108
C. I. R. Braem, BSc. The Philips wearable biosensor in transcatheter aortic valve implantation treatment workflow March 22 nd , 2019 Usability and feasibility of the wearable biosensor

The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

C. I. R. Braem, BSc.

The Philips wearable biosensor in transcatheter aortic valve implantation

treatment workflow

March 22nd, 2019

Usability and feasibility of the wearable biosensor

Page 2: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 3: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

The Philips wearable biosensor in transcatheter aortic valve implantation treatment workflow.

For the title of Master of Science in Technical MedicineCarlijn I. R. Braem, BSc.22-03-2019 Grudation committeeMedical supervisor: Dr. M.M. VisDaily supervisor: Dr. M.S. van MourikTechnical supervisor: Prof. Dr. H.J. HermensProcess supervisor: Drs. P. A. van KatwijkExternal member: Dr. E. J. F. M. Ten Berge

University of TwenteTechnical MedicineFaculty of Science and Technology Postbus 217 7500 AE EnschedeThe Netherlands

Usability and feasibility of the wearable biosensor

Page 4: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 5: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

“Not everything that counts can be counted, and not everything that can be counted counts.”

- Albert Einstein

Page 6: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 7: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

v

Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve implantation treatment workflow. – Usability and feasibility of the wearable biosensor’, describing the rationale and first results of the TELE-TAVI study. This study examines a patch sensor and additional phone receiver for tele-monitoring in-TAVI patients. The thesis has been written to obtain the master’s degree in Technical Medicine at the University of Twente. It is the result of my graduation internship at the Amsterdam UMC, location AMC, from February 2018 to March 2019.

The TELE-TAVI study concept was designed by Martijn van Mourik and Marije Vis, who started the collaboration with Philips and medical ethical approval, which took over a year. My activities within this study was the realization, execution and (first) analysis of the TELE-TAVI. It was a long, bumpy road, but the study is almost at its end.

I would like to thank all my supervisors who helped me accomplish this thesis. My supervisors at the Amsterdam UMC, Martijn van Mourik and Marije Vis, your advice and supervision helped me to complete this thesis. Hermie Hermens and Ainara Garde-Martinez from the University of Twente, you assisted me in the technical difficulties of the project and widen my scope. Paul van Katwijk, thank you for the guidance in this process. Erik ten Berge, I am delighted that you are part of my gradation committee. I also wish to thank all of the participants, without whom I would not have been able to conduct this study.

To my other colleagues at the Amsterdam UMC, I would like thank you for your insights into research. I want to thank the members of my group intervision for the helpful meetings and advices.

Finally I would like to thank my family, friends, and especially my love, for being helpful and supportive during my internship and time studying Technical Medicine at the University of Twente.

Carlijn Braem

Amsterdam, March 13th, 2019

Preface

Page 8: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

vi

Page 9: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

vii

Transcatheter aortic valve implantation (TAVI) is currently standard care for intermediate to high risk patients in patients with aortic valve stenosis, which is associated with aging and has a high burden on health care. Current screening tools however are insufficient as frailty is not included. A wearable sensor could allow for an in-depth analysis for screening TAVI patients. Besides, post-procedure monitoring and TAVI follow-up could benefit from extended monitoring. This thesis reviews the usability and feasibility of the Philips wearable biosensor for TAVI workflow.

The TELE-TAVI study is an observational, prospective, investigator initiated pilot, started in June 2018 in the Amsterdam UMC, location AMC (Amsterdam, the Netherlands). The wearable biosensor (Philips Medical Systems, Andover, Massachusetts, USA) is a lightweight, wireless, wearable medical-grade biosensor, that can measure vital signs and detects posture for up to 4 days. Healthy volunteers and patients in work-up for TAVI were included. Healthy volunteers were enrolled and received one biosensor to test the system technically. TAVI patients received the wearable biosensor thrice; before the TAVI procedure (T0), directly post TAVI procedure on the cardiac care unit (CCU) (T1) and 6-weeks after the TAVI procedure (T2). The reliability of the biosensors vital signs was compared to a standard care monitor (Philips MP70 monitor, Philips Healthcare, Eindhoven, the Netherlands). Posture detection reliability was tested with walking exercises and compared to diaries and data collection reliability was assessed. TAVI patient experience with the system was reviewed with the post study system usability questionnaire (PSSUQ) and a custom made questionnaire. Activity was estimated with the integral of the modulus of the accelerometer output (IMA), and with thresholds activity classification and daily activity levels were computed.

At February 22nd of 2019, a total of 6 healthy and 24 TAVI patients were enrolled in the TELE-TAVI study. Eighteen and eight TAVI patients completed measurements at T1 and T2, respectively and ten patients dropped out. The TAVI population is 76.6 (± 4.8) years old, 75% male. Vital signs limit of agreement was between -3.9 -7.0 and -8.1 and 7.8 for heart and respiratory rate respectively. Walking was detected by the biosensor if the gait speed was higher than 0.7 m/s. After one day, posture detection diverged substantially. Of the 96 recording hours, 56.2% is recorded with no gaps in the data. 45 wearability questionnaires were received and the PSSUQ showed an overall system satisfaction of 63.2% (± 30.7%). Sensor wear was comfortable, but the sensor fell off in 31% of the patients IMA correlates with gait speed (r2 = 0.8 and p<0.01). No, low, medium and high activity levels are 63.3%, 25.2%, 10.9% and 0.6%, respectively and daily activity levels are 40.8%, 28.7%, 12.2% and 0.6%.

Patients tolerated wearing the biosensor well and activity classification can give insight in patient activity patterns. The feasibility of the wearable biosensors is as of yet insufficient, as the reliability of the biosensor is deficient compared to the predefined criteria. Data collection reliability is low and posture detection is unusable, as detection deteriorates within a day. . The usability of the wearable biosensor shows great promise to improve TAVI work flow and encourages research in sensor technology in elderly.

Abstract

Page 10: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

viii

Page 11: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

ix

Preface

Abstract

Contents

Abbreviations

Chapter 1: Introduction 1.1. Aortic valve stenosis .............................................................................................................. 1

1.2. Transcatheter aortic valve implantation ................................................................................ 1

1.3. Frailty ..................................................................................................................................... 2

1.4. Wearable biosensor ............................................................................................................... 3

1.5. TELE-TAVI study ..................................................................................................................... 3

Chapter 2: Research Questions 2.1. Research question ................................................................................................................. 5

2.2. Outline thesis......................................................................................................................... 5

Contents

Page 12: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

x

Contents

Chapter 3: Feasibility and usability of a wearable patch sensor in monitoring vital signs and activity in transcatheter aortic valve implantation patients: design and rationale of the TELE-TAVI study.

3.1. Introduction ........................................................................................................................... 9

3.2. Method ................................................................................................................................ 103.2.1. General design ........................................................................................................................................... 103.2.2. Devices ....................................................................................................................................................... 10

Biosensor ........................................................................................................................................... 10Research Kit ....................................................................................................................................... 13Philips MP50 Monitor ....................................................................................................................... 13

3.2.3. Study population ........................................................................................................................................ 13Healthy subjects ................................................................................................................................. 13TAVI patients ...................................................................................................................................... 17

3.3. Discussion ............................................................................................................................ 17

Chapter 4: Patient characteristics before and after transcatheter aortic valve implantation: First results of the TELE-TAVI study.

4.1. Introduction ......................................................................................................................... 19

4.2. Method ................................................................................................................................ 194.2.1. Devices ....................................................................................................................................................... 194.2.2. Study population ........................................................................................................................................ 19

Healthy subjects ................................................................................................................................. 19TAVI patients ...................................................................................................................................... 20

4.2.3. Statistical analysis ....................................................................................................................................... 21

4.3. Results ................................................................................................................................. 224.3.1. Healthy subjects ......................................................................................................................................... 224.3.2. TAVI patients............................................................................................................................................... 22

4.4. Discussion ............................................................................................................................ 24

Chapter 5: Reliability of the Philips wearable biosensor in monitoring vital signs and activity

5.1. Introduction ......................................................................................................................... 27

5.2. Method ................................................................................................................................ 285.2.1. Vital signs ................................................................................................................................................... 30

Signal analysis .................................................................................................................................... 305.2.2. Posture detection ....................................................................................................................................... 325.2.3. Data collection............................................................................................................................................ 32

5.3. Results ................................................................................................................................. 345.3.1. Vital signs ................................................................................................................................................... 345.3.2. Posture detection ....................................................................................................................................... 355.3.3. Data collection............................................................................................................................................ 35

5.4. Discussion ............................................................................................................................ 365.4.1. Posture detection ....................................................................................................................................... 375.4.2. Data collection............................................................................................................................................ 38

5.5. Conclusion ........................................................................................................................... 38

Page 13: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Contents

xi

Chapter 6: End-user experience of the Philips wearable biosensor and Research Kit in the TELE-TAVI study

6.1. Introduction ......................................................................................................................... 41

6.2. Method ................................................................................................................................ 416.2.1. Devices ....................................................................................................................................................... 416.2.2. PSSUQ ......................................................................................................................................................... 426.2.3. Custom made questionnaire ...................................................................................................................... 426.2.4. Statistical analysis ....................................................................................................................................... 42

6.3. Results ................................................................................................................................. 42

6.4. Discussion ............................................................................................................................ 43

6.5. Conclusion ........................................................................................................................... 44

Chapter 7: Discussion7.1. Implications ......................................................................................................................... 47

7.1.1. Recommendations ..................................................................................................................................... 48

7.2. Future perspectives ............................................................................................................. 487.2.1. TAVI ............................................................................................................................................................ 497.2.2. Wearable sensors ....................................................................................................................................... 49

Chapter 8: Objective physical activity levels with the wearable biosensors in patients undergoing transcatheter aortic valve implantation: first results

8.1. Introduction ......................................................................................................................... 51

8.2. Method ............................................................................................................................... 518.2.1. Data analysis............................................................................................................................................... 528.2.2. IMA ............................................................................................................................................................. 528.2.3. Activity levels.............................................................................................................................................. 528.2.4. Daily activity levels ..................................................................................................................................... 52

8.3. Results ................................................................................................................................. 538.3.1. IMA ............................................................................................................................................................ 538.3.2. Activity levels ............................................................................................................................................. 538.3.3. Daily activity levels ..................................................................................................................................... 54

8.4. Discussion ............................................................................................................................ 54

8.5. Conclusion ........................................................................................................................... 56

Chapter 9: References

Page 14: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

xii

Contents

Chapter A: AppendixA.1. Overview of data collected during TELE-TAVI study ...........................................................A-I

A.2. Supplementary data on the reliability of the Philips wearable biosensor ..........................A-IIA.2.1. Distribution difference data ......................................................................................................................A-IIA.2.2. Precision and bias of the wearable biosensor ......................................................................................... A-IVA.2.3. Walking detection of the wearable biosensor ...........................................................................................A-XA.2.4. Measurement length ...............................................................................................................................A-XII

A.3. Supplementary data of the wearability questionnaires ..................................................A-XVIA.3.1. Wearability questionnaires ................................................................................................................... A-XVIA.3.2. Results from PSSUQ for every study participant .................................................................................... A-XXA.3.3. Results custom made questionnaire for every measurement moment ................................................ A-XXI

A.4. Supplementary data of the activity results ....................................................................A-XXIIA.4.1. Results of activity analysis .................................................................................................................... A-XXIIA.4.2. Results of daily activity ........................................................................................................................ A-XXIII

Page 15: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

xiii

Page 16: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

xiv

Page 17: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

xv

Abbreviation

6MWT Six minute walk test

AF Atrial fibrillation

AoS Aortic valve stenosis

BLE Bluetooth Low-Engery

BMI Body mass index

Bpm Beats per minute

Brpm Breaths per minute

CCS Canadian Cardiovascular Society grading of angina pectoris

COPD Chronic obstructive pulmonary disease

ECG Electrocardiogram

EuroSCORE European system for cardiac operative risk evaluation

G Gravitional force (9.81 m/s^2)

HR Heart rate

IMA Integral of the modulus of acceleration

IQR Interquartal range

METS Metabolic equivalent score;

mG Micro Gravition force (9.81 *10^-3 m/s^2)

mV Micro Voltage

NYHA New York Heart Association functional classification.

PSSUQ Post-study system usability questionnaire

RespR Respiratory rate

RPC Reproducibility coefficient

R-R interval R peak to R-Peak interval in the ECG

SAVR Surgical aortic valve replacement

SD Standard deviation

SF36 Short From (36) Health Servey

SPPB Short physical performance battery

STS Society of Thoracic Surgery predicted risk of mortality

T0-T2 Measurement times:; T0: pre-operative, T1: direct post-TAVI, T2: follow-up

TAVI Transcatheter aortic valve implantation

TELE-TAVI Observational pilot study to assess usability of a wearable patch sensor in monitoring vital signs and activity in TAVI patients.

Abbreviations

Page 18: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

xvi

Page 19: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

1

1.1. Aortic valve stenosisThe Western world is aging, as

the life expectancy increases and baby boomers are becoming of age. With this, the number of people with cardiovascular diseases rises, including heart valve disorders.

The heart consist of two smaller and two larger chambers, which are separated with four valves. In aortic valve stenosis (AoS) the valve separating the left ventricle from the aorta, is affected (Figure 1.1) [1]. The leaflets of the aortic valve deteriorate and develop calcifications, which leads to obstruction of left ventricular blood outflow. This results in inadequate cardiac output, decreased exercise capacity, heart failure, and when left untreated death from these cardiovascular causes. The prevalence of AoS increases with age, in which 3.4% of the elderly (>75 years) has severe AoS [2]. It is associated with high burden on health care and patients quality of life [2], [3].

1.2. Transcatheter aortic valve implantationNowadays, an aortic valve replacement is indicated when AoS is severe and the patient experiences

symptoms. The aortic valve can be replaced during cardiac surgery; surgical aortic valve replacement (SAVR). An alternative for SAVR is transcatheter aortic valve implantation (TAVI), in which a bio-prosthetic aortic valve is implanted with a catheter inserted in the femoral artery, subclavian artery, ascending aorta, or via the cardiac apex. Of these, the femoral approach is preferred as it is the least invasive and associated with the lowest risk (Figure 1.2) and can be performed without general anesthesia [4].

In the decision making between SAVR or TAVI patients are screened for surgical risk. The European System for Cardiac Operative Risk Evaluation (EuroSCORE) or Society of Thoracic Surgeons (STS) score are used for screening [6], [7]. Both are risk stratification models for mortality after 30 days after cardiac surgery. If patients score high to moderate on these scales (STS or EUROscore II ≥ 4%), they are considered

Chapter 1: Introduction

Figure 1.1 Anatomy of the heart and valves. The left shows a cross-sec-tion of the heart, showing the left ventricle and aorta, separated by the aortic valve. In the left upper panel, a healthy aortic valve is shown. The leaflets can open and close fully. In the lower right panel, stenotic aortic valves are shown. Due to the deterioration and calcifications the valves cannot open and close properly [1].

Page 20: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

2

Chapter 1

eligible for TAVI [8]. However, both scores are not ideally suited for TAVI procedures and TAVI risk algorithm is needed [4], [9], [10].

Next to the 30-day mortality risk, other clinical patient characters are evaluated for the decision between SAVR and TAVI, such as age and frailty. The TAVI procedure is preferred in patients above 75 years of age. Even so, calendar age is an insensitive and non-specific measure for preoperative risk assessment [11].

1.3. FrailtyFrailty on the other hand is closely related to adverse surgical outcomes, such as mortality, morbidity,

and functional decline [11]–[14]. However, it remains difficult to define, due to its multi-factorial and multi-expressional nature. Frailty involves a decline of feedback complexity in physiologic systems, followed by a loss of homeostatic reserves resulting in vulnerability to external stressors. It can be observed as weight loss, muscle weakness, poor endurance and energy, slowness and low physical activity levels [15]. The prevalence of frailty is high among elderly (22.7%) and increases even further with age [12], [13].

There is a vast amount of frailty assessment tools, as a result of the complex and debated definition of frailty. Generally, there are two types of tests, qualitative questionnaires involving frailty phenotype related questions and physical performance tests [16], [17]. Despite the enormous collection of diagnostic tools for frailty, there is no international standard and it is generally acknowledged that there is a need for a standard validated objective frailty assessment tool [14], [16], [18].

Multiple studies attempt to provide a generalized frailty assessment with technology. For example, the use of inertial sensors or accelerometers in a phone or wearable sensors. Hereby several outcomes are analyzed, such as gait, balance, physical performance or activity. These are used individually or in combination with conventional measures [19]. Wearable sensors can also be used to assess the impairment of cardiac autonomic nervous control by analyzing heart rate (HR) and heart rate variability (HRV) characteristics [20]. HRV changes as a consequence of loss in physiologic complexity in frail patietns.

In summary, severe AoS results in inadequate cardiac output and loss of exercise capacity. TAVI has become standard intervention for moderate to high risk patients. Frequently used risk scores are limited in predicting 30-day mortality after TAVI. Frailty, however is closely related to surgical outcome, but a standardized frailty tool is lacking. Frailty and AoS are both associated with declined physical activity and impaired (cardiac) functioning. To date, no research is known attempting to combine physical activity and cardiac physiological parameters as a screenings measure for patients with AoS.

Figure 1.2 Transcatheter aortic valve implantation. A) Under radiological guidance, a catheter is placed into the left ventricle via the femoral artery and aorta. A balloon is threaded through the diseased aortic valve, after which balloon valvuloplasty is performed to dilate the diseased aortic valve. B) A balloon device with the new valve attached is thread through the diseased aortic valve. C) During pacing of the right ventricle by an external pacemaker wire, the new valve is unfolded using inflation of the balloon. D) The balloon is deflated, after which the new aortic valve will directly function. Modified from the ISAR Heart Centre Munich [5].

Page 21: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 1

3

1.4. Wearable biosensorIn 2016, Philips launched the wearable biosensor (Philips Medical Systems, Andover, Massachusetts,

USA) a lightweight, wireless, wearable medical-grade biosensor. The sensor records mono-lead ECG, respiratory rate (RespR), skin temperature and tri-axial acceleration for step count and postural positions with a battery life of four days. Next to these computed parameters, the raw data of the sensor can be extracted, such as tri-axial accelerometer, mono-lead ECG and thermometer. The combination of physiological outcome parameters, HR and RespR, and posture detection could allow for an integrated analysis, which can be used for screening TAVI patients.

However, no study is published on the wearable biosensor and only few on its predecessor, the HealthPatch MD (VitalConnect, San Jose, California, USA), who has the same firm- and hardware [21]–[23]. Therefore, reliability of the biosensor is as of yet, largely debated. Also little data is available on the patient perception of the wearable biosensor [24].

1.5. TELE-TAVI studyIn the Amsterdam UMC (location AMC, Amsterdam, The Netherlands), a pilot study was set-up to

investigate usability and feasibility of the wearable biosensor to monitor cardiac condition and assess frailty and treatment effects in TAVI-patients (TELE-TAVI). In this study two groups of participants were included; healthy subjects and TAVI-patients.

Data collected from the healthy volunteers will be used for algorithm development, reliability assessment and provides the researchers with experience of the biosensor system. In TAVI-patients, the biosensor is attached before, directly after and six weeks after the TAVI procedure, so data on the screening, monitoring and follow-up is collected. Frailty and functional status was assessed pre-procedural and at six weeks follow-up.

The realization of TELE-TAVI entitles a great part of this master graduation. Data from the TELE-TAVI study is used for this master thesis, to assess the usability of the wearable biosensor in the TAVI workflow.

Figure 1.3 Philips wearable biosensor, on a patient [64].].

Page 22: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

4

Page 23: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

5

In Chapter 1 we showed that mortality risk assessments for transcatheter aortic valve implantation is limited. The wearable biosensor could be usable in transcatheter aortic valve implantation workflow, using physiological parameters and activity. However, the reliability and end-user perception is largely unknown. Also, an algorithm for assessment and quantification of physical activity is not yet available for the wearable biosensor.

2.1. Research questionThis leads to the following research question:

What is the usability and feasibility of the Philips wearable biosensor for transcatheter aortic valve implantation workflow?

The following sub questions will be addressed in this thesis:

1. How reliable are the Philips wearable biosensors vital signs, posture detection and data collection?

2. How do TAVI-patients experience the use of the Philips wearable biosensor and additional systems?

3. How can we objectively measure physical activity with the Philips wearable biosensor?

2.2. Outline thesisThese research questions are part of the aforementioned TELE-TAVI study. This study is a prospective,

investigator initiated study.

The general outline of the master thesis is given in Figure 2.1. The rationale and design of the TELE-TAVI study will be given in Chapter 3. Preliminary results of the TELE-TAVI study are presented in Chapter 4. Data from the TELE-TAVI study is used to assess the reliability of the wearable biosensor, given in Chapter 5. Next, the biosensor user experience during the TELE-TAVI study is presented in Chapter 6. Hereafter, activity classification with the wearable biosensor are given in Chapter 7. Overall implications, conclusions and future perspectives are given in Chapter 8.

Chapter 2: Research Questions

Page 24: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

6

Chapter 2

Figure 2.1 Overview of the chapters in this master thesis

Chapter 3TELE-TAVI Study design

and rationale

Chapter 2Research questions

Chapter 1Introduction

Chapter 8Discussion

Chapter 5Reliability biosensor

Research question 2Research question 1 Research question 3

Chapter 6User experience

biosensor

Chapter 7Activity analysis

Chapter 4First results TELE-TAVI

Page 25: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

7

Page 26: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

8

Page 27: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

9

3.1. IntroductionSevere aortic valve stenosis (AoS) occurs in 3.4% of the elderly (>75 years) in which heart valve

replacement is standard treatment [2]. For patients with high or moderate surgical risk, a minimally invasive transcatheter aortic valve replacement (TAVI) is currently standard care [25].Frequently used risk scores are limited in predicting 30-day mortality after TAVI, as major risk factors, such as frailty, are not included [25]. Therefore there is a need for a new pre-surgical assessment [9], [10]. Remote monitoring in a home situation has the potential to objectively assess frailty, as well as to give a better understanding of the patients (cardiac) condition. Compared to a single evaluation in the clinical setting, home monitoring can provide an unbiased evaluation over multiple days. Obtained information could aid the decision making of AoS treatment; surgical, TAVI or conservative treatment.

Monitoring after TAVI is critical, as one major complication is the onset of new conduction disturbances. Therefore telemetry monitoring is now mandatory, which confides patients to the hospital. Home monitoring could extend the monitoring period, without lengthening the patients hospital stay. It could also help in the challenging first part of the rehabilitation, as well as providing objective and reliable information of the post procedural effect of the treatment and status of rehabilitation.

Currently there are wearable devices available on the market that aid monitoring and diagnosing of patients inside and outside the hospital [19], [26]–[28]. These devices measure different vital signs like heart- and respiration rate and more, for example, activity level. The potential of such devices are tremendous, but in daily practice it remains minimal, as more experience with these systems is needed.

The aim of this study is to investigate usability and feasibility of the wearable biosensor to monitor cardiac condition and assess frailty and treatment effects in TAVI-patients. Three potential cases were selected, to show were a wearable sensor could improve TAVI workflow:

T0: Pre-operative screening: a home measurement of vital signs and physical activity. Acquired data will be used to calculate and objectify frailty, clinical symptoms of AoS and give a presurgical evaluation of the patient condition.

T0: Direct post-procedural monitoring: by means of a wearable patch, a patient can be ambulatory monitored to detect early deterioration after the TAVI procedure. Especially cardiac conduction disorders can be monitored.

T0: Follow-up measurement: for analysis of objective clinical results of TAVI patients, extended home measurement can be performed and compared to the obtained pre-operative baseline.

Chapter 3: Feasibility and usability of a wearable patch sensor in monitoring vital signs and activity in transcatheter aortic valve implantation patients: design and rationale of the TELE-TAVI study.

Page 28: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

10

Chapter 3

3.2. Method3.2.1. General design

To examine the usability and wearability of a patch sensor and additional phone receiver in TAVI patients the tele-monitoring in-TAVI patients (TELE-TAVI) study was conducted. TELE-TAVI is an observational, prospective, investigator initiated pilot. The study started in June 2018 and is still including eligible patients in the Amsterdam UMC, location AMC (Amsterdam, the Netherlands). The trial was in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

3.2.2. DevicesIn the TELE-TAVI study, the Philips wearable biosensor and matching Research Kit (Philips Medical

System, Andover, Massachusetts, USA) was used. A standard care monitor of Philips (Philips MP70 monitor, Philips Healthcare, Eindhoven, the Netherlands) was utilized as reference. All devices were provided by Philips.

BiosensorThe Philips wearable biosensor is a wireless, lightweight, medical-grade sensor designed for long-term

monitoring of vital signs. The single-use patch contains two ECG-electrodes, a tri-axial accelerometer, a thermistor, a zinc-air battery and a Bluetooth Low-Energy (BLE) transceiver. The biosensor is placed on one of two allocated places on the chest (Figure 3.1) and can measure vital signs for up to four days. Specifications of the measured vital signs can be found in Table 3.1.

ECGTwo ECG-electrodes compute a continuous single-lead ECG at a sample rate of 125 Hz, of which heart

rate and R-R peak interval is derived. QRS complexes are automatically detected from the single-lead ECG, with a validated algorithm using wavelet transforms [29], [30]. The R-R interval is computed as the time duration between two detected consecutive QRS complexes. The heart rate (HR) is instantaneously computed as the reciprocal of the R-R interval. To smooth the obtained heart rate signal, a 10-beat low-pass filter is applied. [22]

Figure 3.1 Two allocated sites for wearable biosensor placement, upper left chest and rib cage below chest. [26]

Table 3.1 Criteria of the Philips wearable biosensors fall de-tection, as reported by Chan et al. [22]

Criteria fall detection

1. Detection of impact and free-fall

2. Large differences in acceleration in a small time window

3. Change from vertical to horizontal posture

4. Low activity for a specified duration after posture change

bpm indicates beats per minute; brpm, breaths per minute; ECG, electrocardiogram; mV, micro Voltage

Figure 3.2 - Next page - Biosensor placement and accelerometer directions: A: Showing biosensor place on a patient with the normal force (Fnormal). The window shows the biosensor and the raw, not calibrated axes (x, y, z). The raw x-axis is pointing downward, perpendicular to the length of the biosensor. The y-axis pointed parallel to the length of the biosensor. The z-axis points from the bio-sensor up or ‘through’ the paper. Under the patient, the calibrated axes (posterior to inferior, lateral to medial and inferior to posterior (vertical)) are shown in dark blue. B: Raw accelerometer data in three axes of the transition from sitting to walking (100 steps/min) of Pin 101. First few seconds, the subject is sitting, where after a swing from sitting to standing is seen. The end shows the subject walking (at 15:51:05). C: Shows the modulus of the accelerometer, the magnitude of the forces detected by the accelerometer. Visible in sitting time is a magnitude of 1G, which is the normal force (Fnormal). D: Gives the step count as biosensor output. After 5 seconds walking, the step count increases; first fast, where after, the increase slows down. The first big increase compensates for the 5 seconds lost for determination of the posture change, visible as the (estimated) dashed line.

Page 29: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 3

11

-101 -101 -101 0123

BiosensorRaw accelerometer (G)

x-axis y-axis z-axis

Modulus accelerometer (G)

√ x2+y

2+z

2

BiosensorStep count

xy

z

F norm

al F norm

al

Bio

sen

sor

axes

, sit

tin

gA

B C D

Sitt

ing

Wal

king

15:5

1:00

15:5

1:05

15:5

1:10

15:5

1:15

0102030St

ep c

ount

Estim

ate

of s

tep

coun

t

Ante

rior

Superior

Late

ral

Calib

rate

d ax

es

F norm

al

Page 30: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

12

Chapter 3

Tri-axial accelerometerThe wearable biosensor contains a tri-axial accelerometer. An accelerometer measures a change of

speed compared to the free fall state. When an object is in free fall, the gravitational force only influences the object. This means, that an accelerometer cannot detect gravitational acceleration itself. When the accelerometer is placed on a surface, it will measure the ground reaction force or normal force. Therefore, when an accelerometer is placed on a surface, it measures a force of one gravity unit (g0 = 9.81 m/s2) upwards. A tri-axial accelerometer has three perpendicular placed accelerometers, measuring acceleration in three orthogonal directions. As none of the axes is aligned with the normal vector, all accelerometers show some of the normal force when placed on a horizontal surface. This can be seen in Figure 3.2, where a person is sitting up and all the axes a part of the normal vector induces a displacement. Therefore, the wearable biosensor automatically calibrated to obtain a vertical, anterior-posterior and left-right lateral axis. With the calibrated accelerometer falls and posture is detected. Posture is classified as upright, leaning, lying, walking or unknown within five seconds of postural change. Static postures (upright, leaning, lying) are detected based on the angle of the thorax of the individual. Walking is detected based on a threshold in vertical acceleration and the ability to count steps. By peak picking regular peaks in the vertical axis, steps are counted. Falls are detected when several criteria are met, see Table 3.1. The calibrated accelerometer signal is not recorded or available for further analysis.

RespirationRespiration is not directly measured by the wearable biosensor, but estimated using ECG characteristics

and the calibrated accelerometer signal. The R-R interval computed from the ECG modulates due to respiration, which is known as the respiratory sinus arrhythmia. The autonomic nervous system induces a HR increase during inspiration and decrease in expiration, respectively. Next to that, chest movement of the respiration causes movements of the heart and its axis. This is visible as a modulation of the voltage or height of the QRS complex. Chest movements, although small, induce a change in thoracic angle, which can be detected by the accelerometer. Obtaining the respiratory rate (RespR) from ECG or acceleration extensive smoothing and filtering is needed. Hereafter the RespR is estimated from the picked peaks over a

Table 3.2 Overview of the Philips wearable biosensor output characteristics.

Data name Method Unit Fs (Hz) Range Ref

Single-lead ECG micro Voltage 125 [-10 10] mV

Heart Rate ECG derived Beats/minute 0.25* [30 200] bpm [24]

Raw accelerometer (x,y,z) mG 50 [-4 4] G

Position (Posture) Accelerometer derived - 1 [0 11]

Step Count Accelerometer derived Steps 1* [0 65535] steps

Fall detection Accelerometer derived - - - [22]

Respiratory rateECG and accelerometer

derived Breaths/minute 0.25 + brpm [23]

bpm indicates beats per minute; brpm, breaths per minute; ECG, electrocardiogram; G, gravitation; mV, micro Voltage

Table 3.3 Overview of patient monitor output characteristics.

Data name Unit Fs (Hz) Range Accuracy Ref

5-lead ECG micro Voltage 125 [34]

Heart Rate Beats/minute 0.98 [15 300] bmp ± 1% of range

Impedance pneumography mG 62.5 [34]

Respiratory rate Breaths/minute 0.98 [0 120] brpm at 0 to 120 brpm: ±1 brpmbpm indicates beats per minute; brpm, breaths per minute; ECG, electrocardiogram; mV, micro Voltage

Page 31: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 3

13

window of 45 seconds, shifting every 4 seconds. [23]

For each individual method, a separate RespR is computed, where after these are combined dependent on the quality of the underlying signals. The algorithm favors the computed breathing rate to be regular, because than it is more likely to reflect the true breathing rate. A quality metric is calculated from the regularity features of the three computed breathing rates. This quality metric is used to make a weighted average of the final estimated breathing rate. [23]

ECG derived respiration has a major limitation, as it can only can detect breathing rates up to half the heart rate, due to aliasing. Therefore, RespR is solely computed with the accelerometer data when the breathing rate is higher than half the heart rate and the accelerometer signal quality is sufficient. [23]

Research KitThe Research Kits intended use is for clinical researchers to gather, analyze and review patient data

retrospectively and conduct offline analysis with the Philips wearable biosensor. The Research Kit contains a smart phone (Kyocera BRIGADIER) preinstalled with the Research Kit mobile application as well as an offline desktop application. The mobile phone with app captures data acquired by wearable biosensor via Bluetooth and saves the data encrypted on the SD-card. The app also provides guidance for the set-up of the biosensor; connecting biosensor to mobile phone, skin adherence of the biosensor, set-up Bluetooth connection sensor and phone, accelerometer calibration, alerts the user of error conditions and gives suggestion to resolve these problems.

Philips MP50 Monitor As reference, the Philips MP70 monitor (Philips Healthcare, Eindhoven, the Netherlands) was used.

Data were retrieved from the monitor with the ixTrend software (ixellence, Wildau, Germany). Signals measured are six ECG derivations (I, II, III, aVR, aVF and a thoracic electrode), impedance pneumography, HR and RespR. Specifications of the monitor can be found in Table 3.3. Descriptions of the heart and respiratory rate derivation from the ECG and impedance signal, respectively, are not disclosed by Philips.

3.2.3. Study populationHealthy subjects

Healthy volunteers were enrolled in the study to test the system technically, address data quality verification and for algorithm development. Subjects older than 18 years and able to follow instruction of the smartphone were asked to voluntarily cooperate in the study. Exclusion criteria were subjects with heart disease, or other severe chronic illnesses, implanted cardiac devices, an allergy to silicone of hydrocolloid adhesives or damaged or very vulnerable skin at the patch location.

After written informed consent was obtained, the biosensor was applied to the chest of the subject. Hereafter the reference monitor was applied to the subject, followed by a series of exercises, including breathing exercises and posture changes. The breathing exercises involved spontaneous and metronome breathing at 10, 15, 20 and 25 breaths per minute in random order, as well as one long and two short breath stops of 15 and 8 seconds, respectively. Next, the subject performed position tasks, involving 2 minute blocks of sitting in bed, sitting, and standing, all separated by 2 minutes of lying down. Hereafter the reference was detached and walking exercises were performed containing blocks of two-minute walking at 50, 75 and 100 steps per minute. Walking pace was set by audible ques from a metronome. Blocks of walking were separated by one-minute blocks of sitting. Furthermore, physical functioning and frailty was examined with grip strength, distance travelled in 6 minutes (6MWT) and the short physical performance battery (SPPB). The SPPB is a simple test examining, balance, walking speed and repeated chair-stands and is a measure of self-resilience. Additionally, 12 leads ECG was recorded and Edmonton frail scale and Short Form (36) Healthy Survey (SF36) were filled in. An overview of the protocol used for healthy subjects can be found in Figure 3.3.

After these measurements, subjects resumed daily activities, while the wearable biosensor was still attached for the remaining battery life. Subjects were asked to keep a diary of their activities, and specifically note sleeping, sitting, walking, stair climbing, sports, transport and when they would feel palpitations, become unwell or fell. When the measurement was finished, subjects filled in the post-study system usability questionnaire (PSSUQ)[31] and a custom-made questionnaire to assess the wearability of the wearable biosensor and its system (Appendix A.3.1).

Page 32: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

14

Chapter 3

4 da

ys

~ 30

min

utes Breathing

exercises

Posture exercises

Walkingexercises

Frailty assessments

Home

Bios

enso

r

Refe

renc

e

6MWTGrip strengthSPPB

SF-36Edmonton frail scale

Sitting Walking (50, 75, 100 steps/min)

Stair climbing (50, 100 stairs/min)

Keep diary noting: -Sleeping -Sitting -Walking -Falls -Sports -Other daily activities

Rest

Metronome breathing (10, 15, 20, 25 breaths/min)

Breathing stops (15, 8 and 8 seconds)

Lying prone (left and right)

Lying on backSitting

Lying on back Standing

2 min -

8 min -

4 min -

1 min -

2 min -2 min -

2 min -2 min -

1 min -2 min-

2 min -

2 times

2 times

Figure 3.3 Protocol overview of healthy subjects, including breathing exercises, posture test, exercises, frailty assessment.6MWT: six-minute walk test; SF36: Short Form (36) Health Survey; SPPB: short physical performance battery

Page 33: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 3

15

Hom

e

± 4

wee

ksScreening

T0: Inclusion and baseline measurement

Frailty assessment

6MWTGrip strengthSPPBSF-36Edmonton frail scale

4 da

ys

TAVI

T1:Direct post-TAVI

Telemetric events

8 da

ys

2 da

ys

± 6

wee

ks

Checkup

T2: Follow-up

4 da

ysBi

osen

sor

Refe

renc

e

CCU

Hom

e

Frailty assessment

6MWTGrip strengthSPPBSF-36Edmonton frail scale

War

d

Figure 3.4 Overview of TAVI patients’ protocol. T0: Inclusion and baseline measurement; T1: Direct post-TAVI; T2: Follow-up. At the end of all measurements the PSSUQ and custom-made wearability questionnaire was filled in 6MWT: six minute walk test; SF36: Short Form (36) Health Survey; SPPB: short physical performance battery; TAVI: transcatheer aortic valve implantation

Page 34: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

16

Chapter 3

Table 3.4 Inclusion and exclusion criteria for the TELE-TAVI study for healthy subjects and TAVI patients.

Inclusion criteria

All subjects Healthy subjects TAVI patients

Older than 18 years old

Able to follow instructions of smartphone and measurement set-up

Able to provide written consent

In work-up for TAVI procedure

Independent at home or helped by an informal care giver

Exclusion criteria

All subjects Healthy subjects TAVI patients

Subjects with implanted devices, such as a pacemaker or implantable cardioverter-defibrillator

Subjects with a damaged or very vulnerable skin around the patch location

Subjects known with allergy to silicone or hydrocolloid adhesives

Subjects with a cognitive impairment or inability to understand and follow-up instructions from the researcher and smartphone

Subjects with known heart disease

Subjects with severe chronic illnesses Unlikely to get transfemoral TAVI, due to anatomical variations

Table 3.5 Overview TELE-TAVI study parameters

T0 T1 T2

Pre-TAVI Direct post-TAVI 6 weeks after TAVI

Informed consent x

Demographics x

Medical history x

Risk scores x

Biosensor monitoring 4 days 8 days 4 days

High-end monitoring x

Frailty x x

- Grip strength x x

- SPPB x x

- 6MWT x x

- Edmonton frailty scale x x

SF-36 x x

Wearability questionnaire x x x

- PSSUQ x x x

- Custom made x x x

Page 35: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 3

17

TAVI patientsEligible patients registered for TAVI workup were asked by phone if they would participate in the study.

Patients who were considered for the study were independent at home or helped by an informal caregiver and able to follow instructions of the smartphone. Exclusion criteria were patients with implanted devices, such as pacemakers and implantable cardioverter, damaged or vulnerable skin around the patch location, allergy to silicone or hydrocolloid adhesive materials were excluded from the study. All in- and exclusion criteria are summarized in Table 3.4. Patients’ enrolment was on the day of the TAVI workup diagnostic computed tomography, after written informed consent was obtained.

The biosensor patch was applied at three different time-points; pre-operative (T0), direct post-TAVI after the procedure (T1) and at follow-up (T2) (Figure 3.4):

T0: The pre-operative measurement started the day the informed consent of the patient was obtained. After application of the biosensor, physical functioning and frailty was examined with grip strength, 6MWT and the SPPB. Hereafter, patients resumed their daily activities and wore the biosensor until the battery ran out or biosensor fell off with a maximum of four days.

T0: Direct post-procedural measurements started when patients were admitted to the cardiac care unit after the TAVI-procedure. First, the Philips biosensor was attached to the patient and the reference monitor was connected for about two hours by the researchers. Patients received an additional biosensor and were asked to replace the biosensor themselves or with help from an informal caregiver.

T0: Four to twelve weeks after TAVI a biosensor was applied by one of the researchers. Physical functioning and frailty was again examined with the grip strength, 6MWT and the SPPB.

At every measurement time-point, patients received a small kit containing supplies for the study; a phone charger, an instruction booklet, a return envelope, and post-measurement wearability questionnaires (PSSUQ and custom made). At T1, the kit contained an additional replacement biosensor and an alcohol wipe. After each measurement the phone, charger, (un)used biosensor(s) and questionnaires were mailed back to the researchers. An overview of all the TAVI patients study parameters can be found in Table 3.5.

3.3. DiscussionThe TELE-TAVI study is designed to assess the feasibility and usability of the Philips biosensor in TAVI

patients. TAVI workflow could be improved by remote patient monitoring with a patch, by enhancing screening, post-TAVI monitoring and TAVI follow-up. Analysis should provide whether the wearable biosensor can be used for objective frailty outcomes, clinical symptoms of AoS, activity and cardiac monitoring.

To this day, no clinical study reviewed the usability of sensor technology in TAVI patients. However, the added value of a wearable system for post-TAVI monitoring is described by, Hermans et al.[32]. Recently a study was started to implement remote patient monitoring post-TAVI, as addition to standard care[33]. Still, the TELE-TAVI study is the first study in TAVI patients to address screening and follow-up improvements with sensor technology. Next to that, the TELE-TAVI study likely provide useful feedback for further development of the Philips wearable biosensor.

The TELE-TAVI study is a pilot to address the feasibility of using the wearable biosensor in a larger study population. Statistical power of this study will consequently be small and most of the results will be used to form hypotheses. After careful evaluation of the TELE-TAVI studies results, an additiaonal study will be needed to prove the clinical value of remote patient monitoring.

Page 36: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

18

Page 37: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

19

4.1. IntroductionTranscatheter aortic valve implantation (TAVI) is an established intervention for intermediate to high

mortality risk patients with severe aortic valve stenosis [25]. Currently, routinely used screening scores are not sufficient for screening TAVI patients, as frailty is not included [35], [36]. Next to that, nowadays, telemetry after TAVI confides the patient to the hospital. Furthermore only limited support is available after discharge from the hospital. Sensor technology could improve TAVI workflow in screening, monitoring and follow-up of TAVI patients.

The TELE-TAVI study is started to evaluate the usability and feasibility of a wearable sensor for TAVI workflow: screening, monitoring and follow-up of TAVI patients (Chapter 3). This chapter will present the first results of the TELE-TAVI study.

4.2. MethodThe TELE-TAVI study is prospective investigator initiated study, in the Amsterdam UMC (location AMC,

Amsterdam, The Netherlands). The study, included healthy volunteers and patients in work-up for TAVI. The local ethics committee evaluated and approved the study design to national and international standards. The design and rationale of the TELE-TAVI is elaborately described in Chapter 3. Here a brief summary of the study will be given.

4.2.1. DevicesThe Philips wearable biosensor (Philips Medical System, Andover, Massachusetts, USA) with the

Research Kit (Philips Medical System, Andover, Massachusetts, USA) was used for remote patient monitoring. The biosensor is a lightweight, medical degree, patch that can measure vital signs and posture for up to four days. The Research Kit, consist of a mobile phone (Kyocera BRIGADIER), pre-installed with a dedicated application. Reliability of the biosensors vital signs was compared to a standard care monitor of Philips (Philips MP70 monitor, Philips Healthcare, Eindhoven, the Netherlands.

4.2.2. Study populationHealthy subjects

Healthy volunteers were enrolled in the study to test the system technically and to gain experience with the system. Subjects older than 18 years and able to follow instruction of the smartphone were asked to voluntarily cooperate in the study. Exclusion criteria were subjects with heart disease, or other severe chronic illnesses, implanted cardiac devices, an allergy to silicone of hydrocolloid adhesives or damaged or very vulnerable skin at the patch location.

After written consent was obtained, a wearable biosensor was adhered to the chest. Hereafter, breathing, posture and walking exercises were performed. Also physical functioning was tested with the 6 minute walk test (6MWT), short physical performance battery (SPPB) and grip strength. Additionally, 12 leads ECG was recorded and Edmonton frail scale and Short Form (36) Healthy Survey (SF36) were conducted.

Chapter 4: Patient characteristics before and after transcatheter aortic valve implantation: First results of the TELE-TAVI study.

Page 38: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

20

Chapter 4

TAVI patientsEligible patients registered for TAVI workup were asked by phone if they would participate in the

study. Patients who were considered for the study were independent at home or helped by an informal caregiver and able to follow instructions of the smartphone. Exclusion criteria were implanted devices, such as pacemakers and implantable cardioverter-defibrillator, damaged or vulnerable skin around the patch location, allergy to silicone or hydrocolloid adhesive. Patients’ enrolment was on the day of the TAVI workup diagnostic computed tomography, after written informed consent was obtained.

Table 4.1 Drop-out reasons for ten TAVI-patients

Subject number Reason Timing drop out

206 Difficulties with measurements (sensor detachment and problems charging phone), resulting that participation was too much effort After T1

208 Excluded by researchers as patient detached sensor prematurely after TAVI During T1

210 Experienced to many, disturbing audible warnings by the phone After T0

211 Not willing to participate after deterioration post-TAVI After T1

212 Burden of study too high in combination with transaortic TAVI After T0

214 Impact of wearing phone too high During T0

219 Pacemaker implantation During T1

221 Pacemaker implantation During T1

223 No symptoms AoS and therefore no TAVI After T0

224 No symptoms AoS and therefore no TAVI After T0

TAVI indicates transcatheter aortic valve implantation

Figure 4.1 . Study flow diagram of the patients screened for the study. One patient is waiting for a TAVI procedure and five patients are waiting for a follow-up appointment

Patients screened for studyn = 90

Excluded: n = 24Declined to participate: n = 42

T0: pre-TAVI

Included n = 24

Drop-out: 3Excluded: 2 - No TAVI, because no AoS symptomsWaiting on T1: 1T1: direct post-TAVI

n = 18

T2: 6 weeks follow-up

n = 8

Drop-out: 3Excluded: 2 - Pacemaker implantationWaiting on T2: 5

Page 39: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 4

21

The biosensor patch was applied at three different moments; pre-operative (T0), direct post-TAVI after the procedure (T1) and at follow-up (T2), as shown in Figure 3.4. Physical functioning was assessed at T0 and T2 with the 6MWT, SPPB, grip strength, SF36 and Edmonton frail scale. At every measurement moment, patients received a small kit containing supplies for the study, as well as user-experience questionnaires (PSSUQ and custom made). After each measurement the phone, charger, used biosensor(s) and filled in questionnaires were mailed back.

4.2.3. Statistical analysisBaseline characteristics of the included TAVI-patients is compared with all screened patients during

the inclusion period (June 2018 and February 2019). Grip strength is separated into three groups, stronger, average and weaker, correcting for age and gender following Dodds et al. [37]. Difference in the mean between the groups are tested with a student t-test for continuous variables. To compare the difference in the mean in the proportions of medical history a z-test is used. Characteristics of TAVI-patients who completed measurement T0 and T2 are compared for significant differences and continuous variables are analyzed with the paired t-test. Categorical variables (CCS, NYHA) are tested with Chi-squared test. For the TELE-TAVI study a significance level of 0.05 is chosen.

Table 4.2 Averaged result of the biosensors precision and bias for HR and RespR, compared to unfiltered and filtered reference monitor, in the healthy subjects and TAVI population with and without atrial fibrillation.

Characteristics TELE-TAVI All TAVI-patients

n = 24 n = 176 p-value

Age 76.6 (± 4.8) 78.8 (± 7.5) 0.15

Male 18 (75 %) 87 (50 %) 0.02*

BMI 28.0 (± 5.6) 27.4 (± 5.0) 0.57

Medical history

- Hypertension 16 (67 %) 99 (56 %) 0.33

- Atrial fibrillation 6 (25 %) 51 (29 %) 0.98

- COPD 1 (4 %) 13 (7 %) <0.001*

- Diabetes 6 (25 %) 50 (28 %) 0.73

NYHA: <0.001*

- I 2 (10 %) 18 (10 %)

- II 12 (60 %) 42 (24 %)

- III 6 (30 %) 106 (60 %)

- IV 0 (0 %) 10 (6 %)

CCS: <0.001*

- No angina 17 (77 %) 2 (1 %)

- Grade I-II 4 (18 %) 119 (68 %)

- Grade III 0 (0 %) 21 (12 %)

- Grade IV 1 (5 %) 24 (14 %)

METS 6.3 (± 1.4) 5.6 (± 1.5) 0.03*

Logistic EuroSCORE I 8.8 (± 7.2) 13.7 (± 9.3) 0.01*

EuroSCORE II 1.8 (± 1.7) 4.2 (± 3.3) 0.001*

STS (mortality score) 2.0 (± 1.2) 4.2 (± 2.6) <0.001*

BMI indicated body mass index; NYHA, New York Heart Association; CCS, Canadian Cardiovascular Society grading of angina pectoris; COPD, chronic obsturctive pulmonary disease; EuroSCORE European System for Cardiac Operative Risk Evaluation;METS, Metabolic equivalent score; STS, Society of Thoracic Surgery predicted risk of mortality; * significant difference

Page 40: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

22

Chapter 4

4.3. Results4.3.1. Healthy subjects

During the period of June to September 2018, six healthy subjects volunteered to participate in the study and all completed the study protocol. The subjects are averaged 41 (29-35) years old, have a BMI of 26 (22-30) and 50% is male. In one subject (103), an error of the Research Kits mobile app occurred; the mobile phone was unable to reconnect to the biosensor. As solution, a new phone was given. In three subjects (102, 103 and 106), the patch detached within three days, in which two subjects (103 and 106) attached a new sensor.

4.3.2. TAVI patientsAt February 22nd, 2019, 90 patients in work-up

for TAVI were screened of which 24 were included (Figure 4.1). Ten subjects dropped out of the study, for varying reasons, which can be found in Table 4.1. In total, 50 measurements are made, of which two data set are lost (206-T0 and T1), due to accidental deletion of the files and loss of biosensors ECG. In 18 patients, a measurement with the reference is available, in which for one patient (208) data were not recorded for two leads and excluded from further analysis. In 2 other patient (213 and 217), lead aVL was recorded instead of aVF. An overview of the retrieved data can be found in Appendix A.1.

Table 4.2 summarizes the patient characteristics at baseline (T0), compared to the all patients screened for TAVI during the inclusion period. The study population is 76.6 (± 4.8) years old, the screened TAVI-population 78.8 (± 7.5). Patients included are predominantly male (77%), which differs significantly from the general TAVI-population in which half is male. Included TAVI patients significantly have lower mortality risk scores, compared to the general population. Patients have higher METS scores, compared to overall TAVI-screening group. Other baseline characteristics are summarized in Table 4.3 The SPPB score show that 46% of the population is not frail and 50% is pre-fail and 4% is frail. The average distance walked in 6 minutes is 409 (± 89) meter. The grip strength is reduced in 8% of the TAVI patients and overall SF36 result is 62%.

Table 4.4 shows the study parameters for patients who completed both the screening (T0) and follow-up (T2) measurement. The grip strength is significantly reduced after the TAVI, from 38 (± 9) to 34 (± 11) kg. Significant increase is found in the SF36 health change domain, which compares the current perceived health with a year before.

Table 4.3 Baseline characteristics, SF36 and frailty measures, for included TELE-TAVI patients.

Characteristics TELE-TAVI

n = 24

Quality of life (SF36)

- Overall 62% (± 14%)

- Physical functioning 61% (± 24%)

- Social functioning 78% (± 23%)

- Limitation of function 56% (± 43%)

- Limitation of emotions 65% (± 44%)

- Vitality 73% (± 19%)

- Mental health 57% (± 18%)

- Pain 80% (± 19%)

- General health 54% (± 18%)

- Health change 33% (± 20%)

6MWT (distance, m) 409 (± 89 m)

Grip strength 34.0 (± 89.1)

- Male average (kg) 37.7 (± 7.6)

- Female average (kg) 22.7 (± 3.8)

- Stronger+ 5 (21 %)

- Average+ 17 (71 %)

- Weaker+ 2 (8 %)

Edmonton frail scale 3.1 (± 2.4)

- Not frail (0 - 5) 15 (79 %)

- Vulnerable (6 - 7) 4 (21 %)

- Mild frail (8 - 9) 0 (0 %)

- Moderate frail (6 - 7) 0 (0 %)

- Severe frail (12 - 17) 0 (0 %)

SPPB 9 (± 2.7)

- Not frail ( > 9) 11 (46 %)

- Pre-frail (4 - 9) 12 (50 %)

- Frail ( < 4) 1 (4 %)

6MWT indicates six minute walk test; kg, kilogram; SF36, Short Form (36) Health survey; SPPB, Short Physical Performance Battery. + Classified by Dodds et al. [37]

Page 41: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 4

23

Table 4.4 Pre- and post-study variations in patients included in TELE-TAVI that completed follow-up (T2).

Characteristics T0 T2

n = 8 n = 8 p-value

NYHA: a

- I 2 (10 %) 2 (33 %)

- II 12 (60 %) 3 (15 %)

- III 6 (30 %) 1 (17 %)

- IV 0 (0 %) 0 (0 %)

CCS: a

- No angina 17 (77 %) 7 (100 %)

- Grade I-II 4 (18 %) 0 (0 %)

- Grade III 0 (0 %) 0 (0 %)

- Grade IV 1 (5 %) 0 (0 %)

Quality of life (SF36)

- Overall 62 % (± 14 %) 61 % (± 12 %) 0.56

- Physical functioning 70 % (± 20 %) 63 % (± 33 %) 0.77

- Social functioning 70 % (± 22 %) 67 % (± 30 %) 0.68

- Limitation of function 50 % (± 50 %) 38 % (± 46 %) 0.70

- Limitation of emotions 52 % (± 50 %) 79 % (± 40 %) 0.33

- Vitality 53 % (± 14 %) 57 % (± 26 %) 0.39

- Mental health 66 % (± 21 %) 73 % (± 26 %) 0.57

- Pain 76 % (± 14 %) 67 % (± 30 %) 0.50

- General health 48 % (± 16 %) 53 % (± 23 %) 0.34

- Health change 28 % (± 16 %) 56 % (± 32 %) 0.04*

6MWT (distance, m) 391 (± 108) 445 84 0.13

Grip strength 38 (± 9) 34 (± 11) 0.02*

- Male average (kg) 41 (± 6) 38 (± 10) 0.09

- Female average (kg) 27 (± 1) 22 (± 3) 0.32

- Stronger 3 (37 %) 3 (37 %)

- Average 5 (63 %) 3 (37 %)

- Weaker 0 (0 %) 2 (26 %)

Edmonton frail scale 2.5 (± 2) 3.5 (± 2.7) 0.39

- Not frail (0 - 5) 5 (83 %) 5 (83 %)

- Vulnerable (6 - 7) 1 (17 %) 0 (0 %)

- Mild frail (8 - 9) 0 (0 %) 1 (17 %)

- Moderate frail (6 - 7) 0 (0 %) 0 (0 %)

- Severe frail (12 - 17) 0 (0 %) 0 (0 %)

SPPB 10 (± 2) 10 (± 1) 0.70

- Not frail ( > 9) 4 (57 %) 5 (71 %)

- Pre-frail (4 - 9) 3 (43 %) 2 (29 %)

- Frail ( < 4) 0 (0 %) 0 (0 %)

6MWT indicates six minute walk test; CCS, Canadian Cardiovascular Society grading of angina pectoris; NYHA in-dicates New York Heart Association; SF36, Short Form (36) Health survey; SPPB, Short Physical Performance Battery. + Classified by Dodds et al. [37] * Siginificant change a No p-value unavailable as one category has zero patients in both T0 and T2 (NYHA IV, CCS Grade III)

Page 42: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

24

Chapter 4

4.4. DiscussionThe TELE-TAVI study is designed to evaluate the Philips wearable biosensors usability and feasibility in

TAVI patients, pre- and post-TAVI as well as at follow-up. It is the first non-commercial study of the Philips wearable biosensor. This chapter presents the first results of the TELE-TAVI study.

The TELE-TAVI is a small pilot study, but less patients agreed to participate than expected. The study was designed to minimize the participation burden, as additional hospital visits are not needed. Next to that, the wearable biosensor is considered non-invasive. However, low inclusion rates are likely caused as patients were not experienced with mobile phone use and and find handling a phone a daunting taks.

Overall, the population in this study is younger, includes more men, which have less comorbidities, a better functional status and lower risk scores, compared to other reports (81-83 years old; 32.5-46.6% men; 67.6-81.6% hypertension, 34.6-41.2% atrial fibrillation, 24.6% COPD; STS-score of 4.0-6.1 and EUROSCORE II of 3.7-5.1) [38], [39]. When compared to all patients that are screened for TAVI, included patients have lower mortality risk and better functional capacity, but age is not significantly different. The reported lower age is likely caused by a shift to younger TAVI patients, as reported by Kesteren et al. [38].

Less data is available to compare the health surveys (SF36 and Edmonton frail scale), functional status (SPPB) and physical function (6MWT and grip strength), to other studies with aortic valve stenosis patients. However, patients in this report performed much better on all domains (SF36 score of 26.1-43.3% [40], 24% frail TAVI patients by the Edmonton frail scale [41]; SPPB score of 8.2 +- 3.2 [42] and 6MWT distance of 72-240 meter [40]). Notable is that the patient population in all reports are older, which could account for some of the differences. However, it seems that only patients with a better health status were willing to participate in the study. This is cause for careful interpretation of the study results, as the study population is not representative for the general TAVI population.

Surprising result was the significant decrease of grip strength post-TAVI. Grip strength is indicator for muscle mass in elderly. The decrease in grip strength, could be attributed that patients were not yet fully recovered from the procedure. Next to that, a siginifant improvement was found in the health change domain of the SF36, indicating that patients are feeling better than compared to a year ago. However, there was no significant change found in the general health domain and therefor this should be carefully interpreted.

Although, the population does not represent a real-world TAVI population, much can be learned from the TELE-TAVI study. Biosensors parameters are collected at screening, monitoring and at follow-up of TAVI patients and will show potential possibilities, improvements and limitations of the Philips wearable biosensor in TAVI workflow

Page 43: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

25

Page 44: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

26

Page 45: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

27

5.1. IntroductionWearable wireless sensing is an upcoming technology for measuring vital signs continuously and

unobtrusively. They can change the field of patient monitoring, screening and follow-up, at home and in-hospital. It possibly allows for sooner hospital discharge, earlier deterioration detection and can play a role in personalized health. However, implementing this technology in every day healthcare remains a challenge as knowledge of the reliability, involving the accuracy of measured data and consistency of continuous data collection, of these systems is limited [43].

Only few studies review the reliability of wearable sensors [43]. Not a single study is published on the reliability of the biosensor (Philips Medical System, Andover, Massachusetts, USA) and only few on its predecessor, the HealthPatch MD (VitalConnect, San Jose, California, USA), which has the same firm- and hardware [21]–[23].

Only one independent in-hospital study is published, reviewing the HealthPatch reliability of the measured heart and respiratory rate (HR, RespR, respectively), in 25 patient on a surgical step down unit [21]. They showed the HealthPatch can accurately measure HR (within 10% error), but the RespR was not reliable compared to bedside impedance pneumography. With a 15 minute median filter, the accuracy of the RepsR improved. Similar results were found by an in-house research [44]. Data were available in 94% of the time and in halve of the patients there was no data interruption.

Validation research of HR, RespR and activity measured by the HealthPatch was done by VitalConnect in 25 healthy people [22]. Accurate HR was measured in rest and during activities of daily living. Also RespR was accurate during metronome breathing, compared to a capnograph. Yet it must be noted only a third of the RespR data was compared and only at the places the breathing rate was most constant. The error in RespR increased in spontaneous breathing and during activities of daily living. Nonetheless, it showed that the combined respiratory rate estimation with ECG and acceleration (as described in Chapter 3) was far superior than one modality estimation [23].

On the accuracy of the posture and activity of HealthPatch only little data is published. Research from VitalConnect showed that posture detection was 80.1% accurate in an test environment [22], but details on the study protocol are missing. Additionally, step count analysis of the Health Patch was similar to manually counted steps, but appeared sensitive to ‘step-like’ moves, for example cycling. Falls were detected in 92.5% in simulated falls and no false positive falls were detected during other simulated daily activities.

In short, research on the reliability of the wearable biosensor is limited and focuses more on vital signs, such as HR and RespR. It appears that HR is measured accurately by the HealthPatch, but RespR validation showed varying results. Very limited research is reported on the validity of the activity or posture detection and needs more attention. This chapter validates the accuracy of the HR and RespR, the reliability of posture detection and continuous data collection of the wearable biosensor.

Chapter 5: Reliability of the Philips wearable biosensor in monitoring vital signs and activity

Page 46: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

28

Chapter 5

5.2. MethodReliability assessment of the Philips wearable biosensor is part of the TELE-TAVI study, which reviews

telemedicine in transcatheter aortic valve implantation intervention workflow. In the TELE-TAVI study 6 healthy volunteers and 24 TAVI-patients participate. All patients provided written informed consent. The protocol is extensively described in Chapter 3 and a relevant summary will be given here.

TAVI patients received the wearable biosensor thrice; before the TAVI procedure (T0), directly after the TAVI procedure on the cardiac care unit (CCU) (T1) and 6-weeks after the TAVI procedure (T2). At T0 and T2 patients performed the functional test (six minute walk test (6MWT), short physical performance battery and grip strength), while they were wearing the sensor. After the procedure (T1), patients were connected to a standard care monitor (Philips MP70 monitor, Philips Healthcare, Eindhoven, the Netherlands) as reference for the biosensor.

After signed written informed consent, the Philips wearable biosensor was attached to the healthy subjects. Hereafter, patients completed a protocol including breathing, posture and walking exercises. During the breathing and posture exercises, the reference monitor was connected. Next to that, functional test (6MWT, SPPB and grip strength) were performed. Hereafter, healthy subject resumed daily activities and kept a diary on their activities.

Figure 5.1 Flowchart of vital signs datasets used for analysis of biosensors HR and RespR precision.

Healthy subjectsn = 6

HR analysis RespR analysis

TAVI patients totaln = 18

Total vital signsn = 24

Vital signs datasetsn = 19

Excluded: - Decryption error biosensor: n = 4 - Data lost in reference monitor: n = 1

Healthy subjects: n = 6TAVI patietns: n = 8 - Sinus rhythm: n = 6 - AF: n = 2

Excluded: - Interference reference signal: n = 5

Healthy subjects: n = 6TAVI patietns: n = 13 - Sinus rhythm: n = 7 - AF: n = 6

Figure 5.2 -Next page- Three examples, showing the HR and RespR of the biosensor and reference, not filtered and filtered. A: shows data of a healthy subject during the breathing exercises and posture changes, RespR is only shown during breathing exercises. B; gives the HR and RespR of TAVI patient directly post-TAVI (T1). Hereafter the RespR has a larger variance for reasons unknown. C: Shows HR and RespR data of TAVI-patient known with AF.

Page 47: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 5

29

A: Healthy subject - 101

60

80

100

120

HR

14:52 15:00 15:07 15:14 15:21 15:28 15:360

10

20

30

Resp

R

50

60

70

HR

14:38 14:52 15:07 15:21 15:36 15:50 16:04 16:19 16:33 16:48 17:020

10

20

30

Resp

R

80

100

120

HR

11:02 11:16 11:31 11:45 12:00 12:14 12:28 12:43 12:57 13:12 13:260

10

20

30

Resp

R

B: TAVI patient with sinus rhythm - 213

C: TAVI patient with AF - 219

BiosensorReferenceReference �ltered

Page 48: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

30

Chapter 5

5.2.1. Vital signsTo assess the reliability of the biosensors

vital signs, HR and RespR from the biosensor was compared to the reference monitor. In healthy subjects the data from breathing exercises and posture test was used. Breathing exercises contains breathing normal , breathing at the rate of metronome and breathing stops, whilst sitting. During posture exercises, healthy volunteers were lying, sitting or standing for 2 minutes, which induced orthostatic heart rate changes. For the analysis of the RespR, only the breathing exercises were applied, as the reference experienced too much perturbations from the posture changes. From TAVI-patients, data of the direct post-TAVI measurement (T1) is used, during the time they were connected with the reference monitor. TAVI patients were subdivided into two groups, sinus rhythm or atrial fibrillation (AF), by reviewing telemetry results at the time the reference monitor was connected.

Signal analysisData from the biosensor was decrypted using the Research Kit (Philips Medical System, Andover,

Massachusetts, USA) and were retrieved in CSV format. From the reference monitor, data were saved in CSV format with ixTrend (ixellence GmbH, Wildau, Germany). All data were stored and processed using Matlab (The MathWorks, Natick, Massachusetts, USA). Missing data were interpolated with empty values (not-a-number).

Data of the two modalities were synchronized with a perturbation in the data induced by the researcher. Hereafter outliers bigger than 100 times the average were removed from the biosensors data. For the reference data outliers bigger than 10 or 5 times were removed in the HR and RespR, respectively.

The biosensor and reference monitor have different filter settings (Chapter 3), which will induce an error in the reliability analysis. As the biosensor has more extensive filtering, the data from the reference monitor was filtered more to mimic the biosensors filter settings. The biosensor HR has a 10-beat low-pass filter [22], which cannot be reproduced in the obtained data, since the reference HR is sampled at a regular frequency. To mimic this filter, first the HR of the reference monitor was averaged over the full recording. With this averaged HR, a window length was calculated that on average should contain 10 heart beats. The calculated averaging window was used in a moving average filter over the reference data. If more than 60% of the data were available, the HR was averaged over the filtered window, otherwise the sample was replaced by an empty value. Hereafter, the data of the reference was interpolated to the biosensors timestamp and empty values were discarded.

The biosensors RespR is averaged over a 45 second window [23]. This was imitated by computing the RespR for every biosensors sample with the previous 45 seconds. When 20% of the data were missing, the RespR was not computed but replaced with an empty value and discarded hereafter. The unfiltered reference RespR was interpolated to the biosensors timestamp.

The primary outcome was bias and precision of HR and RespR of the biosensor compared to the high-end monitor. Precision is computed as the reproducibility coefficient (RPC; 1.96*SD or 1.45*IQR) and limit of agreement (LOA), in which LOA is the bias ± RPC. This reference standard reports an accuracy for HR of ±1% beats/min and ±1% breaths/min for RespR [34]. HR and RespR were considered to be acceptable for clinical purposes if within ±10% or ±3 breaths/min of the reference monitor, respectively.

Distribution of the difference data (biosensor – reference) was checked with skewness and kurtosis, to determine whether parametric statistical analysis could be used. Skewness is a measure of the asymmetry of the data around the sample mean. If skewness is negative, the data are spread out more to the left of the mean. The skewness of the normal distribution (or any perfectly symmetric distribution) is zero. Kurtosis is a measure of how outlier-prone a distribution is and for a normal distribution equals 3. Distributions that are more outlier-prone than the normal distribution have kurtosis greater than 3; distributions that are less outlier-prone have kurtosis less than 3.

Table 5.1 Skewness and kurtosis (± SD) of difference of the biosensors and reference filtered and unfiltered data.

Skewness Kurtosis

HR 1.4 (± 4.6) 10.9 (± 80.1)

HR filtered 1.9 (± 5.3) 48.9 (± 101.0)

RespR -0.5 (± 3.3) 5.1 (± 3.3)

RespR filtered -0.1 (± 2.0) 4.0 (± 2.0)

HR indicates heart rate; RespR, respiratory rate

Page 49: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 5

31

Table 5.2 Averaged result of the biosensors precision and bias for HR and RespR, compared to unfiltered and filtered reference monitor, in the healthy subjects and TAVI population with and without atrial fibrillation.

n Bias RPC Lower LOA Upper LOA

Beats/min (%) Beats/min Beats/min Beats/min

HR

Healthy 6 0.5 ± 1.0 3.1 ± 2.9 4.4 ± 4.5 -2.6 ± 1.6 3.5 ± 2.5

TAVI 13 0.0 ± 3.2 2.9 ± 10.7 3.7 ± 9.3 -2.9 ± 7.2 2.9 ± 12.8

- Sinus 7 0.0 ± 0.0 1.5 ± 1.2 2.6 ± 1.9 -1.5 ± 1.2 1.5 ± 1.2

- AF 6 3.3 ± 3.0 12.2 ± 6.7 12.6 ± 6.1 -8.8 ±7.5 15.5 ± 9.2

All 19 1.5 ± 3.3 5.4 ± 5.4 6.5 ± 5.9 -3.9 ± 4.5 7.0 ± 7.7

HR fi

ltere

d

Healthy 6 0.5 ± 0.5 2.4 ± 2.4 3.5 ± 3.8 -1.9 1.3 2.9 ± 3.6

TAVI 13 0.2 ± 3.1 1.8 ± 8.9 2.4 ± 7.9 -1.6 ± 5.8 2.0 ± 12.1

- Sinus 7 0.0 ± 0.1 1.0 ± 0.4 1.7 ± 0.7 -0.9 ± 0.5 1.0 ± 0.4

- AF 6 3.2 ± 3.5 10.4 ± 6.0 10.7 ± 5.1 -7.2 ± 7.5 13.6 ± 8.2

All 19 1.5 ± 3.2 4.7 ± 4.5 5.7 ± 5.0 -3.1 ± 4.0 6.2 ± 6.8

Breaths/min (%) Breaths/min Breaths/min Breaths/min

Resp

R

Healthy 6 -1.5 ± 0.9 8.1 ± 1.5 50.6 ± 10.6 -9.6 ± 2.1 6.6 ± 1.4

TAVI 8 0.8 ± 1.6 7.9 ± 1.5 57.3 19.2 -7.0 ± 1.4 8.7 ± 2.8

- Sinus 6 0.0 ± 0.8 7.4 ± 1.5 54.5 ± 21.0 -7.4 ± 1.4 7.4 ± 2.0

- AF 2 3.2 ± 0.5 9.1 ± 0.2 65.6 ± 14.1 -5.9 ± 0.6 12.3 ± 0.3

All 14 -0.2 ± 1.8 7.9 ± 1.5 54.4 ± 15.9 -8.1 ± 2.1 7.8 ± 2.5

Resp

R fil

tere

d

Healthy 6 -1.5 ± 0.9 6.2 ± 2.4 40.1 ± 14.3 -7.7 ± 2.8 4.6 ± 2.2

TAVI 8 0.8 ± 1.6 6.4 ± 1.3 43.9 ± 12.5 -5.7 ± 1.2 7.2 ± 2.6

- Sinus 6 0.0 ± 0.7 6.0 ± 1.2 41.7 ±13.7 -6.0 ± 1.1 6.0 ± 1.7

- AF 2 3.1 ± 0.4 7.6 ± 0.4 50.7 ± 5.2 -4.5 ± 0.8 10.7 ± 0.0

All 14 -0.2 ± 1.7 6.3 ± 1.7 42.3 ± 12.9 -6.5 ± 2.2 6.1 ± 2.7

RespR indicates respiratory rate; HR, heart rate, SD=Standard deviation; AF,Atrial fibrillation; TAVI,transcatheter aortic valve stenosis patients

Figure 5.3 Shows the averaged Bland-Altman figures comparing the biosensor with the reference, for all the subjects. Left HR is compared with the unfiltered reference HR and right RespR is compared with the filtered reference. The squares represent the averaged bias per patient, in which orange are healthy subjects and dark and light blue are TAVI patients with sinus rhythm and AF, respectively. The small faint dots are individual data points. In the left panel, the HR of the TAVI patients with AF is given.

0 50 100 150HR Reference

-30

-20

-10

0

10

20

30

HR

Bios

enso

r - R

efer

ence

7.0

1.5

-3.9

0 10 20 30RespR Reference �ltered

-30

-20

-10

0

10

20

30

Resp

R Bi

osen

sor -

Ref

eren

ce �

ltere

d

7.8

-0.2

-8.1

Page 50: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

32

Chapter 5

Figure 5.4 Flowchart of 6MWT for walking detection analysis

Healthy subjectsn = 6

T0: TAVI patients n = 24

T2: TAVI patients n = 8

TAVI patients totaln = 32

Total 6MWTn = 38

Analysed 6MWTn = 32

Excluded: - Decryption error biosensor: n = 3 - Wheelchair bound: n = 2 - Standing still (30 sec.): n = 1

For all data, HR and RespR, filtered and unfiltered, the lag between the biosensor and reference data was estimated using cross-correlation.

5.2.2. Posture detectionDetermining the reliability of the posture detection is more difficult as no gold standard is available.

Therefore, posture detection is validated with observation. There are only a few postures detected by the biosensor, as shown in Figure 5.6. Of these postures, only walking was easily reviewed. Percentage of walking detected during the 6MWT was analyzed of healthy subjects and TAVI patients. Also the walking exercises of the healthy subjects were used. Walking should be detected more than 80% of the time. Other detection of posture was compared to the diaries of the healthy subjects. Hereof, multiple examples are selected to show the reliability of the biosensors posture detection.

5.2.3. Data collectionTo examine the data collection reliability, biosensors data of the healthy subjects and the pre-TAVI (T0)

and follow-up (T2) measurement was used. With this, the measurement reliability of one biosensor can be assessed. Measures for data collection reliability are, first point of data loss, total measured time and percentage of data loss. First point of data loss is defined as the first time point of a gap longer than 2 minutes, 15 minutes, 1 hour or 4 hours occurs. For this analysis the electrocardiogram (ECG) was used, as this has the highest sample frequency of the biosensors output, giving it the most accurate representation of the data loss due to connection failure. Total measurement time is the length of the biosensor measurement, which is expected to be 96 hours. Of the measurement time, at least 80% should be usable data, so recorded without any data loss.

Figure 5.5 -Next page- Shows acceleration modulus and activity classification for several activities as noted in the diary, during multiple days of subject 101. Panel A and B displays sitting, at day 1 and 2. Panel C and D, show standing at day 1 and 4. Panel E shows the data of walking and running. Panel F, illustrates that activity detection of cycling alternates between walking and lying

Page 51: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 5

33

0.9

1

1.1

1.2

Acc

. mod

ulus

A: Sitting day 1

14:52 14:53

B: Sitting day 2

12:14

0.9

1

1.1

1.2

Acc

. mod

ulus

C: Standing day 1

21:28 21:29

D: Standing day 4

18:53 18:54

0

1

2

3

Acc

. mod

ulus

E: Walking and running

10:16

F: Cycling day 3

17:17 17:18

12:13Lying

Leaning

Up

Walking

Unknown

Lying

Leaning

Up

Walking

Unknown

Lying

Leaning

Up

Walking

Unknown

Page 52: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

34

Chapter 5

Table 5.3 Averaged results of the walking detection

n Walking

Detected Speed

(%) (m/s)

Heal

thy

100 steps/min 6 99.4 ± 1.5 1.19 ± 0.08

75 steps/min 6 74.8 ± 34.3 0.77 ± 0.09

50 steps/min 6 0.8 ± 1.2 0.43 ± 0.12

6MWT 6 99.8 ± 0.4 1.57 ± 0.13

TAVI

6MWT 25 96.5 ± 6.1 1.15 ± 0.26

6MWT indicates six minute walk tes; SD,standard deviation; km/h,kilometer per hour; TAVI, transcatheter aortic valve stenosis patients; n, number of subject

5.3. Results5.3.1. Vital signs

In total, 24 datasets were available for the vital signs reliability analysis, of which 6 and 18 were datasets from the healthy subjects and TAVI patients, respectively. From the 18 TAVI patient datasets, 5 were not analyzed, as the biosensor data were not available due to decryption problems (4) or due to data loss in the reference (1). In the remaining TAVI-patients, 6 showed atrial fibrillation or flutter (AF) on the telemetry during hospital stay. Of all the TAVI-patients, 5 datasets were excluded in the analysis of the RespR, due to electrical interference of the respiratory impedance signal. Overall, there are 19 datasets for analysis, of which 6 from healthy subjects and 7 with AF. For RespR analysis, 14 datasets were used, containing 6 from healthy subjects and 2 AF TAVI-patient. An overview of the used data can be found in Figure 5.1

Figure 5.2 shows an example of the data used for analysis of the biosensors reliability for a healthy subject and two TAVI patients. In the healthy subject (Figure 5.2 A), shows large fluctuations in the second halve, caused by the orthostatic changes in HR. A large peak is visible in the biosensors HR at 15:21. The biosensors RespR shows lag, compared to the unfiltered reference (dark blue). There is no lag visible between the filtered reference and biosensors RespR. Lag was detected in the RespR of 4 TAVI-patients, which in 3 cases disappeared after filtering. Figure 5.2 B shows HR and RespR data of a TAVI patient with sinus rhythm. HR has a small range, but the biosensor HR are visually very similar to the reference HR. Around 15:50, the reference monitor RespR data is lost and unexpected peak is visible. This is probably caused by a movement artefact. The HR of the TAVI patient with AF (Figure 5.2 C) the HR is higher and variance is The HR variance in the reference is larger than of the biosensor.

Skewness and kurtosis was computed for the difference data (biosensor-reference), shown in Table 5.1. The kurtosis for HR is on average 10.9 (± 80.1) and 48.9 (± 101.0) difference data, for the unfiltered and filtered reference HR. Appendix A.2.1 had more detailed information on the distribution of the difference data.

Table 5.2 shows bias and precision, comparing the biosensor with the reference monitor. The bias in the unfiltered HR was 1.5 and LOA ranged between -3.9 (± 4.5) and 7.0 (± 7.7). The agreement between the biosensor and reference monitor is broader for TAVI patients with AF ([-8.8 ± 7.5 to 15.5 ± 9.2]). Filtering the reference HR did not change the outcome of agreement. The unfiltered RespR resulted in bias -0.2 and the LOA ranged from 8.1 (± 2.1) to 7.8 (± 2.5). Filtering the RespR narrowed the LOA to the range of -6.5 (± 2.2) and 6.1 (± 2.7). Figure 5.3 shows the averaged Bland-Altman figures for the unfiltered HR and filtered RespR. The bias and dispersion of individual data points is larger in AF subjects, than healthy subjects. The difference in biosensor and reference RespR disperses for all groups similarly. Appendix A.1.2 gives more detailed data on the precision and bias of the wearable biosensor.

Page 53: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 5

35

5.3.2. Posture detectionIn total, all 6 healthy subjects completed the walking protocol and 6MWT. Of the 24 TAVI patients,

8 were seen for a T2 measurement. One TAVI patient was not able to perform the 6MWT at T0 and T2, as she was wheelchair bounded. Of the 36 6MWT, 3 were not analyzed as biosensor data were lost. The reliability analysis assumes participants walk constantly during the 6MWT. One patient stood still for about half a minute during the 6MWT and was therefore excluded for this analysis. In total 32 6MWT were used to analyze the walking detection of the wearable biosensor. An overview of the included 6MWT is visible in Figure 5.4

Walking was detected 99.8% and 96.5% of the time during the 6MWT in healthy subjects and TAVI patients, respectively. Walking detection declined for slower paces in the walking exercises, at a pace of 0.43 m/s, 0.8% (± 1.2%) was detected as walking. Appendix A.1.3, shows all results of the 6MWT and walking exercises.

Figure 5.5 shows multiple examples from the biosensor posture classification from a healthy subject, compared to the diary. Panel A-B and C-D show sitting and standing, respectively. The first day, both are sitting and standing is correctly classified as ‘upright’ (Figure 5.5 A and C), the acceleration looks visually similar. However, a day or few days later both are detected as laying down. Walking and running are both classified as walking (Figure 5.5 E), but walking has a much smaller acceleration variance than running. Cycling is detected as ‘lying down’ or ‘walking’ (Figure 5.5 F).

The biosensors activity classification for every day and night was made (Figure 5.6). The days and nights were separated on basis of the dairy notes. The first day and night show an appropriate portion of ‘upright’ and ‘lying down’, respectively. In the following days however, the biosensor detected mostly ‘lying down’ during the day, which does not coincide with the subjects diary. As well as during the night, more ‘upright’ and ‘leaning back’ is detected, which was not confirmed by the diary.

5.3.3. Data collectionIn total 35 datasets were analyzed to determine the reliability of the data collection (Table 5.4), 12

(34%) measurements had data available for more than 80%. On average, 56.2% of the 96 hours of data is usable and the measurement length was 60:51:58 (± 32:10:07). A total of 13 (37%) had uninterrupted data, but in 22 (63%) data sets, data loss was apparent, ranging from a few minutes to 59 hours. Figure 5.7 shows the survival analysis for ‘time to first data failure’. It shows that 29% had usable data at when eliminating 2 minute gaps. Appendix A.1.4 gives the tables and survival graphs for data collection and first point of data loss.

Figure 5.6 Illustrates the activity classification of the biosensor during days and nights of subject 101. The days and nights were separated on basis of the dairy notes

5%5%

79%

11%

81%

3%2%

9%4%

79%

13%3%4%

74%

12%

5%9%

89%

9%

81%

5%11%

65%

14%

21%

61%

19%

19%

44% 41%

15%

Day 1 Day 2 Day 3 Day 4 Day 5

Night 1 Night 2 Night 3 Night 4

LyingLeaningUpWalkingUnknownNo data

Page 54: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

36

Chapter 5

Table 5.4 Average length of the measurements, data loss and usable data of the wearable biosensor, in the groups’ healthy subjects and TAVI patients.

n Length measurement Data loss Data not lost Usable data Usable data

(hh:mm:ss) (hh:mm:ss) (hh:mm:ss)

Healthy 6 62:39:50 06:45:01 92.5% 55:54:48 58.2%

(± SD) (± 37:29:21) (± 16:14:56) (± 17.5) (± 34:37:36) (± 36.1)

TAVI 29 60:29:39 06:55:36 90.4% 53:34:03 55.8%

(± SD) (± 31:41:52) (± 16:24:38) (± 21.4) (± 32:22:33) (± 33.7)

All 35 60:51:58 06:53:47 90.8% 53:58:11 56.2%

(± SD) (± 32:10:07) (± 16:07:08) (± 20.6) (± 32:15:16) (± 33.6)

n indicates number of subjects; SD, standard deviation; hh:mm:ss, hours:minutes:seconds.

5.4. DiscussionThis chapter investigated the reliability of the Philips wearable biosensor, as part of the TELE-TAVI

study. In this research the reliability of the biosensor vital signs, posture detection and data collection are investigated. In all three domains, there is room for improvement.

Vital signsAssessment of the vital signs accuracy, HR and RespR of the biosensor was compared to the reference

monitor. Preliminary results show that the biosensor can likely accurately measure HR within 10% deviation in people with sinus rhythm. However, in patients with AF this accuracy is not met and filtering HR did not change this. The difference is probably present due to the different filter settings of the biosensor and reference. Additional filtering was not able to adjust for this error. RespR, on the other hand, seems inaccurate in both research groups. Although, filtering the RespR improves the results, as the LOA decreases and lag disappears.

These results are comparable to earlier work with the similar VitalConnect HealthPatch [21]. They hypothesized that poor agreement in the RespR was due to frequent outliers and implausible variability in the RespR values as a result of movement artifacts. This could also be the case in this TAVI-population of this study. However in the healthy subjects, the respiratory signal was mostly undisturbed (see Figure 5.2 A), as patients were sitting during the breathing exercises. However, the biosensors RespR did not coincide well with the reference. As noted by Breteler et al., RespR might still be usable for slow trend monitoring [21].

A limit of this study is that the range of the HR is narrow (40-120 beats/min), and the accuracy of the HR is only assessed in this limited range. Of this limited range, the higher HR frequencies is from the TAVI population with AF, in which the accuracy was not met. Therefor the accuracy of the biosensor within 40-90 beats/min can only be called accurate.

Another limitation of this study is the use of the impedance pneumography as reference for the RespR. It is widely considered that capnography is the gold standard for RespR. Impedance pneumography was used in this study as it was easier to implement the standard care routine. Unfortunately, also a lot of datasets were lost due to technical problems, which made the sample size for the RespR considerably smaller.

The kurtosis in the difference data (biosensor – reference) was very high for HR, which implies the difference data is outlier prone. This is likely caused by the sudden spikes in HR of the biosensor, as can be seen in Figure 5.2. On closer inspection of the data, these spikes in HR are likely caused by disturbances in the ECG, which are not sufficiently filtered out. This unbalanced distribution will influence the parametric results greatly and therefor the HR bias and precision is estimated non-parametrically.

Inaccuracy of the HR in AF patients is likely caused by different settings of the wearable biosensor and reference monitor. The reference HR was filtered to mimic the HR from the biosensor. However, it did not alter the outcome. The filters from the biosensor could not be replicated and the reference monitor

Page 55: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 5

37

has own filters, which most likely causes these differences. There is still a need for reliability analysis in AF patients, as about 40% of the patients getting TAVI have AF [38], [39].

Surprisingly, the biosensors HR agreed better in TAVI-patients with a sinus rhythm, than in the healthy subjects. Appendix A.1.2 shows that in healthy subject, the LOA is very large. Upon closer inspection of the data, this is likely caused by an inaccuracy of the reference monitor. During the breathing exercise, the reference HR dipped from 60 beats/min to 30 beats/min. During this dip also some data was missing. This indicates the reference monitor was probably inaccurate at this moment.

In conclusion, preliminary results show that the biosensors HR seems accurate for sinus rhythm and can be used for further data analysis. The RespR is likely to be inaccurate compared to impedance pneumography. However, further research should determine the cause of this inaccuracy.

5.4.1. Posture detectionThis is the first independent research into the activity classification of the wearable biosensor.

For this purpose, walking detection was analyzed for different walking speeds and posture was compared to the activity diaries. Result shows that the walking detection is accurate above 0.7 m/s, but

Figure 5.7 Gives the Kaplan-Meier figure for the survival of usable data, due to measurement length or data loss, with gaps of 2 minutes, 15 minutes, 1 hour and 4 hours. Every plus sign represents the end of a measurement. Under the figure, the number at risk are shown, every 10 hours.

0 10 20 30 40 50 60 70 80 90

Time (hour)

0

0.2

0.4

0.6

0.8

1

Surv

ival

of u

sabl

e da

ta

Kaplan-Meier

Length of data loss(hh:mm)

Length ofdata loss(hh:mm)

04:0001:0000:1500:02

Number recordings:

31313131

28272727

22211917

19171513

16151512

15141411

151413 9

111111 7

9996

7775

04:0001:0000:1500:02

Page 56: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

38

Chapter 5

degrades fast when walking slower. Posture detection is limited, as only a few outcomes are possible and posture classification deteriorates within a day.

This report has limitations, as no validated analysis are used for the posture detection, because there were none available. Therefore, quality assessment were made by comparing it to exercises and diaries. The precision of the diary, differed greatly between subjects, which made it hard to make a generalized analysis. In further research, diaries could be made more homogenous by giving stricter instruction how and what to note. Determining the walking detection accuracy was better controlled as it was overseen by a researcher.

Despite this limitations, it is clear that walking is detected well above 0.7 m/s. Implications of this threshold seem to be of minor influence in TAVI patients, as only one TAVI patient (209) walked slower. This indicates that in most TAVI patients the biosensor is able to detect walking accurately. However, one has to consider that the cohort in this study has a better functional status than compared to other studies reporting over a general TAVI population (Chapter 4).

More surprising is the result that the posture classification of the biosensor deteriorates much after one day, which makes it unusable in practice. It is hypothesized, that the signal analysis is somehow affected during the night. However, it is not known how posture is exactly detected (Chapter 3), therefore the real error cannot be pinpointed.

The biosensors algorithm for posture detection, unfortunately does not include a detection for running. In Figure 5.5 shows that the difference between walking and running is clearly visible in the acceleration data. The ability to classify running, would especially be valuable when the biosensor is used in sports and sport revalidation. Detection of running is not essential for TAVI patients, as it unlikely these patient will run frequently.

In conclusion, data from the posture classification is not reliable and unusable for further analysis. Walking and step count could be used in activity analysis, but is limited as a standalone parameter.

5.4.2. Data collectionData collection reliability is an essential part for the usability of the wearable biosensor. To our

knowledge, this is the first study reporting data collection reliability in an at home setting with the current set-up. The found reliability is low due to a combination of short recording times and large gaps in the data.

The data collection reliability found in this study is worse, than in other work in a hospital step-down unit [21]. The divergence in results is likely caused by the different study environments. In this study, biosensors were attached in hospital, where after study participants resumed their daily lives. As a result, there was less support for the patients in this study compared to an in-hospital research. Also, when the sensor or receiver malfunctioned, we could not replace it quickly.

The suboptimal data collection reliability ensues further analysis greatly, as only a few datasets have enough data eligible for in depth analysis. Therefore, measurement length should be improved and data gaps should be diminished. Reasons for short measurement lengths and data gaps will be reviewed by means of wearability questionnaires in Chapter 6.

5.5. ConclusionThis chapter investigated the reliability of the Philips wearable biosensor for vital signs, posture

detection and data collection in the TELE-TAVI study. Preliminary outcomes suggests that HR is the most reliable vital sign, but needs to be further analyzed in patients with AF. The RespR seems to be unreliable compared to the reference monitor, but could maybe be used for slow trend monitoring. Posture detection was limited and deteriorates fast over time. Data collection was below the predetermined limit and complicates further in-depth analysis of the biosensor data.

Page 57: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

39

Page 58: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

40

Page 59: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

41

6.1. IntroductionThe population is aging and mobile health applications are becoming more common in the health care

of the elderly. Acceptance of technology by elder patients relays heavily on the ease of use [45]. Elderly encounter different usability problems, as a result of physical decline, aging characteristics and disease complexities [46].

To assess the usability of a mobile system, the TELE-TAVI study is conducted. It is hypothesized that a wearable sensor can aid the transcatheter aortic valve implantation (TAVI) workflow. The system can only provide benefits, when its use is accepted by the intended population; elderly TAVI patients. Therefore, the user experience of the biosensors system is investigated as part of the TELE-TAVI study.

6.2. MethodTo test the usability of the Philips wearable biosensor and additional systems, data from the TELE-TAVI

study is used (Amsterdam UMC, location AMC, Amsterdam, The Netherlands). Patients with severe aortic valve stenosis, received the system at three moments: two weeks pre-TAVI (T0), post-TAVI (T1) and at 6 weeks follow-up (T2). After every measurement a questionnaire was filled in by the study participant about the used devices. This questionnaire contained the post study system usability questionnaire (PSSUQ) and a custom made questionnaire, targeting specific questions on the use of the wearable biosensor and the mobile phone.

6.2.1. DevicesThe system used in the TELE-TAVI study is a wearable patch sensor (Biosensor, Philips Medical System,

Andover, Massachusetts, USA) and Research Kit (Philips Medical System, Andover, Massachusetts, USA), which contained a mobile phone (Kyocera Brigadier, Android version 5.1.1) with smart phone application. The mobile phone with application captured data acquired by the biosensor and provided guidance in English for the set-up of the biosensor and gave feedback during the measurement. Set-up guidance involved giving instruction for preparing the skin, activating the biosensor, connecting the biosensor with the mobile phone, and accelerometer calibration. After the initial set-up, the application gave information on the battery life of the biosensor and connection of the signal. The screen also suggested to stay within 10 meter distance of the phone. Visible and audible warnings were given, in case the biosensors signal is not detected or the sensors adherence failed.

In the TELE-TAVI study, the researcher set-up and adhered the biosensors first, in all time points (T0-2). During this process, patients were engaged in participating and working with the phone and app. Patients received information booklets about the biosensor and its use, which were approved by the local ethics committee. Directly post-TAVI (T1), an extra sensor was supplied and patients had to change the sensor at home by themselves, or with the help of an informal caretaker. When patients were still admitted to the academic hospital, a researcher helped with changing the biosensor.

Chapter 6: End-user experience of the Philips wearable biosensor and Research Kit in the TELE-TAVI study

Page 60: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

42

Chapter 6

6.2.2. PSSUQOne of the questionnaires used to assess the user experience of system was the post study system

usability questionnaire (PSSUQ) [31]. This 19-item questionnaire evaluates the perceived satisfaction after system use. The questionnaire is divided in four categories; overall system use (SysUse), information quality (InfoQual) and interface quality (IntQuality). Answers were based on a 7-point Likert scale, from strongly agree (1) to strongly disagree (7) and a “not applicable” option. Missing data or not applicable data were interpolated. Hereafter, scores were calculated for every category as an average percentage of agreement (100% strongly agree, 0% strongly disagree), as suggested by the questionnaire developers.

6.2.3. Custom made questionnaireA custom made questionnaire, similar to the PSSUQ, was added to evaluate specific biosensor system

related questions (Table 6.1). It had 9 questions, with a Likert scale between 1 and 7 and a “not applicable” option. Questions involved the wearability, burden of the biosensor, as well as a specific question about the difficulty of changing sensors. All questions with accompanying Likert scale answers can be found in Table 6 1. Missing values were not used in the analysis. The custom made questionnaire was followed by an empty page, were further feedback, supplementary notes and practical restrictions could be entered. The used custom made questionnaire is included in Appendix A.3.1.

6.2.4. Statistical analysisPrimary outcomes are PSSUQ percentages and custom made questionnaire answers. Furthermore, a

paired t-test was used to evaluate if there is significant difference in the PSSUQ over the three measurements for all four categories. Next to that, we examined whether the PSSUQ results were related to the length of usable data (Pearson correlation). Usable data were defined as the data measurement length without gaps, due to data loss.

6.3. ResultsIn total 45 (90%) questionnaires were received, after measurement with the wearable biosensor. Of

these, 21, 16 and 8 were received from the pre-TAVI, post-TAVI and follow-up measurement, respectively. Of the received questionnaires, 3 PSSUQ were not filled in and 2 custom made questionnaires were missing. In 28 times feedback was given in the open comment section.

The PSSUQ results (in Table 6.2), show that system usability satisfaction is 63.2% (± 30.7). Appendix A2.1, gives the PSSUQ results per patient. Here, it is visible that the in-patient variation between the four domains is low, but the interpatient variation is large. There was no significant difference found between the answers of T0, T1 and T2, on all PSSUQ domains. Also no correlation was found between the length of usable data and PSSUQ outcomes. No significant difference was found between the answers at T0, T1 or T2.

The overall results of the custom made questionnaire are shown inFigure 6.1. Most patients experienced no discomfort (74%), itch (91%), and skin irritation (95%) and were not limited in daily activities by the measurement (79%). Removing the sensor from the skin was easy (83%) and not painful (86%). Patients

Table 6.1 Questions of the custom made questionnaire, with answers for Likert scales 1 and 7.

Questions 1 7

1 How was wearing the biosensor? Comfortable Uncomfortable

2 Did you experience itchiness caused by the biosensor? No itch Much itch

3 Did you experience skin irritation? No irritation Much irritation

4 Did the biosensor remained adhered to the skin? Stayed adhered Came off

5 Was the biosensor easily removable from the skin? Easy Much difficulty

6 Was removing the biosensor painful? Not painful Very painful

7 Were you activities limited because of the measurement? No limitation Much limitation

8 Did have to phone with you? Always Never

9 Changing the biosensor was easy. Strongly agree Strongly disagree

Page 61: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 6

43

said that they always carried there phone around (93%). In 31% of the TAVI patients, the biosensor fell off. Changing the sensor at T1 was easy for 56%, but 19% did not attempt to change the sensor.

In the open feedback page, the same 14 (31%) patients wrote that the sensor fell off. Another frequent note was that the audible warning signaling that the sensor signal was out of range, was often heard, even when the phone was nearby. As patient 204 commented after T0 (translated from Dutch):

“Very many (too many) warnings of connection failure, while he was in the pocket of my shirt or trousers. And that day and night!”

Also several commented that the changing the biosensor was rather challenging. For example, Pin 209 wrote:

“New adherence of the biosensor causes some problems. Needed to sort out the workings of a couple of things. I can imagine that not everybody can do this alone”.

Only one patient (Pin 211) described having difficulty with the foreign language.

6.4. DiscussionThis chapter explored the end-user experience of the wearable biosensor and additional phone and

application, in elderly TAVI patients in context of the TELE-TAVI study. Results demonstrate that patients experienced the wearable biosensor as comfortable and easy to use. However, the phone and applications experience varied much, but overall was lower than expected.

Limitation of this part of the TELE-TAVI study is that only a part of the screened TAVI patients (35%, see Chapter 3) participated. The participants in the study were probably patients open to and experienced with mobile phones and technology. Therefore, this study will probably overestimate the positive results of the overall TAVI population. The results of the PSSUQ are limited as patients experienced difficulty with the complicated sentence structure. The custom made questionnaire on the other hand was easier, but is not validated. Missing in both the questionnaires is comfortability with own phone use of participants. It seems that this played a major role in the satisfaction and ease of system use. Although the research methods are limited, the results provide a great and quantified insight in the use of the wearable biosensor and the additional systems.

The varying experience with the phone and mobile application has probably multiple reasons. One is of course, the skill of the patient with a smartphone. Also the applications language was English, while patients first language was Dutch. Surprisingly, only one patient complained about language difficulty. More patients explained that they could not understand English during the inclusions. But most patients seemed to have enough understanding of the language to handle the phone. Also there were a great number of steps and procedures for setting up the biosensor and phone, which made the system complicated. Therefore, the author suggest it is better to simplify the system, preferably without using a phone.

Wearing the biosensor was experienced as overall comfortable, but in almost a third of the patients the sensor fell off. Early detachment of the sensor has had major impacts on the clinical usability of the data obtained with the biosensor. Therefore, one could argue the adherence layer of the biosensor should be improved. Only, this could influence the wearability negatively. This makes for a difficult balance; patient

would probably be more reluctant to wear the sensor for long periods of time.

There was no significant relation found between the PSSUQ outcomes at T0, T1 and T2. A positive relation could have indicated that there is a learning curve for patients in using the system. Also, no relation was found between the PSSUQ outcomes and the amount of usable data. If the PSSUQ results also correlated with the patients ability to handle the system, it could have suggested that the data loss is not caused by the user, but rather a system error.

An unexpected finding was the feedback on the phones

Table 6.2 Overview of answers PSSUQ

Mean (± SD)

1 Overall 63.2 % (±30.7)

2 SysUse 65.0 % (±31.9)

3 InfoQual 61.7 % (±34.7)

4 IntQual 61.7 % (±31.5)

Mean (2-4) 62.8 % (±32.7)

PSSUQ indicates post-study system usability questionnaire, SD, standard deviation

Page 62: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

44

Chapter 6

audible warnings on the signal strength. Although patients kept the phone close to the sensor, they received a lot of warnings on the phone. This was experienced as very disturbing and in one case even let to discontinuation of study participation (Chapter 4). Patients were given the option to turn of the audible warnings. Disadvantage hereof, was that patients could not receive the warnings and forget the phone. Preferably, the audible warnings should be broadcasted by the sensor instead of the phone. Which gives also the possibility to delay the warning, so warnings are only broadcasted when the signal is lost for several minutes.

6.5. ConclusionIn conclusion, this chapter shows the user experience of the wearable biosensor and system, in an

elderly TAVI population. The wearability of the sensor is good, but probably compromises the longevity of the adhesiveness. The use of the phone was below expectation and audible warnings from the phone were experienced as very bothersome. A simplification of the system, would probably solve many of the experienced issues.

Figure 6.1 Shows results from custom questionnaire averaged for T0, T1 and T2

How was wearing the biosensor?

Did you experience itchiness caused by the biosensor?

Did you experience skin irritation?

Did the biosensor remained adhered to the skin?

Was the biosensor easily removable from the skin?

Was removing the biosensor painful?

Were you activities limited because of the measurement?

Did have to phone with you?

Changing the biosensor was easy.

Answers

Uncomfortable

A lot of itch

Irritation

Detached

Great di�culty

Painful

Hinderence

Always

Totally agree

Easy

Comfortable

No itch

No irritation

Adhered

Not painful

No hinderence

Never

Totally disagree50%

Not applicable

NA

33%

79%

70%

74%

74%

29% 60%

86%

81%

26% 49%

PositiveNegative

1 2 3 4 5 6 7

Page 63: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

45

Page 64: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

46

Page 65: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

47

7.1. ImplicationsIn current medicine, transcatheter aortic valve implantation (TAVI) is a widely accepted intervention

in patients with severe symptomatic aortic valve stenosis with intermediate to high surgical risk. However, used risk scores are not sensitive for TAVI outcomes. Frailty is closely related to surgical outcomes, but no standardized assessment is available. Recently, the use of wearable sensors is explored in objective frailty assessment, but none use an integrated approach of physical activity and cardiac physiology. The Philips wearable biosensor (Philips Medical Systems, Andover, Massachusetts, USA) measures vital signs and posture for up to 4 days, which could allow for integrated screening tool. Currently no publication is available on the biosensor.

This master thesis was part of the TELE-TAVI study, a prospective trial investigating the possibilities of remote patient monitoring in TAVI patients (Chapter 3). However, little is known on the reliability and end-user experience of the wearable biosensor, which should be addressed first. The aim of this master thesis was to find an objective measure with the Philips wearable biosensor for screening of TAVI patients. Hereafter, an objective measure for physical activity was investigated. First preliminary results in Chapter 4 show that the study population has a good functional status compared to overall TAVI population and other TAVI study populations.

Chapter 5 discusses the reliability of the wearable biosensors outputs and data collection. Vital signs were compared to the high-end patient monitor and preliminary results show that the heartrate (HR) is reliable during sinus rhythm. However, in patients with atrial fibrillation (AF), the HR differed more from the reference. The RespR seems to be unreliable compared to the reference monitor, but could be used for hourly trend monitoring. The biosensors posture was deemed impractical, as the accuracy deteriorated within a day. Most surprisingly was that only somewhat more than halve of the 96 hours of data were stored. The reliability of the data collection and posture detection had consequences for other parts of this thesis.

The user experience of TAVI-patients, was addressed in Chapter 6. The use of the phone and application received mixed results, some patients were satisfied with the systems, others less. Even though, patients included in the study were probably more accustomed and open to technology, which overestimated the results. The wearability of the biosensor was good, as most experienced no discomfort of the sensor and patients found the burden of wearing a sensor low. However, in a third of the patients the sensor fell off before the measurement ended.

Finally, an objective physical activity measure was researched in Chapter 7. Physical activity can be estimated with the raw acceleration recorded by the wearable biosensor. With this, daily activity profiles could be made. However, more research should focus on finding a robust algorithm for physical activity.

In conclusion, implementation of the wearable biosensor is feasible in the TAVI workflow. However, one of the major setbacks was the reliability of the wearable biosensor., as much data were lost without a known cause. Also the posture classification was unworkable, which let to exploration of estimating physical

Chapter 7: Discussion

Page 66: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

48

Chapter 7

activity. Daily physical activity profiles could be made, which shows the potential of the biosensor usability. Next to that, patient tolerated wearing the biosensor well and encourages research in sensor technology in elderly.

7.1.1. RecommendationsDuring this master thesis, data collection reliability had the most impact on analysis of the biosensors

data. Because of the many hours of data loss, only limited analysis were possible. Most notable, is that only 1 out of the 8 patients, who completed all measurements, had more than 80% of the data available at pre-TAVI screening, directly after TAVI and at six weeks follow-up. Consequently, data collection reliability must be improved, prior to new and larger study with the wearable biosensor.

Also user experience of the phone and application should be further improved. Most difficult for the TAVI patients was setting-up the sensor and phone connection and working with the phone. For this reason, it could be suggested to integrate the data collection module into the wearable sensor as it could also solve the data collection reliability. However, this will increase the costs of the sensor. Next to that, some potential additional functionalities of system will be lost, for example, manual data entry, such as measured blood pressure, would not be possible. However, with the wearable biosensors current functionalities, the data collection module could be integrated in the sensor.

Currently, the patient interface, which includes the phone and application is a disadvantage of the system. Improvement of the system should include end-users, in this case TAVI-patients [59]. New software for the application is available and addressed the issues with the frequent audio warnings when the connection between the phone and sensor was lost. A delay of 5 minutes between loss of connection and the warning was added, so short connection interruptions would not trigger a warning. However, the added value of the audio warning, is reduced, as patients could be moved too far away from the phone in this time period. Another addition of the new software, was that the interface of the app was secured with an 8-letter password. This will probably only decrease the usability of the phones application.

Before starting a new study with the wearable biosensor in TAVI-patients, the abovementioned problems should be addressed. Additionally, length of measurements could be improved, by developing an improved adhesive layer.

These recommendations do not imply that further research is not possible before these flaws are addressed. The data gathered during the TELE-TAVI study is rich and many algorithms could be developed for further use of the wearable sensors.

First, an algorithm that can detect physical activity is needed. Chapter 7 showed a promising start, however this is only an example of accelerometry analysis. Analysis of physical activity is a large research field and further profound investigation of possible algorithms is needed. Also, HR as a measure for screening should be considered. Yet, the standard heart rate variability (HRV) is less suitable in this patient population, as almost 29% of the TAVI population has atrial fibrillation, in which this analysis is impossible.

The end goal is to have an integrated analysis of vital functions and physical activity, as it would likely give an in-depth view into the patient’s physical and cardiac condition and possible improvement after TAVI. However, a challenge will be handling the different parameters’ resolution. Activity can be estimated every 10 seconds, but RespR is probably only reliable in slow hourly trends. Somehow, a smart algorithm should account for these differences.

Thus, before a new study is initiated, the wearable sensors data collection reliability, user interface and adhesive layer have to be improved. However, development of algorithms integrating physical activity and vital signs is possible with collected data. First focus should be making an algorithm for physical activity analysis.

7.2. Future perspectivesAlthough, the Philips wearable biosensor is not yet usable in all TAVI-patients, it is expected to be

useful in the near future. In the meantime, both TAVI and wearable sensing will further develop.

Page 67: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 7

49

7.2.1. TAVIThree possible cases are defined where a wearable sensor could aid TAVI workflow; screening, post-

procedure monitoring and follow-up. As TAVI is a relatively new intervention, management of TAVI workflow is changing constantly.

First, TAVI was only indicated in high risk patients with severe symptomatic aortic valve stenosis. The indication for TAVI is expanding, as now patients with intermediate mortality risks are also accepted. In addition, studies are currently exploring TAVI in low risk patients [60]. It is even suggested to study the early intervention of asymptomatic patients and prevents patient deterioration and irreversible myocardial damage [61], [62]. However, TAVI does come with peri- and post-procedural complication, which needs to considered. Wearable sensors could help identify the right timing and patient for TAVI, by monitoring activity and HR for long periods.

The use of wearable sensor is already acknowledged for monitoring patients post-TAVI [32]. Recently, a study started exploring the feasibility of ECG home-monitoring [33]. Next to monitoring, a wearable sensor could also benefit cardiac rehabilitation at home, by monitoring HR and physical activity, and engage exercise with serious interactive gaming [63].

Follow-up after TAVI could be expanded with wearable sensors, it can objectify patients improvement and give more insight in what happens between doctor’s visits. Objectively measuring symptom improvement could be used as outcomes randomized control trials and could reduce number of subjects as could earlier detect patient improvement.

7.2.2. Wearable sensorsThis thesis mainly focuses on the use of a wearable sensor in TAVI-patients, but more fields could

benefit from wearable sensing. Generally speaking, wearable sensing has potential to improve diagnostics and monitoring of a variety of diseases. Next to that, long term monitoring could give insight in a person’s health condition. With this, diseases could be found before symptoms are noticed or even prevented.

To make this happen, dedicated algorithms must be made. Currently, wearable sensors can measure a variety of outputs, but most are nonspecific without context. For example, HR can easily be measured, but without context an increase in HR could be caused by exercise or patient deterioration. Algorithm development could for example be established with machine learning, but these algorithms need vast amounts of data.

Functionality and wearability of wearable sensors is still being improved. Accuracy of sensor outputs are improving and more data can be recorded and stored. Battery life could be extended with body energy harvesting. Sensors are getting more flexible and miniaturized, which will improve wearability and extend wear-time. All together this will lead to more data collection of vital signs.

Page 68: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 69: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

51

8.1. IntroductionAortic valve stenosis (AoS) is the most prevalent heart valve condition and is associated with aging.

Nowadays, transcatheter aortic valve implantation is standard care for patients with intermediate to high risk at mortality. However, often used risk models poorly predict TAVI outcome and a specific TAVI model is needed [47], [48].

Adding frailty evaluation has shown to improve the already pre-existing scores [49]. Frailty is a syndrome, and can be observed as weight loss, muscle weakness, poor endurance and energy, slowness and low physical activity levels [15]. Physical performance tests are often utilized for assessing frailty [16], [17]. Physical performance can for example be determined with the six minute walk test (6MWT), short physical performance battery (SPPB) or gait speed and is related to the outcome of TAVI [50]–[52]. However, all give only a snapshot of the patient physical functioning in an estranged out-of-home environment.

Daily activity measured with sensor technology could extend physical performance analysis. Multiple studies are known to assess physical frailty with sensor technology [19]. Combining an objective frailty measure with heart rate, could allow for an extensive evaluation of cardiac condition, improving TAVI screening. The wearable biosensor (Philips Medical System, Andover, Massachusetts, USA), can measure vital signs and tri-axial acceleration and would allow for such evaluation. However, an activity algorithm is not embedded in the system and the posture detection is unreliable (Chapter 5).

The aim of this chapter is to find an objective physical activity measure with the wearable biosensor for TAVI patients.

8.2. MethodDevelopment of an objective physical activity measure is part of the TELE-TAVI study, a prospective,

investigator initiated study, exploring telemedicine in TAVI-patients. The design of the study is extensively described in Chapter 3. For now, a short description of study will be given.

In the TELE-TAVI study, six healthy volunteers and 24 TAVI-patients participated. Patients received a wearable sensor pre-TAVI (T0), directly after the TAVI (T1) and at 6 weeks follow-up (T2). The Philips biosensor (Philips Medical System, Andover, Massachusetts, USA), is a light-weight sensor measuring vital signs and tri-axial acceleration for up to 4 days. At T0 and T2, patients also underwent multiple physical function tests, with a six minute walk test (6MWT), short physical performance battery (SPPB) and grip strength. Healthy subjects, received the biosensor once, and performed the same physical analysis. Next to that, the healthy study population did a walking exercise, where they walked for 2 minutes at 100, 75 and 50 steps per minute with a metronome. The distance traveled during each 2 minute cycle was noted. Walking was alternated with one minute of sitting. After the protocol, the healthy study population maintained a diary in which they noted activities.

Chapter 8: Objective physical activity levels with the wearable biosensors in patients undergoing transcatheter aortic valve implantation: first results

Page 70: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

52

Chapter 8

8.2.1. Data analysisThe biosensors raw tri-axial acceleration data were decrypted with the Research Kit (Philips Medical

System, Andover, Massachusetts, USA) and retrieved in CSV format. All data were stored and processed using Matlab (The MathWorks, Natick, Massachusetts, USA). Missing data were interpolated with empty values (not-a-number).

8.2.2. IMAActivity was estimated with the integral of the modulus of the accelerometer output (IMA), as described

by Bouten et al. [53], [54]. IMA is dimensionless and defined as:

In which t is the time, T is the time period in which the IMA is calculated, a is the acceleration in direction x, y and z.

Acceleration is first filtered with a bandpass filter from 0.11 to 20 Hz to remove off-set caused by gravity and vibrational components not caused by human motion [54]. Filtering the biosensors acceleration was complicated by many small and larger gaps were present in the data. Therefore, small gaps (< 10 samples) were linearly interpolated. Hereafter, acceleration was filtered in between the stretches with longer missing data. Finally, the IMA was computed with a window (T) of 10 seconds. When 10% of the data in the window was missing, the IMA was replaced with an empty value.

The IMA was averaged during 6MWT of the TAVI-patients and healthy subjects. Also the averaged IMA was computed during the metronome walking exercises of 100, 75 and 50 steps/min in the healthy subjects. Linear regression was computed between the walking speed and IMA of during the above mentioned walking exercises and 6MWT.

8.2.3. Activity levelsFrom the computed IMA, activity levels were determined by dividing the IMA into four categories; no,

low, medium and high activity. The categories were determined by defining three thresholds (low, medium and high), separating the four categories. With this, activity profiles, which are percentages of detected activity levels, could be made for the whole measurement.

8.2.4. Daily activity levelsFull measurement activity profiles are largely influenced by the amount of sleep during the

measurement, which does not reflect the activity pattern during the day. Therefore, daily activity profiles were made. Daily activity profiles are computed only for complete days, which are days that were recorded from waking up to going to bed.

First, awake hours must be separated from the whole measurement. For this, the calculated IMA was filtered with an averaging filter of two hours. With this, slow patterns are visible in the activity without losing

𝐼𝐼𝐼𝐼𝐼𝐼 = ∫ |𝑎𝑎𝑥𝑥(𝑡𝑡)|𝑑𝑑𝑡𝑡𝑡𝑡0+𝑇𝑇

𝑡𝑡=𝑡𝑡0+ ∫ |𝑎𝑎𝑦𝑦(𝑡𝑡)|𝑑𝑑𝑡𝑡

𝑡𝑡0+𝑇𝑇

𝑡𝑡=𝑡𝑡0+ ∫ |𝑎𝑎𝑧𝑧(𝑡𝑡)|𝑑𝑑𝑡𝑡

𝑡𝑡0+𝑇𝑇

𝑡𝑡=𝑡𝑡0

Figure 8.1 Gives the estimated power spectrum of the unfiltered (dark blue) and filtered (yellow) acceleration data (x-direction). Acceleration is filtered with a bandpass filter of 0.11 and 20 Hz (grey areas).

0 5 10 15 20 25Frequency (Hz)

-50

0

50

100

Pow

er (d

B)

Raw acc.Filtered acc.Filter

2 3

Page 71: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 8

53

to much resolution. Sleeping periods were easily determined with a threshold (Figure 8.2). However, naps or little activity during the day sometimes interfered and needed to be discarded. Thus, first an estimate of the number of nights was made. After this, the smallest awake or sleep parts were ignored till the number of nights was equal to the estimated number of nights. Finally, only days that were separated by nights were used to make a daily activity profile. With this, only daily activity profiles were only computed for completely measured days.

8.3. ResultsThe walking protocol and 6MWT was completed by all 6 healthy subjects. From the 24 patients, 8

completed the protocol and received a sensor at T2. One TAVI patient was not able to perform the 6MWT at T0 and T2, as the patient was wheelchair bound and, 1 was excluded from analysis, because patient stood for 30 seconds and 3 were not analyzed as biosensor data were lost. In total, 32 6MWT could be compared with the IMA (Figure 5.1).

8.3.1. IMAThe estimated power spectrum of the raw and filtered acceleration is given Figure 8.1. In the areas

where is not filtered, the signal strength is unchanged. In the filtered frequencies, the acceleration is sufficiently damped. In the power spectrum, peaks can be found at 2 and 3 Hz.

Table 8.1 shows the average and range of the IMA on a full measurement, for healthy subjects and TAVI patients. The averaged IMA was higher for healthy subjects with 50.5 (± 17.6), compared to TAVI patients with 32.2 (±8.3). Also the IMA ranges were larger in healthy subjects (3.9 – 703.9) than in TAVI patients (3.9 – 348.3). Individual study subjects’ results can be found in Appendix A.3.1. Figure 8.3, gives the linear regression between walking speed and calculated IMA. A significant correlation was found, between IMA and walking speed (rs2 = 0.80 and p<0.001).

8.3.2. Activity levelsThe thresholds to determine no, low, medium and high activity, were chosen on the basis of Figure 8.3.

Table 8.1 IMA and activity characteristics over full measurements.

Healthy TAVI OverallIM

AMean 50.5 ± 17.6 32.2 ± 8.3 35.7 ± 12.5

Range Min 3.9 ± 0.8 3.9 ± 0.5 3.9 ± 0.6

Max 703.9 ± 299.2 348.3 ± 55.3 415.0 ± 191.8

Activ

ity

No (%) 54.5 ± 9.0 65.3 ± 9.6 63.3 ± 10.3

Low (%) 28.2 ± 4.1 24.6 ± 6.8 25.2 ± 6.5

Medium (%) 15.0 ± 5.2 10.0 ± 3.8 10.9 ± 4.5

High (%) 2.4 ± 2.2 0.2 ± 0.3 0.6 ± 1.2

IMA indicates integrated modulus of acceleration; TAVI, transcatheter aortic valve implantation.

Table 8.2 IMA and activity characteristics during the day.

Healthy TAVI Overall

Mean IMA daily 52.4 ± 10.1 33.8 ± 8.7 37.9 ± 11.8

Activ

ity

No (%) 31.8 ± 4.3 43.3 ± 7.8 40.8 ± 8.6

Low (%) 35.6 ± 7.2 26.8 ± 5.6 28.7 ± 6.9

Medium (%) 16.1 ± 5.1 11.1 ± 4.3 12.2 ± 4.8

High (%) 2.1 ± 1.0 0.2 ± 0.4 0.6 ± 0.9

IMA indicates integrated modulus of acceleration; TAVI, transcatheter aortic valve implantation.

Page 72: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

54

Chapter 8

Walking was considered medium activity in TAVI patients. Most TAVI patients had an averaged IMA between 100 and 250 during walking of the 6MWT, which made the thresholds for medium activity. Averaged IMA values were not computed for lower activities. Therefore, the threshold between no and low activity was based on the diaries of healthy subjects. A threshold of 20 was selected, which was able to differentiate lying and sitting activities from standing. An example of the activity classification can be found in Figure 8.4. It shows the filtered acceleration and IMA. Every IMA window is classified in no, low, medium or high activity levels, which result in activity pattern, shown in the lower panel.

Over a complete measurement, TAVI-patients were more often not active than healthy subjects (65.3% (± 9.6%) and 54.5% (± 9.0%), respectively). High activity levels were sporadically seen in TAVI-patients (0.2% ± 0.3%), and very little in healthy subjects (2.4% ± 2.1%).

8.3.3. Daily activity levelsIn 4 and 15 datasets from healthy and TAVI-patients respectively, a daily activity profile could be made.

In Figure 8.2 the algorithm for day and night classification is visible. Longer inactive periods under the thresholds are classified as nights. Short periods under the cut-off are not detected as nights. Table 8.2, gives the overall result of the daily activity profiles. Averaged IMA during the day is 50.5 (± 17.6) and 32.2 (± 8.3). The percentage of no activity during the day was less compared to the complete measurement.

In one patient daily activity profiles could be made for T0, T1 and T2, which are displayed in Figure 8.5. No high activity was measured in all measurement. Similar portions of activity are found for T0 and T2, with low and medium activity of 25-37% and 15-22%, respectively. Directly post-TAVI (T1), less activity is detected, and low and high activity portions are 3-5% and 15-19%, respectively.

8.4. DiscussionAn objective measure for physical

functioning, can improve patient screening before TAVI. Physical functioning is estimated with the integrated modulus of tri-axial acceleration and gave a good correlation with gait speed. With this measure, daily physical activity profiles could be made, which could give more insight in patients physical performance.

Previous extensive work showed that IMA was related to daily energy expenditure [53]. In

Figure 8.2 Shows the data used for sleep detection. IMA (grey line) is filtered with a 2 hour averaging filter (red line). Hereafter sleeping periods (dark areas) are detected when the filtered IMA is below the (yellow) threshold.

06/21 06/22 06/23 06/240

100

200

300

400

IMA �lteredIMA

Cut-o� nightNight

IMA

(cou

nts)

Figure 8.3 Correlation between IMA and gait speed. Dark and light blue squares are the 6MWT results for the TAVI-patients and healthy subjects, respectively. Orange, green and red squares resulted from the metronome walking of 100, 75 and 50 steps/min, respectively. Correlation found between IMA and gait is r2=0.8 (p<0.001).

)

0

50

100

150

200

250

300

350

400

IMA

(cou

nts)

r 2 = 0.8p < 0.001

0.5 1.0 1.5 2.0

Speed (m/s

Page 73: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 8

55

Figure 8.4 Overview of the algorithm for detecting activity levels. The IMA is divided in 4 categories (high, me-dium, low and no activity) by thresholds. Hereafter, activity levels are computed every 10 seconds, given in the lower panel. Acc. indicates acceleromter data, IMA, integrated modulus of acceleration.

19:50 19:55 20:000

100

200

300

400

20

100

250

Acc.IMAHighMediumLow

20:05

19:50 19:55 20:00

Act

ivity

20:05

High activity Medium activity Low activity No activity

Figure 8.5 Gives the activity levels per day, pre-TAVI, directly after TAVI and at 6 weeks follow-up for patient 201.The legend is given underneath. At T0 and T1 3 complete days are analyzed, at T2 4 days could be computed. There was no high activity detected at T0, T1 or T2.

46%

32%

22%

48%

31%

20%

60% 23%

18%

T0: pre-TAVI

53%

29%

16%

1%

53%25%

22%

48%

37%

15%

T2: 6 week follow-up

Complete day 1 2 3

High activity Medium activity Low activity No activity

78%

19%

3%

82%

15%

3%

76%

19%

5%

81%

15%

4%

T1: direct post-TAVI

4

Page 74: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

56

Chapter 8

this paper a tri-axial accelerometer was attached to the lower back. The wearable biosensor is attached to the upper torso, so torso movement is also measured. Accordingly, IMA should again be validated to energy expenditure.

This study only shows some first ideas for using objective physical activity levels. But one of the major shortcomings here, is that the threshold selection was arbitrary. With the current thresholds, high activity was barely visible and minimal differences were visible between healthy subjects and TAVI patients. The thresholds should be optimized to show differences between patients with AoS. Other papers even suggest to analyze accelerometry without the use of cut-points, but rather using the raw data [55], [56]. It is believed this will improve physical activity characterization to estimate energy expenditure. Figure 8.1 shows already some promise, as there are two frequency peaks visible, in the walking frequency range [57]. The peak frequency could give insight into the gait and the power of the frequency could indicate how much a patient is walking. The physical activity classification research field has grown much the last two decades and an extensive study is needed to find a physical activity classification for the Philips wearable biosensor to provide essential information.

The work reported here is limited, but the use of activity levels shows great promise. Not only could it change TAVI screening, but also follow-up. Post-procedural, patients’ activity could be monitored and coached to exercise more. Exercise training is known to have positive benefits on the exercise capacity and quality of life [58]. Also, pre- and post- TAVI activity could be compared as an objective outcome measure of TAVI. Moreover, regular objective assessment of activity levels could detect deterioration, as consequence of worsening AoS

The wearable biosensor also records vital signs, such as heart rate and respiratory rate. A further in-depth analysis of the patients cardiac function could made possible by combining these vital signs and activity analysis.

8.5. ConclusionThis chapter showed that objective activity levels could be estimated with the integrated modulus of

the tri-axial acceleration. This measure shows promise to improve TAVI screening, as well as, monitoring and screening.

Page 75: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

57

Page 76: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

58

Page 77: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

59

[1] A. Patel and A. J. Kirtane, “Aortic Valve StenosisNarrowed Aortic ValveJAMA Cardiology Patient Page,” JAMA Cardiol., vol. 1, no. 5, p. 623, Aug. 2016.

[2] R. L. J. Osnabrugge, D. Mylotte, S. J. Head, N. M. Van Mieghem, V. T. Nkomo, C. M. Lereun, A. J. J. C. Bogers, N. Piazza, and A. P. Kappetein, “Aortic stenosis in the elderly: Disease prevalence and number of candidates for transcatheter aortic valve replacement: A meta-analysis and modeling study,” J. Am. Coll. Cardiol., vol. 62, no. 11, pp. 1002–1012, 2013.

[3] B. Iung, G. Baron, E. G. Butchart, F. Delahaye, C. Gohlke-Bärwolf, O. W. Levang, P. Tornos, J. L. Vanoverschelde, F. Vermeer, E. Boersma, P. Ravaud, and A. Vahanian, “A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on valvular heart disease,” Eur. Heart J., vol. 24, no. 13, pp. 1231–1243, 2003.

[4] S. V Arnold, S. M. O’Brien, S. Vemulapalli, D. J. Cohen, A. Stebbins, J. M. Brennan, D. M. Shahian, F. L. Grover, D. R. Holmes, V. H. Thourani, E. D. Peterson, and F. H. Edwards, “Inclusion of Functional Status Measures in the Risk Adjustment of 30-Day Mortality After Transcatheter Aortic Valve Replacement: A Report From the Society of Thoracic Surgeons/American College of Cardiology TVT Registry,” JACC Cardiovasc. Interv., vol. 11, no. 6, pp. 581–589, 2018.

[5] I. H. Center, “Transcatheter aortic valve implantation (TAVI).” .[6] S. A. M. Nashef, F. Roques, L. D. Sharples, J. Nilsson, C. Smith, A. R. Goldstone, and U. Lockowandt,

“Euroscore II,” Eur. J. Cardio-thoracic Surg., vol. 41, no. 4, pp. 734–745, 2012.[7] R. S. D’Agostino, J. P. Jacobs, V. Badhwar, F. G. Fernandez, G. Paone, D. W. Wormuth, and D. M. Shahian,

“The Society of Thoracic Surgeons Adult Cardiac Surgery Database: 2018 Update on Outcomes and Quality,” Ann. Thorac. Surg., vol. 105, no. 1, pp. 15–23, 2018.

[8] V. Falk, H. Baumgartner, J. J. Bax, M. De Bonis, C. Hamm, P. J. Holm, B. Iung, P. Lancellotti, E. Lansac, D. R. Muñoz, R. Rosenhek, J. Sjögren, P. Tornos Mas, A. Vahanian, T. Walther, O. Wendler, S. Windecker, and J. L. Zamorano, “2017 ESC/EACTS Guidelines for the management of valvular heart disease,” Eur. J. Cardiothorac. Surg., vol. 52, no. 4, pp. 616–664, 2017.

[9] B. E. Stähli, H. Tasnady, T. F. Lüscher, C. Gebhard, F. Mikulicic, L. Erhart, I. Bühler, U. Landmesser, L. Altwegg, M. B. Wischnewsky, J. Grünenfelder, V. Falk, R. Corti, and W. Maier, “Early and Late Mortality in Patients Undergoing Transcatheter Aortic Valve Implantation: Comparison of the Novel EuroScore II with Established Risk Scores,” Cardiology, vol. 126, no. 1, pp. 15–23, 2013.

[10] E. Durand, B. Borz, M. Godin, C. Tron, P.-Y. Litzler, J.-P. Bessou, J.-N. Dacher, F. Bauer, A. Cribier, and H. Eltchaninoff, “Performance Analysis of EuroSCORE II Compared to the Original Logistic EuroSCORE and STS Scores for Predicting 30-Day Mortality After Transcatheter Aortic Valve Replacement,” Am. J. Cardiol., vol. 111, no. 6, pp. 891–897, 2013.

[11] H. S. Lin, J. N. Watts, N. M. Peel, and R. E. Hubbard, “Frailty and post-operative outcomes in older surgical patients: A systematic review,” BMC Geriatr., vol. 16, no. 1, 2016.

[12] K. Rockwood, S. E. Howlett, C. Macknight, B. L. Beattie, H. Bergman, D. B. Hogan, C. Wolfson, and I. Mcdowell, “Prevalence , Attributes , and Outcomes of Fitness and Frailty in Community-Dwelling Older

Chapter 9: References

Page 78: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

60

Chapter 9

Adults : Report From the Canadian Study of Health and Aging,” vol. 59, no. 12, pp. 1310–1317, 2018.[13] X. Song, A. Mitnitski, and K. Rockwood, “Prevalence and 10-Year outcomes of frailty in older adults in

relation to deficit accumulation,” J. Am. Geriatr. Soc., vol. 58, no. 4, pp. 681–687, 2010.[14] A. Sepehri, T. Beggs, A. Hassan, C. Rigatto, C. Shaw-Daigle, N. Tangri, and R. C. Arora, “The impact of

frailty on outcomes after cardiac surgery: A systematic review,” J. Thorac. Cardiovasc. Surg., vol. 148, no. 6, pp. 3110–3117, 2014.

[15] L. P. Fried, C. M. Tangen, J. Walston, A. B. Newman, C. Hirsch, J. Gottdiener, T. Seeman, R. Tracy, W. J. Kop, and G. Burke, “Fenotipo_Frailty,” vol. 56, no. 3, pp. 146–157, 2001.

[16] E. Dent, P. Kowal, and E. O. Hoogendijk, “Frailty measurement in research and clinical practice: A review,” Eur. J. Intern. Med., vol. 31, pp. 3–10, 2016.

[17] F. Buckinx, J. Y. Reginster, J. Petermans, J. L. Croisier, C. Beaudart, T. Brunois, and O. Bruyère, “Relationship between frailty, physical performance and quality of life among nursing home residents: the SENIOR cohort,” Aging Clin. Exp. Res., vol. 28, no. 6, pp. 1149–1157, 2016.

[18] A. W. Schoenenberger, S. Stortecky, S. Neumann, A. Moser, P. Jüni, T. Carrel, C. Huber, M. Gandon, S. Bischoff, C. M. Schoenenberger, A. E. Stuck, S. Windecker, and P. Wenaweser, “Predictors of functional decline in elderly patients undergoing transcatheter aortic valve implantation (TAVI),” Eur. Heart J., vol. 34, no. 9, pp. 684–692, 2013.

[19] L. Dasenbrock, A. Heinks, M. Schwenk, and J. M. Bauer, “Technology-based measurements for screening, monitoring and preventing frailty,” Z. Gerontol. Geriatr., vol. 49, no. 7, pp. 581–595, 2016.

[20] S. Parvaneh, C. L. Howe, N. Toosizadeh, B. Honarvar, M. J. Slepian, M. Fain, J. Mohler, and B. Najafi, “Regulation of Cardiac Autonomic Nervous System Control across Frailty Statuses: A Systematic Review,” Gerontology, vol. 62, no. 1, pp. 3–15, 2015.

[21] M. J. M. Breteler, E. Huizinga, K. Van Loon, L. P. H. Leenen, D. A. J. Dohmen, C. J. Kalkman, and T. J. Blokhuis, “Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: A clinical validation study,” BMJ Open, vol. 8, no. 2, pp. 1–10, 2018.

[22] A. M. Chan, N. Selvaraj, N. Ferdosi, and R. Narasimhan, “Wireless patch sensor for remote monitoring of heart rate, respiration, activity, and falls,” Proc. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. EMBS, pp. 6115–6118, 2013.

[23] A. Chan, N. Ferdosi, and R. Narasimhan, “Ambulatory respiratory rate detection using ECG and a triaxial accelerometer,” Conf. Proc. IEEE Eng. Med. Biol. Soc., vol. 2013, pp. 4058–4061, 2013.

[24] N. Selvaraj, “Long-term remote monitoring of vital signs using a wireless patch sensor BT - 2014 IEEE Healthcare Innovation Conference, HIC 2014, October 8, 2014 - October 10, 2014,” pp. 83–86, 2014.

[25] Y. Ou, Y. Li, and M. Chen, “The interpretation of 2017 ESC/EACTs and AHA/ACC guidelines for the management of valvular heart disease,” Chinese J. Evidence-Based Med., vol. 17, no. 11, pp. 1260–1264, 2017.

[26] S. Suave Lobodzinski and M. M. Laks, “New devices for very long-term ECG monitoring,” Cardiol. J., vol. 19, no. 2, pp. 210–214, 2012.

[27] M. A. Rosenberg, M. Samuel, A. Thosani, and P. J. Zimetbaum, “Use of a noninvasive continuous monitoring device in the management of atrial fibrillation: A pilot study,” PACE - Pacing Clin. Electrophysiol., vol. 36, no. 3, pp. 328–333, 2013.

[28] M. Hernandez-Silveira, K. Ahmed, S. S. Ang, F. Zandari, T. Mehta, R. Weir, A. Burdett, C. Toumazou, and S. J. Brett, “Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs,” BMJ Open, vol. 5, no. 5, pp. 1–10, 2015.

[29] J. F. F. Lima, L. R. Coura, J. P. Filho, S. D. S. Cruz, and J. G. F. Dos Santos, “Dimensionamento E Operação Em Escala Real De Um Sistema Híbrido De Tratamento Uasb E Lagoas Facultativas – Para Os Efluentes Domésticos E Industriais Da Xerox Do Nordeste , Bahia,” Ix Silubesa, vol. 42, no. 1, p. 8, 2000.

[30] J. P. Martínez, R. Almeida, S. Olmos, A. P. Rocha, and P. Laguna, “A Wavelet-Based ECG Delineator Evaluation on Standard Databases,” IEEE Trans. Biomed. Eng., vol. 51, no. 4, pp. 570–581, 2004.

Page 79: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 9

61

[31] J. R. Lewis, “Psychometric Evaluation of the PSSUQ Using Data from Five Years of Usability Studies,” Int. J. Hum. Comput. Interact., vol. 14, no. 3–4, pp. 463–488, 2002.

[32] M. C. Hermans, M. S. Van Mourik, H. J. Hermens, J. Baan Jr, and M. M. Vis, “Remote Monitoring of Patients Undergoing Transcatheter Aortic Valve Replacement: A Framework for Postprocedural Telemonitoring,” JMIR Cardio, vol. 2, no. 1, p. e9, 2018.

[33] M. Natarajan, “Remote ECG Monitoring to Reduce Complications Following Transcatheter Aortic Valve Implantations,” 2019. .

[34] Philips, Instructions for use: IntelliVue Patient Monitor. Germany: Philips, 2006.[35] K. Ogan, V. A. Master, and D. J. Canter, “Perioperative Assessment of Elderly Surgical Patients,” Curr.

Transl. Geriatr. Exp. Gerontol. Rep., vol. 2, no. 2, pp. 45–50, 2013.[36] S. Kim, A. K. Brooks, and L. Groban, “Preoperative assessment of the older surgical patient: Honing in

on geriatric syndromes,” Clin. Interv. Aging, vol. 10, pp. 13–27, 2014.[37] R. M. Dodds, H. E. Syddall, R. Cooper, M. Benzeval, I. J. Deary, E. M. Dennison, G. Der, C. R. Gale, H. M.

Inskip, C. Jagger, T. B. Kirkwood, D. A. Lawlor, S. M. Robinson, J. M. Starr, A. Steptoe, K. Tilling, D. Kuh, C. Cooper, and A. A. Sayer, “Grip strength across the life course: Normative data from twelve British studies,” PLoS One, vol. 9, no. 12, pp. 1–15, 2014.

[38] F. van Kesteren, M. S. van Mourik, E. M. A. Wiegerinck, J. Vendrik, J. J. Piek, J. G. Tijssen, K. T. Koch, J. P. S. Henriques, J. J. Wykrzykowska, R. J. de Winter, A. H. G. Driessen, A. Kaya, R. N. Planken, M. M. Vis, and J. Baan, “Trends in patient characteristics and clinical outcome over 8 years of transcatheter aortic valve implantation,” Netherlands Hear. J., vol. 26, no. 9, pp. 445–453, 2018.

[39] H. Reichenspurner, A. Schaefer, U. Schäfer, D. Tchétché, A. Linke, M. S. Spence, L. Søndergaard, H. LeBreton, G. Schymik, M. Abdel-Wahab, J. Leipsic, D. L. Walters, S. Worthley, M. Kasel, and S. Windecker, “Self-Expanding Transcatheter Aortic Valve System for Symptomatic High-Risk Patients With Severe Aortic Stenosis,” J. Am. Coll. Cardiol., vol. 70, no. 25, pp. 3127–3136, 2017.

[40] N. Straiton, K. Jin, R. Bhindi, and R. Gallagher, “Functional capacity and health-related quality of life outcomes post transcatheter aortic valve replacement: A systematic review and meta-analysis,” Age Ageing, vol. 47, no. 3, pp. 478–482, 2018.

[41] L. Koizia, S. Khan, A. Frame, G. W. Mikhail, S. Sen, N. Ruparelia, N. Hadjiloizou, I. S. Malik, and M. B. Fertleman, “Use of the Reported Edmonton Frail Scale in the assessment of patients for transcatheter aortic valve replacement. A possible selection tool in very high-risk patients?,” J. Geriatr. Cardiol., vol. 15, no. 6, pp. 463–466, 2018.

[42] S. Fukui, M. Kawakami, Y. Otaka, A. Ishikawa, K. Mizuno, T. Tsuji, K. Hayashida, T. Inohara, F. Yashima, and M. Liu, “Physical frailty in older people with severe aortic stenosis,” Aging Clin. Exp. Res., vol. 28, no. 6, pp. 1081–1087, 2016.

[43] M. Joshi, H. Ashrafian, L. Aufegger, S. Khan, S. Arora, G. Cooke, and A. Darzi, “Wearable sensors to improve detection of patient deterioration.,” Expert Rev. Med. Devices, vol. 00, no. 00, pp. 1–10, 2018.

[44] M. Thesis, “Remote postprocedural monitoring of patients undergoing Transcatheter Aortic Valve Implantation Exploring possibilities,” 2015.

[45] C. Lee and J. F. Coughlin, “PERSPECTIVE: Older Adults’ Adoption of Technology: An Integrated Approach to Identifying Determinants and Barriers,” J. Prod. Innov. Manag., vol. 32, no. 5, pp. 747–759, 2015.

[46] G. A. Wildenbos, M. W. M. Jaspers, M. P. Schijven, and L. W. Dusseljee-Peute, “Mobile health for older adult patients: Using an aging barriers framework to classify usability problems,” Int. J. Med. Inform., vol. 124, no. May 2018, pp. 68–77, 2019.

[47] P. F. Ludman, N. Moat, M. A. de Belder, D. J. Blackman, A. Duncan, W. Banya, P. A. MacCarthy, D. Cunningham, O. Wendler, and D. Marlee, “Transcatheter aortic valve implantation in the United Kingdom: temporal trends, predictors of outcome, and 6-year follow-up: a report from the UK Transcatheter Aortic Valve Implantation (TAVI) Registry, 2007 to 2012,” Circulation, vol. 131, no. 13, pp. 1181–1190, 2015.

Page 80: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

62

Chapter 9

[48] A. Halkin, A. Steinvil, G. Witberg, A. Barsheshet, M. Barkagan, A. Assali, A. Segev, P. Fefer, V. Guetta, and I. M. Barbash, “Mortality prediction following transcatheter aortic valve replacement: a quantitative comparison of risk scores derived from populations treated with either surgical or percutaneous aortic valve replacement. The Israeli TAVR Registry Risk Model Accuracy Assessment (IRRMA) study,” Int. J. Cardiol., vol. 215, pp. 227–231, 2016.

[49] S. Stortecky, A. W. Schoenenberger, A. Moser, B. Kalesan, P. Jüni, T. Carrel, S. Bischoff, C.-M. Schoenenberger, A. E. Stuck, S. Windecker, and P. Wenaweser, “Evaluation of Multidimensional Geriatric Assessment as a Predictor of Mortality and Cardiovascular Events After Transcatheter Aortic Valve Implantation,” JACC Cardiovasc. Interv., vol. 5, no. 5, p. 489 LP-496, May 2012.

[50] P. Green, D. J. Cohen, P. Généreux, T. McAndrew, S. V. Arnold, M. Alu, N. Beohar, C. S. Rihal, M. J. Mack, S. Kapadia, D. Dvir, M. S. Maurer, M. R. Williams, S. Kodali, M. B. Leon, and A. J. Kirtane, “Relation between six-minute walk test performance and outcomes after transcatheter aortic valve implantation (from the PARTNER Trial),” Am. J. Cardiol., vol. 112, no. 5, pp. 700–706, 2013.

[51] S. Kano, M. Yamamoto, T. Shimura, A. Kagase, M. Tsuzuki, A. Kodama, Y. Koyama, T. Kobayashi, K. Shibata, N. Tada, T. Naganuma, M. Araki, F. Yamanaka, S. Shirai, K. Mizutani, M. Tabata, H. Ueno, K. Takagi, A. Higashimori, T. Otsuka, Y. Watanabe, and K. Hayashida, “Gait Speed Can Predict Advanced Clinical Outcomes in Patients Who Undergo Transcatheter Aortic Valve Replacement: Insights From a Japanese Multicenter Registry.,” Circ. Cardiovasc. Interv., vol. 10, no. 9, Sep. 2017.

[52] R. Bagur, J. Rodés-Cabau, É. Dumont, R. De Larochellière, D. Doyle, P. Pibarot, M. Côté, M.-A. Clavel, J. Villeneuve, M. Gutiérrez, P. Poirier, and O. F. Bertrand, “Performance-based functional assessment of patients undergoing transcatheter aortic valve implantation,” Am. Heart J., vol. 161, no. 4, pp. 726–734, 2011.

[53] C. V Bouten, K. R. Westerterp, M. Verduin, and J. D. Janssen, “Assessment of energy expenditure for physical activity using a triaxial accelerometer,” Med. Sci. Sports Exerc., vol. 26, no. 12, p. 1516—1523, Dec. 1994.

[54] S. Bosch, R. Marin-Perianu, P. Havinga, and M. Marin-Perianu, “Energy-efficient assessment of physical activity level using duty-cycled accelerometer data,” Procedia Comput. Sci., vol. 5, pp. 328–335, 2011.

[55] R. P. Troiano, J. J. McClain, R. J. Brychta, and K. Y. Chen, “Evolution of accelerometer methods for physical activity research,” Br. J. Sports Med., vol. 48, no. 13, p. 1019 LP-1023, Jul. 2014.

[56] S. Liu, R. Gao, and P. Freedson, “Computational methods for estimating energy expenditure in human physical activities,” Med. Sci. Sports Exerc., vol. 44, no. 11, p. 2138, 2012.

[57] K. Chen and B. Yu, “Certain progress of clinical research on Chinese integrative medicine,” Chin Med J, vol. 112, no. 10, pp. 934–937, 1999.

[58] A. Pressler, J. W. Christle, B. Lechner, V. Grabs, B. Haller, I. Hettich, D. Jochheim, J. Mehilli, R. Lange, S. Bleiziffer, and M. Halle, “Exercise training improves exercise capacity and quality of life after transcatheter aortic valve implantation: A randomized pilot trial.,” Am. Heart J., vol. 182, pp. 44–53, Dec. 2016.

[59] G. A. Wildenbos, M. W. M. Jaspers, M. P. Schijven, and L. W. Dusseljee-Peute, “Mobile health for older adult patients: Using an aging barriers framework to classify usability problems,” Int. J. Med. Inform., vol. 124, no. September 2018, pp. 68–77, 2019.

[60] H. G. H. Thyregod, “Five-Year Outcomes From the All-Comers Nordic Aortic Valve Intervention Randomized Clinical Trial in Patients with Severe Aortic Valve Stenosis,” Acc, 2018.

[61] M. F. Eleid and P. A. Pellikka, “Asymptomatic Severe Aortic Stenosis: What Are We Waiting For?∗,” J. Am. Coll. Cardiol., vol. 66, no. 25, pp. 2842–2843, 2015.

[62] M. Banovic, B. Iung, J. Bartunek, M. Asanin, B. Beleslin, B. Biocina, F. Casselman, M. da Costa, M. Deja, H. Gasparovic, P. Kala, L. Labrousse, Z. Loncar, J. Marinkovic, I. Nedeljkovic, M. Nedeljkovic, P. Nemec, S. D. Nikolic, M. Pencina, M. Penicka, A. Ristic, F. Sharif, G. Van Camp, M. Vanderheyden, W. Wojakowski, and S. Putnik, “Rationale and design of the Aortic Valve replAcemenT versus conservative treatment in Asymptomatic seveRe aortic stenosis (AVATAR trial): A randomized multicenter controlled event-driven trial,” Am. Heart J., vol. 174, pp. 147–153, 2016.

Page 81: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter 9

63

[63] M. Varnfield, M. K. Karunanithi, A. Särelä, E. Garcia, A. Fairfull, B. F. Oldenburg, and D. L. Walters, “Uptake of a technology-assisted home-care cardiac rehabilitation program,” Med. J. Aust., vol. 194, no. S4, pp. S15–S19, Feb. 2011.

[64] Philips Medical Systems, Research Kit - Instruction for use. Andover, MA, USA: Philips Medical Systems, 2017.

Page 82: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

64

Page 83: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-I

A.1. Overview of data collected during TELE-TAVI study

Figure A.1 Overview TAVI-patient collected data and missing data. Data loss in 206 T0, occurred as files were deleted. Raw ECG data was missing in 206-T1 and in 208-T1 several leads were missing from the reference data.

Chapter A: Appendix

201202203204205206207208209210211212213214215216217218219220221222223224

Pin

T0 T1 T2

Bios

enso

r

Mon

itor

Bios

enso

r

Bios

enso

r

Data available

Patient drop-out

Not measured

Data lost

Decryption error

Leads aVF and aVL interchanged

Page 84: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-II

Chapter A

A.2. Supplementary data on the reliability of the Philips wearable biosensorA.2.1. Distribution difference dataTable A.1 Skewness and kurtosis of the difference between the biosensor and reference, unfiltered and filtered for HR and RespR. Skewness and kurtosis of normal distribution is 0 and 3, respectively

HR HR filtered RespR RespR filtered

Pin s k s k s k s k

Heal

thy

101 4.4 35.9 4.4 37.8 0.6 3.7 -0.2 2.1

102 0.8 11.1 1.1 14.9 -0.9 5.3 -0.4 3.6

103 6.5 64.2 7.3 75.4 -0.4 3.4 0.9 3.6

104 2.3 15.2 3.8 27.7 -0.1 2.6 1.0 4.9

105 2.0 6.0 2.1 6.3 0.0 3.0 0.0 2.3

106 2.2 17.9 2.4 20.8 -0.3 3.8 0.3 6.2

TAVI

Sinu

s

201 15.3 344.4 18.2 434.4 -0.2 3.5 -0.5 4.0

204 3.6 38.6 3.8 37.2 -0.5 4.1 -0.6 4.4

211 -2.9 30.8 -1.6 21.1 * * * *

213 1.8 20.9 2.8 23.0 0.2 4.8 0.2 5.5

218 0.6 4.9 0.9 5.7 -1.5 6.7 -0.4 2.7

219 0.9 8.0 1.3 9.4 -1.5 11.3 -0.4 2.7

220 -8.9 138.8 -10.4 169.8 -1.1 8.7 -0.7 5.0

AF

203 1.6 11.7 2.2 14.6 * * * *

205 -0.1 3.2 0.0 3.3 * * * *

209 -1.6 6.8 -2.1 9.1 * * * *

215 -2.1 9.5 -2.1 8.9 * * * *

217 0.5 3.1 0.6 3.3 -0.4 4.8 0.2 3.3

221 0.6 5.5 0.7 6.2 -1.0 4.9 -0.9 5.1

Mean 1.4 40.9 1.9 48.9 -0.5 5.1 -0.1 4.0

SD 4.6 80.1 5.3 101.0 3.3 3.3 2.0 2.0

s indicates skewness; k, kurtosis; AF, atrial fibrillation; TAVI, transcatheter aortic valve implantation; HR, heart rate; RespR, respiratory rate, SD, standard deviation

* no data

Page 85: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-III

Figure A.2 Distribution of the difference between the biosensor and reference, unfiltered and filtered for HR and RespR, for all subjects. .

-50 0 500

0.05

0.1

0.15

0.2

0.25

0.3

0.35HR

Skewness = -0.2Kurtosis = 17.0

-50 0 500

0.05

0.1

0.15

0.2

0.25

0.3

0.35HR �ltered

Skewness = -0.2Kurtosis = 17.3

-20 0 20 40 600

0.05

0.1

0.15RespR

Skewness = 0.4Kurtosis = 5.8

-20 0 20 40 600

0.05

0.1

0.15Resp rate �ltered

Skewness = -0.0Kurtosis = 4.1

Normal distribution Normal distribution

Normal distribution Normal distribution

Den

sity

Den

sity

Den

sity

Den

sity

Reference - Biosensor Reference - Biosensor

Reference - BiosensorReference - Biosensor

Page 86: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-IV

Chapter A

A.2.2. Precision and bias of the wearable biosensorTable A.2 Bias and precision of the wearable biosensors HR compared to the reference monitor for all subjects.

Pin Mean reference

Mean biosensor Bias RPC RPC Lower

LOAUpper

LOA

Beats/min Beats/min Beats/min Beats/min (%) Beats/min Beats/min

Heal

thy

101 65.2 66.0 0.0 2.9 4.5 -2.9 2.9

102 75.0 76.0 0.9 3.2 4.3 -2.2 4.1

103 68.1 71.0 1.3 4.4 6.5 -3.1 5.6

104 66.0 66.0 0.0 0.0 0.0 0.0 0.0

105 70.0 75.0 1.0 11.3 17.0 -10.3 12.3

106 72.0 72.0 0.0 1.5 2.0 -1.5 1.5

Median 69.0 71.5 0.5 3.1 4.4 -2.6 3.5

IQR 6.0 9.0 1.0 2.9 4.5 1.6 4.2

TAVI

Sinu

s

201 55.0 55.0 0.0 0.0 0.0 0.0 0.0

204 52.0 52.0 0.0 1.5 2.7 -1.5 1.5

211 69.0 69.0 0.0 2.0 2.9 -2.0 2.0

213 57.0 57.0 0.0 1.0 1.7 -1.0 1.0

218 79.0 79.0 0.0 2.9 3.7 -2.9 2.9

219 64.0 65.0 0.0 1.5 2.6 -1.5 1.5

220 75.0 75.0 0.0 0.5 0.7 -0.5 0.5

Median 64.0 65.0 0.0 1.5 2.6 -1.5 1.5

IQR 18.0 18.0 0.0 1.2 1.9 1.2 1.2

AF

203 80.0 81.0 0.8 8.0 9.9 -7.1 8.8

205 110.3 124.0 14.0 11.8 10.7 2.2 25.8

209 96.0 101.0 3.8 16.1 16.6 -12.3 20.0

215 86.4 88.0 0.1 12.6 14.4 -12.5 12.7

217 99.0 102.0 3.0 7.7 7.9 -4.7 10.7

221 90.7 95.0 3.7 14.7 16.0 -11.0 18.4

Median 93.3 98.0 3.3 12.2 12.6 -8.8 15.5

IQR 12.6 14.0 3.0 6.7 6.1 7.5 9.2

Median 79.0 79.0 0.0 2.9 3.7 -2.9 2.9

IQR 29.7 33.5 3.2 10.7 9.3 7.2 12.8

Median 75.2 77.3 1.5 5.4 6.5 -3.9 7.0

IQR 15.5 18.1 3.3 5.4 5.9 4.5 7.7

IQR indicates inter quartile range; RPC, reproducibility coefficient (1.45*IQR, 1.96*SD); LOA, limit of agreement (1.45*IQR, 1.96*SD)

Page 87: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-V

Table A.3 Bias and precision of the wearable biosensors HR filtered compared to the reference monitor for all subjects.

Pin Mean reference

Mean biosensor Bias RPC RPC Lower

LOAUpper

LOA

Beats/min Beats/min Beats/min Beats/min (%) Beats/min Beats/min

Heal

thy

101 65.6 66.0 0.2 2.2 3.4 -2.0 2.4

102 75.2 76.0 0.7 2.6 3.6 -1.9 3.4

103 68.3 71.0 1.4 3.9 5.9 -2.5 5.3

104 66.4 66.0 0.0 1.0 1.5 -1.0 1.0

105 70.7 75.0 0.7 10.3 15.4 -9.6 11.1

106 71.7 72.0 0.3 1.5 2.1 -1.2 1.8

Median 69.5 71.5 0.5 2.4 3.5 -1.9 2.9

IQR 5.3 9.0 0.5 2.4 3.8 1.3 3.6

TAVI

Sinu

s

201 55.1 55.0 0.0 0.8 1.4 -0.8 0.8

204 52.2 52.0 0.0 1.3 2.4 -1.3 1.3

211 69.0 69.0 0.0 1.3 1.8 -1.3 1.3

213 56.7 57.0 0.1 0.8 1.5 -0.8 0.9

218 79.4 79.0 0.0 1.8 2.3 -1.8 1.9

219 64.4 65.0 0.2 1.0 1.7 -0.8 1.1

220 75.3 75.0 0.0 0.9 1.2 -0.9 0.9

Median 64.4 65.0 0.0 1.0 1.7 -0.9 1.0

IQR 18.3 18.0 0.1 0.4 0.7 0.5 0.4

AF

203 80.4 81.0 0.5 6.5 8.1 -6.0 7.0

205 110.4 124.0 13.9 9.3 8.5 4.6 23.2

209 96.3 101.0 4.1 12.5 12.9 -8.5 16.6

215 86.4 88.0 0.6 11.5 13.2 -11.0 12.1

217 98.9 102.0 3.0 5.8 5.9 -2.8 8.9

221 90.7 95.0 3.4 13.6 14.8 -10.2 17.0

Median 93.5 98.0 3.2 10.4 10.7 -7.2 13.6

IQR 12.5 14.0 3.5 6.0 5.1 7.5 8.2

Median 79.4 79.0 0.2 1.8 2.4 -1.6 2.0

IQR 29.6 33.5 3.1 8.9 7.9 5.8 12.1

Median 75.4 77.3 1.5 4.7 5.7 -3.1 6.2

IQR 15.5 18.1 3.2 4.5 5.0 4.0 6.8

IQR indicates inter quartile range; RPC, reproducibility coefficient (1.45*IQR, 1.96*SD); LOA, limit of agreement (1.45*IQR, 1.96*SD)

Page 88: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-VI

Chapter A

Table A.4 Bias and precision of the wearable biosensors RespR compared to the reference monitor for all subjects.

Pin Mean reference

Mean biosensor Bias RPC RPC Lower

LOAUpper

LOA

Breaths/min

Breaths/min

Breaths/min

Breaths/min (%) Breaths/

minBreaths/

min

Heal

thy

101 15.5 13.5 -1.9 7.8 54.9 -9.7 6.0

102 15.9 13.4 -2.5 8.1 44.6 -10.6 5.6

103 15.5 13.7 -1.7 6.9 40.7 -8.6 5.2

104 17.0 16.3 -0.7 8.7 59.8 -9.4 8.0

105 16.9 14.9 -2.1 10.7 64.4 -12.7 8.6

106 16.2 16.2 -0.1 6.3 39.2 -6.4 6.2

Median 16.2 14.7 -1.5 8.1 50.6 -9.6 6.6

SD 0.7 1.3 0.9 1.5 10.6 2.1 1.4

TAVI

Sinu

s

201 19.3 18.2 -1.0 6.1 32.7 -7.1 5.0

204 16.4 15.9 -0.6 5.7 35.0 -6.2 5.1

211 16.4 17.1 0.8 7.4 78.1 -6.6 8.1

218 14.7 15.6 1.0 7.4 49.2 -6.5 8.4

219 14.3 14.5 0.0 9.9 81.8 -9.8 9.9

220 15.9 15.7 -0.1 8.2 50.2 -8.4 8.1

Median 16.2 16.2 0.0 7.4 54.5 -7.4 7.4

SD 1.8 1.3 0.8 1.5 21.0 1.4 2.0

AF

217 17.6 21.2 3.6 9.0 75.5 -5.4 12.5

221 18.0 20.9 2.9 9.2 55.6 -6.3 12.2

Median 17.8 21.1 3.2 9.1 65.6 -5.9 12.3

SD 0.3 0.2 0.5 0.2 14.1 0.6 0.3

Median 16.6 17.4 0.8 7.9 57.3 -7.0 8.7

SD 1.7 2.5 1.6 1.5 19.2 1.4 2.8

Median 16.4 16.2 -0.2 7.9 54.4 -8.1 7.8

SD 1.3 2.5 1.8 1.5 15.9 2.1 2.5

ISD indicates standard deviation; RPC, reproducibility coefficient (1.45*IQR, 1.96*SD); LOA, limit of agreement (1.45*IQR, 1.96*SD)

Page 89: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-VII

Table A.5 Bias and precision of the wearable biosensors RespR filtered compared to the reference monitor for all subjects.

Pin Mean reference

Mean biosensor Bias RPC RPC Lower

LOAUpper

LOA

Breaths/min

Breaths/min

Breaths/min

Breaths/min (%) Breaths/

minBreaths/

min

Heal

thy

101 15.8 13.5 -1.9 4.8 34.5 -6.8 2.9

102 15.9 13.4 -2.5 5.8 35.2 -8.3 3.2

103 15.3 13.7 -1.6 4.5 29.6 -6.1 3.0

104 17.1 16.3 -0.8 7.6 60.9 -8.4 6.8

105 17.6 14.9 -2.1 10.2 54.9 -12.3 8.1

106 16.1 16.2 0.0 3.9 25.5 -4.0 3.9

Median 16.3 14.7 -1.5 6.2 40.1 -7.7 4.6

SD 0.9 1.3 0.9 2.4 14.3 2.8 2.2

TAVI

Sinu

s

201 19.3 18.2 -1.0 5.0 25.9 -6.0 4.0

204 16.4 15.9 -0.5 4.2 25.0 -4.7 3.6

211 16.6 17.1 0.5 6.2 55.3 -5.7 6.7

218 14.7 15.6 0.9 6.2 45.5 -5.3 7.2

219 14.2 14.5 0.0 7.5 56.1 -7.5 7.5

220 15.9 15.7 -0.1 6.9 42.4 -7.1 6.8

Median 16.2 16.2 0.0 6.0 41.7 -6.0 6.0

SD 1.8 1.3 0.7 1.2 13.7 1.1 1.7

AF

217 17.8 21.2 3.4 7.3 54.4 -3.9 10.7

221 18.0 20.9 2.9 7.9 47.1 -5.0 10.8

Median 17.9 21.1 3.1 7.6 50.7 -4.5 10.7

SD 0.2 0.2 0.4 0.4 5.2 0.8 0.0

Median 16.6 17.4 0.8 6.4 43.9 -5.7 7.2

SD 1.7 2.5 1.6 1.3 12.5 1.2 2.6

Median 16.5 16.2 -0.2 6.3 42.3 -6.5 6.1

SD 1.4 2.5 1.7 1.7 12.9 2.2 2.7

SD indicates standard deviation; RPC, reproducibility coefficient (1.45*IQR, 1.96*SD); LOA, limit of agreement (1.45*IQR, 1.96*SD)

Page 90: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-VIII

Chapter A

Figure A.3 Bland-Altman figures comparing HR from the biosensor to the reference monitor (unfiltered on the left, filtered on the right), for healthy subjects (top panels), TAVI patients with sinus rhythm (middle panels) and AF (bottom panels).

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80

HR

Bios

enso

r-Re

fere

nce

3.50.5

-2.6

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80

HR

Bios

enso

r-Re

fere

nce

1.50.0

-1.5

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80

HR

Bios

enso

r-Re

fere

nce

15.5

3.3

-8.8

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80

HR

Bios

enso

r-Re

fere

nce

2.90.5

-1.9

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80H

R Bi

osen

sor-

Refe

renc

e

1.00.0

-0.9

50 100 150HR Reference

-80

-60

-40

-20

0

20

40

60

80

HR

Bios

enso

r-Re

fere

nce

13.6

3.2-7.2

HR HR �ltered

Hea

lthy

subj

ects

TAVI

pat

ient

s: s

inus

rhyt

hmTA

VI p

atie

nts:

AF

Page 91: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-IX

Figure A.4 Bland-Altman figures comparing RespR from the biosensor to the reference monitor (unfiltered on the left, filtered on the right), for healthy subjects (top panels), TAVI patients with sinus rhythm (middle panels) and AF (bottom panels).

RespR RespR �ltered

Hea

lthy

subj

ects

TAVI

pat

ient

s: s

inus

rhyt

hmTA

VI p

atie

nts:

AF

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60Re

spR

Bios

enso

r-Re

fere

nce

6.59-1.48-9.55

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60

Resp

R Bi

osen

sor-

Refe

renc

e

4.6-1.5-7.7

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60

Resp

R Bi

osen

sor-

Refe

renc

e

4.6-1.5-7.7

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60Re

spR

Bios

enso

r-Re

fere

nce

6.0-0.0-6.0

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60

Resp

R Bi

osen

sor-

Refe

renc

e

12.3

3.2

-5.9

0 20 40 60 80RespR Reference

-60

-40

-20

0

20

40

60

Resp

R Bi

osen

sor-

Refe

renc

e

10.7

3.1

-4.5

Page 92: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-X

Chapter A

A.2.3. Walking detection of the wearable biosensor

Table A.10 Detection of walking by the wearable biosensor during the walking exercises of the healthy subjects.

100 steps/min 75 steps/min 50 steps/min

Percentage walking Speed Percentage

walking Speed Percentage walking Speed

Pin (%) (m/s) (%) (m/s) (%) (m/s)

101 100,0 1.25 95,5 0.83 1,8 0.48

102 96,2 1.19 98,1 0.82 0,0 0.49

103 100,0 1.11 97,2 0.81 2,8 0.22

104 100,0 1.10 78,3 0.77 0,0 0.47

105 100,0 1.29 71,2 0.82 0,0 0.52

106 100,0 1.20 8,6 0.60 0,0 *

Mean 99,4 1.19 74,8 0.77 0,8 0.43

SD 1,5 0.08 34,3 0.09 1,2 0.12

SD indicates standard deviation; min, minute, m/s, meter per second

* no data available

Page 93: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XI

Table A.6 Detection of walking by the wearable biosensor during the walking exercises of the healthy subjects.

Walking detection Speed

Pin (%) (m/s)

Heal

thy

101 100.0 1.60

102 99.4 1.55

103 100.0 1.41

104 99.1 1.70

105 100.0 1.44

106 100.0 1.74

Mean 99.8 1.57

SD 0.4 0.13

TAVI

201 T0 91.6 1.41

201 T2 100.0 1.50

202 T0 98.3 1.21

203 T0 91.7 0.90

203 T2 92.9 0.79

204 T0 91.7 0.91

204 T2 100.0 1.21

207 T0 100.0 1.45

207 T2 100.0 1.46

208 T0 93.2 1.06

209 T0 73.5 0.60

209 T2 97.7 1.19

210 T0 99.7 1.06

211 T0 100.0 1.06

212 T0 100.0 1.39

213 T0 100.0 1.17

215 T0 99.1 1.14

215 T2 100.0 1.25

216 T0 85.0 0.71

217 T0 100.0 1.13

218 T0 100.0 1.19

219 T0 97.4 0.94

221 T0 100.0 1.38

222 T0 98.6 0.89

223 T0 99.4 1.24

224 T2 99.4 1.64

Mean 96.5 1.15

SD 6.1 0.26

Mean 97.1 1.23

SD 5.6 0.29

SD indicates standard deviation; m/s, meter per second

* no data available

Page 94: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XII

Chapter A

A.2.4. Measurement length

Figure A.5 Gives the Kaplan-Meier figure for the survival to first data loss, with gaps of 2 minutes, 15 minutes, 1 hour and 4 hours. Every plus sign represents the end of a measurement. Under the figure, the number at risk are shown, every 10 hours.

0 10 20 30 40 50 60 70 80 90

Time (hour)

0

0.2

0.4

0.6

0.8

1

Surv

ival

to d

ata

loss

Kaplan-Meier

Length of data loss(hh:mm)

Length ofdata loss(hh:mm)

04:0001:0000:1500:02

Number recording:

29292929

26252525

20191716

17161513

16151512

15141411

151413 9

111111 7

9996

7775

04:0001:0000:1500:02

Page 95: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XIII

Figure A.6 Gives survival of the measurement of the wearable biosensor in the TELE-TAVI study. Under the figure, the number at risk are shown, every 10 hours.

0 10 20 30 40 50 60 70 80 90

Time (hour)

0

0.2

0.4

0.6

0.8

1

Surv

ival

mea

sure

men

t len

gth

Kaplan-Meier

Number recording:

31 30 26 23 21 20 19 15 10 9

Page 96: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XIV

Chapter A

Table A.7 Measurement length, data loss and usable data for all subjects. Detection of walking by the wearable biosensor during the walking exercises of the healthy subjects.

Pin Length mea-surement Data loss Data not lost Usable data Usable data

hh:mm:ss hh:mm:ss (%) hh:mm:ss (%)

Heal

thy

101 95:55:39 00:47:29 99.2 95:08:10 99.1

102 05:52:23 00:01:57 99.4 05:50:26 6.1

103 46:10:51 00:04:59 99.8 46:05:52 48.0

104 95:56:12 00:00:12 100.0 95:55:59 99.9

105 40:35:40 00:01:09 100.0 40:34:30 42.3

106 91:28:13 39:34:20 56.7 51:53:52 54.1

Mean 62:39:50 06:45:01 92.5 55:54:48 58.2

SD 37:29:21 16:04:56 17.5 34:37:36 36.1

TAVI

201 T0 92:14:32 00:01:36 100.0 92:12:56 96.1

201 T2 91:39:04 00:11:44 99.8 91:27:19 95.3

202 T0 75:02:01 00:11:38 99.7 74:50:23 78.0

203 T0 95:54:26 50:07:03 47.9 45:47:22 47.7

203 T2 93:27:29 00:00:13 100.0 93:27:16 97.3

204 T0 66:23:06 00:02:21 99.9 66:20:44 69.1

204 T2 58:13:13 20:27:23 64.5 37:45:49 39.3

205 T0 78:11:44 00:02:50 99.9 78:08:54 81.4

205 T2 43:33:33 00:00:04 100.0 43:33:28 45.4

207 T0 70:09:22 48:10:39 31.1 21:58:43 22.9

207 T0 71:05:05 11:56:04 83.0 59:09:00 61.6

208 T0 11:12:10 00:04:25 99.3 11:07:45 11.6

209 T0 35:49:16 01:26:56 96.0 34:22:19 35.8

209 T2 05:47:52 00:00:04 100.0 05:47:48 6.0

210 T0 64:16:00 00:01:07 100.0 64:14:53 66.9

211 T0 92:27:50 00:26:52 99.5 92:00:58 95.9

212 T0 26:50:52 01:59:21 92.6 24:51:31 25.9

213 T0 21:39:36 00:07:09 99.4 21:32:26 22.4

214 T0 10:35:13 00:01:14 99.8 10:33:59 11.0

215 T0 74:14:33 59:14:40 20.3 14:59:53 15.6

215 T2 93:34:31 04:30:38 95.3 89:03:53 92.8

216 T0 67:48:32 00:37:47 99.1 67:10:45 70.0

217 T0 18:21:17 00:00:03 100.0 18:21:13 19.1

218 T0 92:04:21 00:01:36 100.0 92:02:44 95.9

219 T0 91:31:47 00:01:33 100.0 91:30:13 95.3

221 T0 10:51:45 00:00:00 100.0 10:51:45 11.3

222 T0 91:31:13 00:00:15 100.0 91:30:57 95.3

223 T0 23:18:07 01:06:52 95.1 22:11:14 23.1

224 T0 86:31:24 00:00:07 100.0 86:31:17 90.1

Mean 60:29:39 06:55:36 90.4 53:34:03 55.8

SD 31:41:52 16:24:38 21.4 32:22:33 33.7

Mean 60:51:58 06:53:47 90.8 53:58:11 56.2

SD 32:10:07 16:07:08 20.6 32:15:16 33.6

hh:mm:ss indicates hours:minutes:seconds

Page 97: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XV

Table A.8 First point of data loss for every subject in the TELE-TAVI study.

Pin Point of first data loss

2 min 15 min 1 hour 4 hoursHe

alth

y101 17:06:11 17:06:11

102

103

104

105

106 10:17:14 10:17:14 10:17:14 10:17:14

TAVI

201 T0

201 T2 31:14:41

202 T0 28:16:10

203 T0 09:35:51 09:35:51 09:35:51 09:35:51

203 T2

204 T0

204 T2 16:52:49 16:52:49 16:52:49 16:52:49

205 T0

205 T2

207 T0 01:51:06 02:52:50 02:52:50 02:52:50

207 T0 04:01:19 04:01:19 04:01:19 61:07:28

208 T0

209 T0 14:34:01 14:34:01

209 T2

210 T0

211 T0 28:48:31

212 T0 13:24:47 14:39:51

213 T0 18:35:13

214 T0

215 T0 11:59:37 11:59:37 11:59:37 11:59:37

215 T2 53:06:42 89:15:37 89:15:37 89:15:37

216 T0 58:24:58 58:24:58

217 T0

218 T0

219 T0

221 T0

222 T0

223 T0 22:09:26 22:14:44

224 T0

hh:mm:ss indicates hours:minutes:seconds

Page 98: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XVI

Chapter A

A.3. Supplementary data of the wearability questionnairesA.3.1. Wearability questionnaires

As given to patients (in Dutch)

Pilot proefpersonen meting: draadloze monitoring in TAVI patiënten, V1.1, 30-05-2018

Evaluatie systeem en sensor

Op de volgende bladzijde gaat u een vragenlijst invullen over het gebruik van het systeem; de draagbare biosensor, met bijbehorende telefoon en app. Deze vragenlijst moet ingevuld worden door diegene die het meest het systeem heeft gebruikt. Deze vragenlijst bestaat uit 16 stellingen waarbij U kunt aangeven of deze stelling overeenkomt. U kunt kiezen uit zeven opties; van helemaal eens (1) tot helemaal niet mee eens (7) of niet van toepassing (N.v.t.). Denk vooral niet te lang na over de antwoorden.

Hierna volgen nog eens 9 vragen over het comfort tijdens het dragen en het vervangen van de sensor.

Deze vragenlijsten zijn ingevuld door (aankruisen wat van toepassing is):

□ Patiënt/proefpersoon □ Mantelzorger □ Anders, namelijk:

Page 99: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XVII

Pilot proefpersonen meting: draadloze monitoring in TAVI patiënten, V1.1, 30-05-2018

Evaluatie systeem

1 2 3 4 5 6 7 N.v.t.

1 Over het algemeen, ben ik tevreden met het gebruiksgemak.

○ ○ ○ ○ ○ ○ ○ ○

2 Het was eenvoudig om het systeem te gebruiken.

○ ○ ○ ○ ○ ○ ○ ○

3 Ik kon de taken en scenario’s snel afmaken in de app.

○ ○ ○ ○ ○ ○ ○ ○

4 Ik voelde me op mijn gemak tijdens het gebruik van het systeem.

○ ○ ○ ○ ○ ○ ○ ○

5 Het was gemakkelijk om te leren hoe ik het systeem moest gebruiken.

○ ○ ○ ○ ○ ○ ○ ○

6 Ik denk dat ik met dit systeem snel in de praktijk kan werken.

○ ○ ○ ○ ○ ○ ○ ○

7 Ik kreeg foutmeldingen waaruit bleek hoe ik een probleem kon oplossen.

○ ○ ○ ○ ○ ○ ○ ○

8 Als ik een foutje maakte, kon ik dit snel en gemakkelijk recht zetten.

○ ○ ○ ○ ○ ○ ○ ○

9 De informatie (zoals de berichten op het scherm en de handleiding) was duidelijk.

○ ○ ○ ○ ○ ○ ○ ○

10 Het was gemakkelijk om de juiste informatie te vinden.

○ ○ ○ ○ ○ ○ ○ ○

11 De informatie die ik op het scherm kreeg hielp bij het gebruiken van de app en sensor.

○ ○ ○ ○ ○ ○ ○ ○

12 De indeling van de informatie op het scherm was duidelijk.

○ ○ ○ ○ ○ ○ ○ ○

13 De interface zag er aangenaam uit (Met de interface wordt bedoeld alle onderwerpen die je kunt zien die nodig zijn om het platform te gebruiken. Bijvoorbeeld: het scherm, de plaatjes en de taal.)

○ ○ ○ ○ ○ ○ ○ ○

14 Ik vond het fijn om de interface te gebruiken (invullen van gegevens op de diverse schermen).

○ ○ ○ ○ ○ ○ ○ ○

15 Dit systeem heeft alle functies en mogelijkheden die ik ervan verwacht.

○ ○ ○ ○ ○ ○ ○ ○

16 Over het algemeen, ben ik tevreden met het systeem.

○ ○ ○ ○ ○ ○ ○ ○

Helemaal niet mee eens Helemaal eens

Page 100: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XVIII

Chapter A

Pilot proefpersonen meting: draadloze monitoring in TAVI patiënten, V1.1, 30-05-2018

Evaluatie sensor

1. Hoe was het dragen van de biosensor?

Comfortabel Oncomfortabel N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

2. Hebt u last gehad van jeuk door de biosensor?

Geen jeuk Veel jeuk N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

3. Hebt u last gehad van huid irritatie?

Geen irritatie Veel irritatie N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

4. Bleef de biosensor goed op de huid vastplakken?

Bleef volledig plakken Is losgekomen N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

5. Kwam de biosensor gemakkelijk los van de huid?

Gemakkelijk Veel moeite N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

6. Was het verwijderen van de sensor pijnlijk?

Niet pijnlijk Erg pijnlijk N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

7. Was u beperkt in uw activiteiten door de meting?

Niet beperkt Erg beperkt N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

8. Had u de telefoon altijd bij u? Altijd Nooit N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

9. Het verwisselen van de sensor was gemakkelijk.

Helemaal eens Helemaal oneens N.v.t.

○ ○ ○ ○ ○ ○ ○ ○

Page 101: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XIX

Pilot proefpersonen meting: draadloze monitoring in TAVI patiënten, V1.1, 30-05-2018

Opmerkingen, aanvullingen en praktische beperkingen:

Page 102: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XX

Chapter A

A.3.2. Results from PSSUQ for every study participantTable A.11 PSSUQ results for every study participant

Pin Overall SysUse InfoQual IntQual Mean SD

201 T0 85.6 88.9 80.6 88.9 86.1 4.8

201 T1 96.4 94.4 97.2 100.0 97.2 2.8

201 T2 90.5 94.4 88.9 83.3 88.9 5.6

202 T0 55.6 75.0 0.0 33.3 36.1 37.6

202 T1 55.6 75.0 0.0 33.3 36.1 37.6

203 T0 84.7 86.7 83.3 83.3 84.4 1.9

203 T1 83.3 83.3 83.3 83.3 83.3 0.0

203 T2 92.9 93.3 91.7 94.4 93.1 1.4

204 T0 57.1 53.3 83.3 50.0 62.2 18.4

204 T1 84.4 80.6 88.9 83.3 84.3 4.2

204 T2 46.7 44.4 33.3 66.7 48.1 17.0

205 T1 88.9 91.7 83.3 87.5 5.9

205 T2 66.7 66.7 66.7 0.0

206 T1

207 T0 15.6 25.0 13.9 4.2 14.4 10.4

207 T1 44.8 33.3 52.8 50.0 45.4 10.5

207 T2 26.0 25.0 33.3 16.7 25.0 8.3

208 T0 12.5 16.7 0.0 25.0 13.9 12.7

209 T0 38.1 33.3 38.9 44.4 38.9 5.6

209 T1 50.0 60.0 40.0 50.0 50.0 10.0

209 T2 37.8 44.4 30.6 38.9 38.0 7.0

211 T0 40.6 38.9 47.2 33.3 39.8 7.0

212 T0 73.8 75.0 75.0 66.7 72.2 4.8

213 T0 87.2 86.7 87.5 87.5 87.2 0.5

213 T1 94.8 100.0 91.7 91.7 94.4 4.8

213 T2 100.0 100.0 100.0 100.0 100.0 0.0

214 T0

215 T0 40.9 46.7 38.9 33.3 39.6 6.7

215 T1

215 T2 91.1 100.0 77.8 100.0 92.6 12.8

216 T0 100.0 100.0 100.0 100.0 100.0 0.0

217 T0 37.8 66.7 16.7 22.2 35.2 27.4

217 T1 2.8 2.8 2.8 0.0

218 T0 97.8 100.0 94.4 100.0 98.1 3.2

218 T1 4.8 10.0 0.0 4.2 4.7 5.0

219 T0 77.1 91.7 83.3 45.8 73.6 24.4

219 T1 16.7 16.7 16.7 16.7 16.7 0.0

220 T0 26.4 16.7 20.0 50.0 28.9 18.4

220 T1 45.5 20.0 66.7 43.3 33.0

221 T0 88.5 94.4 88.9 79.2 87.5 7.7

221 T1 34.4 27.8 41.7 33.3 34.3 7.0

222 T0 100.0 100.0 100.0 100.0 100.0 0.0

222 T1 100.0 100.0 100.0 100.0 100.0 0.0

223 T0 81.0 72.2 100.0 50.0 74.1 25.1

224 T0 100.0 100.0 100.0 100.0 100.0 0.0

Mean 63.2 65.0 61.7 61.7 62.8 9.3

SD 30.7 31.9 34.7 31.5 32.7 10.4

Overall system use (SysUse), information quality (InfoQual) and interface quality (IntQuality), SD indi-cates standard deviation

Page 103: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XXI

A.3.3. Results custom made questionnaire for every measurement moment

Figure A.7 Results of custom made questionnaire for T0, T1 and T2.

How was wearing the biosensor?

Did you experience itchiness caused by the biosensor?

Did you experience skin irritation?

Did the biosensor remained adhered to the skin?

Was the biosensor easily removable from the skin?

Was removing the biosensor painful?

Were you activities limited because of the measurement?

Did have to phone with you?

Changing the biosensor was easy.

Answers

Uncomfortable

A lot of itch

Irritation

Detached

Great di�culty

Painful

Hinderence

Always

Totally agree

Easy

Comfortable

No itch

No irritation

Adhered

Not painful

No hinderence

Never

Totally disagree

79%

70%

74%

74%

86%

81%

49%

PositiveNegative

How was wearing the biosensor?

Did you experience itchiness caused by the biosensor?

Did you experience skin irritation?

Did the biosensor remained adhered to the skin?

Was the biosensor easily removable from the skin?

Was removing the biosensor painful?

Were you activities limited because of the measurement?

Did have to phone with you?

Changing the biosensor was easy.

Answers

Uncomfortable

A lot of itch

Irritation

Detached

Great di�culty

Painful

Hinderence

Always

Totally agree

Easy

Comfortable

No itch

No irritation

Adhered

Not painful

No hinderence

Never

Totally disagree

79%

70%

74%

74%

86%

81%

49%

PositiveNegative

How was wearing the biosensor?

Did you experience itchiness caused by the biosensor?

Did you experience skin irritation?

Did the biosensor remained adhered to the skin?

Was the biosensor easily removable from the skin?

Was removing the biosensor painful?

Were you activities limited because of the measurement?

Did have to phone with you?

Changing the biosensor was easy.

Answers

Uncomfortable

A lot of itch

Irritation

Detached

Great di�culty

Painful

Hinderence

Always

Totally agree

Easy

Comfortable

No itch

No irritation

Adhered

Not painful

No hinderence

Never

Totally disagree

79%

70%

74%

74%

86%

81%

49%

PositiveNegative

44%

19% 19% 63%

75%

19% 19% 56%

19% 56%

73%

88%

81%

31% 44%

1 2 3 4 5 6 7

26%

95%

16% 63%

84%

84%

47%

89%

84%

32% 53%

47%

74%

Not applicable

NA

74%

Not applicable

NA

19%

29%

25% 75%

25% 75%

88%

88%

63%

75%

75%

25% 50%

Not applicable

NA

57%

T0 - Pre-TAVIn = 21

T1 - Direct post-TAVIn = 16

T2 - Six week post-TAVIn = 8 1 2 3 4 5 6 7

1 2 3 4 5 6 7

Page 104: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

A-XXII

Chapter A

A.4. Supplementary data of the activity resultsA.4.1. Results of activity analysisTable A.9 IMA and activity characteristics of a full measurement, for all study participants.

IMA Activity levels

Pin Mean SD Range Empty No Low Medium High

(%) (%) (%) (%) (%)

Heal

thy

101 44.4 72.2 3.6 767.4 1.1 55.2 31.2 12.2 1.4

102 85.3 119.2 5.5 1214.5 0.8 38.4 30.8 24.5 6.3

103 35.2 54.6 3.4 394.1 0.6 64.4 24.0 10.9 0.7

104 45.1 68.4 3.6 588.0 0.1 57.7 28.1 11.0 3.2

105 47.3 61.3 3.6 806.0 0.2 51.3 32.5 15.0 1.2

106 45.7 63.9 3.6 453.1 43.8 59.7 22.4 16.5 1.4

Mean 50.5 73.3 3.9 703.9 7.8 54.5 28.2 15.0 2.4

SD 17.6 0.8 299.2 17.7 9.0 4.1 5.2 2.1

TAVI

201 T0 31.1 48.8 4.0 359.3 0.1 71.6 17.2 11.0 0.1

201 T2 34.4 48.3 3.6 347.6 0.3 66.9 20.5 12.5 0.1

202 T0 34.5 43.2 3.7 310.9 0.4 61.8 27.8 10.4 0.1

203 T0 28.9 41.3 4.0 313.0 52.2 68.6 22.5 8.7 0.1

203 T2 20.0 30.4 3.9 333.4 0.1 79.0 16.4 4.5 0.0

204 T0 20.7 29.9 3.1 305.3 0.3 75.9 20.4 3.6 0.1

204 T2 22.8 34.6 3.5 279.7 39.5 75.9 18.0 6.1 0.0

207 T0 35.3 46.3 4.5 312.1 69.0 60.7 28.3 10.9 0.1

207 T2 39.8 50.4 3.3 332.8 17.1 57.6 28.2 14.1 0.1

208 T0 38.6 44.7 4.4 297.6 0.8 54.1 31.8 14.0 0.1

209 T0 26.1 31.0 3.7 269.8 4.4 63.9 30.9 5.3 0.0

209 T2 44.9 50.6 5.4 384.4 0.3 48.0 37.0 15.0 0.0

210 T0 42.2 57.3 3.5 497.0 0.1 57.2 29.0 13.1 0.8

211 T0 31.2 41.8 3.4 495.6 0.7 64.4 26.7 8.7 0.2

212 T0 41.2 60.0 3.8 364.3 7.5 66.1 17.6 15.5 0.8

213 T0 26.5 43.1 3.7 402.3 0.7 72.5 19.1 8.3 0.1

215 T0 41.6 51.3 3.9 344.0 79.7 55.1 29.2 15.6 0.1

215 T2 32.2 45.8 3.6 370.5 4.8 70.5 17.7 11.6 0.2

216 T0 19.7 29.9 3.8 310.4 1.0 80.8 14.2 5.0 0.0

217 T0 22.0 34.2 4.7 295.2 0.1 78.8 15.7 5.5 0.0

218 T0 30.5 43.2 4.0 355.0 0.1 65.7 24.8 9.3 0.1

219 T0 20.5 30.0 4.0 334.4 0.1 76.8 18.6 4.6 0.0

221 T0 46.7 52.3 4.8 323.8 0.0 46.4 38.2 15.2 0.1

222 T0 29.8 40.7 3.6 377.1 0.1 65.9 25.7 8.3 0.1

223 T0 35.9 43.6 3.6 347.1 5.1 58.3 30.4 11.2 0.1

224 T0 40.9 53.9 3.8 392.8 0.0 55.2 32.6 11.0 1.2

Mean 32.2 43.3 3.9 348.3 10.9 65.3 24.6 10.0 0.2

SD 8.3 0.5 55.3 22.5 9.6 6.8 3.8 0.3

Mean 35.7 48.9 3.9 415.0 10.3 63.3 25.2 10.9 0.6

SD 12.5 0.6 191.8 21.5 10.3 6.5 4.5 1.2

TAVI indicates transcatheter aortic valve implantation; IMA, integrated modulus of acceleration; SD, standard deviation; T, measurement at T0, T1, or T2.

Page 105: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

Chapter A

A-XXIII

A.4.2. Results of daily activity

Table A.12 IMA and activity characteristics of full days, for all study participants.

IMA Activity levels

Pin Days Mean SD No Low Medium High

(%) (%) (%) (%)

Heal

thy

101 3 48.0 1.9 34.9 32.7 12.8 2.0

103 2 49.2 0.7 27.1 33.7 17.5 1.0

104 3 45.1 0.1 35.9 29.8 11.4 3.3

105 1 67.2 0.1 29.5 46.1 22.6 1.9

Mean 2.3 52.4 0.7 31.8 35.6 16.1 2.1

SD 1.0 10.1 4.3 7.2 5.1 1.0

TAVI

201 0 3 41.2 0.1 42.2 23.6 16.4 0.2

201 2 3 38.0 0.6 42.0 25.3 14.7 0.1

202 0 2 36.1 0.6 34.9 28.7 12.7 0.1

203 2 3 20.4 0.0 58.6 18.9 5.0 0.0

204 0 2 24.4 0.3 42.0 29.1 4.4 0.0

207 2 1 49.0 0.0 47.2 33.8 18.8 0.2

210 0 2 43.9 0.1 31.5 29.6 14.6 0.9

211 0 3 30.8 0.5 44.1 27.0 9.3 0.1

215 0 3 31.8 0.4 50.8 17.8 12.4 0.1

216 0 2 22.0 0.0 51.0 19.6 6.0 0.0

218 0 3 34.0 0.2 40.0 30.2 11.4 0.0

219 2 3 26.6 0.1 49.5 24.6 8.1 0.1

222 0 3 31.2 0.1 42.8 29.6 9.4 0.1

224 0 3 44.0 0.0 30.2 37.4 12.6 1.4

Mean 2.6 33.8 0.2 43.3 26.8 11.1 0.2

SD 0.6 8.7 7.8 5.6 4.3 0.4

Mean 2.5 37.9 0.3 40.8 28.7 12.2 0.6

SD 0.7 11.8 8.6 6.9 4.8 0.9

TAVI indicates transcatheter aortic valve implantation; IMA, integrated modulus of acceleration; SD, standard deviation; T, measurement at T0, T1, or T2..

Page 106: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 107: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve
Page 108: The Philips wearable biosensor in transcatheter aortic ...essay.utwente.nl/77157/1/Braem_MA_TNW.pdf · Before you lies ‘The Philips wearable biosensor in transcatheter aortic valve

C. I. R. Braem, BSc.

The Philips wearable biosensor in transcatheter aortic valve implantation

treatment workflow

March 22nd, 2019

Usability and feasibility of the wearable biosensor