NVLAP CODE 201045-0 Shenzhen Anbotek Pengcheng Compliance Laboratory Limited Page 1 of 13 ADD: Floor 1, Building C, Gold Power Industrial Park, Julongshan Grand Industrial Zone, Pingshan District, Shenzhen, Guangdong, China. TEL: (86)755-26066440 FAX: (86)755-26014772 EMAIL: [email protected] WEB: www.anbotek.com TEST REPORT IES LM-79-08 For Builders Pack. No.701, BuildingA, Shenfubao Industry Park, Futian District, Shenzhen PRC 518017. Report No.: SZANL181217002-01 Product Description: Outdoor Non-Cutoff and Semi-Cutoff Wall-mounted Area Luminaires Electrical Rating: 120-277 VAC, 60 Hz, 72 W Model No.: Model Difference: All construction and rating are the same except for CCT. Test Date: 2018-12-17 to 2018-12-20 Report Date: 2018-12-21 Test Standard: LM-79-08 Test Laboratory: Shenzhen Anbotek Pengcheng Compliance Laboratory Limited Tested by Reviewed by Dick Xiao / Helen Li / Note: This test report is prepared for the customer shown above and for the device described herein. It may not be duplicated or use in part without prior written consent from Shenzhen Anbotek Pengcheng Compliance Laboratory Limited. This report must not be used by the customer to claim product certification, approval or endorsement by NVLAP, NIST or any agency of the Federal Government. GL02-70W-30K-SNN-A GL02-70W-50K-SNN-A

TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

NVLAP CODE 201045-0

Shenzhen Anbotek Pengcheng Compliance Laboratory Limited Page 1 of 13 ADD: Floor 1, Building C, Gold Power Industrial Park, Julongshan Grand Industrial Zone, Pingshan District, Shenzhen, Guangdong, China.

TEL: (86)755-26066440 FAX: (86)755-26014772 EMAIL: [email protected] WEB: www.anbotek.com

TEST REPORT

IES LM-79-08

For

Builders Pack.

No.701, BuildingA, Shenfubao Industry Park,

Futian District, Shenzhen PRC 518017.

Report No.: SZANL181217002-01

Product Description: Outdoor Non-Cutoff and Semi-Cutoff Wall-mounted Area Luminaires

Electrical Rating: 120-277 VAC, 60 Hz, 72 W

Model No.:

Model Difference: All construction and rating are the same except for CCT.

Test Date: 2018-12-17 to 2018-12-20

Report Date: 2018-12-21

Test Standard: LM-79-08

Test Laboratory: Shenzhen Anbotek Pengcheng Compliance Laboratory Limited

Tested by Reviewed by

Dick Xiao / Helen Li /

Note: This test report is prepared for the customer shown above and for the device described herein. It may not be

duplicated or use in part without prior written consent from Shenzhen Anbotek Pengcheng Compliance Laboratory

Limited. This report must not be used by the customer to claim product certification, approval or endorsement by

NVLAP, NIST or any agency of the Federal Government.

GL02-70W-30K-SNN-AGL02-70W-50K-SNN-A

Page 2: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 2 of 13

General Information

Applicant: Builders Pack.

Applicant Address: No.701, BuildingA, Shenfubao Industry Park, Futian District, Shenzhen PRC 518017

Manufacturer: Builders Pack.

Manufacturer Address: No.701, BuildingA, Shenfubao Industry Park, Futian District, Shenzhen PRC 518017

Brand Name: N/A

Tested Model: GL02-70W-30K-SNN-AGL02-70W-50K-SNN-A

Page 3: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 3 of 13

1 Test Condition

1.1 Air Temperature

The ambient temperature in which measurements are being taken shall be maintained at 25°C±1°C, measured at a point not more than 1 m from the SSL product and at the same height as the SSL product. The temperature sensor shall be shielded from direct optical radiation from the SSL product and optical radiation from any other source. If measurements are performed at other than this recommended temperature, this is a non-standard condition and shall be noted in the test report.

1.2 Thermal Conditions for Mounting SSL Products

The method of mounting can be the primary path for heat flow away from the device and can affect measurement results significantly. The SSL product under test shall be mounted to the measuring instrument so that heat conduction through supporting objects causes negligible cooling effects. If the SSL product under test is provided with a support structure that is designated to be used as a component of the luminaire thermal management system, the product shall be tested with the support structure attached. Any such support structure included in the measurement shall be reported.

1.3 Air Movement

The incidence of air movements on the surface of a SSL product under test may substantially after electrical and photometric values. Air flow around the SSL product being tested should be such that normal convective air flow induced by device under test is not affected.

1.4 Waveshape of AC Power Supply

The AC power supply, while operating the SSL product, shall have a sinusoidal voltage wave shape at the prescribed frequency (typically 60 Hz or 50 Hz) such that the RMS summation of the harmonic components does not exceed 3 percent of the fundamental during operation of the test item.

1.5 Voltage Regulation

The voltage of an AC power supply (RMS voltage) or DC power supply (instantaneous voltage) applied to the device under test shall be regulated to within ±0.2 percent under load.

1.6 Seasoning

For the purpose of rating new SSL products, SSL products shall be tested with no seasoning.

1.7 Stabilization

Before measurements are taken, the SSL product under test shall be operated long enough to reach stabilization and temperature equilibrium. The time required for stabilization depends on the type of SSL products under test. The stabilization time typically ranges from 30 min to 2 or more hours for large SSL products.

1.8 Operating Orientation

The SSL product under test shall be evaluated in the operating orientation recommended by the manufacturer for an intended use of the SSL product. Stabilization and photometric measurements of SSL products shall be done in such operating orientation.

Page 4: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 4 of 13

2 Test Method

2.1 Integrating Sphere Measurement

The integrating sphere system includes AC power source, digital power meter, DC power supply, spectrophotometer, and integrating sphere. The system is calibrated by standard lamp before measurement weekly. The standard lamp has been calibrated regularly and traced to the National Primary Standard.

The 4π geometry was used to measure total luminous, luminous efficacy, chromaticity coordinates, correlated color temperature, and color rendering index, the spectral radiant flux measurements taken at 1 nm intervals over the range of 380 to 780 nm. The product was operated in its intended orientation and was recorded in the report.

2.2 Goniophotometer Measurement

The goniophotometer system is calibrated by standard lamp before measurement weekly. The standard lamp has been calibrated regularly and traced to National Primary Standards.

Type C goniophotometer was used for measuring total luminous flux, luminous efficacy, luminous intensity distribution, and color angular uniformity, which were calculated from the software taken at 1° vertical intervals and 22.5° horizontal intervals. The product was operated in its intended orientation and was recorded in the report.

2.3 Electrical Measurement

According to ANSI C82.77-2002, the measurement was made using a digital power meter and power supply, the SSL product under test was operated at rated voltage and stabilized enough before measurement. The total harmonic distortion of current and power factor can be calculated from the digital power meter. The digital power meter was calibrated regularly and traced to National Primary Standards.

Page 5: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 5 of 13

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor Power

(W)

119.98 60 0.5894 0.9926 70.19

Luminous Flux (lm)

Radiant Flux (W)

CCT (K)

Duv Luminous Efficacy

(lm/W)

8342.10 25.06 3117 0.0001 118.85

Ra x y u’ v’

82.5 0.4290 0.4014 0.2466 0.5192

R1 R2 R3 R4 R5

81 91 97 80 80

R6 R7 R8 R9 R10

88 84 61 9 77

78 R12 R13 R14 R15

78 66 83 99 74

3 Test Result

3.1 Integrating Sphere Test at 25°C for GL02-70W-30K-SNN-A

Page 6: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 6 of 13

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor Power

(W)

120.05 60 0.5874 0.9925 69.99

Luminous Flux (lm)

Radiant Flux (W)

CCT (K)

Duv Luminous Efficacy

(lm/W)

8657.80 26.75 5021 0.0039 123.70

Ra x y u’ v’

84.0 0.3452 0.3595 0.2085 0.4885

R1 R2 R3 R4 R5

82 92 95 80 82

R6 R7 R8 R9 R10

87 86 67 14 78

R11 R12 R13 R14 R15

78 61 85 98 78

3.2 Integrating Sphere Test at 25°C for GL02-70W-50K-SNN-A

Page 7: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 7 of 13

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor Power

(W)

120.03 60 0.5920 0.9925 70.49

Luminous Flux (lm)

CBCP (cd)

Zone/Spacing Criteria (80-90°)

Luminous Flux (lm) (0-90°)

Luminous Efficacy (lm/W) (0-90°)

8438.38 1803 9.20% 7365.24 104.49

Luminous Intensity Distribution

Zonal Lumen Summary

3.3 Goniophotometer Test at 25°C for GL02-70W-30K-SNN-A

Page 8: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 8 of 13

Luminous Intensity (cd) Distribution Data

0 22.5 45 67.5 90 112.5 135 157.5 180 202.5 225 247.5 270 292.5 315 337.5 360

0 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803 1803

1 1755 1801 1804 1830 1842 1773 1772 1787 1770 1787 1772 1773 1842 1830 1804 1801 1755

2 1742 1767 1742 1836 1856 1809 1749 1666 1648 1666 1749 1809 1856 1836 1742 1767 1742

3 1870 1896 1785 1795 1862 1805 1661 1675 1673 1675 1661 1805 1862 1795 1785 1896 1870

4 1899 1918 1911 1750 1848 1725 1703 1582 1491 1582 1703 1725 1848 1750 1911 1918 1899

5 1990 1962 1906 1738 1831 1625 1607 1575 1570 1575 1607 1625 1831 1738 1906 1962 1990

10 2060 1969 1960 1913 1853 1543 1499 1351 1314 1351 1499 1543 1853 1913 1960 1969 2060

15 2189 2182 2101 1965 1793 1446 1239 957 848 957 1239 1446 1793 1965 2101 2182 2189

20 2336 2332 2123 2029 1753 1387 894 708 637 708 894 1387 1753 2029 2123 2332 2336

25 2708 2586 2160 1891 1624 1170 719 489 398 489 719 1170 1624 1891 2160 2586 2708

30 3238 2439 2448 1905 1510 945 543 316 268 316 543 945 1510 1905 2448 2439 3238

35 3008 2692 2491 1801 1350 763 372 294 306 294 372 763 1350 1801 2491 2692 3008

40 3013 3061 2533 1720 1173 606 276 314 312 314 276 606 1173 1720 2533 3061 3013

45 3162 3042 2612 1763 1037 449 251 264 195 264 251 449 1037 1763 2612 3042 3162

50 3017 3142 2632 1710 902 347 237 173 139 173 237 347 902 1710 2632 3142 3017

55 3197 2990 2538 1652 770 271 226 135 141 135 226 271 770 1652 2538 2990 3197

60 3423 3227 2335 1592 634 218 180 113 95 113 180 218 634 1592 2335 3227 3423

65 3321 3209 2263 1510 508 175 124 68 49 68 124 175 508 1510 2263 3209 3321

70 2912 2759 2142 1344 402 137 87 30 16 30 87 137 402 1344 2142 2759 2912

75 2821 2652 1864 1160 294 113 53 3 2 3 53 113 294 1160 1864 2652 2821

80 2584 2302 1518 904 209 89 37 3 2 3 37 89 209 904 1518 2302 2584

85 2362 2204 1342 689 150 70 32 3 3 3 32 70 150 689 1342 2204 2362

90 2133 1965 1135 557 120 63 29 3 4 3 29 63 120 557 1135 1965 2133

95 1799 1657 854 412 109 57 25 4 4 4 25 57 109 412 854 1657 1799

100 1438 1182 651 286 100 53 23 4 5 4 23 53 100 286 651 1182 1438

105 795 671 487 218 85 55 19 5 5 5 19 55 85 218 487 671 795

110 211 253 325 203 74 51 16 5 5 5 16 51 74 203 325 253 211

115 229 223 236 136 67 42 12 5 6 5 12 42 67 136 236 223 229

120 215 216 181 110 56 33 10 5 6 5 10 33 56 110 181 216 215

125 190 203 127 113 46 26 8 5 6 5 8 26 46 113 127 203 190

130 160 177 121 77 38 21 7 5 5 5 7 21 38 77 121 177 160

135 170 184 117 57 30 17 5 5 5 5 5 17 30 57 117 184 170

140 159 152 91 42 21 11 4 5 5 5 4 11 21 42 91 152 159

145 132 141 95 41 11 5 4 5 5 5 4 5 11 41 95 141 132

150 109 107 79 44 5 3 4 4 5 4 4 3 5 44 79 107 109

155 67 76 80 15 6 4 4 4 4 4 4 4 6 15 80 76 67

160 63 68 54 6 4 3 4 4 4 4 4 3 4 6 54 68 63

165 38 39 7 3 3 3 3 4 4 4 3 3 3 3 7 39 38

170 4 4 3 3 3 3 3 3 3 3 3 3 3 3 3 4 4

175 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3

180 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3

Page 9: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 9 of 13

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor Power

(W)

120.02 60 0.5910 0.9926 70.35

Luminous Flux (lm)

CBCP (cd)

Zone/Spacing Criteria (80-90°)

Luminous Flux (lm) (0-90°)

Luminous Efficacy (lm/W) (0-90°)

8781.03 1930 9.10% 7687.33 109.27

Luminous Intensity Distribution

Zonal Lumen Summary

3.4 Goniophotometer Test at 25°C for GL02-70W-50K-SNN-A

Page 10: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 10 of 13

Luminous Intensity (cd) Distribution Data

0 22.5 45 67.5 90 112.5 135 157.5 180 202.5 225 247.5 270 292.5 315 337.5 360

0 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930 1930

1 1816 1811 1823 1870 1937 1929 1906 1892 1898 1892 1906 1929 1937 1870 1823 1811 1816

2 1957 1926 1832 1836 1951 1893 1873 1859 1836 1859 1873 1893 1951 1836 1832 1926 1957

3 1991 1989 1968 1826 1957 1887 1801 1729 1744 1729 1801 1887 1957 1826 1968 1989 1991

4 2003 1999 2008 1853 1947 1864 1755 1757 1728 1757 1755 1864 1947 1853 2008 1999 2003

5 2202 2108 2009 1912 1946 1806 1765 1568 1590 1568 1765 1806 1946 1912 2009 2108 2202

10 2203 2069 2088 2009 1956 1686 1507 1456 1426 1456 1507 1686 1956 2009 2088 2069 2203

15 2311 2286 2248 2151 1895 1572 1368 1070 965 1070 1368 1572 1895 2151 2248 2286 2311

20 2435 2441 2238 2017 1857 1498 1006 765 696 765 1006 1498 1857 2017 2238 2441 2435

25 3042 2540 2203 2032 1741 1283 782 546 458 546 782 1283 1741 2032 2203 2540 3042

30 3374 2485 2587 2013 1579 1028 602 342 282 342 602 1028 1579 2013 2587 2485 3374

35 3105 2794 2688 1897 1431 827 411 296 311 296 411 827 1431 1897 2688 2794 3105

40 3142 3077 2803 1822 1268 659 299 304 339 304 299 659 1268 1822 2803 3077 3142

45 3315 3128 2787 1823 1110 501 246 294 215 294 246 501 1110 1823 2787 3128 3315

50 3239 3238 2758 1835 952 379 247 187 147 187 247 379 952 1835 2758 3238 3239

55 3320 3015 2637 1766 823 292 227 139 146 139 227 292 823 1766 2637 3015 3320

60 3473 3247 2426 1606 674 231 204 121 109 121 204 231 674 1606 2426 3247 3473

65 3362 3406 2388 1568 541 187 131 77 55 77 131 187 541 1568 2388 3406 3362

70 3014 2869 2316 1392 427 143 92 35 25 35 92 143 427 1392 2316 2869 3014

75 2881 2712 1915 1196 313 120 61 3 2 3 61 120 313 1196 1915 2712 2881

80 2661 2351 1573 940 213 95 39 2 2 2 39 95 213 940 1573 2351 2661

85 2346 2297 1378 731 151 74 33 3 3 3 33 74 151 731 1378 2297 2346

90 2252 1983 1186 575 123 66 30 3 4 3 30 66 123 575 1186 1983 2252

95 1874 1656 888 422 115 59 27 4 5 4 27 59 115 422 888 1656 1874

100 1451 1137 658 293 103 56 24 5 5 5 24 56 103 293 658 1137 1451

105 798 703 499 228 90 57 21 5 5 5 21 57 90 228 499 703 798

110 240 246 339 210 78 53 17 5 6 5 17 53 78 210 339 246 240

115 225 225 242 137 69 44 13 5 6 5 13 44 69 137 242 225 225

120 222 222 183 115 57 34 10 5 6 5 10 34 57 115 183 222 222

125 200 204 130 118 47 27 9 5 6 5 9 27 47 118 130 204 200

130 168 182 128 82 39 22 7 5 5 5 7 22 39 82 128 182 168

135 184 192 118 63 31 17 5 5 5 5 5 17 31 63 118 192 184

140 160 158 96 46 22 11 4 5 5 5 4 11 22 46 96 158 160

145 138 147 98 41 11 5 4 5 5 5 4 5 11 41 98 147 138

150 112 108 86 29 6 4 4 5 5 5 4 4 6 29 86 108 112

155 74 88 92 13 6 4 4 4 4 4 4 4 6 13 92 88 74

160 86 86 49 6 4 3 4 4 4 4 4 3 4 6 49 86 86

165 34 29 4 2 3 3 3 4 4 4 3 3 3 2 4 29 34

170 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3

175 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3

180 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3

Page 11: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 11 of 13

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor THDi

277.06 60 0.2861 0.9073 15.80%

Input Voltage (V)

Frequency (Hz)

Current (A)

Power Factor THDi

277.05 60 0.2857 0.9070 15.84%

3.5 Electrical Test for GL02-70W-30K-SNN-A

3.6 Electrical Test for GL02-70W-50K-SNN-A

Page 12: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 12 of 13

4 Test Equipment

Equipment Name Manufacturer Model No Reference No Calibration Due Date

Integrating Sphere (1.5m)

SENSING SPR-600M SE-451 Before Use

Integrating Sphere (2.0m)

EVERFINE YF-1000 SE-599 Before Use

Standard Lamp EVERFINE D062 SE-606 2019-06-05

Digital Power Meter YOKOGAWA WT210 SE-074 2019-06-05

Goniophotometer System

SENSING GMS-3000 SE-450 Before Use

Standard Lamp SENSING 110V/100W SE-465 2019-06-05

Digital Power Meter YOKOGAWA WT310 SE-381 2019-06-05

AC Power Source HUAYANG HY9010 SE-114 Before Use

DC Power Source EVERFINE WY605 SE-605 2019-06-05

Temperature Sensor

WALVICO HG126D SE-616 2019-06-05

Statement of Traceability: Shenzhen Anbotek Pengcheng Compliance Laboratory Limited attests that all calibration has been performed using suitable standards traceable to national primary standards and International System of Unit (SI).

Page 13: TEST REPORT · 2019. 12. 10. · Report No.: SZANL181217002-01 LM-79 REPORT Page 4 of 13 2 Test Method 2.1 Integrating Sphere Measurement The integrating sphere system includes AC

Report No.: SZANL181217002-01

LM-79 REPORT Page 13 of 13

Attachment A – Product Photo

*****END OF TEST REPORT*****