Upload
mohamednibin
View
255
Download
0
Embed Size (px)
Citation preview
7/23/2019 Teledyne Analyser
1/20
Series 6000
Ultraviolet Photometric Analyzer
Built for Reliability and Performance
TELEDYNEANALYTICAL INSTRUMENTS
T
eledynes Series 6000 Microprocessor Controlled
Ultraviolet Photometer is designed to provide
continuous on-line analytical data of critical processconditions and eliminate time consuming laboratory
analysis. This analyzer can be configured to
measure at any wavelength within the ultraviolet
or visible portions of the electromagnetic energy
spectrum.
Since numerous gases and liquids have a useful
absorbance peak in the ultraviolet region, a wide
variety of applications are possible.
Advantages Dependable microprocessor-controlled analysis
Auto calibration and auto ranging
Continuous real time analysis
Eliminates time-consuming lab analysis
Features
Microprocessor based, high resolution, ultra stable
electronics (no potentiometers)
Three user-selectable ranges
Signal & Range ID output: 0-1 VDC and 4-20 mADC
(isolated)
Programmable auto ranging
Range ID contacts (quantity 4) Form A normally open
contacts, 3A@250 VAC resistive
Two adjustable concentration alarm set points with
programmable relay function Form C contacts,
3A@250 VAC resistive
Programmable auto calibration with mode ID Form A
normally open contacts
Remotely operated calibration (customer suppliedvalves and 24 VDC signal)
Self diagnostics with Form C failure contacts
Full duplex RS232 communication link
Alphanumeric VF display for set-up and diagnostics
Power: 115/230 VAC 50/60 Hz
Principle of Operation
The analyzers optical path (figure 1) consists of a
UV source, sample cell, and detector. A filter wheel
located in front of the detector continuously rotates
a reference and measuring filter in and out of the
optical path.
The material of interest continuously flows through
the sample cell where the optical energy is absorbed
at the various wavelengths in accordance with its
composition. The analyzer utilizes the reference
wavelength to compensate for extraneous
phenomena (turbidity, dirt on optics, source drift,
etc.) thereby producing a high degree of stability
over long periods.
7/23/2019 Teledyne Analyser
2/20
Flourescence ProbeOil-in-Water Analyzer
MODEL 6650Flourescence ProbeOil-in-Water Analyzer
TELEDYNEANALYTICAL INSTRUMENTS
Built for Reliability and Performance
Teledynes Model 6650is comprised ofa versatile transmitter and probe
for monitoring oil-in-water in
various reneries, power plants,
and offshore platforms. The
system incorporates state-of-
the-art electronics and a ber
optic based, in-situ probe thatdetects oil-in-water.
Detection of oil-in-water
in process boilers, cooling
towers, reneries, and offshore
platforms is an important
measurement from an economic
and environmental standpoint.
This new technology from Teledyne makes
this measurement simple, and reduces the
cost of ownership. Because of its in-situ
measurement capabilities, the Model 6650
analyzer offers real time analysis within a
second.
Features
* Direct measurement; no interference from
turbidity
No spare part replacements required for
3 years
Long life UV lamp
Virtually no drift
No inner lter error effects
Fiber Optic Probe
The Model 6650 uses a ber optic
based, in-situ probe that monitors actual
concentrations of oil-in-water using UV
uorescence technology. TAI also offers an automatic
retractor as an option for cleaning the probe,
depending upon the application.
Model 6650 Transmitter
Teledynes 6650 uses a highly versatile
transmitter that measures the signal fromour probe and sends the signal back to the
transmitter. This reading is based on the
amount of optical attenuation from the ber
optic in-situ probe.
The transmitter is compact and designed
to connect to the probe via two ber optic
cables. A local digital display is provided and the
transmitter offers a 4-20 mA output signal to other
devices. In addition,
the transmitter
is equipped with
automatic calibration
checking. A reference
lter is inserted into
the measuring beam
either locally or
remotely to verify operation of the transmitter.
7/23/2019 Teledyne Analyser
3/20
Model 6650 Oil-In-Water Analyzer
2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD
Warranty
Instrument is warranted for 1 year against defects in materialor workmanship
NOTE: Specifications and features will vary with application. The above are establishedand validated during design, but are not to be construed as test criteria for every product.All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
User Display & Control
Type of display: LED display
Display format: 3-1/2 digits in user defined engineering
units
Electrical
Power requirement: 24 VDC (9 - 32 VDC) if 110 / 220 VAC
is available (optional) AC / DC power
supply
Power consumption: 0.48 Watts
Analog outputs: 4 - 20 mA isolated
Analog loop resistance: 500 Ohms, nominally @ 24V
Alarms: Optional
Area classification: Z-purge for Class I, Division 2;
X-purge for Class I, Division 1
Mechanical
Transmitter weight: 1.5 lbs
Enclosure construction: Extruded aluminum (NEMA
enclosures optional)
8 H x 3 - 7/8 W x 71 - 1/2 D
Probe
Materials: 316 SS or Hastelloy C
Temperature rating: 315.5 C (600 F)Pressure rating: 10,000 psig
Probe options: 1. 6, 12 or 24 in length
2. Temperature controlled
3. Temperature thermistor
4. Pressure strain gauge
APPLICATIONS
Pollution abatement programs involving effluent
from refineries, chemical / petrochemical plants,
oil fields, steel mills, automotive production, food
processing, and other industries
Offshore drilling platforms, produced water, oil
field water flooding, steam injection operations On-board shipping (fast analysis required
7/23/2019 Teledyne Analyser
4/20
MODEL 6600Oil in Water Analyzer System
Redesigned with modularity and lower cost in mind
Rack or cabinet mounted
General purpose or explosion proof
Teledynes Model 660/661 has been the industry standard inoil-in-water analysis for 30 years. The microprocessor based6600 / 6610 improves upon our proven analog circuitry and the
sample handling system. The new, compact modern design
incorporates many new features at a significantly reduced cost.
Critical features include:
Folded optical system optimized for excellent signal-to-noise
ratio and stability
Increased reliability by eliminating many moving parts
New sonicator homogenizes oil (in-line) in water instantly (less
than one second)
Improved auto-zero filter system prepares zero fluid
for calibration and zeroing without long sparging cycle
Specially developed software with auto calibration features
Automatic self-cleaning sample cell suitable for high pressure
and temperature
Manual calibrations done easily with collected zero
fluid from sample system using the gravity fed reservoir
Manual or automatic span sensitivity easily checked with
internal span flag once field calibrated
Faster response time
Modular design configured to meet specific application
requirements
The 6600 utilizes the ultraviolet absorption principle to detect
and continuously measure oil concentration in water with a
reproducibility that exceeds standard laboratory techniques.
Teledyne Analytical Instruments
Oil in Water Analyzer System
The automated analysis system includes a single beam,
dual wavelength, UV photometric analyzer that compensates
for turbidity, algae, cell window coatings, and other optical
attenuation.
To assure accuracy, a unique conditioning system delivers a
sample representative of the true oil content of the stream.
A continuous ultrasonic homogenizer (figure 2) disperses all
suspended oil droplets and oil absorbed into foreign matter so
the sample, to the analyzer, appears to be uniform and in true
solution.
OIL DETECTION AND ANALYSIS
The basic oil detection UV analyzers offered by other
manufacturers are single wavelength photometers a "go / no
go" instrument without a sample conditioning system.
Teledyne feels these lack the sophistication necessary for reliable
oil detection monitoring, preferring a dual wavelength design.Attenuation of energy transmission by turbidity and cell window
deposits can then be distinguished from absorption despite the
presence of oil.
With sample conditioning (figure 1) and automatic zeroing, our
analyzers produce a truly representative solution for analysis and
can compensate for background absorption found in most oil-in-
water applications.
DUAL WAVELENGTH
Recognizing the importance of eliminating the effects of turbidity
and inherent electro-optical interferences, Teledyne developed a
single sample cell, chopped beam, dual-wavelength photometric
analyzer unmatched in reliability and accuracy.The 6600 has been extensively field tested to show excellent
correlation with the EPA gravemetric reference method 413.1.
The system utilizes optical filters to isolate and make energy
measurements at two specific wavelength bands. The UV
measuring band (centered at 254 nanometers) is extremely
narrow to avoid side-band interference.
A reference signal is generated at a selected wavelength region
where oil does not absorb significantly. This signal is effected by
the aforementioned interference as well as by absorption due to
the presence of oil.
Components contributing to total absorption in a typical water
effluent sample are shown in figure three. Turbidity and electro-optical interferences are removed by the analyzers design, and
non-oil background is zeroed out.
Since both the measuring and reference wavelengths are
generated at the end of the transmission system, electro-optical
interference (cell window coating, variations in source lamp,
aging of detector and other electronics) effects each signal
equally. The output signal is obtained by electronically generating
the logarithm of the ratio of the measuring signal to the reference
signal. Therefore, the output signal, linear with absorption, shows
little or no affect from turbidity. Additionally, electro-optical stability
is enhanced by the dual wavelength design.
7/23/2019 Teledyne Analyser
5/20
APPLICATIONS
Pollution abatement programs involving effluents from refineries,
chemical / petrochemical plants, oil fields, steel mills, automotive
production, food processing, and other industries
Offshore drilling platforms, produced water, oil field water flooding,
steam injection operations
On-board shipping (fast analysis required < 10 seconds)
On-shore deballasting discharges and ballast treatment facilities
Boiler return, feedwater, steam condensate, cooling water, leak
detection
Monitoring of airport runoff, municipal water treatment plants
Wastewater and sewage treatment plants
Process stream monitoring
Aromatic hydrocarbons like benzene, toluene, styrene, xylenes in
water
FEATURES
Specific to oil (correlates to EPA 413.1)
Microprocessor based electronics
Auto calibration
Remotely initiated cal / zero via customer supplied 24 VDC signal
Self-diagnostics with Form C relay contacts
Ratio measurement
Continuous measurement
Total oil readout standard, undissolved oil readout possible (with
autozero filtering option)
Integral alarms
Self-cleaning cell
Automatic turbidity compensation
Full duplex RS-232 communication link
Process pressure up to 150 PSIG (higher pressure possible
application dependent)
Sample temperature to 120C (higher temperatures possible,
application dependent)
Response time under 5 seconds possible (application dependent)
NEMA-4x (stainless steel standard) mounted control unit
AUTOMATIC ZERO MODE
The 6600 offers an automatic zeroing feature, permitting a continuous
analysis specific for oil without interference. The zeroing mode
technique (application dependent) is accomplished by the automatic
preparation (filtering) and periodic analysis of an oil-free fluid which
contains the compounds contributing to background absorption. When
this fluid is analyzed, the background absorption is measured and
then zeroed out. The cycle is automatically repeated once each hour
to update the oil analysis by correcting for fluctuations. This allows for a
more accurate analysis of oil and grease content specifically separated
from other non-oil organics that many times are also present in effluents.
For very clean processes like steam or boiler return condensates, only
oil free nitrogen (N2) gas or air can be used for electro-optical energy
background variations such as from normal source or detector anomalies
over time.
UNHOMOGENIZED HOMOGENIZED
FIGURE 2
Homogenizer Function The effect of homogenization on the sample is
illustrated above. Both the particulate oil and the compounds representing
turbidity are of random size. Homogenization causes the oil droplets to be
dispersed to such a particle size that, to the analyzer, they appear to be in
true solution. Oil absorbed upon or contained within the portion represented
by turbidity is also placed into solution along with the particulate oil. Turbidity
has no characteristic absorption and is ratioed to zero by the analyzer.
NON-OILORGANIC
BACKGROUND
DISSOLVED OIL
UNDISSOLVED OIL
TURBIDITY
ZERO FLUID
"ZERO"
PURE
H2
0
HOMOGENIZED
SAMPLE
FIGURE 3
Fractions contributing to total absorption in typical water effluen
The homogenized sample contains all components of the effluent stream
The UV analyzer measures the stream fractions as non-oil background
dissolved oil, and non-dissolved oil. Turbidity will be ratioed out and give
no reading. The analyzer will measure the zero fluid as the non-oil fraction
Turbidity and oil have been removed from the zero fluid by coarse and ultra
fine filtering. Coarse filtering removes the non-dissolved oil, while ultra-fine
filtering removes the lighter dissolved oils. During the zero cycle, the
analyzer is automatically adjusted to read zero. The non-oil UV absorbing
background is subtracted from the total UV reading.
7/23/2019 Teledyne Analyser
6/20
SAMPLE CONDITIONING
Factors associated with undissolved oil and background absorption are
eliminated by sample conditioning. The 6600 incorporates conditioning
components that prepare the sample so both undissolved and dissolved
oils are detected and measured.
A new, in-line, high frequency (20 kHz) ultrasonic homogenizer
minimizes the size of the oil droplets suspended while dispersing
them into a more uniform molecular level in the sample so they can
be measured accurately. For any photometric analytical technique
measuring oil in water above 20 ppm oil requires this homogenization.Oil absorbed onto the particulate matter in the sample is also dispersed
by the homogenizer.
Since the response of particulate matter as a possible attenuator is
automatically neutralized by the optical system and ratioed to zero by
the electronics of the photometer, the characteristic absorption of the
total oil can be measured. The output signal or meter readout is linearly
proportional with total oil concentration.
OIL AND GREASE
Many environmental applications (as monitored by regulatory agencies)
allow certain minimal levels of dissolved oil to remain in the water
measured. The 6600 (when ordered specifically to do so) can monitor
for total oil and grease recoverable or be easily switched to monitor
the undissolved oil portion only. This minimizes the customers
costly requirement in energy to process water below a minimum oil
concentration as dictated by local regulations before out-flowing the
water as an effluent.
COMPENSATION FOR BACKGROUND
Many applications require compensation for background absorption
due to the presence of non-oil organic and inorganic compounds that
absorb UV energy. Furthermore, the concentration of these compounds
will likely fluctuate, specifically in applications with multiple non-oil
organics and inorganic compounds present in the background. In
order to obtain an analysis specific to oil, the background absorption
contributed by these compounds must be eliminated and zeroed out.
The 6600 does exactly this.
Sample Conditioning System A sample is fed through a V1 sample / utility valve, then through a split tee; one side as a bypass loop, the other through
an inline ultrasonic homogenizer. This disperses any oil in the sample before measurement including small or large oil droplets and oil absorbed on foreign
particles. A portion of the stream is conditioned to remove all oil, both dissolved and undissolved, without altering the background (boiler additives or other
inorganic compounds.) When this portion is delivered to the analyzer (through SV3), the analyzer subtracts the background from the total and reads out
total oil only. The analyzer is manually calibrated with a known standard on a one time basis through the indicated zero and span inlet reservoir. The
specially designed sample cell can interface with process fluids up to 150 psig and 120C (248F)
BY-PASSF2BY-PASS
FLOWMETER0-110 LPH
V10HOMOGENIZER
FC1
FLOW
CONTROLLER
F1
SAMPLE
FLOWMETER
0-30 LPH
SAFETY BLOCK
VALVE
ZERO & SPAN IN
C C C C
V5
V7
V6
CELL BLOCK
V13-WAYVALVE
C
C
V4
V8V9C
IN
BACK PRESSURE
REGULATOR
20-50 PSI
PUMP(2-5 LPM)30-80 GPH
FILTER0,2 MICRON
FILTER3.0 MICRON
V3
IN
PV2
BPR1
SAMPLE IN(2-5 LPM)30-80 GPH
15' WATER VACUUM
UTILITY
GRABSAMPLE
N2 OR AIR
90 TO 100 PSIG
RETURN
SAFE VENT
SV3
SV4
F2 F3
F1
IN
OUT
SV1SV2
NOTES: UNLESS OTHERWISE SPECIFIED.
1. USE HOMOGENIZER AND FILTER ASSEMBLIES FOR HIGH RANGE,
HIGH BACKGROUND
2. USE FILTER ASSEMBLY FOR LOW RANGE, HIGH BACKGROUND
3.
USE BACKFLUSH SOLENOIDS FOR LOW RANGE, VERY LOW BACKGROUND
4. USE PUMP ASSEMBLY FOR LOW PRESSURE OR NO GRAVITY FEED
FIGURE 1
7/23/2019 Teledyne Analyser
7/20
OPTIONAL MODULES
Homogenizer Module A high frequency (20 kHz) ultrasonic
homogenizer is used to minimize the size of the oil droplets
while dispensing them in the sample so they can be measured
accurately. Measuring samples above 20 ppm typically require this
module.
Filter Assembly Module required for applications with highbackground interference
Pump Module required for applications with low pressure or no
gravity feed
Purge Module Z purge unit or X purge unit
CONFIGURATIONS
General purpose
Class I, Div. 2, Groups B, C & D
Class I, Div. 1, Groups B, C & D
CE mark, CENELEC certified
OIL IN WATER ANALYZER SYSTEM SERIES 6600-6610
SPECIFICATIONS
Range: 0-10 ppm up to 0-200 ppm oil in water
(application dependent)
Noise: Less than 1% of full scale
Zero drift: Corrected by automatic zero
Accuracy: Dependent upon variability of oil composition;
2% of full scale when calibrated against
same oil as being analyzed
Reproducibility: 1% of full scale or better
Linearity: 1% of full scale or better
Power 115 VAC 50/60 Hz; requirements: 230 VAC,
50/60Hz (specify)
Ambient 0 to 50C; may be heated or cooled to
temperature: operate in other temperatures
Sample 1 to 120C, nominal (non-freezing),temperature: lower temperatures possible
with salt waters
Sample 0 to 150 PSIG (an optional pump can be
pressure: supplied when no gravity feed or
sample pressure is available)
Signal output: Analytical measurement, 0-1 VDC and
isolated 4-20 mADC
Alarms: Dual fully adjustable concentration alarm set
points with programmable relay function.
Form C contacts, 3A resistive; self-diagnostics with Form C failure alarm contact
Data lines: Bi-directional RS-232C serial interface
Readout: Two line alphanumeric vacuum florescent
display0-10 ppm oil-in-water
monitoring system for boiler /
steam condensate application
2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD
WarrantyInstrument is warranted for 1 year against defects in material
or workmanshipNOTE: Specifications and features will vary with application. The above are establishedand validated during design, but are not to be construed as test criteria for every product.All specifications and features are subject to change without notice.
TELEDYNEANALYTICAL INSTRUMENTSA Teledyne Technologies Company
16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
7/23/2019 Teledyne Analyser
8/20
Features
Ranges are 0-50 ppb to 0-20 ppm, user selectable
Dual ranges and auto ranging
Microprocessor controlled for versatility
Multi-tasking software allows viewing test variables while
operating
Continuous self checking with alarms
Dual bi-directional RS-232 ports for remote operation
(optional RS-485 or Ethernet)
Digital status outputs indicate instrument operating condition
Adaptive signal filtering optimizes response time
Temperature and pressure compensation
Internal Zero & Span check (optional)
Internal data logging with 1 min to 365 day multiple
averages
Critical orifices provide flow stability
Teledyne Analytical Instruments
6400
E
U V F l u o r e s c e n c e S O 2 A n a l y z e r
Teledynes 6400E uses a proven UV fluorescenceprinciple coupled with state-of-the-art microprocessortechnology to provide accurate and dependable
measurements of low level S02.
Exceptional stability is achieved with the use of an opticalshutter to compensate for PMT drift and a reference
detector to correct for changes in UV lamp intensity.
A hydrocarbon kicker and advanced optical design
combine to prevent inaccuracies due to interferents. The
multi-tasking software gives real time indication of many
operational parameters and provides automatic alarms if
diagnostic limits are exceeded.
Advanced Performance
The 6400E includes a built-in data acquisition capability
using the analyzers own internal memory. This allows the logging
of multiple parameters including averaged or instantaneousconcentration values, calibration data and operating parameters
such as flow, pressure and lamp intensity.
Stored data are easily retrieved through the serial port or optional
Ethernet port via our software or from the front panel, allowing
operators to perform predictive diagnostics and enhanced data
analysis by tracking parameter trends.
The 6400E combines lighter weight, rugged construction, ease
of use, powerful diagnostics, modular design and outstanding
performance to meet air monitoring requirements.
Applications
Detect low trace levels of SO2 in medical air and medical
oxygen to satisfy USA and European pharmacopeiarequirements.
Detect trace levels of H2S in methane-rich Bio-Gas streams
used as the feed for fuel cell power generation devices. The
H2S impurities, which can poison the platinum based catalyst,
are converted into SO2 on a 1:1 basis via Teledynes THC
converter. The resultant converted sample gas is then fed to the
6400E for continuous analysis purposes.
6400E Schematic6400E Schematic
7/23/2019 Teledyne Analyser
9/20
6400E FLUORESCENT SO2 ANALYZER
SPECIFICATIONS
Ranges: 0-50 ppb to 0-20,000 ppb full scale, user
selectable; dual ranges and auto ranging
supported
Units: ppb, ppm, g/m3, mg/m3
Zero noise:
7/23/2019 Teledyne Analyser
10/20
TELEDYNEANALYTICAL INSTRUMENTS
Total
Sulphides
Analyzer
FOR CO2 Purity Quality Monitoring
7/23/2019 Teledyne Analyser
11/20
The need to continuously detect for suldes
in carbon dioxide has become increasingly
important within the food and beverage market.
Gas suppliers must ensure that the CO2 they provide
for use in the preservation of food and drink related
products meet todays exacting standards. The
contamination of carbon dioxide can emanate from
many sources and therefore it is essential that the
CO2 used for such purposes be monitored at theCO2 generator facility, at the beverage producer
facility, or both.
PRINCIPLE OF OPERATION
The Model 6200E Total Suldes Analyzer utilizes
proven UV uorescent technology to continuously
detect suldes found in inert gas streams. An
internal, quartz catalytic converter is employed to
convert the suldes, when mixed with scrubbed
ambient air, into SO2 via high temperature oxidation.
An internal vacuum pump is employed to draw boththe sample and the ambient air into the converter.
The converted sample gas is fed to the uorescence
chamber where it is then exposed to ultra-violet
radiation.
The measurement process in the reaction cell is
described by the following equation:
SO2 + hu1dSO2*
SO2*dSO2 + hu2
Where hu1 is the incident UV energy at 214nm andhu2 is the resultant uorescence, which is directly
proportional to the SO2 concentration in the reaction
cell.
The uorescence emission is in turn sensed by
the photo-multiplier tube (PMT) and amplied to
provide an analog output. The result is a continuous
measurement of total suldes (i.e. H2S, CS2, COS,
mercaptans, etc.) as SO2 from 0-50 ppb to 0-20
ppm.
Model 6200E basic configuration
7/23/2019 Teledyne Analyser
12/20
ELECTRONICS PLATFORM
The 6200E combines the proven UV uorescence
principle with state of the art microprocessor
technology to provide accurate and dependable
measurement of trace levels of total suldes. Long
term stability is obtained by the use of an optical
shutter to compensate for zero drift, coupled with
lamp reference detector to correct for lamp drift.
The multi-tasking microprocessor allows easy eld
ranging from 50 ppb to 20 ppm as well as providing
on-line indication of instrument status. The 6200E
is continuously checking operating parameters and
provides alarms on out of specication conditions,
including electrical and optical parameters.
The easy to read display, intuitive menu structure,
and ease of operation combine to make the
6200E the instrument of choice for Total Suldes
applications.
The instrument includes a built-in data acquisitioncapability, utilizing the analyzers internal RAM.
This allows logging of the measured variable
concentration as well as other parameters such as
calibration data, ow rates, lamp intensities or a
variety of other congurable parameters.
This provides the operator the ability to perform
predictive diagnostics by tracking parameters over
time. This stored data is available through an RS-
232C port on either an automatic timed or manual
manner.
THE COMPONENTS
The 6200E comes complete with three 19 inch relay
rack modules - the Sample Converter Module, the
Analysis Module, and a Calibration Module.
The Sample Converter Modulecomes with an
SO2 scrubber, sample / ambient air lters, high
temperature dilution control orice box, and Total
Reduced Sulfur (TRS) converter.
The Analysis Moduleconsists of the UV
uorescence analyzer, pressure and ow
transducers, and vacuum pump.
The Calibration Moduleconsists of restrictor /
orice ow regulators and ow control system,
all provided in a temperature controlled setting to
ensure accurate performance.
The two inlet ports allow use of two independent
calibration gases which may be diluted to provide
the required span gas concentration. Dilution ratios
from 20:1 to 300:1 are manually set from the frontpanel.
FEATURES
High sensitivity with user selectable ranges
from 0-50 ppb to 0-20 ppm
Microprocessor controlled, providing
advanced user interface capabilities
Built-in data acquisition capability using the
analyzers own RAM
Extremely high-end self diagnostic capability
providing continuous self checking with
continuous warning alarms
Bi-directional RS-232C for remote operation
Critical orices provide ow sensitivity
High temperature TRS converter
Digital status outputs provide instrument
condition
Easy to read display
Proven UV uorescence technique
7/23/2019 Teledyne Analyser
13/20
2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 05/06LD
Warranty
Instrument is warranted for 1 year against defects in materialor workmanship
NOTE: Specifications and features will vary with application. The above are establishedand validated during design, but are not to be construed as test criteria for every product. Allspecifications and features are subject to change without notice.
6200E Total Sulphides Analyzer
SPECIFICATIONS
Ranges: 0 - 50 ppb to 0 - 20,000 ppb full-scale, user
selectable
Dual ranges and auto-ranging supported
Zero noise: < 0.2 ppbSpan noise: < 0.5% of reading (above 50 ppb)
LDL: 0.4 ppb
Zero drift**: < 0.5 reading / 24hrs; 1 ppb / 7 days
Span drift**: < 0.5% reading / 24 hrs; 1% of reading / 7
days
Lag time: 20 secs
Rise time (95%): < 120 secs
Linearity: 1% of full scale
Sample flow rate: 700 cc/min 10%
Converter temperature: 1000C
Converter efficiency: > 98% above 0C dewpoint
Converter life: 3000 ppm hours
SOx scrubber efficiency: > 98%
SOx scrubber life: > 1000 ppm hours
Analyzer operating temperature: 5 to 40C
Power: 100 Vac 240 Vac, 50/60 Hz (user
specified)
Outputs: Analog: 10 V, 5 V, 1 V, 100 mV (selectable)
(4-20mA isolated output optional)
RS232 (I/O): Standard DB9
Digital status: 12 contact closure outputs
TELEDYNE
ANALYTICAL INSTRUMENTSA Teledyne Technologies Company
16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com
Size (H x W x D):
Model 6200E analyzer:
(7 x 17 x 23.5)
(177.8 x 432 x 596.9 mm)
Dilutor / Converter module:
(7 x 17 x 23.5)(177.8 x 432 x 596.9 mm)
Calibration module:
(5.25x 17 x 23.5)
(133.3 x 432 x 596.9 mm)
**At constant conditions
ADDITIONAL OPTIONS
Fluorocarbon zero / span valves
19 rack mount brackets
19 rack mount with chassis sliders
2 to 5 year warranty
Contact Teledyne for details on our Carbon Dioxide
Quality Control (CDQC), which is a turn-key
system capable of detecting:
Total Sulphides (UV
Total Hydrocarbons Trace Moisture (AI2 O3)
Trace Oxygen (Electrochemical)
CO2 Purity (NDIR)
7/23/2019 Teledyne Analyser
14/20
Process Photometers
PUV3402 and PIR3502
AnalyzeIT
FieldIT
ControlIT
EngineerIT
FieldIT
InformIT
OperateIT
PowerIT
IndustrialIT
Continuous Measurements
VistaNET Connectivity
Multiple Component Samples
Measures Vapor or Liquid Samples
Operates in IR, NIR, UV and VIS Regions
Fiber Optic Option for NIR Applications
Multiple Interference Compensation Capability
7/23/2019 Teledyne Analyser
15/20
UV/VIS/NIR Fiber Optics
Option
The PFO3502 Multiwave offers a fiber optic option
for applications that require remote sampling capa-
bility. With this option, light is transmitted via one
waveguide to the remote sample cell. A second
waveguide is used to return the sample-modified
light to the detector.
This Fiber Optics option is effective in
UV/Visible/NIR applications where:
The sample stream is highly toxic.
Corrosive products are analyzed.
Fast response time is required.
The sample is at high pressure.
Cell-In-Oven Sample
System
The Multiwave Cell-In-Oven designprovides a simple and effective pack-age for high temperature applications.The major benefit of this design is thatthe sample cell and sample handlingcomponents are together in one com-
mon temperature controlled enclosure.The elimination of heat trace require-ments for sampling lines minimizesthe risk of cold spots. This designalso offers excellent accessibility forfast cell removal during maintenance.
Performance Specifications
Precision: 1% of full scale
Noise:
1% of full scale at 0.02 absorption units 0.5% of full scale at 0.20 absorption units
Linearity:Standard 2% of full scale
Zero Drift: 0.5% of full scale per day for IR; 1.0% of full scale per/day for UV
Response Time:Programmable
Ambient Electronic Stability: 1% of full scale for 18F (10C) in 4 hours
Operating Specifications
Wavelength Range:Ultraviolet 200400 nm
Visible 400800 nmNear Infrared 8002500 nmFundamental Infrared 2.5014.50 um.
Ambient Temperature Range:32 to 113F (045C)
Max. Cell Heat:302F (150C)
Electric Cell Heat:Power Consumption 450 watts maximum(temperature is application dependant)
Sample Flow Rate:Typical 20500 cc/min for vapors;5120 cc/min for liquids
Sample Pressure:0500 PSIG (034 bar)
Voltage Input Variation:10% fluctuation without causing anoutput variation of 0.05% of full scale
Electrical Power To Analyzer:100/115/220/230 VAC, 4566 HZ,150W, Maximum Power Consumption600W With Electric Heat
Instrument Air Supply for Enclosure Purge
(Safety):Pressure 4080 psi (36 bars)Flow Rate 0.5 CFM (15L/min.)
Optical Purge, typically nitrogen:Pressure @ 1530 psi (12 bar)
Flow Rate 1015 cc/min
General Installation
Protect the instrument from direct sunlight andrain at operating temperature between 32113 F(045C)
Dimensions
Weight:
approximately 80 lbs (36.28 kg)Height:13.5 inches (342.9 mm)
Depth:10.5 inches (266.7 mm)
Width:10 inches (254 mm) per module
Overall Length:(based on cell pathlength)
Minimum26.8 inches (681.4 mm) for 0.5 to 16 mm cell
Maximum65.5 inches (1665 mm) for 1 meter cell
Tube Fittings
Sample Inlet/Outlet:
Metallic Cells Teflon Cell
Size: 1/4" 1/8"
Brand: Standard Gyrolok Gallek or Fluorocarb(Swagelok available)
Material: 316 SS TeflonHastelloy C, Monel
Purge Inlet and Outlet:Metal Cell Size: 1/4" NPT-F
Brand: Gyrolok
Material: 316 SS
Power:Size: 18 AWG, 3/4" Conduit Hub
Type: 3 conductors each
Output Signals
Analog Outputs:4 ea. 4-20 mA isolated into 600 ohms max.
Contact Closures:2 ea. Relay, 3W at 0.25 A or 28 VDC, 5 ea.isolated solid state. Both relay and solid statecontact closures N.O. or N.C.
Digital Inputs:8 ea., 2 ea. are dedicated
Digital Outputs:4 ea., 110 VDC, 25 watts ea., max. standard,110 VDC, 24 VDC optional.
Area Classification
NEC Class I, Groups B, C, D, Division 2 withoutenclosure purge; Division 1 with type Y-Purge;CSA Class I, Groups B, C, D, Division 2 with typZ-Purge;Division1 withtype X-Purge (CSA 15549CE Zone 2, II3G; EEx pz IIB+H2 T4 to T2;
CE Zone 1, II2G; EEx pd [ib] IIB+H2 T4 to T2(LCIE 03 ATEX 6007X)
7/23/2019 Teledyne Analyser
16/20
Process Photometers
PUV3402andPIR3502...
The Next Generation of On-Line
Process Photometers.
MultiwaveProcess Photometers
are designed to provide on-line meas-
urements of gas or liquid components,
in simple or complex process streams,
for process control, product quality
assurance, safety, catalyst protection
and area monitoring.
The Multiwave is a fixed filter photo-
meter that utilizes optical filters to
make continuous measurements. The
single beam, dual wavelength concept
used in the Multiwave compensates
for source and detector aging and the
obstruction of cell windows, while
allowing the sample cell to be isolated
from the electronics.
The Multiwave design takes the single
beam, dual wavelength concept one
important step further, by adding up
to eight filters to the filter wheel. Thisimproved method provides more meas-
urement solutions than a conventional
single component process photometer.
The use of multiple wavelengths ena-
bles the Multiwave to compensate
for numerous interferences and per-
form multiple component applications.
The newMultiwaveProcess Photo-
meters offer even more performance,
operating efficiency and versatility to
the user. The Multiwave line features
two basic models, each offering a wide
range of applications capability.
PIR3502 Multiwave Photometers
can be applied to Infrared and Near
Infraredmeasurements.
PUV3402 Multiwave Photometers
are designed for theUltraviolet and
Visible spectral regions.
Optical Schematic
Multiwave Photometers can be con-
nected to ABB AnalyticalsVistaNET
Process Analyzer Network, enabling
data interchange from the analyzer tothe DCS, seamless connectivity to the
Plant LAN and remote user access.
With the addition of VistaNET, all
operator functions may be performed
at the analyzer or from a remote PC.
7/23/2019 Teledyne Analyser
17/20
MultiwaveTM Process Photometer
VistaNET
Connectivity
The VistaNET Process Analyzer
Network is a local area network that
supports data interchange from
process analyzers to the Distributed
Control System (DCS), in a dedicated
and secured manner. VistaNET also
provides seamless connectivity to the
plant operating system (Plant LAN).
Through the Multiwave Remote
User Interface, the user can configure,
operate, or troubleshoot the operation
of the analyzer from a remote PC.
With this remote access capability all
operator functions may be performed.
Remote configuration and observa-
tion of the Multiwave photometer,
enabling technicians to use their
time more efficiently.
Graphic display of Absorbance vs.
Concentration data, allowing the
user to check the validity and
linearity of calibration samples.
Remotemaintenance viamodem
multiple users can view datasimultaneously and work together
in troubleshooting activities.
Reports and Tables can be printed
and reviewed at remote locations.
And VistaNet is designed for use
with widely used and accepted PC
hardware and operating systems,
including Windows 95, 98 and NT,
TCP/IP and other protocols. Connect-
ing to other network operating sys-
tems is simple and very economical*.
*Consult our VistaNET Brochure for more details.
Multiwave Spectral Regions
Infrared (IR) 2.5014.5 um
Near Infrared (NIR) 8002500 nmVisible (VIS) 400800 nmUltraviolet (UV) 200400 nm
Multiwave Process Photometers Leading Applications*
PIR3502
Infrared: Isocyanate in chloroaromatic solventAmbient air monitoringMulticomponent monomers in polymer process
Near Infrared: Measurement of Hydrogenic Compounds (C-H, N-H, O-H)Water in organics, such as Ethylene Dichloride (EDC)Caustic in Acid Gas Scrubbers
PUV3402
Visible: Color Measurements (ASTM, APHA, Saybolt Color Units)Ultraviolet: %T @ Multiple Wavelengths in Glycol Process
Chlorine in Phosgene
*Consult our Multiwave Applications Brochure
for more details.
7/23/2019 Teledyne Analyser
18/20
MultiwaveTM Process Photometer
Designed for Reliability
and Performance
PUV3402 and PIR3502 Multiwave Process Photo-meters are designed to meet all the significantchallenges of the process environment, includingambient temperature variation, moisture, corrosiveand explosive stream compositions, electricalhazards, dust and vibration.
Accessibility Every major component may
be easily removed and replaced, significantly
reducing maintenance time.
Solid State Detectors The PIR3502 and PUV3402
use solid state detectors. These detectors are
thermally stable and insensitive to vibration.
They require no mechanical adjustment and
minimal interconnecting circuitry. They provide
excellent linearity and long service life.
Brushless Chopper Motor Provides good
mechanical reliability and long service life.
Linearizer Circuit with Up to 8 Points Ensures
a 2% of full scale linearity and allows the user
to select the best accuracy for a specific range.
Isolated Sample Cell Prevents flammable andcorrosive streams from contacting electronics,
permits cell heating to optimum sample condi-
tions, allows easy access and optimum optical
pathlength selection.
Self-Aligning Optics Saves time during
maintenance operations.
Temperature Controlled Filter Assembly Eliminates temperature effects on filters,
improving long term stability.
Electric Cell Heating Concept The first ever
Division 1/Zone 1 electrically heated sample cell.
Employs a unique heat pipe concept that reduces
temperature gradients across the cell, providing
a more stable output and accurate temperature
control.
Data validation to ensure the reliability of
Multiwave data.
Self-diagnostics to aid in troubleshooting
operations.
Functional User Interface Direct user interface
at the analyzer is simple and easy using the
Multiwave's front panel keyboard and display
screen. And with built-in VistaNET capability,
all operator functions may also be performed
from a remotePC.
7/23/2019 Teledyne Analyser
19/20
ABB Inc.
843 N. Jefferson Street
Lewisburg
WV 24901
USA
Office: 304.647.4358
Fax: 304.645.4236
The IndustrialIT
wordmark and all mentioned product names
in the form XXXXXXIT
are registered or pending trademarks of ABB.
ABB has Sales & Customer Support expertise in over
100 countries worldwide.
The Companys policy is one of continuous
product improvement and the right
is reserved to modify the information
contained herein without notice.
2005 ABB Inc. BUASC 3-2-105
www.abb.com/analytical
7/23/2019 Teledyne Analyser
20/20
Series 6000
Ultraviolet Photometric Analyzer
Series 6000
Ultraviolet Photometric Analyzer
CONFIGURATIONS
6000: General purpose wall mount
6020: Totally explosion proof suitable for use in
Class I, Division 1, Groups B, C & D
Alternate configurations are available; contact the
factory.
SPECIFICATIONS
Noise: < 1%
Drift: Less than 1% per day
Diurnal: Less than 1% per 20F
Accuracy: 1% of full scale
Ambient: 32F to 122F
(0-50C) non-condensing
Sample cell: Stainless steel with quartz windowstandard. Other materials available.
Cell length: .01 to 40 inches
(application dependent)
Flow rate: 50 to 1500 cc/min
Light source: Tungsten lamp, mercury, deuterium
arc (application dependent)
Power rating: 115/230 VAC, 50/60Hz volts,approximately 200 watts
Sensitivity: .02 to 3 absorbance units
Optical System (figure 1)
SOURCELENSSAMPLE IN
FILTER WHEEL
DETECTOR
SAMPLE OUT
MOTOR
Reproducibility: 1% of scale or better
Electronic response: 90% in 1 second
Filter wavelength: 210 to 1000 nano-meters
Sample pressure: Quartz window - 30 psig
Sapphire - 40 psig
Enclosure:
(6000)General purpose - sheet steel
(6020) Hazardous duty - condulet
Readout: 2-line alphanumeric vacuum
fluorescent display (VFD)
2005 Teledyne Analytical Instruments, A Teledyne Technologies Company.All rights reserved. Printed in the USA. 11/05LD
Warranty
Instrument is warranted for 1 year against defects in materialor workmanship
NOTE: Specifications and features will vary with application. The above are establishedand validated during design, but are not to be construed as test criteria for every product.All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTSA Teledyne Technologies Company16830 Chestnut StreetCity of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: [email protected]
www.teledyne-ai.com