9

The KAALCPROFCA Personal Breathalyzer is a hand held

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The KAALCPROFCA Personal Breathalyzer is a hand held
Page 2: The KAALCPROFCA Personal Breathalyzer is a hand held

The KAALCPROFCA Personal Breathalyzer is a hand held

Professional Alcohol Analyzer. It provides stable, accurate and

reliable measuring technology.

Please wash your mouth with water and abstain from smoking or

drinking alcohol for at least 10 minutes before use.

· Slide back the battery cover on the back of the device.

· Install 3 x 1.5V alkaline batteries (AAA) into the battery

compartment.

· When battery power is not sufficient enough for testing, the battery

indicator will flash. In this circumstance, please remove the drained

batteries and replace with a fresh set of batteries immediately.

Power Supply

Measurement

Page 3: The KAALCPROFCA Personal Breathalyzer is a hand held

Parts

Blow

No Driving

Unit

Wait ReadyBlood alcohol content above alert level

Battery Indication

Page 4: The KAALCPROFCA Personal Breathalyzer is a hand held

· Press the 'ON' key once to turn on the device. When the machine

is powered on, 'ALCO' will be shown on the LCD display. Press

the 'OK' key. The sensor preheating will be activated. 'PH' will

flash three times and begin to count down (from 'PH99' to 'PH90-

00'). 'READY' & “BLOW” will flash alternately with beep tones

when the machine is ready to take a breath sample.

Note: the device will count down for 30 seconds. Please take a

test within this time.

· When ready to take a test, the user should blow continuously

through the mouthpiece for about 5-7 seconds. Stop blowing

once you hear a 'beep' tone. If the sampling is successful, 'Anly”

will flash upon the display, indicating an analysis of the sample

is underway. A reading will then display upon the screen.

If the blowing time is too short, the airflow from the user is too

soft or is interrupted, 'ERR' will be shown on the LCD. (indicating

an error) The user may press the 'OK' key for a new test. Test

results are displayed on the LCD within 10 seconds.

Press OK KEYTo start

Operation

Page 5: The KAALCPROFCA Personal Breathalyzer is a hand held

Note: with the LCD screen displaying the test result, as shown

above, the user may press the 'ON' key to recall the last test

result record.

Page 6: The KAALCPROFCA Personal Breathalyzer is a hand held

· When the device records its maximum available tests, '500' and

'MAX TEST' will display and flash on the LCD. Users will need to

refresh/calibrate the device so that the device can take a test with

correct result reading. 'MAX TEST' will display and blink each time

when you enter into the test mode of the device. To continue testing,

you will have to wait for 5 seconds. To discontinue the display of

“MAX TEST”, you will need to visit an authorized service center and

calibrate your equipment.

· When unused for a long time or at a low temperature, the machine

will take longer to prime itself / warm up.

· After 5 consecutive tests, allow the device to rest for 5 minutes.

· Please store in a dry environment at room temperature. Avoid

extreme heat or long-term sun exposure.

· When the battery indicator flashes, please replace the batteries.

· If the difference between two successive readings is >0.1%, the user

should test again.

· Do NOT block the outlet during sampling.

· When “Err” flashes, press 'OK' to restart the test. The device will

revert to “ALCO” mode once “Err” flashes for 5 seconds.

Important Notes

Page 7: The KAALCPROFCA Personal Breathalyzer is a hand held

· Alcohol Sensor: Fuel cell Sensor

· Blowing time: 5 to 7 seconds

· Breath power: 10L/min

· Display Range: 0.0~400.0mg/100ml or 0.00~4.00‰BAC;

or 0.000~0.400% BAC or 0.000mg/L~2.0mg/L

(if ‰ to mg/L convection factor is 2.0)

· Accuracy: +/- 0.06‰BAC at 0.5‰BAC

· Use temperature: +5°C to +45°C / Storage temperature: 0°C to

+55°C

· Test counter with recalibration alarm

· Calibration: Every 500 tests.

· Breath pressure and interruption detection

· Power Input: 3 x AAA alkaline battery.

· Display: Digital LCD display with bi-colour backlighting.

Specifications

Page 8: The KAALCPROFCA Personal Breathalyzer is a hand held

· Auto Power Off: approximately 30 seconds

· Dimension: 125 x 62 x 18 mm.

· Weight: 80g.

· Agency Approval: CE

· Accessories: 2 mouth pieces.

· Your blood alcohol content (BAC) may change within minutes

· Do not use this device to determine whether or not it is safe for you to

drive.

· You cannot use these test results in Court.

· Your safest bet is not to mix drinking with driving at all.

· Avoid Binge drinking; Remain under the ALERT LEVEL, regardless of

whether you're driving.

· Do not share with friends. This unit is designed for personal use, with

accuracy and hygienic considerations taken into account.

· This unit may require a few minutes of standby between

measurements.

· Store according to specifications. Avoid long term sun exposure and

humidity.

· Handle with care, as a fragile item.

Precautions

Page 9: The KAALCPROFCA Personal Breathalyzer is a hand held

The tester can remain accurate for 500 uses. We recommend

recalibration after this many measurements to ensure maximum

accuracy.

Please contact an Authorized service center.

NB* It is important to observe the recommended 20 minute pause

without consuming any alcohol, cigarettes or anything else before

testing. Failure to do this can shorten the life of the sensor. In such

cases, the sensor can be permanently damaged or may require

more frequent calibration. This user manual should be read

carefully for information on caring for the unit sensor.

Many factors (e.g. environment gas residue, smoke &various smells in the human mouth, etc…) may affectthe BAC level, so the readings are For ReferenceOnly. Kogan does not accept any liability arising fromthe use of the unit. Don’t drink before you drive!

How often should I re-calibrate the tester?