The Design of Barrel Electromagnetic Calorimeter (BEMC)
October 13 2001
---Lu Jun-guang
Barrel Electromagnetic Calorimeter (BEMC)
1. General Consideration
2. Choice of Calorimeter Type
3. How to arrange L3 BGO crystals
4. Structure of BEMC
5. Readout
6. Calibration and monitoring
7. MC Simulation
8. Expected Performance
9. R&D plan, schedule
10. Summary
Excellent photon energy and angular resolution,
coupled with good low energy detection efficiency
is essential to the physics of the BESIII
General physics requirements• Energy response region:
10 MeV to 2 GeV. The key energy region< 500MeV
• Reconstruction of π0 and η• Contributes to e/πand e/μ separation• Provide neutral energy trigger
1. General Consideration
2. Choice of Calorimeter Type
• Comparison of the scintillating fiber calorimeter and BGO crystal calorimeter options, L3-BGO crystals have been selected
to use as the BESIII calorimeter:
☆ a very big benefit for energy and angular resolution at BGO crystal option.
• Energy resolution: % /√E (GeV)
2.5 (BGO) 6 (scint_fiber)
• Spatial resolution : σ,~ /√E (GeV)
0.3cm (BGO) 1cm (scint_fiber)
3.How to use L3 BGO crystals
* L3 BGO crystals
Rin(barrel) ~52 cm, Lin(barrel) ~100 cm
7680 (barrel) + 3072 (end-cap)
2cm x 2cm, 2.6x2.6~2.9x2.9 cm2
Crystal length : 24 cm ( 22X0 )
Density 7.13 g/cm3
Radiation length 1.12 cm
dE/dX 9 MeV/cm
dE/E 2%/√E(GeV)
a. Cut BGO crystals (10752) 22 X0: ~13X0 (14cm) + 8.5 X0(9.5cm)
making new piece by gluing: 9.5cm + 4.5 cm=14cm 8800 (front face: ~2.42cm2 ) / L: 14 cm 1952 (front face: ~2.02cm2 ) / L: 14 cm 8800 (front face: ~2.02cm2 ) / L: 9.5 cm 1952 (front face: ~2.42cm2 ) / L: 9.5 cm 2x 1952=3904 / L: 4.5 cm 2x192=384 / L: 4.5 cm
b. produce new BGO crystal piece: 4512/L: 4.5cm ~130 kcc total crystals: 8800 (~2.42cm2)+10560(~2.02cm2) =19360
4. Structure of BEMC
* to rebuilt a barrel EMC: Rin ~77cm , Lin ~194cm cosθ=0.78
44 rings (240 piece: front face~2.02cm2)
44 rings (200 piece: front face~2.42cm2)
10560(~2.02cm2) + 8800 (~2.42cm2)=19360
* grinding and polishing of the BGO crystals Insert one newly machined crystal for every two old crystals in both θ and φ
5. Readout
• Two photodiodes per crystal in L3
Hamamatsu PD-S2662, Ssens=20x7.5mm2
Total PDs:
2x18160=36320 piece (S2662)
1200 piece (S3950/Ssens=10x10mm2)
• Two charge amplifier and one main amplifier
Total electronics: 19360 chs.
6. Calibration and monitoring
• Different response of crystals due to:
*different crystal gluing, shape, quality
*different PD and its coupling
*different radiation damage• Several effects can change the response of the c
alorimeter:
radiation damage, temperature variations,
optical changes of the crystal-PD glue joint,
variations of electronics.
BGO crystal scanning using γ source Cs137 (662 KeV) For the glued crystal, light output reduced by ~25%, energy resolution worsened by ~30%
BGO crystal scanning using γ source Cs137 (662 KeV) For the glued crystal, light output reduced by ~25%, energy resolution worsened by ~30%
γ source Cs137 (662 KeV) at glue section , crystal:4cm+7cm
γ source Cs137 (662 KeV) at glue section , crystal:4cm+7cm
• Calibration methods *single crystal (before assembly) # optical testing
# radiation source scan
# cosmic rays testing (after gluing PD)
*whole calorimeter(after assembly) # try to use a γ ( >4 MeV) source
# test beam
# cosmic rays testing
*In-situ Calibration
# light pulser (Xenon -optical fiber) monitoring
# temperature monitoring
# electronic calibration
# physics processes (e+e-→ββ,γγ,μμ, γββ, and πo )
7. MC Simulation
* Energy resolution: 14cm length (~13X0)
Eγ : 0.1~1.5 GeV, The fluctuation of deposited
energy (σ) is from 10% to 3%.
* Energy leakage: 1~5%
* Transverse spread of shower
Eγ=200MeV, shown the transversal spread
is within 5 pieces of BGO crystals.
* The deposited energy of π and μ
* The deposited energy of π and e
e and π+,π- at 1 GeV e and π+,π- at 1 GeV
8. Expected Performance
• Energy resolution: 2.5~3%/√E(GeV) • Spatial resolution:
,≤ 3 mm/√E(GeV) • 2γ angular resolution: 2 = 50 mr Mass of 0 →γγ: 0 = 6 MeV• Minimum energy of the detectable :
Eγ=10 MeV
• Solid angle: 73%×4π
9. R&D and schedule
• Machining of the BGO crystals
* BGO crystal arrangement
* how to cut, grind and polish crystals• Mechanical support of crystals
* one or several crystals for each cell
* material of the wall: carbon fiber or Al?• Hardware systems for calibration and monitoring
* xenon light pulser system
* temperature monitoring
* γ source testing
* Test beam
Schedule of Project 2001/6——2002/12: shipping and R&D2003/1——2003/12: machining and crystal testing
crystal cutting, grinding, polishing, gluing and painting
* measuring properties of the BGO crystal* manufacture of mechanical support structure * production of the preamplifier
2004/1——2004/12: * crystals machining continue* crystal assembly and testing with readout modules
2005/1—2005/6
*inserting the crystal elements into the support system
* testing the whole crystal calorimeter
*assembling the Calorimeter to BES spectrometerⅢ 2005/6: Testing and tuning the BES spectrometerⅢ
• Preliminary Cost Estimate
L3 BGO crystals 25.50 M ¥
Supplemental new BGO crystals ~130 kcm3 11.00 M ¥
Silicon photodiodes 20000 pieces 3.40 M ¥
BGO crystal cutting, grinding, polishing 3.00 M ¥
Structure of the crystal elements 1.00 M ¥
Monitor system (xenon –light and temperature) 1.20 M ¥
Support product and assembly equipment 5.00 M ¥
Measurement equipment
(normal and cosmic ray test system) 1.20 M ¥
Total 51.30 M ¥
10. Summary• A basic design of BEMC is to use L3 BGO crystals
after cutting, grinding and polishing, with nearly 13X0 in length
• Building BEMC with a size: R~77cm, L~ 194cm• Readout: adopt two PD S2662 in each crystal,total chan
nels: 19360 • Single crystal calibration will adopt γ source and Xenon flusher for monitoring• MC: E/E ≤ 3%/√E, Mπ0 ~ 6 MeV • Expected performance:
E/E ≤ 3%/√E , , ≤ 3mm/√E