Time Resolution Measurements for 8” MCP:Noise Filter and Shape Fitting

Alexander Vostrikov

University of Chicago

[email protected]

January 29, 2013

Fast Timing (PSEC) Group Tuesday Regular Meeting

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 1 / 12

Page 2: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Introduction

Goal

Improve differential time resolution.

Approaches

High frequency noise filtering.

Pulse shape fitting.

Data

8” MCP April 5, 2012.

Scope data: 10 Gsample/sec (100 psec).External trigger.Two opposite channels of the one stripline.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 2 / 12

Page 3: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Motivation for noise filtering

Differential time resolutiondepends on signal-to-noiseratio.

A lot of efforts to reduce thenoise were done at the APStesting facility.

Suggested approach

Cut the noise in frequencydomain during the dataanalysis.

Figure: Differential time resolution asa function of noise-to-signal ratio.8” MCP data collected onApril 5, 2012. [Andrey Elagin]

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 3 / 12

Page 4: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Motivation for pulse shape fitting

Figure: Time resolution versus thenumber of primary photo-electrons fordifferent timing techniques:single/multiple thresholds, CFD, fitting.[NIM A607: 387–393, 2009]

Figure: The same plot with the abscissaexpanded to cover from 40 to 110photo-electrons and the ordinateexpanded to cover from 1 to 9 psec.[NIM A607: 387–393, 2009]

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 4 / 12

Page 5: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Time and frequency domains

10 12 14 16 18 20 22 24Time, nsec

−80

−60

−40

−20

0

20

Volt

ag

e, m

V

Original

Filtered@ 800 MHz

Figure: Time domain.

0 1 2 3 4 5Frequency, GHz

−70

−60

−50

−40

−30

−20

−10

0

Fou

rier

Tra

nsf

orm

, d

BFigure: Frequency domain.

Noise filter removes all frequencies in Fourier transformation higherthan frequency cut level.

Here frequency cut level of 800 MHz selected.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 5 / 12

Page 6: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Events selection

0 1 2 3 4 5 6Full signal width (nsec)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Eve

nts / (100 psec)

Channel #1: 14064 events

Figure: Noise filter is OFF.

0 1 2 3 4 5 6Full signal width (nsec)

0

1000

2000

3000

4000

5000

6000

7000

Eve

nts / (100 psec)

Channel #1: 14005 events

Figure: Noise filter is ON.

Full signal width is defined as difference between CFD timing at 10%of falling and rising edges.

Selection cut is 1.3 nsec.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 6 / 12

Page 7: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Time resolution: Noise filter effect

0 50 100 150Time difference, psec

0

20

40

60

80

100

120

140

160

Eve

nts / (2 psec)

sigma=11.94 psec2035 events

Figure: Noise filter is OFF.

0 50 100 150Time difference, psec

0

50

100

150

200

250

Eve

nts / (2 psec)

sigma=6.90 psec1988 events

Figure: Noise filter is ON.

Pulses with amplitude between 110 and 130 mV selected.

Time difference distribution is fitted with Gaussian.

Time resolution is σ-parameter of the fit.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 7 / 12

Page 8: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Pulse shape fitting

Figure: Noise filter is OFF.

Pulse shape is fitted with Gaussian between two black lines.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 8 / 12

Page 9: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Time resolution: Pulse shape fitting effect

0 50 100 150Time difference, psec

0

20

40

60

80

100

120

140

160

Eve

nts / (2 psec)

sigma=11.94 psec2035 events

Figure: Spline. Noise filter is OFF.

0 50 100 150Time difference, psec

0

20

40

60

80

100

120

140

160

180

Eve

nts / (2 psec)

sigma=10.21 psec2035 events

Figure: Gaussian fit. Noise filter is OFF.

Pulses with amplitude between 110 and 130 mV selected.

Parameter t0 of Gaussian fit to the pulse shape is timing for the fitresults.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 9 / 12

Page 10: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Time resolution: Pulse shape fitting effect

0 50 100 150Time difference, psec

0

50

100

150

200

250

Eve

nts / (2 psec)

sigma=6.90 psec1988 events

Figure: Spline. Noise filter is ON.

0 50 100 150Time difference, psec

0

50

100

150

200

250

Eve

nts / (2 psec)

sigma=6.86 psec1988 events

Figure: Gaussian fit. Noise filter is ON.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 10 / 12

Page 11: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Differential time resolution vs. Pulse amplitude

0 50 100 150 200 250Pulse amplitude, mV

0

5

10

15

20

25

30

35

40

Differe

ntial time reso

lution, pse

c

CFD

Fit

CFD @ 800 MHz

Fit @ 800 MHz

Figure: Differential time resolution as afunction of pulse amplitude.

0 5 10 15 20 25 30Inverse ppulse amplitude, 1/V

0

5

10

15

20

25

30

35

40

Differential time resolution, psec

CFD

Fit

CFD @ 800 MHz

Fit @ 800 MHz

Figure: Differential time resolution as afunction of inverse pulse amplitude.

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 11 / 12

Page 12: Time Resolution Measurements for 8' MCP: Noise Filter and … · 2013-01-29 · Time Resolution Measurements for 8" MCP: Noise Filter and Shape Fitting Alexander Vostrikov University

Time Resolution Measurements for 8” MCP:Noise Filter and Shape Fitting

Alexander Vostrikov

University of Chicago

[email protected]

January 29, 2013

Fast Timing (PSEC) Group Tuesday Regular Meeting

A. Vostrikov (UChicago) Time Resolution Measurements for 8” MCP January 29, 2013 12 / 12