Upload
evilasio-souza
View
218
Download
0
Embed Size (px)
Citation preview
8/10/2019 TM Opaque Abril03 2014. Nilson
1/15
1
Pulsed Photothermal Mirror Technique: Characterization of
Opaque Materials
O. A. Capeloto1, G. V. B. Lukasievicz
1,2,3, V. S. Zanuto
1,2,3, L. S. Herculano,N. E. Souza
Filho4,A. Novatski
5, L. C. Malacarne
1, S. E. Bialkowski
2, and N. G. C. Astrath
1,2,3,*
1 Departamento de Fsica, Universidade Estadual de Maring, Maring, PR 87020-900, Brazil
2 Department of Chemistry and Biochemistry, Utah State University, Logan, UT 84322-0300, USA
3
CAPES Foundation, Ministry of Education of Brazil, Braslia, DF 70040-020, Brazil4Departamento de Engenharia Acstica, Universidade Federal de Santa Maria, Santa Maria, RS 97105-900, Brazil
5Departamento de Fsica, Universidade Estadual de Ponta Grossa, Ponta Grossa, PR 84030-900, Brazil
ABSTRACT:
In this work, the time-resolved Thermal Mirror technique is developed under pulsed laser
excitation for measuring thermal and mechanical properties of opaque materials. The heat
diffusion and thermoelastic equations are solved analytically considering pulsed excitation
assuming surface absorption and instantaneous pulse. The analytical results for the temperature
change in the sample is compared to all numerical solutions using finite element method analysis
taking the pulse width laser into account. Experiments are realized in order to validate theoretical
model and regression fitting is performed to obtain the thermal diffusivity and the linear thermal
expansion coefficient of the samples. The values obtained for these properties are in good
agreement with literature data. The experimental technique combined with the theoretical model
is shown to be useful for quantitative determinations of the physics properties of metals with highthermal diffusivity.
PACS numbers: 61.82.Bg; 65.60.+a; 78.20.nb
*E-mail address: [email protected]
8/10/2019 TM Opaque Abril03 2014. Nilson
2/15
2
INTRODUCTION
Photothermal Mirror spectrometry has been developed both experimentally and
theoretically over the past few years [1-10]. The thermoelastic effect has been known and used
for material characterization for decades, and most recently applied as a time-resolved
quantitative method for investigating transparent and opaque materials under continuous [1-8]
and pulsed laser excitation [9,10]. Advances in theoretical treatment of this complex dynamic
thermoelastic problem made possible to predict for effects of heat coupling between sample and
the surrounding fluid [7-9], in addition to account for different time-distribution pulses [10]. Theaccuracy of the theoretical descriptions have been, in most cases, tested against all numerical
modeling by comparing the analytical solutions under various assumptions with finite element
method analysis.
The Photothermal Mirror effect arises from photo-induced heat generation in a sample,
followed by the expansion or contraction of the surface. This deformation acts, in this case, as an
optical element that can be probed by analyzing the wavefront distortion of a second weak laser
reflected off of the surface. The surface deformation then focus or defocus the probe beam,
depending on the expansion coefficient of the sample, and the central portion of the probe beam
intensity is monitored in the far field. This configuration is mostly referred to as Thermal Mirror
(TM).
The TM transient depends direct on thermo-physical properties of the sample. For
instance, the maximum amplitude of the surface expansion or contraction is related to the linear
expansion coefficient and absorption coefficient, and the relaxation time - the characteristic
transient time - brings direct information on the thermal diffusivity of the sample. These
characteristics made this method a powerful tool for quantitative analysis of solid materials. Due
to the high values of the thermal diffusivity of most metals, the TM transient formation time is
much smaller than that observed in glasses for instance. The use of continuous laser excitation in
these cases could induce large errors due to the shutter opening time.
Here, we extend the theoretical and experimental capabilities of the TM technique to
studying opaque materials under pulsed laser excitation. First, it is presented a theoretical model
to describe the temperature, the surface displacement and the TM signal. Then, the model is used
to study a series of opaque metals, providing quantitative information on the thermal diffusivity
8/10/2019 TM Opaque Abril03 2014. Nilson
3/15
3
and thermophysical properties of the samples.
THEORY
In the TM experimental apparatus explored in this work, the sample is excited by a pulsed
TEM00 Gaussian laser beam of radius 0ew . The TM signal arises from optical absorption in a
sample. Sample heating occurs due to non-radiative decay processes. The Gaussian profile
produces a radial temperature gradient perpendicular to the propagation of the laser beam, which
in turn results in a surface deformation of the sample. During the pulse, the deformation tends to
follow the temperature rise distribution and then stress and strain govern the deformation. The
curvature of the deformation acts as an optical element to a second beam reflected off of the
surface. A cw TEM00Gaussian beam (probe beam) is assumed to have a radius 0Pw at its waist,
located at a distance 1Z from the sample, and a radius 1Pw at the sample surface. The TM affects
the propagation of the probe beam, resulting in a change in its intensity profile measured in the
far field. The central portion of the reflected probe laser intensity variation gives information on
physical properties of the sample.
The propagation of the excitation and probe beams is along the direction z , and the sample
surface is located at 0z= . A semi-infinite sample is considered here. The radial dimensions of
the sample are assumed to be large compared with the excitation beam radius to avoid edge
effects. The absorption of light for the opaque sample is assumed to be completely superficial and
can be safely represented by a Dirac delta function as ( )zd . The schematic of the mode-
mismatched TM technique is showed in Fig. 1.
FIG. 1. Schematic of the geometric positions of the beams in a TM technique.
8/10/2019 TM Opaque Abril03 2014. Nilson
4/15
4
The temperature change at the surface of the sample and within it is described by the
solution of the heat conduction equation [11],
( )( ) ( ) ( ) ( )2 2 20 0
, ,., , = exp 2 e
T r z t T Q z Q t D r z t Q r w
t
D D
- -
(1)
PD k Cr= is the thermal diffusivity of the sample, k , PC and r are the thermal conductivity,
specific heat, and mass density, respectively. ( ) ( )Q t td= , and the Dirac delta function in time
( )td represents the pulsed excitation. The initial condition for this problem is ( ), , 0 0T r zD = , and
the boundary conditions are ( ), , 0T z tD = and ( ) 0, , 0zT r z t z =
D = . The later assumes that
there is no heat conduction from the sample to the surrounding fluid; this is a good approximation
when air is the surrounding fluid [8, 9]. ( ) ( )2 eQ z z Ad= and eA is the optical absorption
coefficient at the excitation beam wavelength. ( ) 20 0 02 1e p eQ E A R C w f r p= - ; 0E is the pulse laser
energy, R is the surface reflectivity, and f is a heat yield parameter accounting for energy loss
due to luminescence. The mathematical procedure for solving Eq. (1) has been explored in other
reports [2,3,10], and consists of applying Laplace, Hankel and Fourier cosine transforms of the t-
time, r - and z -dependent temperature change, respectively, yielding
( )
2 2 2 20 0
0
2exp exp
1 22, , .
1 2
e e
c c
ce e c
r w z w
t t t t QT r z t
t tA w t t p
- - + D =
+ (2)
The characteristic thermal time constant is 20 4c et w D= .
The non-uniform temperature change in the sample causes a non-uniform surface
contraction or expansion that evolves following the Navier-Stokes thermoelastic equation, in the
quasistatic approximation, as [12-14]
( ) ( ) ( ) ( ) ( )u u21 2 , , . , , 2 1 , , .Tr z t r z t T r z t n n a - + = + D (3)
u is the displacement vector, n is the Poisson's ratio, and Ta is the linear thermal expansion
coefficient. The boundary conditions at the free surface are on the normal stress components,
0 0rz zs = = and 0 0zz zs = = . The problem presents axial symmetry due to the radial nature of the
8/10/2019 TM Opaque Abril03 2014. Nilson
5/15
5
Gaussian pulsed heat source, and is treated in cylindrical coordinates. The solution has been
proposed by introducing scalar potential, as described in details in Refs. [5,12], and can bewritten in terms of the Hankel transform of the temperature change, yielding at the surface
displacement
( ) ( ) ( ) ( ) ( )2
0 0 2 20 0 00
, 0, 2 1 exp 8 Erfc 2 J .4
ez T e e c
e
Q wu r t w w t t r d
An a a a a a a
= - + - (4)
( )nJ x is the Bessel's function of the first kind and ( )Erfc x is the complementary error function.
Eq. (4) represents the surface deformation induced on an opaque sample by a Gaussian profile
pulsed laser beam.
Assuming an absorption completely superficial using ( )zd in addition to the Dirac delta
time representation, ( )td , simplify considerably the numerical calculations of Eq.(4), although
leading to divergent results for the temperature change at 0t= and discrepant results in
nanosecond time scales because of finite absorption coefficient. These points were investigated
by solving Eq. (1) numerically using finite element method (FEM) analysis, employing Comsol
Multiphysics 4.2a software. This software provides numerical solutions to the heat transfer and
thermoelastic equations with the realistic boundary conditions imposed by the experiment. We
considered the absorbance to follow the Beer-Lambert law as ( ) ( )exp eQ z A z = - in the source
term of the heat diffusion equation and, in addition, we represent the time-dependence of the
pulse as [10] ( ) ( ) ( )2 22 exp 1 Erf Q t t x t t p x t
= - - + , where t is pulse width, x is the
time to the maximum irradiance for the Gaussian pulse, ( )Erf x is the error function. The
temperature change is obtained and compared to the solution given by Eq. (2).
UNS C10100 Copper Alloy was the test sample, and the parameters used in the
simulations are listed in the caption of Fig. 2. We assumed the mechanical and thermal properties
of the sample are independent of temperature. This assumption limits the applicability of the
solution obtained to a small range of temperature change. Fig. 2 (a) shows the calculated
temperature change on the surface at the center of the excitation beam ( 0z= and 0r= ),
obtained with Eq. (2) and simulations using numerical FEM considering different pulse widths.
Fig. 2 (b) shows the dependence of temperature with different optical absorption coefficients. The
figures reveal excellent agreement between both solutions for500
t ns
>, when used pulses
8/10/2019 TM Opaque Abril03 2014. Nilson
6/15
6
widths shorter than 50nsand samples with optical absorption bigger than 5 15 10eA m-> .
0
2
4
6
8
10
0 100 200 300 400 5000
1
2
3
4
5
Pulse width: = 15ns
(t)
= 5ns
= 15ns
= 30ns
= 50ns
Temperature(C)
(a) Different pulse width
Superficial Absorption
(b) Different absorption coefficients
Analytical model: (t) and (z)
Superficial Absorption
Ae= 1. 107m
-1
Ae= 2. 106m
-1
Ae= 1. 106m
-1
Ae= 5. 105m
-1
Ae= 2. 105m
-1Temperature(C)
Time (ns)
FIG. 2. Physical properties of copper used for the simulations: 4 2 11.14 10D m s- -= , 1 1391k Wm K - -= ,
1 1385c J kg K - -= , 38940kg mr -= , 6 117 10T Ka - -= , 0.31n= , 0 100ew mm= and
50 1 10eQ A K m
-= .
The surface displacement of the sample act as a concave/convex mirror for the probe
beam, causing a phase shift given by ( )2 2 ( , 0, ) (0, 0, )p z zu r t u t p l= -F [1-3]. By Eq. (4), we get
the TM phase shift as
( ) ( ) ( ) ( )20 2 2
0 00 00exp 8 Erfc 2 J, 1 ,eTM e e
c
ecg t w
w w t t m w dt
gq a a a a a - - -F = (5)
8/10/2019 TM Opaque Abril03 2014. Nilson
7/15
7
where 0(1 ) (1 )TTM PE R kq a n f l= - + , ( )2
1 0P em w w= , 1 2( / )Pwg r= , and pl is the probe
beam wavelength. The complex electric field of TEM00 Gaussian probe beam reflected by the
deformed surface of the sample is given by [1, 15]
2 2
1 21 1
( , ) exp ,pp p P
r rU r Z B i
R w
p
l
= - + - F (6)
where 11
12 / exp 2 /P PPB w P i Z p p l- = - . 1pR and pP are the radius of curvature of the probe
beam and the reflected probe beam power at 1Z , respectively. Considering only the center of the
beam reflected at the detector's plane we can use diffraction Fresnel theory to calculate the
complex amplitude of the electrical field of the probe beam. In cylindrical coordinates it is [15]
1 2 0( , ) exp (1 ) ( , ) ,U Z Z t C iV g i g t dg
+ = - + - F (7)
in which, ( ) ( ) ( )1 2 12
1c c cV Z Z Z Z Z Z
= + + , 22 1 2/exp( 2 )P P PC B i i Z Zp l pw l= - , cZ is the probe's
beam confocal distance, and 2Z is the distance from the sample to the detector plane. The
normalized intensity of the electric field at the detector's plane is given [15]
2
0
2
0
exp (1 ) ( , )
( )
exp (1 )
iV g i g t dg
I t
iV g dg
- + - =
- +
F
(8)
The fit of the experimental data with Eq. (8) gives the thermal diffusivity and the
parameter TMq , which is related with the optical, thermal and mechanical properties of the
samples.
EXPERIMENTAL RESULTS AND DISCUSSION
The experimental set-up for the Thermal Mirror system for opaque samples is shown in
Fig. 3. A Q-switched pulsed Nd:YAG laser (Brilliant B, Quantel), operating at the second
harmonic wave with a wavelength of 532nm and pulse width of 15ns, is used as pumping source.
8/10/2019 TM Opaque Abril03 2014. Nilson
8/15
8
Part of the laser beam is reflected by a beam sampler (ThorLabs, Model BSF05-A) and then
focused in the sample by a biconvex lens of focal length 0.75f m= . The transmitted beamthrough the beam sampler is used as reference for energy measurement by a Pyroelectric Energy
Sensor (Thorlabs, Model ES120C) plugged in a digital energy meter console (Thorlabs, Model
PM100D). A continuous TEM00Gaussian He-Ne laser at 632.8nm (Newport, Model R-32734)
was arranged almost collinear to the excitation beam and used to probe the TM effect. After
being reflected off of the sample surface, the probe beam propagates to the photomultiplier
detector P1(Hamamatsu, Model R928), which was assembled with a pinhole and a laser line filter
(ThorLabs, Model FL632.8-10). The PMT is biased with a high voltage power supply (Newport,
Model 70706). Only the central part of the probe beam is detected by the photomultiplier tube
and recorded by a digital oscilloscope (Tektronix, Model DPO 4102B). The oscilloscope is
triggered by P2 (Newport, Model 818BB-22). The apparatus is mounted on an optical table
(TMC, size 2.0x1.0m).
FIG. 3. Schematic diagram of the TM experimental setup used: Mi stands for mirrors, Li stands for lenses,
PMT stands for the photomultiplier tube and P2stands for the photodiode.
The geometric parameters of the excitation and probe beams are: 0 123ew mm= ,
1 1467pw mm= , 1.47cZ cm= , 1 41.1Z cm= , 2 387.1Z cm= , 30.93V = , and 142.2m= . The probe
radius along the z-axis is measured with a Beam Profiler (Thorlabs, BP104-UV) to determine the
8/10/2019 TM Opaque Abril03 2014. Nilson
9/15
9
confocal distance. The radius of the excitation and probe beam is measured with a beam profile
camera (Coherent, LaserCam-HR) at the sample position.
TM measurements are performed in four opaque samples. The samples are UNS
C10100/C10200 Copper Alloy (Copper) [16], UNS A96351 Aluminum Alloy [16, 17], UNS
C36000 Copper Alloy (Brass) [16], and UNS C90300 Cooper Alloy (Bronze SAE 620) [16]. The
cylindrical samples are 15.0mm thicker with diameter of 25.4mm . All the samples are polished
with polycrystalline diamond compound in order to ensure minimum roughness on the surface.
The values of surface reflectivity are highly dependent of the surface polishing [18].
FIG. 4. Normalized TM signal ( ) / (0)I t I for copper sample. Circles: experimental data; Solid
lines: best fit curves.
Figures 4 and 5 show the normalized TM signal ( ) ( )0I t I for Copper and Aluminum
Alloy samples, respectively. More than 100 transients were averaged to produce each transient.
The repetition rates of excitation beam are 5Hz for copper and 2Hz for aluminum. The
normalization is chosen by the minimum value of the transient in order to discriminate all curves
in function of the energy in the graph. The detector response time is less than 2 sm . The surface
deformation produced by the excitation beam is outward, acting as convex mirror due the positive
linear thermal expansion coefficient of these samples, which decreases instantaneously the signal
at the detector and eventually returning to the initial value as the heat diffuses along the sample.
8/10/2019 TM Opaque Abril03 2014. Nilson
10/15
10
FIG. 5. Normalized TM signal ( ) / (0)I t I for aluminum sample. Circles: experimental data; Solid lines:
best fit curves.
The same characteristics in the TM signals are observed for brass and bronze samples as
shown in Figs. 6 and 7, respectively. The excitation beam repetition is 2Hz for both samples.
Note that the time to reach the steady-state differs for each sample, which is directly related to the
thermal diffusivity of the samples.
FIG. 6. Normalized TM signal ( ) / (0)I t I for brass sample. Circles: experimental data; Solid lines: best fit
curves.
The theoretical model presented in this paper is in good agreement with the experimental
8/10/2019 TM Opaque Abril03 2014. Nilson
11/15
11
data. Regression using the function "NonLinearModelFit" is used to fit experimental data, which
provides the thermal diffusivities and the parameter TMq . To speed up the regression, the
experimental data is smoothed and about 30 points is taken from the decay part of the dataset.
FIG. 7. Normalized TM signal ( ) / (0)I t I
for bronze sample. Circles: experimental data; Solidlines: best fit curves.
TMq yields information about optical, thermal and mechanical properties of the sample.
Figure 8 shows TMq versus 0E , the slope value of the linear fit gives (1(1 ) )T PkR a n l+- . Here
1f= , i.e., all the energy absorbed by sample is converted into heat. With the additional values of
R, and k, the linear thermal expansion coefficient can be obtained from 0TM Eq . Poisson's ratio
and thermal conductivity are available in Refs. [16, 17]. The surface reflectivity at 532nm were
measured in a spectrophotometer (Perkin Elmer, Model Lambda 1050) assembled with
integrating sphere (Labsphere, Model# 150MM RSA ASSY). Table 1 shows reference properties
of the samples and the average parameters obtained from regression. The obtained properties are
in excellent agreement with those find in the literature.
8/10/2019 TM Opaque Abril03 2014. Nilson
12/15
12
0 10 20 30 40 50 60 70
0
2
4
6
8
10
12
Bronze
Brass
Aluminum
Copper
Linear Fit
TM(
s)
E0(J)
FIG. 8. The fittedTMq as a function of excitation energy. Open symbols are experimental data
obtained from TM signals fitted and solid lines are best linear curve fit.
Table 1 - Comparative table with thermal, optical, and mechanical properties obtained from TM
experiment and literature data [16, 17].
Samples D
(Measured)
5 2( 10 )m s-
D
(Literature)
5 2( 10 )m s-
TM/E0
(Measured) 1W-
R@532nm
(Measured)
T
(Measured)
6 1( 10 )K- -
T
(Literature)
6 1( 10 )K- -
Bronze 2.4(0.1) 2.2 0.1415(0.001) 0.719 17.6 18.0
Brass 3.4(0.2) 3.6 0.077(0.002) 0.810 22.5 20.5
Aluminum 7.8(0.3) 7.3 0.0472(0.003) 0.857 27.6 23.4
Copper 12.6(0.8) 11.4 0.029(0.002) 0.619 14.4 17.0
8/10/2019 TM Opaque Abril03 2014. Nilson
13/15
13
CONCLUSIONS
In conclusion, we have presented a simple mathematical model to describe the induced
effect when a Gaussian pulsed laser shines on a high absorbing material. The temperature change
in the sample was calculated and compared with FEM simulations for different pulse widths and
optical absorption coefficients. Experiments in four opaque metals were performed to validate the
model and determine the thermal diffusivity of these samples. The results obtained by regression
have yielded good values for the fitted parameters when compared with literature data. The TM
technique using pulsed excitation has been shown be very useful to measure thermal and
mechanical properties of opaque materials with high thermal diffusivity, since, the continuous
excitation presents difficulties to measure such samples.
8/10/2019 TM Opaque Abril03 2014. Nilson
14/15
14
REFERENCES
1. N. G. C. Astrath, L. C. Malacarne, P. R. B. Pedreira, A. C. Bento, M. L. Baesso, J. Shen,
Appl. Phys. Lett. 91, 191908 (2007).
2. L. C. Malacarne, F. Sato, P. R. B. Pedreira, A. C. Bento, R. S. Mendes, M. L. Baesso, N. G.
C. Astrath, J. Shen, Appl. Phys. Lett. 92, 131903 (2008).
3. F. Sato, L. C. Malacarne, P. R. B. Pedreira, M. P. Belancon, R. S. Mendes, M. L. Baesso,
N. G. C. Astrath, and J. Shen, J. Appl. Phys. 104, 053520 (2008).
4. V. S. Zanuto, L. S. Herculano, M. L. Baesso, G. V. B. Lukasievicz, C. Jacinto, L. C.
Malacarne, N. G. C. Astrath. "Thermal mirror spectrometry: An experimental investigation
of optical glasses". Optical Materials. 2013. 35(5):1129-1133.
5. N. G. C. Astrath, L. C. Malacarne, V. S. Zanuto, M. P. Belancon, R. S. Mendes, M. L.
Baesso, C. Jacinto. Finite-size effect on the surface deformation thermal mirror method.
J. Opt. Soc. Am. B. 2011. 28(7): 1735-1739.
6. N. G. C. Astrath, F. B. G. Astrath, J. Shen, J. Zhou, C. E. Gu, L. C. Malacarne, P. R. B.
Pedreira, A. C. Bento, M. L. Baesso. "Top-hat cw laser induced thermal mirror: a complete
model for material characterization", Appl. Phys. B: Lasers Opt. 2009. 94(3):473-481.
7. L. C. Malacarne, N. G. C. Astrath, G. V. B. Lukasievicz, E. K. Lenzi, M. L. Baesso, S. E.
Bialkowski. Time-Resolved Thermal Lens and Thermal Mirror Spectroscopy with
Sample-Fluid Heat Coupling: A Complete Model for Material Characterization. Appl.
Spectrosc. 2011. 65(1): 99-104.
8. G. V. B. Lukasievicz, L. C. Malacarne, N. G. C. Astrath, V. S. Zanuto, L. S. Herculano, S.
E. Bialkowski. "A Theoretical and Experimental Study of Time-Resolved Thermal Mirror
with Non-Absorbing Heat Coupling Fluids". Appl. Spectrosc. 2012. 66(12): 1461-1467.9. N. G. C. Astrath, L. C. Malacarne, G. V. B. Lukasievicz, M. P. Belancon, M. L. Baesso, P.
R. Joshi, S. E. Bialkowski. Pulsed-laser excited thermal lens spectroscopy with sample-
fluid heat coupling. J. Appl. Phys. 2010. 107(8): 083512.
10. G. V. B. Lukasievicz, N. G. C. Astrath, L. C. Malacarne, L. S. Herculano, V. S. Zanuto, M.
L. Baesso, S. E. Bialkowski. "Pulsed-laser time-resolved thermal mirror technique in low-
absorbance homogeneous linear elastic materials". Appl. Spectrosc. 2013. 67(10): 1111-
1116.
8/10/2019 TM Opaque Abril03 2014. Nilson
15/15
15
11. H. S. Carslaw and J. C. Jaeger. Conduction of heat in solids (Clarendon Press, Oxford,
1959) Vol. 1, p. 78.12. W. Nowacki, Thermoelasticity (Pergamon, Oxford, 1982) Vol. 3, p. 11.
13. G. L. Bennis, R. Vyas, R. Gupta, S. Ang, W. D. Brown. "Thermal diffusivity measurement
of solid materials by the pulsed photothermal displacement technique". J. Appl. Phys.
1998. 84(7): 3602-3610.
14. B. C. Li. "Three-dimensional theory of pulsed photothermal deformation". J. Appl. Phys.
1990. 68(2): 482-487.
15. J. Shen, R. D. Lowe, R. D. Snook. A model for cw laser induced mode-mismatched dual-
beam thermal lens spectrometry. Chem. Phys. 1992. 165(2-3): 385-396.
16. Metals Handbook, Vol.2 - Properties and Selection: Nonferrous Alloys and Special-
Purpose Materials, ASM International 10th Ed. 1990.
17. MatWeb Material Property Data, "Aluminum 6351-T54"; available from
http://www.matweb.com/search/SpecificMaterial.asp?bassnum=NLN201.
18. M. Auinger et al: J. Europ. Opt. Soc. Rap. Public. 9, 14004 (2014).