Technical documentationin the new MDR

Annex I/IIAnnex I/II

Gabriele Catignoli, Andreas Purde

TÜV SÜD Product Service GmbH Slide 1TÜV SÜD Product Service GmbH 16-07-06

Page 2: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

MDR – Annex I, II

1 Core changes, Essential Requirements

2 Role of Common Specifications

3 Technical Documentation3 Technical Documentation

4 Interfaces

TÜV SÜD Product Service GmbH Slide 216-07-06

Page 3: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

1. Core changes, Essential Requirements Annex I

Annex I of the MDR

ER 1a / 2 Requirements for risk management (similar to ISO 14971)(hint: warnings are a risk mitigation measure also MDR wise)

ER 6 How to deal with risk/benefit of Annex XV devices (e.g. cosmetic devices) ER 7.4 Substances in direct contact: less than 0.1% weight or justification: carcinogenic,

mutagenic, toxic to reproduction, endocrine disrupting (link to other regulations) ER 7.1 Impact of processes on material properties ER 7.6 Size and properties of particles / nano materials ER 11.4 Adjustment, calibration and maintenance safe and effective ER 11 5 Sh tibilit d i t bilit ER 11.5 Show compatibility and interoperability ER 13.4aa Reference to 2013/59/EURATOM for ionizing radiation

TÜV SÜD Product Service GmbH Slide 316-07-06

Page 4: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

1. Core changes, Essential Requirements Annex I

Annex I of the MDR

ER 14.3a IT security (IT security measures, protect against unauthorized access) ER 14 3 Mobile computing platforms (size contrast of screen noise) ER 14.3 Mobile computing platforms (size, contrast of screen, noise) ER 15.8 Minimize unauthorized access ER 18 Use by lay persons ER 18 Use by lay persons ER 19.1 Information supplied has to be available on the web site of the manufacturer

Remember: All your TDs have to show compliance with the new essential requirements! As shown: many new or changed requirements!

TÜV SÜD Product Service GmbH Slide 416-07-06

You will have to generate additional evidence!

Page 5: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

2. Role of Common Specifications Article 7

Harmonized Standards Common specifications

Mandatory for devices without medical purpose (annex XV)

When harmonized standards seem insufficient

TÜV SÜD Product Service GmbH Slide 56-Jul-16

Page 6: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Article 8

Shall be kept 10 / 15 years after the last piece is placed on the market Shall be kept 10 / 15 years after the last piece is placed on the market

The term design dossier will not exist anymore and is substituted by “Technicalg y yDocumentation”

Specific content of the technical documentation can be found in dedicated annex (II IIa) Specific content of the technical documentation can be found in dedicated annex (II, IIa)

TÜV SÜD Product Service GmbH Slide 616-07-06

Page 7: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Annex II

Structure is now defined

Device description and specification Information supplied by the manufacturer Design and manufacturing information General safety and performance requirements Risk management Risk management Product verification and validation

TÜV SÜD Product Service GmbH Slide 716-07-06

Page 8: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Annex II

Device description and product specification

Basic UDI DI Patient population Principle of operation (scientifically demonstrated) Novel features Key functional elements Key functional elements Raw materials and technical specifications Reference to previous and similar devices (own devices and other devices) Reference to previous and similar devices (own devices and other devices)

TÜV SÜD Product Service GmbH Slide 816-07-06

Page 9: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Annex II

Design and manufacturing information

Information about the design stages Manufacturing processes and their validation D t il ti it i d fi l d t t ti Details on continuous monitoring and final product testing Identification of all sites (suppliers, sub-contractors)

TÜV SÜD Product Service GmbH Slide 916-07-06

Page 10: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Annex II

Product verification and validation

Results and critical analysis of all verification and validation tests Pre-clinical and clinical data Additi l i f ti i ifi ( di i l d t d ti / ll i l d) Additional information in specific cases (e.g. medicinal products and tissues/cells involved)

Remember: Structure of the TD changed New evidence documents required

TÜV SÜD Product Service GmbH Slide 1016-07-06

New evidence documents required Again: you have to rework your TDs!!!

Page 11: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Article 26

Summary of safety and clinical performance

It shall be submitted to the Notified Body

Th i i t t i l l d fi d i thThe minimum content is clearly defined in theregulation

The commission may set out form and presentationof the data elements to be included in the summary

It is only applicable for class III and implantabledevicesdevices

TÜV SÜD Product Service GmbH Slide 1116-07-06

Page 12: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Article 26

Summary of safety and clinical performance

It shall be made available to the public via Eudamed

It shall be written in a way that it is clear to the usery(and the patient, where applicable)

Its location in terms of availability shall be mentionedIts location in terms of availability shall be mentionedin the accompaying documents

TÜV SÜD Product Service GmbH Slide 1216-07-06

Page 13: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

3. Technical Documentation Article 17 / Annex III

Declaration of Conformity

Reference to Common Specifications applied ismandatory (not standards!)

It shall claim conformity also to other directives,when applicable

It shall mention the basic UDI-DI and the SingleRegistration Number

Is it in an official union language or languagesrequired by the Member State in which the device is

d il blmade available

TÜV SÜD Product Service GmbH Slide 1316-07-06

Page 14: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

4. Interfaces

Interfaces to other parts of the MDR

Post market surveillance results (annex IIa) are part of the technical documentation Clinical investigations (annex XIV) are part of the Technical Documentation Annex I defines requirements for risk management Annex I defines requirements for risk management

Assessment of the TD: Sampling approach for class IIa / IIb IIb implantable (with exceptions) and III every device

During unannounced audits (min. once every five years): „verify that the manufactured device is in conformity with the TD“

Unannounced audits may include sampling from the market: „verify that the manufactured device is in conformity with the TD“

TÜV SÜD Product Service GmbH Slide 1416-07-06

Page 15: Technical documentation in the new MDR - tuev-sued.de · Technical documentation in the new MDR Annex I/II Gabriele Catignoli, Andreas Purde TÜV SÜD Product Service GmbH 16-07-06

Open Discussion