Cervical DistractorSurgical Technique
Image intensifier control
This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended.
Processing, Reprocessing, Care and MaintenanceFor general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to:http://emea.depuysynthes.com/hcp/reprocessing-care-maintenanceFor general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance
Cervical Distractor Surgical Technique DePuy Synthes 1
Table of Contents
AO Spine Principles 2
Introduction Set Composition 3
Product Information Instruments 4
Handling Instructions 6
Bibliography 9
coronalaxial
sagittal
2 DePuy Synthes Cervical Distractor Surgical Technique
The four principles to be considered as the foundation for proper spine patient management underpin the design and delivery of the Curriculum: Stability – Alignment – Biology – Function.1,2
StabilityStabilization to achieve a specifi c therapeutic out-come
BiologyEtiology, pathogenesis, neural protection, and tissue healing
AlignmentBalancing the spine in three dimensions
FunctionPreservations and resto-ration of function to pre-vent disability
AO Spine Principles
Copyright © 2012 by AOSpine
1 Aebi et al (1998)2 Aebi et al (2007)
Cervical Distractor Surgical Technique DePuy Synthes 3
Set Composition:187.780 Instrument Set for Cervical Distractors
Item no. Units
687.780 Vario Case for Cervical 1 Distractors size ½, with Lid, without Contents, including:
687.780.001 Bottom, size ½, for Vario Case No. 687.780
687.781 Lid, size ½, for Vario Case No. 687.780
Introduction
396.395
396.396
396.967
396.970
396.971
4 DePuy Synthes Cervical Distractor Surgical Technique
Instruments
Item no. Units
396.395 Cervical Distractor, left, 1 with adjustable angle
396.396 Cervical Distractor, right, 1 with adjustable angle
396.967 Screwdriver, hexagonal, 1 length 200 mm, for Screw for Cervical Distractor
396.970 Screw B 2.7 mm, length 14 mm, 2 for Cervical Distractor, Stainless Steel
396.971 Screw B 2.7 mm, length 16 mm, 2 for Cervical Distractor, Stainless Steel
Product Information
311.440
396.968
Cervical Distractor Surgical Technique DePuy Synthes 5
Optional
Item no. Units
311.440 T-Handle with Quick Coupling 1
396.968 Drill Bit B 1.8 mm, length 132/12 mm, 1 for Cervical Distractor, for Quick Coupling
Additionally available
687.780.001 Bottom, size ½, for Vario Case No. 687.780 (automatically included when ordering no. 687.780)
687.781 Lid, size ½, for Vario Case No. 687.780 (automatically included when ordering no. 687.780)
6 DePuy Synthes Cervical Distractor Surgical Technique
The following handling instructions for the Synthes Cer-vical Distractor are intended for illustration of the basic functioning of these instruments. For further information about the surgical steps, please consult the respective surgical technique(s).
To insert the 2.7 mm Screws for Cervical Distractor (396.970 or 396.971), mount them on the self-holding, hexagonal Screwdriver for Cervical Distractor (396.967).
Should insertion prove difficult, pre-drill a hole for the screws using the 1.8 mm Cervical Drill Bit (396.968) mounted on the T-Handle with Quick Coupling (311.440).
Place the left Cervical Distractor (396.395) or right Cervical Distractor (396.396) over two of the 2.7 mm Screws (396.970 or 396.971). The distractor fits the screws loosely to facilitate placement and removal, and is not meant to lock on the screws.
Handling Instructions
2
1
Cervical Distractor Surgical Technique DePuy Synthes 7
To allow ease of placement of the cervical distractor over the screws for cervical distractor regardless of their ori-entation, ensure that the angle-setting tab (1, below) of the cervical distractor is initially unlocked.
Once positioned on the screws, push the angle-setting tab (1) (left) to the locked position (2, below) to ensure parallel distraction of the two screws.
Even with a locked angle-setting tab (2), the distractor still allows variable angulation in one plane to accommo-date different patient heights.
4
3
8 DePuy Synthes Cervical Distractor Surgical Technique
Achieve distraction between the two screws by turning the distraction tab (3) in a clockwise direction. The dis-traction can be adjusted by turning the distraction tab (3) counterclockwise, or released by pressing and hold-ing the release tab (4).
After removal of the cervical distractor, remove the screws using the selfholding hexagonal Screwdriver for Cervical Distractor (396.967).
Handling Instructions
Cervical Distractor Surgical Technique DePuy Synthes 9
Aebi M, Arlet V, Webb JK (2007) AOSPINE Manual (2 vols), Stuttgart, New York: Thieme
Aebi M, Thalgott JS, Webb JK (1998) AO ASIF Principles in Spine Surgery. Berlin Heidelberg New York: Springer
Bibliography
0123
Synthes GmbHEimattstrasse 34436 OberdorfSwitzerlandTel: +41 61 965 61 11Fax: +41 61 965 66 00www.depuysynthes.com ©
DeP
uy S
ynth
es S
pine
, a d
ivis
ion
of S
ynth
es G
mbH
. 201
7.
All
right
s re
serv
ed.
036.
000.
715
DS
EM
/SP
N/0
717/
0708
08
/17
Not all products are currently available in all markets.
This publication is not intended for distribution in the USA.
All surgical techniques are available as PDF files at www.depuysynthes.com/ifu