The Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer [email protected] Artificial Pancreas Center Department of Anesthesiology Jefferson Medical College Thomas Jefferson University Philadelphia, PA

The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer [email protected] Artificial Pancreas Center Department of

Embed Size (px)

Citation preview

Page 1: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

The Utility of Simultaneous Glucose Sensor Measurements

Brian Hipszer [email protected]

Artificial Pancreas CenterDepartment of Anesthesiology

Jefferson Medical CollegeThomas Jefferson University

Philadelphia, PA

Page 2: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

In-hospital Glycemic Management

Clinical glycemic management for a 63 y.o. white male with type 2 diabetes undergoing an esophagectomy (subject D2)

0

50

100

150

200

250

300

350

06:00 12:00 18:00 00:00 06:00 12:00 18:00 00:00 06:00 12:00 18:00 00:00

Time

Glu

cose

0

5

10

15

20

25

30

35

Insu

lin

Accu-Chek (mg/dl)Central Lab Glucose (mg/dl)D5 1/2 NSS (ml/hr)Tube Feeding (ml/hr)IV Insulin Infusion (U/hr)SC Insulin Bolus (U)IV Insulin Bolus (U)s

urgery

Page 3: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

In-hospital Glycemic Management

Percentage of time per patient a nurse must devote to frequent blood glucose monitoring4-8%

Ratio of patient to nurse in ICU (general floor)2:1 (4:1)

Average time to obtain blood glucose reading with conventional point-of-care device†~5 minutes

Average time between blood glucose tests1-2 hours

† Aragon D. Evaluation of nursing work effort and perceptions about blood glucose testing in tight glycemic control. Am J Crit Care. 2006 Jul;15(4):370-7.

Page 4: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

CGM in the Hospital

• Rationale– Improve glycemic management– Avoid hypoglycemia– Reduce workload and cost

• Requirements– Accurate– Reliable– User-friendly

Page 5: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

In-hospital CGM Evaluation• Assessment of two glucose-sensing technologies

– Interstitial fluid glucose sensors (TGMS)– Intravenous glucose sensor (VGMS)

TGMS

VGMSData collected under the project entitled “Artificial Pancreas for Control of BG and Insulin Levels in Hospitalized Patients with Diabetes and Stress

Hyperglycemia” sponsored by the Technologies in Metabolic Monitoring (TMM) Initiative with additional support from Medtronic Diabetes

Page 6: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

In-hospital CGM Evaluation• 10 patients studied in the perioperative period• Each patient studied for a maximum of 60 hours

transhiatal esophagectomy5.6‡2.936.463MD2hepatic resection cancelled due to metastatic cancer7.51.322.368MC2pancreatic resection cancelled due to metastatic cancer6.50.931.673MB2panniculectomy and exploratory laparotomy7.31.9#46.153FA2whipple procedure 5.0‡0.827.856MF3whipple procedure 5.51.220.751FE3whipple procedure 5.61.029.559MD3whipple procedure 6.12.122.958MC3whipple procedure 5.30.822.755MB3whipple procedure 6.10.719.147FA3

%ng/mlkg/m2yearsProcedureHbA1c†C-Peptide*BMIAgeSexSubject ID

* reported normal range is 0.8-3.5ng/ml unless otherwise noted# reported normal range is 0.8-3.1ng/ml† reported normal range is 3.6-6.9% unless otherwise noted‡ reported normal range is 4-6%

T2D

M

Page 7: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Reference Data

• Arterial (q20 min) and venous (q60 min) glucose, lactate, blood gases and electrolytes levels

• Capillary blood glucose levels (q3 hr)

• Arterial insulin and fatty acids levels• Urine analysis every hour• Recorded vital signs, IV infusion rates, body

position, sedation, meals and medications

Page 8: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

TGMS Sensors

• Guardian RT® sensors (Medtronic Diabetes, Northridge, CA)

• Grouped into two three-sensor arrays

• Inserted into the arm, chest or thigh

• Modified transmitters wirelessly transmit every minute

Photograph of TGMS sensor arrays implanted in subject B2

Page 9: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Sensor/Reference Data

Arterial (red circle) and venous (blue squares) reference glucose measurements (in mg/dl)

TGMS glucose sensor signals (in nA) smoothed with a 7th order FIR filter (sensors 1-6 are red, yellow, green, cyan, blue and magenta, respectively).

Data from subject D2

Page 10: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Sensor Performance

-1

-0.8

-0.6

-0.4

-0.2

0

0.2

0.4

0.6

0.8

1

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

ND

Venous

Subject

Reference

NDT2DMT2DM

ArterialVenousArterial

*Not pictured: arterial and venous outlying MARD values for subject D3, sensor 1 (1.29 and 2.27) and sensor 3 (3.17 and 1.39)

RMARD*

The mean absolute relative difference (MARD) and Pearson Correlation Coefficient (R) were calculated from paired reference/sensor values. Statistics were computed separately for arterial and venous reference measurements. Sensor data were smoothed using a 7th order FIR filter and recalibrated every six hours using a one-point calibration with a fixed offset after a two-hour run-in period.

NDNDT2DMT2DMSubject

VenousArterialVenousArterialReference

Page 11: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Sensor Combination Schemes

• Mean– Average all six sensor values

• Trimmed Mean– Rank the six sensor values and average the

second, third, forth and fifth ranked values• Median

– Rank the six sensor values and average the third and forth ranked values

Page 12: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Combined sensor measures (mean, trimmed mean and median) are displayed in red, green and blue.

Combined Sensor Performance

0.140.170.250.26 (0.16 - 0.85)venous

0.090.100.120.16 (0.08 - 0.34)arterialMARD

MedianTrimmed MeanMeanmedian (range)

Combined MeasuresIndividual SensorReferenceStatistic

Data from subject D2

Page 13: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Combined Sensor PerformanceA

rteria

l MA

RD

*Not pictured: arterial MARD values for subject D3, sensor 1 and sensor 3 (1.29 and 3.17)

0.160.120.110.14 (0.10 - 0.44)F30.240.230.220.26 (0.09 - 0.51)E30.780.260.180.39 (0.13 - 3.17)D30.150.120.120.13 (0.10 - 0.83)C30.160.170.170.17 (0.16 - 0.21)B30.130.120.120.15 (0.14 - 0.20)A30.120.100.090.16 (0.08 - 0.34)D20.280.270.300.29 (0.20 - 0.49)C20.100.110.110.18 (0.10 - 0.47)B20.110.110.110.11 (0.09 - 0.15)A2MeanTrimmed MeanMedianmedian (range)

Combined MeasuresIndividual SensorsSubject

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

Individual SensorsMedianTrimmed MeanMean

A2 B2 C2 D2 A3 B3 C3 D3* E3 F3T2DM

Page 14: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Combined Sensor PerformanceV

enou

s M

AR

D

*Not pictured: venous MARD values for subject D3, sensor 1 and sensor 3 (2.27 and 1.39)

†Insufficient venous reference glucose measurements to complete analysis0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

Individual SensorsMedianTrimmed MeanMean

A2 B2† C2† D2 A3 B3 C3 D3* E3 F3T2DM

0.160.140.130.19 (0.12 - 0.36)F30.270.250.230.29 (0.19 - 0.53)E30.660.180.140.28 (0.16 - 2.27)D30.200.130.130.13 (0.13 - 0.76)C30.170.170.170.18 (0.16 - 0.23)B30.160.160.160.17 (0.15 - 0.19)A30.250.170.140.25 (0.16 - 0.85)D2

††††C2††††B2

0.180.180.180.19 (0.16 - 0.20)A2MeanTrimmed MeanMedianmedian (range)

Combined MeasuresIndividual SensorsSubject

Page 15: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Multiple Sensors

• Robust estimate of blood glucose level (accuracy)

• Identify and replace failed sensor without interruption of data (reliability)

Page 16: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

AcknowledgementsDepartment of the Army’s Technologies in Metabolic Monitoring (TMM) Initiative

Investigators and Research Personnel

Neil Seligman, MDJen Soares, RNAnne Marlay, RN

Teresa Campo, RNJason McConomy, RNAnn Liotino, RN

Sean McShane, RNElise Dorr-Dorynek, RNAmy Callahan, RN

Joanne VesciSarah Buckley, RNElisabeth McNeal, RNAdrianne Moore, RN

Waleed ShahRebecca Brown, RN Eileen Gleason Donnelly, RNJennifer Lessin, RN

Paul DidomenicoPatty McGovern, RNDyllan Siemann, RNKevin Furlong, MD

Patrick ShumLisa Wus, RNDawn Gillespie, RNInna Chervoneva, PhD

Matthew MufflyLarissa Lightstone, RNDawn Fisher, RNBarry Goldstein, MD, PhD

Jonathan TannebaumKerin Perry, RN Cindy Trappler, RNZvi Grunwald, MD

Garry PowellKathleen O'Malley, RNCarrie Christiansen, RNCharles Yeo, MD

Cheryl StarrettKate Passey, RNCarleo Naluan, RNJames Heitz, MD

Carin KozlowskiKate Ashburn, RNBonnie Grady, RNDavid Maguire, MD

Amanda FurlongJulia Snyder, RNBrook Redeyoff, RNDavid Gratch, DO

Michael Picone, MDJohn Furlong, RNAshley Benedict, RNJeffrey Joseph, DO

Medtronic Diabetes

Page 17: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of
Page 18: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of
Page 19: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Subject E3

Page 20: The Utility of Simultaneous Glucose Sensor … Utility of Simultaneous Glucose Sensor Measurements Brian Hipszer brian.hipszer@jefferson.edu Artificial Pancreas Center Department of

Subject B2