The Disposable Infusion Pump Instruction

Basal + PCA Bolus ModelBasal + PCA Bolus Model

Page 2: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Characteristics

The Disposable Infusion Pump is a safe, effective and The Disposable Infusion Pump is a safe, effective and durable micro injector for safe analgesia. It is driven by the durable micro injector for safe analgesia. It is driven by the balanced contraction of the medical silicone rubber balanced contraction of the medical silicone rubber reservoir, with the flow rate controlled by the Flow Rate reservoir, with the flow rate controlled by the Flow Rate Controlled TubingControlled Tubing. .

The pump is mainly used in the analgesia for all types of The pump is mainly used in the analgesia for all types of post-operative local pain including childbirth, in the post-operative local pain including childbirth, in the therapy of chronic pains, in the analgesia for malicious therapy of chronic pains, in the analgesia for malicious tumor as well as some other therapies that involve the self-tumor as well as some other therapies that involve the self-controlled micro injections.controlled micro injections.

Page 3: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Material

The Disposable Infusion Pump contains a silicone rubber rThe Disposable Infusion Pump contains a silicone rubber reservoir, together with PVC tubing, ABS tubing connector. eservoir, together with PVC tubing, ABS tubing connector. (Warning: certain solution may be incompatible with the m(Warning: certain solution may be incompatible with the materials used in the Disposable Infusion Pump. Before use aterials used in the Disposable Infusion Pump. Before use consult the instruction of the fluid or drug to be used.).consult the instruction of the fluid or drug to be used.).

   The Disposable Infusion Pump is assembled with a standarThe Disposable Infusion Pump is assembled with a standar

d male luerd male luer ,, which can connect with standard infusion tuwhich can connect with standard infusion tubing. bing.

Page 4: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Specification (Basal + PCA Bolus Model)

ModelCapacity of

Reservoir (ml)Basal Rate

(ml/hr)PCA Bolus

(ml/min)WZ-6523C-2 100ml 1.0ml/hr 0.5ml/30minWZ-6523C-3 100ml 1.0ml/hr 0.5ml/15minWZ-6523C-4 100ml 2.0ml/hr 0.5ml/15minWZ-6523C-4.25 100ml 0.5ml/hr 0.5ml/8minWZ-6523C-5 100ml 1.0ml/hr 1.0ml/15minWZ-6523C-6 100ml 2.0ml/hr 1.0ml/15minWZ-6523C-7 100ml 3.0ml/hr 1.0ml/15minWZ-6523C-8 100ml 4.0ml/hr 1.0ml/15minWZ-6523C-9 100ml 5.0ml/hr 1.0ml/15minWZ-6523C-10 100ml 2.0ml/hr 2.0ml/15minWZ-6523C-13 100ml 5.0ml/hr 2.0ml/15minWZ-6523C-02 275ml 1.0ml/hr 0.5ml/30minWZ-6523C-04 275ml 2.0ml/hr 0.5ml/15minWZ-6523C-05 275ml 3.0ml/hr 0.5ml/15minWZ-6523C-09 275ml 5.0ml/hr 1.0ml/15minWZ-6523C-013 275ml 5.0ml/hr 2.0ml/15minWZ-6523C-016 275ml 8.0ml/hr 2.0ml/15min

The Disposable Infusion Pump List (II)

Page 5: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Usage: Before filling the reservoir

Pick out a pump from its packaging,Pick out a pump from its packaging,

confirm that a sliding clip is fitted on the confirm that a sliding clip is fitted on the button of PCA administration set. button of PCA administration set.

Close the tubing with a clip. Close the tubing with a clip.

Fill a syringe with a volume of Fill a syringe with a volume of solution to be placed in the reservoir.solution to be placed in the reservoir.

Page 6: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Usage: Filling the reservoir

Fill in the reservoir through the filtFill in the reservoir through the filter and check valve. After filling, reer and check valve. After filling, remove the filter and close the check move the filter and close the check valve with a screw cap.valve with a screw cap.

Although about 0.5ml air bubble wAlthough about 0.5ml air bubble will be remained in the reservoir afteill be remained in the reservoir after filling, the air bubble can be remor filling, the air bubble can be removed automaticlly through the silicoved automaticlly through the silicone reservior in 1-2 hours,ne reservior in 1-2 hours, owing to owing to the character of silicone rubber matthe character of silicone rubber material.erial.

Page 7: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Usage: Quickly get rid of the air bubbles from the tubing

Remove the clip from tubingRemove the clip from tubing

Open the adding portOpen the adding port

Fit a syringe to the filter and inject 1ml Fit a syringe to the filter and inject 1ml solution into the adding port to quickly solution into the adding port to quickly eliminate the air bubbles from the eliminate the air bubbles from the tubing tubing

Do not forget to close the adding port Do not forget to close the adding port with the screw cap after the above with the screw cap after the above operation.operation.

Page 8: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Usage: After filling

Remove the sliding clip from Remove the sliding clip from the PCA administration set;the PCA administration set;

Attention: Attention: This step is very important, This step is very important,

please do not forget it. please do not forget it. Otherwise, the PCA Bolus Otherwise, the PCA Bolus will be out of control.will be out of control.

During the use, when a patient feel pain or discomfort, During the use, when a patient feel pain or discomfort, he can press the button of PCA administration set, and he can press the button of PCA administration set, and the pump will give the patient an additional dose of the pump will give the patient an additional dose of medication according to the PCA bolus rate. medication according to the PCA bolus rate.

Page 9: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Attention:Attention: Keep the reservoir and the PCA ad-Keep the reservoir and the PCA ad-

ministration set on the same level when ministration set on the same level when the infusion pump starts working. Do the infusion pump starts working. Do not hang the reservoir on a higher not hang the reservoir on a higher place. Otherwise, the flow rate will be place. Otherwise, the flow rate will be increased. increased.

The male lure at the end of the tubing The male lure at the end of the tubing may connect with epidural catheter may connect with epidural catheter connector or other standard female lure. connector or other standard female lure. Please ensure that no air bubble existing Please ensure that no air bubble existing in the tubing before connecting catheter in the tubing before connecting catheter with the male lure. with the male lure.

Page 10: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Other function: Withdrawing function

When necessary, doctors can withdraw When necessary, doctors can withdraw some solution from the reservoir with some solution from the reservoir with an extractor.an extractor.

Fit the extractor to a syringe and keep Fit the extractor to a syringe and keep that the joint is tight. Once the extractor that the joint is tight. Once the extractor is fitted to check valve, the solution will is fitted to check valve, the solution will flow into the syringe immediatelyflow into the syringe immediately..

Page 11: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

Other function: Adding dosage function

Besides Basal infusion and the PCA bolus Besides Basal infusion and the PCA bolus infusion, if the doctors want to give the patients infusion, if the doctors want to give the patients extra or higher doses, they can inject the extra or higher doses, they can inject the medication from the adding portmedication from the adding port..

Attention: Attention: 1. Please keep on pressing the button of PCA 1. Please keep on pressing the button of PCA

administration set when injecting from the administration set when injecting from the adding port, otherwise, the PCA administration adding port, otherwise, the PCA administration set will be damaged.set will be damaged.

2. Do not forget to use a filter when injecting2. Do not forget to use a filter when injecting 3. Do not forget to close the adding port with 3. Do not forget to close the adding port with

screw cap after injection.screw cap after injection.

Page 12: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

The infusion pump is sterilized by E.O. gas and it is The infusion pump is sterilized by E.O. gas and it is valid for three years. Do not use after the expiration valid for three years. Do not use after the expiration date.date.

Do not use if the packaging has been previously opened Do not use if the packaging has been previously opened or damaged. Single use only. or damaged. Single use only.

Read the instruction carefully before use it.Read the instruction carefully before use it. When operating it an aseptic technique should be When operating it an aseptic technique should be

maintained. maintained. Dilute the solution as required before filling it into the Dilute the solution as required before filling it into the

Reservoir.Reservoir. Do not overfill the reservoir. Do not overfill the reservoir.

Attention

Page 13: The Disposable Infusion Pump Instruction Basal + PCA Bolus Model

The accuracy of flow rate control is within The accuracy of flow rate control is within ±± 10%. 10%.

The basal rate and the PCA bolus indicated on the packaging are The basal rate and the PCA bolus indicated on the packaging are determined basing on the concentration of determined basing on the concentration of 0.9% sodium 0.9% sodium chloride chloride under standard room temperature. Owing to the under standard room temperature. Owing to the different concentration of solution and temperature condition as different concentration of solution and temperature condition as well as other factors, the actual flow rate will be little different.well as other factors, the actual flow rate will be little different. Please try to use the pump under the temperature around 21-Please try to use the pump under the temperature around 21-2525°C.°C.

The doctors should prescribe the dosage and use proper models The doctors should prescribe the dosage and use proper models of the infusion pump for different patients. of the infusion pump for different patients.

When patients feel pain or discomfort, please press the button of When patients feel pain or discomfort, please press the button of PCA administration set to manage additional doses. However, PCA administration set to manage additional doses. However, please don't press the button if they are feeling too sleepy. please don't press the button if they are feeling too sleepy.