GUS Training Guide For Transesophageal (TEE) and Vaginal/Rectal (VR) Ultrasound Probes 1 2 3 4 5 Dirty Transportation Pre-Clean High Level Disinfection Clean Transportation Storage CONFIDENT COMPLIANCE GUS®

Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

GUS Training GuideFor Transesophageal (TEE) and Vaginal/Rectal (VR) Ultrasound Probes

1

2

3

4

5

Dirty Transportation

Pre-Clean

High Level Disinfection

Clean Transportation

Storage

C O N F I D E N T C O M P L I A N C E

GUS®

Page 2: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

Supplies and Environment

Supplies for reprocessing are readily available:

PPE, as defined by HLD manufacturer and facility procedures.

Pre-Clean/Drying Supplies, as defined by probe manufacturer and facility procedures.

High-level disinfection supplies compatible with GUS:

HLD Solution - OPA, Glutaraldehyde, or Hydrogen Peroxide

MRC test strips

Area should be clean and free from clutter as defined by facility procedures.

Daily Checks / Prior to Operation of GUS

Visually inspect electrical connections, fan operation, and exterior of the unit daily.

Transportation of Dirty Probe

Post-exam, transport ultrasound probe according to the manufacturer’s IFU.

For TEE probes using CIVCO’s transport tray, place the dirty probe in the clean tray & flip lid to dirty. Transport probe to reprocessing location and dispose of tray.

Pre-clean

Reference your ultrasound probe manufacturers’ IFU for pre-cleaning instructions.

Visually inspect entire ultrasound probe to ensure it is clean.

CIVCO offers dry and enzymatic sponges, disinfecting wipes, and lint-free cloths to aid a compliant pre-clean process.

High-Level Disinfection

Make sure the GUS Disinfection Soak Station is ON before opening door and removing caps.

Fill the GUS containers with an approved HLD.

Take temperature using thermometer. Log results onto GUS Disinfection Soak Station Log attached to the end of this training guide.

Simplify your TEE probe cleaning process by contacting 800.445.6741.

Pre-Cleaning Sponges & Wipes

REF #610-1351

REF #610-1352

REF #610-1353

REF #610-1354

REF #610-1418 and #610-1471

Thermometers for High-Level Disinfection

REF #610-1290

Page 3: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

Dip one MRC test strip into the HLD, holding it submerged for length specified in the HLD manufacturer’s IFU. Remove the test strip.

After waiting the specified time indicated on the manufacturer’s bottle, compare the test strip to the color chart on the manufacturer’s bottle. Log results onto GUS Disinfection Soak Station Log attached to the end of this training guide.

NOTE: A FAIL result requires a new bottle of HLD. See Disinfectant Disposal Section of this training guide.

Gently insert probe into HLD container by following the directions in the operator’s manual for proper insertion.

Use timer to calculate soak time. Soak time is dependent on your high-level disinfectant manufacturer’s instructions for use.

For TEE probes only, conduct leak testing according to manufacturer’s instructions. Log results onto GUS Disinfection Soak Station Log attached to the end of this training guide.

After soak time, the first rinse is performed in the “initial rinse” container in the GUS station to avoid off-gassing and drips.

Some disinfectants require more than one rinse. Follow the probe and disinfectant manufacturer’s rinsing instructions.

After rinsing, carefully remove and dry the probe according to the manufacturer’s instructions.

Countdown Timer for GUS Disinfection Soak Stations

REF #610-2049

Select Your Leak Test KitTo perform leak testing with GUS, you need a leakage tester meter and conductivity probe, and an ultrasound leakage tester adapter compatible with your ultrasound system.

Leak Test Meter: #610-2054

Conductivity Probe for Manual Reprocessor: #610-2244

TrueCare Sterile Lint-Free Wipes

REF #610-1354

Transportation of Clean Probe

Post disinfection, transport ultrasound probe according to the manufacturer’s IFU.

For TEE probes using CIVCO’s transport tray, use a clean tray to transport the disinfected probe to its storage cabinet.

Simplify your TEE probe cleaning process by contacting 800.445.6741.

REF #610-1418 and #610-1471

Page 4: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

Storage of Clean Probe

REF #610-1323 REF #610-1327

“Store the device in a manner that will protect from damage or contamination… Hanging vertically in a cabinet and storing in a clean environment.” – The Joint Commission

Store ultrasound probe according to the manufacturer’s IFU

GUS Maintenance

Disinfectant Disposal

While the GUS is ON, follow the HLD manufacturer’s instructions for use for neutralizing the HLD prior to disposal.

Wash and rinse container thoroughly following the cleaning instructions in the Operator’s Manual to remove any residual neutralizer.

Refill HLD at the GUS station while it is ON. Pour HLD into the container up to the maximum fill level established.

Test Strip Quality Control

Quality Control must be performed when a new bottle of test strips is opened. Test strips can be used for the number of days after opening specified by the manufacturer or the expiration date, whichever occurs first. Discard any test strip bottle left open longer than allowed by manufacturer’s instructions.

Filter Replacement

Change Filter every 6 months following directions in Operator’s Manual.

DISCLAIMER:

The following information highlights topics and recommendations presented during this training.

Please refer to your GUS Disinfection Soak Station for TEE Probes or GUS Disinfection Soak Station for Endocavity Probes Operators Manual for specific details related to the care and reprocessing of your ultrasound probes.

Storage Cabinets

TEE Probe Cabinet: #610-1323

General Purpose & Endocavity Probe Cabinet: #610-1327

Storage Cabinet replacement HEPA filter: #610-1328

REF #610-2307

REF #610-1337

REF #610-714

Page 5: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

GUS® DISINFECTION SOAK STATION™ LOGFor Ultrasound Probes

GUS Model Number: GUS Serial Number: Location/Dept:

GUS Filter Install Date: GUS Filter Change Date: QC Test Strip Result (Initials/Date):

Date Test Strip Bottle Opened: Expiration Date/Open Shelf Life: Test Strip Bottle Lot #:

Date Disinfectant Opened: Expiration Date/Open Shelf Life: Disinfectant Bottle Lot #:

Disinfectant Test Date

Disinfectant Test Time Temperature Initials MRC Test:

Pass/FailElectrical Leak Test: Pass/Fail Comments

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Pass Fail Pass Fail

Page 6: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

Operator Training Log

GUS Operator ID / Facility ID Name Trained Date Trained By Notes

GUS®

Page 7: Training Guide - CIVCO · e sure the GUS Disinfection Soak Station is Mak ON before opening door and removing caps. Fill the GUS containers with an approved HLD. ake temperature using

CUSTOMER, NOT CIVCO, IS RESPONSIBLE FOR CORRECTLY PERFORMING THE STEPS DOCUMENTED IN THIS TRAINING. IN NO EVENT WILL CIVCO BE LIABLE FOR ANY SPECIAL, INCIDENTAL, CONSEQUENTIAL, PUNITIVE, EXEMPLARY OR OTHER INDIRECT LOSSES OR DAMAGES OF ANY KIND, INCLUDING BUT NOT LIMITED TO LOSS OF USE, LOSS OF BUSINESS OR PROFITS, OR LOSS OF OR DAMAGE TO DATA, HOWEVER CAUSED, WHETHER BASED ON CONTRACT, TORT, STRICT LIABILITY, OR OTHER LEGAL THEORY, WHETHER FORESEEABLE OR NOT, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. CIVCO’S TOTAL LIABILITY FOR DIRECT DAMAGES OR OTHERWISE WILL NOT EXCEED THE TOTAL AMOUNTS PAID OR PAYABLE FOR THE PRODUCT(S) THAT GAVE RISE TO THE CLAIM. THESE LIMITATIONS WILL APPLY NOTWITHSTANDING THE FAILURE OF THE ESSENTIAL PURPOSE OF ANY LIMITED REMEDY. BUYER ACKNOWLEDGES THAT THESE LIMITATIONS OF LIABILITY ARE A MATERIAL PART OF THE BARGAIN BETWEEN THE PARTIES AND ARE REFLECTED IN PRODUCT PRICING, WHICH WOULD BE HIGHER WITHOUT THESE LIMITATIONS.

COPYRIGHT © 2020. ALL RIGHTS RESERVED. CIVCO IS A REGISTERED TRADEMARK OF CIVCO MEDICAL SOLUTIONS. GUS IS A TRADEMARK OF CIVCO. ALL OTHER TRADEMARKS ARE THE PROPERTY OF THEIR RESPECTIVE OWNERS. ALL PRODUCTS MAY NOT BE LICENSED IN ACCORDANCE WITH CANADIAN LAW. PRINTED IN USA. 2019R-3150 REV 1

C I V C O. C O M | I N F O @ C I V C O. C O M | 1 . 8 0 0 . 4 4 5 . 6 7 4 1