Considerations

• Acquire FRF with a single accelerometer

• Mount a second accelerometer next to the

first and re-acquire FRF

• Compare for measurable differences

13

Page 14: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Test Setup Considerations

• Understand goals/reasons for performing

experimental modal analysis

– Troubleshooting or failure analysis

– Finite element model verification

– Finite element model correction

– Component substructure / system modeling

14

Page 15: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Test Setup Considerations

• Recognize the 4 primary assumptions of

experimental modal analysis

– Observability

– Time Invariance (Stationarity)

– Linearity

– Maxwell’s Reciprocity

15

Page 16: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Observability Assumption

• Response DOF must have adequate

spatial resolution to represent the modes

of interest

16

Graphics from Agilent

Application Note 243-3

Page 17: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Observability Assumption

17

First bending – beam with seven accelerometer measurement points

Page 18: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Observability Assumption

18

If data acquired only at endpoints… bending is not observable

Page 19: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Observability Assumption

19

Page 20: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Observability Assumption

• Forcing function(s) applied at input

location(s) must adequately excite the

modes of interest

20

Graphic from Agilent

Application Note 243-3

AVOID NODES!

Page 21: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Modally-Tuned Impact Hammers as Pre-

Test Tool for Evaluating Structures...

21

Page 22: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

… for Optimizing Reference

Locations and Ensuring Observability

22

Freq Resp 2:1 82798009.DAT

40dBg/lb

-60

Mag (dB)

kHz10 Hz

Freq Resp 2:1 82798229.DAT

Good Input Location

Bad Input Location

Page 23: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

… for Determining Optimal

Frequency Range

23

Empty

40dBg/lb

-60

Mag (dB)

kHz1.60 Hz

Freq Resp 2:1

• Assure adequate spatial resolution to observe:

- Important, dominant modes

- Necessary modal density

Page 24: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

… for Testing Boundary Conditions

24

Graphic from Agilent

Application Note 243-3

Rule of Thumb: 5-10x separation between rigid

body and flexible modes

Page 25: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Time Invariance Assumption

• Test article (and its boundary conditions) must

exhibit stationarity

– Parameter estimation algorithms assume consistent

global modal properties throughout data set

– Environmental changes during data acquisition cause

shifts in stiffness/damping properties resulting in

measurable shifts in resonant frequencies

– Roving accelerometers to acquire data set results in

variable mass loading on test article

25

Page 26: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Time Invariance Assumption

• DATA CONSISTENCY

• DATA CONSISTENCY

• DATA CONSISTENCY

• i.e. acquire entire data set simultaneously

(single “snapshot”) or at least as fast as

possible

26

Page 27: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Time Invariance Assumption

• Test methodology to achieve best data

consistency

– Simultaneous MIMO/SIMO testing

– Automated bankswitching

– Manual bankswitching

– Roving accelerometers

– Impact testing

27

Page 28: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

28

Benefits of Bank-Switching

Bank-switch, 288 to 112 ch

No. of Test

Configurations

PreSetup

Time

Acquisition

Time

Cost

Estimate

1 9 hrs 17 min 60%

2 17 min

3 17 min

4 17 min

Total Time

Allotted

10 hrs

8 min

Roving, 112 ch

No. of Test

Configurations

PreSetup

Time

Acquisition

Time

Cost

Estimate

1 3 hrs 6 hrs 17 min 40%

2 6 hrs 17 min

3 6 hrs 17 min

4 6 hrs 17 min

Total Time

Allotted

28 hrs8 min

Simultaneous, 288 ch

No. of TestConfigurations

PreSetupTime

AcquisitionTime

CostEstimate

1 9 hrs 5 min 100%

2 5 min

3 5 min

4 5 min

Total Time

Allotted

9 hrs

20 min

Page 29: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Bank-Switching Example

• Inputs: 2 vertical, 1 lateral, 1 skewed

29

Page 30: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Bank-Switching Example

• Response points: 17 patch panels, each

bank of 16 accelerometers

30

Page 31: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Bank-Switching Example

• Bank-switch patches of data (3 x 96 ch)

into smaller data acquisition system

31

Page 32: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Modular Cabling / Patch Panel

System for Clean Setup

32

• Eases setup troubleshooting

• Eliminates messy “rat’s nest” of cables

• Economical multi-conductor cabling

Page 33: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Time Invariance Assumption

• Roving accelerometers results in inconsistent

global resonant frequencies due to variable

mass loading on test structure

33

Graphic from Agilent

Application Note 243-3

Page 34: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Linearity Assumption

• Input and output characteristics remain

proportional within measurement range

• Confirm using precisely controlled inputs

from shaker(s) across a range force levels

• Impact testing technique poorly suited

when dealing with nonlinear test

structures

34

Page 35: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Electrodynamic Modal Shakers as

Excitation Source for MIMO• Allows best control of

input forcing function to optimize frequency content and signal-to-noise ratio

• Through-hole armature greatly simplifies setup attachment to test structure

35

Page 36: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Through-Hole Armature

Eases Setup

• Traditional shakers

with tapped armature

connection leave

little tolerance since

setup has tapped

connection at both

ends

36

Page 37: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Reciprocity Assumption

• Maxwell’s Theory of Reciprocity states

that FRF matrix is symmetric

• FRF between input A and output B is the

same as output A and input B

• Confirm using multiple shaker locations

and impedance heads for driving point

measurement

37

Page 38: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Impedance Heads for Verifying

Reciprocity Assumption

38

Freq Resp 2:1

30dBg/lb

-70

Mag (dB)

Hz500 0 Hz

Freq Resp 2:1

Accelerometer built

into preload stud of

force transducer

Page 39: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Other Pre-Test Considerations

• Free Boundary Conditions

– Shock Cord

– Foam Rubber

– Air Suspension

39

Page 40: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Other Pre-Test Considerations

• Fixed Boundary Conditions

• Realistic Boundary Conditions

• Match Impedance(s) at Boundaries

• Mass Loaded Boundary Conditions

40

Page 41: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Other Pre-Test Considerations

• Transducer selection

– Single axis vs triaxial package

– Sensitivity, measurement range & resolution

– Frequency range & mass

41

Page 42: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Transducer Electronic Data Sheet

(TEDS, IEEE 1451.4)

42

TEDS

Memory

To ICP Supply

4ma

-5 V

• Identifies transducer (type, serial number, location)

• Stores calibration data

• Automates book keeping, reducing errors

Uses reverse bias scheme

to access digital memory

Page 43: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Other Transducer Setup

Considerations

• Use PDA scanner with bar-coded TEDS

transducers to ease bookkeeping

43

Page 44: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

44

Final Channel Setup Definition

• Combine Data From Geometry, PDA, and

TEDS

• Complete Test Set-up Information

Defined in Universal Files– Virtual Channel Table (1807)

– Channel Table (1808)

– Geometry (15)

ParameterStored In

TEDS

Stored In

PDA

Stored On

Host (PC)

Calibration X

Model / Serial No. X X

Direction X

Node No. X X

Meas. Ch. X

Geometry X

Page 45: Transducer Mounting and Test Setup Configurations Test Setup.pdf · Transducer Mounting and Test Setup Configurations 1 ... mass loading on test structure 33 ... • Identifies transducer

Thank you for your time.