85
TECHNICIAN MANUAL Prevacuum Table - top Autoclave Models 2340, 2540 EPV

Technical_Manual_2340_-_2540__EVP[1]

Embed Size (px)

DESCRIPTION

medical

Citation preview

  • TECHNICIANMANUAL

    Prevacuum Table - top AutoclaveModels 2340, 2540 EPV

  • CAT. No. MAN205-0062 Rev. ACOPYRIGHT

  • TECHNICIANMANUAL

    Prevacuum Table - top AutoclaveModels 2340, 2540 EPV

    Cat. No. MAN205-0062 Rev. A Tuttnauer Europe b.v., P.O. Box 7191, 4800GD Breda, The Netherlands. + 31/76-5423510

  • 1TABLE OF CONTENTSPARAGRAPH PAGE NO.

    1. INTRODUCTION 42. SYMBOL DESCRIPTION 43. INSTALLATION INSTRUCTIONS 5

    3.1 Connections 54. TESTS 6

    4.1. Installation Tests 64.2 Periodical Tests 6

    5. WATER QUALITY 75.1 Water for Generating Steam 75.2 Reverse Osmosis 7

    6. DESCRIPTION OF THE CONTROL SYSTEM. 86.1 Digital Board DIG T2 86.2 Analog Board ANL-T2 106.3 AC - Board - AC-T1 11

    7. CALIBRATION OF TEMPERATURE AND PRESSURE 127.1 Method of Calibration 127.2 Temperature Calibration Procedure 137.3 Pressure Calibration Procedure 15

    8. TEST POINTS 169. CHANGING PARAMETERS 17

    9.1 General 179.2 Changing Parameters 179.3 Resetting the Back-Up Memory 239.4 Test Mode 23

    10. REPLACEMENT PROCEDURES 2410.1 Dismantling the Outer Covers of the Autoclave 2410.2 Replacing the Safety Valve 2510.3 Replacing the electronics board (control panel) 2610.4 Replacing the Electronics Box 2610.5 Replacing Heating elements 2810.6 Replacing the Temperature Safety Thermostat 2910.7 Replacing the Cut-Off Thermostat 3010.8 Cleaning and Replacing the Water Level Electrodes 3110.9 Replacing the Drain Valve 32

  • 2TABLE OF CONTENT (Cont.)PARAGRAPH PAGE NO.

    10.10 Replacing the Pressure Gauge 3310.11 Replacing the Locking Device 3410.12 Replacing the Printer 3510.13 Replacing the Door Switch 3610.14 Replacing the circuit breaker 3710.15 Vacuum Pump 3810.16 The Water Reservoir 3910.17 Replacement of the Door Cover 41

    11. TROUBLESHOOTING 4211.1 Displayed Messages and Reference to Trouble Shooting 4211.2 Pre-process malfunction 4311.3 In Process Malfunction 4911.4 Components Malfunction 5411.5 Mechanical malfunction 67

    12. LIST OF SPARE PARTS 6913. VALVES NUMBERING 79

  • 3TABLE OF CONTENT (Cont.)DRAWINGS PAGE NO.

    Rear View 5Control System Block Diagram 8DIG- T2 Board 9ANL T2 Board 10AC-T1 Board 11Electronic Box 27Water Reservoir 40Front View 74Vessel Assembly 75Door Tightening Bolt Assembly 76Outer Cabinet Assembly 77Water Outlet Strainer 78Piping Drawing 81Electrical Drawing 82

  • 41. INTRODUCTIONThis manual, together with the operators manual, forms the complete editionof the Operation and Maintenance instructions. This manual is intended for theuse of the technician. It is forbidden for unqualified and unauthorizedpersonnel to service the autoclave in accordance with the instructions in thismanual. Any unauthorized service may result in the invalidation of themanufacturers guarantee.The qualified technician shall be an authorized electrician with the rightqualifications in electronics and shall be familiar with the localtechnical/electrical regulations.

    2. SYMBOL DESCRIPTION

    Caution! Consult accompanying documents

    Caution! Hot surface.

    Caution! Hot steam.

    Protective earth (Ground)

    Stand by

  • 53. INSTALLATION INSTRUCTIONSNetwork and connection should comply with the device consumption. It mustcomply with local installation and safety rules and regulations.The voltage supplied to the device must comply with the label + 5%.In order to avoid any injury by electrical hazard, it is mandatory for thecustomer to have installed an earth leakage relay in the electrical board towhich the autoclave is connected.This relay disconnects all the poles of the electric power line in case ofaccidental contact with the instrument metal enclosure, by the operator oranother person, leading to a dangerous leakage current.Set the hind legs of the autoclave to give the body an inclination of 2 towardsthe front; this ensures that water is completely drained out of the chamberthrough the opening on the bottom in front of the chamber. To check pour aglass of water into the bottom of the autoclave ensuring water flows out of thechamber.Ensure while placing the autoclave, to leave space around the machine, to givethe technician access to service the machine also making sure the autoclave ison a leveled and rigid surface.Ensure the following connections to the utilities sources and drain (refer to thedrawing REAR VIEW below).3.1 Connections

    Connect the exhaust from pump to the drain by 5/16 tube. Connect the drain from the chamber, by a 1/4 tube.Note: All the connections listed above will be executed with flexiblehoses, which are supplied with the autoclave.Connect the power cord of the autoclave to a power outlet, 230V-16A.

    REAR VIEW

  • 64. TESTS4.1. Installation Tests

    The service technician shall perform the followingpreliminary checks before operating the autoclave:

    a. Leveling CheckCheck that the autoclave is leveled.

    b. Leakage current test;Check the precise operation of the earth leakage relay.

    c. Continuity CheckCheck the continuity of the grounding connection.

    d. Safety CheckCheck the safety elements; safety valve and the door lockingmechanisms.

    e. Programs CheckRun basic programs of the autoclave and check the operationsequences, the sterilization parameters etc.

    f. Integrity CheckPerform a visual check to verify that there are no dents,scratches, broken gauges, etc.

    After the above steps are performed, the autoclave is ready for operation.4.2 Periodical Tests

    PERIOD TEST1 month Test the safety valve by operating it.6months

    Remove the cover of the autoclave, tighten the screws of the heaters and theelectrical connections at the heaters, valves and connectors in the control box.Check the continuity of the grounding connections.Check the temperature and pressure calibration.Perform validation of the autoclave.Check the precise operation of the earth leakage relay.Check that the autoclave is levelled.Check the safety elements; safety valve, safety and cut-off thermostats doorlocking mechanisms.Run basic programs of the autoclave and check the operation sequences, thesterilization parameters etc.Check the water reservoir, piping, plastic parts and electric wires.check and tighten the piping joints to avoid leakage.check and tighten all screw connections in the control box, heaters andvalves and instrumentation.

    Year

    Observe the closing device for excessive wear5 years Observe the closing device for excessive wearSafety tests (pressure vessel, efficiency, electrical) shall be performed in accordancewith local rules or regulations, by an authorised inspector.

    An authorized technician shall perform the 6-months and yearly tests!

  • 75. WATER QUALITY5.1 Water for Generating Steam

    The distilled or mineral free water supplied to the sterlizer shall beaccording to the table below:

    Physical Characteristics and Maximum acceptable contaminantslevels in water or steam, for sterlizers

    (In compliance with ISO 11134 and ISO 13683).Evaporate residue < 15 mg/lSilica < 2 mg/lIron < 0.2mg/lCadmium < 0.005 mg/lLead < 0.05 mg/lRest of heavy metals < 0.1 mg/lChloride < 3 mg/lPhosphate < 0.5 mg/lConductivity < 50 s/cmpH 6.5 to 8Appearance colourless, clean, without sedimentHardness < 0.1 mmol/l

    Compliance with the above data should be tested in accordance withacknowledged analytical methods, by an authorized laboratory.

    Attention:The use of water that do not comply with the table above may havesevere impact on the working life of the sterilizer and can invalidatethe manufacturers guarantee.

    5.2 Reverse OsmosisA Reverse Osmosis system may be used to improve the quality of thewater used to generate steam in the autoclave chamber. The use ofmineral free will contribute to better performance and longer life of theautoclave.

  • 86. DESCRIPTION OF THE CONTROL SYSTEM.(See CDR Control diagram).The control system is based on 3 electronic boards designed according to theautoclave requirements, the digital board DIG-T2 containing themicro-controller memories, buffers and digital ICs and the analog boardANL-T2 which performs the processing of signals coming from the sensorsand switches. The AC-T1 board consists of filters and AC drivers. DCvoltages 12V and 5V, supplied by a switching type power supply, powers the3 electronic boards.

    The system is provided with communication interfaces RS 232 to PC and toparallel printer.

    6.1 Digital Board DIG - T2 The digital board is connected to the keypad panel, to the parallel

    printer and to the analog board ANL-T2. The board contains the micro-controller (U15) type 80C32 that

    runs the software program of the system. On the board, are three types of memories:

    1) EPROM memory (U17), part no. 27512 storing the programcodes.

    2) RAM memory (U21) with a capacity of 32KB for thetemporary data during the running of the program.

    3) EPROM memory (U22) that is a fixed serial memory with anelectrical writing and erasing.

    This EPROM serves as a non-volatile memory, enabling thesystem to change follow-up tables during running of programcodes, and ensuring this data is not lost in case of power failure.

    D S1

    BZ1

    97-05

    C222

    DIG T3

    1

    JP1

    D1

    RP1

    U18

    U8

    C2

    C3

    C4

    81 JP3

    R4

    U3

    Q1

    R2

    U4

    U19

    U1

    R1

    20

    19

    R3

    C16

    Y1

    C15R9

    U14U11

    37

    1

    SKT

    U17

    U16

    U12

    C17U21

    C9

    C7

    C6

    C5

    U15

    +

    +

    DIG-T2 BOARD

    3433

    C24C23

    U22C25

    37

    19

    20

    1

    PLUG

    JP1

    1

    JP81

    JP12

    1

    JP9

    1

    JP3

    1

    JP10

    1

    JP61JP7

    P1

    12

    1JP5

    JP2

    1

    JP4

  • 9 The board contains an optimal Real Time Clock (U12), whichserves as a clock to the system, including a back-up battery, whichensures the clock, runs continuously even when the autoclave isnot powered.This component is optimal because it is related to the operation ofthe printer, which is also an optional item.

    The board contains a watchdog, which detects any faulty situationin the running program code. It performs an automatic reset of themicro-controller and stops all the commands to avoid anuncontrolled activation of any of the heating elements or thevalves.

    The board functions as an MMI (Man-Machine Interface). It isconnected to a LCD display of two rows with 16 characters oneach row and to the following light indicators: START (autoclave in process), FAIL (the process failed), WATER (no water in the reservoir), DOOR (the door is not closed).

    A keyboard connected to the digital board, serves as a control panelcontaining the command and programming keys. The digital inputs and outputs are transferred to the system, as

    follows: Through the digital board to the analog board by means of buffers

    74HC377. RS232 interface is performed on the board by the U13 component,

    the signal is transmitted to the communication connector locatedon the ANL-T2 board.

    The printer is connected directly to this board, connector DS1enables to connect the DPU-20 printer manufactured by SEIKO.The printer receives the data and the supply voltage directly fromthis connector.

    The layout of the DIG-T2 board components is provided below

    C 1

    D S

    B Z 1

    97-05

    C 22

    D IG

    1

    C 1

    J P 1

    D 1

    R P

    U 1

    R PU 7U 8

    C 1

    C 2

    C 3

    C 4

    81 J P 3

    R 4

    JP

    C 1

    U 3

    Q 1

    R 1

    +

    R 2

    U 4

    U 1

    U 1

    U 2

    U 6

    R 1

    20

    19R 8 R 7 R 6 R 5

    R 3

    U 1

    C 16

    Y 1

    C 1R 9

    U 5

    U 1

    U 1

    U 1

    37

    1

    S KT

    U 1

    U 1

    U 1

    C 1

    R 1C 2

    1

    U 21

    C 9

    C 8

    C 7

    C 6

    C 5

    C 11

    U 1

    +

    +

    R 1C 2

    C 1

    C 1

    3433

    C 2C 2

    U 2C 2

    U 2

    H 2

    DIG- T2 BOARD

  • 10

    6.2 Analog Board ANL-T2The analog board contains the drivers of the valves, heaters and pump.It contains the sensors circuits connected to the control system, andserves as a junction to the autoclave connections.The input is 12VDC & 5VDC and the output is for all the signals to theautoclave. The power supply provides the DC voltages 12V and 5V through

    the connector JP3 to the board and further to the DIG-T2 board. The computer is connected to the board through the connector P1,

    by RS232 interface. The analog and digital inputs, from the temperature sensor,

    electrode door switch, etc. are connected through connector JP6,the conversion circuits of the sensors are located on this board.

    The pressure sensor MPX2200 that measures the chamber pressureis connected to the board through connector JP8, to the conversioncircuit of the sensor that is located on this board.

    The analog to digital conversion circuit A/D (U5) and the analogmultiplexer (U6) for 8 analog inputs (temperature, pressure, andelectrode) are located on this board and transfer digital signal tothe DIG-T2 board.

    All the drivers and power circuits to the solenoids (valves, doorlocking) are included on this board.

    Layout diagram of the ANL-T2 is provided below.ANL - T2 BOARD

    JP6

    12 1 1

    1

    JP11

    1

    JP2

    37

    19

    JP7

    20

    1

    JP9

    1

    PLUG

    JP1

    JP5

    JP3

    ANL T1 BOARD

    JP4

    1

    P1

    JP10

    1

    1

    JP12 JP8

    1 1

  • 11

    6.3 AC - Board - AC-T1This board receives command signals from ANL-T2 board designatedto activate AC devices. It converts them to AC mode.The AC-T1 board provides AC filtering for the power supply to thecontrol system and protection from sharp current fluctuations.The AC-T1 board includes: AC drivers for the AC devices (pump, heaters, etc.) up to 6A. AC - input connector JP3. Filter circuit based on L1C5, the power transfer from the filter to

    power supply via JP2. Spike protection - RV13. Three - 230/115 VAC output channels, triac control circuit.

    Q1 - 16Amp - JP4/1,2Q2 - 8 Amp - JP7/1Q3 - 8 Amp - JP7/2

    R8 C2

    JP7

    R7 C1

    1JP1

    R4

    R1

    R3

    U1

    ETYPEAC-97

    R2

    Q1

    RV1

    R12

    R5

    C3

    Q3

    R10U3

    R11

    R9

    L1

    RV3 RV2

    JP4 JP3

    1

    JP5

    C5

    R6U2

    1

    JP6

    JP2

    Q2

    (JP3,JP4,JP7 ON THE BOTTOM SIDE OF THE BOARD)

    CAT TECH.

    AC-T1 BOARD

  • 12

    7. CALIBRATION OF TEMPERATURE AND PRESSURE7.1 Method of Calibration:

    The calibration of temperature and pressure is performed digitally.The temperature and pressure measuring circuits are designed withcomponents having 1% accuracy.The temperature circuit is linear and has an output of 100mV2400mVfor a temperature range 20C150C.The pressure circuit is also linear and has an output of100mV2400mV for a pressure range 0400 kPa.The measuring at the A/D is limited for values higher than 2400mV orbelow 100mV.It is obvious that although the accuracy of the circuit components is1%, the sum deviation may reach 5%. Therefore calibration isnecessary.The system has a non-volatile memory in which the offset and gaindata of the sensors are stored.Calibration is performed by use of the keyboard (or the communicationport for computer use).Following is a description of the calibration of temperature andpressure using the keyboard. The calibration procedure is identical toall autoclave models.Definition of Gain and Offset:Every straight line can be defined by the equation ax+b=y. a is thedeviation from y=0 and b is the slope of the line. In our case a=Gainand b=Offset.The system is based on calculation of offset and gain of two points asin the following exampleIf the actual measured pressures are 100 kPa & 300 kPa and thedisplayed temperatures are 90 kPa & 310 kPa respectively, the inputVs pressure graph will be as follows.

    The calibration method enables to input this data into the system inorder to perform automatic correction of OFFSET and GAIN.It is necessary to define the two points, which have to be input to thesystem, prior to performing calibration.

    aX

    Y

    INPUT FROM SENSOR

    PRES

    URE

    V1

    310300

    V2

    10090

    READ

    ACTUAL

  • 13

    7.2 Temperature Calibration Procedure

    Press the PROGRAM key. The system enables to set the systems date the time. This operation is skipped by pressing PROGRAM key

    six times. CODE: will be displayed. Select Technician Code (011) by the UP/DN keys and press the

    PROGRAM key.

    Select CALIB CODE: 107 to calibrate the system by means of theUP/DN keys and press the PROGRAM key again untilTEMP CAL: will be displayed.

    Set calibration code to TEMP CAL: 1 Press the PROGRAM key again and do not change the

    code.

    The following data will be displayed:UPPER ROW: A 60.0 R 60.0LOWER ROW: A 130.0 R130.0

    Enter the actual data (A) and the displayed data (R) for two knownpoints as in the following example:The actual temperatures are 70C & 120C and the displayedtemperatures are 68C & 120C respectively. theProceed as follows: On the upper row, change A60.0 to A 70.0 by means of UP/DN

    keys. Press the PROGRAM key.

    The cursor on the upper row will move to R60.0, change to R68.0by means of UP/DN keys.

    Press the PROGRAM key. The cursor moves to the lower row to A 130.0, change it to A

    120.0 by means of UP/DN keys. Press the PROGRAM key.

    The cursor on the lower row will move to R130.0, change toR123.0 by means of keys UP/DN, then press thePROGRAM key

    This completes calibration procedure for temperature.

  • 14

    Remarks: When cursor is set under any figure, pressing the START/STOP

    key stops calibration.

    When the cursor is set under any figure, it enables to feed thisvalue to the reading of temperature by pressing thePARAMETERS key.

    Example:If the cursor is set under R68.0 on the upper row and the actualtemperature of the autoclave is now 66.0C, by pressing thePARAMETERS key, the temperature reading will be corrected fromR68.0 to R 66.0.This calibration method enables the technician to change a certain pointwithout affecting a second point.Example:When the working temperatures are 134C and 121C but the actualtemperature are 132C and 121CThe measured temperature for 134C must be corrected withoutchanging the temperature at 121C,Considering the above example, the setting of data can be done asfollows:

    A 121.0 R 121.0A 132.0 R 134.0

    The system will calibrate the new offset and gain and retain them intothe non-erasable memory.As a result, while running a 134C program the correction isautomatically made, while at 121C no change is felt.

  • 15

    7.3 Pressure Calibration ProcedureTo perform the pressure calibration two points have to be definedallowing the calculation of new OFFSET and GAIN values. Press the PROGRAM key. Skip date and the time setting by pressing the PROGRAM

    key six times. CODE: will be displayed. Select the Technician Code (011) by the UP/DN keys and press the

    PROGRAM key. Select CALIB CODE: 107 to calibrate the system by means of the

    UP/DN keys and press the PROGRAM key again until thesystem displays: PRES CAL:

    Press the PROGRAM key.The following is displayed:

    Point 1 Upper Row A 100 R 100Point 2 - Lower Row A 300 R 300

    Open the door, do not alter A 100. Press the PROGRAM key, the cursor will move to R 100. Press the PARAMETERS key. The system performs

    readingpressure and displays it instead of R100. E.g. R93 (i.e. the systemreads out 93kPa, when the door is open).

    Press the PROGRAM key. The cursor moves to the lower row to A300.

    If the system reads out 280kPa, do not change A300. Press thePROGRAM key to move to R300 and change it to theactualpressure value 280kPa by means of the keys UP/DN.

    Press the PROGRAM key again.The system calculates the new values for offset and gain and storesthese values in the non-volatile memory.

    This completes the calibration procedure for the pressure.

  • 16

    8. TEST POINTSThe test point list provides testing points on the junctions on board to assist inlocating the malfunction.

    NU TP FUNCTION VALVETP1 GND TP2 +5V DC TP3 +12V DC TP4 OUT - HEATERS ON-5V; OFF-0VTP5 OUT-VAC. PUMP ON-5V; OFF-0VTP6 TP7 OUT-AIR-RELEASE VALVE ON-5V; OFF-0VTP8 OUT-EXH-VALVE ON-5V; OFF-0VTP9 OUT-WATER TO RES. ON-5V; OFF-0V

    TP10 OUT-VACUUM VALVE ON-5V;OFF-0VTP11 OUTDOOR LOCK ON-5V; OFF-0VTP12 OUT-AIR RELEASE ON-5V; OFF-0VTP13 TP14 TP15 IN FLOAT HIGH. ON-5V; OFF-0VTP16 IN DOOR SWITCH ON-5V; OFF-0VTP17 IN THERMOSTAT ON-5V; OFF-0VTP 18 TP19 TP20 TP21 IN TEMPERATURE 134C 1.97VTP22 TP23 TP24 IN PRESSURE 2.0B 2.031VTP25 TP26 IN ELECTRODE HIGH 0V-YES; 2.5V NO

  • 17

    9. SOFTWARE PROGRAMMING PROCEDURES9.1 General

    The software for the control system of the EPV autoclave is EpvEn320.The software contains a table of parameters of which part of themdefines the autoclave, and part defines the processes in the autoclave.This document describes parameters and their task in the software.

    9.2 Changing ParametersAccess to the program is achieved by inserting the access code.Access code 1 This access code enables change of the sterilization

    time, sterilization temperature and the drying time.Access code 11 changes that may be done by the supervisor.

    Enables to change a few parametersAccess code 13 changes that may be done by the technician.

    Enables to change all the parametersTo change the parameters listed below, proceed as follows:a. Select the cycle.b. Enable the operation by entering the ACCESS CODE (11).c. Set the parameter that has to be modified, by pressing successively

    the PROGRAM key.d. Set the desired data by means of the UP/DN keys.e. Pressing the PROGRAM key enters the modified data into

    memory, and moves to the next parameter.The pre-set parameters values will replace the last default values andbecome the updated default values.9.2.1 STER Temp Required sterilization temperature for the

    processEntry Code 1Resolution 1CMinimum value for the change 105CMaximum value for the change 137C

    Cycle 1 2 3 4 5 6 7 8Set value Value 105 121 134 121 134 121 134 0

    9.2.2 STER Time Required sterilization time for the processEntry code 1Resolution 1 minuteMinimum value 3 minutesMaximum value 59 minutes

    Cycle 1 2 3 4 5 6 7 8Set value Value 10 15 7 15 18 30 3.5 0

  • 18

    9.2.3. Heat Standby Operation of the heating elements inStandby ModeAccess Code 1Resolution 1Value 0 or 1 If Heat Standby = 0, the heating elements will not

    operate in the Standby Mode. If Heat Standby = 1, the heating elements will operate in

    the Standby mode, in a shoot mode of 1 second on,59 seconds off.

    9.2.4 Dry Time Required drying time for the processEntry Code 1Resolution 1 minuteMinimum Value minutesMaximum Value 99 minutes

    Cycle 1 2 3 4 5 6 7 8Set value Value 1 1 10 30 30 0 2 09.2.5 Water Time Time for entering water to the autoclave

    This value defines the entry time of the water to the autoclave tolocate the electrode touching the water.This time will change from process to process and even fordifferent autoclaves.When the autoclave is vertical and the defined time is seconds, the water valve will not open at all.Entry Code 11Resolution 1 secondMinimum Value 1 secondMaximum Value 90 seconds

    Cycle 1 2 3 4 5 6 7 8Set value Value 7 7 7 7 7 7 7 39.2.6 Heat T.O. - The maximum time allowed for the heating stage

    (testing)After this time, if the system does not enter the sterilization stage,it will receive a Low Heat message and the process will abort.Entry Code 11Resolution 1minMinimum Value 20minMaximum Value 53 min

    Cycle 1 2 3 4 5 6 7 8Set value Value 3000 3000 3000 3000 3000 3000 3000 0

  • 19

    9.2.7 End Temp - The ending temperature of the processIf at the end of the cycle the temperature is higher than thistemperature, the process will not end and the door will remainclosed.Entry Code 11Resolution 1CMinimum Value 40CMaximum value 137C

    Cycle 1 2 3 4 5 6 7 8Set value Value 130 130 120 120 134 95 120 1349.2.8 PulsNum. No. of pulses in the prevacuum stage

    Entry Code 11Resolution 1Minimum Value 1Maximum Value 5

    Cycle 1 2 3 4 5 6 7 8Set value Value 1 2 2 2 2 2 2 19.2.9 VacDip1 Vacuum value in the first pulse

    Defines the vacuum value in pulse no.1 of the prevacuum stage.Entry Code 11Resolution 1kpaMinimum Value 1kpaMaximum Value 90kpaSet value 15 kpa

    9.2.10 VacDip2 Vacuum value in the remaining pulsesThis value defines the vacuum in the remaining pulses in theprevacuum stage (not just pulse no. 2)Entry Code 11Resolution 1kpaMinimum Value 1kpaMaximum Value 90kpaSet value 65 kpa

    9.2.11 Vac Time 1 Vacuum Time in the First PulseThis value defines the time the system will continue to maintainvacuum after reaching Vac Dip1 for the first pulse.Entry Code 11Resolution 30secMinimum Value 1secMaximum Value 1800secSet value 120sec

  • 20

    9.2.12 Vac Time2 Vacuum Time in the remaining pulses.This value defines the time the system will continue to maintainvacuum after reaching Vac Dip2 for the remaining pulses.Entry Code 11Resolution 30secMinimum Value 1secMaximum Value 1800secSet value 240sec

    9.2.13 Ex Mode The method for exhausting the steam at the end ofthe processEntry Code 11

    Ex ModeFast Ex 1N.A. 2Slow Ex (Waste) 3Slow Ex (Liquids) 4

    Cycle 1 2 3 4 5 6 7 8Set value Value 1 1 1 1 1 4 1 1

    9.2.14 LimitP add The allowable addition to the maximum pressurein every processIf the maximum pressure is 225kpa, the system will not heatbeyond this pressure. If LimitP add will be , the maximumpressure will remain 225kpa. If LimitP add equals 10, themaximum pressure will be 235kpa.Entry Code 11Resolutio 1kpaMinimum Value kpaMaximum Value 30kpaSet value 0kpa

    9.2.15 Calib Code Calibration CodeThis allows entry to the calibration mode.It must be set on 107 in order to allow the code to be entered.Entry Code 11Resolution 1Minimum Value Maximum Value 255Set value 100

  • 21

    9.2.16 Req Prs+This parameter defines the required addition to the sterilizationpressure in kpa.For example, for a sterilization temperature of 121C the requiredpressure is 204kpa. Since the system controls the sterilizationprocess according to pressure and temperature, if Req Prs+ equals, the system will maintain the pressure at 204kpa. If the valueis at 5kpa, the system will be maintained at 209kpa, and so on.Entry Code 11Resolution 1Minimum Value 0kpaMaximum Value 20kpa

    Cycle 1 2 3 4 5 6 7 8Set value Value 9 6 6 6 6 6 6 6

    9.2.17. Atmos. Pres - This parameter enables setting the pressureaccording to the ambient Atmospheric Pressure at the place inwhich the autoclave is operating.This value is derived from the altitude of the area. Theatmospheric pressure at Sea Level is 100kPa. For every 100mabove sea level, the atmospheric pressure drops 1kPa, and forevery 100m below sea level, the atmospheric pressure increases1kPa.Although the weather can affect the atmospheric pressure by1-2kPa, this will not affect the accuracy of the system.Entry Code 13Resolution 1kPaMinimum Value 55kPa (for +4500m)Maximum Value 108kPa (for 800m)*Set value 93kPa

    * Although altitude 800m does not exist, this parameter allowsfor the lowest altitude, plus adjustment range due to weatherchanges.

    9.2.18. Dry Mode rate of breaking vacuumThis parameter is used to set the rate of breaking the vacuum bycontrolling the air valve opening. 0 is closed valve and 5 iscompletely opened valve.Access Code 13Resolution 1Minimum Value 0Maximum Value 5

    Cycle 1 2 3 4 5 6 7 8Set value Value 1 1 0 0 1` 0 1 1

  • 22

    9.2.19. Puls Press pulse pressure during pre-vacuum stage.This parameter is used to set the maximum pressure in eachpulse of the pre-vacuum stage.Access Code 13Resolution 10Minimum Value 110Maximum Value 200Set value 140kPa

    9.2.20. Heat Step heating rate during sterilizationThis parameter is used to set the sterilization heating-rate. Theheating rate is constant until the pressure in the chamber reachesa pressure of 30 kPa below the required sterilization pressure.From this point on heating is performed in pulses. Each pulselasts 45 seconds and increases the pressure in the chamber byX kPa. Increasing the value of X will increase thechambers heating rate and decreasing the value of X willdecrease the heating rate.Access Code 13Resolution 1Minimum Value 1Maximum Value 10Set value 5

    9.2.21. Auto. Num Autoclave serial numberThis parameter is used to set the serial number in the facility.This number enables the operator to identify each autoclave.This number is printed on the autoclaves printout.Access Code 13Resolution 1Minimum Value 1Maximum Value 99

  • 23

    9.3 Resetting the AutoclaveWhenever certain data, stored in the back-up memory, must be erasedthe system must be reset, in order to restore the system to normaloperation or for situations that follow: When the machine is operated for the first time or after a long

    interruption, and disorders appear in the operations sequence. When operation was stopped in the middle of the cycle, as a result

    of a power failure or manual stop and the cycle has not beencompleted.

    To reset the system; proceed as follows: Turn the main power switch OFF. Turn the main power switch ON, while pressing the SEL. CYCLE

    key, until the program parameters are displayed.After resetting the autoclave, the autoclave is in stand-by mode

    9.4 Test ModeTurn on the main power switch, while pressing the UP keycontinuously, to enter the Test mode.The output test will start (valves, heater, pump), each output will beactivated for one second and its name will be displayed. To switchfrom one output to another, press each time the UP key.Keep pressing the UP key, the inputs test is performed, pressing eachtime the UP key the next input is selected.The same way the temperature and pressure values are displayed.

  • 24

    10. REPLACEMENT PROCEDURES 10.1 Dismantling the Outer Covers of the Autoclave

    Caution!Before starting, disconnect the instrument from the power sourceand ensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.1. Remove the screws holding the rear cover (1).2. Remove the screws holding the cover to the base (2).3. Dismantle the air filter from the service opening cover (3).4. Remove the screws holding the filter.5. Remove the grounding wires from the cover.6. Pull the cover upwards.

    3

    21

  • 25

    10.2 Replacing the Safety ValveCaution!Before starting, be sure that the electric cord is disconnected and thatthere is no pressure in the autoclave.1. Remove the autoclave cover (see para. 10.1 Dismantling the

    Outer Covers of the Autoclave).2. Remove the water reservoir cover.3. Unscrew the safety valve and remove it from the safety valve base.4. Replace the valve with a new safety valve (install an original

    only!). use Teflon tape for sealing the thread.5. Perform one cycle and verify that the valve operates correctly.

    Safety valve

    Safety valve base

  • 26

    10.3 Replacing the electronics board (control panel)Caution!Make sure that the power cord is disconnected!If the electronic box was damaged by any liquid that entered the box,do not attempt to replace it. In this case only the factory techniciansmay repair the autoclave.Allow the autoclave to cool before removing outer covers.1. Remove the autoclave cover (see para. 10.1 Dismantling the

    Outer Covers of the Autoclave).2. Remove the screws that connected the plastic panel with the digital

    electronic board to the autoclave (2).3. Disconnect the flat cable from DIG-T1 to ANL-T1 (4).4. Disconnect the flat cable of communicator from P1-ANL-T1 (6).5. Disconnect the grounding cable shoe (green cable) (5).6. On autoclave model 2540 disconnect the main switch (7).7. Remove the plastic panel, including the electronics board (control

    panel), from the autoclave.8. If replacement of the control panel is required, unscrew four

    fastening screws located on the rear of the panel. Remove thecontrol panel from the plastic panel and replace it with a new one.

    10.4 Replacing the Electronics BoxCaution!Make sure that the power cord is disconnected!If the electronic box was damaged by any liquid that entered the box,do not attempt to replace it. In this case only the factory techniciansmay repair the autoclave.Allow the autoclave to cool before removing outer covers.1. Remove the autoclave cover (see para. 10.1 Dismantling the

    Outer Covers of the Autoclave).2. Remove two screws from the bottom side and from the top (1, 3).3. Disconnect all the electric system connectors from the electronics

    base (including the MPX connectors).4. Remove electronics box and replace with new one.5. Re-assemble the instrument. Verify that the grounding connections

    are connected correctly.6. Test any cycle and verify that the autoclave operates as required.

  • 27

    Electronic box

    632

    4

    5

    71 1

    Electronic boxControl panel

    72

  • 28

    10.5 Replacing Heating elementsCaution:Before starting, be sure that the electric cord is disconnected from thepower source and that there is no pressure in the autoclave chamber.1. Remove the autoclave cover (see para. 10.1 Dismantling the

    Outer Covers of the Autoclave).2. Remove the insulation.3. Remove the heating element tightening bolts (1).4. Release the two terminal wires from the heating element.5. Replace the heating element (s)

    5.1 Verify that the upper thermo-couple tube is connected to theupper (cut-off) thermostat and that the lower thermo-coupletube is connected to the lower (safety) thermostat.

    6. Ensure that the heating element strap is well tightened to theautoclave body, ensuring proper heat dissipation from the heatingelement.

    7. Re-assemble the autoclave insulation and cover.8. After replacing one or more heating elements run a cycle and

    verify that it operates as required.9. Retighten the fixing screws of the heaters. This operation is done

    with the autoclave hot, but with the power cord disconnected fromthe power outlet and autoclave door open, to ensure that thechamber is not pressurized.

    10. Test the autoclave by performing a full cycle.

  • 29

    10.6 Replacing the Temperature Safety ThermostatThe autoclave is supplied with a temperature thermostat, whichprotects the heaters and autoclave against overheating, during the drycycle.This device reconnects automatically when the chamber cools down.CautionBefore starting, disconnect the instrument from the power source andensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.The temperature safety thermostat is located on the lower side of thefuse and socket panel on the rear of the autoclave.1. Remove the rear cover.2. Loosen the heating band.3. Unscrew the thermostat and replace it with a new one.

  • 30

    10.7 Replacing the Cut-Off ThermostatThis thermostat cuts out power to the autoclave, in the event that allother safety means do not function.For example: If the safety thermostat is defective and the temperaturecontinues to rise, then the cut-off thermostat cuts out the power to theautoclave. In order to restart the operation press the Reset Button. If theautoclave is operated according to the instructions, and the thermostatagain cuts out, the Cut-Off Thermostat must be replaced.CautionBefore starting, disconnect the instrument from the power source andensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.The Cut-Off Thermostat is located on the upper side of the fuse andsocket panel on the rear of the autoclave.1. Remove the rear cover.2. Loosen the heating band.3. Unscrew the thermostat and replace it with a new one.

  • 31

    10.8 Cleaning and Replacing the Water Level ElectrodesThe water level electrode is located at the rear bottom area of thechamber.

    10.8.1 ReplacingCautionBefore starting, disconnect the instrument from the powersource and ensure that there is no pressure in theautoclave.Allow the autoclave to cool before removing outer covers.

    1. Remove the rear cover.2. Disconnect the wire from the electrode connection.3. Open the nuts that tighten the electrodes.4. Insert a new electrode and tighten the fixing bolt to

    avoid leakage.5. Reconnect the wire to the electrodes.6. Test the unit.

    10.8.2 Cleaning1. Pull out the trays and tray holder.2. Clean the electrode tip.3. Test the unit.

  • 32

    10.9 Replacing the Drain ValveCaution!Before starting, disconnect the instrument from the power source andensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.1. Remove the autoclave cover (see para. 10.1 Dismantling the

    Outer Covers of the Autoclave).2. Disconnect the drainpipe from the valve, using a 9/16 wrench.3. Remove the nut (3) and the ring for drain valve (2).3. Remove the drain valve (1) from the panel.4. Install a new valve according to the drawing below.5. Verify that there is no leakage.

    CMT240-0020 CMT240-0003 VLV170-0066

    Item Cat No.1 SRV000-02242 SRV000-0232

  • 33

    10.10 Replacing the Pressure GaugeCaution!Before starting, disconnect the instrument from the power sourceand ensure that there is no pressure in the autoclave.1. Remove the door cover.2. Remove the pressure gauge from the door.3. Install the new pressure gauge using Teflon tape for sealing the

    thread.4. Operate the autoclave and verify that there is no leakage.5. Install the door cover.

  • 34

    10.11 Replacing the Locking DeviceCaution!Before starting, disconnect the instrument from the power sourceand ensure that there is no pressure in the autoclave.1. Remove the door cover.2. Disconnect the wires from the connector.3. Unscrew the tightening nut (cat. No. CMT201-0003).4. Remove the pin, the spring and the solenoid.5. Replace the damaged items and reinstall according to drawing

    below.6. Operate the autoclave and verify that the pin is retracted when the

    cycle reaches stand-by stage.

  • 35

    10.12 Replacing the PrinterCaution!Before starting, disconnect the instrument from the power source.

    1. Remove the autoclave cover (see para. 10.1 Dismantling theAutoclaves Outer Covers).

    3. Disconnect the contactor of the flat cable (1) connecting theelectronic box to the printer.

    4. Unscrew the two screws (2) fastening the printer to the panel.5. Remove the printer and install the new one.6. Verify connection of power by performing a self-test.7. Run a cycle and verify that the printer operates correctly.

    1

    2

  • 36

    10.13 Replacing the Door SwitchCaution!Before starting, disconnect the instrument from the power sourceand ensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.

    1. Remove the autoclave cover (see para. 10.1 Dismantling theOuter Covers of the Autoclave).

    2. Disconnect the wires (1), (2) from the door switch (3).3. Remove the microswitch and replace it with a new one.4. Reconnect the wires the microswitch. Verify that the wire is placed

    on the isolating cover (4) and does not touch the chamber.5. Reassemble the door cover.6. Test the connection with an ohmmeter. In open position the

    ohmmeter shows disconnection and in close position theohmmeter shows connection.

    3

    2

    1

    4

  • 37

    10.14 Replacing the circuit breakerCaution!Before starting, disconnect the instrument from the power source.

    1. Remove the autoclave cover (see para. 10.1 Dismantling theOuter Covers of the Autoclave).

    2. Disconnect the wires from the circuit breaker.3. Remove the four screws connecting the circuit breaker to the panel

    (1).4. Replace the circuit breaker with a new one.5. Reconnect the electrical wires.6. Reassemble the cover.7. Turn on the autoclave and verify it operates correctly.8. Move the circuit breakers lever to the tripped position and

    verify that the autoclave turns off.

  • 38

    10.15 Vacuum PumpFor generation of vacuum in the prevacuum and dry stages, thementioned models are equipped with a diaphragm vacuum pumphaving the following technical specifications:To improve the operation of the vacuum pump a condenser has beenfitted in the vacuum circuit, as described in detail in the next paragraph.In order to reduce the noise produced by the operation of the pump, asilencer delivered with the autoclave, can be mounted at the exhaustoutlet of the pump and connected to the drain by silicone flexibletubing.

    Caution:Make sure the silencer is never submerged in water andis leveled at least equal or below the exhaust outlet fromthe pump.

  • 39

    10.16 The Water Reservoir10.16.1 description

    The water reservoir is a one-compartment reservoir thatsupplies water to the chamber to generate steam. Thereservoir is equipped with two outlets at the bottom. Oneoutlet is connected to the chamber to supply water togenerate steam. The second outlet is connected to the drainvale to enable draining the reservoir. The reservoir must besupplied with water conforming to requirements specified inparagraph 5.1 (Water for Generating Steam).A float switch is mounted on the wall of the reservoir. Thefloat stops the cycle when the water level drops below thefloat, to avoid working without water.Flexible tubing connects the aeration inlet of the vacuumpump to the upper side of the pump reservoir.

    10.16.2 ReplacementIt is possible to replace damaged items on the reservoir or toreplace the complete assembled reservoir. Since the items onthe reservoir tend to deteriorate after a long period, it isrecommended to replace the complete assembly.Caution!Before starting, disconnect the instrument from thepower source.1. Drain the reservoir2. Remove the autoclave cover (see para. 10.1

    Dismantling the Autoclaves Outer Covers).3. Disconnect the wires from the reservoir.4. Disconnect the piping from the reservoir.5. Remove the reservoir from the autoclave.6. If required, replace the damaged item on the reservoir7. Reassemble the repaired reservoir or a new reservoir,

    as appropriate.8. Reconnect wiring and piping.9. Reassemble the cover.10. Turn on the autoclave and verify that there are no leaks

    and that it operates correctly

  • 40

    Water reservoir

  • 41

    10.17 Replacement of the Door CoverCaution:Before starting, be sure that the electric cord is disconnected from thepower source and that there is no pressure in the autoclave chamber.1. Unscrew the four screws attaching the door cover and remove the

    door cover. Since the screw pressing the door microswitchincludes two washers and a spring, be aware not to lose them.

    2. Reassemble the new cover.3. Insert screw (1) until dimension A is approximately 15 mm.4. Perform final adjustment of the screw as follows:

    4.1 While the autoclave is disconnected from electricity turn onthe circuit breaker.

    4.2 Connect the electrical plug to a multi-meter.4.3 Press the microswitch and verify that the microswitch

    functions as required.4.4 Close the door and verify that the microswitch operates.4.5 If the microswitch does not operate unscrew the screw one

    turn counter-clockwise and check per para. 4.4. Repeat untilmicroswitch operates.

    4.6 Connect the autoclave to electricity.4.7 Close the door until the microswitch indicates that the door is

    closed. Operate the autoclave and verify that there is no steamor pressure leak.

    4.8 If there is a steam leak stop the autoclaves operation, reducesteam pressure, open the door and turn the screw one turnclockwise and check per para. 4.7. Repeat until leakage

    ceases.3

    122

    A

    4

    5

  • 42

  • 42

    11. TROUBLESHOOTING11.1 Displayed Messages and Reference to Trouble Shooting

    Message Refer to trouble shooting sectionPara.11.2

    Para.11.3.1

    Para.11.3.2

    Para.11.3.3

    Para.11.3.4

    Para.11.4

    Para.11.5

    Low Temp. 1 15, 20

    Low Heat 1 4, 5 12, 16 3

    High Temp. 1 1 1 6, 7, 16,19, 21 1

    Low Pres. 16, 7, 16,19, 20,21

    3

    High Pres. 1 11,16,18, 19 1

    Add Water 7 3, 25

    Door Unlock 2 24

    Low Water 4 31, 6, 7,22, 25,26, 27

    1, 2

  • 43

    Corr

    ectio

    ns1.1

    Repla

    ce da

    mage

    d cord

    or re

    pair f

    aulty

    conn

    ection

    s.1.2

    Reset

    cut-o

    ff.

    1.3Re

    turn t

    he ci

    rcuit b

    reake

    r leve

    r to th

    eO

    N po

    sition

    after

    fixing

    the c

    ause

    forthe

    short

    circu

    it.1.4

    Repla

    ce the

    blow

    n fuse

    after

    fixing

    the

    cause

    for th

    e sho

    rt circ

    uit.

    Fa

    sten c

    ontac

    ts.

    Re

    store

    powe

    r sou

    rce.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts1.1

    Check

    the p

    ower

    cord

    and t

    heco

    nnect

    ions

    at the

    autoc

    lave i

    nput.

    1.2Ch

    eck th

    e cut-

    out th

    ermost

    at.If

    opera

    tion c

    ontin

    ues a

    fter r

    esettin

    g the

    cut-o

    ut. fin

    d out

    what

    caused

    the

    therm

    ostat

    to sw

    itch-o

    ff.1.3

    Check

    the c

    ircuit

    break

    er. If

    it trip

    ped,

    there

    is a s

    hort-

    circu

    it.1.4

    Check

    the f

    use of

    the p

    ower

    supply

    . Ifthe

    fuse

    is blo

    wn, c

    heck

    if on

    e of th

    eva

    lve co

    ils or

    one o

    f the e

    lectro

    nic bo

    ards

    causes

    a sho

    rt circ

    uit.

    1.5Ch

    eck th

    e AC

    volta

    ges.

    Th

    e inp

    ut sup

    plies

    volta

    ge at

    the

    contr

    ol sys

    tem - S

    ee Ele

    ctrica

    l Wirin

    gDi

    agram

    .

    Switc

    h off

    the po

    wer o

    f the u

    nit.

    Se

    t the m

    ultite

    ster o

    n the

    AC

    volta

    ge

    measu

    ring r

    ange

    .

    Conn

    ect th

    e neg

    ative

    prob

    e at p

    in 3 t

    oAC

    -INPU

    T con

    necto

    r and

    the p

    ositiv

    e pro

    be to

    pin 1

    or 2

    on A

    C-T1

    board

    .

    Turn

    on th

    e pow

    er. Th

    e ind

    icatio

    n of

    the in

    strum

    ent m

    ust be

    the s

    ame a

    s the

    powe

    r netw

    ork vo

    ltage

    . if no

    t:

    Main

    powe

    r swi

    tch\R

    ocke

    r swi

    tch

    conta

    cts m

    ay be

    loose

    .

    Electr

    ical p

    ower

    source

    may

    be

    faulty

    .

    11.

    TROU

    BLES

    HOOT

    ING

    11.2

    P

    re-pro

    cess m

    alfun

    ction

    Symp

    tom1.

    No re

    spon

    se wh

    en th

    esy

    stem

    is tur

    ned o

    n

  • 44

    Corr

    ectio

    ns

    Fix

    conn

    ection

    s.

    Re

    place

    faulty

    cable

    s.

    Re

    place

    faulty

    conn

    ector.

    Re

    place

    faulty

    conn

    ector.

    Re

    place

    faulty

    powe

    r swi

    tch

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts

    It is p

    ossibl

    e tha

    t the d

    ual p

    ower

    suppli

    es ou

    tputs

    12V,

    but th

    e 5V

    circu

    it is f

    aulty

    . To i

    denti

    fy the

    fault

    y ite

    m pro

    ceed a

    s foll

    ows:

    Ch

    eck th

    e con

    necti

    on be

    tween

    the

    two e

    lectro

    nic bo

    ards A

    NL-T

    1 and

    DI

    G-T2

    .

    Check

    the c

    able,

    whic

    h tran

    sfers

    the

    volta

    ge fr

    om th

    e AC-

    T1/JP

    2 co

    nnect

    or to

    the po

    wer s

    upply

    .

    Check

    the i

    nput

    volta

    ge of

    the e

    lectric

    box,

    at the

    AC-

    INPU

    T con

    necto

    r, on

    the A

    C-T1

    board

    /JP3.

    Ch

    eck if

    there

    is no

    short

    -circu

    it at th

    eAC

    -OUT

    PUT c

    onne

    ctor o

    n the

    AC

    -T1 b

    oard

    (JP2).

    Ch

    eck if

    there

    is no

    short

    -circu

    it at th

    einp

    ut co

    nnect

    or (JP

    3) or

    at the

    outpu

    t co

    nnect

    or (JP

    5) on

    the A

    NL-T

    1 bo

    ard.

    1.6Ch

    eck th

    e DC

    volta

    ges s

    upply

    to th

    esys

    tem.

    Ch

    eck m

    ain po

    wer s

    witch

    .1.6

    .1Ch

    eck th

    e grou

    nding

    Plu

    g con

    necto

    r 26 p

    in FL

    AT of

    the

    test b

    oard

    into c

    onne

    ctor J

    P5 m

    ounte

    don

    board

    ANL

    -T1.

    Co

    nnect

    the n

    egati

    ve pr

    obe o

    f the

    multit

    ester

    to TP

    1 on t

    he te

    st bo

    ard.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.2

    Pre-

    proces

    s malf

    uncti

    on (c

    ont.)

    Symp

    tom1.

    (cont.

    )

  • 45

    Corr

    ectio

    ns

    Tig

    hten l

    oose

    screw

    s.

    Reco

    nnect

    grou

    nding

    cable

    .

    Re

    pair o

    r rep

    lace t

    he fa

    ulty i

    tem.

    Re

    place

    faulty

    powe

    r sup

    ply.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts

    Check

    with

    the p

    ositiv

    e prob

    e of

    the m

    ultite

    ster th

    e volt

    age a

    t diff

    erent

    groun

    ding p

    oints

    of the

    autoc

    lave,

    on

    the A

    NL-T

    1 and

    DIG

    -T2 b

    oards

    , on

    the en

    closur

    e of th

    e elec

    tric bo

    x and

    the

    main

    grou

    nding

    screw

    of th

    e ma

    chine

    .Th

    e volt

    age d

    iffere

    nces

    must

    be le

    ss tha

    n 5mV

    .Hi

    gher

    volta

    ge m

    ay in

    dicate

    that

    ANL-

    T1 bo

    ard fa

    stenin

    g scre

    ws ar

    e loo

    se or

    that th

    e pow

    er sup

    ply ca

    ble is

    loose.

    1.6.2

    Check

    the 1

    2VDC

    volta

    ge.

    Co

    nnect

    the n

    egati

    ve pr

    obe o

    f the

    multit

    ester

    to TP

    1 (GN

    D).

    Co

    nnect

    the p

    ositiv

    e prob

    e of th

    e mu

    ltitest

    er to

    TP3.

    The i

    ndica

    tion o

    f the

    instr

    umen

    t must

    be 10

    .8-12

    .5V.

    If

    indica

    tion i

    s not

    as spe

    cified

    , dis

    conn

    ect th

    e outp

    ut co

    nnect

    or of

    thepo

    wer s

    upply

    . If th

    e righ

    t valu

    e is

    achiev

    ed th

    er is

    a sho

    rt circ

    uit.

    Ide

    ntify

    and l

    ocate

    the s

    hort c

    ircuit

    .If

    instru

    ment

    indica

    tion i

    s belo

    w or

    abov

    e spe

    cified

    rang

    e the

    powe

    r sup

    ply is

    fault

    y.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.2

    Pre-

    proces

    s malf

    uncti

    on (c

    ont.)

    Symp

    tom1.

    (cont.

    )

  • 46

    Corr

    ectio

    ns

    1.6.4

    calibr

    ate vo

    ltage

    to 5.

    2V by

    adjus

    ting

    the in

    terna

    l pote

    ntiom

    eter. T

    his op

    eratio

    nmu

    st be

    done

    with

    great

    care

    to av

    oidele

    ctrica

    l sho

    ck.

    Re

    place

    faulty

    coil a

    nd re

    set fu

    se.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts1.6

    .3If

    neith

    er 12

    V no

    r 5V

    volta

    ges a

    reou

    tput:

    C

    heck

    the v

    oltag

    e tran

    sfer to

    the

    powe

    r sup

    ply th

    rough

    the c

    onne

    ctor

    JP2 o

    n the

    AC

    - T1 b

    oard.

    1.6.4

    Check

    the 5

    VDC

    volta

    ge.

    Co

    nnect

    the n

    egati

    ve pr

    obe o

    f the

    multit

    ester

    to TP

    1 (GN

    D).

    Co

    nnect

    the p

    ositiv

    e prob

    e of th

    e mu

    ltitest

    er to

    TP2.

    The i

    ndica

    tion o

    f the

    instr

    umen

    t must

    be 4.

    95.3

    V.

    Check

    if the

    fuse

    of the

    powe

    r sup

    ply

    is blo

    wn. It

    is po

    ssible

    that

    one o

    f the

    valve

    coils

    or th

    e elec

    tronic

    board

    cau

    se a s

    hort c

    ircuit

    .

    Disco

    nnect

    the p

    ower

    from

    the

    AC-T

    1 boa

    rd/co

    nnect

    or JP

    2, an

    d sw

    itch o

    n the

    unit.

    If in

    no-lo

    ad

    cond

    ition t

    he po

    wer s

    upply

    deliv

    ers

    12V

    and 5

    V vo

    ltage

    s, it c

    an pr

    obab

    ly be

    that

    there

    is a p

    roblem

    on on

    e of

    the bo

    ards, o

    ne of

    the i

    nputs

    or

    outpu

    ts.

    Conn

    ect th

    e AV-

    T1/JP

    2 con

    necto

    r ba

    ck to

    the b

    oard

    and d

    iscon

    nect

    the

    conn

    ectors

    JP6 (

    inputs

    ), JP5

    (ou

    tputs)

    , pres

    sure s

    ensor

    .

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.2

    Pre-

    proces

    s malf

    uncti

    on (c

    ont.)

    Symp

    tom1.

    (cont.

    )

  • 47

    Corr

    ectio

    ns

    Re

    place

    faulty

    unit.

    Re

    place

    faulty

    board

    .

    2.1Re

    conn

    ect ca

    ble.

    2.2Re

    place

    the D

    IG-T

    2 boa

    rd

    3.1Re

    place

    DIG-

    T2 bo

    ard.

    4.1Re

    conn

    ect ca

    ble.

    4.1Re

    place

    the ke

    yboa

    rd.4.2

    Repla

    ce the

    DIG

    -T2 b

    oard.

    5.1Ca

    librat

    e sen

    sor an

    d che

    ck ca

    librat

    ionacc

    uracy

    durin

    g the

    opera

    tion.

    5.2Th

    e sen

    sor is

    fault

    y. Re

    place

    sensor

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts

    Reap

    ply th

    e pow

    er fro

    m the

    ANL

    -T1

    board

    .

    If the

    volta

    ges 1

    2V an

    d 5V

    reapp

    ear,

    the pr

    oblem

    is on

    e of th

    e exte

    rnal

    eleme

    nts co

    nnect

    ed to

    the s

    ystem

    (va

    lves, s

    ensor

    s, etc.

    ).

    If vo

    ltage

    s 12V

    and 5

    V do

    not a

    ppear

    ,on

    e of th

    e boa

    rds is

    dama

    ged.

    Disco

    nnect

    board

    DIG

    -T2 a

    nd ch

    eck

    board

    ANL

    -T1.

    2.1Co

    mmun

    icatio

    n cab

    le is

    disco

    nnect

    ed.

    2.2If

    the ot

    her f

    uncti

    ons a

    re OK

    , the d

    isplay

    is pro

    bably

    dama

    ged.

    3.1DI

    G-T2

    board

    is fa

    ulty.

    4,1Di

    gital

    cable

    is dis

    conn

    ected

    .4.1

    Keyb

    oard

    is fau

    lty.

    4.2DI

    G-T2

    board

    is fa

    ulty.

    5.1Ch

    eck if

    the di

    splay

    ed va

    lue is

    near

    theatm

    osphe

    ric pr

    essure

    .5.2

    If the

    displ

    ayed

    value

    is si

    gnific

    antly

    differ

    ent fr

    om th

    e atm

    osphe

    ric pr

    essure

    proba

    bly th

    at the

    re is

    a prob

    lem w

    ith th

    epre

    ssure

    measu

    ring.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.2

    Pre-

    proces

    s malf

    uncti

    on (c

    ont.)

    Symp

    tom1.

    (cont.

    )

    2.Sy

    stem

    is on

    and d

    isplay

    isno

    t lit.

    3.Sy

    stem

    on an

    d erro

    neou

    sor

    fragm

    ented

    digit

    s.4.

    No re

    spon

    se fro

    m on

    e of

    the ke

    ys.

    5.Th

    e disp

    layed

    read

    out fo

    rthe

    atmo

    sphe

    ric pr

    essure

    ,wh

    en th

    e doo

    r of th

    ech

    ambe

    r is op

    en, d

    oes n

    otco

    nform

    to th

    e actu

    alab

    solut

    e pres

    sure.

  • 48

    Corr

    ectio

    ns6.1

    See s

    ympto

    m s 2

    3 & 24

    in pa

    ra. 11

    .4.

    7.1Se

    e sym

    ptom

    s 25 i

    n para

    . 11.4

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts6.1

    Incorr

    ect re

    ading

    of th

    e doo

    r swi

    tchpo

    sition

    7.1Inc

    orrect

    read

    ing of

    the r

    eservo

    ir floa

    tsw

    itch p

    ositio

    n.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.2

    Pre-p

    rocess

    malf

    uncti

    on (c

    ont.)

    Symp

    tom6.

    Press

    ing th

    e STA

    RT ke

    ydo

    es no

    t star

    t the p

    rocess

    ,de

    spite

    the d

    oor is

    clos

    ed.

    7.Pr

    essing

    the S

    TART

    key

    does

    not s

    tart th

    e proc

    essan

    d the

    error

    mess

    age

    ADD

    WAT

    ER is

    disp

    layed

    desp

    ite th

    e wate

    r rese

    rvoir

    being

    full.

  • 49

    Corr

    ectio

    ns1.1

    If req

    uired

    chan

    ge W

    ater T

    ime

    param

    eter.

    2.1Se

    e sym

    ptoms

    29 in

    para.

    11.4.

    2.2Inc

    rease

    the Pr

    ogram

    inter

    val. A

    dd 1-

    2sec

    onds

    to the

    wate

    r inlet

    .3.1

    Clean

    the p

    iping

    or th

    e valv

    e.

    4.1Cl

    ean th

    e pipi

    ng..

    4.2Re

    place

    faulty

    valve

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts1.

    After

    vacu

    um st

    age i

    s com

    pleted

    messa

    ge W

    ater

    will a

    ppear

    on th

    edis

    play a

    nd w

    ater e

    nters

    the ch

    ambe

    run

    til it R

    eache

    s the

    elect

    rode.

    In thi

    ssta

    ge a

    coun

    ter is

    displ

    ayed

    in th

    ebo

    ttom

    row of

    the d

    isplay

    . Afte

    rnu

    llifica

    tion o

    f the c

    ounte

    r stop

    the c

    ycle

    manu

    ally.

    Wait

    until

    proces

    s can

    celing

    have

    been

    comp

    leted

    and o

    pen t

    he do

    or.Ve

    rify t

    hat th

    e wate

    r qua

    ntity

    is 60

    0 ml

    (20 oz

    s).If

    the au

    toclav

    e doe

    s not

    opera

    te as

    descr

    ibed a

    bove

    , proc

    eed as

    follo

    ws:

    2.1Inc

    orrect

    read

    ing of

    the w

    ater s

    ensin

    gele

    ctrod

    e,2.2

    Progra

    mmed

    inter

    val a

    fter th

    e elec

    trode

    senses

    wate

    r surf

    ace is

    very

    short.

    3.1Pro

    bably

    wate

    r pipi

    ng or

    valve

    isclo

    gged

    . This

    lead

    s to a

    long

    fillin

    g tim

    e,

    4.1Pip

    ing is

    bloc

    ked

    4.2Th

    e wate

    r inlet

    valve

    is fa

    ulty.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.3

    In Pr

    ocess

    Malf

    uncti

    on

    11.3

    .1 W

    ater I

    nlet S

    tage

    Symp

    tom1.

    If the

    re is

    any m

    alfun

    ction

    relate

    d to i

    nlet w

    ater

    quan

    tity (L

    ow W

    ater, H

    ighTe

    mp, L

    ow Pr

    ess or

    Low

    Heat)

    , che

    ck th

    e auto

    matic

    water

    fillin

    g.

    2.Th

    e wate

    r inlet

    phase

    issh

    ort an

    d the

    wate

    r leve

    l inthe

    cham

    ber is

    low.

    3.Th

    e wate

    r inlet

    time i

    spro

    longe

    d and

    the q

    uanti

    tyof

    water

    is no

    t acc

    eptab

    le.4.

    The p

    rocess

    fails

    on Lo

    wW

    ater in

    the w

    ater in

    letsta

    ge, w

    ater d

    oes n

    ot en

    terthe

    cham

    ber.

  • 50

    Corr

    ectio

    ns

    4.1Re

    locate

    the c

    oolin

    g coil

    and r

    elease

    float

    or cle

    an th

    e floa

    t.

    4.2Re

    place

    float.

    4.3Fix

    disco

    nnec

    tion.

    4.4Re

    place

    plun

    ger o

    r com

    plete

    solen

    oid.

    4.5Cl

    ean w

    ater in

    let pi

    pe4.6

    See s

    ympto

    m s 3

    & 6

    in pa

    ra. 11

    .4.

    Repla

    ce w

    ater in

    let va

    lve.

    4.7Re

    place

    plun

    ger o

    r com

    plete

    solen

    oid.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    stsIf

    no w

    ater e

    nters

    the ch

    ambe

    r, the

    re is

    no w

    ater

    in the

    reser

    voir a

    nd di

    splay

    indic

    ates t

    hat th

    ereis

    water

    :4.1

    Float

    is stu

    ck. C

    heck

    if the

    cooli

    ng co

    ilha

    s mov

    ed an

    d prev

    ents

    the flo

    at fro

    mmo

    ving o

    r if flo

    at is

    stuck

    due t

    o dirt.

    4.2Flo

    at is

    faulty

    .Th

    ere is

    wate

    r in th

    e rese

    rvoir a

    nd th

    e disp

    layind

    icates

    that

    there

    is wa

    ter. D

    espite

    this,

    wate

    rdo

    es no

    t ente

    r the c

    hamb

    er.4.3

    No vo

    ltage

    at so

    lenoid

    inpu

    t.Ve

    rify c

    omma

    nd at

    Test

    Point

    TP9 f

    romele

    ctron

    ic bo

    x.4.4

    solen

    oid fa

    ulty.

    4.5W

    ater in

    let pi

    pe is

    clog

    ged.

    4.6W

    ater in

    let va

    lve (2

    1) is

    faulty

    .

    4.7Va

    lves

    solen

    oid is

    fault

    y.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.3

    In Pr

    ocess

    Malf

    uncti

    on (c

    ont.)

    1

    1.3.1

    Wate

    r Inle

    t Stag

    e (co

    nt.)

    Symp

    tom5.

    Wate

    r doe

    s not

    enter

    the

    cham

    ber d

    uring

    FIL

    L WAT

    ER

    stage

    .

  • 51

    Corr

    ectio

    ns1.1

    See s

    ympto

    ms 14

    , 16 a

    nd 16

    in pa

    ra.11

    .3.2.

    2.1Cl

    ose th

    e doo

    r more

    tightl

    y or s

    et the

    door

    switc

    h more

    stead

    ily.

    3.1If

    amou

    nt of

    water

    is no

    t corr

    ect, c

    lean

    the el

    ectrod

    e and

    set th

    e wate

    r leve

    l

    3.2Se

    e sym

    ptom

    21 in

    para.

    11.4.

    3.3Th

    e the

    rmost

    at mu

    st be

    set a

    t a hi

    gher

    switc

    hing p

    oint a

    nd th

    e tim

    e afte

    r the

    electr

    ode t

    ouch

    es the

    wate

    r be p

    rolon

    ged.

    3.4Cl

    ean or

    repla

    ce wa

    ter in

    let va

    lve an

    dex

    haust

    valve

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts1.1

    The p

    roblem

    is th

    e inc

    orrect

    read

    ing of

    the te

    mpera

    ture s

    ensor

    .

    2.1Th

    e pres

    sure i

    n the

    cham

    ber r

    ises d

    uring

    the he

    ating

    stag

    e and

    is pr

    essing

    on th

    edo

    or. If

    the d

    oor s

    witch

    is lig

    htly p

    ressed

    it cou

    ld be

    turne

    d off.

    3.1

    The p

    rocess

    fails

    if the

    elect

    rode i

    s abo

    vethe

    wate

    r surf

    ace an

    d the

    safet

    ythe

    rmost

    at sw

    itche

    s off.

    Check

    if the

    amou

    nt of

    water

    in th

    ech

    ambe

    r is ov

    er 35

    0cc.

    3.2If

    the w

    ater q

    uanti

    ty rem

    ains i

    nsuffi

    cient,

    the pr

    oblem

    is a

    wron

    g elec

    trode

    sign

    al.3.3

    Proces

    s fail

    s desp

    ite co

    rrect

    quan

    tity of

    water

    .

    3.4Pre

    sence

    of dir

    t insid

    e the

    wate

    r inlet

    valve

    may

    cause

    slow

    down

    of th

    e flow

    .

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.3

    In Pr

    ocess

    Malf

    uncti

    on (c

    ont.)

    11

    .3.2

    Hea

    ting S

    tage

    Symp

    tom1.

    Proc

    ess fa

    ils w

    hile

    initia

    ting t

    he he

    ating

    stag

    e.me

    ssage

    HIG

    H TE

    MP i

    sdis

    playe

    d.2.

    Mess

    age D

    OOR

    UNLO

    CKis

    displa

    yed d

    uring

    the

    heati

    ng st

    age,

    altho

    ugh t

    hedo

    or is

    closed

    .3.

    LOW

    WAT

    ER m

    essag

    e is

    displa

    yed d

    uring

    the

    heati

    ng st

    age a

    nd pr

    ocess

    fails.

  • 52

    Corr

    ectio

    ns

    4.1.1

    Clean

    or re

    place

    electr

    ode.

    4.1.2

    Leve

    l auto

    clave

    .

    4.2Se

    e sym

    ptoms

    12 in

    para.

    11.4.

    5.1Re

    place

    gaske

    t and

    close

    well

    .5.2

    Repla

    ce do

    or ga

    sket.

    5.3Ch

    eck an

    d tigh

    ten pi

    ping c

    onne

    ction

    s.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts4.1

    The h

    eating

    time i

    s very

    long

    and t

    hehe

    ating

    rate

    is no

    rmal.

    4.1.1

    Exces

    s qua

    ntity

    of wa

    ter in

    the

    cham

    ber d

    ue to

    dirty

    elect

    rode,

    loose

    electr

    ode c

    onne

    ction

    s or

    faulty

    elect

    rode.

    4.1.2

    Exces

    s qua

    ntity

    of wa

    ter in

    the

    cham

    ber d

    ue to

    un-le

    velle

    dau

    toclav

    e.4.2

    Heati

    ng el

    emen

    ts da

    mage

    d.

    5.1Le

    akag

    e at th

    e wate

    r filte

    r.5.2

    Steam

    leak

    age f

    rom th

    e doo

    r.5.3

    Piping

    leak

    age.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.3

    In Pr

    ocess

    Malf

    uncti

    on (c

    ont.)

    1

    1.3.2

    Hea

    ting S

    tage (

    cont.

    )

    Symp

    tom4.

    Mess

    age L

    OW H

    EAT i

    sdis

    playe

    d duri

    ng th

    ehe

    ating

    stag

    e.

    5.LO

    W V

    ACUU

    M m

    essag

    eis

    displa

    yed d

    uring

    the

    heati

    ng st

    age a

    nd th

    epro

    cess

    fails,

    or du

    ring t

    heva

    cuum

    test.

    .

  • 53

    Corr

    ectio

    ns5.1

    Calib

    rate t

    he sy

    stem.

    5.1Fix

    elect

    rode o

    r rep

    lace v

    alve.

    5.2Re

    place

    air tra

    p.

    5.3Pin

    point

    the p

    ressur

    e sen

    sor (S

    eesym

    ptoms

    18 &

    19 in

    para.

    11.4)

    or th

    etem

    perat

    ure se

    nsor a

    nd re

    place

    thede

    fectiv

    e sen

    sor.

    2.1Re

    place

    exha

    ust va

    lve or

    clean

    pipe

    .

    Corr

    ectio

    ns1.1

    Repla

    ce fus

    e.1.2

    increa

    se tem

    perat

    ure ad

    justm

    ent b

    ytur

    ning t

    he po

    tentio

    meter

    screw

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts5.1

    Gene

    rally

    these

    proble

    ms ar

    e a re

    sult o

    fthe

    inco

    mpati

    bility

    betw

    een te

    mpera

    ture

    and p

    ressur

    e.5.2

    Insuff

    icien

    t qua

    ntity

    of wa

    ter in

    the

    cham

    ber d

    ue to

    short

    circu

    it elec

    trode

    orfau

    lty in

    let va

    lve.

    5.2Th

    e air t

    rap je

    t is cl

    ogge

    d, the

    air3e

    limina

    tion i

    s insu

    fficie

    nt.5.4

    One o

    f the s

    ensor

    s is d

    efecti

    ve.

    2.1Ex

    haust

    valve

    is fa

    ulty o

    r pipe

    isclo

    gged

    .

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts1.1

    DRY

    heate

    rs fus

    e burn

    t.1.2

    Check

    if the

    therm

    ostat

    switc

    hes o

    ff at

    atoo

    low

    tempe

    rature

    .

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.3

    In Pr

    ocess

    Malf

    uncti

    on (c

    ont.)

    1

    1.3.3

    Ster

    ilizati

    on st

    age

    Symp

    tom1.

    proce

    ss fai

    ls on

    HIG

    HTE

    MP,

    LOW

    TEM

    P,HI

    GH PR

    ES, o

    r LOW

    PRES

    durin

    g the

    steril

    isatio

    n stag

    e.

    2.Th

    e fast

    exha

    ust o

    perat

    ionis

    too sl

    ow.

    11.3

    .3 D

    ry sta

    ge

    Symp

    tom1.

    The d

    ry op

    eratio

    n is n

    oteff

    ectiv

    e.

  • 54

    Corr

    ectio

    ns

    1.1Cl

    ean or

    repla

    ce the

    valve

    .

    1.2Re

    place

    the bo

    ard.

    1.3Fix

    the c

    onne

    ction

    s.

    1.4Th

    e valv

    e must

    be fix

    ed or

    repla

    ced.

    1.5Re

    place

    ANL-

    T1 bo

    ard.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    stsTh

    e prob

    lem m

    ay be

    a me

    chan

    ical fa

    ilure

    of the

    valve

    , the v

    alve d

    river

    locate

    d on A

    NL-T

    1bo

    ard or

    the c

    ontro

    l circ

    uit on

    board

    DIG

    -T2.

    1.1Cl

    ogge

    d or f

    aulty

    valve

    .1.2

    Check

    TP8.

    If the

    volta

    ge is

    high

    er tha

    n4V

    and t

    here

    is co

    ntrol

    signa

    l to th

    eva

    lve, D

    IG-T

    2 boa

    rd is

    faulty

    .If

    the vo

    ltage

    is 0V

    DC th

    e valv

    e is o

    pen,

    thepro

    blem

    is no

    t the c

    ontro

    l circ

    uit bu

    t som

    e othe

    rele

    ment

    of the

    syste

    m.

    1.3Ch

    eck th

    e con

    necti

    on be

    tween

    the

    conn

    ector

    JP5/4

    and t

    he va

    lve.

    1.4Re

    move

    the c

    over

    protec

    ting t

    heter

    mina

    ls of

    the co

    il and

    check

    the

    volta

    ge w

    hen t

    he so

    lenoid

    is co

    nnect

    ed.

    If the

    valve

    is op

    en de

    spite

    the co

    ilvo

    ltage

    less

    than 2

    V, th

    ere is

    ame

    chan

    ical p

    roblem

    .1.5

    If all

    items

    check

    ed ab

    ove c

    onfor

    m to

    requir

    emen

    ts, A

    NL-T

    1 boa

    rd is

    faulty

    .

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.4

    Comp

    onen

    ts Ma

    lfunc

    tion

    Symp

    tom1.

    Exha

    ust v

    alve i

    s alw

    ays

    open

    .

  • 55

    Corr

    ectio

    ns

    2.1Re

    place

    board

    DIG

    -T2.

    2.2Fix

    the c

    onne

    ction

    s.

    2.3Re

    place

    the so

    lenoid

    . If th

    is do

    es no

    the

    lp, re

    pair t

    he m

    echan

    ical p

    art of

    the

    valve

    or re

    place

    the va

    lve.

    3.1Cl

    ean or

    repla

    ce the

    valve

    .3.2

    Repla

    ce the

    board

    .

    3.3Fix

    the c

    onne

    ction

    s.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    stsTh

    e prob

    lem m

    ay be

    caus

    ed by

    the m

    echa

    nical

    part o

    f the v

    alve o

    r the d

    river

    locate

    d on b

    oard

    ANL-

    T1 or

    the c

    ontro

    l circ

    uit on

    board

    DIG

    -T2.

    2.1Ch

    eck TP

    8. If

    the vo

    ltage

    is co

    nstan

    tly0V

    DC, w

    hile t

    he va

    lve m

    ust be

    open

    ,the

    n the

    prob

    lem is

    on th

    e DIG

    -T2 b

    oard.

    2.2Ch

    eck th

    e con

    necti

    on be

    tween

    conn

    ector

    JP5/4

    and t

    he va

    lve.

    2.3Ch

    eck th

    e coil

    inpu

    t volt

    age w

    ith th

    esol

    enoid

    conn

    ected

    to th

    e valv

    e. If

    thevo

    ltage

    is hi

    gher

    than 1

    0VDC

    and t

    heva

    lve do

    es no

    t ope

    n, the

    solen

    oid or

    the

    mech

    anica

    l part

    of th

    e valv

    e are

    faulty

    .Th

    e prob

    lem m

    ay be

    caus

    ed by

    the m

    echa

    nical

    part o

    f the v

    alve,

    the dr

    iver lo

    cated

    on th

    eAN

    L-T1

    board

    or th

    e con

    trol c

    ircuit

    on th

    eDI

    G-T2

    board

    .3.1

    Clog

    ged o

    r fau

    lty va

    lve.

    3.2Ch

    eck TP

    9. If

    the vo

    ltage

    is hi

    gher

    than

    4V, th

    e con

    trol o

    f the v

    alve i

    s OK,

    and

    board

    DIG

    -T2 i

    s fau

    lty.

    If the

    volta

    ge is

    0V an

    d valv

    e is o

    pen,

    thepro

    blem

    is no

    t the c

    ontro

    l circ

    uit bu

    t othe

    rele

    ments

    of th

    e sys

    tem.

    3.3Ch

    eck th

    e con

    necti

    on be

    tween

    conn

    ector

    JP5/5

    and t

    he va

    lve.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.4

    Comp

    onen

    ts Ma

    lfunc

    tion

    Symp

    tom2.

    Exha

    ust v

    alve i

    s alw

    ays

    closed

    .

    3.W

    ater v

    alve i

    s alw

    ays o

    pen

  • 56

    Corr

    ectio

    ns3.3

    Repa

    ir or r

    eplac

    e the

    valve

    .

    3.4Re

    place

    ANL-

    T1 bo

    ard.

    4.1Re

    place

    board

    DIG

    -T2.

    4.2Fix

    the c

    onne

    ction

    s.

    4.3Re

    place

    the so

    lenoid

    . If th

    is do

    es no

    tsol

    ve th

    e prob

    lem, re

    pair o

    r rep

    lace t

    heva

    lve

    4.4Re

    place

    ANL-

    T1 bo

    ard.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    sts3.3

    Check

    the s

    oleno

    id co

    il inp

    ut vo

    ltage

    with

    the so

    lenoid

    conn

    ected

    . remo

    ve th

    eco

    ver th

    at pro

    tects

    the te

    rmina

    ls of

    theco

    il. If

    the va

    lve is

    open

    despi

    te the

    volta

    ge be

    ing le

    ss tha

    n 2VD

    C, th

    ere is

    ame

    chan

    ical p

    roblem

    .3.4

    If all

    items

    check

    ed ab

    ove c

    onfor

    m to

    requir

    emen

    ts, A

    NL-T

    1 boa

    rd is

    faulty

    .

    The p

    roblem

    may

    be th

    e mec

    hanic

    al pa

    rt of th

    eva

    lve, th

    e sole

    noid,

    the d

    river

    on A

    NL-T

    1 boa

    rdor

    the co

    ntrol

    circu

    it on D

    IG-T

    2 boa

    rd.4.1

    Check

    TP7.

    If vo

    ltage

    is co

    nstan

    tly 0V

    atthe

    stag

    e whe

    n the

    valve

    must

    be op

    enthe

    n DIG

    -T2 b

    oard

    is fau

    lty.

    4.2Ch

    eck th

    e con

    necti

    on be

    tween

    conn

    ector

    JP5/3

    and t

    he va

    lve.

    4.3Ch

    eck th

    e volt

    age c

    oming

    to th

    e sole

    noid

    coil w

    hile t

    he so

    lenoid

    is co

    nnect

    ed to

    the va

    lve. If

    the v

    alve i

    s clos

    ed de

    spite

    the vo

    ltage

    being

    high

    er tha

    n 10V

    DC, th

    esol

    enoid

    is fa

    ulty,

    or the

    re is

    ame

    chan

    ical p

    roblem

    with

    the v

    alve.

    4.4If

    all ite

    ms ch

    ecked

    abov

    e con

    form

    toreq

    uirem

    ents,

    ANL

    -T1 b

    oard

    is fau

    lty.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.4

    Comp

    onen

    ts Ma

    lfunc

    tion

    Symp

    tom3.

    (cont.

    )

    4.Ai

    r valv

    e is a

    lway

    s clos

    ed.

  • 57

    Corr

    ectio

    ns

    5.1rep

    lace D

    IG-T

    2 boa

    rd.

    5.2rep

    lace t

    he A

    C-T1

    board

    , or th

    eco

    mpon

    ents

    Q2 (T

    RIAC

    ) or U

    2 (MO

    C).

    6.1Cl

    ean or

    repla

    ce the

    valve

    .6.2

    Repla

    ce bo

    ard D

    IG-T

    2.

    6.3Fix

    the c

    onne

    ction

    s.

    6.4Re

    pair o

    r rep

    lace t

    he va

    lve.

    6.5Re

    place

    ANL-

    T1 bo

    ard.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    stsGe

    neral

    ly thi

    s situ

    ation

    is ca

    used

    by a

    contr

    olfai

    lure o

    n the

    DIG

    -T2 o

    r a fa

    ult at

    the d

    river

    locate

    d on t

    he bo

    ard.

    5.1Ch

    eck th

    e volt

    age a

    t TP5

    , if it

    isco

    nstan

    tly hi

    gher

    than 4

    VDC,

    the f

    ault i

    son

    the D

    IG-T

    2 boa

    rd.5.2

    If the

    volta

    ge at

    TP5 i

    s 0VD

    C an

    d the

    pump

    work

    s the

    failu

    re is

    on A

    C-T1

    ;Th

    e prob

    lem co

    uld be

    the m

    echa

    nical

    part o

    f the

    valve

    , the d

    river

    locate

    d on A

    NL-T

    1 boa

    rd or

    the co

    ntrol

    circu

    it on D

    IG-T

    2 boa

    rd.6.1

    Clog

    ged o

    r fau

    lty va

    lve.

    6.2Ch

    eck TP

    7. If

    the vo

    ltage

    is hi

    gher

    than

    4VDC

    and t

    here

    is a c

    ontro

    l sign

    al to

    theva

    lve, th

    e prob

    lem is

    on D

    IG-T

    2 boa

    rd.If

    the vo

    ltage

    is 0V

    DC an

    d the

    valve

    is op

    en,

    the pr

    oblem

    is no

    t the c

    ontro

    l circ

    uit bu

    t one

    ofthe

    othe

    r elem

    ents

    of the

    syste

    m.6.3

    Check

    the c

    onne

    ction

    betw

    een co

    nnect

    orJP

    5/3 an

    d the

    valve

    .6.4

    Check

    the v

    oltag

    e at th

    e sole

    noid

    coil,

    while

    the s

    oleno

    id is

    conn

    ected

    to th

    eva

    lve, b

    y rem

    oving

    the c

    over

    which

    protec

    ts the

    term

    inals

    of the

    coil.

    If the

    valve

    is op

    en de

    spite

    the vo

    ltage

    being

    lower

    than 2

    VDC,

    there

    is a

    mech

    anica

    lpro

    blem

    with

    the va

    lve.

    6.5If

    all ite

    ms ch

    ecked

    abov

    e con

    form

    toreq

    uirem

    ents,

    ANL

    -T1 b

    oard

    is fau

    lty.

    11.

    TROU

    BLES

    HOOT

    ING

    (cont.

    ) 1

    1.4

    Comp

    onen

    ts Ma

    lfunc

    tion

    Symp

    tom5.

    Vacu

    um pu

    mp is

    activ

    ated

    conti

    nuou

    sly 6

    .Ai

    r valv

    e is a

    lway

    s ope

    n.

  • 58

    Corr

    ectio

    ns

    7.1Cl

    ean or

    repla

    ce the

    valve

    .7.2

    Repla

    ce bo

    ard D

    IG-T

    2.

    7.3Fix

    the c

    onne

    ction

    s.

    7.4Re

    pair o

    r rep

    lace t

    he va

    lve.

    7.5Re

    place

    ANL-

    T1 bo

    ard.

    8.1Re

    place

    board

    DIG

    -T2.

    Possi

    ble ca

    use c

    heck

    -up a

    nd te

    stsTh

    e prob

    lem m

    ay be

    with

    the m

    echa

    nical

    part

    of the

    valve

    , the s

    oleno

    id, th

    e driv

    er on

    ANL

    -T1

    board

    or th

    e con

    trol c

    ircuit

    on D

    IG-T

    2 boa

    rd.7.1

    Clog

    ged o

    r fau

    lty va

    lve.

    7.2Ch

    eck TP

    10. If

    the v

    oltag

    e is h

    igher

    than

    4VDE

    and t

    here

    is a c

    ontro

    l sign

    al to

    theva

    lve th

    e prob

    lem is

    on D

    IG-T

    2 boa

    rd.If

    the vo

    ltage

    is 0V

    DC an

    d the

    valve

    is op

    en,

    the pr

    oblem

    is no

    t the c

    ontro

    l circ

    uit bu

    t one

    ofthe

    othe

    r elem

    ents

    of the

    syste

    m.7.3

    Check

    the c

    onne

    ction

    betw

    een co

    nnect

    orJP

    5/6 an

    d the

    valve

    .7.4

    Check

    the s

    oleno

    id co

    il inp

    ut vo

    ltage

    while

    the s

    oleno

    id is

    conn

    ected

    to th

    eva

    lve, b

    y rem

    oving

    the c

    over

    that

    protec

    ts the

    term

    inals

    of the

    coils

    . If th

    eva

    lve is

    open

    , desp

    ite th

    e volt

    age b

    eing

    less t

    han 2

    VDC,

    there

    is a

    mech

    anica

    lpro

    blem.

    7.5If

    all ite

    ms ch

    ecked

    abov

    e con

    form

    toreq

    uirem

    ents,

    ANL

    -T1 b

    oard

    is fau

    lty.

    The p