gke Steri-Record ® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact Process Challenge Devices (Bio-C-PCDs ® ) Application The gke Steri-Record ® Mini-Bio-Plus self-contained biological indicators (SCBI) are used for validation and routine monitoring of steam-, ethylene oxide-, formaldehyde- and hydrogen peroxide/plasma- sterilization processes. After sterilization the SCBIs can be incubated in the CSSD by the user without using an external microbiological laboratory. The Mini-Bio-Plus SCBIs can be used for routine monitoring inside a package, container, or inside a gke Steri-Record ® Bio-Compact Process Challenge Device (Bio-C-PCD ® ) without losing sensitivity. They have been designed with a minimal internal vol- ume. The internal volume of conventional SCBIs is too large and would dramatically decrease the sensi- tivity of class 2 indicators and cannot be used in hol- low load PCDs to check internal lumens of MIS in- struments. gke offers for the first time a special test system worldwide consisting of a hollow process challenge device (Bio-C-PCD ® ) with self-contained biological indicators inside. The combination of the gke Steri- Record ® Bio-C-PCD ® with the Mini-Bio-Plus SCBI is the first class 2 indicator which simulates minimal invasive surgical (MIS) hollow instruments. In addi- tion Bio-C-PCD ® s with different air removal charac- teristics are now available. The sensitivity of these Bio-C-PCD ® s can be utilized to meet the requirement of the load. Bio-C-PCD ® s are available with five different air re- moval and steam penetration characteristics. The validation of the Bio-C-PCD ® according to the load can be achieved by using the test method de- scribed in DIN 58921. gke Bio-C-PCD ® s can be used in all four steriliza- tion processes mentioned above, if the correct SCBIs are inserted. There are various hydrogen peroxide/plasma steriliza- tion processes in the market with different penetration characteristics of the sterilization agent. Depending on the process used and on the load configuration an appropriate Bio-C-PCD ® should be selected. Product description The gke Steri-Record ® Mini-Bio-Plus SCBI is de- signed using a plastic vial with a small internal vol- ume containing a biological indicator spore plate and a glass ampoule with a growth medium and pH- indicator inside. For steam-, ethylene oxide- and for- maldehyde-sterilization processes, filter paper is used as a carrier and as a closing filter below the cap. For hydrogen peroxide/plasma-sterilization processes plastic material is used. For a better recognition of the different SCBI versions they each have different col- oured caps (see table). The labels of all SCBIs for steam sterilization proc- esses contain a class 1 chemical indicator according to EN ISO 11140-1 to check, if the SCBI has passed a sterilization process. The self-adhesive label may be detached from the SCBI and adhered on a documenta- tion sheet. An additional chemical indicator class 5 according EN ISO 11140-1 allows that the result of steam ster- ilization processes can be instantly read at the end of the sterilization process. Therefore, it is not necessary to wait for the result of the SCBI incubation since class 5 indicators are equivalent to biological indica- tors according to the above standard. EO VH2O2 STEAM FORM

STEAM gke Steri-Record Mini-Bio-Plus EOfoxtrot-25.srv.hosting.fi/files/Mini-Bio-plus.pdfgke Steri-Record® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact

  • Upload
    others

  • View
    22

  • Download
    0

Embed Size (px)

Citation preview

Page 1: STEAM gke Steri-Record Mini-Bio-Plus EOfoxtrot-25.srv.hosting.fi/files/Mini-Bio-plus.pdfgke Steri-Record® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact

gke Steri-Record® Mini-Bio-Plus

self-contained biological indicators (SCBI)

and Bio-Compact Process Challenge Devices (Bio-C-PCDs®)

Application

The gke Steri-Record® Mini-Bio-Plus self-contained

biological indicators (SCBI) are used for validation

and routine monitoring of steam-, ethylene oxide-,

formaldehyde- and hydrogen peroxide/plasma-

sterilization processes. After sterilization the SCBIs

can be incubated in the CSSD by the user without

using an external microbiological laboratory.

The Mini-Bio-Plus SCBIs can be used for routine

monitoring inside a package, container, or inside a

gke Steri-Record® Bio-Compact Process Challenge

Device (Bio-C-PCD®) without losing sensitivity.

They have been designed with a minimal internal vol-

ume. The internal volume of conventional SCBIs is

too large and would dramatically decrease the sensi-

tivity of class 2 indicators and cannot be used in hol-

low load PCDs to check internal lumens of MIS in-

struments.

gke offers for the first time a special test system

worldwide consisting of a hollow process challenge

device (Bio-C-PCD®) with self-contained biological

indicators inside. The combination of the gke Steri-

Record® Bio-C-PCD® with the Mini-Bio-Plus SCBI

is the first class 2 indicator which simulates minimal

invasive surgical (MIS) hollow instruments. In addi-

tion Bio-C-PCD®s with different air removal charac-

teristics are now available. The sensitivity of these

Bio-C-PCD®s can be utilized to meet the requirement

of the load.

Bio-C-PCD®s are available with five different air re-

moval and steam penetration characteristics.

The validation of the Bio-C-PCD® according to the

load can be achieved by using the test method de-

scribed in DIN 58921.

gke Bio-C-PCD®s can be used in all four steriliza-

tion processes mentioned above, if the correct SCBIs

are inserted.

There are various hydrogen peroxide/plasma steriliza-

tion processes in the market with different penetration

characteristics of the sterilization agent. Depending on

the process used and on the load configuration an

appropriate Bio-C-PCD® should be selected.

Product description

The gke Steri-Record® Mini-Bio-Plus SCBI is de-

signed using a plastic vial with a small internal vol-

ume containing a biological indicator spore plate and

a glass ampoule with a growth medium and pH-

indicator inside. For steam-, ethylene oxide- and for-

maldehyde-sterilization processes, filter paper is used

as a carrier and as a closing filter below the cap. For

hydrogen peroxide/plasma-sterilization processes

plastic material is used. For a better recognition of the

different SCBI versions they each have different col-

oured caps (see table).

The labels of all SCBIs for steam sterilization proc-

esses contain a class 1 chemical indicator according to

EN ISO 11140-1 to check, if the SCBI has passed a

sterilization process. The self-adhesive label may be

detached from the SCBI and adhered on a documenta-

tion sheet.

An additional chemical indicator class 5 according

EN ISO 11140-1 allows that the result of steam ster-

ilization processes can be instantly read at the end of

the sterilization process. Therefore, it is not necessary

to wait for the result of the SCBI incubation since

class 5 indicators are equivalent to biological indica-

tors according to the above standard.

EO

VH2O2

STEAM

FORM

Page 2: STEAM gke Steri-Record Mini-Bio-Plus EOfoxtrot-25.srv.hosting.fi/files/Mini-Bio-plus.pdfgke Steri-Record® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact

The specially designed and patented gke Steri-

Record® Bio-C-PCD® construction consists of a large

internal volume with a stainless steel tubing system

inside and a minimal capsule volume at the closed end.

It can be used only with the specially designed Mini-

Bio-Plus SCBIs as described before, to create a high

sensitive Hollow Load PCD which cannot be com-

bined with conventional SCBIs.

For incubation of the SCBIs gke Steri-Record® incu-

bators with two different temperatures and program-

ming versions are available (see separate data sheet).

Performance characteristics

Self-contained biological indicators:

All gke biological indicators comply with the stan-

dard EN ISO 11138 series. For hydrogen peroxide

sterilization processes the standard prEN ISO 11138-6

draft is currently under development and not yet avail-

able.

gke Mini-Bio-Plus SCBIs for hydrogen peroxide

sterilization processes are supplied with a D-Value that

is tested under the defined sterilization conditions

described in the certificate. The SCBIs for Low Tem-

perature Steam Formaldehyde (LTSF) sterilization

processes** contain in the growth medium substances,

decomposing remaining absorbed formaldehyde, so

that the pretreatment with Na2SO3 according to

EN ISO 11138-5 is not required.

The incubation time of all Mini-Bio-Plus SCBIs has

been optimized, so that steam-, formaldehyde- and

H2O2-Mini-Bio-Plus SCBIs can be fully interpreted

within 24 hours and EO-Mini-Bio-Plus SCBIs within

48 hours incubation time. The SCBIs do not contain

additional enzymes and do not require fluorescent light

for evaluation. If the incubation time exceeds the time

mentioned, the colour of the media does not change

back, this could happen with some conventional

SCBIs. If the sterilization process is unable to kill the

spores, in most cases the colour change will already

occur after 8 hours.

For steam sterilization processes an additional class 5

indicator according to EN ISO 11140-1 is inserted.

The indicator enables the user to make an interpreta-

tion of the result immediately at the end of the process

and the batch can be released or retained before the

incubation time has finished. Therefore it is not neces-

sary to wait 3 hours or longer for incubation to take

place so the batch can be released. The result of a class

5 indicator provides a much higher probability of ste-

rility instead of using the result of a SCBI incubation

after 3 hours where only a probability of > 97 % is

achieved.

The Mini-Bio-Plus SCBIs meet the performance char-

acteristics published in the current United States Phar-

macopeia (USP) and European Pharmacopeia (EP).

Operational description

The SCBIs must be placed in the most difficult pene-

tration condition of packs or containers. If hollow de-

vices are to be sterilized SCBIs cannot be placed in-

side MIS instruments. In this case Bio-C-PCD®s with

SCBIs inside are used to simulate MIS instruments.

This alternative offers for the first time the possibility

to check sterility inside hollow instruments. SCBIs in

PCDs can be directly incubated at the end of the cycle

without opening packs or containers.

After the sterilization process the SCBIs are taken out

of the load or the Bio-C-PCD® and are activated by

crushing the glass ampoule. After they are incubated

correctly according to the directions for use, the colour

of the liquid will indicate a pass or fail of the steriliza-

tion process. If the colour has not changed, this indi-

cates a pass of the process, a colour change indicates a

fail.

Special test devices (gke Steri-Record®

Bio-C-PCD®s) where SCBIs can be used as a detec-

tor inside:

Bio-C-PCD®s with SCBIs inside can simulate porous

loads and hollow devices simultaneously. A special

Bio-C-PCD® to simulate the hollow load device de-

scribed in EN 867-5 is available as well.

Additional Bio-C-PCD®s are available to simulate

different load configurations and may be validated

with the test method, described in DIN 58921.

Page 3: STEAM gke Steri-Record Mini-Bio-Plus EOfoxtrot-25.srv.hosting.fi/files/Mini-Bio-plus.pdfgke Steri-Record® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact

Art.-No.* Product code Quantity/

pack Pop.

Sterilization

process

Colour

of cap

Colour change of

Incubation

temperature

Biological

indicator

spores

Indicators on label growth media in SCBIs

after sterilization and

incubation Before After

Sterilization sterile unsterile

324-501

B-S-MBP-10-5

10

105

Steam

Light blue

Yellow

Brown

Purple Yellow-

green 55-60°C G. Stearo-

thermophilus

324-505 50

324-510 100

324-551

B-S-MBP-R-10-5**

10

Steam***

Rapid Indicator 324-555 50

324-550 100

324-605 B-S-MBP-10-6

106

Steam Dark blue 50

324-655 B-S-MBP-R-10-6** Steam***

Rapid Indicator

325-601

B-F-MBP-10-6**

10

Formaldehyde Yellow Yellow 325-605 50

327-601

B-V-MBP-10-6

10 Hydrogen peroxide

White Purple Pink

327-605

50

326-605

B-E-MBP-10-6 Ethylene oxide Red Blue Green Green Yellow-

orange 33-37°C B. Atrophaeus

326-610 100

* To all article numbers a 3-digit alpha code is added. The additional letter code refers to the language and/or customized version. It is only added on the outside label, the inside of the pack is identical to the article numbers on the above tables.

** Under development, please ask gke order processing about delivery time.

*** Additional Class 5 chemical indicator according to EN ISO 11140-1

Order information

Mini-Bio-Plus self-contained biological indicators ( SCBI )

Benefits

First worldwide PCD-system (class 2 indicator)

with a special internal volume design hosting the

gke Steri-Record® Bio-C-PCD®s, which simulate

hollow devices.

Immediate release at the end of the steam steriliza-

tion process by checking the colour change of the

class 5 chemical indicator. In addition the proof of

the biological indicator result will be available after

incubation.

Quick readout within 12-24 hours for final inter-

pretation, immediate read out of SCBIs for steam

sterilization process without waiting 3 hours for

incubation.

Cost reduction using SCBIs with direct incubation

instead of using an external microbiological labora-

tory.

Special Mini-Bio-Plus SCBIs with minimize inter-

nal volume usable in Bio-C-PCD®s for steam-,

ethylene oxide-, formaldehyde and hydrogen per-

oxide/plasma-sterilization processes.

Mini-Bio-Plus SCBIs can be incubated on comple-

tion of each cycle without having to open any

packs or containers when used in Bio-C-PCD®s.

Art.-No.* Product code Product description Quantity

224-002 I-C Crusher for SCBI activation if no gke Steri-Record® incubator is used. 1

Simple handling and traceability of test

results.

The labels on the SCBI are self-adhesive and can

be removed for external documentation.

The combination of Mini-Bio-Plus SCBI and spe-

cially designed Bio-C-PCD®s allows correct testing

of hollow lumen instruments.

Several Bio-C-PCD®s are available to simulate

different loads.

A special oval and round Bio-C-PCD design for

use in small table-top or large sterilizers.

Cost-effective due to multiple use of the

Bio-C-PCD® without loosing sensitivity. All im-

portant parts are made of stainless steel or thermal

resistant polymers to guarantee long-term durabil-

ity.

The Mini-Bio-Plus SCBI can be easily removed

and evaluated immediately on completion of each

cycle because all outside parts consist of highly

thermal resistant polymers that protect hands from

high temperatures.

SCBIs also can be used for LTSF-formaldehyde-

sterilization processes without pretreatment of the

chemo-indicators with NO2SO3, described in

DIN EN ISO 11138-5.

Page 4: STEAM gke Steri-Record Mini-Bio-Plus EOfoxtrot-25.srv.hosting.fi/files/Mini-Bio-plus.pdfgke Steri-Record® Mini-Bio-Plus self-contained biological indicators (SCBI) and Bio-Compact

gke GmbH Your local gke sales partner:

Auf der Lind 10

65529 Waldems-Esch / Germany +49 61 26 - 94 32- 0

+49 61 26 - 94 32- 10

[email protected]

http://www.gke.eu 700-015 EN V07 12/2011

Different gke Steri-Record® process challenge devices (Bio-C-PCD®s) for Mini-Bio-Plus SCBIs

Art.-No.* Product code PCD-Version** Penetration Characteristics***

300-011 B-PM-OCPCD-1 oval

minimal requirements for air removal

300-016 B-PM-RCPCD-1 round

300-012 B-PM-OCPCD-2 oval

low requirements for air removal

300-017 B-PM-RCPCD-2 round

300-013 B-PM-OCPCD-3 oval air removal less difficult than Hollow load test

according to EN 867-5 (draft EN ISO 11140-6) 300-018 B-PM-RCPCD-3 round

300-014 B-PM-OCPCD-4 oval air removal equal to Hollow load test

according to EN 867-5 (draft EN ISO 11140-6) 300-019 B-PM-RCPCD-4 round

300-015 B-PM-RCPCD-5 round air removal more difficult than Hollow load test

according to EN 867-5 (draft EN ISO 11140-6)

Incubators****

Please order always 1 incubator version and aluminum block together which are assembled delivered.

Also order the correct plug on the power cord.

Art.-No.* Product code Description of incubator

610-109 I-37 Incubation temperature: 37°C fixed

610-110 I-57 Incubation temperature: 57°C fixed

610-111 I-V Variable temperature selection

610-112 I-V-T Variable temperature selection and programming of the incubation time

610-113 I-AB-MBP Aluminum block for all gke Steri-Record® SCBIs

610-114 I-AB-AMP Aluminum block for gke Steri-Record® stearo ampoules

**** For details see separate data sheet.

* To all article numbers a 3-digit alpha code is added. The additional letter code refers to the language and/or customized version. It is only

added on the outside label, the inside of the pack is identical to the article numbers on the above tables.

** Round versions are used for large and oval versions for small sterilizers.

*** PCDs for routine monitoring need to be validated according to the load using the test method DIN 58921.