    System Recovery

    System Recovery

    By executing system recovery, you can return the system of the E5070B/E5071B (the

    Windows operating system and the firmware) to the factory state (at the time of

    purchase*1).

    Notes on executing system recovery

    Executing system recovery causes the following:

    In addition to the Windows operating system and the firmware, the following settings

    of the E5070B/E5071B are returned to the factory state.

    Network setting

    GPIB setting

    Printer setting

    The driver for the supported printer installed after purchase is deleted.

    You need to execute initial registration again.

    Files you created using the save function (files in the D drive) are not affected, but we

    recommend backing them up before executing system recovery for precautionary

    purposes. For more information on backup, refer to Making Backup Files on page 418.

    *1.If the hard disk failed and has been replaced after purchase, the state when the replacement

    was performed is recovered.

    http://-/?-http://-/?-
  • 8/13/2019 sys_rec_e

    2/4

    Chapter 12 319

    Setting and Using the Control and Management Functions

    System Recovery

    g

    Functions

    Procedure to execute system recovery

    NOTE You need the keyboard for this operation.

    Step 1. Shut down the E5070B/E5071B.

    Step 2. Connect the keyboard to the E5070B/E5071B.

    Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.

    Step 4. Press the standby switch of the E5070B/E5071B to turn it on.

    Step 5. When the screen as shown in the figure below appears, press and hold of the

    keyboard until this screen disappears.

    NOTE After several seconds, the next screen appears automatically even if you do not press any

    key, so do not miss it.

    If the above message does not appear, the instrument is at fault; contact your local Agilent

    customer center listed at the end of this manual or distributor.

  • 8/13/2019 sys_rec_e

    3/4

    320 Chapter 12

    Setting and Using the Control and Management Functions

    System Recovery

    Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not

    selected, select it with of the keyboard), and press of the keyboard.

    Step 7. The message as shown below appears. Press of the keyboard. If you want to cancel the

    system recovery, press here.

    Agilent Technologies System Utilities

    Recovery & Backup Options (for the E5070/71B)

    Choose One of the following:

    _________________________________________________________________

    1. Recover Factory Backup Image

    2. Exit

    _________________________________________________________________

    Enter a Choice: _

    NOTE If the above message does not appear, the instrument or the disk for the system recovery is

    at fault; contact your local Agilent customer center listed at the end of this manual or

    distributor.

  • 8/13/2019 sys_rec_e

    4/4

    Chapter 12 321

    Setting and Using the Control and Management Functions

    System Recovery

    g

    Functions

    Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the

    system recovery, press here.

    You chose to Restore your system by installing the original factory

    installed OS and system software.

    WARNING: Press C to Continue only if you are sure that you wantto proceed. The C: Drive will be completely overwritten with no

    chance of recovering any data. Use Option 1 to recover the system

    from a serious malfunction caused by corrupted or inadvertently

    deleted files on the system's primary C: partition.

    Press C to Continue or E to Exit: _

    Step 9. The message as shown below appears. Press of the keyboard to start the system

    recovery. If you want to cancel the system recovery, press here.

    CAUTION! Interrupting this process may leave the system in an

    unstable state. Allow the software to complete the

    backup and recovery process. This may take up to 20

    minutes depending on the system configuration.

    Press C to Continue or E to Exit: _

    CAUTION Never turn off the power during the system recovery because doing so may cause

    serious damage to the E5070B/E5071B.

    Step 10. The system recovery will be complete in about 5 minutes. When the system recovery is

    complete, the message as shown below appears. Press , , and of the

    keyboard at the same time to restart.

    Remove the disk and Press CLT+ALT+DEL to restart your

    system.

    NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent

    customer center listed at the end of this manual or distributor.

    Step 11. After restart, the screen for initial registration appears. Execute initial registration. For

    information on the execution procedure, refer to Installation/Quick Start Guide.

    Step 12. For the E5070B/E5071B equipped with the Option 016 touch screen, execute the

    calibration of the touch screen. For information on the execution procedure, refer to

    Calibration of the Touch Screen on page 322.