Upload
maria-andersson
View
214
Download
1
Embed Size (px)
Citation preview
Research Article
Synthesis and bioanalytical evaluation ofmorphine-3-O-sulfate and morphine-6-O-sulfate in human urine and plasma usingLC-MS/MS
The aim of this work was to synthesize morphine-3-O-sulfate and morphine-6-O-sulfate
for use as reference substances, and to determine the sulfate conjugates as possible heroin
and morphine metabolites in plasma and urine by a validated LC-MS/MS method.
Morphine-6-O-sulfate and morphine-3-O-sulfate were prepared as dihydrates from
morphine hydrochloride, in overall yields of 41 and 39% with product purities of 499.5%
and 498%, respectively. For bioanalysis, the chromatographic system consisted of a
reversed-phase column and gradient elution. The tandem mass spectrometer was oper-
ated in the positive electrospray mode using selected reaction monitoring, of transition
m/z 366.15 to 286.40. The measuring range was 5–500 ng/mL for morphine-3-O-sulfate
and 4.5–454 ng/mL for morphine-6-O-sulfate in plasma. In urine, the measuring range
was 50–5000 ng/mL for morphine-3-O-sulfate and 45.4–4544 ng/mL for morphine-6-O-
sulfate. The intra-assay and total imprecision (coefficient of variation) was below 11% for
both analytes in urine and plasma. Quantifiable levels of morphine-3-O-sulfate in
authentic urine and plasma samples were found. Only one authentic urine sample
contained a detectable level of morphine-6-O-sulfate, while no detectable morphine-6-O-
sulfate was found in plasma samples.
Keywords: Human / LC-MS/MS / Morphine-3-O-sulfate / Morphine-6-O-sulfate /SynthesisDOI 10.1002/jssc.201100739
1 Introduction
Morphine is a natural product with historical use as
therapeutic and abused drug. The two glucuronidated
metabolites morphine-3-O- (M3G) and morphine-6-O-
glucuronide (M6G) have received considerable attention
ever since methods for their determination in body fluids
became available in the 1980 s [1–3]. M3G is the main
metabolite excreted in urine, while M6G is pharmacologi-
cally active and may accumulate during long-term treatment
[4, 5].
The corresponding metabolites morphine-3-O-sulfate
(M3S) and morphine-6-O sulfate (M6S) have received much
less attention and are assumed to be less abundant than the
glucuronides. Morphine ethereal sulfate was reported as a
morphine metabolite in human adults and in cats but this
was based on tentative identification [6–8]. Choonara et al.
detected M3S as a urinary morphine metabolite in children
using HPLC [9]. However, no sulfate metabolites were
detected from heroin in urine [10], or as morphine meta-
bolite in children in plasma [9].
Since glucuronidation capacity is low at birth the sulfate
metabolites are of interest when studying morphine meta-
bolism in children [9]. Further reasons for interest in sulfate
conjugates of morphine is the polymorphism in glucu-
ronidation enzymes [11] making sulfation potentially more
important in some individuals, and the potential pharma-
cologic activity of morphine-O-sulfate conjugates [12].
A fundamental obstacle for this work was the lack of
commercially available M3S and M6S reference substances.
The available synthetic routes to morphine-6-O-sulfate
(M6S, Scheme 1, 3) and morphine-3-O-sulfate (M3S,
Scheme 2, 6) rely on manipulation of 3-acetylmorphine
[13–15] or 6-acetylmorphine [13], respectively. However,
Maria Andersson1
Tomasz Janosik2
Hamid Shirani2
Johnny Slatt3
Andreas Fischer3
Olof Beck4
1Department of LaboratoryMedicine, Division of ClinicalPharmacology, KarolinskaUniversity Hospital, Stockholm,Sweden
2Department of Biosciences andNutrition, Karolinska Institute,Novum, Huddinge, Sweden
3Division of Inorganic Chemistry,Royal Institute of Technology,Stockholm, Sweden
4Department of Medicine,Division of ClinicalPharmacology, KarolinskaUniversity Hospital, Stockholm,Sweden
Received August 19, 2011Revised October 31, 2011Accepted November 7, 2011
Abbreviations: M-d3, morphine-d3; M3G, morphine-3-O-glucuronide; M6G, morphine-6-O-glucuronide; M3S,morphine-3-O-sulfate; M6S, morphine-6-O-sulfate; SRM,selecting reaction monitoring
Correspondence: Maria Andersson, Division of Clinical Pharma-cology C1:68, Huddinge Karolinska University Hospital, 141 86Stockholm, SwedenE-mail: [email protected]: 146-8-58581050
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
J. Sep. Sci. 2012, 35, 367–375 367
these approaches (which all rely on the Welsh acetylation of
morphine) [16] gave, in our hands, products of insufficient
purity, as was revealed by product characterization with
LC-MS/MS. Therefore alternative synthetic procedures were
developed and are included in this report.
Previous methods used for M3S and M6S analysis
comprise TLC, GC and HPLC [8, 9], while LC-MS and
LC-MS/MS are commonly used in more recent work on
other morphine metabolites. In this study we aimed at
developing an LC-MS/MS method for determination and
quantification of M3S and M6S in human urine and
plasma. The purpose is to apply the method to the study
of morphine pain treatment of neonates (www.neoopioid.
eu/) and for study of abnormal metabolic patterns of heroin
addicts [17].
2 Material and methods
2.1 Chemicals
Morphine-d3 (M-d3) was obtained from Cerilliant (Round
Rock, TX, USA). Morphine HCl was from Apoteket AB
(Stockholm, Sweden). Formic acid (pro analysis quality) and
acetonitrile (LiChrosolv isocratic grade for Liquid Chroma-
tography) were from Merck KGaA (Darmstadt, Germany).
Methanol (HiPerSolv CHROMANORM for HPLC-gradient
grade) was purchased from VWR (Radnor, PA, USA).
Dimethyl sulfoxide was from J. T. Baker (Philipsburg,
Netherlands). Ultra-pure water was produced in-house by a
Milli-Q Millipore Water system. All reagents and solvents
for synthesis originated from commercial sources, and were
used as received, except pyridine, which was stored over
sodium hydroxide pellets. Column chromatography was
performed using silica gel (particle size 40–63 mm).
2.2 Synthesis
2.2.1 Morphine-6-O-sulfate (3, Scheme 1)
Morphine hydrochloride (1) (3.22 g, 10 mmol) was
suspended at 0–51C (ice-water bath) under vigorous stirring
in a mixture of CH2Cl2 (50 mL) and saturated aqueous
NaHCO3 (50 mL). Acetic anhydride (2.83 mL, 30 mmol) was
then added over ca. 5 min at 0–51C. The cooling bath was
removed, and the mixture was allowed to warm to room
temperature over 1 h. The layers were separated, and the
aqueous phase was extracted with CH2Cl2 (3� 20 mL). The
combined organic extracts were washed with brine, dried
(MgSO4), and evaporated in vacuum, leaving morphine-3-O-
acetate as a colorless glass. This material was dissolved
immediately in anhydrous pyridine (25 mL), and sulfur
trioxide–pyridine complex (3.50 g, 22 mmol) was added in
one portion. The resulting mixture was stirred at 55–601C
for 4 h. After cooling, most of the solvent was removed
in vacuum, and the foamy tan residue was treated
with ice-cold water (30 mL) and CHCl3 (20 mL). The
organic layer was removed using a pipette, and the solid
material which precipitated in the aqueous layer was
collected by filtration, washed with several portions of
cold water, and dried in vacuum, providing the intermediate
2 as a pinkish solid (2.54 g). The material above was
added in one portion to a 5% solution of NaOH in
methanol (50 mL). The mixture was stirred at room
temperature for 1 h and 45 min. After concentration of the
mixture at reduced pressure to ca. 20% of the original
volume, 5% aqueous solution of acetic acid was added,
adjusting the pH to ca. 5–6. Upon standing at room
temperature for 24 h, crystals separated from the
solution. This material was collected by filtration, washed
with water, and recrystallized twice from water (with
filtration while hot), to give M6S (3) as a dihydrate (1.66 g,
41% over three steps) as colorless needles. The 1H-NMR
data were in good agreement with those published
previously [18]. The purity of the material produced using
this procedure was typically 499.5% as determined by
LC-MS/MS. The structure was further confirmed by X-ray
diffraction (Fig. 1A), which also revealed that 3 crystallizes
as a dihydrate. Crystal data: C17H21NO8S, orthorhombic,
Scheme 1. Reagents and conditions: (a) Ac2O, NaHCO3 (aq),CH2Cl2, 0–51C to room temperature, 1 h; (b) pyridine �SO3,pyridine, 55–601C, 4 h; (c) NaOH, MeOH, room temperature,1 h, 41% from 1.
HO
HO
O
HNMe
HCl
a,b
AcO
TBSO
O
HNMe
c
HO
TBSO
O
HNMe
d
O3SO
HO
O
HN
H
Me
41
5 6
Scheme 2. Reagents and conditions: (a) Ac2O, NaHCO3 (aq),CH2Cl2, 0–51C to rt, 1 h; (b) TBSCl, imidazole, DMF, roomtemperature, 72 h; (c) Na2CO3, MeOH, 48 h, 69% from 1; (d)pyridine �SO3, pyridine, 55–601C, 4 h, 56% from 5.
J. Sep. Sci. 2012, 35, 367–375368 M. Andersson et al.
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
P212121, a 5 9.799(4), b 5 11.688(11), c 5 15.50(4) A,
V 5 1775(5) A3, Z 5 4, 3443 reflections, 250 L.S. parameters,
R1 5 0.070, wR2 5 0.147, Flack parameter x 5 0.18(19). A
CIF file with the crystallographic data is available as
Supporting Information.
2.2.2 Morphine-3-O-sulfate (6, Scheme 2)
Morphine-3-O-acetate was prepared as above on half the
scale, starting from morphine hydrochloride (1) (1.61 g,
5 mmol) and acetic anhydride (1.42 mL, 15 mmol). After
work-up as above, the crude product was dissolved in dry
DMF, followed by addition of tert-butyldimethylsilyl chloride
(TBSCl) (1.51 g, 10 mmol), and imidazole (1.36 g, 20 mmol).
The resulting mixture was stirred at room temperature for
72 h, whereupon it was partitioned between CHCl3 (50 mL)
and water (100 mL). The layers were separated, and the
aqueous phase was extracted with CHCl3 (3� 25 mL). The
combined organics were dried over MgSO4, and concen-
trated in vacuum, leaving crude 4 as a colorless viscous oil.
The product 4 was dissolved in methanol (100 mL), and
Na2CO3 (3.20 g, 30 mmol) was added. The resulting mixture
was stirred at room temperature for 36 h. Removal of the
solvent in vacuum gave a solid residue, which was treated
with water (100 mL), followed by addition of 10% aqueous
acetic acid until pH 6. The mixture was extracted with
CHCl3 (4� 25 mL). The combined organic extracts were
washed with water, dried over MgSO4, and concentrated in
vacuum to give a solid cream-colored residue. This was
subjected to flash column chromatography, initially using
CH2Cl2 as the eluent, followed by 5% methanol in CH2Cl2containing 0.5% triethylamine, to provide the TBS-protected
intermediate 5 (1.38 g, 69% over 3 steps) as a white solid
([M1H]1 m/z 400). During some runs, a di-silylated deri-
vative (identified by LC-MS (ESI); m/z 514 [M1H]1),
originating from unreacted morphine from the first step,
was present as an impurity even after chromatography
(elutes just before the product). This can be easily removed
by trituration of 5 with Et2O, leading to high purity material.
A mixture of 5 (195 mg, 0.49 mmol) and sulfur trioxide-
pyridine complex (400 mg, 2.51 mmol) in pyridine (10 mL)
was heated at 55–601C for 4 h. After cooling, the mixture was
evaporated in vacuum at 30–401C until completely dry. Water
was added to the residue (50 mL), followed by saturated
aqueous NaHCO3, adjusting the pH to 6. This mixture was
heated until a clear solution was obtained. After filtration while
warm, the solution was left standing at room temperature for
18 h, resulting in crystallization of the product as colorless
needles. The crystals were collected by filtration, washed with
several portions of water. This material was recrystallized from
water, to provide M3S (6) as a dihydrate (110 mg, 56%) as
colorless needles. MS (ESI) m/z 364 [M–H]�. The 1H-NMR
data were in good agreement with those published previously
[18]. The purity of the product obtained using this procedure
was typically 498% as determined by LC-MS/MS. The overall
yield from morphine hydrochloride (1) was 39%. The single
crystal X-ray analysis (Fig. 1B) demonstrated that 6 was
obtained as a dihydrate. Crystal data: C17H21NO8S, ortho-
rhombic, P212121, a 5 6.9041(8), b 5 13.564(2), c 5 18.867(2) A,
V 5 1766.8(4) A3, Z 5 4, 3225 reflections, 248 L.S. parameters,
R1 5 0.054, wR2 5 0.149, S 5 1.06, Flack parameter
x 5 –0.09(16). A CIF file with the crystallographic data is
available as Supporting Information.
2.3 Patient samples
Patient samples were de-coded surplus samples obtained
from the routine flow of clinical samples sent to the
laboratory for analysis.
2.4 Sample preparation
2.4.1 Urine
Urine samples were prepared for analysis by dilution
with ultra-pure water containing internal standard M-d3
(340 ng/mL). A 25-mL aliquot of urine was added to 100 mL
of internal standard solution in an autosampler vial and
mixed for about 10 s.
Figure 1. The crystal structures of (A) morphine-6-O-sulfate(M6S, 3) dihydrate (B) morphine-3-O-sulfate (M3S, 6) dihydrate.In the case of 6, the water molecules are omitted, as the locationof their hydrogen atoms could not be determined.
J. Sep. Sci. 2012, 35, 367–375 Liquid Chromatography 369
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
2.4.2 Plasma
Whole blood was sampled in heparinized vacuum tubes, the
samples were centrifuged at 3000� g for 10 min and the
plasma was collected. The plasma samples were then
prepared by acetonitrile precipitation. A 50-mL aliquot of
plasma and 100 mL of acetonitrile containing the internal
standard M-d3 (500 ng/mL) were mixed in Eppendorf tubes
for about 10 s, and centrifuged for 10 min at 16 000� g. The
resulting supernatant was transferred to a glass-tube and
evaporated to dryness under nitrogen at 401C. Finally, the
residue was reconstituted with 30 mL of 0.1% aqueous
formic acid and transferred into an autosampler vial.
2.5 Analysis
The instrument used was an ACQUITY UPLC system
connected to a Quattro Premier XE tandem mass spectro-
meter (Waters, Milford, MA, USA). All data were collected
and analyzed with the MassLynx 4.1 software. The analytical
column used was an ACQUITY UPLC HSS T3,
2.1� 100 mm, 1.8 mm particle size, kept at 601C. For the
analysis of plasma extracts, the analytical column was
combined with an ACQUITY HSS T3 VanGuard Pre-
column, 2.1� 5 mm, 1.8 mm particle size. The injection
volume was 2 mL for the urine extracts (using partial loop
with needle overfill) and 3 mL for the plasma extracts (full
loop). The injector loop volume was 5 mL.
A gradient elution was used using mobile phase A: 0.1%
formic acid and mobile phase B: methanol. The gradient for
the urine extracts started at 9.5% mobile phase B for
0.50 min, increased to 30% B for 1.60 min, then further
increased to 95% B for 1.40 min and was held at 95% B for
0.40 min. The column was then equilibrated for 1.60 min at
9.5% B. The mobile phase flow was 0.20 mL/min. The
gradient for the plasma extracts was modified to be held at
100% B for 2.00 min and the column was equilibrated for
1.00 min at 9.5% B. This made the run time for plasma
extracts 1 min longer. Total run times were 5.5 min for
urine and 6.5 min for plasma samples.
The tandem mass spectrometer was operated in the
positive electrospray mode using selecting reaction moni-
toring (SRM). The capillary voltage was 1 kV, extractor
voltage 4 V and the RF lens was set to 0 V. The source
temperature was 1301C and the desolvation temperature
was 3501C. The cone gas flow rate, N2, was 50 L/h, desol-
vation gas, N2, flow rate was 900 L/h and the collision gas,
Ar2, flow rate was 0.30 mL/min. Table 1 shows the indivi-
dual settings for each analyte. One transition was monitored
for quantification, while recording in negative mode was
used for qualification because of lack of a second prominent
transition ion. The analysis of M, M3G and M6G was
performed separately according to an earlier reported
procedure [19].
2.6 Standards and quality controls
Calibration curves were constructed from peak area ratios
between SRM transitions for M3S and M6S and the internal
standard, M-d3. The weighting factor used for the linear
regression calculation was 1/x. Standards were prepared in
both blank urine and blank plasma from stock solutions in
dimethyl sulfoxide. Seven concentration levels were used for
plasma ranging from 5 to 500 ng/mL for M3S and 4.5 to
454 ng/mL for M6S and 8 concentration levels were used for
in urine ranging from 50 to 5000 ng/mL for M3S and 45.4 to
4544 ng/mL for M6S. Quality controls were prepared in the
same manner at three levels, 15, 50 and 300 ng/mL for M3S
in plasma and 13.6, 45.4 and 272 ng/mL for M6S in plasma.
The three concentration levels prepared in urine were 150,
500 and 3000 ng/mL for M3S and 136, 454 and 2720 ng/mL
for M6S. The calibrators and quality controls were stored at
�201C.
2.7 Validation
The limit of detection (LOD) was estimated by using spiked
blank plasma and urine. The lower limit of quantification
(LLOQ) was estimated based on the total CVo20% criterion
and accuracy 80–120%. Accuracy and precision were studied
by analyzing a replicate of five of each quality control level
during five days (n 5 25). The recovery of analytes during
plasma protein precipitation was calculated according to:
recoveryð%Þ ¼ 100� peak area of plasma spiked before precipitation
peak area of spiked blank plasma extract
Ion suppression was studied using two experimental
designs. The first experiment was post column infusion of a
solution containing both analytes (10 mg/mL) at a rate of
10 mL/min. Extracts of three different blank patient urine
and three different blank patient plasma samples were
thereafter injected while monitoring the SRM transition
366.15 to 286.40. In the second experiment, M3S and M6S
Table 1. MS/MS acquisition parameters used for the SRM mode and retention time
Analyte Precursor ion (m/z) Product ion (m/z) Dwell time (s) Cone voltage (V) Collision energy (eV) Retention time (min)
M3S 366.15 286.40 0.150 40 24 1.80
M6S 366.15 286.40 0.150 40 24 2.55
M-d3 289.35 201.35 0.050 45 25 2.17
J. Sep. Sci. 2012, 35, 367–375370 M. Andersson et al.
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
were added to 12 patient plasma samples at the level of 50
and 45.4 ng/mL, respectively, and 15 patient urine samples
at the level of 1000 and 909 ng/mL, respectively. The area
response of analytes was then compared with the area
response of analytes from a reference solution prepared at
the same concentration levels in mobile phase A. The matrix
effect was then calculated according to:
matrix effectð%Þ ¼ 100� peak area of spiked plasma extract
peak area of spiked mobile phase A� 100
The stability in plasma and urine was studied for 24 h in
room temperature and at 41C. The stability at �201C was
studied for 16 days for plasma and 2.5 month for urine.
Stability of extracts on the autosampler was studied for 12 h
at 121C. Stock solution stability was studied for 5 months in
dimethyl sulfoxide at �201C. The three-cycle freeze and
thaw stability was also studied.
3 Results
3.1 Synthesis of M3S and M6S
Upon screening of various conditions of the preparation of
3-acetylmorphine, we noted that treatment of morphine
hydrochloride (1) with acetic anhydride in a two-phase
mixture consisting of saturated aqueous NaHCO3 and
CH2Cl2 at 0–51C resulted in efficient transformation into
the desired product, without co-formation of detrimental
levels of side-products. The resulting 3-acetylmorphine was
directly subjected to sulfatation [13, 14], providing the
known intermediate 2 (Scheme 1). Cleavage of the acetyl
group of 2 was performed using NaOH in methanol [13],
leading to morphine-6-O-sulfate (3). The high product
purity (499.5%) and overall yield (41%) were very
consistent between different runs. Recrystallization of this
product from water gave 3 as a dihydrate, which was
revealed by X-ray crystallography (Fig. 1). In this context, it
should also be noted that the crystal structure of the
intermediate 2 has previously been reported [20].
M3S (6) has previously been prepared from 6-ace-
tylmorphine following a similar path involving installation
of the sulfate moiety, followed by deacetylation [13].
However, since 6-acetylmorphine is not readily available as
synthetic starting material, a different protecting strategy
was used, where morphine hydrochloride (1) was initially
converted into 3-acetylmorphine, which was immediately
treated with TBSCl in DMF in the presence of imidazole,
providing the silyl ether 4 (Scheme 2). Subsequent deace-
tylation of 4 with sodium carbonate in methanol produced
the 6-O-silylated derivative 5 in 69% overall yield from 1.
The corresponding tert-butyldiphenylsilyl-protected deriva-
tive has previously been used as an intermediate towards 3-
amino-3-desoxy derivatives, and was prepared by a selective
desilylation of 3,6-bis-tert-butyldiphenylsilyl ether at the
phenolic hydroxyl group [21]. Exposure of 5 to sulfur triox-
ide-pyridine complex as in the sequence above was followed
by removal of the pyridine in vacuo. Upon addition of water,
the acidic medium caused the final desilylation, to provide
morphine-3-O-sulfate (M3S, 6) as a dihydrate in 56% yield.
The overall yield from morphine hydrochloride (1) was 39%.
The purity of the material obtained using this protocol was
typically 498%. The identity of 6 was finally rigorously
proven by a single crystal X-ray study, which showed that 6was isolated as a dihydrate.
3.2 Method development
The SRM transitions as well as cone voltage and the
collision energy were optimized by direct infusion of each
analyte including the internal standard, M-d3. A compound
solution of M3S and M6S was infused at a concentration of
10 mg/mL and at a rate of 15 mL/min. Only one useful SRM
transition was found that could be used for both M3S and
M6S. However, a second SRM transition was obtained for
identification by data recording in negative mode.
Two analytical columns were evaluated during the
method development of the LC system. The ACQUITY
UPLC HSS T3, 2.1� 100 mm, 1.8 mm and the ACQUITY
BEH C18 1.0� 100 mm, 1.7 mm. The HSS T3 column is
designed to separate and retain polar compounds in the
reversed-phase mode. There was good separation between
M3S and M6S on both columns but, in addition, separation
between M, M3G and M6G was best on the HSS T3.
Acetonitrile and methanol were both evaluated as mobile
phase B during development of the chromatographic
system. Methanol gave the best separation. The best choice
of internal standard would have been isotopic-labeled M3S
and M6S but these were not available. Instead, M3G-d3,
M6G-d3 and M-d3 were tested for M3S and M6S. M-d3 was
selected because it gave the best calibration curves, i.e.
correlation coefficients for both M3S and M6S.
Figure 2A shows a chromatogram from a plasma cali-
brator containing 100 ng/mL M3S and 91 ng/mL M6S. M3S
and M6S are well separated from each other. The chroma-
togram also shows were M3G, M and M6G elutes in the
chromatographic system. M3S, M and M6G are barely
separated at the baseline. However, this is not necessary
because they are monitored by different transitions in the
mass spectrometer. There is no contribution at the M3S and
M6S channel from the M3G, M6G or M. In the M channel,
there is a small contribution from M3S and M6S transition
and the M3G and M6G transition due to in-source frag-
mentation. A chromatogram obtained from analysis of a
urine calibrator is shown in Fig. 2B. The capacity factor (k0)for the first eluting compound M3G is 0.38, k0 for M3S is
0.62 and k0 for M6S is 1.3.
3.3 Validation
A linear relationship between peak area ratio of analyte to
internal standard and the analyte concentrations were
J. Sep. Sci. 2012, 35, 367–375 Liquid Chromatography 371
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
observed in the range of 5–500 ng/mL for M3S and
4.5–454 ng/mL for M6S in plasma (r 5 0.999, n 5 3). In
urine, the linearity was observed in the range of
50–5000 ng/mL for M3S and 45.4–4544 ng/mL for M6S
(r 5 0.999, n 5 3). LOD (s/n 5 3) was estimated to 6 ng/mL
for M3S and 4 ng/mL for M6S in urine. In plasma
LOD (s/n 5 3) was estimated to 0.6 ng/mL for M3S and
0.3 ng/mL for M6S. The LLOQ was according to the
criterion CV o20% and accuracy 80–120% for the lowest
concentration levels in the calibration curves. Thus, the
LLOQ for M3S in plasma was 5 ng/mL (CV 5 7.4%, n 5 5),
for M6S in plasma 4.54 ng/mL (CV 5 14.8%, n 5 5), for
M3S in urine 50 ng/mL (CV 5 7.0%, n 5 5) and for M6S in
urine 45.4 ng/mL (CV 5 3.7%, n 5 5). The difference of
LLOQ between matrices was due to the dilution factor. A
summary of intra-assay and total imprecision and accuracy
data are shown in Table 2, a and b for both urine and
plasma. All the CV values were o11% and the accuracy was
within 98–111% from their nominal value for all analytes in
both urine and plasma.
The average recovery of M3S during plasma protein
precipitation (n 5 3) was 97.2% at the concentration level of
50 ng/mL, and the average recovery of M6S was 96.7%
(n 5 3) at the concentration level of 45.4 ng/mL. The first
study of matrix effect on ionization was an infusion
experiment with injection of blank extracts of both urine
and plasma. The SRM signal was markedly reduced for both
M3S and M6S following injection of blank urine and plasma
extracts after elution of the void volume. Both M3S and M6S
elutes before the recovery of the signal. The second ion
suppression experiment M3S and M6S were spiked in 12
different plasma samples and 15 different urine samples
and in mobile phase A. The matrix effect data are
summarized in Table 3.
Inte
nsity
(cp
s)
0
500000
1.35 1.8 2.25 2.7Time (min)
M3S
M6S
M-d3
M
M3G
M6G
Inte
nsity
(cp
s)
0
45000
2 2.5 31.5
Time (min)
m/z 289.35 201.35
m/z 366.15 286.40
m/z 289.35 201.35
m/z 366.15 286.40
M-d3
M3S
M6S
A
B
Figure 2. Chromatograms obtained from the LC-MS/MS analysis(A) a plasma calibrator containing 100 ng/mL M3S and 91 ng/mLM6S. Peaks for M3G, M and M6G are also shown in thechromatogram at the concentration level of 100 ng/mL for eachanalyte; (B) a urine calibrator containing 500 ng/mL M3S and454 ng/mL M6S.
Table 2. Intra-assay, total imprecision and accuracy data for M3S and M6S in (a) urine and (b) plasma
M3S M6SConc. added (ng/mL)
n Accuracy% CV%
Conc. added (ng/mL)
n Accuracy% CV%
(a) Urine
Intra-assay Intra-assay
150 25 98 5.4 136 25 105 3.9
500 25 98 6.9 454 25 107 6.4
3000 25 101 5.2 2720 25 111 4.9
Total imprecision Total imprecision
150 25 98 5.4 136 25 105 8.3
500 25 98 9.7 454 25 107 10.1
3000 25 101 7.1 2720 25 111 8.5
(b) Plasma
Intra-assay Intra-assay
15 25 99 8.3 13.6 25 117 10.5
50 25 96 5.2 45.4 25 107 6.6
300 25 98 5.5 272 25 110 6.5
Total imprecision Total imprecision
15 25 99 8.3 13.6 25 117 10.5
50 25 96 5.2 45.4 25 107 10.9
300 25 98 5.5 272 25 110 10.7
The experiment was performed with five replicates for five days.
J. Sep. Sci. 2012, 35, 367–375372 M. Andersson et al.
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
The matrix effect was greater from plasma extracts than
from urine extracts. The data presented in Table 3 was
obtained from 12 different plasma and 15 different urine
specimens and documents the variability between indivi-
duals.
The stability study was performed on triplicate samples at
the level of 50 ng/mL for M3S and 45.4 ng/mL for M6S in
plasma. Freshly prepared calibrators were used as the refer-
ence. In plasma, M3S and M6S were stable (97–106%) at all
studied storage conditions. In urine, M3S and M6S were stable
(107–116%) at all studied storage conditions. Autosampler
stability of M3S and M6S was documented in urine extracts
(98–101%) and in plasma extracts (90–92%) for 12 h at 121C.
Stock solutions of both M3S and M6S were found to be stable
(101–102%) for 5 months in dimethyl sulfoxide at �201C
using freshly prepared stock solutions as reference.
3.4 Application
Nine authentic plasma samples were analyzed of which
eight samples contained quantifiable levels of M3S ranging
from 5.8 to 12.9 ng/mL (Table 4). No detectable levels of
M6S in any of the plasma patient samples could be found. A
representative chromatogram from one plasma patient
sample is shown in Fig. 3A. This sample contained
9.6 ng/mL M3S but no detectable M6S. This patient sample
also contained M3G, M and M6G, see Table 4.
A number of 18 patient urine samples were analyzed
(Table 4) selected on the basis of the presence of the heroin-
specific metabolite, 6-acetylmorphine (6-AM). All these
samples contained detectable levels of M3S, 13 samples were
above the quantification level ranging from 69 to 1500 ng/mL.
Only one urine patient sample contained a detectable level of
M6S together with 101 ng/mL M3S (Fig. 3B). Identification
was confirmed by monitoring the SRM transition 364.2 to
364.2 in negative mode (dwell time 0.100 s, cone voltage to
40 V and collision energy 10 eV). This sample also
contained high levels of M, M3G, M6G and 6-AM. Table 4 also
shows the ratio between M3S and M3G. The ratio between
M3S and M3G appears to be higher in plasma as compared
with urine.
Table 3. Results from matrix effect experiment
Urine n 5 15 Plasma n 5 12
M3S M6S M3S M6S
Matrix effect% �4.4710.4 14.577.6 �47.875.8 �37.474.8
Table 4. De-coded surplus samples collected at random from the routine flow of clinical samples sent to the laboratory for analysis
Patient sample M3S (ng/mL) M6S (ng/mL) Morphine (ng/mL) M3G (ng/mL) M3S/M3G %
Plasma 1 12.9 a) 5.0 2200 0.59
Plasma 2 6.25 a) a) a) No ratio
Plasma 3 a) a) a) 230 No ratio
Plasma 4 6.5 a) a) 510 1.3
Plasma 5 5.8 a) a) 3.0 193
Plasma 6 8.2 a) 50 60 13.7
Plasma 7 7.3 a) 60 80 9.1
Plasma 8 9.6 a) 60 140 7.0
Plasma 9 8.2 a) 20 100 8.2
Urine 1 101 b) 15 780 78 900 0.13
Urine 2 b) a) 890 12 100 No ratio
Urine 3 81 a) 3870 33 800 0.24
Urine 4 b) a) 780 11 200 0.17
Urine 5 74 a) 2560 38 600 0.19
Urine 6 155 a) 10 210 71 100 0.22
Urine 7 1500 a) 57 960 230 800 0.65
Urine 8 b) a) 2660 11 200 No ratio
Urine 9 69 a) 1560 26 600 0.26
Urine 10 93 a) 3530 29 600 0.31
Urine 11 493 a) 18 530 108 700 0.45
Urine 12 94 a) 2670 44 300 0.21
Urine 13 73 a) 7320 54 100 0.13
Urine 14 499 a) 38 760 199 100 0.25
Urine 15 b) a) 2450 31 900 No ratio
Urine 16 460 a) 24 800 111 500 0.41
Urine 17 b) a) 610 6400 No ratio
Urine 18 83 a) 3530 29 600 0.28
(a) Not detectable (M3S o 0.6 ng/mL and M6S o 0.3 ng/mL in plasma and M3S o 6 ng/mL and M6S o 4 ng/mL in urine).
(b) Detectable.
J. Sep. Sci. 2012, 35, 367–375 Liquid Chromatography 373
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
4 Discussion
4.1 Synthesis of M3S and M6S
With M6S (3) as the first target, the conditions described by
Welsh for the acetylation of morphine were used (acetic
anhydride in the presence of saturated aqueous NaHCO3)
[16]. The material obtained from these experiments was
thereafter exposed to the sulfur trioxide–pyridine complex,
followed by cleavage of the acetyl group, in line with
previous syntheses [13, 14]. However, we found that the
quality of the resulting M6S (3) was not satisfactory, as it
was contaminated with morphine residues. We attributed
this issue to the fact that, in our hands, the Welsh
acetylation procedure gave 3-acetylmorphine containing
the side-product heroin, as well as some unreacted
morphine, as judged from TLC and LC-MS analyses. In
addition, it was noted that 3-acetylmorphine is rather labile,
and even carefully purified samples underwent degradation
within days into mixtures containing morphine and heroin
upon cold storage in the dark. This implied that
3-acetylmorphine must be used immediately after prepara-
tion, thereby avoiding problems with impurities and side-
reactions in subsequent steps. Very similar approaches to
those mentioned above have been employed very recently
for synthesis of both M6S (3) and M3S (6) [18], but
unfortunately, neither the product yields, nor the purities
were accounted for. Consequently, as highly pure samples
of M6S (3) and M3S (6) were required for reference
purposes, it was decided that modified synthetic approaches
had to be employed.
4.2 Bioanalytical aspect
A validated bioanalytical method for accurate determination of
M3S and M6S in plasma and urine was successfully
developed. One further improvement could be to include
stable isotope labeled analogues of M3S and M6S as internal
standards to better compensate the matrix effect, but these
substances are not yet available. We considered it important to
obtain separation between morphine and metabolites in both
the chromatographic system as well as in the mass spectro-
meter. This is mainly due to the production of morphine from
the conjugates by in-source fragmentation. The availability of
M3S and M6S as a reference substance is of value in the
validation of possible interference in any analytical method for
morphine determination in biological samples.
Our study confirms that the morphine sulfate conju-
gates are less abundant than the glucuronides. We have for
the first time been able to demonstrate the presence of M3S
in urine and plasma as morphine metabolites using highly
selective mass spectrometry technique. In future work, the
upper limit of the measuring range may be lowered fivefold
for both plasma and urine. In the work by Choonara et al.,
M3S could be detected using HPLC UV in urine, but not in
plasma [9]. This could be explained by a more than
25-fold increased sensitivity in our method. Regarding M6S,
a signal indicating its presence was observed in only one
urine sample. Therefore, it still remains to more convin-
cingly demonstrate M6S as a morphine metabolite in
humans.
This project was supported by the European Community’sSeventh Framework Programme under grant agreement no.223767 and through the regional agreement on medical trainingand clinical research (ALF) between Stockholm County Counciland Karolinska Institutet, GA 20090490.
The authors have declared no conflict of interest.
5 References
[1] Svensson, J.-O., Rane, A., Sawe, J., Sjokvist, F.,J. Chromatogr. 1982, 230, 427–432.
[2] Svensson, J.-O., J. Chromatogr. 1986, 375, 174–178.
[3] Milne, R. W., Nation, R. L., Reynolds, G. D., Somogyi,A. A., Van Crugten, J. T., J. Chromatogr. 1991, 565,457–464.
[4] Glare, P. A., Walsh, T. D., Ther. Drug Monit. 1991, 13,1–23.
0
450000
1.5 2 2.5 3Time (min)
Inte
nsity
(cp
s)
M-d3
M3S
m/z 289.35 201.35
1.50
42000
2 2.5 3
Inte
nsity
(cp
s)
Time (min)
M-d3
M3S
M6S
m/z 366.15 286.40
m/z 289.35 201.35
m/z 366.15 286.40
A
B
Figure 3. Chromatograms obtained from the LC-MS/MS analysisof (A) an authentic plasma sample containing 9.6 ng/mL M3Sand no detectable peak for M6S, the transition m/z 366.15 to286.40 was multiplied 50 times; (B) an authentic urine samplecontaining 101 ng/mL M3S and a detectable peak for M6S, thetransition m/z 366.15 to 286.40 was multiplied five times.
J. Sep. Sci. 2012, 35, 367–375374 M. Andersson et al.
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com
[5] Coller, J. K., Christrup, L. L., Somogyi, A. A., Eur. J. Clin.Pharmacol. 2009, 65, 121–139.
[6] Yeh, S. Y., Chernov, H. I., Woods, L. A., J. Pharm. Sci.1971, 60, 469–471.
[7] Yeh, S. Y., Fed. Proc. 1973, 32, 763.
[8] Yeh, S. Y., J. Pharmacol. Exp. Ther. 1974, 192, 201–210.
[9] Choonara, I. A., Ekbom, Y., Lindstrom, B., Rane, A., Br.J. Clin. Pharmacol. 1990, 30, 897–900.
[10] Yeh, S. Y., McQuinn, R. L., J. Pharm. Sci. 1977, 66,201–204.
[11] Klepstad, P., Dale, O., Skorpen, F., Borchgrevink, P. C.,Kaasa, S., Acta Anaesthesiol. Scand. 2005, 49, 902–908.
[12] Brown, E. C., Roerig, S. C., Burger, V. T., Cody, R. B.,Fujimoto, J. M., J. Pharm. Sci. 1985, 74, 821–824.
[13] Mori, M., Oguri, K., Yoshimura, H., Shimomura, K.,Kamata, O., Ueki, S., Life Sci. 1972, 11, 525–533.
[14] Crooks, P. A., Kottayil, S. G., Al-Ghananeem, A. M.,Byrn, S. R., Butterfield, D. A., Bioorg. Med. Chem. Lett.2006, 16, 4291–4295.
[15] Preechagoon, D., Brereton, I., Staatz, C., Prankerd, R.,Int. J. Pharm. 1998, 163, 177–190.
[16] Welsh, L. H., J. Org. Chem. 1954, 19, 1409–1415.
[17] Beck, O., Bottcher, M., J. Anal. Toxicol. 2006, 30,73–79.
[18] Varadi, A., Gergely, A., Beni, S., Jankovics, P., Noszal,B., Hosztafi, S., Eur. J. Pharm. Sci. 2011, 42, 65–72.
[19] Gustafsson, E., Andersson, M., Stephanson, N., Beck,O., J. Mass Spectrom. 2007, 42, 881–889.
[20] Pratt Brock, C., Kottayil, S., Butterfield, D. A., Crooks,P. A., Acta Crystallogr., Sect. C: Cryst. Struct. Commun.1996, C52, 122–125.
[21] Wentland, M. P., Duan, W., Cohen, D. J., Bidlack, J. M.,J. Med. Chem. 2000, 43, 3558–3565.
J. Sep. Sci. 2012, 35, 367–375 Liquid Chromatography 375
& 2012 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.jss-journal.com