TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
InternationalHOORA Pharma
Meditec
InstrumentsAM sale & service Ali Traders
1Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4
2 Aditional benefit = 2 2
latest incometax paid = 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 6 6
4 Sale refrance to the products = 8 3 0 0
5 Overall reputation to the products = 10 0 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 0 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 12 0
Spare parts = 8 8
At least 5 technical persons = 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 10 6
Total = 100 (Req 65 %) 67 34 40 0 0
Remarks: Graded Not Graded Not Graded Not Graded Not Graded
1. PH Analyzer
3
8
9
incompelete offer
make and model is
not propper
mention in their
technical offer no
warrenty and
validity mention in
their technical offer
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
InternationalHoora Pharma
Meditec
InstrumentsAM sale & service Ali Traders
1 PH/Blood Gas Ananlyzer (Automatic)
MODEL BG100 348 Easy State Not mention CCA TS
MANUFACTURERConvergent
TechologiesSiemens
Medica
CorporationNova Opti Medical
COUNTRY OF ORIGIN Germany Europe USA USA USA
AUTHORIZATION Yes
Quality Cirtificate Yes
Warrenty YesMicro-Processor Controlled micro blood gas/pH measuring system.
Automatic calibration ' Automatic measurement of P02, pH, PC02. Yes
Automatic calculation of base excess, 02 Sat Act HC03 standard HC03 and total
C02. complete with standard and operational accessories, including built n
display and printer. Compatible online UPS. Electrolyte (Na/K/CL/ Hemoglobin). Power
source 220 V 50 Hz AC Imported from EEC. US, JapanYes
RecommendationTechnically
Approved
Techincally Not
Approved.Technically Not
Approved
Technically Not
Approved
Technically Not
Approved
1. PH Analyzer
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a.w
arre
nty
no
t m
en
tio
n.
Inco
mp
ele
te o
ffe
r 3
rd p
arty
dis
trib
uti
on
an
d q
uo
ted
war
ren
ty is
on
e y
ear
inst
ead
of
3 y
ear
s.
Inco
mp
ele
te o
ffe
r m
ake
an
d m
od
el i
s
no
t p
rop
pe
r m
en
tio
n in
th
eir
tech
nic
al o
ffe
r n
o w
arre
nty
an
d
valid
ity
me
nti
on
in t
he
ir t
ech
nic
al
off
er.
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a.W
arre
nty
no
t m
en
tio
n.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
International
Quintex
Medical
Sameer
Enterprises
Su
Enterprises
Scientific
Technical
Corporation
Ali TradersSheikhupura
Surgical
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 8 6 7 8
4 Sale refrance to the products = 8 3 5 7 7 6
5 Overall reputation to the products = 10 0 6 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10 4 10 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 12 12 0 0 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 10 10 0 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 5 3 8 6
Total = 100 (Req 65 %) 67 72 36 60 70
Remarks: Graded Graded Not Graded Not Graded Graded
2. Water bath
3
8
9
Incomeplet
offer 3rd
party
distribution
one year
warrenty
instead of
three years.
Incomeplet
offer model and
make of quoted
items not
mention.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
International
Quintex
Medical
Sameer
Enterprices
Su
Enterprises
Scientific
Technical
Corporation
Ali TradersSheikhupura
Surgical
2 Water Bath
MODEL WNB 29 BW-20H BM-30 Precisdig WSB-30 BM30 Not Mention
MANUFACTURER Memmert Medline Nuve JP Selecta Witage NUVE Not Mention
COUNTRY OF ORIGIN Germany UK Turkey Spain Germany Turkey Not Mention
AUTHORIZATION Yes Yes Yes Yes Yes No
Quality Cirtificate Yes Yes Yes Yes Yes Yes
Warrenty 3Y 3Y 2Y 1Y 3Y 1Y
Thermostatic control, temperature upto 100 C+/- 1 C, inner stainless steel, with
thermometer gabelled, cover, time switch . Capacity of 30 liters. 220 V, 50Hz. AC
Complete accessories, imported from EEC. US, Japan
Yes
Offered
model is
having 20
litter capacity
instead of 30
as required
by tender.
Do not fulfill
grading
criteria. Also
the offered
model is not
from USA,
Europe,
Japan
warrenty if
two years
instead of
three years.
Did not fulfill
assessment
criteria
quoted
warrenty is
one year
instead of
three years.
Capacity is 12
liters instead
of 20
Country of
origin is korea
instead of
Germany
WITAGE is not
orignal
equipment
manufacturer.
Incomeplet
offer 3rd
party
distribution
one year
warrenty
instead of
three years.
Country of
origin is
turkey that is
not come in
europe.
RecommendationTechnically
Approved
Technically
Not
Approved.
Technically
Not
Approved,
Technically
Not
Approved
Technically
Not Approved
Technically
Not
Approved
Technically Not
Approved
Incomeplet
offer model and
make of quoted
items not
mention.
2. Water Bath
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/N
o.Assessment Criteria Sameer Enterprises SU Enterprises
Popular
Internatinoal (Pvt)
Ltd.
Scientific Technical
CorporationAli Traders Hospitech Sheikhupura Surgical
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 7 8 8 8
4 Sale refrance to the products = 8 7 7 7 6 0
5 Overall reputation to the products = 10 0 7 0 1
6In one continent = 4 or
In two continents = 6 (Max 10)4 10 10 10 10
7 Products specialization = 10 0 0 0
Repair workshop as per manufacturer = 12 0 0 12 12 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 0 10 10 10 0
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
3 8 5 6 4
Total = 100 (Req 65 %) 36 60 75 70 51
Remarks: Not Graded Not Graded Graded Graded Not Graded
3. Centrifuge
Incomeplet offer
model and make of
quoted items not
mention.
Incompelete offer
3rd party distribution
and quoted warrenty
is one year instead of
3 years.
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sameer Enterprises SU Enterprises
Popular
International
Scientific Technical
CorporationAli Traders Hospitech healthcare Sheikhupura Surgical
MODEL NF 200 EBA-20 Humax 4K 2420 NV200 Microcen 23
MANUFACTURER NUVE Hettich Human KUBOTA NUVE Ortoalresa
COUNTRY OF ORIGIN Turkey Germany Germany Japan Turkey Spain
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 2Y 1Y 3Y 3Y 1 year
Speed 5000-6000 rpm. capacity 8x15ml swing out
rotor, built in speedometer, vanable speed
control, lid' locking and braking device, time
switch 0-60 minutes. Standard accessories, 223V. 50
Hz. AC. Imported from EEC. US, Japan
Do not fulfill
assment criteria.
Quoted warrenty is
2years inatesad of
3years.
Also offered model is
not from
USA,Europe, Japan
Did not fulfill
assessment criteria.
Quoted warrenty is
1year inatead of
3years.
Yes4000 rpm instead of
5000
Did not fullfill
assessment criteria.
Quoted warrenty is
one year instead of
three years.Fixed
angle rotor
Technically Not
Approved.
Technically Not
Approved.
Technically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Incomeplet offer
model and make of
quoted items not
mention.
3. Centrifuge
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of 3
years.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
InternationalQuintex Medical
Sameer
EnterprisesSU Enterprises
Scientific
Technical
Corporation
Sheikhupura
Surgical
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 8 6 7 8
4 Sale refrance to the products = 8 3 5 7 7 6
5 Overall reputation to the products = 10 0 6 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10 4 10 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 12 12 0 0 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 10 10 0 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 5 3 8 6
Total = 100 (Req 65 %) 67 72 36 60 70
Remarks: Graded Graded Not Graded Not Graded Graded
4. Hot Air Oven
3
8
9
Incomeplet offer
model and make of
quoted items not
mention.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
InternationalQuintex Medical
Sameer
EnterprisesSU Enterprises
Scientific
Technical
Corporation
Sheikhupura
Surgical
MODEL UN75 A3-217-535 FN 55 ED-115 WOF-105
MANUFACTURER Memmert Titanox NUVE Binder WITAGE
COUNTRY OF ORIGIN Germany Italy Turkey Germany Germany
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 3Y 3Y 2Y 1Y 3Y
Thermostatically controlled, temperature range max upto 220 C+/- 2 C -
‘‘ted with thermostat. Thermometer, digital timer, internal stainless
steel, 220 V. 50 Hz AC.
Yes Yes
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar in
ate
ad o
f 3
year
s.
Co
un
try
of
ori
gin
is k
ore
a in
ste
ad o
f G
erm
any
WIT
AG
E is
no
t o
rign
al e
qu
ipm
en
t m
anu
fact
ure
r.
Comp'ete with standard accessories, capacity 60 liters. Imported from EEC. US .JapanYes Yes
RecommendationTechnically
Approved
Technically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Inco
me
ple
t o
ffe
r m
od
el a
nd
mak
e o
f q
uo
ted
ite
ms
no
t m
en
tio
n.
4. Hot Air Oven
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
International
Sameer
Enterprises
SU
Enterprises
Popular
International
Scientific Technical
Corporation
Hospitech
Healthcare
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 7 8
4 Sale refrance to the products = 8 3 7 6
5 Overall reputation to the products = 10 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10 10
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 12 0 12
Spare parts = 8 8 8 8
At least 5 technical persons = 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 8 6
Total = 100 (Req 65 %) 67 60 70
Lab Items For DHQ Hospital Gujranwala Remarks: Graded Not Graded Graded
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
International
Sameer
Enterprises
SU
Enterprises
Popular
International
Scientific Technical
Corporation
Hospitech
Healthcare
5 Roller Mixture
MODEL Movil-Rod RM-500 7000912 HumaRoll ATRL-2 F 406
MANUFACTURER JP Selecta Teknolabo Jp Selecta Human A&E LAB Not Mention
COUNTRY OF ORIGIN Spain Italy Spain Germany U.K Not Mention
AUTHORIZATION
Quality Cirtificate
Warrenty
Imported from EEC. US, Japan. Complete accessories
Recommendation:
5. Roller Mixer
May be retender with generalized specifications.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
InternationalQuintex Medical Sameer Enterprises SU Enterprises
Popular
International
Scientific Technical
CorporationAli Traders
Hospitech
Healthcare
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 7 8
4 Sale refrance to the products = 8 3 7 6
5 Overall reputation to the products = 10 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10 10
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 12 0 12
Spare parts = 8 8 8 8
At least 5 technical persons = 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 8 6
Total = 100 (Req 65 %) 67 60 70
Lab Items For DHQ Hospital Gujranwala Remarks: Graded Not Graded Graded
3
8
9
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma
International Quintex Medical Sameer Enterprises SU Enterprises
Popular
International
Scientific Technical
Corporation Ali Traders
Hospitech
Healthcare
6 Vortex Mixture
MODEL Movil-Vib AAH-360115 300-100 ZX3 HumaTwist VM-10 VX 200 MIX 20
MANUFACTURER JP Selecta Medline Teknolabo Velp Human Witage Labnet Falsi Insturment
COUNTRY OF ORIGIN Spain UK Italy Italy Germany Germany USA Italy
AUTHORIZATION
Quality Cirtificate
Warrenty
Imported from EEC. US, Japan. Complete
accessoriesMay be retender with generalized specifications.
6. Vortex Mixture
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma International Sameer Enterprises SU EnterprisesScientific Technical
CorporationHospitech Sheikhupura Surgical
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 6 7 8 8
4 Sale refrance to the products = 8 3 7 7 6 0
5 Overall reputation to the products = 10 0 0 0 1
6In one continent = 4 or
In two continents = 6 (Max 10)10 4 10 10 10
7 Products specialization = 10 0 0 0
Repair workshop as per manufacturer = 12 12 0 0 12 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 10 0 10 10 0
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 3 8 6 4
Total = 100 (Req 65 %) 67 36 60 70 51
Lab Items For DHQ Hospital Gujranwala Remarks: Graded Not Graded Not Graded Graded Not Graded
7. LAB Refrigrator
3
8
9
Incomeplet offer
model and make of
quoted items not
mention.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma International Sameer Enterprises SU Enterprises
Scientific Technical
CorporationHospitech Healthcare Sheikhupura Surgical
MODEL LR 440 BLCRF 425W LFF660 AF70V/2 Labor 7002T
MANUFACTURER Evermed Evermed Arctiko Frimed Not Mention
COUNTRY OF ORIGIN Italy Italy Denmark Italy Not Mention
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 3Y 2Y 1Y 3Y 1Y
For storage of specimens, diagnostic kits, reagents and anti-serum. Temperature
variable. Control panel includes audible alarm, defrost, door & over temperature
indicator lights, digital temperature select dial with adjustable shelves in the hair of
cab-net and roll out baskets with safety cathes in the other half, two double pan
glass dcors with heating :7m, lock and key. S.S. interior, interior light, automatic
safety device to shut down when temperature is below -2 C or
0 to 15 degree C Yes
above +20 C Cap: 15-20 Cuft. Complete unit operatable on 220 V 50 Hz AC.
Imported from EEC. US, Japan. Complete accessories
Yes Yes
RecommendationTechnically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically
Approved
Technically Not
Approved.
Technically Not
Approved.
Inco
me
ple
t o
ffe
r m
od
el a
nd
mak
e o
f q
uo
ted
ite
ms
no
t m
en
tio
n.
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar in
ate
ad o
f 3
year
s.
Did
no
t fu
llfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
on
e y
ear
inst
ead
of
thre
e y
ear
s. F
ixe
d a
ngl
e r
oto
r
7. Laboratory Refrigerator/Freezer
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
InternationalQuintex Medical Sameer Enterprises SU Enterprises
Scientific Technical
Corporation
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 8 6 7 8
4 Sale refrance to the products = 8 3 5 7 7 6
5 Overall reputation to the products = 10 0 6 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10 4 10 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 12 12 0 0 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 10 10 0 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 5 3 8 6
Total = 100 (Req 65 %) 67 72 36 60 70
Remarks: Graded Graded Not Graded Not Graded Graded
3
8
9
8. Electronic Balance
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
International
Quintex
Medical
Sameer
EnterprisesSU Enterprises
Scientific Technical
Corporation
MODEL FA-2204B MD 8104 FA-2004B PA 214 NBL 214E
MANUFACTURERJP Selecta Branded by
ComectaMedline JP Selecta Ohaus ADAM Equp
COUNTRY OF ORIGIN Spain UK Spain USA UK
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 3Y 3Y 2Y 1Y 3Y
Capscity 40/200 Mg. Sensitivity 0.1 mg, 220 V 50 Hz AC complete ready to use with standard
accessories. Imported from EEC. US, Japan. Yes
Off
er
Sen
siti
vity
is 1
mg
inst
ead
of
0.1
mg
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar in
ate
ad o
f 3
year
s.
Yes
RecommendationTechnically
Approved
Technically Not
Approved.Technically Not
Approved.
Technically Not
Approved.Technically
Approved
8. Electronic Balance
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Gulf Marketing
International
HOORA
Pharma
Mian
Scientific
Popular
International
Ghrasa
Labartory
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 68 6 6 8 6
4 Sale refrance to the products = 8 0 0 0 7 0
5 Overall reputation to the products = 10 0 0 0 7 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 0 10 10 10
7 Products specialization = 10 0 0 0 0 0
Repair workshop as per manufacturer = 12 0 0 0 12 0
Spare parts = 8 0 8 8
At least 5 technical persons = 10 0 10 0 10 0
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
0 10 0 5 2
Total = 100 (Req 65 %) 28 36 26 77 36
Remarks: Not Graded Not Graded Not Graded Graded Not Graded
3
8
9
9. Spectrophotometer
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Gulf Marketing
InternationalHoora Pharma Mian Scientific
Popular
International
Ghrasa
Labartory
MODEL Tecno 786 Plus MC15 5V3 Humalyzer 5010V5+
MANUFACTURERaudit
diagnostic
Diasys
DiagnosticERBA Chem Human Robert Riele
COUNTRY OF ORIGIN ireland Europe German Germany Germany
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 2Y Not Mention 3Y 3Y 3Y
Microprocessor control UV/ Visible wavelength range 200-900 nm banwicth 2-5 nm better 220" V, 50 Hz AC. Imported irom EcC US, Japan. Complete accessories. Compatible online UPS
Did not fulfill
assessment
criteria.
Quoted
warrenty is 2Y
instead of 3Y.
Did not fulfill
assessment
criteria
warrenty not
mention.
Did not fulfill
assessment
criteria.
Yes
Did not fulfill
assessment
criteria.
Recommendation
Technically
Not
Approved
Technically
Not
Approved
Technically
Not
Approved
Technically
Approved
Technically
Not
Approved
9. Spectrophotometer
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sameer Enterprises SU Enterprises HOORA PharmaScientific Technical
Corporation
Scientific Technical
Corporation
Scientific Technical
Corporation
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4 4
2 Aditional benefit = 2 2 2 2 2
latest incometax paid = 4 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 7 6 8 8 8
4 Sale refrance to the products = 8 7 7 0 6 6 6
5 Overall reputation to the products = 10 0 0 0 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)4 10 0 10 10 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 0 0 0 12 12 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 0 10 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
3 8 10 6 6 6
Total = 100 (Req 65 %) 36 60 34 70 70 70
Remarks: Not Graded Not Graded Not Graded Graded Graded Graded
3
8
9
10. Dilutor
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sameer
EnterprisesSU Enterprises Hoora Pharma
Scientific Technical
Corporation
Scientific Technical
Corporation
Scientific Technical
Corporation
MODEL Repette-HW 263-029 Varispenser+ 520.002 520.005 520.05
MANUFACTURER JP Selecta Jencons Eppendrof Socorax Socorax Socorax
COUNTRY OF ORIGIN Spain UK Germany Switzerland Switzerland Switzerland
AUTHORIZATION Yes Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes Yes
Warrenty 2Y 1Y No mention 3Y 3Y 3Y
For repetitive dispensing of samples from 0.2-30 ml with wide neck
reservoir Syringe size 10 ml. reservoir 5000 ml. Imported from EEC US, Japan. Complete
sccesscres.
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a.
Qu
ote
d w
arre
nty
is 1
year
inat
ead
of
3ye
ars.
Did not fulfill
assessment criteria
warrenty not
mention.
0.25 to 2 1 to 5 5 to 50
Recommendation
Technically Not
Approved.
Technically Not
Approved.Technically Not
Approved.
Subject to Sample
evaluation
Subject to Sample
evaluation
Subject to Sample
evaluation
10. Dilutor
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Quintex Medical Sameer Enterprises SU Enterprises HOORA PharmaPopular Internatinoal
(Pvt) Ltd.
Scientific Technical
CorporationAli Traders
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4 4
2 Aditional benefit = 2 2 2
latest incometax paid = 4 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 68 6 7 6 8 8
4 Sale refrance to the products = 8 5 7 7 0 7 6
5 Overall reputation to the products = 10 6 0 0 7 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 4 10 0 10 10
7 Products specialization = 10 0 0 0
Repair workshop as per manufacturer = 12 12 0 0 0 12 12
Spare parts = 8 8 8 8 8 8
At least 5 technical persons = 10 10 0 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
5 3 8 10 5 6
Total = 100 (Req 65 %) 72 36 60 34 75 70
Remarks: Graded Not Graded Not Graded Not Graded Graded Graded
3
8
9
Incompelete offer 3rd
party distribution and
quoted warrenty is one
year instead of 3 years.
11. Automatic Pipettes
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Quintex Medical
Sameer
Enterprises SU Enterprises Hoora Pharma
Popular
International
Scientific Technical
CorporationAli Traders
MODEL AO-10 .. 5000 Research + fixed & adjs
MANUFACTURER Medline Diamate Ohaus Eppendrof Human Socorax HTL
COUNTRY OF ORIGIN UK UK USA Germany Germany Switzerland Poland
AUTHORIZATION Yes Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes Yes
Warrenty 3Y 2Y 1Y Not Mention 3Y 3YAdjustable size 10-100 ul Complete accessories. Imported from EEC ,US , Japan
Yes Yes 10 to 100
Adjustable size 100-500 ul Complete accessories. Yes Yes 100 to 1000Fixed volume 10 ul with tip ejector. Complete accessorie Imported from EEC.
US. Japan. Yes Yes Yes
Fixed volume 20 ul with tip ejector. Complete accessories Imported from; EEC.
US, Japan Yes Yes Yes
Fixed volume 50 ul with tip ejector. Complete accessories Imposed from EEC,
US, Japan Yes Yes Yes
Fixed volume 100 u! with tip ejector. Complete accessories Imported from. EEC
US. Japan Yes Yes Yes
Technically Approved
Technically Not
Approved.
Do not fulfill assment
criteria
Technically Not
Approved
Technically Not
ApprovedTechnically Approved Technically Approved
Technically Not
Approved
11. Automatic Pipettes
Incompelete offer 3rd
party distribution and
quoted warrenty is
one year instead of 3
years.
Did not fulfill
assessment criteria
warrenty not mention.
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar in
ate
ad o
f 3
year
s.
Do
no
t fu
lfill
ass
me
nt
crit
eri
a.
Qu
ote
d w
arre
nty
is 2
year
s
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma
International
Seico Scientific
Traders
Gulf Marketing
InternationalHOORA Pharma
Popular
Internatinoal (Pvt)
Ltd.
Ghrasa Labartory Ghrasa Labartory Ali Traders Ali Traders
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 8 8 6 8 6 6
4 Sale refrance to the products = 8 3 5 0 0 7 0 0
5 Overall reputation to the products = 10 0 0 0 0 7 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 4 10 0 10 10 10
7 Products specialization = 10 0 0 0 0 0 0
Repair workshop as per manufacturer = 12 12 0 0 0 12 0 0
Spare parts = 8 8 0 8 8 8
At least 5 technical persons = 10 10 10 0 10 10 0 0
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 10 0 10 5 2 2
Total = 100 (Req 65 %) 67 45 26 34 75 36 36
Remarks: Graded Not Graded Not Graded Not Graded Graded Not Graded Not Graded
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
12. Auto Chemistry
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM
Sigma
InternationalSeico scientific
Gulf Marketing
InternationalHoora Pharma
Popular
InternationalGhrasa Labartory Ghrasa Labartory Ali Traders Ali Traders
MODEL Fully pictus basic accent 300 XL-180 Humaster 200 Alcor Sphera Mocrolab 300 Selecta Pro S
MANUFACTURER BSI diatron pz cormay ERBA Mannhein Human EDIF EDIF
COUNTRY OF ORIGIN Italy hungary Poland Germany Germany Italy Italy Neatherland Neatherland
AUTHORIZATION Yes Yes Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes Yes Yes
Warrenty 3Y 1Y 2Y Not Mention 3Y 3Y 3Y
Automatic Analyzer Computer controlled,
programmable Complete accessories foe all
chemistries Open System randcmaccess . Display
Monitor, Builtin Printer. The manufacturers must
have a range of standards, reagents about they
must give a surity regarding their availability With
standard operational accessories Power source
220 V. 50 Hz. AC. Compatible online UPS. Import
crom EEC. US, Japan
Yes
Do not fulfill
assment criteria.
Quoted warrenty
is 1Y instaed of 3Y
& UPS not quoted.
Also offered
model is not fully
automatic
Did not fulfill
assessment
criteria. Quoted
warrenty is 2Y
instead of 3Y.
Did not fulfill
assessment
criteria warrenty
not mention.
Yes
Did not fulfill
assessment
criteria.
Did not fulfill
assessment
criteria.
RecommendationTechnically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
12. Auto Chemistry
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Seico Scientific
Traders
Gulf Marketing
International
Popular
Internatinoal (Pvt)
Ltd.
Scientific
Technical
Corporation
AM sale & service Ali Traders Hospitech HOORA Pharma
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4 4
2 Aditional benefit = 2 2 2 2 2 2 2
latest incometax paid = 4 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 68 8 8 8 8 6
4 Sale refrance to the products = 8 5 0 7 6 0 0
5 Overall reputation to the products = 10 0 0 7 0 1 0
6In one continent = 4 or
In two continents = 6 (Max 10)4 10 10 10 10 0
7 Products specialization = 10 0 0 0 0 0 0
Repair workshop as per manufacturer = 12 0 0 12 12 12 0
Spare parts = 8 0 8 8 8
At least 5 technical persons = 10 10 0 10 10 0 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
10 0 5 6 4 10
Total = 100 (Req 65 %) 47 28 77 70 53 36
Lab Items For DHQ Hospital Gujranwala Remarks: Not Graded Not Graded Graded Graded Not Graded Not Graded
13. Hamatology Analyzer
3
8
9
Incompelete offer
3rd party
distribution and
quoted warrenty is
one year instead of
3 years.
incompelete offer
make and model is
not propper
mention in their
technical offer no
warrenty and
validity mention in
their technical
offer
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Seico Scientific Gulf Marketing
Popular
International
Scientific
Technical
Corporation
AM sale & service Ali TradersHospitech
HealthcareHOORA Pharma
13 Hematology Anaylizer
MODEL abacus 380 mythic 18 Humacount 60TS Norma Icon Not mention Medonic M SeriesSwelab Alpha
Basic 16PDrew 3
MANUFACTURER diatron orphee s.a Human Norma Icon Not mention Boule Medical Drew Scientific Inc
COUNTRY OF ORIGIN hungary germany Germany Austria Not mention Sweeden Swedan USA
AUTHORIZATION
Quality Cirtificate
Warrenty 1Y 2Y 3Y 3Y Not Mention 1Y Not Mention
Electronic resistence methode atlest RBC, WBC,HCT, MCH, HGB, PLT,
MCV, MCHC parameter with digital display and printer. Sample volume
according to individual design. Compelete accessories. Power source
220V, 50Hz. Compatible onlone UPS imported.
Do not fulfill
assment criteria.
Quoted warrenty
is 1Y instaed of 3Y
& UPS not quoted.
Also offered
model is not fully
automatic
Did not fulfill
assessment
criteria. Quoted
warrenty is 2Y
instead of 3Y.
Yes Yes
Inco
mp
ele
te o
ffe
r m
ake
an
d
mo
de
l is
no
t p
rop
pe
r m
en
tio
n in
the
ir t
ech
nic
al o
ffe
r n
o
war
ren
ty a
nd
val
idit
y m
en
tio
n
in t
he
ir t
ech
nic
al o
ffe
r.
Did not fullfill
assessment
criteria. Quoted
warrenty is one
year instead of
three years. Fixed
angle rotor
Did not fulfill
assessment
criteria warrenty
not mention.
RecommendationTechnically Not
Approved.
Technically Not
Approved.
Technically
Approved
Technically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Incompelete offer
3rd party
distribution and
quoted warrenty
is one year instead
of 3 years.
13. Hamatology Analyzer
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma InternationalPopular Internatinoal
(Pvt) Ltd.AM sale & service Ali Traders
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4
2 Aditional benefit = 2 2
latest incometax paid = 4 4 4
Balance sheet = 8 or
Bank statement = 66 8
4 Sale refrance to the products = 8 3 7
5 Overall reputation to the products = 10 0 7
6In one continent = 4 or
In two continents = 6 (Max 10)10 10
7 Products specialization = 10 0 0
Repair workshop as per manufacturer = 12 12 12
Spare parts = 8 8 8
At least 5 technical persons = 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 5
Total = 100 (Req 65 %) 67 75
Remarks: Graded Graded
14. Coagulation Analyzer
3
8
9
incompelete offer make
and model is not
propper mention in
their technical offer no
warrenty and validity
mention in their
technical offer
Incompelete offer 3rd
party distribution and
quoted warrenty is one
year instead of 3 years.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma International
Popular
InternationalAM sale & service Ali Traders
MODEL CA4000 Humaclot Duo + Not mention Medonic M Series
MANUFACTURERConvergent
TechnologiesHuman SAS Wienerlab
COUNTRY OF ORIGIN Germany Germany France Argentina
AUTHORIZATION Yes Yes
Quality Cirtificate Yes Yes
Warrenty 3Y 3YAutomatic 2 channel, complete with electric pipett,. Random access system. Minimum capacity of 30 cuvettes 3-reagents bottles.. Complete unit ready to use with
standard operational accessories with all test and ratio. Power source 220 V, 50 Hz, AC. Compatible online UPS. Imported from EEC. US, Japan
Yes Yes
Remarks Technically approved Technically approved Technically Not
Approved
Technically Not
Approved
14. Coagulation Analyzer
Incompelete offer 3rd
party distribution and
quoted warrenty is one
year instead of 3 years.
Incompelete offer
make and model is not
propper mention in
their technical offer no
warrenty and validity
mention in their
technical offer.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Hoora Pharma AM sale & service
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
HOORA Pharma
2 Aditional benefit = 2 4
latest incometax paid = 4
Balance sheet = 8 or
Bank statement = 64
4 Sale refrance to the products = 8 6
5 Overall reputation to the products = 10 0
6In one continent = 4 or
In two continents = 6 (Max 10)0
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 0
Spare parts = 8 0
At least 5 technical persons = 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
10
Total = 100 (Req 65 %) 24
Remarks: Not Graded
15. URINE STRIP
incompelete offer make and model
is not propper mention in their
technical offer no warrenty and
validity mention in their technical
offer
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Hoora Pharma AM sale & service
MODEL Not mention
MANUFACTURER Uro-Dispckeck 240e Electronika
COUNTRY OF ORIGIN ERBA Mannheim Hungary
AUTHORIZATION Germany
Quality Cirtificate
Warrenty
Imported from EEC. US, Japan. Complete accessories.
Recommendation Technically Not Approved. Technically Not Approved.
15. URINE STRIP
Incompelete offer make and
model is not propper mention in
their technical offer no warrenty
and validity mention in their
technical offer.
Did not fulfill assessment criteria
warrenty not mention.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma International Sameer Enterprises SU Enterprises HOORA Pharma HOORA PharmaScientific Technical
CorporationAli Traders
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4 4
2 Aditional benefit = 2 2 2 2
latest incometax paid = 4 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 6 7 6 6 8
4 Sale refrance to the products = 8 3 7 7 0 0 6
5 Overall reputation to the products = 10 0 0 0 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 4 10 0 0 10
7 Products specialization = 10 0 0 0 0
Repair workshop as per manufacturer = 12 12 0 0 0 0 12
Spare parts = 8 8 8 8 8
At least 5 technical persons = 10 10 0 10 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 3 8 10 10 6
Total = 100 (Req 65 %) 67 36 60 34 34 70
Remarks: Graded Not Graded Not Graded Not Graded Not Graded Graded
16. Microscope 2Head
16. Microscope 2Head
Incompelete offer 3rd
party distribution and
quoted warrenty is
one year instead of 3
years.
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma International Sameer Enterprises SU Enterprises Hoora Pharma Hoora Pharma
Scientific Technical
CorporationAli Traders
MODEL B-192 G-302 MCX 100 BB4260 Novex-81.600 LX400 DM 500+2
MANUFACTURER Optika Optima Micros Blue Line EUROMAX Labomed Leica
COUNTRY OF ORIGIN Italy Japan Australia Holland Holland USA Germany
AUTHORIZATION Yes Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes Yes
Warrenty 3Y 2Y 1Y 3Y 3Y 3Y
Binocular. Quadruple revolving
nosepiece Built in mechanical
stage with coaxial and vernier
scale Abbes two lens condernser
and diaphragm (Moveable),
filter holder. Rack & Pinion,
Standard filters. Coaxial,
coarse & fine focusing
adjustment. Built in
illuminator. Paired Eyepiece:
10X, Arch: Objectives:
Scanning, 10 x, 40 x,. 100 x,
(spring loasded oil immersion).
Carrying case with lock . Dust
cover. Imported from EEC. US.
Japan. Power source 220 V, 50
Hz, AC. Complete accessories.
YesD
o n
ot
fulf
ill a
ssm
en
t cr
ite
ria.
Qu
ote
d w
arre
nty
is 2
year
s in
ate
sad
of
3ye
ars.
Als
o o
ffe
red
mo
de
l is
no
t fr
om
USA
,Eu
rop
e, J
apan
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar
inat
ead
of
3ye
ars.
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a w
arre
nty
no
t m
en
tio
n.
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a w
arre
nty
no
t m
en
tio
n.
cou
ntr
y o
f o
rigi
n o
f th
e q
uo
ted
mo
de
l an
d m
ake
r is
Ind
ia in
ste
ad
of
USA
.
RecommendationSubject to
Demostration.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Inco
mp
ele
te o
ffe
r 3
rd p
arty
dis
trib
uti
on
an
d q
uo
ted
war
ren
ty is
on
e y
ear
inst
ead
of
3 y
ear
s.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sheikhupura Surgical
16. Microscope 2Head
16. Microscope 2Head
Incomeplet offer
model and make of
quoted items not
mention.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sheikhupura Surgical
Technically Not
Approved.
Inco
me
ple
t o
ffe
r m
od
el a
nd
mak
e o
f q
uo
ted
ite
ms
no
t m
en
tio
n.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma International SU Enterprises Ali Traders Sheikhupura Surgical
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4
2 Aditional benefit = 2 2 2
latest incometax paid = 4 4 4
Balance sheet = 8 or
Bank statement = 66 7
4 Sale refrance to the products = 8 3 7
5 Overall reputation to the products = 10 0 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 10
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 12 0
Spare parts = 8 8 8
At least 5 technical persons = 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 8
Total = 100 (Req 65 %) 67 60
Remarks: Graded Not Graded Not Graded Not Graded
17. Microscope 4 Head
3
8
9
Incomeplet offer model and
make of quoted items not
mention.
Incompelete offer 3rd party
distribution and quoted
warrenty is one year instead of 3
years.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma International SU Enterprises Ali Traders Sheikhupura Surgical
MODEL B193 MCX500 Lilly DM 500+4
MANUFACTURER Optika Micros Leica
COUNTRY OF ORIGIN Italy Austria Germany
AUTHORIZATION Yes Yes
Quality Cirtificate Yes Yes
Warrenty 3Y 1Y
Tetranocuar. Quadruple revolving nosepiece Built in mechanical
stage with coaxial and vernier scale Abbes two lens condernser
and diaphragm (Moveable), filter hoider. Rack & Pinion,
Standard filters. Coaxial, coarse & fine focusing adjustment.
Built in illuminator. Paired Eyepiece: 10X, Arch: Objectives:
Scanning, 10 x, 40 x,, 100 x. (spring loasded -oil immersion).
Carrying case with lock . Dust cover. Imported from EEC US.
Japan. Power source 220 V. 50 Hz.AC Complete accessories
Yes
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
1ye
ar in
ate
ad o
f 3
year
s.
Recomendtaion Technically Approved Technically Not Approved Technically Not Approved Technically Not Approved
17. Microscope 4 Head
Incomeplet offer model and
make of quoted items not
mention.
Inco
mp
ele
te o
ffe
r 3
rd p
arty
dis
trib
uti
on
and
qu
ote
d w
arre
nty
is o
ne
ye
ar in
ste
ad
of
3 y
ear
s.
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sigma International Sameer EnterprisesScientific Technical
CorporationHospitech HOORA Pharma
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4 4 4 4
2 Aditional benefit = 2 2 2
latest incometax paid = 4 4 4 4 4 4
Balance sheet = 8 or
Bank statement = 66 6 8 8 6
4 Sale refrance to the products = 8 3 7 6 0 0
5 Overall reputation to the products = 10 0 0 0 1 0
6In one continent = 4 or
In two continents = 6 (Max 10)10 4 10 10 0
7 Products specialization = 10 0 0 0 0
Repair workshop as per manufacturer = 12 12 0 12 12 0
Spare parts = 8 8 8 8 8
At least 5 technical persons = 10 10 0 10 0 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
8 3 6 4 10
Total = 100 (Req 65 %) 67 36 70 51 34
Remarks: Graded Not Graded Graded Not Graded Not Graded
3
8
9
18. Distilation Plant
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sigma International
Sameer
Enterprises
Scientific Technical
CorporationHospitech Healthcare Hoora Pharma
MODEL L-4B SA-300VL BMS-4CE DIS 4 Plant-15lr
MANUFACTURER JP Select Sturdy BMS Falc Insturment Siemens Water
Technologies
COUNTRY OF ORIGIN Spain Japan Canada Italy Ultra Clear
AUTHORIZATION Yes Yes Yes Yes Yes
Quality Cirtificate Yes Yes Yes Yes Yes
Warrenty 3Y 2Y 2Y 1Y Not Mention
All borosilicate glass, capacity 4 liters with ground glass,
joints, complete. Imported from EEC. US, Japan.
Power source 220 V. 50 Hz AC. Complete accessories.
Yes
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year
s.
Als
o o
ffe
red
mo
de
l is
no
t
fro
m U
SA,E
uro
pe
, Jap
an
Co
un
try
of
ori
gin
is c
anad
a
inst
ead
of
USA
/ E
UR
OP
/
JAP
AN
.
Did
no
t fu
llfill
ass
ess
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
on
e y
ear
inst
ead
of
thre
e
year
s. F
ixe
d a
ngl
e r
oto
r
Did
no
t fu
lfill
ass
ess
me
nt
crit
eri
a w
arre
nty
no
t
me
nti
on
.
RecommendationTechnically
Approved
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
Technically Not
Approved.
18. Distilation Plant
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Seico Scientific TradersPopular Internatinoal
(Pvt) Ltd.Ali Traders
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4 4
2 Aditional benefit = 2
latest incometax paid = 4 4 4
Balance sheet = 8 or
Bank statement = 68 8
4 Sale refrance to the products = 8 5 7
5 Overall reputation to the products = 10 0 7
6In one continent = 4 or
In two continents = 6 (Max 10)4 10
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 0 12
Spare parts = 8 0 8
At least 5 technical persons = 10 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
10 5
Total = 100 (Req 65 %) 45 75
Remarks: Not Graded Graded
19. ESR System
Incompelete offer 3rd
party distribution and
quoted warrenty is one
year instead of 3 years.
3
8
9
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Seico Scientific Popular International Ali Traders
MODEL Vesmatic 20
MANUFACTURER LENA Humased 100mix Diesse Diagnostic
COUNTRY OF ORIGIN Linear Human Germany
AUTHORIZATION Spain Germany
Quality Cirtificate
Warrenty 1Y
Imported from EEC. US. Japan. Compete accessories. May be retnder with generalized and broad specifications.
19. ESR System
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Scientific Technical Corporation
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4
2 Aditional benefit = 2 2
latest incometax paid = 4 4
Balance sheet = 8 or
Bank statement = 68
4 Sale refrance to the products = 8 6
5 Overall reputation to the products = 10 0
6In one continent = 4 or
In two continents = 6 (Max 10)10
7 Products specialization = 10
Repair workshop as per manufacturer = 12 12
Spare parts = 8 8
At least 5 technical persons = 10 10
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
6
Total = 100 (Req 65 %) 70
Remarks: Graded
3
8
9
20. Autoclave Electric and Gas operated
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Scientific Technical Corporation
20 Autoclave Electric
MODEL WAC-47
MANUFACTURER WITAGE
COUNTRY OF ORIGIN Germany
AUTHORIZATION Yes
Quality Cirtificate Yes
Warrenty 3Y
Vertical type, electrically heated elements, complete with all accessories, rubber liner. Temp range max 127 +/-
2 C. Capacity 40 liters. Power source 220 V, 50 Hz. AC. Imported from EEC. US, Japan. Complete accessories.
Country of origin is korea instead of Germany
WITAGE is not orignal equipment manufacturer.
Recommendation Technically Not Approved
20. Autoclave Electric and Gas operated
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
S/No. Assessment Criteria Sameer Enterprises
1
Product cirtificate (CE/MHLW/CE/MDD)
for any one =4
any two = 8
4
2 Aditional benefit = 2
latest incometax paid = 4 4
Balance sheet = 8 or
Bank statement = 66
4 Sale refrance to the products = 8 7
5 Overall reputation to the products = 10 0
6In one continent = 4 or
In two continents = 6 (Max 10)4
7 Products specialization = 10 0
Repair workshop as per manufacturer = 12 0
Spare parts = 8 8
At least 5 technical persons = 10 0
Each diploma =1 and
Each Bsc Engineer = 2 and
Each Msc engineer = 5 (Max 10)
3
Total = 100 (Req 65 %) 36
Remarks: Not Graded
3
8
9
21. Diffrential Counter
TECHNICAL EVALUATION REPORT OF BIDS OF ELECTROMEDICAL lAB EQUIPMENTS PURCHAES FOR DHQ HOSPITAL GUJRANWALA 2014-15
Sr.
NONAME OF FIRM Sameer Enterprises
MODEL 7121431
MANUFACTURER JP selecta
COUNTRY OF ORIGIN Spain
AUTHORIZATION Yes
Quality Cirtificate Yes
Warrenty 2YDiffererntial counter, microprocessor controlled, 12 channel unit with overlay for differentiating
all types of cells, digital readout on all 12 channels plus the grand total. Power sourcp-220 V^
50 Hz. AC. Imported from EEC. US, Japan. Complete accessories
Recomendation Technically Not Approved.
Do
no
t fu
lfill
ass
me
nt
crit
eri
a. Q
uo
ted
war
ren
ty is
2ye
ars
inat
esa
d o
f 3
year