Upload
others
View
8
Download
0
Embed Size (px)
Citation preview
55
The Measurement of Temperature
Temperature is an intuitive concept that tells whether a body is “hot” or
“cold.” The two temperature scales are the Fahrenheit and Celsius scales
The relation between scales can be shown in figure (1).
Fig.(1) Relationship between Fahrenheit and Celsius temperature scales
Temperature Measurement by Mechanical Effects
The liquid-in-glass thermometer is one of the most common types of
temperature measurement devices. The construction details of such an
instrument are shown in Figure (2). A relatively large bulb at the lower
portion of the thermometer holds the major portion of the liquid, which
expands when heated and rises in the capillary tube, upon which are
etched appropriate scale markings.
55
Fig.(2) Schematic of a mercury-in-glass thermometer
Mercury-in-glass thermometers are generally applicable up to about
600◦F (315◦C), but their range may be extended to 1000◦F (538◦C) by
filling the space above the mercury with a gas like nitrogen. This
increases the pressure on the mercury, raises its boiling point, and thereby
permits the use of the thermometer at higher temperatures.
A very widely used method of temperature measurement is the bimetallic
strip. Two pieces of metal with different coefficients of thermal
expansion are bonded together to form the device shown in Fig.(3). When
the strip is subjected to a temperature higher than the bonding
temperature, it will bend in one direction; when it is subjected to a
temperature lower than the bonding temperature, it will bend in the other
direction. Eskin and Fritze have given calculation methods for bimetallic
strips. The radius of curvature r may be calculated as:
55
(1)
where
t = combined thickness of the bonded strip, m or ft
m = ratio of thicknesses of low- to high-expansion materials
n = ratio of moduli of elasticity of low- to high-expansion materials
α1 = lower coefficient of expansion, per ◦C
α2 = higher coefficient of expansion, per ◦C
T = temperature, ◦C
T0 = initial bonding temperature, ◦C
Fig.(3) The bimetallic strip.
The thermal-expansion coefficients for some commonly used materials
are given in Table (1). Bimetallic strips are frequently used in simple on-
off temperature-control devices (thermostats). Movement of the strip has
sufficient force to trip control switches for various devices. The bimeta-
llic strip has the advantages of low-cost, negligible maintenance expense,
and stable operation over extended periods of time. Alternate methods of
construction can use a coiled strip to drive a dial indicator for tempera-
tures.
Table (1) Mechanical properties of some commonly used thermal materials
55
Ex: Curvature and deflection of bimetallic strip
A bimetallic strip is constructed of strips of yellow brass and Invar
bonded together at 30◦C. Each has a thickness of 0.3 mm. Calculate the
radius of curvature when a 6 cm strip is subjected to a temperature of
100◦C.
Solution
We use Eq. (1) with properties from Table (1).
T − T0 = 100 − 30 = 70◦C
m = 1.0
n = 147/96.5 = 1.52
α1 = 1.7 × 10−6
◦C−1
α2 = 2.02 × 10−5
◦C−1
t = (2)(0.3 × 10−3
) = 0.6 × 10−3
m
Thus,
r = (0.6 × 10−3
)[(3)(2)2 + (1 + 1.52)(1 + 1/1.52)]/6(2.02 − 0.17)
(10−5)(70)(2)2 =0.132 m
Fluid-expansion thermometers represent one of the most economical,
versatile, and widely used devices for industrial temperature measure-
ment applications. The principle of operation is indicated in Fig.(4). A
bulb containing a liquid, gas, or vapor is immersed in the environment.
The bulb is connected by means of a capillary tube to some type of
pressure-measuring device, such as the bourdon gage shown.
55
Fig.(4) Fluid-expansion thermometer
An increase in temperature causes the liquid or gas to expand, thereby
increasing the pressure on the gage; the pressure is thus taken as an
indication of the temperature. Fluid-expansion thermometers are usually
low in cost, stable in operation, and accurate within ±1◦C.
Thermoelectric Effects (Thermocouples)
The most common electrical method of temperature measurement uses
the thermocouple.
There are three emfs present in a thermoelectric circuit: the Seebeck
emf, caused by the junction of dissimilar metals; the Peltier emf, caused
by a current flow in the circuit; and the Thomson emf, which results
from a temperature gradient in the materials.
Fig.(5) Junction of two dissimilar metals indicating thermoelectric effect.
When the two dissimilar materials are connected to a measuring device,
there will be another thermal emf generated at the junction of the
materials and the connecting wires to the voltage-measuring instrument.
56
This emf will be dependent on the temperature of the connection, and
provision must be made to take account of this additional potential.
Two rules are available for analysis of thermoelectric circuits:
1. If a third metal is connected in the circuit as shown in Fig. (6), the
net emf of the circuit is not affected as long as the new connections
are at the same temperature. This statement may be proved with the
aid of the second law of thermodynamics and is known as the law
of intermediate metals.
Fig.(6) Influence of a third metal in a thermoelectric circuit; law of intermediate
metals.
2. Consider the arrangements shown in Fig. (7). Simple thermocouple
circuits are constructed of the same materials but operate between
different temperature limits.
Fig.(7) Circuits illustrating the law of intermediate temperatures.
The circuit in Figure (7-a) develops an emf of E1 between temperatures
T1 and T2; the circuit in Fig. (7-b) develops an emf of E2 between
temperatures T2 and T3. The law of intermediate temperatures states
that this same circuit will develop an emf of E3 = E1 + E2 when operating
between temperatures T1 and T3, as shown in Figure (7-c).
56
The arrangement in Figure (8-a) maintains both thermocouple wires at a
reference temperature of 32◦F, whereas the arrangement in Figure(8-b)
maintains only one at the reference temperature.
Fig.(8) Conventional methods for establishing reference temperature in
thermocouple circuit. Iron-constantan thermocouple illustrated.
It is common to express the thermoelectric emf in terms of the potential
generated with a reference junction at 32◦F (0◦C). Standard thermocouple
tables have been prepared on this basis, and a summary of the output
characteristics of the most common thermocouple combinations is given
in Table (1). These data are shown graphically in Fig. (9). The output
voltage E of a simple thermocouple circuit is usually written in the form
(2)
56
where T is the temperature in degrees Celsius and E is based on a
reference junction temperature of 0◦C. The constants A, B, and C are
dependent on the thermocouple material
The sensitivity, or thermoelectric power, of a thermocouple is given by
(3)
Table(1-a) Thermal emf in absolute millivolts for commonly used thermocouple
combinations, (Reference junction of 0◦C).
Table(1-b)
56
Table(1-c)
Fig.(9) Emf temperature relations for thermocouple materials, positive electrode
listed first.
Legend:
1 Chromel-constantan (type E) A Rhenium-molybdenum
2 Iron-constantan (type J) B Rhenium-tungsten
3 Copper-constantan (type T) C Iridium-iridium rhodium
4 Chromel-alumel (type K) D Tungsten-tungsten rhenium
5 Platinum-platinum rhodium (type R) E Plat. rhodium-plat. 10% rhodium
6 Platinum-platinum rhodium (type S)
7 Nicosil-Nisil (type N)
Thermopiles:
Thermopile is a multi-thermocouples connected in series to provide
more sensitive circuit as shown in figure (10). For a three-junction
56
situation the output would be three times that of a single thermocouple
arrangement provided the temperatures of the hot and cold junctions are
uniform.
Fig.(10) Thermopile
The thermopile arrangement is useful for obtaining a substantial
emf for measurement of a small temperature difference between the
two junctions. When a thermopile is installed, it is important to ensure
that the junctions are electrically insulated from one another.
Consider the series thermocouple arrangement shown in Fig. (11). The
four junctions are all maintained at different temperatures and connected
in series. Since there are an even number of junctions, it is not
necessary to install a reference junction because the same type of
metal is connected to both terminals of the potentiometer.
Fig.(11) Series connection of thermocouples
The total emf measured at the potentiometer terminals is
E=eA-eB+eC-eD
55
Effect of Cold Junction Level of Thermopile
Ex: A thermopile consisting of five junction pairs of Chromel-constantan
is used to measure a temperature difference of 50◦C with the cold
junctions at 25◦C. Determine
1. The voltage output of the thermopile.
2. Suppose the cold-junction temperature is incorrectly stated as 75◦C
(in reality, the hot-junction temperature). What error in
temperature-difference measurement would result from this
incorrect statement?
Solution
From Table 1-a the data needed are:
E25 = 1.495 mV
E75 = 4.657 mV
E125 = 8.054 mV
all referenced to 0◦C. For the correct statement of the cold junction at
25◦C, with a hot-junction temperature of 75◦C, the voltage output reading
of the thermopile would be:
EΔT = Number of thermocouples× (E75-E25)= (5)×(4.657 − 1.495) =
15.810 mV
By incorrectly stating the cold-junction temperature as 75◦C, we have not
changed the actual reading value of 15.810 mV; however, the reading
now indicates a different value of ΔT. To obtain the new high
temperature we would calculate a new high-temperature emf as
ET = 4.657 + 15.810/5= 7.819 mV
From table 1-a
6.319mV 100oC
7.819mV X
9.789 150oC
55
Or
From which X=121.61oC
And
Ex. An iron-constantan thermocouple is connected to a potentiometer
whose terminals are at 25◦C. The potentiometer reading is 3.59 mV.
What is the temperature of the thermocouple junction?
Solution
The thermoelectric potential corresponding to 25◦C is obtained from
Table 1-a as: E25 = 1.277 mV
The emf of the thermocouple based on a 0◦C reference temperature is
thus
ET=E25+EPT
ET = 1.277 + 3.59 = 4.867 mV
From Table (1-a), the corresponding temperature is
75oC 3.918mV
X 4.867
100
5.26
Or
X-75=17.73
Or X=92.7 oC
(Q)A certain mercury-in-glass thermometer has been calibrated for a
prescribed immersion depth. The thermometer is immersed too much,
such that the extra depth is equal to a distance of 10◦ on the scale. The
true temperature reading may be calculated with
55
( )
where:
Tind = indicated temperature
Tamb = ambient temperature of the exposed stem
D = extra immersion depth of the thermometer past the correct mark
Calculate the thermometer error for an indicated temperature of 210◦F
and an ambient temperature of 70◦F.
Solution
( )
( )
Error=|Ttrue-Tind|= | | 4 oF
(Q) A chromel-alumel thermocouple is exposed to a temperature of
1560◦F. The potentiometer is used as the cold junction, and its
temperature is estimated to be 83◦F. Calculate the emf indicated by the
potentiometer.
Solution
Since 83oF can be converted to
oC as
Or (83-32)×5/9=28.33oC
And 1560 oF=848.8888
oC
From table(1) E83 can be determined by interpolation as
1.133mV and ET=E1560=35.28mV
ET=E83+EPT
35.28=1.133+EPT
EPT=35.28-1.133=34.147mV
55
(Q) Result of Installation Mistake
A heat-exchanger facility is designed to use type J thermocouples to
sense an outlet gas temperature. A safety device is installed to shut down
the flow heating system when the gas temperature reaches 800◦C. During
a periodic maintenance inspection, the thermocouple is judged to need
replacement because of oxidation. By mistake, a type K thermocouple is
installed as the replacement. What may be the results of such an
installation?
Solution
The voltage output of a type J thermocouple at 800◦C (from table 1-a)is
E800(J)=45.494mV. For this same voltage output from a type K
thermocouple the corresponding temperature can be evaluated as:
41.276mv 1000oC
45.494mV X
48.838 1200oC
Or X=1111.55oC
It can be concluded that the safety device would not be activated until a
temperature 311.5◦C higher than the design value is reached. This could
easily result in material failure of parts of the equipment.
(Q) 1. State the law of intermediate metals for thermocouples.
2. What is the Seebeck effect?
3. State the law of intermediate temperatures for thermocouples.
4.Why is a reference temperature necessary when using thermo-
couples?
5. Where might one use a bimetallic strip thermometer?