Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Mike Al locco, Soham Chakrabor ty, Lesl ie Havens, Daniel le Koch, Andrew Mi l ler, Kr isteen Yee, Stephanie Zambito
TEAM 14026 – MSD I
PRESENTATION
P14026_Detailed Design Review 12/05/13
Problem Statement
Engineering Needs and Requirements
Functional Decomposition
Week 6 System Design
Current System Design
Connection to PEV
Assist & CPR Mode
Tubing
Trachea
Lung
AGENDA
Sensor Placement
Pressure Sensors
Flow Sensors
Bil l of Materials
Budget
Test Plan
Risk Assessment
Plan for MSD II
P14026_Detailed Design Review 12/05/13
Current State
1st generation of a validation system
Desired State
Simulate the respiratory system and measure flow, pressure, volume,
and oxygen concentration
Project Goals/Opportunities
Collect data to evaluate 13027’s PEV
Constraints
ANSI Z79.7
ISO 5469:1967
ASTM F1100-90
ISO 80601-2-12
PROBLEM STATEMENT
P14026_Detailed Design Review 12/05/13
MATRIX
P14026_Detailed Design Review 12/05/13
FUNCTIONAL DECOMPOSITION
P14026_Detailed Design Review 12/05/13
DESIGN PROPOSED WEEK 9
P14026_Detailed Design Review 12/05/13
Testing the LTV
Connection to the MediResp-III
Breathing into the LTV
WHY THE DESIGN CHANGED
P14026_Detailed Design Review 12/05/13
Breaths per Minute
Breaths per Minute at 0.9 Lpb
WHY THE DESIGN CHANGED
MediResp-III LTV Display Observed
4.6 100 17
7 41 23
MediResp-III LTV Display
4.6 26
7 35
10 46
P14026_Detailed Design Review 12/05/13
The LTV takes readings at the beginning and end of the tubing
Using the LTV would require the team to reverse engineer the
LTV
Would be outside the time constraint of the project
Potentially, would be outside the budget constraint of the project
WHY THE DESIGN CHANGED
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
CONNECTION TO PEV
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
Actuation is an
excellent way to make
a process repeatable.
The rotational energy
of a small motor would
have to be converted to
linear pushing/pulling.
An easy -to-execute choice
was to use the rack-and-
pinion
A lead screw was also
considered, but these are
less efficient
ASSIST MODE COMPONENTS
P14026_Detailed Design Review 12/05/13
The syringe is 150mL and we expect to move roughly 100mL
at a time, based on a typical inhalation.
The rack part of the rack -and-pinion must make a solid
connection with the plunger
ASSIST MODE COMPONENTS
100 mL
Dist = x
P14026_Detailed Design Review 12/05/13
The rack will need support to avoid slipping out of place.
The typical solution is to fit the rack into a short channel to
constrain its path.
ASSIST MODE COMPONENTS
Characteristics of the Rack as Designed:
1) Keep rack on a level with the pinion.
2) Leave space (0.01 in) on either side.
3) Fairly wide to allow for easier
connection to plunger of the syringe.
P14026_Detailed Design Review 12/05/13
THREE CARDINAL RULES OF GEAR DESIGN:
Meshing gears must have the same height,
Distance between gears’ teeth (“pitch”) should be as close to same
as possible
Different number of teeth is mechanically advantageous for torque -
to-turn ratio and for speed of rotation.
ASSIST MODE COMPONENTS
P14026_Detailed Design Review 12/05/13
ASSIST MODE COMPONENTS
-35-45 HR
-Pressure Angle 20deg
-Circular Pitch 2mm
-300mm Long
-Pressure Angle 20deg
-Circular Pitch 2mm
-Bore Size to Order
-23 Teeth
SH2MS2B250F7A RM-2M-300
P14026_Detailed Design Review 12/05/13
The gear connection to the stepper motor relies on the motors
default pin being able to find purchase in the gear bore.
The round pin of the motor is 5mm. We expect to order the
gear at a NEMA#23 single-flat bore type.
ASSIST MODE COMPONENTS
P14026_Detailed Design Review 12/05/13
SIZING OF THE STEPPER MOTOR
The maximum expected resistance to motion is a volume of 100mL
of air.
We would like to be able to apply the pressure of roughly half a breath (17 cmH2O or 1.667 kPa)
We would like a fairly quick reaction time of 0.2 sec.
Recall that Work/Time is Power, so analysis yielded:
Moreover, stepper motors are roughly 50% efficient, so the stepper
motor selected was sized at 2 Watts of power.
The stepper motor is also bidirectional to allow for retraction.
ASSIST MODE COMPONENTS
P14026_Detailed Design Review 12/05/13
Originally, the team planned on using the syringe, stepper
motor, and actuation to simulate a patient receiving
compressions
At first team 13027 counted compressions by sensing
pressure
Team 13027 has narrowed the scope of their project
CPR MODE
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
Why PVC?
PVC has many standard sizes and variations for connections.
Will be easier to implement.
PVC has smooth Inner Diameter walls.
Turbulent Vs. Laminar Flow?
If Re<2300 then we can assume flow is laminar:
𝑅𝑒 =ρ𝑉𝐷
μ, for air at STP: ρ = 1.205
𝑘𝑔
𝑚3 , and μ = 1.983 ∗ 10−5𝑃𝑎 ∗ 𝑠𝑒𝑐
Where 𝑄𝑚𝑎𝑥 = 32 Lpm, 𝐴𝑝𝑖𝑝𝑒 = Π(𝑑
2)2 & 𝑄 = 𝑉𝐴, 𝑉𝑚𝑎𝑥 = 0.027
𝑚
𝑠,
𝑅𝑒𝑚𝑎𝑥 =1.205∗.027∗.0127
1.983∗10−5 = 20.83
20.83<<2300 therefore we can easily assume flow is laminar.
TUBING
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
Airway Resistance Basics
𝑅𝑒𝑠𝑖𝑠𝑡𝑎𝑛𝑐𝑒 ≡Δ𝑃
𝑄 [
𝑐𝑚𝐻2𝑂𝐿
𝑆𝑒𝑐
] (1)
Orifice Plate Feasibi l i ty
ṁ =𝐴2
1−(𝐴2𝐴1
)22ρΔ𝑃 (2)
plugging in ρQ=ṁ and solving for 𝐴2 results in,
𝐴2 =(ρ𝑄)2
2ρΔ𝑃+(ρ𝑄)2(1
𝐴1)2
where ΔP comes from Equation (1) (3)
Which results in orifice plate diameters as follows:
These orifice plate ΔPs will need to be validated; diameters may need to be adjusted based on empirical correction factors and experimental results.
TRACHEA RESISTANCE
Target Resistance
(cmH2O/L/Sec) d UB flow (mm) d LB flow (mm) davg (mm)
5 5.652 4.258 4.955
25 3.81 2.854 3.332
50 3.207 2.401 2.804
100 2.697 2.019 2.358
150 2.438 1.825 2.1315
200 2.27 1.698 1.984
P14026_Detailed Design Review 12/05/13
TRACHEA RESISTANCE CONT’D.
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
How can we vary lung compliance with only one test lung?
Add a weight
Use a rubber band
LUNG COMPLIANCE
P14026_Detailed Design Review 12/05/13
Bezier Curves are splines that are controlled through the
manipulation of control points
BEZIER CURVES
P14026_Detailed Design Review 12/05/13
BEZIER ANIMATION
P14026_Detailed Design Review 12/05/13
Uses Bezier curves to model an ideal breath curve and
generate simulated data based on this curve
We “force” the program to generate a curve that we want and
use it to predict how dif ferent factors will af fect our results
5 th degree Bezier Curve gives more control
(1 − 𝑡 ) 5 𝑃0 + 5 (1 − 𝑡 ) 4 𝑡 𝑃1 + 1 0(1 − 𝑡 ) 3 𝑡 2 𝑃2 + 1 0(1 − 𝑡 ) 2 𝑡 3 𝑃3 + 5 (1 − 𝑡 )𝑡 4 𝑃4 + 𝑡 5 𝑃5
This equation gives an x or y value for a specific t value.
0≤t≤1
Rearranging the equation allows us to get control points from
given coordinates.
Gives us a useful estimation of data
COMPLIANCE PROGRAM
P14026_Detailed Design Review 12/05/13
COMPARISON TO REAL DATA
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
SENSOR PLACEMENT
P14026_Detailed Design Review 12/05/13
SM5470 -Product of Silicon Microstructures Incorporated
Low Pressure Scale 0-3 PSI
Gauge Pressure Sensor
Accuracy is .25%
Used within Medical Instrumentation and Monitoring
PRESSURE READING
P14026_Detailed Design Review 12/05/13
SENSOR PLACEMENT
P14026_Detailed Design Review 12/05/13
TSI 4000 Series Flow Meter
Flow Measurement:
Accuracy +/- 2% of reading
Range 0.01-99.9L
Response: 4ms
Pressure Measurement:
Accuracy +/- 1 kPa
Range 50-199kPa
Computer Cable (mini -DIN to 9 pin D-Sub) allows interface between device and computer.
Outcome: Data logging for flow and volume measurements
*Note – same device used by 13027 to calibrate MediResp -IV
FLOW METER
P14026_Detailed Design Review 12/05/13
Tested Mike exhalation of breath
Connected tubing to flow meter
Read in real time data
PRELIMINARY FLOW TESTING
P14026_Detailed Design Review 12/05/13
SENSOR PLACEMENT
P14026_Detailed Design Review 12/05/13
4320T15 Miniature PVDF Relief Valve
Temp. Range: 26F to 255F
Use with air
Relieve excess pressure in air lines
Valve body is PVDF
Seal is Buna-N
RELEASE VALVE
P14026_Detailed Design Review 12/05/13
CURRENT SYSTEM DESIGN
P14026_Detailed Design Review 12/05/13
ELECTRICAL SCHEMATIC
P14026_Detailed Design Review 12/05/13
EMANT300 LOW COST USB 24 DAQ
MODULE
Up to 6 channels of differential multiplexed ADC
Single channel 22 bit @ 10 samples/sec
Single channel 16 bit waveform @ 2500 samples/sec
1 channel of 8-bit D/A conversion
8 digital IO channels
One 16-bit general -purpose counter or 16-bit PWM
USB connection
25 pin D-Sub connects to physical world
Application adaptors with instructional guides
LabView compatible
Cost = $99.00 http://www.emant.com/251004.page
P14026_Detailed Design Review 12/05/13
PCB LAYOUT
P14026_Detailed Design Review 12/05/13
BILL OF MATERIALS
Purchasing Req # Part Name Part Number Vendor Quantity, Amount
1 PVC Tubing 6.11942E+11 Home Depot
1, 10 ft, 1/2 inch
diameter
2 Syringe SYRMJ140LL syringepump.com 1, 140 mL
3 Stepper Motor STP-MTR-17040 automationdirect.com 1, 2 W
4 Computer Dell Precision T1700 Dell 1, use lab computer
5 Pressure Sensor 987-sm5470-015-g-b mouser.com 2, 0-1.5 psi
6 TSI Mass Flowmeter 4040 1980339, Rev G TSI 1, 0-300 Std L/min
7 Aluminium Sheet 53538520052 hardwareandtools.com 1
8 Bolts TBHS-104 monsterfastener.com 4
9 O-Rings AS568-021 oringsonline.com 2
10 PVC Flanges 854-005 zorotools.com 2
11 PVC Cement/Primer kit KT1010 spadepot.com 1
12 Valve (release, ball) 4320T15 1, 1 psi
13 Rack+Pinion Gears KWGDL qtcgears.com 1
14 DAQ Board
EMANT300 USB 24 bit
DAQ Module EMANT 1,
15 LabView 1, use lab computer
16 Siemen Lung 6006832 Medical Support Products 1, have
17 PVC Caulk T9AB598814 globalindustrial.com 1, 10.3 oz.
18
Computer Cable (mini-DIN to 9-bin
D-Sub) 1303583 TSI 1
19 RS232 Serial Command Set Manual 1980340 TSI 1
20 D-sub 25 Socket 1125FEA-ND digikey.com 1
21 Op-Amp (buffer) 512-KA2904DTF mouser.com 2
22 PCB oshpark.com 3
23
CKD Series PVC Check Valve for
low pressure CKD050V-PV Plast-O-Matic (amazon.com) 1
P14026_Detailed Design Review 12/05/13
BUDGET
Purchasing Req
# Part Name Sub Total Shipping Total Price
1 PVC Tubing $5.70 $0.00 $5.70
2 Syringe $4.95 $4.95
3 Stepper Motor $18.00 $6.00 $24.00
4 Computer $779.00 $0.00 $0.00
5 Pressure Sensor $30.08 $6.99 $37.07
6 TSI Mass Flowmeter 4040 $925.00 $0.00 $0.00
7 Aluminium Sheet $13.80 $7.29 $21.09
8 Bolts $1.40 $0.00 $1.40
9 O-Rings $7.36 $7.50 $14.86
10 PVC Flanges $9.80 $5.79 $15.59
11 PVC Cement/Primer kit $9.95 $3.95 $13.90
12 Valve (release, ball) $27.78 $27.78
13 Rack+Pinion Gears $40.00 $5.00 $45.00
14 DAQ Board $99.00 $99.00
15 LabView $500.00 $0.00 $0.00
16 Siemen Lung $199.00 $0.00 $0.00
17 PVC Caulk $12.75 $5.00 $17.75
18 Computer Cable (mini-DIN to 9-bin D-Sub) $45.00 $0.00 $0.00
19 RS232 Serial Command Set Manual $0.00 $0.00 $0.00
20 D-sub 25 Socket $5.33 $5.33
21 Op-Amp (buffer) $1.14 $6.99 $8.13
22 PCB $25.00 $0.00 $25.00
23 CKD Series PVC Check Valve for low pressure $36.91 $0.00 $36.91
$2,796.95 $403.46
P14026_Detailed Design Review 12/05/13
Leak Test
Test sequentially over build of assembly using flow meter
If lose 1% of flow, then look into forming better seals
Calibration Test of Pressure Sensors
Use manometer to validate pressure sensors
If within 5% of the sensor, then proceed
Calibration Test of Flow Meter
Use validated pressure sensors over a known distance to validate flow meter
Take into account the discrepancy between analytical and experimental values
Timing of Actuation
Use stopwatch or switch to time and validate actuation is working as expected
See mitigation plan (in future slide)
TEST PLAN FOR OUR SYSTEM
P14026_Detailed Design Review 12/05/13
RISK ASSESSMENT
P14026_Detailed Design Review 12/05/13
RISK ASSESSMENT
P14026_Detailed Design Review 12/05/13
RISK ASSESSMENT
Failure Mode Trigger Mitigation Plan Owner Failure of communication
between electrical and
mechanical system
(1) System is unable to record or
store data Problemshoot the connections between
sensors and electrical system. Mike,
Stephanie
Software errors (i.e.
compilation errors, run
time error, inputs do not
match outputs)
(1) Receive errors from the
computer, (2) data does not
make sense with the given inputs
Check code and logic behind the code.
Look up the error that the software is
returning.
Mike
Actuation failure (1) If during week 8 actuation is
still a high concern of the team,
(2) expert advice is to disregard
actuation
Change team focus to manually moving
the syringe in standardized fashion.
Leave actuation of the syringe as future
work for a future team.
Danielle
Failure to measure
outputs accurately (1) Data is inconsistent, (2) data
does not make sense Check programming, sensors and
follow test plan. If failure mode
continues, seek expert advice to see
where system is failing.
Leslie
Poor/ineffective
communication between
team members
(1) group member complains, (2)
a group member shows no
improvement in the next phase
after receiving a bad peer review
Team Lead will talk to the team
member or team members to discuss
what changes must occur. If failure
mode continues the team will talk to
the guide.
Kris
P14026_Detailed Design Review 12/05/13
PLAN FOR MSD II
P14026_Detailed Design Review 12/05/13
QUESTIONS?
P14026_Detailed Design Review 12/05/13
BACK-UP SLIDES
P14026_Detailed Design Review 12/05/13
First and last points given to the program wil l be the “anchors” for the spl ine, P0 and P5.
Solve for the rest of the handles using
matrices
1
𝑑𝑒𝑡𝑒𝑟𝑚
𝐴 𝐵 𝐶 𝐷𝐸 𝐹 𝐺 𝐻𝐼 𝐽 𝐾 𝐿
𝑀 𝑁 𝑂 𝑃
𝑒𝑓𝑔ℎ
=
𝑃1𝑃2𝑃3𝑃4
Where c,d,e and f are the other four given points and A through P are the terms from the original equation
MATH BEHIND COMPLIANCE PROGRAM
Control Point x y
P0 0 0
P1 2.250644857 46.93264
P2 0.306596939 1203.165
P3 4.860655676 2243.364
P4 4.454849913 -257.988
P5 6 0
Point 1 0 0
Point 2 1.85 700
Point 3 2.85 1000
Point 4 4 850
Point 5 5 300
Point 6 6 0
P14026_Detailed Design Review 12/05/13
Corrugated pipe can be assumed smooth at corrugation
heights and wavelengths much less than pipe radius.
“Instabilities
of Flow in
Corrugated
Pipes” - Tao
SMOOTH PIPE ASSUMPTION
P14026_Detailed Design Review 12/05/13