THIS IS GOING TO BE TORQUED: TORQUE NEWTON CENTIMETER (NCM) CREATECH Bridge, implant level, straight channel 35 NCM Bridge, abutment level, straight channel 15 NCM Bridge, implant or abutment level, angulated channel 15 NCM ABUTMENT GROUP TORQUE CROWN/BRIDGE ON THE ABUTMENT IS TORQUED TO: EXCEPTION SynOcta e.g., 048;601,605,612-618 35 NCM 15 NCM Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both the abutment and the zirconia crown shall be torqued to 35 NCM. Variobase family 35 NCM n/a Does not include the Variobase coping for SRA. It shall be torqued to 15 NCM. It utilizes an occlusal screw and not a basal screw. SRA 35 NCM 15 NCM Cares all materials 35 NCM n/a Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both the abutment and the zirconia crown shall be torqued to 35 NCM. Anatomical ZrO2 and Ti 35 NCM n/a Torque control device

Torque control device - Straumannnordic.straumann.com/content/dam/internet/straumann_nordic... · Using the torque control device Torque control device. Being used together with the

  • Upload
    dinhthu

  • View
    237

  • Download
    6

Embed Size (px)

Citation preview

Page 1: Torque control device - Straumannnordic.straumann.com/content/dam/internet/straumann_nordic... · Using the torque control device Torque control device. Being used together with the

THIS IS GOING TO BE TORQUED:

TORQUE NEWTON CENTIMETER (NCM)

CREATECH

Bridge, implant level, straight channel

35 NCM

Bridge, abutment level, straight channel

15 NCM

Bridge, implant or abutment level, angulated channel 15 NCM

ABUTMENT GROUP TORQUECROWN/BRIDGE ON THE ABUTMENT IS TORQUED TO:

EXCEPTION

SynOcta e.g., 048;601,605,612-618 35 NCM 15 NCM Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both

the abutment and the zirconia crown shall be torqued to 35 NCM.

Variobase family

35 NCM n/a Does not include the Variobase coping for SRA. It shall be torqued to 15 NCM. It utilizes an occlusal screw and not a basal screw.

SRA

35 NCM 15 NCM

Caresall materials 35 NCM n/a Cares ZrO2 RN comes with a synocta 1,5 (048.601). In this case both

the abutment and the zirconia crown shall be torqued to 35 NCM.

AnatomicalZrO2 and Ti 35 NCM n/a

Torque control device

Page 2: Torque control device - Straumannnordic.straumann.com/content/dam/internet/straumann_nordic... · Using the torque control device Torque control device. Being used together with the

Using the torque control device

Torque control device.Being used together with the SCS screwdriver in order to torque abutments and restorations from Straumann®.

Storing the torque control deviceThe torque control must be disassembled for cleaning and sterilization. It shall be stored disassembled and only assembled before use.

Using the torque control devicePlace a finger on the SCS screwdriver for stability. Push/pull the drop shaped handle only, until it has reached the correct torque marking. See page 1 for details. Observe! The arrow shall always point clockwise.

Abutment The abutment is inserted into the implant and tightened by hand first. With the use of the torque control device you torque the abutment to the correct torque with one uninterrupted movement. The screw channel can be sealed with e.g., guttapercha/silicontube and composite.

Restoration.The restoration is inserted on the abutment/into the implant and tightened by hand first. With the use of the torque control device you torque the it to the correct torque with one uninterrupted movement.The screw channel can be sealed with e.g., guttapercha/silicontube and composite.

The torque in this guide is only valid for Straumann® Original or Createch prosthetics on Straumann® implants. Regarding prosthetic components or restorations which is not manufactured by Straumann®. Please contact the supplier/manufacturer for the correct torque it is the abutment/screw that decides the torque not the implant.