Su M Tu W Th F Sa Day July 2020 Days Hours 28 29 30 1 2 H 4 6 5 6 7 8 9 10 11 31 11 13 14 15 16 17 18 11.5-Month Staff Returns to Work 07/16/2020 19 20 21 22 23 24 25 11 -Month Staff Returns to Work 08/03/2020 26 27 28 29 30 31 Session 1-July 6-31 Credit Recovery/Missouri & US Constitution/CPR 9th -12th 1 August 2020 Days Hours 2 3 4 5 6 7 8 10.5-Month Staff Returns to Work 08/10/2020 9 10 11 12 13 14 15 10 - Month Staff Returns to Work 08/17/2020 16 17 18 19 20 21 22 3 Session 2 - Literacy Aug. 3-14 23 24 25 26 27 28 29 14 Session 3 - Academic Gear-Up Aug. 3-14 PK4, 6th, 7th, 8th 30 31 17 Site-Based PD 18-19 District PD 20-21 Classroom Prep 31 September 2020 Days Hours 1 2 3 4 5 7 Holiday–Labor Day; District Closed 6 H 8 9 10 11 12 18 Interim Progress Reports for Quarter 1 13 14 15 16 17 18 19 25 Site-Based PD - 1/2 day 20 20 21 22 23 24 25 26 27 28 29 30 1 2 3 October 2020 Days Hours 4 5 6 7 8 9 10 16 Quarter 1 ends 11 12 13 14 15 16 17 19 Second Quarter begins 18 SQ 20/PTC 21/PTC 22 23 24 20-21 Parent-Teacher Conferences 25 26 27 28 29 30 31 22 1 2 3 4 5 6 7 November 2020 Days Hours 8 9 10 H 12 13 14 3 Election Day - Site-based PD; no students 15 16 17 18 19 20 21 11 Veteran's Day 22 23 24 NS* H H 28 13 Interim Progress Reports for Quarter 2 29 30 25 *No School for non 12-month staff /Schools Closed 26-27 Holiday–Thanksgiving; District Closed 16 1 2 3 4 5 December 2020 Days Hours 6 7 8 9 10 11 12 18 Record Keeping 13 14 15 16 17 18 19 21 - 31 Winter Break - no school, school staff only; 12-month works 20 WB WB WB H H 26 24-25 Holiday - District Closed 13 27 WB WB WB WB H 2 January 2021 Days Hours 3 TQ 5 6 7 8 9 1 Holiday–New Year's Day; District Closed 10 11 12 13 14 15 16 4 Classes resume; third quarter begins 17 H 19 20 21 22 23 15 Site-Based PD 24 25 26 27 28 29 30 18 Holiday–Dr. Martin Luther King, Jr. Day; District Closed 18 31 1 2 3 4 5 6 February 2021 Days Hours 7 8 9 10 11 12 13 12 Interim Progress Reports for Quarter 3 14 H 16 17 18 19 20 12 Site-Based PD/District Based PD 21 22 23 24 25 26 27 15 Holiday–Presidents’ Day; District Closed 18 28 1 2 3 4 5 6 March 2021 Days Hours 7 8 9 10 11 12 13 12 Quarter 3 ends 14 FQ 16/PTC 17/PTC 18 19 20 15 Fourth Quarter Begins 21 SB SB SB SB SB 27 16-17 Parent-Teacher Conferences 17 28 29 30 31 19 Site-Based PD 22-26 Spring Break 1 2 3 April 2021 Days Hours 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 22 25 26 27 28 29 30 1 May 2021 Days Hours 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 31 Memorial Day 23 24 25 26 27 28 29 30 H 20 1 2 3 4 5 June 2021 Days Hours 6 7 8 9 10 11 12 3 Last Day of School for Students 13 14 15 16 17 18 19 4 Record Keeping 3 20 21 22 23 24 25 26 7 Summer School PD 27 28 29 30 8 9 1 2 3 July 2021 Days Hours 4 H 6 7 8 9 10 2 Last Day of Summer School 11 12 13 14 15 16 17 4 Holiday–Independence Day; District Closed 18 19 20 21 22 23 24 25 26 27 28 29 30 31 175 8 Professional Development 175 Student Days 2 Classroom Prep 210 Teacher Days 2 Record Days Last Day for 11.5-Month Employees 06/30/2021 Classroom Preparation Last Day for 10-Month Employees 06/04/2021 Last Day for 10.5-Month Employees 06/11/2021 Last Day for 11-Month Employees 06/18/2021 First Day of Summer School Saint Louis Public School District 2020-2021 Academic Year Calendar 08/31/2020 through 06/03/2021 - FINAL (updated 8/5/20) 1st day 2020 Summer School (Session 1) Last day 2020 Summer School (Session 1) First Class Day for All Students 6

The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

Technical Overview

AuthorsMelanie Metzlaff and Patric Hörth Agilent Technologies, Inc.

AbstractThe Agilent 1260 Infinity II Flexible Pump provides the highest flexibility for running conventional applications on 4.6 mm columns or fast runs on 2.1 and 3.0 mm id columns with no compromise regarding precision and accuracy.

This Technical Overview demonstrates the outstanding performance of the 1260 Infinity II Flexible Pump. Precision of retention times under demanding conditions was evaluated and compared to the Agilent 1260 Infinity II Binary Pump. In addition, the composition accuracy, delay volume, and flow accuracy of both pumps were evaluated by running step gradients.

The Agilent 1260 Infinity II Flexible Pump

The Benchmark for High Accuracy and Precision in Low Pressure Mixing Gradient Pumps

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0Time (min)

0

100

200

300

400

500

600

700

800

mAU

Page 2: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

2

IntroductionThe Agilent 1260 Infinity II Flexible Pump is a low-pressure mixing gradient pump, and is capable of 800 bar operating pressure. The integrated multipurpose valve enables convenient features such as automatic purge, switching an additional mixer in and out, and filter backflush.

To demonstrate the superior performance of the 1260 Infinity II Flexible Pump, different demanding applications were chosen. Results obtained with the 1260 Infinity II Flexible Pump were compared to the results obtained with the Agilent 1260 Infinity II Binary Pump.

Experimental

InstrumentationFor the experiments, the following modules were used:

• Agilent 1260 Infinity II Flexible Pump (G7104C)

• Agilent 1260 Infinity II Binary Pump (G7112B)

• Agilent 1260 Infinity II Multisampler (G7167A), equipped with an integrated sample cooler (option #100)

• Agilent 1260 Infinity II MCT (G7116A)

• Agilent 1260 Infinity II Diode Array Detector WR (G7115A), equipped with a standard flow cell: 10 mm path length,13 µL volume

SoftwareAgilent OpenLAB CDS 2.2 (M8413A)

ChemicalsAll solvents were LC grade. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with a 0.22 µm membrane point-of-use cartridge (Millipak). Acetonitrile and acetone were purchased from

Merck (Darmstadt, Germany). Trifluoroacetic acid (TFA) and formic acid (FA) were purchased from Sigma-Aldrich (Steinheim, Germany).

Methods

Table 1. Chromatographic conditions for step gradients.

Parameter Value

Column Restriction capillary

Solvent A) 0.3 % Acetone in water B) Water

Gradient

Between 20 and 45 minutes in 1 % steps from 100 to 94 %A Between 45 and 55 minutes in 2 % steps from 94 to 90 %A Between 55 and 80 minutes in 10 % steps from 90 to 51 %A Between 80 and 90 minutes in 1 % steps from 51 to 49 %A Between 90.01 and 110 minutes in 10 % steps from 40 to 10 % Between 110.01 and 135 minutes in 2 % steps from 6 to 0 %A

Stop time 140 minutes

Flow rate 1 mL/min

Column temperature 30 °C

DAD 273/4 nm, Ref. 360/100 nm, 10 Hz

Table 2. Chromatographic conditions for shallow gradient runs.

Parameter Value

Compounds Polyphenolic mix: Catechin, caffeic acid, epicatechin, epigallocatechin (all purchased from Sigma-Aldrich)

Column Agilent ZORBAX StableBond C18, 2.1 × 50 mm, 1.8 µm (p/n 857700-902)

Solvent A) 0.1 % FA in water B) 0.1 % FA in acetonitrile

Gradient 0 minutes–6 %B 8 minutes–12 %B

Stop time 8 minutes

Post time 3 minutes

Flow rate 0.6 mL/min

Injection volume 2 µL with 3 seconds needle wash (60 % acetonitrile in water)

Sample temperature 10 °C

Column temperature 30 °C

DAD 280/4 nm, Ref. 360/50 nm, 40 Hz

Table 3. Chromatographic conditions for low organic percentage runs.

Parameter Value

Compounds Sulfa drugs: Sulfanilamid, sulfathiazole, sulfamerazine, sulfachloropyridazine, sulfamethazine (all purchased from Sigma-Aldrich)

Column Agilent ZORBAX StableBond C18, 2.1 × 50 mm, 1.8 µm (p/n 857700-902)

Solvent A) 0.1 % TFA in water B) 0.1 % TFA in acetonitrile

Gradient 0 minutes–1 %B 6 minutes–25 %B

Stop time 8 minutes

Post time 3 minutes

Flow rate 0.6 mL/min

Injection volume 2 µL with 3 seconds needle wash (60 % acetonitrile in water)

Sample temperature 10 °C

Column temperature 45 °C

DAD 254/4 nm, Ref. 360/80 nm, 20 Hz

Page 3: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

3

Results and Discussion

Experiment 1: Step gradientStep gradients, including a tracer in solvent A, were used to determine the system delay volume, the composition accuracy, mixing noise, and flow accuracy. For the flow accuracy evaluation, a flow accuracy meter (FAM) was used.

Step gradients under identical conditions were applied to the 1260 Infinity II Flexible Pump, with and without the Jet Weaver mixer, and to the 1260 Infinity II Binary Pump. Figure 1 shows an overlay of the three step gradients running from 100 %A (water + tracer) to 0 %A (water + tracer) in 1, 2, and 10 % steps. Both pumps showed excellent performance over the complete gradient range. The gradient steps were perfectly resolved even for the 1 % steps at low % solvent B. These results clearly demonstrate an equivalent gradient performance of the 1260 Infinity II Flexible Pump compared to the 1260 Infinity II Binary Pump, even for demanding ranges like 1 to 5 %B.

Table 4. Chromatographic conditions for high organic percentage runs.

Parameter Value

Compounds Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich)

Column Agilent ZORBAX Eclipse Plus C18, 3.0 × 50 mm, 1.8 µm (p/n 959941-302)

Solvent A) Water B) 95 % Acetonitrile in water

Gradient 0 minutes–90 %B 4 minutes – 100 %B

Stop time 4 minutes

Post time 2 minutes

Flow rate 1 mL/min

Injection volume 1 µL with 3 seconds needle wash (acetonitrile)

Sample temperature 10 °C

Column temperature 40 °C

DAD 490/20 nm, no Ref., 20 Hz

Table 5. Chromatographic conditions for fast gradient runs.

Parameter Value

Compounds Agilent RRLC checkout sample (p/n 5188-6529)

Column Agilent ZORBAX Eclipse Plus C18, 3.0 × 50 mm, 1.8 µm (p/n 959941-302)

Solvent A) Water B) Acetonitrile

Gradient 0 minutes–30 %B 1 minutes–100 %B

Stop time 2 minutes

Post time 1 minutes

Flow rate 2 mL/min

Injection volume 1 µL with 3 seconds needle wash (60 % acetonitrile in water)

Sample temperature 10 °C

Column temperature 60 °C

DAD 254/4 nm, Ref. 360/60, 80 Hz

Figure 1. Step gradients from 100 to 0 % tracer on an Agilent 1260 Infinity II Flexible Pump with and without a mixer, and on an Agilent 1260 Infinity II Binary Pump.

min20 40 60 80 100 120

Norm.

-100

0

100

200

300

400

500

600

Agilent 1260 Infinity II Flexible Pump with the 380 µL Jet Weaver Mixer Agilent 1260 Infinity II Flexible Pump without the mixerAgilent 1260 Infinity II Binary Pump

min20 25 30 35 40 45

Norm.

565

570

575

580

585

Page 4: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

4

Composition accuracy was calculated for both pumps based on the step gradients. Figure 2 shows the composition accuracy for the 1260 Infinity II Flexible Pump (with and without the 380 µL Jet Weaver mixer) and for the 1260 Infinity II Binary Pump. The specifications for composition accuracy were defined as ±0.35 % for the 1260 Infinity II Binary Pump and ±0.4 % for the 1260 Infinity II Flexible Pump. The composition accuracy for both pumps was inside the specified range.

Table 7 summarizes the system delay volume, mixing noise, and flow accuracy. As expected, the 1260 Infinity II Flexible Pump without the Jet Weaver mixer had the lowest delay volume, approximately 310 µL. Switching the 380 µL Jet Weaver mixer into the flowpath increased the delay volume to only 460 μL. For the 1260 Infinity II Binary Pump, a delay volume of 610 µL was calculated. All pump delay volumes include the volume of the capillaries, restriction capillary, and detector cell. The 1260 Infinity II Multisampler was not part of the flowpath for these measurements.

The mixing noise was calculated for every step over the complete gradient. Table 7 compares the values at three different steps: 95 %, 49 %, and 4 % tracer. The 1260 Infinity II Flexible Pump and the 1260 Infinity II Binary Pump showed similar results. However, the 1260 Infinity II Flexible Pump with the 380 µL Jet Weaver mixer provided the best results regarding the mixing noise. With the additional mixer, the delay volume of the 1260 Infinity II Flexible Pump increased slightly, but facilitates

Com

posi

tion

accu

racy

(%)

Setpoint solvent B (%)

Agilent 1260 Infinity II Flexible Pump with a

380 µL Jet Weaver mixerAgilent 1260 Infinity II

Flexible Pump without a mixerAgilent 1260 Infinity II

Binary Pump

+0.35 %

-0.35 %

0 20 40 60 80 100 0 20 40 60 80 100 0 20 40 60 80 100

-1.0

-0.8

-0.6

-0.4

-0.2

0

0.2

0.4

0.6

0.8

1.0

Figure 2. Composition accuracy for the Agilent 1260 Infinity II Flexible Pump with and without a mixer, and for the Agilent 1260 Infinity II Binary Pump.

Table 7. Step gradients results.

Agilent 1260 Infinity II Flexible Pump with the 380 µL

Jet Weaver mixerAgilent 1260 Infinity II

Flexible PumpAgilent 1260 Infinity II

Binary Pump

Delay volume ~460 µL ~310 µL ~610 µL

Mixing noise at 95 %A 0.015 % 0.048 % 0.023 %

Mixing noise at 49 %A 0.024 % 0.040 % 0.045 %

Mixing noise at 4 %A 0.013 % 0.034 % 0.012 %

Mean flow rate (mL/min) 0.998 0.998 1.003

Mean ∆ flow rate (%) -0.164 -0.155 0.270

a more stable UV signal due to a better mixing performance. These results clearly demonstrate the outstanding performance of the Jet Weaver mixer with efficient mixing and smart design to keep the delay volume as low as possible.

By adding a FAM to the LC stack configuration, the actual flow over the step gradient runs was monitored.

For the 1260 Infinity II Flexible Pump, the real flow rate (0.998 mL/min) was slightly lower than the programmed flow rate of 1 mL/min. The real flow rate for the 1260 Infinity II Binary Pump showed slightly higher values with 1.003 mL/min. However, both pumps performed excellently, and were inside the specification range of ±1 %.

Page 5: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

5

Experiment 2: Retention time precision for shallow gradientsTo determine the retention time precision under shallow gradient conditions, a mix of four polyphenolic compounds was analyzed. The gradient went from 6 %B to 12 %B in eight minutes, representing a change of solvent B of 0.75 % per minute. Figure 3 shows an overlay of two chromatograms, one acquired with the 1260 Infinity II Flexible Pump (without additional mixer) and the other with the 1260 Infinity II Binary Pump. The retention time precisions (Table 8) for the four polyphenolic compounds were similar on both pumps. This experiment demonstrates the high mixing precision of the 1260 Infinity II Flexible Pump even under shallow gradient conditions, where a binary pump is preferred.

Experiment 3: Retention time precision for low organic percentageA mix of five sulfa drugs was analyzed, running a gradient from 1 to 25 % organic (Figure 4). For this low organic percentage range, a binary pump is advantageous. The 1260 Infinity II Flexible Pump is specified for a composition range of 0 to 100 %. To evaluate the performance at this low organic percentage range for both pumps, the retention time precision was calculated from eight consecutive runs (Table 9).

Both pumps showed excellent performance regarding retention time precision. For the two early eluting compounds sulfanilamide and sulfathiazole, the 1260 Infinity II Flexible Pump even outperformed the 1260 Infinity II Binary Pump.

Agilent 1260 Infinity Flexible Pump

Catechin Caffeic acid Epicatechin Epigallocatechin

Average RT (min) 2.019 2.372 3.979 4.426

Standard deviation RT 0.0018 0.0015 0.0042 0.0036

RSD RT 0.087 0.061 0.106 0.081

Agilent 1260 Infinity II Binary Pump

Average RT (min) 2.348 2.710 4.927 5.491

Standard deviation RT 0.0035 0.0023 0.0032 0.0033

RSD RT 0.151 0.085 0.066 0.060

Table 8. Retention time precision for the analysis of polyphenolic compounds. Eight consecutive runs were used for the calculation.

Time (min)

mAU

Agilent 1260 Infinity II Flexible Pump

Agilent 1260 Infinity II Binary Pump

Figure 3. Analysis of polyphenolic compounds using a shallow gradient. Elution order: catechin, caffeic acid, epicatechin, then epigallocatechin.

Figure 4. Analysis of sulfa drugs at low organic percentage. Elution order: sulfanilamide, sulfathiazole, sulfamerazine, sulfachloropyridazine, then sulfamethazine.

Time (min)

mAU

Agilent 1260 Infinity II Flexible Pump

Agilent 1260 Infinity II Binary Pump

Page 6: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

6

Experiment 4: Retention time precision for high organic percentageFor this experiment, Sudan Red 1, 2, 3, and 4 were analyzed using a gradient from 90 to 100 % organic (Figure 5 and Table 10). The results show that the 1260 Infinity II Flexible Pump is well suited for this kind of application, and performs better than the 1260 Infinity II Binary Pump.

Table 9. Retention time precision for the analysis of sulfa drugs.

Agilent 1260 Infinity Flexible Pump

Sulfanilamid Sulfathiazole Sulfamerazine Sulfachloropyridazine Sulfamethazine

Average RT (min) 0.505 2.926 3.607 3.935 4.586

Standard deviation RT

0.0005 0.0009 0.0012 0.0013 0.0005

RSD RT 0.092 0.032 0.033 0.033 0.010

Agilent 1260 Infinity II Binary Pump

Average RT (min) 0.499 3.468 4.331 4.671 5.355

Standard deviation RT

0.0009 0.0044 0.0022 0.0022 0.0024

RSD RT 0.174 0.127 0.050 0.047 0.045

Figure 5. Analysis of a mix of Sudan Red compounds at high organic percentage. Elution order: Sudan Red 1, Sudan Red 2, Sudan Red 3, then Sudan Red 4.

Time (min)

mAU

Agilent 1260 Infinity II Flexible Pump

Agilent 1260 Infinity II Binary Pump

Table 10. Retention time precision for the analysis of Sudan Red compounds. Eight consecutive runs were used for the calculation.

Agilent 1260 Infinity Flexible Pump

Sudan I Sudan II Sudan III Sudan IV

Average RT (min) 0.643 1.085 1.497 2.656

Standard deviation RT 0.0004 0.0008 0.0007 0.0014

RSD RT 0.055 0.070 0.047 0.053

Agilent 1260 Infinity II Binary Pump

Average RT (min) 0.630 1.058 1.468 2.640

Standard deviation RT 0.0009 0.0018 0.0027 0.0044

RSD RT 0.147 0.173 0.187 0.166

Page 7: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

7

Experiment 5: Retention time precision for fast runs To enable shorter runs, a high flow rate of 2 mL/min was applied. The gradient time was set to one minute, and went from 30 to 100 % organic (Figure 6). For the 1260 Infinity II Binary Pump, the standard configuration and the low delay configuration were tested. For the low delay configuration, the damper and mixer were bypassed. The 1260 Infinity II Flexible Pump was used with and without the mixer. The pressure for all system configurations was approximately 540 bar.

The 1260 Infinity II Binary Pump with the low delay configuration provided the shortest run time, followed by the 1260 Infinity II Flexible with the standard configuration.

To demonstrate the retention time precision for these fast chromatographic runs (Table 11), three representative peaks were chosen. Even under these demanding conditions, the retention time precision for both pumps and their corresponding configuration showed excellent values, with <0.09 %RSD.

In an additional experiment, the flow rate was increased from 2 mL/min to 2.6 mL/min, and the resulting system pressure was approximately 750 bar. For this experiment, only the 1260 Infinity II Flexible Pump in standard configuration was used, given that this pump has an operating pressure range of up to 800 bar, whereas the 1260 Infinity II Binary Pump is limited to 600 bar.

Figure 6. Analysis of the RRLC checkout sample using a high flow rate.

Time (min)

mAU

Agilent 1260 Infinity II Binary Pump low delay configuration

Agilent 1260 Infinity II Binary Pumpstandard configuration

Agilent 1260 Infinity II Flexible Pumpstandard configuration

Agilent 1260 Infinity II Flexible Pump with the 380 µL Jet Weaver mixer

Table 11. Retention time precision for the analysis of the RRLC checkout sample. Three representative peaks were chosen. Eight consecutive runs were used for the calculation.

Agilent 1260 Infinity II Flexible Pump—standard configuration

Peak 3 Peak 6 Peak 8

Average 0.697 0.977 1.166

Standard deviation 0.0004 0.0005 0.0005

RSD 0.051 0.047 0.040

Agilent 1260 Infinity II Flexible Pump with 380 µL Jet Weaver mixer

Average 0.766 1.135 1.341

Standard deviation 0.0005 0.0004 0.0000

RSD 0.060 0.031 0.003

Agilent 1260 Infinity II Binary Pump—standard configuration

Average 0.813 1.112 1.290

Standard deviation 0.0005 0.0005 0.0006

RSD 0.066 0.048 0.050

Agilent 1260 Infinity II Binary Pump low delay configuration

Average 0.614 0.850 1.023

Standard deviation 0.0005 0.0006 0.0005

RSD 0.084 0.074 0.047

Page 8: The Agilent 1260 Infinity II Flexible Pump...Sudan red mix: Sudan I, Sudan II, Sudan III, Sudan IV (all purchased from Sigma-Aldrich) Column Agilent ZORBAX Eclipse Plus C18, 3.0 ×

www.agilent.com/chem

This information is subject to change without notice.

© Agilent Technologies, Inc. 2018 Printed in the USA, March 1, 2018 5991-9031EN

By increasing the flow rate to 2.6 mL/min, the gradient run time was reduced from 1 to 0.77 minutes (Figure 7). All peaks eluted below one minute, with an outstanding precision of retention times, below the specification of 0.15 %RSD. The last compound eluted earlier (0.967 minutes) with the 1260 Infinity II Flexible Pump than with the 1260 Infinity II Binary Pump (1.095 minutes) with a flow rate of 2 mL/min in low delay configuration.

ConclusionThe Agilent 1260 Infinity II Flexible Pump provided excellent performance regarding retention time precision for various challenging LC applications. The applied step gradients showed excellent composition accuracy and mixing noise for the 1260 Infinity II Flexible Pump and the Agilent 1260 Infinity II Binary Pump. Real-time flow rates were measured and monitored for both pumps, showing that the flow accuracy was always below 0.3 %.

The system delay volume of the 1260 Infinity II Flexible Pump is approximately 310 µL, and 460 μL with the 380 µL Jet Weaver mixer. For the 1260 Infinity II Binary Pump, the system delay volume adds up to 610 µL.

This Technical Overview shows that both pumps demonstrated excellent performance. In addition, the 1260 Infinity II Flexible Pump showed excellent results for demanding applications where a binary pump is normally advantageous.

Compound Average RT (min) RSD RT (%)

1. Acetanilide 0.171 0.103

2. Acetophenone 0.351 0.071

3. Propiophenone 0.546 0.037

4. Butyrophenone 0.668 0.021

5. Valerophenone 0.710 0.019

6. Hexanophenone 0.763 0.025

7. Heptanophenone 0.842 0.025

8. Octanophenone 0.909 0.024

9. Benzophenone 0.967 0.024

0.1

12

3 46 7

8 9

5

0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5Time (min)

×102

00.51.01.52.02.53.03.54.04.55.0

mAU

Figure 7. Overlay of eight chromatograms applying a flow rate of 2.6 mL/min at 750 bar backpressure.