Upload
others
View
17
Download
0
Embed Size (px)
Citation preview
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
1 de 13
FINAL REPORT
Supplementary Comparison
CALIBRATION OF MASS AND CONVENTIONAL MASS
OF WEIGHTS
BETWEEN INACAL - CENAM - CESMEC
(2 kg and 10 kg)
SIM.M.M-S13
September 2017
FINAL REPORT
SIM.M.M-S13
Index Page 1 Introduction 2 2 List of participants NMIs 2 3 Travelling weight standards 3 4 Circulation Schedule 3 5 Surface damages of the standard 4 6 Measurement conditions 4 7 Procedures and measurement methods 5 8 Measurement results 5 9 Results analysis 8 10 Results for mass 10 11 Results for conventional mass 11 12 Conclusions 12 13 Acknowledgments 12 14
References 13
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
2 de 13
FINAL REPORT
SUPPLEMENTARY COMPARISON
CALIBRATION OF MASS AND CONVENTIONAL MASS OF WEIGHTS
BETWEEN INACAL - CENAM - CESMEC
(2 kg and 10 kg)
SIM.M.M-S13
Luz Cori Almonte1, Donny Taipe Araujo1, Aldo Quiroga Rojas 1, Luis Omar Becerra2, Luis Manuel Peña2,
Fernando Garcia3, Francisco Garcia3.
1 Instituto Nacional de Calidad, INACAL / Perú 2 Centro Nacional de Metrología, CENAM / México
3 Centro de Estudios de Medición y Certificación de Calidad, CESMEC / Chile
1. Introduction
This report describes the results of the supplementary comparison of 2 kg and 10 kg stainless steel mass standard between INACAL-PERÚ, CENAM-MEXICO and CESMEC-CHILE.This comparison was undertaken by INACAL, CENAM and CESMEC, was based on a decision taken on August, 2012. The supplementary comparison was piloted by INACAL- PERÚ. This comparison was registered as an official supplementary comparison under the designation SIM.M.M-13. The reference values were provide by CENAM-MEXICO. The travelling weight standards belonged to INACAL-. They were OIML E2 weights. The travelling weight standards were prepared by INACAL. INACAL measured the volume of the travelling standard of 10 kg and PTB measured the volume of the travelling standard of 2 kg. The travelling weight standards were circulated between the NMIs. Each NMI determined their mass and uncertainty using their procedures and methods. The measurements were carried out from September 2012.
2. List of participant NMIs
The participating laboratories and their respective technical contacts are listed below: • Instituto Nacional de Calidad - INACAL, Perú (*)
Luz Cori Almonte Donny Taipe Araujo Aldo Quiroga Rojas
(*) Before June 01st 2015, the Peruvian NMI was the Instituto Nacional de la Competencia y de la Protección de la Propiedad
Intelectual - INDECOPI
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
3 de 13
• Centro Nacional de Metrología - CENAM, México
Luis Omar Becerra Luis Manuel Peña
• Centro de Estudios de Medición y Certificación de Calidad - CESMEC, Chile
Francisco García Fernando García
Raúl Hernández 3. Travelling weight standards
The travelling standards used were OIML Class E2 weights belonging to INACAL, with the following nominal values of 2 kg and 10 kg. The shape and material of travelling weight standards are shown in Table 1.
Table 1
Nominal value Accuracy Class Material Shape
2 kg E2 Stainless Steel Cylindrical w ith a lif ting knob
10 kg E2 Stainless Steel Cylindrical w ith a lif ting knob
In August 2012, the volume of the mass of 10 kg was determined in the laboratory INACAL. The volume of the mass of 2 kg was measured in the PTB. These results were provided to participating laboratories and are shown in Table 2.
Table 2: Information about of the travelling weight standards
Identification Manufacturer Nominal Value Volume at
20 °C cm
3
Uncertainty k=2
cm3
50703712 SARTORIUS 2 kg 254,73 0,04
212 KERN 10 kg 1265,9 0,8
4. Circulation Schedule
The circulation schedule of the travelling weight standards are shown in tables 3 and 4. INACAL started the measurements in September 2012 and the last measurements were also made by INACAL in September 2014. The original schedule was modified due to difficulties in the transportation and retention of the standards at customs in some countries.
Table 3: 2 kg weight
Laboratory Period of measurements
Instituto Nacional de Calidad - INACAL 1, Perú 2012-09
Centro Nacional de Metrología - CENAM, México 2012-09
Centro de Estudios de Medición y Certif icación de Calidad CESMEC, Chile
2013-02
Instituto Nacional de Calidad - INACAL 2, Perú 2013-03
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
4 de 13
Table 4: 10 kg weight
Laboratory Period of
measurements
Instituto Nacional de Calidad – INACAL1, Perú 2012-10
Centro de Estudios de Medición y Certif icación de Calidad - CESMEC, Chile
2013-02
Centro Nacional de Metrología - CENAM, México 2014-01
Instituto Nacional de Calidad – INACAL 2, Perú 2014-12
5. Surface damages of the standards:
The traveling standards were examined by each NMI at the reception and departure of the weights in order to register all marks and damages during circulation. The participating laboratories sent to the pilot laboratory the forms used to register the superficial conditions of the traveling standards, at both reception and departure. The data sent by the participating laboratories showed that there was no significant damage on the traveling standards
6. Measurement conditions
The environmental conditions for determination of the mass and conventional mass reported by the participating laboratories are listed in table 5.
Table 5
INACAL 1 CENAM CESMEC INACAL 2
2 kg 10 kg 2 kg 10 kg 2 kg 10 kg 2 kg 10 kg
T (º C) 22,70 21,00 21,09 20,02 23,65 23,4 20,45 20,81
ΔT/h (ºC) 0,08 0,23 0,20 0,16 0,1 0,0 0,25 0,32
H / tr 51,85 47,33 9,56 °C 10,12 °C 56,85 58,6 57,0 56,7
ΔH / Δtr (51,7-52,0)% (48,0-46,7)% (9-10) °C (10-11) °C (57-57)% (57-60)% (57-57) % (56-57) %
ρa(kg/m3) 1,1655 1,1760 0,9545 0,9580 1,1005 1,1009 1,1695 1,1685
The resolutions of the weighing instruments, used by the different laboratories, are showed in Table 6.
Table 6
Units in mg INACAL 1 CENAM CESMEC INACAL 2
2 kg 0,1 0,2
0,01 0,1 0,01 0,2
10 kg 0,2
5 0,01 1
0,2
5
The resolutions of the environmental conditions, used by the different laboratories, are shown in table 7.
Table 7
INACAL 1 CENAM CESMEC INACAL 2
TEMPERATURE 0,01 °C 0,001 °C 0,1 °C 0,01 °C
PRESURE 0,1 mbar 1 Pa 1 pa 0,1 mbar
HUMIDITY 0,1 % 0,1 % 0,1 % 0,1 %
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
5 de 13
Table 8 shows the traceability of the mass standards used by the NMIs, to indicate the possible correlation of the traceability source of the standards.
Table 8
INACAL 1 CENAM CESMEC INACAL 2
2 kg CEM CENAM PTB CEM
10 kg NRC CENAM PTB CENAM
The 10 kg mass standard used by INACAL was traceable to NRC (INACAL 1) and traceable to CENAM (INACAL 2). The dates of calibration of the mass standards used for the measurements are listed in table 9. An overdue calibration date could introduce a drift that may affect the results of the calibration.
Table 9
INACAL 1 CENAM CESMEC INACAL 2
2 kg 2012 2012 2011 2014
10 kg 2012 2014 2011 2014
7. Procedures and measurement methods
The measurement method used by laboratories CENAM and INACAL was the subdivision method. CESMEC used the direct comparison method. INACAL uses a least squares method to obtain the results by means of the Gauss-Markov estimation, using weights of different nominal value. This method involves solving an oversized system of equations. CENAM used the subdivision procedure, starting from one kilogram and generating a set of independent measurements that is modeled by a system of linear equations and the solution to this set of equations was found by the technique of least squares. The mass values and their respective uncertainties for the travelling weight standards were obtained by subdivision method. Matrix solution by weighted least squares was applied to the weighing scheme. The density of the air was calculated by equation CIPM-2007 and the approximation formula for air density, according to table 10.
Table 10
Laboratory Equation
INACAL CIPM-20071
CENAM CIPM-20071
CESMEC Approximation formula
for air density2
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
6 de 13
8. Measurements results
Table 11 shows the mass correction found by the NMIs and its associated uncertainty for a confidence level of 95,45 %. 1Rev ised f ormula f or the density of moist air (CIPM-2007) [2] 2OIML R111, E.3. [1]
Table 11: Mass correction
Participant
laboratory
2 kg 10 kg
Reference
Value
mg
mA
mg
UA (k=2)
mg
Reference
Value
mg
mA
mg
UA (k=2)
mg
INACAL 1
5,93
5,88 0,23
18,82
18,0 4,6
CENAM 5,93 0,124 18,82 1,54
CESMEC 8,1 1,0 21,7 5,0
INACAL 2 5,79 0,31 17,0 4,8
Table 12 shows the conventional mass correction found by the NMIs and its associated uncertainty for a confidence level of 95,45 %.
Table 12: Conventional Mass correction
Participant laboratory
2 kg 10 kg
Reference
Value
mg
mA
mg
UA
(k=2)
mg
Reference
Value
mg
mA
mg
UA
(k=2)
mg
INACAL 1
0,255
0,21 0,23
-0,27
-1,1 4,6
CENAM 0,255 0,124 -0,27 1,54
CESMEC 2,3 1,0 2,6 5,0
INACAL 2 0,12 0,31 -2,1 4,8
The average value was obtained by the pilot laboratory considering the drift obtained, measuring the mass correction at the beginning and end of the comparison. For this the following equations were used:
2
12 PLPLPL
mmm
(8.1)
where: mPL1: Mass (or conventional mass) correction value measurement 1, by INACAL the pilot
laboratory mPL2: Mass (or conventional mass) correction value measurement 2, by INACAL the pilot
laboratory
2
122
2
32)()(
PLPL
PLPL
mmmumu (8.2)
)()( PLPL mukmU (8.3)
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
7 de 13
Tables 13 and 14 shows the average value of pilot laboratory
Table 13: Mass correction
Pilot laboratory
2 kg 10 kg
mA
mg
U (k=2)
mg
mA
mg
U (k=2)
mg
INACAL 5,84 0,31 17,5 4,8
Table 14: Conventional Mass Correction
Pilot
laboratory
2 kg 10 kg
mA
mg
U (k=2)
mg
mA
mg
U (k=2)
mg
INACAL 0,16 0,31 -1,6 4,8
Uncertainty analysis has been made for all NMIs according to GUM. The uncertainty contributions assigned by the NMIs for each nominal value are shown in tables 15 and 16 respectively, according to the format established in Annex D of the Technical Protocol.
Table 15: Uncertainty contributions reported by the NMIs for 2 kg weight
NOMINAL VALUE 2 kg INACAL 1 CENAM CESMEC INACAL 2
INFLUENCE MAGNITUDE Xi
UNIT u
k=1 u(xi)
u k=1 u(xi)
u k=1 u(xi)
u k=1 u(xi)
Mass of standard, P mg 0,0280 0,057 0,15 0,0800
Drift of standard, δP mg 0,0179 0,0012 0,0294
Volume of standard, Vp cm3 0,0233 0,0234
Density of mass standard 0,0048 0,003881
Volume of sample Vm cm3 0,0233 0,0234
Density of sample m 0,019 0,12
Air density, ρa mg/cm3 0,0056 0,00074 0,004 0,0056
Sensibility of the balance 0,010
Resolution of balance mg 0,0749 0,0041 0,028 0,0687
Repetibility mg 0,0702 0,0033 0,02 0,0548
Weighing process (fitt ing) mg 0,0212 0,0050 0,0902
Reproducibility cm3 0,0022
Combined uncertainty mg 0,115 0,062 0,5 0,156
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
8 de 13
Table 16: Uncertainty contributions reported by the NMIs for 10 kg weight
NOMINAL VALUE 10 kg INACAL 1 CENAM CESMEC INACAL 2
INFLUENCE MAGNITUDE Xi
UNIT u
k=1 u(xi)
u k=1 u(xi)
u k=1 u(xi)
u k=1 u(xi)
Mass of standard, P mg 0,11000 0,60 0,8 0,2200
Drift of standard, δP mg 0,09238 0,29 0,0981
Volume of standard, Vp cm3 0,01764 0,0175
Density of mass standard mg/cm3 0,025 0,019684
Volume of sample Vm cm3 0,47049 0,4674
Density of sample m mg/cm3 0,38 0,019150
Air density, ρa mg/cm3 0,01740 0,0032 0,015 0,0169
Sensibility of the balance mg 0,010
Resolution of balance mg 2,04257 0,0041 0,3 2,0429
Repetibility mg 0,81125 0,0074 0,4 0,8837
Weighing process (fitt ing) mg 0,64847 0,0063 0,5994
Reproducibility cm3 0,0071
Combined uncertainty mg 2,34 0,77 2,5 2,36
Most of the laboratories considered the same influence magnitudes (uncertainty contributions) for the estimation of the uncertainty. This guarantees a greater homogeneity in the determination of the combined uncertainty.
9. Results Analysis
The aim of the comparison is to demonstrate the consistency between the results provided by the laboratories and the reference values. The reference values for this comparison were determined by the CENAM with an expanded uncertainty (k=2). The results were sent directly to the pilot laboratory
(INACAL). The uncertainty values associated with the mass measurements, reported by the participating laboratories, are listed in Table 17.
Table 17
NOMINAL
VALUE
2 kg 10 kg
U (k=2) mg
U (k=2) mg
INACAL 0,31 4,8
CENAM 0,12 1,5
CESMEC 1,0 5,0
9.1 Drift uncertainty Ud
A drift uncertainty Ud for the mass instability of the travelling weight standards is
taken into account with the following equation:
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
9 de 13
2
12
32
PLPL
d
mmkU (9.1)
where: mPL1: Mass (or conventional mass) correction value measurement 1, by INACAL the pilot laboratory mPL2: Mass (or conventional mass) correction value measurement 2, by INACAL the pilot laboratory
Note: The uncertainty due to instability, Ud, was obtained from the measurements made by the pilot laboratory, INACAL, at the beginning and at the end of the comparison.
In order to evaluate the degrees of equivalence of the results given by the participating laboratories, it was necessary to calculate the normalized error of measurement in relation to the reference value, and its associated uncertainty, according to the equation (9.2).
222RdA
RAn
UUU
mmE
(9.2)
En Normalized error
mR The reference value, CENAM
UR The expanded uncertainty of the reference value with k = 2 , CENAM
mA the NMI value
UA the expanded uncertainty of the NMI value with k = 2
9.2 Evaluation of the consistency of the comparison
This dimensionless value should be smaller than 1. It allows comparison of the results of the participants with different uncertainties and allows the appreciation of the consistency of the obtained measurement results ( IEnI ≤ 1).
Table 18, expresses the normalized error calculated according to equation (9.2) from the results of the NMIs for each nominal value in mass.
Table 18
NMIs NOMINAL
VALUE
|𝑬𝒏|
mg
|𝑬𝒏| ≤ 𝟏
INACAL 2 kg
0,3 YES
CESMEC 2,1 NO
INACAL 10 kg
0,2 YES
CESMEC 0,5 YES
Table 18, shows that the IEnI value for CESMEC is greater than 1 for 2 kg .
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
10 de 13
The difference between the laboratory result mA and the reference value m
R is
calculated by:
Di = m
A- m
R (9.3)
This parameter Di is called the difference from the reference value.
Every Di is associated with an uncertainty given by the following formula,
u²(Di) = uA² +ud – uR
²
(9.4)
The same notation is used for both mass and conventional mass results in Sections 10 with the meaning set by the context.
10. Results for mass
10.1 2 kg
The results for the transfer standard of 2 kg are contained in the following table:
Table 19
NMIs NOMINAL
VALUE 𝑫𝒊 mg
𝑼(𝑫𝒊) mg
CENAM
2 kg
0,00 0,124
INACAL -0,09 0,34
CESMEC 2,17 1,01
The values D
n and their corresponding uncertainties U(D
n) (k = 2) are shown in the
following graph 1:
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
11 de 13
Graph 1
10.1 10 kg The results for the travelling standard of 10 kg are given in the following table:
Table 20
NMIs NOMINAL
VALUE 𝑫𝒊 mg
𝑼(𝑫𝒊)
mg
CENAM
10 kg
0,0 1,54
INACAL -1,3 5,1
CESMEC 2,9 5,2
The values D
n and their corresponding uncertainties U(D
n) (k = 2) are shown in the
following graph:
Graph 2
11. Results for conventional mass
11.1 2 kg
The results for the travelling standard of 2 kg are given in the following table:
Table 21
NMIs NOMINAL
VALUE 𝑫𝒊 mg
𝑼(𝑫𝒊)
mg
CENAM
2 kg
0,00 0,124
INACAL -0,09 0,34
CESMEC 2,05 1,01
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
12 de 13
The values Dn
and their corresponding uncertainties U(Dn) (k = 2) are shown in
the following graph:
Graph 3
11.2 10 kg
The results for the travelling standard of 10 kg are given in the following table:
Table 22
NMIs NOMINAL
VALUE 𝑫𝒊 mg
𝑼(𝑫𝒊)
mg
CENAM
10 kg
0,00 1,54
INACAL -1,3 5,1
CESMEC 2,9 5,2
The values Dn
and their corresponding uncertainties U(Dn) (k = 2) are shown in
the following graph:
SUPPLEMENTARY COMPARISON CALIBRATION OF MASS AND
CONVENTIONAL MASS SIM.M.M-S13
13 de 13
Graph 4
12. Conclusions
The comparison was successful. There was no significant deviation in the measured mass of any of the travelling weight standards. The participant results were reported according to the protocol. For laboratory CESMEC, the normalized error for 2 kg weight is greater than 1, this is shown in the table 18. In graphs 1, 2, 3 and 4 are shown the difference of the participating laboratories with respect to the reference laboratory, CENAM, and their expanded uncertainties with k=2.
13. Acknowledgments
We would like to thank to Gregorio Alvarez Clara from CENAM for his valuable contribution on the measurements of the travelling weight standards involved in this comparison.
14. References
[1] OIML R111, Weights of classes E1, E2, F1, F2, M1, M2, M3, 2007 [2] Revised formula for the density of moist air (CIPM-2007); Metrologia 45 (2008)
149-155. [3] JCGM 100:2008, Evaluation of measurement data - Guide to the expression of
uncertainty in measurement, 2008. [4] http://www.bipm.org/utils/common/documents/CIPM-MRA/CIPM-MRA-D-05.pdf [5] EA-4/02, Expression of the Uncertainty of Measurement in Calibration, 1999
http://www.european-accreditation.org/n1/doc/ea-4-02.pdf [6] Wolfang Wöger, Remarks on the En – Criterion Used in Measurement
Comparison, PTBMitteilingen 109 1/99, Internationale Zusammenarbeit [7] ISO/IEC 17043:2010, Conformity assessment -- General requirements for
proficiency testing.