Surgical ProtocolImplant Placement

Surgical ProtocolCover Screw Insertion

No-Touch™ MountlessOSSEOTITE Certain

Implant Delivery System

Surgical ManualOSSEOTITE ® Certain® Implant Systems

And OSSEOTITE External Hex Implant Systems

Page 2: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

Icon Key:

OSSEOTITE® Certain® Internal Connection Implant System:

OSSEOTITE External HexConnection Implant System:

OSSEOTITE Certain Internal and OSSEOTITE External HexConnection Implant Systems:

How To Use The Icon Key:

The icons represent the connection type of the implantsystem. Both internal and external connection types arerepresented in this manual. In the fully illustrated protocols,each icon is present by each step. When a dark blue icon anda light blue icon are present together, the dark blue indicateswhich system is illustrated. When both icons are dark blue,then both systems are illustrated together.

Instructions For Use:

OSSEOTITE, OSSEOTITE XP®, OSSEOTITE Certain, Certain PREVAIL™,OSSEOTITE NT ®, Miniplant®, Microminiplant ™

This manual applies to dental implants and associated surgical components.

Description: 3i Dental Implants are manufactured from biocompatible titaniumand titanium alloy. 3i Dental Implants include various surface treatments andcoatings. For specific product description and net quantity refer to individualproduct labels.

Indications for Use: 3i Dental Implants are intended for surgical placement in theupper or lower jaw to provide a means for prosthetic attachment in single toothrestorations and in partially or fully edentulous spans with multiple single teeth, oras a terminal or intermediary abutment for fixed or removable bridgework and toretain overdentures.

In addition, when a minimum of four dental implants ≥ 10mm in length, areplaced in the mandible and splinted in the anterior region, immediate loadingis indicated.

Contraindications: Placement of dental implants may be precluded by patientconditions that are contraindications for surgery. 3i Dental Implants should notbe placed in patients where the remaining jaw bone is too diminished to provideadequate implant stability.

Storage and Handling: Devices should be stored at room temperature. Refer toindividual product labels and this manual for special storage or handlingconditions.

Warnings: Excessive bone loss or breakage of dental implant may occur whenan implant is loaded beyond its functional capability. Physiological and anatomicconditions may negatively affect the performance of dental implants. This shouldbe taken into consideration when placing dental implants with the following:Poor bone qualityPoor oral hygieneMedical conditions such as blood disorders or uncontrolled hormonal conditions

Mishandling of small components inside the patients mouth carries a risk ofaspiration and/or swallowing.

Forcing the implant into the osteotomy deeper than the depth established by thedrills can result in: stripping the driver hex interface inside the implant, strippingthe driver, cold-welding of the mount-driver interface to the implant, or strippingthe walls of the osteotomy that may prevent an effective initial implant fixation.

Clinical data have demonstrated enhanced performance of OSSEOTITE Implantsas compared to other 3i Dental Implants in patients with poor quality bone.

Precautions: For safe and effective use of 3i Dental Implants and other surgicaldental accessories, these products or devices should only be used by trainedprofessionals. The surgical techniques required to properly utilize these devicesare highly specialized and complex procedures. Improper technique can lead toimplant failure, loss of supporting bone, restoration fracture, screw looseningand aspiration.

Sterility: All dental implants are supplied sterile and are sterilized by anappropriate validated method. Refer to individual product labels for sterilizationinformation; all sterile products are labeled ‘STERILE.’ All products sold sterileare for single use before the expiration date printed on the product label. Do notuse sterile products if the packaging has been damaged or previously opened.Do not re-sterilize or autoclave except where instructions to do so are providedon the product label, in this manual or in any additional marketing literature forthat product. Products provided non-sterile must be cleaned and sterilizedaccording to the directions found in ART630 or this manual prior to use.

Procedural Precautions, Surgery: During the planning phase it is important todetermine the vertical dimension, the actual space available between the alveolarcrest and the opposing dentition, in order to confirm that the available space willaccommodate the proposed abutment and the final crown restoration. Thisinformation varies with each patient and abutment; therefore it should becarefully evaluated before placing any dental implant. The final prosthesis shouldbe designed prior to the placement of the dental implant. Utilize continuousirrigation with a cool, sterile irrigating solution to avoid excessive damage to thesurrounding tissue and to prevent compromising osseointegration. This ismandatory during all procedures. Avoid excessive pressure during preparation ofthe bone site. As the drilling speed varies based on the instrument and thesurgical procedure, recommendations for speed can be found in this manual.Only sharp instruments of the highest quality should be used for any bonesurgical procedure. Minimizing trauma to the bone and surrounding tissueenhances the potential for successful osseointegration. In order to eliminatecontaminants and other sources of infection, all non-sterile devices should becleaned and/or sterilized prior to use, per the instructions on the individualproduct labels.

Potential Adverse Events: Potential adverse events associated with the use ofdental implants may include:- Failure to integrate- Loss of integration- Dehiscence requiring bone grafting- Perforation of the maxillary sinus, inferior border, lingual plate, labial plate,

inferior alveolar canal, gingiva- Infection as reported by: abscess, fistula, suppuration, inflammation,

radiolucency- Persistent pain, numbness, paresthesia- Hyperplasia- Excessive bone loss requiring intervention- Implant breakage or fracture- Systemic infection- Nerve injury

Caution: U.S. Federal Law restricts this device to sale by or on the order of alicensed dentist or physician.

Page 3: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3

Table Of Contents

Introduction And Treatment Planning . . . . . . . . . . . . . . . . 1

Preoperative Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Top-Down Treatment Planning . . . . . . . . . . . . . . . . . . . . . . 3

Surgical Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Cleaning And Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . 5

3i Depth Marking System . . . . . . . . . . . . . . . . . . . . . . . . 6–10Parallel Walled Implants

Countersink Drill Marking System . . . . . . . . . . . . . . . . . . 11

Mountless Delivery Guidelines. . . . . . . . . . . . . . . . . . . . . 12OSSEOTITE® Certain® Implant

OSSEOTITE Parallel Walled Implant Surgical Tray . . . 13Coordinating The Use Of The Surgical Tray With The SurgicalManual Illustrations

Subcrestal Implant Placement Protocol -Quick ReferenceCertain PREVAIL™ 4/3mm Diameter Implant . . . . . . . . . . . . . . 17Certain PREVAIL 5/4mm Diameter Implant . . . . . . . . . . . . . . . 18Certain PREVAIL 3/4/3mm Diameter Implant . . . . . . . . . . . . . . 14Certain PREVAIL 4/5/4mm Diameter Implant . . . . . . . . . . . . . . 14Certain PREVAIL 5/6/5mm Diameter Implant . . . . . . . . . . . . . . 15OSSEOTITE Certain and OSSEOTITE External Hex

MicroMiniplant™ 3.25mm Diameter Implant . . . . . . . . . . . . 16OSSEOTITE External Hex Standard 3.75mm And OSSEOTITE

XP® 3/4mm Diameter External Hex Implant. . . . . . . . . . . . . 16OSSEOTITE Certain And OSSEOTITE External Hex

4.0mm Diameter Implant. . . . . . . . . . . . . . . . . . . . . . . . . . . 17OSSEOTITE Certain And OSSEOTITE External Hex

5.0mm Diameter Implant. . . . . . . . . . . . . . . . . . . . . . . . . . . 18OSSEOTITE Certain And OSSEOTITE External Hex

6.0mm Diameter Implant. . . . . . . . . . . . . . . . . . . . . . . . . . . 19OSSEOTITE XP 4/5mm And 5/6mm Diameter

External Hex Implant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Subcrestal Implant Placement ProtocolCertain PREVAIL 4/3mm Diameter Implant . . . . . . 34–35, 41–43Certain PREVAIL 5/4mm Diameter Implant . . . . . . 36–37, 41–43Certain PREVAIL 3/4/3mm Diameter Implant . . . . . 21–22, 28–29Certain PREVAIL 4/5/4mm Diameter Implant . . . . . 23–24, 28–29Certain PREVAIL 5/6/5mm Diameter Implant . . . . . . . . . . . 25–29OSSEOTITE Certain and OSSEOTITE MicroMiniplant

3.25mm Diameter External Hex Implant . . . . . . 30–31, 41–43OSSEOTITE External Hex Standard 3.75mm And OSSEOTITE

XP 3/4mm External Hex Diameter Implant . . . . 32–33, 41–43OSSEOTITE Certain And OSSEOTITE External Hex

4.0mm Diameter Implant. . . . . . . . . . . . . . . . . . 34–35, 41–43

OSSEOTITE Certain® And OSSEOTITE® External Hex5.0mm Diameter Implant. . . . . . . . . . . . . . . . . . 36–37, 41–43

OSSEOTITE Certain And OSSEOTITE External Hex6.0mm Diameter Implant. . . . . . . . . . . . . . . . . . . . . . . . 38–43

OSSEOTITE XP® 4/5mm Diameter External HexImplant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23–24, 28–29

OSSEOTITE XP 5/6mm Diameter External Hex Implant . . . 25–29

OSSEOTITE NT® Implant System . . . . . . . . . . . . . . . . . . . 44Why Tapered Implants Are Different

Quad Shaping Drill (QSD)Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Key Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

OSSEOTITE NT Depth Indicator (NTDI) . . . . . . . . . . . . . . 47Introduction

OSSEOTITE NT Implant Taps . . . . . . . . . . . . . . . . . . . . . . . 48Introduction

OSSEOTITE NT Surgical Trays . . . . . . . . . . . . . . . . . . . . . . 49Coordinating The Use Of The Surgical Tray With The SurgicalManual Illustrations (NTDI/QSD)

Subcrestal Implant Placement Protocol -Quick ReferenceOSSEOTITE NT Implants Using QSD/NTDI. . . . . . . . . . . . . . . . 50

Subcrestal Implant Placement ProtocolOSSEOTITE NT Certain And OSSEOTITE NT External Hex

3.25mm Diameter Implant Using QSD/NTDI . . . 51–52, 56–58OSSEOTITE NT Certain And OSSEOTITE NT External Hex

4.0mm Diameter Implant Using QSD/NTDI . . . . . . . 53, 56–58OSSEOTITE NT Certain And OSSEOTITE NT External Hex

5.0mm Diameter Implant Using QSD/NTDI . . . . . . . 54, 56–58OSSEOTITE NT Certain And OSSEOTITE NT External Hex

6.0mm Diameter Implant Using QSD/NTDI . . . . . . . . . . 55–58

Subcrestal Implant Placement ProtocolIn Soft Type IV BoneOSSEOTITE NT Certain And OSSEOTITE NT External Hex

4.0mm Diameter Implant. . . . . . . . . . . . . . . . . . 59–60, 66–68OSSEOTITE NT Certain And OSSEOTITE NT External Hex

5.0mm Diameter Implant. . . . . . . . . . . . . . . . . . 61–62, 66–68OSSEOTITE NT Certain And OSSEOTITE NT External Hex

6.0mm Diameter Implant. . . . . . . . . . . . . . . . . . . . . . . . 63–68

Single-Stage Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . 69OSSEOTITE Implants

Bone Profiling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70OSSEOTITE Implants

Surgical Indexing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71–72OSSEOTITE Implants

Page 4: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

1

These instructions were designed to serve as a reference

guide for the dental practitioner to utilize 3i Implants and

surgical instruments to their maximum potential. 3i’s

Implant System was developed to meet the diverse needs

of the patient and to offer the practitioner a choice of

surgical techniques customized to meet each patient’s

individual requirements.

3i’s Unique Designs enable the practitioner to place implants

in edentulous or partially edentulous mandibles or maxillae in

order to serve as support abutments for fixed and removable

bridgework or single tooth crowns and to provide the

stabilization needed for securing overdentures. 3i’s System

uses proven surgical procedures to properly secure the

implant in the osseous tissue, thus achieving the

physiological phenomenon referred to as osseointegration.

General Information:

These instructions will instruct practitioners in the use

of 3i’s Implant Systems. The success of any dental implant

system depends upon proper use of the components and

instrumentation. This manual is not intended for use as a

substitute for professional training and experience.

Treatment Planning:

Patient Evaluation And Selection

Several important factors must be considered when

evaluating a patient prior to implant surgery. The presurgical

evaluation must include a cautious and detailed assessment

of the patient’s general health, current medical status,

medical history, oral hygiene, motivation and expectations.

Factors such as heavy tobacco use, chewing patterns and

alcohol consumption should also be considered. In addition,

the clinician should determine if the case presents an

acceptable anatomical basis conducive to implant placement.

An extensive intraoral examination should be undertaken to

evaluate the oral cavity for any potential bone or soft-tissue

pathology. The examiner should also determine the perio-

dontal status of the remaining teeth, the health of the soft

tissue, or the presence of occlusal abnormalities such as

bruxism or crossbite. The presence of other conditions

that could adversely affect any existing natural dentition

or healthy tissue surrounding the implant should also

be evaluated.

Diseases of the mucous membrane and connective tissues,

pathologic bone disease and severe malocclusion could

affect the determination of whether the patient is a suitable

implant candidate.

The use of anticoagulants and the existence of metabolic

diseases, such as diabetes, allergies, chronic renal or cardiac

disease and blood dyscrasia could significantly influence the

patient’s ability to successfully undergo implant procedures.

If the patient’s medical history reveals an existing condition

or signals a potential problem that may compromise treat-

ment and/or the patient’s well-being, consultation with a

physician is recommended.

Introduction AndTreatment Planning

Page 5: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

22

Preoperative Planning

Radiographic Marking Balls (RMB30)

The vertical height of the bone is best determined

radiographically. Accurate measurement of the vertical

dimension on the radiograph facilitates the selection of the

appropriate implant length. This helps to avoid implant

placement into the maxillary sinus, the floor of the nose or the

mandibular canal and prevents perforation of the inferior aspect

of the mandible. Measurements can be made directly on the

panoramic radiograph using a millimeter ruler. Corrections

should be made for the degree of enlargement produced by the

particular radiographic equipment.

Radiographic marking balls of a known dimension can be

embedded in a plastic template prior to radiographic examination.

Once the radiograph is taken and the metal marking balls are

visible on the image, measurements can be taken to determine

the amount of bone available for implant placement.

To calculate the distortion factor, a simple formula can be

utilized: (5 ÷ A) x B = amount of actual bone available.

Formula Key =

• Radiographic marking ball = 5mm in diameter.

• A = Size of marking ball image on radiograph.

• B = Length in millimeters on the radiograph of available

bone between the crest of the ridge and the inferior

alveolar nerve canal.Example:

A = 6.5mmB = 14mmTherefore: (5 ÷ 6.5) x 14 = 10.76mm actual bone available

NOTE: A 2mm margin of safety, from the apical end ofthe implant to the adjacent vital structure, shouldbe considered.

Preoperative Planning:

Proper treatment planning, as well as the selection of the

proper implant length and diameter, are crucial to the long-

term success of the implant and restoration. Before an

implant can be selected, the anatomical foundation available

to receive the implant must be carefully assessed. Several

steps should be taken to complete the evaluation:

1. Clinical examination of the oral cavity can provide

important information about the health of the soft

tissue at the proposed implant site. Tissue tone and the

state of the superficial tissues should be evaluated. In

addition, the patient should demonstrate an adequate

dimension of attached mucosa or keratinized tissue at

the site selected for implantation. In partially edentulous

cases, the periodontal status of the remaining dentition

should be assessed and interaction between the implant

restoration and the adjacent natural dentition should

be considered.

2. The bony foundation and ridge need to be clinically

analyzed to ensure the presence of proper dimensions

and the amount of bone for implant placement. At least

one millimeter of bone should be present at the buccal and

lingual aspects of the implant following placement. During

the planning state, it is useful to measure the existing

bone foundation.

Inferior AlveolarNerve Canal

Marking Ball Image(6.5mm on this radiograph)

A

B

Page 6: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

7.5

3.54

5.5 5

8 9 5 5.55

8 58 5

3

In its simplest form, top-down treatment planning refers

to a protocol whereby the desired restorative result is

considered first, leading to consideration of the ideal

prosthetic platform and subsequent implant selection

based on bony anatomy.

A top-down treatment planning methodology will provide

maximum biomechanical stability and allow for soft tissue

flaring by utilizing an implant with a prosthetic platform

slightly smaller in diameter than the emergence diameter of

the tooth being replaced. 3i’s wide selection of implants

allows clinicians to match the size of the prosthetic platform

to the restoration it will eventually support, while allowing

for different bone volumes and anatomical features at the

implant site. Implant and healing abutment selections are

based upon the relationship of several key measurements:

• The emerging dimension of the crown in relation to

the diameter of the prosthetic platform of the implant

• The height and diameter of the intended restoration

at the tissue exit point

• The bone volume at the implant site in relation to the

diameter of the implant body

The Emergence Profile (EP®) Healing Abutment System

consists of healing abutments of various diameters and

heights for shaping the soft tissue to replicate the geometry

and gingival contours of natural dentition.

Top-Down Treatment Planning

5/6mm 6mm 4mm 4mm 5mm

5/6mm 6mm 4mm 5mm 4/5mm 4/3/4mm

CrownDiameter

ImplantBody

Diameter

3/4/3mm 5/4/5mm

3.25mm

Implants depicted arerepresentative of thebreadth of 3i Implantproducts.

Page 7: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

Clinical Considerations

True bone contours can only be evaluated after tissue flaps

have been reflected at the time of surgery. Even if bone

dimensions are painstakingly measured prior to surgery, the

doctor and patient must accept the possibility that inadequate

bone anatomy might be discovered during surgery and

preclude implant placement.

During the presurgical planning phase, it is important to

determine the vertical dimension - the actual space available

between the alveolar crest and the opposing dentition - to

confirm that the available space will accommodate the

proposed abutment and the final crown restoration. The

height required by the abutment may vary with the type of

abutment; therefore, the surgeon and restorative dentist

should carefully evaluate the abutment size. The final

prosthesis should be designed prior to the placement of

the implant.

Study models should be used preoperatively to evaluate the

residual ridge and to determine the position and angulation

of all implants. These models allow the clinician to evaluate

the opposing dentition and its effect on the implant position.

A surgical guide stent, which is critical for determining the

precise position and angulation of the implant, can be

constructed on the study model.

To prevent damage to the bone tissue and to prevent

compromising osseointegration, abundant and continuous

irrigation with a cool, sterile, irrigating solution is mandatory

during all drilling procedures. The application of excessive

pressure during preparation of the bone site must be avoided.

Bone surgery utilizes a high-torque electric drilling unit

that can be operated in forward and reverse modes at speeds

ranging from 0 to 1500rpm, depending on the surgical

requirements. Sharp instruments of the highest quality

should be utilized during implant site preparation to reduce

possible overheating and trauma to the bone. Minimizing

trauma enhances the potential for successful osseointegration.

The time elapsed between surgical placement of the implant

and final abutment placement is referred to as the healing

period. Healing periods can vary or be modified, depending

on the quality of the bone at the implantation site, bony

response to the implant surface and other implanted materials

and the surgeon’s assessment of the patient’s bone density

at the time of the surgical procedure. Extreme care must be

taken to avoid excessive force being applied to the implant

during the healing period.

44

Surgical Precautions

Page 8: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5

Single use drills/burs are supplied sterile and should be

properly disposed of after each procedure. Reusable

drills/burs and instrumentation are supplied nonsterile and

must be sterilized prior to use. Nonsterile items must be

removed from the packaging before sterilization.

Multiple sterilizations may affect the flow of fluid through

internally irrigated drills. The drills should be checked

following each sterilization cycle to determine if fluid flows

through the irrigation ports. Although the surgical drills are

constructed of stainless steel, these should be adequately

dried prior to packaging for sterilization and again after the

sterilization cycle.

To extend the useful life of 3i’s instruments, certain proce-

dures should always be followed:

Cleaning:

1. After use, place drills into a beaker of plain water,

mild soap or specialized cleaning solution.

2. Rinse with tap water for a minimum of two minutes

while brushing with a soft bristled brush to remove

visible debris. Clean the interior lumen with a thin wire

to remove any remaining debris.

3. Place instruments in an ultrasonic bath containing

enzymatic detergent for five minutes.* Scrub the

instruments again with a soft bristled brush and ream

interior lumen to remove any remaining debris.

4. Rinse and flush the instruments for one minute

using tap water.

5. Inspect visually for any remaining bone fragments or

debris and scrub as necessary.

Sterilization:

6. Remove the bur block from the surgical tray. Scrub the

surgical tray and block with a soft bristled brush and

mild soap. Rinse thoroughly.

7. Place the components into the surgical tray and pour

ethyl alcohol (do not use rubbing alcohol) over the burs

and tray to remove soap residue and minerals from the

water. This step is important to help prevent corrosion

and spotting. Let the components dry before wrapping.

8. Wrap the surgical tray in paper or autoclave-approved

bags twice to prevent a tear of the outer packaging from

contaminated instruments.

9. Steam gravity sterilize for forty minutes at a temperature

of 270–̊275˚F (132–̊135˚C).

Notes:

1. Multiple sterilizations may affect the flow of fluid

through internally irrigated burs. After each use, ream

the burs individually with wire to remove any bone

fragments or debris that will prevent the flow of water.

This is done prior to the sterilization cycle.

2. Do not remove drills, instrumentation or surgical tray

from the autoclave until the “dry cycle” is complete.

Very Important!

3. These guidelines DO NOT apply to the cleaning and

sterilization of your powered instrumentation.

Please follow your powered instrumentation

manufacturer’s instructions.

Please refer to ART630 for complete instructions on the

sterilization and care of stainless steel.

*ENZOL enzymatic detergent was used to validate this process, per the manufacturer’sdilution recommendation.

Cleaning And Sterilization

Page 9: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

66

3i Depth Marking SystemParallel Walled Implants

The 3i Depth Marks measurement system provides a markon the drill that corresponds to the placement of the implantvia a well-established procedure. 3i’s Original Protocolfollows the principles of protecting the implant frompremature loading by placing the implant subcrestally.

NOTE: A drill extension for areas of limited access is available.

Drill DiameterITD/DTN/DT

Drill Tip LengthACT™

Drill Tip Length2.00mm 0.6mm 0.6mm2.30mm 0.7mm N/A2.75mm 0.8mm 0.9mm3.00mm 0.9mm 0.9mm3.15mm 1.0mm 1.0mm3.25mm 1.0mm 1.0mm3.85mm N/A 1.2mm4.25mm 0.4mm 1.3mm4.85mm N/A 1.3mm5.25mm 0.5mm 1.2mm

The length of the drill tip is not included in the depthmark measurement. The drill tip length should beconsidered when preparing the osteotomy.

The length of the drill tip varies with the diameter ofthe drill.

The drill depth marks do not indicate implant lengths.Rather, the marks represent the length of the implantwith the cover screw in place. As a result, to place animplant and cover screw subcrestally requires drillingto the middle of the single line depth mark or thebeginning or end of the broad band depth mark onACT Drills. For crestal placement, drill halfway beforethe corresponding depth mark for the implant length.For supracrestal placement, the drill depth markshould remain above the bone by 1mm for the coverscrew plus the implant collar height. Refer to thediagram on the bottom of page 10 for moreinformation.

Traditional Subcrestal Protocol

Implant Anatomy

ImplantBody

Collar Height 1.25mm (Collar H) .75mm (Std. Collar H)

ImplantBody

Collar Height

External Hex

3i OSSEOTITE®

Certain®

3i OSSEOTITEExternal Hex

Drilling Depth

The drilling depth with the Twist Drillwill vary depending on the type ofplacement related to the bone crest.

The depth marks are specific forsubcrestal implant placement only.There are no specific depth marks onthe drills for crestal or supracrestalplacement. ACT3015

Certain® External Hex

ACT3015

2.0mmTwist Drill

Depth GaugeImplant WithCover Screw

Implant Platform Implant Platform

1-Subcrestal2-Crestal

3-Supracrestal

11.5mmImplant

Drill Tip max 1.3mm

11.5mmImplant

Drill Tip max 1.3mm

1-Subcrestal2-Crestal 3-Supracrestal

Page 10: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

7

Drill DiameterITD/DTN/DT

Drill Tip LengthACT

Drill Tip Length

2.00mm 0.6mm 0.6mm

2.30mm 0.7mm N/A

2.75mm 0.8mm 0.9mm

3.00mm 0.9mm 0.9mm

3.15mm 1.0mm 1.0mm

3.25mm 1.0mm 1.0mm

3.85mm N/A 1.2mm

4.25mm 0.4mm 1.3mm

4.85mm N/A 1.3mm

5.25mm 0.5mm 1.2mm

ITD Reusable Drills

• Internal Irrigation Lumen• All Thin Lines

Types Of Twist Drills

DTN Disposable Drills

• Without Internal Irrigation Lumen• Bands

Drill Marks

8.5mm

7mm

Drill Tip

10mm

11.5mm

13mm

15mm

max 1.3mm

The length of the drill tip is notincluded in the depth markmeasurement. The drill tiplength should be consideredwhen preparing the osteotomy.

The length of the drill tip varieswith the diameter of the drill.

The center of the drill’s singleline depth marks and thebeginning or end of the broadband indicates subcrestalplacement for the correspon-ding length implant. Tip Dimensions

PREVAIL™, Certain®, External Hex

Drill Marks And ImplantsThe standard cover screwsare 1mm in height for bothCertain and External HexImplants.

1mm

IIOS4511 IOSS411 OSS411 OS4511

Certain

Regardless of the implant type, thelength of the implant is the same fromthe top of the implant collar (platform),to the tip of the implant.

IIOS4511 IOSS411 OSS411 OS4511

Certain Ex Hex

The depth marks on the drill representimplant placement where the top of the

cover screw (when height of thecover screw is 1mm) is flush withthe crest of the bone (subcrestal).

ACT™ Reusable Drills

• No Hub• Without Internal Irrigation• Alternating Lines And Bands

Ex Hex

3i Depth Marking SystemParallel Walled Implants

Page 11: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

88

Labeled vs. Actual Lengths

9.1mm

7.6mm

10.6mm

12.1mm

13.6mm

15.6mm

LabeledLengths

ACT3015

The center of the drill’s singleline depth marks and thebeginning or end of the broadband indicate the length ofthe implant with a standardcover screw in place.

Drill Tipmax 1.3mm

1mm

The actual implant lengths from the top of the implantcollar (platform) to the tip of the implant are shorterby 0.4mm than the labeled length.

The landmarks (grooves) on the CertainImplant Driver Tip and the Ex Hex ImplantMount act as references during implantplacement.

Ex Hex Certain Ex Hex

Certain Ex Hex

Actual Implant LengthsWith Cover Screw ON

15mm

13mm

11.5mm

10mm

8.5mm

7mm

LabeledLengths

Actual Implant LengthsWith Cover Screw OFF

15mm

13mm

10mm

8.5mm

7mm

Certain®

*From point on drill at which maximum diameter starts. (Drill depth mark is .5mm wide) Drill length listed in chart does not include drill tip.

NOTE: A 2mm margin of safety from the apical end of the implant to the adjacent vital structure should be considered.

OSSEOTITE NT®

Implant Length(Label)

Actual OSSEOTITE NT

Length

Parallel WalledImplant Length

(Label)

Actual ParallelWalled Implant

LengthCover Screw

Height Actual Drill LengthTo Subcrestal Mark

7.0mm 6.6mm 1.0mm 7.6mm8.5mm 8.1mm 8.5mm 8.1mm 1.0mm 9.1mm

10.0mm 9.6mm 10.0mm 9.6mm 1.0mm 10.6mm11.5mm 11.1mm 11.5mm 11.1mm 1.0mm 12.1mm13.0mm 12.6mm 13.0mm 12.6mm 1.0mm 13.6mm15.0mm 14.6mm 15.0mm 14.6mm 1.0mm 15.6mm

18.0mm 17.6mm 1.0mm 18.6mm20.0mm 19.6mm 1.0mm 20.6mm

12.6mm

14.6mm

8.1mm

6.6mm

9.6mm

11.1mm11.5mm

Page 12: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

9

11.5mm implants

Subcrestal

Drill Tip max 1.3mm

8.5mm7mm

10mm

Crestal Placement

Subcrestal Placement

• The implant platform will be 1mm (or more) below the bone crest.• Mostly used in the anterior region for aesthetics.

Drill Tip max 1.3mm

8.5mm7mm

1mm

11.5mm implants

Subcrestal

10mm

• The implant platform will be at the bone crest.

11.5mm implants

Crestal

Drill Tip max 1.3mm

8.5mm7mm

11.5mm

Drill Tip max 1.3mm

8.5mm7mm

11.5mm

11.5mm implants

Crestal

10mm1mm

1mm

Certain® External Hex

Certain External Hex

3i Depth Marking SystemParallel Walled Implants

Bone Crest Bone Crest

10mm

11.5mm11.5mm

Bone Crest Bone Crest

For Subcrestal Certain and External Hex implant placement,Drill to the drill depth mark that corresponds to the labeledimplant length.

A Countersink Drill is used when placing a 4mm diameterimplant subcrestally to prepare the bone to accept the coverscrew. Remember, for a 4mm diameter implant, the implantplatform is 4.1mm in diameter and the cover screw flares to4.5mm in diameter.

1mm

A Countersink Drill is generally not needed becausethe cover screw for the 4mm diameter implant willstay above the bone crest. It may be needed in hardbone due to the shape of the implant collar.

For Crestal Certain and External Hex implant placement, stopdrilling 1mm before the drill depth mark that corresponds tothe labeled implant length (1mm equals cover screw height).

Page 13: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

Subcrestal Crestal

11.5mm Certain ImplantsDrill Tip max 1.3mm

1010

11.5mm implants

Supracrestal

Drill Tip max 1.3mm

8.5mm7mm

10mm

Supracrestal Placement

11.5mm

8.5mm7mm

11.5mm10mm 0.75mm collar H

Drill Tip max 1.3mm

8.5mm7mm

11.5mm

11.5mm implants

Supracrestal

10mm

Drill Tip max 1.3mm

For Supracrestal Certain implant placement, stop drilling2.25mm before the drill depth mark that corresponds to thelabeled implant length (2.25mm equals 1mm cover screwheight plus 1.25mm Certain Implant collar height).

For Supracrestal External Hex implant placement, stop drilling1.75mm before the drill depth mark that corresponds to thelabeled implant length (1.75mm equals 1mm cover screw heightplus 0.75mm External Hex Standard Implant collar height).

• The height of the implant platform above the bonevaries with the specific implant type being placed.

Implant Type Collar Height

MicroMiniplant™ Implants 1.50mm

Certain Implants 1.25mm

OSSEOTITE NT® Implants 1.25mm

External Hex Standard Implants 0.75mm3.75 & 4.0mm diameter

External Hex Wide Implants 0.50mm5.0 & 6.0mm diameter

A Countersink Drill is not needed.

• The implant collar will be above the bone crest.• The height of the implant platform above the bone crest varies with the implant type being placed–see diagram below.

Implant Collar Height

Certain® External Hex

Placement Comparison Diagram

1mm cover screw H

Bone Crest

Supracrestal

Bone Crest Bone Crest1mm cover screw H

1.25mm collar H

Page 14: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

11

Important Information Concerning Countersink Drills CD500 And CD600:A second depth mark has been added to the 5mm and 6mmdiameter Countersink Drills (CD500 & CD600). The bottomline (closest to the apex) is positioned where the original

single line has traditionally been. The top line (closest to theshank) has been added to accommodate the OSSEOTITE®

Certain® Implant.

Crest of Bone

3i ExternalHex Implant

Subcrestal Implant Placement

Countersink Drill MarkingSystem

3i OSSEOTITECertain Implant

3i ExternalHex Implant

Crestal Implant Placement

3i OSSEOTITECertain Implant

CrestalPlacement

SubcrestalPlacement

3i ExternalHex Implant

CD500 orCD600

CD500 orCD600

Important Information Concerning Countersink DrillsDC100, CD100, CD4500 And CD5600:

DC100 orCD100

CD4500 CD5600

Page 15: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

1212

Pick-Up And Delivery Of Implant Care must be taken when inserting the Implant PlacementDriver tip into the implant. A very low RPM must be used asyou approach the internal connection of the implant with thedriver tip to properly align the internal hex of the implant with the external hex of the driver. Press down firmly to engage theimplant securely.

NOTE: The Certain MicroMiniplant™ (3.25mm dia.) requires theuse of a dedicated MicroMiniplant Driver Tip (IMPDTS orIMPDTL) that is marked with a purple band on the shank. Theinternal hex configuration of the MicroMiniplant is smaller thanthe standard Certain Internal Hex Connection (4, 5 and 6mmdiameter implants). The item numbers can be identified on theside of the driver tip.

Pick-Up And Delivery Of Cover Screw Or Healing AbutmentThe 0.048 inch tip of the Certain Implant Placement Driver can be used to pick up and place the cover screw or thehealing abutment.

NOTE: When using the Internal Connection Implant Driver(IIPDTS or IIPDTL), reduce the torque setting on the drillingunit to 10Ncm.

The cover screw replica portion of the driver allows for visualverification of the cover screw position, making subcrestaland crestal placement of the implant predictable.

NOTE: Periodic O-Ring replacement is required for the CertainDriver Tips.

Implant pick-up

Implant and driver hex design

Hex

Cover screw pick-up

Subcrestal PlacementCrestal Placement

Mountless Delivery GuidelinesOSSEOTITE® Certain® Implant

Page 16: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

13

OSSEOTITE® Parallel Walled ImplantSurgical Tray

Close-up view of Surgical Tray illustrating numbering sequence.

Coordinating The Use Of The Surgical Tray With The Surgical Manual Illustrations:

The OSSEOTITE Parallel Walled Implant SurgicalTray is numbered to indicate the appropriate stepsof the implant placement protocol. The followingillustrated implant placement protocol uses thesame numbering sequence.

Close-up view of Surgical Tray illustrating numbering sequence.

Page 17: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

1414

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

C’sink DrillCD4500

RoundDrill

RD100

4/5/4mm ImplantIIOS4511

4/5mm C’sinkDepth Indicator

CDI4500

Cover ScrewICSF41

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

2.75 or 3.0mmTwist Drill

C’sink DrillICD100

RoundDrill

RD100

3/4/3mm ImplantIIOS3411

4mm C’sinkDepth Indicator

ICDI100Cover Screw

IMCSF34

Subcrestal Implant PlacementProtocol - Quick Reference

Certain® PREVAIL™ 3/4/3mm Implant

Certain PREVAIL 4/5/4mm Implant

See page 21 for detailed instructions.

See page 23 for detailed instructions.

Page 18: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

15

Subcrestal Implant PlacementProtocol - Quick Reference

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

C’sink DrillCD500

RoundDrill

RD100

5/6/5mm ImplantIIOS5611

4.25mmTwist Drill

5/6mm C’sinkCD5600

Certain® PREVAIL™ 5/6/5mm Implant

See page 25 for detailed instructions.

5/6mm C’sinkDepth Indicator

CDI5600Cover Screw

ICSF50

Page 19: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

1616

OSSEOTITE® Certain® And OSSEOTITE External Hex MicroMiniplant™ 3.25mm Implant

OSSEOTITE External Hex Standard 3.75mm And OSSEOTITE XP® 3/4mm Implant

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

2.75 or 3.0mmTwist Drill

C’sink DrillCD100

RoundDrill

RD100

Cover ScrewCS375

3.75mm ImplantOSS311

RoundDrill

RD100

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

2.75 or 3.0mmTwist Drill

Cover ScrewIMMCS1

Cover ScrewMMCS1

3.25mm ImplantIOSM311

3.25mm ImplantOSM311

See page 30 for detailed instructions.

See page 32 for detailed instructions.

Page 20: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

17

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

C’sink DrillICD100

RoundDrill

RD100

Cover ScrewICS375

4.0mm ImplantIOSS411

Certain® PREVAIL™ 4/3mm And OSSEOTITE® Certain 4.0mm Implant

OSSEOTITE External Hex 4.0mm Implant

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

C’sink DrillCD100

RoundDrill

RD100

Cover ScrewCS375

4.0mm ImplantOSS411

Cover ScrewIMCSF34

4/3mm ImplantIIOS4311

See page 34 for detailed instructions.

See page 34 for detailed instructions.

Subcrestal Implant PlacementProtocol - Quick Reference

Page 21: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

5mm Pilot/C’sink DrillCD500

RoundDrill

RD100

Cover ScrewCS500

5.0mm ImplantOSS511

4.25mmTwist Drill

1818

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

5mm Pilot/C’sink DrillCD500

RoundDrill

RD100

Cover ScrewICSF41

5/4mm ImplantIIOS5411

5.0mm ImplantIOSS511

4.25mmTwist Drill

Certain® PREVAIL™ 5/4mm And OSSEOTITE® Certain 5.0mm Implant

OSSEOTITE External Hex 5.0mm Implant

Cover ScrewICS500

3.0 or 3.25mmTwist Drill

See page 36 for detailed instructions.

See page 36 for detailed instructions.

Page 22: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

19

OSSEOTITE® Certain® 6.0mm Implant

OSSEOTITE External Hex 6.0mm Implant

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

5mm Pilot/C’sink DrillCD500

6mm Pilot/C’sink DrillCD600

RoundDrill

RD100

6.0mm ImplantIOSS611

4.25mmTwist Drill

5.25mmTwist Drill Cover Screw

ICS600

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

5mm Pilot/C’sink DrillCD500

6mm Pilot/C’sink DrillCD600

RoundDrill

RD100

6.0mm ImplantOSS611

4.25mmTwist Drill

5.25mmTwist Drill

Cover ScrewCS600

See page 38 for detailed instructions.

See page 38 for detailed instructions.

Subcrestal Implant PlacementProtocol - Quick Reference

Page 23: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

4.25mmTwist Drill

5mm Pilot/C’sink DrillCD500

C’sink DrillCD5600

RoundDrill

RD100

Cover ScrewCS600

3.0 or 3.25mmTwist Drill

XP 5/6mm ImplantOS5611

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.0 or 3.25mmTwist Drill

5mm Pilot/C’sink DrillCD4500

RoundDrill

RD100

XP 4/5mm ImplantOS4511

Cover ScrewCS500

2020

OSSEOTITE XP® 4/5mm External Hex Implant

OSSEOTITE XP 5/6mm External Hex Implant

See page 23 for detailed instructions.

See page 25 for detailed instructions.

Page 24: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

21

For a quick reference guide to implant placement,refer to page 14 for OSSEOTITE® Certain PREVAIL3/4/3mm Implant.

1. Once the implant site has been determined, mark thesite with a Round Drill and penetrate the cortical bone.The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill to approximately7mm, then verify the direction with the thin portion ofthe Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of the preparationby inserting the thin portion of the Direction Indicatorinto the osteotomy. Thread dental floss through hole toprevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 3/4/3mmDiameter Implant

1.5

2.4

4.1

3.4

Page 25: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2222

4. Use the Pilot Drill to shape the coronal aspect of theimplant site. Drill to the depth mark. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using the DirectionIndicator, proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in normal densebone. Proceed with the 2.75mm Twist Drill to the desireddepth for implant placement in soft Type IV Bone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.0mm Twist Drill for normal dense bone(ITD315, DT315, DTN315 or ACT3015)2.75mm Twist Drill for soft Type IV Bone (ITD2715, DT2715, DTN2715 or ACT2715)

6. Use the Certain® Countersink Drill to shape thecoronal aspect of the osteotomy. Drill to the top of thedepth mark for subcrestal placement. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:Certain Countersink Drill (ICD100)

NOTE: Drilling to the bottom of the depth mark on theICD100 will prepare the osteotomy for crestal placementand not subcrestal placement of implant.

7. Verify the direction and position of the preparationby inserting the Countersink Depth Indicator into theosteotomy.

• Instruments needed:Countersink Depth Indicator (ICDI100)

Proceed to step 10a on page 28 for Certain PREVAIL™Implant Placement.

Page 26: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

23

For a quick reference guide to implant placement, refer topage 14 for OSSEOTITE® Certain PREVAIL 4/5/4mm Implantand to page 20 for OSSEOTITE XP 4/5mm Implant.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill to approximately7mm, then verify the direction with the thin portion ofthe Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of the preparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 4/5/4mm AndOSSEOTITE XP® 4/5mm Diameter

External Hex Implant

1.5 1.5

2.6

4.8

4.1

2.6

5.0

2.7

Page 27: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2424

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. Therecommended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.25mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm Twist Drill for normal dense bone(ITD3215, ACT3215, DT3215 or DTN3215)3.0mm Twist Drill for soft Type IV Bone (ITD315, ACT3015, DT315 or DTN315)

6. Use the flared cutting edge on the CountersinkDrill to shape the coronal aspect of the osteotomy. Drillto the top of the depth mark for subcrestal placement.The recommended drill speed is 1200–1500rpm.

• Instruments needed:4/5 Countersink Drill (CD4500)

NOTE: Drilling to the bottom of the depth mark on theCD4500 will prepare the osteotomy for crestal placementand not subcrestal placement of implant.

7. Verify the direction and position of thepreparation by inserting the CDI4500 4/5 CountersinkDepth Indicator into the osteotomy.

• Instruments needed:4/5 Countersink Depth Indicator (CDI4500)

Proceed to step 10a on page 28 for Certain® PREVAIL™and OSSEOTITE XP® External Hex Implant Placement.

Page 28: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

25

For a quick reference guide to implant placement, refer topage 15 for OSSEOTITE® Certain PREVAIL 5/6/5mm Implantand to page 20 for OSSEOTITE XP 5/6mm Implant.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill to approximately7mm, then verify the direction with the thin portion ofthe Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

1.5

3.1

5.8

5.0Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 5/6/5mm AndOSSEOTITE XP® 5/6mm Diameter

External Hex Implant

1.5

3.1

6.02.7

Page 29: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2626

4. Use the Pilot Drill to shape the coronal aspectof the implant site. Drill to the depth mark. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.25mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm Twist Drill for normal dense bone(ITD3215, ACT3215, DT3215 or DTN3215)3.0mm Twist Drill for soft Type IV Bone (ITD315, ACT3015, DT315 or DTN315)

6. Use the 5mm Countersink Drill to shape thecoronal aspect of the osteotomy. Drill to the center ofthe top depth mark for subcrestal placement.The recommended drill speed is 1200–1500rpm.

• Instruments needed:5/6/5 Countersink Drill (CD500)

NOTE: Drilling to the center of the bottom depth mark on the CD500 will prepare the osteotomy for crestalplacement and not subcrestal placement of implant.

7. Proceed with the 4.25mm Twist Drill to the desired depth for implant placement in normal densebone. Proceed with the 3.85mm Twist Drill for implantplacement in soft Type IV Bone. The recommended drillspeed is 1200–1500rpm.

• Instruments needed:4.25mm Twist Drill for normal dense bone(ITD423, DTN423, DT423 or ACT4213)3.85mm Twist Drill for soft Type IV Bone (ACT3815)

Page 30: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

27

8. Use the flared cutting edge on the 5/6/5Countersink Drill to shape the coronal aspect of theosteotomy. Drill to the top of the depth mark forsubcrestal placement. The recommended drill speed is900rpm.

• Instruments needed:5/6/5 Countersink Drill (CD5600)

NOTE: Drilling to the bottom of the depth mark on theCD5600 will prepare the osteotomy for crestal placementand not subcrestal placement of implant.

9. Verify the direction and position of the preparationby inserting the 5/6 Countersink Depth Indicator into theosteotomy.

• Instruments needed:5/6 Countersink Depth Indicator (CDI5600)

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 5/6/5mm AndOSSEOTITE XP® 5/6mm Diameter

External Hex Implant

Page 31: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

2828

No-Touch™ Delivery System

10a. Remove contents from the implant box.

10b. A nonsterile assistant should peel back the traylid and drop the No-Touch Implant Tray onto the steriledrape.

10c. Place the No-Touch Implant Tray into the appropriate location on the surgical tray.

10d. Peel back the tray lid to expose the implant andcover screw

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ And OSSEOTITE XP®

External Hex Implant Placement

Page 32: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

29

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ And OSSEOTITE XP®

External Hex Implant Placement11. Pick up the implant from the surgical tray using the

Certain Implant Placement Driver Tip (IIPDTS or IIPDTL).NOTE: The Certain MicroMiniplant™ (3.25mm dia.) requiresthe use of a dedicated MicroMiniplant Driver Tip (IMPDTSor IMPDTL) that is marked with a purple band on the shank.The Internal Hex configuration of the MicroMiniplant issmaller than the standard Certain Internal Hex Connection(4, 5 and 6mm dia. implants). The item numbers can beidentified on the side of the driver tip.

orPick up the implant from the surgical tray using

the Handpiece Connector (MDR10). Carry the implant to the mouth facing upward to

prevent accidental dislodging.

12. Place the implant in the prepared site atapproximately 15–20rpm. The ICE™ Design will allowsmooth and precise implant placement without tappingin all but the most dense bone. The recommended drilltorque setting for placing an implant is 40–50Ncm.

13. To remove the Certain Implant Placement DriverTip from the implant, lift straight up and out.

orTo remove the implant mount, place the Open-

End Wrench onto the mount. Loosen the screw at thetop of the mount with a Large Hex Driver or the LargeHex Driver Tip inserted into the Right-Angle Driver androtate counter-clockwise. After the screw is loosened,rotate the Open-End Wrench counter-clockwise slightlybefore removing the mount. The mount may be carriedfrom the mouth with the Open-End Wrench.

• Instruments needed:Open-End Wrench (CW100), Large Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with CADD1) or a Large Hex Driver (PHD02N)

14. Pick up the cover screw from the surgical trayusing the Certain Implant Placement Driver Tip (IIPDTSor IIPDTL) or Large Hex Driver (PHD02N) and placeonto the implant.NOTE: When using the Certain Implant Placement Driver,reduce the torque setting on the drilling unit to 10Ncm.

or

Pick up the cover screw from the surgical trayusing the Small Hex Driver (PHD00N) and place onto theimplant. The Small Hex Driver may be used for allExternal Hex Cover Screws. Final hand-tightening of thecover screw should be done with the Small Hex Driver.NOTE: At this step, a temporary healing abutment may beplaced for single-stage surgery instead of a cover screw.

15. Close tissue and suture.

Page 33: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3030

Refer to page 16 for a quick reference guide to OSSEOTITECertain and OSSEOTITE External Hex MicroMiniplant ImplantPlacement Protocols.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

OSSEOTITE® Certain® And OSSEOTITEExternal Hex MicroMiniplant™

3.25mm Diameter Implant

1.5

3.4

2.4

1.5

2.4

.7

3.42.5

Page 34: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

31

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.0mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 2.75mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.0mm Twist Drill for normal dense bone(ITD315, DT315, DTN315 or ACT3015)2.75mm Twist Drill for soft Type IV Bone (ITD2715, DT2715, DTN2715 or ACT2715)

Proceed to step 10a on page 41 for OSSEOTITE Certainand OSSEOTITE External Hex Implant Placement.

Subcrestal Implant PlacementProtocol

OSSEOTITE® Certain® And OSSEOTITEExternal Hex MicroMiniplant™

3.25mm Diameter Implant

Page 35: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3232

For a quick reference guide to implant placement, refer topage 16 for OSSEOTITE Standard 3.75mm Implant andOSSEOTITE XP 3/4mm Implant.

1. Once the implant site has been determined, mark thesite with a Round Drill and penetrate the cortical bone.The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill to approximately7mm, then verify the direction with the thin portion ofthe Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of the preparationby inserting the thin portion of the Direction Indicatorinto the osteotomy. Thread dental floss through hole toprevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

OSSEOTITE® Standard 3.75mm AndOSSEOTITE XP® 3/4mm Diameter

External Hex Implant2.3

4.1

.75

2.7.7

Page 36: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

33

4. Use the Pilot Drill to shape the coronal aspect of theimplant site. Drill to the depth mark. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using the DirectionIndicator, proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in normal densebone. Proceed with the 2.75mm Twist Drill to the desireddepth for implant placement in soft Type IV Bone.The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.0mm Twist Drill for normal dense bone(ITD315, DT315, DTN315 or ACT3015)2.75mm Twist Drill for soft Type IV Bone (ITD2715, DT2715, DTN2715 or ACT2715)

6. Using the Countersink Drill, prepare the bone toaccept the 4.5mm flared cover screw of the 3.75mmdiameter implant for subcrestal placement. Drill to thecenter of the depth mark for subcrestal placement. Therecommended drill speed is 1200–1500rpm.

• Instruments needed:Countersink Drill (CD100)

Proceed to step 10a on page 41 for OSSEOTITECertain® and OSSEOTITE XP External Hex ImplantPlacement.

Subcrestal Implant PlacementProtocol

OSSEOTITE® Standard 3.75mm AndOSSEOTITE XP® 3/4mm Diameter

External Hex Implant

Page 37: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3434

Refer to page 17 for a quick reference guide to CertainPREVAIL 4/3mm, OSSEOTITE Certain and OSSEOTITEExternal Hex 4.0mm Implant Placement Protocols.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the cortical bone.The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 4/3mm, OSSEOTITE®

Certain And OSSEOTITE External Hex4.0mm Diameter Implant

1.0

2.6

4.14.1

2.6 2.75

4.1

.75

2.7.7

3.4

Page 38: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

35

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.25mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm Twist Drill for normal dense bone(ITD3215, ACT3215, DT3215 or DTN3215)3.0mm Twist Drill for soft Type IV Bone (ITD315, ACT3015, DT315 or DTN315)

6. Using the Countersink Drill, prepare the bone toaccept the 4.5mm flared cover screw of the 4.0mmdiameter implant for subcrestal placement.The recommended drill speed is 1200–1500rpm.

• Instruments needed:

Countersink Drill (ICD100) - Certain (Drill to the top of the laser line for subcrestal placement)

Countersink Drill (CD100) - External Hex (Drill to the center of the laser line for subcrestalplacement)

Proceed to step 10a on page 41 for Certain PREVAIL,OSSEOTITE Certain and OSSEOTITE External HexImplant Placement.

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 4/3mm, OSSEOTITE®

Certain And OSSEOTITE External Hex4.0mm Diameter Implant

Page 39: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3636

Refer to page 18 for a quick reference guide to CertainPREVAIL 5/4mm, OSSEOTITE Certain and OSSEOTITEExternal Hex 5.0mm Implant Placement Protocols.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 5/4mm, OSSEOTITE®

Certain And OSSEOTITE External Hex5.0mm Diameter Implant

1.25

3.03.0

5.0

3.0

.5

5.0

2.7.75.0

4.1

Page 40: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

37

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.25mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm Twist Drill for normal dense bone(ITD3215, ACT3215, DT3215 or DTN3215)3.0mm Twist Drill for soft Type IV Bone (ITD315, ACT3015, DT315 or DTN315)

6. Use the 5mm Countersink/Pilot Drill to shape thecoronal aspect of the implant site. For subcrestalplacement, drill to the center of the top depth mark forOSSEOTITE Certain Implant placement or drill to thecenter of the bottom depth mark for an OSSEOTITEExternal Hex Implant placement. The recommended drillspeed is 900rpm.

• Instruments needed:5.0mm Countersink/Pilot Drill (CD500)

7. Once the coronal aspect of the osteotomy hasbeen prepared, proceed with the 4.25mm Twist Drill tothe desired depth for implant placement in normal densebone. Proceed with the 3.85mm Twist Drill to the desireddepth for implant placement in soft Type IV Bone. Therecommended drill speed is 1200–1500rpm.

• Instruments needed:4.25mm Twist Drill for normal dense bone(ITD423, ACT4213, DT423 or DTN423)3.85mm Twist Drill for soft Type IV Bone (ACT3815)

Proceed to step 10a on page 41 for Certain PREVAIL,OSSEOTITE Certain and OSSEOTITE External HexImplant Placement.

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™ 5/4mm, OSSEOTITE®

Certain And OSSEOTITE External Hex5.0mm Diameter Implant

Page 41: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3838

Refer to page 19 for a quick reference guide to OSSEOTITECertain and OSSEOTITE External Hex 6.0mm ImplantPlacement Protocols.

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol

OSSEOTITE® Certain®And OSSEOTITEExternal Hex 6.0mm Diameter Implant

1.25

4.0

6.0

4.0

.5

6.02.7

.7

Page 42: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

39

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Once proper alignment is verified using theDirection Indicator, proceed with the 3.25mm Twist Drillto the desired depth for implant placement in normaldense bone. Proceed with the 3.0mm Twist Drill to thedesired depth for implant placement in soft Type IVBone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm Twist Drill for normal dense bone(ITD3215, ACT3215, DT3215 or DTN3215)3.0mm Twist Drill for soft Type IV Bone (ITD315, ACT3015, DT315 or DTN315)

6. Advance the 5mm Countersink/Pilot Drill to thefirst cutting groove to widen the coronal aspect of theosteotomy, allowing the 4.25mm Twist Drill to enter theosteotomy. The recommended drill speed is 900rpm.

• Instrument needed:5.0mm Countersink/Pilot Drill (CD500)

7. Once the coronal aspect of the osteotomy hasbeen prepared, proceed with the 4.25mm Twist Drill tothe desired depth. The recommended drill speed is900rpm.

• Instruments needed:4.25mm Twist Drill (ITD423, ACT4213, DT423or DTN423)

Subcrestal Implant PlacementProtocol

OSSEOTITE® Certain®And OSSEOTITEExternal Hex 6.0mm Diameter Implant

Page 43: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

4040

8. Use the 6.0mm Countersink/Pilot Drill to shapethe coronal aspect of the implant site. For subcrestalplacement, drill to the center of the top depth mark forOSSEOTITE® Certain® Implant placement or drill to thecenter of the bottom depth mark for an OSSEOTITEExternal Hex Implant placement. The recommended drillspeed is 900rpm.

• Instruments needed:6.0mm Countersink/Pilot Drill (CD600)

9. Once the coronal aspect of the osteotomy hasbeen prepared, proceed with the 5.25mm Twist Drill tothe desired depth for implant placement in normal densebone. Proceed with the 4.85mm Twist Drill to the desireddepth for implant placement in soft Type IV Bone.The recommended drill speed is 900rpm.

• Instruments needed:5.25mm Twist Drill for normal dense bone(ITD523, ACT5213, DT523 or DTN523)4.85mm Twist Drill for soft Type IV Bone (ACT4813)

Proceed to step 10a on page 41 for OSSEOTITE Certainand OSSEOTITE External Hex Implant Placement.

Page 44: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

41

No-Touch™ Delivery System

10a. Remove contents from the implant box.

10b. A nonsterile assistant should peel back the traylid and drop the No-Touch Implant Tray onto the steriledrape.

10c. Place the No-Touch Implant Tray into the appropriate location on the surgical tray.

10d. Peel back the tray lid to expose the implant andcover screw.

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™, OSSEOTITE® CertainAndOSSEOTITE External Hex Implant Placement

Page 45: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

4242

11. Pick up the implant from the surgical tray usingthe Certain® Implant Placement Driver Tip (IIPDTS orIIPDTL).

NOTE: The Certain MicroMiniplant™ (3.25mm dia.)requires the use of a dedicated MicroMiniplant Driver Tip(IMPDTS or IMPDTL) that is marked with a purple band onthe shank. The Internal Hex configuration of theMicroMiniplant is smaller than the standard CertainInternal Hex Connection (4, 5 and 6mm dia. implants). Theitem numbers can be identified on the side of the driver tip.

or

Pick up the implant from the surgical tray usingthe Handpiece Connector (MDR10).

Carry the implant to the mouth facing upward toprevent accidental dislodging.

12. Place the implant in the prepared site atapproximately 15–20rpm. The ICE™ Design will allowsmooth and precise implant placement without tappingin all but the most dense bone. The recommended drilltorque setting for placing an implant is 40–50Ncm.

13. To remove the Certain Implant Placement DriverTip from the implant, lift straight up and out.

or

To remove the implant mount, place the Open-End Wrench onto the mount. Loosen the screw at thetop of the mount with a Large Hex Driver or the LargeHex Driver Tip inserted into the Right-Angle Driver androtate counter-clockwise. After the screw is loosened,rotate the Open-End Wrench counter-clockwise slightlybefore removing the mount. The mount may be carriedfrom the mouth with the Open-End Wrench.

• Instruments needed:Open-End Wrench (CW100), Large Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with CADD1) or a Large Hex Driver (PHD02N)

Page 46: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

43

14. Pick up the cover screw from the surgical trayusing the Certain Implant Placement Driver Tip (IIPDTSor IIPDTL) or Large Hex Driver (PHD02N) and place ontothe implant.

NOTE: When using the Certain Implant Placement Driver,reduce the torque setting on the drilling unit to 10Ncm.

or

Pick up the cover screw from the surgical trayusing the Small Hex Driver (PHD00N) and place onto theimplant. The Small Hex Driver may be used for allExternal Hex Cover Screws. Final hand-tightening of thecover screw should be done with the Small Hex Driver.

NOTE: At this step, a temporary healing abutment may beplaced for single-stage surgery instead of a cover screw.

15. Close tissue and suture.

Subcrestal Implant PlacementProtocol

Certain® PREVAIL™, OSSEOTITE® CertainAndOSSEOTITE External Hex Implant Placement

Page 47: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

4444

Due to the geometrical differences that exist between atapered and a parallel walled implant, there are severalimportant technique adjustments that are required.

In all tapered implant placement procedures, the surgeonshould determine the appropriate vertical position of theimplant (supracrestal, crestal or subcrestal) at the timeof osteotomy preparation. The surgeon should preparethe tapered osteotomy so that when the implant is fullyseated, the implant seating surface is at the desiredposition. The OSSEOTITE NT Depth Indicators (NTDI) weredesigned to confirm OSSEOTITE NT Implant position priorto seating the implant. After preparation of the osteotomywith the final shaping drill, suction out the osteotomy toremove debris. Select the corresponding NTDI and placethe tapered end into the osteotomy. Check the platformposition (crestal or subcrestal) of the NTDI in relation tothe adjacent bone. This position locates where the platformof the OSSEOTITE NT Implant will be positioned whenproperly placed. If during placement with the power drill,the OSSEOTITE NT Implant platform is higher in relation tothe bone than was the NTDI platform, the clinician shouldconsider using a hand ratchet to complete the implantplacement so that the tapered portion of the implant bodyconforms correctly with the tapered portion of theosteotomy (Figure 1. Proper Subcrestal Placement).

Over Preparing the osteotomy depth and then placing theimplant at a crestal level may result in a conical spacearound the apical and coronal aspects of the taperedimplant with minimal thread engagement (Figure 2. OverPrepared Subcrestal Placement). This placement positionmay result in decreased implant to osteotomy contact, withcontact occurring only along the parallel coronal portion ofthe implant, resulting in decreased stability of the implant.

Under Preparing the osteotomy depth and then placing theimplant more apical relative to the prepared depth mayresult in increased pressure along the tapered portion ofthe osteotomy and on the collar contact areas of theimplant profile (Figure 3. Under Prepared SubcrestalPlacement), which may result in pressure necrosis of thesurrounding bone. The clinician may consider undersizingthe osteotomy in extremely soft bone density (e.g. Type IV).

Figure 2. Over Prepared Subcrestal Placement Of 11.5mm Implant

Figure 1.Proper Subcrestal Placement Of 11.5mm Implant

15

13

11.5

10

8.5

7

Figure 3.Under Prepared Subcrestal Placement Of 11.5mm Implant

OSSEOTITE NT ® Implant SystemWhy Tapered Implants Are Different

15

13

11.5

10

8.5

7

15

13

11.5

10

8.5

7

Page 48: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

45

What Are The Features Of The New Quad Shaping Drills?

• The depth references on the QSD Drills are ledges or stepsand the hub has been eliminated. The hubless featureimproves access and flow of external irrigation.

• The QSD Drills have four flutes to improve cutting efficiencyand reduce drill chatter.

• The QSD Drills have end cutting capability.

• The QSD Drill design eliminates laser lines, producing a drillthat is more resistant to corrosion in those areas.

• The operating speed for the QSD Drills is 1200–1500rpmto correspond to the speed of other drills used during theprocedure (such as the Round Bur and Initial Twist Drill).

*Gingival Depth Marks - These depth marks are not used in thesurgical procedure covered in this manual.

Subcrestal

Supracrestal

Crestal

The Quad Shaping Drills (QSDs) are used to prepare theosteotomy for placement of OSSEOTITE NT® Implants.

NTDI QSD

NOTE: Depth landmarks on the Quad Shaping Drill versus correspon-ding depth landmarks on the OSSEOTITE NT Depth Indicator.

*{

Quad Shaping Drill (QSD)Introduction

Page 49: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

4646

Shaping Drill Speed:

Quad Shaping Drills should operate between 1200–1500rpm.

Quad Shaping Drills cut efficiently; reducing the downwardforce will allow the drill to cut without chatter.

Shaping Drill Technique:

• For either crestal or subcrestal implant placement, drill tothe top of either the crestal or subcrestal depth landmarkson the Quad Shaping Drill (full depth - see illustration).

• Do not pump the shaping drill as you might pump a twistdrill when creating the osteotomy. The shaping drill shouldbe advanced once to full depth, then be removed withoutany pumping action.

• Once the shaping drill has reached full depth, pull it out ofthe site without running the drill. If the drill does not pullout easily, tap the foot pedal while pulling out. In additionto preserving the integrity of the osteotomy site, thistechnique maximizes autogenous bone recovery from theshaping drill flutes.

• When placing an OSSEOTITE NT® Implant into softerType IV Bone, the surgeon should consider undersizingthe osteotomy by one shaping drill size (i.e. if placing anOSSEOTITE NT 5mm diameter X 10mm length implant inType IV Bone, stop at the 4mm diameter X 10mm lengthshaping drill and directly place the implant). For moredetailed information on implant placement in soft bone,please refer to page 59 of this manual.

• It is recommended that the clinician should tap theosteotomy when placing an OSSEOTITE NT Implant intoharder Types I and II Bone.

The Quad Shaping Drills have been designed with geometricaldepth landmarks to assess proper depth rather than laseretched markings. The clinician should become familiar withthese depth landmarks to prevent over or under preparationof the osteotomy site.

NOTE: During preparation of the osteotomy, the shaping drillshould advance into the osteotomy using light pressure. Theneed to push heavily on the shaping drill may indicate theneed to replace the shaping drill, the need to tap or that theprevious drill depth was inadequate.

Quad Shaping Drill (QSD)Key Points

Subcrestal

Crestal

Supracrestal (notch)

Parallel Walled Cutting Flutes

Tapered Cutting Flutes

Drilling Sequence And Recommended RPM

Assess Bone Quality

Adjust Drilling Speed

1. Round Drill 1200–1500rpm

2. Initial Twist Drill 1200–1500rpm

3. Pilot Drill (optional) 1200rpm

4. Quad Shaping Drill 1200–1500rpm

5. NT Tap

5. Implant Placement 15–20rpm

Bone Quality

Bone Type

Denser

1

15–20rpm

Dense

2

Dense

3

Soft

4

N/A

Apical End Cutting

Page 50: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

47

The OSSEOTITE NTDepth Indicators are used to simulate theimplant platform position prior to placing the implant.

OSSEOTITE NT ®

Depth Indicator (NTDI)Introduction

When using the NTDI, verify the osteotomy is clear of bonedebris.

Verify the NTDI platform position in reference to the crest ofthe bone. (This also verifies the depth of the osteotomy thathas been created) The NTDI platform should be at the levelyou desire the implant platform to attain. If the NTDI platformis too high versus the desired position, then re-drilling to theappropriate depth is required. If the NTDI platform is too deepversus the desired position, this indicates some degree ofosteotomy over preparation has taken place. To ensure properengagement of the implant, it must be seated to the depthdemonstrated by the NTDI. A longer implant can be considered.The clinician may consider verifying the position of the NTDIwith a radiograph.

When placing the implant, the implant platform should reachthe same position that the NTDI platform previously attained.If the implant platform is positioned higher in relation to thecrest of the bone than the platform of the NTDI previouslydemonstrated, then final hand ratcheting of the implant intothe osteotomy is required. If the surgical motor stalls priorto full placement of the implant due to insufficient torquegeneration, the clinician should consider hand ratchetingfor final positioning.

These guidelines will help ensure good bone-to-implantcontact and primary stability of the implant.

Page 51: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

4848

OSSEOTITE NT ® Implant TapsIntroduction

8.5mmOsteotomy

13mmOsteotomy

11.5mmOsteotomy

When placing an OSSEOTITE NT Implant in dense (Types I andII) bone, the clinician should consider tapping the osteotomyprior to implant placement.

The OSSEOTITE NT Implant Taps are all 8.5mm in length andare designed to tap the coronal aspect of the osteotomy for allimplant lengths. For both crestal and subcrestal OSSEOTITE NTImplant placement, the tap is advanced into the osteotomyuntil the junction of the colored mount and the tap body islevel with the crest of the bone.

Page 52: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

49

OSSEOTITE NT ® NTDI/QSDSurgical Tray

Coordinating The Use Of The Surgical Tray With The Surgical Manual Illustrations:The OSSEOTITE NT NTDI/QSD Surgical Tray is numbered to indicate theappropriate steps of the implant placement protocol. The following illustratedimplant placement protocol uses the same numbering sequence.

Close-up view of Surgical Tray illustrating numbering sequence.

Page 53: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5050

OSSEOTITE NT® Certain® And OSSEOTITE NT External Hex Implants Using Quad Shaping Drills - QSD/NTDI

2.0 or 2.3mmTwist Drill

Pilot DrillPD100

3.25mmQuad DrillQSD32XX

3.25mmNTDI

RoundDrill

RD100

INT32XX

4.0mmQuad DrillQSD4XX

5.0mmQuad DrillQSD5XX

6.0mm Quad DrillQSD6XX

INT4XX INT5XX INT6XX

NT32XX NT4XX NT5XX NT6XX

OR

Subcrestal Placement

Crestal Placement

DirectionIndicator (NTDI)

Quad ShapingDrill (QSD) See pages 51–68 for detailed instructions.

Subcrestal Implant PlacementProtocol - Quick Reference

OSSEOTITE NT CertainImplants

With Cover Screws

4.0mmNTDI

5.0mmNTDI

6.0mmNTDI

OSSEOTITE NTExternal Hex Implants

With Cover Screws

Page 54: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

51

Subcrestal Implant Placement ProtocolOSSEOTITE NT® Certain® And OSSEOTITE NT External

Hex 3.25mm Diameter Implant Using QSD/NTDI

1. Once the implant site has been determined, markthe site with a round drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. Set the drill speed at approximately 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100 or DI2310)Gelb Radiographic Depth Gauge (XDGXX)

4. With hard cortical bone, use the Pilot Drill toshape the coronal aspect of the implant site. Drill to thedepth mark for placement of an OSSEOTITE NT Implant.The apical end of a 3.25mm Shaping Drill is smaller indiameter than the osteotomy created by the 2.3mmInitial Twist Drill. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

Refer to page 50 for a quick reference guide to OSSEOTITE NTImplant Placement Protocol using Quad Shaping Drills.

Page 55: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5252

Final Shaping Drill Step For 3.25mm OSSEOTITE NT® Implant:

5a. Proceed with the 3.25mm QSD Shaping Drill thatis the same length as the OSSEOTITE NT Implant to beplaced. The recommended drill speed is 1200–1500rpm.

• Instruments needed:3.25mm QSD Shaping Drill (QSD32XX)

5b. Using suction, remove drilling debris from theosteotomy before proceeding with the NTDIDepth/Direction Indicator.

5c. Insert the tapered end of the 3.25mm NTDIDepth/Direction Indicator (purple) that corresponds tothe length of the OSSEOTITE NT Implant to be placed.This will simulate the position of the implant platform inrelation to the crest of the bone. If the position of theNTDI Depth/Direction Indicator does not indicate properosteotomy depth, adjust the depth of the osteotomy withthe corresponding 3.25mm QSD Shaping Drill orconsider a longer length implant if the site has been overprepared. Re-evaluate with a proper length NTDIDepth/Direction Indicator.

NOTE: If placing a 3.25mm OSSEOTITE NT Implant inType I or Type II Bone, tapping with the NTAP32 isrecommended.

Continue on for placement of 4.0, 5.0 and 6.0mmOSSEOTITE NT Implants.

Proceed to step 9a on page 56 for OSSEOTITE NTImplant Placement.

1.25

3.4

2.0

3.4

1.25

2.5.7

2.0

Page 56: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

53

Second Shaping Drill Step For Placement Of 4.0,5.0 And6.0mm OSSEOTITE NT Implant. Final Shaping Drill StepFor 4.0mm OSSEOTITE NT Implant.

6a. Continue preparing the osteotomy with the4.0mm QSD Shaping Drill that is the same length as theOSSEOTITE NT Implant to be placed. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:4.0mm QSD Shaping Drill (QSD4XX)

NOTE: If placing a 4.0mm OSSEOTITE NT Implant in TypeIV Bone, undersizing the osteotomy by one shaping drillsize (diameter) is recommended.

6b. Using suction, remove drilling debris from theosteotomy before proceeding to the NTDI.

6c. Insert the tapered end of the 4.0mm NTDIDepth/Direction Indicator (blue) that corresponds to thelength of the OSSEOTITE NT Implant to be placed. Thiswill simulate the position of the implant platform inrelation to the crest of the bone. If the position of theNTDI Depth/Direction Indicator does not indicate properosteotomy depth, adjust the depth of the osteotomywith the corresponding 4.0mm QSD Shaping Drill orconsider a longer length implant if the site has beenover prepared. Re-evaluate with a proper length NTDIDepth/Direction Indicator.

NOTE: If placing a 4.0mm OSSEOTITE NT Implant in Type Ior Type II Bone, tapping with the NTAP4 is recommended.

6d. Countersink the osteotomy with an ICD100 toaccommodate the cover screw for the 4mm implant.The recommended drill speed is 1200–1500rpm.

IMPORTANT NOTE: When placing a 4.0mm diameterOSSEOTITE NT Implant subcrestally, you must use anICD100 Countersink Drill to prepare the ridge beforeplacing the implant. The ICD100 is not required forcrestal and supracrestal placement of a 4.0mm diameterOSSEOTITE NT Implant.

Continue on for placement of 5.0 and 6.0mmOSSEOTITE NT Implants.

Proceed to step 9a on page 56 for OSSEOTITE NTImplant Placement.

4.1

2.5

1.25 1.25

2.5

2.7.7

4.1

Subcrestal Implant Placement ProtocolOSSEOTITE NT® Certain® And OSSEOTITE NT External

Hex 4.0mm Diameter Implant Using QSD/NTDI

Page 57: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5454

Third Shaping Drill Step For Placement Of 5.0 And 6.0mmOSSEOTITE NT Implant. Final Shaping Drill Step For 5.0mmOSSEOTITE NT Implant.

7a. Resume preparing the osteotomy with the5.0mm QSD Shaping Drill that is the same length as theOSSEOTITE NT Implant to be placed. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:5.0mm QSD Shaping Drill (QSD5XX)

NOTE: If placing a 5.0mm OSSEOTITE NT Implant in TypeIV Bone, undersizing the osteotomy by one shaping drillsize (diameter) is recommended.

7b. Using suction, remove drilling debris from theosteotomy before proceeding to the NTDI.

7c. Insert the tapered end of the 5.0mm NTDIDepth/Direction Indicator (yellow) that corresponds tothe length of the OSSEOTITE NT Implant to be placed.This will simulate the position of the implant platform inrelation to the crest of the bone. If the position of theNTDI Depth/Direction Indicator does not indicate properosteotomy depth, adjust the depth of the osteotomywith the corresponding 5.0mm QSD Shaping Drill orconsider a longer length implant if the site has beenover prepared. Re-evaluate with a proper length NTDIDepth/Direction Indicator.

NOTE: If placing a 5.0mm OSSEOTITE NT Implant in Type Ior Type II Bone, tapping with the NTAP5 is recommended.

Continue on for placement of 6.0mm OSSEOTITE NTImplant.

Proceed to step 9a on page 56 for OSSEOTITE NTImplant Placement.

5.0

3.1

1.25 1.25

2.75.0

.7

3.1

Subcrestal Implant Placement ProtocolOSSEOTITE NT® Certain® And OSSEOTITE NT External

Hex 5.0mm Diameter Implant Using QSD/NTDI

Page 58: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

55

Final Shaping Drill Step For 6.0mm OSSEOTITE NT Implant:

8a. Resume preparing the osteotomy with the6.0mm QSD Shaping Drill that is the same length as theOSSEOTITE NT Implant to be placed. The recommendeddrill speed is 1200–1500rpm.

• Instruments needed:6.0mm QSD Shaping Drill (QSD6XX)

NOTE: If placing a 6.0mm OSSEOTITE NT Implant in TypeIV Bone, undersizing the osteotomy by one shaping drillsize (diameter) is recommended.

8b. Using suction, remove drilling debris from theosteotomy before proceeding to the NTDI.

8c. Insert the tapered end of the 6.0mm NTDIDepth/Direction Indicator (green) that corresponds tothe length of the OSSEOTITE NT Implant to be placed.This will simulate the position of the implant platform inrelation to the crest of the bone. If the position of theNTDI Depth/Direction Indicator does not indicate properosteotomy depth, adjust the depth of the osteotomywith the corresponding 6.0mm QSD Shaping Drill orconsider a longer length implant if the site has beenover prepared. Re-evaluate with a proper length NTDIDepth/Direction Indicator.

NOTE: If placing a 6.0mm OSSEOTITE NT Implant in Type Ior Type II Bone, tapping with the NTAP6 is recommended.

Proceed to step 9a on page 56 for OSSEOTITE NTImplant Placement.

3.6

6.0

1.25 1.25

2.7

3.6

6.0

.7Subcrestal Implant Placement ProtocolOSSEOTITE NT® Certain® And OSSEOTITE NT External

Hex 6.0mm Diameter Implant Using QSD/NTDI

Page 59: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5656

No-Touch™ Delivery System

9a. Remove contents from the implant box.

9b. The nonsterile assistant should peel back thetray lid and drop the No-Touch Implant Tray onto thesterile drape.

9c. Place the No-Touch Implant Tray into the appropriate location on the surgical tray.

9d. Peel back the tray lid to expose the implant andcover screw.

Subcrestal Implant PlacementProtocol

OSSEOTITE NT® Certain®And OSSEOTITE NTExternal Hex Implant Placement

Page 60: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

57

Subcrestal Implant PlacementProtocol

OSSEOTITE NT® Certain®And OSSEOTITE NTExternal Hex Implant Placement

10. Pick up the implant from the surgical tray usingthe Certain Implant Placement Driver Tip.NOTE: The Certain MicroMiniplant™ (3.25mm dia.)requires the use of a dedicated MicroMiniplant Driver Tip(IMPDTS or IMPDTL) that is marked with a purple bandon the shank. The Internal Hex configuration of theMicroMiniplant is smaller than the standard CertainInternal Hex Connection (4, 5 and 6mm dia. implants).The item numbers can be identified on the side of thedriver tip.

or

Pick up the implant from the surgical tray usingthe Handpiece Connector.

Carry the implant to the mouth facing upward toprevent accidental dislodging.

• Instruments needed:Implant Placement Driver Tip (IIPDTS or IIPDTL)Handpiece Connector (MDR10)

11. Place the implant in the prepared site atapproximately 15–20rpm. It is not uncommon for thehandpiece to stall before the implant is completelyseated. The implant position must match what wassimulated with the NTDI or there is a risk of a poor fitbetween the implant and osteotomy. In dense bone, it isrecommended to tap the site prior to implant placement.

12. Final seating of the implant may require the useof the Ratchet Extension and the Ratchet Wrench.

• Instruments needed:Ratchet Wrench (WR150)Certain Ratchet Extension (IRE100 or IRE200)External Hex Ratchet Extension (RE100 or RE200)

Page 61: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

5858

13. To remove the Certain® Implant Placement DriverTip from the implant, lift straight up and out.

To remove the implant mount, place the Open-End Wrench onto the mount. Loosen the screw at thetop of the mount with a Large Hex Driver or the LargeHex Driver Tip inserted into the Right-Angle Driver androtate counter-clockwise. After the screw is loosened,rotate the Open-End Wrench counter-clockwise slightlybefore removing the mount. The mount may be carriedfrom the mouth with the Open-End Wrench.

• Instruments needed:Open-End Wrench (CW100), Large Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with CADD1) or a Large Hex Driver (PHD02N)

14. Pick up the Cover Screw from the No-Touch™Implant Tray with the Implant Driver or Large HexDriver and place onto the implant.

NOTE: When using the Certain Implant Placement Driver,reduce the torque setting on the drilling unit to 10Ncm.

• Instruments needed:Implant Placement Driver Tip (IIPDTS or IIPDTL)Large Hex Driver (PHD02N)

15. Pick up the Cover Screw from the No-TouchImplant Tray with the Small Hex Driver and place ontothe implant.

• Instruments needed:Small Hex Driver (PHD00N)

NOTE: At this step, a temporary healing abutment may beplaced for single-stage surgery instead of a cover screw.

16. Close the tissue and suture.

Page 62: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

59

The drilling protocol for the Subcrestal Placement of the 4.0 x 11.5mm OSSEOTITE NT Certain Implant was chosenas a representative protocol to provide a detailed implantplacement synopsis. For the drilling protocol specific toimplant diameter and design, please refer to the ImplantPlacement Protocol Quick Reference section of this manual.

NOTE: When placing an OSSEOTITE NT Implant in soft Type IVBone, the osteotomy should be undersized by one shaping drillsize (diameter).

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100, DI2310 or NTDIXXX)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex 4.0mm Diameter Implant

Soft Type IV Bone Protocol

4.1

2.5

1.25 1.25

2.5

2.7.7

4.1

Page 63: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

6060

4. Use the Pilot Drill to shape the coronal aspect ofthe implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Proceed with the 3.25 x 11.5mm Quad ShapingDrill. Drill to the top depth mark on the QSD forsubcrestal placement. The recommended drill speed is1200–1500rpm.

• Instruments needed:3.25mm QSD (QSD32XX)

6. Finish the osteotomy with the Countersink Drill.The recommended drill speed is 1200–1500rpm.

• Instruments needed:Countersink Drill (ICD100)

IMPORTANT NOTE: When placing a 4.0mm diameterOSSEOTITE Certain® Implant subcrestally, you must usean ICD100 Countersink Drill to prepare the ridge before placing the implant.

Proceed to step 11a on page 66 for OSSEOTITE NT®

Implant Placement in soft Type IV Bone.

Page 64: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

61

The drilling protocol for the Subcrestal Placement of the 5.0 x 11.5mm OSSEOTITE NT Certain Implant was chosenas a representative protocol to provide a detailed implantplacement synopsis. For the drilling protocol specific toimplant diameter and design, please refer to the ImplantPlacement Protocol Quick Reference section of this manual.

NOTE: When placing an OSSEOTITE NT Implant in soft Type IVBone, the osteotomy should be undersized by one shaping drillsize (diameter).

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100, DI2310 or NTDIXXX)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex 5.0mm Diameter Implant

Soft Type IV Bone Protocol

5.0

3.1

1.25 1.25

2.75.0

.7

3.1

Page 65: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

6262

4. Use the Pilot Drill to shape the coronal aspectof the implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Proceed with the 3.25 x 11.5mm Quad ShapingDrill. Drill to the top depth mark on the QSD forsubcrestal placement. The recommended drill speed is1200–1500rpm.

• Instruments needed:3.25mm QSD (QSD32XX)

6. Continue preparing the osteotomy with the4.0mm QSD. Drill to the top depth mark on the QSD.The recommended drill speed is 1200–1500rpm.

• Instruments needed:4.0mm QSD (QSD4XX)

7. Finish with a 5.0 x 8.5mm QSD. This will createan osteotomy of proper dimension in the dense corticalbone to receive the implant, but will slightly undersizethe osteotomy in the cancellous region to allow forsome compression. The recommended drill speed is1200–1500rpm.

• Instruments needed:5.0mm QSD (QSD585)

Proceed to step 11a on page 66 for OSSEOTITE NT®

Implant Placement in soft Type IV Bone.

Page 66: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

63

The drilling protocol for the Subcrestal Placement of the 6.0 x 11.5mm OSSEOTITE NT Certain Implant was chosenas a representative protocol to provide a detailed implantplacement synopsis. For the drilling protocol specific toimplant diameter and design, please refer to the ImplantPlacement Protocol Quick Reference section of this manual.

NOTE: When placing an OSSEOTITE NT Implant in soft Type IVBone, the osteotomy should be undersized by one shaping drillsize (diameter).

1. Once the implant site has been determined, markthe site with a Round Drill and penetrate the corticalbone. The recommended drill speed is 1200–1500rpm.

• Instruments needed:Round Drill (RD100 or DR100)

2. Proceed with the Initial Twist Drill toapproximately 7mm, then verify the direction with thethin portion of the Direction Indicator.

Continue to penetrate the bone to the desired depth. The recommended drill speed is 1200–1500rpm.

• Instruments needed:2.0 or 2.3mm Twist Drill (ITD215, DT215, DTN215 or ACT2015)

3. Verify the direction and position of thepreparation by inserting the thin portion of the DirectionIndicator into the osteotomy. Thread dental flossthrough hole to prevent accidental swallowing.

At this step, a Gelb Radiographic Depth Gauge may alsobe used.

• Instruments needed:Direction Indicator (DI100, DI2310 or NTDIXXX)Gelb Radiographic Depth Gauge (XDGXX)

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex 6.0mm Diameter Implant

Soft Type IV Bone Protocol3.6

6.0

1.25 1.25

2.7

3.6

6.0

.7

Page 67: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

6464

4. Use the Pilot Drill to shape the coronal aspectof the implant site. Drill to the depth mark. The recom-mended drill speed is 1200–1500rpm.

• Instruments needed:Pilot Drill (PD100 or DP100)

5. Proceed with the 3.25 x 11.5mm Quad ShapingDrill. Drill to the top depth mark on the QSD forsubcrestal placement. The recommended drill speed is1200–1500rpm.

• Instruments needed:3.25mm QSD (QSD32XX)

6. Continue preparing the osteotomy with the4.0mm QSD. Drill to the top depth mark on the QSD.The recommended drill speed is 1200–1500rpm.

• Instruments needed:4.0mm QSD (QSD4XX)

7. Resume preparing the osteotomy with the5.0mm QSD that is the same length as the implant tobe placed. The recommended drill speed is1200–1500rpm.

• Instruments needed:5.0mm QSD (QSD5XX)

Page 68: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

65

8. Using suction, remove drilling debris from theosteotomy before proceeding to the NTDI.

9. Insert the tapered end of the 5.0mm NTDIDepth/Direction Indicator (yellow) that corresponds tothe length of the OSSEOTITE NT Implant to be placed.This will simulate the position of the implant platform inrelation to the crest of the bone. If the position of theNTDI Depth/Direction Indicator does not indicate properosteotomy depth, adjust the depth of the osteotomywith the corresponding 5.0mm QSD or consider alonger length implant if the site has been over prepared.Re-evaluate with a proper length NTDI. The recommended drill speed is 1200–1500rpm.

10. Finish with a 6.0 x 8.5mm Quad Shaping Drill.This will create an osteotomy of proper dimension inthe dense cortical bone to receive the implant, but willslightly undersize the osteotomy in the cancellousregion to allow for some compression. Therecommended drill speed is 1200–1500rpm.

• Instruments needed:6.0mm QSD (QSD685)

Proceed to step 11a on page 66 for OSSEOTITE NTImplant Placement in soft Type IV Bone.

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex 6.0mm Diameter Implant

Page 69: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

6666

No-Touch™ Delivery System

11a. Remove contents from the implant box.

11b. The nonsterile assistant should peel back thetray lid and drop the No-Touch Implant Tray onto thesterile drape.

11c. Place the No-Touch Implant Tray into the appropri-ate location on the surgical tray.

11d. Peel back the tray lid to expose the implant andcover screw.

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex Implant Placement

Page 70: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

67

Subcrestal Implant PlacementProtocol In Soft Type IV Bone

OSSEOTITE NT® Certain® And OSSEOTITE NTExternal Hex Implant Placement

12. Pick up the implant from the surgical tray usingthe Certain Implant Placement Driver Tip.NOTE: The Certain MicroMiniplant™ (3.25mm dia.)requires the use of a dedicated MicroMiniplant Driver Tip(IMPDTS or IMPDTL) that is marked with a purple bandon the shank. The Internal Hex configuration of theMicroMiniplant is smaller than the standard CertainInternal Hex Connection (4, 5 and 6mm dia. implants).The item numbers can be identified on the side of thedriver tip.

or

Pick up the implant from the surgical tray usingthe Handpiece Connector.

Carry the implant to the mouth facing upward toprevent accidental dislodging.

• Instruments needed:Implant Placement Driver Tip (IIPDTS or IIPDTL)Handpiece Connector (MDR10)

13. Place the implant in the prepared site atapproximately 15–20rpm. It is not uncommon for thehandpiece to stall before the implant is completelyseated. The implant position must match what wassimulated with the NTDI or there is a risk of a poor fitbetween the implant and osteotomy. In dense bone, it isrecommended to tap the site prior to implant placement.

14. Final seating of the implant may require the useof the Ratchet Extension and the Ratchet Wrench.

• Instruments needed:Ratchet Wrench (WR150)Certain Ratchet Extension (IRE100 or IRE200)External Hex Ratchet Extension (RE100 or RE200)

Page 71: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

6868

15. To remove the Certain® Implant Placement DriverTip from the implant, lift straight up and out.

To remove the implant mount, place the Open-End Wrench onto the mount. Loosen the screw at thetop of the mount with a Large Hex Driver or the LargeHex Driver Tip inserted into the Right-Angle Driver androtate counter-clockwise. After the screw is loosened,rotate the Open-End Wrench counter-clockwise slightlybefore removing the mount. The mount may be carriedfrom the mouth with the Open-End Wrench.

• Instruments needed:Open-End Wrench (CW100), Large Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with CADD1) or a Large Hex Driver (PHD02N)

16. Pick up the Cover Screw from the No-Touch™Implant Tray with the Implant Driver or Large HexDriver and place onto the implant.

NOTE: When using the Certain Implant Placement Driver,reduce the torque setting on the drilling unit to 10Ncm.

• Instruments needed:Implant Placement Driver Tip (IIPDTS or IIPDTL)Large Hex Driver (PHD02N)

17. Pick up the Cover Screw from the No-TouchImplant Tray with the Small Hex Driver and place ontothe implant.

• Instruments needed:Small Hex Driver (PHD00N)

NOTE: At this step, a temporary healing abutment may beplaced for single-stage surgery instead of a cover screw.

18. Close the tissue and suture.

Page 72: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

69

Several advantages can be realized by utilizing a two-stageimplant system in a single-stage treatment protocol.Attaching a one-piece or two-piece healing abutmentimmediately following implant placement eliminates theneed for a second-stage surgery. Eliminating the secondsurgical procedure reduces trauma and decreases treatmenttime, while the two-stage implant design maintainsrestorative flexibility.

1. After the implant is fully seated in the osteotomy,remove the implant mount from the External Hex Implant.

2. Select the appropriate one-piece healingabutment or Encode® Healing Abutment depending uponthe implant seating surface, tissue depth and desiredEP® Dimension.

or

Select the appropriate one or two-piece healingabutment or Encode Healing Abutment depending uponthe implant seating surface, tissue depth and desired EPDimension.

Bone profiling of the osteotomy may benecessary to fully seat the healing abutment onto theimplant.

3. Tighten the one or two-piece healing abutmentscrew to 20Ncm and suture the soft tissue around thehealing abutment.

Single-Stage TreatmentsOSSEOTITE® Implants

Page 73: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

7070

EP® Healing Abutments A Bone Profiling Pin, Bone Profilers and corresponding EPBone Profilers are available to contour the bone that is toreceive the EP Healing Abutment. These tools are especiallyhelpful in a single-stage surgical protocol when the implant isplaced subcrestally.

If the implant is placed subcrestally and use of an EP HealingAbutment is indicated, the coronal aspect of the osteotomymust be prepared to receive the flare of the EP HealingAbutment.

NOTE: Non-EP, straight healing abutments and impression copings are available if bone profiling is not preferred at eitherstage one or stage two surgery.

OSSEOTITE Certain® Two-Piece Bone Profiling Pin (IBPGP)The OSSEOTITE Certain Implant requires a dedicated BoneProfiling Pin, which is used with the existing EP BoneProfilers. This new two-piece design allows the pin to engagethe internal hex of the implant. The hex engagement preventsthe pin from tightening into the implant during profiling, making it easy to remove. Lubricating the top of the pin withan appropriate lubricant, such as tetracycline ointment,is recommended. Do not exceed 50rpm when using BoneProfilers Pins.

Bone Profiling Technique

OSSEOTITE CertainTwo-Piece

Bone Profiler Pin

OSSEOTITE ExternalHex One-Piece

Bone Profiler Pin

• EP Bone Profiler slides over the two-Piece Bone Profiler Pin.

• EP Bone Profiler creates a flare in the crest of bone.

• Flare of EP Abutment matchesthe flare of the corresponding EP Bone Profiler.

• EP Healing Abutment seatedproperly onto the implant insubcrestal placement.

EP Bone Profilers correspond to sizes of EP Healing Abutments

Bone ProfilingOSSEOTITE® Implants

Page 74: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

71

Surgical IndexingOSSEOTITE® Implants

Surgeon 1. For surgical implant placement of a 3i Implant, follow

normal protocol as described in the previous sections.

Surgical Indexing2. A surgical index may be made at stage one or stage two

to facilitate the fabrication of a provisional restoration.This can be accomplished by using a Pick-UpImpression Coping or a Hexed Temporary Cylinder withretention and a waxing screw and a medium-to-heavybody impression material.

Creating A Surgical Index3. Select the proper Pick-Up Impression Coping by

matching the diameter of the implant platform.

Activate the fingers using the QuickSeat® ActivatorTool. Place the Pick-Up Impression Coping or theTemporary Cylinder into the implant, line up the hexand press firmly until feeling the tactile click.

Place the Pick-Up Impression Coping or theTemporary Cylinder on the implant and engage the hex.

Thread the Pick-Up Impression Coping Screw or waxingscrew into the implant until finger tight. Tighten thescrew using the Large Hex Driver.

4. If a flapless surgery is performed or if the index is madeat stage two surgery, radiograph the interface to verifycomplete seating of the coping on the implant. Place thefilm perpendicular to the interface of the coping on theimplant.

Seated Not Seated

Page 75: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

7272

5. Syringe a medium-to-heavy body impression materialaround the impression coping or temporary cylinderand over the occlusal surfaces of the adjacent teethapproximately 1.5 teeth on either side. Allow theimpression material to set per the manufacturer’sinstructions. Once the material has set, remove theimpression coping screw or waxing screw using theLarge Hex Driver. Remove the surgical index from themouth. Send the index to the restorative clinician so itmay be included in the package to the laboratory.

6. Select a healing abutment by matching the implantplatform, preferred EP® diameter and collar height. Thecollar height should be selected by measuring from theimplant platform to the highest crest of the gingivaltissue and add 1mm.

Page 76: Surgical Manual - Biomet 3i · PDF fileSurgical Protocol Implant Placement Surgical Protocol Cover Screw Insertion No-Touch™ Mountless OSSEOTITE Certain Implant Delivery System Surgical

3i and design, Certain, Encode, EP, OSSEOTITE, OSSEOTITE NT, OSSEOTITE XP and QuickSeat are registered trade-marks and ACT, ICE, MicroMiniplant, No-Touch and PREVAIL are trademarks of Implant Innovations, Inc., a BIOMETCo., BIOMET is a registered trademark of BIOMET, Inc. © 2006 Implant Innovations, Inc. All rights reserved. REV E 5/06

Implant Innovations, Inc.Global Headquarters4555 Riverside DrivePalm Beach Gardens, FL 334101-800-342-5454Outside The U.S.: +1-561-776-6700Fax: +1-561-776-1272www.3i-online.com

CATSM

SUBSIDIARIES

AUSTRALIAPhone: +61-2-9855-4444Fax: +61-2-9888-9900

BELGIUMPhone: +32-2-5410290Fax: +32-2-5410291

BRAZILPhone: +55-11-5081-4405Fax: +55-11-5081-7484

CANADAPhone: +514-956-9843Fax: +514-956-9844

FRANCEPhone: +33-1-41054343Fax: +33-1-41054340

GERMANYPhone: +49-721-255177-10Fax: +49-721-255177-73

IRELANDPhone: +35-31-477-3925Fax: +35-31-402-9590

MEXICOPhone: +52-55-5679-1619Fax: +52-55-5684-8098

THE NETHERLANDSPhone: +31-35-646-2830Fax: +31-35-646-2666

NORDIC REGIONPhone: +46-40-17-6090Fax: +46-40-17-6099

PORTUGALPhone: +351-21-000-1645Fax: +351-21-000-1675

SPAINPhone: +34-93-470-59-50Fax: +34-93-372-11-25

SWITZERLANDPhone: +41-1-3804646Fax: +41-1-3834655

U.K.Phone: +44-1628-829314Fax: +44-1628-820182

DISTRIBUTORS

ARGENTINADentalmax, SAPhone: +541-1482-71001Fax: +541-1482-67373

AUSTRIAWieladentPhone: +43-7672-93901Fax: +43-7672-93903

CHILECybel, SAPhone: +56-2-2321883Fax: +56-2-2330176

COLOMBIAImplantes y ComponentesPhone: +571-612-9362Fax: +571-620-5450

COSTA RICAImplantec S.A.Phone: +506-2-256411Fax: +506-2-247620

EL SALVADORDentimerc SA de CVPhone: +503-263-6350Fax: +503-263-6676

GREECEKostas Kornisorlis and Co.Phone: +302310-269-079Fax: +302310-555-573

ISRAELH.A. SystemsPhone: +972-3-6138777Fax: +972-3-6138778

ITALYBiomax, srl.Phone: +39-0444-913410Fax: +39-0444-913695

JAPANImplant Innovations JapanPhone: +81-66-868-3012Fax: +81-66-868-2444

KOREAJungsan Biomed Corp.Phone: +82-2-516-1808Fax: +82-2-514-9434

LEBANONTamer Freres s.a.l.Phone: +961-1-485690Fax: +961-1-510233

PANAMAOdontomedica, S.A.Phone: +507-2-239622Fax: +507-2-239621

PARAGUAYAndres H. Arce y Cia SRLPhone: +595-21-208185Fax: +595-21-496291

POLANDDental DepotPhone: +48-71-341-3091Fax: +48-71-343-6560

RUSSIACom-dentalPhone: +7095-797-6686Fax: +7095-242-9567

SINGAPOREAsia Implant Support & ServicesPhone: +65-6223-2229Fax: +65-6220-3538

TAIWANKuo Hwa Dental Suppliers Co., Ltd.Phone: +886-2-2226-1770Fax: +886-2-2226-8747

THAILANDPT Endeavour Co., Ltd.Phone: +662-264-2574Fax: +662-264-2573

UKRAINEOlimpex, Ltd.Phone: +38-044-209-5713Fax: +38-044-482-3678

URUGUAYPro3implant S.R.L.Phone: +598-2-4034163Fax: +598-2-4034163