Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
2014
International Summer Water Resources Research School
Dept. of Water Resources Engineering, Lund University
Study of estrogenic compounds in water using
POCIS
By
Cajsa-Lisa Ivarsson
The 8th Ling Feng Summer Research School
Xiamen University 2014
1
Abstract
Different passive measuring methods have been generated during the last decades in order to develop
the ways of water sampling beyond grab sampling. One of these is the Polar Organic Chemical
Integrative Sampler (POCIS). A POCIS contains an absorbent which in water will absorb a time weighted
average of the concentration of different types of polar compounds over that time span. This study
focus on measurements of estrogenic compounds with POCIS. The concentration of estrogenic
compounds are increasing in our environment and has been shown to have a negative effect on living
species. The reproduction system seems to be affected and especially the one of male fishes.
The study contains the uptake dynamics of estrogenic compounds using POCIS, an analyze of the
sampling rate of the different estrogenic compounds and finally a comparison of the sampling rate for
each estrogenic compound under different concentrations exposure.
15 POCIS device were left in water of 1 µg/L concentration of estrogenic compounds for 1, 3, 5, 7, and
10 days. 3 plates were gathered for each time scale. 3 plates additional POCIS were placed in water of 2
µ/L concentration of estrogenic compounds for 10 days to see if the concentration of the water will
affect the sampling rate. The samples were then washed and processed in a High performance liquid
chromatography mass spectrometer, HPLC-MS.
The result showed that the estrogenic compounds have a linear uptake in 10 days exposure. The
estrogenic compound E3 had been absorbed in the highest concentration after 10 days, had the highest
sampling rate. The sampling rate of estrogenic compounds is not affected by the exposure
concentration and they seems well fit to be measured with this kind of equipment.
Keywords: POCIS, Estrogenic compounds, water samples, HPLC-MS
Supervisor Yue Yun, me and my fellow students Huiping Zhang and Dan Lin outside the Department of Environmental Science
The 8th Ling Feng Summer Research School
Xiamen University 2014
2
Table of content
Abstract ........................................................................................................................................................ 1
Table of content ........................................................................................................................................... 2
1. Introduction .......................................................................................................................................... 3
2. Objectives ............................................................................................................................................. 3
3. Background and theory ........................................................................................................................ 4
3.1 POCIS .................................................................................................................................................. 4
3.1.1 Function ....................................................................................................................................... 4
3.1.2 Performance Referent Compound (PRC) .................................................................................... 5
3.1.3 Calculations ................................................................................................................................. 6
3.2 Estrogens ............................................................................................................................................ 7
4. Method and material ........................................................................................................................... 8
4.1 Material .............................................................................................................................................. 8
4.2 Preparing of POCIS device and monitoring procedure ...................................................................... 9
4.3 Pretreatment of POCIS samples ....................................................................................................... 10
4.3 Analyze of absorption ....................................................................................................................... 11
5. Results ................................................................................................................................................ 12
5.1 Time series uptake data ................................................................................................................... 12
5.2 Calculation of the sampling rate, RS ................................................................................................. 14
5.3 Compare RS after 10 days with different concentrations ................................................................. 14
6. Discussion ........................................................................................................................................... 14
7. Sources of errors ................................................................................................................................ 15
8. Conclusions ......................................................................................................................................... 15
9. Acknowledgments .............................................................................................................................. 16
10. References ...................................................................................................................................... 17
11. Appendix ............................................................................................................................................... 18
The 8th Ling Feng Summer Research School
Xiamen University 2014
3
1. Introduction
The study of water has always been of interest and different aquatic system surround us in our everyday
life as rivers, lakes, ponds and so on. If a pollutant reaches an aquatic system it will spread wildly and
quick. Most aquatic studies used to be based on grab sampling which contains of collecting bottle
samples to bring back to the lab or a 24 hours average of water at a given time. The use of grab sampling
are sometimes not convenient in the study of pollutants since these are mostly found as trace amounts
and the concentration usually vary over time. This conclude that with grab samples pollutants can easily
be overlooked or wrongly estimated in concentration (Grabic, Jurcikova, Tomejova, Ocelka, & Halirova,
2009).
Over the past two decades passive sampling methods as Semipermeable membrane devices (SPMDs)
and more recently the Polar Organic Chemical Integrative Sampler (POCIS) has been developed to
overcome some of these difficulties. These passive samplers will produce a time weight average (TWA)
of the concentration of lipophilic and hydrophilic chemicals, respectively. After positioned for a time in
contaminated water, they will obtain the same concentration of the waterborne contaminants, on ultra-
trace detection level (Mazzella, o.a., 2009). Estrogen is a pollutant which is increasing in our
environment and can be hazardous in smaller amounts than what can be revealed by grab sampling.
Detection of estrogens in the environment has raised concerns in recent years because of their potential
to affect both wildlife and humans (Caldwell, o.a., 2009). Among other things they are most likely to
affect the endocrine system of living species. This study will focus on the procedure and preciseness of
measurements of estrogens compounds in water using the passive sampler POCIS.
2. Objectives
The main objectives for this study was to investigate how the concentration in water of estrogenic
compounds, BPA, DES, HEX, E1, E2, E3 and EE2, can be measured with POCIS. This was followed through
with three criteria:
1. To study the uptake dynamics of estrogenic compounds using POCIS. That is carried through by
plotting the absorbed concentration of the different estrogenic compounds after the chosen
time spans of 1, 3, 5, 7 and 10 days in water with a concentration of 1 µg/L estrogenic
compounds.
2. To analyze the sampling rate (RS) for each estrogenic compounds in quite water with a
concentration of 1µg/L estrogenic compounds. The sampling rate is the rate of the volumes of
water that are purified per units of time (L/days) which also corresponds with the absorption
time for the different compounds.
3. To compare the sampling rate for the estrogenic compounds under different concentration
exposures. This is undertaken to see if the concentration effects the sampling rate.
The 8th Ling Feng Summer Research School
Xiamen University 2014
4
3. Background and theory
3.1 POCIS
3.1.1 Function The Polar Organic Chemical Integrative Sampler (POCIS) were developed by researchers with the United
States Geological Society (USGS) in Columbia, Missouri. It is used in water sampling to monitor
hydrophilic compounds such as medical drugs, surfactants (nonylphenols), triazines and caffeine
(Mi´ege, o.a., 2011). There foremost quality is to present a time weighted average of pollutant in
contrast to grab sampling which are only showing the pollutant which existed in that blink of an eye
when the sample is gathered. With this passive sampler is also possible to detect smaller amounts of a
matter than what is possible to find with grab samples.
The actual monitoring system consists of two stainless steel plates, two isolation silicon plates, two
membrane and a powder for absorption of the pollutants is placed in the middle (Figure 1).
Figure 1. The different parts of the POCIS (Mi´ege, o.a., 2011)
The type of absorbent can vary depending on which matter that is to be traced. It is beneficial to use an
absorbent with relatively high surface area (>∼300 m2/g) to get a linear uptake (Mazzella, o.a., 2009).
For this study the absorbent Oasis-HLB where used which has a surface area of around 800 m2/g and is
common to use for measurements of estrogen.
The POCIS are to be submerged in an aquatic environment. When in water the hydrophilic membrane
acts as a filter and the absorbent as an accumulator of the contaminants. The process of accumulation in
the POCIS is essentially adsorption into the internal adsorbent after contaminants passively diffuse
through the membrane. The sampling rate differs between contaminants and is also effected by the
sampler design and environmental conditions which are described further down. The time for the
absorption to receive the same concentration of a matter as the surrounding water can take from hours
to days. Figure 2 shows the faces that take place as the absorbent reaches the equilibrium regime. This
The 8th Ling Feng Summer Research School
Xiamen University 2014
5
study is not made during enough time to reach the equilibrium regime and will only show the
integrative kinetic regime in the first part of the graph.
Figure 2. How the sample reaches the same concentration as the surrounding water (Mi´ege, o.a., 2011)
3.1.2 Performance Referent Compound (PRC)
POCIS can be used for measurements both in quiet water; in-lab, as in moving water i.e. a river or lake,
so called in situ. Detection of trace levels from polar compounds in surface water and wastewater has
been proved to give reliable results (Mi´ege, o.a., 2011). When using these devices for measurements in
situ is it important to bear in mind the effects of environmental factors such as flow and temperature,
why these should be measured at regular intervals (Harman, Reid, & Thomas, 2011). Furthermore, the
main physico-chemical parameters of the target water should be measured (i.e., dissolved organic
carbon, conductivity, pH, particulate suspended matter) as they can trace water quality changes over
time and also might influence the POCIS-available fraction (Mi´ege, o.a., 2011).
An approach to calculate the effect from these outsider sources is to use of Performance Referent
Compound (PRC) the dissipation of which is equally affected by changes in deployment conditions
(Harman, Reid, & Thomas, 2011). A PRC is a compound which is added to the POCIS during construction
and are gradually lost to the surrounding water over time. Determination of the lost amount of PRC
provides an environmental adjustment factor to correct laboratory-derived sampling rates RS for the
site-specific environmental factors. Initial studies indicate careful selection of PRC. The absorbent is also
used to allow measureable loss of chemicals. This study aimed to include PRC but there was not enough
time to carry that through.
The 8th Ling Feng Summer Research School
Xiamen University 2014
6
3.1.3 Calculations When the exposure time is over the adsorbents are collected from the POCIS and the volumes of
absorbed compounds are extracted and detected. The mean contaminant concentration of the water is
then calculated as described in equation (1):
𝐶𝑤𝑎𝑡𝑒𝑟 =𝐶𝑃𝑂𝐶𝐼𝑆 𝑀𝑃𝑂𝐶𝐼𝑆
𝑅𝑆 𝑡 (1)
Cwater – mean contaminant concentration (over the sampling period) in the ambient water (ng/L)
CPOCIS – concentration in the POCIS (ng/g)
MPOCIS – mass of absorbent phase in the POCIS (g)
RS – sampling rate (L/d) which corresponds to the volume of water purified per unit of time
t – total exposure time (days)
The PRC samples are used to correct the RS value calculated in equation (1). RS corrected is calculated by
equation (2) and (3):
First the elimination constant, ke for both in lab and in situ is calculated:
𝑙𝑛𝑚𝑡
𝑚𝑜= 𝑘𝑒𝑡 (2)
mt – mass of PRC after time t (ng)
m0 – mass of PRC at time t = 0 (ng)
ke – elimination constant
t – total exposure time (days)
The different between ke in lab and in situ gives the corrected RS:
𝑅𝑆 𝑐𝑜𝑟𝑟𝑒𝑐𝑡𝑒𝑑 =𝑘𝑒 𝑖𝑛 𝑠𝑖𝑡𝑢
𝑘𝑒 𝑖𝑛 𝑙𝑎𝑏𝑏∗ 𝑅𝑆 (3)
ke in lab – elimination constant calculated in the lab
ke in situ – elimination constant calculated in the field
RS – sampling rate (L/d), calculated in equation (1)
Finally, Rs corrected value is substituted for Rs in equation (1) in order to obtain the corresponding time-
weighted average Cwater.
The 8th Ling Feng Summer Research School
Xiamen University 2014
7
3.2 Estrogens
The female sex hormone estrogen is found naturally in humans and animals. It is primarily to affect the
development, maturation and function of the female reproduction system but in recent years it has
been showing up more frequently all over the environment. The two major natural occurring of these
hormone steroids are estrone (E1) and estradiol (E2). There are today also a lot of man-mad estrogens
such as 17a-ethynylestradiol (EE2), diethylstilbestrol (DES) etc. which are used in drugs like birth control
pills or to ease troubles with menopause and also in industrial products like plasticizers and surfactants,
and in agriculture (University of Wisconsin Aquatic Sciences Center, 2007). A major part of the steroids
are used in livestock to estrous cycle, treat reproductive disorders and induce abortion and will follow
the livestock urine to water (Kralchevska, Milanova, Hristova, & Todorovsky, 2013).
Figure 3. Molecular structures of 17β-estradiol (E2, a), 17α-ethynylestradiol (EE2, b), estrone (E1, c) and estriol (E3, d) (Kralchevska, Milanova, Hristova, & Todorovsky, 2013)
Estrogens are included among the Endocrine Disrupting Compounds (EDC), which has received
considerable attention the past decades since they in 1994 were first detected to affect the fishes by a
wastewater treatment plant in the UK. EDCs are defined as exogenous substances or mixtures that alter
functions of the endocrine system and consequently cause adverse health effects in an intact organism,
its offspring, or (sub)populations (Grabic, Jurcikova, Tomejova, Ocelka, & Halirova, 2009).
Both natural and synthetic EDCs that end up in the environment have been proved to produce harmful
effects in aquatic organisms, such as feminization of the males and hermaphroditism causing
reproductive disturbance to humans and wildlife. Especially the reproduction system of male fishes
seems to be affected. These compounds can be extremely potent even at low concentrations; for
instance, less than 1 ng/l EE2 can induce vitellogenin (egg yolk protein usually associated with adult
females) production in male rainbow trout and 4 ng/l caused male fathead minnows to fail to develop
normal secondary sexual characteristics (Kralchevska, Milanova, Hristova, & Todorovsky, 2013).
The 8th Ling Feng Summer Research School
Xiamen University 2014
8
There are ways of cleaning the wastewater from most of the EDCs, for example with sand infiltration
aerobic pretreatment (University of Wisconsin Aquatic Sciences Center, 2007). The amounts of these
matters differs a lot over the world is also known to vary significantly over time. As a consequence it is
difficult to acquire an accurate picture of how much estrogen the environment accumulate over a longer
time-period (Vermeisster, Burkhardt-Holm, Körner, Schönenberger, & Suter, 2005). There is there for of
most important to find an easy and reliable way to measure these concentrations to find leakage in
early stage so that quick action can be taken to protect our ecosystems.
4. Method and material
4.1 Material
The following material was used in the study:
21 POCIS plates á diameter? Each consisting of 2 stainless steel plates and two isolating silicon
plates, and 3 screws with mutts
Membrane with diameter of 4.7 cm and poor size of 0.1µm. 2 á POCIS
HLB absorbent, á 100 mg/POCIS
Solution of estrogens 1 ppm of BPA, DES, HEX, E1, E2, E3 and EE2, which was diluted to a
concentration of 1 µg/L and 2 g/L
Methanol (CH3OH) for cleaning of the equipment, activation of the absorbent and solvent
Chemical regent, NaN3, to prevent bacterial growth in the beakers
Beakers to store the POCIS plates with the solvent of estrogen
Nonylphenol for internal standard
An injector and acetate, hexane and dichloromethane to clean the injector
Ultrasonic cleaner
Rotary evaporator
Pear shaped bottles
0.22 μm membrane to remove any left of the absorbent when transferring to bottles
High performance Liquid Chromatography Mass Spectrometer, HPLC-MS
The 8th Ling Feng Summer Research School
Xiamen University 2014
9
4.2 Preparing of POCIS device and monitoring procedure
Each part of the POCIS device was first cleaned with an Ultrasonic cleaner and there after wiped with
methanol to dissolve any trace of estrogens that could be there from us humans for instant. They were
put together and 100 mg of the HLB absorbent was scaled and placed in the center of each between the
membranes. The membranes had different surfaces, and the shine side where to be placed facing out.
The two plates where finally tightly screwed together. 18 plates where produced and placed for 10
minutes in methanol to activate the absorbent.
When the POCIS devises where finish they were places in totally 4 beakers and the solution of dissolved
estrogen was added. One beaker would contain 3 plates and the solution with a concentration of 2 µg/L
of each estrogenic compound. The rest of the plates where spread out in the remaining beakers and the
1 µg/L solution was poured in until the plates where fully covered. The chemical regent NaN3 which will
prevent bacterial growth was added to all of the beakers, and the containers were covered with
aluminum foil. At last they were put in a dark space to prevent any estrogen being destroyed by the
sunlight.
Since the POCIS will absorb the estrogens by diffusion the solution needed to be changed daily to keep a
steady estrogen concentration at 1 µg/L. Three plates at a time were removed after 1, 3, 5, 7 and 10
days. The three plates in the 2 µg/L PRC container were removed after 10 days.
After a weak there were recognized that no PRC had been prepared. 4 more samples were produced
with PRC which contained of E2- d4, EE2- d3 and BPA – d10, and sorbent (HLB). These matter were
chosen since they represent the three different types of estrogens (natural, most common man-made
and others). Since the PRC were put in so late, they will not be used in this study.
Figure 4. The material for making POCIS Figure 5. Finish POCIS in exposure
The 8th Ling Feng Summer Research School
Xiamen University 2014
10
4.3 Pretreatment of POCIS samples
When all the plates had been taken out they were dried and then
opened so that the absorbent between the membranes could be
removed and placed in sampling tubes. The pretreatment was
preformed to separate the estrogenic compounds from the
absorbent.
To detect how much of the compounds that is lost during the
pretreatment, an internal standard is added. For this project
Nonylphenol was used which is a type of estrogenic compound
that is not inncluded in this study. The internal standards is
added in a known amount and afterward detected to get a value
of how much is lost during the pretreatment. The internal
standard were added with an injector which was washed with
ethyl acetate, hexane and dichloromethane in between each
matter.
10 mL methanol was added to each sampling tube and they were
placed in an Ultrasonic cleaner for 15 minutes, Figure 6. The estrogenic compounds from each sample
were to dissolve in the methanol which was now pipetted over to a pear-shaped bottle. This procedure
was repeated with 5 mL and 5 mL methanol respectively.
A rotary evaporator was used to remove the methanol. Each pear-shaped bottle now contained only the
gathered estrogenic compounds. 1 mL methanol were added to it dissolve the estrogenic compounds
and a 1 ml pipette was used to collect it all. The solution was the transferred to a small bottle while
filtration through 0.22 μm membrane to remove any left of the absorbent, Figure 7.
The samples were then refrigerated to await determination.
Figure 7. Samples of estrogenic compounds
Figure 6. Samples in Ultrasonic cleaner
The 8th Ling Feng Summer Research School
Xiamen University 2014
11
4.3 Analyze of absorption
The measurements of the amounts of estrogenic compounds in the samples were made in a High
performance Liquid Chromatography Mass Spectrometer, HPLC-MS. Where the HPLC separates the
estrogens and the MS detect them. 6 more bottles with known concentrations of estrogenic compounds
had also been prepared to determine if the machine measured correctly. The concentration of the
estrogenic compounds in these bottles were 1 ppb, 10 ppb, 100 ppb, 200 ppb, 500 ppb and 1 ppm.
The collected data from the MS was sorted in the program Agilient MassHunter Quantative Analyzis.
This program corrected the samples after the added amount of internal standard. It presented the data
of the average concentration of each estrogenic compound absorbed by the sorption after 1, 3, 5, 7, and
10 days (Table 2 page 10) and the data from the 6 bottles that were made to see the accuracy of the
total measurement. The data was then analyzed with Excel.
Figure 8. The samples are being processed in the HPLC-MS
The 8th Ling Feng Summer Research School
Xiamen University 2014
12
5. Results
The data from the measurements of the 6 bottles of known concentration is shown in Table 1. These
were used to see how well the machine preformed. If the samples would have been made by perfection
and the HPLC-MS would measure with absolute accuracy - the number for each standard concentration
would equal the number for each estrogenic compound.
Table 1. The measurements of the known standard concentrations
Standard
conc. (ppb)
BPA
(ppb)
HEX
(ppb)
E1
(ppb)
E2
(ppb)
E3
(ppb)
EE2
(ppb)
DES
(ppb)
1 2,6 2,0 3,0 - - 14,3 2,0
10 15,6 14,2 20,4 11,4 - 20,6 14,2
100 95,4 95,5 97,5 101,3 94,9 95,4 95,5
200 182,3 189,1 192,1 193,1 232,2 193,3 189,1
500 502,5 505,2 472,5 506,0 589,2 471,3 505,2
1000 1002,7 - 1015,5 998,2 949,8 1016,0 -
The data from the comparing samples where plotted to see if they were linear and there for would
indicate a trustworthy measurement from the machine. The R2 from these graphs are presented in Table
2 and their graphs are in appendix.
Table 2. R2 values of the corresponding estrogen concentrations are all close to 1 wich indicate a good correlation between measured an known value.
BPA HEX E1 E2 E3 EE2 DES
R2 0,999 0,999 0,999 0,999 0,985 0,998 0,999
5.1 Time series uptake data The measured amount of the different estrogenic compounds that had been absorbed during the test
period of totally 10 days is presented in Table 3. As can be seen the estrogenic compound E3 was not
absorbed at all in measurable amounts in the start but had the highest concentration after 10 days.
The 8th Ling Feng Summer Research School
Xiamen University 2014
13
Table 3. Concentration of the different estrogenic compounds that had been absorbed after 1, 3, 5, 7, and 10 days.
Days from start: BPA
(ppb)
DES
(ppb)
HEX
(ppb)
E1
(ppb)
E2
(ppb)
E3
(ppb)
EE2
(ppb)
1 4,2 6,3 5,7 6,2 2,2 - 10,6
3 7,6 6,6 6,0 14,6 7,6 - 15,2
5 11,9 6,9 6,7 21,0 17,9 24,8 20,3
7 17,7 7,1 7,5 27,4 30,2 76,9 26,2
10 22,5 7,6 8,5 41,3 52,1 145,0 36,9
The data was plotted to get the time series uptake data of POCIS for the different estrogenic
compounds, Figure 9. As can be seen in Figure 9 when looking at the R2 values the concentration of
estrogenic compounds found in the sample follow an almost linear increase during the first ten days.
Figure 9. Time series uptake for each estrogenic compound. As can be seen the E3 has the fastest increase.
R² = 0,9907
R² = 0,9877R² = 0,9877
R² = 0,9741
R² = 0,9986
R² = 0,9909
-100
-50
0
50
100
150
200
0 2 4 6 8 10 12
Co
nce
ntr
atio
n (
pp
b)
Days
Time series uptake data of POCIS for Estrogenic compounds
BPA 结果 DES 结果 HEX 结果 E1 结果 E2 结果 E3 结果 EE2 结果
The 8th Ling Feng Summer Research School
Xiamen University 2014
14
5.2 Calculation of the sampling rate, RS
The sampling rate for each estrogenic compound was calculated as described in equation (1) page 6.
Following data were used for the calculations:
Cwater = 1 µg/L
CPOCIS = see Table 1, page 10
MPOCIS = 0, 01 g
t = total exposure time in days
Table 4. The sampling rate (RS) for each estrogenic compound. Described in average value and mean distribution.
RS BPA DES HEX E1 E2 E3 EE2
Median
distribution (L/d)
0,0033
±0,001
0,0034
±0,003
0,0031
±0,003
0,0051
±0,001
0,0037
±0,002
0,0097
±0,005
0,0071
±0,007
Average (L/d) 0,0028 0,0023 0,0022 0,0047 0,0036 0,0101 0,0054
The result is presented in Table 4. Both mean and median distribution is shown and they match fairly
well except for EE2.
5.3 Compare RS after 10 days with different concentrations
Table 5. Sampling rate (days/L) for each estrogenic compound after 10 days in 1 µg/L in compare to 2 µg/L
RS (d/L) BPA DES HEX E1 E2 E3 EE2
1 µg/L 0,0023 0,0008 0,0008 0,0041 0,0052 0,0145 0,0037
2 µg/L 0,0023 0,0004 0,0005 0,0041 0,0052 0,0141 0,0036
As can be seen in Table 5 the sampling rate is the about same in both concentrations except for DES and
HEX where the values are around the double.
6. Discussion
The R2 values from the 6 samples with known concentrations were close to 1 so the graphs are
considered to be linear and the equation were approaching x=y. This indicates that the measurements
The 8th Ling Feng Summer Research School
Xiamen University 2014
15
from the machine are very close to the known values, and instrument seamed therefore to have been
working correct and the result should be considered credible.
After 10 days of exposure to water, none of the compounds hade reach their equilibrium stage, where
they will not continue to increase in concentration. E3 had been absorbed in the highest concentration
after and it make sense that it also had the highest sampling rate of 0, 0097 or 0, 0101 L/day. During the
first 3 days no E3 at all was traced in the absorbents. This was the same in the 6 bottles with known
concentrations (Table 1) which indicates that E3 is difficult to measure at low levels. EE2 is the
compound that varies the most in sampling rate which can be seen by the large span of values (±0,007)
and it also differs the most between mean and median value.
The concentration had increased for DES and HEX but not in reverse correlation with increased
concentration of the water. For the other compounds no such increase was detected. From earlier
studies it is concluded that the absorption of estrogenic compounds is not concentration dependent. It
should be very hard to measure a concentration dependent compound in situ since that would make it
impossible to produce a standard data for that compound. Since we detected so low concentration it
seems possible that the increase is just due to wrong measurements.
7. Sources of errors
There is possible to believe that some errors could be exciting due to mistake during the making of the
samples. There were a lot of different small measurements that had to be taken and the solutions were
transferred many times to different containers which could result in losses. To reduce this kind of
influence 3 plates were used for every time span but unfortunately a lot of the membrane were broken
during the drying and which resulted in samples that could not be used.
Since data from the 6 samples of known concentration were close to linear it is likely to believe that the
amount measured with the machine were accurate.
8. Conclusions
The estrogenic compounds, BPA, DES, HEX, E1, E2, E3 and EE2, have a linear uptake in 10 days exposure.
It take more than 10 days for the estrogenic compounds to reach their equilibrium stage. E3 had been
absorbed in the highest concentration after 10 days, had the highest sampling rate and is most likely to
reach equilibrium stage in the shortest time. The measurements of the bottles with known
concentration gave a content result and showed that the HPLC-MS seams to measure with good
accuracy. The sampling rate of estrogenic compounds shown not to be concentration dependent and
seems well fit to be measured with this kind of equipment. For further knowledge the study would
preferably be made until all compounds reaches their equilibrium stage.
The 8th Ling Feng Summer Research School
Xiamen University 2014
16
9. Acknowledgments
Firstly I would like to thank my Chinese supervisor Yue Yun, for being so kind and for the excellent
support she gave me throughout this project. I want to give my best wishes you to my fellow Chinese
students Huiping Zhang and Dan Lin who’ve been great partners during the laboratory work and to
Professor Xinhong Wang for taking initiative for this study. I would also like to thank all the Chinese
students who made me feel at home in Xiamen and have been helping out so much. I want to send a
special thought to the eight Swedish students for sharing this experience and especially to Josefin
Lindström who never cease making my day. A great thanks to my Swedish professor Linus Zhang and
Lund University who made it possible for me to attend this course. I’m also very grateful to all the staff
at Xiamen University for taking care of us during our stay, especially the guards, and the staff at CEE Café
whom have been offering a wonderful place to study. And finally, I would like to thank Thyréns and
Sweco who made it possible for us to travel to China with their foundations.
All the Swedish students of Ling Feng Swedish Summer Research School 2014. From left; Beatrice Nordlöf, Tove Juhl Andersen, Ida Arvidsson, Josefin Lindström, Sofia Akhlaghi, Cajsa-Lisa Ivarsson (me), Karin Lindberg and Erik Sönegård
The 8th Ling Feng Summer Research School
Xiamen University 2014
17
10. References
Caldwell, D. J., Mastrocco, F., Nowak, E., Johnston, J., Yekel, H., Pfeiffer, D., . . . Anderson, P. D. (2009).
An Assessment of Potential Exposure and Risk from Estrogens in Drinking Water. Environ Health
Perspect. Mar 2010; 118(3), 338–344.
Grabic, R., Jurcikova, J., Tomejova, S., Ocelka, T., & Halirova, J. (2009). Passive sampling methods for
monitoring edocrine distruption in the Svartka and Svitava rivers in the Czech Republic. Environmental
Toxicology and Chemistry, Vol. 29, No. 3, 550–555.
Harman, C., Reid, M., & Thomas, K. V. (2011). In Situ Calibration of a Passive Sampling Device for
Selected Illicit Drugs and Their Metabolites in Wastewater, And Subsequent Year-Long Assessment of
Community Drug Usage. Environ. Sci. Technol. 2011, 45, 5676–5682.
Kralchevska, R. P., Milanova, M. M., Hristova, I. L., & Todorovsky, D. S. (2013). Some endocrine
disrupting compounds in the environment and possibilities for their removal / degradation. Bulgarian
Chemical Communications, Volume 45, Number 2, 131 – 143.
Mazzella, N., Lissalde, S., Moreira, S., Delmas, F., Mazellier, P., & Huckins, J. N. (2009). Evaluation of the
Use of Performance Reference Compounds in an Oasis-HLB Adsorbent Based Passive Sampler for
Improving Water Concentration Estimates of Polar Herbicides in Freshwater. Environ. Sci. Technol. 2010,
14, 1713–1719.
Mi´ege, C., Budzinski, H., Jacquet, R., Soulier, C., Peltec, T., & Coquerya, M. (2011). Polar organic
chemical integrative sampler (POCIS): application for monitoring organic micropollutants in wastewater
effluent and surface water. J. Environ. Monit., 2012, 14, 626–635.
University of Wisconsin Aquatic Sciences Center. (2007, spring). Care for a glass of H2O? Aquatic
Sciences Chronicle, p. 3.
Vermeisster, E., Burkhardt-Holm, P., Körner, O., Schönenberger, R., & Suter, M. (2005). Characterization
of Environmental Estrogens in River Water Using a Three Pronged Approach: Active and Passive Water
Sampling and the Analysis of Accumulated Estrogens in the Bile of Caged Fish”,. ENVIRONMENTAL
SCIENCE & TECHNOLOGY, 39, 21, 8191-8198.
The 8th Ling Feng Summer Research School
Xiamen University 2014
18
11. Appendix
The graphs from the bottles of know concentration measured with the HPLC-MS.
y = xR² = 0,9999
0
500
1000
1500
0 200 400 600 800 1000 1200Mea
sure
d c
on
c.
Supposed concentration
E2
Linjär ()
y = 1,004x - 2,8718R² = 0,9985
-500
0
500
1000
1500
0 500 1000 1500Mea
sure
d c
on
c.
Supposed concentration
E1
Serie1
Linjär (Serie1)
y = 0,9782x + 15,77R² = 0,9853
0
500
1000
1500
0 500 1000 1500
Mea
sure
d c
on
c.
Supposed concentration
E3
Serie1
Linjär (Serie1)
The 8th Ling Feng Summer Research School
Xiamen University 2014
19
y = 1,0048x - 1,7756R² = 0,999
-200
0
200
400
600
0 200 400 600Mea
sure
d c
on
c.
Supposed concentration
DES
Serie1
Linjär (Serie1)
y = 1,0048x - 1,7756R² = 0,999
-200
0
200
400
600
0 200 400 600Mea
sure
d c
on
c.
Supposed concentration
HEX
Serie1
Linjär (Serie1)
y = 1,004x - 2,8718R² = 0,9995
-500
0
500
1000
1500
0 5 0 0 1 0 0 0 1 5 0 0MEA
SUR
ED C
ON
C.
SUPPOSED CONCENTRATION
BPA
BPA 结果
Linjär (BPA 结果)
y = x - 1E-13R² = 0,9981
0
500
1000
1500
0 500 1000 1500Mea
sure
d c
on
c.
Supposed concentration
EE2
Serie1
Linjär (Serie1)