SVS 2000 Signal Processor
• Unpacking &Inspection
• Mounting• Wiring• Sealing Openings• Installing optional
PCBs
SVS 2000 Signal Processor
SVS 2000™
32 4
ZERO
TARE
NET
GROSS
FUNC
76 8
0
1
5
9 ENTER
Kistler-Morse R
SVS 2000Default Procedure 1
• Use CAUTION when performing this procedure (Press ENTER after each step)– From RUN Select Function 94– DEFAULT SYSTEM– DEFAULT NVRAM?– ARE YOU SURE?– Perform normal “Quick Config” procedure.
SVS 2000Default Procedure 2
• Use CAUTION when performing this procedure– Hold the ENTER key– Cycle Power– Release the ENTER key– Perform normal “Quick Config” procedure.
SVS 2000Signal Processor
• Quick Config– Complete Table 3-1 Information– Default Unit– Use ENTER key to cycle through steps and
enter required information– Perform live load (Auto Cal) calibration when
installation constraints permit.
SVS 2000Signal Processor
• Display Menu [10]– [11] Disp Format– [12] Disp Countby– [13] Disp Units– [14] Bargraph Span (only available in Gross
Mode)• Special Use of ENTER Key
SVS 2000Signal Processor
• Cal (calibration) Menu [20]– [21] Auto Cal– [22] Manual Cal– [23] Linearization– [24] Zero Tolrance– [29] Cal Disply
SVS 2000Signal Processor
Setpoint Menu [30]– [36] Slct Digtl #4– [37] Slct Digtl #5– [38] Slct Digtl #6– [39] Digtl Status– Setpoint Test
– [31] Slct Relay #1– [32] Slct Relay #2– [33] Slct Digtl #1– [34] Slct Digtl #2– [35] Slct Digtl #3
This section of pipe contains 20 lbs of material whichcontinues to flow into vessel after valve closes
Relay energizes at 8,980 lbsRelay de-energizes at 7,980 lbs
Relay de-energizes at 2,500 lbs
Relay energizes at 2,000 lbs
Setpoint 2 Value = 2,000 lbs (Lo)
Figure 7-2 Setpoint Example
Deadband = 500 lbs
Deadband = 1,000 lbs
Setpoint 1 Value = 9,000 lbs (Hi)Preact = 20 lbs
ValveLocation
SVS 2000Signal Processor
– [41] IOut Range– [42] IOut: 4/0mA– [43] IOut: 20mA– [44] IOut Track
– [45] Iout Failsafe– [46] Iout Enable– [48] Current Adjst– [49] Test Current
I-Output (current Output) Menu [40]16 bit D/A
SVS 2000Signal Processor
• Com (communications) Menu [50]– [51] Serial Addres– [52] Serial Baud– [53] PLC Menu
• A-B RIO Menu (available)• DeviceNet Menu (available 1st qtr 99)• Profibus Menu (future)• Modbus Menu (future)
SVS 2000Signal Processor
• Sensor Menu [60]– [61] Sensor Adj Ex– [62] Resolution– [63] Sensor Gain– [69] Sensor Disply
[61] Sensor Adj Ex
• Adjustable from 13 V to 5 V in one volt increment.– 12 V -- Default for K-M Sensors– 10 V -- Default for Metal Foil Sensors
Jumper configuration for half bridge to full bridge
[62] Resolution
Conversion TimeEffective Resolution(bits) (mS) Hertz16 16.67 6017 20.00 5018 25.00 4019 33.33 3020 50.00 2021 100.00 10
[63] Sensor GainSensor Gain Sensor Input
Voltage @12V EXC.1 ±3.02 ±1.54 ±0.758 ±0.37516 ±0.187532 ±0.0937564 ±0.046875
128 ±0.0234375
Max Sensor input = [exc ÷ (4 x Sensor Gain)]
[69] Sensor Disply(display)
FilteredValues
ApplyLinearization Display
Values(for display
inRun Mode)
Apply Averaging,DSP Filter,
and TrackingOutputValues
(for current output,setpoints, and
serialcommunications)
RawValues
ApplyCountby
Figure 10-2. Raw, Filtered, Output, and Display Values
SVS 2000Signal Processor
• Filter Menu [70]– [71] Averaging– [72] DSP Enable– [73] DSP Step– [74] DSP Qualify– [75] DSP Factor– [79] DSP Step Mon (Press ZERO key to
reset)
DSP FilterFunction
ReferenceMedian
DSPStep
DSPStep
Weight3rd successiveweight on sameside of median
TimeExample 2 - DSP Qualify =3,DSP Factor = 80%
New median - reference median+ 0.8 x (reference median - last weight)
ReferenceMedian
Filtered Weight
ReferenceMedian
DSPStep
DSPStep
WeightWeight falls outside ofStep envelope
New median = last weight
TimeExample 1 - DSP Qualify =3,DSP Factor = 80%Raw Weight
SVS 2000Signal Processor
• Tracking Menu [80]– [81] Rate Threshld– [82] Sample Intrvl– [83] Zero Trak Win– [84] Drift Limit– [85] Mat Trak Enab– [89] Rate Monitor
Accumulateddrift withinMaterialTrackingdrift limit Accumulated
drift withinMaterialTrackingdrift limit
Rate ofchange withinthreshold --material notmoving
Rate ofchangeexceedsthreshold --materialmoving
Rate of changewithin threshold-- material notmoving
Rate of changeexceedsthreshold --material moving
Rate ofchange withinthreshold --material notmoving
Drift withinZero TrackingWindow
Window
Window Time
Weight
Corrected weight
Raw weight
Zerolive load
SVS 2000Signal Processor
• Service Menu [90]– [91] User Access– [92] Keyboard Lock– [93] Keyboard Test– [94] Default Systm– [99] Diagnostics