Pitfalls of Hemodynamic Monitoring in Patients with Trauma Richard R. McNeer, MD, PhD a,b, *, Albert J. Varon, MD, MHPE, FCCM a,c INTRODUCTION Resuscitation of the patient with trauma with ongoing blood loss has historically been geared toward maintaining normal or higher than normal blood pressure. Investiga- tions of the role of hypotensive resuscitation have been conducted with mixed results. Funding sources: None. Conflict of interest: None. a Division of Trauma Anesthesiology, Department of Anesthesiology, University of Miami Miller School of Medicine, PO Box 016370-M820, Miami, FL 33101, USA; b Department of Biomedical Engineering, College of Engineering, University of Miami, PO Box 248294, Coral Gables, FL 33146, USA; c Ryder Trauma Center at Jackson Memorial Hospital, 1800 NW 10th Avenue, Miami, FL 33136, USA * Corresponding author. Division of Trauma Anesthesiology, University of Miami Miller School of Medicine, PO Box 016370-M820, Miami, FL 33101. E-mail address: [email protected] KEYWORDS Trauma Hemodynamic monitoring Hemorrhagic shock Blood pressure measurement Dynamic response Natural frequency End-tidal carbon dioxide Assessment of cardiac output KEY POINTS Hypotensive resuscitation imposes additional burdens on the anesthesiologist to maintain adequate blood pressure while also avoiding blood pressure overshoots or sustained hypertension. Accuracy and interpretation of blood pressure monitoring in trauma depend on complex interactions between patient physiology and factors intrinsic to measurement devices. Oscillometry overestimates systolic blood pressure in hypotensive patients. The impact of hemorrhagic shock on reflection wave behavior may limit the reliability or confound interpretation of both noninvasive and invasive blood pressure measurements. Adequate dynamic response is required for proper recording of an arterial waveform, and dynamic response requirements increase with tachycardia. The only way to determine dynamic response is to conduct a flush test. End-tidal CO 2 is correlated with cardiac output; however, this correlation is affected following cardiac or respiratory arrest. Anesthesiology Clin - (2013) -- http://dx.doi.org/10.1016/j.anclin.2012.11.005 anesthesiology.theclinics.com 1932-2275/13/$ – see front matter Ó 2013 Elsevier Inc. All rights reserved.

The Environmental Services Grou p

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The Environmental Services Grou p

Pitfal ls of HemodynamicMonitoring in Patients withTrauma

Richard R. McNeer, MD, PhDa,b,*, Albert J. Varon, MD, MHPE, FCCMa,c

KEYWORDS

� Trauma � Hemodynamic monitoring � Hemorrhagic shock� Blood pressure measurement � Dynamic response � Natural frequency� End-tidal carbon dioxide � Assessment of cardiac output

KEY POINTS

� Hypotensive resuscitation imposes additional burdens on the anesthesiologist to maintainadequate blood pressure while also avoiding blood pressure overshoots or sustainedhypertension.

� Accuracy and interpretation of blood pressure monitoring in trauma depend on complexinteractions between patient physiology and factors intrinsic to measurement devices.

� Oscillometry overestimates systolic blood pressure in hypotensive patients.

� The impact of hemorrhagic shock on reflection wave behavior may limit the reliability orconfound interpretation of both noninvasive and invasive blood pressure measurements.

� Adequate dynamic response is required for proper recording of an arterial waveform, anddynamic response requirements increase with tachycardia.

� The only way to determine dynamic response is to conduct a flush test.

� End-tidal CO2 is correlated with cardiac output; however, this correlation is affectedfollowing cardiac or respiratory arrest.

INTRODUCTION

Resuscitation of the patient with trauma with ongoing blood loss has historically beengeared toward maintaining normal or higher than normal blood pressure. Investiga-tions of the role of hypotensive resuscitation have been conducted with mixed results.

Funding sources: None.Conflict of interest: None.a Division of Trauma Anesthesiology, Department of Anesthesiology, University of Miami MillerSchool of Medicine, PO Box 016370-M820, Miami, FL 33101, USA; b Department of BiomedicalEngineering, College of Engineering, University of Miami, PO Box 248294, Coral Gables, FL33146, USA; c Ryder Trauma Center at Jackson Memorial Hospital, 1800 NW 10th Avenue,Miami, FL 33136, USA* Corresponding author. Division of Trauma Anesthesiology, University of Miami Miller Schoolof Medicine, PO Box 016370-M820, Miami, FL 33101.E-mail address: [email protected]

Anesthesiology Clin - (2013) -–-http://dx.doi.org/10.1016/j.anclin.2012.11.005 anesthesiology.theclinics.com1932-2275/13/$ – see front matter � 2013 Elsevier Inc. All rights reserved.

Page 2: The Environmental Services Grou p

McNeer & Varon2

A review of 52 animal studies of mortality during induced hemorrhage found that hypo-tensive resuscitation reduced the risk of death.1 However, in human studies,decreased blood loss and transfusion requirements have only been confirmed inpenetrating injury.2 Other studies3 are currently in progress and may add evidenceto support a paradigm shift in the management of blood pressure in the patient withtrauma. It may no longer be sufficient to maintain blood pressure at an adequate level,and added vigilance may be needed to prevent sustained overshoots in blood pres-sure. It is therefore ironic that the monitoring strategies meant to inform the anesthe-siologist of a patient’s hemodynamic status are limited in the context of hypotensionand the deranged physiology associated with hemorrhagic shock.This article discusses the complex interplay between alterations in the physiology of

a patient with trauma, the monitoring devices engineered to measure these alter-ations, and interpretation of these data by the anesthesiologist. Focus is placed onthe early stages of caring for the patient with trauma before surgical hemostasis is ob-tained, when hemodynamic assessment must be made quickly and accurately tomaintain sufficient cardiac output and blood pressure without exceeding levels thatmay worsen outcome. Hemodynamic monitoring and resuscitation end points ofpatients in the period after surgical hemostasis have been reviewed elsewhere4 andare only mentioned here. For example, monitoring strategies based on arterial pres-sure waveform variation, venous oxygen saturation (Svo2) and central venous oxygensaturation (ScvO2), and laboratory parameters such as base excess and lactate havebeen, or are increasingly becoming, a part of the trauma anesthesiologist’s armamen-tarium, but it is often not feasible to implement these strategies during the initial,frenetic stages of care. The 3 bread-and-butter monitors heavily relied on during theinitial assessment and resuscitation of the unstable patient with trauma are arguablythe end-tidal pressure of CO2 (PETCO2), noninvasive blood pressure (NIBP), and inva-sive arterial blood pressure (IABP). The underpinnings and limitations of these moni-tors in the patient with trauma are discussed in reverse order starting with invasiveblood pressure, acknowledging that the other two may be the only hemodynamicmonitors available before an arterial catheter can be inserted. Wherever possible,figures are included to illustrate important points from the text.

Invasive Blood Pressure Monitoring

When using IABP to guide the management of patients with trauma, it is incumbent onthe physician to discern true signal (the physiologic waveform) from artifact (a dis-torted waveform). The physiologic waveform should be processed with high fidelityto obtain an accurate recording. Several characteristics of the patient with traumaand of the recording device can lead to the generation of distorted (or what appearsto be distorted) waveformmorphology. Patients with trauma are likely to exhibit tachy-cardia, hypotension, decreased cardiac output, and high (or occasionally low)systemic vascular resistance (SVR). These derangements in patient physiology canexceed the processing capability of the catheter-tube-transducer (CTT) systemused to invasively measure blood pressure. Factors that affect waveform morphologycan lead to the display of inaccurate blood pressure values and potentially confoundclinical interpretation. Two of the main factors are pulse wave reflection within arterialvasculature and the dynamic response characteristics inherent to the CTT system.To understand why these factors are important, it is useful to recall that any physio-

logic waveform is decomposable into several sine waves of varying amplitude,frequency, and phase (Fig. 1). The reverse process is also true, and sign waves canbe selected and summed to generate a simulated waveform. It takes just 3 sine wavesto give a persuasive approximation of the typical arterial waveform.5 Of the 3 sinewaves

Page 3: The Environmental Services Grou p

Fig. 1. Generation of simulated arterial waveform from sine waves. Summation of 3 sinewaves (A), 1 fundamental (sine wave 1) with frequency equal to pulse rate and 2 harmonics(sine waves 2 and 3) produces a good approximation of an arterial waveform at a pulse rateof 150 beats per minute (B). Sine waves of higher frequency relative to pulse rate arerequired to produce detailed features in the waveform such as the dicrotic notch and steepsystolic upstroke; change in pressure (dP)/change in time (dt) (arrows). The simulation wasperformed using custom scripts written in MATLAB, and sine wave frequencies and ampli-tudes from Szockik and colleagues5 (2000). (Data from Szockik JF, Barker SJ, Tremper KK.Fundamental principles of monitoring instrumentation. In: Miller RD, editor. Anesthesia.5th edition. Philadelphia: Churchill Livingstone; 2000. p. 1053.)

Pitfalls of Hemodynamic Monitoring in Trauma 3

(see Fig. 1A), the first has the lowest frequency, which corresponds with the heart rate.The secondand third sinewaveshave successively higher frequencies and lower ampli-tudes, which give the simulated waveform finer features such as the dicrotic notch. Thesecond and third sinewaves are required to simulate the systolic upstroke (seeFig. 1B).

Page 4: The Environmental Services Grou p

McNeer & Varon4

Steeper upstrokes require the summation of sinewaves of even higher frequencies. TheCTTsystemmust be able toprocess these higher frequencies to avoidwaveformdistor-tion (discussed later). The concept of waveform summation is also applicable to themorphology of the physiologic arterial waveform, which is determined by the overlapof pulse waves generated by the heart and waves reflected from the arterioles.

Pulse wave reflectionReflection waves are not abnormal, but are important contributors to the arterial wave-form in most animals.6 When blood is ejected from the heart, pulse pressure wavesgenerated in the ascending aorta propagate toward the periphery at a velocity thatis faster than the flow wave. On reaching distal arteriolar beds, the incident pulsewave is reflected, generating a retrograde pulse wave.6 The incident and reflectedpulse waves overlap (summate) to form a new composite waveform (Fig. 2). However,

Fig. 2. Demonstration of the effect of pulse wave reflection on arterial waveform mor-phology. The simulated arterial waveform from Fig. 1 is shownwith its reflection (A). Summa-tion of the incident and reflected wave produces a composite arterial waveformwith systolicanddiastolicwaves (B).Whenpulsewavevelocity is increased,as inhypertension, the reflectedwave overlaps sooner (C) yielding a characteristic late systolic peak (arrow) and diminisheddiastolic wave. Hypotension is associated with slower pulse wave velocities, which leads tooverlap of reflected waves later in the cardiac cycle (D), and large diastolic waves are seen(arrow). These simulations were performed using custom scripts written in MATLAB.

Page 5: The Environmental Services Grou p

Pitfalls of Hemodynamic Monitoring in Trauma 5

in the vasculature, the final arterial waveform is the summation of reflected waves fromall arteriolar beds, with reflected waves sometimes being rereflected several times.Wave reflection is important in maintaining normal blood pressure and organ perfu-sion. Three factors influence the effect of wave reflection on the final arterial waveform:distance to reflection points (arteriolar beds), pulse wave velocity, and degree of vaso-constriction. The typical distance a pulse wave has to travel before it is reflecteddepends on the location of the arteriolar bed. For example, the distance from the heartto the brachial bed is about 0.5 m. In a normal adult, pulse waves travel about 10 m/s,with the velocity varying directly with the arterial blood pressure.7 Reflected wavestypically have amplitudes of 80% of the originating waveform amplitude.8 In youngpatients, diastolic blood pressure (DBP) is increased by reflected waves overlappingduring diastole, and this helps maintain coronary perfusion. However, in elderly andhypertensive patients, pulse wave velocity is increased because of blood pressureand decreased arterial distensibility.6,9 This increased velocity allows the reflectedwave to arrive sooner and overlap with the incident wave, leading to increased systolicblood pressure (SBP), decreased DBP, and the formation of a characteristic latesystolic peak in the arterial waveform (see Fig. 2C, arrow).6

Pulse wave reflection is particularly important in explaining the arterial waveform inthe setting of hemorrhagic shock, which is associated with hypotension, tachycardia,and increased arteriolar vasoconstriction. At low blood pressures, pulse wavevelocity can be significantly decreased, whereas the duration of systole is decreasedin tachycardia. Vasoconstriction is a common physiologic response to shock and isoften manipulated with pharmacotherapy by the anesthesiologist. Conditions thatdecrease SVR, such as sepsis or pharmacologic vasodilation, minimize the degreeof wave reflection. However, as noted earlier, increased vasoconstriction in responseto shock or administration of peripheral vasoconstrictors maximizes wave reflection.6

The net effect of hypotension, tachycardia, and vasoconstriction contributes to thegeneration of a composite arterial waveform with a characteristic large diastolicwave (see arrow in Fig. 2D and waveform in Fig. 3B), and, at times, a slowed systolicupstroke.6

Central aortic pressure may be overestimated by up to 20% if SBP is measured witha peripheral catheter.10 This is because, in addition to the factors already discussed,waveform morphology depends strongly on the location of measurement, as shownby the observed differences in waveform appearance at the dorsalis pedis relativeto the radial artery catheter site. These differences are attributed to overlap of incident,reflected, and rereflected waves from various arteriolar beds, and to the contributionof reflected waves to systolic pressure occurring earlier as measurements arerecorded further from the heart.8 Therefore, many factors contribute to the mor-phology of the arterial waveform (Table 1), and the anesthesiologist must be carefulto consider these factors when interpreting information obtained from arterial CTTsystems. In particular, SBP overestimation should be avoided by considering SBPin conjunction with mean arterial pressure (MAP), because MAP is not significantlyaffected by reflection waves6 and measurement site.The impact of trauma pathophysiology on the contribution of reflected waves to the

arterial waveform can lead to incorrect identification of waveform distortion. Contraryto popular opinion, it is not possible to determine whether a recorded arterial wave-form is overdamped or underdamped simply by casually observing waveformmorphology (see Fig. 3).11 The only way to determine whether a recorded waveformis distorted is to assess the dynamic response of the CTT system.11,12 This assess-ment can be accomplished at the bedside by performing a flush test,13 or in simula-tions by introducing a square wave11 and analyzing the resulting oscillating waves.13

Page 6: The Environmental Services Grou p

Fig. 3. Two arterial waveform recordings typically seen in trauma. The first waveform couldbe incorrectly considered to be underdamped (A), whereas the second might be consideredto be overdamped (B). However, the dynamic response characteristics underlying eachrecording were satisfactory.

Table 1Factors affecting invasive and NIBP measurement

CTT Oscillometry

Systolic Mean Systolic Mean

Distal measurement site O — — —

Hypotensive (shock) V — O Va

Wave reflection V — Vb Vb

Overdamped U — n/a n/a

Underdamped O — n/a n/a

Abbreviations: n/a, not applicable; O, overestimated; U, underestimated; V, variable effect.a Possible unreliability caused by group-specific differences in amplitude envelope.b The algorithm used in oscillometry to calculate blood pressures is affected in part by arterial

waveform morphology and pulse rate.

McNeer & Varon6

Page 7: The Environmental Services Grou p

Pitfalls of Hemodynamic Monitoring in Trauma 7

Dynamic response of the IABP monitoring systemCTT systems have been shown to behave like underdamped second-order systems.14

As such, the dynamic response of the CTT can be characterized in terms of its naturalfrequency and damping coefficient. This dynamic response depends on the mass offluid in the tubing, the compliance of the tubing, and friction to fluid movement inthe tubing. Clinical CTT systems are designed with short, thin, noncompliant tubingwith the goal of obtaining an optimal dynamic response. The anesthesiologist candetermine these values for a particular CTT system by performing a flush test. Theresulting oscillating waveform is then analyzed to obtain the natural frequency of oscil-lation and the amplitude ratio of the first 2 oscillation peaks (for details see Mark13

[1998]). Next, a plot of natural frequency versus damping coefficient (Fig. 4) is refer-enced to determine whether the dynamic response is adequate to allow proper pro-cessing of the physiologic waveform.11 If the dynamic response is not adequate,the measured waveform may be distorted. The fallacy in attempting to determinedynamic response by casual observation of the waveform was shown by Gardner11

(1981) on analysis of waveforms from 37 patients. He showed that it is possible tohave a waveform that appears to be underdamped when its dynamic response isdetermined to be overdamped, and vice versa.11

Introduction of a bubble into the tubing or the use of a resonance overshoot elimi-nator (ROSE) damping device (Spectramed, Oxnard, CA) are strategies commonlyused to correct an underdamped CTT system. However, introduction of an air bubbleincreases damping and simultaneously decreases natural frequency,11,15 which maydo little to improve, or may even worsen, the dynamic response of an underdampedsystem (see a in Fig. 4). Furthermore, an optimal dynamic response can inadvertentlybe made suboptimal (see b in Fig. 4), increasing the likelihood that the SBP measure-ment will be inaccurate. The ROSE damping device has been shown to increase the

Fig. 4. How natural frequency, damping coefficient, and patient physiology interact todefine optimum dynamic response requirement. As either heart rate or the slope of thesystolic upstroke (dP/dt) increases, the natural frequency necessary to allow proper process-ing of the arterial waveform increases and the range of damping coefficient associated withoptimum dynamic response decreases (compare vertical arrow bars). Presence of a bubbleincreases damping coefficient and decreases natural frequency (a and b).

Page 8: The Environmental Services Grou p

McNeer & Varon8

damping coefficient while having little effect on natural frequency, and therefore mayimprove measurement accuracy of an underdamped system. However, resonancemay still occur at high heart rates (a common occurrence in patients with trauma)and with steep systolic upstroke.15

In the case in which dynamic response is suboptimal because of low naturalfrequency, the recorded waveform may become distorted because of resonanceand appear to be underdamped. A recorded waveform appears to resonate whenthe frequencies of sine waves that compose the physiologic waveform approachthe natural frequency of the CTT system. The effect is increased in an underdampedsystem. A simple demonstration helps to explain the interaction between the naturalfrequency of the CTT system and the sine waves that make up the arterial waveform.Fig. 5 shows the effect of resonance on the simulated arterial waveform constructedfrom 3 sine waves from Fig. 1. The frequencies of these sine waves increase withheart rate. For example, at a heart rate of 60 beats per minute (bpm), sine waves 1,2, and 3, have frequencies of 1, 2, and 4 Hz, respectively. These frequencies areincreased to 2.5, 5.0, and 10 Hz, respectively, when the heart rate is 150 bpm. Atthe higher heart rate, sine waves 2 and 3 resonate in a CTT system with naturalfrequencies of 5 (see Fig. 5A) and 10 Hz (see Fig. 5B), respectively. The effect of reso-nance is to distort the recorded waveform such that SBP is overestimated by 10 to 30mmHg and the diastolic pressure is underestimated, whereas the effect on MAP is notsignificant.11

The dynamic response plot published by Gardner11 (1981) (see Fig. 4) was derivedfrom analysis of only 2 physiologic waveforms. One waveform was thought to repre-sent the general patient population and had a heart rate of 94 bpm. The other wave-form (118 bpm) was chosen because it required the greatest dynamic responseattributed to its steep systolic upstroke. Although there are no experiments thataddress this in the literature, it is feasible that physiologic waveforms obtained frompatients with trauma in hemorrhagic shock with heart rates of 150 bpm or more andincreased ventricular contractility may exceed the capability of the CTT system. Futureresearch may help to determine whether this is the case and, if so, consideration forthe design of CTT systems with natural frequencies more relevant to the traumasetting may be warranted. This possibility is especially germane when consideringthe potential role of hypotensive resuscitation in the management of patients withtrauma. In the meantime, anesthesiologists should rely more on monitoring MAPthan SBP, because MAP measurements are the least affected by the dynamicresponse characteristics of the CTT system.Current CTT systems are claimed to have natural frequency between 10 and 20 Hz.5

At our institution, we have taken steps to optimize the dynamic response of our CTTsystems. We now perform flush tests on the CTT system when we use invasive moni-toring. We have found that the natural frequency is usually between 10 and 20 Hz, asadvertised (Fig. 6). When extension tubing is used (eg, dorsalis pedis site), the naturalfrequency is reduced about 50% (on occasion, we have observed natural frequenciesof less than 5 Hz). Based on our observations, we now instruct residents not to useextension tubing, but instead to place the CTT transducer close to the catheter site.However, in considering that the pulse waveform of unstable patients with traumamay demand higher dynamic response than is advertised on out-of-the-box CTTsystems, we have changed our practice further by removing any unnecessary tubingbetween the transducer and the catheter insertion site. By performing this modifica-tion, the natural frequency is usually doubled (see Fig. 6). We are currently investi-gating whether these modifications improve the fidelity of recorded waveforms inthe setting of hemorrhagic shock.

Page 9: The Environmental Services Grou p

Fig. 5. The effect of resonance on arterial waveform morphology. Resonance occurs whena harmonic frequency of the physiologic waveform approaches the natural frequency ofa particular CTT system. Two examples are shown using the first (A) and second (B) harmonicsfrom Fig. 1. In both cases, amplification of the harmonic wave is observed in the recordedwaveform, and SBPs and DBPs are overestimated and underestimated, respectively.

Pitfalls of Hemodynamic Monitoring in Trauma 9

NIBP Monitoring

NIBP monitoring is likely to be relied on to manage the patient with acute trauma inprehospital and emergency room settings. In the operating room, NIBP can be valu-able in assessing hemodynamic status and effects of fluid and pharmacologic treat-ments until invasive monitoring is established. NIBP is easy to implement as long asan extremity is available for cuff placement. It is common for a patient with traumato sustain injury in both upper extremities and, in such cases, cuff placement at thethigh or ankle provides MAP values that correlate well with values obtained at the

Page 10: The Environmental Services Grou p

Fig. 6. The effect of tube length on CTT system natural frequency. At our institution, naturalfrequency is typically between 10 and 20 Hz (ORIGINAL). The effect of adding extensiontubing (LENGTHENED) or removing unnecessary tubing (SHORTENED) can significantlyaffect CTT natural frequency (unpublished data, McNeer RR, 2013).

McNeer & Varon10

arm, although measurements with thigh cuff placement take the longest tocomplete.16 NIBP has been reported as an effective means of tracking blood pressurechanges and hypotension in unstable patients.17 However, it is usually desirable toestablish an arterial line as soon as other procedures have been performed, such assecuring the airway and obtaining large-bore intravenous access. Limitations to theuse of NIBP include placing an inappropriately sized cuff,18,19 presence of dysrhyth-mias,20 and patient movement or movement associated with the surgical procedure(including surgeons). However, more relevant for the patient with trauma is the limita-tion of obtaining accurate blood pressure values during hypotension.21

Most intraoperative NIBP measurements are performed with automated devices,which use the principle of oscillometry.20 In this technique, the amplitude of oscillationsin an air-filled occlusive cuff is measured at cuff pressures that start at more than SBPand extend to less than DBP. A plot of oscillation amplitude versus cuff pressure gener-ates the oscillation amplitude envelope (Fig. 7). The maximum amplitude algorithm(MAA) is used bymost devices to estimateMAP,which is considered to be the pressurein the cuff at maximum oscillation amplitude.22 The SBP andDBP are calculated basedon the shape of the amplitude envelope. However, the algorithm used is proprietary,not standardized, and varies with manufacturer.18,20 Clinical studies have shown thatNIBP correlates well with IABP for pressure values in the normal range, whereas ithas been shown to be less accurate at extremes of blood pressure. When comparedwith sphygmomanometry, oscillometric SBP values between 90 and 110 mm Hg

Page 11: The Environmental Services Grou p

Fig. 7. The maximum amplitude algorithm for measuring MAP. The pressure in the occlusivecuff at maximum oscillation amplitude is considered to be the MAP (solid line curve). Factorssuch as arterial waveform morphology and pulse rate can affect the shape of the amplitudeenvelope such that the pressure corresponding with maximum oscillation amplitude is noteasy to identify.31 For example, the ambiguous amplitude envelope illustrated by thedashed curve has been observed in hypertensive patients, where the maximum amplitudecannot be identified (?).

Pitfalls of Hemodynamic Monitoring in Trauma 11

overestimate SBP (119 � 6 mm Hg vs 103 � 4 mm Hg). The overestimation is moresignificant for SBP values less than 90 mm Hg (106 � 6 mm Hg vs 80 � 2 mm Hg)21

In contrast, oscillometric MAP has been shown to be consistent with IABP MAP forvalues less than 65 mm Hg,16 which may seem appropriate because MAP is consid-ered to be a directly measured parameter in oscillometry.Although it is commonly assumed that the MAP as measured by the MAA algorithm

is more accurate than SBP and DBP, all 3 parameters have been shown in simulationstudies to be influenced by other factors independent of MAA, such as pulse pressure,arterial pressure/volume relationship, and pulse wave morphology and heart rate.22–25

In addition, errors in estimates of SBP, DBP, and MAP have been shown in simulationstudies to depend on interactions between the arterial waveform amplitude, waveformmorphology, and arterial wall mechanical characteristics.23 Furthermore, there may betwice as much error associated with MAP than with SBP and DBP measurements.26

Another simulation study suggested that there is no direct relationship between theMAP estimate obtained with MAA and the actual MAP.22

There is no standardization of the oscillometric algorithm used by manufacturers,and most validation studies have been conducted in healthy patients with hyperten-sive and hypotensive conditions being absent or underrepresented. These factsmay add an additional layer of error to oscillometric measurements in certain patientpopulations. Group-specific differences between oscillometry and the auscultatorymethod have been reported in pregnant women,27,28 diabetics,29 and patients withdecreased arterial elasticity.30 For example, hypertensive patients have been ob-served to display an oscillation amplitude envelope with an ambiguous maximumamplitude31 (see Fig. 7) leading to significant errors in estimated MAP. Although theeffect of hypotension on the oscillation amplitude envelope has not been investigatedin the setting of hemorrhagic shock, patient physiology and device factors may explainthe unreliability of oscillometric measurements in unstable patients with trauma.Given themany factors (see Table 1) that contribute to the unreliability of oscillomet-

ric monitoring, it is sobering to acknowledge that patients being managed with a hypo-tensive resuscitation strategy are sometimes monitored with these devices. Future

Page 12: The Environmental Services Grou p

McNeer & Varon12

investigations on hypotensive resuscitation should indicate the methods used forblood pressure measurement to allow clinical interpretation of study results.

PETCO2 as a Measure of Cardiac Output

Initial hemodynamic assessment and monitoring of patients with acute trauma canbe difficult because of the frenetic and at times chaotic operating room environmentin these cases. The act of placing monitors on the patient can be challenging, espe-cially because many of the standard monitors are either inaccurate or do not functionduring low-flow circulatory and hypotensive states. Initial intraoperative hemody-namic assessment of the patient needs to be attained in a timely and efficientmanner. A simple way to assess whether a patient has cardiac output on arrival tothe operating room is to ventilate the patient and observe the end-tidal CO2 level(PETCO2).

4 Although this can be accomplished with a colorimetric CO2 detector(eg, in the emergency room or bay), capnography provides numerical values forPETCO2 and is easily established on attaching the patient to the ventilator. However,special considerations must be taken into account when using this approach in thepatient with trauma to avoid misdiagnosis or missed detection of severe hemody-namic instability.The relationship between PETCO2 and cardiac output has been investigated in

animals32–36 and humans.37–39 Cardiac output and PETCO2 have been shown to corre-late during low-flow circulatory states including cardiogenic, septic, and hemorrhagicshock.32 The relationship between PETCO2 and cardiac output has been shown to bea logarithmic curve.34 A study in patients being weaned from cardiopulmonary bypassreported that when PETCO2 was more than 30 mm Hg, the cardiac output was greaterthan 4 L/min. However, at PETCO2 levels of more than 34 mm Hg, further increases incardiac output did not correlate with comparable changes in PETCO2.

37

The correlation between PETCO2 and cardiac output in low-flow states can beexplained based on the pathophysiology of shock. As cardiac output to the lungsdecreases, less CO2 is available for transfer into alveoli, there is an increase in theWest zone 1, which contributes to dead space–like ventilation (ie, VQ mismatch),and absolute dead space secondary to pulmonary vessel collapse can occur.An additional explanation can be obtained by considering 2 fundamental expres-

sions from respiratory physiology that describe the relationship between alveolarventilation (VA), CO2 elimination (VCO2), and partial pressure of alveolar CO2 (PACO2),(VA5 VCO2/PACO2 � K), and the relationship between CO2 elimination, cardiac output,and arterial and venous CO2 content (Cvco2) or Fick principle (VCO2 5 Q � (CaCO2 �Cvco2)).

40 After assuming normal lung function and that PETCO2, PACO2, and partialpressure of arterial CO2 (PaCO2) are equal, Maslow and colleagues37 (2001) rearrangedthese expressions to obtain the following:

PETco2yA _QPPvco2

_VA1A _QP

where Qp is pulmonary blood flow (ie, cardiac output), Pvco2 is partial pressure ofmixed venous CO2, and A is CO2 solubility in blood. A limitation of this equation isthe assumption that lungs are homogeneous; however, in the setting of shock, VQmismatch and absolute dead space are likely to be present. Nonetheless, the equationis useful in showing that PETCO2 is affected by several interdependent factors.During periods of apnea or hypoventilation, Pvco2 (as well as PaCO2) increases

because of decreased or absent pulmonary elimination of CO2.40 Reestablishment

of ventilation in this circumstance reveals a PETCO2 that is normal or greater than

Page 13: The Environmental Services Grou p

Pitfalls of Hemodynamic Monitoring in Trauma 13

normal even if cardiac output is unchanged. In contrast, in cardiac arrest or lowcardiac output states Pvco2 increases because less CO2 is delivered to the lungs.In this situation, PaCO2 does not increase as much as Pvco2 and may decrease,12

whereas the Pvco2-PaCO2 difference increases.Restoration of cardiac output after cardiac arrest or low-flow state results in

increased PETCO2 level. However, in this circumstance, increased PETCO2 does notnecessarily reflect adequate cardiac output. Experimental evidence in support ofthis phenomenon was shown in a study in which hemorrhagic shock was induced insheep while mechanical ventilation rate was held constant.12 Arterial and venousCO2 were measured during baseline, hemorrhagic shock, and postresuscitationperiods. During the shock period, cardiac output decreased (from 2.3 to 0.6 L/min),resulting in significant increase in Pvco2, modest decrease in PaCO2, and decreasein PETCO2. The decrease in PETCO2 was predominantly caused by the decreasedcardiac output. After resuscitation and return of normal cardiac output, PaCO2increased to more than baseline levels and the Pvco2-PaCO2 difference decreased.Although PETCO2 was not measured in the postresuscitation period, it is possible toinfer that it would have been increased to more than the baseline, assuming that thePaCO2-PETCO2 difference remained the same between baseline and postresuscitationperiods.Applying this analysis to patients with trauma, there is the prospect that PETCO2 can

reach near-normal levels with small increases in cardiac output after a period of shockor cardiac arrest. These relationships have not yet been quantified experimentally inpatients with trauma. Nonetheless, it is prudent to consider prehospital events (ie,cardiac or respiratory arrest) that increase Pvco2 when using PETCO2 to assess hemo-dynamic status.

SUMMARY

The ultimate goal of resuscitation in the setting of uncontrolled hemorrhagic shock isto ensure adequate organ perfusion until surgical hemostasis is obtained. Tissueperfusion in this setting is usually indirectly assessed by monitoring blood pressurewith either NIBP or IABP and cardiac output by monitoring PETCO2 level. A generaltheme of this article is that factors common in the trauma setting and characteristicsof current monitoring devices can often interfere with accurate assessment of hemo-dynamic status. Overestimation of SBP as can occur with oscillometry and IABP isarguably one of the most important monitoring pitfalls to be avoided. More accurateassessment can be accomplished by considering SBP in conjunction with MAPmeasurements. However, oscillometric MAP has potential limitations in trauma, espe-cially considering that the amplitude envelope is affected by factors such as arterialwaveform shape and heart rate. Research and development need to be appliedtoward improving algorithms such that oscillometry is more reliable in patient subpop-ulations including trauma. However, the current lack of standardization is not alignedwith this goal. It may be easier to improve the reliability of IABP monitoring, becauseimproving the dynamic response of the CTT system can ensure that the arterial wave-form is properly processed and not distorted. However, the extent to which thenatural frequency should be increased to provide adequate dynamic response inthe setting of trauma is unknown. We are currently conducting studies to addressthis question.The unreliability of blood pressure monitoring is likely to be a significant confounding

variable in studies investigating the role of hypotensive resuscitation. Themost reliablymeasured blood pressure parameter is currently MAP, and the most reliable method

Page 14: The Environmental Services Grou p

McNeer & Varon14

for MAP measurement is invasively with a CTT system independent of dynamicresponse (see Table 1). The reliability of oscillometric MAP measurements is unclear,because most evidence suggesting its unreliability comes from theoretic and simula-tion studies. Nonetheless, it is prudent to control for the methods used tomonitor MAPin future studies.

REFERENCES

1. Mapstone J, Roberts I, Evans P. Fluid resuscitation strategies: a systematicreview of animal trials. J Trauma 2003;55(3):571–89.

2. Bickell WH, Wall MJ Jr, Pepe PE, et al. Immediate versus delayed fluid resuscita-tion for hypotensive patients with penetrating torso injuries. N Engl J Med 1994;331(17):1105–9.

3. Morrison CA, Carrick MM, Norman MA, et al. Hypotensive resuscitation strategyreduces transfusion requirements and severe postoperative coagulopathy intrauma patients with hemorrhagic shock: preliminary results of a randomizedcontrolled trial. J Trauma 2011;70(3):652–63.

4. McNeer RR, Varon AJ. Monitoring the trauma patient. In: Varon AJ, Smith CE,editors. Essentials of trauma anesthesia. 1st edition. New York: CambridgeUniversity Press; 2012. p. 116–29.

5. Szockik JF, Barker SJ, Tremper KK. Fundamental principles of monitoring instru-mentation. In: Miller RD, editor. Anesthesia. 5th edition. Philadelphia: ChurchillLivingstone; 2000. p. 1053.

6. O’Rourke MF, Yaginuma T. Wave reflections and the arterial pulse. Arch InternMed 1984;144(2):366–71.

7. O’Rourke MF. Vascular impedance in studies of arterial and cardiac function.Physiol Rev 1982;62(2):570–623.

8. Henneman EA, Henneman PL. Intricacies of blood pressure measurement: reex-amining the rituals. Heart Lung 1989;18(3):263–71.

9. Weber T, O’Rourke MF, Ammer M, et al. Arterial stiffness and arterial wave reflec-tions are associated with systolic and diastolic function in patients with normalejection fraction. Am J Hypertens 2008;21(11):1194–202.

10. Bridges ME, Middleton R. Direct arterial vs oscillometric monitoring of blood pres-sure: stop comparing and pick one (a decision-making algorithm). Crit CareNurse 1997;17(3):58–66, 68–72.

11. Gardner RM. Direct blood pressure measurement–dynamic response require-ments. Anesthesiology 1981;54(3):227–36.

12. Garnett AR, Glauser FL, Ornato JP. Hypercarbic arterial acidemia following resus-citation from severe hemorrhagic shock. Resuscitation 1989;17(1):55–61.

13. Mark JB. Atlas of cardiovascular monitoring. New York: Churchill Livingstone;1998.

14. Shapiro GG, Krovetz LJ. Damped and undamped frequency responses of under-damped catheter manometer systems. Am Heart J 1970;80(2):226–36.

15. Kleinman B, Powell S. Dynamic response of the ROSE damping device. J ClinMonit 1989;5(2):111–5.

16. Lakhal K, Macq C, Ehrmann S, et al. Noninvasive monitoring of blood pressure inthe critically ill: reliability according to the cuff site (arm, thigh, or ankle). Crit CareMed 2012;40(4):1207–13.

17. Lakhal K, Ehrmann S, Runge I, et al. Tracking hypotension and dynamic changesin arterial blood pressure with brachial cuff measurements. Anesth Analg 2009;109(2):494–501.

Page 15: The Environmental Services Grou p

Pitfalls of Hemodynamic Monitoring in Trauma 15

18. Ng KG, Small CF. Changes in oscillometric pulse amplitude envelope with cuffsize: implications for blood pressure measurement criteria and cuff size selection.J Biomed Eng 1993;15(4):279–82.

19. Jilek J, Stork M. Cuff width alters the amplitude envelope of wrist cuff pressurepulse waveforms. Physiol Meas 2010;31(7):N43–9.

20. Smulyan H, Safar ME. Blood pressure measurement: retrospective and prospec-tive views. Am J Hypertens 2011;24(6):628–34.

21. Davis JW, Davis IC, Bennink LD, et al. Are automated blood pressure measure-ments accurate in trauma patients? J Trauma 2003;55(5):860–3.

22. Baker PD, Westenskow DR, Kuck K. Theoretical analysis of non-invasive oscillo-metric maximum amplitude algorithm for estimating mean blood pressure. MedBiol Eng Comput 1997;35(3):271–8.

23. Forster FK, Turney D. Oscillometric determination of diastolic, mean and systolicblood pressure–a numerical model. J Biomech Eng 1986;108(4):359–64.

24. Ursino M, Cristalli C. A mathematical study of some biomechanical factorsaffecting the oscillometric blood pressure measurement. IEEE Trans BiomedEng 1996;43(8):761–78.

25. Raamat R, Talts J, Jagomagi K, et al. Mathematical modelling of non-invasive os-cillometric finger mean blood pressure measurement by maximum oscillationcriterion. Med Biol Eng Comput 1999;37(6):784–8.

26. Raamat R, Talts J, Jagomagi K, et al. Errors of oscillometric blood pres-sure measurement as predicted by simulation. Blood Press Monit 2011;16(5):238–45.

27. Arfeen ZU, Maran NJ, Simon EJ, et al. A comparison of non-invasive methods ofblood pressure measurement in normotensive and hypertensive pregnantwomen. Int J Obstet Anesth 1996;5(3):168–71.

28. Penny JA, Shennan AH, Halligan AW, et al. Blood pressure measurement insevere pre-eclampsia. Lancet 1997;349(9064):1518.

29. van Ittersum FJ, Wijering RM, Lambert J, et al. Determinants of the limits of agree-ment between the sphygmomanometer and the SpaceLabs 90207 device forblood pressure measurement in health volunteers and insulin-dependent diabeticpatients. J Hypertens 1998;16(8):1125–30.

30. van Popele NM, Bos WJ, de Beer NA, et al. Arterial stiffness as underlying mech-anism of disagreement between an oscillometric blood pressure monitor anda sphygmomanometer. Hypertension 2000;36(4):484–8.

31. SunTech Medical: the rationale for clinical monitors and validation of the oscillo-metric method. Available at: http://www.suntechmed.com/downloads/OEM/white_papers/OEM_Whitepaper_0002.pdf. Accessed October 23, 2012.

32. Jin X, Weil MH, Tang W, et al. End-tidal carbon dioxide as a noninvasive indicatorof cardiac index during circulatory shock. Crit Care Med 2000;28(7):2415–9.

33. Weil MH, Bisera J, Trevino RP, et al. Cardiac output and end-tidal carbon dioxide.Crit Care Med 1985;13(11):907–9.

34. Ornato JP, Garnett AR, Glauser FL. Relationship between cardiac output and theend-tidal carbon dioxide tension. Ann Emerg Med 1990;19(10):1104–6.

35. Isserles SA, Breen PH. Can changes in end-tidal PCO2 measure changes incardiac output? Anesth Analg 1991;73(6):808–14.

36. Dubin A, Murias G, Estenssoro E, et al. End-tidal CO2 pressure determinantsduring hemorrhagic shock. Intensive Care Med 2000;26(11):1619–23.

37. Maslow A, Stearns G, Bert A, et al. Monitoring end-tidal carbon dioxide duringweaning from cardiopulmonary bypass in patients without significant lungdisease. Anesth Analg 2001;92(2):306–13.

Page 16: The Environmental Services Grou p

McNeer & Varon16

38. Shibutani K, Muraoka M, Shirasaki S, et al. Do changes in end-tidal PCO2 quanti-tatively reflect changes in cardiac output? Anesth Analg 1994;79(5):829–33.

39. McManus JG, Ryan KL, Morton MJ, et al. Limitations of end-tidal CO2 as an earlyindicator of central hypovolemia in humans. Prehosp Emerg Care 2008;12(2):199–205.

40. West JB. Respiratory physiology: the essentials. 7th edition. Philadelphia: Lippin-cott Williams & Wilkins; 2005.