189
Harald Becher . Peter N Burns Handbook of Contrast Echocardiography Left ventricular function and myocardial perfusion

the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite I

Page 2: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

With a foreword by

Sanjiv Kaul, MD Professor of CardiologyDirector of Cardiovascular Imaging CenterUniversity of Virginia

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite III

Page 3: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Although the technology required for the successful application of contrast echocardiography hasevolved rapidly over the past few years, the technique has not yet gained widespread clinicalacceptance. One important reason for the lack of clinical acceptance is the relative complexity ofthe technique, particularly in respect to myocardial perfusion imaging. The interaction betweenmicrobubbles and ultrasound is an entire field by itself, as is the coronary microvasculature.

It is in this regard that practicing echocardiographers, cardiologists in training, radiologists, sono-graphers, and students will find ‘A Handbook of Contrast Echocardiography’ particularly useful.Written by two leaders in the field who have presented illustrative cases not only from their ownlaboratories but also from others around the world, this volume is a lucid, concise, and practicalguide for the day-to-day use of contrast echocardiography.

Dr Peter Burns has been involved in almost all the technical advances in the imaging methodsthat have made it possible to detect opacification of the left ventricular cavity and myocardiumfrom a venous injection of microbubbles. He has been responsible to a large degree for advancingour understanding of the interaction between microbubbles and ultrasound, which he describesin clear and easy to understand terms in this book. Dr Harald Becher has been active in the clincalapplication of contrast echocardiography for several years and has gained considerable experiencewith many imaging techniques and microbubbles, which he describes in this volume in somedetail.

The reader will find ‘A Handbook of Contrast Echocardiography’ very useful to get them startedin contrast echocardiography. The answer to almost any question that may come up in the initialperiod of one’ s training in this method will be found in this book. The authors and publishersshould be congratulated for making available a very useful and welcome addition to the armamen-tarium required to make one a successful practitioner in the art and science of contrast echocardio-graphy.

Sanjiv Kaul, MD

Charlottesville, Virginia, USAApril, 2000

Foreword

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite V

Page 4: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

The injection of a contrast agent forms a routine part of clinical x-ray, CT, MR and radionuclideimaging of the heart and vascular system. Yet in spite of the obvious significance of the vascularcomponent of the echocardiographic examination, and in spite of widespread experimentationwith contrast agents for ultrasound, ultrasound imaging of the heart has only just begun toexploit the potential benefit of contrast enhancement. Why?

A typical response from echocardiographers new to contrast is that intravascular injections woulddetract from one of ultrasound’s major attractions, that it is noninvasive. Yet, if it can be shownthat the additional diagnostic information obtained with contrast enhancement will spare thepatient from a more invasive procedure, it would be doing them no favour to deny them a pain-less venous injection. A more fundamental consideration appeals to the nature of the ultrasoundimage itself, which benefits from an intrinsically high contrast between blood and solid tissue. Itcould be argued that, unlike x-ray angiography, ultrasound does not need a contrast agent andan associated subtraction imaging method to ‘see’ blood. If we are interested in visualising wherethe blood flows, colour Doppler imaging offers a powerful and effective tool, providing the addi-tional ability to quantify hemodynamic parameters such as the direction and velocity of flow.

In the early days of echocardiography, a B-mode image could depict the cavity of a cardiac chamberbut not whether blood was flowing across a septal defect. Free gas bubbles were injected into thecavity and their trajectory imaged: this was really the first use of a contrast agent in ultrasound.Yet this invasive procedure, which requires catheterisation, is often no longer necessary. ColourDoppler imaging has largely supplanted the intra-arterial injection of bubbles in cardiac diagnosisand its capabilities now define the flow information obtainable in echocardiography.

It is precisely these capabilities that the new generation of ultrasound contrast agents for cardiacdiagnosis has extended, redefining the role of echocardiography and acting at all levels of theexamination. In imaging modes, contrast agents can help delineate the endocardial border,improving the success rate of wall motion studies. In Doppler modes, the agents boost the echofrom blood, allowing detection of epicardial vessels that would otherwise fail. Finally, and mostexciting of all, the agents make it possible for ultrasound to achieve entirely new objectives, themost striking of which is the ability for the first time to image perfusion of the myocardium inreal time. This books aims to provide both a guide and a reference for the practical use of contrastagents for these new indications.

In Chapter 1, the fundamental principle of contrast enhancement for echocardiography is describedand the mode of action of currently available microbubble agents explained. The various imaging

Preface

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite VI

Page 5: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

VII Preface

methods that have been developed to make their use more effective are discussed along with guide-lines for their clinical application. Chapter 2 addresses the use of contrast echocardiography inassessing left ventricular function. The specific indications for contrast are proposed and its roleset in the context of other clinical examinations of systolic and diastolic function. Practical, step-by-step instructions are given for using contrast in Doppler, wall motion and ejection fractionstudies and their incorporation into the stress echo examination.

Chapter 3 is devoted to imaging blood perfusion at the tissue level with ultrasound, the newestand most demanding application of contrast. The physiology and pathophysiology of myocardialperfusion are reviewed and the role of the echo examination discussed. Protocols are provided forthe new techniques of harmonic imaging, harmonic power angio, pulse inversion and powerpulse inversion and the use of the different agents explained. Although these technologies areevolving rapidly, current recommendations for specific instrument settings are included for eachexamination, with a careful explanation of their significance together with guidelines for keepingthem up to date. The indications for myocardial perfusion and coronary flow reserve contraststudies are discussed. Illustrative cases are included from a selection of centres experienced withthe techniques as well as from the authors’ own practice. Considerable attention is given to inter-pretation of the images and commonly encountered artifacts.

The final chapter introduces the more advanced topics of quantitative analysis of contrast images,including the estimation of relative myocardial vascular volume, flow velocity and perfusion rate.Examples are provided using software tools currently available for clinical use. An extensive indexand a glossary of some of the many new terms used in contrast echocardiography are provided.

We believe that contrast agents have the potential to add an entirely new dimension to the roleof ultrasound imaging in cardiology. If this book helps stimulate the reader to consider their use,it will have served its purpose.

Harald BecherPeter N Burns

Bonn and Toronto, May 2000

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite VII

Page 6: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Foreword ........................................................................................................................ V

Preface ............................................................................................................................. VII

Table of Contents ........................................................................................................ VII

Acknowledgements ..................................................................................................... XIV

Chapter 1Contrast agents for echocardiography: Principles and Instrumentation

1.1 The need for contrast agents in echocardiography .............................................. 21.1.1 B-mode imaging ............................................................................................... 21.1.2 Doppler ............................................................................................................ 31.1.3 Doppler examination of small vessels ................................................................ 4

1.2 Contrast agents for ultrasound .............................................................................. 51.2.1 Contrast agent types ......................................................................................... 5

1.2.1.1 Blood pool agents1.2.1.2 Selective uptake agents

1.2.2 Using a contrast agent ...................................................................................... 81.2.2.1 Preparation1.2.2.2 Preparing for injection

1.2.3 Administration methods ................................................................................... 121.2.3.1 Bolus1.2.3.2 Infusion

1.3 Mode of action ....................................................................................................... 161.3.1 Bubble behaviour and incident pressure ........................................................... 17

Table of Contents

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite VIII

Page 7: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

IXTable of Contents

1.3.1.1 The Mechanical Index (MI)1.3.2 I – Linear Backscatter: Doppler enhancement .................................................. 18

1.3.2.1 Enhancement studies with conventional imaging1.3.3 II – Nonlinear backscatter: harmonic imaging .................................................. 20

1.3.3.1 The need for bubble-specific imaging1.3.3.2 Harmonic imaging1.3.3.3 The impact of harmonic imaging1.3.3.4 Tissue harmonic imaging1.3.3.5 Pulse inversion imaging1.3.3.6 Power pulse inversion imaging

1.3.4 III – Transient disruption: intermittent imaging ............................................... 361.3.4.1 Triggered imaging1.3.4.2 Intermittent harmonic power Doppler

1.3.5 Summary .......................................................................................................... 39

1.4 Safety considerations .............................................................................................. 40

1.5 New developments in contrast imaging ................................................................ 411.5.1 Tissue specific bubbles ...................................................................................... 411.5.2 Bubbles for therapy ........................................................................................... 41

1.6 Summary ................................................................................................................ 42

1.7 References ............................................................................................................... 42

Chapter 2Assessment of Left Ventricular Function by Contrast Echo

2.1 Physiology and pathophysiology of LV function ................................................. 48

2.2 Available methods – The role of contrast ............................................................. 492.2.1 Systolic function: the need for endocardial border definition ............................ 492.2.2 Diastolic function: the need for pulmonary venous flow recordings ................. 50

2.3 Indications and selection of methods .................................................................... 512.3.1 Indications for contrast echo ............................................................................. 51

2.3.1.1 LV border delineation2.3.1.2 Detection, determination of size and shape of LV thrombi2.3.1.3 Pulmonary venous flow

2.3.2 Selection of patients .......................................................................................... 532.3.3 Selection of imaging methods ........................................................................... 54

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite IX

Page 8: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

X Table of Contents

2.4 How to perform an LV contrast study ................................................................... 562.4.1 Fundamental B-mode ....................................................................................... 56

2.4.1.1 Contrast agent dose2.4.1.2 Image acquisition and interpretation2.4.1.3 Pitfalls and troubleshooting

2.4.2 Fundamental pulsed wave (PW)Doppler .......................................................... 592.4.2.1 Contrast agent dose2.4.2.2 Image acquisition and interpretation2.4.2.3 Pitfalls and troubleshooting

2.4.3 Harmonic B-mode ............................................................................................ 632.4.3.1 Contrast agent dose2.4.3.2 Image acquisition and interpretation2.4.3.3 Pitfalls and troubleshooting

2.4.4 Harmonic imaging of LV thrombi .................................................................... 672.4.5 Harmonic colour/power Doppler for LVO ....................................................... 68

2.4.5.1 Continuous imaging for evaluation of LV wall motion2.4.5.2 Contrast agent dose2.4.5.3 Image acquisition and interpretation2.4.5.4 Pitfalls and troubleshooting

2.4.6 Triggered imaging for assessment of end-systolicand end-diastolic volumes and ejection fraction ................................................ 72

2.4.6.1 Contrast agent dose2.4.6.2 Image acquisition and interpretation2.4.6.3 Pitfalls and troubleshooting

2.5 Summary ......................................................................................................... 77

2.6 References ........................................................................................................ 78

Chapter 3Assessment of Myocardial Perfusion by Contrast Echo

3.1 Physiology and pathophysiology of myocardial perfusion ........................... 823.1.1 Normal perfusion ............................................................................................. 823.1.2 Acute myocardial infarction .............................................................................. 833.1.3 Chronic ischemic heart disease ......................................................................... 83

3.2 Currently available methods for myocardial perfusion imaging .................. 843.2.1 Stress ECG ....................................................................................................... 843.2.2 Stress echo ........................................................................................................ 84

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite X

Page 9: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

XITable of Contents

3.2.3 Coronary flow reserve (CFR) ............................................................................ 853.2.4 Myocardial scintigraphy .................................................................................... 853.2.5 Myocardial contrast echo .................................................................................. 86

3.3 Indications and selection of methods ............................................................ 863.3.1 Indications ........................................................................................................ 863.3.2 Selection of patients and contraindications ....................................................... 883.3.3 Selection of the imaging method ...................................................................... 88

3.3.3.1 Harmonic power Doppler (HPD)3.3.3.2 Harmonic B-Mode3.3.3.3 Pulse inversion imaging3.3.3.4 Power pulse inversion (PPI)

3.4 Special considerations for myocardial contrast ............................................. 923.4.1 Impact of the scanplane .................................................................................... 923.4.2 Triggered imaging ............................................................................................. 94

3.4.2.1 Incremental triggered imaging3.4.2.2 Double or multiple trigger3.4.2.3 Flash echo3.4.2.4 Power pulse inversion flash echo 3.4.2.5 Systolic versus diastolic trigger

3.5 Choice of agent and method of administration ............................................ 973.5.1 Continuous infusion versus bolus injection ...................................................... 973.5.2 Preparation of contrast infusion ........................................................................ 993.5.3 Adjustment of infusion rate .............................................................................. 99

3.6 Instrument settings ......................................................................................... 1003.6.1 Harmonic power Doppler ................................................................................ 100

3.6.1.1 Setting the trigger point3.6.2 Harmonic B-mode ............................................................................................ 1033.6.3 Pulse inversion imaging .................................................................................... 1063.6.4 Power pulse inversion ....................................................................................... 108

3.7 Image acquisition ............................................................................................ 108

3.8 Stress testing during myocardial contrast echo ..................................................... 1093.8.1 Exercise and dobutamine stress ......................................................................... 1093.8.2 Vasodilator stress ............................................................................................... 1093.8.3 Combined assessment of wall motion and myocardial perfusion ...................... 111

3.9 Reading myocardial contrast echocardiograms ..................................................... 112

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite XI

Page 10: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

XII Table of Contents

3.9.1 Visual assessment of unprocessed recordings ..................................................... 1123.9.1.1 Normal perfusion3.9.1.2 Perfusion Defect

3.9.2 Visual assessment of post-processed recordings ................................................. 1193.9.3 Report of visual judgement ............................................................................... 119

3.10 Clinical profiles/interpretation of myocardial contrast echo ............................... 1213.10.1 Acute myocardial infarction .............................................................................. 1213.10.2 Scar or fibrosis versus viable myocardium ......................................................... 1233.10.3 Coronary artery stenosis ................................................................................... 124

3.11 Pitfalls and troubleshooting .................................................................................. 1273.11.1 Inadequate myocardial contrast ........................................................................ 1273.11.2 Contrast shadowing .......................................................................................... 1293.11.3 Blooming .......................................................................................................... 1313.11.4 Wall motion artifacts ........................................................................................ 1313.11.5 The bubble depletion artifact ............................................................................ 133

3.12 Coronary flow reserve and myocardial contrast echo ........................................... 1333.12.1 What is coronary flow reserve? .......................................................................... 134

3.13 Available methods – need for contrast enhancement ........................................... 134

3.14 Coronary flow reserve: indications and selection of methods ............................. 1353.14.1 Selection of patients .......................................................................................... 136

3.15 How to perform a CFR study ................................................................................ 1373.15.1 Intravenous lines ............................................................................................... 1373.15.2 Contrast agent .................................................................................................. 1373.15.3 Protocols to induce hyperemia .......................................................................... 1373.15.4 Image orientation for visualising blood flow in the LAD .................................. 1393.15.5 Instrument settings ........................................................................................... 1393.15.6 Combination with myocardial contrast echo (MCE) ........................................ 141

3.16 Image acquisition and interpretation .................................................................... 1423.16.1 Significance of an LAD stenosis ........................................................................ 1433.16.2 Follow-up of an LAD stenosis after PTCA ....................................................... 143

3.17 Pitfalls and troubleshooting .................................................................................. 1433.17.1 Angle of the vessel to the beam ......................................................................... 1433.17.2 Displacement of the sample volume ................................................................. 1443.17.3 Flow in mammary artery .................................................................................. 144

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite XII

Page 11: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

XIIITable of Contents

3.17.4 Bubble noise ..................................................................................................... 1453.17.5 Interpretation problems: impact of preload ....................................................... 1463.17.6 Impact of blood pressure .................................................................................. 1463.17.7 Vasodilation during hyperemia ......................................................................... 147

3.18 Summary ................................................................................................................. 147

3.19 References ............................................................................................................... 147

Chapter 4Methods for quantitative Analysis

4.1 Basic tools for image quantification ...................................................................... 1544.1.1 Single image quantification ............................................................................... 1554.1.2 Cineloop quantification .................................................................................... 156

4.1.2.1 Reduction of data4.1.2.2 Positioning Regions of Interest (ROI)4.1.2.3 Automatic analysis

4.2 Advanced image processing: cases & examples ..................................................... 1594.2.1 Single image quantification ............................................................................... 1594.2.2 Background subtraction .................................................................................... 1604.2.3 Analysing the time-course of enhancement ....................................................... 164

4.2.3.1 Intravenous bolus technique 4.2.3.2 Negative bolus technique 4.2.3.3 Real-time negative bolus technique

4.3 Summary ................................................................................................................ 170

4.4 References ............................................................................................................... 170

Glossary .................................................................................................................. 172

Index .................................................................................................................. 180

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite XIII

Page 12: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

This book arose out of a series of seminars on contrast echocardiography organised by SabinoIliceto of the University of Cagliari. We are grateful to Sabino and his colleagues, especially CarloCaiati, for this and their contribution to the section on coronary flow reserve. We are also indebtedto Danny Skyba and to Damien Dolimier, both pioneers of quantitative analysis of contrast echo-cardiograms, for their material which comprises much of Chapter 4. At home, we both acknowl-edge the invaluable help of our own colleagues: in Bonn, of Melanie Hümmelgen, StefanieKuntz, Heyder Omran, Birgit Piel, Rami Rabahieh, Rainer Schimpf, Klaus Tiemann; and inToronto, of Cash Chin, Stuart Foster, Kasia Harasewicz, David Hope Simpson, Jennifer LeeWah, Zvi Margaliot, Carrie Purcell, Howard and Sylvie.

Considerable effort was expended to produce a book which is easy to read and whose technicalquality does justice to that of its many clinical images. The evident success is entirely that ofRaffaela Kluge of Schering AG and Kay Bengelsdorf of Agentur B&S, who carried the produc-tion editing, typesetting and printing in Berlin, virtually single-handed. Their extraordinaryefforts went far beyond these traditional roles and were supported by Luis Reimer Hevia ofSchering and Walter Schimmel of B&S, to whom we are also grateful.

Finally, credibility is lent to a new clinical technique only when it is mastered by a large numberof independent investigators; the authors are fortunate that the field of echocardiography isoccupied by clinicians who are not only skilled and experienced, but open to innovation anddedicated to teaching. In this spirit a large number of them have contributed their own cases asillustrations to the contrast techniques described here. This handbook could not have been writtenwithout them, and we thank them all for their generosity:

Acknowledgements

Luciano AgatiUniversity of Rome, Italy

Gian Paolo BezanteUniversity of Genoa, Italy

Carlo CaiatiUniversity of Cagliari, Italy

Damien Dolimier ATL Ultrasound, Bothell WA, USA

Steve FeinsteinRush University, Chicago, USA

Kathy FerraraUniversity of California, Davis, USA

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite XIV

Page 13: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

XVAcknowledgment

Christian FirschkeUniversity of Munich, Germany

Francesco GentileUniversity of Milan, Italy

Sabino IlicetoUniversity of Cagliari, Italy

Sanjiv KaulUniversity of Virginia, USA

Heinz LambertzDeutsche Klinik für Diagnostik, Wiesbaden,Germany

Roberto LangUniversity of Chicago, USA

Jonathan LindnerUniversity of Virginia, USA

Thomas MarwickUniversity of Brisbane, Australia

Gerd P MeyerUniversity of Hanover, Germany

Mark MonaghanUniversity of London, UK

Sharon MulvaghMayo Clinic, USA

Thomas PorterUniversity of Nebraska, USA

Ricardo Ronderos Universidad Nacional de La Plata, Argentina

Jiri SklenarUniversity of Virginia, USA

Danny SkybaATL Ultrasound, Bothell WA, USA

Folkert Ten CateThorax Centre, Rotterdam, Netherlands

Jim ThomasCleveland Clinic, USA

Neil WeissmanWashington Hospital,Washington DC,USA

Eric YuUniversity of Toronto, Canada

J Luis ZamoranoUniversity of Madrid, Spain

Kardio-Titelei for pdf 06.06.2000 20:05 Uhr Seite XV

Page 14: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

Peter N BurnsProfessor of Medical Biophysics and RadiologyUniversity of TorontoImaging ResearchSunnybrook and Women’s Health Science Centre2075 Bayview AvenueToronto, OntarioCanada M4N [email protected]

Harald BecherProfessor of CardiologyUniversity of BonnRheinische Friedrich-Wilhelms-UniversitätMedizinische Universitätsklink und Poliklink IIKardiologie/PneumologieSigmund-Freud-Straße 2553105 [email protected]

Copyright © 2000 by Harald Becher and Peter N Burns. This book is protected by copyright. All rights are reserved, whether the wholepart of material is concerned, specifically the rights of translation, reprinting, reuse of illustrations, recitation, broadcasting, reproductionon microfilm.

This electronic copy was downloaded under the conditions of the End User License Agreement which accompanies it. Use and storageof this document signifies agreement with the terms of the Agreement. The files may not be altered without prior written permission ofthe copyright owners. No text, figures, tables or images may be displayed or reproduced, except for personal use, in any form or by anymeans, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission of the copyright owners.

The use of general descriptive names, registered names, trademarks, etc. in this publication does not imply, even in the absence of a specificstatement, that such names are exempt from the relevant protective laws and regulations and therefore free for general use.

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 1

Page 15: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 2

Page 16: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

1 Contrast Agents for Echocardiography:Principles and Instrumentation

3

Shall I refuse my dinner because I do not fullyunderstand the process of digestion?

Oliver Heaviside, 1850–1925

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 3

Page 17: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Introduction

Contrast agents for ultrasound are unique inthat they interact with, and form part of, theimaging process. Contrast imaging cannot beperformed effectively without a basic under-standing of this interaction and how it isexploited by the new imaging modes that havebecome available on modern ultrasoundsystems. In this chapter we consider how echo-cardiography might benefit from a contrastagent, describe currently available agents andexplain their mode of action. We discuss theimpact of ultrasound contrast on echocardio-graphic techniques and instrumentation andconclude with the most recent developmentsin this rapidly evolving field.

1.1 The need forcontrast agents in echocardiography

1.1.1 B-mode imaging

It is well known that blood appears ‘black’ on ultrasound imaging. This is not becauseblood produces no echo, it is simply that thesound scattered by red blood cells at the lowdiagnostic frequencies is very weak, about1,000–10,000 times weaker than that fromsolid tissue, so lies below the displayed dynamicrange of the image. Amongst the roles that theimage plays in the ultrasound examination ofthe heart is the identification of boundaries,especially of those between the blood and thewall of the cavity. Identification of the entiremargin of the endocardium in a view of the leftventricle, for example, is an important compo-nent of any wall motion study. Although in

some patients this boundary is seen clearly, inmany the endocardial border is poorly definedbecause of the presence of spurious echoeswithin the cavity. These echoes, which arefrequently a result of reverberation of the ultra-sound beam between the transducer and thechest wall and aberration of the beam in itspath between the ribs, result in a reduction of useful contrast between the wall and thecavity – the blurred haze that is familiar tomany ultrasonographers. By enhancing the echofrom blood in these patients by using a contrastagent, the blood in the cavity can be renderedvisible above these artifacts. Because it is morehomogeneous than the wall and because it isflowing, this echo does not suffer from thesame artifacts and a clear boundary is seen,revealing the border of the endocardium(Figure 1).

If the echo from blood is enhanced by a con-trast agent, the signals obtained from a duplexDoppler examination of the vessel will besimilarly enhanced (Figure 2). A further rolefor such an agent immediately suggests itself. Ifthe echo from large blood vessels can be

2 Handbook of Contrast Echocardiography

Fig. 1 Contrast enhanced harmonic image of theleft ventricle shows the endocardial border clearly.

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 2

Page 18: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

enhanced by an agent, what effect will it haveon the small volume of blood that is in themicrovessels of the myocardium? The muscleitself appears bright on an ultrasound image, sowe can expect the additional brightness due tothe agent in the myocardial vessels to be verysmall, yet if it is detectable, it would open thepossibility of using ultrasound to map therelative perfusion volume of blood in themuscle itself. As we shall see, it is normally notpossible to detect this tiny echo with existingtechniques, but with the aid of new methodssuch as harmonic and pulse inversion imagingand Doppler, myocardial perfusion volumeimaging becomes possible with ultrasound.

1.1.2 Doppler

Doppler forms an essential part of all echo-cardiography examinations. It is used both todetect large volumes of blood moving slowly,such as in the cavities, and small volumes ofblood moving fast, such as in stenotic valvularjets. It is also used in many vascular beds tovisualise the flow at the parenchymal level ofthe circulation, where blood vessels lie belowthe resolvable limit of the image. The detectionof such ‘unresolved’ flow using Dopplersystems can be demonstrated simply by using aduplex scanner to create a power Dopplerimage of an abdominal organ such as thekidney, in which vessels that are not seen onthe greyscale image become visible using theDoppler mode. These vessels are the arcuateand interlobular branches of the renal artery,whose diameter is known to be less than100 µm and therefore below the resolutionlimit of the image. However, as we progressdistally in the arterial system, the blood flowsmore slowly as the rate of bifurcation increases,producing lower Doppler shift frequencies.The quantity of blood in a given volume oftissue also decreases, weakening the back-scattered echo. Eventually, a point is reached atwhich the vessel cannot be visualised and theDoppler signals cannot be detected. The myo-cardial perfusion bed lies beyond this point.

3Principles and Instrumentation

Conditions for successful Dopplerdetection of blood flow

• Velocity of flow must be sufficient to give detectable Doppler shift

• Strength of blood echo must be sufficient for detection

• Tissue motion must be sufficiently slow that its Doppler shift may be distinguished from that of blood flow

Fig. 2 Enhancement of Doppler signals by amicrobubble contrast agent. The spectral displayshows an increase in intensity with the arrival ofthe agent in an artery following intravenous injec-tion.

Contrast in echocardiography: why?

• To enhance Doppler flow signals from thecavities and great vessels

• To delineate the endocardium by cavity opacification

• To image perfusion in the myocardium

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 3

Page 19: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

1.1.3 Doppler examination of small vessels

To understand how contrast agents may helpto image myocardial perfusion we need toexamine these performance limits for theDoppler detection of blood flow. Two condi-tions must be satisfied before a Doppler signalcan be detected: first, the velocity of bloodmust be sufficient to produce a Doppler shiftfrequency that is distinguishable from that pro-duced by the normal motion of tissue, andsecond, the strength of the echo must providea sufficient signal at the transducer to allowdetection above the acoustic and electricalnoise of the system. For fast moving blood,such as that in a stenotic jet, it is the secondcondition that determines if a Doppler inter-rogation – in spectral or colour modes – fails toobtain a signal. If the echo is too weak, theexamination fails. The role of the contrastagent is to enhance the blood echo, therebyincreasing the signal-to-noise ratio and hencethe success rate of a Doppler examination.Figure 3 shows how with the addition of a con-trast agent, a colour Doppler system can showthe apical coronary vessels from a transthoracicview.

For small blood vessels, on the other hand, thediagnostic objective in using an ultrasoundcontrast agent is to detect flow in the circula-tion at a level that is lower than would other-wise be possible in Doppler or greyscale modes.In cardiac applications the target small vesselsare those supplying the myocardium. Theechoes from blood associated with such flow –at the arteriolar level for example – exist in themidst of echoes from the surrounding solidstructures of the organ parenchyma, echoeswhich are almost always stronger than even thecontrast-enhanced blood echo. Thus, in order

to be able to image flow in the myocardium, acontrast agent is required that either increasesthe blood echo to a level that is substantiallyhigher than that of the surrounding tissue, or amethod must be conceived for suppressing theecho from non-contrast-bearing structures.Although Doppler is an effective method forseparating the echoes from blood and tissue, itrelies on the relatively high velocity of theflowing blood compared to that of, say, thecardiac wall. Although this distinction – whichallows us to use a highpass (or ‘wall’) filter toseparate the Doppler signals due to bloodflowfrom those due to the wall itself – is valid forflow in large vessels or across the cardiac valves,it is not valid for the myocardium, where themuscle is moving much faster than the bloodwhich flows in its vasculature. Thus theDoppler shift frequency from the movingmuscle is comparable to or higher than that ofthe moving blood itself. Because the wall filtercannot be used without eliminating both theflow and the muscle echoes, the use of Dopplerin such circumstances is defeated by the over-whelming signal from tissue movement: the

4 Handbook of Contrast Echocardiography

Fig. 3 Contrast enhanced Doppler examiniationof the apical coronary vessels.

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 4

Page 20: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

5Principles and Instrumentation

‘flash’ artifact in colour or the ‘thump’ artifactin spectral Doppler. Thus myocardial flow can-not be imaged with conventional Doppler,with or without intravenous contrast agents. Anew Doppler modality is necessary: as we shallsee, harmonic and pulse inversion imagingprovide this (Figure 4).

1.2 Contrast agents for ultrasound

The principal requirements for an ultrasoundcontrast agent are that it should be easily intro-ducible into the vascular system, be stable forthe duration of the diagnostic examination,have low toxicity and modify one or moreacoustic properties of tissues which determinethe ultrasound imaging process. Although it isconceivable that applications may be found forultrasound contrast agents which will justifytheir injection directly into arteries, the clinicalcontext for contrast ultrasonography requires

that they be capable of intravenous administra-tion. As we shall see, these constitute ademanding specification for a drug, one thatonly recently has been met. The technologyuniversally adopted is that of encapsulatedbubbles of gas which are smaller than redblood cells and therefore capable of circulatingfreely in the body. These are the so-called‘blood pool’ agents. Agents have also beenconceived that are taken up by a chosen organsystem or site, as are many nuclear medicinematerials.

1.2.1 Contrast agent types

Contrast agents might act by their presence inthe vascular system, from where they are ulti-mately metabolised (‘blood pool’ agents) or bytheir selective uptake in tissue after a vascularphase. Of the properties of tissue that influencethe ultrasound image, the most important arebackscatter coefficient, attenuation and acous-tic propagation velocity (1). Most agents seekto enhance the echo by increasing the back-scatter of the tissue that bears them as much aspossible, while increasing the attenuation inthe tissue as little as possible, thus enhancingthe echo from blood.

An ideal ultrasound contrast agent

• Non-toxic

• Intravenously injectable, by bolus or infusion

• Stable during cardiac and pulmonary passage

• Remains within the blood pool or has a well-

specified tissue distribution

• Duration of effect comparable to that of the

imaging examination

Fig. 4 Myocardial perfusion imaged using inter-mittent harmonic power Doppler and Levovist.

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 5

Page 21: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

1.2.1.1 Blood pool agentsFree gas bubblesGramiak and Shah first used injected bubblesto enhance the blood pool echo in 1968 (2).They injected saline into the ascending aortaduring echocardiographic recording and notedstrong echoes within the normally echo freelumen of the aorta and the chambers of theheart. Subsequent work showed that thesereflections were the result of free bubbles of airwhich came out of solution either by agitationor by cavitation during the injection itself. Inthis early work, many other fluids were foundto produce a contrast effect when similarlyinjected (3, 4). The intensity of the echoes pro-duced varied with the type of solution used:the more viscous the solution, the more micro-bubbles were trapped in a bolus for a sufficientlength of time to be appreciated on the image.Agitated solutions of compounds such as indo-cyanine green and renografin were also used.

Most of the subsequent research into the appli-cation of these bubbles as ultrasound contrastagents focused on the heart, including evalua-tion of valvular insufficiency (5, 6), intracardiacshunts (7) and cavity dimensions (8). Thefundamental limitation of bubbles produced inthis way is that they are large, so that they areeffectively filtered by the lungs, and unstable,so that they go back into solution within asecond or so. Hence this procedure was invasiveand, except by direct injection, unsuitable forimaging of left-sided cardiac chambers, thecoronary circulation and the systemic arterialtree and its organs.

Encapsulated air bubblesTo overcome the natural instability of free gasbubbles, attempts were made to encapsulategas within a shell so as to create a more stableparticle. In 1980 Carroll et al (9) encapsulatednitrogen bubbles in gelatin and injected theminto the femoral artery of rabbits with VX2tumours in the thigh. Ultrasound enhance-ment of the tumour rim was identified.However, the large size of the particles (80 µm)precluded administration by an intravenousroute. The challenge to produce a stable encap-sulated microbubble of a comparable size tothat of a red blood cell and which could survivepassage through the heart and the pulmonarycapillary network was first met by Feinstein etal in 1984 (10), who produced microbubblesby sonication of a solution of human serumalbumin and showed that it could be visualisedin the left heart after a peripheral venous injec-tion. This agent was subsequently developedcommercially as Albunex® (MallinckrodtMedical Inc, St Louis, MO).

A burgeoning number of manufacturers havesince produced forms of stabilised micro-bubbles that are currently being assessed for

6 Handbook of Contrast Echocardiography

Fig. 5 The principle of Levovist. Air adheres tothe surface of galactose microparticles which 'size'the resulting bubbles to have a median diameter of about 4 µm. Upon dissolution, the bubbles arecoated with a thin, permeable shell comprisingpalmitic acid.

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 6

Page 22: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

use as intravenous contrast agents for ultra-sound. Several have passed through ‘Phase 3’clinical trials and gained regulatory approval inEurope, North America and, more recently,Japan. Levovist® (Schering AG, Berlin,Germany), is a dry mixture comprising 99.9percent microcrystalline galactose micro-particles, and 0.1 percent palmitic acid. Upondissolution and agitation in sterile water, thegalactose disaggregates into microparticleswhich provide an irregular surface for theadherence of microbubbles 3 to 4 µm in size(Figure 5). Stabilisation of the microbubblestakes place as they become coated with palmiticacid, which separates the gas – liquid interfaceand slows their dissolution (11). These micro-bubbles are highly echogenic and are suf-ficiently stable for transit through the pulmon-ary circuit. The estimated median particle sizeis 1.8 µm and the median bubble diameterapproximately 2 µm with the 97th centileapproximately 6 µm (12). The agent is chemi-cally related to its predecessor Echovist®

(Schering AG, Berlin, Germany) a galactoseagent that forms larger bubbles and which hasbeen used extensively without suggestion oftoxicity. Both preclinical (13) and clinical (14)studies with Levovist demonstrate its capacityto traverse the pulmonary bed in sufficientconcentrations to enhance both colourDoppler and, in some instances, the B-modeimage itself.

Low solubility gas bubbles

The ‘shells’ which stabilise the microbubblesare extremely thin and allow a gas such as air todiffuse out and go back into solution in theblood. How fast this happens depends on anumber of factors which have been seen tovary not only from agent to agent, but frompatient to patient. After venous introduction,however, the effective duration of the two

agents described above is of the order of a fewminutes. Because they are introduced as abolus and the maximum effect of the agent isin the first pass, the useful imaging time isusually considerably less than this. Newer(sometimes referred to as ‘second generation’)agents, designed both to increase backscatterenhancement further and to last longer in thebloodstream are currently under developmentand early clinical use. Instead of air, many ofthese take advantage of low solubility gasessuch as perfluorocarbons, the consequent lowerdiffusion rate increasing the longevity of theagent in the blood. However, a price may bepaid for this stability in the reduced acousticresponsiveness of the agent (see §1.3.4.2).Optison® (Mallinckrodt Medical Inc, St Louis,MO) is a perfluoropropane filled albumin shellwith a size distribution similar to that ofAlbunex (Figure 6). The stability of the smallerbubbles in its population is the probable causeof the greater enhancement observed with thisagent. Echogen® (Sonus Inc, Bothell WA) is anemulsion of dodecafluoropentane dropletswhich undergo a phase change, ‘boiling’ tocreate gas which is stabilised into bubbles by anaccompanying surfactant. In practice, this

7Principles and Instrumentation

Fig. 6 Optison microbubbles photographed invitro with red blood cells (Courtesy Mallinckrodt Inc).

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 7

Page 23: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

agent requires external preparation of the mix-ture (eg, the ‘popping’ of a syringe by forcedwithdrawing of the plunger against its closedoutlet) to create the bubble population beforeinjection. SonoVue™ (Bracco Inc, NJ) usessulphur hexafluorane in a phosphorlipid shell.Definity™ (also known as DMP115, DuPontInc, Boston MA) comprises a perfluoropropanemicrobubble coated with a particularly flexiblebilipid shell which also shows improved stabili-ty and high enhancements at low doses (15)(Figure 7). Other agents are under aggressivedevelopment (see Table).

1.2.1.2 Selective uptake agentsA perfect blood pool agent displays the sameflow dynamics as blood itself, and is ultimatelymetabolised from the blood pool. Agents canbe made, however, that are capable of provid-ing ultrasound contrast during their meta-

bolism as well as while in the blood pool.Colloidal suspensions of liquids such as per-fluorocarbons (16) and certain agents withdurable shells (17) are taken up by the reticulo-endothelial system from where they ultimatelyare excreted. There they may provide contrastfrom within the liver parenchyma, demarkingthe distribution of Kupffer cells (18). In thefuture, such agents with a cell-specific pathwaymay be used as a means both to detect and todeliver therapeutic agents to a specific site inthe cardiovascular system.

1.2.2 Using a contrast agent

There is no question that the majority of dif-ficulties that occur when a contrast agent isfirst used in a clinical echo laboratory can beattributed to problems with the preparationand administration of the injected materialitself. Many physicians and nurses used toadministering pharmaceutical drugs, includingx-ray contrast, cannot understand why the pre-paration and injection itself are so critical for

8 Handbook of Contrast Echocardiography

Fig. 7 Definity microbubbles seen flowing in ablood vessel. Intravital microscopy shows a capil-lary following intravenous administration of fluores-cent-labeled agent. Two microbubbles traverse thevessel with adjacent red blood cells (arrows) seenfaintly under the concomitant low-level trans-illumination. Courtesy Jonathan Lindner, University of Virginia, VA,USA

Characteristics

Could not traversecardiopulmonary beds

Successful trans-pulmonary passage

Improved stability

Controlled acousticproperties

Formulation

Free gas bubbles

Encapsulated airbubbles

Encapsulated lowsolubility gas

bubbles

‘Particulate’(eg polymer shell)

gas bubbles

‘Generation’

0

1

2

3

The evolution of contrast agentsfor ultrasound

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 8

Page 24: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

9Principles and Instrumentation

Some ultrasound contrast agents

Manufacturer Name Type Development Stage

Acusphere AI-700 Polymer/perflourocarbon Early Development

Alliance/Schering Imavist™ Surfactant/ Clinical Development

perfluorohexane-air

Bracco SonoVue™ Phospholipid/ Late Clinical Development

Sulphur hexafluoride

Byk-Gulden BY963 Lipid/air Not commercially developed

Cavcon Filmix™ Lipid/air Pre-clinical Development

DuPont Definity™ Liposome/perfluoropropane Late Clinical Development

Molecular Optison® Cross-linked human serum Approved in EU, US,

Biosystems/ albumin/perfluoropropane Canada

Mallinckrodt

Molecular Albunex® Sonicated albumin/air Approved in EU, US

Biosystems/

Mallinckrodt

Nycomed Sonazoid™ Lipid/perfluorocarbon Late Clinical Development

Point Biomedical Bisphere™ Polymer bilayer/air Clinical Development

Porter PESDA Sonicated dextrose albumin/ Not commercially

perfluorocarbon developed

Quadrant Quantison™ Spray-dried albumin/air Pre-clinical Development

Schering Echovist® Galactose matrix/air Approved in EU, Canada

Schering Levovist® Lipid/air Approved in 69 countries,

including EU, Canada, Japan (not US)

Schering Sonavist™ Polymer/air Clinical Development

Sonus Echogen® Surfactant/ Late Clinical Development

Pharmaceutical dodecafluoropentane

Kardio-Kap 1 for pdf 06.06.2000 16:55 Uhr Seite 9

Page 25: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

10 Handbook of Contrast Echocardiography

an ultrasound contrast study. Yet although theyare classified as drugs, all current agents arereally physical suspensions of bubbles in aninactive medium such as saline. The bubblesare stabilised by shells that are sometimes onlytens of nanometres in thickness: they are physi-cally delicate and particularly susceptible todestruction by pressure or shear stress. The gasalso diffuses out of the bubbles with time andsometimes this process is faster outside thebody than once injected. In addition, thebubbles have a tendency to float and henceseparate from the solution that holds themover a period of time. As long as it is realisedthe bubble suspensions cannot be handled likean ordinary drug and require special care, it isrelatively easy to prepare the echo laboratoryfor their handling.

1.2.2.1 PreparationIt should be appreciated that no two agents arealike in the way that they are prepared for injec-tion. Some vials contain bubbles that simply arereconstituted by the addition of saline, othersrely on the user to manufacture the bubblesduring the preparation process. These requiremechanical agitation or a more elaborate mixingprocedure that must be followed carefully.

Levovist is prepared by injection of sterile waterinto a vial containing a sugar/lipid powderfollowed by vigorous shaking of the vial byhand. The agent can be administered in one ofthree concentrations (200, 300 or 400 mg/ml),upon which the volume of water to be useddepends. It should be left to stand for about 2 minutes after mixing. The more concentratedsuspensions are somewhat viscous, so thatbubble flotation is not a practical problem withthis agent. The bubbles are, however, sensitiveto the increase of pressure within the vial thatwould result from the injection of the waterinto a closed space, so the vial is vented for theaddition of water and withdrawal of the agentby the use of a special cap provided with eachdose (Figure 8). The agent should be usedwithin about 30 minutes of preparation.

Optison is kept in a refrigerator from which itshould be removed to bring it to roomtemperature before use. It is prepared bysimple shaking by hand and withdrawal from avented vial through an 18-gauge needle.Optison bubbles are buoyant and have atendency to rise quickly to the surface of the syringe unless it is gently but constantlyrotated before injection.

SonoVue is prepared by simple mixing withsaline, which is provided in the measuredsyringe kit (Figure 9). The bubbles are also

Fig. 8 Preparation of Levovist. Sterile water ismeasured according to the concentration requiredand added to a vial of dry powder through a specialventing cap. The mixture is agitated by hand andleft to stand for two minutes before slow withdrawalinto the injecting syringe through the cap.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 10

Page 26: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

11Principles and Instrumentation

buoyant and the preparation of quite lowviscosity, so the bubbles also float rapidly.SonoVue can be infused effectively using aninfusion pump, but care should be taken toensure that the direction of the infused outputis vertical, so that bubbles do not becometrapped.

Definity, like Optison, is already mixed whenthe vial is manufactured. However, the bubblesare formed only after the vial has been agitatedfor 45 seconds in a mechanical shaker, which ismost conveniently located in the examinationroom (Figure 10 a, b). The agent is withdrawnfrom the vial by venting with a second 18-gauge needle and may be injected by syringe(Figure 10 c) or infused by simple injection intoa drip bag (Figure 10 d). In this form Definityis unusually stable and the bubbles neutrallybuoyant, so that flotation is not a problem.

1.2.2.2 Preparing for injectionWith all agents, the use of smaller diameterneedles should be avoided because the bubblesare subjected to large pressure drop due to theBernoulli effect as the fluid exits the tip of thelumen. The faster the injection and the smallerthe diameter, the larger the pressure drop andthe greater the likelihood of damage to thebubbles. A 22-gauge or larger needle is best. Itshould be borne in mind that the smaller the

Fig. 9 Preparation of SonoVue. The agent ispackaged as a dry powder and reconstituted usingsaline measured in a the syringe supplied.

a) b)

c) d)

Fig. 10 Preparation of Definity. The vial is agitatedby a mechanical device supplied specifically for thisagent (a, b). It is then withdrawn from the vial forbolus injection (c) or injected directly into a drip bagfor infusion (d).

Preparing a contrast agent: tips

• Establish machine settings and scan patient before mixing the agent

• Use vents: never withdraw from or inject agent against a closed space

• Watch for flotation of bubbles: gently shake the vial and syringe after preparation

• Follow the manufacturer’s directions carefully

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 11

Page 27: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

needle lumen, the slower the injection shouldbe to avoid inadvertent bubble destruction.Bubbles can also be destroyed by pushing thesyringe plunger against a closed line: if thishappens accidentally, the dose should be dis-carded. Because the total injection volume forsome studies can be less than 1ml, a flush isneeded to push the agent into the centralvenous stream. 5–10 ml saline administeredthrough a 3-way stopcock at the end of a shortline which allows free movement of the syringefor mixing is often best. It is good practice forthe saline to traverse the right angle bend of thestopcock, not the agent, which should beconnected along the direct path, again to avoidbubble destruction (Figure 11).

1.2.3 Administration methods

1.2.3.1 BolusFigure 12 shows dose-response measurementsmade using a dedicated Doppler probe posi-tioned on the brachial artery of a patient afteran injection of Optison. A first-pass peakenhancement of 30 dB is followed by a steady,exponential wash-out of about 3 minutes’duration, which is typical of the kinetics of ablood pool microbubble agent. The hugeenhancement at the peak usually causes over-

loading of the Doppler receiver, creating spec-tral or colour blooming. Sonographers com-monly evade this by waiting until the wash-outphase, in which a more suitable enhancementcan be obtained for a longer period. It is clear,however, that this is an inefficient way to usethe agent because the majority of bubbles donot contribute to the collection of diagnosticinformation. Decreasing the bolus volumedoes not necessarily help. If the dose ofOptison is increased progressively from

12 Handbook of Contrast Echocardiography

Administering a contrast agent: tips

• Establish the venous line before preparing the agent

• Use a 22-gauge or larger cannulation; cubital vein is often best

• Use a 3-way stopcock if preferred but never push filled syringe against a closed valve

• Watch for bubble flotation: keep the syringe moving between injections

• Inject slowly: follow the manufacturer’s recommendations

Fig. 12 Example of typical time enhancementcurve measured in the brachial artery of a patientafter injection with Optison. Note that the effectiveduration of enhancement in this case is about 3minutes.

Fig. 11 A three way stopcock, line and flush readyfor injection of Optison.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 12

Page 28: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

13Principles and Instrumentation

Fig. 13 The enhancement curves of Optison with increasing dose. Note that increasing the dose in-creases both the peak enhancement and the wash-out time of the agent.

Fig. 14 Peak enhancement as a function ofadministered dose. This graph shows that the effectof the agent at its peak does not increase withlarge doses. This may be due to 'bolus spreading'.

Fig. 15 Duration of enhancement as a function ofadministered dose. Note that the effect of largerdoses is to extend the duration of enhancement,rather than increase the height of the peak effect.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 13

Page 29: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

14 Handbook of Contrast Echocardiography

10–200 µl/kg (Figure 13), we see the peakenhancement increase slightly and the wash-out time increase more dramatically. Note thatincreasing the dose by a factor of 10 does nothave the same effect on the peak enhancement(Figure 14). Instead, it is the wash-out time ofthe agent, reflected by the duration of enhan-cement that is increased (Figure 15). The areaunder these curves represents the product of

the enhancement and the time of enhancementand may be regarded as a crude measure of theintegrated effect of the agent. If this is plottedagainst the dose we see (Figure 16) that increas-ing the volume of a bolus causes this value torise in a roughly linear manner. It is on thisobservation that the effort to infuse agents isbased.

1.2.3.2 InfusionAlthough there is a definite role for bolus injec-tion where short duration, maximum effect isrequired from the agent (eg in LV opacificationfor the wall motion studies of §2.4), infusing acontrast agent offers the possibility of enhance-ment for a length of time that more closelymatches that of the imaging examination.Agents may be infused by slow injection, by useof an injection pump, by dripping the dilutedagent through an IV line, or by use of an infu-sion pump of the kind used to titrate IV drugs.Figure 17 compares SonoVue administered as abolus (Figure 17 a) and as an infusion throughan IV pump (Figure 17 b). Figure 18 shows theresults of a slow manual injection of Levovist,with an enhancement of 15 dB effectivelylasting for more than 9 minutes. It can be seenthat the effect of infusion is to maintain anenhancement comparable to that of the peakbolus for a period which matches more closelythat of the clinical ultrasound examination.There is, then, a clear advantage to infusingcontrast agents, not least of which is that theimproved efficiency translates into an economicsaving on the cost of the agent. Against thismust be weighed a few practical considerationswhich vary from agent to agent.

First, if a slow or mechanically controlled injec-tion is to be used, the agent must remain stablein the syringe for the duration of the infusion.Levovist, which is quite viscous after mixing,

Bolus: pros and cons

Pro Con

Easy to perform Contrast effect short lived

Highest peak Contrast effect changing enhancement during study

Wash-in and Timing of bolus difficultwash-out visible

Agent is used quickly: Comparative contrast no stability problems studies difficult

Fig. 16 Integrated fractional enhancement (definedas the area under the curves of Figure 13) plottedas a function of dose for Optison. A steadily increas-ing response is seen as the administered volume ofthe agent is increased.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 14

Page 30: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

provides this more easily than, say Optison, inwhich emulsion the bubbles tend to float.Agitating the preparation while it is in thesyringe is in practice difficult. On the otherhand, some perfluorocarbon agents such asDefinity are effective in very high dilution,

rendering them ideal for injection into a dripbag and drip infusion. This cannot be attempt-ed with Levovist because of the adverse effect of dilution on the bubble population. Finally,some agents which are effective at very lowvolumes require whole body doses of much less

15Principles and Instrumentation

Fig. 17 Bolus vs infusion. (a) A bolus injection of SonoVue gives a wash-out time of about 1 minute.(b) Infused, the agent delivers uniform enhancement over a period of about 7 minutes.

a) b)

Fig. 18 The enhancement provided by a slowmanual injection of Levovist. Enhancement is seento last for approximately nine minutes.

Infusion: pros and cons

Pro Con

Extends time More complex to perform:of enhancement may require pump

Provides consistent Titration of infusion effect takes time and effort

Avoids blooming/ Stability of agent overartifacts infusion period

can be a problem

Dose is optimised:agent is used more

efficiently

Allows negative bolus method for quantitationof perfusion (see Ch4)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 15

Page 31: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

than 1ml, rendering infusion difficult and slowinjection virtually impossible. Others, whichare fundamentally unstable out of the body,such as Echogen, offer no obvious means forinfusion or slow injection. Clearly, the manu-facturers’ advice should be sought beforeattempting infusion in a clinical setting. TheTable below shows when an infusion issessential to contrast examination.

1.3 Mode of action

The interaction of an ultrasound beam with apopulation of bubbles is a process whose sub-tlety and complexity has only recently beenrecognised. Understanding what happenswhen a microbubble contrast agent is exposedto an ultrasound beam is the key to understand-ing – and using – new clinical methods forcontrast imaging and thus the key to interpret-ing a clinical contrast echocardiographic study.

A sound field comprises a train of travellingwaves, much like ripples on a pond. The fluidpressure of the medium (in this case tissue)changes as the sound propagates through it. A

gas bubble is highly compliant and hence issquashed when the pressure outside it is raisedand expanded when the pressure is lowered. Ata typical clinical frequency of 2 MHz, forexample, a bubble sitting in an acoustic fieldundergoes this oscillatory motion two milliontimes per second. As it moves in this way, thebubble becomes a source of sound that radiatesradially from its location in the body, as would

ripples on a pond from a small object movingat a point on its surface (Figure 19). The soundthat reaches the transducer from this bubble,combined with that from all of its neighbours,is what constitutes the scattered echo from acontrast agent. Characterising this echo so that

16 Handbook of Contrast Echocardiography

When is an infusion necessary?

Infusion mandatory Bolus sufficient

Myocardial perfusion: Myocardial perfusion: rest and stress studies, rest studies only,

quantitative analysis qualitative assessment

Coronary flow reserve LVO: rest or stress,thrombus detection

Pulmonary veinDoppler enhancement

Fig. 19 A bubble in an acoustic field responds tothe changes in pressure which constitute the soundwave by changing in size. The radius oscillates atthe same rate as the incident sound, radiating anecho.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 16

Page 32: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

17Principles and Instrumentation

it can be differentiated from those of ordinarytissue such as the cardiac muscle is the basis ofcontrast specific imaging modes such as pulseinversion Doppler or harmonic imaging.

1.3.1 Bubble behaviour and incident pressure

Unlike tissue, a bubble does not scatter in thesame way if it is exposed to weak (that is lowamplitude) sound, than to strong, high ampli-tude sound. Instead, there are three broadregimes of scattering behaviour that depend onthe peak pressure of the incident sound fieldproduced by the scanner (see Table). These areused in different ways in contrast imaging ofthe heart.

Looking at the Table, we see that at low inci-dent pressures (corresponding to low transmitpower of the scanner), the agents producelinear backscatter enhancement, resulting in anaugmentation of the echo from blood. This isthe behaviour originally envisaged by the

contrast agent manufacturers for their firstintended clinical indication: Doppler signalenhancement. It is this for which they obtainedapproval from the regulatory authorities, anindication which does not require any changeto ultrasound imaging and Doppler instru-mentation. As the transmit intensity control of the scanner is increased and the pressureincident on a bubble goes beyond about50–100 kPa, which is still below the level usedin most diagnostic scans, the contrast agentbackscatter begins to show nonlinear characte-ristics, such as the emission of harmonics. It isthe detection of these that forms the basis ofcontrast specific imaging modes such asharmonic and pulse inversion imaging andDoppler. Finally, as the peak pressure passesabout 1 MPa, near the maximum emitted by atypical echo imaging system, many agents exhibit transient nonlinear scattering. Thisforms the basis of triggered imaging and moststrategies for detection of myocardial perfusion.The Table shows these three regimes togetherwith the imaging methods developed to exploitthem. It should be noted that in practice,

Three regimes of bubble behaviour in an ultrasound field

Peak pressure Mechanical Bubble Acoustic Clinical(approx) Index (MI) behaviour behaviour Application

@ 1MHz

< 100 kPa < 0.1 Linear Backscatter Coronary artery Doppler,oscillation enhancement PV Doppler, fundamental

B-mode LVO

100 kPa–1 MPa 0.1 – 1.0 Nonlinear Harmonic Coronary artery Doppler, oscillation backscatter harmonic B-mode LVO,

real-time perfusion imaging

> 1 MPa > 1.0 Disruption Transient Power Doppler LVO, harmonic echoes intermittent perfusion

imaging

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 17

Page 33: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

because of the different sizes present in a real-istic population of bubbles (19), the bordersbetween these behaviours are not sharp. Norwill they be the same for different agent types,whose acoustic behaviour is strongly depend-ent on the gas and shell properties (20). In thenext section we examine the clinical imagingmethods which rely on each of these behav-iours in turn.

1.3.1.1 The Mechanical Index (MI)For reasons unrelated to contrast imaging,ultrasound scanners marketed in the US arerequired by the Food and Drugs Adminis-tration (FDA) to carry an on-screen label ofthe estimated peak negative pressure to whichtissue is exposed. Of course, this pressurechanges according to the tissue through whichthe sound travels as well as the amplitude andgeometry of the ultrasound beam: the higherthe attenuation, the less the peak pressure intissue will be. A scanner cannot ‘know’ whattissue it is being used on, so the definition ofan index has been arrived at which reflects theapproximate exposure to ultrasound pressure atthe focus of the beam in an average tissue. TheMechanical Index (or ‘MI’) is defined as thepeak rarefactional (that is, negative) pressure,divided by the square root of the ultrasoundfrequency. This quantity is related to theamount of mechanical work that can be per-formed on a bubble during a single negativehalf cycle of sound (21). In clinical ultrasoundsystems, this index usually lies somewherebetween 0.1 and 2.0. Although a single value isdisplayed for each image, in practice the actualMI varies throughout the image. In theabsence of attenuation, the MI is maximal atthe focus of the beam. Attenuation shifts thismaximum towards the transducer. In a phasedarray, steering reduces the intensity of theultrasound beam so that the MI is also less at

the edges of the sector. Furthermore, because itis a somewhat complex procedure to calculatethe index, which is itself only an estimate ofthe actual quantity within the body, the indicesdisplayed by different machines are notprecisely comparable. Thus, for example, morebubble disruption might be observed at a dis-played MI of 1.0 using one machine than onthe same patient using another. For this reason,recommendations of machine settings for aspecific examination will include MI valuespeculiar to a given ultrasound manufacturer’sinstrument. Nonetheless, the MI is one of themost important machine parameters in acontrast echo study. It is usually controlled bymeans of the ‘output power’ setting of the

scanner.

1.3.2 I – Linear backscatter: Doppler enhancement

Although bubbles of a typical contrast agentare smaller than red blood cells, and theirvolume concentration in the blood followingintravenous injection is a fraction of a percent,the amplitude of the echo from the micro-bubble agent eclipses that of the blood itself.

18 Handbook of Contrast Echocardiography

The Mechanical Index (MI)

• Defined by MI = , where Pneg is the peak negative ultrasound pressure, f the ultrasound frequency

• Reflects the normalised energy to which a target (such as a bubble) is exposed in an ultrasound field

• Is defined for the focus of the ultrasound beam• Varies with depth in the image (lessens with

increasing depth)• Varies with lateral location in the image (lessens

towards the sector edges)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 18

Page 34: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

19Principles and Instrumentation

This is because the weakness of the echofrom blood originates from the cells them-selves, which are poor scatterers of ultrasound.Their acoustic impedance is almost identicalto that of the surrounding plasma. A bubblecontaining compressible gas, on the otherhand, presents a strong discontinuity inacoustic impedance and hence acts as a strongreflector. Size for size, a bubble is about onehundred million million times more effectiveat scattering ultrasound. Thus the injection ofa relatively sparse population of bubbles intothe bloodstream results in a substantial enhance-ment of the blood echo. In a Doppler examina-tion, the arrival of the contrast agent, someseconds after peripheral venous injection, inthe portion of the systemic vasculature underinterrogation is marked by a dramatic increasein signal strength. In spectral Doppler this isseen as an intensifying of the greyscale of thespectrum (Figure 2), whose enhancement isrelated to dose and whose duration depends onthe method of administration. A 10 ml bolusof Levovist at a concentration of 400 mg/ml,for example, provides a peak enhancement ofabout 25 dB with a usable wash-out period ofabout 2–4 minutes. Spectral Doppler examina-tions of, for example, the pulmonary vein, thatfail because of lack of signal strength, can be‘rescued’ by the contrast examination, detect-ing signals that would be otherwise obscuredby noise.

1.3.2.1 Enhancement studies with conventional imaging

In the initial studies which were carried out toobtain regulatory approval of ultrasoundcontrast agents, the sole indication was tosalvage a nondiagnostic Doppler examination.These demonstrated the capacity of contrastagents to increase the technical success rate ofDoppler assessments of aortic stenosis, mitral

regurgitation and pulmonary venous flow, inwhich latter case it rose from 27 percent to 80percent (22).

Echo enhancement caused by the agent may beconsidered in other ways too. For example,given a satisfactory transthoracic colourDoppler study, one use of the agent might besimply to enable a higher ultrasound frequencyto be used, exploiting the agent to counter thehigher tissue attenuation. In such a case, thecontrast enhancement translates into higherspatial resolution (Figure 3). Alternatively, thecolour system may be set to use fewer pulsesper scan line (that is a lower ensemble length)while still achieving the same sensitivity toblood flow by means of the contrast enhance-ment. The agent will then provide the userwith a higher frame rate.

With conventional greyscale imaging, enhance-ment might be seen in lumina of the ventriclesor large vessels if the concentration of thebubbles is sufficiently high, as it is for per-fluorocarbon gas agents such as Optison andDefinity. The contrast is not, however, normal-ly seen in the small vessels within the muscle ofthe myocardium itself. This is because the10–25 dB of enhancement provided by theagent still leaves the blood echo some10–20 dB below that of the echogenic tissue ofthe heart wall. In order to enhance the visiblegrey level, either higher concentrations ofbubbles must be achieved or specific, newimaging strategies employed. A greater numberof bubbles, however, inevitably leads to higherattenuation of the sound beam as it traversesthe cavities, so perfusion imaging cannot beeffective using conventional imaging (23).Thus new, bubble specific imaging methodsare necessary.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 19

Page 35: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

1.3.3 II – Nonlinear backscatter: harmonic imaging

1.3.3.1 The need for bubble-specific imagingIn many of the potential applications of contrastagents, one might ask whether it is possible tocontinue to increase the amount of agentsinjected and obtain progressively strongerechoes from blood; to the point, for example,where the myocardium becomes visible on agreyscale image. Unfortunately, attenuation ofthe ultrasound beam by the agent in the cavityalso increases with bubble concentration, withthe result that shadowing occurs distal to theagent and the myocardium disappears alto-gether. Because of this limitation of the useableconcentration of the agent, we are generally leftwith small enhancements in the myocardiumecho that must be identified against the strongbackground echo from the solid tissue itself. X-ray angiography, which is faced with a similarproblem after dye is injected into the blood-stream, deals with these ‘clutter’ components ofthe image by simple subtraction of a pre-injec-tion image. What is left behind reveals flow inindividual vessels or the ‘blush’ of perfusion atthe tissue level. If, however, we subtract twoconsecutive ultrasound images of a solid organ,we get a third ultrasound image of the sameorgan, produced by the decorrelation of thespeckle pattern between acquisitions. In order toshow parenchymal enhancement due to theagent, speckle variance must first be reduced byfiltering, with a consequent loss of spatial ortemporal resolution. Even if the speckle pro-blem could be overcome, subtraction would stillbe poorly suited to the dynamic and interactivenature of cardiac ultrasound imaging. InDoppler modes, the problem of the moving wallinterference (the clutter) prevents the smallerecho from the blood itself being detected, asdiscussed in §1.1.3.

How then might contrast agents be used toimprove the visibility of blood in movingvascular structures such as the myocardium?Clearly, a method that could identify the echofrom the contrast agent and suppress the echofrom solid tissue would provide both a realtime ‘subtraction’ mode for contrast-enhancedB-mode imaging, and a means of suppressingDoppler clutter without the use of a velocity-dependent filter in spectral and colour modes.Nonlinear imaging aims to provide such amethod, and hence the means for detection offlow in smaller vessels than is currently poss-ible.

1.3.3.2 Harmonic imagingExamining the behaviour of contrast-enhancedultrasound studies reveals two important piecesof evidence. First, the size of the echo enhance-ment at very high dilution following a smallperipheral injection (7 dB from as little as0.01 ml/kg of Levovist, for example (24)) ismuch larger than would be expected from suchsparse scatterers of this size in blood. Second,investigations of the acoustic characteristics ofseveral agents (25) have demonstrated anapproximately linear dependence of back-scattered coefficient on numerical density ofthe agent at low concentrations, as expected,but a dependence of attenuation on ultrasoundfrequency different to that predicted by theRayleigh law, which describes how the echo-genicity of normal tissue changes with fre-quency. Instead, peaks exist which are depen-dent on both ultrasound frequency and the sizeof the microbubbles, suggestive of resonancephenomena. This important observationsuggests that the bubbles resonate in the ultra-sound field. As the ultrasound wave – whichcomprises alternate compressions and rarefac-tions – propagates over the bubbles, they ex-perience a periodic change in their radius in

20 Handbook of Contrast Echocardiography

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 20

Page 36: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

21Principles and Instrumentation

sympathy with the oscillations of the incidentsound. Like vibrations in other structures,these radial oscillations have a natural – orresonant – frequency of oscillation at whichthey will both absorb and scatter ultrasoundwith a peculiarly high efficiency. Consideringthe linear oscillation of a free bubble of air inwater, we can use a simple theory (1) to predictthe resonant frequency of radial oscillation of abubble of 3 µm diameter, the median diameterof a typical transpulmonary microbubbleagent. As Figure 20 shows, it is about 3 MHz,

approximately the centre frequency of ultra-sound used in a typical echocardiographyscan. This extraordinary – and fortunate –coincidence explains why ultrasound contrastagents are so efficient and can be administeredin such small quantities. It also predicts thatbubbles undergoing resonant oscillation in an ultrasound field can be induced to emit harmonics, the basis of harmonic imaging.

One consequence of this extraordinary coinci-dence is that bubbles undergoing resonantoscillation in an ultrasound field can be inducedto nonlinear motion. It has long been recog-nised (26) that if bubbles are ‘driven’ by theultrasound field at sufficiently high acousticpressures, the oscillatory excursions of thebubble reach a point where the alternateexpansions and contractions of the bubble’ssize are not equal. Lord Rayleigh, the origina-tor of the theoretical understanding of soundupon which ultrasound imaging is based, wasfirst led in 1917 to investigate this by hiscuriosity over the creaking noises that histeakettle made as the water came to the boil.The consequence of such nonlinear motion is

Fig. 20 Microbubbles resonate in a diagnosticultrasound field. This graph shows that the reson-ant – or natural – frequency of oscillation of abubble of air in an ultrasound field depends on itssize. For a 3.5 µm diameter, the size needed for anintravenously injectible contrast agent, the reson-ant frequency is about 3 MHz.

Fig. 21 Higher incident pressures induce non-linear behaviour in a bubble. Asymmetric oscillationof a resonant bubble in an ultrasound field givesrise to an echo with even harmonics.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 21

Page 37: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

that the sound emitted by the bubble, anddetected by the transducer, contains harmonics,just as the resonant strings of a musical instru-ment, if plucked too vigorously, will produce a ‘harsh’ timbre containing overtones (themusical term for harmonics). The origin of thisphenomenon is the asymmetry which beginsto affect resonant oscillation as the amplitudebecomes large. As a bubble is compressed bythe ultrasound pressure wave, it becomes stifferand hence resists further reduction in itsradius. Conversely, in the rarefaction phase ofthe ultrasound pulse, the bubble becomes lessstiff, and, therefore, enlarges much more(Figure 21). Figure 22 shows the frequencyspectrum of an echo produced by a micro-bubble contrast agent following a 3.75 MHzburst. The particular agent is Levovist, though

many microbubble agents behave in a similarway. Ultrasound frequency is on the horizontalaxis, with the relative amplitude on the verticalaxis. A strong echo, at -13 dB with respect tothe fundamental, is seen at twice the transmit-ted frequency, the second harmonic. Peaks inthe echo spectrum at sub- and ultraharmonicsare also seen. Here, then, is one potentialmethod to distinguish bubbles from tissue:excite them so as to produce harmonics anddetect these in preference to the fundamentalecho from tissue. Key factors in the harmonicresponse of an agent, which varies frommaterial to material, are the incident pressureof the ultrasound field, the frequency, as well asthe size distribution of the bubbles and themechanical properties of the bubble capsule.A stiff capsule, for example, will dampen the

22 Handbook of Contrast Echocardiography

Fig. 22 Harmonic emission from Levovist. A sample of a contrast agent is insonated at 3.75 MHz and theecho analyzed for its frequency content. It is seen that most of the energy in the echo is at 3.75 MHz, butthat there is a clear second peak in the spectrum at 7.5 MHz, as well as a third at 1.875 MHz. The secondharmonic echo is only 13 dB less than that of the main, or fundamental echo. Harmonic imaging andDoppler aims to separate and process this signal alone. The smaller peak is the first subharmonic.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 22

Page 38: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

23Principles and Instrumentation

oscillations and attenuate the nonlinear re-sponse.

Harmonic B-mode imagingA new real-time imaging and Doppler methodbased on this principle, called harmonic imag-ing (27), is now widely available on most

modern echocardiography ultrasound scanners.In harmonic mode, the system transmitsnormally at one frequency, but when in har-monic mode is tuned to receive echoes pre-ferentially at double that frequency, where theechoes from the bubbles lie. Typically, thetransmit frequency lies between 1.5 and3 MHz and the receive frequency is selected bymeans of a bandpass filter whose centrefrequency is at the second harmonic between 3and 6 MHz (Figure 23). Harmonic imaginguses the same array transducers as conventionalimaging and for in most of today’s ultrasoundsystems involves only software changes. Echoesfrom solid tissue, as well as red blood cellsthemselves, are suppressed. Real-time har-monic spectral Doppler and colour Dopplermodes have also been implemented (some-times experimentally) on a number of com-mercially available systems. Clearly, an excep-tional transducer bandwidth is needed tooperate over such a large range of frequencies.Fortunately much effort has been directed inrecent years towards increasing the bandwidthof transducer arrays because of its significantbearing on conventional imaging performance,so harmonic imaging modes do not require theadditional expense of dedicated transducers.

Harmonic Doppler imagingIn harmonic images, the echo from tissuemimicking material is reduced – but not elim-inated, reversing the contrast between theagent and its surrounding (Figure 24). Thevalue of this effect is to increase the conspicuityof the agent when it is in blood vessels normal-ly hidden by the strong echoes from tissue. Inspectral Doppler, one would expect the sup-pression of the tissue echo to reduce the tissuemotion ‘thump’ that is familiar to all echo-cardiographers. Figure 25 shows spectralDoppler applied to a region of the aorta in

Fig. 23 The principle of harmonic imaging andDoppler. A conventional phased array is used, withthe receiver tuned to double the transmittingfrequency. The tissue and blood give an echo at thefundamental frequency, but the contrast agentundergoing nonlinear oscillation in the sound fieldemits the harmonic which is detected by the har-monic system.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 23

Page 39: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

which there is wall motion as well as bloodflow within the sample volume. The conven-tional Doppler image of Figure 25 a shows thethump artifact due to clutter, which is almostcompletely absent in the harmonic Dopplerimage of Figure 25 b. All instrument settings,including the filters, are identical in theseimages; we have merely switched betweenmodes. In vivo measurements from spectral

Doppler show that the signal-to-clutter ratio isimproved by a combination of harmonic imag-ing and the contrast agent by as much as 35 dB(28). Figure 26 shows harmonic colour imagesof an aorta, this time with flash artifact fromrespiratory motion. The harmonic imagedemonstrates the flow without the flashartifact. The potential application of this diag-nostic method is to detect blood flow in small

24 Handbook of Contrast Echocardiography

Fig. 24 Harmonic imaging. An vitro phantom shows a finger-like void in tissue equivalent material, filledwith dilute contrast agent, barely visible in conventional mode (a). Harmonic imaging is seen to reversethe contrast between the microbubble agent and the surrounding material (b).

Fig. 25 Clutter rejection with harmonic spectral Doppler. (a) The abdominal aorta of an animal is ex-amined with harmonic spectral Doppler. In conventional mode, clutter from the moving wall causes thefamiliar artifact which also obscures diastolic flow. (b) In harmonic mode, the clutter is almost completelysuppressed, so that flow can be resolved. The settings of the filter and other relevant instrument para-meters are identical.

a) b)

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 24

Page 40: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

25Principles and Instrumentation

Fig. 26 Clutter rejection with harmonic colour Doppler. (a) Flow in the abdominal aorta is superimposedon respiratory motion, producing severe flash artifact. (b) In harmonic mode at the same point in therespiratory cycle, the flash artifact disappears. Flow from a smaller vessel (the cranial mesenteric artery)is visualised. All instrument settings are the same.

vessels surrounded by tissue which is moving:in the branches of the coronary arteries (29) orthe myocardium itself (30), as well as in theparenchyma of abdominal organs (31).

Harmonic power Doppler imaging

In colour Doppler studies using a contrastagent, the effect of the arrival of the agent in acolour region of interest is often to produce‘blooming’ of the colour image, wherebysignals from major vascular targets spread outto occupy the entire region. Thus, althoughflow from smaller vessels might be detectable,the colour images can be swamped by artifac-tual flow signals. The origin of this artifact isthe amplitude thresholding that governs mostcolour displays in conventional (or ‘velocity’)mode imaging. Increasing the backscatteredsignal power simply has the effect of displayingthe velocity estimate, at full intensity, over awider range of pixels around the detected loca-tion. A display in which the parameter mappedto colour is related directly to the backscatteredsignal power, on the other hand, has the advan-tage that such thresholding is unnecessary andthat lower amplitude Doppler shifts, such as

those which result from sidelobe interference,are displayed at a lower visual amplitude,rendering them less conspicuous. Echo enhanc-ed flow signals, in contrast, will be displayed at a higher level. This is the basis of the powerimaging map (also known as colour powerangiography, or colour Doppler energy map-ping). Power Doppler can help eliminate someother limitations of small vessel flow detectionwith colour Doppler. Low velocity detectionrequires lowering the Doppler pulse repetitionfrequency (PRF), which results in multiplealiasing and loss of directional resolution. Adisplay method that does not use the velocityestimate is not prone to the aliasing artifact,and therefore allows the PRF to be lowered andhence increase the likelihood of detection ofthe lower velocity flow from smaller vessels.

Because it maps a parameter directly related tothe acoustic quantity that is enhanced by thecontrast agent, the power map is a naturalchoice for contrast enhanced colour Dopplerstudies. However, the advantages of the powermap for contrast enhanced detection of smallvessel flow are balanced by a potentially devas-

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 25

Page 41: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

26 Handbook of Contrast Echocardiography

tating shortcoming: its increased susceptibilityto interference from clutter. Clutter is bothdetected more readily, because of the powermode’s increased sensitivity, and displayedmore prominently, because of the high intensitydisplay of high amplitude signals. Further-more, frame averaging has the additional effectof sustaining and blurring the flash over thecardiac cycle, so exacerbating its effect on theimage. This is the reason that conventionalpower mode, while quite popular in radiologi-cal and peripheral vascular ultrasound imaging,has almost no role in echocardiography.

At the small expense of some sensitivity, amplycompensated by the enhancement caused bythe agent, harmonic mode effectively over-comes this clutter problem (Figure 27).Combining the harmonic method with powerDoppler produces an especially effective toolfor the detection of flow in the small vessels ofthe organs of the abdomen which may bemoving with cardiac pulsation or respiration.In a study in which flow imaged on contrastenhanced power harmonic images was

compared with histologically sized arterioles inthe corresponding regions of the renal cortex(24), it was concluded that the method iscapable of demonstrating flow in vessels of lessthan 40µm diameter; about ten times smallerthan the corresponding imaging resolutionlimit, even as the organ was moving withnormal respiration. Recent studies of thispower mode method in the heart (32) showthat flow can be imaged in the myocardiumwith an agent such as Levovist (33).

1.3.3.3 The impact of harmonic imagingHarmonic imaging succeeds in identifyingmicrobubble contrast agent in the tissue vascu-lature by means of its echo ‘signature’. In doingso it helps tackle some old problems in ultra-sound, such as the rejection of the tissue echoin Doppler modes designed to image movingblood, and creating a ‘subtraction’ mode with-out sacrificing the real time nature of theexamination. Figure 28 summarises graphicallythe effect of contrast agents and harmonicdetection on the Doppler process. In conven-tional Doppler, the signal from blood is larger

Fig. 27 Reduction of the flash artifact in harmonic power Doppler. The harmonic contrast method helpsovercome one of the principal shortcomings of power Doppler, its increased susceptibility to tissue motion.(a) Aortic flow in power mode with flash artifact from cardiac motion of wall. (b) In harmonic mode at thesame point in the cardiac cycle, the flash is largely suppressed. All instrument settings are the same.

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 26

Page 42: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

27Principles and Instrumentation

than the clutter signal from tissue. In contrastenhanced Doppler, the signal from blood israised, sometimes to near that of tissue. Withharmonic mode, the signal from blood is raisedbut that from the tissue reduced, so reversingthe contrast between tissue and blood. Another

way of looking at the harmonic method is that,because of its greater sensitivity to small quan-tities of agent, a given level of enhancementdue to a bolus will last longer (Figure 29). Thisis the reason that a number of investigatorshave found that cardiac contrast imaging inany mode is generally more effective with theharmonic method (29, 30). Harmonic imaginghas been shown to render possible the detec-tion of microvessels containing contrast agentseven in the presence of highly echogenic tissue,such as the liver (31) and myocardium (34).

Harmonic imaging demands exceptional per-formance from the transducer array and systembeamformer. Its implementation forces impli-cit compromises between, for example, imageresolution and rejection of the non-contrastagent echo. It places unusual demands on thebandwidth performance of transducers, as wellas the flexibility of the architecture of the imag-ing system. The ease with which it has beendeveloped on modern instruments, however,

Fig. 28 Quantifying the impact of harmonicDoppler. Clutter and spectral Doppler signal levelsmeasured in conventional, contrast enhanced, andharmonic contrast enhanced modes.

Fig. 29 The improved sensitivity of the harmonic method translates into increased useful imaging timefrom a contrast agent bolus. Time-enhancement curve caused by circulation of contrast agent bolus showshow harmonic imaging, by reducing the detection threshold, increases the imaging lifetime of the agent.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 27

Page 43: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

28 Handbook of Contrast Echocardiography

reflects this flexibility and augurs well for thefuture of the method. More significantly,contrast agents are now being developed speci-fically with nonlinear response as a designcriterion. Entirely new agents present oppor-tunities for entirely new detection strategies(17). Our laboratory measurements show thatsome new contrast agents are capable of creat-ing an echo with more energy in the secondharmonic than at the fundamental: that is,they are more efficient in harmonic than con-ventional mode. With such agents, nonlinearimaging is the preferred clinical method fortheir detection.

1.3.3.4 Tissue harmonic imagingIn second harmonic imaging, an ultrasoundscanner transmits at one frequency and receivesat double this frequency. The resultingimproved detection of the microbubble echo is

due to the peculiar behaviour of a gas bubblein an ultrasound field. However, any source ofa received signal at the harmonic frequencywhich does not come from the bubble, willclearly reduce the efficacy of this method.

Such unwanted signals can come from non-linearities in the transducer or its associatedelectronics, and these must be tackled effec-tively in a good harmonic imaging system.However, tissue itself can produce harmonicswhich will be received by the transducer. Theyare developed as a wave propagates throughtissue. Again, this is due to an asymmetry: thistime the fact that sound travels slightly fasterthrough tissue during the compressional partof the cycle (where it is denser and hence morestiff ) than during the rarefactional part.Although this effect is very small, it is sufficientto produce substantial harmonic componentsin the transmitted wave by the time it reachesdeep tissue, so that when it is scattered by alinear target such as the myocardium, there is aharmonic component in the echo, which isdetected by the scanner along with the har-monic echo from the bubble (35). This is thereason that solid tissue is not completely darkin a typical harmonic image. The effect is toreduce the contrast between the bubble andtissue, rendering the problem of detecting per-fusion in the myocardium more difficult.

Tissue harmonics, though a foe to contrastimaging, are not necessarily a bad thing. Infact, an image formed from tissue harmonicswithout the presence of contrast agents hassome properties which recommend it over con-ventional imaging. These follow from the factthat tissue harmonics are developed as thebeam penetrates tissue, in contrast to the con-ventional beam, which is generated at thetransducer surface (36). Artifacts which accrue

Application

LVO; myocardial perfusion (offline subtraction necessary)

Coronary flow reserve;pulmonary veinDoppler enhancement

Coronary vesselimaging

LVO; myocardialperfusion

Usual Format

Greyscale displayof harmonicechoes

Spectral Dopplerwith tissue motionsuppression

Colour Dopplerwith tissue sup-pression, conven-tional B-modebackground

Power Dopplerwith tissue motionsuppression,conventional B-mode background

Method

HarmonicB-mode

Harmonic spectral Doppler

Harmonic colour Doppler

Harmonic power Doppler

Harmonic imaging modalities

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 28

Page 44: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

29Principles and Instrumentation

from the first few centimetres of tissue, such asreverberations, are reduced by using tissue har-monic imaging. Sidelobe and other low-levelinterference is also suppressed, making tissueharmonic imaging the routine modality ofchoice for many echocardiographers (Figure 30).

For contrast studies, the tissue harmonic limitsthe visibilty of bubbles within tissue in a B-mode harmonic image and therefore can beconsidered an artifact. In contemplating howto reduce it, it is instructive to bear in minddifferences between harmonics produced by

tissue propagation and by bubble echoes. First,tissue harmonics require a high peak pressure,so are only evident at high MI. Reducing theMI leaves only the bubble harmonics. Second,tissue harmonics become greater at greaterdepths, whereas harmonics from bubbles aredepth independent. Finally, harmonics fromtissue at high MI are continuous and sustained,whereas those from bubbles are transient innature as the bubble disrupts.

1.3.3.5 Pulse inversion imagingThe method we have described above for har-monic imaging imposes some fundamentallimitations on the imaging process whichrestrict its clinical potential in organ imaging.First, in order to ensure that the higher fre-quencies are due only to harmonics emitted bythe bubbles, the transmitter must be restrictedto a band of frequencies around the funda-mental (Figure 31 a). Similarly, the receivedband of frequencies must be restricted to thoselying around the second harmonic. If these tworegions overlap (Figure 31 b), the result will bethat the harmonic filter will receive echoes

Potential origin of echoesat harmonic frequency

• Nonlinear scattering from bubbles• Nonlinear propagation of echoes from solid tissue• Inadvertent transmission of harmonics by non-

linearities in transmitter or transducer• Inadvertent production of harmonics by non-

linearities in receiver or transducer• Deliberate transmission at second harmonic

frequency to improve bandwidth of image

Fig. 30 Tissue harmonic imaging of the heart. Note that in comparison to the fundamental image (a), thetissue harmonic image (b) has a reduced level of artifacts associated with reverberation and sidelobe inter-ference.

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 29

Page 45: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

30 Handbook of Contrast Echocardiography

from ordinary tissue, thus reducing the con-trast between the agent and tissue. However,restricting the receive bandwidth degrades theresolution of the resulting image, thus framinga fundamental compromise in harmonic imag-ing between contrast and resolution. For opti-mal detection of bubbles in the microvascula-ture, this compromise must favour contrast, sothat the most sensitive harmonic images aregenerally of low quality. In other applicationswhere the bubble echo has a higher intensity,such as in the ventricular cavities, higher band-widths can be tolerated. A further drawback ofthe filtering approach to harmonic imaging isillustrated in Figure 31 c. There it can be seenthat if the received echo is weak, the over-lapping region between the transmit andreceived frequencies becomes large relative tothe whole received signal. This means thatcontrast in the harmonic image is dependenton how strong the echo is from the bubbles,which is determined by the concentration ofbubbles and the intensity of the incident ultra-sound pulse. In practice, this dictates thatscanning in harmonic mode must be per-formed with a high mechanical index, that is,using the maximum transmit power of thesystem. This results in the transient andirreversible disruption of the bubbles (37).Consequently, as the bubbles enter the scanplane of a real-time ultrasound image, theyprovide an echo but then disappear. Thus ves-sels that lie within the scan plane are not vis-ualised as tubular structures in a typical har-monic image, but instead have a punctateappearance.

Principle of pulse inversion Pulse inversion imaging overcomes the conflictbetween the requirements of contrast andresolution in harmonic imaging and providesgreater sensitivity, thus allowing low incident

Fig. 31 a) The compromises forced by harmonicimaging: In harmonic imaging, the transmittedfrequencies must be restricted to a band aroundthe fundamental, and the receive frequencies mustbe limited to a band around the second harmonic.This limits resolution.

Fig. 31 b) If the transmit and receive bandwidthsare increased to improve resolution, some funda-mental echoes from tissue will overlap the receivebandwidth and will be detected, reducing contrastbetween agent and tissue.

Fig. 31 c) When the harmonic echoes are weak,due to low agent concentration and/or low incidentpulse intensity, this overlap will be especially large,and the harmonic signal may be largely composedof tissue echoes.

Kardio-Kap 1 for pdf 06.06.2000 16:56 Uhr Seite 30

Page 46: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

31Principles and Instrumentation

power, nondestructive, continuous imaging ofmicrobubbles in an organ such as the liver. Themethod also relies on the asymmetric oscillati-on of an ultrasound bubble in an acoustic field,but detects all nonlinear (or, ‘even’) compo-nents of the echo, over the entire bandwidth ofthe transducer. In pulse inversion imaging, twopulses are sent in rapid succession into thetissue. The second pulse is a mirror image ofthe first (Figure 32). That is, it has undergonea 180° phase change. The scanner detects theecho from these two successive pulses andforms their sum. For ordinary tissue, whichbehaves in a linear manner, the sum of twoinverted pulses is simply zero. For an echo withnonlinear components, such as that from abubble, on the other hand, the echoes produ-ced from these two pulses will not be simplemirror images of each other, because of the

asymmetric behaviour of the bubble radiuswith time. The result is that the sum of thesetwo echoes is not zero. Thus, a signal is detec-ted from a bubble but not from tissue. It canbe shown by mathematical analysis that thissummed echo contains the nonlinear ‘even’harmonic components of the signal, includingthe second harmonic (38).

One advantage of pulse inversion over the filterapproach to detect harmonics from bubbles isthat it no longer suffers from the restriction ofbandwidth. The full frequency range of soundemitted from the transducer can be detected inthis way, and a full bandwidth, that is highresolution, image of the nonlinear echoes frombubbles may be formed in real time. The pricepaid, using current technology, is a reductionof the effective frame rate by a factor of two.

Fig. 32 Principle of Pulse Inversion Imaging. A pulse of sound is transmitted into the body and echoesare received from agent and tissue. A second pulse, which is an inverted copy of the first pulse, is thentransmitted in the same direction and the two echoes are summed. Linear echoes from tissue will beinverted copies of each other and will cancel to zero. The microbubble echoes are distorted copies of eachother, and the nonlinear components of these echoes will reinforce each other when summed, producinga strong harmonic signal.

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 31

Page 47: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

32 Handbook of Contrast Echocardiography

One application for this advantage is in tissueimaging without contrast agents. In conventio-nal imaging, reverberation, phase aberrationand sidelobe artifacts caused by the linearsound beam can reduce image contrast,especially in hypoechoic structures such as theventricular cavities. Tissue harmonic imagingcan reduce these artifacts and improve imagecontrast, but with a somewhat degraded resolu-tion. Figure 37 illustrates how pulse inversionimaging provides better suppression of linearechoes than harmonic imaging and is effectiveover the full bandwidth of the transducer,showing improvement of image resolution overharmonic mode.

Because this detection method is a more effi-cient means of isolating the bubble echo,weaker echoes from bubbles insonated at low,nondestructive intensities, can be detected.Figure 33 shows a dramatic improvement incontrast sensitivity obtained over both funda-

mental and harmonic modes by pulse inver-sion imaging in a phantom. For LVO, pulseinversion imaging provides improved resolu-tion of the endocardium (Figure 34). At highMI, bubbles are disrupted and their contents

Fig. 33 Demonstration of pulse inversion imaging. In vitro images of a vessel phantom containingstationary Optison surrounded by tissue equivalent material (biogel and graphite). a) Conventional image,MI = 0.2. b) Harmonic imaging, MI = 0.2, provides improved contrast between agent and tissue. c) Pulseinversion imaging, MI = 0.2. By suppressing linear echoes from stationary tissue, pulse inversion imagingprovides better contrast between agent and tissue than both conventional and harmonic imaging.

Fig. 34 Pulse inversion imaging with Levovistshowing LV opacification. Notice that perforatingbranches of the coronary vessels are visible.Courtesy Eric Yu, University of Toronto, Canada

a) b) c)

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 32

Page 48: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

33Principles and Instrumentation

Fig. 35 Pulse inversion imaging showing coronary vessels using Levovist at two frames per heart beat.(a) Agent enters cavity. (b) Perfusion, 6 seconds later. (c) One second later, vessels are seen entering themyocardium.

Fig. 36 The appearance of blood vessels in harmonic and pulse inversion imaging in a liver. In a harmonic image of a liver following an injection of Optison, large vessels have a punctate appearanceas the high MI ultrasound disrupts the bubbles as they enter the scan plane (a). In the pulse inversionimage of the same liver, a lower MI can be used so that continuous vessels are now seen. Improved resolu-tion of pulse inversion imaging demonstrates 4th order branches of the portal vein (b).

a) b) c)

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 33

Page 49: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

34 Handbook of Contrast Echocardiography

Fig. 37 Tissue harmonic imaging. In conventionalimage (a), reverberation, phase aberration andsidelobe artifacts degrade depiction of cavityborders. Tissue harmonic image (b) shows reducedartifacts, but with a somewhat degraded resolution.Pulse inversion image (c) gives better suppressionof artifacts than harmonic imaging and uses fullbandwidth of the transducer, thus improving resolu-tion.

dispersed following insonation (see §1.3.4).Perforating coronary vessels, however, whichhave a sufficiently high flow velocity to refillbetween frames, become visible on pulse inver-sion images with Levovist (Figure 34–35). In the liver, low MI imaging allows long lengths of vessels to become visible. Figure 36compares the appearance of large hepaticvessels imaged in harmonic and pulse inversionmodes following a peripheral venous injectionof Optison. The high MI harmonic image(Figure 36 a) shows the punctate appearance ofvessels in which the agent is destroyed; thelower MI pulse inversion image (Figure 36 b)shows fourth order branches of the portal veinimaged with great clarity. Finally, because of its

extended bandwidth, pulse inversion imagingis an ideal method to use for high MI tissueharmonic imaging. Figure 37 compares thefundamental (Figure 37 a), harmonic (Figure37 b) and pulse inversion images (Figure 37 c)of the same view. Note that artifact suppressi-on is achieved in both the harmonic and pulseinversion images, but that the resolution of thepulse inversion image is superior.

1.3.3.6 Power pulse inversion imagingIn spite of the improvements offered by pulseinversion over harmonic imaging for suppress-ing stationary tissue, the method is somewhatsensitive to echoes from moving tissue. This isbecause movement of tissue causes linear

a) b)

c)

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 34

Page 50: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

35Principles and Instrumentation

Fig. 38 The principle of pulse inversion Doppler (also known as power pulse inversion). The method issimilar to pulse inversion imaging, except that a sequence of more than two pulses are transmitted withalternating phase. The echoes from successive pulses are recombined in a way that eliminates the effectof steady displacement of the target due to tissue motion. The method allows suppression of movingtissue without the need to disrupt the bubble, hence achieving real time, low MI imaging. The exampleshown here of three pulses illustrates a principle that can be applied to arbitrarily long ensembles ofpulses, improving bubble-to-tissue sensitivity.

echoes to change slightly between pulses, sothat they do not cancel perfectly. Furthermore,at high MI, nonlinear propagation also causesharmonic echoes to appear in pulse inversionimages, even from linear scattering structuressuch as solid tissue. While tissue motion arti-facts can be minimised by using a short pulserepetition interval, nonlinear tissue echoes canmask the echoes from bubbles, reducing theefficacy of microbubble contrast, especially ininterval delay imaging when a high MI is used.A recent development seeks to address theseproblems by means of a generalisation of thepulse inversion method, called pulse inversionDoppler (38). This technique – which is alsoknown as power pulse inversion imaging –combines the nonlinear detection performanceof pulse inversion imaging with the motion

discrimination capabilities of power Doppler.Multiple transmit pulses of alternating polarityare used and Doppler signal processing tech-niques are applied to distinguish betweenbubble echoes and echoes from moving tissueand/or tissue harmonics, as desired by theoperator. In a typical configuration, the echoesfrom a train of pulses are combined in such away that signals from moving tissue – whichpose a problem to pulse inversion imaging –are eliminated (Figure 38). This method offersimprovements in the agent to tissue contrastand signal to noise performance, though at thecost of a somewhat reduced framerate. Themost dramatic manifestation of this method’sability to detect very weak harmonic echoes hasbeen its first demonstration of real-time per-fusion imaging of the myocardium (39)

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 35

Page 51: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Handbook of Contrast Echocardiography36

(Figure 39). By lowering the MI to 0.1 or less,bubbles undergo stable, nonlinear oscillation,emitting continuous harmonic signals. Becauseof the low MI, very few bubbles are disrupted,so that imaging can take place at real-timerates. Because sustained, stable nonlinear oscil-lation is required for this method, perfluoro-carbon gas bubbles work best.

1.3.4 III - Transient disruption: intermittent imaging

As the incident pressure to which a resonatingbubble is exposed increases, so its oscillationbecomes more wild, with radius increasing insome bubbles by a factor of five or more duringthe rarefaction phase of the incident sound.Just as a resonating violin string, if bowed over-zealously, will break, so a microbubble, if drivenby intense ultrasound, will suffer irreversibledisruption of its shell. A physical picture ofprecisely what happens to a disrupted bubble isonly now emerging from high speed videostudies (Figure 40). It is certain, however, thatthe bubble disappears as an acoustic scatterer(not instantly, but over a period of time deter-mined by the bubble composition), and that asit does so it emits a strong, brief nonlinearecho. It is this echo whose detection is the basisof intermittent myocardial perfusion imagingwith ultrasound contrast agents.

1.3.4.1 Triggered imagingIt was discovered during the early days of har-monic imaging that by pressing the ‘freeze’button on a scanner for a few moments, andhence interrupting the acquisition of ultra-sound images during a contrast study, it is pos-sible to increase the effectiveness of a contrastagent. So dramatic is this effect that it can raise

Fig. 40 Fragmentation of contrast agent observedwith a high-speed camera. The frame images (a–g)are captured over 50 nanoseconds. The streakimage (similar to an M-mode) shows the variationsin bubble diameter with a temporal resolution of 10 nanoseconds. The bubble is insonified with2.4 MHz ultrasound with a peak negative pressureof 1.1 MPa (MI~0.7). (a) The bubble is initially 3 µmin diameter. (b) The first large expansion. (c–d) Thebubble fragments during compression after the firstexpansion. (e–f ) Fragments are seen during expan-sion. Resulting bubble fragments are not seen afterinsonation, because they are either fully dissolvedor below the optical resolution.Courtesy of James Chomas, Paul Dayton, KathyFerrara, University of California, Davis CA, USA

Fig. 39 Pulse inversion Doppler (also known as‘power pulse inversion’) showing real-time myo-cardial perfusion. The contrast agent is Optison,the mechanical index is 0.15, the frame rate is 15 Hz.

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 36

Page 52: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

37Principles and Instrumentation

Fig. 41 Bubble disruption demonstrated in a laboratory phantom. a) Bubbles are suspended in a waterbath containing a block of tissue mimicking material. b) The block is scanned at high mechanical indexusing a linear array scanner. A bright echo is seen from the bubbles as they experience high peak pres-sure insonation. c) An image taken a moment later reveals a swirl of echo-free fluid, corresponding to thescanplane which no longer bears intact bubbles. This explains the apical 'swirl' seen in many LVO studiesat high MI.

the visibility of a contrast agent in the myo-cardial circulation to the point that it can beseen on a harmonic B-mode image, above theecho level of the normal heart muscle (40).This is a consequence of the ability of theultrasound field, if its peak pressure is suf-ficiently high, to disrupt a bubble’s shell andhence destroy it (37, 41) A phantom studyillustrating this effect is shown in Figure 41. Asthe bubble is disrupted, it releases energy, socreating a strong, transient echo, which is richin harmonics. This process is sometimes mis-leadingly referred to as ‘stimulated acousticemission’. The fact that this echo is transient innature can be exploited for its detection. Thefirst, and simplest method, is to subtract froma disruption image a baseline image obtainedeither before or (more usefully) immediatelyafter insonation. The residual echoes can thenbe attributed to the bubbles which weredisrupted by the imaging process. Such amethod requires offline processing of storedultrasound images, together with softwarewhich can align the ultrasound images beforesubtraction. Chapter 4 contains more discus-sion and examples of this technique. A more

effective method is to compare the echoes fromtwo or more consecutive high intensity pulses,separated by a fraction of a millisecond. This isthe principle of intermittent harmonic powerDoppler.

1.3.4.2 Intermittent harmonic power Doppler Power Doppler imaging (also known as colourpower angio, or energy imaging) is a techniquedesigned to detect the motion of blood or oftissue. It works by a simple, pulse-to-pulsesubtraction method (42), in which two ormore pulses are sent successively along eachscan line of the image. Pairs of received echotrains are compared for each line: if they areidentical, nothing is displayed, but if there is achange (due to motion of the tissue betweenpulses), a colour is displayed whose saturationis related to the amplitude of the echo that haschanged. This method, though not designedfor the detection of bubble disruption, is ideallysuited for high MI ‘destruction’ imaging. Thefirst pulse receives an echo from the bubble,the second receives none, so the comparisonyields a strong signal. In a sense, powerDoppler may be thought of as a line-by-line

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 37

Page 53: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

subtraction procedure on the radiofrequencyecho detected by the transducer.

A crucial question is how long one needs towait between pulses. If the two pulses are tooclose together in time, the bubble’s gaseouscontents, which are dispersed after disruptionof the shell by a process of diffusion andfragmentation, will still be able to provide anecho, so reducing the effectiveness of the detec-tion. If the two pulses are too far apart in time,the solid tissue of the myocardium will havemoved, so that the detection process will showthem as well. Of course, the amplitude of theecho from a block of muscle is much strongerthan that from a micron-sized bubble within it,so this signal will mask that from the bubble.There are two solutions. First, by using harmo-nic detection, some – but not all – of themoving tissue ‘clutter’ can be rejected. Second,a bubble can be designed to disrupt quickly sothat rapid (that is, high pulse repetitionfrequency (PRF)) imaging may be used. Such a

bubble will have a gas content that is highlydiffusible and soluble in blood. In this respect,air is perfect. Diffusion of air after acoustic dis-ruption is about 40 times faster than suchdiffusion of a perfluorocarbon gas from a simi-lar bubble (37), so that air based agents such asLevovist are easiest to image in this mode(Figure 42). For perfluorocarbon bubbles,longer insonation periods are necessary, but byusing multiple pulses for each line (4–8 is typi-cal) and careful filter design, very effective per-fusion imaging can be achieved with theseagents too. It is upon such considerations thatthe machine settings for perfusion imagingrecommended in Chapter 3 are based.

Blood flow velocities in the vessels of the myo-cardium are very low, so that as soon as highMI imaging is commenced, bubbles are dis-rupted before they have time to refill themicrovessels. Pausing the imaging processbetween acquisitions allows new agent to washin and fill the vascular bed. The next imaging

38 Handbook of Contrast Echocardiography

Fig. 42 Air bubbles disrupt more rapidly than do perfluorocarbon bubbles under high mechanical indexinsonation. In this experiment, bubbles from an air agent (a) and a perfluoropropane agent (b) areinsonated under identical conditions. The half-life of the air bubble is approximately 20 ms, while the per-fluropropane bubble is still providing an echo after 0.5 sec. The rapid response of the air agents at highMI is a strong advantage when they are imaged using intermittent harmonic power Doppler.

a) b)

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 38

Page 54: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

pulse results in a strong echo, which is sub-stantially stronger than the normal echo fromthe agent, and causes further destruction of thebubbles. This imaging strategy is sometimesreferred to as ‘intermittent’ or ‘transient’ imag-ing. We (32) and others (30) have found thatwith harmonic imaging, this single acquisitionimaging gives even more contrast. By combin-ing intermittent harmonic imaging with powerDoppler, one can assemble the most sensitiveimaging mode so far available for contrastagents, with which have been demonstratedthe first Doppler images of myocardial perfusi-on (32, 33). It is now commonplace to image

a single frame over one, two, or up to eightheartbeats (Figure 43), so as to follow the timefor the agent to refill the smallest vessels of themyocardium, into which the blood flows atonly about 1 mm/s. Chapter 4 explores theopportunities for quantitation that thispresents.

1.3.5 Summary

There are three regimes of behaviour ofbubbles in an acoustic field, which depend onthe intensity of the transmitted ultrasoundbeam. In practice, this intensity is best moni-tored by means of the mechanical index (MI)displayed by the scanner. At low MI, thebubbles act as simple, but powerful echoenhancers. This regime is most useful for spec-tral Doppler enhancement of, for example,pulmonary venous flow signals. At slightlyhigher intensities (the bottom of the range ofthose used diagnostically), the bubbles emitharmonics as they undergo nonlinear oscilla-tion. These harmonics can be detected by har-monic and pulse inversion imaging, and formthe basis of B-mode imaging for left ventricu-lar opacification. By using the newer, moresensitive techniques such as power pulse

39Principles and Instrumentation

Fig. 43 Harmonic power Doppler image of myo-cardial perfusion, made using a trigger internal offour heartbeats.

Three methods for perfusion imaging

MI Mode Technique Framerate(typical)

High Harmonic B-mode or Intermittent imaging 1/5 Hzpulse inversion imaging Offline subtraction of pre-contrast image

High Harmonic power Doppler Intermittent imagingRecord first frame after disruption 1/5 Hz

Low Power pulse inversion Nondestructive, real-time imaging 16 Hz

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 39

Page 55: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

40 Handbook of Contrast Echocardiography

inversion imaging, these echoes from bubblescan be separated from those due to tissue togive us real-time perfusion imaging of themyocardium. Finally, at the highest intensitysetting of the machine used in routine scann-ing, the bubbles are disrupted, emitting astrong, transient echo. Detecting this echo withharmonic power Doppler remains the mostsensitive means we have to image bubbles invery low concentration, but it comes at theprice of destroying the bubble. Because of thelong reperfusion periods of myocardial flow,intermittent imaging at framerates as low as oneframe per eight heartbeats is then necessary.

1.4 Safety considerations

Contrast echocardiography exposes patients toultrasound in a way that is identical to that ofa normal echocardiography examination. Yetthe use of ultrasound pulses to disrupt bubbleswhich sit in microscopic vessels raises somenew questions about the potential for hazard.When a bubble produces the brief echo whichis associated with its disruption, it releasesenergy which it has stored during its exposureto the ultrasound field. Can this energydamage the surrounding tissue? At higherexposure levels, ultrasound is known toproduce bioeffects in tissue, the thresholds forwhich have been studied extensively. Do thesethresholds change when bubbles are present inthe vasculature? Whereas the safety of ultra-sound contrast agents as drugs has been estab-lished to the satisfaction of the most stringentrequirements of the regulating authorities in anumber of countries, it is probably fair to saythat there is much left to learn about the inter-action between ultrasound and tissue whenbubbles are present.

The most extreme of these interactions isknown as cavitation. Cavitation refers to theformation, growth and collapse of a gas cavityin fluid as a result of ultrasound exposure. Ithas been studied extensively prior to thedevelopment of microbubble contrast agents(43). In fact, most of the mathematical modelsused to describe contrast microbubbles wereoriginally developed to model cavitation(44–46). When sound waves of sufficientintensity travel through a fluid, the rarefactio-nal half-cycle of the sound wave can actuallytear the fluid apart, creating spherical cavitieswithin the fluid. The subsequent rapid collapseof these cavities during the compressional halfcycle of the sound wave can focus largeamounts of energy into a very small volume,raising the temperature at the centre of thecollapse to thousands of degrees Kelvin, form-ing free radicals, and even emitting electromag-netic radiation (sonoluminescence) (44–46).

The concern over potential cavitation-inducedbioeffects in diagnostic ultrasound has lead tothe development of several indices to describethe relative likelihood of cavitation from clini-cal instruments, of which the MechanicalIndex described in §1.2.1.1 is now widely dis-played on the monitors of most clinical ultra-sound systems. With exception of one studywhich showed cavitation-induced hemorrhagein a mouse lung exposed under conditions thatdiffer substantially from those found clinically(47), no evidence of bioeffects from conventio-nal imaging at these levels have been reported.A number of experienced groups have carriedout experiments to assess whether the presenceof contrast microbubbles can act as cavitationseeds, potentiating cavitation-induced bio-effects (48–54). While much of this work hasshown that adding contrast agents to bloodincreased cavitation effects (e.g. peroxide for-

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 40

Page 56: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

41Principles and Instrumentation

mation, acoustic emissions) and related bio-effects (e.g. hemolysis, platelet lysis), all signifi-cant bioeffects occurred with either very highagent concentration, sound pulse duration orMI, or with hematocrit well below the physio-logical range. In experiments in whichclinically relevant values of these parameterswere used (agent concentration <0.2percent,pulse duration <2 µs, MI<1.9, hematocrit40–45percent), no significant bioeffects havebeen reported to date (49, 55).

1.5 New developments in contrast imaging

1.5.1 Tissue specific bubbles

As manufacturers of the agents gain morecontrol over the rather difficult process ofproducing stable microbubbles, the possibili-ties to tailor their behaviour in the bodyincrease. One of the most significant efforts isto target agents to specific sites in the body.Agents have been developed that adhere tothrombus, for example (56). Initial in vitrostudies show a 6 dB enhancement of clot bythe agent when imaged with intravascular ultra-sound (57). The potential for such materials isclear, though the question remains as to theproportion of injected material that will reachthe desired target and the dose needed to attainthe desired contrast.

This dose can be minimised if the targetedagent itself contains gas. As described above,agents that are phagocytosed and remain intactin the Kupffer cells offer the prospect of a newkind of ultrasound liver imaging. However,until recently only particles have successfullyperformed this function. Recently a gas based

agent (SHU 563, Schering AG, Berlin) hasbeen developed which comprises a polymerparticle with a gas filled interior. The meanparticle size is approximately 1µm and theagent acts as an efficient ultrasound contrastagent in its blood pool phase, with extremelystrong harmonic emission (55). After 15–20minutes the agent is taken up, largely intact,into the reticuloendothelial system. Whenimaged with colour Doppler at high peak pres-sures, the bubbles are disrupted in the liver. Thelarge change in signal level between two conse-cutive pulses is interpreted by the Dopplersystem as a random change in phase, and arandom Doppler shift is assigned to this strongsignal. The result is that the liver, thoughstationary, can be imaged with colour Doppler.Defects in the Kupffer cells brought about by, for example, a cancer, show as an absence of these Doppler signals. The method is sosensitive that we have shown that individualbubbles can be detected in this way; extremelylow doses (3 µl/kg, for example) are all that isnecessary. Using this technique, an isoechoiccancer of a diameter less than 2 mm has beenimaged in an animal liver (17). This example oftissue-specific imaging using specially a designedcontrast agent is likely to represent just thebeginning of a completely new approach to theuse of diagnostic ultrasound.

1.5.2 Bubbles for therapy

Future development of targeted agents canyield even more exciting prospects. By combin-ing the targeting of an agent with the ability ofdiagnostic ultrasound to disrupt bubbles in aselected region in the body, a new method ofdrug delivery might become possible (58).Drugs could be delivered so that they attaintoxic levels only in desired volumes of tissue.

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 41

Page 57: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

42 Handbook of Contrast Echocardiography

Growth factors that enhance revascularisationmay be potentiated by bubble-bound deliveryand insonation over ischemic regions of tissue.Genetic material itself might be transportedwithin a microbubble into a cell and releasedby acoustic means (59, 60).

1.6 Summary

Unlike contrast agents for other imagingmodalities, microbubbles are modified by theprocess used to image them. Understanding thebehaviour of bubbles while exposed to an ultra-sound imaging beam is the key to performing aneffective contrast echocardiography examina-tion. The appropriate choice of a contrastspecific imaging method is based on thebehaviour of the agent and the requirements ofthe examination. The Mechanical Index (MI) isthe major determinant of the response ofcontrast bubbles to ultrasound. Low MI har-monic and pulse inversion imaging offer real-time B-mode methods for left ventricular opaci-fication. Harmonic spectral and colour Dopplermay be used to suppress tissue motion in flowstudies of coronary vessels. High MI harmonicpower Doppler is ideally suited to intermittentimaging of myocardial perfusion with contrastagents, while power pulse inversion imaging atvery low MI allows real-time visualisation ofmyocardial perfusion using perfluorocarbonagents.

1.7 References

1. Ophir J, Parker KJ. Contrast agents in diagnostic ultrasound. Ultrasound Med Biol 1989; 15:319–33.

2. Gramiak R, Shah PM. Echocardiography of the aortic root. Invest Radiol 1968; 3:356–366.

3. Ziskin MC, Bonakdapour A, Weinstein DP, Lynch PR. Contrast agents for diagnostic ultrasound. Invest Radiol 1972; 6:500–505.

4. Kremkau FW, Carstensen EL. Ultrasonic detection of cavitation at catheter tips. Am J Roentgenol 1968; 3:159–167.

5. Kerber RE, Kioschos JM, Lauer RM. Use of an ultrasonic contrast method in the diagnosis of valvular regurgitation and intracardiac shunts.Am J Cardiol 1974; 34:722–727.

6. Reid CL, Kawanishi DT, McKay CR. Accuracy of evaluation of the presence and severity of aortic and mitral regurgitation by contrast2-dimensional echocardiography. Am J Cardiol 1983; 52:519–524.

7. Sahn DJ, Valdex-Cruz LM. Ultrasonic contraststudies for the detection of cardiac shunts. J Am Coll Cardiol 1984; 3: 978-985.

8. Roelandt J. Contrast echocardiography. Ultrasound Med Biol 1982; 8:471.

9. Carroll BA, Turner RJ, Tickner EG, Boyle DB, Young SW. Gelatin encapsulated nitrogen micro-bubbles as ultrasonic contrast agents. Invest Radiol 1980; 15:260–266.

10. Feinstein SB, Shah PM, Bing RJ, et al. Micro-bubble dynamics visualised in the intact capillary circulation. J Am CollCardiol 1984; 4:595–600.

11. Schlief R. Echo enhancement: agents and techniques – basic principles. Adv Echo-Contrast 1994; 4:5–19.

12. Fritzsch T, Schartl M, Siegert J. Preclinical and clinical results with an ultrasonic contrast agent. Invest Radiol 1988; 23:5.

13. Goldberg BB, Liu JB, Burns PN, Merton DA, Forsberg F. Galactose-based intravenous sono-graphic contrast agent: experimental studies.J Ultrasound Med 1993; 12:463–70.

14. Fobbe F, Ohnesorge O, Reichel M,Ernst O, Schuermann R, Wolf K. Transpulmonary contrast agent and color-coded duplex sonography: first clinical experience. Radiology 1992; 185(P):142.

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 42

Page 58: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

43Principles and Instrumentation

15. Unger E, Shen D, Fritz T, et al. Gas-filled lipid bilayers as ultrasound contrast agents. Invest Radiol 1994; 29:134–136.

16. Mattrey RF, Scheible FW, Gosink BB, Leopold GR, Long DM, Higgins CB. Perfluoroctylbromide: a liver/spleenspecific and tumor-imaging ultrasound contrast material. Radiology 1982; 145:759–62.

17. Fritzsch T, Hauff P, Heldmann F, Lüders F, Uhlendorf V, Weitschies W. Preliminary results with a new liver specific ultrasound contrast agent. Ultrasound Med Biol 1994; 20:137.

18. Mattrey RF, Leopold GR, van Sonnenberg E,Gosink BB, Scheible FW, Long DM. Perfluorochemicals as liver- and spleenseeking ultra-sound contrast agents. J Ultrasound Med 1983; 2:173–6.

19. Chin CT, Burns PN. Predicting the Acoustic Response of a Microbubble Population for Contrast Imaging. Ultrasound Med Biol 2000; In press.

20. de Jong N. Physics of Microbubble Scattering. In: Nanda NC, Schlief R, Goldberg BB, eds. Advances in Echo Imaging Using Contrast Enhancement. Dubai: Kluwer Academic Publishers, 1997:39–64.

21. Apfel RE, Holland CK. Gauging the Likelihood of Cavitation From Short-Pulse, Low-Duty Cycle Diagnostic Ultrasound. Ultrasound Med Biol 1991; 17:175–185.

22. von Bibra H, Sutherland G, Becher H, Neudert J, Nihoyannopoulos P. Clinical evaluation of left heart Doppler contrast enhancement by a saccharide-based transpulmonary contrast agent. The Levovist Cardiac Working Group. J Am CollCardiol 1995; 25:500–8.

23. Uhlendorf V. Physics of ultrasound contrast imaging: scattering in the linear range. IEEE Trans UFFC 1994; 41:70–79.

24. Burns PN, Powers JE, Hope Simpson D, et al. Harmonic power mode Doppler using microbubble contrast agents: an improved method for small vessel flow imaging. Proc IEEE UFFC 1994:1547–1550.

25. Bleeker H, Shung K, Barnhart J. On the application of ultrasonic contrast agents for blood flowmetry and assessment of cardiac perfusion. J Ultrasound Med 1990; 9:461–71.

26. Neppiras EA, Nyborg WL, Miller PL. Nonlinear behaviour and stability of trapped micron-sized cylindrical gas bubbles in an ultrasound field. Ultrasonics 1983; 21:109–115.

27. Burns PN, Powers JE, Fritzsch T. Harmonic imaging: a new imaging and Doppler method for contrast enhanced ultrasound. Radiology 1992; 185(P):142 Abstr).

28. Burns PN, Powers JE, Hope Simpson D, Uhlendorf V, Fritzsch T. Harmonic contrast enhanced Doppler as a method for the elimination of clutter – In vivo duplex and color studies. Radiology 1993; 189:285.

29. Mulvagh SL, Foley DA, Aeschbacher BC,Klarich KK, Seward JB. Second harmonic imaging of an intravenously administered echocardiographic contrast agent: Visualization of coronary arteries and measurement of coronary blood flow. J Am Coll Cardiol 1996; 27:1519–25.

30. Porter TR, Xie F, Kricsfeld D, Armbruster RW. Improved myocardial contrast with second harmonic transient ultrasound response imaging in humans using intravenous perfluorocarbon-exposed sonicated dextrose albumin. J Am Coll Cardiol 1996; 27:1497–501.

31. Kono Y, Moriyasu F, Yamada K, Nada T, Matsumura T. Conventional and harmonic grey-scale enhancement of the liver with sonication activa-tion of a US contrast agent. Radiology 1996; 201.

32. Burns PN, Wilson SR, Muradali D, Powers JE, Fritzsch T. Intermittent US harmonic contrast enhanced imaging and Doppler improves sensitivity and longevity of small vessel detection. Radiology 1996; 201:159.

33. Becher H. Second harmonic imaging with Levovist: initial clinical experience, Second European Symposium on Ultrasound Contrast Imaging. Rotterdam, 1997. Erasmus Univ.

34. Kaul S. Myocardial contrast echocardiography in coronary artery disease: potential applications using venous injections of contrast. Am J Cardiol 1995; 75:61–68.

35. Hamilton MF, Blackstock DT. Nonlinear Acoustics. San Diego: Academic Press, 1998.

36. Burns PN, Hope Simpson D, Averkiou MA. Nonlinear Imaging. Ultrasound Med Biol 2000;In press.

37. Burns PN, Wilson SR, Muradali D, Powers JE, Greener Y. Microbubble destruction is the origin of harmonic signals from FS069. Radiology 1996; 201:158.

38. Hope Simpson D, Chin CT, Burns PN. Pulse Inversion Doppler: A new method fordetecting nonlinear echoes from microbubble

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 43

Page 59: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

44 Handbook of Contrast Echocardiography

contrast agents. IEEE Transactions UFFC 1999; 46:372–382.

39. Tiemann K, Lohmeier S, Kuntz S, et al. Real-time contrast echo assessment of myocardial perfusion at low emission power: first experimental and clinical results using power pulse inversion imaging. Echocardiography 1999; 16:799–809.

40. Porter TR, Xie F. Transient myocardial contrast after initial exposure to diagnostic ultrasound pressures with minute doses of intravenously injected microbubbles. Demonstration and potential mechanisms. Circulation 1995; 92:2391–5.

41. Uhlendorf V, Scholle F-D. Imaging of spatial distribution and flow of microbubbles using nonlinear acoustic properties. Acoustical Imaging 1996; 22:233–238.

42. Taylor KJ, Burns PN, Wells PNT. Clinical Applications of Doppler Ultrasound. New York: Raven Press, 1996.

43. Brennan CE. Cavitation and Bubble Dynamics. New York: Oxford University Press, 1995.

44. Poritsky H. The collapse or growth of a spherical bubble or cavity in a viscous fluid, First U.S. National Congress on Appl. Mech, 1951.

45. Rayleigh L. On the Pressure Developed in a Liquid During the Collapse of a Spherical Cavity. Philosophy Magazine 1917; Series 6:94–98.

46. Plesset MS. The dynamics of cavitation bubbles.J. Appl. Mech. 1949; 16:272–282.

47. Child SZ, Hartman CL, Schery LA, Carstensen EL. Lung Damage from Exposure to Pulsed Ultrasound. Ultrasound in Medicine and Biology 1990; 16:817–825.

48. Miller MW, Miller DL, Brayman A. A Review of in Vitro Bioeffects of Inertial Ultrasonic Cavitation From a Mechanistic Perspective. Ultrasound Med Biol 1996; 22:1131–1154.

49. Miller DL, Gies RA, Chrisler WB. Ultrasonically Induced Hemolysis at High Cell and Gas Body Concentrations in a Thin-Disk Exposure Chamber. Ultrasound Med Biol 1997; 23:625–633.

50. Miller DL, Thomas RM, Williams AR. Mechanisms for Hemolysis By Ultrasonic Cavitation in the Rotating Exposure System. Ultrasound Med Biol 1991; 17:171–178.

51. Miller DL, Thomas RM. Ultrasound Contrast Agents Nucleate Inertial Cavitation in Vitro. Ultrasound Med Biol 1995; 21:1059–1065.

52. Holland CK, Roy RA, Apfel RE, Crum LA. In Vitro Detection of Cavitation Induced by a Diagnostic Ultrasound System. IEEE Trans. IEEE Transaction on Ultrasonics, Ferroelectrics, and Frequency Control 1992; 29:95–101.

53. Everbach EC, Makin IRS, Francis CW, Meltzer RS. Effect of Acoustic Cavitation on Platelets in the Presence of an Echo-Contrast Agent. Ultrasound Med Biol 1998; 24:129–136.

54. Williams AR, Kubowicz G, Cramer E. The Effects of the Microbubble Suspension SHU 454(Echovist) on Ultrasound-Induced Cell Lysis in a Rotating Tube Exposure System. Echocardiography 1991; 8:423–433.

55. Uhlendorf V, Hoffmann C. Nonlinear acoustical response of coated microbubbles in diagnostic ultrasound. Proc IEEE Ultrasonics Symp 1994:1559–1562.

56. Lanza GM, Wallace KD, Scott MJ, et al. Initial description and validation of a novel site targeted ultrasonic contrast agent. Circulation 1995;92:I–260.

57. Christy DH, Wallace KD, Lanza GM, et al. Quantitative intravascular ultrasound: demonstrationusing a novel site targeted acoustic contrast agent. Proc IEEE Ultrasonics Symp 1995:1125–1128.

58. Unger EC. Drug delivery applications of ultrasound contrast agents, Second European Symposium on Ultrasound Contrast Imaging. Rotterdam, 1997. Erasmus Univ.

59. Greenleaf WJ, Bolander ME, Sarkar G, Goldring MB, Greenleaf JF. Artificial Cavitation Nuclei Significantly Enhance Acoustically Induced Cell Transfection. Ultrasound Med Biol 1998;24:587–595.

60. Unger EC, McCreery TP, Sweitzer RH. Ultrasound Enhances Gene Expression of Liposomal Transfection. Investigative Radiology 1997; 32:723–727.

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 44

Page 60: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

45Principles and Instrumentation

Kardio-Kap 1 for pdf 06.06.2000 16:57 Uhr Seite 45

Page 61: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

Peter N BurnsProfessor of Medical Biophysics and RadiologyUniversity of TorontoImaging ResearchSunnybrook and Women’s Health Science Centre2075 Bayview AvenueToronto, OntarioCanada M4N [email protected]

Harald BecherProfessor of CardiologyUniversity of BonnRheinische Friedrich-Wilhelms-UniversitätMedizinische Universitätsklink und Poliklink IIKardiologie/PneumologieSigmund-Freud-Straße 2553105 [email protected]

Copyright © 2000 by Harald Becher and Peter N Burns. This book is protected by copyright. All rights are reserved, whether the wholepart of material is concerned, specifically the rights of translation, reprinting, reuse of illustrations, recitation, broadcasting, reproductionon microfilm.

This electronic copy was downloaded under the conditions of the End User License Agreement which accompanies it. Use and storageof this document signifies agreement with the terms of the Agreement. The files may not be altered without prior written permission ofthe copyright owners. No text, figures, tables or images may be displayed or reproduced, except for personal use, in any form or by anymeans, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission of the copyright owners.

The use of general descriptive names, registered names, trademarks, etc. in this publication does not imply, even in the absence of a specificstatement, that such names are exempt from the relevant protective laws and regulations and therefore free for general use.

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 1

Page 62: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 46

Page 63: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

2 Assessment of Left Ventricular Function by Contrast Echo

And they shall be accounted poet kingsWho simply tell the most heart-easing things

John Keats 1795–1821

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 47

Page 64: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

2.1 Physiology and pathophysiology of LV function

Accurate and reproducible quantification ofleft ventricular (LV) function is a crucialcomponent of clinical cardiology. LV functionshould be assessed in each cardiac patient.Earlier, research and development of diagnosticmeasures concentrated on systolic LV function,because diastole was thought to be a passiveprocess, which is now known not to be the case.However, it is still easier to evaluate systolicthan diastolic function. In the physiology lab,LV function is assessed using calculationsderived from pressure and volume recordings;in clinical cardiology measurement of LV wallmotion has become the usual method forroutine invasive and non-invasive diagnostics(1). We distinguish between global and regionalsystolic function; in order to assess globalsystolic LV function, quantification of end-diastolic and end-systolic volumes (LVED,LVES) and ejection fraction (EF) is essential.LV volume and ejection fraction are powerfulprognostic measures in patients with coronaryand valvular heart disease (2).

Regional LV function is assessed by imagingindividual segments of the left ventricle. Wallthickening and the extent of active wall motionare evaluated. Regional function is crucial inthe diagnosis of ischemic heart disease. Sincereduced perfusion results in wall motionabnormality, impaired regional function can beused as an indicator of perfusion deficits.However, there is a perfusion threshold abovewhich wall motion abnormalities becomeapparent. Animal experiments have shown thatwall motion is normal if less than 20 percent ofthe thickness of the myocardium is involved orif the circumferential extent of the ischemia(infarction) is less than 12 percent of the entirecircumference. Transmural infarctions involvingless than 5 percent of the left ventricular massmay be below the threshold to produce detect-able wall abnormality (3).

Diastolic LV function is determined by thefilling of the left ventricle by means of acomplex process involving passive and activecomponents. Since reliable measurement ofwall motion in diastole is more difficult than insystole, information on intracardiac pressures isneeded. For non-invasive studies, Dopplervelocities at the mitral valve and in the pulmon-ary vein (PV) are correlated to direct hemo-dynamic pressure recordings (4) (Figure 1).Diastolic dysfunction is increasingly recog-nised as a key cause of heart failure symptomsin many patients whose systolic function isnormal. Abnormalities in diastolic functionoccur within seconds of the onset of coronaryocclusion. Shortly afterwards, wall thickeningand endocardial motion are impaired (5). Inchronic ischemia or myocardial disease, diastolicdysfunction may be found without systolicdysfunction; patients are not correctly treatedif only systolic dysfunction is evaluated.

48 Handbook of Contrast Echocardiography

Assessment of systolic LV functionby echocardiography

Regional LV function

Wall thickening/endocardial movement

Apical and parasternalviews: ASE standard,6 segments/scanplanes

Usually qualitative

Global LV function

End-diastolic/end-systolicvolumes, ejection fraction

2 or 3 apical scanplanes

Mostly quantitative

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 48

Page 65: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

49Assessment of LV Function

2.2 Available methods – The role of contrast

2.2.1 Systolic function: the need for endocardial border definition

Two-dimensional echocardiography is themethod of choice for evaluating systolic LVfunction. Visual assessment of wall thickeningand inward motion of the endocardium ismostly used to assess regional and global LVfunction. The dynamics of regional wall motionare usually assessed visually using semiquanti-tative grading (dyskinetic, akinetic, hypokineticor normal). Similarly global LV function isassessed visually (normal, low, medium orsevere). LV volumes and ejection fraction arenot measured in routine clinical practice due todifficulties in obtaining reliable data. The pre-requisite for reliable measurements is visibilityof the entire endocardium in cross section,which can present a problem, particularly onstill frames. Dropouts and noise in the image

may impair endocardial definition. For thesame reasons, the application of automaticquantification and three-dimensional (3 D) re-construction is still limited. Endocardialborders can be assessed more satisfactorily withmoving images, therefore many institutions usea semiquantitative visual scale rather thanmanually tracing the endocardium and calculat-ing the numerical volumes (6). Nonetheless,accurate and reproducible quantification of leftventricular ejection fraction should be the aimof every study which evaluates LV function.

Transesophageal echocardiography (TEE)provides excellent endocardial definition inmost patients, but off-axis imaging may presentdifficulty (7). Because reliable less invasive andalternatives are available, the indication for a TEE is rarely a simple assessment of LV function, except in mechanically ventilatedpatients in the intensive care unit. Moreover,stress studies are impractical and inconvenientusing the transesophageal approach.

Non-echocardiographic methods suffer fromthe same limitations of sensitivity in detectingwall motion abnormalities (8–10). The advan-tage of scintigraphy (and of MRI and angio-graphy) is that the acquisition of the images ismuch less observer-dependent than in echo-

Fig. 1 Schema of normal relation of mitral flow(top) and pulmonary venous flow (bottom). E = earlydiastolic LV filling, A = filling following atrial contrac-tion, S = systolic antegrade flow, D = diastolic ante-grade flow, A = retrograde flow during atrial contrac-tion.

Methods for measurement of systolic LV function

• Fundamental/harmonic B-mode imaging

• Transesophageal echocardiography (TEE)

• Pulsed wave Doppler echocardiography

• Contrast echocardiography

• Nuclear gated blood-pool imaging

• Magnetic resonance imaging (MRI)

• Cine left ventriculography

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 49

Page 66: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

50 Handbook of Contrast Echocardiography

cardiography. Moreover, non-echocardiographicmethods work in patients for whom ultra-sound methods fail because of a poor acousticwindow. Scintigraphy and angiography havebeen validated in numerous trials. Nucleargated blood-pool imaging is considered the goldstandard for assessment of LV function becauseit provides a 3 D data set for quantification ofLV volumes and ejection fraction. Never-theless, gated blood pool scintigraphy isrelatively costly for the data obtained and it isnow used more as a reference method tovalidate new echocardiographic methods (11).In clinical practice there is no real alternative toechocardiography, which is widely available,portable and repeatable. The introduction ofcontrast media has further reduced the needfor non-echocardiographic methods. Opacifica-tion of the left ventricle has been shown toenhance endocardial border definition in aseries of well documented studies (12–16). Asignificant number of suboptimal exams havebeen salvaged using intravenous contrastinjections. In combination with stress echo-cardiography, contrast agents greatly enhancequantification of LV function. It is moredifficult to obtain satisfactory images duringstress even in patients with good acousticwindows. In several studies both inter- andintra-observer variability decreased significantlywith the addition of contrast agents.

With conventional imaging methods, as manyas 20 percent of echocardiographic studies maybe suboptimal for reliable subjective assess-ment of left ventricular function. With theadvent of tissue harmonic imaging, this figurehas fallen to 5–10 percent (17). Ongoingimprovement of ultrasound systems will resultin further improvement of image quality, sothere will be less need in future for contrast toimprove visual assessment of LV wall motion.

However, automatic border delineation worksmuch better with contrast enhanced images,especially in colour or power Doppler modes.

2.2.2 Diastolic function: the need for pulmonary venous flow recordings

Diastolic function is still neglected in manyroutine studies. If measured at all, flow at themitral valve level is recorded and the ratio of Eto A wave amplitudes is used. It is well knownthat this approach is simplistic and pulmonaryvenous flow data are needed to avoid misinter-pretation of findings derived exclusively frompulsed wave Doppler tracings of the mitralflow (18). In addition to diastolic LV function,the combined use of mitral and pulmonaryvenous flow curves has proved useful for distin-guishing between restrictive and constrictiveLV dysfunction, allowing analysis of theseverity of mitral regurgitation and detectionof atrial parasystole after heart transplantation.

The problem with transthoracic recordings ofpulmonary venous flow is that these tracingscannot satisfactorily be obtained in many adultpatients. Weak and noisy signals are oftenfound in patients with poor acoustic windowsdue to obesity or emphysema. TEE is analternative to a certain extent, but it is semi-invasive, time-consuming and is usually notperformed only to assess diastolic function.Contrast enhancement to rescue weak Dopplersignals is now the method of choice for obtain-ing satisfactory PV flow recordings in patientswho are difficult to image (19, 20). In thefuture, tissue Doppler may be an alternative,but at least for the moment, moderately goodacoustic windows are needed for reliable tissueDoppler studies. The strength of this method

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 50

Page 67: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

51Assessment of LV Function

lies in the segmental analysis of LV diastolicfunction. Recent studies give reference valuesfor this new method, but its clinical import-ance compared to traditional Doppler methodshas yet to be established.

2.3 Indications and selection of methods

2.3.1 Indications for contrast echo

2.3.1.1 LV border delineationA simple three step visual score can be used tograde the quality of endocardial borderdefinition (Figure 2). Using the 22 segmentmodel of the American Society of Echocardio-graphy (ASE), a suboptimal B-mode echo-cardiogram is defined as a study in which morethan two segments are classified as 1 or 0.Contrast echocardiography should always beperformed if the recordings obtained with theusual basic techniques or tissue harmonicimaging – if available – are suboptimal(Figure 3). There is sufficient evidence thatendocardial delineation is improved using anultrasound contrast agent both for fundamen-tal echocardiography and harmonic imaging(11–16, 21, 22). However, the use of contrastshould only be contemplated if the anticipatedimprovement in endocardial delineationchanges the way in which the patient is furthertreated. Not every patient with incomplete

endocardial delineation deserves a contrastexamination. In the case of an 80-year oldpatient, for example, apical hypokinesia byitself is often indicative even when the lateralwall is poorly displayed.

In clinical practice, there are two groups ofpatients for whom contrast echocardiographyis of immediate diagnostic benefit: the first arepatients on intensive care wards where areliable assessment of LV function is vital.These patients are often difficult to imagebecause of their limited mobility or mechanicalventilation. Transesophageal echocardiographyis often necessary in such situations. Contrastechocardiography is a cheap and time-savingalternative to the considerably lengthier andmore invasive TEE examination, particularly asthere is always access to a vein in these patientsand contrast echocardiography prolongs theusual transthoracic ultrasound examination byonly a few minutes.

Stress echocardiography is another ultrasoundexamination in which the administration ofcontrast does not prolong the examinationunduly but in which there may be a useful gainin diagnostic confidence. In the case of

Endocardial border visualisationscore

0 Not visible1 Barely visible2 Well visualised

Fig. 2 Four chamber view showing differentdegrees of endocordial border delineation for thesingle myocardial segments, judgment using a threestep LV endocardial delineation score.

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 51

Page 68: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

52 Handbook of Contrast Echocardiography

pharmacological stress, access to a vein is madethrough which the contrast agent can beadministered. It is a waste of time to try toanalyze changes in wall motion – which areoften marginal – with special equipment andtime consuming protocols in patients withsuboptimal images! Only when endocardialborders are well delineated can reproducibleand useful results be expected with this method(6). If a patient does not have a good acousticwindow, stress echocardiography should not beperformed today without contrast enhance-ment. The total costs of a stress echo withcontrast agent are well below those of a myo-cardial scintigram which is used to diagnoseischemia and vitality when patients are difficultto image.

2.3.1.2 Detection, determination of size and shape of LV thrombi

Recent observations, though few in number,have demonstrated convincingly the usefulnessof LV opacification for delineation of thrombi

(23, 24). It is difficult to offer general recom-mendations for a technique based on thesestudies. Our own experience, however, suggeststhat contrast enhanced pulse inversion imagingor harmonic power Doppler will often enhancethe visibility of LV thrombi if fundamentalor tissue harmonic B-mode are equivocal(Figure 4).

Fig. 3 Suboptimal non-contrast four chamber view (a), with poorly visualised endocardial border of thelateral wall. Following 4 ml Levovist (400 mg/ml), the left ventricle is fully opacified and endocardial bordersare clearly defined.

Contrast for LV opacification

When to use contrast

• Assessment of LV function in patients with suboptimal recordings in 2–6 segments

• Automatic LV boundary detection and 3 D reconstruction

• Delineation of LV thrombi

When not to use contrast

• If anticipated image improvement will not alter patient management

• If native recordings are adequate• If native recordings are irredeemably inadequate

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 52

Page 69: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

53Assessment of LV Function

2.3.1.3 Pulmonary venous flowAdministration of contrast to enhance pulmon-ary venous Doppler should be considered forall patients presenting with heart failure andnormal systolic LV function, for whom nativeDoppler curves are non-diagnostic. Suspectedtamponade and restrictive or constrictivecardiac disease are rare but important indica-tions (18). Systolic function may be normalunder these conditions and other Dopplerparameters not indicative. In this case werecommend contrast enhanced Doppler forevery patient who gets contrast for LVopacification or myocardial perfusion. Thestudy can be performed during washout andimproves recordings and diagnostic confidenceeven in those patients who already havediagnostic baseline recordings.

2.3.2 Selection of patients

Contrast cannot rescue LV delineation studieswhen native images are totally inadequate. Themaximum yield of contrast enhancement hasbeen demonstrated in patients in whom endo-cardial borders of 2–6 segments are notdelineated satisfactorily. However, with a

Fig. 4 Apical four chamber view, noise in the apex using harmonic B-mode without contrast (left), delinea-tion of a thrombus impossible; the right panel shows a nice delineation of an apical thrombus using triggeredharmonic power Doppler and injection of 0.5 ml Optison.Courtesy of Heinz Lambertz, Deutsche Klinik für Diagnostik Wiesbaden, Germany

When to use contrast to enhance pulmonary venous Doppler

In patients with suboptimal PV Doppler waveforms

• Patients with heart failure but normal systolic function• Patients with suspected restrictive or constrictive

disease• Patients with suspected tamponade

In patients who receive contrast for LVO or myocardialperfusion

• All patients. PV signals obtained during wash-out of contrast are improved in all cases

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:15 Uhr Seite 53

Page 70: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

54 Handbook of Contrast Echocardiography

reasonable baseline image, it is still worthwhileusing contrast if more than six segments arepoorly visualised. No incremental diagnosticvalue has been found in patients with adequatenative recordings except when automatic LVboundary detection or 3 D reconstruction isperformed.

2.3.3 Selection of imaging methods

Several ultrasound methods are currentlyavailable for LV contrast studies; none is clearlyoptimal for measurement of LV function.Contrast studies therefore frequently includedifferent modes, much as in non-contrastimaging where B-mode, colour and spectralDoppler imaging are integrated. Switchingfrom one mode to another is as quick and easyin contrast imaging as it is in conventionalimaging.

As a contrast agent is injected, one shouldconsider the different sensitivities of the variousimaging methods and time their use according-ly. Often a poor display of endocardial bordersis found concurrently with poor Dopplerrecordings. With a conventional duplex scanner,contrast may be used for LV opacification andenhancement of pulmonary venous Dopplersignals (Figure 5). Since the latter method needsless contrast, LV opacification is assessed first.When LV contrast opacification begins tosubside and cavity filling becomes incomplete,there is still enough contrast in the blood toimprove the signal-to-noise ratio of a PVDoppler study.

The seminal step in LV contrast studies wasmarked by the advent of harmonic imaging. Forthe first time, intense and complete opacifica-tion could be achieved in the majority ofpatients. Fundamental (that is, conventional)

Fig. 5 Selection of the imaging method for contrast enhanced evaluation of LV function for machine withonly fundamental imaging capabilities.

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 54

Page 71: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

55Assessment of LV Function

imaging should not be used if harmonicimaging is available. LV motion should firstlybe assessed using continuous imaging (Figure6). In most patients a bolus injection ofLevovist, Optison or Definity will result in suf-ficient opacification and border delineationusing continuous imaging. As with fundamentalimaging, the wash-out period can be used forimaging modalities which provide additionalinformation but require a lower concentrationof microbubbles. Two methods are useful forevaluation of LV function: triggered colour orpower Doppler mode for assessing ejectionfraction, and pulsed wave Doppler for record-ing PV flow. Triggered colour or powerDoppler images provide the best delineation ofLV cavity, which makes manual or automatictracing of LV borders easy and highly reproduci-ble. The wash-out period following a bolusinjection is long enough to allow acquisition ofboth colour Doppler data for ejection fraction

and spectral Doppler for assessment of PV flowvelocities.

If harmonic B-mode does not provideadequate imaging, harmonic colour or powerDoppler can be used in continuous mode(Figure 6). Although these modalities provide

Fig. 6 Selection of the imaging method for contrast enhanced evaluation of LV function for machine withharmonic imaging capabilities.

Before you inject …

The scanner should be adjusted carefully before

injection of contrast. Initial settings can be stored on

the machine as presets. They should be checked first

by scanning the patient without contrast and making

minor adjustments. Be sure that your storage medium

(magneto-optical (MO) disc or tape) is ready for use.

Only then are you ready to give a contrast injection!

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 55

Page 72: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

56 Handbook of Contrast Echocardiography

excellent delineation of endocardial borders,they are limited by a reduced frame ratecompared to B-mode harmonics and by wallmotion artifacts. For these reasons harmonicB-mode is first choice for assessment of LV wallmotion. If continuous colour Doppler is stillnot able to opacify LV cavity completely, usingthis method in triggered mode usually providescomplete LV opacification and at least providesinformation on ejection fraction. In the rareevent that all transthoracic ultrasound methodsfail, TEE or scintigraphy should be used.

2.4 How to perform an LV contrast study

2.4.1 Fundamental B-mode

Fundamental (conventional) imaging shouldonly be used when harmonic imaging is notavailable. While perfluorocarbon agents suchas Optison, Definity or SonoVue provideadequate signal to produce opacification infundamental mode, Levovist does not (Figure

Fundamental B-mode for LVO: settings

Initial settings for all systems (preset):

Scanhead frequency 2.5 – 3.5 MHzTransmit power Mechanical index (MI) < 0.6Dynamic range Low – mediumLine density Medium or highCompression Medium – highPersistence Disabled

Individual adjustment of instrument controls:

Transmit power Try reduction of initial level if apical swirling (due to bubble destruction) is seen

Scanplanes Conventional planesFocus Below mitral valve (apical planes), below

posterior wall (parasternal view), to reduce bubble destruction in the near field

Time gain compensation As in fundamental imagingLateral gain (HP only) As in fundamental imagingReceive gain Slightly reduce gain to decrease the grey

levels in the myocardium before injection of contrast

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 56

Page 73: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

57Assessment of LV Function

7). Harmonic imaging allows lower dose, withconsequently less attenuation artifact andlonger duration (Figure 8). Conversion of non-diagnostic to diagnostic echocardiograms withcontrast is possible in many patients, but stillabout 30 percent of suboptimal studies cannotbe salvaged with fundamental contrast echo.The contrast dose for fundamental imaging isabout 3 times higher than for harmonicimaging. As with harmonic imaging, adjust-ment of transmit power is crucial for optimalcontrast effect. When switching from non-contrast imaging to a contrast study, transmitpower should be reduced to limit bubbledisruption. The receive gain should be reducedslightly to lower the grey levels in the myo-cardium before injection of contrast. Thisenhances the delineation of the cavities. For

opacification of the left ventricle, the presenceof contrast is more important than differencesbetween contrast signal amplitudes. Thereforedynamic range can be set low and thecompression raised slightly.

2.4.1.1 Contrast agent doseOptison: 1.0 ml as a bolus, followed by a rapidflush of saline (5 ml). The injection rate shouldnot exceed 1 ml per second. This dose regimenusually allows sampling of at least 5 cardiaccycles in several imaging planes. If the contrastenhancement is inadequate after the initialdose, a second dose of 2 ml may be injectedintravenously. The maximum total dose shouldnot exceed 8.7 ml.Definity: 10 µl/kg as a bolus. During stressechocardiography, additional injections are

Fig. 7 Apical four chamber view using fundamental B-mode before and after injection of 10 µl/kg ofDefinity. Poor endocardial endocardial definition without contrast (a), good opacification and delineation ofthe cavity following contrast injection (b). Courtesy of Neil J Weissman, Washington Hospital Center, Washington DC, USA

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 57

Page 74: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

58 Handbook of Contrast Echocardiography

necessary at peak stress or when the patientbecomes symptomatic. Use the same dose as atbaseline.

2.4.1.2 Image acquisition and interpretationStart scanning from an apical window torecord the four chamber view. With bolusinjections attenuation is usually observedinitially. When attenuation subsides and theentire cavity is opacified, check for apicalbubble disruption and make a final adjustmentto the transmit power. After optimisation, sam-

ple at least 5 cardiac cycles (storing to cineloopand/or MO disk) and then record at leastanother apical view. Parasternal views shouldalways be recorded later, because attenuationby the RV often impairs LV opacification.Evaluation of the stored images is performed aswith non-contrast echocardiography.

2.4.1.3 Pitfalls and troubleshootingThe problems and their management are thesame as for harmonic B-mode (see §2.4.3).

Fig. 8 Apical two chamber view acquired at end-diastole with native frames (left hand panel) and contrastenhanced frames (right hand panel) following intravenous injection of 5 ml Levovist (400 mg/ml). The topimages are acquired during fundamental imaging, the bottom images during harmonic imaging. Note the mar-ked improvement of contrast strength of harmonic imaging compared to fundamental imaging.

5 mlLevovist

5 mlLevovist

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 58

Page 75: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

59Assessment of LV Function

2.4.2 Fundamental pulsed wave (PW)Doppler

Venous injection of contrast leads to betterpulmonary venous flow signals recorded bytransthoracic Doppler (Figure 9, 10). AEuropean multicentre study with Levovist hasshown that non-diagnostic noisy recordingscould be improved by contrast enhancement toenable quantitative assessment (19). There areconsiderably more transthoracic studies whichclearly display the retrograde flow during atrialsystole with contrast than without. The peakantegrade and retrograde velocities obtained bytransthoracic contrast enhanced Doppler arecomparable to those of the TEE techniquewithout an echo-enhancing agent. The doseneeded for enhancement of Doppler signals islow compared to that used for LV opacificati-on or perfusion imaging.

Doppler studies for assessment of diastolic LVfunction include recordings of transmitral flowand pulmonary venous flow. Usually the PWDoppler study at the mitral valve provides

adequate signals without contrast enhancementeven in patients with poor windows. Contrastenhancement may cause spectral blooming andlead to an unreliable spectral envelope.Therefore the mitral flow signals should beobtained immediately before or late afterapplication of echo contrast. Contrast enhance-ment will improve both colour Doppler andspectral Doppler signals of pulmonary venousflow. Colour Doppler imaging helps to find thepulmonary veins and align the beam to thevein, but quantitative information is obtainedfrom the pulsed Doppler spectra.

Contrast enhanced studies can be carried outwith all commercially available scanners withPW capability. Harmonic imaging is not usual-ly necessary, because tissue motion is rarely aproblem. Low frequency transducers (2.5 MHzor less) are preferable because of their sensitivityto flow, but 3.5 MHz transducers have beenshown to be effective too. Enhancement ofDoppler signals is the contrast method whichrequires the fewest instrument adjustments.Only the receive gain needs to be adjusted; a

Fundamental pulsed Doppler for pulmonary venous flowenhancement: settings

Initial settings for all systems (preset):

Scanhead Lowest frequency

Transmit power Lowest setting consistent with good signals

Scanplanes Normal

PRF Normal setting

Sample volume Normal setting

Scanplanes Normal setting, no change necessary for contrast

Receive gain Usually reduction necessary

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 59

Page 76: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

60 Handbook of Contrast Echocardiography

Fig. 9 Contrast enhanced colour Doppler recordings of pulmonary venous flow (four chamber view): nocolour Doppler signals in the pulmonary veins without contrast (a), following bolus injection of 5 ml Levovist(300 mg/ml) flow signals in two pulmonary veins, LLPV = left lower pulmonary vein, RUPV = right upperpulmonary vein (b).

Fig. 10 PW-Doppler-recordings before and after contrast enhancement with 2 ml Levovist (300 mg/ml): in thepre-contrast recording (a) the Doppler spectrum is noisy and not completely displayed, whereas the contrastenhanced recording (b) provides the complete envelope of the Doppler spectrum with high signal to noise ratio.Courtesy of Heinz Lambertz, Deutsche Klinik für Diagnostik Wiesbaden, Germany

a) b)

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 60

Page 77: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

61Assessment of LV Function

reduction is usually necessary to avoid spectralblooming and saturation.

According to the Canadian ConsensusRecommendations for the Measurement andReporting of Diastolic Dysfunction, the rightupper pulmonary vein is evaluated using the four chamber view (18). After parallel align-ment of the beam to the vein in colourDoppler mode PW Doppler interrogation isinitiated adjusting the gain in the same way aswithout contrast. The receive gain should beadjusted in most patients because the optimaldose for the individual patient cannot bepredicted. The recommended dose of echocontrast is sufficient to achieve signal enhance-ment even in patients with very poor windows.The criteria for an optimal contrast enhancedDoppler spectrum are the same as for non-contrast Doppler studies. Since the samplevolume for interrogation of pulmonary venousflow is far from the skin, local acoustic powerusually is not high enough to cause significantbubble disruption noise. Therefore no reduc-tion in transmit power is normally needed.

2.4.2.1 Contrast agent doseIf pulmonary venous flow is evaluated togetherwith LV opacification or myocardial perfusion,the wash-out phase can be used for Dopplerenhancement and no additional contrastinjection is needed.

If enhancement of PV flow is the primaryindication, bolus injections of Levovist (3 ml,300 mg/ml concentration), Optison (0.3 ml)or Definity (5–10 µl/kg) can be injected as a bolus followed by a flush of 5 ml saline.Usually the Doppler gain has to be reduced tokeep the grey level of the Doppler spectrumwithin the normal ranges.

2.4.2.2 Image acquisition and interpretationThis is the same as for non-contrast Dopplerstudies. When adequate Doppler spectra arevisible, recordings are stored in cineloop or onvideotape. Peak antegrade and retrogradevelocities and velocity-time integrals as well asduration of retrograde flow are measured.Mitral flow curves are recorded beforeinjection of contrast or after the wash-out andevaluated for peak A and E wave velocities, thevelocity-time integrals and the duration of theA-wave. The average of at least 3 cardiac cyclesis taken.

2.4.2.3 Pitfalls and troubleshootingSpectral bloomingIn a spectral Doppler system, the arrival of theagent has the effect of increasing the grey levelintensity of the display. The peak Doppler shiftfrequency, which corresponds to the maximumflow velocity in the sample volume, remainsunaffected. If, however, the power of theDoppler signal tapers steadily to zero at highDoppler shift frequencies, as is commonly thecase (25), another form of the blooming artifactoccurs. In Figure 11 one can see that the arrivalof the contrast agent is accompanied by anapparent increase in the peak systolic frequency.The blood is not, in fact, flowing any faster; itis simply that the enhanced sensitivity of thedetection is causing the weaker signals at thehigher Doppler shift frequencies to be displayed.

Pitfalls in contrast enhanced PW-Doppler

• Signals too strong: spectral blooming • Gain too high: spectral blooming • Signals too weak: inadequate effect• Contrast dosage too low: inadequate effect• Contrast dosage too high: attenuation• Sample volume not properly placed within the

vein: no effect

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 61

Page 78: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

62 Handbook of Contrast Echocardiography

Put more concisely, if the dynamic range of thedisplay is not as great as the dynamic range ofthe signal, the agent will cause the weakerechoes to be displayed preferentially. In mostspectral Doppler systems, the same effect can beobtained simply by increasing the Dopplerreceiver gain setting. Conversely, the artifactcan be obviated simply by reducing theDoppler gain setting as the enhancementoccurs. The clear conclusion here is thatcontrast agents should be used to bring un-detectable signals into the dynamic range of thedisplay, not to push detectable signals beyondthe dynamic range of the display!

Attenuation

Transthoracic imaging of flow in the rightupper pulmonary vein is highly dependent onattenuation because of the long distancebetween transducer and sample volume. Thisfact often reduces the signal-to-noise ratio innon-contrast Doppler and can be found incontrast enhanced studies if there are too manymicrobubbles in the left ventricle and atrium.Attenuation is easily identified by switching toharmonic or fundamental B-mode, which willshow good opacification in the nearfield but nocontrast signals in the deeper parts, particularly

in the left atrium. Better penetration of ultra-sound can be achieved when the LV contrastappears faint and incomplete. The problem isresolved simply by waiting until further wash-out has occurred or, if applicable, by reducingthe rate of the contrast infusion.

Displaced sample volume

Positioning of the sample volume in contrastenhanced pulsed wave Doppler recordings canbe guided by colour Doppler which providesan excellent display of the proximal segment ofthe right or left upper vein (Figure 9). Whenthe Doppler spectrum is inadequate, the first

Fig. 11 Spectral blooming of PW Doppler recordings in the right upper pulmonary vein due to inadequategain setting (a), satisfactory Doppler recording after reduction of gain (b).

Avoiding artifacts in spectral Doppler contrast studies

• Use the agent only if the signal is inadequate

because of attenuation by solid tissue. Use the

minimum dose needed

• Reduce the Doppler transmitted power to the

minimum setting consistent with obtaining a good

contrast enhanced signal

• During enhancement, reduce the receive gain to

optimise the brightness of the spectral display

• Make velocity measurements only from the

optimised display

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 62

Page 79: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

63Assessment of LV Function

step should be to check in colour Dopplermode whether the sample volume has beendisplaced. Underdosing of contrast is rarely thereason when the recommended doses are used.Too low a dosing can be seen in colourDoppler mode (no display of the flow in thepulmonary vein and left atrium) and in B-mode (almost no contrast visible).

2.4.3 Harmonic B-mode

Before a contrast study is considered, non-contrast imaging using tissue harmonic imag-ing should be tried. In this mode high transmitpower is used to enhance tissue signals. Forcontrast studies, the effect of high transmitpower is ambivalent: on the one hand, har-monic response is enhanced and LV contrastsignals are more intense; on the other, myo-cardial grey levels increase, reducing thecontrast between LV cavity and myocardialtissue and thus potentially worsening thedelineation of endocardial borders. Moreover,bubble disruption is dependent on transmitpower and needs to be taken into accountwhen transmit power is adjusted. The follow-ing recommendations attempt to balance theseopposing effects and have been validated inseveral studies by different institutions, includ-ing ours (16, 21).

2.4.3.1 Contrast agent doseLevovist: use 4.0 g vials and the 400 mg/mlconcentration. Half of the vial is injected as abolus, followed by a rapid flush of saline.Optison: 0.5 ml bolus followed by a rapid flushof saline. The injection rate should not exceed1 ml per second.These dose regimens usually allow sampling ofat least 5 cardiac cycles in several imagingplanes. If contrast enhancement is inadequate

after the initial dose, a second dose may beinjected. The maximum total dose should notexceed 6 vials of Levovist or 8.7 ml Optison.During stress echocardiography additionalinjections are necessary at peak stress or whenthe patient becomes symptomatic. Use thesame dose as for baseline.

2.4.3.2 Image acquisition and interpretationStart scanning from an apical window to recordthe four chamber view. With bolus injectionsattenuation is usually observed initially. Whenattenuation subsides and the entire cavity isopacified, check for apical bubble disruptionand make a final adjustment to the transmitpower. After optimisation, sample at least 5cardiac cycles (storing to cineloop and/or MOdisk) and then record at least another apicalview (Figure 12). Parasternal views shouldalways be recorded later, because attenuationby the RV often impairs LV opacification.Evaluation of the stored images is performed aswith non-contrast echocardiography.

During dobutamine stress the low amounts ofOptison and Definity can be injected into theline without interrupting the infusion ofdobutamine. The agent is driven by thedobutamine infusion and provides prolongedLV opacification. For Levovist higher volumes

Pitfalls in contrast harmonic B-mode

• Attenuation• Blooming• Bubble dissolution• Nearfield defects• Inhomogeneous opacification, swirling• Systolic reduction/disappearance of contrast• LVOT defect in aortic regurgitation• Anti-contrast: myocardium and cavity almost

same grey level

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 63

Page 80: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

64 Handbook of Contrast Echocardiography

are necessary, therefore we recommend inter-ruption of the dobutamine infusion and veryslow injection (at least 5 seconds) to avoidadministration of a dobutamine bolus.

2.4.3.3 Pitfalls and troubleshootingBubble disruptionNearfield defects, inhomogeneous opacifica-tion, systolic reduction or disappearance ofcontrast are all due to bubble disruption. Due

to the higher local acoustic pressures close tothe transducer, more bubbles are destroyed inthe nearfield and during systole where thehigher pressures enhance bubble disruption. Inapical views swirling may be seen, reflectingmixing of apical blood containing low amountsof microbubbles with blood from deeperregions with more microbubbles (Figure 13).Inadequate contrast effect in the left ventricu-lar outflow tract (LVOT) may be found in

Harmonic B-mode for real-time LVO: settings

Initial settings for all systems (preset):

Imaging mode Pulse inversion (ATL) or harmonic B-modeDynamic range Low – mediumCompression Medium – highLine density MediumPersistence DisabledTransmit power MI < 0.3 for pulse inversion (ATL)

MI < 0.6 for all other systems

Individual adjustment of instrument controls:

Transmit power Reduction if apical defect or swirling is seen, try to increase if global contrast is too weak

Scanplanes Same as conventional planesFocus Below mitral valve (apical planes) or below

posterior wall (parasternal view). Reducesbubble destruction in the near field

Receive gain Slightly reduce gain to decrease the greylevels in the myocardium before injectionof contrast

Time gain compensation As for non-contrast studiesLateral gain (HP only) As for non-contrast studies

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 64

Page 81: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

65Assessment of LV Function

patients with aortic insufficiency, which alsocauses increased bubble dissolution. First checkto see whether the settings listed in the Tableare helpful. The most frequent error is leavingthe transmit power on the same level as intissue harmonic mode and not changing thefocus position. Adjusting these parametersoften gives satisfactory images. If the instru-ment controls are correctly set, attempts shouldbe made to reduce the transmit power further.If this measure does not produce better LVopacification, administration of a higher doseof contrast should be considered.

The receive gain is less important. It should bereduced to a level where myocardial tissue onlyproduces weak signals. This can furtherenhance the differentiation between contrastsignals in the cavity and myocardial tissue.However by doing this, the baseline imagebecomes worse than with a normal non-contrast gain setting. It is important to bear inmind the special distinction between thetransmit power and the receive gain controls ina contrast study. The transmit power controlsthe energy delivered by the scanhead, whichdetermine bubble behaviour, whereas the

Fig. 12 Harmonic B-mode, apical four chamber view acquired at end-diastole and end-systole with nativeframes (top images) and contrast enhanced frames (bottom images) following intravenous injection of 0.4 mlOptison.Courtesy of Sharon Mulvagh, Mayo Clinic, Rochester MN, USA

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 65

Page 82: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

66 Handbook of Contrast Echocardiography

Fig. 14 Apical four chamber view using harmonic B-mode before and after injection of 4 ml Levovist(400 mg/ml). Poor endocardial definition without contrast (a), good opacification and delineation of thecavity following contrast injection. Note that the LV is not completely opacified, particularly below the mitralvalve due to contrast shadowing (b).

Fig. 13 Apical four chamber view using harmonic B-mode before and after injection of 4 ml Levovist(400 mg/ml). Systolic swirling due to apical bubble destruction (a) can be reduced by decreasing transmitpower of the scanhead (lower mechanical index) and increasing receive gain, complete opacification duringdiastole (b).

receive gain simply controls the sensitivity ofthe scanner to the resulting echo (see §1.3).

Attenuation

Attenuation is often observed following bolusinjections of contrast. For a short time thereare so many microbubbles in the cavity that thetransmitted ultrasound is completely scattered

and absorbed by the microbubbles in the near-field and no ultrasound penetrates into thedeeper regions of the image. Strong signals arefound in the nearfield whereas the deeper partsshow no signals (Figure 14). Attenuationsubsides with wash-out and waiting for wash-out is the best way to resolve this problem.

a) b)

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 66

Page 83: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

67Assessment of LV Function

Anti-contrast effect

A specific limitation of all greyscale techniquesis the fact that the tissue signals are displayedtogether with the contrast signals in the LVcavity, from which they cannot sometimes bedistinguished. For effective delineation ofendocardial borders the amplitudes of thecontrast echoes in the cavity should exceed thatof the tissue echoes. However, if grey levelswithin the cavities are comparable to those inthe myocardium, an ‘anti-contrast’ effect isobserved (Figure 13 left). The borders of theseptum may particularly be obscured becausethe normal septum is echogenic. In individualstill frames, the endocardial border is usually

less clear than it is in moving greyscale images.Injection of a higher dose may be considered inorder to increase LV contrast in such cases.

2.4.4 Harmonic imaging of LV thrombi

Besides endocardial border delineation contrastechocardiography is useful for delineation of LVthrombi which may be found in conjunctionwith global or regional wall motion abnormali-ties. Harmonic imaging without contrast islimited in the nearfield of the transducerbecause tissue harmonics are weak in the near-field. High transmit power and nearfield focus-ing of the beam can be used to enhance display,but sometimes hazy recordings impair reliableconfirmation or exclusion of an apical LVthrombus. Real-time pulse inversion usingcontrast and low transmit power (MI < 0.3) isthe method of choice, because it provides thebest spatial resolution and apical bubbledestruction can be minimised. The other imag-ing modalities need higher transmit powers withthe risk of enhanced apical bubble destructionresulting in false positive findings. With real-time harmonic B-mode, power should be re-duced until contrast swirling is absent. Besidesapical swirling, the shape of an apical cavitydefect reveals whether it is due to apical bubbledestruction or an intracavitary mass (Figure 15).Symmetric convex defects suggest bubbledestruction rather than thrombi and reductionof the transmit power or higher dose of contrastshould be performed.

Harmonic power Doppler works best indestruction mode, which uses high transmitpower. Triggered imaging is then necessary tofill the cavity completely. One or two framesper cardiac cycle can be acquired. With all

Impact of transmit power on contraststudies for LV opacification (LVO)

Low Power

• Low tissue harmonics (myocardium becomes less visible)

• Low contrast signal intensity• Microbubble destruction reduced (homogenous

opacification of the entire LV)

High Power

• Strong tissue harmonic (myocardium becomes more visible)

• High contrast signal intensities• Microbubble destruction enhanced (apical filling

defects, swirling)

How to avoid artifacts in harmonic B-mode contrast studies

• Contrast shadowing: wait for wash-out of contrast or lower dose

• Apical bubble destruction: reduce transmit power• Insufficient cavity contrast at all depths: increase

dose or use more sensitive imaging method

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 67

Page 84: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

68 Handbook of Contrast Echocardiography

three imaging modalities the region with asuspected thrombus should be recorded withchanging transducer positions, because throm-bi are often better displayed with modifiedscanplanes. Moreover, repeated views have tobe obtained during the different phases of LVopacification after injection of the contrastbolus, because high amounts of contrast mayobscure thrombi.

2.4.5 Harmonic colour/power Doppler for LVO

2.4.5.1 Continuous imaging for evaluation of LV wall motion

Although harmonic colour and power Dopplerprovide the best LV delineation, thesetechniques are not the first choice for assessmentof LV wall motion when non-contrast studieshave proved suboptimal. The lower frame rate(typically 10–15 Hz) compared to harmonic

B-mode (as high as 50 Hz) and wall motion artifacts limit their use. The frame rate ofcurrent colour Doppler systems is simply notsufficient for evaluation of wall motion analysis,particularly during stress with high heart rates.On the other hand, both Doppler techniquesare much more sensitive to contrast, so lessagent is required if the investigation is per-formed using Doppler techniques (26–28).

In order to delineate LV cavities, the contrastagent should ideally be visible within the LVcavity but not within the intramyocardialvessels. Using the recommended dose ofcontrast and machine settings, there should beno significant filling of the intramyocardialvessels (Figure 16, 17). However, the highsensitivity of some newer instruments maycause colour or power Doppler signals withinthe myocardium. Flow in larger vessels mayusually be visible, for instance in the septalperforators, but contrast in smaller subendo-cardial vessels cannot be displayed withinjections of Levovist or Optison using con-tinuous imaging. Thus the endocardial defini-tion is excellent with contrast.

Since continuous colour or power Doppler isusually initiated when harmonic B-mode issuboptimal, the change from one mode to theother should be performed quickly to avoidadditional injections of contrast. In mostmachines you just have to press a button, butoptimisation of a few instrument controls iscrucial: these settings can be stored as a preset,so they are immediately available (see B-modesettings). If different scanheads are available forharmonic imaging, a lower frequency is usuallypreferred for Doppler studies because of itsgreater sensitivity, unlike B-mode, wherehigher frequencies often work better.

Imaging LV thrombi with contrastecho

Methods

1. Real-time pulse inversion using low transmit power (MI < 0.3) (Method of choice: best spatial resolution with minimal apical bubble destruction)

2. Real-time harmonic B-mode or triggered harmonic power Doppler (1:1–2)

Technique

• Avoid apical destruction of microbubbles by reducing MI

• Record region with suspected thrombus, changing the transducer position

• Use modified scanplanes to improve visualisation of thrombus

• Avoid blooming: record region with suspected thrombus during wash-out of agent

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 68

Page 85: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

69Assessment of LV Function

Fig. 15 Harmonic B-mode recordings of a 29-year-old male 2 weeks following an anterior wall myocardialinfarction. The study was performed to assess left ventricular performance. Apical four chamber view demon-strates an echo density at the LV apex suspicious for thrombus (a). A multi-lobed apical thrombus is clearlydelineated following a bolus injection of Optison (b).Courtesy of Roberto M Lang, James Bednarz, Victor Mor-Avi and Kirk T Spencer, University of Chicago, Chicago IL, USA

a) b)

Fig. 16 38-year old patient referred for assessment of left ventricular function prior to the administration ofchemotherapy. This patient was unable to complete the assessment of ventricular function by RNA due toclaustrophobia and was subsequently referred to the echocardiographic laboratory for the quantitation of leftventricular function. The baseline apical four chamber view is shown using tissue harmonic imaging (THI) (a).Notice that despite the use of THI, endocardial border definition remains poor. After the intravenous injectionof 400 mg/ml of Levovist a significant improvement in endocardial border definition is found (b). Courtesy of Eric Yu, University of Toronto, Canada

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 69

Page 86: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

70 Handbook of Contrast Echocardiography

Colour Doppler provides information on flowvelocity and variance. This information is notneeded when colour Doppler is used for LVopacification. The presence of microbubbles inthe cavity is of greater interest than theirvelocity (which will be misestimated if there isbubble destruction anyway). Colour Dopplersignals can therefore be displayed in a modifiedmap which uses just one colour. The use of amonochromatic map helps the visual assess-ment of endocardial borders, particularly inaneurysms.

2.4.5.2 Contrast agent doseColour or harmonic power Doppler studies are

usually performed during wash-out of contrastinjected for harmonic B-mode studies. WhenDoppler studies are initiated, as soon as LVcontrast becomes insufficient for B-mode, no additional contrast has to be injected.Otherwise, additional boluses of contrast maybe injected (for instance 5 ml of Levovist at300 mg/ml concentration).

2.4.5.3 Image acquisition and interpretationStart scanning with apical windows as forharmonic B-mode. Check for apical bubbledestruction and wall motion artifacts. Afteroptimisation, sample at least five cardiac cycles.Then record at least another apical view and

Fig. 17 Endocardial border delineation in a patient with poor acoustic window, comparison of four imaging modes: (a) = fundamental B-mode, (b)= tissue harmonic imaging, (c) = harmonic B-mode andLevovist, (d) = harmonic colour Doppler and Levovist.Courtesy of of Gian Paolo Bezante, University of Genoa, Italy

a) b)

c) d)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 70

Page 87: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

71Assessment of LV Function

Continuous harmonic colour or power Dopplerfor LVO: settings

Initial settings for all systems (preset):

Scanhead Lowest frequencyTransmit power Mechanical Index (MI) 1.0

Acuson Sequoia: -6 dB from maximumDisplay mode Colour Doppler

Standard variance mode (gives a colour mosaic look)Unidirectional map (set colour baseline to one flow direction extreme)Power Doppler with standard map

Persistence DisabledLine density Highest valueSensitivity MediumWall filter MediumPRF Highest level

Individual adjustment of instrument controls:

Scanplanes Same as conventional planes. Optimise in tissue harmonic mode then reduce B-mode receive gain

Power Doppler Adjust box so that the entire LV-myocardium is included

Focus Below mitral valve (apical planes), or below posterior wall (parasternal view)

Transmit power Reduction if apical defect or swirling, try to increase MI if global contrast is too weak

Receive gain Default, reduce if wall motion artifacts are displayed

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 71

Page 88: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

72 Handbook of Contrast Echocardiography

parasternal views in the same way. Evaluationof the stored images is performed as for non-contrast echocardiography.

2.4.5.4 Pitfalls and troubleshootingSame as for triggerd imaging see 2.4.5.8

2.4.6 Triggered imaging for assessment of end-systolic and end-diastolic volumes and ejection fraction

End-diastolic and end-systolic volumes can bemeasured with continuous imaging byanalyzing end-diastolic and end-systolic still

frames. If these frames show completeopacification of the LV cavity, reliable measure-ments can be performed. For those patientswith inadequate LV opacification, triggeredimaging usually provides opacification of theentire cavity (26, 27).

Triggered or intermittent imaging means thattransmission of ultrasound is confined to shorttime intervals – usually the time to create onesingle frame. During this period, the micro-bubbles in the imaging field exhibit theirharmonic response and then are rapidlydissolved. Before the next frame is sampled, themicrobubbles have to be replenished fromareas outside the imaging field. The best way todo this is to interrupt the transmission of ultra-sound. Triggered imaging increases the pre-valence of microbubbles in the imaging field.The amount of bubbles increases with theduration of the non-transmission periodsbetween the imaging frames. For LV opacifica-

Ejection fraction by contrast echo

High quality, machine-derived LV volume and LVEF

measurements can be made in almost all patients

using triggered harmonic power or colour Doppler

technique.

Fig. 18 Two-dimensional image of apparent left ventricular apical aneurysm with suspicious echoes notedat distal lateral wall (a). Harmonic power Doppler contrast image demonstrating flow into a distal lateralpseudoaneurysm (b).Courtesy of Roberto M Lang, James Bednarz, Victor Mor-Avi and Kirk T Spencer, University of ChicagoChicago IL, USA

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 72

Page 89: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

73Assessment of LV Function

tion studies only short intervals are necessaryand end-systolic and end-diastolic frames canbe sampled in the same cardiac cycle.

Bubble destruction presents no problem whentriggered imaging is used. Bubbles which aredestroyed during imaging produce strongsignals which form the basis of the image.

Increasing the transmitted power is actuallybeneficial to these studies, because it enhancesmicrobubble disruption. At high MI, opacifica-tion of the entire LV cavity can be performedeven in patients with large aneurysms.

A secondary consequence of a high concentra-tion of microbubbles is the display of intra-

Triggered harmonic colour or power Doppler for LVO:settings

Initial settings for all systems (preset):

Scanhead Lowest frequencyTransmit power MI > 1.2Display mode Colour Doppler

Standard variance mode (gives colour mosaic appearance)Unidirectional map (set colour baseline to one flow directionextreme)Power Doppler with standard map

Persistence DisabledLine density Highest valueSensitivity MediumWall filter MediumPRF Highest level

Individual adjustment of instrument controls:

Scanplanes Same as conventional planes. Optimise in tissue harmonic mode then reduce B-mode receive gain

ECG-trigger End-diastole and end-systole trigger intervalevery cardiac cycle (use M-mode to set trigger points)

Power Doppler Adjust box so that entire LV-myocardium is includedFocus Below mitral valve (apical planes) or below posterior wall (para

sternal view)Transmit Power MI > 1.2. Do not adjust!Receive gain Reduced slightly if myocardial contrast is too high

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 73

Page 90: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

74 Handbook of Contrast Echocardiography

myocardial contrast. For reliable contour find-ing of LV volumes, myocardial contrast shouldbe as low as possible. There should be nodetectable microbubbles, at least in the smallersubendocardial vessels. Display of larger vesselsin the mid or epicardial layers does not, how-ever, interfere with the tracing of the cavity.Because of the short trigger intervals used inthe recommended protocol there is unlikely tobe visible contrast in the smaller intramyo-cardial vessels. In addition, the signals derivedfrom smaller intramyocardial vessels are muchweaker than the signals found in the cavity.Thus segmentation is possible by limiting thesensitivity of the ultrasound device.

2.4.6.1 Contrast agent doseTriggered imaging is usually performed in con-junction with continuous harmonic B-mode orharmonic colour/power Doppler, using a bolusinjection of contrast. During wash-out there isstill enough contrast for triggered colour orpower Doppler studies. If assessment of LVejection fraction is the primary indication, thesame dosing scheme is recommended as forassessment of regional wall motion. The benefitof a contrast enhanced segmental wall motionanalysis should be exploited in harmonic B-mode, before triggered imaging is initiated.

2.4.6.2 Image acquisition and interpretationAt least two apical views should be recorded. Ineach view end-systolic and end-diastolic framesof at least five cardiac cycles are recorded andstored on tape or disc. Shallow breathing orbreathholding is recommended to acquirecomparable planes. Off-line analysis is per-formed using the software packages imple-mented in the ultrasound systems. In eachview manual tracing of the endocardial bordersis performed in at least three cardiac cycles andaveraged (Figure 19). Then LVED and LVES

volumes are calculated using Simpsons methodor the area length method.

2.4.6.3 Pitfalls and troubleshootingWall motion artifact

The movement of the myocardial tissue mayproduce Doppler signals which are differentfrom cavity blooming and from the signalsoriginating from contrast in myocardial vessels(Figure 20). Usually the high pass (wall) filterof the ultrasound system rejects slow-movingstructural echoes. However, fast movement ofthe heart during deep breathing and coughingor exaggerated cardiac motion during physicalor dobutamine stress result in wall motionvelocities which exceed the threshold of thewall filter. Wall motion artifacts can easily berecognised as flash like coloured areas (chang-ing from frame to frame) which often overlayepi- or pericardial layers of the lateral wall (fourchamber view), but may be found in otherregions such as near the mitral valve and theseptum as well. Structure with high backscatterand fast movement parallel to the beam areprone to produce wall motion artifacts, there-fore wall motion artifacts are often found inthe fibrotic pericardial layers and the peri-valvular tissue.

Note that wall motion artifacts can be seenwithout contrast! Proper adjustment of themachine settings before contrast injection sup-presses most of the signals caused by wallmotion. High PRF and wall filter are crucial.Performing the studies with shallow breathingis usually sufficient to prevent wall motionartifacts. If not, patients should hold theirbreath for 5–10 s. If these measures are noteffective, a slightly different scanplane maybring the echodense structures into a positionwhere their movement is more perpendicularto the beam.It is more difficult to suppress wall

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 74

Page 91: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

75Assessment of LV Function

motion artifacts during stress. It may be possi-ble to control breathing but the wall motionartifacts caused by increased inotropy may stillhinder wall motion analysis.

Myocardial contrastThis is not a real pitfall. With the recommend-ed settings, generally bigger intramyocardialvessels like perforators may be displayed, butthese should not be mistaken for wall motionartifacts (see Chapter 3). The reticular or

patchy pattern of intramyocardial vessels isvery different from more extended wall motionartifacts which often are not confined to myo-cardial tissue.

AttenuationWhen harmonic colour or power Doppler areused during wash-out of a study performed forLV opacification or myocardial perfusion,attenuation will rarely be a problem. In thosepatients for whom colour Doppler was the

Fig. 19 Tracing of LV borders using harmonic power Doppler imaging. (a) = apical four chamber view acquired at end-diastole, (b) = turning the grey levels to zero, (c)= the contour can easily be traced, (d) = LVcavity contour with Simpsons slices.

a) b)

c) d)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 75

Page 92: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

76 Handbook of Contrast Echocardiography

primary method following injection orinfusion of contrast, attenuation may beobserved. As with B-mode, strong signals arefound in the nearfield whereas the deeper partsshow no signals (Figure 21). Attenuationsubsides with wash-out and waiting for wash-out is the best way to handle this problem.

Colour blooming

The ultrasound beam used to form an image isnot of a precise width, like that of a laser lightbeam, but tapers in intensity at its edges, morelike a flashlight beam. Just as when the intensityof a flashlight beam is increased the illuminatedspot it creates appears to get bigger, so when anultrasound beam encounters a strong scatterer

Fig. 20 Myocardial signals during harmonic power Doppler studies for assessment of LV function: end-diastolic (a) and endsystolic (b) frames, endocardial border definition possible despite myocardial signalsdue to wall motion artifacts (arrows).

Fig. 21 Harmonic power Doppler: contrast shadowing due to overdosing of contrast (a) with no powerDoppler signals at greater depth, better penetration and endocardial border delineation after reduction ofinfusion rate (b). Note inhomogeneous contrast due to enhanced bubble dissolution in the nearfield.

a) b)

a) b)

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 76

Page 93: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

77Assessment of LV Function

it detects it near the edge of the beam, so inter-preting it to be bigger than its actual size.Because of the threshold criterion for colourDoppler images previously described, thiseffect is especially noticeable in a colour image,where vessels already detected on colour willappear to ‘bloom’ when the agent arrives. Acomparable effect occurs in the axial direction,again because the axial sensitivity of the pulsedDoppler resolution cell is not sharp, but tapersat its proximal and distal boundaries. Theeffect can be to smother the image with colour,destroying its spatial resolution (Figure 21). Inparticular, the arrival of contrast in the cavitycan cause colour to ‘bleed’ into the myo-cardium. It is important that this is not mis-interpreted as ‘enhanced’ detection of flow inthe small vessels of the myocardium, nor leadsone to overestimate the size of the cavity. Usingthe minimum dose of the agent that is requiredto visualise the endocardium is one way to

resolve this artifact. Another is to reduce theeffective sensitivity of the colour system byeither reducing the receiver gain, or, moreconstructively perhaps, decreasing the en-semble length of the colour detection schemeso that more lines can be made in a shortertime, potentially increasing both the spatialand temporal resolution.

Bubble disruptionSimilar to harmonic B-mode, bubble destruc-tion may be a problem in harmonic Dopplermodes (which is assessed when apical fillingdefects are visible or swirling of contrast isobserved). The transmit power should then bereduced.

2.5 Summary

Intravenous contrast agents are now approvedand available for definition of the left ventricu-lar border and to salvage Doppler recordings inpatients who are difficult to image. Extensivestudies have demonstrated improvement ofimage quality and diagnostic confidence, andthis is now widely accepted. In spite of this, thethe optimal way in which to use contrast deser-ves consideration . Contrast studies for LV opa-cification are best performed using contrastspecific imaging modalities such as harmonicB-mode, pulse inversion and harmonic powerDoppler. Harmonic B-mode provides the hig-hest frame rate which is useful for stress echo,whereas spatial resolution is best using pulseinversion, which enhances delineation of intra-cardiac masses. Excellent segmentation of thetissue from the cavities on still frames makesharmonic power or colour Doppler the idealmethod for measurement of LV volumes andejection fraction. The protocols and machine

Avoiding artifacts in power/colourDoppler contrast studies

• Use the minimum dose of agent needed

• Eliminate colour blooming by reducing the

Doppler receiver gain

• Avoid shadowing by imaging after the peak of a

bolus or by appropriate titration of the infusion rate

• If possible, do not image through the right ventricle

Pitfalls in contrastenhanced harmonic power Doppler

• Wall motion artifacts

• Myocardial contrast

• Attenuation

• Blooming

• Bubble disruption

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 77

Page 94: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

78 Handbook of Contrast Echocardiography

settings in this chapter have been extensivelyvalidated. For imaging problems which stillarise reliable troubleshooting algorithms havebeen developed.

2.6 References

1. Schiller NB, Shah PM, Crawford M et al. Recom-mendations for quantification of the left ventricle by 2-dimensional echocardiography; American Society of Echocardiography committee on standards, sub-committee on quantification of 2-dimensional echo-cardiograms. J Am Soc Echocardiogr 1989; 2:358–367

2. White HD, Norris RM, Brown MA, Brandt PWT, Whitlick PML, Wild CJ. Left ventricular end-systolic volume as the major determinant of survival recovery from myocardial infarction. Circulation1987; 76:44–51

3. Cannon CP, Thompson B, McCabe et al. Predictors of non Q wave acute myocardial infarction in patients with acute ischemic syndroms. An ananlysis from the TIMI III trials . Am J Cardiol 1995; 75:977–981

4. Cohen GC, Pietrolungo DO, Thomas JD, Klein AL. A practical guide to assessment of ventricular diastolic function using Doppler echocardiography. J Am Coll Cardiol 1996; 27:1753–60

5. Weyman AE , Franklin TD, Egenes KM et al. Correlation between extent of abnormal regional wall motion and myocardial infarct size in chronically infarcted dogs. Circulation 1997; 56:72–78

6. Hoffmann R, Lethen H, Marwick T et al. Analysis of interinstitutional observer agreement in the inter-pretation of dobutamine stress echocardiography. J Am Coll Cardiol 19996; 27:330–336

7. Panza J, Laurienzo M, Curiel RV, Quyumi AA, Cannon RO. Transesophageal dobutamine stress echocardiography for evaluation of patients with coronary artery disease. J Am Coll Cardiol 1994; 24:1260–1267

8. Nagel E, Lehmkuhl HB, Bocksch W et al. Non-invasive diagnosis of ischemia-induced wall motion abnormalities with the use of high-dose dobutamine stress MRI: comparison with dobutamine stress echo-cardiography. Circulation 1999; 99:763–770

9. Ashburn Wl, Schelbert Hr, Verba JW. Left ventricular ejection fraction – a review of several radionuclide angiographic approaches using the scintillation camera. Proc cardiovasc Dis 1978; 20:267–284

10. Schad N. The intermittend phased injection of con-trast material into the heart. Am J Roentgen 1968; 104:464

11. Yu E, Sloggett C, Iwanochko RM, Rakowski H, Siu S. Feasibility and accuracy of Left ventricular ejection fraction determination by fundamental, tissue harmonic, and Levovist enhanced contrast imaging.J Am Soc Echocardiogr 2000 (in press)

12. Cohen JL, Cheirif J, Segar DS, Gillam LD, Gottdiener JS, Hausnerova E, Bruns DE. Improved left ventricular endocardial border delineation and opacification with Optison, a new echocardiographic contrast agent. Results of a Phase III multicenter trial. J Am Coll Cardiol 1998; 32 (3):746–752

13. Crouse LJ, Cheirif J, Hanly DE et al. Opacification and border delineation improvement in patients with suboptimal endocardial border definition in routine echocardiography: results of the phase III Albunex multicenter trial. J Am Coll Cardiol 1993;22:1494–1500

14. Feinstein SB, Cheirif J, Ten Cate FJ et al. Safety and efficacy of a new transpulmonary ultrasound contrast agent: initial multicenter clinical results. J Am Coll Cardiol 1990; 16:316–324

15. Grayburn PA, Weiss JL, Hakc TC et al. Phase 3 multicenter trial comparing the efficacy of 2 % dodeca-fluoropentane emulsion (EchoGen) and sonicated 5 % human albumin (Albunex) as ultrasound contrast agents in patients with suboptimal echocardiograms. J Am Coll Cardiol 1998; 32:230–236

16. Mulvagh SL, Rainbird AJ, Al-Mansour HA et al. Harmonic dobutamine stress echocardiography: Does contrast make a difference in diagnostic feasability and accuracy? Circulation 1998; 12.527

17. Becher H, Tiemann K, Schlosser T, Pohl C, Nanda NC, Averkiou M, Powers J, Lüderitz B. Improvement of endocardial border delineation using tissue harmonic imaging. Echocardiography 1998;15(5):511–516

18. Rakowski H, Appleton C, Chan KW et al. Canadien consensus recommendations for the measurement and reporting of diastolic dysfunction by echocardio-graphy. J Am Soc Echocardiogr 1996; 9:736–760

19. von Bibra H, Sutherland G, Becher H, Neudert J, Nihoyannopoulos P. Clinical evaluation of left heart Doppler contrast enhancement by a saccharide-based transpulmonary contrast agent. The Levovist Cardiac Working Group. J Am Coll Cardiol 1995; 25(2):500–508

20. Lambertz H, Schumacher U, Tries HP, Stein T. Improvement of pulmonary venous flow Doppler

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 78

Page 95: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

79Assessment of LV Function

signal after intravenous injection of Levovist. J Am Soc Echocardiogr 1997; 10: 891–8

21. Thanigaraj, S, Perez JE. Image enhancement with second generation contrast agents improve diagnostic accuracy and reduces additional testing. J Am Soc Echocardiogr 1999; 12:366

22. Porter TR, Xie F, Kricsfeld A, Chiou A, Dabestani A. Improved endocardial border resolution during dobutamine stress echocardiography with intravenous sonicated dextrose albumin. J Am Coll Cardiol 1994; 23:1440–1443

23. Thanigaraj S, Schlechtman KB, Perez JE. Improved echocardiographic delineation of left ventricular thrombus with use of intravenous second-generation-contrast image enhancement. J Am Soc Echocardiogr 1999; 12:1022–1026

24. Roberto M. Lang, M.D., James Bednarz, B.S., RDCS,Victor Mor-Avi, Ph.D., Kirk T. Spencer, M.D.J Am Soc Echocardiogr 1999; 12:871–5

25. Burns, P.N. (1995). Hemodynamics. In K. J. W. Taylor, P. N. Burns, & P. N. T. Wells (Eds.), Clinical Applications of Doppler Ultrasound (pp. 35–55). New York: Raven Press

26. Bezante G, Schwarz KQ, Chen X, Bertolazzi M, Spallarosa P, Villa G, Brunelli C. Automatic LV volume and ejection fraction by contrast harmonic color Doppler imaging compared to radionuclide ventriculography Eur Heart J 1999; 3628

27. Caiati C, Bina A, Zedda N et al. Contrast-enhanced harmonic colour Doppler reduces variability in assess-ing left ventricular volumes. Eur Heart J 1998; 19 (abstract supplement):2350

28. Becher H, Tiemann K, Schlief R, Nanda N. Harmonic Power Doppler Contrast Echocardio-graphy: preliminary clinical results (1997) Echo-cardiography 14; 6:637–642

Kardio-Kap 2 for pdf 06.06.2000 17:16 Uhr Seite 79

Page 96: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

Peter N BurnsProfessor of Medical Biophysics and RadiologyUniversity of TorontoImaging ResearchSunnybrook and Women’s Health Science Centre2075 Bayview AvenueToronto, OntarioCanada M4N [email protected]

Harald BecherProfessor of CardiologyUniversity of BonnRheinische Friedrich-Wilhelms-UniversitätMedizinische Universitätsklink und Poliklink IIKardiologie/PneumologieSigmund-Freud-Straße 2553105 [email protected]

Copyright © 2000 by Harald Becher and Peter N Burns. This book is protected by copyright. All rights are reserved, whether the wholepart of material is concerned, specifically the rights of translation, reprinting, reuse of illustrations, recitation, broadcasting, reproductionon microfilm.

This electronic copy was downloaded under the conditions of the End User License Agreement which accompanies it. Use and storageof this document signifies agreement with the terms of the Agreement. The files may not be altered without prior written permission ofthe copyright owners. No text, figures, tables or images may be displayed or reproduced, except for personal use, in any form or by anymeans, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission of the copyright owners.

The use of general descriptive names, registered names, trademarks, etc. in this publication does not imply, even in the absence of a specificstatement, that such names are exempt from the relevant protective laws and regulations and therefore free for general use.

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 1

Page 97: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 2

Page 98: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

3 Assessment of Myocardial Perfusionby Contrast Echocardiography

So much for the circulation! If it is either hindered or perverted oroverstimulated, how many dangerous kinds of illnesses and surprisingsymptoms do not ensue?

William Harvey, De Motu Cordis et Sanguinis, 1578–1657

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 3

Page 99: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Myocardial PerfusionImaging

No other application of ultrasound contrastagents elicits the enthusiasm evoked by myo-cardial perfusion imaging. We believe that inthe forseeable future contrast echo will replacescintigraphy as the reference method for per-fusion imaging. This optimism is based onwidespread experience with intracoronary con-trast echocardiography where optimal imagingconditions are found. For intravenous contrastperfusion imaging, we hope that the followingsections will demonstrate that the technologyis ready to yield images of diagnostic quality,but that optimisation and standardisation ofclinical approaches are still necessary. Thischapter aims to provide a first step.

3.1 Physiology and pathophysiology of myocardial perfusion

3.1.1 Normal perfusion

The vascular compartments in the myocardiumcomprise the larger arteries, the arterioles, thecapillary network and the smaller and largerintramyocardial veins (1) (Figure 1). Followingan intracoronary bolus injection of micro-bubble contrast, an effect is subsequently seenin the different compartments. All currentlyavailable echo contrast agents are pure intra-vascular tracers which traverse the capillary bedof the myocardium following intravenousinjection. For assessment of perfusion it isnecessary to detect blood in the capillarycompartment which contains more than 90percent of the intramyocardial blood volume.

Quantitation of perfusion is aimed at measu-ring the intravascular blood volume and thevelocity of blood flow through the vessels, fromwhich flow rate can be derived. Specific tech-niques are discussed in Chapter 4.

The distribution of intramyocardial vessels is not uniform in the left ventricular (LV)myocardium. Vessel density is highest in thesubendocardial layers. Myocardial oxygenconsumption of the endocardial layers is higherthan that of the epicardial layers, because the endocardium contributes more to wallthickening and is also subject to the highestintramyocardial pressures (2). This spatial non-uniformity of myocardial perfusion is furthercomplicated by the temporal changes in myo-

82 Handbook of Contrast Echocardiography

Fig. 1 Scale in the vascular system. 90 percent ofthe intramyocardial blood volume lies within thecapillaries, and this is the objective of the contrastperfusion study. Capillary flow is characterised byvelocities of 1 mm/s or less.

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 82

Page 100: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

83Perfusion Imaging

cardial blood flow (MBF) and myocardialblood volume (MBV). The cyclic changes ofMBF are similar to the changes found in theepicardial vessels. The increase in global myo-cardial blood flow during stress or pharma-cological intervention is described in §3.2.Briefly, blood flow through the myocardiumcan be increased by up to three or four times,provided there is no obstruction in the epicar-dial or intramyocardial arteries.

3.1.2 Acute myocardial infarction

Transmural myocardial infarction is character-ised by areas of significantly reduced or non-existent flow, which has caused necrosis of themyocytes. Myocyte necrosis is associated withnecrosis of the vascular cells and local loss ofmicrovascular integrity. During transmural

myocardial infarction, necrosis occurs first inthe subendocardial myocardium (3). Withlonger occlusions, a front of necrosis movesprogressively across the wall involving thetransmural thickness of the ischemic zone(Figure 2). The resultant infarct size dependsnot only on the perfusion bed of the occludedvessel and on the onset of reperfusion, but alsoon the presence of collateral vessels. Aftersuccessful recanalisation of an occluded infarctvessel, necrotic zones still have zero flow (noreflow), whereas salvaged myocardium withpreserved microvascular integrity is reperfused.It takes some time before mechanical functionrecovers after an ischemic event. Stunned myo-cardium is defined as salvaged myocardiumwhich is still akinetic but will regain itscontractility within subsequent weeks. Theonly current diagnostic procedure for earlyassessment of reperfusion is myocardial scinti-graphy.

3.1.3 Chronic ischemic heart disease

At rest, up to 90 percent of stenoses of epicardialvessels do not result in a change in overall MBFin the perfusion bed because of compensatoryvasodilation of peripheral vessels. There are,however, differences between the epicardialand subendocardial layers (4). Subendocardialintramural vessels are maximally dilated,whereas subepicardial vessels are not. A vasodi-lator stimulus will augment transmural flow dueto dilation of subepicardial vessels and decreaseof the driving pressure. Because subendocardialvessels are already maximally dilated, the fall indriving pressure will result in a fall of flow,whereas subepicardial flow improves and totalflow is maintained or increased. Thus the trans-mural steal phenomena from stenosed tonormal vessels may be created. However, the

Nonnecrotic, salvaged, butstunned myocardium

Necrotic myocytes withoutgross microvascular damage

Myocyte necrosis withgross microvascular damage

Viable subepicardium

Nonischemic tissue

LV Cavity

RISKZONE

Fig. 2 Acute myocardial infarction showing thesubendocardial necrotic zone and viable subepi-cardial tissue. The extent of wall necrosis increaseswith length of coronary occlusion and absence ofcollaterals. Modified from Braunwald (3).

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 83

Page 101: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

84

transmural gradient is not uniform for a wholehypoperfused segment. Ischemic myocardiumcan show a heterogeneous pattern with normallyperfused areas supplied by collaterals near scarsor hypoperfused tissue.

Coronary stenoses may affect the volume ofblood within the perfusion bed as well as itsblood flow (5). Myocardial blood flow isthought to decrease in the stenosed bed in thepresence of hyperemia (6). This decrease mayexplain the differences in maximal signal inten-sities between normally and hypoperfusedmyocardial segments on contrast echo. Thereduced blood volume is thought to be broughtabout by a decrease in capillary density,possibly reflecting the tendency to regulate fora constant capillary perfusion pressure.

3.2 Currently available imaging methods for myo-cardial perfusion imaging

Perfusion imaging has two major objectives –assessment of ischemia and assessment of myo-

cardial viability. In clinical cardiology there arethree established methods to achieve this.

3.2.1 Stress ECG

The probability of a positive stress electrocardio-gram (ECG) is only 50 percent in patients withsingle vessel disease (stenoses > 70 percent) (7,8). Stress ECG provides only limited informa-tion on the extent of ischemia. However, it isthe initial method of choice provided there areno contraindications, because it is a simple andinexpensive way of demonstrating induciblemyocardial ischemia (Figure 3). If positive,stress ECG can be used for control studies afterintervention.

3.2.2 Stress echo

Stress echocardiography has become the mostimportant diagnostic method, in spite of thefact that by imaging left ventricular wall motion,it addresses perfusion only indirectly. Thesensitivity and specificity are between 80 and90 percent according to a recent meta-analysis(9). Even under optimal imaging conditions,

Handbook of Contrast Echocardiography

Fig. 3 Traditionally diagnostic path for assessmentof inducible ischemia in patients with exertionalchest pain and risk factors (University of Bonn).

Clinical methods for assessmentof myocardial ischemia

Stress ECG Limited sensitivity

Stress echocardiography Wall motion as indirectmarker of perfusion

Doppler echocardiography Limited to LADperfusion bed

Myocardial scintigraphy Still the gold standard

Contrast echocardiography The future …

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 84

Page 102: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

85Perfusion Imaging

analysis of LV wall motion is inferior to thedisplay of a perfusion mismatch which precedesthe wall abnormality in the ischemic cascade(Figure 4). Classically, flow must be reduced to 50 percent in at least 5 percent of the myo-cardium to detect new wall motion abnormal-ities (10). Thus, an echocardiographic per-fusion method should enable us to detect myo-cardial ischemia earlier, providing higher sensi-tivity in the detection of coronary arterydisease. Moreover, there are some importantclinical situations in which wall motion cannotbe used as an indicator of perfusion. One is theevaluation of patients following thrombolysisof an acute myocardial infarction. Stunned butreperfused myocardium has the same mechan-

ical properties as an unperfused segment, sothat clinical decisions cannot be based on wallmotion analysis alone.

3.2.3 Coronary flow reserve (CFR)

Doppler echocardiographic measurement offlow at rest and during exercise is a well-vali-dated method for assessment of flow reserve inthe left anterior descending (LAD) territory –particularly if contrast enhanced Dopplertracings are used. Coronary flow reserve (CFR)should be included in every perfusion studyinvolving the LAD. However, because of thelimitations of the method (§3.12–17) and inorder to evaluate the perfusion provided by the circumflex and right coronary arteries,additional methods are needed.

3.2.4 Myocardial scintigraphy

Myocardial scintigraphy is the only establishedclinical method which directly addresses myo-cardial perfusion. However, scintigraphy isexpensive and is not as widely available as echo-cardiography. In many European countriesscintigraphic studies are not performed in thecardiology department, whereas echocardio-graphic studies are always carried out in anexperienced cardiological setting. Perfusionscintigraphy is known to be less compromisedby submaximal stress because perfusion mis-match comes earlier (11, 12). In many centresstress echo is the first choice and scintigraphicstudies are confined to patients with poorechocardiographic windows. Theoretically,myocardial scintigraphy should be moresensitive than stress echo in the detection ofinducible myocardial ischemia (see Figure 4).

Fig. 4 The ischemic cascade

• Reporting is usually qualitative

• Perfusion changes preceed wall motion

abnormalities

• Sensitivity and specificity are limited

• Induction of ischemia is necessary

• Risk of submaximal stress

Why are wall motion abnormalitiesnot sufficient for the diagnosis of

coronary artery disease?

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 85

Page 103: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

86

In fact, the difference between both methods ismarginal due to the technical limitations ofscintigraphy (9). Thus the potential of per-fusion imaging by contrast ultrasound is notthwarted by scintigraphy. Nevertheless, scinti-graphy is widely considered to be the clinical‘gold standard’ for perfusion imaging, againstwhich contrast echo will initially be judged. Inthe future, the roles of the two methods arelikely to become complementary.

3.2.5 Myocardial Contrast Echo

There is a clear role for a reproducible, non-invasive, real-time method for imaging myo-cardial perfusion. Myocardial contrast echo(MCE) offers a pure intravascular tracer, betterspatial resolution, the potential for quantita-tion and real-time imaging during rest, stressand interventional studies. Ultrasound is morewidely available, portable and less costly thanother methods and remains in the hands of thecardiologist. It is likely to offer complementarydiagnostic information to existing methods.

3.3 Indications and selection of methods

3.3.1 Indications

The potential of contrast echo has beendemonstrated in numerous animal studies andhumans using intracoronary injections ofcontrast. However, only few data are availablefrom clinical studies using intravenous injec-tions of contrast. The following guidelines arebased on a review of the current literature andon our clinical experience of intravenouscontrast echo performed in more than 300patients over the last three years (13–19). Sofar there are two clinical situations where wehave begun to establish and validate themethod:

1. Acute transmural infarctionContrast echo can be used to determine theinfarction size and to assess reperfusion. Thisindication was the first objective of intra-coronary contrast echo, because it can be treatedas a yes/no decision (contrast demonstrableversus no contrast) and may be performed with-out further quantification. Several studies haveshown that the area of risk correlates well withthat demonstrated using thallium imaging (13).Infarct size and myocardial salvage followingreperfusion can be monitored with contrastechocardiography (14). The presence of contrastindicates myocellular viability in patients withrecent myocardial infarction (15).

2. Coronary artery stenosisDetection and functional assessment has beenevaluated in preliminary clinical studies (19).Reduction of contrast in a region of interestand reduced velocity of the contrast micro-bubbles can be used to detect significant

Handbook of Contrast Echocardiography

• Better spatial resolution: transmural distribution

of perfusion is shown

• Real-time control of imaging

• Uses a pure intravascular tracer

• Inexpensive, portable, uses no ionising radiation

• Patients stay in a specialised cardiological

environment.

Why myocardial perfusion with echo?

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 86

Page 104: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

87Perfusion Imaging

Prior/post or without coronary angiography:

• Infarct size MCE

• Reperfusion/no reflow MCE

Clinical indications for myocardial contrast echocardiography (MCE)and coronary flow reserve (CFR)

Acute Myocardial Infarction

Chronic Ischemic Heart Disease

Without coronary angiography:

• Detection of myocardial ischemia MCE

(selection for coronary angiography)

Exclusion of myocardial ischemia*

In conjunction with coronary angiography:

• Functional significance of LAD stenosis MCE and/or CFR

• Functional significance of other coronary arteries MCE

• Normal coronary arteries: ‘false’ positive stress test CFR

to confirm cardiac origin of chest pain

• Microvascular disease CFR

* Optimal display of the entire myocardium is necessary for this indication, a prerequisite not usually fulfilled with current technology.

** No clinical trials have yet been performed.

stenoses of the epicardial vessels during physicaland pharmacological stress.

Myocardial contrast echocardiography stillsuffers from limited display of the myocardium(few scanplanes available, poor imaging condi-tions in certain regions, see §3.11). It maytherefore be questioned whether it is possibleto exclude significant stenoses using currenttechnology. However, an abnormal finding inan MCE study, if performed properly, can betaken to represent myocardial ischemia with areasonable level of confidence. For the timebeing, MCE should be used in combinationwith a regular stress echo protocol assessingwall motion abnormalities when detection of

inducible myocardial ischemia is the indica-tion. Thus trainees can learn the new methodand take advantage of the better sensitivity ofMCE compared to wall motion analysis.Should they fail, there still is the data of theestablished methods. With the introduction ofreal-time perfusion imaging, wall motion andperfusion can be assessed together, without thepresent time-consuming combination of imag-ing modes. This will help facilitate the accept-ance of myocardial contrast into the clinicalstress echo lab.

Measurement of coronary flow reserve com-pletes an MCE study when the significance ofan LAD stenosis has to be evaluated. In patients

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 87

Page 105: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

88

with normal coronary arteries and chest pain,CFR measurements can be used to confirm orexclude the cardiac origin of the chest pain.Coronary flow reserve provides a quantitativemeasure whereas reading of MCE studiesgenerally relies on visual assessment. For didac-tic reasons assessment of CFR is described in aseparate section (§3.12). In clinical practice,evaluation of CFR of the LAD should be part ofa global perfusion study which includes tissueperfusion. It should be noted that the quanti-tative approaches for MCE described inChapter 4 are in rapid development.

3.3.2 Selection of patients and contraindications

With currently available ultrasound equipmentand approved contrast agents, we are stillworking at the margins of the machine’s abili-ty to obtain adequate contrast in the myocar-dial tissue. If the acoustic window is poor,myocardial contrast usually is suboptimal. Thesuccess of a contrast study can often be anti-

cipated from the image quality at baseline.Unlike LV opacification studies, where contrastis indicated for suboptimal baseline imaging,only those patients in whom image quality is good should be accepted for a myocardialcontrast study. Using tissue harmonic B-mode,patients should have an image quality whichwould enable a non-contrast stress echo to besuccessful. For the contrast agent itself, contra-indications might rarely result from the specificformulation of the shell or surfactant. For per-fusion stress echo, the contraindications areidentical to those of a non-contrast study.

3.3.3 Selection of the imaging method

Intravenous administration of contrast resultsin a very low concentration of bubbles in themyocardium which can be evaluated only byusing contrast specific imaging modalities. Fourcontrast specific imaging modes are available(see Table and Chapter 1). All have strengthsand weaknesses: some are better suited to per-

Handbook of Contrast Echocardiography

Contrast specific imaging methods for assessment of myocardial perfusion

Harmonic Harmonic Pulse Power pulseB-mode power Doppler inversion inversion

Bubble-to-tissue Moderate Very good Good Very good

sensitivity

Off-line background Yes No Yes No

subtraction needed

LV-myocardium Poor Good Moderate Good

delineation

Wall motion artifacts None Can be severe Moderate Few

Real-time imaging No No No Yes

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 88

Page 106: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

89Perfusion Imaging

fluorocarbon, some to air-filled microbubbles:some are not available in all scanners. Becauseintramyocardial vessels make up less than 10percent of myocardial volume, even withcontrast specific imaging methods, myocardialcontrast signals are weak compared to those inthe LV cavity. Dissolution and disruption ofmicrobubbles by intramural pressure and ultra-sound exposure further reduces myocardialcontrast. Thus for every contrast methodoptimal adjustment of the echo machine iscrucial to gain maximum sensitivity.

3.3.3.1 Harmonic power Doppler (HPD)Harmonic power Doppler (HPD) is the currentmethod of choice for myocardial perfusionstudies using intravenous infusions of an airbased agent such as Levovist (20–24).Harmonic power Doppler is supplied bydifferent manufacturers under different names(harmonic power angio, loss of correlation imag-ing, etc). In the following chapters, the term is used to represent the whole group which

function in a technically similar manner. Thephysics of harmonic power Doppler is de-scribed in §1.3.3.2; the specific settings foreach system are discussed in §3.6. When usingHPD with air-filled contrast agents, themethod relies on the detection of bubbleechoes which are undergoing rapid change as aresult of disruption by the ultrasound beam. Itis therefore a high Mechanical Index (MI)method. Since these signals can be obtainedfrom stationary or slowly moving micro-bubbles, the entire myocardial microcircula-tion can be detected. The appearance of contrastin the myocardium can easily be visualisedsince pre-contrast recordings exhibit no HPDin the myocardium (Figure 5). No backgroundsubtraction is needed and this is a majoradvantage over B-mode techniques. Anotheradvantage is the better segmentation of myo-cardium from the cavities, which is a prerequi-site for further quantitative analysis. The signalintensities in the LV cavity and myocardiumusually differ by at least 15 dB. The only

Fig. 5 Harmonic power Doppler recordings (four chamber view) before (a) and during (b) Levovist infusion.Myocardial contrast is clearly visible.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 89

Page 107: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

90 Handbook of Contrast Echocardiography

clinically relevant limitations are wall motionartifacts, which can be avoided by carefuladjustment of the scanplane and the instru-ment controls in most patients.

3.3.3.2 Harmonic B-modeHarmonic greyscale images offer higher framerate and no motion artifacts (19, 25, 26). Theproblem with greyscale techniques (harmonicB-mode and pulse inversion) is that at highmechanical index myocardial tissue givesstrong tissue harmonic echoes and the contrasteffect must be extracted from the sum of base-line grey level and the additional effect of thecontrast agent (Figures 6, 7). The small increasein grey level which can be achieved with intra-venous contrast is hardly visible in areas wherethe tissue grey level is high (Figure 7). Forreliable evaluation, a sophisticated and time-consuming procedure, offline background sub-traction, is necessary. Background subtractionof non-contrast recordings from contrastimages is the prerequisite for quantification ofmyocardial contrast using the greyscale modal-

ities, and is discussed in §3.9.2 and Chapter 4.Even with this, background subtraction inechocardiography remains less satisfactory thanother imaging methods such as digital angio-

Fig. 6 Harmonic B-mode (four chamber view) before (a) and during (b) Levovist infusion. Because the tissueharmonic is strong in the septum, perfusion enhancement is better seen in the lateral wall.

Fig. 7 Rationale for harmonic power Doppler in per-fusion imaging. Left: in harmonic B-mode, contrastenhancement of the perfusing blood must exceed thealready strong echo from the tissue harmonic inorder to be detected. Right: with harmonic powerDoppler, the echo from the tissue is eliminated,except for tissue motion artifact, so that the samedose of agent produces a greater degree of perfusionenhancement.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 90

Page 108: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

91Perfusion Imaging

graphy or scintigraphy. Only a few studies havebeen successful in using harmonic B-mode forassessment of fixed perfusion defects and in-ducible ischemia (19).

3.3.3.3 Pulse inversion imagingPulse inversion imaging offers the improvedresolution and sensitivity compared toharmonic B-mode (27). However, at high MI,the tissue harmonic imposes the same limit onthe pulse inversion image as it does on theharmonic image (Figure 8). The simplestscheme of pulse inversion (see §1.3.3) uses twoconsecutive pulses. If the tissue is stationary,only bubbles will create a different echobetween the two pulses. If the tissue movesbetween pulses, its location between pulses willcreate a different signal similar to that of thebubble, rendering the method susceptible tomotion artifact. Like harmonic B-mode, pulseinversion creates a need for background subtrac-tion and offers limited segmentation of myo-cardial tissue from the cavities. If, however,motion artifacts can be overcome by appro-

priate setting of the trigger and adjustment ofthe pulse repetition frequency, pulse inversionis probably the modality of choice for offlinesubtraction (See Figure 6 in Chapter 4).

3.3.3.4 Power pulse inversion (PPI)Recently, pulse inversion Doppler, also knownas power pulse inversion (PPI) has been intro-duced as the latest contrast specific imagingmethod (Figure 9). The new technique providesthe ability to image the myocardial bloodvolume at acoustic powers sufficiently low to cause little bubble destruction. Real-timeperfusion imaging is therefore possible (28).The display of cavity and myocardial contrastis similar to that of power Doppler, so that back-ground subtraction is not required (Figure 9).With this method, it is possible to displaymyocardial thickening, wall motion and per-fusion simultaneously. If the encouraging pre-liminary clinical results are confirmed in dailyroutine patients, power pulse inversion andrelated low MI real-time techniques will formthe method of choice in the future.

Fig. 8 Pulse inversion imging at high MI before (a) and during (b) Levovist infusion. The baseline signals inthe septum are higher than in the lateral wall, making enhancement more difficult to see.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 91

Page 109: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

92 Handbook of Contrast Echocardiography

3.4 Special considerations for myocardial contrast

There are some general requirements whichapply to all perfusion imaging studies. Theserequirements are described here; particularsettings for the different imaging techniqueswill follow in §3.6. Stress testing, reading andinterpretation of the findings are common toall imaging modes and are discussed in§3.7–3.9.

3.4.1 Impact of the scanplane

The adjustment of the scanplane has a substan-tial impact on the success of a myocardialcontrast study. First the standard apical planesshould be found and optimised as usual(Figure 10). For myocardial contrast studiesthe grey level in the lateral and anterior wall isa good indicator of how well these segmentscan be filled with contrast. With regular adjust-

Fig. 9 Power pulse inversion at low MI before (a) and during (b) Definity infusion. Almost no colour is seenwithin the myocardium at baseline. During infusion intense enhancement is seen within the LV cavity and theentire myocardium. MI = 0.1, framerate = 10 Hz.

Scanplanes

(§3.4.1)

Triggered

imaging

(§3.4.2)

Contrast

administration

(§3.5)

Machine settings

(§3.6)

At least two apical views

Consider modified scanplanes

to improve display of

lateral/anterior wall

Mandatory, except power

pulse inversion technique

Continuous infusion

recommended

Different from non-contrast

imaging

Special considerations for myocardial contrast studies

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 92

Page 110: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

ment of the scanplane, the lateral wall (fourchamber view) and anterior wall (two chamberview) often have weak or almost no grey withinthe myocardium. This indicates that local

acoustic power is far less than for instance inthe septum with high grey levels (Figure 11).The reason for the weaker signals in thesesegments is attenuation of ultrasound causedby the interposition of pulmonary tissue orribs. For non-contrast imaging this hetero-geneity in grey levels is not a problem as longas endocardial borders are delineated. Forcontrast studies, however, a reduction in localacoustic power results in reduced or even nocontrast signals, which may be misinterpretedas a perfusion defect. Scanplanes should there-fore be adjusted in such a way that lateral andanterior walls are displayed with grey myo-cardium. This can be achieved by slightlychanging the scanhead position and movingthe lateral or anterior wall further to the centreof the imaging sector (Figure 11). Adjustmentof the scanplane should be carried out beforeinjection of contrast. During infusion ofcontrast the scanplanes should not be changed.

Parasternal views are less suitable for myo-cardial contrast studies than apical views.Usually the contrast within the right ventricle

93Perfusion Imaging

Fig. 10 Positioning the forearm for extendedperiods when the sonographer is seated to the rightof the patient. In order to avoid changing the scan-plane during acquisition of triggered images, theforearm is laid on the bed. Scanning without supportof the forearm is very strenuous.

Fig. 11 Adjustment of the four chamber view for a myocardial contrast study. With regular scanning thelateral wall is almost black (a). A slight modification of the scanhead position results in good grey levels inthe lateral wall (b).

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 93

Page 111: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

94 Handbook of Contrast Echocardiography

attenuates the propagation of ultrasound to thedeeper parts of the heart, so that inferior andposterior segments cannot be evaluated in manypatients. In some patients it is possible tomodify the scanplane in such a way that theright ventricle is out of the plane.

3.4.2 Triggered imaging

The disruption of microbubbles by continuousexposure to ultrasound prevents contrastappearing in the myocardium during real-timescanning at high MI. Intermittent or triggeredimaging refers to the interruption of ultrasoundexposure between the acquisition of one or more imaging frames (Figure 12). During thisinterval, the disrupted contrast within the myo-cardium is replenished and there are again

Fig. 12 Triggered imaging: the scanhead only trans-mits ultrasound for acquisition of single high powerframes (black bars), which can be chosen by settingthe trigger. The time interval between the acquiredframes can be varied by changing the number ofcardiac cycles. This schematic illustration demon-strates triggering on each, then on every 7th cardiaccycle.

Fig. 13 Harmonic power Doppler, four chamber view, different degrees of myocardial contrast duringinfusion of Levovist: real time imaging (a), triggered imaging once every cardiac cycle (b), every third cardiaccycle (c) and every 5th cardiac cycle (d).

a) b)

c) d)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 94

Page 112: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

95Perfusion Imaging

enough microbubbles to be imaged for thenext frame. The frames which are sampled canbe selected by trigger markers on the ECG(Figure 12). The sampled image is frozen onthe screen of the echo machine until acquisi-tion of the following frame. The introduction ofintermittent imaging signalled the initial break-through in intravenous myocardial contrastechocardiography (29, 30) and is explained in§1.3.4. In order to replenish the arterial andcapillary bed of the myocardium completely, atleast 10 seconds are needed. If the percentageof contrast replenishment is plotted againsttime interval between the imaging frames, anexponential curve is found, the asymptote ofwhich represents full replenishment (seeChapter 4). This means that most replenish-ment of contrast is found within a periodwhich clinically corresponds to pauses of 5 to 7cardiac cycles. Therefore trigger rates higherthan once every 7th cardiac cycle are usually notnecessary.

3.4.2.1 Incremental triggered imagingIt is not possible to predict the optimal triggerintervals in individual patients. The ‘best’triggering interval is dependent on cardiac out-put, dose of echo contrast and the actualattenuation. The recommended protocolstherefore include varying trigger intervals witha maximum of one frame every fifth cardiaccycle (Figure 13). In our experience this is longenough to provide a myocardial contrast strongenough to distinguish between perfusionabnormalities in most patients. The incrementalpulsing technique is not only suitable for find-ing the ‘best’ individual trigger but it can alsohelp quantifying the severity of stenosis.

3.4.2.2 Double or multiple triggerIn double or multiple triggering, the singledestruction image is replaced by a short series

of imaging frames, acquired in rapid succession(Figure 14). As before, the first frame after thereperfusion interval destroys all of the bubblesin the scanplane and creates the ‘perfusion’image. The second and subsequent frames donot show perfusion of the myocardium, but doshow cavity flow, as sufficient time has elapsedto allow the faster moving cavity blood to washnew bubbles into the scan plane. Motion arti-facts from tissue will also be present in the sub-sequent frames. One may therefore deducethat only those echoes present in the first frameand absent in subsequent frames are due to per-fusion. This is the main use of multiple trigger-ing. Some systems show the triggered framesside by side, which is especially helpful forinterpretation (Figure 15).

3.4.2.3 Flash echo During intermittent imaging and acquisition ofsingle frames which are frozen until the nextframe is sampled, the sonographer does not havecontrol of the scanplane and may lose it becauseof respiratory or other thoracic motion. Aningenious solution to this problem is to scancontinuously in a ‘scout’ mode during the

Fig. 14 Double frame triggering: short insonifica-tion for acquisition of two frames alternating withperiods with no transmission of ultrasound (one ormore cardiac cycles).

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 95

Page 113: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

interval with low emission power (MI < 0.2)which does not disrupt the bubbles (Figure 16).With low MI harmonic imaging, contrast canbe displayed in the cavities but not in the myo-cardium; cardiac structures can be displayed forcorrection of the scanplane if the patient

breathes or moves. Destruction frames at highMI are triggered as before, creating the perfusionimage. Multiple triggering may also be used.

3.4.2.4 Power pulse inversion flash echoThe superior sensitivity of power pulse inversion

96 Handbook of Contrast Echocardiography

Fig. 15 The double trigger technique: The top panel shows a four chamber view using harmonic Doppler andmultiple frame triggering. Every six beats, two high power ultrasound frames are made to disrupt the bubbleswithin the myocardium. The right panel shows the second frame, where some of the myocardial signal is nowabsent. This corresponds to bubbles that have been destroyed, an indirect sign of perfusion. The lower twoimages are from the same patient at a single beat trigger. Imaging every beat does not give enough time tofill the myocardium with bubbles. Therefore, the coloured myocardium is not a sign of perfusion. Persistenceof myocardial harmonic power Doppler signals in the second frame suggests motion artifact rather than per-fusion.Courtesy of J Luis Zamorano, University of Madrid, Spain

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 96

Page 114: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

97Perfusion Imaging

(PPI) in detecting bubbles at low MI means thatintermittent imaging is not needed for powerpulse inversion to achieve display of myocardialcontrast, because low transmit power results inonly minor destruction of microbubbles.However, even with PPI, triggering may stillprovide a benefit. During continuous imagingof perfusion at low MI, a single high powerframe may be used to disrupt the contrast agentwithin the myocardial tissue and then to assessreperfusion in real-time (see §4.2.3.3). Realtime dynamics of myocardial contrast replenish-ment provides additional information to thesteady signal intensities seen during infusion.

3.4.2.5 Systolic versus diastolic trigger

Contrast imaging is often improved whenframes are sampled during systole rather thanduring diastole. The small systolic cavity causes

less attenuation compared to diastole and thethickened myocardium facilitates further quan-titative analysis, for instance by allowing regionsof interest to be positioned without touchingthe cavity. Moreover, the myocardium movesto the centre of the imaging sector where sensi-tivity to contrast is better than in the lateralsectors of the imaging field. However, myo-cardial blood flow is highest during diastoleand especially subendocardial vessels aresqueezed during systole. Since myocardialcontrast echo evaluates the blood volume andthe replenishment which takes several cardiaccycles, it is probably irrelevant which portionof the cardiac cycle is chosen for data acquisi-tion. Positioning the trigger point is easy withharmonic B-mode, where just the frame at thetop of T-wave should be selected. In powerDoppler, wall motion artifacts may be present,requiring the sonographer to take specialmeasures to suppress these artifacts or look forother frames with fewer artifacts (for specificsuggestions see §3.6.1.1).

3.5 Choice of agent and method of administration

3.5.1 Continuous infusion versus bolus injection

Until recently, bolus injections have been wide-ly used for myocardial contrast echo. They stillmay be used for a rest study in acute trans-mural myocardial infarction, providing qualita-tive information as to whether there is a per-fusion defect or reperfusion. For all otherindications, the contrast agent should beadministered as an infusion (Figure 17, 18).Although administration by bolus saves sometime, there are a number of disadvantages,

Fig. 16 Flash imaging: continuous scanning withlow transmit power using low transmit power is per-formed to adjust the scanplane without display ofmyocardial contrast. Single frames are acquired withhigh transmit power to image myocardial contrast.The power pulse inversion method uses the samesequence of frames allowing display of myocardialcontrast signals in real-time at low transmit power.Single or multiple periods of insonification may beused to destroy the myocardial microbubbles and toassess myocardial contrast replenishment.

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 97

Page 115: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

most notably the comparatively short time fordata acquisition, frequent artifacts and greatlyreduced ability to quantify the resulting data.Wei et al. offer the following reasons for usingan infusion (31, 32): 1. It is relatively easy to adjust the dose of

contrast agent to the patient’s particular imaging conditions.

2. Blooming and contrast shadowing, regular occurrences with administration by bolus,can be considerably reduced by titrating the infusion rate for each individual.

3. Loosing the scanplane – not an uncommon occurrence in intermittent imaging – does not result in loss of the study.Recordings can be repeated under compar-

98 Handbook of Contrast Echocardiography

Fig. 17 Infusion of Levovist for myocardial contrast echocardiography. A PULSAR system (Medrad Inc.) filledwith two vials of 4 g Levovist (a) with a short tube connected with the venous cannula in a cubital vein (b).

Fig. 18 Set-up for infusion of undiluted perfluorocarbon agents.The undiluted agent is infused at low speed,a saline infusion at high speed prevents settlement of the microbbubbles within the tubes and the vein. Thisallows continuous infusion of Optison, but agitating the pump is necessary during infusion (left). For Definity thisset-up is recommended only for real-time perfusion imaging, but agitation is not necessary (right).

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:25 Uhr Seite 98

Page 116: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

99Perfusion Imaging

able conditions or additional, modified planes can be imaged.

4. The infusion can be started or modified by the same person who is performing the ultrasound recordings. Repeated adminis-trations of bolus usually require an extra pair of hands.

5. A simple quantification of myocardial contrast effect can only be achieved by using a stable contrast infusion.

3.5.2 Preparation of contrastinfusion

For infusion of Levovist, a venous line shouldbe introduced into a cubital vein. For other

agents and vasodilator stress, forearm veins canbe used. Preparation of the contrast agentshould not begin before the pre-contrastscanning has shown that the patient is suitablefor a myocardial contrast study (good imagequality, no contraindications). When harmonicpower Doppler is used, the triggered framesshould show no wall motion artifacts and good‘spontaneous contrast’ in the LV cavity.

3.5.3 Adjustment of infusion rate

Optimisation of infusion rate is performed byvisual evaluation of contrast intensities in theapical and basal segments using triggered imag-ing (one frame every 5th beat is best to begin

Fig. 19 Impact of the infusion rate on the display of myocardial contrast (harmonic power Doppler, modifiedfour chamber view with optimal display of the lateral wall, infusion of Levovist). a) Dose too low, fullopacification of LV, but weak contrast signals within the myocardium. b) Dose too high, contrast shadowing,strong signals in the nearfield but no signals at greater depth. c) Dose too low, septum signal intensities arelow compared to the lateral wall suggesting a perfusion deficit. d) Optimal dose with similar signal intensitiesin septum and lateral wall.

a) b)

c) d)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 99

Page 117: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

100 Handbook of Contrast Echocardiography

Contrast Contents Dilution Required Cannula Loading Initial

agent of vial number size bolus infusion

of vials rate

Levovist 4 g granulate add water to 2 vials 18 gauge 2 ml 1.5– 5 ml/min*

400 mg/ml

Optison** Suspension undiluted 1 vial 20 gauge 0.3 ml 7 ml/h

for injection (3 ml vial)

Definity Suspension 1.3 ml in 1 vial 18–20 0.5 ml 6 ml/min

for injection 50 ml gauge

(1.3 ml vial) saline

* Depending on ultrasound imaging system

** Not stable in a regular infusion pump, special set-up necessary (Figure 18) (33)

Dosing of ultrasound contrast agents for myocardial perfusion

with). Weak signals in the entire myocardiumindicate underdosing of contrast (Figure 19).An optimal infusion results in homogeneouscontrast enhancement in the apical and basalsegments. In this situation cavity contrast satu-rates and strong cavity signals can be seen withinthe left atrium as well as the left ventricle.Intense signals in the nearfield but weak orabsent signals at a greater depth indicatecontrast shadowing: the infusion rate shouldbe reduced. Guidelines as to how to adjust theinfusion rate for some different agents arelisted in the Table. Because of the time neededto reach a steady state, adjustment of doseshould not be performed in intervals of lessthan one minute.

3.6 Instrument settings

For successful studies correct adjustment of the imaging system to the specific acousticproperties of the contrast agent is essential. In

contrast echo some parameters which are nottouched in routine B-mode and Doppler echo-cardiography become crucial. The trainee inmyocardial contrast echo must learn how tomanipulate some controls of the echo machinewhich he or she has never used before. Carefuladjustment is necessary, as even minor devia-tions from the optimal setting may result in adramatic reduction in sensitivity. First, initialsettings are given which may be stored asscanner ‘presets’. Next, we provide guidelinesfor optimising settings for each individualpatient’s conditions. This adjustment of instru-ment controls is performed before infusion ofcontrast, so that during the contrast infusionitself no further changes are usually necessary.

3.6.1 Harmonic power Doppler

In harmonic power Doppler, the aim is toachieve complete disruption of the contrastagent in the myocardium and to display the

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 100

Page 118: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

101Perfusion Imaging

power Doppler signal without also displayingtissue motion artifact. The most importantmachine settings are listed below. Please notethat the effect of machine settings varies withindividual machine architecture. The recom-mendations that follow are for initial settingsand should be read in conjunction with thosesuggested by the manufacturer.

3.6.1.1 Setting the trigger pointDuring spontaneous respiration the triggerpoint is set at the top of the T-wave andintermittent imaging is performed with oneframe each cardiac cycle. After recording ofseveral cardiac frames in triggered mode, avisual check should be made as to whether wallmotion artifacts are superimposed on the myo-cardium (Figure 20). It is essential that nocolour signals should be visible within themyocardium before injection of contrast.Within the LV cavity there may be some spon-taneous harmonic power Doppler contrastthroughout the entire cardiac cycle. Wallmotion artifacts must be evaluated during

triggered imaging, as artifacts in continuousimaging will be different.

If wall motion artifacts are visible, the follow-ing manoeuvres may be undertaken:

1. Changing the trigger point by advancing ordelaying in single increments. This will beeffective in most patients. If no systolic framecan be found without wall motion artifacts,diastolic triggering can be set with a triggershortly before the onset of the P-wave.

2. Controlled breathing or breathold. In patientswith wall motion artifacts induced by respira-tory movements, holding the breath can betolerated only for frames with a high triggerrate. For longer trigger intervals ‘controlledbreathing’ is recommended: the patient isasked to breathe shallowly and arrest inspira-tion when the imaging frame is obtained. Thepatient should have some practice with thistechnique before baseline and contrast record-ings are performed.

Fig. 20 Selection of the trigger position for intermittent imaging: four chamber view, two consecutive framesacquired during baseline: a) shows optimal trigger point with no wall motion artifacts, b) is not usefulbecause of wall motion artifacts in the lateral wall and the septum. Note intracavitary harmonic powerDoppler signals in both frames.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 101

Page 119: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

102 Handbook of Contrast Echocardiography

3. Increasing the PRF. By reducing the periodbetween, this reduces motion artifact. It alsoreduces sensitivity to the contrast agent.

4. Increasing wall filter setting. This also elim-inates some of the contrast signal, so should beused carefully.

5. Increasing the colour threshold. This is asomewhat perilous measure that will eliminate

any of the contrast signal whose strength isbelow that of the moving tissue. Not recom-mended!

6. Decreasing the colour gain. A last resort,which can often reduce sensitivity to thecontrast agent to the point that it will not beseen on infusion. If a gain adjustment must bemade, a satisfactory level to eliminate artifactshould be established before contrast, but

Harmonic power Doppler perfusion studies: settings

Initial settings for all systems (preset):

Scanhead Lowest frequency

Transmit power Mechanical Index (MI) > 1.2

Receive gain Default

Display-Mode Standard monochromatic map

Dynamic range Maximum

Persistence Disabled

Line density Lowest value

Sensitivity Medium

Wall filter High

PRF 1500 kHz (Optison/Definity), 2500 kHz (Levovist)

Scanplane Conventional planes

Optimise in tissue harmonic mode (see text) then reduce

receive gain for tissue harmonic

Imaging field Adjust box so that the entire LV-myocardium is included

Focus Below mitral valve (apical planes),

Below posterior wall (parasternal view)

HP system: move towards apex if apical defect present

Trigger Systole (peak of T-wave)

Double trigger, if available

Once every cardiac cycle (increase during infusion of contrast)

Carefully adjust to avoid wall motion artifacts (see text)

Individual adjustment of instrument controls:

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 102

Page 120: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

103Perfusion Imaging

implemented only after arrival of the contrast,when its effect on the study can be seen.

If double frame acquisition is available, bothconsecutive frames should be checked for wallmotion artifacts, as they are acquired at slightlydifferent points in the cardiac cycle.

3.6.2 Harmonic B-mode

Harmonic B-mode may be used if the availableinstrument does not provide harmonic powerDoppler or if the offline subtraction method hasbeen chosen. Harmonic response increases withtransmit intensity of ultrasound, so a high MI isrecommended (see Table). The user should beprepared, however, to see strong pre-contrastechoes from tissue harmonics. In fact, a good

Harmonic B-mode for myocardial perfusion: settings

Initial settings for all systems (preset):

Individual adjustment of instrument controls:

Scanhead Lowest frequency

Transmit power Mechanical index (MI) > 1.2

Receive gain Default

Compression None

Dynamic range Maximum

Persistence Disabled

Line density Lowest value

Scanplanes Find conventional planes then reduce

receive gain (see text)

Focus Below mitral valve (apical planes),

Below posterior wall (parasternal view)

Transmit power Reduce initial level if apical defect or swirling is seen

Increase MI if global contrast is too weak

Receive gain May be reduced slightly to suppress myocardial

tissue echo

TGC As in standard imaging

Lateral gain (HP) As in non-contrast imaging

Trigger Systole (peak of T-wave)

Once every cardiac cycle (increase during infusion

of contrast)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 103

Page 121: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

104 Handbook of Contrast Echocardiography

Mode Harmonic power angio Also known as: coded harmonic angio;

colour power angio; harmonic power Doppler.

Transducer Lowest, eg 1.5/3.0 MHz Lower frequencies give better bubble

frequency disruption and hence sensitivity,

but slightly poorer resolution.

Output power High: MI > 1.0 High MI is essential for bubble disruption.

(MI) Usually the maximum is best.

Colour box Embrace entire MI reduces at edges of sector, making

size/position myocardium of interest imaging there less reliable.

Focus Level of mitral valve Focus affects uniformity of exposure

conditions in image. With some systems

(eg HP Sonos 5500), it is necessary to

move focus to apex to see perfusion there.

Pulse repetition 2.5–4.0kHz Also known as: Doppler scale. Highest possible.

frequency (PRF) Lower PRF increases sensitivity to both contrast

but also tissue motion. Air agents (eg Levovist)

work better at high PRF than perfluorocarbon

agents (eg Optison or Definity).

Trigger Mid T-wave (initial setting) Note that trigger often affects MI. Adjust trigger

to minimise motion artifacts and adjust other

controls with trigger active.

Frames 2–4 This controls the number of frames

acquired at each trigger. The first frame

shows contrast combined with artifacts,

subsequent frames show artifacts only.

Dual display On (if available) With multiframe trigger, a dual image display

showing first and a subsequent frame can help

interpretation of contrast perfusion study.

Colour gain Decrease to the point that Adjust gain last, after trigger, PRF, filter

motion artifacts are just and other colour settings have been

visible in pre-contrast image adjusted to minimise motion artifact.

B-mode gain Sufficient to see Excessively high B-mode gain can cause grey

endocardium to overwrite colour in image.

Control Setting Comments

Understanding machine settings for harmonic power Doppler perfusion studies

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 104

Page 122: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

105Perfusion Imaging

way to predict whether adequate contrast signalswill be obtained in the lateral wall is to checkwhether clear tissue harmonic echoes can beseen in these areas. Thus one first ensures thatthere is a good tissue harmonic echo using anormal gain setting, after which the receive gainis reduced in readiness for the contrast study. Atbaseline the myocardium should be almostblack and only the endocardial borders shouldbe visible.

As with power Doppler, the position of thetransmit focus and the line density affectbubble disruption. Reduction of line densitycompared to that used for non-contrast imag-ing and placing the focus below the mitralvalve helps to make insonation more homo-geneous. Dynamic range should be as high aspossible to ensure that small changes in contrastsignals are not overlooked. Compression andother forms of nonlinear processing which

Colour priority Maximum Also known as: Angio priority. Forces

colour to overwrite grey level in image.

Colour threshold Minimum Removes low level contrast signals:

must be set to zero

Line density Minimum Reducing line density limits inadvertent

bubble destruction (that is, bubble

destruction that does not contribute to

the image). Increasing frame rate setting

may also decrease line density.

Frame rate Medium – High Higher frame rates may decrease

line density

Persistence Zero Also known as: frame averaging. Contrast

signals are typically present in a single

frame. Averaging may reduce their value.

Ensemble length Medium – High (8 pulses) Also known as: Doppler sensitivity, packet length.

This determines the number of pulses sent along

each scan line. A high number improves both

bubble disruption rate and sensitivity to the agent.

Lower settings can be used for Levovist.

Dynamic range High Too low dynamic range will create an

‘on/off’ display of colour; too high will increase

background noise and blooming from the cavity

Wall Filter Medium – Low Also known as: low velocity reject.

Eliminates signals from slowly moving structures

such as tissue. At too high a level, will also reject

contrast echoes. Use the lowest setting possible,

while still eliminating motion artifact from tissue.

Control Setting Comments

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 105

Page 123: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

106 Handbook of Contrast Echocardiography

Pulse inversion imaging for myocardial perfusion: settings

Initial settings for all systems (preset):

Transmit power Mechanical Index (MI) = 0.3 for triggered imaging

Receive gain Default

Compression None

Dynamic range Maximum

Persistence Disabled

Line density Regular

Individual adjustment of instrument controls:

Scanplanes Corresponding to conventional planes.

Optimise before injection of contrast (see text), then

reduce receive gain

Focus Below mitral valve (apical planes),

Below posterior wall (parasternal view)

Receive gain Adjust to be just above the noise level

TGC As in standard imaging

Trigger Systole (peak of T-wave)

Once every cardiac cycle (increase during infusion of contrast)

Adjust to reduce motion artifact

enhance the display of diagnostic informationin non-contrast echocardiography serve only tocorrupt the interpretation of regional myo-cardial perfusion echoes. Therefore no com-pression should be used for myocardialcontrast echo. Positioning of the trigger is easywith harmonic B-mode and the double trigger(frame) technique is not required to check forartifacts. However, the double frame techniquemay be used for subtraction during contrastinfusion.

3.6.3 Pulse inversion imaging

Similar adjustments are necessary for the pulseinversion method as for harmonic B-mode. Todate, there is limited experience with thismethod, so the recommendations listed in theTable are preliminary. The pulse inversionmethod has a remarkable sensitivity forcontrast at low transmit power settings and ismuch more effective than harmonic powerDoppler or harmonic B-mode at low MI. It istherefore recommended to work with lowerMIs with pulse inversion imaging than withharmonic B-mode, as the reduction in tissue

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 106

Page 124: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

107Perfusion Imaging

harmonic improves myocardial perfusioncontrast. With lower MI, a lower infusion ratemay also be more effective. For quantitativeanalysis, which should be the aim of everymyocardial contrast study, triggered imaging isstill as necessary as it is with the other imagingmethods. With triggered imaging the contrasteffect increases with increasing transmit power.Spatial resolution is much better with pulseinversion than with power Doppler or har-

monic B-mode. The potential problem ofmotion artifact in greyscale pulse inversionimaging means that special attention should bepaid to the trigger point and to the pulserepetition frequency, as with harmonic powerDoppler. In some systems, increasing the pulserepetition frequency can be achieved simply byincreasing the framerate, which will tend toreduce motion artifact at the expense of someloss of sensitivity to the agent.

Power pulse inversion for real-time myocardial perfusion: settings

Transmit power Mechanical index (MI) = 0.15

PRF 2500 Hz

Dynamic range Low

Sensitivity Medium

Persistence Disabled

Line density Low (increases frame rate)

Penetration depth 12.7 cm

Focus 10 cm

Receive gain Default

Individual adjustment of instrument controls:

Transmit power Good window : MI = 0.09–0.11

Moderate window: MI = 0.11–0.14

Scanplanes Corresponding to conventional planes

optimise before injection of contrast (see text)

TGC Depth 1 ~ 30 optimise to suppress tissue harmonics

Depth 2 > 50 optimise to suppress tissue harmonics

Depth 3–8 maximum

Trigger* Optional – peak of the R wave

Once every 15th cardiac cycle

Double frame (MI = 0.5)

Initial settings for all systems (preset):

* can be used to destroy microbubbles in myocardium for real-time destruction-reperfusion measurement (see §4.2.3)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 107

Page 125: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

108 Handbook of Contrast Echocardiography

3.6.4 Power pulse inversion

This is the only method which currently allowsreal-time perfusion imaging. Preliminarysettings are listed in the Table. In practice,‘real-time’ means frame rates of up to about26 Hz. Very low transmit power (MI = 0.09–0.15) is the essential prerequisite for visualisingmyocardial contrast in real-time. With currenttechnology effective myocardial contrast canonly be achieved with perfluorocarbon agents.Higher doses are needed: bolus injections of0.5–1.0 ml of Optison or infusion rates of0.5 ml/min provide good myocardial contrast.Single or double frames with high transmitpower may be used to destroy the microbubbleswithin the myocardium and to assess re-plenishment in real-time (see §4.2.3.3). Notriggering is necessary.

3.7. Image acquisition

As with conventional stress echocardiography,all recordings should be digitally stored andanalysed offline. Capabilities for digital image

storage are provided in all harmonic echomachines, using direct connection to aDICOM network or by magneto-optical(MO) discs. Before each contrast study, acheck must be made to see whether there isenough space to store the study. Videotape is auseful backup and also forms a continuousrecord. During data acquisition shallowbreathing is recommended. Holding the breathis not useful for triggered imaging, becausemany cardiac patients are unable to do so forsufficiently long to cover several frames at longtrigger intervals. It usually takes at least twominutes before the myocardial microbubbleconcentration has reached a steady state. Thuscontrast recordings should be started threeminutes after contrast infusion has beeninitiated (Figure 21). The recommendedrecordings for all imaging modalities exceptpower pulse inversion imaging are shown inthe Table.

Power pulse inversion imaging can be per-formed in real-time. At least 5 cardiac cyclesshould be sampled in each view. Breatholdingis often useful to obtain reproducibleconsecutive cycles.

Recommended image files to document a myocardial contrast study

• four chamber view • four chamber view • four chamber view• two chamber view • two chamber view • two chamber view• long axis • long axis • long axis

• Each file should include at least 5 frames, triggered 1:1, 1:3 and 1:5.

• With real-time imaging at least 3 cardiac cycles should be recorded, followed by the destruction bursts and the variable period of myocardial contrast replenishment which is completed within 10 cardiac cycles.

• Baseline recordings are necessary in PPI and HPD to prove, that tissue harmonics and wall motion harmonics are eliminated. With B-mode harmonic and pulse inversion imaging baseline recordings areneeded for background subtraction.

RestBaseline Stress

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 108

Page 126: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

109Perfusion Imaging

Fig. 21 Protocol for resting and stress studies.

3.8 Stress testing during myocardial contrast echo

3.8.1 Exercise and dobutamine stress

During myocardial contrast echo, stress can beperformed with any of the available methods,though the special imaging conditions makepharmacological stress more suitable. Exercisestress (exercise or treadmill) provides thegreatest myocardial oxygen consumption com-pared to other methods (10). However, scann-ing during and after exercise is difficult becauseof increased respiratory and cardiac motion. At peak stress it is sometimes a problem tokeep or reproduce the same scanplane, becausethe patient may be unable to hold his or herbreath. Because myocardial contrast echorequires recording a series of consecutive trig-gered frames, physical exercise is a challengingprocedure for the echocardiographer. Withpulse inversion imaging or harmonic powerDoppler, wall motion artifacts are accentuatedfollowing an increase in heart rate and inotro-

pic state leading to further reduction of imagequality. Finally, the need to adjust the triggerpoint to the varying heart rate further compli-cates the method.

Dobutamine infusion only slightly reducesthese difficulties. At peak stress there are prob-lems similar to those found during physicalexercise, where the enhanced cardiac andrespiratory motion limits scanning, particularlywith harmonic power Doppler and pulse inver-sion. Thus vasodilator stress is our method ofchoice for myocardial stress contrast echo-cardiography.

3.8.2 Vasodilator stress

During dipyridamole and adenosine infusionthe changes in heart rate are moderate and theinotropic state is not altered significantly. As aresult, oxygen consumption is not increased.Ischemia only develops with additional hori-zontal or vertical steal (Figure 22), which isfound in up to 50 percent of patients with sig-

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 109

Page 127: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

110 Handbook of Contrast Echocardiography

nificant stenoses. Only these patients developwall motion abnormalities. Thus wall motionabnormalities are found less frequently withvasodilator stress than with exercise or dobuta-mine. However, blood flow in the perfusionbed supplied by a significantly stenosed arteryincreases only marginally (or decreases whensteal phenomena develop) compared to thethree to four fold increase in areas supplied bynon-stenotic arteries, and a perfusion mis-match can be displayed with high sensitivity(34, 35).

The myocardial segments not affected bycoronary artery stenosis will show an increasein signal intensities corresponding to theirnormal flow reserve whereas hypoperfusedareas will show no increase in signal intensities

or even a decrease due to a steal. Evaluation ofvasodilator stress images can be performed byvisual judgement or by using post-processingand quantitative analysis (see §4.2). Sincevasodilator stress is not widely used for con-ventional stress echocardiography, the doseregimen and side effects and contraindicationsare reviewed in the Tables. Side-effects arecommon, particularly for adenosine stress, butthey are benign. Dipyridamole and adenosinehave extra-cardiac effects which limit scanningduring infusion of contrast. The most import-ant is dyspnea which causes deep breathingand threatens loss of the imaging plane. Thepatient must be informed about this phenom-enon before starting the adenosine infusion.Many patients can tolerate the discomfort ifthey are informed before the examination.

Fig. 22 Effects of adenosine on intramyocardial vessels in normal epicardial coronary arteries and in thepresence of a stenosis. At baseline (a) the blood flow in the perfusion bed of the stenosed coronary artery isnot different from areas without stenoses due to maximal dilatation of the arterioles and collateral flow. Withadenosine (b) arteriolar resistance decreases in the normally perfused territory resulting in a 3- to 4-foldincrease of blood flow in the normal myocardium. In the presence of a significant stenosis blood flow doesnot increase, but may decrease due to reduced collateral flow (steal phenomenon).

baseline

myocardium myocardium

adenosinea) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 110

Page 128: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

3.8.3 Combined assessment ofwall motion andmyocardial perfusion

Myocardial contrast echocardiography can beperformed in conjunction with a routine stressecho protocol for wall motion analysis (see§3.3.1). With real-time perfusion imagingsimultaneous assessment of perfusion and wall

motion is possible using a constant infusion ofcontrast (Figure 23). With triggered imagingcontrast infusion should be confined to MCErecordings for two reasons: contrast is notneeded for left ventricular opacification (LVO)because good acoustic windows are recom-mended for perfusion studies. Furthermore,the doses for MCE using triggerd imaging arelower than for real-time LVO studies and

111Perfusion Imaging

Ease of Induced ischemia Modality of

scanning choice

Physical exercise difficult needed 2 D harmonic PPI

Dobutamine difficult needed 2 D harmonic PPI

Dipyridamole easier not necessary all modalities

Adenosine easier not necessary all modalities

Choosing a stress modality for harmonic power Doppler

Fig. 23 How to integrate an MCE study using triggered imaging into a routine stress echo protocol for wallmotion analysis. With adequate baseline images contrast infusion can be confined to MCE recordings.

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 111

Page 129: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

112 Handbook of Contrast Echocardiography

adjusting infusion speeds to LVO or MCE isdifficult with the time constraints of a stress test.

3.9 Reading myocardial contrast echocardiograms

Reading MCE studies involves steps which aresimilar to those of a regular stress echo forevaluation of LV wall motion, which mayparallel a perfusion study. In each view thesegments defined by the American Society ofEchocardiography (ASE) are evaluated and thecorresponding segments at rest and duringstress are compared. Reading and interpreta-tion of perfusion studies needs to be performedoffline. Visual assessment of perfusion, how-ever, is mandatory during the stress phase of thecontrast study: if perfusion defects developunder stress, the stress should be stopped. Thefollowing guidelines for reading and inter-pretation apply to all imaging techniques. Theevaluation of triggered images is in principlenot different to that of images made with real-time perfusion imaging. There are several waysto deal with the stored recordings:• Visual judgement of original single frames

and cine loops,• Visual judgement of post-processed images,• Quantitative analysis of original or

postprocessed images Original recordings can be used for harmonicpower Doppler and power pulse inversion,whereas evaluation of recordings obtained withharmonic B-mode is more difficult. When usinga greyscale technique at high MI, the contrastsignals need to be separated from the tissue har-monic signals present at baseline. Because this isso difficult to appreciate subjectively from thecompressed grey levels seen on the screen, back-ground subtraction and colour coding must be

applied before visual judgement is attemptedfrom harmonic B-mode recordings.

3.9.1 Visual assessment of unprocessed recordings

Visual assessment is practicable with powerDoppler and power pulse inversion recordings.High MI greyscale images should not be as-sessed in this way. However, once softwaretools for background subtraction and colourcoding of the myocardial contrast signals havebeen applied, visual judgement of processedgreyscale images uses the same criteria as thosefor the assessment of unprocessed powerDoppler images. Visual judgement mainlyrelies on the signal intensities of myocardialcontrast, which reflects relative myocardialblood volume. In order to get an idea of myo-cardial blood flow, one needs to evaluate thechanges in the signal intensities with time. Avery rough estimate of flow is provided by thetime needed to achieve visible myocardial

Visual grading of myocardial contrast: separate scores for

each of the six ASE myocardial segments of the scanplane

0 No contrast enhancement

1 Poor contrast enhancement, incomplete

filling of segment

2 Moderate contrast enhancement,

complete filling

3 Strong contrast enhancement

X Unsuccessful (artifacts, attenuation,

blooming, etc)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 112

Page 130: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

contrast following the disruption of myo-cardial contrast by high power frames (36).With the power pulse inversion method thisreplenishment can be recorded in real-time.Using the other methods and triggered imaging,one can determine lowest trigger interval atwhich myocardial contrast becomes visible.

3.9.1.1 Normal perfusionThe dose regimen and machine settings listedin §3.6 provide moderate or strong contrastenhancement at the longer trigger intervals(1:3–1:5). At shorter intervals (1:1–1:3), thecontrast pattern is patchy or reticular, repre-senting the larger intramyocardial vessels.More homogeneous opacification is foundwith intervals above 1:3 (Figure 24). Withoptimal display, myocardial contrast involvesthe entire myocardial thickness (Figures 25,26). Isolated epicardial contrast is caused bydisplay of epicardial vessels and may be foundduring wash-in of contrast following a bolusinjection but not during infusion and triggeredimaging. In power Doppler mode, isolated epi-

cardial colour signals represent wall motionartifacts rather than vascular contrast (see§3.11.4).

Before MCE studies are evaluated, the readershould judge the corresponding scanplanes forwall motion. At rest all segments without wallmotion abnormalities should show contrastenhancement. Missing or incomplete myo-cardial contrast in a normally contractingmuscle is an artifact! Incomplete myocardialcontrast enhancement may often be found innormals – particularly in the basal segments.These basal dropouts represent attenuationrather than real perfusion defects – especiallywhen the defect does not end in the myo-cardium. During stress these segments often fillin. Coronary stenoses usually involve mid andapical segments, so that isolated basal defectsseem to have only minor clinical impact.

The variation of exposure to transmitted ultra-sound in different parts of the image shouldalways be taken into account in a visual

113Perfusion Imaging

Fig. 24 Example of successful visualisation of all segments in modified four chamber view during Levovistinfusion.

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 113

Page 131: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

114 Handbook of Contrast Echocardiography

Fig. 25 Normal perfusion: SPECT, horizontal axis (a) and harmonic power Doppler, four chamber view (b). Courtesy of Francesco Gentile, University of Milan, Italy

Fig. 26 Harmonic power Doppler, four chamber view, Optison infusion. Two frames showing normal findingswith excellent delineation of cavity from myocardial signals. Note the intense spots within the myocardiumwhich represent intramyocardial vessels. In the right frame the myocardial contrast scores of the six segmentsare shown.Courtesy of Joanne Sandelski and Steve Feinstein, Rush University, Chicago IL, USA

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 114

Page 132: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

assessment. When perfusion is normal, signalsof varying intensity are found in the individualmyocardial segments (Figures 26, 27). This isbecause the effective mechanical index is notthe same at all points in the image plane. Inlateral regions, the transducer transmits lessintense pulses; in deep regions attenuationreduces the intensity progressively (Figure 28).

Dropouts in the entire lateral wall (four cham-ber view) and the anterior wall (two chamberview) should be rare when the imaging guide-lines are followed. If lateral or anterior drop-outs cannot be eliminated, the diagnostic valueof an MCE study is limited. However, in somecircumstances, it is not necessary to achievegood contrast signals in the entire myocardium.For instance, in a perfusion study for the as-sessment of the significance of an LAD stenosis,lateral dropout can be tolerated.

During stress an increase in myocardial signalintensity is observed. Due to an increase of

myocardial blood flow, myocardial opacifica-tion is seen at lower trigger intervals than in therest study. With real-time imaging, a corres-ponding reduction of contrast replenishmenttime is found. At rest it usually takes three ormore cardiac cycles to fill the myocardium,

115Perfusion Imaging

Fig. 27 Mean Doppler power in patients with nor-mal myocardial perfusion (four chamber view): notethe differences between the myocardial segmentsduring infusion of Levovist.

Fig. 28 Normal perfusion, harmonic power Doppler, two chamber view (infusion of Levovist). a) shows earlyLV filling during wash-in of contrast with some wall motion artifacts in the inferior wall. Note the lower greysignal intensities in the anterior wall compared to the inferior wall because of lower local acoustic power. Withoptimised dosing of contrast infusion (b), intense contrast signals are found in the entire inferior wall. Thedisplay of contrast in the anterior wall is excellent in the apical and mid segments but not in the basal segment.Courtesy of Gerd P Meyer, University of Hanover, Germany

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 115

Page 133: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

whereas myocardial contrast is seen within asingle cardiac cycle during adenosine infusion.

A side-by-side display of the baseline and stressframes or loops is recommended when readingperfusion stress studies. The best approach forvisual assessment is the comparison of the samesegments at baseline and during stress. If weakor no signals are seen in a segment, one shouldcheck for echoes in the deeper-lying segments.With septal and inferior defects, the oppositewall can also be compared. Note that this pro-cedure does not work in reverse. If no strongsignals are detectable in the compared seg-ments, the diagnosis of a perfusion defect isnot safe.

3.9.1.2 Perfusion defectTo date, myocardial scintigraphy has been theonly clinical method to image myocardial per-fusion. Myocardial contrast echocardiographyaims to replace scintigraphic techniques andpartly uses the same kind of reading. However,display of myocardial perfusion is different forthe two modalities. Often scintigraphic defectsdo not match the size of MCE defects. Thepoor spatial resolution and the special process-ing of myocardial scintigraphy result in per-fusion deficits which usually extend over theentire thickness of the LV wall. It is known,though, that perfusion abnormalities ofteninvolve subendocardial layers with preservedflow in subepicardial layers. Due to its betterspatial resolution, MCE has provided for thefirst time display of subendocardial ischemia(Figure 29). It too is capable of showing patchyand transmural patterns of perfusion defects,however (Figure 30).

As with myocardial scintigraphy, perfusiondefects with contrast echo are classified as fixedor reversible. A fixed defect is marked by a

relative decrease in signal intensity comparedto the adjacent myocardium, visible at rest andduring exercise. It may be subendocardial ortransmural. The apparent degree of a perfusiondefect depends on the amount of contrast pre-sent in the vessels of normally perfused myo-cardium compared to the vessels within thehypoperfused area. Even in a scar after a trans-mural infarction, it will be seen that somevessels are visible. High amounts of contrastwithin these vessels may cause intense signalsand even blooming which may obscure thevisual delineation of the hypoperfused areas.Since the amount of myocardial contrast isdependent on the trigger interval and the infu-sion rate of the contrast agent, the display of aperfusion deficit is variable (Figure 31).Perfusion defects may decrease or even dis-appear with increasing contrast dose or triggerinterval! This is particularly true for harmonicpower Doppler studies where blooming ismore evident than with the other imagingmodalities. Careful evaluation of all imagessampled at different trigger intervals is manda-tory: frames with blooming or saturation mustnot be used for evaluation. Due to these limita-tions, the absolute size of a perfusion defectshould not be estimated. It is sufficient tospecify the segments in which contrast patternslook abnormal.

Reversible defects appear normal at rest butbecome visible during stress, where a markedincrease of signal intensity is observed in normalmyocardium, compared to a slight increase oreven a decrease in the defect area. Thus, duringpeak stress reversible defects show a relativedecrease in signal intensity compared toadjacent normal myocardium. Usually thedelineation is enhanced at lower trigger inter-vals, when normally perfused tissue exhibits amoderate or strong contrast enhancement. The

116 Handbook of Contrast Echocardiography

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 116

Page 134: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

117Perfusion Imaging

Fig. 29 Harmonic B-mode, four chamber view, continuous infusion of Levovist, post apical myocardialinfarction. a) Subendocardial perfusion defect. b) Transmural defect in corresponding plane of ThalliumSPECT at rest.Courtesy of Ricardo Ronderos, Universidad Nacional de La Plata, Argentina

Fig. 30 Real-time resting myocardial contrast studies using power pulse inversion imaging following 0.3mlOptison. (a) Normal patient. (b) Patient with an extensive resting apical perfusion defect due to previousinfarction. Power pulse inversion frame rates were 15Hz at an MI of 0.1– 0.15.Courtesy of Thomas Porter and Feng Xie, University of Nebraska, NE, USA

a) b)

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 117

Page 135: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

118 Handbook of Contrast Echocardiography

Fig. 31 Harmonic power Doppler, four chamber view, infusion of Levovist. Example of shorter pulsing inter-val revealing apical perfusion defect. A defect is hardly visualised at 5:1 pulsing interval (a), but clearly visibleat 1:1 (b).

flow in perfusion beds distal to a significantcoronary stenosis is not changed by stress,therefore longer intervals are needed to opacifythe myocardium. With longer trigger intervalsthe hypoperfused volume of tissue stillcontains less microbubbles than the adjacentnormally perfused areas. However, the differ-ences in bubble concentration between normaland hypoperfused areas may not be sufficientlyhigh to provide adequate visualisation of theperfusion deficit, which can therefore be over-looked. Real-time perfusion imaging of replen-ishment following one or several destructionframes, provides multiple frames with different

concentrations of myocardial contrast. There-fore, the opportunity to create an optimal dis-play of a perfusion defect should be higherthan with triggered imaging. Clinical studiesare in progress to investigate this hypothesis.

Evaluation of the perfusion defects is furthercomplicated by effect of respiratory andscanhead movement on the signal intensities.Visual judgement should always be made usinga series of frames rather than single ones.Because of the high number of frames neededfor-side-by-side comparisons of rest and stressimages, arrangement of digitally stored data ona computer is most convenient. Having therecordings on a computer further facilitates theuse of advanced methods for quantitativeanalysis, which may complete the visual evalu-ation (see Chapter 4). Since conventional stressecho is often performed in conjunction with aperfusion study, acquisition analysis of MCErecordings is ideally implemented into thestress echo recording package on the scanner.

Calling a perfusion defect

A visually evident contrast defect is considered pre-

sent when there is a relative decrease in contrast

enhancement in one region compared with other

adjacent regions, which have the same or worse

imaging conditions.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 118

Page 136: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

119Perfusion Imaging

3.9.2 Visual assessment of post-processed recordings

Post processing of myocardial perfusion imagesis necessary when harmonic B-mode or pulseinversion imaging is used (Figure 32). This isbecause of the small incremental enhancementprovided by contrast over the backgroundtissue harmonic echo. Several offline imaginganalysis tools are available for this task, whichrequires digital recording of both baseline andcontrast enhanced images. Before subtractiontwo tasks must be carried out. First, grey levelsin the image, which are logarithmicallymapped from the echo level received by thescanner, must be converted to a linear scale. Ifthis is not done – for example if two imagesrecorded on videotape are simply subtracted –

the resulting image will depend on machinesettings, attenuation and other factors un-related to the amount of contrast in the myo-cardium, and essentially be meaningless.Second, the images must be aligned so that thesubtracted quantities correspond to the sameanatomic area. Because the heart has complexmovement and because the baseline andcontrast images are acquired at different times,this is not a trivial challenge. The process isdescribed in detail in §4.2.2.

3.9.3 Reporting visual judgement

We have developed reporting forms for per-fusion studies similar to those used in conven-tional stress echocardiography (Figure 33).

Normal perfusion Moderate or strong Moderate or strong signals

contrast enhancement at low trigger interval (1:1)

(score 2 or 3) at least at higher

trigger intervals (1:3 or 1:5) Increase of signal

(contrast enhancement in intensities at higher

basal segments may be poor) trigger intervals

Fixed perfusion defect No or poor contrast No or poor contrast

(Subendocardial or transmural) enhancement enhancement

(score 0 or 1)* (score 0 or 1)*

Reversible perfusion defect Moderate or adequate Poor signals at low

(subendocardial or transmural) contrast enhancement trigger interval (1:1)*

(score 2 or 3) at least

at higher trigger

intervals (1:3 or 1:5)

* Moderate or strong enhancement in adjacent segments with similar or worse imaging conditions

Rest Stress

Criteria for visual assessment of MCE studies

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 119

Page 137: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

120 Handbook of Contrast Echocardiography

First, the segments in each scanplane are evalu-ated for contrast enhancement using the threestep scoring as shown in §3.9.1. If a segmentcannot be evaluated because of wall motionartifacts or attenuation, it is marked with an‘X’. This is repeated for the rest and stressimages, which are then compared. To describethe changes between baseline and stress record-ings, a judgement is made as to whether thereis increased (+), decreased (-), no change (0) incontrast compared to baseline. If no signals arevisible in a segment at both baseline and stress,this segment cannot be used for assessment ofstress changes and is marked with an ‘X’.

Visual judgement is summarised using thefollowing statements:

1. Which segments can be evaluated? It is notalways necessary to have the information for theentire myocardium. If, for instance, the studyis performed to assess the hemodynamic rel-evance of an LAD stenosis, it is sufficient to seeseptal and anterior segments. In this casenormal or abnormal findings only pertain tothe LAD territory. This is different from myo-cardial scintigraphy, where perfusion is alwaysevaluated in the entire myocardium.

2. Judgement as to whether those segmentsseen are normal or abnormal according to thecriteria listed in the following Tables.

3. The results of the advanced methods foranalysis of myocardial contrast studies may be

Fig. 32 B-mode harmonic recordings (four chamber view) after background subtraction and colour coding,comparison with SPECT. Effect of increasing pulsing intervals of an anteroapical (a) and inferoposterior (b)myocardial infarction.Courtesy of Jonathan Lindner and Sanjiv Kaul, University of Virginia, Charlottesville, VI, USA

a)

b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 120

Page 138: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

121Perfusion Imaging

When is a perfusion defect real?

Criteria for a real perfusion defect (four chamber view)

Decrease of contrast enhancement in: compared to:

Apical septum Mid and basal septum, lateral wall

Mid septum Mid and basal lateral wall

Basal septum Basal lateral wall

Apical lateral wall Mid and basal lateral wall

Mid lateral wall Basal lateral wall

Basal lateral wall Not available

Criteria for a real perfusion defect (two chamber view)

Decrease of contrast enhancement in: compared to:

Apical inferior wall Mid and basal inferior wall, anterior wall

Mid inferior wall Mid and basal anterior wall

Basal inferior wall Basal anterior wall

Apical anterior wall Mid and basal inferior wall

Mid anterior wall Basal inferior wall

Basal anterior wall Not available

added to the report as soon there are referencevalues (see Chapter 4).

4. A brief indication as to which coronaryartery is thought to be involved.

3.10 Interpretation of myocardial contrast echo: clinical profiles

3.10.1 Acute myocardial infarction

In acute transmural myocardial infarctionthere is a high grade or complete reduction of

perfusion in a particular area of the myo-cardium. Therefore no, or only slight contrastwill be found in this area following intravenousinjection of the contrast agent (Figure 34). Aslong as there is complete occlusion of the epi-cardial vessel, the entire wall will show eitherno contrast signals or reduced filling if collateralblood flow is present. After restoration of bloodflow, contrast will fill in those areas with pre-served microvascular integrity, which is a pre-requisite for myocardial viability (Figure 35).Thus areas with complete necrosis will showno refilling with contrast after successful re-opening of the occluded coronary artery.Preliminary clinical studies have demonstratedthe usefulness of contrast echo in the evalua-tion of patients with acute myocardial infarc-tion (14, 15, 17, 36–42). Figure 36 shows a

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 121

Page 139: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

122 Handbook of Contrast Echocardiography

Fig. 33 Report form for visual assessment of a myocardial perfusion study.

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 122

Page 140: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

123Perfusion Imaging

Fig. 34 Harmonic power Doppler (apical two chamber view, Levovist infusion). Apical and anterior wall per-fusion defect (arrows) shown in anterolateral myocardial infarction.Courtesy of Folkert Ten Cate, Thoraxcenter, Rotterdam, Netherlands

flow chart for the use of contrast echo in thesetting of acute myocardial infarction.Although only preliminary clinical studies areavailable, the lack of effective alternatives forassessment of reperfusion may be sufficientjustification for using contrast echo if theguidelines for exclusion of false defects aretaken into account.

3.10.2 Scar or fibrosis versus viable myocardium

Necrotic tissue and the evolving scar are dis-played as fixed defects. They have no or mini-mal rest perfusion, because the microvascularnetwork has been damaged irreversibly duringischemia (Figures 37, 38, 39). In contrast, viablemyocardium – stunned or hibernating – still hasits microvascular network, which can be imaged

Fig. 35 Harmonic Power Doppler, four chamber view, Levovist infusion. Large apical defect in acute apico-septal infarction (a) and after successful reperfusion (b).Courtesy of Luciano Agati, University of Rome, Italy

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 123

Page 141: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

using contrast echo. It should be rememberedthat contrast echo provides information onmicrovascular integrity, not myocyte function.Absence of signal in an infarcted area repre-sents no reflow or a fibrotic scar in the chronicphase. Whether the presence of contrast in themyocardium is an indicator of viabilityremains to be established. Studies using intra-arterial injections of contrast do suggest thatthe presence of signal in areas of damagedmyocardium indicates viability (15), but littledata are available for intravenous contrastimaging. Until trials establish ranges, evalua-tion of viability should be confined to differen-tiating between a definite lack of contrast anda strong and convincing contrast effect. Studiesmay be used in combination with othermethods such as low dose dobutamine stress orpositron emission tomography (PET).

It has been shown that the appearance ofcontrast and full filling may often be delayedmarkedly because the area of interest is sup-plied mainly by collaterals. Thus quantitativeanalysis should include longer trigger intervals(e. g. 1:10) to detect late filling.

3.10.3 Coronary artery stenosis

Perfusion defects during exercise, but normalperfusion at rest are the typical findings in thevisual assessment of significant coronarystenoses (see §3.9.1.2). Figures 40–43 showexamples with different imaging methods.Only with very severe stenosis is resting bloodflow reduced and the defect visible at rest. Inmoderate epicardial stenosis (50–85 percent),myocardial blood flow is maintained at restbecause the post-stenotic arterioles dilate.During exercise or pharmacological stress per-fusion defects can be displayed because theability to increase blood flow is reduced inregions supplied by coronary arteries with sig-nificant stenosis.

Reduction of myocardial blood flow is associ-ated with a reduction of myocardial bloodvolume. Since ultrasound contrast agents are

124 Handbook of Contrast Echocardiography

Fig. 36 Use of myocardial contrast echo in themanagement of acute transmural myocardial infarc-tion.

Fig. 37 Extended inferior infarction due to no reflowfollowing reopening of the right coronary artery, twochamber view, harmonic power Doppler. Dotted linerepresents endocardial border. Note the absence ofsignals in the papillary muscle and entire inferiorwall, compared to the signals in the anterior wall.There are only a few small epicardial vesselsdisplayed on the inferior wall (arrows).

Kardio-Kap 3 for pdf 06.06.2000 18:26 Uhr Seite 124

Page 142: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

125Perfusion Imaging

pure intravascular tracers, the myocardialcontrast signals have been used as an estimateof myocardial blood volume. The prerequisitefor using contrast signals as an estimate of localblood volume is that the concentration ofmicrobubbles in the blood is not different inthe areas which are compared. This is true witha non-destructive imaging mode such as PPI.With a destructive bubble imaging techniquelike HPD myocardial contrast strength repre-sents myocardial blood volume only if longtrigger intervals are used (>1:5, see Chapter 4).

With shorter trigger intervals the speed ofreplenishment of the disrupted microbubblesheavily determines myocardial contrast signalstrength. During stress flow rate is increased in

Fig. 38 Apicoseptal scar following myocardial infarc-tion (harmonic power Doppler, four chamber view,Levovist infusion). Decreased signals in a thin suben-docardial layer (arrow) compared to subepicardiallayers.

Fig. 39 Myocardial perfusion imaging at rest in apatient with remote inferior myocardial infarction.Left panel shows venous myocardial contrast echo-cardiography, harmonic B-mode, Levovist infusion,pulsing rate 1:7, background subtraction and colourcoding. In the two chamber view a contrast defect ofthe inferior wall is visible. Corresponding verticallong axis on MIBI SPECT (b) shows missing traceruptake in the same region.Courtesy of Christian Firschke, Deutsches Herz-zentrum, Technische Universität München, Germany

Fig. 40 Schematic changes in fixed and reversibleperfusion defects in a two chamber view. Relativereduction of signal intensities is seen in the apicalmyocardium compared to the neighbouring segmentsat rest and during stress in a fixed defect, whereasthe reversible defect is displayed only during stress.Note the increase in myocardial opacification ob-served during stress in normal myocardium.

baselinerest

normal

reversible defect

fixed defect

contrastrest

contraststress

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 125

Page 143: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

regions supplied by normal epicardial coronaryarteries and disrupted microbubbles are re-plenished during shorter time intervals than atrest. This is not possible in areas supplied bycoronary arteries with significant stenosis.

Thus, at short trigger intervals myocardialcontrast reflects flow, and is reduced in ische-mic segments compared to areas supplied bynormal coronary arteries.

126 Handbook of Contrast Echocardiography

Fig. 41 Normal response to adenosine stress (harmonic power Doppler, four chamber view). Note increasein myocardial signal intensities in the adenosine recording (b) over the baseline image (a).

Fig. 42 Myocardial perfusion imaging at rest andwith stress in a patient with a 58 percent diameterstenosis of mid LAD (segment 8) considered ofuncertain functional relevance. Small reversibletracer uptake defect of the apical myocardium(arrow) on TC99m Sestamibi SPECT: horizontal longaxis at rest (a) and with exercise (b). Correspondinghomogeneous myocardial contrast during intraveousinfusion of Levovist in the four chamber view at rest(c) and small contrast defect (arrow) during infusionof adenosine (d).Courtesy of Christian Firschke, Deutsches Herz-zentrum, Technische Universität München, Germany

a) b)

c) d)

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 126

Page 144: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

127Perfusion Imaging

Fig. 43 Visual judgement of rest and adenosine stress images (harmonic power Doppler, four chamber view,trigger 1:3) in a patient with exertional chest pain using the report form presented in Fig. 29. Reduction ofapical contrast signal intensity and no increase in septal signal strength compared to marked enhancementin the lateral wall. Corresponding findings in a Sestamibi SPECT, coronary angiography revealed 90 percentproximal LAD stenosis.

3.11 Pitfalls andtroubleshooting

3.11.1 Inadequate myocardial contrast

Weak or absent myocardial contrast does notalways mean impaired myocardial perfusion,

but may be caused by technical problems. Forreliable evaluation of perfusion, a minimumquantity of microbubbles need to be present inthe myocardial vessels. However, completeabsence of myocardial contrast despite cavityenhancement is often caused by inappropriatemachine settings. Inadequate contrast doseshould be considered after scanplane and

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 127

Page 145: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

128 Handbook of Contrast Echocardiography

instrument controls have been re-checked (seeTable). Strong myocardial and cavity contrastsignals in the nearfield but inadequate signalsat greater depth are often caused by attenuationby the agent, which can be obviated by reducingthe rate of agent infusion. In some instances,however, focus position in the nearfield canresult in a similar contrast image.

Weak or absent myocardial contrast after aninitially adequate display may also be due todisplacement of the scanhead. Triggered imag-ing is a challenge for both the investigator andthe patient. Dyspnea during vasodilator stressand atypical chest pain cause deep breaths andloss of the imaging plane. In this situation it isnecessary to leave the triggered mode and touse continuous scanning to readjust the scan-plane (with harmonic power Doppler, har-monic B-mode should be used for optimisingthe scanplane). Beware that at high MI, thisprocess depletes the systemic level of the agentduring the infusion, to the extent that onemust wait for about two minutes for the con-trast level to recover before starting to scanagain. After readjustment of the scanplane,triggered imaging can be initiated to completethe protocol, discarding the first frame.

Appparent defects in the lateral wall often donot represent a real perfusion deficit, but in-adequate myocardial contrast. Similarly, false-positive defects may be found in the anteriorwall. Extracardiac attenuation by ribs and pul-

monary tissue may be the reasons for loweramounts of acoustic power being delivered tothe lateral wall in the four chamber view and theanterior wall in the two chamber view.Sometimes this is so severe that completeshadowing of the deeper-lying structuresoccurs. Shadowing is recognised by the abruptweakening of echoes distal to a point in thescan line. This shadowing line is often foundlaterally of the endocardial border and so issometimes only identifiable with the contrast

Weak/absent contrast in the entire myocardium

1. Check scanplane in harmonic B-mode

2. Check machine settings

3. Check venous line, stopcock

4. Check infusion pump: running, correct setting

If all are set correctly, increase infusion speed

Weak/absent myocardial contrast at greater depth

but strong signals in the nearfield

1. Check whether focus position is correct

2. Check for contrast shadowing, reduce infusion

speed if necessary

Weak/absent myocardial contrast after initial

adequate display

Displacement of the scanhead

Readjust during continuous harmonic B-mode

Check for systemic depletion of bubbles following

real-time imaging

Weak/absent myocardial contrast in the lateral

sectors of the imaging field

1. Rule out extra-cardiac shadowing (ribs,

pulmonary tissue)

2. Check myocardial grey level in harmonic B-mode

3. Adjust scanplane to show lateral structures in

centre of sector

4. Exclude attenuation by cavity contrast (reduced

intensity of cavity signals close to the lateral wall)

Troubleshooting for inadequatemyocardial contrast

• Inadequate myocardial contrast

• Contrast shadowing

• Blooming

• Wall motion artifacts

Pitfalls in myocardial contrast echocardiography

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 128

Page 146: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

129Perfusion Imaging

agent. If shadowing is detected, the scanplaneshould be modified using continuous har-monic B-mode so that the lateral wall iscentred in the imaging field. The aim is for thelateral and anterior walls to be displayed withgrey myocardium. However, difficulty with thelateral wall is common even when thesemeasures have been taken. An apparent isolateddefect in the lateral wall should thereforealways be treated with caution.

3.11.2 Contrast shadowing

Weak or absent myocardial contrast at greaterdepth when there are strong signals in the near-field is probably caused by cavity bubblesshadowing distal regions (Figures 44, 45).Shadowing by the contrast agent is a particularproblem for myocardial contrast echo, becausehigher concentrations of bubbles are needed.Those present in the nearfield attenuate thesound to the point that bubble disruption doesnot occur at greater depth. Thus the basalseptum (four chamber view), basal inferior wall(two chamber view) and particularly basal

lateral wall (four chamber view) can have‘pseudo’ defects in healthy subjects. Theproblem is more serious for parasternal thanfor apical views, because the ultrasound mustpenetrate both the RV and LV cavities to reachthe inferior or posterior LV wall. The RV mayattenuate more, because it contains largerbubbles wich have not been filtered by thelungs. So far there is no way to compensate forattenuation, but there are some guidelines forits reduction.

Most important is to use the lowest effectivedose of contrast. Particularly for quantitativeanalysis, it is an advantage not to have a brightmyocardium with large quantities of micro-bubbles. When the patient is scheduled for astress study a dramatic increase in signal inten-sity must be anticipated. The protocols pro-vided in the tables on pages 103, 104 and 107,108 aim for the highest sensitivity with thelowest doses of contrast. Since the optimal dosefor the individual cannot be predicted, itshould be determined empirically, and adjust-ment of a contrast infusion is the best way todo this. Note that changes in infusion rate takeup to two minutes to become effective, so

Fig. 44 Contrast shadowing due overdosing of contrast (a). With no power Doppler signals at greater depth,better penetration after reduction of infusion rate (b).

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 129

Page 147: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

130 Handbook of Contrast Echocardiography

Fig. 45 False defect due to attenuation in the basal posterolateral wall using harmonic power Doppler mode(long axis view). (a) Defect using standard scanplane. After changing transducer position with posterolateralwall located closer to the centre of the sector, the false defect disappeared (b).

Fig. 46 Perfusion imaging artifacts: 63-year-old man with recent anterior MI and mid LAD occlusion, doubleframe triggered harmonic power Doppler, four chamber view. Notice the different scanplane with a larger LVcavity in frame two compared to frame one. Although consecutive frames are acquired, there has beenconsiderable cardiac movement in between due to the low frame rate of harmonic power Doppler. The basalseptal defect is caused by attenuation, the infarcted area is hardly visualised because of blooming or wallmotion artifacts.Courtesy of Jim Thomas, Cleveland Clinic, OH, USA

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 130

Page 148: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

131Perfusion Imaging

should be avoided during data acquisition. Thecontrast dose is better adjusted before startingthe acquisition protocol. However, the dra-matic increase in signal intensity followingvasodilator administration may cause attenua-tion problems despite optimal recordings atrest. In this situation it is better to completethe protocol and repeat the acquisition of thevasodilator images at a lower infusion rate. Thecomparison of rest and stress studies performedwith different doses of contrast is permissible,when relative measurements are used instead ofabsolute intensity measurements.

3.11.3 Blooming

Blooming describes the appearance of contrastsignals which originate from the myocardialtissue but spread into neighbouring compart-ments. This phenomenon occurs when cavitysignals are strong and is found both in har-monic power Doppler and greyscale tech-niques (harmonic B-mode and pulse inver-sion). There are two types of blooming: withcavity blooming, the cavity signals exceed theendocardial borders to a certain extent andresemble myocardial opacification (Figure 46).As intense opacification of the cavities of theheart is always detectable with intravenousadministration of contrast agent, bloomingmust be expected. Sceptics of myocardialcontrast echocardiography maintain that themyocardial signals are largely due to bloomingafter intravenous administration of contrastagent. However, this speculation is easily dis-missed. Following an IV bolus, there is a cleardelay between the appearance of signal in theleft ventricle and its appearance in the myo-cardium. Furthermore, the more macular andreticular distribution pattern – correspondingto the intramyocardial vessels – at low trigger

rates cannot be accounted for by bloomingartifacts. With blooming alone, a homo-geneous contrast effect would be expected inthe subendocardial layers. While bloomingdoes occur, it can be distinguished from real per-fusion.

Cavity blooming is caused by strong echoeswhich lie above or below the scan plane. Strongechoes can particularly be formed in theseptum from blooming of the right-ventricularcavity. Such septal blooming then occurssimultaneously with the maximum contrasteffect in the right ventricle, clearly before thetime at which myocardial perfusion is displayed.This phenomenon can be seen during bolusinjections. With a proper rate of contrast agentinfusion, cavity blooming is less common. Theimmediate subendocardial area of 1 mm shouldbe omitted if a region of interest is sited forquantitative analysis in the myocardium.

Blooming of the intramyocardial vessels alsooccurs (48). The vessels displayed in the ultra-sound image appear much wider than in reality(Figure 47). This is mainly due to the physicaland technical limitations of ultrasound equip-ment, whose resolution is currently 1–2 mm.Most intramyocardial vessels have a consider-ably smaller diameter so that an exact repre-sentation is not possible. Blooming of intra-myocardial vessels may be of benefit for visualassessment because the very small vessels areappreciated more easily. They may, however,bias quantitative assessment if the region-of-interest is too small and contains such a vessel.

3.11.4 Wall motion artifacts

Motion artifacts are most important in har-monic power Doppler. Wall motion artifacts

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 131

Page 149: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

132 Handbook of Contrast Echocardiography

have already been mentioned in §3.6.1.1 as apitfall during contrast enhanced power Dopplerfor LV opacification, where they usually do nothave a great impact on endocardial borderdelineation. During assessment of myocardialperfusion, wall motion artifacts can distort thedisplay of myocardial contrast. They canmimic tissue perfusion signals if they are noteliminated by the pre-contrast settings.However, even after successful suppression ofwall motion artifacts at baseline some may beencountered during contrast enhancement.These may be due to unexpected respiratory orthoracic movement and occur with the contrastsignals from intramyocardial vessels. Suchartifacts are often confined to a few frames,which should be excluded from furtherprocessing.

Using short trigger intervals only the largervessels are filled with contrast. Their character-istic pattern is easy to distinguish from wallmotion artifact, which often is not confined tomyocardial tissue (Figure 48). At longer triggerintervals, perfusion contrast becomes morehomogeneous and recognition of wall motionartifacts more difficult. They can be recognisedas flash-like coloured areas (changing fromframe to frame) which often overlay epi- orpericardial layers of the lateral wall (four cham-ber view), but may be found in other regionssuch as near the mitral valve and the septum.Structures with high backscatter and fastmovement parallel to the beam are prone toproduce motion artifacts, which are oftentherefore found in fibrotic pericardial layersand perivalvular tissue.

Fig. 47 Vessel blooming in harmonic power Doppler in a patient with previous apicoseptal infarction (fourchamber view). (a) shows recording at a trigger rate of 1:1 with display of blooming intramyocardial vessels –particularly in the lateral wall. (b) A homogenous myocardial contrast pattern is found in the lateral wall using1:4 trigger rate due to additional filling of small intramyocardial vessels. Notice the reduced contrast in theseptum compared to the lateral wall.

a) b)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 132

Page 150: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

133Perfusion Imaging

Fig. 48 Apical wall motion artifacts to respiratory motion (a), complete disappearance during shallowrespiration (b).

In single contrast enhanced frames it may bedifficult to distinguish between wall motionartifact and perfusion signals. Here, reviewingconsecutive frames acquired with the sametrigger intervals is very helpful: any abruptchange in signal pattern should make onethink of a wall motion artifact. It is alwaysnecessary to review a series of frames (for exam-ple, five) taken at the same machine settingduring infusion of the agent.

3.11.5 The bubble depletion artifact

The intention of an infusion is to maintain asteady level of enhancement in the myo-cardium. However, the imaging frame used todestroy the agent in the myocardium also dest-roys agent in the cavities. As more destructionframes are used at shorter trigger intervals, thesystemic pool of the agent begins to becomedepleted and its concentration drop. Shortertrigger intervals have the ability to reduce thesystemic level of contrast to the extent that themyocardial signal is lowered artifactually. Insome cases, this can result in an incorrect

reperfusion curve and a misinterpretation ofthe images. Inadvertant bubble depletion canalso occur after a period of high MI continuousimaging (for example to adjust the scanplaneduring a triggered study). Power pulse inver-sion, which uses an MI so low that the agent isnot destroyed as it is imaged, is not susceptibleto this artifact.

Coronary Flow Reserve✳

3.12 Coronary flow reserve and myocardial contrast echo

Measurement of coronary flow reserve is a pro-cedure to assess myocardial ischemia in a quan-titative way. Since coronary flow and tissue per-fusion are linked, changes in myocardial tissueperfusion following a vasodilator stimulusshould provide complemetary information.However, myocardial contrast echo has thesame limitations as SPECT in detectingbalanced hypoperfusion in, for example, three-

a) b)

✳ This section courtesy of Carlo Caiati, University of Cagliari, Italy

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 133

Page 151: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

134 Handbook of Contrast Echocardiography

vessel disease, small vessel disease, leftventricular hypertrophy and assessment ofgradual changes in perfusion. Here, we discussa technique which has been validated in a seriesof clinical studies. It provides reliable measure-ments of coronary flow reserve in the LADusing pulsed wave Doppler echocardiography.While corresponding quantitative measure-ments are being developed for myocardialcontrast echocardiography, assessment of CFRhas an important role in the diagnosis of myo-cardial ischemia.

3.12.1 What is coronary flow reserve?

Absolute coronary flow reserve represents themaximal possible increase of flow that canoccur in a single coronary vessel (49).Coronary flow reserve is usually expressed asthe ratio of maximal to basal flow. Since flow isdirectly related to the driving pressure andinversely related to resistance (Poiseuille’s law),maximal flow can be elicited by drugs such asadenosine and dipyridamole which lowerresistance by vasodilating the coronary tree.

These drugs act in two ways. First, a directendothelium-independent mechanism, vaso-dilates the coronary microvessels of less than170 µm in diameter, where 70 percent of totalcoronary resistance occurs. Second, a flow-mediated endothelium dependent mechanismvasodilates the coronary arterioles proximal to170 µm in diameter, where 25 percent of totalcoronary resistance occurs (50).

3.13 Available methods – need for contrast enhancement

In the clinical setting CFR is usually measuredusing one of two approaches: either invasiveintracoronary Doppler (51, 52) by catheter or transducer-tipped angioplasty guide wire, or non-invasively by quantitative positronemission tomography (PET) (51). The formermeasures velocity while the latter measuresabsolute myocardial flow. Both methods areprecise, but limited in their clinical applicability.The intracoronary Doppler flow wire is perfor-mable only with catheterisation, implies a cer-tain risk for the patient and is expensive. Forthese reasons it is not easily performable andespecially not repeatable. The major limita-tions of PET scanning are its scarcity, its costand radiation exposure to the patient. Becauseof the clinical importance of CFR there is aneed for a simple, non-invasive, repeatable andinexpensive tool capable of this functionalevaluation.

Transesophageal Doppler echocardiography(TEE) was proposed some years ago as a semi-invasive tool for measuring CFR by blood flowvelocity in the proximal LAD at baseline andafter maximally lowered coronary resistancewith dipyridamole (52). There is no doubt that

Definition: Ratio of maximal to basal flow

Normal values: More than 2.0

Causes of reduction: 1. Coronary artery stenosis

> 50 percent lumen

reduction

2. No coronary stenosis, eg

Syndrome X

Hypertensive heart disease

Hypercholesteremia

Diabetes

Coronary flow reserve

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 134

Page 152: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

135Perfusion Imaging

TEE evaluation of CFR offers considerablediagnostic potential. However, broader appli-cability of this new technique has been ham-pered by three major limitations: 1) inability toobtain an adequate Doppler signal in allpatients; 2) only pre-stenotic CFR is measur-able (sampling is obtained few millimetresbehind the main left coronary artery bifurca-tion); 3) the semi-invasiveness of esophagealintubation, which makes the approach quiteunpopular with patients.

The first of the three above mentioned limita-tions has been overcome to a great extent bythe utilisation of echocontrast agent capable ofincreasing Doppler signal intensity and signalto noise ratio not only in cardiac cavities, butalso in coronary arteries after intravenous in-jection (53, 54). The second has been solved bythe combined use of second harmonic imagingand contrast agents (55), which together makecoronary flow velocity and flow reserve feasibleby the transthoracic approach (56–60).Diagnostic signals are assessable in only 55percent of patients before contrast and inalmost all patients after contrast enhancement.The transthoracic measurement of CFR in theLAD concurs very closely with Doppler flowwire CFR measurements. The reproducibilityof CFR measurements is high. The 95 percentconfidence interval for variation between thetwo measurements is -0.32 to +0.32. Thus,contrast enhanced transthoracic Doppler pro-vides feasible, reliable and reproduciblemethod for assessment of CFR in the LAD.

3.14 Coronary flow reserve: indications and selection of methods

Since CFR measurements are confined to midand distal LAD, isolated use of CFR toevaluate myocardial ischemia is not warrantedin patients with an unknown coronary angio-gram. CFR may be used in conjunction withmyocardial contrast echo. Both techniquesneed the same kind of preparations and use thesame stress protocols (infusions of contrast anda vasodilator). The additional time for sampl-ing the LAD flow is a minor issue compared tothe more time consuming recording of tissueperfusion. However, if an LAD lesion is pre-sent, diagnostic confidence is improved whentwo different approaches are used.

CFR measurements have special relevance aftercoronary angiography. In the case of an LADstenosis it can be used for integrating anatomicinformation by catheter or for serial examina-tion of patients undergoing angioplasty of theLAD (60). For this indication isolated use ofCFR might be considered to save contrastagent, as it requires somewhat less agent thanperfusion measurements. One of the mainadvantages of this method over the traditionalnon-invasive stress tests such as stress echo orstress myocardial scintigraphy in following-upan LAD angioplasty is its quantitative informa-tion. As a consequence, even sub-clinical,subtle modification of the functional status ofthe dilated LAD might be unveiled.

Normal coronary arteries in patients with chestpain and/or positive stress test are anothersituation where CFR should be assessed. Theclinical syndrome of ‘coronary insufficiencywith normal coronary arteriogram’ is found in

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 135

Page 153: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

136 Handbook of Contrast Echocardiography

approximately 10–20 percent of patients withexercise-induced coronary insufficiency. Anabnormal CFR finding confirms the cardiacorigin of the chest pain and rules out falsepositive stress testing. CFR can be impaired bya variety of disorders of the microcirculation.Pharmacologic interventions of these disorderscan be easily assessed by CFR measurement.Diabetes, arterial hypertension or hyperchole-steremia, for example, are associated withimpaired CFR (61–63).

3.14.1 Selection of patients

A suboptimal ultrasound window, a majorconcern in any ultrasound approach, does not

reduce the feasibility of CFR measurements ifthe combination of second harmonic imagingand a contrast agent is used. In our experiencethe feasibility of almost 100 percent has beenconfirmed in large series of patients includingthose with large body habitus (59, 62).

Contrast enhancement should always be usedeven when good Doppler quality is achieved atbaseline without enhancement. Contrast helpsmaintain good monitoring of the Dopplersignal from the LAD during hyperemia, wherevasodilators induced hyperpnea and tachycar-dia make an adequate blood flow Dopplerrecording more difficult. In cases with goodsignal at baseline a very small amount ofcontrast can be used just for the hyperemicportion of the study.

Clinical methods to assess CFR

Positron emission Non-invasive, global and Scarce

tomography (PET) regional myocardial perfusion Radiation exposure

reserve objectively assessable Expensive

Intracoronary Precise post-stenotic CFR

Doppler flow wire assessement

Evaluation of relative and Invasive and risky

fractional coronary flow reserve Expensive

Difficult to repeat

Transesophageal Widely available Semi-invasive

Doppler Limited prediction of

echocardiography (TEE) coronary stenosis

Limited to LAD

Transthoracic Non invasive, widely available Limited to LAD

contrast Doppler Optimal prediction of coronary

echocardiography stenosis

Technically feasible with harmonic

contrast imaging

Pros ConsMethodology

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 136

Page 154: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

137Perfusion Imaging

3.15 How to perform a CFR study

3.15.1 Intravenous lines

Two intravenous lines are needed for studiesusing adenosine as vasodilator. An indwellingcannula (20 gauge or less) is inserted into acubital vein for the infusion of the contrastagent. Another line is necessary for infusion ofadenosine, this line can have a smaller size (22gauge). Interactions between adenosine andcontrast agents are unlikely, but have not beensystematically evaluated so far. Thus infusingadenosine and contrast agent via the same linecannot be recommended. If dipyridamole isused for induction of hyperemia, only onevenous line is needed, because the vasodilatoreffect persists after dipyridamole has beeninfused and simultaneous infusion with theecho contrast agent is not necessary.

3.15.2 Contrast agent

All available contrast agents can be used toenhance Doppler signals. Infusion of thecontrast agent is necessary to maintain thesame level of contrast enhancement duringbaseline and hyperemia. This method ofadministration has the advantage of maintain-ing enhancement over several minutes, withoutaffecting the peak intensity that is attained. In comparison to myocardial contrast echo,doses of echo contrast are much lower.Anexample of a dose regimen for Levovist is:one vial prepared using a concentration of300 mg/ml. The agent is administered byinfusion using an infusion pump connectedover a special 50-cm connector tubing. Theinitial infusion rate is 1 ml/min. This rate canbe increased to a maximum of 2 ml/min ordecreased to a minimum of 0.5 ml/min,according to the quality of the Doppler signalobtained.

3.15.3 Protocols to induce hyperemia

The dipyridamole and adenosine protocolswhich are established for regular stress echo andSPECT are useful for assessment of CFR too(see Figures 49, 50). For dipyridamole, only onevenous line is necessary but more contrastagent is needed: because of the delay betweenthe start of the drug and achieving completevasodilatatory effect (6–8 minutes for the low

Indications for transthoracicDoppler of LAD flow

• Before coronary angiography

Suspected LAD stenosis (previous anterior

infarction, stress ECG, stress echo, myocardial

scintigraphy)

Myocardial contrast echo

• After coronary angiography – LAD stenosis

Significance of stenosis

(in combination with myocardial contrast)

• After coronary angiography – normal coronary

arteries

To confirm cardiac origin of chest pain

To control therapeutic interventions

CFR in the ‘difficult’ echo patient

CFR measurements can be performed in patients

with poor windows where myocardial contrast

echocardiography is not recommended

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 137

Page 155: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

138 Handbook of Contrast Echocardiography

dose and 12–14 min for the high dose), twoseparate contrast infusions need to be given.This also implies that LAD flow needs to be re-imaged during dipyridamole as the spot isusually lost at the end of the contrast enhance-ment for the baseline acquisition.

Intravenous adenosine (140 µg/kg/min over 6 minutes) seems superior to dipyridamole.First, adenosine acts rapidly, achieving peakeffect in about 55 seconds compared to about100 seconds with dipyridamole (64). Thisshortens the hyperemic part of the study to2–3 minutes so that a single contrast infusionof 11 ml volume is sufficient for all of theexamination. Second, because the enhance-ment of the Doppler signal is obtained withoutinterruption, the Doppler recording is moreeasily and accurately performed. Third, unlikedipyridamole, adenosine in doses >100 µg/kghas been shown to be nearly equivalent topapaverine to produce maximal coronary vaso-

dilation. Fourth, the short duration of theexamination combined with the promptreversibility of the side effects, if any, aftertermination of the infusion make this drug safeand acceptable to the patient. The only short-comings of adenosine are 1) the need of twointravenous lines, 2) the higher cost, and 3) thefrequently induced hyperpnea. Hyperpnea,however, may be disturbing for the patient, butrarely causes degradation of image quality.

Steps to perform a CFR study

1. Insert an indwelling cannula (20 gauge) into a

cubital vein. Connect the line for contrast infusion

2. Insert an indwelling cannula (22 gauge or more)

into a forearm vein (only for adenosine). Connect

the line with adenosine infusion

3. Prepare contrast agent and adenosine infusion

4. Search for LAD flow using fundamental colour

Doppler

5. Start contrast infusion

6. Search for LAD flow using harmonic colour Doppler

7. Switch to fundamental (harmonic) PW Doppler,

record LAD flow at rest

8. Start adenosine infusion

9. Record LAD flow during hyperemia

Fig. 49 Protocol for dipyridamole stress,DIP1 = 0.56 mg/kg over 4 minutes, DIP2 = 0.28 mg/kg over 2 minutes

Fig. 50 Adenosine protocol for CFR studies

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 138

Page 156: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

3.15.4 Image orientation for visualising blood flow in the LAD

A systematic attempt must be made to recordflow in the distal or in the mid part of the LAD(Figure 51). The approach for the distal partconsists first in obtaining a short axis of the leftventricular apex and of the anterior groove.Thereafter, a search for coronary flow in theanterior groove is started. When a diastolicflow, circularly-shaped vessel is recognised inthe anterior groove area, it is brought into thecentral part of the ultrasound field by anglinglaterally and slightly above the central ray ofthe scan plane. At this point, the transducershould be rotated counterclockwise to obtainthe best long axis colour view. Alternatively, inorder to achieve a better alignment betweenflow and the ultrasound beam, a modified

foreshortened two chamber view can be ob-tained by sliding the transducer superiorly andmedially from an apical two chamber position.In this way visualisation of the epicardial partof the anterior wall is obtained with less pul-monary interference. The mid part of the LADis visualised by a low parasternal short axis viewof the base of the heart modified by a slightclockwise rotation of the beam in order tobring into view the anterior groove area andthereby the mid portion of the LAD that runsover the left border of the right ventricular out-flow tract (post-pulmonary tract). Pulsed waveflow Doppler recording in the LAD is firstattempted at baseline in fundamental mode,and after contrast in harmonic mode.

3.15.5 Instrument settings

All the setting adjustments described here havebeen established using the Acuson Sequoiasystem. First, regular tissue harmonic modewith the default settings is used to obtain ashort axis of the left ventricular apex and of theanterior groove. Then fundamental colourDoppler is used to search the LAD. As soon asLAD flow has been identified, the contrastinfusion is started and harmonic colourDoppler mode initialised. The harmoniccolour Doppler setting should be adjusted to maximise scanning sensitivity withoutreducing the frame rate. The pulse repetitionfrequency is decreased by changing the colour Doppler velocity scale (between 11 and25 cm/s), gate size should be maximallyincreased, gain increased until noise begins toappear faintly and colour Doppler box sizekept small. Colour transmit power is turned upto the maximum level. Increasing ultrasoundenergy to maximum level (destruction mode) isimportant during contrast administration in

139Perfusion Imaging

Fig. 51 Plane orientation for acquisition of LADflow: the mid part of the LAD is visualised by a lowparasternal short axis view of the base of the heartmodified by a slight clockwise rotation of the beam inorder to bring into view the anterior groove area.

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 139

Page 157: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

140 Handbook of Contrast Echocardiography

order to maximise harmonic production by themicrobubbles and so for optimising coronaryflow recording with harmonic colour Doppler.

With fundamental spectral Doppler, gain andtransmit power are slightly reduced (-4 and -7 dB respectively) in order to avoid excessive

spectral blooming during contrast administra-tion. During contrast administration thissetting does not generally need to be modifiedexcept on some occasions where, duringspectral recording transmit power needs to belowered a little in order to reduce disturbingvertical spikes caused by ultrasound breaking

Transducer NormalScanning modality Tissue harmonic Transmit power Maximum (0 dB), as optimised for tissue harmonicsGeneral gain Default (0 dB)TGC Keep low; (high B-mode gain hampers colour flow display)Depth Reduce depth (to increase frame rate)

Transducer NormalScanning modality Fundamental velocity colour Doppler (2.5 MHz)Transmit Power Maximum (0 dB)Sample volume size Maximum (3) (to increase sensitivity)Gain Set high, at the level at which noise begins to appear faintly Colour Doppler box size Keep as small as possible (to increase frame rate)Pulse repetition frequency Decrease to improve sensitivity (velocity range 14-18 cm/s) (PRF)

Transducer NormalScanning modality Harmonic colour Doppler (1.7 MHz transmit/3.5 MHz receive)Transmit Power Maximum (0 dB) (destruction mode)Gain Unchanged or slightly reducedColour Doppler box size UnchangedPulse repetition frequency Unchanged or a little increased (to reduce flash artifacts(PRF) caused by the increased heart rate)

B-mode and colour Doppler for CFR: settings

B-Mode:

Colour Doppler before contrast:

Colour Doppler after contrast:

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 140

Page 158: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

141Perfusion Imaging

bubbles, the so called ‘bubble noise’ phenom-enon (see §3.17.4). To measure coronary flow,colour flow imaging is used as a guide. Samplevolume is positioned at the location of thevessel. Because of the absence of ventricularcontraction in diastole, the position of the leftanterior descending coronary artery is morestable in this part of the cardiac cycle, thusfacilitating sample volume positioning.

3.15.6 Combination with myocardial contrast echo (MCE)

CFR measurements may be made in conjunc-tion with myocardial contrast echocardio-graphy. For this purpose the MCE study has tobe performed with an adenosine infusion. After

the MCE protocol has been completed, theadenosine infusion and contrast infusion aremaintained and LAD flow is recorded. Usuallytransmit power needs to be reduced to avoidblooming and saturation of the LAD flowsignals. Afterwards adenosine infusion isstopped but the contrast infusion continuedand 20 seconds later the baseline LAD flow isrecorded. This procedure can be performedwithin one minute, if the right plane for inter-rogating the LAD is quickly found. With somepractice the sonographer will be able to easilyfind the LAD, because the colour and PWsignals are very strong during contrastenhancement. The time to search for the LADcan further be reduced if the sonographer hasidentified the imaging plane during the base-line study.

Transducer 2–3 MHz sectorScanning modality Fundamental (2.5 MHz)Sample volume size 4 mmTransmit power From –4 to –7 dBGain From –4 to –7 dBDynamic range Default (50 dB)

Transducer 2–3 MHz sectorScanning modality Fundamental (2.5 MHz)Sample volume size 4 mmTransmit Power Unchanged or slightly reduced (to reduce “bubble

noise”)Gain Unchanged or a slightly reduced (to reduce blooming)Dynamic range Default (50 dB)

Spectral Doppler for CFR: settings

Spectral Doppler before contrast:

Spectral Doppler after contrast:

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 141

Page 159: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

142 Handbook of Contrast Echocardiography

3.16 Image acquisition and interpretation

Once adequate Doppler spectra are visible,recordings are started. It is useful to store ashort period in colour mode with the cursorand sample volume in the LAD, to establishthe Doppler angle, which must remain con-stant. For PW Doppler, at least five framesshould be stored on videotape and/or cineloop.Recording of hyperemic flow velocity by PWDoppler is started as soon as the colour signalshows an increase in velocity, or in any case,

within two minutes of the beginning of theadenosine infusion. It is continued until the 5th

minute. If the Doppler angle or LAD segmentappears different during the infusion to thatseen at baseline, a new baseline velocity record-ing must be made. This can be done fiveminutes after the end of the infusion. It is bestto maintain monitoring throughout the wash-out phase.

The recordings are analysed offline using thetools available in every Doppler system (Figure52). Velocity-time integrals (VTI) are obtainedby tracing the envelope of the Doppler spectra.

Fig. 52 Contrast enhanced pulsed wave Doppler tracings of LAD flow at rest (top) and during adenosineinfusion (bottom). CFR is assessed by tracing the envelope of the Doppler spectra and calculating the velocity-time integrals. The ratio of adenosine to baseline data represents CFR.Courtesy of Heinz Lambertz, Deutsche Klinik für Diagnostik Wiesbaden, Germany

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 142

Page 160: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

143Perfusion Imaging

An average of at least 3 cardiac cycles is made.The ratio of VTI during hyperemia to the VTIat baseline is the coronary flow reserve. If aCFR > 2.0 is considered normal and CFR < 2.0 abnormal, the sensitivity and specificityfor detecting a >75 percent diameter LADstenosis is 91 and 76 percent respectively.

3.16.1 Significance of an LADstenosis

Coronary angiography provides little insightinto the physiological significance of coronarystenoses, especially those of intermediateseverity (65). Understanding functionalimpact of a stenosis is important for clinicaldecision making, as PTCA can be deferredsafely in patients with intermediate stenosisbut an adequate CFR value (66) as shown onFigure 53.

Coronary angiography (Figure 53 a) showed anapparently significant stenosis in the mid-LAD. In order to confirm its functional sig-nificance, a TTE CFR was performed (Figure53 b) which showed a normal CFR. This was

in agreement with adenosine/baseline 99 mTc-Sestamibi SPECT that did not show anyreversible perfusion defect.

3.16.2 Follow-up of an LAD stenosis after PTCA

After PTCA, CFR recovers dramatically,reaching the lower limit of the normal range.The baseline diastolic/systolic ratio also returnsto normal. The normalisation of diastolic/systolic ratio after PTCA could be a morespecific index for successful recanalisation thanCFR since CFR can still be depressed immedi-ately after intervention (so-called ‘micro-circulation stunning’) (67). Figure 54 is a typi-cal example. Before PTCA the CFR was flat; inaddition the ratio of the diastolic to systolicvelocity is 1 (which is abnormal). After PTCAthe ratio increased.

3.17 Pitfalls and trouble-shooting

3.17.1 Angle of the vessel to the beam

One of the basic assumptions of the method isthat the Doppler angle is kept constantthroughout the examination. One way to reachthis goal is to maintain the same view through-out the examination with no interruption ofimaging. In this respect the use of adenosineprotocol has eased the method enormously.However, in some cases there can still be someuncertainty. It is a good idea to check theconstancy of the Doppler angle by verifyingthat the depth of the sample volume is thesame throughout the examination. This onlymakes sense if the tomographic plane does not

Fig. 53a Coronary angiogram in a patient with pro-longed chest pain: 60 percent stenois in the mid LAD(arrow)

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 143

Page 161: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

change. If during hyperemia the Doppler angleor the LAD segment in the colour imageappears different to that visualised at baseline,a new baseline velocity recording should bemade in the wash-out phase. This manouvreworks better with adenosine because of itsrapid reversibility.

3.17.2 Displacement of the sample volume

It is important that a CFR measurement ismade in the post-stenotic territory. Post-stenoticCFR is much more accurate than pre-stenoticCFR in assessing the residual vasodilatorycapacity of the vascular bed caused by a nar-rowed coronary artery (68). In fact, coronaryflow reserve measured in regions with branchesproximal to the lesion is unreliable becauseflow is concurrently assessed for regions ofvarying vasodilatatory reserve, so that the resultreflects a weighted average of these potentiallydisparate zones.

The measurement of CFR can also be invali-dated if it is taken at the stenosis site itself.Worse still, one can measure blood flow vel-ocity in one segment of the artery at baselineand, with the effect of tachycardia and hyper-pnea, in a slightly different segment duringhyperemia. If a stenotic jet is inadvertentlyimaged only during stress, a false-normal CFRcan result. To avoid these problems, it is usefulat baseline to explore a tract of the artery as longas possible and check that the flow is homo-geneous. If abnormally high flow is detected inone portion of the artery, the sample volumeshould be placed as distally as possible from thestenosis site, making sure that in the new loca-tion flow velocity is in the normal range. Thisposition should be maintained throughout theexamination.

3.17.3 Flow in the mammary artery

Flow in mammary artery can be mistaken forcoronary flow. It can, however, easily be recogn-

144 Handbook of Contrast Echocardiography

Fig. 53b Transthoracic contrast enhanced PW Doppler: Normal increase of LAD flow during adenosine infusion

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 144

Page 162: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

145Perfusion Imaging

Fig. 54 CFR measurements before and one week after PTCA of a subocclusive proximal LAD stenosis. BeforePTCA (top panel) the spectra of LAD flow were flat with a ratio of the systolic to the diastolic velocities of one.Almost no change was found between baseline (left) and adenosine recordings (right). After PTCA (bottompanel) the typical pattern of normal coronary flow is recorded with higher velocities in diastole than in systole.

ised by its lack of cardiac motion and by the pre-valent systolic velocity component of itsDoppler signal.

3.17.4 Bubble noise

Small bubbles enhance the Doppler signalwithout affecting the echo statistics, so that thecharacter of the spectral Doppler sound doesnot change. Larger bubbles, on the other hand,

pass through the sensitive volume as single,highly reflective scatterers and create transientswhich are readily heard as ‘popping’ sounds.They give rise to the familiar bidirectional‘spike’ in the Doppler spectrum (Chapter 1).The same effect is also caused by bubbleswhich are disrupted by the ultrasound pressureas they pass through the sample volume.Agents such as Levovist, as they disintegrate inthe blood stream, appear to aggregate intolarger collections of gas, which are capable of

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 145

Page 163: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

146 Handbook of Contrast Echocardiography

producing this artifactual echo appearance inthe wash-out phase. The PW Doppler record-ings in Figure 55 show the typical spikes of anair based agent which is breaking up as a resultof its passage through the heart and lungs.Bubble noise can be avoided by reducing thetransmit power (that is the MI, not the gain!).

3.17.5 Interpretation problems: impact of preload

Increase of preload caused by aortic or mitralregurgitation, anemia, high cardiac output, etc.

reduces CFR by increasing basal flow throughan increase in oxygen consumption. On theother hand, an increased heart rate reducesCFR by increasing basal flow and at same timedecreasing maximal flow (69). Thus, anyreduction of absolute CFR must always beinterpreted in the light of the hemodynamicconditions. A general increase of basal flowvelocit, is often detectable in these circum-stances.

3.17.6 Impact of blood pressure

Blood pressure has little effect on CFR sincealterations in aortic pressure within the limitsof autoregulation produce parallel increases inbasal and maximal coronary flow, so that CFRremains the same (69). This is true if bloodpressure variations affect both baseline andhyperemia in a uniform way. However, a dropin blood pressure can occur only in the hyper-emic part of the study as a consequence of theadenosine effect on systemic arterial resistance.When this occurs, even small blood pressure

Pitfalls in CFR measurement

Technical problems

• Angle of the vessel to the beam

• Displacement of the sample volume

• Inadvertant scanning of internal mammary artery

• Bubble noise

• Impact of preload, blood pressure, vasodilation

Fig. 55 Bubble noise: the Doppler spectrum is composed of a series of spikes which represent disruption ofthe microbubbles.

Interpretation problems

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 146

Page 164: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

variations can induce large variation of maxim-al coronary blood flow since, with the removalof autoregulation by adenosine, the relationbetween blood flow and blood pressurebecomes very steep (69). CFR can therefore beaffected profoundly by adenosine-inducedblood pressure variations. Blood pressure mustalways be monitored carefully even though sig-nificant adenosine-induced blood pressurevariation, in our experience, affects less than 10percent of patients. A variation of more than30 mm Hg in blood pressure during adenosinewith respect to baseline invalidates the CFRmeasurement.

3.17.7 Vasodilation during hyperemia

One reason for errors in assessing coronaryflow reserve with this Doppler method andadenosine could be flow-mediated dilation ofthe epicardial vessels during hyperemia (70). Ifthis phenomenon should occur, CFR as thesimple ratio of two velocities would underesti-mate true coronary flow reserve, since maximalvelocity would be reduced as consequence ofvasodilatation. Current knowledge suggeststhat this effect is limited in patients witharteriosclerosis and intravenous application ofadenosine and dipyridamol, but further studiesare needed to shed more light on this issue.

3.18 Summary

For echocardiography, contrast agents presenta unique opportunity to add assessment ofmyocardial perfusion to the examination of the patient with ischemic disease. From atechnical point of view, the examination is achallenging one which requires both training

and experience. There is no question that asthe technology of both the contrast agents andthe imaging instruments improve, so theexamination will become easier to perform.The advent of real-time, low MI perfusionimaging is one of the most significant advancesin this regard. Triggered harmonic powerDoppler remains the most sensitive availablemethod, but it is hampered by very low frame-rates and the challenge of recognising andminimising motion artifact. The integration ofthese techniques with the stress echo is funda-mental to its success and initial experience isvery promising. There is every reason to expectthat real-time contrast perfusion stress echowill be a routine procedure in the near future.Quantitative assessment of coronary flowreserve from contrast enhanced measurementof velocity in apical vessels is a well-validatedstress procedure that can provide informationthat is more limited in anatomic domain buteasier to acquire than perfusion imaging. Withboth these methods, contrast ultrasound offersthe opportunity to provide quantitative per-fusion information previously unavailable in asimple, real-time procedure.

3.19 References

1. Marcus ML. Anatomy of coronary vasculature. In: Marcus ML (editor) The coronary circulation in health and disease. McGraw Hill, New York 1983: 3–21

2. Myers JH, Stirling MC, Choy M et al. Direct measurement of inner and outer wall thickening dynamics with epicardial echocardiography: Circulation 1986, 74: 164–172

3. Braunwald E, Kloner RA. Myocardial reperfusion:A double-edge sword? J Clin Invest 1985; 76:1715

4. Epstein SE, Cannon RO, Talbot TL. Hemodynamic principles in the control of coronary blood flow.Am J Cardiol 1985; 56:8E

147Perfusion Imaging

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 147

Page 165: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

148 Handbook of Contrast Echocardiography

5. Kaul S, Jayaweera AR. Coronary and myocardial blood volumes: noninvasive tools to assess the coronary microcirculation? Circulation 1997; 96: 719–724

6. Wei K, Jayaweera Ar, Firoozan S et al. Quantification of myocardial blood flow using ultrasound-induced destruction of microbubbles administered as a constant venous infusion. Circulation 1998; 97:473–483

7. Schlant RC, Blomquist CG, Brandenburg RO et al. Guidelines for exercise testing: A report of the JointAmerican College of Cardiology-American Heart Association Task Force on Assessment of cardio-vascular procedures. Circulation 1986; 74 (suppl III):653A

8. Gianrossi R, Detrano R, Mulvihill D et al. Exercise-induced ST depression in the diagnosis of coronaryartery disease. A meta analysis. Circulation1989; 80:87

9. Armstrong WF, Mueller TM, Kinney EL,Tickner EG, Dillon JC, Feigenbaum H. Assessment of myocardial perfusion abnormalities with contrast-enhanced two-dimensional echocardiography. Circulation 1982; 66(1):166–173

10. Marwick TH, Nemecc JJ, Pashkow FJ et al.Accuracy and limitations of exercise echocardiography in a routine clinical setting. J Am Coll Cardiol 1992; 19:74–81

11. Elhendy A, Geleijnse ML, Roelandt JRTC, et al. Dobutamine-induced hypoperfusion without transient wall motion abnormalities: less severe ischemia or less severe stress? J Am Coll Cardiol 1996; 27:323–9

12. Geleijnse ML, Salustri A, Marwick TH et al. Should the diagnosis of coronary artery disease be based on the evaluation of myocardial fucntion or perfusion? Eur Heart J 1997; 18 (suppl D):D68–D77

13. Kaul S. Myocardial contrast echocardiography.15 years of research and development. Circulation 1997; 96:3745–3760

14. Kaul S, Pandian NG, Okada RD et al. Contrast echocardiography in acute myocardial ischemia: I.In vivo determination of total left ventricular ‘area ar risk’. J Am Coll Cardiol 1984; 4:1272–1282

15. Ragosta M, Camarano GP, Kaul S, Powers E, Gimple LW. Microvascular integrity indicates myocellular viability in patients with recent myo-cardial infarction: new insights using myocardial contrast echocardiography. Circulation 1994; 89:2562–2569

16. Distante A, Rovai D. Stress echocardiography and myocardial contrast echocardiography in viability assessment. European Heart Journal 1997;18:714–715

17. Sabia PJ, Powers ER, Jayaweera AR, Ragosta M, Kaul S. Functional significance of collateral blood flow in patients with recent myocardial infarction: a study using myocardial contrast echocardiography. Circulation 1992; 85: 2080–2089

18. Cheirif J, Desir RM, Bolli R, Mahmarian JJ,Zoghbi WA, Verani MS, Quinones MA. Relation of perfusion defects observed with myocardial contrast echocardiography to the severity of coronary stenosis: correlation with thallium-201 single-photon emission tomography. J Am Coll Cardiol 1992; 19(6):1343–1349

19. Kaul S, Senior R, Dittrich H, Raval U, Khattar R, Lahiri A. Detection of coronary artery disease with myocardial contrast echocardiography: comparison with 99mTc-sestamibi single-photon emission computed tomography. Circulation 1997; 96(3):785–792

20. Agrawal DI, Malhotra S, Nanda NC, Truffa CC, Agrawal G, Thakur Ac, Jamil F, Taylor GW, Becher H. Harmonic Power Doppler Contrast Echocardiography: Preliminary experimental results. Echocardiography 1997; 14,6:631–635

21. Becher H, Tiemann K, Schlief R, Lüderitz B,Nanda NC. Harmonic Power Doppler Contrast Echocardiography: Preliminary clinical results. Echocardiography 1997; 14,6:637–642

22. Tiemann K, Becher H, Bimmel, Schlief R, Nanda N. Stimulated acoustic emission. Nonbackscatter contrast effect of microbubbles seen with harmonic power Doppler imaging. Echocardiography 1997;14:65–69

23. Becher H. Power Doppler. In: Rush-Presbyterian-St.Luke's Clinical Use of Contrast Echocardio-graphy. (1998) Am J Cardiol – a continuing education

24. Becher H, Tiemann K, Powers J. Power Harmonic Imaging – clinical application in contrast echo-cardiography. Medica Mundi 1999; 43(2):26–30

25. Porter TR, Li S, Kilzer K et al. Effect of significant two-vessel versus one-vessel coronary artery stenosis on myocardial contrast effects observed with inter-mittent harmonic imaging after intravenous contrast injection during dobutamine stress echocardio-graphy. J Am Coll Cardiol 1997; 30:1399–1406

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 148

Page 166: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

149Perfusion Imaging

26. Marwick TH, Brunken R, Meland N et al. Accuracy and feasability of contrast echocardiography fordetection of perfusion defects in routine practice: comparison with wall motion and technetium-99m sestamibi SPECT. J Am Coll Cardiol 1998; 38,5:1260–1269

27. Hope-Simpson D, Chin CT, Burns PN. Pulse Inversion Doppler: A new method for detecting non-linear echoes from microbubble contrast agents. IEEE Transactions UFFC 1999; 46:376–382

28. Tiemann K, Lohmeier S, Kintz S et al. Real-time contrast echo assessment of myocardial perfusion at low emission power: First experimental and clinical results using power pulse inversion imaging. Echocardiography 1999; 16:799–809

29. Porter TR, Xie F. Transient myocardial contrast after initial exposure to diagnostic ultrasound pressures with minute doses of intravenously injected micro-bubbles: demonstration and potential mechanisms. Circulation 1995; 92:2391–2395

30. Porter TR, Li S, Jiang L, Grayburn P, Deligonul U. Real-time visualization of myocardial perfusion and wall thickening in human beings with intravenous ultrasonographic contrast and accelerated inter-mittent harmonic imaging. J Am Soc Echocardiogr 1999; 12(4):266–271.

31. Wei K, Jayaweera AR, Firoozan S et al. Basis for detection of stenosis using venous administration of microbubbles during myocardial contrast echo-cardiography: bolus or continuous infusion? J Am Coll Cardiol 1998; 32:252–260

32. Albrecht T, Urbank A, Mahler M et al. Prolongation and optimization of Doppler enhancement with a microbubble US contrast agent by using continuous infusion: preliminary experience. Radiology 1998; 207:339–347

33. Miller JJ, Tiemann K, Podell S, Becher H, Doerr Stevens JK, Kuvelas T, Greener Y, Killam AAL, Goenechea J, Dittrich HC. Infusion of Optison for left ventricular opacification and myocardial contrast echocardiography. J Am Soc Echocardiogr 1999 (in press)

34. Wilson RF, Wyche K, Christensen BV et al. Effects of adenosine on human coronary arterial circulation. Circulation 1990; 82:1595–1606

35. Johnston DL, Daley JR, Hodge DO, Hopfenspirger MR, Gibbons RJ. Hemodynamic responses and adverse effects associated with adenosine and dipyridamole pharmacologic stress testing: a comparison in 2000 patients. Mayo Clin Proc 1995; 70: 331–336

36. Wei K, Jayaweera AR, Firoozan S et al. Quantification of myocardial blood flow with ultra-sound induced destruction of microbubblesadministered as a continuous infusion. Circulation 1998; 97:473–483

37. Ito H, Tomooka T, Sakai N et al. Lack of myo-cardial perfusion immediately after successful thrombolysis: A predictor of poor recovery of left ventricular function in anterior myocardial infarction. Circulation 1992; 85:1699–1705

38. Agati L, Funaro S, Veneroso G, Lamberti A, Altieri C, Bilotta F, Benedetti G, De Castro S, Autore C, Fedele F. Assessment of myocardial reperfusion after acute myocardial infarction using harmonic power Doppler and Levovist: Intravenous versus intracoronary contrast injection. Eur Heart J 1999; 20 (abstract supplement):362

39. Iliceto S, Galiuto L, Marchese A et al. Functional role of microvascular integrity in patients with infarct-related artery patency after acute myocardial infarction. Eur Heart J 1997; 18:618–624

40. Kloner RA; Ganote CE, Jennings RB.The ‘no-reflow’ phenomen after tempory coronary occlusion in the dog. J Clin Invest 1974; 54:1496–1508

41. Firschke C, Basini R, Patiga J et al. Contrast echocardiographic evaluation of myocardial salvage after reperfusion therapy in acute myocardial infarction. Eur Heart J 1999; 20 (abstract supplement):363

42. Horcher J, Blasini R, Martinoff S, Schwaiger M, Schömig A, Firschke C. Myocardial perfusion in acute coronary syndrome. Circulation1999; 99:e1

43. Firschke C, Lindner JR, Wei K, Goodman NC, Skyba DM, Kaul S. Myocardial perfusion imaging in the setting of coronary artery stenosis and acute myocardial infarction using venous injection of a second-generation echocardiographic contrast agent. Circulation 1997; 96(3):959–967.

44. Macioch JE, Sandelski J, Ostoic TA, Johnson MS,In M, Thew ST, Liebson PR, Becher H, Holoviak M,Feinstein SB. Clinically reproducible myocardial perfusion studies with contrast echo in 82 consecutivepatients:correlation to cardiac cath or nuclear imaging in 36 patients. Echocardiography 1998;15[6, part2]:98

45. Porter TR, Xie F, Kilzer K, Deligonul U. Detection of myocardial perfusion abnormalities during dobutamine and adenosine stress echocardiography with transient myocardial contrast imaging after minute quantities of intravenous perfluorocarbon-exposed sonicated dextrose albumin. J Am Soc Echocardiogr 1996; 9(6):779–786

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 149

Page 167: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

150 Handbook of Contrast Echocardiography

46. Porter TR, Li S, Kricsfeld D, Armbruster RW. Detection of myocardial perfusion in multiple echo-cardiographic windows with one intravenous injection of microbubbles using transient response second harmonic imaging. J Am Coll Cardiol 1997; 29(4):791–799.

47. Firschke C. Venous myocardial contrast echocardio-graphy can guide clinical decision making in the catheterization laboratory. Circulation 1998;

48. Lentz C, Tiemann, K, Köster J, Schlosser T, Goenechea J, Becher H. Does color blooming limitate assessment of myocardial perfusion using harmonic power Doppler? Eur Heart J 1999; 20 (abstract supplement):359

49. Gould KL, Lipscomb K, Hamilton GW. Physiologic basis for assessing critical coronary stenosis. Instantaneous flow response and regional distributionduring coronary hyperemia as measures of coronary flow reserve. Am.J.Cardiol. 1974; 33:87–94.

50. Doucette JW, Corl PD, Payne HM, Flynn AE,Goto M, Nassi M, Segal J. Validation of a Doppler guide wire for intravascular measurement of coronary artery flow velocity. Circulation 1992; 85:1899–1911.

51. Demer LL, Gould KL, Goldstein RA, Kirkeeide RL, Mullani NA, Smalling RW, Nishikawa A, Merhige ME. Assessment of coronary artery disease severity by positron emission tomography. Comparison with quantitative arteriography in 193 patients. Circulation 1989; 79:825–835

52. Iliceto S, Marangelli V, Memmola C, Rizzon P. Transesophageal Doppler echocardiography evaluation of coronary blood flow velocity in base-line conditions and during dipyridamole-induced coronary vasodilation. Circulation 1991; 83:61–69

53. Iliceto S, Caiati C, Aragona P, Verde R, Schlief R, Rizzon P. Improved Doppler signal intensity in coronary arteries after intravenous peripheral injection of a lung-crossing contrast agent (SHU508A) . J.Am.Coll.Cardiol. 1994; 23:184–190.

54. Caiati C, Aragona P, Iliceto S, Rizzon P. Improved Doppler detection of proximal left anterior descending coronary artery stenosis after intravenous injection of a lung-crossing contrast agent: a trans-esophageal Doppler echocardiographic study. J.Am.Coll.Cardiol. 1996; 27:1413–1421

55. Powers JE, Burns PN, Souquet J. Imaging instrumentation for ultrasound contrast agents. In: Nanda NC, Schlief R, Goldberg BB, eds. Advances in echo imaging using contrast enhancement. Kluwer Academic Publishers, 1997:139–170

56. Caiati C, Montaldo C, Zedda N, Bina A, Iliceto S. New noninvasive method for coronary flow reserveassessment: contrast-enhanced transthoracic second harmonic echo Doppler. Circulation 1999; 99:771–778.

57. Lambertz H, Tries HP, Stein T, Lethen H. Noninvasive assessment of coronary flow reserve with transthoracic signal-enhanced Doppler echo-cardiography. J.Am.Soc.Echocardiogr. 1999; 12:186–195.

58. Mulvagh SL, Foley DA, Aeschbacher BC, Klarich KK, Seward JB. Second harmonic imaging of an intravenously administered echocardiographic contrast agent: Visualization of coronary arteries and measurement of coronary blood flow. J.Am.Coll.Cardiol. 1996; 27:1519–1525.

59. Caiati C, Montaldo C, Zedda N, Montisci R, Ruscazio M, Lai G, Cadeddu M, Meloni L, Iliceto S. Validation of a new non-invasive method (contrast-enhanced transthoracic second harmonic echo Doppler) for the evaluation of coronary flow reserve. Comparison with intra-coronary Doppler flow wire. J.Am.Coll.Cardiol. 1999; 34:1193–1200

60. Caiati C, Zedda N, Montaldo C, Montisci R,Iliceto S. Contrast-enhanced transthoracic second harmonic echo Doppler with adenosine: a non-invasive, rapid and effective method for coronary flow reserve assessment. J.Am.Coll.Cardiol. 1999; 34:122–130.

61. Gould KL, Martucci JP, Goldberg DI, Hess MJ, Edens RP, Latifi R, Dudrick SJ. Short-term cholesterol lowering decreases size and severity of perfusion abnormalities by positron emission tomo-graphy after dipyridamole in patients with coronary artery disease. A potential noninvasive marker of healing coronary endothelium. Circulation 1994; 89:1530–1538.

62. Arora GD, Reeves WC, Movahed A. Alteration of coronary perfusion reserve in hypertensive patients with diabetes. J.Hum.Hypertens. 1994; 8:51–57.18

63. Pitkänen OP, Nuutila P, Raitakari OT, Rönnemaa T, Koskinen PJ, Iida H, Lehtimäki TJ, Laine HK, Takala T, Viikari JS, Knuuti J. Coronary flow reserve is reduced in young men with IDDM. Diabetes 1998; 47:248–254.

64. Rossen JD, Quillen JE, Lopez AG, Stenberg RG, Talman CL, Winniford MD. Comparison of coronary vasodilation with intravenous dipyridamoleand adenosine. J.Am.Coll.Cardiol. 1991; 18:485–491

65. White CW, Wright CB, Doty DB, Hiratza LF, Eastham CL, Harrison DG, Marcus ML. Does visual

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 150

Page 168: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

151Perfusion Imaging

interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N.Engl.J.Med. 1984; 310:819–824

66. Kern MJ, Donohue TJ, Aguirre FV, Bach RG, Caracciolo EA, Wolford T, Mechem CJ, Flynn MS, Chaitman B. Clinical outcome of deferring angio-plasty in patients with normal translesional pressure-flow velocity measurements. J.Am.Coll.Cardiol. 1995; 25:178–187.

67. Wilson RF, Johnson MR, Marcus ML, Aylward PE, Skorton DJ, Collins S, White CW. The effect of coronary angioplasty on coronary flow reserve. Circulation 1988; 77:873–885.

68. Donohue TJ, Kern MJ, Aguirre FV, Bach RG, Wolford T, Bell CA, Segal J. Assessing the hemo-dynamic significance of coronary artery stenoses: analysis of translesional pressure-flow velocity relations in patients. J.Am.Coll.Cardiol. 1993; 22:449–458.

69. McGinn AL, White CW, Wilson RF. Interstudy variability of coronary flow reserve. Influence of heart rate, arterial pressure, and ventricular preload. Circulation 1990; 81:1319–1330.

70. Takeuchi M, Nohtomi Y, Kuroiwa A. Does coronary flow reserve assessed by blood flow velocity analysis reflect absolute coronary flow reserve? [see comments]. Cathet.Cardiovasc.Diagn. 1996; 38:251–254.

Kardio-Kap 3 for pdf 06.06.2000 18:27 Uhr Seite 151

Page 169: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

1

Harald Becher . Peter N Burns

Handbook of Contrast EchocardiographyLeft ventricular function and myocardial perfusion

Peter N BurnsProfessor of Medical Biophysics and RadiologyUniversity of TorontoImaging ResearchSunnybrook and Women’s Health Science Centre2075 Bayview AvenueToronto, OntarioCanada M4N [email protected]

Harald BecherProfessor of CardiologyUniversity of BonnRheinische Friedrich-Wilhelms-UniversitätMedizinische Universitätsklink und Poliklink IIKardiologie/PneumologieSigmund-Freud-Straße 2553105 [email protected]

Copyright © 2000 by Harald Becher and Peter N Burns. This book is protected by copyright. All rights are reserved, whether the wholepart of material is concerned, specifically the rights of translation, reprinting, reuse of illustrations, recitation, broadcasting, reproductionon microfilm.

This electronic copy was downloaded under the conditions of the End User License Agreement which accompanies it. Use and storageof this document signifies agreement with the terms of the Agreement. The files may not be altered without prior written permission ofthe copyright owners. No text, figures, tables or images may be displayed or reproduced, except for personal use, in any form or by anymeans, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission of the copyright owners.

The use of general descriptive names, registered names, trademarks, etc. in this publication does not imply, even in the absence of a specificstatement, that such names are exempt from the relevant protective laws and regulations and therefore free for general use.

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 1

Page 170: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 2

Page 171: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

4 Methods for quantitative Analysis

Wann überhaupt, wenn nicht jetzt?(When at all, if not now?)

Albert Einstein, 1879–1955

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 3

Page 172: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Introduction✳

It is known that the strength of the echoenhancement in a region perfused by bloodcarrying microbubbles is approximately pro-portional to the number of bubbles present.Though attenuation and other technicalfactors preclude estimation of the absoluteconcentration of bubbles from the echo ampli-tude alone, relative changes in bubbleconcentration in a region of the image can be estimated from the echo signal itself.Knowing this value enables a number ofimportant clinical and physiological estimatesto be made, including those of ejection frac-tion, relative vascular volume, flow velocityand relative perfusion rate. Quantitative report-ing of contrast echo studies is likely to becomeas routine a part of the examination as it is innuclear medicine.

Today, as contrast echocardiography continuesto gain clinical acceptance, the interpretationof images remains essentially qualitative.Techniques for deriving quantitative informa-tion from contrast studies are still primarilyused in the research rather than the clinicalsetting (1–5). Hesitation to use image process-ing in contrast echo may reflect its lack of rolein other applications of echocardiography.Until recently, contrast echo image processinghas required the use of a video frame grabberand custom designed computer software andhardware. Common difficulties such as off-linevideotape transfer, calibration, data com-pression, and lack of a standard computeroperating system has limited its clinical poten-tial. Therefore, users of myocardial contrastecho may be resistant to perform time-consuming off-line analysis and thereforeconfine themselves to visual judgement ofunprocessed recordings.

In response to the need for accessible imageprocessing for clinical contrast echocardio-graphy, several commercially available softwarepackages have emerged from, among others,Sklenar at the University of Virginia, ATLUltrasound, Agilent and GE/Vingmed. Thesepackages are either integrated into the ultra-sound systems or use standard personal com-puter (PC) hardware, and comprise basic soft-ware routines which allows users to analysecontrast echo information quickly. Althoughimage processing may at first appear intimidat-ing, the basic operations when taken singly areeasy to learn and perform.

4.1 Basic tools for image quantification

The simplest form of contrast echo analysisrequires only a single image, for example tomeasure a cavity volume or compare enhance-ment in different regions of the myocardium.Following the time course of enhancementneeds cineloop analysis; this also forms thebasis of flow calculations in myocardial per-fusion studies.

Basic tools used in contrast echo imageprocessing range from simple drawing and stat-istical calculations to sophisticated editing,image alignment, and graphing tools, all ofwhich are aimed at measuring the effect of thecontrast agent on the backscattered signal fromblood. These tools are applied sequentially tosingle frames or to series of frames (cineloops).Therefore, learning the steps involved is assimple a becoming familiar some basic tools,and a flowchart for their use.

154 Handbook of Contrast Echocardiography

✳ This section courtesy of Danny M Skyba and Damien Dolimier, Bothell WA, USA

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 154

Page 173: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

4.1.1 Single image quantification

After selecting a representative contrastenhanced frame, the user may select a pixellocation (‘Pick’), draw a region-of-interest(ROI) encompassing an area, or select a linearROI or ‘Profile’. Figure 1 shows a single imagealong with the results generated by a ‘pick’ anda ROI. The result of a pick or ROI is typicallythe intensity value of the pixel selected, or themean intensity of the pixels contained withinthe ROI. For a profile, a vector or graph ofintensities along the line is displayed. Intensitydata within a ROI can also be displayed as anarray of information or as a histogram, rather

than as a single mean intensity. Figure 1 showsexamples of these data.

155Quantitative Methods

Steps involved in processing singleimage

1. Select a representative frame

2. Select the desired tool for collecting intensity

information, such as a ‘pick’, ‘profile’, ROI or

histogram

3. Draw or ‘pick’ a region or pixel

4. Display and interpret the resulting amplitude data

5. Repeat for different regions in the image, if

desired

Fig. 1 Analysis of a single imageselected from a cineloop. Fourdifferent regions have been drawn inthe image: a pick region, a line (R0),and a free-form ROI (R1). Below areexamples of results which can begenerated using single frame quantifi-cation. The pick result is a dynamicstatistical list to which entries areadded each time as user ‘picks’ a pixel.A profile is generated along the line(R0). A histogram displays the amplitu-de data in the free-form ROI (R1).

‘pick list’ profile along RO histogram in R1

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 155

Page 174: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

4.1.2 Cineloop quantification

Although single image quantification mayallow a user to determine and compare inten-sity values between various myocardial beds orchambers, temporal information is ignored.Since ultrasound is a real-time imaging modal-ity, it is the interpretation of image intensitydata as a function of time that forms the basisfor the quantification of myocardial perfusionfrom contrast echo images. Fortunately, quan-tification of an entire cineloop of images is notmuch more difficult than that of a singleimage. Time behaviour of a bolus is one aspectof a contrast study that can be analysed usingsuch an approach.

Prior to processing cineloop data, the usermust review the images carefully and take noteof the quality of the images and the data. Doesthe loop capture the information required? Isthe image size and quality satisfactory? Arethere artifacts or other image frames whichneed to be excluded? Which frame is bestsuited for placement of a representative ROI?Quick mental notes during review can save lotsof time during the editing, quantification anddata interpretation steps that follow. Addi-tionally, if the user performing the analysis hasdifferent image preferences from the sono-grapher, aesthetic information such as the greymap and colour map can often be changedwithout changing the underlying intensitydata. This can be especially beneficial in creat-ing results that are similar in presentation tothose of nuclear cardiology images. If the dataanalysis is not performed immediately duringthe echo exam, information regarding thepatient is usually stored within the image filesand can be used later to organise stages of astudy.

4.1.2.1 Reduction of dataThe first step in quantifying a cineloopinformation is to reduce the data. Usually,many more images are collected during theultrasound examination than are required foranalysis. Data reduction speeds up imagequantitation routines and also prevents arti-factual data from being displayed as a result.Simple software editing tools allow the user toselect a new first and last frame in the cineloopfor processing. Additionally, the user maysingle out individual image frames and ‘cut’them out of the loop. Frames containing imageartifacts such as ectopic beats or post-extra-systolic beats should be excluded, as shouldframes with severe wall motion artifacts. Othercommon artifacts include arrhythmias, exagger-ated patient breathing, momentarily poortransducer contact or unwanted transducertranslation. ECG information is often includedin the cineloop allowing the user to justify datareduction.

4.1.2.2 Positioning Regions of Interest (ROI)Once relevant image frames have been selected,the next processing step is to specify a ROIwithin which measurements will be made. Arepresentative frame in the cineloop is selectedand the user may choose a particular shape, orcreate a custom ROI. ROIs are often drawnusing the standard segmental model of theAmerican Society of Echocardiography. OneROI should be placed within a single segmentor perfusion territory, and the size of the ROIshould cover most of the segment. ROIs withina segment should be as large as possible to mini-mise statistical error within the ROI, as well asto reduce scatter of the single measurements forconsecutive frames (6). To minimise tedium, theROI is usually drawn only once in the represen-tative frame, and is automatically applied to thesame position in every cineloop frame.

156 Handbook of Contrast Echocardiography

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 156

Page 175: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

Fig. 2 (a) Reference image with a well positioned ROI. (b) Unaligned frame with the ROI falling into the LVcavity. (c) The user performs manual alignment by positioning the unaligned frame (green) over the referenceframe (magenta). (d) Optimal alignment is indicated by a mostly grey image.

a) b)

c) d)

157Quantitative Methods

It is important that the cineloop be reviewed tovalidate the position of every ROI in eachframe to be evaluated. The borders of the ROIsmust not touch the endocardial borders, other-wise the measurements in the ROI may becorrupted by the strong contrast signals withinthe cavity. ROIs which ‘fall out’ of the myo-cardium and into the LV cavity will skew themean intensity value for that frame. To correctROI position problems, image alignment mustbe performed. The specific methods used forimage alignment differ among the software

available, but options for image or regionadjustment are generally available. The firstoption is image alignment, where the positionof the cineloop images can be changed tomatch the position of the fixed ROI. (Figure 2)This is the traditional method of correcting forheart motion and translation in contrast echoimage analysis (7–8). The second option isROI adjustment, where the ROI can be indi-vidually repositioned in each frame to match theanatomy of the unaligned images (Figure 3).When these operations are performed man-

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 157

Page 176: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

ually, both are equally reliable in correcting forsimple motion, such as in plane movement ofthe sonographers hand position. When track-ing an ROI over the entire cardiac cycle, ROIadjustment is preferable, since image align-ment is difficult with dynamically changingimages. In contrast, when images are triggeredat a particular phase of the cardiac cycle, suchas end systole or end diastole, manual imagealignment may be faster. Image alignment alsoallows placement of multiple ROIs and alignsthe image only once for all ROIs.

4.1.2.3 Automatic analysisSome software packages offer options for auto-matic image alignment using cross-correlationtechniques. If available, automatic alignmentroutines should be tried first. When successful,automatic alignment algorithms greatly speeddata analysis. An advantage of image process-ing software packages which capture directdigital data rather than frame grabbed intensitydata is that colour ultrasound information canbe accurately quantified. Because of the im-proved segmentation provided by colour imag-ing techniques, harmonic power Doppler(HPD) images may be superior to B-mode

greyscale images in delineating myocardiumfrom LV cavity in contrast frames. This differ-ence can be visually appreciated using a colour-coded display which is already implementedon commercially available ultrasound systems.High power signals originate from contrastwithin the cavities and are displayed in a singlebright colour when they reach a variablethreshold. Low power signals are caused bycontrast within the myocardium. Differenthues of a complementary colour provide thevisual information on the concentration ofmicrobubbles in the sample volume. Thissegmentation of the LV cavity from the myo-cardium can facilitate ROI placement. Thecontrast in colour along the border of the LV also provides an excellent ‘edge’ for cross-correlation based alignment techniques.Unfortunately, without some additional post-processing, some users of HPD techniquesreport ‘bleeding’ intensity signal from the LVcavity into the myocardium which may resultin artificially high myocardial intensity data.These users prefer to operate on B-mode grey-scale images where ROI placement and imagealignment must often be performed manually,and cautiously.

158 Handbook of Contrast Echocardiography

Fig. 3 The same case as Figure 2 where the user repositions the ROI in each frame so that the ROI repre-sents the same anatomic region. In the case of ROI adjustment, the new ROI appears dashed, indicating thatit has been repositioned; further editing of the ROI is performed on a frame-by-frame basis.

R0 is well positionedin this frame

R0 is not well positionedin a different frame

after adjusting the ROI,R0 is well positioned

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 158

Page 177: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

The final step in basic image processing ofcineloops is to quantify and display the data. Inthe case of a ROI, it is the mean image inten-sity contained within the ROI that is plottedfor every frame in the cineloop, usually as afunction of time. Software that requires cali-bration takes this into account at the time ofresult calculation. Software packages that areintegrated into ultrasound hardware, or useproprietary image formats perform internalcalibration automatically. Tools for rescalingdata results are also generally available.

4.2 Advanced image processing:cases & examples

The current state-of-the-art for advanced pro-cessing of contrast echocardiographic images is

really nothing more than a combination ofbasic image processing routines with a fewtechnical and mathematical enhancements. Inthis section, we describe some commonlyemployed routines through a series of experi-mental and clinical examples.

4.2.1 Single image quantification

The simplest form of quantification is on asingle still image taken from a cineloop.Typically, a single representative frame can bechosen from a baseline cineloop, where apatient is at rest, and another image can betaken from a cineloop during peak stress.Visually disparate information between thetwo images can be validated quantitatively bymeasuring intensity levels in various perfusionbeds before and after stress. The tools of choicein this case are the intensity ‘pick’, or the ROItool with comparison of the mean image inten-sity under the ROI. Figure 4a shows normalmyocardium at rest. Two regions have beencarefully selected in the LAD and LCx beds,respectively, and pick results as well as themean intensity under the ROI is shown forboth ROIs. In Figure 4b, a moderate stenosishas been applied to the LCx. The same ROIshave been overlaid on this image and the pickand mean intensity values are shown. In thissimple case, it is visually apparent that there isa lack of contrast in the region supplied by thestenosed LCx. The statistical informationprovided by the image processing softwareprovides a means of putting a value on thedecrease in intensity. In experimental models,it has been shown that the normalised decreasein intensity can be correlated to a decrease inmyocardial flow. Therefore, in the case of amild stenosis or where visual information is notas clear in the example given here, computer

159Quantitative Methods

Steps involved in processing cineloops:

1. Load and review the cineloop visually . Note the

need for loop editing and alignment

2. Edit the cineloop by selecting new first and last

frames (if necessary)

3. Remove individual frames from the cineloop that

may contain artifacts

4. Select a representative frame from the cineloop

5. Draw one or more ROIs on the image in the

desired perfusion territories

6. Visually review the edited cineloop, carefully

noting the position of the ROI(s) in each frame

7. Align the images or the individual ROIs. Try

automatic routines (if available) before resorting

to manual alignment

8. Review aligned ROIs. Re-adjust individual ROIs

manually if necessary

9. Display and scale the results

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 159

Page 178: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

aided statistical processing offers a method forapproximate assessment of perfusion.

4.2.2 Background subtraction

The use of background subtraction has beenchampioned by Kaul, et. al. since the earliestdays of contrast echocardiographic image pro-cessing (7–8). While the concept is not diffi-

cult to understand, it is very important ifharmonic B-mode or pulse inversion are usedfor imaging the myocardium. This is becausethe level of contrast enhancement at high MI isoften only a few decibels above the backgroundlevel created by the tissue harmonic.Subtraction assumes that image signal otherthan the enhancement due to the contrastagent is ‘background’ information, or noise.Additionally, the amount of background infor-

160 Handbook of Contrast Echocardiography

Fig. 4 (a) A normally perfused myocardium. A statistical intensity ‘pick’ has been applied to several pointsin both beds, and the results are displayed in the window. A simple ROI has also been drawn and the meanintensity for both beds is shown in the window (in dB, amplitude and linear scales). (b) The same myocardialplane with a moderate stenosis in the LCx bed. The results show a decrease in ‘pick’ and mean intensity

a) b)

‘pick list’ ‘pick list’

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 160

Page 179: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

161Quantitative Methods

mation, or noise-floor, may change slightly dueto subtle changes in the position of the trans-ducer during an echo examination. Since theonly signal of interest is that due to the agent,the background intensity information is sub-tracted from the total signal and discarded.When this concept is applied to the B-modeimage, areas of increased perfusion appearbrighter, while areas which lack perfusion, hencecontrast enhancement, appear darker. Sincethe signal due to the contrast may be subtle,this simple technique provides the clinicalreader with an enhanced visual perception ofperfusion (Figure 5).

In practice background subtraction shouldalways be performed with multiple imagegroups where the result of the subtraction is,more or less, an average subtracted image. Allimages should be acquired at the same point inthe cardiac cycle and they should be dividedinto two groups of three or more images: thebackground group and the contrast-enhancedgroup. For each pixel position within a group,an image is created containing mean intensityper pixel position. Then the mean values of thetwo groups are subtracted as shown above, ona pixel by pixel basis. An example of this isshown in Figure 6. Figure 6a shows three pre-contrast images and the mean pre-contrastimage. Figure 6b shows three post-contrastimages and the mean post-contrast image.Figure 6c shows the result of the subtraction ofthe pre-contrast image from the post-contrastimage, where the intensity is primarily attri-buted to contrast agent. In this single image wehave captured only the information of primaryinterest, thus making a more rapid interpreta-tion possible. When combined with curve-fitting techniques, the technique of backgroundsubtraction provides the data for the statisticalanalysis of perfusion.

Figure 7 shows an example of automated postprocessing used to evaluate adenosine inducedchanges in myocardial perfusion volume in apatient with a 99 percent left anterior descend-ing artery lesion. After recording pre- and post-adenosine baseline images (Figure 7a, b), theanalysis software performs automatic alignmentof the images combined to produce averagedimages (Figure 7c, d). Some reduction in imageintensity, representing a reduction in myo-cardial blood volume and redistribution ofmyocardial perfusion is seen at the apex. Thepost adenosine averaged image is subtractedfrom the post adenosine image and the result-ing difference displayed with a colour map(Figure 7e). Since the acquired images havelogarithmic grey scales, the effect of subtrac-tion is to display their ratio. Myocardialsegments that are perfused by coronary bedswith good flow reserve should increase in myo-cardial blood volume following adenosine andare shown in red shades. Segments supplied bydiseased vessels become relatively underper-fused and the resulting reduction is colourcoded blue. In practice, adenosine can cause ahyperventilatory response in some patientswhich can make image acquisition difficult.

Fig. 5 (a) 44-year old woman with infero-lateralMI who underwent obtuse marginal branch PTCA.(b) Subtracted contrast images show improved per-fusion in the infero-posterior wall following PTCA. Courtesy of Thomas Marwick and Brian Haluska,University of Brisbane, Australia

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 161

Page 180: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

162 Handbook of Contrast Echocardiography

Fig. 6 (a) Three pre-contrast images and the mean pre-contrast image. (b) Three post-contrast images andthe mean post-contrast image. (c) The result of the subtraction of the pre-contrast image from the post-contrast. The technique has reduced the amount of background ‘noise’ and highlights the intensity attributedto contrast agent.

a)

b)

c)

background frames

average of the background frames

average of the contrast frames

contrast frames

linear substraction

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 162

Page 181: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

163Quantitative Methods

Fig. 7 Post processing used to evaluate adenosine induced changes in myocardial perfusion volume in apatient with a 99 percent left anterior descending artery lesion. End systolic, 2 beat interval intermittent,pulse inversion images are acquired during an injection of Sonazoid before (a) and after (b) a short adenosineinfusion. The images are transferred to an off-line workstation for processing using software developed by Dr Morten Eriksen (Nycomed Amersham), which performs automatic alignment of the images which areaveraged to produce pre- (c) and post- (d) adenosine images. Some reduction in image intensity, represent-ing a reduction in myocardial blood volume and redistribution of myocardial perfusion is seen at the apex.The post adenosine averaged image is subtracted from the post adenosine image and the resulting differencedisplayed in colour (e). Since the acquired images have logarithmic grey scales, the effect of subtraction is todisplay their ratio. Good flow reserve areas give high signal ratios and are shown in red; poor flow reserveareas with low ratios are shown in blue.Courtesy of Mark J Monaghan, King’s College Hospital, London, UK and Morten Eriksen, NycomedAmersham, Oslo, Norway.

a)

b)

c)

d)

e)

Kardio-Kap 4 for pdf 06.06.2000 19:12 Uhr Seite 163

Page 182: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

164 Handbook of Contrast Echocardiography

Fig. 8 (a) A series of four images taken in time sequence from a cineloop. The first shows the image priorto bolus injection. The second shows the LV cavity filled with contrast, but no agent is visible in the myocar-dium. The third image shows the agent in the myocardium at its peak intensity level. Image four shows theslowly decreasing intensity level as the agent ‘washes-out’ of the vascular system. (b) The mean backgroundsubtracted intensity data from the ROI corresponding to all end-systolic images in the cineloop. The mathe-matical best-fit curve is overlayed on the data.

However, this technique clearly shows thepotential of the combination of three newtechnologies: adenosine echo contrast perfusion,pulse inversion imaging and colour-coded sub-traction post processing.

4.2.3 Analysing the time-course of enhancement

4.2.3.1 Intravenous bolus techniqueTracking changes in the intensity level due tocontrast agents over the time-course of a bolusinjection yields information which is related tothe flow of blood to the myocardium (8–10).While deriving information by visually track-

ing the intensity due to a bolus is difficult,computer aided image processing and statisti-cal curve-fitting are well suited to this task. Theresult of computed curve-fitting is a singleparameter which in certain circumstances canbe related to the perfusion rate of blood to themyocardium.

The technique is usually performed as follows.Real-time information in the form of cineloopof images is acquired just prior to, and duringthe injection of a rapid bolus of contrast agent.Figure 8a shows a series of four images taken insequence from a cineloop during this timecourse. The first shows the image prior to bolusinjection. The second shows the LV cavity

a)

b)

priot to bolus injection LV cavity filled with contrastno agent visible

in the myocardium

agent in the myocardium atits peak intensity level

mean background substracted intensity datamathematical best-fit curve is overlayed on the data

agent ‘wash-out’

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 164

Page 183: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

165Quantitative Methods

filled with contrast, with no agent is visible inthe myocardium. The third image shows theagent in the myocardium at its peak intensitylevel, and the fourth shows the slowly decreas-ing intensity levels as the agent washes out ofthe vascular system. Figure 8b shows the dataas a series of points taken from the full cine-loop of information. Each data point is themean intensity under the ROI for each frameof the cineloop. In this example, the data werereduced so that only end-systolic frames areshown. Each data point is subtracted from thelevel of background intensity, taken as an aver-age of a few representative frames prior to theinjection of contrast. The data usually are thenfitted to a curve predicted by a model such asthe gamma-variate model. This process reducesthe data to only two parameters: the amplitude

scale and the mean-transit time. The meantransit time has been shown to correlate closelywith myocardial flow in experimental modelsunder very specific conditions and may there-fore be relevant to clinical perfusion studies.Many commercial image processing packageshave the provision for this curve-fitting pro-cedure. However, a number of difficulties withthis technique, especially the lack of knowledgeof the precise form of the bolus that enters themyocardial bed, limit its usefulness in clinicalpractice, for which the negative bolus methodoffers many advantages.

4.2.3.2 Negative bolus techniquePrinciples

In this technique, a continuous infusion ofcontrast agent during intermittent triggered

Fig. 9 Negative bolus destruction-reperfusion study using harmonic power Doppler. Regions of interestare placed in the apex, lateral wall and septum and repeated measurements made at each of 1, 2, 4 and 8heartbeat intervals. The graph shows reperfusion curves for each area. All fit well to a monoexponentialfunction, as predicted by the model described in the text.

a)

b) c)

d)

1:2 1:4

1:8

1:1

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 165

Page 184: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

166 Handbook of Contrast Echocardiography

imaging is used to overcome the limitations ofthe bolus injection model. The method isbased on the simple principle of destruction-reperfusion (11,12). The transducer ispositioned over the region of interest in themyocardium and a steady infusion of the agentis established. Using ECG-gated, high MIimaging in harmonic power Doppler mode, aseries of perfusion images is made of the myo-cardium. At one trigger per heartbeat, theseimages will reflect the amount of blood thathas reperfused the myocardium in a singleheartbeat (Figure 9 a). If the interval is increas-ed to two heartbeats, a higher intensity is seenas bubbles from smaller vessels refill the myo-cardium after each disruptive pulse (Figure8 b). At a four beat interval, the echo level of

the myocardium increases more (Figure 9 c). Byan eight beat interval, the myocardium attainsits maximal brightness, because sufficient timehas passed for all bubbles, even those movingmost slowly, to refill the myocardium (Figure9d). It can readily be appreciated that thebrightness of the final image reflects the totalnumber of bubbles present in the region ofinterest (the total vascular volume), whereasthe rate at which this level is attained reflectsthe speed (or mean velocity) of flow into theregion. The product of these two quantities isthe flow or perfusion rate. By plotting themean echo level against time (Figure 9 e), theflowrate can be estimated by fitting a simpleexponential curve. The principle is identical tothat of indicator dilution theory used in, for

Fig. 10 Principle of negative bolus (or destruction-reperfusion) indicator dilution measurement. During asteady infusion of agent, the bubbles are disrupted in the scan plane. The rate at which they wash back in tothe image plane is proportional to flow rate. The relative number of bubbles present at incremental intervalsfrom disruption is measured by a further imaging frame, triggered to occur at varying times after the destruc-tion. Real-time power pulse inversion imaging can image the reperfusion process in real time.

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 166

Page 185: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

167Quantitative Methods

example, the thermodilution calculation ofcardiac output, but here the bolus is the de-struction of the agent, hence the name of thetechnique. It should be noted that because theecho level from the region of interest isinfluenced by quantities other than the bubblenumber (for example, the cavity attenuation),this method only yields a relative flow value.Figure 10 shows the principle of the methodand the results of mathematical analysis of thebolus behaviour. Analysis of a negative bolusstudy is straightforward: all that is required isto find the asymptotic (that is the plateau)value (A) of the enhancement following reper-fusion, and the exponent (β) of a simple expo-nential function fitted to the reperfusion curve.The value A is related to the vascular volume,the value β to the reperfusion velocity. Theirproduct yields an estimate of relative perfusionrate.

Clinical techniqueUsing a steady infusion, ECG gated images areacquired at different triggering intervals on theultrasound system. Individual images in thecineloop can be grouped according to thetriggering interval. A ROI is positioned as pre-

Negative bolus measurement of myocardial perfusion

1. Establish an infusion and set up imaging (see §1.2.3)

2. Acquire a set of images at each trigger interval

3. Freeze and store image data

4. Position ROI in myocardium and review to ensure that cavity is excluded

5. Plot reperfusion curve and use analysis software to fit exponential

6. Compare exponent (related to velocity) and asymptote (related to volume) between regions of interest

Fig. 11 Negative bolus destruction-reperfusion study using harmonic B-mode. Five representative framesare taken at every 1, 3, 5, 9 and 12 ECG triggering intervals (a). The grouped and averaged data from the entirecineloop are shown in (b) with the mathematical best-fit.

a)

b)

group 1every 1 ECG trig. interval

group 2every 3 ECG trig. interval

group 3every 5 ECG trig. interval

group 4every 9 ECG trig. interval

group 5every 12 ECG trig. interval

mean intensity for each groupmathematical best-fit overlayed on data

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 167

Page 186: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

168 Handbook of Contrast Echocardiography

viously described, and the mean intensity datais averaged with data from within its group.Curve fitting is then performed, using a mono-exponential function. Mathematically, thetechnique is appealing since all of the pertinentinformation can be captured with a single

expression containing three variables: the back-ground intensity level (the constant), the rateof rise (the initial slope), and a final asymptoticvalue (the plateau). It has been shown that theslope may relate to the rate of reperfusion ofthe myocardium after contrast destruction by

Fig. 12 Parametric images of a heart with experimental stenosis of the LAD coronary artery, displayingcalculated parameters from the negative bolus method. These ‘parametric’ images show (a) Blood volume,(b) blood velocity, (c) blood flow and (d) goodness of fit of the exponential model. Images were digitised fromvideo tape, selected, aligned, averaged, and masked. Then an exponential function was fitted for everyselected pixel and parameters defining the local best fit curve displayed using different colour maps.Courtesy of Jiri Sklenar, University of Virginia, USA. Processing was done on a Macintosh PC using softwaredeveloped at the University of Virginia and by Yabko, LLC.

a) b)

c) d)

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 168

Page 187: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

169Quantitative Methods

ultrasound, and that the plateau value mayrelate to the volume of the myocardial bed.Figure 11 illustrates this type of analysis per-formed on harmonic B-mode images. Figure12 shows how the processing can be taken astep further, producing parametric images

whose colours display the quantities calculatedby the model. These include relative bloodvolume, blood velocity and relative blood flow.An image can also be made (Figure 12 d) show-ing the extent to which the data conform to themodel.

Fig. 13 Real-time myocardial contrast during a slow bolus injection at low MI using Power Pulse Inversionimaging (a). High power frames destroy all of the contrast agent in the scanplane (b). The agent begins toreperfuse the scanplane (c–d), and fully reperfuses the myocardium in a few cardiac cycles (e). Real-timedata collection begins just after the high power frames. The data is reduced to include only end diastolic fra-mes, and a monoexponential curve is fitted (f ).

a)

d) e)

f )

b) c)

full myocardial contrastduring a slow bolus injection

high power ultrasound destroysthe contrast agent in the scanplane

the agent begins to reperfusethe scanplane

the agent begins to reperfusethe scanplane

reduced data fit to a 1-exponential function

the agent reperfuse the myocardiumin a few cardiac cycles

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 169

Page 188: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

170 Handbook of Contrast Echocardiography

4.2.3.3 Real-time negative bolus techniqueThe new ultrasound technology of power pulseinversion (see §1.3.3.6) allows the rate ofreperfusion of contrast to be determined inreal-time, rather than with intermittenttriggered imaging. The scanner is set to imagecontinuously at low MI (typically 0.1 or less).The trigger then determines the point at whicha single burst of high MI frames are introducedto disrupt the bubbles. Reperfusion is thenimaged continuously at low MI. Figure 13shows the real-time reperfusion of contrast intothe myocardium after such a series of high MIframes have disrupted all of the contrast agentin the scanplane. The monoexponential func-tion is applied to the entire data set, or to thereduced data. Because it works in real-time,this quantitative method is most likely to evol-ve into technique integrated into a scanner’ssoftware.

4.3 Summary

As the importance of contrast ultrasoundgrows in the field of cardiology, so will the needfor automated routines to provide robust andconsistent data to the cardiologist. Based onthe interest and investment in the field ofcontrast echocardiography, it appears likelythat the image processing routines discussedhere will soon become incorporated into thedaily routine of the clinical lab. The impor-tance of providing the research clinician with auser-friendly set of tools can be appreciated.Simultaneously, research in contrast echo-cardiography is still evolving rapidly. The soft-ware developed for contrast echo must have thecapacity for future enhancement and expan-sion. Fortunately, this type of growth can beprovided through the use of expandable soft-

ware modules. Tools are likely to mature to thepoint that they will be simple to use, stan-dardised and perhaps seamlessly incorporatedinto ultrasound systems themselves, graduatingfinally from their modest origins in an off-linepersonal computer.

4.4 References

1. Ong K, Maurer G, Feinstein S, Zwehl W, Meerbaum S, Corday E. Computer methods for myocardial contrast two-dimensional echocardio-graphy. J Am Coll Cardiol 1984; 3:1212–1218.

2. Skorton DJ, Collins SM, Garcia E, Geiser EA, Hillard W, Koppes W, Linker D. Schwartz G. Digital signal and image processing in echocardiography. The American Society of Echocardiography. American Heart Journal.1985; 110:1266–1283.

3. Ten Cate FJ, Cornel JH, Serruys PW, Vletter WB, Roelandt J. Mittertreiner WH. Quantitative assessment of myocardial blood flow by contrast two-dimensional echocardiography: initial clinical observations. American Journal of Physiologic Imaging 1987; 2:56–60.

4. Kemper AJ, Force T, Kloner R, Gilfoil M, Perkins L, Hale S, Alker K, Parisi AF. Contrast echocardio-graphicestimation of regional myocardial blood flow after acute coronary occlusion. Circulation 1985; 72:1115–24.

5. Sklenar J, Jayaweera AR, Kaul S. A computer-aided approach for the quantitation of regional left ventricular function using two-dimensional echo-cardiography. J Am Soc Echocardiography 1992; 5:33–40.

6. Jayaweera AR, Skyba DM, Kaul S. Technical factors that influence the determination of microbubble transit rate during contrast echocardiography. J Am Soc Echocardiography 1995; 8:198–206.

7. Jayaweera AR, Matthew TL, Sklenar J, Spotnitz WD, Watson DD, Kaul S. Method for the quantitation of myocardial perfusion during myocardial contrast two-dimensional echocardio-graphy. J Am Soc Echocardiography 1990; 3:91–98.

8. Jayaweera AR, Sklenar J, Kaul S. Quantification of images obtained during myocardial contrast echo-cardiography. Echocardiography 1994; 11:385–396.

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 170

Page 189: the-eye.euthe-eye.eu › public › Books › BioMed › Handbook of Contrast... · Although the technology required for the successful application of contrast echocardiography has

171Quantitative Methods

9. Vandenberg BF, Kieso R, Fox-Eastham K, Chilian W, Kerber RE. Quantitation of myocardial perfusion by contrast echocardiography: analysis of contrast grey level appearance variables and intra-cyclic variability. J Am Coll Cardiol 1989; 13:200–206.

10. Jayaweera AR, Sklenar J, Kaul S. Quantification of images obtained during myocardial contrast echo-cardiography. Echocardiography 1994; 11:385–396.

11. Wei K, Skyba D M, Firschke C, Jayaweera AR, Lindner J R, Kaul S. Interactions between micro-bubbles and ultrasound: in vitro and in vivo observations. J Am Coll Cardiol 1997; 29:1081–1088.

12. Wei K, Jayaweera AR, Firoozan S, Linka A,Skyba DM, Kaul S. Quantification of myocardial blood flow with ultrasound-induced destruction of microbubbles administered as a continuous venous infusion. Circulation 1998; 97:473–483.

Kardio-Kap 4 for pdf 06.06.2000 19:13 Uhr Seite 171