• Forsomemanufacturers,onlyaportionoftheGCinletisactuallyatthetemperaturesetpoint;asignificantthermalgradientexistsbothaboveandbelowthiszone.
• ThethermalprofileofoneGCinletcanvaryfromothersimilarinlets—andvarydramaticallybetweendifferentstyles.
• RemovalordamagetoGCinsulationcanhavealargeeffectontheinlet’sthermalprofile.
InjectingaliquidsampleintoahotGCinletisadynamicandcomplexevent.Ofthemanyparametersthataffectthesuccessofaninjection,inlettemperatureisoneofthemostsignificant.Raisingorloweringtheinlettemperaturesetpointcanhaveaprofoundeffectonhowmuchsampleistransferredontothecolumndependingonsamplevolatilityandthermalsensitivity.But,oncetheinlettemperatureisset,howmuchoftheinletisactuallykeptatthatsetpoint?Moreover,howmightthermalprofileschangebetweeninlets?
Temperature Varies Within and Between Similar InletsThemotivationforthisworkcamefromaquestionabouttheactualtemperatureofanO-ringinstalledinanAgilentsplit/splitlessinletatagiveninlettemperaturesetpoint.(SeeFigure1toidentifythecomponentsofaGCinlet.)InsteadofjustmeasuringthetemperatureinsidealinerneartheO-ring’slocation,weusedathermocoupletomeasuretemperaturealongtheentirelengthofthelinerataconstantinlettemperaturesetpointof250°C.Theresultingthermalprofileconfirmedthatatemperaturegradientexistswithintheinlet.*
Inpreviouswork(www.restek.com/hotseptum),wealsodiscussedthisgradientwithinGCinletsandnotedthatinletthermalpro-filescanvarygreatlybetweenmanufacturers,butwouldtheyvarybetweensimilarinletsfromthesamemanufacturer?Wecheckedanothersimilarinlettocomparethethermalprofilesandfoundthatthesecondinletexhibitedadifferentthermalprofilefromthefirst.AftermeasuringseveralmoreAgilentGCinlettemperatureprofiles,wefoundinlet-to-inletvariationinallcases,eveninostensiblyidenticalinlets(Figure2).
Technical Article
It’s a Matter of Degrees, but Do Degrees Really Matter?
An Observation of GC Inlet Temperature Profile and Inlet-to-Inlet Temperature Variability
By Scott Grossman
www.restek.comInnovative Chromatography Products
* For these experiments, we only measured the thermal profile of the liner inside the inlet, not the entire inlet.
www.restek.com 2
Figure 1: Considering how little of the GC inlet is actively heated by the heating element, it’s no surprise a tem-perature gradient exists—especially if insulation is missing from the top or bottom.
Figure 2: A temperature gradient exists within a GC inlet, and temperature profiles can vary between similar inlets. These variations increase dramatically with the absence of insulation.
Oven Wall
Open Air
Oven
Column
Reducing Nut
Nut Warmer Cup
Inlet Seal
Thermal Nut
AluminumHeater Block
Heating Element
Heater Sensor
LinerPoint of Injection
O-Ring
Inlet BodyPerforated Disk
Insulation
1.2 cm
www.restek.com3
Insulation is Crucial to Minimizing Temperature VariationWedidobserveonesplit/splitlessinletwithsignificantlylowertemperaturesatthetopandbottom.Afterinvestigating,wediscov-eredthatthetopringofinsulation,whichsitsjustbelowtheperforateddiskoftheAgilent6890split/splitlessinletweldment,wasmissing.Someoftheinsulationatthebottomoftheinlet,alongwiththethermalnut,wasalsonotinstalled.Simplyplacinginsula-tioninthetopcavityandinstallingthethermalnutcausedthetemperatureoftheinletlinertomorecloselymatchtheotherinlets(Figure2).Thistestwasavaluablereminderoftheneedtocarefullyreconstructtheinletwhenevertheinsulationisdisturbed.
Wheninsulationismissingatthetopoftheinlet,theinletlosesheattotheopenair;however,wheninsulationismissingatthebottom,theGCoveninfluencesthetemperatureinbothdirections(Figure3).Becausecolumninstallationcanbemorechallengingwiththeinsulatednutwarmercupinstalled,analystsmaybetemptedtoleaveitinadrawer,buttheeffectonyourinlettemperaturecanbesignificant.
Figure 3: Installing the nut warmer cup can help minimize the effects of oven temperature on the actual tempera-ture of the inlet. (Inlet shown below was set to a constant 250 °C.)
Temperature Can Vary Drastically Between Dissimilar InletsThenewlyintroducedAgilentMultimodeInlet(MMI)issaidtobecapableofperformingbothhotsplitandhotsplitlessinjectionslikeanormalsplit/splitlessinlet.But,whenwemeasuredthethermalprofilesfortwoMMIinlets,itwasinterestingtonotehowdifferenttheMMIthermalprofileswerefromasplit/splitlessinlet—adropofover190°Cfromthesetpointtothetopoftheinletasopposedtoaround100°Cforthesplit/splitlessinlets(Figure4).Thisvariationshowsthatchangingequipmentmayalsochangeyourresults,eveniftheequipmentisnominallyabletodothesameanalysis.
Figure 4: The Multimode Inlets (MMIs) we measured experienced almost twice the temperature drop (190 °C) of a standard split/splitless inlet between the inlet setpoint to the top of the liner.
The Effects of Inlet Temperature Variations on ChromatographyAsdemonstratedhere,thermalgradientsexistwithinasingleGCinlet,andtemperatureprofilescanvarybetweensimilar,aswellasbetweendissimilar,inlets.Thesevariations affect the vaporization of a liquid sample (and, thus, the overall success of the analysis).
Initially published in the Restek® Advantage
Questions? Contact us or your local Restek representative (www.restek.com/contact-us).Restek patents and trademarks are the property of Restek Corporation. (See www.restek.com/Patents-Trademarks for full list.) Other trademarks in Restek literature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered in other countries. To unsubscribe from future Restek communications or to update your preferences, visit www.restek.com/subscribe To update your status with an authorized Restek distributor or instrument channel partner, please contact them directly.