Analytica Chimica Acta 820 (2014) 76–83 Contents lists available at ScienceDirect Analytica Chimica Acta journal homepage: www.elsevier.com/locate/aca Total microcystins analysis in water using laser diode thermal desorption-atmospheric pressure chemical ionization-tandem mass spectrometry Audrey Roy-Lachapelle a , Paul B. Fayad a , Marc Sinotte b , Christian Deblois c , Sébastien Sauvé a,a Department of Chemistry, Université de Montréal, Montréal, QC, Canada b Direction du Suivi de l’État de l’Environnement (DSEE), Ministère du développement durable, de l’environnement, de la faune et des parcs (MDDEFP), Québec, QC, Canada c Centre d’expertise en analyse environnementale (CEAEQ), Ministère du développement durable, de l’environnement, de la faune et des parcs (MDDEFP), Québec, QC, Canada highlights Total microcystins are analyzed in water using the MMPB oxidation product. Samples are subjected to a simple oxidation and analyzed in 15 s. Limits of detection for total microcystins are 0.2 gL 1 in field-collected water samples. The calibration curve showed good linearity (R 2 > 999). Interday and intraday variation coefficients were below 15%. graphical abstract article info Article history: Received 1 December 2013 Received in revised form 13 February 2014 Accepted 15 February 2014 Available online 19 February 2014 Keywords: Total microcystins LDTD APCI Mass spectrometry Cyanobacterial bloom abstract A new approach for the analysis of the cyanobacterial microcystins (MCs) in environmental water matri- ces has been developed. It offers a cost efficient alternative method for the fast quantification of total MCs using mass spectrometry. This approach permits the quantification of total MCs concentrations without requiring any derivatization or the use of a suite of MCs standards. The oxidation product 2-methyl-3- methoxy-4-phenylbutyric acid (MMPB) was formed through a Lemieux oxidation and represented the total concentration of free and bound MCs in water samples. MMPB was analyzed using laser diode thermal desorption-atmospheric pressure chemical ionization coupled to tandem mass spectrometry (LDTD-APCI-MS/MS). LDTD is a robust and reliable sample introduction method with ultra-fast analysis time (<15 s sample 1 ). Several oxidation and LDTD parameters were optimized to improve recoveries and signal intensity. MCs oxidation recovery yield was 103%, showing a complete reaction. Internal calibration with standard addition was achieved with the use of 4-phenylbutyric acid (4-PB) as internal standard and showed good linearity (R 2 > 0.999). Limits of detection and quantification were 0.2 and 0.9 gL 1 , respec- tively. These values are comparable with the WHO (World Health Organization) guideline of 1 gL 1 for total microcystin-LR congener in drinking water. Accuracy and interday/intraday variation coefficients were below 15%. Matrix effect was determined with a recovery of 91%, showing no significant signal sup- pression. This work demonstrates the use of the LDTD-APCI-MS/MS interface for the screening, detection and quantification of total MCs in complex environmental matrices. © 2014 Elsevier B.V. All rights reserved. Corresponding author at: Université de Montréal, Departement of Chemistry, CP 6128 Centre-Ville, Montréal, Qc, H3C 3J7, Canada. Tel.: +1 514 343 6749; fax: +1 514 343 7586. E-mail addresses: [email protected], [email protected] (S. Sauvé). http://dx.doi.org/10.1016/j.aca.2014.02.021 0003-2670/© 2014 Elsevier B.V. All rights reserved.

Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

Tds

ASa

b

Qc

Q

h

a

ARRAA

KTAMC

f

h0

Analytica Chimica Acta 820 (2014) 76–83

Contents lists available at ScienceDirect

Analytica Chimica Acta

journa l homepage: www.e lsev ier .com/ locate /aca

otal microcystins analysis in water using laser diode thermalesorption-atmospheric pressure chemical ionization-tandem masspectrometry

udrey Roy-Lachapellea, Paul B. Fayada, Marc Sinotteb, Christian Debloisc, Sébastienauvéa,∗

Department of Chemistry, Université de Montréal, Montréal, QC, CanadaDirection du Suivi de l’État de l’Environnement (DSEE), Ministère du développement durable, de l’environnement, de la faune et des parcs (MDDEFP),uébec, QC, CanadaCentre d’expertise en analyse environnementale (CEAEQ), Ministère du développement durable, de l’environnement, de la faune et des parcs (MDDEFP),uébec, QC, Canada

i g h l i g h t s

Total microcystins are analyzed inwater using the MMPB oxidationproduct.Samples are subjected to a simpleoxidation and analyzed in 15 s.Limits of detection for totalmicrocystins are 0.2 �g L−1 infield-collected water samples.The calibration curve showed goodlinearity (R2 > 999).Interday and intraday variationcoefficients were below 15%.

g r a p h i c a l a b s t r a c t

r t i c l e i n f o

rticle history:eceived 1 December 2013eceived in revised form 13 February 2014ccepted 15 February 2014vailable online 19 February 2014

eywords:otal microcystins LDTDPCIass spectrometry

yanobacterial bloom

a b s t r a c t

A new approach for the analysis of the cyanobacterial microcystins (MCs) in environmental water matri-ces has been developed. It offers a cost efficient alternative method for the fast quantification of total MCsusing mass spectrometry. This approach permits the quantification of total MCs concentrations withoutrequiring any derivatization or the use of a suite of MCs standards. The oxidation product 2-methyl-3-methoxy-4-phenylbutyric acid (MMPB) was formed through a Lemieux oxidation and represented thetotal concentration of free and bound MCs in water samples. MMPB was analyzed using laser diodethermal desorption-atmospheric pressure chemical ionization coupled to tandem mass spectrometry(LDTD-APCI-MS/MS). LDTD is a robust and reliable sample introduction method with ultra-fast analysistime (<15 s sample−1). Several oxidation and LDTD parameters were optimized to improve recoveries andsignal intensity. MCs oxidation recovery yield was 103%, showing a complete reaction. Internal calibrationwith standard addition was achieved with the use of 4-phenylbutyric acid (4-PB) as internal standard andshowed good linearity (R2 > 0.999). Limits of detection and quantification were 0.2 and 0.9 �g L−1, respec-

−1

tively. These values are comparable with the WHO (World Health Organization) guideline of 1 �g L fortotal microcystin-LR congenerwere below 15%. Matrix effectpression. This work demonstraand quantification of total MC

∗ Corresponding author at: Université de Montréal, Departement of Chemistry, CP 6128ax: +1 514 343 7586.

E-mail addresses: [email protected], [email protected] (S.

ttp://dx.doi.org/10.1016/j.aca.2014.02.021003-2670/© 2014 Elsevier B.V. All rights reserved.

in drinking water. Accuracy and interday/intraday variation coefficients

was determined with a recovery of 91%, showing no significant signal sup-tes the use of the LDTD-APCI-MS/MS interface for the screening, detection

s in complex environmental matrices.© 2014 Elsevier B.V. All rights reserved.

Centre-Ville, Montréal, Qc, H3C 3J7, Canada. Tel.: +1 514 343 6749;

Sauvé).

Page 2: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

tica Ch

1

dfbidotk9[atttvsriopttw

icmftaccwabatetmpaatatftmtb

A. Roy-Lachapelle et al. / Analy

. Introduction

Microcystins (MCs) are cyclic heptapeptide hepatotoxins pro-uced by cyanobacteria (blue-green algae), and are the mostrequently observed cyanobacterial toxins [1]. A cyanobacterialloom can occur in surface waters rich in nutrients and the tox-

ns can be released from the cells to the natural water or even inrinking water reservoirs [2,3]. The cyclic structure of MCs consistsf uncommon amino acids, of which two vary from one specieso another (X and Z in Fig. 1) thus potentially generating over 80nown MCs structures [2,4]. The �-amino acid Adda (3-amino--methoxy-2,6,8-trimethyl-10-phenyldeca-4(E), 6(E)-dienoic acid5]) is the only part common to all MCs hepatotoxin congeners,nd is responsible for their toxicity. This toxicity is caused by theoxin’s ability to inhibit specific protein phosphatases in animalissues, and it is aggravated by its tendency to bioaccumulate inhe liver, thus potentially leading to liver failure [6,7]. MCs LD50alues range from 45 to 1000 �g kg−1 for mice, depending of thepecific congener tested [8]. The World Health Organization (WHO)ecommends an upper limit of 1 �g L−1 for microcystin-LR in drink-ng water, the most studied and thought to be the most frequentlyccurring MC [9]. In a recent study, it was mentioned that it wasossible to observe significant cyanobacterial bloom within a waterreatment facilities, suggesting significant risks of toxin release,hus potentially contributing to high MC concentrations in drinkingater [10].

There are many techniques available for the screening of MCsn natural water. Solid phase extraction for preconcentration andleanup and quantification using high-performance liquid chro-atography coupled with mass spectrometry (HPLC-MS) is used

or the specific analysis of several MCs species [7,11–13]. However,his analysis is limited to the MCs for which MCs standards are avail-ble. Currently, HPLC-MS analysis can detect roughly 10 differentongeners over more than 80 possible forms. Gas chromatographyoupled with mass spectrometry is used for the analysis of total MCsith the oxidation product 2-methyl-3-methoxy-4-phenylbutyric

cid (MMPB) [14–16]. Since the toxin itself is not volatile enough toe separated by gas chromatography, an oxidation reaction prior tonalysis is necessary to obtain the volatile compound MMPB fromhe Adda structure common to all the different MC congeners. How-ver, the technique requires an extra derivatization step to makehe MMPB more volatile using an ester addition. Traditional chro-

atography techniques are also time consuming, requiring samplere-treatment, solid phase extraction and a chromatographic sep-ration which takes several minutes for each analysis, as wells relying on a restricted number of available standards and notaking into account other MCs congeners. In mass spectrometrynalysis, MALDI-TOF-MS has also been used recently for the iden-ification of microcystins. However, the technique can be only usedor qualitative purposes and the possibility for automated rou-

ine HPLC-MALDI interface is still limited for the quantification of

icrocystins [17]. Finally, the most commonly used technique forhe nearly instant detection of MCs is enzyme-linked immunosor-ent assays (ELISA) [12,18–20]. This technique is fast and sensitive,

Fig. 1. Reaction scheme of Lemieux oxidation reaction to pr

imica Acta 820 (2014) 76–83 77

but encounters cross-selectivity which tends to overestimate MCsconcentration. The ELISA method mostly uses the microcystin-LRstandard, so the quantification assumes that every congener isreacting like MC-LR and results are expressed in MC-LReq.

The laser diode thermal desorption-atmospheric pressurechemical ionization interface coupled to tandem mass spectrom-etry (LDTD-APCI-MS/MS) is an alternative technique proposedfor the quantitative analysis of total MCs. The LDTD is a sampleintroduction method using thermal desorption, thus eliminatingthe use of chromatographic separation prior to mass spectrome-try detection. This approach results in ultra-fast sample analysis(<15 s sample−1) with simple sample preparation, reducing timeand material costs otherwise required for chromatography whilealso reducing solvent consumption. The LDTD method was previ-ously developed for the analysis of a cyanobacterial neurotoxin,anatoxin-a, in water matrices [21]. It was also developed for severalpharmaceutical and pesticide compounds in different environ-mental matrices including wastewater, sludge, sediments and soilsamples [22–26]. The schematic and assembly of the LDTD-APCIsource apparatus have been detailed previously [27].

The aim of this study is to develop and validate a new methodusing the LDTD-APCI-MS/MS apparatus allowing for a simple, rapidand high-throughput detection and quantification of total MCs incomplex environmental water matrices. MCs oxidation and MMPBextraction were optimized as a function of the MCs reaction yieldand MMPB recovery yield. Several LDTD parameters were stud-ied in order to optimize the thermal desorption and enhance thecompound signal: the solvent position, the deposition volume,the laser power, the laser pattern and the carrier gas flow. Themethod was validated using 4-phenylburytic acid (4-PB) as theinternal standard. The method validation was done by evaluat-ing the detection and quantification limits (MDL and MQL), lineardynamic range, accuracy, precision and matrix effect. This ultrafasttechnique could be used for the quicker screening and quantifica-tion of MCs applied for environmental and public health purposes.

2. Materials and methods

2.1. Chemicals, reagents and stock solutions

MMPB (2-methyl-3-methoxy-4-phenylbutyric acid) sodiumsalt standards, 94%, were purchased from Wako Pure Chemi-cal Industries, Ltd. (Osaka, Japan). 4-PB (4-phenylbutyric acid)standard, 99%, was obtained from Sigma-Aldrich (Oakville, ON,Canada). Sodium (meta) periodate (purity ≥ 99,0), sodium bisulfate(A.C.S. reagent), potassium carbonate (purity ≥ 99.0) and sulfu-ric acid standard solution (1.000 mol L−1) were obtained fromSigma–Aldrich (Oakville, ON, Canada). Potassium permanganate(A.C.S. reagent) was obtained from Biopharm (Montreal, QC,Canada). Methanol (MeOH) and ethyl acetate (EtAc) were of ana-

lytical grade purity from Fisher Scientific (Whitby, ON, Canada).Deionized/distilled water (dd-H2O) was used for dilution. Indi-vidual stock standard solutions were prepared in MeOH at aconcentration of 100 mg −1 and kept at −20 ◦C for 12 months for the

oduce the MMPB used for total microcystin analysis.

Page 3: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

7 tica Ch

MmF

2

plopauprLqrM

2

cm0wbaTtimtafatutuatt2L

2

pem4dc4trisPa

The method validation was done according to the recommen-dation of validation protocol for environmental chemistry analysisfrom the Quebec’s MDDEFP guidelines [28]. Data analysis was

Table 1Selected reaction monitoring (SRM) optimal parameters for the MMPB analysis byAPCI-MS/MS in negative ionization mode (NI) −→ [M]− .

Compound Precursor ion −→production (m/z)

Collision energy(V)

Tube lens(V)

MMPB 207.1 −→ 131.1a −15 −39207.1 −→ 175.1b −13

8 A. Roy-Lachapelle et al. / Analy

MPB standards with no significative degradation (P > 0.5), and aaximum of 6 months for the 4-PB standards (see supplementary,

ig. A-1).

.2. Cyanobacterial bloom samples

Environmental samples were provided from the monitoringrogram realized by the Ministère du Développement Durable, de

’Environnement, de la Faune et des Parcs (MDDEFP – The provincef Quebec Ministry of the Environment) (Québec, Canada). The sam-les were collected from lakes encountering cyanobacterial bloomsnd were sampled in glass bottles that were kept in the dark at 4 ◦Cntil analysis. The lakes were targeted in different regions from therovince of Québec, Canada, and were chosen for their high occur-ence of cyanobacterial blooms. Microcystins (HilR, HtyR, LA, LF,R, LR (D-Asp3), LW, LY, RR, RR (D-Asp3), WR, YR) were previouslyuantified by LC-MS/MS by the Centre d’Expertise en Analyse Envi-onnementale du Québec (CEAEQ), the analytical services of theDDEFP.

.3. Optimization of Lemieux oxidation and MMPB extraction

Microcystins (MCs) were obtained from a cyanobacterial bloomulture where the supernatant was sampled. The cells were sub-itted to freeze–thaw lysis and then filtered with nylon filters of

.2 �m obtained from Whatman (Florham Park, NJ, USA). Solutionsere kept at −20 ◦C for 6 months. The solutions were analyzed

y LC-MS/MS and the MCs (MC-RR, MC-YR, MC-LR, MC-LY, MC-LWnd MC-LF) were quantified with a total concentration of 22 mg L−1.hese MCs were used to optimize the Lemieux oxidation parame-ers. The reaction scheme of the Lemieux oxidation is presentedn Fig. 1. The MCs solutions were spiked into a river water blank

atrix to obtain a total MCs concentration of 150 �g L−1. All reac-ions were conducted with KMnO4 (50 mM) and NaIO4 (50 mM) inlkaline conditions (pH 9) by adding K2CO3, at room temperatureor 1 h, in the absence of light. The reaction was quenched withsaturated sodium bisulfite solution until all the purple color of

he solution disappeared. A solution of 10% sulfuric acid was addedntil the pH became acidic (∼2) [15]. Each reaction test consisted ofhree replicates and all solutions were conducted with a total vol-me of 1 mL. A volume of 100 �L of a saturated NaCl solution wasdded in the aqueous phase to enhance the liquid–liquid extrac-ion. The liquid–liquid extraction was done with 1 mL of EtAc, andhen the solutions were submitted to a vortex at 2500 rpm formin. The supernatants were then used directly for analysis byDTD-APCI-MS/MS.

.4. Oxidation recovery yield and matrix effect

The sample oxidation recovery yield was measured as the pro-ortion of MCs converted to MMPB and extracted by liquid–liquidxtraction. Recovery was determined by spiking blank environ-ental water matrix with a known amount of MCs (120 �g L−1 and

83 �g L−1) and of internal standard (4-PB – 150 �g L−1) before oxi-ation and compared to the same blank samples spiked with theorresponding amount of MMPB (25 �g L−1 and 100 �g L−1) and-PB (150 �g L−1) after the oxidation and the liquid–liquid extrac-ion with ethyl acetate. The concentration ratios were expressed asecovery percentage. The matrix effect was determined by compar-

ng blank environmental water matrix samples with pure solventamples both spiked with MMPB (25 �g L−1 and 100 �g L−1) and 4-B (150 �g L−1). The concentration ratios gave the signal recoverynd were expressed as percentage.

imica Acta 820 (2014) 76–83

2.5. LDTD-APCI-MS/MS

Desorption and ionization of MMPB and 4-PB were achievedwith the T-960 LDTD-APCI interface model, controlled by theLazSoft 4 Software (Phytronix Technologies, Quebec, Qc, Canada)integrated with the Excalibur 2.0 software (Thermo Fisher Scien-tific, Waltham, MA, USA). Negative mode ionization was achievedwith the APCI source. Compound detection was performed using aQuantum Ultra AM triple quadrupole mass spectrometer (ThermoFisher Scientific, Waltham, MA). Sample solution aliquots for theextraction supernatants (2 �L) were spotted into a LazWell 96-wellplate containing inserts made of a thin proprietary stainless steelalloy. The samples were heated at 40 ◦C for 5 min until completeevaporation of solvent. Details on the operation of the LDTD appa-ratus [27] (Fig. A-2) were described in previous studies [21,23,27].Briefly, an infrared laser (980 nm, 20 W, continuous) heats the backof the hexagonal shaped concave well, and a thermal desorptionof the compounds occurs. The uncharged sample analytes, in gasphase, are transported through a transfer tube by a carrier gas(medical grade purified air) and are then ionized in the APCI coronadischarge. The ionized analytes are then transported to the MSdetector.

The optimization of the LDTD-APCI parameters was conductedby adding 2 �L of a MMPB (250 �g L−1) and 4-PB (150 �g L−1) solu-tions made from blank environmental water matrix subjected tooxidation and extraction to take into account matrix effects. TheLDTD-APCI desorption parameters were optimized to the follow-ing settings: the maximum laser power used was 35% for the laserpattern (2 s at 0%, 2 s ramping from 0% to 35%, 0.1 s from 35% to0%, and 2 s at 0% for a total of 6 s), the carrier gas flow rate was2 L min−1 with a temperature of 50 ◦C. The ion source parameterswere set with the following settings: ion sweep gas 0.3 (arbitraryunit) and sheath gas, auxiliary gas, skimmer offset and vaporizertemperature were set to 0. Capillary temperature was set to 350 ◦Cwith a discharge current of 4 �A in NI mode.

Selected reaction monitoring (SRM) mode was used, and twoproduct ions for MMPB and 4-PB were selected as quantificationand confirmation ions. The quantification ion, with the highestsignal intensity, was used to determine the limits of detectionand quantification. The relative intensity ratios of product ions,with constant values, were used for confirmation to the targetedcompound and to avoid false positives. Optimization of MS/MSparameters was done with the following settings: collision gas (Ar)pressure at 1.5 mTorr, resolution for Q1/Q3 was set at 0.7 amu andscan time was set at 0.005 s. The different SRM transitions andtheir optimized parameters for the quantification and confirma-tion of MMPB are shown in Table 1. MMPB and 4-PB structures arepresented with their respective fragments in Fig. A-3.

2.6. Method validation

4-PB 163.1 −→ 91.1a −16 −22163.1 −→ 119.1b −12

a Quantification ion.b Confirmation ion.

Page 4: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

tica Ch

pWmfLalttrsqPpaobao

eft

3

3

tl

Fcr

A. Roy-Lachapelle et al. / Analy

erformed with LCQuan 2.5 software (Thermo Fisher Scientific,altham, MA, USA). Internal calibration with standard additionethod was used with a seven-point calibration curve, ranging

rom 1 to 500 �g L−1 in triplicates with a bloom matrix submitted toemieux oxidation, liquid–liquid extraction and spiked with MMPBnd 4-PB. Method detection limit (MDL) and method quantificationimit (MQL) were evaluated as 3 and 10 times the standard devia-ion of the mean value of 10 matrix samples spiked with about 10imes the estimated detection limit (we targeted 10 �g L−1). Accu-acy was determined by the relative error (RE – %) using the MCpiked bloom sample previously quantified by LC-MS/MS and thenuantified by internal calibration with standard addition method.recision (interday/intraday) and accuracy were expressed as theercentage relative standard deviation (RSD), and were then evalu-ted using concentrations of 25 �g L−1, 100 �g L−1 and 250 �g L−1

f MMPB as quality control (QC). The solutions were spiked in alank river water matrix submitted to oxidation and extraction andnalyzed 6 times with internal standard (IS) with a concentrationf 150 �g L−1.

For a statistical comparison, the Statistical Package for Social Sci-nce (SPSS 17.0, Chicago, IL) for Windows was used, when needed,or the ANOVA test as well as the Tukey’s post-hoc test with statis-ical significance defined as P < 0.05.

. Results and discussion

.1. Microcystins oxidation and MMPB extraction optimization

All the Lemieux oxidation parameters were optimized in ordero have the best reaction efficiency. It was shown that weakly alka-ine conditions (pH ∼9) were the most efficient for MC oxidation.

ig. 2. Different MCs oxidation parameters optimized in order to enhance peak area of theoncentrations (potassium permanganate and sodium periodate) between 10 mM and 1epresent standard deviations from the mean (n = 6).

imica Acta 820 (2014) 76–83 79

The different pH conditions during the oxidation reaction weretested between 2 and 10.5; pH 9 was confirmed as the best value(Fig. 2a). Potassium permanganate and sodium periodate concen-trations, as well as reaction time, were optimized in order to havethe best reaction yield. Each oxidant was tested with a concentra-tion between 10 mM and 100 mM, and for each concentration thereaction lasted between 0.5 h and 5 h. The optimal concentrationwas set at 50 mM for a reaction time of 1 h (Fig. 2b). The use ofa higher oxidant concentration could induce compound degrada-tion and provoke higher signal variability. Between 1 and 3 h foreach concentration, no significant reaction yield enhancement wasobserved (P > 0.05) and it was decided to choose the fastest reac-tion time in order to minimize sample preparation time. It wasalso important to choose a reaction time that would not inducecompound degradation, which was observed after 3 h with higheroxidant concentrations (Fig. 2b). Finally, in several studies, it wasshown that the temperature was not a significant parameter whenmaintained higher than 25 ◦C for the Lemieux oxidation reactionyield [29]. These results were similar to those of several previousstudies [15,16,30,31]. Thus, the optimal reaction conditions usedfor the determination of total MCs in natural water matrices wereas follows: an alkaline condition of pH 9, 50 mM of KMnO4 andNaIO4 each, and a total reaction time of 1 h at room temperature.

A liquid–liquid extraction was used after MCs oxidation toextract the formed MMPB. Several organic solvents were testedaccording to their non-miscibility with water; EtAc, hexane, cyclo-hexane, heptane, diethyl ether, methylene chloride and chloroform.

However, only EtAc and hexane showed good recovery values forMMPB extraction. Finally, EtAc was chosen as a compromise withthe optimal deposition solvent for the LDTD (Fig. 3). To enhanceextraction recovery of MMPB and 4-PB, the addition of a solution of

produced MMPB (a) effect of pH between 2 and 10.5, (b) effect of oxidative reagent00 mM depending on the reaction time between 0.5 h and 5 h. Vertical error bars

Page 5: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

80 A. Roy-Lachapelle et al. / Analytica Ch

FLs

ssc[2tp

3

otl

−1

F

ig. 3. Effect of using different solvents for the deposition of the analytes in theDTD Lazwell plate wells with MMPB at 250 �g L−1. Vertical error bars representtandard deviations from the mean (n = 6).

aturated NaCl in the aqueous phase was tested (Fig. A-4). By addingaturated salt in the aqueous phase of the reaction solution, neutralompounds are less soluble and the organic extraction is enhanced32]. In addition, since the oxidative solution was adjusted to pHfollowing the quenching reaction, the compounds were now in

heir neutral form which improved the migration from the aqueoushase to the organic phase.

.2. Optimization of LDTD/APCI parameters

In order to obtain optimal LDTD desorption and APCI ionizationf the target compounds, several LDTD parameters were optimizedo achieve signal enhancement while minimizing variability. Theaser power, laser pattern, deposition solvent, deposition volume

ig. 4. Effect of LDTD laser power on average peak area of the target compound, MMPB at

Fig. 5. Illustration of the LDTD laser pattern with its corresponding SR

imica Acta 820 (2014) 76–83

and carrier gas flow were all evaluated. All LDTD parameters wereoptimized in negative ionization (NI) mode by spiking 250 �g L−1

of MMPB in river water matrix that was subjected to the Lemieuxoxidation and liquid–liquid extraction and sample aliquots weredeposited in Lazwell plates. Each sample was analyzed 6 times(n = 6), in SRM mode using the optimized MS/MS parameters cor-responding to parent and fragment m/z with indicated optimizedtube lens and collision energies (Table 1).

The laser power along with the laser pattern are compounddependent parameters that will have an impact on how the com-pounds are desorbed when subjected to heat, thus affecting signalintensity. The laser efficiently heats the metal back of the well andsome of that energy is transferred to the compounds which arethermally desorbed and transferred by the gas flow to the APCIionization. The amount of energy transferred is dependent on thelaser power setting of the instrument (%), a fairly critical parametersince it must give enough energy to desorb the compound but notso much so as to cause its degradation or fragmentation. Usually,a stable working laser power is obtained with the range of 5–65%and this is estimated to correspond to generated well tempera-ture between approximately 50 and 220 ◦C [33]. Depending of thenature of the matrix, increasing laser power can cause a decrease insignal to noise response because of a higher amount of interferingcompounds being desorbed and transferred to the corona dischargealong with the compounds of interest [23]. By adjusting the laserramp pattern and the hold time at maximum laser power, it ispossible to increase the amount of desorbed compounds and con-trol the amount of simultaneously desorbed interfering analytes.

For MMPB, these parameters were evaluated by spiking 250 �g Lof MMPB in blank river water matrix submitted to oxidation andextraction in order to properly consider any matrix interference.The tested parameters were optimized using 6 replicates. Laser

250 �g L−1. Vertical error bars represent standard deviations from the mean (n = 6).

M signal and peak shape obtained using APCI-MS/MS analysis.

Page 6: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

A. Roy-Lachapelle et al. / Analytica Ch

Fig. 6. Average peak area of the target compound, MMPB at 250 �g L−1, as influ-enced by: (a) the deposition volume in the LDTD (1–8 �L), and (b) LDTD gas flow((

pplaicap0isib

tucwpssita

TV

1–8 L min−1). Vertical error bars represent standard deviations from the meann = 6).

ower was evaluated between 10 and 60% (Fig. 4). The optimal laserower for maximum compound desorption was set at 35%. Lower

aser power did not allow for maximum compound desorption andhigher laser power caused higher variability and decreased signal

ntensity which we presume can be attributed to heat-promotedompound degradation. Optimized laser pattern was 2 s at 0%, withramp of 2 s from 0 to 35%, with no hold time required at maximumower, and a direct decrease of 0.1 s from 35 to 0% with finally 2 s at%. Fig. 5 shows graphically the laser pattern with its correspond-

ng signal. Overall, with a 6 s LDTD desorption, a chromatogram-likeignal is observed covering a time span of less than 15 s. This signals subsequently quantified the same way a chromatogram woulde, illustrating the fast analysis time the LDTD can offer.

Using an adequate deposition solvent is an important parame-er affecting compound desorption [23]. Depending on the solventsed for deposition, the nanocrystals formed after drying in the wellan have a different effect upon crystallization, repartition into theell and adherence to the metal surface of the well [33]. The com-ound solubility in a specific solvent can also be a factor of differenturface well coverage after dryness. If a specific compound is less

oluble in a chosen solvent, after deposition in the hexagonal well,t will crystallize first as a donut around the well, therefore, whenhe laser hits the center of the well, it could be possible that notll analytes will be desorbed because of the heterogeneous crystal

able 2alidation parameters using internal calibration with three different concentrations (n = 6

QC concentrations(�g/L) Accuracy RE(%) Intraday RSD(%) Interday RSD(%

25 10 5 13100 8 6 11250 6 4 9

imica Acta 820 (2014) 76–83 81

formation. Also, depending of the solvent, the shape and super-position of nanocrystals formed after dryness could be different,thus creating several crystal layers with bound analytes hinderingtotal desorption. Many solvents were tested for deposition, and inorder to use the same solvent for the optimal MMPB extraction afteroxidation, the use of ethyl acetate was deemed most appropriate(Fig. 3).

The sample deposition volume in the wells will influence theamount of analytes thermally desorbed by LDTD. The depositionvolume will therefore influence the effectiveness of the APCI ioniza-tion and the MS signal intensity of the target compound. Adding alarge amount of sample means a higher amount of compounds thatwould reach ionization, increasing the signal, therefore, increas-ing sensitivity and enhancing method detection limit (MDL). Morecompounds in the ionization area can also induce a charge compe-tition during the APCI discharge and also a chance of trapping thecompounds into nonvolatile matrix products, all of which wouldaffect signal intensity [23]. The deposition volume was testedbetween 1 and 8 �L in steps of 1 �L with the spiked matrix. Fig. 6aillustrates the MMPB signal intensity as a function of the depositionvolume. The signal response progressively decreases by increasingthe volume above 2 �L. These results suggest a signal suppressionby matrix effect, and this is in concordance with previous work[22,23]. As a result, 2 �L was set as the optimal deposition volumewith significantly higher signal intensity (P < 0.05).

Carrier gas flow rates have an impact on the transfer of thethermally desorbed molecules to the corona discharge region forionization. It also allows thermalization of the desorbed and vapor-ized analytes, thus reducing thermal degradation [34]. Carrier gasflow was optimized between 1 and 8 L min−1 (Fig. 6b). This param-eter is also optimized for sensitivity enhancement. Carrier gas flowwas optimal at 2 L min−1 with a significantly higher signal response(P < 0.05), and a significant signal drop occurs above 3 L min−1.These results are in concordance with a previous study of BTEXanalysis in air where carrier gas flow had an influence on the APCIionization rate [35]. In fact, it was shown that compound signalintensity was dependent of the residence time in the corona dis-charge region where ionization occurs. It was also discussed thatoptimal flow was between 2 and 3 L min−1, as high as possible tomaximize the mass of analytes but not so fast as to prevent ioniza-tion from occurring, as discussed earlier [21–24,34]. This parameteris not expected to be compound-dependent but rather a physi-cal factor necessary to optimize the ionization in the APCI source.Since LDTD-APCI ionization occurs in dry conditions, the amountof available reactive water molecules becomes an important factorin terms of signal enhancement. In previous work, the possibilityof increasing water concentration in the corona discharge area wastested, but had no significant impact above a minimal level [36]. Itis therefore important to control the time of residence in the ion-ization area to optimize ionization rate, thus, allowing sensitivityenhancement.

3.3. Method validation

We used 4-phenylbutyric acid as IS because of the unavailabilityof isotopically-labeled MMPB, however, 4-PB has been success-fully used in previous work [37]. We used a seven-point internalcalibration curve based on a standard addition method of MCs

).

) Oxidation recovery yield (%) Signal recovery from matrix effect (%)

109 87102 90

97 97

Page 7: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

82 A. Roy-Lachapelle et al. / Analytica Chimica Acta 820 (2014) 76–83

Table 3Comparison of total microcystins analysis of lakes samples using LDTD-APCI-MS/MS and LC–MS/MS.

No. Sample Location Date Total MC via MMPB(�g L−1)a

(RSD – %)

Total MC with standards(�g L−1)b

MC isomer withoutstandards (�g/L)c

Total MC(�g L−1)d

Percentage of MCwith standards(%)e

1 Estrie 20130614 425 (9) 70 340 (RR)f 410 162 Estrie 20130614 1.0 (5) 0.15 0.5 (RRf 0.65 153 Saguenay 20130620 5.4 (7) 1.6 3.3 (YRf 4.9 304 Saguenay 20130620 4.7 (8) 1.8 2.5 (YRf 4.3 385 Abitibi-Temiscamingue 20130624 ND ND ND ND –6 Laurentides 20130731 37.4 (7) 35.4 ND 35.4 957 Abitibi-Temiscamingue 20130731 0.9 (8) ND ND ND –8 Abitibi-Temiscamingue 20130731 2.7 (7) 0.59 ND 0.59 229 Monteregie 20130801 ND 0.1 ND 0.1 –

ND – Not detected.a Total microcystins determined via MMPB using LDTD-APCI-MS/MS.b Microcystins determined via the summation of all microcystins for which standards allowed detection and quantification (i.e., HilR, HtyR, LA, LF, LR, LR (D-Asp3), LW, LY,

RR, RR (D-Asp3), WR, YR) using LC–MS/MS.c Microcystin isomer for which identification and concentration could not be certified due to the absence of appropriate standards using LC–MS/MS. Although their

characteristics are slightly different from the available standard, there is a very high probability that these are isomers of LC-RR and LC-YR.d ich st

t00.

sem5ss%a0clVtctwonenonemitTtasplMwpMMstotsvl

Total microcystins determined via the summation of all microcystins for whentatively identified using LC–MS/MS.

e Percentage is calculated as: (MC isomer with standards/total MC via MMPB) × 1f Suspected isomer of the specified congener.

piked in river water submitted to the optimized oxidation andxtraction so as to account for matrix effects and validate theethod. Each point of the linear dynamic range (between 1 and

00 �g L−1) was analyzed in triplicates. Correlation coefficientshowed a good linearity with R2 > 0.999. The precision of the mea-urements, expressed in terms of relative standard deviation (RSD –) was deemed satisfactory with 9%. Method detection limit (MDL)nd method quantification limit (MQL) were respectively 0.2 and.9 �g L−1. Fig. A-5 shows an example of the LDTD signal at MQLoncentration (1 �g L−1). The MDL is comparable to other simi-ar analytical techniques ranging from 0.02 to 15 �g L−1 [30,38].alidation parameters were determined with three concentra-

ion levels (25 �g L−1, 100 �g L−1 and 250 �g L−1) with calibrationurves and results are shown in Table 2. Accuracy, expressed inerms of relative error (RE – %), was below 10%. Intraday precisionas below 6% and interday precision was below 11%. The Lemieux

xidation reaction yield is 103% with a completed reaction and sig-al recovery of MMPB at 91%, thus showing no significant matrixffect in the analysis. The precision of the reaction yield and the sig-al recovery were 7% and 6% respectively. The method was testedn MCs present in a real cyanobacterial bloom where the super-atant was sampled and submitted to a cell lysis, filtered to removexcess cells debris and quantified in parallel by LC-MS/MS. Thisethod was applied to determine the total microcystins present

n nine natural cyanobacterial bloom samples during the prolifera-ion season in order to assess their levels in several water sources.able 3 shows the different results obtained from the quantifica-ion of total microcystins using LDTD-APCI-MS/MS. These resultsre compared with previous analysis of the same samples usingtandard LC-MS/MS analysis. We observe that for 5 out of 6 sam-les for which results from both methods were above detection

imits, the LC–MS/MS method which is relying on only the 12C standards available, significantly underestimated total MCshen compared to LDTD-APCI-MS/MS. For these five samples, theercentage of MC certified and quantified with standards using LC-S/MS by CEAEQ represents between 15 and 38% of the total MC viaMPB (Table 3). Although CEAEQ had tentatively identified what

eems to be isomers of MC-RR and MC-YR, they could not certifyhem in the absence of the specific standards. When concentrationsf MC and MC isomers are added to give total MC via LC-MS/MS,

he results of both methods are quite similar exception made ofample no. 8 for which total MC still only represents 22% of MCia MMPB since no congener isomers were detected by the CEAEQaboratory. Our results show that the presence of isomers of MC

andards allowed detection and quantification and also for the suspected isomer

or other congeners for which standards are not available could bemissed altogether, potentially leading to substantial underestima-tions of MCs in water samples. Overall, the LDTD-APCI-MS/MS is arobust method that allows high throughput analysis with less than15 s sample−1, and could be a powerful tool for a fast screeningof cyanobacterial toxins including some MC congeners that wouldotherwise go undetected.

4. Conclusion

A new method for the screening of total microcystins in naturalwater has been demonstrated and validated using LDTD-APCI-MS/MS. This method uses an oxidation step but this is a simplesample pre-treatment that has been kept relatively easy to imple-ment. The MS analysis is simple, fast and allows high-throughputanalysis (<15 s sample−1). Oxidation and instrumental parameterswere optimized in order to enhance sample treatment efficiencyand signal response. An internal calibration with standard additionof MMPB and using 4-PB as IS gave an excellent linearity with a cor-relation coefficient (R2) above 0.999. MDL and MQL were measuredto be 0.2 and 0.9 �g L−1 respectively. Accuracy and precision werebelow 10%. Interday and intraday precision were evaluated below15%. The limit of quantification is below the WHO microcystin-LRguideline for drinking water making the proposed method a use-ful approach for a rapid total microcystins screening in drinkingwater, surface waters and water treatment facilities. It could givea fast quantification response prior to applying any LC–MS anal-ysis for characterization of the MCs congeners, eliminating costlyand lengthy analysis while providing more selectivity and accuracythan ELISA detection which does not utilize MS/MS detection. Themethod is based on total MCs and could therefore also identify con-taminated samples related to MC congeners which are not usuallyincluded in the LC–MS analytical protocol, thus preventing somefalse negative. The ultrafast analysis combined with a robust quan-tification and relatively simple sample pre-treatment makes thismethod a usable alternative for environmental safety and healthconcerns caused by the presence of microcystins in cyanobacterialblooms occurring in surface waters.

Acknowledgments

The Fond de Recherche Québec Nature et Technologies andthe Natural Sciences and Engineering Research Council of Canada(NSERC) are acknowledged for financial support. We thank Thermo

Page 8: Total microcystins analysis in ... - webdepot.umontreal.ca€¦ · Audrey Roy-Lachapelle a, Paul B. Fayad , Marc Sinotteb, Christian Debloisc, Sébastien Sauvéa,∗ a Department

tica Ch

FaG

A

t

R

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[

[hepatotoxic cyclic peptide microcystins in animal tissues, Toxicon 47 (7) (2006)734–741.

[38] R. Kaushik, R. Balasubramanian, Methods and approaches used for detection of

A. Roy-Lachapelle et al. / Analy

isher Scientific and Phytronix Technologies for their support. Welso thank Morgan Solliec, Khadija Aboulfadl, Lucilla Coral andabriela Vázquez for their technical help and scientific support.

ppendix A. Supplementary data

Supplementary data associated with this article can be found, inhe online version, at http://dx.doi.org/10.1016/j.aca.2013.12.001.

eferences

[1] I.R. Falconer, A.R. Humpage, Health risk assessment of cyanobacterial (blue-green algal) toxins in drinking water, International Journal of EnvironmentalResearch and Public Health 2 (1) (2005) 43–50.

[2] G.A. Codd, L.F. Morrison, J.S. Metcalf, Cyanobacterial toxins: risk managementfor health protection, Toxicology and Applied Pharmacology 203 (3) (2005)264–272.

[3] I.R. Falconer, An overview of problems caused by toxic blue-green algae(cyanobacteria) in drinking and recreational water, Environmental Toxicology14 (1) (1999) 5–12.

[4] I. Chorus, J. Bartram, Toxic Cyanobacteria in Water: A Guide to their PublicHealth Consequences Monitoring and Management, E & FN Spon, London, NewYork, 1999, p. 416.

[5] K.L. Rinehart, et al., Nodularin, microcystin, and the configuration of Adda,Journal of the American Chemical Society 110 (25) (1988) 8557–8558.

[6] R.W. Mackintosh, et al., The cyanobacterial toxin microcystin binds covalentlyto cysteine-273 on protein phosphatase-1, FEBS Letters 371 (3) (1995) 236–240.

[7] M.E. van Apeldoorn, et al., Toxins of cyanobacteria, Molecular Nutrition & FoodResearch 51 (1) (2007) 7–60.

[8] A. Zamyadi, et al., Fate of toxic cyanobacterial cells and disinfection by-productsformation after chlorination, Water Research 46 (5) (2012) 1524–1535.

[9] World Health Organization, Guidelines for Drinking-Water Quality, fourth ed.,World Health Organization, Geneva, 2011, p. 541.

10] A. Zamyadi, et al., Monitoring of potentially toxic cyanobacteria using an onlinemulti-probe in drinking water sources, Journal of Environmental Monitoring14 (2) (2012) 579–588.

11] J. Meriluoto, Chromatography of microcystins, Analytica Chimica Acta 352(1–3) (1997) 277–298.

12] D.O. Mountfort, P. Holland, J. Sprosen, Method for detecting classes of micro-cystins by combination of protein phosphatase inhibition assay and ELISA:comparison with LC–MS, Toxicon 45 (2) (2005) 199–206.

13] J. Wang, et al., An ultra-performance liquid chromatography-tandem massspectrometry method for determination of microcystins occurrence in surfacewater in Zhejiang Province, China, Toxicon 49 (8) (2007) 1120–1128.

14] K. Harada, et al., Mass spectrometric screening method for microcystins incyanobacteria, Toxicon 34 (6) (1996) 701–710.

15] K. Kaya, T. Sano, Total microcystin determination using erythro-2-methyl-3-(methoxy-d(3))-4-phenylbutyric acid (MMPB-d(3)) as the internal standard,Analytica Chimica Acta 386 (1, 2) (1999) 107–112.

16] K. Tsuji, et al., Analysis of microcystins in sediments using MMPB method,Toxicon 39 (5) (2001) 687–692.

17] M. Welker, et al., Applications of MALDI-TOF MS analysis in cyanotoxinresearch, Environmental Toxicology 17 (4) (2002) 367–374.

18] J.S. Metcalf, G.A. Codd, Analysis of cyanobacterial toxins by immunologicalmethods, Chemical Research in Toxicology 16 (2) (2003) 103–112.

19] F.M. Young, et al., Production of antibodies against microcystin-RR for the

assessment of purified microcystins and cyanobacterial environmental sam-ples, Toxicon 48 (3) (2006) 295–306.

20] A. Zeck, et al., Highly sensitive immunoassay based on a monoclonal antibodyspecific for [4-arginine]microcystins, Analytica Chimica Acta 441 (1) (2001)1–13.

imica Acta 820 (2014) 76–83 83

21] P. Lemoine, et al., Ultra-fast analysis of anatoxin-A using laser diode thermaldesorption-atmospheric pressure chemical ionization-tandem mass spec-trometry: validation and resolution from phenylalanine, Toxicon 61 (2013)165–174.

22] M. Boisvert, P.B. Fayad, S. Sauvé, Development of a new multi-residue laserdiode thermal desorption atmospheric pressure chemical ionization tandemmass spectrometry method for the detection and quantification of pesti-cides and pharmaceuticals in wastewater samples, Analytica Chimica Acta 754(2012) 75–82.

23] P.B. Fayad, M. Prévost, S. Sauvé, Laser diode thermal desorption/atmosphericpressure chemical ionization tandem mass spectrometry analysis of selectedsteroid hormones in wastewater: method optimization and application, Ana-lycal Chemistry 82 (2) (2010) 639–645.

24] L. Viglino, M. Prévost, S. Sauvé, High throughput analysis of solid-boundendocrine disruptors by LDTD-APCI-MS/MS, Journal of Environmental Moni-toring 13 (3) (2011) 583–590.

25] O. Heudi, et al., Laser diode thermal desorption-positive mode atmosphericpressure chemical ionization tandem mass spectrometry for the ultra-fastquantification of a pharmaceutical compound in human plasma, Journal ofPharmaceutical and Biomedical Analysis 54 (5) (2011) 1088–1095.

26] J.G. Swales, et al., Determination of paracetamol in mouse, rat and dog plasmasamples by laser diode thermal desorption-APCI-MS/MS, Bioanalysis 4 (11)(2012) 1327–1335.

27] J. Wu, et al., High-throughput cytochrome P450 inhibition assays using laserdiode thermal desorption-atmospheric pressure chemical ionization-tandemmass spectrometry, Analytical Chemistry 79 (12) (2007) 4657–4665.

28] Centre d‘Expertise en Analyse Environnementale du Québec, Protocole pour lavalidation d‘une méthode d‘analyse em chimie, DR -12-VMC.

29] X.Q. Wu, et al., Kinetic study of the 2-methyl-3-methoxy-4-phenylbutanoicacid produced by oxidation of microcystin in aqueous solutions, EnvironmentalToxicology and Chemistry 27 (10) (2008) 2019–2026.

30] M.R. Neffling, E. Lance, J. Meriluoto, Detection of free and covalently boundmicrocystins in animal tissues by liquid chromatography-tandem mass spec-trometry, Environmental Pollution 158 (3) (2010) 948–952.

31] X.Q. Wu, et al., Optimal strategies for determination of free/extractable andtotal microcystins in lake sediment, Analytica Chimica Acta 709 (2012) 66–72.

32] V. Gomis, et al., Equilibrium for the ternary-system water plus sodium-chlorideplus ethyl-acetate at 30 ◦C, Journal of Chemical and Engineering Data 38 (4)(1993) 589–590.

33] P. Picard, P. Tremblay, E.R. Paquin, Laser Diode Thermal Desorption Ioniza-tion Source (LDTD): fundamental aspects, in: 56th American Society for MassSpectrometry Conference. 2008, Denver, CO, USA, 2014.

34] P.A. Segura, et al., High-throughput quantitation of seven sulfonamide residuesin dairy milk using laser diode thermal desorption-negative mode atmosphericpressure chemical ionization tandem mass spectrometry, Journal of Agricul-tural and Food Chemistry 58 (3) (2010) 1442–1446.

35] K. Badjagbo, et al., Direct atmospheric pressure chemical ionization-tandemmass spectrometry for the continuous real-time trace analysis of benzene,toluene, ethylbenzene, and xylenes in ambient air, Journal of the AmericanSociety for Mass Spectrometry 20 (5) (2009) 829–836.

36] I. Beattie, et al., Evaluation of laser diode thermal desorption (LDTD) coupledwith tandem mass spectrometry (MS/MS) for support of in vitro drug discoveryassays: increasing scope, robustness and throughput of the LDTD technique foruse with chemically diverse compound libraries, Journal of Pharmaceutical andBiomedical Analysis 59 (2012) 18–28.

37] J.L. Ott, W.W. Carmichael, LC/ESI/MS method development for the analysis of

cyanotoxins in environmental samples: a review, Critical Reviews in Environ-mental Science and Technology 43 (13) (2013) 1349–1383.