Reporting Organization Location of Samples Report Period Comparison Method Used PCR Kit Used Positive Control Value of the Kit Used in symptomatic cases (Ct Avarage (Min.-Max.)) Positive Control Value of the Kit Used in asymptomatic cases (Ct Avarage (Min.-Max.))
: Medlifelab : Accredited laboratory centres of Medlifelab (İstanbul, Tekirdağ, Bolu, Düzce) : 01.10.2020/04.10.2020 : Real Time PCR Method : EasyprepFast Realtime PCR Kit : 25,62 (24,3-27,2)
: 27,63 (24,0-30,4)
E-Mail: [email protected]: 0380 512 0 500 / Fax: 0380 514 3 999
Kiremit Ocağı Mh. Hacı Bilal Sok. No:3 (Eski Devlet Hastanesi Karşısı ) / DÜZCE
1
T.C. SAĞLIK BAKANLIĞI
ÖZEL MEDLİFE LABORATUVARIKLİNİK KARŞILAŞTIRMA SONUÇLARI
2
E-Mail: [email protected]: 0380 512 0 500 / Fax: 0380 514 3 999
Kiremit Ocağı Mh. Hacı Bilal Sok. No:3 (Eski Devlet Hastanesi Karşısı ) / DÜZCE
3
E-Mail: [email protected]: 0380 512 0 500 / Fax: 0380 514 3 999
Kiremit Ocağı Mh. Hacı Bilal Sok. No:3 (Eski Devlet Hastanesi Karşısı ) / DÜZCE
4
E-Mail: [email protected]: 0380 512 0 500 / Fax: 0380 514 3 999
Kiremit Ocağı Mh. Hacı Bilal Sok. No:3 (Eski Devlet Hastanesi Karşısı ) / DÜZCE
General Manager Signature-StampAnalyst Signature-Stamp
Sensitivity for Symptomatic Positive Cases 202/207 97,58%Sensitivity for Asymptomatic Positive Cases 217/223 97,30%
Detected Positive
Detected Negative
Total
Symptomatic Positive Cases Asymptomatic Positive Cases
202
5
207
217
6
223
Medlife Loboratories from four di�erent sites in Turkey (Istanbul, Tekirdag, Bolu and Düzce) have tested n=430 true positive samples for the Sars Covid-19-2 virus with the Unibioscience rapid antigen test of the manufacturer Unioninvest Ltd. (Skyport Residence No.: 1/64 Beylikduzu, Istanbul).
In the group of samples of symptomatic cases (n=207), the sensitivity for the detection of SARS Covid-19-2 at an average cycled-threshold (Ct-value) of 25.62 was 97,58%. (95% CI: 94.45% to 99.21%)
In the group of samples of asymptomatic cases (n=223), the sensitivity for the detection of SARS Covid-19-2 at an average cycled-threshold (Ct-value) of 27.63 was 97,30%.(95% CI: 94.24% to 99.01%)